Introduction
The analysis of bodily fluids for trace
elements is common to monitor
individuals’ exposure to toxic elements
and assess nutritional deficiencies.
To realize these goals, urine, whole
blood, and blood serum are commonly
analyzed for a variety of trace elements.
Although these matrices are similar,
there could be wide variation within
a matrix type among individuals. This work describes a case study involving the
analysis of selenium in blood serum, where the ability to remove the gadolinium
(Gd) doubly-charged interference is demonstrated.
Experimental
Sample Preparation
Ten serum samples and two QC samples (UTAK Serum, Normal and High) were
prepared by 10x dilution in 1% nitric acid. Germanium (10 µg/L) was added as an
internal standard. Quantitative results were determined with Additions Calibrations,
which involves preparing the calibration standards in a representative sample and
measuring all other samples against this calibration curve.
ICP-MS
A P P L I C A T I O N N O T E
Authors
Peter Dickenson1
, Gavin Robinson2
1
PerkinElmer Australia
Melbourne, Victoria 3150
Australia
2
Robinson Scientific
Cambridge 3434
New Zealand
The Benefits of
NexION 300D ICP-MS’
Reaction Mode in
Removing the Gd+2
Interference on
Selenium in Serum
2
Instrumental Conditions
All analyses were performed on a PerkinElmer NexION®
300D
ICP-MS in both Collision and Reaction modes – Table 1 shows
the conditions used. In Collision mode, 78
Se was measured,
while the use of methane in Reaction mode allowed both 78
Se
and 80
Se to be monitored.
Table 1. ICP-MS Conditions
	
	 Parameter	Condition	
Instrument	 NexION 300D ICP-MS
Nebulizer	 Glass concentric
Spray chamber	 Glass cyclonic
Sample uptake rate	 0.25 mL/min
RF power	 1500 W
Analytes	 78
Se, 80
Se
Internal standard	 Ge at 10 ppb
Dwell time	 50 ms
Collision mode	 He = 4 mL/min
Reaction mode	 CH4
= 0.65 mL/min
		 RPq = 0.75
First, a mass scan in Collision mode was performed on a sample that had good agreement between Collision and Reaction modes,
as shown in Figure 1. The peaks appear as expected, indicating that any interferences on 78
Se are being removed in Collision mode.
Next, a mass scan of a sample with poor agreement between Collision and Reaction modes was performed in Collision mode,
as shown in Figure 2. This scan revealed a number of peaks from m/z 77-79, indicating an interference not being removed by
Collision mode. Scans having this appearance generally indicate the presence of species present at half-masses (i.e. 77.5 amu),
which are not being resolved with the typical mass resolution used for analyses (0.7 amu). Half-masses result from doubly-
charged ions being formed in the plasma. The only doubly-charged element that would give peaks in this mass range of
selenium is gadolinium, so this was the suspected cause.
Table 2. Quantitative Results for QC Samples (units in µg/L)
Table 3. Quantitative Results from Sample Analysis (units in µg/L)
Figure 1. Mass scan in Collision mode of a sample which showed good
agreement between Collision and Reaction mode results.
	
	 Sample	 Collision Mode	 Reaction Mode	 Reaction Mode	 Certified Range
			 78
Se	 78
Se	 80
Se	 	
Utak Normal	 101	 98	 97	 86 – 144
Utak Elevated	 294	 257	 256	 235 - 353
	
	 Sample	 Collision Mode	 Reaction Mode	 Reaction Mode
			 78
Se	 78
Se	 80
Se
1		 115	 97	 97
2		 96	 86	 86
3		 61	 56	 56
4		 92	 84	 83
5		 52	 48	 49
6		 37	 34	 34
7		 1235	 36	 26
8		 61	 28	 28
9		 54	 48	 48
10	72	 64	 65
Results
The results for the QC samples are displayed in Table 2, while Table 3 shows the results for both Collision and Reaction modes for ten
samples and the QC samples. The accuracy of the results is evident from the QC samples, which recovered within the certified range.
For most samples, there is good agreement between the results in Collision mode and the two Se isotopes measured in Reaction mode,
except for samples 7 and 8, which show a large positive bias in Collision mode. This discrepancy required further investigation.
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2013, PerkinElmer, Inc. All rights reserved. PerkinElmer®
is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
011184_01	
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA	
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
Figure 2. Mass scan in Collision mode of a sample with poor agreement
between Collision and Reaction mode results.
The green bars in Figure 2 show the natural isotopic abundance
fingerprint of Gd+2
, which closely corresponds to the peaks seen in
the sample. The match is not perfect due to the presence of Br in
the reagent blank at m/z 79 and the inability of Collision mode to
completely remove Ar2
+
at m/z 80. Therefore, it appears that the
false high readings in Collision mode are the result of Gd+2
, which
is not completely removed. However, the good agreement of the
78
Se and 80
Se results in Reaction mode indicate that Gd+2
is being
more efficiently removed than in Collision mode. Since gadolinium
is used for MRI procedures, its presence in serum is not surprising
and is indicative of a patient who underwent an MRI.
Conclusion
This work demonstrates the ability of the NexION 300D ICP-MS to measure selenium in serum in both Collision and Reaction modes.
However, Reaction mode is more effective since it efficiently removes the gadolinium doubly-charged interference in serum. This is
important to avoid false high readings in samples from patients who have undergone MRI procedures and may have gadolinium in
their systems.

