SlideShare une entreprise Scribd logo
1  sur  205
Sound 
• Sound beam is similar to x ray beam as both waves 
transmit energy
DIAGNOSTIC 
ULTRASOUND 
X RAYS 
Wave type Longitudinal waves Electromagnetic 
waves 
transmission Elastic medium No medium 
generation Stressing the 
medium 
Accelerating 
electric charges 
velocity Depends on the 
medium 
constant 
Similar waves Seismic, acoustic Radio, light
Longitudinal Waves 
• Ultrasound pulses are 
transmitted, as longitudinal 
waves, ie; the motion of 
particles in the medium is 
parallel to direction of wave 
propagation. 
• They produce bands of 
compression and rarefaction 
• Wave length is distance 
between two bands of 
compression or rarefaction
“Ultra”…….sound? 
• Audible range is 20 to 20,000 cycles per 
second 
• Ultrasound has frequency greater than 
20,000 cycles per second
Velocity of Sound 
• It is independent of frequency , 
• but depends on material through which it 
propagates 
• Depends on 
– Compressibility 
– Density
Velocity of Sound 
1. Compressibility 
– velocity is inversely proportional 
– liquids and solids propagate sound more rapidly than 
gases (easily compressed) 
2. Density 
– denser materials have greater inertia - so, decreased 
velocity
ULTRASOUND CHARACTERISTICS 
• Frequency 
• Velocity 
• Wavelength 
• Amplitude 
• Intensity and Power
Frequency 
• The frequency of sound is 
determined by the source. 
(2-20MHz) 
v = f λ 
• In ultrasonic frequency range, the 
velocity of sound is constant in 
any particular medium. 
f  
1 
 
• If the frequency increases then 
the wave length must decrease as 
they are inversely proportional to 
each other.
Velocity 
• The significance of 
ultrasound velocity is 
that it is used to 
determine the depth 
location of structures in 
the body 
• The period (T) is the 
time required for one 
vibration cycle. It is the 
reciprocal of the 
frequency
Amplitude 
• The amplitude of an 
ultrasound pulse is the 
range of pressure 
excursions , related to the 
energy content. 
• In diagnostic applications, 
it is usually necessary to 
know only the relative 
amplitude of ultrasound 
pulses. 
• Units of decibels (dB).
Intensity 
• Also called loudness 
• Determined by amplitude of oscillation 
• As amplitude is increased , intensity also 
increases
Intensity 
• Intensity is the rate at which ultrasound energy is 
applied to a specific tissue location within the 
patient's body. 
• It is the quantity that must be considered with 
respect to producing biological effects and safety 
• The intensity of most diagnostic ultrasound 
beams at the transducer surface is on the order 
of a few milliwatts per square centimeter.
History 
• First successful application – SONAR in world 
war 2 (SOund Navigation And Ranging)
Successful medical application – 1940s 
Uses of ultrasonic energy in the 1940s. Left, in gastric ulcers. Right, in arthritis 
Ultrasonic 
therapy 
generator, the 
"Medi-Sonar" 
in the 1950s. 
A British 
ultrasonic 
apparatus for 
the treatment 
of Meniere's 
disease in the 
late 1950s
History 
Denier's Ultrasonoscopic apparatus with 
ultrasound generator, emitter transducer 
and oscilloscope. This can be adapted for 
both therapeutic and diagnostic purposes The first hand-held imaging instrument was 
developed by John Wild and John Reid in the 
early 1950's
The Present
Components 
1. Transmitter 
2. Transducer 
3. CPU 
4. Display 
5. Key board / cursor 
6. Disc storage device. 
7. Printer
Transducer 
• Instrument which converts one form 
of energy to other 
• The conversion of electrical pulses to 
mechanical vibrations and the 
conversion of returned mechanical 
vibrations back into electrical energy. 
Electrical Energy Mechanical Energy
SELECTION OF TRANSDUCER 
• Superficial vessels and organs within 1 to 3cms 
depth and intra operative imaging – 
• 7.5 to 15 Hz 
• Deeper structures in abdomen and pelvis within 
12 to 15cms – 
• 2.25 to 3.5Hz
Transducer - Parts 
• A simple single-element, 
plane-piston source 
transducer has major 
components including the 
– Piezoelectric material, 
– Sensor electrodes, 
– Insulated layer, 
– Backing block, 
– Acoustic insulator 
– Insulating cover, and 
– Transducer housing.
Piezoelectric Element 
• The active element is basically a piece of polarized 
material - a piezoelectric ceramic sandwiched 
between electrodes 
• The piezoelectric element converts electrical signals 
into mechanical vibrations (transmit mode) and 
mechanical vibrations into electrical signals (receive 
mode).
Piezoelectric materials 
• Natural – Quartz 
• Artificial 
– most of USG materials 
– ferroelectrics 
• barium titanate 
• PZT (lead zirconate titanate) 
ADVANTAGE – they can be 
formed into different 
shapes
Ceramic  Piezoelectric Crystal? 
• The piezoelectric attributes are attained after a 
process of 
- Molecular synthesis, (dipoles) 
- Heating, (Curie Temperature) 
- Orientation of internal dipole structures with an 
applied external voltage, (Poling) 
- Cooling to permanently maintain the dipole 
orientation, and 
- Cutting into a specific shape.
Ceramic  Piezoelectric Crystal? 
• Once the material has cooled, the dipoles 
retain their alignment. 
• Heating the crystal above this temp reduces 
its usefulness. So, transducers should not be 
autoclaved.
Piezoelectric crystal 
• At equilibrium, there is no 
net charge on ceramic 
surfaces. 
• When compressed, an 
imbalance of charge 
produces a voltage between 
the surfaces - piezoelectric 
effect 
• Similarly, when a voltage is 
applied between electrodes 
attached to both surfaces, 
mechanical deformation 
occurs - electrostriction
Piezoelectric crystal – how thick? 
• The thickness of the 
active element is 
determined by the 
desired frequency of 
the transducer 
• Piezoelectric crystals 
are cut to a thickness 
that is 1/2 the desired 
radiated wavelength
Backing/Damping Block 
• The rear face of the 
piezoelectric crystal 
material is usually 
supported by a backing 
material which is 
tungsten loaded araldite, 
so that the vibrations in 
the piezoelectric material 
are rapidly damped after 
the initial excitation.
Backing/Damping Block 
• This component also dampens the transducer vibration 
to create an ultrasound pulse with a short spatial pulse 
length, which is necessary to preserve detail along the 
beam axis (axial resolution).
Couplant 
• Material (usually liquid) 
that facilitates the 
transmission of 
ultrasonic energy from 
the transducer into the 
test specimen. 
• Necessary to overcome 
the acoustic impedance 
mismatch between air 
and solids.
Modes of Vibration 
2 TYPES: 
1. thickness mode 
• most common 
• Used in medical 
crystals 
2. radial mode
Resonant Frequency 
• Natural frequency to which the transducer is 
sensitive 
• Resonant frequency determined by thickness of 
crystal 
• Thick crystal – low frequency sound 
• Natural frequency – one that produces internal 
wavelengths that are twice the thickness of crystal 
• Frequency corresponding to half the wavelength is-fundamental 
resonant frequency
Transducers 
Resonance Transducers Non Resonance Transducers
Resonance Transducers 
• They are manufactured to operate in a 
“resonance” mode, whereby a voItage 
(commonly 150 V) of very short duration (a 
voltage spike of 1 msec) is applied, causing 
the piezoelectric material to initially contract, 
and subsequently vibrate at a natural 
resonance frequency. 
• The operating frequency is determined from 
– the speed of sound in, and 
– the thickness of, the piezoelectric material.
Resonance Transducers 
• Higher frequencies are 
achieved with thinner 
elements, and lower 
frequencies with thicker 
elements.
Nonresonance (Broad-Bandwidth) 
“Multifrequency” Transducers 
• Modern transducer design coupled with digital signal 
processing enables “multifrequency or “multihertz” 
transducer operation, whereby the center frequency 
can be adjusted in the transmit mode.
Unlike the resonance transducer design, the piezoelectric element is 
intricately machined into a large number of small “rods,” and then filled 
with an epoxy resin to create a smooth surface.
Nonresonance (Broad-Bandwidth) 
“Multifrequency” Transducers 
• Excitation of the multifrequency transducer is 
accomplished with a short square wave burst 
of 150 V with one to three cycles, unlike the 
voltage spike used for resonance transducers. 
• This allows the center frequency to be 
selected within the limits of the transducer 
bandwidth.
Nonresonance (Broad-Bandwidth) 
“Multifrequency” Transducers 
• Likewise, the broad bandwidth response 
permits the reception of echoes within a wide 
range of frequencies. 
• For instance, ultrasound pulses can be 
produced at a low frequency, and the echoes 
received at higher frequency.
• “Harmonic imaging” is a recently introduced 
technique that uses this ability. 
• Lower frequency ultrasound is transmitted 
into the patient, and the higher frequency 
harmonics (e.g., two times the transmitted 
center frequency) created from the interaction 
with contrast agents and tissues, are received 
as echoes.
Spatial Pulse Length 
• The length of the sonic pulse. 
• The number of waves multiplied by their wavelengths 
• The backing block is incorporated to quench the 
vibrations and to shorten the sonic pulse.
Transducer Q Factor (Q = Quality) 
Refers to two characteristics of crystal 
• purity of their sound and 
• length of time the sound persists
Transducer Q Factor 
• The “Q factor” describes the bandwidth of the sound 
emanating from a transducer as: 
f 
Q o  
Bandwidth 
• where fo = center frequency 
bandwidth = width of the frequency distribution.
• The interval between initiation of the wave 
and complete cessation of vibrations is called 
the “ring down time”. 
• Dampening of the vibration lessens the purity 
of the resonance frequency and introduces a 
broadband frequency spectrum.
푄 = 
푓0 
퐵푎푛푑푤푖푑푡ℎ 
∝ 푆푝푎푡푖푎푙 푃푢푙푠푒 퐿푒푛푔푡ℎ 
• A “high Q” transducer has a narrow bandwidth 
(i.e., very little damping) and a corresponding 
long spatial pulse length – organ imaging 
• A “low Q” transducer has a wide bandwidth and 
short spatial pulse length – doppler
Characteristics of Ultrasound Beam 
• A single vibrating point sets out waves in all 
directions 
• Waves move away as concentric circles
When two sound waves interact , they 
cancel each other or reinforce each other
Beam Properties 
• The ultrasound beam propagates as a longitudinal 
wave from the transducer surface into the 
propagation medium, and exhibits two distinct 
beam patterns: 
– a slightly converging beam out to a distance specified 
by the geometry and frequency of the transducer (the 
near field), and 
– a diverging beam beyond that point (the far field).
The Near Field 
• The near field, also known as the Fresnel zone, 
is adjacent to the transducer face and has a 
converging beam profile. 
• Beam convergence in the near field occurs 
because of multiple constructive and 
destructive interference patterns of the 
ultrasound waves from the transducer surface.
• Huygen’s principle describes a large 
transducer surface as an infinite number of 
point sources of sound energy where each 
point is characterized as a radial emitter. 
• By analogy, a pebble dropped in a quiet pond 
creates a radial wave pattern.
Near Field Length………. 
• The near field length for an unfocused, single-element 
transducer is dependent on the transducer frequency and 
diameter: 
Near Field Length = 
풓ퟐ 
 
= 
풓ퟐ 
풗 
풇 
= 
풓ퟐ.풇 
풗
• For multiple transducer element arrays, an 
“effective” transducer diameter is determined 
by the excitation of a group of’ transducer 
elements. 
• Because of the interactions of each of the 
individual beams and the ability to focus and 
steer the overall beam, the formulas for a 
single-element, unfocused transducer are not 
directly applicable.
Near Field Length = 
푟2.푓 
푣
• Lateral resolution (the ability of the system to 
resolve objects in a direction perpendicular to 
the beam direction) is dependent on the 
beam diameter and is best at the end of the 
near field for a single-element transducer. 
• Lateral resolution is worst in areas close to 
and far from the transducer surface.
• Pressure amplitude characteristics in the near 
field are very complex, caused by the 
constructive and destructive interference 
wave patterns of the ultrasound beam. 
• Peak ultrasound pressure occurs at the end of 
the near field, corresponding to the minimum 
beam diameter for a single-element 
transducer.
• Pressures vary rapidly from peak compression 
to peak rarefaction several times during 
transit through the near field. 
– Only when the far field is reached do the 
ultrasound pressure variations decrease 
continuously.
Far Field 
• The far field is also known as the 
Fraunhofer zone, and is where the beam 
diverges.
• For a large-area single-element transducer, 
the angle of ultrasound beam divergence, , 
for the far field is given by 
 
d 
sin 1.22 
• where d is the effective diameter of the transducer and 
 is the wavelength; both must have the same units of 
distance.
sin 휃 = 1.22 
 
