INTRODUCTION
PerkinElmer Spotlight on Applications e-Zine – Volume 3
PerkinElmer knows that the right training, methods, applications, reporting and
support are as integral to getting answers as the instrumentation. That’s why
PerkinElmer has developed a novel approach to meet the challenges that today’s
labs face – that approach is called EcoAnalytix™, delivering to you complete
solutions for your applications challenges.

In this effort, we are pleased to share with you our Spotlight on Applications
e-zine, delivering a variety of topics that address the pressing issues and analysis
challenges you may face in your application areas today.

Our Spotlight on Applications e-zine consists of a broad range of applications
you’ll be able to access at your convenience. Each application in the table of contents
includes an embedded link that will bring you directly to the appropriate page
within the e-zine.




PerkinElmer
CONTENTS


Energy
• Determination of Methanol in Crude Oils According to ASTM D7059-04 Using the
  Clarus 680 GC with S-Swafer Micro-Channel Flow Technology
• Determination of Low-Level Oxygenated Compounds in Gasoline Using the Clarus 680 GC
  with S-Swafer Micro-Channel Flow Technology
• Fatty Acid Methyl Ester Contamination of Aviation Fuel by GC/MS
• Curing Determination of EVA for Solar Panel Application by DSC




Environmental
• Determination of Mercury in Wastewater by Inductively Coupled Plasma-Mass Spectrometry
• Qualitative Analysis of Evolved Gases in Thermogravimetry by Gas Chromatography/Mass
  Spectrometry




Food & Beverage
• The Determination of Metals in Dietary Supplements
• Solid Phase Extraction and GC/MS Analysis of Melamine Adulteration in Dairy Products




Pharmaceuticals & Nutraceuticals
• Fast Analysis of Fat-Soluble Vitamins Using Flexar FX-10 UHPLC and Chromera CDS
• Assuring Safety of Traditional Chinese Herbal Medicines by Monitoring Inorganic Impurities
  using ICP-MS
• Polymorphism in Acetaminophen Studied by Simultaneous DSC and Raman Spectroscopy
• Pressure-Balanced Headspace for the Determination of Class I, II and III Residual Solvents in
  Pharmaceuticals by USP Chapter <467> Methodology




                                                                                                  PerkinElmer
a p p l i c at i o n n o t e


                                              Gas Chromatography

                                              Author
                                              Andrew Tipler
                                              PerkinElmer, Inc.
                                              Shelton, CT 06484 USA




Determination of         Introduction
                         The gas chromatographic (GC) analysis of crude oil
Methanol in Crude Oils   is a challenging undertaking. Samples are viscous,
                         making them difficult to handle, and they contain
According to ASTM        hundreds of different compounds with carbon
                         numbers up to or even above C120, making a complete
D7059-04 Using the       chromatographic separation effectively impossible.

Clarus 680 GC with       ASTM® D7059-04 is an established method that has
                         been well validated for the determination of methanol
S-Swafer Micro-          in crude oils. The method lists five variants of the instru-
                         mentation that have demonstrated compliance with
Channel Flow             this method. The method also allows for alternative

Technology               configurations that will meet the required performance
                         criteria.

                         In this application note, a method based on a
                         PerkinElmer® Clarus® 680 GC with an S-Swafer™
                         micro-channel flow splitting device is described;
                         the data presented here will demonstrate that this
                         method complies with the requirements of ASTM®
                         D7059-04.
Experimental                                                                 will enter the first column, leaving the heavier crude-oil
    The Clarus 680 GC used in this application note is described                 compounds in the liner. The two alcohols enter the first
    in Figure 1 with a diagram, and Figure 2 with a photograph                   non-polar column and elute early in the chromatography.
    of the S-Swafer micro-channel flow splitting device used to                  A chromatogram of a standard mixture is shown in Figure 3.
    perform this analysis. This two-column backflushing configura-
    tion (designated as S6 in the Swafer documentation) enables
    the first column to be backflushed while the analytes are still
    being chromatographed on the second column. A restrictor
    tube is also connected to one of the S-Swafer outlets to enable
    the carrier-gas flow rate to be increased and to allow the
    chromatography to be monitored on the first column by
    connecting the restrictor outlet to the FID. Nitrogen is used
    as the carrier gas throughout this application – it is well
    suited for use with 0.530 mm i.d. columns. Nitrogen, when
    compared to helium, is less expensive, more available, and
    not in limited supply. The use of nitrogen is consistent with
    PerkinElmer initiatives to reduce the use of the declining
    global stocks of helium.
                                                                                 Figure 3. Chromatogram of standard mixture on first column with the
    The crude-oil sample is diluted 50:50 with clean toluene                     restrictor tube connected to the FID.
    solvent containing 1-propanol internal standard to produce
    a final concentration of 500 ppm. 1.0 µL of the diluted
    sample is injected into the programmable split/splitless (PSS)               From Figure 3, it can be seen that the last peak of interest,
    injector which has the liner temperature set to 125 °C.                      the 1-propanol internal standard, elutes at about 3.2 minutes.
    At this temperature, only the volatile fraction of the sample                Anything that elutes later than this time is of no analytical
                                                                                 interest and backflushing should commence soon after
                                                                                 elution of the 1-propanol peak – in this case 3.3 minutes.

                                                                                 The backflushing process occurs when the pressure at the
                                                                                 column inlet is less than that at the column outlet. This
                                                                                 can be achieved by reducing the first-column inlet pressure
                                                                                 at the PSS injector, increasing the (midpoint) pressure at
                                                                                 the S-Swafer or doing both. In this analysis, we want to
                                                                                 continue chromatography on the second column while we
                                                                                 backflush the first column so the only option is to reduce
                                                                                 the pressure at the injector. To enable a large backpressure
                                                                                 to be used, the second column has an inline restrictor
    Figure 1. The S-Swafer system used to determine methanol in crude oil.       connected between it and the S-Swafer. This enables the
                                                                                 midpoint pressure to be increased, yet still allows reasonable
                                                                                 flow rates to be applied to the second column. The reduction
                                                                                 in the inlet pressure is affected through the use of a simple
                                                                                 GC timed event.

                                                                                 When the backflushing commences after a crude-oil sample
                                                                                 has been injected, the heavier fraction of the sample will still
                                                                                 reside within the injector liner which has been held at 125 °C.
                                                                                 At this point, the PSS liner is temperature programmed to
                                                                                 a high temperature to vaporize this less-volatile material.
                                                                                 Because the column is being backflushed, none of this vapor
                                                                                 will enter the column, but will be flushed out of the system
                                                                                 through the split vent. In this way, removal of the heavy
                                                                                 compounds is very efficient and doesn’t expose the columns
                                                                                 to this material, thus prolonging the column life.
    Figure 2. Photograph of installed system showing the S-Swafer connections.




2
Figure 4 shows a chromatogram of a standard solution on the                    backflushed until the oven reaches its initial programmed
first column with backflushing applied at 3.3 minutes. The                     temperature. The initial split-flow rate is set to 100 mL/min
chromatography is now very clean beyond the 1-propanol peak.                   to ensure that the pre-run pressure change equilibrates
                                                                               quickly. The split flow is reset to 10 mL/min by a pre-run
                                                                               event at -0.50 minutes, which occurs once the pressure has
                                                                               stabilized. These split-flow changes serve only to save time.


                                                                                Table 1. Analytical Conditions for the Determination of
                                                                                Methanol in Crude Oil.
                                                                                Gas Chromatograph        PerkinElmer Clarus 680 GC
                                                                                Oven Temperature         125 °C for 1 minute, then 25 °C/min to
                                                                                                         250 °C
                                                                                Injector                 Programmable Split/Splitless (PSS)
                                                                                Injector Temperature     125 °C for 3.3 minutes, then 200 °C/min
                                                                                                         to 400 °C and hold until the end of the run
                                                                                Carrier Gas              Nitrogen
Figure 4. Chromatogram of standard mixture on first column with backflushing    Initial Injector Pressure
at 3.3 minutes with the restrictor tube connected to the FID.                   Setpoint                  2 psig (see text)
                                                                                Initial Injector Split
Figure 5 shows chromatograms from the CP-Lowox column           ®               Flow Rate                100 mL/min (see text)
of three calibration mixtures that cover the calibration range                  Detector                 Flame Ionization (FID)
of this method. Again, the chromatography looks very clean.                     Detector Temperature 325 °C
                                                                                Detector Combustion
                                                                                Gases               Air: 450 mL/min, Hydrogen: 45 mL/min
                                                                                Detector Range           x1
                                                                                Detector Attenuation     x4
                                                                                Backflush System         S-Swafer configured in S6 mode
                                                                                Precolumn                30 m x 0.530 mm x 5 µm PerkinElmer
                                                                                                         Elite™ 1 with 25 cm x 0.250 mm
                                                                                                         deactivated fused silica restrictor
                                                                                                         connected between S-Swafer and column
                                                                                Analytical Column        10 m x 0.530 mm x 10 µm Varian®
                                                                                                         CP-Lowox® with in-line 25 cm x 0.100 µm
                                                                                                         deactivated fused silica restrictor
                                                                                                         connected between S-Swafer and column
                                                                                Restrictor Tubing
Figure 5. Chromatography of three methanol standard solutions containing        between S-Swafer
~500 ppm w/w 1-propanol internal standard.                                      and Detector             30 cm x 0.100 µm deactivated fused silica
                                                                                (Midpoint) Pressure
The full method for this analysis is given in Table 1. The use                  at S-Swafer              20 psig
of timed events in this method needs explanation. The GC                        Timed Events (see text) PSS pressure set to 24 psig at -1.00 min
oven needs to be programmed up to 250 °C in order to                                                     PSS split flow set to 10 mL/min at -0.50 min
elute the analytes from the CP-Lowox® column. At this                                                    PSS pressure set to 2 psig at 3.30 min
temperature, there is some slight stationary-phase bleed from
                                                                                                         PSS split flow set to 100 mL/min at 3.31 min
the thick-film methyl silicone precolumn. The precolumn
remains in the backflush mode during cooling to prevent any                     Sample Preparation       5 g of crude-oil sample mixed with 5 g of
                                                                                                         toluene containing 1-propanol internal
of the bleed from the precolumn at the onset of cooling from
                                                                                                         standard to deliver a final concentration
entering the CP-Lowox® column. The initial injector pressure                                             of 500 ppm w/w
of 2 psig and the pre-run timed event to set the pressure
                                                                                Sample Injection         Normal injection of 1.0 µL of prepared
to 24 psig at -1.00 minutes ensures that the pre-column is
                                                                                                         sample using an autosampler




                                                                                                                                                        3
Discussion                                                                 What should also be noted is that the total chromatography
    The calibration plot obtained for the mixtures of methanol                 time is just six minutes. For most of this time, the injector
    and 1-propanol in toluene are shown in Figure 6. This plot                 liner is being heated at a high temperature and the precolumn
    demonstrates good linearity which is in compliance with                    is being backflushed at a very high flow rate, while the oven
    ASTM® D7059.                                                               is being temperature programmed. This means that, at the
                                                                               end of the run, all of the less-volatile sample compounds
                                                                               have been flushed from the system. All that is then necessary is
                                                                               to cool the oven, reset the inlet pressure and inject the next
                                                                               sample. The Clarus 680 GC used for this analysis has rapid
                                                                               oven cooling and starts to load the autosampler with sample
                                                                               prior to the GC coming ready and so sample throughput
                                                                               exceeds six samples per hour.

                                                                               Table 2 shows the quantitative precision obtained from ten
                                                                               replicate injections of a high-level and a low-level check
                                                                               standard. Considering the complexity of the sample matrix,
                                                                               these results demonstrate the efficacy of the Swafer tech-
                                                                               nology for this type of application. These results greatly
                                                                               exceed the requirements of ASTM® D7059.
    Figure 6. Calibration plot showing response ratio vs. amount ratio for
    methanol:1-propanol internal standard.
                                                                                Table 2. Quantitative precision of low-level and high-level
                                                                                check standards prepared using a sample of light crude oil.
    Figure 7 shows a chromatogram of a low-level check standard
                                                                                Run                   Results for          Results for
    prepared from a sample of light crude oil. The hydrocarbons                                       Check Standard A     Check Standard B
    that elute with the alcohols into the second column before                                        28.2 ppm w/w         1096 ppm w/w
    backflushing is initiated are lightly retained on the CP-Lowox®             1                     33.2                 1063
    column and so quickly elute in the chromatography. The
                                                                                2                     32.2                 1063
    alcohols are much more strongly retained and require an
    extended temperature program to elute and are well separated                3                     32.7                 1065
    from the light hydrocarbons in the crude oil. Note the clean                4                     32.6                 1059
    baseline around the alcohols, which facilitates precise quan-               5                     32.9                 1059
    tification.
                                                                                6                     35.6                 1062
                                                                                7                     33.4                 1065
                                                                                8                     33.9                 1068
                                                                                9                     33.6                 1061
                                                                                10                    33.8                 1064
                                                                                Relative Std Dev %    2.88                 0.26


                                                                               The analytical system was validated using a set of 20 ‘round-
                                                                               robin’ samples prepared by Spectrum Analytical Standards,
                                                                               Sugarland, Texas. These are shown in Figure 8. These sam-
                                                                               ples were prepared five years ago to validate the published
                                                                               ASTM® D7059-04 method. It is believed that these samples
                                                                               are stable and would provide a good basis for validating this
    Figure 7. Chromatogram of 25 ppm w/w check standard prepared in a sample   Swafer system for compliance with the ASTM® method.
    of light crude oil.




4
One of the concerns regarding the GC analysis of crude oil
                                                                is that non-volatile sample residue will accumulate in the
                                                                system giving rise to adsorption, thermolysis or carryover
                                                                effects. This method uses a combination of low injection
                                                                temperature and column backflushing to keep heavier
                                                                sample compounds out of the columns. Figure 9 shows the
                                                                chromatography of crude-oil samples soon after a new liner
                                                                and septum have been installed compared against the chro-
                                                                matography after over 150 crude-oil injections have been
                                                                made into the same liner and septum. Clearly, sample residue
                                                                will accumulate as the number of injections increases (particu-
                                                                larly with heavy crude oils) and eventually the liner and the
                                                                septum will need to be changed. Figure 9 shows that after
                                                                150 injections, this method continues to perform.

Figure 8. ASTM® D7059 round-robin validation samples.



Table 3 summarizes the results from the round-robin sample
set. All are within the ASTM® D7059 limits except the result
for Sample 19, which is just outside the accepted range.
Given the age of these samples, these results are considered
acceptable, and the method does tolerate one result deviation
out of twenty.


  Table 3. Results obtained for ASTM® D7059 round-robin
  validation samples.
  Sample               Experimental               Expected
  1                    2.0                        2.7
                                                                Figure 9. Chromatograms of the same crude-oil sample spiked with 400 ppm
  2                    4.3                        5.8           w/w methanol produced using a single injector liner showing no significant
                                                                degradation in chromatography after over 150 runs.
  3                    7.4                        11.7
  4                    11.4                       14.9
  5                    7.8                        12.9
  6                    29.8                       35.1
  7                    23.8                       27.6
  8                    31.2                       35.7
  9                    58.1                       66.3
  10                   60.4                       67.0
  11                   69.5                       74.5
  12                   105.5                      113.4
  13                   113.4                      133.2
  14                   265.3                      285.2
  15                   393.8                      404.7
  16                   542.8                      592.9
  17                   596.3                      524.2
  18                   873.2                      912.2
  19                   739.8                      826.5
  20                   709.9                      734.7




                                                                                                                                             5
Conclusion
This method, based on the S-Swafer technology, fully
complies with the requirements of ASTM® D7059-04 and
offers the user a number of unique benefits:

•	 The	method	presented	here	will	provide	very	high	throughput	
   and fast delivery of results; with a complete analysis cycle
   time less than ten minutes per sample (more than six
   samples/hour throughput)

•	 Reduced	instrument	maintenance	as	a	result	of	temperature-
   programmed injection which prevents heavy sample
   compounds from entering and degrading columns

•	 Additional	protection	of	the	analytical	columns	and	reduction	
   of run time with the use of backflushing

•	 Single-detector	configuration	simplifies	and	reduces	the	cost	
   of the system

•	 Splitting	restrictor	enables	precolumn	chromatography	to	be	
   monitored, facilitating the setup of backflush timing

•	 Nitrogen	carrier	gas	reduces	costs	associated	with	carrier	
   gas and serves to help preserve helium stocks

•	 Quantitative	performance	exceeding	the	method	specification	
   delivers results that will meet laboratory quality-control
   specifications.




PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com



For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.

009027A_01
a p p l i c at i o n n o t e


                                              Gas Chromatography

                                              Author
                                              Andrew Tipler
                                              PerkinElmer, Inc.
                                              Shelton, CT 06484 USA




Determination of          Introduction
                          The existing ASTM® D4815 method is designed to monitor
Low-Level Oxygenated      oxygenated compounds in gasoline at percentage concen-
                          trations. The method described in this application note is
Compounds in Gasoline     intended to enable these analytes to be monitored down to

Using the Clarus 680 GC   low-ppm concentrations.

                          There is an increased need to monitor the level of oxygen-
with S-Swafer Micro-      ates in reformulated gasoline because of the increased risk
                          of contamination from more abundantly available biofuels.
Channel Flow Technology   Gasoline contamination with oxygenated compounds can
                          lead to more toxic carbonyl emissions into the atmosphere
                          from engines running the fuel.

                          Contamination of feedstocks with oxygenated compounds
                          can cause degradation of expensive catalysts and impact the
                          refining process.
The analytes targeted in this method are those specified in
    ASTM® D4815 and are given in Table 1.


     Table 1. Targeted Oxygenated Compounds.
     Methanol
     Ethanol
     Isopropanol
     tert-Butanol
     n-Propanol
     Methyl tert-butyl ether (MTBE)
     sec-Butanol                                                       Figure 1. Diagram of the S-Swafer micro-channel flow technology used to
                                                                       determine trace-level oxygenated compounds in gasoline.
     Diisopropyl ether (DIPE)
     Isobutanol
     Ethyl tert-butyl ether (ETBE)                                     to one of the S-Swafer outlets to enable the carrier-gas flow
     tert-Pentanol                                                     rate to be increased and to allow the chromatography to be
     1,2-Dimethoxyethane                                               monitored on the first column by connecting the restrictor
                                                                       outlet to the FID. Note the use of nitrogen as the carrier gas.
     n-Butanol
                                                                       This carrier gas is well suited to 0.530 mm i.d. columns and
     tert-Amyl methyl ether (TAME)                                     is consistent with initiatives to reduce the use of the declining
                                                                       global stocks of helium.

    Experimental                                                       The experimental details for this analysis are given in Table 2.
    Figure 1 shows a diagram of the S-Swafer™ micro-channel            The sample of gasoline is injected and chromatographed on
    flow splitting device used to perform this analysis. This two-     the non-polar precolumn. The polar oxygenated compounds
    column backflushing configuration (designated as S6 in the         are quick to elute into the second CP-Lowox® column and
    Swafer documentation) enables the first column to be back-         will precede the bulk of the gasoline hydrocarbons. Once
    flushed while the analytes are still being chromatographed         the oxygenated compounds are in the second column, the
    on the second column. A restrictor tube is also connected          first column is backflushed to remove the hydrocarbons



     Table 2. Analytical Conditions for the Determination of Trace-Level Oxygenated Compounds in Gasoline.
     Gas Chromatograph           PerkinElmer® Clarus® 680 GC           Analytical Column        10 m x 0.530 mm x 10 µm Varian® CP-Lowox®
     Oven Temperature            80 °C for 1 minute then 5 °C/min                               with in-line 25 cm x 0.100 µm deactivated
                                 to 125 °C then 10 °C/min to 230 °C                             fused silica restrictor connected between
                                                                                                S-Swafer and column
     Injector                    Heated split/splitless
                                                                       Restrictor Tubing        30 cm x 0.100 µm deactivated fused silica
     Injector Temperature        250 °C                                between S-Swafer
     Carrier Gas                 Nitrogen                              and Detector
     Initial Injector Pressure   2.0 psig (see text)                   (Midpoint) Pressure      35 psig
                                                                       at S-Swafer
     Injector Split-Flow Rate    15 mL/min
                                                                       Timed Events             Carrier gas pressure set to 45 psig at -1.00 min
     Detector                    Flame Ionization (FID)
                                                                       (see text)               Carrier gas pressure set to 2 psig at 1.52 min
     Detector Temperature        325 °C
                                                                       Sample Preparation       a) AccuStandard® 4815-RT-PAK diluted
     Detector Combustion Gases Air: 450 mL/min,                                                 ~1:1500 in washed gasoline to give 27.4
                               Hydrogen: 45 mL/min                                              to 50 ppm w/w
     Detector Range              x1                                                             b) AccuStandard® 4815-RT-PAK diluted
     Detector Attenuation        x4                                                             ~1:7500 in washed gasoline to give 5.5
                                                                                                to 10 ppm w/w
     Backflush System            S-Swafer configured in S6 mode
                                                                       Sample Injection         Fast injection of 1.0 µL of prepared sample
     Precolumn                   15 m x 0.530 mm x 1 µm PerkinElmer                             using an autosampler
                                 Elite™-1 with 25 cm x 0.250 mm
                                 deactivated fused silica restrictor
                                 connected between S-Swafer and
                                 column

2
from the GC system. While the backflushing is in progress,
the oxygenated compounds are chromatographed on the
highly polar CP-Lowox® column which easily separates them
from any hydrocarbons that also enter the second column.

Precolumn backflushing is initiated by a timed event at
1.52 minutes that reduces the pressure inside the injector
to 2.0 psig so that carrier gas flows backwards. The inlet
pressure is set to 2.0 psig to maintain the backflushing of
the precolumn during oven cooling. A pre-run timed event
sets the inlet pressure to 45 psig so that forward flow is
restored prior to injection of the sample.

Figure 2 shows a chromatogram of gasoline on the pre-
column with no backflushing. The oxygenated compounds
would elute within the first one or two minutes of this
                                                                 Figure 3. Precolumn chromatogram of a mixture of oxygenated compounds
chromatogram and would be totally obscured by the early-
                                                                 at 4% to 7.3% w/w concentration indicating the proposed backflush point.
eluting hydrocarbons in the gasoline. The later-eluting peaks    Injector split flow increased to 100 mL/min.
are not needed in this analysis so they will be the target of
the backflushing step.
                                                                 Figure 4 shows the same sample run under the same conditions
                                                                 as those for Figure 3 but with the detector connected to
                                                                 the end of the analytical column. Good peak shapes and
                                                                 separations of the oxygenated compounds are seen within a
                                                                 20.5-minute run time. On this column, some of the alcohol
                                                                 isomers are not separated. The total cycle time was about
                                                                 25 minutes, which included oven cooldown and equilibration
                                                                 and autosampler loading.




Figure 2. Chromatogram of 87-octane gasoline on the precolumn.



Figure 3 shows a chromatogram of a mixture of the oxygenated
compounds (no solvent) on the precolumn and it can be
seen that all the peaks of interest have eluted from this
column in just less than 1.50 minutes. From this trace, the
backflush point was chosen to be 1.52 minutes.


                                                                 Figure 4. Analytical column chromatogram of a mixture of oxygenated
                                                                 compounds at 4% to 7.3% w/w concentration with precolumn backflushing
                                                                 at 1.52 minutes. Injector split flow increased to 100 mL/min.




                                                                                                                                            3
To test the system on samples of gasoline containing known                     The standard mixture of oxygenated compounds was diluted
    levels of oxygenated compounds, a quantity of gasoline was                     with the washed gasoline to produce low-level mixtures.
    ‘washed’ with water. 5 mL of 87-octane gasoline obtained                       These were chromatographed using the newly developed
    from a local filling station were shaken with 10 mL of deionized               method and the chromatogram is shown in Figure 6.
    water for two minutes and then centrifuged at 5000 rpm for
    ten minutes. The aqueous layer was discarded and the
    process was repeated. 0.5 g of anhydrous sodium sulfate
    was added to the cleaned gasoline and shaken to remove
    any traces of water.

    Figure 5 shows chromatography of the gasoline before
    and after washing. Even though the gasoline contained up
    to 10% ethanol, this and all other polar compounds were
    effectively removed by the washing procedure.
                                                                                   Figure 6. Chromatogram of washed gasoline spiked with low levels of
                                                                                   oxygenated compounds.



                                                                                   The 27.4 – 50 ppm w/w mixture was injected repeatedly to
                                                                                   measure the area and retention-time precisions. These are
                                                                                   summarized in Table 3.

                                                                                   The area precision is around 1% relative standard deviation
                                                                                   which is an excellent result – especially for polar compounds
                                                                                   at low levels in a highly complex sample matrix like gasoline.
                                                                                   The retention-time precision, which ranges from 0.013%
                                                                                   to 0.039% relative standard deviation, is again an excellent
                                                                                   result, considering the nature of this analysis.



    Figure 5. Efficacy of gasoline washing procedure for preparation of standard
    mixtures.



      Table 3. Area and Retention-Time Precision of a Standard Mixture of Oxygenated Compounds (n=9).
      Compound                                Conc. (ppm w/w)             Mean Area       RSD% Area          Mean R.T (min)         RSD% R.T.
      Ethyl tert-Butylether (ETBE)            27.4                        64807           0.69               9.580                  0.028
      Methyl tert-Butylether (MTBE)           27.4                        58557           0.64               9.827                  0.021
      Diisopropylether (DIPE)                 27.4                        53993           0.80               10.044                 0.039
      tert-Amylmethylether (TAME)             50.0                        101209          0.65               11.503                 0.039
      Methanol                                50.0                        31028           1.11               13.642                 0.023
      Ethanol                                 50.0                        60104           0.96               15.663                 0.015
      n & i-Propanol                          50.0/50.0                   142362          1.10               17.109                 0.016
      i, s & t-Butanol                        50.0/50.0/50.0              273656          0.72               18.233                 0.009
      n-Butanol                               50.0                        89299           0.76               18.740                 0.013
      tert-Pentanol                           50.0                        103335          0.63               19.287                 0.013
      1,2-Dimethoxyethane                     41.1                        64233           1.26               19.982                 0.016




4
Conclusion
This method, based on the S-Swafer technology, is able to
separate and quantify the ASTM® D4815 target analytes at
low-ppm levels and offers the following user benefits:

•	 High-throughput	analysis,	with	a	complete	analysis	cycle	
   time of 25 minutes

•	 A	robust	system	–	the	analytical	column	is	protected	from	
   most of the gasoline compounds by backflushing

•	 In-line	restrictor	with	secondary	column	allows	the	primary	
   column to be backflushed during chromatography, reducing
   analysis time

•	 Splitting	restrictor	enables	precolumn	chromatography	to	be	
   monitored – facilitating system setup

•	 Use	of	nitrogen	carrier	gas	to	help	preserve	helium	stocks	
   and reduce operating costs

•	 Excellent	quantitative	precision.




PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com



For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.

009033_01
a p p l i c at i o n n o t e

                                                                   Gas Chromatography/
                                                                   Mass Spectrometry

                                                                   Author
                                                                   Greg Johnson
                                                                   PerkinElmer, Inc.
                                                                   Shelton, CT 06484 USA




Fatty Acid Methyl                             Introduction
                                              The contamination of aviation fuel with fatty acid
Ester Contamination of                        methyl esters (FAMEs) can arise due to the use of
                                              multi-product pipelines for fuel supply and distribution.
Aviation Fuel by GC/MS                        In some countries, the widespread, mandatory intro-
                                              duction of automotive fuels with a bio-material content
                                              means that these pipelines are exposed to both auto-
                                              motive biodiesel with a 5% FAME (BD 5) content as
                                              well as to aviation fuel. FAMEs can adsorb onto the
                                              surface of the pipeline and later desorb, contaminating
                                              whatever fuel that follows, including aviation fuel.

        FAMEs alter the physical properties of the fuel. While presence of FAMEs in aviation fuel is
        of global concern, the United Kingdom is at the forefront of the analysis. The specification
        of aviation fuel in the UK is defined in the Ministry of Defence (MoD) Defence Standard
        91-91. Technical authority for this standard is controlled by the MoD Defence Fuels Group
        with the agreement of the UK Civil Aviation Authority (CAA). This specification has recently
        been amended to formalize the acceptance of FAMEs in aviation fuel up to a maximum of
        5 mg/Kg (ppm) combined total. To measure this specification, the Energy Institute has issued
        a method entitled IP PM DY: Determination of fatty acid methyl esters (FAME) derived from
        biodiesel fuel, in aviation turbine fuel – GC/MS with selective ion monitoring/scan detection
        method. The US specification of aviation fuel is covered by ASTM® D1655; in August of
        2009, in a special airworthiness bulletin (NE-09-25R1), the Federal Aviation Administration
        (FAA) stated its intent to include a similar regulation in its specifications.
Current methods for determining FAME in aviation fuel use              Table 1. Detailed instrument parameters for the GC/MS
    a polar ‘wax’ type GC column with mass spectrometry (MS)               analysis of FAMEs.
    for detection. The polar column will be more retentive for
                                                                           GC Conditions
    FAME compounds relative to the less polar hydrocarbon
                                                                           Column             25 m x 0.32 mm id x 0.25 μm BPX70
    compounds of the fuel. The MS will identify FAMEs based
    on unique spectral data, further distinguishing FAMEs from             Injector           Split/splitless @ 280 ˚C with glass wool packed
    hydrocarbon aviation fuel. The limitation of current methods                              liner
    is that the wax column has a relatively low maximum-                   Carrier Gas        Helium @ 15 psig at injector and 3 psig at Swafer
    temperature limit and high column bleed as it approaches               Injection          1.0 μL splitless for 0.75 minutes then 50 mL/min
    this limit.                                                            Oven               150 ˚C for 2 minutes then 10 ˚C/min to 220 ˚C
                                                                                              and hold for 6 minutes
                                                                           MS Conditions
                                                                           Mode               Electron Ionization (70 eV)
                                                                           Source             180 ˚C
                                                                           Transfer Line      300 ˚C
                                                                           Photomultiplier 550 V



                                                                           Table 2. Calibration summary for FAMEs with GC/MS.

    Figure 1. 0.1 ppm (wt/wt) individual FAME compounds in n-dodecane.     FAME        Concentration Range        Coefficient of
                                                                                                                  Determination (r2)
                                                                           C16:0       0.5 to 10.0 ppm            0.999929
    In this application note, a different type of polar capillary
                                                                           C18:0       0.5 to 10.0 ppm            0.999872
    column is used. This column reduces the temperature necessary
    to elute all compounds, while also providing a higher maxi-            C18:1       0.5 to 10.0 ppm            0.999744
    mum temperature. This will reduce both the wear on the                 C18:2       0.5 to 10.0 ppm            0.999577
    column and the amount of signal in the chromatogram                    C18:3       0.5 to 10.0 ppm            0.999092
    associated with column bleed. Additionally, this method will
    increase sample throughput. The specification of 5 mg/Kg is
    a total for all FAME compounds present and therefore the             Calibration
    limit of detection for individual FAME compounds needs to            A commercially available standard (Supelco®) containing
    be significantly lower than 5 mg/Kg. Figure 1 demonstrates           equal masses of pure C16:0, C18:0, C18:1, C18:2 and
    that 0.1 mg/Kg is possible using this experimental setup.            C18:3 was diluted by weight to 2000 mg/Kg with n-dodecane
                                                                         solvent, as recommended in Institute of Petroleum method
    Experimental                                                         PM-DY/09. Calibration standards were prepared at 0.5, 1.0,
    This application was performed on a PerkinElmer® Clarus®             2.5, 5.0 and 10.0 mg/Kg and spiked with 2.0 µL of C21:0 in
    680 GC/MS with a capillary split/splitless injector. A 1.0 µL        n-dodecane solution used as an internal standard.
    splitless injection was used to introduce the standards and
    samples into an unpacked, 2 mm i.d., quartz liner. The chro-
    matographic separation was achieved on a 25 m x 0.32 mm
    i.d. x 0.25 µm BPX70 column (SGE, Australia). The MS was
    used in single-ion-recording mode (SIR) to provide maximum
    sensitivity and specificity. Complete instrument parameters
    are presented in Table 1.