Contenu connexe

Tendances

housman_mini_4_5 foot PPT poster template 50 percent
housman_mini_4_5 foot PPT poster template 50 percenthousman_mini_4_5 foot PPT poster template 50 percent
housman_mini_4_5 foot PPT poster template 50 percent
Jonathan Oyler
 
Some Biophysical Methods for Demonstrating Comparability of Conformation and ...
Some Biophysical Methods for Demonstrating Comparability of Conformation and ...Some Biophysical Methods for Demonstrating Comparability of Conformation and ...
Some Biophysical Methods for Demonstrating Comparability of Conformation and ...
KBI Biopharma
 
Non aqueous reversed phase liquid chromatography with charged aerosol detecti...
Non aqueous reversed phase liquid chromatography with charged aerosol detecti...Non aqueous reversed phase liquid chromatography with charged aerosol detecti...
Non aqueous reversed phase liquid chromatography with charged aerosol detecti...
AizelFernandez1
 
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Fatemeh Sedaghatpour
 
2009 FAMRI symposium scaled file at 60%
2009 FAMRI symposium scaled file at 60%2009 FAMRI symposium scaled file at 60%
2009 FAMRI symposium scaled file at 60%
Jesus M. Gonzalez PhD
 

Tendances (20)

Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MSQuantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
 
housman_mini_4_5 foot PPT poster template 50 percent
housman_mini_4_5 foot PPT poster template 50 percenthousman_mini_4_5 foot PPT poster template 50 percent
housman_mini_4_5 foot PPT poster template 50 percent
 
2016 Poster
2016 Poster2016 Poster
2016 Poster
 
Some Biophysical Methods for Demonstrating Comparability of Conformation and ...
Some Biophysical Methods for Demonstrating Comparability of Conformation and ...Some Biophysical Methods for Demonstrating Comparability of Conformation and ...
Some Biophysical Methods for Demonstrating Comparability of Conformation and ...
 
Non aqueous reversed phase liquid chromatography with charged aerosol detecti...
Non aqueous reversed phase liquid chromatography with charged aerosol detecti...Non aqueous reversed phase liquid chromatography with charged aerosol detecti...
Non aqueous reversed phase liquid chromatography with charged aerosol detecti...
 
Analyzing Aggregates by Sedimentation Velocity and Light Scattering
Analyzing Aggregates by Sedimentation Velocity and Light ScatteringAnalyzing Aggregates by Sedimentation Velocity and Light Scattering
Analyzing Aggregates by Sedimentation Velocity and Light Scattering
 
Alta Peg X%20 Poster
Alta Peg X%20 PosterAlta Peg X%20 Poster
Alta Peg X%20 Poster
 
Measuring Comparability of Conformation, Heterogeneity and Aggregation with C...
Measuring Comparability of Conformation, Heterogeneity and Aggregation with C...Measuring Comparability of Conformation, Heterogeneity and Aggregation with C...
Measuring Comparability of Conformation, Heterogeneity and Aggregation with C...
 
Validation of factor xa assay for nadroparin calcium nadroparin injection
Validation of factor xa assay for nadroparin calcium nadroparin injectionValidation of factor xa assay for nadroparin calcium nadroparin injection
Validation of factor xa assay for nadroparin calcium nadroparin injection
 
Studying Reversible Self-Association of Biopharmaceuticals using AUC and Ligh...
Studying Reversible Self-Association of Biopharmaceuticals using AUC and Ligh...Studying Reversible Self-Association of Biopharmaceuticals using AUC and Ligh...
Studying Reversible Self-Association of Biopharmaceuticals using AUC and Ligh...
 
