2. NAYAMED TECH TIP #2
Saving Battery
Life in
NayaMed ICDs.
ATP during charging provides painless therapy
without delaying the shock if it’s needed.
However, it could cause battery depletion
equivalent to a delivered shock.
3. NAYAMED TECH TIP #2
In NayaMed devices, if ATP during charging with ChargeSaver
are programmed ON, the device switches to ATP Before
Charging if successful ATP during charging therapy has
been delivered.
A study evaluating ATP in the VF zone reported that in 421
implanted patients, 116 VF episodes occurred in 37 patients.
In 4 patients, 38 charges were saved through delivering ATP
Before Charging 1. If you consider that a full-energy charge
reduces longevity by approximately 40 days 2 , that’s a total of
1520 days of longevity that were saved by ATP Before Charging
for those 4 patients (or 380 days/person).
Schoels, Wolfgang, et al. "Optimizing implantable cardioverter-defibrillator treatment of rapid ventricular
tachycardia: antitachycardia pacing therapy during charging." Heart Rhythm 4.7 (2007): 879-885.
NayaMed ND X4 VR Clinician Manual, Page 279.