Engr Khaleeque Ahmed 
Sir syed university of engineering & 
technology Karachi
One 
• X-ray tube 
Two 
• Operating Console 
Three 
• High voltage generator 
Four 
• Tube head
Five 
• X-ray tube housing 
Six 
• Collimator 
Seven 
• Grid
Eight 
• Patient Table 
Nine 
• Bucky 
Ten 
• Radiographic film
This is a modern rotating anode x-ray tube. It is encased 
completely in a metal protective housing. 
There are two principle parts: 
•The rotating anode(acts as the target for electrons and 
releases x-rays) 
•The cathode(provides a source of electrons) 
Purpose: 
– Controls leakage & off-focus radiation 
– Isolates high voltages 
– Helps to cool the tube
OPERATING 
CONSOLE 
THE CONSOLE IS THE PART OF MACHINE THAT 
CONTROLS THE OPERATION OF X-RAY MACHINE 
ALL MACHINE CONSOLE ARE LITTLE DIFFERENT BUT 
THERE ARE ALWAYS SIMILARITIES. 
THE CONSOLE IS WHERE WE CONTROL THE X-RAY TUBE 
CURRENT,VOLTAGES LINE COMPENSATION.
HIGH VOLTAGE 
GENERATOR 
 A single phase HV generator can take the power 
supplied by AC mains rectify it & transform 
 Rectification process is either in half wave or full 
wave to remove any negative going voltages, which 
would otherwise excite an anode to act as electron 
source.
TUBE HEAD 
 The external structure of x-ray tube consists of 
3 parts 
1.The support system 
2.the protective housing 
3.glass or metal enclosure
PROTECTIVE HOUSING 
 The x-ray photons are 
generated isotropic 
ally or in all directions. 
 The housing is 
designed to limit the 
beam to window.
PROTECTIVE HOUSING – PREVENT 
DAMAGE 
 The housing also provide mechanical support 
and protection from damage. 
 On some tubes, the housing also contains oil 
that provides more insulation and a thermal 
cushion.
Collimators 
Beam restricting device. 
it’s considered the best beam restricting 
device. 
remove unwanted radiations.
A collimator has two or three sets of lead shutters. 
Located immediately below the tube window, the 
entrance shutters limit the x-ray beam. 
One or more sets of adjustable lead shutters are 
located 3 to 7 inches (8 − 18 cm) below the tube.
GRID 
IT’S MADE UP OF LEAD MATERIAL 
ITS INSERTED B/W PATIENT AND THE FILM 
PURPOSE 
REDUCE THE SCATTERED RADIATION 
IMPROVE IMAGE CONTRAST.
X-RAY TABLES OR PATIENT 
TABLES 
CHARACTERISTICS 
•RADIOLUCENT MATERIAL 
•EASILY CLEANED 
•SCRATCH RESISTANT 
•BUCKY TRAY 
•BUCKY SLOT COVER 
•GRID
BUCKY 
A BUCKY IS TYPICALLY USED FOR TABLE 
OR WALL MOUNTED X-RAY SYSTEMS AND 
HOLDS THE X-RAY CASSETTE AND GRID. A 
BUCKY IS A DEVICE FOUND UNDERNEATH 
THE EXAM TABLE, A DRAWER LIKE DEVICE 
THAT THE CASSETTE AND GRID IS SLID 
INTO BEFORE SHOOTING X-RAY.
THANK U…………!!!!!

Contenu connexe

Tendances

PARTS OF X-RAY TUBE
PARTS OF X-RAY TUBEPARTS OF X-RAY TUBE
PARTS OF X-RAY TUBE
Rabia Iqbal
 
Medical imaging
Medical imagingMedical imaging
Medical imaging
sirrainbow
 

Tendances (20)

X ray machine- ppt
X ray machine- pptX ray machine- ppt
X ray machine- ppt
 
PARTS OF X-RAY TUBE
PARTS OF X-RAY TUBEPARTS OF X-RAY TUBE
PARTS OF X-RAY TUBE
 
x ray machine ppt
x ray machine pptx ray machine ppt
x ray machine ppt
 
Introduction to X-Ray Machine
Introduction to X-Ray MachineIntroduction to X-Ray Machine
Introduction to X-Ray Machine
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
1. x ray generation & properties
1. x ray generation & properties1. x ray generation & properties
1. x ray generation & properties
 