Validation of Factor IIa assay for nadroparin calcium or nadroparin injection
Validation of Factor IIa assay for nadroparin calcium or nadroparin injectionValidation of Factor IIa assay for nadroparin calcium or nadroparin injection
Validation of Factor IIa assay for nadroparin calcium or nadroparin injection
 
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03
 
Validation of factor xa assay for dalteparin sodium dalteparin injection
Validation of factor xa assay for dalteparin sodium dalteparin injectionValidation of factor xa assay for dalteparin sodium dalteparin injection
Validation of factor xa assay for dalteparin sodium dalteparin injection
 
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
 
Standard methods
Standard methodsStandard methods
Standard methods
 
Validation of factor xa assay for heparin sodium heparin injection
Validation of factor xa assay for heparin sodium heparin injectionValidation of factor xa assay for heparin sodium heparin injection
Validation of factor xa assay for heparin sodium heparin injection
 
Validation of Factor II and Xa Chromogenic Assays for Dalteparin, Enoxaparin,...
Validation of Factor II and Xa Chromogenic Assays for Dalteparin, Enoxaparin,...Validation of Factor II and Xa Chromogenic Assays for Dalteparin, Enoxaparin,...
Validation of Factor II and Xa Chromogenic Assays for Dalteparin, Enoxaparin,...
 
LCMS-MS Method for Evaluation of PPCPs in Environmental Water
LCMS-MS Method for Evaluation of PPCPs in Environmental WaterLCMS-MS Method for Evaluation of PPCPs in Environmental Water
LCMS-MS Method for Evaluation of PPCPs in Environmental Water
 
2009 FAMRI symposium scaled file at 60%
2009 FAMRI symposium scaled file at 60%2009 FAMRI symposium scaled file at 60%
2009 FAMRI symposium scaled file at 60%
 
Validation of factor xa assay for tinzaparin sodium tinzaparin injection
Validation of factor xa assay for tinzaparin sodium tinzaparin injectionValidation of factor xa assay for tinzaparin sodium tinzaparin injection
Validation of factor xa assay for tinzaparin sodium tinzaparin injection
 

En vedette

En vedette (7)

Determination of Impurities in Organic Solvents used in the Semiconductor Ind...
Determination of Impurities in Organic Solvents used in the Semiconductor Ind...Determination of Impurities in Organic Solvents used in the Semiconductor Ind...
Determination of Impurities in Organic Solvents used in the Semiconductor Ind...
 
Differentiation of Cultivars of Spanish Olive Oils Using Multiple LC/MS Analy...
Differentiation of Cultivars of Spanish Olive Oils Using Multiple LC/MS Analy...Differentiation of Cultivars of Spanish Olive Oils Using Multiple LC/MS Analy...
Differentiation of Cultivars of Spanish Olive Oils Using Multiple LC/MS Analy...
 
Application Note: A Study of Aged Carbon Nanotubes by Thermogravimetric Analysis
Application Note: A Study of Aged Carbon Nanotubes by Thermogravimetric AnalysisApplication Note: A Study of Aged Carbon Nanotubes by Thermogravimetric Analysis
Application Note: A Study of Aged Carbon Nanotubes by Thermogravimetric Analysis
 
Quantification of Bromide by APHA 4500 Method
Quantification of Bromide by APHA 4500 MethodQuantification of Bromide by APHA 4500 Method
Quantification of Bromide by APHA 4500 Method
 
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
 
Cavity-enhanced Absorption Spectroscopy for Trace Gas Detection with Frontier
Cavity-enhanced Absorption Spectroscopy for Trace Gas Detection with FrontierCavity-enhanced Absorption Spectroscopy for Trace Gas Detection with Frontier
Cavity-enhanced Absorption Spectroscopy for Trace Gas Detection with Frontier
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
 

Similaire à Benefits of NexION 300D ICP-MS reaction mode in removing the Gd interference one selenium in serum 011184-01

Poster-mAbChem-ppt file
Poster-mAbChem-ppt filePoster-mAbChem-ppt file
Poster-mAbChem-ppt file
Rongliang Lou
 
Automated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterAutomated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling Poster
Rick Youngblood
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminants
SCIEX
 

Similaire à Benefits of NexION 300D ICP-MS reaction mode in removing the Gd interference one selenium in serum 011184-01 (20)

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
 
Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...
 