푑 
= 1.22 
푣 
푓 
푑 
= 
푣 
푑. 푓 
• Less beam divergence occurs with: 
– High - frequency transducers 
– Large - diameter transducers
Near Field Length = 
• High frequency beams – 
– fresnel zone is longer 
– depth resolution is superior 
• Disadvantage : 
푟2.푓 
푣 
– Tissue absorption is more, leading to deterioration of side 
to side resolution 
Solution: Focused Transducer
Focused Transducers 
• Single-element transducers are 
focused by using 
– a curved piezoelectric element or 
– a curved acoustic lens 
to reduce the beam profile.
Focal Distance 
• The focal distance, the 
length from the 
transducer to the 
narrowest beam width, 
is shorter than the focal 
length of a non-focused 
transducer and is fixed.
Focal Zone 
• The focal zone is defined as the region over 
which the width of the beam is less than two 
times the width at the focal distance; 
– Thus, the transducer frequency and dimensions 
should be chosen to match the depth 
requirements of the clinical situation.
Interactions B/N Ultrasound and Matter 
1. Reflection 
2. Refraction 
3. Absorption
Reflection 
• Diagnostic images are produced by reflected 
portion of the beam. 
• Percentage of beam reflected at tissue 
interfaces depends on 
– Tissue’s acoustic impedance 
– Beam’s angle of incidence.
Acoustic Impedance (Z) 
• The most important tissue property in imaging. 
• This quantity is more properly called the specific 
acoustic impedance of the medium 
• A simplified definition is 
Z = d v 
d = density of the tissue (g/cm3) 
v = velocity of sound (cm/sec) 
• Z – Rayl (g/cm2 x 10-5)
• At most interfaces 
within the body, only 
a portion of the 
ultrasound pulse is 
reflected 
• The pulse is divided 
into two pulses - 
one pulse, the echo, 
is reflected back 
toward the 
transducer and the 
other penetrates into 
the other material.
• The brightness of a 
structure in an 
ultrasound image 
depends on the 
strength of the 
reflection, or echo. 
• This in turn 
depends on the 
difference in 
acoustic 
impedance of the 
two materials.
Angle of incidence 
• Angle b/n sound waves 
and reflecting surfaces 
• The more the angle , 
the less the reflection 
• In medical USG, 
reflected sound is not 
detected when angle is 
greater than 3o
Reflection 
• Two distinct patterns of reflection give rise to 
the echoes that make up an ultrasound image 
– specular reflection and 
– scattering
Specular reflection 
• When an incident ultrasound pulse encounters a large, 
smooth interface between two types of tissue with 
different acoustic impedance values , the result is a 
partially reflected echo that travels back toward the 
transducer and a partially transmitted pulse that travels 
deeper into the patient. 
• Responsible for the bright appearance of fibrous structures 
such as tendons and of boundaries between different 
tissues.
Scattering / Diffuse Reflection 
• If the ultrasound pulse encounters reflectors whose 
dimensions (d) are smaller than the ultrasound 
wavelength (ie, d << λ), scattering occurs. 
• Scattering gives rise to the characteristic texture 
(echo texture) of the image seen within soft tissue.
Refraction 
• Bending of waves as they pass from 
one medium to other 
• The change in wavelength and 
direction of propagation of sound 
occurs, but frequency remains 
constant 
• Artifacts due to refraction are 
– Loss of resolution of image 
– Spatial distortion
Snell’s Law 
풔풊풏 휽풊 
풔풊풏 휽풕 
= 
풗ퟏ 
풗ퟐ
Absorption / Attenuation 
• Refers to conversion of 
ultrasound energy to 
heat energy
Absorption / Attenuation 
• Depends On 
– frequency of sound (increase absorption) 
– viscosity of medium (increase absorption) 
– relaxation time of medium 
– temperature (varies with tissues)
• The transmission of sound waves without 
much energy loss can be done by 
– using mineral oil b/n transducer and patient skin 
– mechanical impedance matching 
• Any process that lessens the reflection. it is done by a 
coupler who has intermediate density b/n transducer 
and tissue 
• The thickness of this matching layer must be equal to 
one fourth the wavelength of sound in the matching 
layer (Quarter-wave matching)
IMAGE DISPLAY 
• Electronic representation of data generated 
from returning signals and displayed on TV 
monitor 
– A MODE 
– M MODE 
– Real time B MODE
AMPLITUDE MODE 
• Echoes are displayed in 
the form of spikes on 
CRO traced along time 
base 
• Amplitude of spike 
measures the echo size 
• Information about the 
depth of the structures 
and the amplitude of 
the returning echo
AMPLITUDE MODE 
• Used in 
– Ophthalmology 
– Echoencephalography 
– Echocardiography. 
• Disadv: 
– 1D information, 
– takes lot of space in CRO
M-MODE & TM-MODE 
• Detects motion of 
structures- cardiac valves 
and of cardiac chambers , 
vessels 
• Echoes are displayed as 
dots of varying intensity 
• M-Mode does not have 
Time factor. 
• It also provides 1D 
information 
• TM-Mode has Time factor
BRIGHTNESS MODE 
• Echo signals as – Line of dots. 
• Intensity of dot gives relative size of echo 
• It provides depth of information and 
variations in direction of beam
REAL TIME B MODE 
• System is fast enough to allow movements to 
be followed. 
• At least 16 frames / sec
GRAY SCALE IMAGING 
• Display variation of amplitudes of echoes 
arising from tissues as varying shades of gray.
Controls
Controls 
• Adjustment of image to obtain Optimal Gray 
scale image done using several controls: 
– TGC 
– Coarse Gain 
– Intensity governs the amplitude 
– Reject 
– Delay 
– Near Gain - diminish near echoes 
– Far Gain - enhance distant echoes
TGC 
• Time Gain Compensator 
• Compensates differences in echo 
strength by adjusting the 
variations in degree of 
amplification 
• The slope of the TGC adjusts the 
degree of amplification 
• The delay control regulates the 
depth at which the TGC begins to 
augment weaker signals
Pulse Rate 
• The number of separate little packets of sound 
that are sent out each second 
• It determines the total number of echoes 
returning to the transducer in a unit of time. 
• High pulse rate is desirable. 
• But then the receiving time decreases. 
• So, the pulse rate must be set to 
accommodate the thickest part that might be 
examined.
Spatial Resolution
• In ultrasound, the major factor that limits the spatial resolution 
and visibility of detail is the volume of the acoustic pulse. 
• The axial, lateral, and elevational (slice thickness) dimensions 
determine the minimal volume element. 
• Each dimension has an effect on the resolvability of objects in 
the image.
Axial Resolution 
• Also known as linear, range, longitudinal, or 
depth resolution 
• Refers to the ability to discern two closely 
spaced objects in the direction of the beam. 
• Achieving good axial resolution requires that 
the returning echoes be distinct without 
overlap.
• The minimal required separation distance between two 
reflectors is one-half of the spatial pulse length (SPL) to avoid 
the overlap of returning echoes, as the distance traveled 
between two reflectors is twice the separation distance.
Lateral Resolution 
• Lateral resolution, also known as azimuthal 
resolution 
• Refers to the ability to discern as separate two 
closely spaced objects perpendicular to the 
beam direction.
• For both single 
element transducers 
and multielement 
array transducers, the 
beam diameter 
determines the lateral 
resolution.
• Since the beam 
diameter varies with 
the distance from the 
transducer in the near 
and far field, the 
lateral resolution is 
depth dependent. 
• The best lateral 
resolution occurs at 
the near field—far 
field face.
• At this depth, the 
effective beam 
diameter is 
approximately equal to 
half the transducer 
diameter. 
• In the far field, the 
beam diverges and 
substantially reduces 
the lateral resolution.
• The typical lateral resolution for an unfocused 
transducer is approximately 2 to 5 mm. 
• A focused transducer uses an acoustic lens (a 
curved acoustic material analogous to an 
optical lens) to decrease the beam diameter at 
a specified distance from the transducer.
Elevational Resolution 
• The elevational or slice-thickness dimension of 
the ultrasound beam is perpendicular to the 
image plane. 
• Slice thickness plays a significant part in image 
resolution, particularly with respect to volume 
averaging of acoustic details in the regions 
close to the transducer and in the far field 
beyond the focal zone.
• Elevational resolution 
is dependent on the 
transducer element 
height in much the 
same way that the 
lateral resolution is 
dependent on the 
transducer element 
width.
Scanning the Ultrasound Beam
Types: 
• Mechanical Scanners 
• Electronic Array Scanners
Mechanical scanners 
• Early US systems relied on 
the operator to manually 
change the position and 
orientation of the 
transducer and scan the 
ultrasound beam through a 
plane in the patient to 
obtain the echo data 
necessary for each image.
• 3 types: 
Mechanical scanners 
– Oscillating Transducer (Unenclosed type) 
– Oscillating Transducer (Enclosed type) 
– Rotating Wheel Transducer
Mechanical scanners 
• This was a time-consuming process but produced images that 
covered large FOVs 
• Image artefacts occurred if the patient moved during the acquisition
Electronic array scanning 
• Most modern US imagers automatically scan 
the ultrasound beam using transducers 
consisting of arrays of many narrow 
piezoelectric elements. 
• The array may consist of as many as 128–196 
elements
• Two modes of activation are used to produce a beam. 
– These are the “linear” (sequential) and “phased” 
activation/receive modes.
Linear 
• The linear array scanners produce sound 
waves parallel to each other and produces 
a rectangular image. 
• The width of the image and number of scan 
lines are the same at all tissue levels. 
• Often used with high frequencies ie 7MHz. 
• Advantage -good near field resolution. 
• Disadvantage is artifacts when applied to a 
curved part of the body creating air gaps 
between skin and transducer.
Sector/Vector 
• Produces a fan like 
image that is narrow 
near the transducer and 
increase in width with 
deeper penetration. 
• The disadvantge is poor 
near field resolution.
Curved 
• Often with frequencies 
of 2 - 5 MHz (to allow 
for a range of patients 
from obese to slender). 
• The array of elements is 
arranged across a 
convex arc (instead of a 
straight line), which 
rapidly scans a larger 
FOV
Endoscopic Ultrasound 
• Dedicated linear-array and radial 
echo-endoscopes for structural 
evaluation of the luminal wall 
adjacent tissues. 
• Flexible shaft with a central wire 
that drives rotation of a 
mechanical transducer at the tip. 
• The transducer is surrounded by 
oil that serves as an acoustic 
interface with tissue, 
providing 360-degree imaging
• With the introduction of 
the curved linear array 
echoendoscope , the 
indications for EUS have 
expanded. 
• Allows the endoscopist to 
perform a whole range of 
interventional applications 
ranging from fine needle 
aspiration (FNA) of lesions 
surrounding the 
gastrointestinal tract to 
celiac plexus block and 
drainage of pancreatic 
pseudocyst
• Ultrasound imaging of vascular and non-vascular lumina 
employing miniaturized high-frequency (20-30 
MHz) transducers inserted into catheters as small as 2 mm 
in external diameter
Recent Innovations in B-mode US 
• Tissue Harmonic Imaging 
• Spatial Compound Imaging 
• Extended FOV Imaging 
• Coded Pulse Excitation 
• 3D and 4D imaging 
• Elastography. 
• Ultrasound contrast media.
Tissue Harmonic Imaging 
• A musical note has three characteristics. 
– Pitch (or frequency) 
– Loudness 
– Quality (or tone). 
• Quality is the audible difference heard between two 
musical notes of the same pitch and loudness. 
• Hence, a piano based C-note does not sound the same 
as a C-note played on a guitar, due to the existence of 
harmonic frequencies. 
• The same note from various instruments has different 
qualities because the sounds are not pure notes i.e. of 
one frequency and these integral multiples of the 
fundamental frequency (overtones) give an instrument 
its characteristic sound.
Tissue Harmonic Imaging 
• Harmonics are frequencies that occur at 
multiples of the fundamental or transmitted 
sonographic frequency. 
• In conventional gray-scale sonography, the same 
frequency spectrum that is transmitted into the 
patient is subsequently received to produce the 
sonographic image. 
• In THI sonography, higher harmonic frequencies 
generated by propagation of the ultrasound 
beam through tissue are used to produce the 
sonogram.
Why harmonics? 
• The ultrasonic pulse gets altered 
with time as it traverses the 
tissues with non-linear motion. 
• The peaks within the pulse 
waveform move faster than the 
troughs because the propagation 
speed is higher in compressed 
regions of tissue than in the areas 
which are expanded by the 
passing pressure wave. 
• The degree of such acoustic 
signal distortion in tissue depends 
on the amplitude of the emitted 
pulse and the distance it has 
travelled in the tissue. 
Drawing of undistorted pressure wave (top) and of 
pressure wave after undergoing nonlinear propagation 
(bottom). Nonlinear effects cause high-pressure regions 
of sound wave to travel faster than low-pressure 
regions and result in progressive distortion of 
transmitted wave with generation of sound at higher 
harmonics of transmit frequency
• As much of the fundamental sonographic signal 
as possible must be removed to make these 
theoretical harmonic improvements a clinical 
reality. 
• This can be done by either frequency- based or 
phase inversion methods. 
• There are 2 basic harmonic imaging methods. 
– conventional frequency-based second-harmonic 
imaging 
– pulse inversion or phase inversion harmonic imaging
Conventional Frequency-based 
Second-harmonic Imaging 
• A narrowband pulse is 
emitted, and then high-pass 
or narrow-bandpass 
filtering is 
applied to the received 
echoes to filter out the 
fundamental echo 
components. 
• This results in reduction 
of both spatial and 
contrast resolution.
Phase Inversion Harmonic Imaging 
• In wideband harmonic imaging 
(also called pulse inversion or 
phase inversion harmonic 
imaging) 
• A train of 2 pulses is emitted, 
with the phase of the second 
inverted relative to the phase 
of the first. 
• When the echoes from the 
transmitted pulses are added, 
the linear components of the 
echoes cancel each other, 
whereas the nonlinear 
components are amplified. 
• This results in superior 
contrast and spatial resolution. 
Two phase-inverted but otherwise identical 
sonographic pulses are transmitted. Summing the 
returning echoes in a buffer cancels most of the 
fundamental and odd harmonic echoes and effectively 
amplifies the second harmonic.
ADVANTAGES OF THI 
1. Improve lateral resolution 
2. Reduce side lobe artifacts 
3. Improve signal to noise ratio 
4. Improved near field and far field image 
quality
ADVANTAGES OF THI 
5. Lesions are clearer & better defined. 
6. Use of higher frequencies improves 
resolution. 
7. Helps differentiate cysts from hypoechoic 
solid masses. 
8. Better clarity of contents. 
9. It is superior to conventional USG in 
visualization of lesions containing highly 
reflective tissues like fat, calcium & air.
ADVANTAGES OF THI 
10.Harmonic imaging is generally considered to 
be most useful for “technically difficult” 
patients with thick and complicated body 
wall
THI is useful in 
• Better defining the borders of pancreas 
• Better visualization of lower pole of left kidney 
• Better visualization of fatty livers 
• Small renal cysts can be easily seen 
• Presence or absence of sludge in GB 
• Metastatic lesions in liver easily seen 
• Minimal fluid present between liver and kidney 
• Provides better images of aortic wall, IVC, portal vein 
and renal arteries
Spatial Compound Imaging 
• Electronic steering of ultrasound beams from 
an array transducer is used to image the same 
tissue multiple times by using parallel beams 
oriented along different directions. 
• The echoes from these multiple acquisitions 
(upto 9 sets) are then averaged together into a 
single composite image.
Disadvantage 
• More time is required for data acquisition 
• The compound imaging frame rate is reduced 
compared with that of conventional B-mode 
imaging.
Advantage 
• Spatial compound images often show reduced 
levels of speckle, noise, clutter, and refractive 
shadows and improved contrast and margin 
definition 
• Enhancement and shadowing artifacts may 
also be reduced, which may be an advantage 
or potential drawback, depending on the 
imaging situation.
Clutter Suppression 
Conventional image: acoustic clutter 