Dr vishnu x ray imaging
Dr vishnu x  ray imagingDr vishnu x  ray imaging
Dr vishnu x ray imaging
 
X ray generation Radiology information by rahul ppt 2
X ray generation  Radiology information by rahul ppt 2X ray generation  Radiology information by rahul ppt 2
X ray generation Radiology information by rahul ppt 2
 
xray tubes
xray tubesxray tubes
xray tubes
 
X ray ppt
X ray pptX ray ppt
X ray ppt
 
Automatic exposure control
Automatic exposure controlAutomatic exposure control
Automatic exposure control
 
Basic of x ray production
Basic of x ray productionBasic of x ray production
Basic of x ray production
 
What is xray
What is xrayWhat is xray
What is xray
 
Medical imaging
Medical imagingMedical imaging
Medical imaging
 
Physics(x rays)
Physics(x rays)Physics(x rays)
Physics(x rays)
 
Fundamentals of x ray imaging i
Fundamentals of x ray imaging iFundamentals of x ray imaging i
Fundamentals of x ray imaging i
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
 
Teletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 MachinesTeletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 Machines
 
Radiation physics
Radiation physicsRadiation physics
Radiation physics
 
Filters
FiltersFilters
Filters
 

Similaire à 2lab xray parts

2-componentsofx-raymachine-150205095657-conversion-gate02.pdf
2-componentsofx-raymachine-150205095657-conversion-gate02.pdf2-componentsofx-raymachine-150205095657-conversion-gate02.pdf
2-componentsofx-raymachine-150205095657-conversion-gate02.pdf
keerthikrishna41
 
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdfEEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
AnbarasuL1
 
CT TUBES AND ITS TYPES.pptx...............
CT TUBES AND ITS TYPES.pptx...............CT TUBES AND ITS TYPES.pptx...............
CT TUBES AND ITS TYPES.pptx...............
justinfan550
 
X ray tube
X ray tubeX ray tube
X ray tube
Rad Tech
 

Similaire à 2lab xray parts (20)

X ray
X rayX ray
X ray
 
X ray machine l
X ray machine lX ray machine l
X ray machine l
 
Rectifier Operations & Maintenance - Don Olson - 2.4.2020
Rectifier Operations & Maintenance - Don Olson - 2.4.2020Rectifier Operations & Maintenance - Don Olson - 2.4.2020
Rectifier Operations & Maintenance - Don Olson - 2.4.2020
 
2-componentsofx-raymachine-150205095657-conversion-gate02.pdf
2-componentsofx-raymachine-150205095657-conversion-gate02.pdf2-componentsofx-raymachine-150205095657-conversion-gate02.pdf
2-componentsofx-raymachine-150205095657-conversion-gate02.pdf
 
Components of X-Ray Machine
Components of X-Ray MachineComponents of X-Ray Machine
Components of X-Ray Machine
 
COMPONENTS OF CO-60 UNIT.pptx
COMPONENTS OF CO-60 UNIT.pptxCOMPONENTS OF CO-60 UNIT.pptx
COMPONENTS OF CO-60 UNIT.pptx
 
X ray tube, cassette and screens
X ray tube, cassette and screensX ray tube, cassette and screens
X ray tube, cassette and screens
 
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdfEEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
 
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdfEEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
EEE- BEE601 - TRANSMISSION AND DISTRIBUTION- Dr. V. Jayalakshmi.pdf
 
Equipments of power transmission
Equipments of power transmissionEquipments of power transmission
Equipments of power transmission
 
CT TUBES AND ITS TYPES.pptx...............
CT TUBES AND ITS TYPES.pptx...............CT TUBES AND ITS TYPES.pptx...............
CT TUBES AND ITS TYPES.pptx...............
 