                                                                         Figure 2. The calibration result from TurboMass™ GC/MS software
                                                                         demonstrating the linear response of FAME C16:0 across a range of
                                                                         0.5-10.0 mg/Kg.



2
The calibration standards were analyzed in triplicate. All data
points were used without exception. The response across the         Table 5. The precision results from the analysis of a rapeseed
                                                                    oil mixture (n=10).
calibration range was very linear (r2 > 0.9991) for all FAME
compounds; complete calibration data is presented in Table 2.                                                             RSD%
Additionally, Figure 3 demonstrates the calibration output                                                       RSD%     Int’l Std.
                                                                         Conc. (%) in            Conc.           Absolute Response
from the PerkinElmer TurboMass software.
                                                                    FAME original mix            (ppm) final     Area     Ratio
Immediately following the instrument calibration, a 5 mg/Kg         C14:0     1.0                0.50            2.02         0.86
standard was analyzed to provide additional precision data.         C16:0     4.0                2.00            1.66         1.18
The results of this precision study are reported in Table 3.        C18:0     3.0                1.50            2.08         0.63
                                                                    C18:1     60.0               30.00           1.52         0.54
 Table 3. System precision evaluation, replicate injections
                                                                    C18:2     12.0               6.00            1.84         0.19
 (n=15) of the 5.0 mg/Kg calibration mixture.
                                                                    C18:3     5.0                2.50            2.34         0.89
 FAME       Concentration    RSD% Absolute       RSD% IS
                             Peak Area           Resp. Ratio        C20:0     3.0                1.50            2.08         0.33
 C16:0      5.0 ppm          1.34                0.78               C20:1     1.0                0.50            3.41         1.71
 C18:0      5.0 ppm          1.34                0.50               C22:0     3.0                1.50            2.83         1.10
 C18:1      5.0 ppm          1.40                0.62               C22:1     5.0                2.50            3.04         1.53
 C18:2      5.0 ppm          1.22                0.86               C24:0     3.0                1.50            3.66         1.86
 C18:3      5.0 ppm          1.55                0.86               Internal Standard = C21:0


The ability of the method to determine a wide range of
FAMEs at various concentrations was verified by analyzing a
commercially available rapeseed oil standard (Supelco®). The
standard was diluted 100 fold (100 mg to 10 g) by weight
in n-dodecane, with a further 200:1 dilution resulting in a
mixture with the composition in Table 4.


 Table 4. Composition of fame mixture used to verify the
 method identification of a mixture of FAMEs.
 FAME       Initial Concentration      Final Concentration
 C14:0      1.0 %                      0.5 ppm
 C16:0      4.0%                       0.8 ppm
 C18:0      3.0%                       0.6 ppm
                                                                  Figure 3. Chromatography of the rapeseed oil reference mixture using SIR.
 C18:1      60.0%                      30.0 ppm
 C18:2      12.0%                      6.0 ppm
                                                                  The final verification of this method was to analyze an
 C18:3      5.0%                       2.5 ppm
                                                                  aviation-fuel sample spiked with a known amount of FAME.
 C20:0      3.0%                       1.5 ppm                    Figure 4 presents the resultant chromatogram from the
 C20:1      1.0%                       0.5 ppm                    analysis of an aviation-fuel sample spiked with between
 C22:0      3.0%                       1.5 ppm                    0.5 and 30 ppm FAME.
 C22:1      5.0%                       2.5 ppm
 C24:0      3.0%                       1.5 ppm


The rapeseed oil was analyzed in ten replicates to verify the
precision of the method when the FAME materials span a
wide concentration range (Table 5). An example chromatogram
from this analysis is shown in Figure 3.




                                                                                                                                              3
Figure 4. The resultant chromatogram of the analysis of aviation fuel spiked
with FAME.



As you can see in Figure 4, the GC conditions provide
adequate resolution of C16:0 FAME from the hydrocarbon
envelope of the aviation fuel, as well as low-level detection
capabilities independent of matrix.


Conclusions
This application note demonstrated the use of the Clarus
680 GC/MS to identify and determine the concentration of
low-level FAMEs. A polar column was used in conjunction
with a short GC run to provide a fast and robust method.
The MS was operated in SIR mode to achieve the highest
levels of sensitivity and specificity.

Also demonstrated in this note is compliance with the newly
released methods set forth by regulatory agencies such as
the Institute of Petroleum. In addition, preliminary results
indicate that the current method used for the analysis of
FAME contamination in aviation fuel can be speeded up very
significantly without sacrificing either sensitivity or precision.




PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com



For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.

009031_01
a p p l i c at i o n n o t e

                                                                     Differential Scanning
                                                                     Calorimetry

                                                                     Authors
                                                                     Ji-Tao Liu
                                                                     Tiffany Kang
                                                                     Peng Ye
                                                                     PerkinElmer, Inc.
                                                                     Shelton, CT 06484 USA




Curing Determination                                Introduction
                                                    Renewable energy has attracted a lot of
of EVA for Solar Panel                              interest due to the limited supply of coal
                                                    and oil and the environmental concern
Application by DSC                                  of carbon dioxide (CO2) emission. There
                                                    are many different forms of renewable
                                                    (green) energy including: solar, wind,
                                                    geothermal, biomass, and so on. Among
                                                    them, solar energy is the fastest-growing
           segment. Increasing manufacturing capacity and decreasing product costs have
           led to significant growth in the solar industry over the past several years. For
           instance, solar photovoltaic (PV) production has been increasing by an average
           of 48% each year since 2002. By the end of 2008, the cumulative PV installation
           reached more than 15 giga-watts globally.

           A solar cell is a device that can convert sunlight directly into electricity. Different
           solar-cell technologies including crystalline silicon, organic photovoltaics, and
           dye-sensitized solar cells have been developed for various solar-cell applications.
           Currently, the most widely commercially available solar cell is based on crystalline-
           silicon technology. This technology is mature compared with the other solar-cell
           technologies and its energy-conversion efficiency is high.
A photovoltaic module or system consists of many jointly             Experimental
    connected solar cells. The solar cells are packaged between          The instrument used here is the PerkinElmer® double-furnace
    a backsheet on the bottom and a tempered-glass window                DSC 8000. It features power-controlled design for direct and
    on the top. The cells are encapsulated by a polymer encapsulant      accurate heat-flow measurements to and from the sample
    (Figure 1). The polymer encapsulant serves many functions –          material. The cooling accessory is an Intracooler 2P mechanical
    it provides mechanical support, electrical isolation, and            refrigerator. Nitrogen is used as the sample purge gas at
    protection against outdoor environmental elements of                 20 mL/min. The instrument was calibrated with two metal
    moisture, UV radiation and temperature stress. Many different        reference materials: indium and zinc were used for temperature
    materials can be used for encapsulation, but one commonly            calibration, and indium was used for heat of fusion for
    used encapsulant for this purpose is EVA (ethylene-vinyl-acetate).   heat-flow calibration. The EVA samples are from a solar PV
                                                                         manufacturer. They were cured at a high temperature and
    EVA, a thermal-set material, is a copolymer elastomer supplied
                                                                         pressure for some time. Each EVA sample weighed approxi-
    in sheet form for use in the encapsulation of PV modules.
                                                                         mately 10 mg. Each EVA sample cured at different times
    It has many desirable properties which make it the material
                                                                         was encapsulated in the standard aluminum pans. The DSC
    of choice for this application.
                                                                         program started from -50 ˚C and heated to 220 ˚C at 10 ˚C/min.
    • It is not adhesive at room temperature for easy handling.
                                                                         Results
    • It makes a permanent and adhesive tight seal in the
                                                                         The raw EVA material exhibits several transitions during
      solar-cell system through crosslinking and enhanced
                                                                         heating, as shown in Figure 2 (Page 3). It was heated in the
      bonding when the film is heated and pressed.
                                                                         DSC from -50 ˚C to 220 ˚C at 10 ˚C/min, and after that it
    • After crosslinking, the EVA has high optical transmittance,        was cooled to the starting temperature quickly at 100 ˚C/min.
      good adhesion to the different module materials – it               It was heated for the second time at the same heating rate.
      provides good dielectric properties and great moisture-            The first heating curve shows an endothermic melting peak
      barrier properties with adequate mechanical compliance             (26 J/g) followed by the exothermic curing peak with the
      to accommodate system thermal stresses due to the                  curing enthalphy of 16.6 J/g. The second heating curve
      different thermal-expansion coefficiencies.                        shows a glass transition (Tg) at -35.6 ˚C; the melting peak is
                                                                         smaller (12 J/g vs. 26 J/g) and there is no detectable curing
    During the PV-package process, the EVA sheet is placed               exothermal peak. So by comparing the first heating curve
    between the solar cells and the backsheet/glass. It is heated,       with the second heating curve, it is clear that the EVA raw
    pressed into place, and cured at a certain high temperature          material is cured completely after first heating it up to 220 ˚C.
    for some time. Since the final cured material’s properties are
    largely dependent on the curing degree, it is important to           For a partially cured EVA sample, the residual curing peak
    know the degree of curing of the EVA so that the encapsu-            during the first heating will be between the curing enthalpy
    lation process is optimized. Differential scanning calorimetry       of raw EVA material and zero for completely curing EVA. So
    (DSC) has been traditionally used for curing studies of thermoset    the residual curing enthalpy can be used as an indicator of
    resins. DSC can study the degree of cure and curing kinetics.        the curing degree of EVA material. A series of EVA samples
    In this note, different EVA materials with different curing          with different curing time are studied by DSC and the results
    times were investigated with PerkinElmer‘s high-end DSC 8000.        are shown in Figure 3 (Page 3). The calculated residual curing
                                                                         enthalpy is tabulated in Table 1 and fitted to a straight line
                                                                         in Figure 4 (Page 3). As can be seen, the residual curing
                                                                         enthalpy can be correlated to the curing time very nicely
                                                                         (R2 = 0.9893).

                                                                           Table 1. Residual curing enthalpy of eight different EVA
                                                                           samples with different curing times.
                                                                           EVA        Curing Time      ΔH (residual curing
                                                                           samples       (min)           enthalpy J/g)
                                                                           EVA-1            1                 11.3572
                                                                           EVA-2            2                 10.7635
                                                                           EVA-3            3                 9.6878
                                                                           EVA-4            4                 7.9689
               Figure 1. Scheme of a crystalline solar panel.              EVA-5            5                 7.5885
                                                                           EVA-6            6                 6.7448
                                                                           EVA-7            8                 4.9335
                                                                           EVA-8            9                 3.9811

2
Figure 2. The first (red) and second (blue) heating curve of raw EVA material.




Figure 3. Partially cured EVA samples with different curing times.


                                                                                                                          Conclusion
                                                                                                                          This study shows that DSC can be used to study the curing
                                                                                                                          degree of the EVA resin by measuring the residual curing
                                                                                                                          enthalpy. The data show that the residual curing enthalpy can
                                                                                                                          be correlated to the curing time in a linear way. The DSC test
                                                                                                                          is quick and easy. The double-furnace PerkinElmer DSC 8000
                                                                                                                          delivers accurate heat-flow data with great reproducibility.
                                                                                                                          The power-controlled design ensures great accuracy and
                                                                                                                          true isothermal measure so that it can be used for both
                                                                                                                          scanning-curing and isothermal-curing studies of EVA resin.


Figure 4. The relationship between residual-curing enthalpy and the curing
time of the eight different EVA samples.




PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com



For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.

009038_01
a p p l i c at i o n n o t e


                                               ICP-Mass Spectrometry

                                               Authors
                                               Jianmin Chen
                                               PerkinElmer, Inc.
                                               Shelton, CT 06484 USA




Determination of           Mercury is ubiquitous in nature, and the human health
                           consequences of mercury exposure were recognized from

Mercury in Wastewater      prehistory to the present. The first emperor of unified China
                           who came to power in 221 B.C., Qin Shi Huang, reportedly

by Inductively Coupled     died of ingesting mercury pills that were intended to give
                           him eternal life. The severity of mercury's toxic effects
Plasma-Mass Spectrometry   depends on the form and concentration of mercury and
                           the route of exposure. Although its potential for toxicity in
                           highly contaminated areas such as Minamata Bay in Japan
                           is well documented, research has shown that mercury can
                           be a threat to the health of people and wildlife in many
                           environments that are not obviously polluted. There is no
                           safe level of mercury for humans. The main toxic effects
                           of mercury are known to negatively affect the neurological,
                           renal, cardiovascular and immunological systems.
Mercury exists in three chemical forms: elemental or metallic,
    organic or methylmercury, and inorganic complexes.                  Table 1. ELAN ICP-MS Instrumental Conditions and
                                                                        Experimental Parameters.
    Mercury has thousands of industrial applications. Some
    common uses for mercury include conducting electricity,             RF power                      1100 W
    measuring temperature and pressure, acting as a biocide,            Plasma gas flow               15 L/min
    preservative, and disinfectant, as well as being a catalyst for     Auxiliary gas flow            1.2 L/min
    reactions. Unlike most other pollutants, mercury is highly
                                                                        Nebulizer gas flow            0.96 L/min
    mobile, non-biodegradable, and bio-accumulative; as a
    result, it must be closely monitored to ensure its harmful          Nebulizer                     MEINHARD® Concentric Type A3
    effects on local populations are minimized.1 Thus, measurement      Spray chamber                 Baffled Quartz Cyclonic
    of mercury in environmental samples, and in particular waste-       Scanning mode                 Peak Hopping
    water, is of great importance as a major tool to protect            Dwell time                    50 ms
    the environment from mercury released through emissions
                                                                        Replicates                    3
    from manufacturing, use, or disposal activities. Currently,
    the prominent methods typically utilized by the environmental       Integration time              1 sec/mass
    community for the determination of mercury generally
    require detection limits as low as 0.5 ng/L (ppt, parts-
    per-trillion).2                                                   Sample Preparation
                                                                      The stability of mercury-containing solutions has been
    Traditionally, mercury is analyzed using Cold Vapor Atomic
                                                                      a topic of concern for all trace analysts performing Hg
    Absorption Spectroscopy (CVAAS) or Cold Vapor Atomic
                                                                      determinations. It is reported that a trace amount of gold
    Fluorescence Spectroscopy (CVAFS). Both of these techniques
                                                                      salt added to HNO3 preserved all forms of mercury. The
    are relatively straightforward to use and can accomplish
                                                                      gold ion acts as a strong oxidizing agent that converts or
    the analytical requirements of detection limits in the low
                                                                      maintains mercury as mercuric ion which remains in solution.3
    ppt range. However, they are generally specific for mercury
                                                                      Thus, a solution of 2% (v/v) HNO3 containing 200 ug/L Au
    analysis only.
                                                                      was used for preparation of all samples and standards.
    In recent years, Inductively Coupled Plasma Mass Spectrometry     Two simulated wastewater certified reference materials
    (ICP-MS) has become one of the most powerful analytical           (Trace Metals Solutions, CWW-TM-A and CWW-TM-C,
    techniques for trace element analysis because of its high         High-Purity Standards, Charleston, SC, USA) were prepared
    sensitivity, wide linear dynamic range, and simultaneous          according the manufacturer’s description using the same
    multi-element detection capability. As a result, ICP-MS has       diluent in this study.
    been increasingly adopted in environmental and biomonitoring
    laboratories for the simultaneous measurement of mercury          Calibration
    with other toxic metals since this technique can offer the        External calibration standards of mercury were at the level
    same analytical performance as CVAAS or CVAFS. This               of 10, 20, 50, 100, 200, 500, 1000 ng/L. Figure 1 shows
    application note describes the application of ELAN® ICP-MS        the calibration curve of 202Hg. The correlation coefficient is
    to the determination of mercury in wastewater.                    0.999973, which allowed the accurate quantitative analysis
                                                                      of mercury at the low ppt levels.
    Instrumentation
    For this study, the PerkinElmer® ELAN DRC™ II ICP-MS
    was used for the analysis of wastewater samples under
    standard mode. The ELAN ICP-MS instrument conditions
    and general method parameters are shown in Table 1.