RBC Poster Final
RBC Poster FinalRBC Poster Final
RBC Poster Final
 
The Influence Study of The Mole Ratio Reactant in Ceftriaxone Sodium Synthesi...
The Influence Study of The Mole Ratio Reactant in Ceftriaxone Sodium Synthesi...The Influence Study of The Mole Ratio Reactant in Ceftriaxone Sodium Synthesi...
The Influence Study of The Mole Ratio Reactant in Ceftriaxone Sodium Synthesi...
 
Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water
 
Practical Implementation of the New Elemental Impurities Guidelines May 2015
Practical Implementation of the New Elemental Impurities Guidelines May 2015Practical Implementation of the New Elemental Impurities Guidelines May 2015
Practical Implementation of the New Elemental Impurities Guidelines May 2015
 
A360105
A360105A360105
A360105
 
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
 
Determination of ascobic acid by coulometer
Determination of ascobic acid by coulometerDetermination of ascobic acid by coulometer
Determination of ascobic acid by coulometer
 
High Sensitivity HPLC Profiling of 2-AB Glycans
High Sensitivity HPLC Profiling of 2-AB GlycansHigh Sensitivity HPLC Profiling of 2-AB Glycans
High Sensitivity HPLC Profiling of 2-AB Glycans
 
Poster-mAbChem-ppt file
Poster-mAbChem-ppt filePoster-mAbChem-ppt file
Poster-mAbChem-ppt file
 
Automated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterAutomated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling Poster
 
Analysis Of Steroids
Analysis Of SteroidsAnalysis Of Steroids
Analysis Of Steroids
 
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminants
 
journals public
journals publicjournals public
journals public
 
Lecture 3 about it governanace and how it works
Lecture 3 about it governanace and how it worksLecture 3 about it governanace and how it works
Lecture 3 about it governanace and how it works
 
Validation of an Automated Glutamate Dehydrogenase Enzyme Activity Assay in S...
Validation of an Automated Glutamate Dehydrogenase Enzyme Activity Assay in S...Validation of an Automated Glutamate Dehydrogenase Enzyme Activity Assay in S...
Validation of an Automated Glutamate Dehydrogenase Enzyme Activity Assay in S...
 
Analytical purity method development and validation by gas chromatography of ...
Analytical purity method development and validation by gas chromatography of ...Analytical purity method development and validation by gas chromatography of ...
Analytical purity method development and validation by gas chromatography of ...
 
MANOJ KUMAWAT -Project HE shade quality
MANOJ KUMAWAT -Project HE shade qualityMANOJ KUMAWAT -Project HE shade quality
MANOJ KUMAWAT -Project HE shade quality
 

Plus de PerkinElmer, Inc.

Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
PerkinElmer, Inc.
 

Plus de PerkinElmer, Inc. (20)

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º Trimestre
 
É A VEZ DO PRIME
É A VEZ DO PRIMEÉ A VEZ DO PRIME
É A VEZ DO PRIME
 
TREC PARA A VIDA
TREC PARA A VIDATREC PARA A VIDA
TREC PARA A VIDA
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o Desempenho
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no Brasil
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGS
 
O que é X Frágil?
O que é X Frágil?O que é X Frágil?
O que é X Frágil?
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedor
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
 
CG Clarus® All In One
CG Clarus® All In OneCG Clarus® All In One
CG Clarus® All In One
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
 
XXI Encontro de Genética
XXI Encontro de GenéticaXXI Encontro de Genética
XXI Encontro de Genética
 

Dernier

🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
Call Girls In Delhi Whatsup 9873940964 Enjoy Unlimited Pleasure
 

Dernier (20)

Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
 
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
 
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
 
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 

Benefits of NexION 300D ICP-MS reaction mode in removing the Gd interference one selenium in serum 011184-01