THE COBALT BOMB
THE COBALT BOMBTHE COBALT BOMB
THE COBALT BOMB
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its care
 
Group 4
Group 4Group 4
Group 4
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Cttubeanddetectors
Cttubeanddetectors Cttubeanddetectors
Cttubeanddetectors
 
Collimation & filtration
Collimation & filtrationCollimation & filtration
Collimation & filtration
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Production of x rays
Production of x rays Production of x rays
Production of x rays
 
Monocrystalline silicon solar cells
Monocrystalline silicon solar cellsMonocrystalline silicon solar cells
Monocrystalline silicon solar cells
 

Plus de Khaleeque Memon (11)

1lab xray and xray tube
1lab xray and xray tube1lab xray and xray tube
1lab xray and xray tube
 
1lab xray and xray tube
1lab xray and xray tube1lab xray and xray tube
1lab xray and xray tube
 
2lab xray parts
2lab xray parts2lab xray parts
2lab xray parts
 
1lab xray and xray tube
1lab xray and xray tube1lab xray and xray tube
1lab xray and xray tube
 
11 lab pet
11 lab pet 11 lab pet
11 lab pet
 
10lab spect
10lab spect10lab spect
10lab spect
 
8labgamma
8labgamma8labgamma
8labgamma
 
7lab components of mri
7lab components of mri7lab components of mri
7lab components of mri
 
5lab components of ct scanner
5lab components of ct scanner5lab components of ct scanner
5lab components of ct scanner
 
4lab ct
4lab ct 4lab ct
4lab ct
 
3lab fluoroscopy
3lab fluoroscopy3lab fluoroscopy
3lab fluoroscopy
 

2lab xray parts

  • 1. Engr Khaleeque Ahmed Sir syed university of engineering & technology Karachi
  • 2. One • X-ray tube Two • Operating Console Three • High voltage generator Four • Tube head
  • 3. Five • X-ray tube housing Six • Collimator Seven • Grid
  • 4. Eight • Patient Table Nine • Bucky Ten • Radiographic film
  • 5. This is a modern rotating anode x-ray tube. It is encased completely in a metal protective housing. There are two principle parts: •The rotating anode(acts as the target for electrons and releases x-rays) •The cathode(provides a source of electrons) Purpose: – Controls leakage & off-focus radiation – Isolates high voltages – Helps to cool the tube
  • 6.
  • 7. OPERATING CONSOLE THE CONSOLE IS THE PART OF MACHINE THAT CONTROLS THE OPERATION OF X-RAY MACHINE ALL MACHINE CONSOLE ARE LITTLE DIFFERENT BUT THERE ARE ALWAYS SIMILARITIES. THE CONSOLE IS WHERE WE CONTROL THE X-RAY TUBE CURRENT,VOLTAGES LINE COMPENSATION.
  • 8. HIGH VOLTAGE GENERATOR  A single phase HV generator can take the power supplied by AC mains rectify it & transform  Rectification process is either in half wave or full wave to remove any negative going voltages, which would otherwise excite an anode to act as electron source.
  • 9. TUBE HEAD  The external structure of x-ray tube consists of 3 parts 1.The support system 2.the protective housing 3.glass or metal enclosure
  • 10. PROTECTIVE HOUSING  The x-ray photons are generated isotropic ally or in all directions.  The housing is designed to limit the beam to window.
  • 11. PROTECTIVE HOUSING – PREVENT DAMAGE  The housing also provide mechanical support and protection from damage.  On some tubes, the housing also contains oil that provides more insulation and a thermal cushion.
  • 12. Collimators Beam restricting device. it’s considered the best beam restricting device. remove unwanted radiations.
  • 13. A collimator has two or three sets of lead shutters. Located immediately below the tube window, the entrance shutters limit the x-ray beam. One or more sets of adjustable lead shutters are located 3 to 7 inches (8 − 18 cm) below the tube.
  • 14. GRID IT’S MADE UP OF LEAD MATERIAL ITS INSERTED B/W PATIENT AND THE FILM PURPOSE REDUCE THE SCATTERED RADIATION IMPROVE IMAGE CONTRAST.
  • 15. X-RAY TABLES OR PATIENT TABLES CHARACTERISTICS •RADIOLUCENT MATERIAL •EASILY CLEANED •SCRATCH RESISTANT •BUCKY TRAY •BUCKY SLOT COVER •GRID
  • 16. BUCKY A BUCKY IS TYPICALLY USED FOR TABLE OR WALL MOUNTED X-RAY SYSTEMS AND HOLDS THE X-RAY CASSETTE AND GRID. A BUCKY IS A DEVICE FOUND UNDERNEATH THE EXAM TABLE, A DRAWER LIKE DEVICE THAT THE CASSETTE AND GRID IS SLID INTO BEFORE SHOOTING X-RAY.
  • 17.