                                                                      Figure 1. External calibration curve of 202Hg. Standard solutions were prepared
                                                                      in 2% HNO3 containing 200 ug/L Au with concentrations ranging from 10 to
                                                                      1000 ng/L.




2
Spike recovery
The spike recovery test was performed using two simulated
wastewater certified reference materials (HPS CWW-TM-A
and CWW-TM-C). The results are presented in Table 2.


   Table 2. Spike recovery test results from two simulated
   wastewater certified reference materials (HPS CWW-TM-A
   and CWW-TM-C).
   Wastewater                  Found               Spike              Measured               Recovery
   Sample                      (ng/L)              Level              (ng/L)                 %
                                                   (ng/L)
   CWW-TM-A                    49                  200                238                    95
   CWW-TM-C                    375                 1000               1370                   99



Memory effect
It is generally viewed that routine determination of mercury
by ICP-MS is affected by a pronounced memory effect in the                                                               Figure 2. Uptake and wash out profile for Hg at 1000 ng/L level. The signal of
sample introduction system. This results in long washout times                                                           four mercury isotopes returns to the background level in around 60 seconds,
for the analyte, which affects the accuracy and reliability of                                                           which is equivalent to the typical rinse time used for many other analytes.
the analytical procedure. To minimize the memory effect, the
system was washed using 2% (v/v) HNO3 containing 200 ug/L
Au between samples. Figure 2 shows the uptake and wash out                                                               Conclusion
profile of Hg. The signal of four mercury isotopes returns to                                                            The ELAN ICP-MS combines high sensitivities and superior
the background level in around 60 seconds, which is equivalent                                                           detection limits with ease of use and high sample throughput.
to the typical rinse time used for many other analytes. Thus,                                                            This allows it to be an excellent tool for trace mercury
the results indicated application of gold salt is effective in                                                           determination in wastewater samples. Mercury can be detected
preventing mercury sorption and vapor buildup within the                                                                 simultaneously with the other trace elements of interest within
sample introduction system.                                                                                              a single sample run.


                                                                                                                         References
                                                                                                                         1. J.O. Nriagu, Global inventory of natural and anthropogenic
                                                                                                                            emissions of trace metals to the atmosphere, Nature 279
                                                                                                                            (1979) 409–411.

                                                                                                                         2. EPA Method 1631. Mercury in Water by Oxidation,
                                                                                                                            Purge and Trap, and Cold Vapor Atomic Fluorescence
                                                                                                                            Spectrometry.

                                                                                                                         3. www.epa.gov/nerlesd1/factsheets/mpt.pdf. Mercury
                                                                                                                            Preservation Techniques.




PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com



For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2009-2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.

008973A_01
a p p l i c at i o n n o t e

                                                              Thermogravimetric Analysis –
                                                              GC Mass Spectrometry

                                                              Author
                                                              Greg Johnson
                                                              PerkinElmer, Inc.
                                                              Shelton, CT 06484 USA




Qualitative Analysis                          Introduction
                                              Thermogravimetric analysis (TGA)
of Evolved Gases in                           measures the change in the weight of
                                              a sample as a function of temperature.
Thermogravimetry by                           A limitation of TGA is that it cannot identify

Gas Chromatography/                           what material is lost at a specific tempera-
                                              ture. The analysis of gases evolved during

Mass Spectrometry                             a TGA experiment by gas chromatography
                                              mass spectrometry (GC/MS) provides
                                              laboratories with a way to identify the
                                              compound or groups of compounds
                                              evolved during a specific weight-loss
                                              event in a TGA analysis.

       This application note discusses the utility of TG-GC/MS with an example
       application – the identification of specific organic acids evolved during TGA
       analysis of switchgrass.
Figure 3 describes in greater detail the pneumatic supply
                                                                            to the S-Swafer device and assists in explaining why this
                                                                            approach is so well suited to interfacing the Pyris 1 TGA
                                                                            to the Clarus 600 GC/MS.




    Figure 1. Clarus 600 GC/MS interfaced to the Pyris 1 TGA.




    Switchgrass (Panicum Irgatum) is a perennial warm-season
    grass native to the northern states of the USA; it is easily
    grown in difficult soils. Switchgrass is potentially useful in          Figure 3. Schematic showing carrier-gas supply to the S-Swafer device.
    the production of biofuels, specifically cellulosic ethanol
    and bio-oil.

    The instrumentation used in this study was a PerkinElmer®               The S-Swafer configuration is optimal because it will ensure
    Pyris™ 1 TGA interfaced to the PerkinElmer Clarus® 600 GC/              a very rapid adjustment of carrier gas to the Swafer device,
    MS with the S-Swafer™ micro-channel flow splitting device               allowing for a rapid switch between backflush of the transfer
    (S4 configuration). The preferred mode of operation of the              line and sampling from the TGA. The samples of evolved gas
    TGA maintains the atmosphere around the sample at ambi-                 are collected by setting a simple parameter in the GC method;
    ent atmospheric pressure. The sample is collected from the              multiple samples can be collected during a TGA analysis.
    TGA by allowing the high vacuum of the MS to create a                   Additionally, backflush of the transfer line will isolate the
    pressure drop across the GC column, causing a flow of gas               GC/MS and enable purge gas at the TGA to be switched
    from the TGA through the transfer line and the analytical               from an inert gas (during analysis) to an air supply for
    column to the MS. During the analysis, there are times when             cleaning of the TGA pan prior to the next sample.
    the TGA inlet will be surrounded by air, rather than an inert
    atmosphere. This would cause air to flow into the GC/MS; this           Oven subambient cooling will be extremely useful in this
    is undesirable as it will cause oxidation to a number of dif-           application, allowing protracted sampling periods to be
    ferent areas of the system. The S-Swafer device (shown in               refocused into a narrow band of analytes on the column.
    Figure 2) is used to switch between backflushing of the
    TGA transfer line during non-sampling time and sampling                 Experimental
    of the TGA environment during analysis.                                 The deactivated fused-silica transfer line used here was
                                                                            1.6 m x 0.32 mm i.d. A few centimeters of the deactivated
                                                                            fused silica protrudes into the sample weighing area of the
                                                                            TGA. Approximately 30 cm of fused silica passes through
                                                                            the injector into the oven environment and is connected to
                                                                            the S-Swafer using specialized SilTite™ nuts and ferrules to
                                                                            ensure a leak-free connection that will not shrink and leak
                                                                            during normal or even extended thermal cycling of the main
                                                                            oven.

                                                                            In all cases, a 30 m x 0.32 mm analytical column was
                                                                            employed as this allows a carrier flow of approximately
                                                                            1 mL/min with the fixed 1.00 atmosphere pressure drop
                                                                            from ambient at the TGA to vacuum at the MS. Data was
    Figure 2. Schematic showing the pneumatic interfacing of the TG-GC/MS   acquired using an Elite™ WAX stationary phase.
    using the S-Swafer.




2
A small quantity of dried and ground switchgrass was              timed events that will be used to sample the evolved gases
placed on the TGA pan and weighed using Pyris software.           onto the GC/MS column. Note that the TGA is held iso-
A rapid TGA analysis based on heating the sample from             thermal for the first 5.0 min at which point heating begins.
30 ˚C to 1000 ˚C at 100 ˚C/min in a nitrogen atmosphere           Simultaneously, the GC/MS analysis is started.
was performed to determine which regions of the weight-
loss curve were to be further studied using the TG-GC/MS          Figure 5 illustrates the TG-GC/MS analysis of the switch-
technique.                                                        grass based on timed events that collect the evolved gases
                                                                  from the main transition shown in Figure 4. The smaller
The primary reason for using such rapid heating, which            earlier transition, also seen in the same figure, was also sampled
reduces the resolution of the weight-loss curve produced          onto the GC/MS but preliminary findings indicate that this
by the TGA, is to transfer the evolved gas quickly into the       is simply evolved water. The major transition produced large
GC column. A quick transfer will improve GC peak shape,           numbers of oxygenated volatile organic compounds (VOCs),
sensitivity and resolution.                                       including some very polar species. Earlier work using a non-
                                                                  polar capillary column had generated extremely smeared-out
After the sample was loaded onto the TGA and the furnace          early-eluting peaks. The chromatogram below was generated
raised, the analysis was started immediately. The first step in   using a thick-film polar Elite WAX column.
the TGA heating program maintained the low initial furnace
temperature for 5 to 10 min. During this time, the furnace
environment is being purged with helium (or nitrogen/argon),
and the carrier-gas pressure of 7.0 psig maintained at Aux 1
(Figures 2 and 3) ensures that no sample can enter the ana-
lytical column. After this initial hold period, the TGA furnace
begins to heat the sample, and simultaneously, the GC/MS
run is started using an external start command.
                                                                  Figure 5. TG-GC/MS analysis of the switchgrass sample on a 30 m x 0.32 mm
Based on previous TGA runs on the same sample, timed
                                                                  x 1 μm Elite WAX column.
events within the GC method switch off the carrier gas
supplied by the Aux 1 PPC module and then close the
solenoid valve (SV3) shown in Figure 2 (Page 2). This begins      The three peaks labeled with asterisks in Figure 5 are identified
the sampling and this procedure is reversed to bring the sam-     as a homologous series of free fatty acids (Figure 6 – Page 4),
pling period to an end. After the sampling is complete, both      based on a library search of their MS spectra (NIST® 2008).
the GC oven-temperature program and MS data acquisition           Usually in GC, a homologous series tends to elute in carbon-
begins. The TGA can now be programmed to switch purge             number order but here, the elution order appears to be
gases to clean the system using oxidation at elevated             acetic, followed by formic, followed by propanonic acid. As
temperature, prior to the next analysis.                          this retention behavior is not typical and in the absence of a
                                                                  literature reference or a similar chromatogram in the public
A typical TGA weight-loss curve for the switchgrass is shown
                                                                  domain, it seemed prudent to analyze a simple retention-
in Figure 4 and reveals a typical weight % loss curve for the
                                                                  time standard to confirm this tentative result. Figure 7 (Page 4)
sample of switchgrass that was tested. In addition, super-
                                                                  shows the same analysis again but with the retention-time
imposed on the weight-loss curve is the derivative of this
                                                                  standard shown in parallel. This standard was a simple
curve which greatly assists the analyst in setting up the GC
                                                                  mixture diluted in water with a small (5 μL) aliquot of this
                                                                  aqueous solution deposited by syringe onto the TGA pan
                                                                  for analysis.




Figure 4. Typical result for the TGA analysis of switchgrass.

                                                                                                                                              3
Conclusions
                                                                                                                              In this application note, we describe the technique of TG-
                                                                                                                              GC/MS through the analysis of switchgrass. TG-GC/MS is
                                                                                                                              demonstrated to be a valuable technique in the separation
                                                                                                                              and identification of complex mixtures of gas evolved during
                                                                                                                              a TGA analysis. The S-Swafer device is demonstrated as a
                                                                                                                              means to interface a TGA to GC/MS. The main benefits are
                                                                                                                              its simplicity and the inertness of the entire sample path.




Figure 6. Mass spectra extracted from the total ion chromatogram of the
switchgrass sample. The spectral data matches that of acetic, formic and
propionic acid respectively available in the 2008 NIST® mass spectral library.




Figure 7. TG-GC/MS analysis of the switchgrass sample (bottom) on a 30 m
x 0.32 mm x 1 μm Elite WAX column and the analysis of a simple retention
time mix (top) under the same TGA and GC conditions.




PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com



For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.

009030_01
a p p l i c at i o n n o t e


                                                                                  ICP-Mass Spectrometry


                                                                                  Authors
                                                                                  Lee Davidowski, Ph.D.

                                                                                  Zoe Grosser, Ph.D.

                                                                                  Laura Thompson

                                                                                  PerkinElmer, Inc.
                                                                                  710 Bridgeport Avenue
                                                                                  Shelton, CT USA




The Determination                          Introduction
                                           Dietary supplements are regulated by the FDA under the general umbrella of foods,
of Metals in Dietary                       but with different regulations than conventional foods. Dietary supplements have been
                                           defined by Congress as materials taken by mouth that include ingredients intended to
Supplements                                provide dietary supplementation. They can be found in various forms, including tablets,
                                           powders, and liquids. They may consist of vitamins, minerals, herbs or other botanicals,
                                           amino acids, and substances such as enzymes, organ tissues, glandulars, and metabolites.
                                           Under the Dietary Supplement Health and Education Act of 1994 (DSHEA), the dietary
                                           supplement manufacturer is responsible for ensuring that a dietary supplement is safe
                                           before it is marketed. FDA is responsible for taking action against any unsafe dietary
                                           supplement product after it reaches the market.1

     One facet of ensuring a safe product is analysis of the final product before distribution. Although organic components often
     form the majority of a supplement, metals may be found due to their inclusion in vitamin structures such as vitamin B-12
     (cobalt) and minerals, such as selenium. They may also be added as a contaminant through natural products or in the manu-
     facturing process. The measurement of toxic metals or metals intended to be present for quality and labeling confirmation
     may be required. Analysis challenges include measurement at low concentrations in a variety of matrices. In this note we
     will use inductively coupled plasma mass spectrometry (ICP-MS) to measure a variety of elements generally considered to
     be hazardous to human health at low to medium concentrations. The four elements generally considered to be hazardous
     and not necessary for nutrition are Pb, Cd, As, and Hg. The elements Se and Cr are often added at low concentrations for
     nutritional purposes and were also included.
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3
Spotlight on Analytical Applications Complete e-Zine - Volume 3

Contenu connexe

En vedette

FR-Méga Nutrition Super Aliment Biologique_Final
FR-Méga Nutrition Super Aliment Biologique_FinalFR-Méga Nutrition Super Aliment Biologique_Final
FR-Méga Nutrition Super Aliment Biologique_Final
Sabine FAURE SA Mlle
 

En vedette (12)

Application Note: Study Rigid Amorphous Fraction in Polymer Nano-Composites b...
Application Note: Study Rigid Amorphous Fraction in Polymer Nano-Composites b...Application Note: Study Rigid Amorphous Fraction in Polymer Nano-Composites b...
Application Note: Study Rigid Amorphous Fraction in Polymer Nano-Composites b...
 
Quantification of Bromide by APHA 4500 Method
Quantification of Bromide by APHA 4500 MethodQuantification of Bromide by APHA 4500 Method
Quantification of Bromide by APHA 4500 Method
 
Water Analysis using LAMBDA Spectrophotometer
Water Analysis using LAMBDA SpectrophotometerWater Analysis using LAMBDA Spectrophotometer
Water Analysis using LAMBDA Spectrophotometer
 
Spotlight on Analytical Applications e-Zine - Volume 9
Spotlight on Analytical Applications e-Zine - Volume 9Spotlight on Analytical Applications e-Zine - Volume 9
Spotlight on Analytical Applications e-Zine - Volume 9
 
Benefits of NexION 300D ICP-MS reaction mode in removing the Gd interference ...
Benefits of NexION 300D ICP-MS reaction mode in removing the Gd interference ...Benefits of NexION 300D ICP-MS reaction mode in removing the Gd interference ...
Benefits of NexION 300D ICP-MS reaction mode in removing the Gd interference ...
 
