1. 1 HPP
Constant Climate Chamber HPP
Technology References
Techn. Data Applications
Equipment Service
Unique selling
propositions
HOME
24.12.2012
2. 2 HPP
Unique Selling Propositions
Extremely economical
For stable and precise long-term tests
Perfect controllability of temperature and humidity
Optimally tuned to the requirements for all ongoing stability
tests (ICH-Q1)
high precision in temperature and humidity control
competitive price / price-performance ratio
ecological requirements
- minimum consumption, energy saving
- almost vibration free/low noise
service-friendly
HOME
24.12.2012
3. 3 HPP
Technical Data
HPP
Volume 108, 749 litres
Temperature range (without light) +5°C to +70°C
Temperature range (with light) +10°C to +40°C
Humidity range (with light) 10% to 85% rh
Humidity range (without light) 10% to 90% rh
Temperature variation in time
≤ +/- 0,1°C
(at 37°C)
Temperature uniformity
≤ +/- 0,3°C
(at 37°C)
Accuracy humidity sensor 0,5% rh
Humidity variation in time ≤ +/- 1,5% rh
Safety class Class3.3
Alarm Visual and accoustic
24.12.2012
4. 4 HPP
Comparison: Energy consumption peltier/compressor
Peltier technology is
particularly economical at
temperatures close to
ambient temperature
Energy is only required if
heating or cooling is needed
Closed Peltier cooling system,
no outside air is exchanged
No de-humidification of the
chamber and heavily reduced
risk of contamination
24.12.2012
5. 5 HPP
Simulation of climatic zones on earth
Climatic zone I: Climatic zone II: Climatic zone III: Climatic zone IV:
21 °C / 45 % rH 25 °C / 60 % rH / Sub-tropical 30 °C / 35 % rH 30 °C / 65 % rH / Hot and
Moderate climate climate / Mediterranien climate Hot and dry climate Humid climate
100
90
Requirements
on ICH Q1A
80
Accelerated test:
IV
70 40°C +/- 2 °C
II
75% +/- 5% rh
60 duration: 6 month
relative humidity %rH
I
50 Long term tests:
25°C +/- 2 °C
40 60% +/-5% rh
III
duration: 12 month
30
Intermediate stability test :
20 30°C +/- 2 °C,
65 % +/- 5 % rh
10
duration: 6 month
0
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95
HOME
temperature °C
24.12.2012
6. 6 HPP
Equipment - standard
2 stainless steel grids
Works calibration certificate at +10°C and +37°C at 60% rh
Software „Celsius“
MEMorycard
USB and printer interface
Capactive humidity sensor
Active microprocessor control for humidifying and dehumidiying
24.12.2012
7. 7 HPP
Equipment - options
Constant Climate Chamber HPP 108 NOW with light
Volume 108 l
Temperature range
without light +5 C up to +70 C
with light +10 C up to +40 C
Humidity range
without light 10 – 90% rh
with light 10 – 85% rh
Light efficiency 10.000 Lux
2 LED- light modules available:
Cold white light 5.500 K
Combination of cold white light
5.500 K and warm white 2.700 K
HOME
24.12.2012