produces spurious echoes within a simple 
cyst (inset), which can be difficult to 
eliminate regardless of incident angle or 
adjustment of system parameters. 
Compound image: compound Imaging 
results in significant clutter suppression in 
the cyst. Also note the good depth of field in 
this single focus compound image, 
comparable to the multi-focus conventional 
image.
Extended FOV Imaging 
• A benefit of early static B-mode scanners that was lost 
with the introduction of mechanical and electronic 
automatic scanning was large imaging FOVs. 
• Extended FOV imaging, has sought to restore this 
capability. 
• The transducer is slowly translated laterally across the 
large anatomic region of interest. During this motion, 
multiple images are acquired from many transducer 
positions. 
• The registered image data are accumulated in a large 
image buffer and then combined to form the complete 
large FOV image
• Extended FOV images are not limited just to B-mode 
acquisitions. 
• Restores the capability of visualizing large 
anatomic regions in a single image and simplifies 
measurements made over these large regions. 
• It is useful in evaluating aneurysms, detecting 
small ligament and tissue damage, and to image 
and calibrate long sections of blood vessels, to 
locate anatomical landmarks for location of 
disease.
Coded Pulse Excitation 
• A fundamental trade-off in US is that between 
imaging depth and spatial resolution. 
• Coded ultrasound pulses help overcome this 
limitation, providing good penetration at the 
higher frequencies necessary for high spatial 
resolution. 
• In this imaging approach, long ultrasound pulses 
are used instead of the very short pulses. These 
long pulses carry greater ultrasonic energy, 
increasing the energy of echoes that return from 
large depths in the patient.
The coded pulses are produced with a very specific, characteristic 
shape, and the resulting echoes will have a similar shape.
The end result is an image with good echo signal and good spatial resolution at large depths. 
Conventional B-mode (a) and coded pulse (b) US images of the liver show the benefits of coded pulse imaging. The spatial 
resolution of the coded pulse image (b) is very comparable with that of the 13-MHz conventional image (a). However, the 
useful imaging depth is about 7.5 cm for the coded pulse image (b) compared with only about 2.8 cm for the conventional 
image (a).
ELASTOGRAPHY 
• Dynamic technique that uses ultrasound to 
provide information on tissue stiffness by 
measuring the degree of distortion under the 
application of an external force - viscoelastic 
properties of tissue 
• Surrogate for that obtained with manual 
palpation. 
• Principle: malignant tissues have more stiffness 
and are harder to distort. On tissue compression 
strain (displacement) within harder tissue is less 
than in softer tissue.
Layers in Jello show soft and hard materials after light compression
Important Quantities 
• Young’s modulus (E) describes longitudinal 
deformation in terms of strain 
• The shear modulus (G) relates to transverse strain to 
transverse stress 
• The bulk modulus (K) of elasticity describes the change 
in volume 
• Poisson ratio (v) which is a ratio of transverse 
contraction per unit breadth divided by longitudinal 
extension per unit length 
Shear and Young’s moduli, are the most suitable elasticity 
parameters to measure.
The tissue is insonified a) before and b) after a small uniform compression. 
In the harder tissues (e.g. the circular lesion depicted) the echoes will be less distorted than in 
the surrounding tissues, denoting thus smaller strain
Elastography - methods 
• Three methods— 
a) spatial correlation method, 
b) the phase-shift tracking method, and 
c) the combined autocorrelation method (CAM)— 
• have been introduced for measuring tissue 
strain at elastography.
Elastography - Uses 
• In the liver it is a very useful method in 
depicting liver hardness and prediction and 
prognostication of liver cirrhosis. 
• Breast imaging, cervix imaging, 
musculoskeletal imaging (especially 
supraspinatus and tendoachilles), thyroid 
lesions assessment and assessment of cervical 
nodes.
3D and 4D ultrasound 
• 3D ultrasound - data set that contains a large number 
of 2D planes (B-mode images). 
• This volume data can be manipulated in different 
planes by rotation. 
• Dissected in any plane, to get multiplanar images 
(similar to multislice CT). 
• Special probes and software are necessary in order to 
perform 3D and 4D imaging. 
• 4D ultrasound is also known as "Real-time 3D 
Ultrasound". The 3D datasets with their multiplanar 
reformations and renderings in real time, give us 
photographic quality images.
Techniques of 3D US 
• Data acquisition 
• Image reconstruction and 
• Image display
Data Acquisition Techniques 
a) Tracked freehand systems, 
b) Untracked freehand systems, 
c) Mechanical assemblies, and 
d) 2D arrays 
• Regardless of which method is used, one must know 
the relative position and angulation of each 2D image 
and must acquire the images rapidly or with gating to 
avoid motion artifacts. If these two criteria are not 
met, the 3D images may be inaccurate.
(a) Tracked Freehand Systems 
• The operator holds an assembly composed of 
the transducer with an attachment and 
manipulates it over the anatomic area being 
evaluated 
• Ensure that there are no significant imaging 
gaps.
• Acoustically tracked 3D scanning 
• Articulated-arm-tracked 3D scanning 
• Magnetic field-tracked 3D scanning
(b) Untracked Freehand 
• 2D images are digitized as the operator moves 
the transducer with a smooth, steady motion 
• Most convenient for the operator 
• Image quality is variable and depends largely on 
how smoothly and steadily the operator moves 
the transducer 
• Geometric measurements may be inaccurate 
because there is no direct information regarding 
the relative position of the digitized images.
• Mechanical movement of the transducer 
across the skin. 
• Tilted about a fixed point on the skin surface. 
• Rotated about its own axis.
(c) Mechanical Assemblies 
• The transducer is propelled or rotated 
mechanically, and 2D images are digitized at 
predetermined spatial or angular intervals . 
• Cumbersome for the user 
• Improve the geometric accuracy 
• To date, their greatest utility has been in 
intracavitary and intraluminal examinations, in 
which the area of interest is relatively small 
and motion artifact is less of a problem.
• Three different scanning techniques can be used: 
– linear scanning, 
– tilt scanning, and 
– rotational scanning.
(d) Two-dimensional Arrays 
• With the first three types of data acquisition 
systems, mechanical motion is used to obtain 3D 
images. 
• An alternative is to keep the transducer stationary 
and use electronic scanning with a 2D transducer 
array generates pyramidal or conical US pulses to 
generate 3D information in real time
Image Reconstruction Techniques 
• Two types: 
– 3D surface model 
– Voxel-based volume model.
Three-dimensional Surface Model: 
• Outline the boundaries of the areas of interest 
on the 2D images manually or with a 
computer algorithm 
• Reduces the amount of 3D data needed 
• Shorter 3D reconstruction times and greater 
efficiency 
• Identification of boundaries can be tedious 
and time-consuming.
Voxel-based Volume Model 
• The computer builds a 3D voxel-based volume 
(3D grid) 
• This process preserves the original 
information 
• Allows a variety of rendering techniques 
• Generates very large data files, which slows 
processing and requires large amounts of 
computer memory.
Image Display Techniques 
• Surface Rendering: operator identifies the 
boundaries of pertinent structures either 
manually or with an algorithm 
• Multiplanar Reformatting: three orthogonal 
planes or texture mapping – image rotated to 
obtain the desired image orientation 
• Combined Surface Rendering and 
Multiplanar Reformatting. 
• Volume Rendering
Artefacts 
• Respiratory motion artifact 
• Artifact caused by incorrect calibration
Advantages of 3D US 
• Many benefits in obstetrical and gynaecological 
scanning where it is important to analyze anatomy. It 
has shown promising results in diagnosing cleft 
lip/palate, spina bifida, polydactyly, club foot, facial 
dysmorphism, low set ears ,fetal cardiac imaging. 
• Excellent tool for demonstrating and accurately 
diagnosing congenital uterine anomalies, pelvic floor 
muscle and sphincters. 
• Considerable interobserver variability in 2D; exact 
relationship between anatomic structures is accurately 
recorded in the 3D image
Advantages of 3D US 
• Unrestricted access to an infinite number of 
viewing planes. 
• Suited for monitoring the effects of therapy 
over a long period of time 
• More accurate quantitative volume estimates
Limitations of 3D US 
• More cumbersome , requiring more user input 
• Data archiving and communication more challenging 
• Waiting for the 3D image to appear can be frustrating 
to users 
• Slow the image interpretation process, especially if 
inexperienced, as the right algorithm needs to be 
chosen. 
• The ability to obtain a good 3-D picture is very much 
dependent on operator skill, the amount of liquor, 
fetus position and the degree of maternal obesity.
Applications for 3D US 
• Fetal Imaging
Applications for 3D US 
• Gynecologic Imaging 
volume data sets to be acquired with both 
transvaginal and transabdominal probes
Applications for 3D US 
• Three-dimensional Power Doppler Imaging 
3D US angiography 
• Prostate Imaging
Applications for 3D US 
• Breast Imaging: demonstrate lesion margins 
and topography 
• Biopsy-related Imaging: needle localization 
and guidance
ULTRASOUND CONTRAST MEDIA 
• Microbubbles of air or other gases which act as echo 
enhancers 
• Microbubbles are less than 10 μm in diameter, and 
• Contrary to most other contrast media which are 
rapidly distributed to the extravascular, extracellular 
space, most microbubbles are confined to the vascular 
space. 
• Microbubbles may produce up to 25 dB (more than 
300-fold) increase in echo strength.
• Stability is increased by 
– external bubble encapsulation (galactose, 
phospholipids, denatured albumin or poly-butyl-cyanoacrylate) 
with or without surfactants 
– using gases with a low diffusion coefficient 
(perfluorocarbons) or 
– a combination of both 
• The gas components of the microbubbles are 
normally eliminated via the lungs. 
• Stabilizing components are eliminated via the 
hepato-renal route.
Contrast Microbubbles 
• The ideal diameter - 
2 μmto 8 μm 
• Enhancement life-time 
of the microbubble, 
often several minutes 
(8–10 min)
Microbubbles – Structure 
• The coatings include 
albumin, gelatin, 
galactose microspheres, 
polyglutaminic acid, 
lipophilic monolayer 
surfactants, and lipid 
bilayers (liposomes). 
• The gas inside the shell 
may be either air or 
various perfluorocarbons, 
which are liquids at room 
temperature but gas at 
body temperature.
The ideal USCA 
• Non-toxic, 
• Injectable intravenously, 
• Capable of crossing the pulmonary capillary 
bed after a peripheral injection, and 
• Stable enough to achieve enhancement for 
the duration of the examination.
Microbubbles - Generations 
• Several "generations" of gas microbubble contrast media 
have evolved; 
• The "1st generation" products do not pass the pulmonary 
vascular bed, and are therefore limited to the venous 
system and the right heart cavities after injection. 
• The "2nd generation" contrast media are both sufficiently 
small and stable to pass into the systemic circulation, and 
these contrast media enhance the doppler signal in various 
arteries after injection. They are short-lived, however, the 
effect is over in a few minutes.
Microbubbles - Generations 
• The "3rd generation" gas microbubble 
contrast media are even more echogenic and 
stable, and are able to enhance the 
echogenicity of parenchyma on B-mode 
images. They may thus show perfusion, even 
in such a difficult region as the myocardium. 
• The various gas microbubble contrast media 
are generally safe with low toxicity in humans.
Ultrasound Contrast Agents - Types 
• Different types of ultrasound contrast agents: 
– tissue contrast agents; 
– contrast agents for vascular enhancement; 
– agents for targeted contrast imaging.
PRINCIPLES 
• The main mechanisms for signal enhancement are 
– backscattering, 
– bubble resonance and 
– bubble rupture. 
• These mechanisms are highly dependent on the acoustic 
power of the transmitted ultrasound, which is reflected by the 
mechanical index (MI). 
• very high reflectivity ( gas content makes them very 
reflective)
Acoustic Excitation of Ultrasound 
Contrast Agents 
• The behavior of these agents under acoustic 
excitation fall into three classes (Frinking et 
al., 1999), depending on the structure of the 
microbubble and the level of the insonifying 
pressure amplitude and frequency: 
– stable linear (low MI), 
– stable nonlinear scattering (medium MI) 
– transient nonlinear scattering (high MI).
Back Scattering 
• At low acoustic power (MI < 0.1), gas microbubbles may be 
regarded as point scatterers, and the mechanism of 
ultrasound reflection is that of Rayleigh Tyndall scattering. 
• The scattering strength of a point scatterer is proportional to 
the sixth power of the particle radius and to the fourth power 
of the ultrasound frequency; the echogenicity of such contrast 
media is therefore highly dependent upon particle size and 
transmit frequency. 
• The backscattered intensity of a group of point scatterers is 
furthermore directly proportional to the total number of 
scatterers in the insonified volume; the concentration of the 
contrast medium is therefore also of importance.
MI- 0.1-0.3 
Reflect and scatter
Bubble Resonance and Harmonics 
• At intermediate acoustic power (0.1 < MI < 0.5) gas 
microbubbles may show strong oscillatory motion provided 
the frequency of the incident ultrasound is close to the 
resonant (fundamental) frequency of the microbubbles. 
MICROBUBBLE AS LINEAR RESONATOR 
• By virtue of their compressibility, microbubbles display unique 
properties in an ultrasound beam, which sets them into 
resonance when there is a match between their diameter and 
the ultrasonic wavelength which for microbubbles 2-7microns 
range at ultrasound frequency 2-10 Mhz.
MI- 0.5 
Resonates
MICROBUBBLE AS NONLINEAR RESONATOR 
• nonlinear behaviour (compression and expansion phases) 
produces echoes that contain frequencies not present in 
the transmitted pulses. 
• An elegant way to extract these nonlinear signals is to send 
a series of pulses down each line, varying their phase and 
amplitude; the returning signals are combined to cancel the 
linear signals from tissue and the remaining bubble-specific 
signals are used to form an image that can be presented as 
a colour overlay on the B-mode image or shown on a side-by- 
side display, all in real-time.
MI- 0.7-0.8 
Oscillates in non harmonic pattern
Bubble rupture 
• At high acoustic power (MI > 0.5), ultrasound at the 
microbubble resonance frequency will cause the bubbles to 
rupture. 
• The result is a transient high-amplitude, broadband signal 
containing all frequencies, not only the harmonics. 
• It will create a transient, strong signal in B-mode, or a 
short-lasting multicoloured, mosaic-like effect in colour 
Doppler sonography. 
• Several terms for the strong, transient signal have been 
proposed: induced or stimulated acoustic emission, loss of 
correlation imaging and sono-scintigraphy.
MI- =/>1 
Disrupt and Burst
Applications 
• Give angiographic capacity to ultrasound ie help in 
demonstrating tissue perfusion, bleeding 
points,vascularity of lesions etc. 
• Widely used in imaging of solid organs, particularly the 
liver 
• Improvement in detection of colour Doppler signal 
from large vessels 
• Microbubble contrast has also found a niche outside 
the vascular compartment in the setting of vesico-ureteric 
reflux in children where a high sensitivity and 
specificity compared with conventional micturating 
cystourography (MCUG) has been demonstrated
Applications 
• Gene therapy and targeted delivery of drugs
Conclusion 
• Modern US equipment is based on many of the same 
fundamental principles employed in the initial devices 
used for human imaging over 50 years ago. 
• US has the characteristics of being relatively 
inexpensive, portable, safe, and real-time in nature, all 
of which make it one of the most widely used imaging 
modalities in medicine. 
• In short, US science, technology, and applications are 
expanding at a brisk pace and are far from mature. 
• Even more exciting developments are on the horizon.
Ultrasound Characteristics and Components