Water Analysis Using LAMBDA UV/Visible Spectrophotometers: Ammonia-Nitrogen D...
Water Analysis Using LAMBDA UV/Visible Spectrophotometers: Ammonia-Nitrogen D...Water Analysis Using LAMBDA UV/Visible Spectrophotometers: Ammonia-Nitrogen D...
Water Analysis Using LAMBDA UV/Visible Spectrophotometers: Ammonia-Nitrogen D...
 
AxION DSA System: A Reference Notebook of DSA Applications--First Edition, 2012
AxION DSA System: A Reference Notebook of DSA Applications--First Edition, 2012AxION DSA System: A Reference Notebook of DSA Applications--First Edition, 2012
AxION DSA System: A Reference Notebook of DSA Applications--First Edition, 2012
 
The Determination of Total Protein Using the LAMBDA UV/Vis Spectrophotometer:...
The Determination of Total Protein Using the LAMBDA UV/Vis Spectrophotometer:...The Determination of Total Protein Using the LAMBDA UV/Vis Spectrophotometer:...
The Determination of Total Protein Using the LAMBDA UV/Vis Spectrophotometer:...
 
Determination of Cholesterol Level in Human Serum
Determination of Cholesterol Level in Human SerumDetermination of Cholesterol Level in Human Serum
Determination of Cholesterol Level in Human Serum
 
Spotlight on Analytical Applications Complete e-Zine Vol. 1
Spotlight on Analytical Applications Complete e-Zine Vol. 1Spotlight on Analytical Applications Complete e-Zine Vol. 1
Spotlight on Analytical Applications Complete e-Zine Vol. 1
 
PerkinElmer Spotlight on Analytical Applications e-Zine – Volume 16
PerkinElmer Spotlight on Analytical Applications e-Zine – Volume 16PerkinElmer Spotlight on Analytical Applications e-Zine – Volume 16
PerkinElmer Spotlight on Analytical Applications e-Zine – Volume 16
 
FR-Méga Nutrition Super Aliment Biologique_Final
FR-Méga Nutrition Super Aliment Biologique_FinalFR-Méga Nutrition Super Aliment Biologique_Final
FR-Méga Nutrition Super Aliment Biologique_Final
 

Similaire à Spotlight on Analytical Applications Complete e-Zine - Volume 3

An Experimental Method for Micro-Particle Ammonium Nitrate (MP-AN)_FINAL
An Experimental Method for Micro-Particle Ammonium Nitrate (MP-AN)_FINALAn Experimental Method for Micro-Particle Ammonium Nitrate (MP-AN)_FINAL
An Experimental Method for Micro-Particle Ammonium Nitrate (MP-AN)_FINAL
Stroma Service Consulting
 
Experimental investigate to obtain the effectiveness of regenerator using Air.
Experimental investigate to obtain the effectiveness of regenerator using Air.Experimental investigate to obtain the effectiveness of regenerator using Air.
Experimental investigate to obtain the effectiveness of regenerator using Air.
IJESFT
 
1.2 Migration Rates of Solute pdf version.pdf
1.2 Migration Rates of Solute pdf version.pdf1.2 Migration Rates of Solute pdf version.pdf
1.2 Migration Rates of Solute pdf version.pdf
AhmadHafizuddin20
 

Similaire à Spotlight on Analytical Applications Complete e-Zine - Volume 3 (20)

Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
An autoignition performance comparison of chemical kinetics models for n-heptane
An autoignition performance comparison of chemical kinetics models for n-heptaneAn autoignition performance comparison of chemical kinetics models for n-heptane
An autoignition performance comparison of chemical kinetics models for n-heptane
 
PLC Based Instrumentation of Closed Loop Gas System for RPC Detectors
PLC Based Instrumentation of Closed Loop Gas System for RPC DetectorsPLC Based Instrumentation of Closed Loop Gas System for RPC Detectors
PLC Based Instrumentation of Closed Loop Gas System for RPC Detectors
 
An Experimental Method for Micro-Particle Ammonium Nitrate (MP-AN)_FINAL
An Experimental Method for Micro-Particle Ammonium Nitrate (MP-AN)_FINALAn Experimental Method for Micro-Particle Ammonium Nitrate (MP-AN)_FINAL
An Experimental Method for Micro-Particle Ammonium Nitrate (MP-AN)_FINAL
 
Calibration of analytical instruments
Calibration of analytical instrumentsCalibration of analytical instruments
Calibration of analytical instruments
 
NASA_AF-M315E
NASA_AF-M315ENASA_AF-M315E
NASA_AF-M315E
 
Scintillations principle, working, merits & demerits & applications
Scintillations  principle, working, merits & demerits & applicationsScintillations  principle, working, merits & demerits & applications
Scintillations principle, working, merits & demerits & applications
 
Determination of Residual Solvents in Flexible Packaging
Determination of Residual Solvents in Flexible PackagingDetermination of Residual Solvents in Flexible Packaging
Determination of Residual Solvents in Flexible Packaging
 
Triplus500 HS working.pptx
Triplus500 HS working.pptxTriplus500 HS working.pptx
Triplus500 HS working.pptx
 
Experimental investigate to obtain the effectiveness of regenerator using Air.
Experimental investigate to obtain the effectiveness of regenerator using Air.Experimental investigate to obtain the effectiveness of regenerator using Air.
Experimental investigate to obtain the effectiveness of regenerator using Air.
 
An experimental method for micro particle ammonium nitrate (mp-an)
An experimental method for micro   particle ammonium nitrate (mp-an)An experimental method for micro   particle ammonium nitrate (mp-an)
An experimental method for micro particle ammonium nitrate (mp-an)
 
PyTeCK: A Python-based automatic testing package for chemical kinetic models
PyTeCK: A Python-based automatic testing package for chemical kinetic modelsPyTeCK: A Python-based automatic testing package for chemical kinetic models
PyTeCK: A Python-based automatic testing package for chemical kinetic models
 
Determination of Oxygen in Anhydrous Ammonia
Determination of Oxygen in Anhydrous AmmoniaDetermination of Oxygen in Anhydrous Ammonia
Determination of Oxygen in Anhydrous Ammonia
 
Sop125v3 1
Sop125v3 1Sop125v3 1
Sop125v3 1
 
Gas permeater
Gas permeaterGas permeater
Gas permeater
 
1.2 Migration Rates of Solute pdf version.pdf
1.2 Migration Rates of Solute pdf version.pdf1.2 Migration Rates of Solute pdf version.pdf
1.2 Migration Rates of Solute pdf version.pdf
 
Example: simulation of the Chlorotoluene chloration with BatchReactor software
Example: simulation of the Chlorotoluene chloration with BatchReactor softwareExample: simulation of the Chlorotoluene chloration with BatchReactor software
Example: simulation of the Chlorotoluene chloration with BatchReactor software
 

Plus de PerkinElmer, Inc.

Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
PerkinElmer, Inc.
 

Plus de PerkinElmer, Inc. (20)

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º Trimestre
 
É A VEZ DO PRIME
É A VEZ DO PRIMEÉ A VEZ DO PRIME
É A VEZ DO PRIME
 
TREC PARA A VIDA
TREC PARA A VIDATREC PARA A VIDA
TREC PARA A VIDA
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o Desempenho
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no Brasil
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGS
 
O que é X Frágil?
O que é X Frágil?O que é X Frágil?
O que é X Frágil?
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedor
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
 
CG Clarus® All In One
CG Clarus® All In OneCG Clarus® All In One
CG Clarus® All In One
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
 

Dernier

IAC 2024 - IA Fast Track to Search Focused AI Solutions
IAC 2024 - IA Fast Track to Search Focused AI SolutionsIAC 2024 - IA Fast Track to Search Focused AI Solutions
IAC 2024 - IA Fast Track to Search Focused AI Solutions
Enterprise Knowledge
 
Histor y of HAM Radio presentation slide
Histor y of HAM Radio presentation slideHistor y of HAM Radio presentation slide
Histor y of HAM Radio presentation slide
vu2urc
 
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptxEIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
Earley Information Science
 

Dernier (20)

IAC 2024 - IA Fast Track to Search Focused AI Solutions
IAC 2024 - IA Fast Track to Search Focused AI SolutionsIAC 2024 - IA Fast Track to Search Focused AI Solutions
IAC 2024 - IA Fast Track to Search Focused AI Solutions
 
The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024
 
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemkeProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
 
Exploring the Future Potential of AI-Enabled Smartphone Processors
Exploring the Future Potential of AI-Enabled Smartphone ProcessorsExploring the Future Potential of AI-Enabled Smartphone Processors
Exploring the Future Potential of AI-Enabled Smartphone Processors
 
Data Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonData Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt Robison
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
 
From Event to Action: Accelerate Your Decision Making with Real-Time Automation
From Event to Action: Accelerate Your Decision Making with Real-Time AutomationFrom Event to Action: Accelerate Your Decision Making with Real-Time Automation
From Event to Action: Accelerate Your Decision Making with Real-Time Automation
 
🐬 The future of MySQL is Postgres 🐘
🐬  The future of MySQL is Postgres   🐘🐬  The future of MySQL is Postgres   🐘
🐬 The future of MySQL is Postgres 🐘
 
Handwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed textsHandwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed texts
 
Histor y of HAM Radio presentation slide
Histor y of HAM Radio presentation slideHistor y of HAM Radio presentation slide
Histor y of HAM Radio presentation slide
 
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
 
Boost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityBoost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivity
 
Tech Trends Report 2024 Future Today Institute.pdf
Tech Trends Report 2024 Future Today Institute.pdfTech Trends Report 2024 Future Today Institute.pdf
Tech Trends Report 2024 Future Today Institute.pdf
 
08448380779 Call Girls In Greater Kailash - I Women Seeking Men
08448380779 Call Girls In Greater Kailash - I Women Seeking Men08448380779 Call Girls In Greater Kailash - I Women Seeking Men
08448380779 Call Girls In Greater Kailash - I Women Seeking Men
 
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptxEIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
 
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot TakeoffStrategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
 
Understanding Discord NSFW Servers A Guide for Responsible Users.pdf
Understanding Discord NSFW Servers A Guide for Responsible Users.pdfUnderstanding Discord NSFW Servers A Guide for Responsible Users.pdf
Understanding Discord NSFW Servers A Guide for Responsible Users.pdf
 
[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf
 
Bajaj Allianz Life Insurance Company - Insurer Innovation Award 2024
Bajaj Allianz Life Insurance Company - Insurer Innovation Award 2024Bajaj Allianz Life Insurance Company - Insurer Innovation Award 2024
Bajaj Allianz Life Insurance Company - Insurer Innovation Award 2024
 
A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)
 