  • 1. Introduction The analysis of bodily fluids for trace elements is common to monitor individuals’ exposure to toxic elements and assess nutritional deficiencies. To realize these goals, urine, whole blood, and blood serum are commonly analyzed for a variety of trace elements. Although these matrices are similar, there could be wide variation within a matrix type among individuals. This work describes a case study involving the analysis of selenium in blood serum, where the ability to remove the gadolinium (Gd) doubly-charged interference is demonstrated. Experimental Sample Preparation Ten serum samples and two QC samples (UTAK Serum, Normal and High) were prepared by 10x dilution in 1% nitric acid. Germanium (10 µg/L) was added as an internal standard. Quantitative results were determined with Additions Calibrations, which involves preparing the calibration standards in a representative sample and measuring all other samples against this calibration curve. ICP-MS A P P L I C A T I O N N O T E Authors Peter Dickenson1 , Gavin Robinson2 1 PerkinElmer Australia Melbourne, Victoria 3150 Australia 2 Robinson Scientific Cambridge 3434 New Zealand The Benefits of NexION 300D ICP-MS’ Reaction Mode in Removing the Gd+2 Interference on Selenium in Serum
  • 2. 2 Instrumental Conditions All analyses were performed on a PerkinElmer NexION® 300D ICP-MS in both Collision and Reaction modes – Table 1 shows the conditions used. In Collision mode, 78 Se was measured, while the use of methane in Reaction mode allowed both 78 Se and 80 Se to be monitored. Table 1. ICP-MS Conditions Parameter Condition Instrument NexION 300D ICP-MS Nebulizer Glass concentric Spray chamber Glass cyclonic Sample uptake rate 0.25 mL/min RF power 1500 W Analytes 78 Se, 80 Se Internal standard Ge at 10 ppb Dwell time 50 ms Collision mode He = 4 mL/min Reaction mode CH4 = 0.65 mL/min RPq = 0.75 First, a mass scan in Collision mode was performed on a sample that had good agreement between Collision and Reaction modes, as shown in Figure 1. The peaks appear as expected, indicating that any interferences on 78 Se are being removed in Collision mode. Next, a mass scan of a sample with poor agreement between Collision and Reaction modes was performed in Collision mode, as shown in Figure 2. This scan revealed a number of peaks from m/z 77-79, indicating an interference not being removed by Collision mode. Scans having this appearance generally indicate the presence of species present at half-masses (i.e. 77.5 amu), which are not being resolved with the typical mass resolution used for analyses (0.7 amu). Half-masses result from doubly- charged ions being formed in the plasma. The only doubly-charged element that would give peaks in this mass range of selenium is gadolinium, so this was the suspected cause. Table 2. Quantitative Results for QC Samples (units in µg/L) Table 3. Quantitative Results from Sample Analysis (units in µg/L) Figure 1. Mass scan in Collision mode of a sample which showed good agreement between Collision and Reaction mode results. Sample Collision Mode Reaction Mode Reaction Mode Certified Range 78 Se 78 Se 80 Se Utak Normal 101 98 97 86 – 144 Utak Elevated 294 257 256 235 - 353 Sample Collision Mode Reaction Mode Reaction Mode 78 Se 78 Se 80 Se 1 115 97 97 2 96 86 86 3 61 56 56 4 92 84 83 5 52 48 49 6 37 34 34 7 1235 36 26 8 61 28 28 9 54 48 48 10 72 64 65 Results The results for the QC samples are displayed in Table 2, while Table 3 shows the results for both Collision and Reaction modes for ten samples and the QC samples. The accuracy of the results is evident from the QC samples, which recovered within the certified range. For most samples, there is good agreement between the results in Collision mode and the two Se isotopes measured in Reaction mode, except for samples 7 and 8, which show a large positive bias in Collision mode. This discrepancy required further investigation.
  • 3. For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2013, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 011184_01 PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com Figure 2. Mass scan in Collision mode of a sample with poor agreement between Collision and Reaction mode results. The green bars in Figure 2 show the natural isotopic abundance fingerprint of Gd+2 , which closely corresponds to the peaks seen in the sample. The match is not perfect due to the presence of Br in the reagent blank at m/z 79 and the inability of Collision mode to completely remove Ar2 + at m/z 80. Therefore, it appears that the false high readings in Collision mode are the result of Gd+2 , which is not completely removed. However, the good agreement of the 78 Se and 80 Se results in Reaction mode indicate that Gd+2 is being more efficiently removed than in Collision mode. Since gadolinium is used for MRI procedures, its presence in serum is not surprising and is indicative of a patient who underwent an MRI. Conclusion This work demonstrates the ability of the NexION 300D ICP-MS to measure selenium in serum in both Collision and Reaction modes. However, Reaction mode is more effective since it efficiently removes the gadolinium doubly-charged interference in serum. This is important to avoid false high readings in samples from patients who have undergone MRI procedures and may have gadolinium in their systems.