Contenu connexe

Tendances

Principle of usg imaging, construction of transducers
Principle of usg imaging, construction of transducersPrinciple of usg imaging, construction of transducers
Principle of usg imaging, construction of transducersDev Lakhera
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifactsansaripv
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging MAMTA PANDA
 
Ultrasound Transducer Types
Ultrasound Transducer TypesUltrasound Transducer Types
Ultrasound Transducer TypesMette Lyng
 
Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasoundu.surgery
 
MRI spin echo pulse sequences
MRI spin echo pulse sequencesMRI spin echo pulse sequences
MRI spin echo pulse sequencesShiva Prakash
 
Color Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsColor Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsSyed Yousaf Gilani
 
Introduction to ultarsound machine and physics
Introduction to ultarsound machine and physicsIntroduction to ultarsound machine and physics
Introduction to ultarsound machine and physicsmanishyadav513
 
Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Prajwith Rai
 
Doppler Effect - Ultrasound
Doppler Effect - UltrasoundDoppler Effect - Ultrasound
Doppler Effect - UltrasoundVictor Ekpo
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkDr pradeep Kumar
 
Ultrasound Basic Knobology & Controls
Ultrasound Basic Knobology & ControlsUltrasound Basic Knobology & Controls
Ultrasound Basic Knobology & ControlsMadhuka Perera
 

Tendances (20)

Principle of usg imaging, construction of transducers
Principle of usg imaging, construction of transducersPrinciple of usg imaging, construction of transducers
Principle of usg imaging, construction of transducers
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging
 
Basics Physics of ultrasound
Basics Physics of ultrasoundBasics Physics of ultrasound
Basics Physics of ultrasound
 
Ultrasound Transducer Types
Ultrasound Transducer TypesUltrasound Transducer Types
Ultrasound Transducer Types
 
CT Artifacts
CT ArtifactsCT Artifacts
CT Artifacts
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
 
Usg diplay modes
Usg diplay  modesUsg diplay  modes
Usg diplay modes
 
Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasound
 
MRI spin echo pulse sequences
MRI spin echo pulse sequencesMRI spin echo pulse sequences
MRI spin echo pulse sequences
 
CT artifact
CT artifact CT artifact
CT artifact
 
Color Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsColor Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physics
 
Introduction to ultarsound machine and physics
Introduction to ultarsound machine and physicsIntroduction to ultarsound machine and physics
Introduction to ultarsound machine and physics
 
Usg artifacts
Usg artifactsUsg artifacts
Usg artifacts
 
Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
Doppler Effect - Ultrasound
Doppler Effect - UltrasoundDoppler Effect - Ultrasound
Doppler Effect - Ultrasound
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
 
Ultrasound Basic Knobology & Controls
Ultrasound Basic Knobology & ControlsUltrasound Basic Knobology & Controls
Ultrasound Basic Knobology & Controls
 
X ray generators
X ray generatorsX ray generators
X ray generators
 

En vedette (20)

Echocardiogram Basics
Echocardiogram BasicsEchocardiogram Basics
Echocardiogram Basics
 
Production of x rays & generators
Production of x rays & generatorsProduction of x rays & generators
Production of x rays & generators
 
Imaging in haemoptysis
Imaging in haemoptysisImaging in haemoptysis
Imaging in haemoptysis
 
Skull base imaging
Skull base imagingSkull base imaging
Skull base imaging
 
Ct skull base
Ct skull baseCt skull base
Ct skull base
 
SKULL BASE IMAGING
SKULL BASE IMAGINGSKULL BASE IMAGING
SKULL BASE IMAGING
 
Pcpndt act
Pcpndt actPcpndt act
Pcpndt act
 
Ct & mr enterography
Ct & mr enterographyCt & mr enterography
Ct & mr enterography
 
Skull base : Development and anatomy.
Skull base : Development and anatomy. Skull base : Development and anatomy.
Skull base : Development and anatomy.
 
Imaging in Skull base
Imaging in Skull baseImaging in Skull base
Imaging in Skull base
 
Ultrasound diagnostics fin
Ultrasound diagnostics finUltrasound diagnostics fin
Ultrasound diagnostics fin
 
Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
 
Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
 
076 cardiac magnetic resonance imaging
076 cardiac magnetic resonance imaging076 cardiac magnetic resonance imaging
076 cardiac magnetic resonance imaging
 
Gre&SE
Gre&SEGre&SE
Gre&SE
 
What is spatial Resolution
What is spatial ResolutionWhat is spatial Resolution
What is spatial Resolution
 
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
 
Mr spectroscopy
Mr spectroscopyMr spectroscopy
Mr spectroscopy
 
Mr fluoroscopy
Mr fluoroscopyMr fluoroscopy
Mr fluoroscopy
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 

Similaire à Ultrasound Characteristics and Components

usgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdfusgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdflittlealphonsa
 
Ultrasound imaging system
Ultrasound imaging system Ultrasound imaging system
Ultrasound imaging system Riadh Al-Haidari
 
Ultrasoundtherapy 150201220726-conversion-gate02 (1)
Ultrasoundtherapy 150201220726-conversion-gate02 (1)Ultrasoundtherapy 150201220726-conversion-gate02 (1)
Ultrasoundtherapy 150201220726-conversion-gate02 (1)Michael Selvaraj
 
Ultrasonography final
Ultrasonography finalUltrasonography final
Ultrasonography finalravikumarbme
 
production of ultrasound and physical characteristics-
production of ultrasound and physical characteristics-production of ultrasound and physical characteristics-
production of ultrasound and physical characteristics-Lushinga Mourice
 
Ultrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfUltrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfghadaElbanby1
 
4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.pptRohit Bansal
 
Lecturer 4 Therapeutic Ultrasound.ppt
Lecturer 4 Therapeutic Ultrasound.pptLecturer 4 Therapeutic Ultrasound.ppt
Lecturer 4 Therapeutic Ultrasound.pptmohammad kraizem
 
Presentation GROUP TWO.pptx
Presentation GROUP TWO.pptxPresentation GROUP TWO.pptx
Presentation GROUP TWO.pptxYasinKidiri
 