Spotlight on Analytical Applications Complete e-Zine - Volume 3

  • 1. SPOTLIGHT ON APPLICATIONS. FOR A BETTER TOMORROW. VOLUME 3
  • 2. INTRODUCTION PerkinElmer Spotlight on Applications e-Zine – Volume 3 PerkinElmer knows that the right training, methods, applications, reporting and support are as integral to getting answers as the instrumentation. That’s why PerkinElmer has developed a novel approach to meet the challenges that today’s labs face – that approach is called EcoAnalytix™, delivering to you complete solutions for your applications challenges. In this effort, we are pleased to share with you our Spotlight on Applications e-zine, delivering a variety of topics that address the pressing issues and analysis challenges you may face in your application areas today. Our Spotlight on Applications e-zine consists of a broad range of applications you’ll be able to access at your convenience. Each application in the table of contents includes an embedded link that will bring you directly to the appropriate page within the e-zine. PerkinElmer
  • 3. CONTENTS Energy • Determination of Methanol in Crude Oils According to ASTM D7059-04 Using the Clarus 680 GC with S-Swafer Micro-Channel Flow Technology • Determination of Low-Level Oxygenated Compounds in Gasoline Using the Clarus 680 GC with S-Swafer Micro-Channel Flow Technology • Fatty Acid Methyl Ester Contamination of Aviation Fuel by GC/MS • Curing Determination of EVA for Solar Panel Application by DSC Environmental • Determination of Mercury in Wastewater by Inductively Coupled Plasma-Mass Spectrometry • Qualitative Analysis of Evolved Gases in Thermogravimetry by Gas Chromatography/Mass Spectrometry Food & Beverage • The Determination of Metals in Dietary Supplements • Solid Phase Extraction and GC/MS Analysis of Melamine Adulteration in Dairy Products Pharmaceuticals & Nutraceuticals • Fast Analysis of Fat-Soluble Vitamins Using Flexar FX-10 UHPLC and Chromera CDS • Assuring Safety of Traditional Chinese Herbal Medicines by Monitoring Inorganic Impurities using ICP-MS • Polymorphism in Acetaminophen Studied by Simultaneous DSC and Raman Spectroscopy • Pressure-Balanced Headspace for the Determination of Class I, II and III Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology PerkinElmer
  • 4. a p p l i c at i o n n o t e Gas Chromatography Author Andrew Tipler PerkinElmer, Inc. Shelton, CT 06484 USA Determination of Introduction The gas chromatographic (GC) analysis of crude oil Methanol in Crude Oils is a challenging undertaking. Samples are viscous, making them difficult to handle, and they contain According to ASTM hundreds of different compounds with carbon numbers up to or even above C120, making a complete D7059-04 Using the chromatographic separation effectively impossible. Clarus 680 GC with ASTM® D7059-04 is an established method that has been well validated for the determination of methanol S-Swafer Micro- in crude oils. The method lists five variants of the instru- mentation that have demonstrated compliance with Channel Flow this method. The method also allows for alternative Technology configurations that will meet the required performance criteria. In this application note, a method based on a PerkinElmer® Clarus® 680 GC with an S-Swafer™ micro-channel flow splitting device is described; the data presented here will demonstrate that this method complies with the requirements of ASTM® D7059-04.
  • 5. Experimental will enter the first column, leaving the heavier crude-oil The Clarus 680 GC used in this application note is described compounds in the liner. The two alcohols enter the first in Figure 1 with a diagram, and Figure 2 with a photograph non-polar column and elute early in the chromatography. of the S-Swafer micro-channel flow splitting device used to A chromatogram of a standard mixture is shown in Figure 3. perform this analysis. This two-column backflushing configura- tion (designated as S6 in the Swafer documentation) enables the first column to be backflushed while the analytes are still being chromatographed on the second column. A restrictor tube is also connected to one of the S-Swafer outlets to enable the carrier-gas flow rate to be increased and to allow the chromatography to be monitored on the first column by connecting the restrictor outlet to the FID. Nitrogen is used as the carrier gas throughout this application – it is well suited for use with 0.530 mm i.d. columns. Nitrogen, when compared to helium, is less expensive, more available, and not in limited supply. The use of nitrogen is consistent with PerkinElmer initiatives to reduce the use of the declining global stocks of helium. Figure 3. Chromatogram of standard mixture on first column with the The crude-oil sample is diluted 50:50 with clean toluene restrictor tube connected to the FID. solvent containing 1-propanol internal standard to produce a final concentration of 500 ppm. 1.0 µL of the diluted sample is injected into the programmable split/splitless (PSS) From Figure 3, it can be seen that the last peak of interest, injector which has the liner temperature set to 125 °C. the 1-propanol internal standard, elutes at about 3.2 minutes. At this temperature, only the volatile fraction of the sample Anything that elutes later than this time is of no analytical interest and backflushing should commence soon after elution of the 1-propanol peak – in this case 3.3 minutes. The backflushing process occurs when the pressure at the column inlet is less than that at the column outlet. This can be achieved by reducing the first-column inlet pressure at the PSS injector, increasing the (midpoint) pressure at the S-Swafer or doing both. In this analysis, we want to continue chromatography on the second column while we backflush the first column so the only option is to reduce the pressure at the injector. To enable a large backpressure to be used, the second column has an inline restrictor Figure 1. The S-Swafer system used to determine methanol in crude oil. connected between it and the S-Swafer. This enables the midpoint pressure to be increased, yet still allows reasonable flow rates to be applied to the second column. The reduction in the inlet pressure is affected through the use of a simple GC timed event. When the backflushing commences after a crude-oil sample has been injected, the heavier fraction of the sample will still reside within the injector liner which has been held at 125 °C. At this point, the PSS liner is temperature programmed to a high temperature to vaporize this less-volatile material. Because the column is being backflushed, none of this vapor will enter the column, but will be flushed out of the system through the split vent. In this way, removal of the heavy compounds is very efficient and doesn’t expose the columns to this material, thus prolonging the column life. Figure 2. Photograph of installed system showing the S-Swafer connections. 2
  • 6. Figure 4 shows a chromatogram of a standard solution on the backflushed until the oven reaches its initial programmed first column with backflushing applied at 3.3 minutes. The temperature. The initial split-flow rate is set to 100 mL/min chromatography is now very clean beyond the 1-propanol peak. to ensure that the pre-run pressure change equilibrates quickly. The split flow is reset to 10 mL/min by a pre-run event at -0.50 minutes, which occurs once the pressure has stabilized. These split-flow changes serve only to save time. Table 1. Analytical Conditions for the Determination of Methanol in Crude Oil. Gas Chromatograph PerkinElmer Clarus 680 GC Oven Temperature 125 °C for 1 minute, then 25 °C/min to 250 °C Injector Programmable Split/Splitless (PSS) Injector Temperature 125 °C for 3.3 minutes, then 200 °C/min to 400 °C and hold until the end of the run Carrier Gas Nitrogen Figure 4. Chromatogram of standard mixture on first column with backflushing Initial Injector Pressure at 3.3 minutes with the restrictor tube connected to the FID. Setpoint 2 psig (see text) Initial Injector Split Figure 5 shows chromatograms from the CP-Lowox column ® Flow Rate 100 mL/min (see text) of three calibration mixtures that cover the calibration range Detector Flame Ionization (FID) of this method. Again, the chromatography looks very clean. Detector Temperature 325 °C Detector Combustion Gases Air: 450 mL/min, Hydrogen: 45 mL/min Detector Range x1 Detector Attenuation x4 Backflush System S-Swafer configured in S6 mode Precolumn 30 m x 0.530 mm x 5 µm PerkinElmer Elite™ 1 with 25 cm x 0.250 mm deactivated fused silica restrictor connected between S-Swafer and column Analytical Column 10 m x 0.530 mm x 10 µm Varian® CP-Lowox® with in-line 25 cm x 0.100 µm deactivated fused silica restrictor connected between S-Swafer and column Restrictor Tubing Figure 5. Chromatography of three methanol standard solutions containing between S-Swafer ~500 ppm w/w 1-propanol internal standard. and Detector 30 cm x 0.100 µm deactivated fused silica (Midpoint) Pressure The full method for this analysis is given in Table 1. The use at S-Swafer 20 psig of timed events in this method needs explanation. The GC Timed Events (see text) PSS pressure set to 24 psig at -1.00 min oven needs to be programmed up to 250 °C in order to PSS split flow set to 10 mL/min at -0.50 min elute the analytes from the CP-Lowox® column. At this PSS pressure set to 2 psig at 3.30 min temperature, there is some slight stationary-phase bleed from PSS split flow set to 100 mL/min at 3.31 min the thick-film methyl silicone precolumn. The precolumn remains in the backflush mode during cooling to prevent any Sample Preparation 5 g of crude-oil sample mixed with 5 g of toluene containing 1-propanol internal of the bleed from the precolumn at the onset of cooling from standard to deliver a final concentration entering the CP-Lowox® column. The initial injector pressure of 500 ppm w/w of 2 psig and the pre-run timed event to set the pressure Sample Injection Normal injection of 1.0 µL of prepared to 24 psig at -1.00 minutes ensures that the pre-column is sample using an autosampler 3
  • 7. Discussion What should also be noted is that the total chromatography The calibration plot obtained for the mixtures of methanol time is just six minutes. For most of this time, the injector and 1-propanol in toluene are shown in Figure 6. This plot liner is being heated at a high temperature and the precolumn demonstrates good linearity which is in compliance with is being backflushed at a very high flow rate, while the oven ASTM® D7059. is being temperature programmed. This means that, at the end of the run, all of the less-volatile sample compounds have been flushed from the system. All that is then necessary is to cool the oven, reset the inlet pressure and inject the next sample. The Clarus 680 GC used for this analysis has rapid oven cooling and starts to load the autosampler with sample prior to the GC coming ready and so sample throughput exceeds six samples per hour. Table 2 shows the quantitative precision obtained from ten replicate injections of a high-level and a low-level check standard. Considering the complexity of the sample matrix, these results demonstrate the efficacy of the Swafer tech- nology for this type of application. These results greatly exceed the requirements of ASTM® D7059. Figure 6. Calibration plot showing response ratio vs. amount ratio for methanol:1-propanol internal standard. Table 2. Quantitative precision of low-level and high-level check standards prepared using a sample of light crude oil. Figure 7 shows a chromatogram of a low-level check standard Run Results for Results for prepared from a sample of light crude oil. The hydrocarbons Check Standard A Check Standard B that elute with the alcohols into the second column before 28.2 ppm w/w 1096 ppm w/w backflushing is initiated are lightly retained on the CP-Lowox® 1 33.2 1063 column and so quickly elute in the chromatography. The 2 32.2 1063 alcohols are much more strongly retained and require an extended temperature program to elute and are well separated 3 32.7 1065 from the light hydrocarbons in the crude oil. Note the clean 4 32.6 1059 baseline around the alcohols, which facilitates precise quan- 5 32.9 1059 tification. 6 35.6 1062 7 33.4 1065 8 33.9 1068 9 33.6 1061 10 33.8 1064 Relative Std Dev % 2.88 0.26 The analytical system was validated using a set of 20 ‘round- robin’ samples prepared by Spectrum Analytical Standards, Sugarland, Texas. These are shown in Figure 8. These sam- ples were prepared five years ago to validate the published ASTM® D7059-04 method. It is believed that these samples are stable and would provide a good basis for validating this Figure 7. Chromatogram of 25 ppm w/w check standard prepared in a sample Swafer system for compliance with the ASTM® method. of light crude oil. 4
  • 8. One of the concerns regarding the GC analysis of crude oil is that non-volatile sample residue will accumulate in the system giving rise to adsorption, thermolysis or carryover effects. This method uses a combination of low injection temperature and column backflushing to keep heavier sample compounds out of the columns. Figure 9 shows the chromatography of crude-oil samples soon after a new liner and septum have been installed compared against the chro- matography after over 150 crude-oil injections have been made into the same liner and septum. Clearly, sample residue will accumulate as the number of injections increases (particu- larly with heavy crude oils) and eventually the liner and the septum will need to be changed. Figure 9 shows that after 150 injections, this method continues to perform. Figure 8. ASTM® D7059 round-robin validation samples. Table 3 summarizes the results from the round-robin sample set. All are within the ASTM® D7059 limits except the result for Sample 19, which is just outside the accepted range. Given the age of these samples, these results are considered acceptable, and the method does tolerate one result deviation out of twenty. Table 3. Results obtained for ASTM® D7059 round-robin validation samples. Sample Experimental Expected 1 2.0 2.7 Figure 9. Chromatograms of the same crude-oil sample spiked with 400 ppm 2 4.3 5.8 w/w methanol produced using a single injector liner showing no significant degradation in chromatography after over 150 runs. 3 7.4 11.7 4 11.4 14.9 5 7.8 12.9 6 29.8 35.1 7 23.8 27.6 8 31.2 35.7 9 58.1 66.3 10 60.4 67.0 11 69.5 74.5 12 105.5 113.4 13 113.4 133.2 14 265.3 285.2 15 393.8 404.7 16 542.8 592.9 17 596.3 524.2 18 873.2 912.2 19 739.8 826.5 20 709.9 734.7 5
  • 9. Conclusion This method, based on the S-Swafer technology, fully complies with the requirements of ASTM® D7059-04 and offers the user a number of unique benefits: • The method presented here will provide very high throughput and fast delivery of results; with a complete analysis cycle time less than ten minutes per sample (more than six samples/hour throughput) • Reduced instrument maintenance as a result of temperature- programmed injection which prevents heavy sample compounds from entering and degrading columns • Additional protection of the analytical columns and reduction of run time with the use of backflushing • Single-detector configuration simplifies and reduces the cost of the system • Splitting restrictor enables precolumn chromatography to be monitored, facilitating the setup of backflush timing • Nitrogen carrier gas reduces costs associated with carrier gas and serves to help preserve helium stocks • Quantitative performance exceeding the method specification delivers results that will meet laboratory quality-control specifications. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 009027A_01
  • 10. a p p l i c at i o n n o t e Gas Chromatography Author Andrew Tipler PerkinElmer, Inc. Shelton, CT 06484 USA Determination of Introduction The existing ASTM® D4815 method is designed to monitor Low-Level Oxygenated oxygenated compounds in gasoline at percentage concen- trations. The method described in this application note is Compounds in Gasoline intended to enable these analytes to be monitored down to Using the Clarus 680 GC low-ppm concentrations. There is an increased need to monitor the level of oxygen- with S-Swafer Micro- ates in reformulated gasoline because of the increased risk of contamination from more abundantly available biofuels. Channel Flow Technology Gasoline contamination with oxygenated compounds can lead to more toxic carbonyl emissions into the atmosphere from engines running the fuel. Contamination of feedstocks with oxygenated compounds can cause degradation of expensive catalysts and impact the refining process.
  • 11. The analytes targeted in this method are those specified in ASTM® D4815 and are given in Table 1. Table 1. Targeted Oxygenated Compounds. Methanol Ethanol Isopropanol tert-Butanol n-Propanol Methyl tert-butyl ether (MTBE) sec-Butanol Figure 1. Diagram of the S-Swafer micro-channel flow technology used to determine trace-level oxygenated compounds in gasoline. Diisopropyl ether (DIPE) Isobutanol Ethyl tert-butyl ether (ETBE) to one of the S-Swafer outlets to enable the carrier-gas flow tert-Pentanol rate to be increased and to allow the chromatography to be 1,2-Dimethoxyethane monitored on the first column by connecting the restrictor outlet to the FID. Note the use of nitrogen as the carrier gas. n-Butanol This carrier gas is well suited to 0.530 mm i.d. columns and tert-Amyl methyl ether (TAME) is consistent with initiatives to reduce the use of the declining global stocks of helium. Experimental The experimental details for this analysis are given in Table 2. Figure 1 shows a diagram of the S-Swafer™ micro-channel The sample of gasoline is injected and chromatographed on flow splitting device used to perform this analysis. This two- the non-polar precolumn. The polar oxygenated compounds column backflushing configuration (designated as S6 in the are quick to elute into the second CP-Lowox® column and Swafer documentation) enables the first column to be back- will precede the bulk of the gasoline hydrocarbons. Once flushed while the analytes are still being chromatographed the oxygenated compounds are in the second column, the on the second column. A restrictor tube is also connected first column is backflushed to remove the hydrocarbons Table 2. Analytical Conditions for the Determination of Trace-Level Oxygenated Compounds in Gasoline. Gas Chromatograph PerkinElmer® Clarus® 680 GC Analytical Column 10 m x 0.530 mm x 10 µm Varian® CP-Lowox® Oven Temperature 80 °C for 1 minute then 5 °C/min with in-line 25 cm x 0.100 µm deactivated to 125 °C then 10 °C/min to 230 °C fused silica restrictor connected between S-Swafer and column Injector Heated split/splitless Restrictor Tubing 30 cm x 0.100 µm deactivated fused silica Injector Temperature 250 °C between S-Swafer Carrier Gas Nitrogen and Detector Initial Injector Pressure 2.0 psig (see text) (Midpoint) Pressure 35 psig at S-Swafer Injector Split-Flow Rate 15 mL/min Timed Events Carrier gas pressure set to 45 psig at -1.00 min Detector Flame Ionization (FID) (see text) Carrier gas pressure set to 2 psig at 1.52 min Detector Temperature 325 °C Sample Preparation a) AccuStandard® 4815-RT-PAK diluted Detector Combustion Gases Air: 450 mL/min, ~1:1500 in washed gasoline to give 27.4 Hydrogen: 45 mL/min to 50 ppm w/w Detector Range x1 b) AccuStandard® 4815-RT-PAK diluted Detector Attenuation x4 ~1:7500 in washed gasoline to give 5.5 to 10 ppm w/w Backflush System S-Swafer configured in S6 mode Sample Injection Fast injection of 1.0 µL of prepared sample Precolumn 15 m x 0.530 mm x 1 µm PerkinElmer using an autosampler Elite™-1 with 25 cm x 0.250 mm deactivated fused silica restrictor connected between S-Swafer and column 2