TRANSDUCER MODES OF OPERATION.pptx
TRANSDUCER MODES OF OPERATION.pptxTRANSDUCER MODES OF OPERATION.pptx
TRANSDUCER MODES OF OPERATION.pptxYasinKidiri
 

Similaire à Ultrasound Characteristics and Components (20)

usgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdfusgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdf
 
Ultrasound Physics Jr.pptx
Ultrasound Physics  Jr.pptxUltrasound Physics  Jr.pptx
Ultrasound Physics Jr.pptx
 
Ultrasound imaging system
Ultrasound imaging system Ultrasound imaging system
Ultrasound imaging system
 
Ultrasoundtherapy 150201220726-conversion-gate02 (1)
Ultrasoundtherapy 150201220726-conversion-gate02 (1)Ultrasoundtherapy 150201220726-conversion-gate02 (1)
Ultrasoundtherapy 150201220726-conversion-gate02 (1)
 
Ultrasound therapy
Ultrasound therapyUltrasound therapy
Ultrasound therapy
 
Us transducers (2)
Us transducers (2)Us transducers (2)
Us transducers (2)
 
USphysics.pptx
USphysics.pptxUSphysics.pptx
USphysics.pptx
 
Ultrasonography final
Ultrasonography finalUltrasonography final
Ultrasonography final
 
Usg physics
Usg physicsUsg physics
Usg physics
 
production of ultrasound and physical characteristics-
production of ultrasound and physical characteristics-production of ultrasound and physical characteristics-
production of ultrasound and physical characteristics-
 
Ultrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfUltrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdf
 
4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt
 
ultrsound.pptx
ultrsound.pptxultrsound.pptx
ultrsound.pptx
 
Lecturer 4 Therapeutic Ultrasound.ppt
Lecturer 4 Therapeutic Ultrasound.pptLecturer 4 Therapeutic Ultrasound.ppt
Lecturer 4 Therapeutic Ultrasound.ppt
 
Presentation GROUP TWO.pptx
Presentation GROUP TWO.pptxPresentation GROUP TWO.pptx
Presentation GROUP TWO.pptx
 
ultrsound.pptx
ultrsound.pptxultrsound.pptx
ultrsound.pptx
 
U ltrasound physiscs
U ltrasound physiscsU ltrasound physiscs
U ltrasound physiscs
 
Ultrasound physics
Ultrasound physicsUltrasound physics
Ultrasound physics
 
TRANSDUCER MODES OF OPERATION.pptx
TRANSDUCER MODES OF OPERATION.pptxTRANSDUCER MODES OF OPERATION.pptx
TRANSDUCER MODES OF OPERATION.pptx
 
Multimedia
MultimediaMultimedia
Multimedia
 

Dernier

LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS  CLASSIFICATIONS.pdfLUNG TUMORS AND ITS  CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS CLASSIFICATIONS.pdfDolisha Warbi
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAAjennyeacort
 
Nutrition of OCD for my Nutritional Neuroscience Class
Nutrition of OCD for my Nutritional Neuroscience ClassNutrition of OCD for my Nutritional Neuroscience Class
Nutrition of OCD for my Nutritional Neuroscience Classmanuelazg2001
 
Clinical Pharmacotherapy of Scabies Disease
Clinical Pharmacotherapy of Scabies DiseaseClinical Pharmacotherapy of Scabies Disease
Clinical Pharmacotherapy of Scabies DiseaseSreenivasa Reddy Thalla
 
Big Data Analysis Suggests COVID Vaccination Increases Excess Mortality Of ...
Big Data Analysis Suggests COVID  Vaccination Increases Excess Mortality Of  ...Big Data Analysis Suggests COVID  Vaccination Increases Excess Mortality Of  ...
Big Data Analysis Suggests COVID Vaccination Increases Excess Mortality Of ...sdateam0
 
Informed Consent Empowering Healthcare Decision-Making.pptx
Informed Consent Empowering Healthcare Decision-Making.pptxInformed Consent Empowering Healthcare Decision-Making.pptx
Informed Consent Empowering Healthcare Decision-Making.pptxSasikiranMarri
 
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATROApril 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATROKanhu Charan
 
SGK HÓA SINH NĂNG LƯỢNG SINH HỌC 2006.pdf
SGK HÓA SINH NĂNG LƯỢNG SINH HỌC 2006.pdfSGK HÓA SINH NĂNG LƯỢNG SINH HỌC 2006.pdf
SGK HÓA SINH NĂNG LƯỢNG SINH HỌC 2006.pdfHongBiThi1
 
Measurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxMeasurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxDr. Dheeraj Kumar
 
Case Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxCase Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxNiranjan Chavan
 
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
PULMONARY EDEMA AND  ITS  MANAGEMENT.pdfPULMONARY EDEMA AND  ITS  MANAGEMENT.pdf
PULMONARY EDEMA AND ITS MANAGEMENT.pdfDolisha Warbi
 
Introduction to Sports Injuries by- Dr. Anjali Rai
Introduction to Sports Injuries by- Dr. Anjali RaiIntroduction to Sports Injuries by- Dr. Anjali Rai
Introduction to Sports Injuries by- Dr. Anjali RaiGoogle
 
Presentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPresentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPrerana Jadhav
 
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdfPULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdfDolisha Warbi
 
world health day presentation ppt download
world health day presentation ppt downloadworld health day presentation ppt download
world health day presentation ppt downloadAnkitKumar311566
 
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisVarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisGolden Helix
 
History and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfHistory and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfSasikiranMarri
 
Apiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptApiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptkedirjemalharun
 
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners
 
ANTI-DIABETICS DRUGS - PTEROCARPUS AND GYMNEMA
ANTI-DIABETICS DRUGS - PTEROCARPUS AND GYMNEMAANTI-DIABETICS DRUGS - PTEROCARPUS AND GYMNEMA
ANTI-DIABETICS DRUGS - PTEROCARPUS AND GYMNEMADivya Kanojiya
 

Dernier (20)

LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS  CLASSIFICATIONS.pdfLUNG TUMORS AND ITS  CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA
 
Nutrition of OCD for my Nutritional Neuroscience Class
Nutrition of OCD for my Nutritional Neuroscience ClassNutrition of OCD for my Nutritional Neuroscience Class
Nutrition of OCD for my Nutritional Neuroscience Class
 
Clinical Pharmacotherapy of Scabies Disease
Clinical Pharmacotherapy of Scabies DiseaseClinical Pharmacotherapy of Scabies Disease
Clinical Pharmacotherapy of Scabies Disease
 
Big Data Analysis Suggests COVID Vaccination Increases Excess Mortality Of ...
Big Data Analysis Suggests COVID  Vaccination Increases Excess Mortality Of  ...Big Data Analysis Suggests COVID  Vaccination Increases Excess Mortality Of  ...
Big Data Analysis Suggests COVID Vaccination Increases Excess Mortality Of ...
 
Informed Consent Empowering Healthcare Decision-Making.pptx
Informed Consent Empowering Healthcare Decision-Making.pptxInformed Consent Empowering Healthcare Decision-Making.pptx
Informed Consent Empowering Healthcare Decision-Making.pptx
 
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATROApril 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATRO
 
SGK HÓA SINH NĂNG LƯỢNG SINH HỌC 2006.pdf
SGK HÓA SINH NĂNG LƯỢNG SINH HỌC 2006.pdfSGK HÓA SINH NĂNG LƯỢNG SINH HỌC 2006.pdf
SGK HÓA SINH NĂNG LƯỢNG SINH HỌC 2006.pdf
 
Measurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxMeasurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptx
 
Case Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxCase Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptx
 
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
PULMONARY EDEMA AND  ITS  MANAGEMENT.pdfPULMONARY EDEMA AND  ITS  MANAGEMENT.pdf
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
 
Introduction to Sports Injuries by- Dr. Anjali Rai
Introduction to Sports Injuries by- Dr. Anjali RaiIntroduction to Sports Injuries by- Dr. Anjali Rai
Introduction to Sports Injuries by- Dr. Anjali Rai
 
Presentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPresentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous System
 
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdfPULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
 
world health day presentation ppt download
world health day presentation ppt downloadworld health day presentation ppt download
world health day presentation ppt download
 
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisVarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
 
History and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfHistory and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdf
 
Apiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptApiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.ppt
 
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
 
ANTI-DIABETICS DRUGS - PTEROCARPUS AND GYMNEMA
ANTI-DIABETICS DRUGS - PTEROCARPUS AND GYMNEMAANTI-DIABETICS DRUGS - PTEROCARPUS AND GYMNEMA
ANTI-DIABETICS DRUGS - PTEROCARPUS AND GYMNEMA
 

Ultrasound Characteristics and Components

  • 1.
  • 2. Sound • Sound beam is similar to x ray beam as both waves transmit energy
  • 3. DIAGNOSTIC ULTRASOUND X RAYS Wave type Longitudinal waves Electromagnetic waves transmission Elastic medium No medium generation Stressing the medium Accelerating electric charges velocity Depends on the medium constant Similar waves Seismic, acoustic Radio, light
  • 4. Longitudinal Waves • Ultrasound pulses are transmitted, as longitudinal waves, ie; the motion of particles in the medium is parallel to direction of wave propagation. • They produce bands of compression and rarefaction • Wave length is distance between two bands of compression or rarefaction
  • 5. “Ultra”…….sound? • Audible range is 20 to 20,000 cycles per second • Ultrasound has frequency greater than 20,000 cycles per second
  • 6. Velocity of Sound • It is independent of frequency , • but depends on material through which it propagates • Depends on – Compressibility – Density
  • 7. Velocity of Sound 1. Compressibility – velocity is inversely proportional – liquids and solids propagate sound more rapidly than gases (easily compressed) 2. Density – denser materials have greater inertia - so, decreased velocity
  • 8.
  • 9.
  • 10. ULTRASOUND CHARACTERISTICS • Frequency • Velocity • Wavelength • Amplitude • Intensity and Power
  • 11. Frequency • The frequency of sound is determined by the source. (2-20MHz) v = f λ • In ultrasonic frequency range, the velocity of sound is constant in any particular medium. f  1  • If the frequency increases then the wave length must decrease as they are inversely proportional to each other.
  • 12. Velocity • The significance of ultrasound velocity is that it is used to determine the depth location of structures in the body • The period (T) is the time required for one vibration cycle. It is the reciprocal of the frequency
  • 13. Amplitude • The amplitude of an ultrasound pulse is the range of pressure excursions , related to the energy content. • In diagnostic applications, it is usually necessary to know only the relative amplitude of ultrasound pulses. • Units of decibels (dB).
  • 14. Intensity • Also called loudness • Determined by amplitude of oscillation • As amplitude is increased , intensity also increases
  • 15. Intensity • Intensity is the rate at which ultrasound energy is applied to a specific tissue location within the patient's body. • It is the quantity that must be considered with respect to producing biological effects and safety • The intensity of most diagnostic ultrasound beams at the transducer surface is on the order of a few milliwatts per square centimeter.
  • 16. History • First successful application – SONAR in world war 2 (SOund Navigation And Ranging)
  • 17. Successful medical application – 1940s Uses of ultrasonic energy in the 1940s. Left, in gastric ulcers. Right, in arthritis Ultrasonic therapy generator, the "Medi-Sonar" in the 1950s. A British ultrasonic apparatus for the treatment of Meniere's disease in the late 1950s
  • 18. History Denier's Ultrasonoscopic apparatus with ultrasound generator, emitter transducer and oscilloscope. This can be adapted for both therapeutic and diagnostic purposes The first hand-held imaging instrument was developed by John Wild and John Reid in the early 1950's
  • 20. Components 1. Transmitter 2. Transducer 3. CPU 4. Display 5. Key board / cursor 6. Disc storage device. 7. Printer
  • 21. Transducer • Instrument which converts one form of energy to other • The conversion of electrical pulses to mechanical vibrations and the conversion of returned mechanical vibrations back into electrical energy. Electrical Energy Mechanical Energy
  • 22. SELECTION OF TRANSDUCER • Superficial vessels and organs within 1 to 3cms depth and intra operative imaging – • 7.5 to 15 Hz • Deeper structures in abdomen and pelvis within 12 to 15cms – • 2.25 to 3.5Hz
  • 23. Transducer - Parts • A simple single-element, plane-piston source transducer has major components including the – Piezoelectric material, – Sensor electrodes, – Insulated layer, – Backing block, – Acoustic insulator – Insulating cover, and – Transducer housing.
  • 24.
  • 25. Piezoelectric Element • The active element is basically a piece of polarized material - a piezoelectric ceramic sandwiched between electrodes • The piezoelectric element converts electrical signals into mechanical vibrations (transmit mode) and mechanical vibrations into electrical signals (receive mode).
  • 26. Piezoelectric materials • Natural – Quartz • Artificial – most of USG materials – ferroelectrics • barium titanate • PZT (lead zirconate titanate) ADVANTAGE – they can be formed into different shapes
  • 27. Ceramic  Piezoelectric Crystal? • The piezoelectric attributes are attained after a process of - Molecular synthesis, (dipoles) - Heating, (Curie Temperature) - Orientation of internal dipole structures with an applied external voltage, (Poling) - Cooling to permanently maintain the dipole orientation, and - Cutting into a specific shape.
  • 28. Ceramic  Piezoelectric Crystal? • Once the material has cooled, the dipoles retain their alignment. • Heating the crystal above this temp reduces its usefulness. So, transducers should not be autoclaved.
  • 29. Piezoelectric crystal • At equilibrium, there is no net charge on ceramic surfaces. • When compressed, an imbalance of charge produces a voltage between the surfaces - piezoelectric effect • Similarly, when a voltage is applied between electrodes attached to both surfaces, mechanical deformation occurs - electrostriction
  • 30.
  • 31.
  • 32. Piezoelectric crystal – how thick? • The thickness of the active element is determined by the desired frequency of the transducer • Piezoelectric crystals are cut to a thickness that is 1/2 the desired radiated wavelength
  • 33. Backing/Damping Block • The rear face of the piezoelectric crystal material is usually supported by a backing material which is tungsten loaded araldite, so that the vibrations in the piezoelectric material are rapidly damped after the initial excitation.
  • 34. Backing/Damping Block • This component also dampens the transducer vibration to create an ultrasound pulse with a short spatial pulse length, which is necessary to preserve detail along the beam axis (axial resolution).
  • 35. Couplant • Material (usually liquid) that facilitates the transmission of ultrasonic energy from the transducer into the test specimen. • Necessary to overcome the acoustic impedance mismatch between air and solids.
  • 36. Modes of Vibration 2 TYPES: 1. thickness mode • most common • Used in medical crystals 2. radial mode
  • 37. Resonant Frequency • Natural frequency to which the transducer is sensitive • Resonant frequency determined by thickness of crystal • Thick crystal – low frequency sound • Natural frequency – one that produces internal wavelengths that are twice the thickness of crystal • Frequency corresponding to half the wavelength is-fundamental resonant frequency
  • 38. Transducers Resonance Transducers Non Resonance Transducers
  • 39. Resonance Transducers • They are manufactured to operate in a “resonance” mode, whereby a voItage (commonly 150 V) of very short duration (a voltage spike of 1 msec) is applied, causing the piezoelectric material to initially contract, and subsequently vibrate at a natural resonance frequency. • The operating frequency is determined from – the speed of sound in, and – the thickness of, the piezoelectric material.
  • 40. Resonance Transducers • Higher frequencies are achieved with thinner elements, and lower frequencies with thicker elements.
  • 41. Nonresonance (Broad-Bandwidth) “Multifrequency” Transducers • Modern transducer design coupled with digital signal processing enables “multifrequency or “multihertz” transducer operation, whereby the center frequency can be adjusted in the transmit mode.
  • 42. Unlike the resonance transducer design, the piezoelectric element is intricately machined into a large number of small “rods,” and then filled with an epoxy resin to create a smooth surface.
  • 43.
  • 44. Nonresonance (Broad-Bandwidth) “Multifrequency” Transducers • Excitation of the multifrequency transducer is accomplished with a short square wave burst of 150 V with one to three cycles, unlike the voltage spike used for resonance transducers. • This allows the center frequency to be selected within the limits of the transducer bandwidth.
  • 45. Nonresonance (Broad-Bandwidth) “Multifrequency” Transducers • Likewise, the broad bandwidth response permits the reception of echoes within a wide range of frequencies. • For instance, ultrasound pulses can be produced at a low frequency, and the echoes received at higher frequency.
  • 46. • “Harmonic imaging” is a recently introduced technique that uses this ability. • Lower frequency ultrasound is transmitted into the patient, and the higher frequency harmonics (e.g., two times the transmitted center frequency) created from the interaction with contrast agents and tissues, are received as echoes.
  • 47. Spatial Pulse Length • The length of the sonic pulse. • The number of waves multiplied by their wavelengths • The backing block is incorporated to quench the vibrations and to shorten the sonic pulse.
  • 48. Transducer Q Factor (Q = Quality) Refers to two characteristics of crystal • purity of their sound and • length of time the sound persists
  • 49. Transducer Q Factor • The “Q factor” describes the bandwidth of the sound emanating from a transducer as: f Q o  Bandwidth • where fo = center frequency bandwidth = width of the frequency distribution.
  • 50. • The interval between initiation of the wave and complete cessation of vibrations is called the “ring down time”. • Dampening of the vibration lessens the purity of the resonance frequency and introduces a broadband frequency spectrum.
  • 51. 푄 = 푓0 퐵푎푛푑푤푖푑푡ℎ ∝ 푆푝푎푡푖푎푙 푃푢푙푠푒 퐿푒푛푔푡ℎ • A “high Q” transducer has a narrow bandwidth (i.e., very little damping) and a corresponding long spatial pulse length – organ imaging • A “low Q” transducer has a wide bandwidth and short spatial pulse length – doppler
  • 52.
  • 53. Characteristics of Ultrasound Beam • A single vibrating point sets out waves in all directions • Waves move away as concentric circles
  • 54. When two sound waves interact , they cancel each other or reinforce each other
  • 55. Beam Properties • The ultrasound beam propagates as a longitudinal wave from the transducer surface into the propagation medium, and exhibits two distinct beam patterns: – a slightly converging beam out to a distance specified by the geometry and frequency of the transducer (the near field), and – a diverging beam beyond that point (the far field).
  • 56. The Near Field • The near field, also known as the Fresnel zone, is adjacent to the transducer face and has a converging beam profile. • Beam convergence in the near field occurs because of multiple constructive and destructive interference patterns of the ultrasound waves from the transducer surface.
  • 57. • Huygen’s principle describes a large transducer surface as an infinite number of point sources of sound energy where each point is characterized as a radial emitter. • By analogy, a pebble dropped in a quiet pond creates a radial wave pattern.
  • 58. Near Field Length………. • The near field length for an unfocused, single-element transducer is dependent on the transducer frequency and diameter: Near Field Length = 풓ퟐ  = 풓ퟐ 풗 풇 = 풓ퟐ.풇 풗
  • 59. • For multiple transducer element arrays, an “effective” transducer diameter is determined by the excitation of a group of’ transducer elements. • Because of the interactions of each of the individual beams and the ability to focus and steer the overall beam, the formulas for a single-element, unfocused transducer are not directly applicable.
  • 60. Near Field Length = 푟2.푓 푣
  • 61. • Lateral resolution (the ability of the system to resolve objects in a direction perpendicular to the beam direction) is dependent on the beam diameter and is best at the end of the near field for a single-element transducer. • Lateral resolution is worst in areas close to and far from the transducer surface.
  • 62. • Pressure amplitude characteristics in the near field are very complex, caused by the constructive and destructive interference wave patterns of the ultrasound beam. • Peak ultrasound pressure occurs at the end of the near field, corresponding to the minimum beam diameter for a single-element transducer.
  • 63. • Pressures vary rapidly from peak compression to peak rarefaction several times during transit through the near field. – Only when the far field is reached do the ultrasound pressure variations decrease continuously.
  • 64. Far Field • The far field is also known as the Fraunhofer zone, and is where the beam diverges.
  • 65. • For a large-area single-element transducer, the angle of ultrasound beam divergence, , for the far field is given by  d sin 1.22 • where d is the effective diameter of the transducer and  is the wavelength; both must have the same units of distance.
  • 66. sin 휃 = 1.22  푑 = 1.22 푣 푓 푑 = 푣 푑. 푓 • Less beam divergence occurs with: – High - frequency transducers – Large - diameter transducers
  • 67. Near Field Length = • High frequency beams – – fresnel zone is longer – depth resolution is superior • Disadvantage : 푟2.푓 푣 – Tissue absorption is more, leading to deterioration of side to side resolution Solution: Focused Transducer
  • 68. Focused Transducers • Single-element transducers are focused by using – a curved piezoelectric element or – a curved acoustic lens to reduce the beam profile.
  • 69. Focal Distance • The focal distance, the length from the transducer to the narrowest beam width, is shorter than the focal length of a non-focused transducer and is fixed.
  • 70. Focal Zone • The focal zone is defined as the region over which the width of the beam is less than two times the width at the focal distance; – Thus, the transducer frequency and dimensions should be chosen to match the depth requirements of the clinical situation.
  • 71. Interactions B/N Ultrasound and Matter 1. Reflection 2. Refraction 3. Absorption
  • 72. Reflection • Diagnostic images are produced by reflected portion of the beam. • Percentage of beam reflected at tissue interfaces depends on – Tissue’s acoustic impedance – Beam’s angle of incidence.
  • 73. Acoustic Impedance (Z) • The most important tissue property in imaging. • This quantity is more properly called the specific acoustic impedance of the medium • A simplified definition is Z = d v d = density of the tissue (g/cm3) v = velocity of sound (cm/sec) • Z – Rayl (g/cm2 x 10-5)
  • 74.
  • 75. • At most interfaces within the body, only a portion of the ultrasound pulse is reflected • The pulse is divided into two pulses - one pulse, the echo, is reflected back toward the transducer and the other penetrates into the other material.
  • 76. • The brightness of a structure in an ultrasound image depends on the strength of the reflection, or echo. • This in turn depends on the difference in acoustic impedance of the two materials.
  • 77. Angle of incidence • Angle b/n sound waves and reflecting surfaces • The more the angle , the less the reflection • In medical USG, reflected sound is not detected when angle is greater than 3o
  • 78. Reflection • Two distinct patterns of reflection give rise to the echoes that make up an ultrasound image – specular reflection and – scattering
  • 79. Specular reflection • When an incident ultrasound pulse encounters a large, smooth interface between two types of tissue with different acoustic impedance values , the result is a partially reflected echo that travels back toward the transducer and a partially transmitted pulse that travels deeper into the patient. • Responsible for the bright appearance of fibrous structures such as tendons and of boundaries between different tissues.