  • 12. from the GC system. While the backflushing is in progress, the oxygenated compounds are chromatographed on the highly polar CP-Lowox® column which easily separates them from any hydrocarbons that also enter the second column. Precolumn backflushing is initiated by a timed event at 1.52 minutes that reduces the pressure inside the injector to 2.0 psig so that carrier gas flows backwards. The inlet pressure is set to 2.0 psig to maintain the backflushing of the precolumn during oven cooling. A pre-run timed event sets the inlet pressure to 45 psig so that forward flow is restored prior to injection of the sample. Figure 2 shows a chromatogram of gasoline on the pre- column with no backflushing. The oxygenated compounds would elute within the first one or two minutes of this Figure 3. Precolumn chromatogram of a mixture of oxygenated compounds chromatogram and would be totally obscured by the early- at 4% to 7.3% w/w concentration indicating the proposed backflush point. eluting hydrocarbons in the gasoline. The later-eluting peaks Injector split flow increased to 100 mL/min. are not needed in this analysis so they will be the target of the backflushing step. Figure 4 shows the same sample run under the same conditions as those for Figure 3 but with the detector connected to the end of the analytical column. Good peak shapes and separations of the oxygenated compounds are seen within a 20.5-minute run time. On this column, some of the alcohol isomers are not separated. The total cycle time was about 25 minutes, which included oven cooldown and equilibration and autosampler loading. Figure 2. Chromatogram of 87-octane gasoline on the precolumn. Figure 3 shows a chromatogram of a mixture of the oxygenated compounds (no solvent) on the precolumn and it can be seen that all the peaks of interest have eluted from this column in just less than 1.50 minutes. From this trace, the backflush point was chosen to be 1.52 minutes. Figure 4. Analytical column chromatogram of a mixture of oxygenated compounds at 4% to 7.3% w/w concentration with precolumn backflushing at 1.52 minutes. Injector split flow increased to 100 mL/min. 3
  • 13. To test the system on samples of gasoline containing known The standard mixture of oxygenated compounds was diluted levels of oxygenated compounds, a quantity of gasoline was with the washed gasoline to produce low-level mixtures. ‘washed’ with water. 5 mL of 87-octane gasoline obtained These were chromatographed using the newly developed from a local filling station were shaken with 10 mL of deionized method and the chromatogram is shown in Figure 6. water for two minutes and then centrifuged at 5000 rpm for ten minutes. The aqueous layer was discarded and the process was repeated. 0.5 g of anhydrous sodium sulfate was added to the cleaned gasoline and shaken to remove any traces of water. Figure 5 shows chromatography of the gasoline before and after washing. Even though the gasoline contained up to 10% ethanol, this and all other polar compounds were effectively removed by the washing procedure. Figure 6. Chromatogram of washed gasoline spiked with low levels of oxygenated compounds. The 27.4 – 50 ppm w/w mixture was injected repeatedly to measure the area and retention-time precisions. These are summarized in Table 3. The area precision is around 1% relative standard deviation which is an excellent result – especially for polar compounds at low levels in a highly complex sample matrix like gasoline. The retention-time precision, which ranges from 0.013% to 0.039% relative standard deviation, is again an excellent result, considering the nature of this analysis. Figure 5. Efficacy of gasoline washing procedure for preparation of standard mixtures. Table 3. Area and Retention-Time Precision of a Standard Mixture of Oxygenated Compounds (n=9). Compound Conc. (ppm w/w) Mean Area RSD% Area Mean R.T (min) RSD% R.T. Ethyl tert-Butylether (ETBE) 27.4 64807 0.69 9.580 0.028 Methyl tert-Butylether (MTBE) 27.4 58557 0.64 9.827 0.021 Diisopropylether (DIPE) 27.4 53993 0.80 10.044 0.039 tert-Amylmethylether (TAME) 50.0 101209 0.65 11.503 0.039 Methanol 50.0 31028 1.11 13.642 0.023 Ethanol 50.0 60104 0.96 15.663 0.015 n & i-Propanol 50.0/50.0 142362 1.10 17.109 0.016 i, s & t-Butanol 50.0/50.0/50.0 273656 0.72 18.233 0.009 n-Butanol 50.0 89299 0.76 18.740 0.013 tert-Pentanol 50.0 103335 0.63 19.287 0.013 1,2-Dimethoxyethane 41.1 64233 1.26 19.982 0.016 4
  • 14. Conclusion This method, based on the S-Swafer technology, is able to separate and quantify the ASTM® D4815 target analytes at low-ppm levels and offers the following user benefits: • High-throughput analysis, with a complete analysis cycle time of 25 minutes • A robust system – the analytical column is protected from most of the gasoline compounds by backflushing • In-line restrictor with secondary column allows the primary column to be backflushed during chromatography, reducing analysis time • Splitting restrictor enables precolumn chromatography to be monitored – facilitating system setup • Use of nitrogen carrier gas to help preserve helium stocks and reduce operating costs • Excellent quantitative precision. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 009033_01
  • 15. a p p l i c at i o n n o t e Gas Chromatography/ Mass Spectrometry Author Greg Johnson PerkinElmer, Inc. Shelton, CT 06484 USA Fatty Acid Methyl Introduction The contamination of aviation fuel with fatty acid Ester Contamination of methyl esters (FAMEs) can arise due to the use of multi-product pipelines for fuel supply and distribution. Aviation Fuel by GC/MS In some countries, the widespread, mandatory intro- duction of automotive fuels with a bio-material content means that these pipelines are exposed to both auto- motive biodiesel with a 5% FAME (BD 5) content as well as to aviation fuel. FAMEs can adsorb onto the surface of the pipeline and later desorb, contaminating whatever fuel that follows, including aviation fuel. FAMEs alter the physical properties of the fuel. While presence of FAMEs in aviation fuel is of global concern, the United Kingdom is at the forefront of the analysis. The specification of aviation fuel in the UK is defined in the Ministry of Defence (MoD) Defence Standard 91-91. Technical authority for this standard is controlled by the MoD Defence Fuels Group with the agreement of the UK Civil Aviation Authority (CAA). This specification has recently been amended to formalize the acceptance of FAMEs in aviation fuel up to a maximum of 5 mg/Kg (ppm) combined total. To measure this specification, the Energy Institute has issued a method entitled IP PM DY: Determination of fatty acid methyl esters (FAME) derived from biodiesel fuel, in aviation turbine fuel – GC/MS with selective ion monitoring/scan detection method. The US specification of aviation fuel is covered by ASTM® D1655; in August of 2009, in a special airworthiness bulletin (NE-09-25R1), the Federal Aviation Administration (FAA) stated its intent to include a similar regulation in its specifications.
  • 16. Current methods for determining FAME in aviation fuel use Table 1. Detailed instrument parameters for the GC/MS a polar ‘wax’ type GC column with mass spectrometry (MS) analysis of FAMEs. for detection. The polar column will be more retentive for GC Conditions FAME compounds relative to the less polar hydrocarbon Column 25 m x 0.32 mm id x 0.25 μm BPX70 compounds of the fuel. The MS will identify FAMEs based on unique spectral data, further distinguishing FAMEs from Injector Split/splitless @ 280 ˚C with glass wool packed hydrocarbon aviation fuel. The limitation of current methods liner is that the wax column has a relatively low maximum- Carrier Gas Helium @ 15 psig at injector and 3 psig at Swafer temperature limit and high column bleed as it approaches Injection 1.0 μL splitless for 0.75 minutes then 50 mL/min this limit. Oven 150 ˚C for 2 minutes then 10 ˚C/min to 220 ˚C and hold for 6 minutes MS Conditions Mode Electron Ionization (70 eV) Source 180 ˚C Transfer Line 300 ˚C Photomultiplier 550 V Table 2. Calibration summary for FAMEs with GC/MS. Figure 1. 0.1 ppm (wt/wt) individual FAME compounds in n-dodecane. FAME Concentration Range Coefficient of Determination (r2) C16:0 0.5 to 10.0 ppm 0.999929 In this application note, a different type of polar capillary C18:0 0.5 to 10.0 ppm 0.999872 column is used. This column reduces the temperature necessary to elute all compounds, while also providing a higher maxi- C18:1 0.5 to 10.0 ppm 0.999744 mum temperature. This will reduce both the wear on the C18:2 0.5 to 10.0 ppm 0.999577 column and the amount of signal in the chromatogram C18:3 0.5 to 10.0 ppm 0.999092 associated with column bleed. Additionally, this method will increase sample throughput. The specification of 5 mg/Kg is a total for all FAME compounds present and therefore the Calibration limit of detection for individual FAME compounds needs to A commercially available standard (Supelco®) containing be significantly lower than 5 mg/Kg. Figure 1 demonstrates equal masses of pure C16:0, C18:0, C18:1, C18:2 and that 0.1 mg/Kg is possible using this experimental setup. C18:3 was diluted by weight to 2000 mg/Kg with n-dodecane solvent, as recommended in Institute of Petroleum method Experimental PM-DY/09. Calibration standards were prepared at 0.5, 1.0, This application was performed on a PerkinElmer® Clarus® 2.5, 5.0 and 10.0 mg/Kg and spiked with 2.0 µL of C21:0 in 680 GC/MS with a capillary split/splitless injector. A 1.0 µL n-dodecane solution used as an internal standard. splitless injection was used to introduce the standards and samples into an unpacked, 2 mm i.d., quartz liner. The chro- matographic separation was achieved on a 25 m x 0.32 mm i.d. x 0.25 µm BPX70 column (SGE, Australia). The MS was used in single-ion-recording mode (SIR) to provide maximum sensitivity and specificity. Complete instrument parameters are presented in Table 1. Figure 2. The calibration result from TurboMass™ GC/MS software demonstrating the linear response of FAME C16:0 across a range of 0.5-10.0 mg/Kg. 2
  • 17. The calibration standards were analyzed in triplicate. All data points were used without exception. The response across the Table 5. The precision results from the analysis of a rapeseed oil mixture (n=10). calibration range was very linear (r2 > 0.9991) for all FAME compounds; complete calibration data is presented in Table 2. RSD% Additionally, Figure 3 demonstrates the calibration output RSD% Int’l Std. Conc. (%) in Conc. Absolute Response from the PerkinElmer TurboMass software. FAME original mix (ppm) final Area Ratio Immediately following the instrument calibration, a 5 mg/Kg C14:0 1.0 0.50 2.02 0.86 standard was analyzed to provide additional precision data. C16:0 4.0 2.00 1.66 1.18 The results of this precision study are reported in Table 3. C18:0 3.0 1.50 2.08 0.63 C18:1 60.0 30.00 1.52 0.54 Table 3. System precision evaluation, replicate injections C18:2 12.0 6.00 1.84 0.19 (n=15) of the 5.0 mg/Kg calibration mixture. C18:3 5.0 2.50 2.34 0.89 FAME Concentration RSD% Absolute RSD% IS Peak Area Resp. Ratio C20:0 3.0 1.50 2.08 0.33 C16:0 5.0 ppm 1.34 0.78 C20:1 1.0 0.50 3.41 1.71 C18:0 5.0 ppm 1.34 0.50 C22:0 3.0 1.50 2.83 1.10 C18:1 5.0 ppm 1.40 0.62 C22:1 5.0 2.50 3.04 1.53 C18:2 5.0 ppm 1.22 0.86 C24:0 3.0 1.50 3.66 1.86 C18:3 5.0 ppm 1.55 0.86 Internal Standard = C21:0 The ability of the method to determine a wide range of FAMEs at various concentrations was verified by analyzing a commercially available rapeseed oil standard (Supelco®). The standard was diluted 100 fold (100 mg to 10 g) by weight in n-dodecane, with a further 200:1 dilution resulting in a mixture with the composition in Table 4. Table 4. Composition of fame mixture used to verify the method identification of a mixture of FAMEs. FAME Initial Concentration Final Concentration C14:0 1.0 % 0.5 ppm C16:0 4.0% 0.8 ppm C18:0 3.0% 0.6 ppm Figure 3. Chromatography of the rapeseed oil reference mixture using SIR. C18:1 60.0% 30.0 ppm C18:2 12.0% 6.0 ppm The final verification of this method was to analyze an C18:3 5.0% 2.5 ppm aviation-fuel sample spiked with a known amount of FAME. C20:0 3.0% 1.5 ppm Figure 4 presents the resultant chromatogram from the C20:1 1.0% 0.5 ppm analysis of an aviation-fuel sample spiked with between C22:0 3.0% 1.5 ppm 0.5 and 30 ppm FAME. C22:1 5.0% 2.5 ppm C24:0 3.0% 1.5 ppm The rapeseed oil was analyzed in ten replicates to verify the precision of the method when the FAME materials span a wide concentration range (Table 5). An example chromatogram from this analysis is shown in Figure 3. 3
  • 18. Figure 4. The resultant chromatogram of the analysis of aviation fuel spiked with FAME. As you can see in Figure 4, the GC conditions provide adequate resolution of C16:0 FAME from the hydrocarbon envelope of the aviation fuel, as well as low-level detection capabilities independent of matrix. Conclusions This application note demonstrated the use of the Clarus 680 GC/MS to identify and determine the concentration of low-level FAMEs. A polar column was used in conjunction with a short GC run to provide a fast and robust method. The MS was operated in SIR mode to achieve the highest levels of sensitivity and specificity. Also demonstrated in this note is compliance with the newly released methods set forth by regulatory agencies such as the Institute of Petroleum. In addition, preliminary results indicate that the current method used for the analysis of FAME contamination in aviation fuel can be speeded up very significantly without sacrificing either sensitivity or precision. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 009031_01
  • 19. a p p l i c at i o n n o t e Differential Scanning Calorimetry Authors Ji-Tao Liu Tiffany Kang Peng Ye PerkinElmer, Inc. Shelton, CT 06484 USA Curing Determination Introduction Renewable energy has attracted a lot of of EVA for Solar Panel interest due to the limited supply of coal and oil and the environmental concern Application by DSC of carbon dioxide (CO2) emission. There are many different forms of renewable (green) energy including: solar, wind, geothermal, biomass, and so on. Among them, solar energy is the fastest-growing segment. Increasing manufacturing capacity and decreasing product costs have led to significant growth in the solar industry over the past several years. For instance, solar photovoltaic (PV) production has been increasing by an average of 48% each year since 2002. By the end of 2008, the cumulative PV installation reached more than 15 giga-watts globally. A solar cell is a device that can convert sunlight directly into electricity. Different solar-cell technologies including crystalline silicon, organic photovoltaics, and dye-sensitized solar cells have been developed for various solar-cell applications. Currently, the most widely commercially available solar cell is based on crystalline- silicon technology. This technology is mature compared with the other solar-cell technologies and its energy-conversion efficiency is high.
  • 20. A photovoltaic module or system consists of many jointly Experimental connected solar cells. The solar cells are packaged between The instrument used here is the PerkinElmer® double-furnace a backsheet on the bottom and a tempered-glass window DSC 8000. It features power-controlled design for direct and on the top. The cells are encapsulated by a polymer encapsulant accurate heat-flow measurements to and from the sample (Figure 1). The polymer encapsulant serves many functions – material. The cooling accessory is an Intracooler 2P mechanical it provides mechanical support, electrical isolation, and refrigerator. Nitrogen is used as the sample purge gas at protection against outdoor environmental elements of 20 mL/min. The instrument was calibrated with two metal moisture, UV radiation and temperature stress. Many different reference materials: indium and zinc were used for temperature materials can be used for encapsulation, but one commonly calibration, and indium was used for heat of fusion for used encapsulant for this purpose is EVA (ethylene-vinyl-acetate). heat-flow calibration. The EVA samples are from a solar PV manufacturer. They were cured at a high temperature and EVA, a thermal-set material, is a copolymer elastomer supplied pressure for some time. Each EVA sample weighed approxi- in sheet form for use in the encapsulation of PV modules. mately 10 mg. Each EVA sample cured at different times It has many desirable properties which make it the material was encapsulated in the standard aluminum pans. The DSC of choice for this application. program started from -50 ˚C and heated to 220 ˚C at 10 ˚C/min. • It is not adhesive at room temperature for easy handling. Results • It makes a permanent and adhesive tight seal in the The raw EVA material exhibits several transitions during solar-cell system through crosslinking and enhanced heating, as shown in Figure 2 (Page 3). It was heated in the bonding when the film is heated and pressed. DSC from -50 ˚C to 220 ˚C at 10 ˚C/min, and after that it • After crosslinking, the EVA has high optical transmittance, was cooled to the starting temperature quickly at 100 ˚C/min. good adhesion to the different module materials – it It was heated for the second time at the same heating rate. provides good dielectric properties and great moisture- The first heating curve shows an endothermic melting peak barrier properties with adequate mechanical compliance (26 J/g) followed by the exothermic curing peak with the to accommodate system thermal stresses due to the curing enthalphy of 16.6 J/g. The second heating curve different thermal-expansion coefficiencies. shows a glass transition (Tg) at -35.6 ˚C; the melting peak is smaller (12 J/g vs. 26 J/g) and there is no detectable curing During the PV-package process, the EVA sheet is placed exothermal peak. So by comparing the first heating curve between the solar cells and the backsheet/glass. It is heated, with the second heating curve, it is clear that the EVA raw pressed into place, and cured at a certain high temperature material is cured completely after first heating it up to 220 ˚C. for some time. Since the final cured material’s properties are largely dependent on the curing degree, it is important to For a partially cured EVA sample, the residual curing peak know the degree of curing of the EVA so that the encapsu- during the first heating will be between the curing enthalpy lation process is optimized. Differential scanning calorimetry of raw EVA material and zero for completely curing EVA. So (DSC) has been traditionally used for curing studies of thermoset the residual curing enthalpy can be used as an indicator of resins. DSC can study the degree of cure and curing kinetics. the curing degree of EVA material. A series of EVA samples In this note, different EVA materials with different curing with different curing time are studied by DSC and the results times were investigated with PerkinElmer‘s high-end DSC 8000. are shown in Figure 3 (Page 3). The calculated residual curing enthalpy is tabulated in Table 1 and fitted to a straight line in Figure 4 (Page 3). As can be seen, the residual curing enthalpy can be correlated to the curing time very nicely (R2 = 0.9893). Table 1. Residual curing enthalpy of eight different EVA samples with different curing times. EVA Curing Time ΔH (residual curing samples (min) enthalpy J/g) EVA-1 1 11.3572 EVA-2 2 10.7635 EVA-3 3 9.6878 EVA-4 4 7.9689 Figure 1. Scheme of a crystalline solar panel. EVA-5 5 7.5885 EVA-6 6 6.7448 EVA-7 8 4.9335 EVA-8 9 3.9811 2
  • 21. Figure 2. The first (red) and second (blue) heating curve of raw EVA material. Figure 3. Partially cured EVA samples with different curing times. Conclusion This study shows that DSC can be used to study the curing degree of the EVA resin by measuring the residual curing enthalpy. The data show that the residual curing enthalpy can be correlated to the curing time in a linear way. The DSC test is quick and easy. The double-furnace PerkinElmer DSC 8000 delivers accurate heat-flow data with great reproducibility. The power-controlled design ensures great accuracy and true isothermal measure so that it can be used for both scanning-curing and isothermal-curing studies of EVA resin. Figure 4. The relationship between residual-curing enthalpy and the curing time of the eight different EVA samples. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 009038_01
  • 22. a p p l i c at i o n n o t e ICP-Mass Spectrometry Authors Jianmin Chen PerkinElmer, Inc. Shelton, CT 06484 USA Determination of Mercury is ubiquitous in nature, and the human health consequences of mercury exposure were recognized from Mercury in Wastewater prehistory to the present. The first emperor of unified China who came to power in 221 B.C., Qin Shi Huang, reportedly by Inductively Coupled died of ingesting mercury pills that were intended to give him eternal life. The severity of mercury's toxic effects Plasma-Mass Spectrometry depends on the form and concentration of mercury and the route of exposure. Although its potential for toxicity in highly contaminated areas such as Minamata Bay in Japan is well documented, research has shown that mercury can be a threat to the health of people and wildlife in many environments that are not obviously polluted. There is no safe level of mercury for humans. The main toxic effects of mercury are known to negatively affect the neurological, renal, cardiovascular and immunological systems.
  • 23. Mercury exists in three chemical forms: elemental or metallic, organic or methylmercury, and inorganic complexes. Table 1. ELAN ICP-MS Instrumental Conditions and Experimental Parameters. Mercury has thousands of industrial applications. Some common uses for mercury include conducting electricity, RF power 1100 W measuring temperature and pressure, acting as a biocide, Plasma gas flow 15 L/min preservative, and disinfectant, as well as being a catalyst for Auxiliary gas flow 1.2 L/min reactions. Unlike most other pollutants, mercury is highly Nebulizer gas flow 0.96 L/min mobile, non-biodegradable, and bio-accumulative; as a result, it must be closely monitored to ensure its harmful Nebulizer MEINHARD® Concentric Type A3 effects on local populations are minimized.1 Thus, measurement Spray chamber Baffled Quartz Cyclonic of mercury in environmental samples, and in particular waste- Scanning mode Peak Hopping water, is of great importance as a major tool to protect Dwell time 50 ms the environment from mercury released through emissions Replicates 3 from manufacturing, use, or disposal activities. Currently, the prominent methods typically utilized by the environmental Integration time 1 sec/mass community for the determination of mercury generally require detection limits as low as 0.5 ng/L (ppt, parts- per-trillion).2 Sample Preparation The stability of mercury-containing solutions has been Traditionally, mercury is analyzed using Cold Vapor Atomic a topic of concern for all trace analysts performing Hg Absorption Spectroscopy (CVAAS) or Cold Vapor Atomic determinations. It is reported that a trace amount of gold Fluorescence Spectroscopy (CVAFS). Both of these techniques salt added to HNO3 preserved all forms of mercury. The are relatively straightforward to use and can accomplish gold ion acts as a strong oxidizing agent that converts or the analytical requirements of detection limits in the low maintains mercury as mercuric ion which remains in solution.3 ppt range. However, they are generally specific for mercury Thus, a solution of 2% (v/v) HNO3 containing 200 ug/L Au analysis only. was used for preparation of all samples and standards. In recent years, Inductively Coupled Plasma Mass Spectrometry Two simulated wastewater certified reference materials (ICP-MS) has become one of the most powerful analytical (Trace Metals Solutions, CWW-TM-A and CWW-TM-C, techniques for trace element analysis because of its high High-Purity Standards, Charleston, SC, USA) were prepared sensitivity, wide linear dynamic range, and simultaneous according the manufacturer’s description using the same multi-element detection capability. As a result, ICP-MS has diluent in this study. been increasingly adopted in environmental and biomonitoring laboratories for the simultaneous measurement of mercury Calibration with other toxic metals since this technique can offer the External calibration standards of mercury were at the level same analytical performance as CVAAS or CVAFS. This of 10, 20, 50, 100, 200, 500, 1000 ng/L. Figure 1 shows application note describes the application of ELAN® ICP-MS the calibration curve of 202Hg. The correlation coefficient is to the determination of mercury in wastewater. 0.999973, which allowed the accurate quantitative analysis of mercury at the low ppt levels. Instrumentation For this study, the PerkinElmer® ELAN DRC™ II ICP-MS was used for the analysis of wastewater samples under standard mode. The ELAN ICP-MS instrument conditions and general method parameters are shown in Table 1. Figure 1. External calibration curve of 202Hg. Standard solutions were prepared in 2% HNO3 containing 200 ug/L Au with concentrations ranging from 10 to 1000 ng/L. 2
  • 24. Spike recovery The spike recovery test was performed using two simulated wastewater certified reference materials (HPS CWW-TM-A and CWW-TM-C). The results are presented in Table 2. Table 2. Spike recovery test results from two simulated wastewater certified reference materials (HPS CWW-TM-A and CWW-TM-C). Wastewater Found Spike Measured Recovery Sample (ng/L) Level (ng/L) % (ng/L) CWW-TM-A 49 200 238 95 CWW-TM-C 375 1000 1370 99 Memory effect It is generally viewed that routine determination of mercury by ICP-MS is affected by a pronounced memory effect in the Figure 2. Uptake and wash out profile for Hg at 1000 ng/L level. The signal of sample introduction system. This results in long washout times four mercury isotopes returns to the background level in around 60 seconds, for the analyte, which affects the accuracy and reliability of which is equivalent to the typical rinse time used for many other analytes. the analytical procedure. To minimize the memory effect, the system was washed using 2% (v/v) HNO3 containing 200 ug/L Au between samples. Figure 2 shows the uptake and wash out Conclusion profile of Hg. The signal of four mercury isotopes returns to The ELAN ICP-MS combines high sensitivities and superior the background level in around 60 seconds, which is equivalent detection limits with ease of use and high sample throughput. to the typical rinse time used for many other analytes. Thus, This allows it to be an excellent tool for trace mercury the results indicated application of gold salt is effective in determination in wastewater samples. Mercury can be detected preventing mercury sorption and vapor buildup within the simultaneously with the other trace elements of interest within sample introduction system. a single sample run. References 1. J.O. Nriagu, Global inventory of natural and anthropogenic emissions of trace metals to the atmosphere, Nature 279 (1979) 409–411. 2. EPA Method 1631. Mercury in Water by Oxidation, Purge and Trap, and Cold Vapor Atomic Fluorescence Spectrometry. 3. www.epa.gov/nerlesd1/factsheets/mpt.pdf. Mercury Preservation Techniques. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2009-2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 008973A_01
  • 25. a p p l i c at i o n n o t e Thermogravimetric Analysis – GC Mass Spectrometry Author Greg Johnson PerkinElmer, Inc. Shelton, CT 06484 USA Qualitative Analysis Introduction Thermogravimetric analysis (TGA) of Evolved Gases in measures the change in the weight of a sample as a function of temperature. Thermogravimetry by A limitation of TGA is that it cannot identify Gas Chromatography/ what material is lost at a specific tempera- ture. The analysis of gases evolved during Mass Spectrometry a TGA experiment by gas chromatography mass spectrometry (GC/MS) provides laboratories with a way to identify the compound or groups of compounds evolved during a specific weight-loss event in a TGA analysis. This application note discusses the utility of TG-GC/MS with an example application – the identification of specific organic acids evolved during TGA analysis of switchgrass.
  • 26. Figure 3 describes in greater detail the pneumatic supply to the S-Swafer device and assists in explaining why this approach is so well suited to interfacing the Pyris 1 TGA to the Clarus 600 GC/MS. Figure 1. Clarus 600 GC/MS interfaced to the Pyris 1 TGA. Switchgrass (Panicum Irgatum) is a perennial warm-season grass native to the northern states of the USA; it is easily grown in difficult soils. Switchgrass is potentially useful in Figure 3. Schematic showing carrier-gas supply to the S-Swafer device. the production of biofuels, specifically cellulosic ethanol and bio-oil. The instrumentation used in this study was a PerkinElmer® The S-Swafer configuration is optimal because it will ensure Pyris™ 1 TGA interfaced to the PerkinElmer Clarus® 600 GC/ a very rapid adjustment of carrier gas to the Swafer device, MS with the S-Swafer™ micro-channel flow splitting device allowing for a rapid switch between backflush of the transfer (S4 configuration). The preferred mode of operation of the line and sampling from the TGA. The samples of evolved gas TGA maintains the atmosphere around the sample at ambi- are collected by setting a simple parameter in the GC method; ent atmospheric pressure. The sample is collected from the multiple samples can be collected during a TGA analysis. TGA by allowing the high vacuum of the MS to create a Additionally, backflush of the transfer line will isolate the pressure drop across the GC column, causing a flow of gas GC/MS and enable purge gas at the TGA to be switched from the TGA through the transfer line and the analytical from an inert gas (during analysis) to an air supply for column to the MS. During the analysis, there are times when cleaning of the TGA pan prior to the next sample. the TGA inlet will be surrounded by air, rather than an inert atmosphere. This would cause air to flow into the GC/MS; this Oven subambient cooling will be extremely useful in this is undesirable as it will cause oxidation to a number of dif- application, allowing protracted sampling periods to be ferent areas of the system. The S-Swafer device (shown in refocused into a narrow band of analytes on the column. Figure 2) is used to switch between backflushing of the TGA transfer line during non-sampling time and sampling Experimental of the TGA environment during analysis. The deactivated fused-silica transfer line used here was 1.6 m x 0.32 mm i.d. A few centimeters of the deactivated fused silica protrudes into the sample weighing area of the TGA. Approximately 30 cm of fused silica passes through the injector into the oven environment and is connected to the S-Swafer using specialized SilTite™ nuts and ferrules to ensure a leak-free connection that will not shrink and leak during normal or even extended thermal cycling of the main oven. In all cases, a 30 m x 0.32 mm analytical column was employed as this allows a carrier flow of approximately 1 mL/min with the fixed 1.00 atmosphere pressure drop from ambient at the TGA to vacuum at the MS. Data was Figure 2. Schematic showing the pneumatic interfacing of the TG-GC/MS acquired using an Elite™ WAX stationary phase. using the S-Swafer. 2
  • 27. A small quantity of dried and ground switchgrass was timed events that will be used to sample the evolved gases placed on the TGA pan and weighed using Pyris software. onto the GC/MS column. Note that the TGA is held iso- A rapid TGA analysis based on heating the sample from thermal for the first 5.0 min at which point heating begins. 30 ˚C to 1000 ˚C at 100 ˚C/min in a nitrogen atmosphere Simultaneously, the GC/MS analysis is started. was performed to determine which regions of the weight- loss curve were to be further studied using the TG-GC/MS Figure 5 illustrates the TG-GC/MS analysis of the switch- technique. grass based on timed events that collect the evolved gases from the main transition shown in Figure 4. The smaller The primary reason for using such rapid heating, which earlier transition, also seen in the same figure, was also sampled reduces the resolution of the weight-loss curve produced onto the GC/MS but preliminary findings indicate that this by the TGA, is to transfer the evolved gas quickly into the is simply evolved water. The major transition produced large GC column. A quick transfer will improve GC peak shape, numbers of oxygenated volatile organic compounds (VOCs), sensitivity and resolution. including some very polar species. Earlier work using a non- polar capillary column had generated extremely smeared-out After the sample was loaded onto the TGA and the furnace early-eluting peaks. The chromatogram below was generated raised, the analysis was started immediately. The first step in using a thick-film polar Elite WAX column. the TGA heating program maintained the low initial furnace temperature for 5 to 10 min. During this time, the furnace environment is being purged with helium (or nitrogen/argon), and the carrier-gas pressure of 7.0 psig maintained at Aux 1 (Figures 2 and 3) ensures that no sample can enter the ana- lytical column. After this initial hold period, the TGA furnace begins to heat the sample, and simultaneously, the GC/MS run is started using an external start command. Figure 5. TG-GC/MS analysis of the switchgrass sample on a 30 m x 0.32 mm Based on previous TGA runs on the same sample, timed x 1 μm Elite WAX column. events within the GC method switch off the carrier gas supplied by the Aux 1 PPC module and then close the solenoid valve (SV3) shown in Figure 2 (Page 2). This begins The three peaks labeled with asterisks in Figure 5 are identified the sampling and this procedure is reversed to bring the sam- as a homologous series of free fatty acids (Figure 6 – Page 4), pling period to an end. After the sampling is complete, both based on a library search of their MS spectra (NIST® 2008). the GC oven-temperature program and MS data acquisition Usually in GC, a homologous series tends to elute in carbon- begins. The TGA can now be programmed to switch purge number order but here, the elution order appears to be gases to clean the system using oxidation at elevated acetic, followed by formic, followed by propanonic acid. As temperature, prior to the next analysis. this retention behavior is not typical and in the absence of a literature reference or a similar chromatogram in the public A typical TGA weight-loss curve for the switchgrass is shown domain, it seemed prudent to analyze a simple retention- in Figure 4 and reveals a typical weight % loss curve for the time standard to confirm this tentative result. Figure 7 (Page 4) sample of switchgrass that was tested. In addition, super- shows the same analysis again but with the retention-time imposed on the weight-loss curve is the derivative of this standard shown in parallel. This standard was a simple curve which greatly assists the analyst in setting up the GC mixture diluted in water with a small (5 μL) aliquot of this aqueous solution deposited by syringe onto the TGA pan for analysis. Figure 4. Typical result for the TGA analysis of switchgrass. 3
  • 28. Conclusions In this application note, we describe the technique of TG- GC/MS through the analysis of switchgrass. TG-GC/MS is demonstrated to be a valuable technique in the separation and identification of complex mixtures of gas evolved during a TGA analysis. The S-Swafer device is demonstrated as a means to interface a TGA to GC/MS. The main benefits are its simplicity and the inertness of the entire sample path. Figure 6. Mass spectra extracted from the total ion chromatogram of the switchgrass sample. The spectral data matches that of acetic, formic and propionic acid respectively available in the 2008 NIST® mass spectral library. Figure 7. TG-GC/MS analysis of the switchgrass sample (bottom) on a 30 m x 0.32 mm x 1 μm Elite WAX column and the analysis of a simple retention time mix (top) under the same TGA and GC conditions. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 009030_01
  • 29. a p p l i c at i o n n o t e ICP-Mass Spectrometry Authors Lee Davidowski, Ph.D. Zoe Grosser, Ph.D. Laura Thompson PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT USA The Determination Introduction Dietary supplements are regulated by the FDA under the general umbrella of foods, of Metals in Dietary but with different regulations than conventional foods. Dietary supplements have been defined by Congress as materials taken by mouth that include ingredients intended to Supplements provide dietary supplementation. They can be found in various forms, including tablets, powders, and liquids. They may consist of vitamins, minerals, herbs or other botanicals, amino acids, and substances such as enzymes, organ tissues, glandulars, and metabolites. Under the Dietary Supplement Health and Education Act of 1994 (DSHEA), the dietary supplement manufacturer is responsible for ensuring that a dietary supplement is safe before it is marketed. FDA is responsible for taking action against any unsafe dietary supplement product after it reaches the market.1 One facet of ensuring a safe product is analysis of the final product before distribution. Although organic components often form the majority of a supplement, metals may be found due to their inclusion in vitamin structures such as vitamin B-12 (cobalt) and minerals, such as selenium. They may also be added as a contaminant through natural products or in the manu- facturing process. The measurement of toxic metals or metals intended to be present for quality and labeling confirmation may be required. Analysis challenges include measurement at low concentrations in a variety of matrices. In this note we will use inductively coupled plasma mass spectrometry (ICP-MS) to measure a variety of elements generally considered to be hazardous to human health at low to medium concentrations. The four elements generally considered to be hazardous and not necessary for nutrition are Pb, Cd, As, and Hg. The elements Se and Cr are often added at low concentrations for nutritional purposes and were also included.