  • 80. Scattering / Diffuse Reflection • If the ultrasound pulse encounters reflectors whose dimensions (d) are smaller than the ultrasound wavelength (ie, d << λ), scattering occurs. • Scattering gives rise to the characteristic texture (echo texture) of the image seen within soft tissue.
  • 81.
  • 82. Refraction • Bending of waves as they pass from one medium to other • The change in wavelength and direction of propagation of sound occurs, but frequency remains constant • Artifacts due to refraction are – Loss of resolution of image – Spatial distortion
  • 83. Snell’s Law 풔풊풏 휽풊 풔풊풏 휽풕 = 풗ퟏ 풗ퟐ
  • 84. Absorption / Attenuation • Refers to conversion of ultrasound energy to heat energy
  • 85. Absorption / Attenuation • Depends On – frequency of sound (increase absorption) – viscosity of medium (increase absorption) – relaxation time of medium – temperature (varies with tissues)
  • 86. • The transmission of sound waves without much energy loss can be done by – using mineral oil b/n transducer and patient skin – mechanical impedance matching • Any process that lessens the reflection. it is done by a coupler who has intermediate density b/n transducer and tissue • The thickness of this matching layer must be equal to one fourth the wavelength of sound in the matching layer (Quarter-wave matching)
  • 87. IMAGE DISPLAY • Electronic representation of data generated from returning signals and displayed on TV monitor – A MODE – M MODE – Real time B MODE
  • 88. AMPLITUDE MODE • Echoes are displayed in the form of spikes on CRO traced along time base • Amplitude of spike measures the echo size • Information about the depth of the structures and the amplitude of the returning echo
  • 89.
  • 90. AMPLITUDE MODE • Used in – Ophthalmology – Echoencephalography – Echocardiography. • Disadv: – 1D information, – takes lot of space in CRO
  • 91.
  • 92. M-MODE & TM-MODE • Detects motion of structures- cardiac valves and of cardiac chambers , vessels • Echoes are displayed as dots of varying intensity • M-Mode does not have Time factor. • It also provides 1D information • TM-Mode has Time factor
  • 93. BRIGHTNESS MODE • Echo signals as – Line of dots. • Intensity of dot gives relative size of echo • It provides depth of information and variations in direction of beam
  • 94. REAL TIME B MODE • System is fast enough to allow movements to be followed. • At least 16 frames / sec
  • 95. GRAY SCALE IMAGING • Display variation of amplitudes of echoes arising from tissues as varying shades of gray.
  • 97. Controls • Adjustment of image to obtain Optimal Gray scale image done using several controls: – TGC – Coarse Gain – Intensity governs the amplitude – Reject – Delay – Near Gain - diminish near echoes – Far Gain - enhance distant echoes
  • 98. TGC • Time Gain Compensator • Compensates differences in echo strength by adjusting the variations in degree of amplification • The slope of the TGC adjusts the degree of amplification • The delay control regulates the depth at which the TGC begins to augment weaker signals
  • 99. Pulse Rate • The number of separate little packets of sound that are sent out each second • It determines the total number of echoes returning to the transducer in a unit of time. • High pulse rate is desirable. • But then the receiving time decreases. • So, the pulse rate must be set to accommodate the thickest part that might be examined.
  • 101. • In ultrasound, the major factor that limits the spatial resolution and visibility of detail is the volume of the acoustic pulse. • The axial, lateral, and elevational (slice thickness) dimensions determine the minimal volume element. • Each dimension has an effect on the resolvability of objects in the image.
  • 102. Axial Resolution • Also known as linear, range, longitudinal, or depth resolution • Refers to the ability to discern two closely spaced objects in the direction of the beam. • Achieving good axial resolution requires that the returning echoes be distinct without overlap.
  • 103. • The minimal required separation distance between two reflectors is one-half of the spatial pulse length (SPL) to avoid the overlap of returning echoes, as the distance traveled between two reflectors is twice the separation distance.
  • 104. Lateral Resolution • Lateral resolution, also known as azimuthal resolution • Refers to the ability to discern as separate two closely spaced objects perpendicular to the beam direction.
  • 105. • For both single element transducers and multielement array transducers, the beam diameter determines the lateral resolution.
  • 106. • Since the beam diameter varies with the distance from the transducer in the near and far field, the lateral resolution is depth dependent. • The best lateral resolution occurs at the near field—far field face.
  • 107. • At this depth, the effective beam diameter is approximately equal to half the transducer diameter. • In the far field, the beam diverges and substantially reduces the lateral resolution.
  • 108. • The typical lateral resolution for an unfocused transducer is approximately 2 to 5 mm. • A focused transducer uses an acoustic lens (a curved acoustic material analogous to an optical lens) to decrease the beam diameter at a specified distance from the transducer.
  • 109. Elevational Resolution • The elevational or slice-thickness dimension of the ultrasound beam is perpendicular to the image plane. • Slice thickness plays a significant part in image resolution, particularly with respect to volume averaging of acoustic details in the regions close to the transducer and in the far field beyond the focal zone.
  • 110. • Elevational resolution is dependent on the transducer element height in much the same way that the lateral resolution is dependent on the transducer element width.
  • 112. Types: • Mechanical Scanners • Electronic Array Scanners
  • 113. Mechanical scanners • Early US systems relied on the operator to manually change the position and orientation of the transducer and scan the ultrasound beam through a plane in the patient to obtain the echo data necessary for each image.
  • 114. • 3 types: Mechanical scanners – Oscillating Transducer (Unenclosed type) – Oscillating Transducer (Enclosed type) – Rotating Wheel Transducer
  • 115. Mechanical scanners • This was a time-consuming process but produced images that covered large FOVs • Image artefacts occurred if the patient moved during the acquisition
  • 116. Electronic array scanning • Most modern US imagers automatically scan the ultrasound beam using transducers consisting of arrays of many narrow piezoelectric elements. • The array may consist of as many as 128–196 elements
  • 117. • Two modes of activation are used to produce a beam. – These are the “linear” (sequential) and “phased” activation/receive modes.
  • 118. Linear • The linear array scanners produce sound waves parallel to each other and produces a rectangular image. • The width of the image and number of scan lines are the same at all tissue levels. • Often used with high frequencies ie 7MHz. • Advantage -good near field resolution. • Disadvantage is artifacts when applied to a curved part of the body creating air gaps between skin and transducer.
  • 119. Sector/Vector • Produces a fan like image that is narrow near the transducer and increase in width with deeper penetration. • The disadvantge is poor near field resolution.
  • 120. Curved • Often with frequencies of 2 - 5 MHz (to allow for a range of patients from obese to slender). • The array of elements is arranged across a convex arc (instead of a straight line), which rapidly scans a larger FOV
  • 121.
  • 122. Endoscopic Ultrasound • Dedicated linear-array and radial echo-endoscopes for structural evaluation of the luminal wall adjacent tissues. • Flexible shaft with a central wire that drives rotation of a mechanical transducer at the tip. • The transducer is surrounded by oil that serves as an acoustic interface with tissue, providing 360-degree imaging
  • 123.
  • 124. • With the introduction of the curved linear array echoendoscope , the indications for EUS have expanded. • Allows the endoscopist to perform a whole range of interventional applications ranging from fine needle aspiration (FNA) of lesions surrounding the gastrointestinal tract to celiac plexus block and drainage of pancreatic pseudocyst
  • 125. • Ultrasound imaging of vascular and non-vascular lumina employing miniaturized high-frequency (20-30 MHz) transducers inserted into catheters as small as 2 mm in external diameter
  • 126. Recent Innovations in B-mode US • Tissue Harmonic Imaging • Spatial Compound Imaging • Extended FOV Imaging • Coded Pulse Excitation • 3D and 4D imaging • Elastography. • Ultrasound contrast media.
  • 127. Tissue Harmonic Imaging • A musical note has three characteristics. – Pitch (or frequency) – Loudness – Quality (or tone). • Quality is the audible difference heard between two musical notes of the same pitch and loudness. • Hence, a piano based C-note does not sound the same as a C-note played on a guitar, due to the existence of harmonic frequencies. • The same note from various instruments has different qualities because the sounds are not pure notes i.e. of one frequency and these integral multiples of the fundamental frequency (overtones) give an instrument its characteristic sound.
  • 128.
  • 129. Tissue Harmonic Imaging • Harmonics are frequencies that occur at multiples of the fundamental or transmitted sonographic frequency. • In conventional gray-scale sonography, the same frequency spectrum that is transmitted into the patient is subsequently received to produce the sonographic image. • In THI sonography, higher harmonic frequencies generated by propagation of the ultrasound beam through tissue are used to produce the sonogram.
  • 130. Why harmonics? • The ultrasonic pulse gets altered with time as it traverses the tissues with non-linear motion. • The peaks within the pulse waveform move faster than the troughs because the propagation speed is higher in compressed regions of tissue than in the areas which are expanded by the passing pressure wave. • The degree of such acoustic signal distortion in tissue depends on the amplitude of the emitted pulse and the distance it has travelled in the tissue. Drawing of undistorted pressure wave (top) and of pressure wave after undergoing nonlinear propagation (bottom). Nonlinear effects cause high-pressure regions of sound wave to travel faster than low-pressure regions and result in progressive distortion of transmitted wave with generation of sound at higher harmonics of transmit frequency
  • 131. • As much of the fundamental sonographic signal as possible must be removed to make these theoretical harmonic improvements a clinical reality. • This can be done by either frequency- based or phase inversion methods. • There are 2 basic harmonic imaging methods. – conventional frequency-based second-harmonic imaging – pulse inversion or phase inversion harmonic imaging
  • 132. Conventional Frequency-based Second-harmonic Imaging • A narrowband pulse is emitted, and then high-pass or narrow-bandpass filtering is applied to the received echoes to filter out the fundamental echo components. • This results in reduction of both spatial and contrast resolution.
  • 133. Phase Inversion Harmonic Imaging • In wideband harmonic imaging (also called pulse inversion or phase inversion harmonic imaging) • A train of 2 pulses is emitted, with the phase of the second inverted relative to the phase of the first. • When the echoes from the transmitted pulses are added, the linear components of the echoes cancel each other, whereas the nonlinear components are amplified. • This results in superior contrast and spatial resolution. Two phase-inverted but otherwise identical sonographic pulses are transmitted. Summing the returning echoes in a buffer cancels most of the fundamental and odd harmonic echoes and effectively amplifies the second harmonic.
  • 134.
  • 135.
  • 136. ADVANTAGES OF THI 1. Improve lateral resolution 2. Reduce side lobe artifacts 3. Improve signal to noise ratio 4. Improved near field and far field image quality
  • 137. ADVANTAGES OF THI 5. Lesions are clearer & better defined. 6. Use of higher frequencies improves resolution. 7. Helps differentiate cysts from hypoechoic solid masses. 8. Better clarity of contents. 9. It is superior to conventional USG in visualization of lesions containing highly reflective tissues like fat, calcium & air.
  • 138. ADVANTAGES OF THI 10.Harmonic imaging is generally considered to be most useful for “technically difficult” patients with thick and complicated body wall
  • 139. THI is useful in • Better defining the borders of pancreas • Better visualization of lower pole of left kidney • Better visualization of fatty livers • Small renal cysts can be easily seen • Presence or absence of sludge in GB • Metastatic lesions in liver easily seen • Minimal fluid present between liver and kidney • Provides better images of aortic wall, IVC, portal vein and renal arteries
  • 140. Spatial Compound Imaging • Electronic steering of ultrasound beams from an array transducer is used to image the same tissue multiple times by using parallel beams oriented along different directions. • The echoes from these multiple acquisitions (upto 9 sets) are then averaged together into a single composite image.
  • 141.
  • 142. Disadvantage • More time is required for data acquisition • The compound imaging frame rate is reduced compared with that of conventional B-mode imaging.
  • 143. Advantage • Spatial compound images often show reduced levels of speckle, noise, clutter, and refractive shadows and improved contrast and margin definition • Enhancement and shadowing artifacts may also be reduced, which may be an advantage or potential drawback, depending on the imaging situation.
  • 144.
  • 145. Clutter Suppression Conventional image: acoustic clutter produces spurious echoes within a simple cyst (inset), which can be difficult to eliminate regardless of incident angle or adjustment of system parameters. Compound image: compound Imaging results in significant clutter suppression in the cyst. Also note the good depth of field in this single focus compound image, comparable to the multi-focus conventional image.
  • 146. Extended FOV Imaging • A benefit of early static B-mode scanners that was lost with the introduction of mechanical and electronic automatic scanning was large imaging FOVs. • Extended FOV imaging, has sought to restore this capability. • The transducer is slowly translated laterally across the large anatomic region of interest. During this motion, multiple images are acquired from many transducer positions. • The registered image data are accumulated in a large image buffer and then combined to form the complete large FOV image
  • 147. • Extended FOV images are not limited just to B-mode acquisitions. • Restores the capability of visualizing large anatomic regions in a single image and simplifies measurements made over these large regions. • It is useful in evaluating aneurysms, detecting small ligament and tissue damage, and to image and calibrate long sections of blood vessels, to locate anatomical landmarks for location of disease.
  • 148.
  • 149. Coded Pulse Excitation • A fundamental trade-off in US is that between imaging depth and spatial resolution. • Coded ultrasound pulses help overcome this limitation, providing good penetration at the higher frequencies necessary for high spatial resolution. • In this imaging approach, long ultrasound pulses are used instead of the very short pulses. These long pulses carry greater ultrasonic energy, increasing the energy of echoes that return from large depths in the patient.
  • 150. The coded pulses are produced with a very specific, characteristic shape, and the resulting echoes will have a similar shape.
  • 151. The end result is an image with good echo signal and good spatial resolution at large depths. Conventional B-mode (a) and coded pulse (b) US images of the liver show the benefits of coded pulse imaging. The spatial resolution of the coded pulse image (b) is very comparable with that of the 13-MHz conventional image (a). However, the useful imaging depth is about 7.5 cm for the coded pulse image (b) compared with only about 2.8 cm for the conventional image (a).
  • 152. ELASTOGRAPHY • Dynamic technique that uses ultrasound to provide information on tissue stiffness by measuring the degree of distortion under the application of an external force - viscoelastic properties of tissue • Surrogate for that obtained with manual palpation. • Principle: malignant tissues have more stiffness and are harder to distort. On tissue compression strain (displacement) within harder tissue is less than in softer tissue.
  • 153. Layers in Jello show soft and hard materials after light compression
  • 154. Important Quantities • Young’s modulus (E) describes longitudinal deformation in terms of strain • The shear modulus (G) relates to transverse strain to transverse stress • The bulk modulus (K) of elasticity describes the change in volume • Poisson ratio (v) which is a ratio of transverse contraction per unit breadth divided by longitudinal extension per unit length Shear and Young’s moduli, are the most suitable elasticity parameters to measure.
  • 155. The tissue is insonified a) before and b) after a small uniform compression. In the harder tissues (e.g. the circular lesion depicted) the echoes will be less distorted than in the surrounding tissues, denoting thus smaller strain
  • 156. Elastography - methods • Three methods— a) spatial correlation method, b) the phase-shift tracking method, and c) the combined autocorrelation method (CAM)— • have been introduced for measuring tissue strain at elastography.
  • 157.
  • 158. Elastography - Uses • In the liver it is a very useful method in depicting liver hardness and prediction and prognostication of liver cirrhosis. • Breast imaging, cervix imaging, musculoskeletal imaging (especially supraspinatus and tendoachilles), thyroid lesions assessment and assessment of cervical nodes.
  • 159. 3D and 4D ultrasound • 3D ultrasound - data set that contains a large number of 2D planes (B-mode images). • This volume data can be manipulated in different planes by rotation. • Dissected in any plane, to get multiplanar images (similar to multislice CT). • Special probes and software are necessary in order to perform 3D and 4D imaging. • 4D ultrasound is also known as "Real-time 3D Ultrasound". The 3D datasets with their multiplanar reformations and renderings in real time, give us photographic quality images.
  • 160. Techniques of 3D US • Data acquisition • Image reconstruction and • Image display
  • 161. Data Acquisition Techniques a) Tracked freehand systems, b) Untracked freehand systems, c) Mechanical assemblies, and d) 2D arrays • Regardless of which method is used, one must know the relative position and angulation of each 2D image and must acquire the images rapidly or with gating to avoid motion artifacts. If these two criteria are not met, the 3D images may be inaccurate.
  • 162. (a) Tracked Freehand Systems • The operator holds an assembly composed of the transducer with an attachment and manipulates it over the anatomic area being evaluated • Ensure that there are no significant imaging gaps.
  • 163. • Acoustically tracked 3D scanning • Articulated-arm-tracked 3D scanning • Magnetic field-tracked 3D scanning
  • 164. (b) Untracked Freehand • 2D images are digitized as the operator moves the transducer with a smooth, steady motion • Most convenient for the operator • Image quality is variable and depends largely on how smoothly and steadily the operator moves the transducer • Geometric measurements may be inaccurate because there is no direct information regarding the relative position of the digitized images.
  • 165. • Mechanical movement of the transducer across the skin. • Tilted about a fixed point on the skin surface. • Rotated about its own axis.
  • 166. (c) Mechanical Assemblies • The transducer is propelled or rotated mechanically, and 2D images are digitized at predetermined spatial or angular intervals . • Cumbersome for the user • Improve the geometric accuracy • To date, their greatest utility has been in intracavitary and intraluminal examinations, in which the area of interest is relatively small and motion artifact is less of a problem.
  • 167. • Three different scanning techniques can be used: – linear scanning, – tilt scanning, and – rotational scanning.
  • 168. (d) Two-dimensional Arrays • With the first three types of data acquisition systems, mechanical motion is used to obtain 3D images. • An alternative is to keep the transducer stationary and use electronic scanning with a 2D transducer array generates pyramidal or conical US pulses to generate 3D information in real time
  • 169. Image Reconstruction Techniques • Two types: – 3D surface model – Voxel-based volume model.
  • 170. Three-dimensional Surface Model: • Outline the boundaries of the areas of interest on the 2D images manually or with a computer algorithm • Reduces the amount of 3D data needed • Shorter 3D reconstruction times and greater efficiency • Identification of boundaries can be tedious and time-consuming.
  • 171. Voxel-based Volume Model • The computer builds a 3D voxel-based volume (3D grid) • This process preserves the original information • Allows a variety of rendering techniques • Generates very large data files, which slows processing and requires large amounts of computer memory.
  • 172. Image Display Techniques • Surface Rendering: operator identifies the boundaries of pertinent structures either manually or with an algorithm • Multiplanar Reformatting: three orthogonal planes or texture mapping – image rotated to obtain the desired image orientation • Combined Surface Rendering and Multiplanar Reformatting. • Volume Rendering
  • 173. Artefacts • Respiratory motion artifact • Artifact caused by incorrect calibration
  • 174. Advantages of 3D US • Many benefits in obstetrical and gynaecological scanning where it is important to analyze anatomy. It has shown promising results in diagnosing cleft lip/palate, spina bifida, polydactyly, club foot, facial dysmorphism, low set ears ,fetal cardiac imaging. • Excellent tool for demonstrating and accurately diagnosing congenital uterine anomalies, pelvic floor muscle and sphincters. • Considerable interobserver variability in 2D; exact relationship between anatomic structures is accurately recorded in the 3D image
  • 175. Advantages of 3D US • Unrestricted access to an infinite number of viewing planes. • Suited for monitoring the effects of therapy over a long period of time • More accurate quantitative volume estimates
  • 176. Limitations of 3D US • More cumbersome , requiring more user input • Data archiving and communication more challenging • Waiting for the 3D image to appear can be frustrating to users • Slow the image interpretation process, especially if inexperienced, as the right algorithm needs to be chosen. • The ability to obtain a good 3-D picture is very much dependent on operator skill, the amount of liquor, fetus position and the degree of maternal obesity.
  • 177. Applications for 3D US • Fetal Imaging
  • 178. Applications for 3D US • Gynecologic Imaging volume data sets to be acquired with both transvaginal and transabdominal probes
  • 179. Applications for 3D US • Three-dimensional Power Doppler Imaging 3D US angiography • Prostate Imaging
  • 180. Applications for 3D US • Breast Imaging: demonstrate lesion margins and topography • Biopsy-related Imaging: needle localization and guidance
  • 181. ULTRASOUND CONTRAST MEDIA • Microbubbles of air or other gases which act as echo enhancers • Microbubbles are less than 10 μm in diameter, and • Contrary to most other contrast media which are rapidly distributed to the extravascular, extracellular space, most microbubbles are confined to the vascular space. • Microbubbles may produce up to 25 dB (more than 300-fold) increase in echo strength.
  • 182. • Stability is increased by – external bubble encapsulation (galactose, phospholipids, denatured albumin or poly-butyl-cyanoacrylate) with or without surfactants – using gases with a low diffusion coefficient (perfluorocarbons) or – a combination of both • The gas components of the microbubbles are normally eliminated via the lungs. • Stabilizing components are eliminated via the hepato-renal route.
  • 183. Contrast Microbubbles • The ideal diameter - 2 μmto 8 μm • Enhancement life-time of the microbubble, often several minutes (8–10 min)
  • 184. Microbubbles – Structure • The coatings include albumin, gelatin, galactose microspheres, polyglutaminic acid, lipophilic monolayer surfactants, and lipid bilayers (liposomes). • The gas inside the shell may be either air or various perfluorocarbons, which are liquids at room temperature but gas at body temperature.
  • 185. The ideal USCA • Non-toxic, • Injectable intravenously, • Capable of crossing the pulmonary capillary bed after a peripheral injection, and • Stable enough to achieve enhancement for the duration of the examination.
  • 186. Microbubbles - Generations • Several "generations" of gas microbubble contrast media have evolved; • The "1st generation" products do not pass the pulmonary vascular bed, and are therefore limited to the venous system and the right heart cavities after injection. • The "2nd generation" contrast media are both sufficiently small and stable to pass into the systemic circulation, and these contrast media enhance the doppler signal in various arteries after injection. They are short-lived, however, the effect is over in a few minutes.
  • 187. Microbubbles - Generations • The "3rd generation" gas microbubble contrast media are even more echogenic and stable, and are able to enhance the echogenicity of parenchyma on B-mode images. They may thus show perfusion, even in such a difficult region as the myocardium. • The various gas microbubble contrast media are generally safe with low toxicity in humans.
  • 188.
  • 189. Ultrasound Contrast Agents - Types • Different types of ultrasound contrast agents: – tissue contrast agents; – contrast agents for vascular enhancement; – agents for targeted contrast imaging.
  • 190. PRINCIPLES • The main mechanisms for signal enhancement are – backscattering, – bubble resonance and – bubble rupture. • These mechanisms are highly dependent on the acoustic power of the transmitted ultrasound, which is reflected by the mechanical index (MI). • very high reflectivity ( gas content makes them very reflective)
  • 191. Acoustic Excitation of Ultrasound Contrast Agents • The behavior of these agents under acoustic excitation fall into three classes (Frinking et al., 1999), depending on the structure of the microbubble and the level of the insonifying pressure amplitude and frequency: – stable linear (low MI), – stable nonlinear scattering (medium MI) – transient nonlinear scattering (high MI).
  • 192. Back Scattering • At low acoustic power (MI < 0.1), gas microbubbles may be regarded as point scatterers, and the mechanism of ultrasound reflection is that of Rayleigh Tyndall scattering. • The scattering strength of a point scatterer is proportional to the sixth power of the particle radius and to the fourth power of the ultrasound frequency; the echogenicity of such contrast media is therefore highly dependent upon particle size and transmit frequency. • The backscattered intensity of a group of point scatterers is furthermore directly proportional to the total number of scatterers in the insonified volume; the concentration of the contrast medium is therefore also of importance.
  • 193. MI- 0.1-0.3 Reflect and scatter
  • 194. Bubble Resonance and Harmonics • At intermediate acoustic power (0.1 < MI < 0.5) gas microbubbles may show strong oscillatory motion provided the frequency of the incident ultrasound is close to the resonant (fundamental) frequency of the microbubbles. MICROBUBBLE AS LINEAR RESONATOR • By virtue of their compressibility, microbubbles display unique properties in an ultrasound beam, which sets them into resonance when there is a match between their diameter and the ultrasonic wavelength which for microbubbles 2-7microns range at ultrasound frequency 2-10 Mhz.
  • 196. MICROBUBBLE AS NONLINEAR RESONATOR • nonlinear behaviour (compression and expansion phases) produces echoes that contain frequencies not present in the transmitted pulses. • An elegant way to extract these nonlinear signals is to send a series of pulses down each line, varying their phase and amplitude; the returning signals are combined to cancel the linear signals from tissue and the remaining bubble-specific signals are used to form an image that can be presented as a colour overlay on the B-mode image or shown on a side-by- side display, all in real-time.
  • 197. MI- 0.7-0.8 Oscillates in non harmonic pattern
  • 198.
  • 199. Bubble rupture • At high acoustic power (MI > 0.5), ultrasound at the microbubble resonance frequency will cause the bubbles to rupture. • The result is a transient high-amplitude, broadband signal containing all frequencies, not only the harmonics. • It will create a transient, strong signal in B-mode, or a short-lasting multicoloured, mosaic-like effect in colour Doppler sonography. • Several terms for the strong, transient signal have been proposed: induced or stimulated acoustic emission, loss of correlation imaging and sono-scintigraphy.
  • 200. MI- =/>1 Disrupt and Burst
  • 201. Applications • Give angiographic capacity to ultrasound ie help in demonstrating tissue perfusion, bleeding points,vascularity of lesions etc. • Widely used in imaging of solid organs, particularly the liver • Improvement in detection of colour Doppler signal from large vessels • Microbubble contrast has also found a niche outside the vascular compartment in the setting of vesico-ureteric reflux in children where a high sensitivity and specificity compared with conventional micturating cystourography (MCUG) has been demonstrated
  • 202. Applications • Gene therapy and targeted delivery of drugs
  • 203.
  • 204. Conclusion • Modern US equipment is based on many of the same fundamental principles employed in the initial devices used for human imaging over 50 years ago. • US has the characteristics of being relatively inexpensive, portable, safe, and real-time in nature, all of which make it one of the most widely used imaging modalities in medicine. • In short, US science, technology, and applications are expanding at a brisk pace and are far from mature. • Even more exciting developments are on the horizon.

Notes de l'éditeur

  1. SOund
  2. (a) Diagram shows the set of ultrasound beam lines used for conventional B-mode imaging. (b, c) Diagrams show two additional sets of beam lines that are used for spatial compound imaging. Data from up to nine such sets of beam lines are acquired and averaged together to form each compound image.
  3. This figure illustrates the concept of spatial compounding of ultrasound images. The conventional B-mode image is constructed with a single scan of the beam generated at 90° to the surface of the beam as shown in panel (b) of the figure. In compound imaging, three to nine views are obtained by steering the beam at different angles, as shown in panels (a), (b) and (c). The frames in panels (a), (b) and (c) are combined to generate a single frame (e). After displaying compound image (frame e), the frame (a) is dropped, a new frame (d) is acquired, and frames (b), (c) and (d) are compounded to form a new image. Since updating of each compound image requires acquisition of only one new frame, compound imaging can be performed in real time, similar to conventional B-mode imaging.