Impact of novel MS/MSALL Acquisition and 
Processing Techniques on Forensic 
Toxicological Screening 
By David M. Cox, Michael Jarvis, Evelyn McClure, 
Adrian M. Taylor 
For Research Use Only. Not for use in diagnostic procedures.
INTRODUCTION 
For research use only, not for use in diagnostic procedures. 
Rapid forensic toxicology screening by high 
resolution mass spectrometry is a powerful 
technique. However, some compounds cannot be 
unambiguously identified with high resolution 
MS1 measurements alone. MS/MS fragmentation 
yields confident identifications of these 
compounds, but how to ensure quality MS/MS of 
these compounds? Data dependent techniques, 
although very powerful, cannot guarantee the 
measurement of all possible MS/MS candidates. 
Targeted MS/MS ensures acquisition of the 
target compounds, but limits the number of 
compounds. Data-independent techniques, such 
as SWATH™ acquisition (the MS/MS of all 
possible candidates), improve identifications 
significantly and enable retrospective analysis of 
the data. The impact of improvements to 
SWATHTM acquisition, including variable 
precursor window sizes, overlapping windows, 
and data processing were evaluated. 
2 © 2014 AB SCIEX 
• Investigate variations of 
SWATHTM acquisition 
(MS/MSALL) 
• Compare identification 
results to best practice 
IDA methods 
• Deconvolution of 
MS/MS from multiple 
precursors
MATERIALS AND METHODS 
Urine was spiked with over 120 drugs and compounds often 
found in forensics screening panels. The data was collected 
on a TripleTOF® 5600 system using one of the following 
methods: 
1) using a TOF-MS survey scan with IDA-triggering of up to 
20 product ion scans or 
2) SWATHTM acquisition. For SWATHTM acquisition, the 
precursor isolation window width was varied for each 
MS/MS experiment, or the windows were overlapped 
between each cycle. Data was processed in PeakView® 
software 2.0, using a research prototype of MasterView™ 
software. 
3 © 2014 AB SCIEX RUO-MKT-10-1442-A
4 © 2014 AB SCIEX 
Information Dependent 
Acquisition 
At any time point during an LC run, 
unit resolution MS/MS (0.7 Da wide 
precursor isolation window) are 
acquired for the top candidates. 
Good method settings ensure that 
almost all potential candidates are 
acquired at least once during the 
run. However, inevitably, there are 
some gaps in the coverage, and 
methods must be modified 
depending on LC peak widths and 
sample complexity to achieve 
optimal coverage.
5 © 2014 AB SCIEX 
SWATHTM Acquisition 
At every time point during an LC 
run, an MS/MS for a window of 
precursor ions is acquired. MS/MS 
for all possible candidates are 
acquired at every point in the LC 
run. Method settings are more 
generic. Individual MS/MS can be 
convolved with multiple precursors. 
Available since Analyst TF 1.6.
6 © 2014 AB SCIEX 
Variable Window 
SWATHTM Acquisition 
Identical to SWATHTM acquisition, 
except the width of the isolation 
window is varied. MS/MS for all 
possible candidates are acquired at 
every point in the LC run. 
Convolution due to multiple 
precursors is reduced. 
Available in Analyst TF 1.7.
7 © 2014 AB SCIEX 
Overlap SWATHTM Acquisition 
• Identical to SWATHTM acquisition, except in each cycle, 
the beginning and end of each window are shifted. 
MS/MS for all possible candidates are acquired at every 
point in the LC run. Convolution due to multiple 
precursors is different from cycle to cycle, due to the 
shifted windows. 
• The differences between shifted windows enable the 
calculation of an MS/MS that would have resulted from 
a narrower window. 
• The data was acquired using prototype software, and 
the results were demultiplexed using custom software 
running on an NVDIA 660 graphics card. 
• Research prototype software.
RESULTS 
8 © 2014 AB SCIEX 
MS1 MS/MS 
Acquired 
Library 
Acquired 
Library 
SWATH 20 Da Window 
Overlap SWATH 
10 Da Effective Window 
Acquired 
Library 
SWATH 10 Da Window 
MS1 MS/MS 
Acquired 
Library 
Acquired 
Library 
SWATH 20 Da Window 
Overlap SWATH 
10 Da Effective Window 
Acquired 
Library 
SWATH 10 Da Window
Advantages of narrower isolation windows 
Two examples demonstrating the advantage of having narrower isolation windows. 
• A) Astemizole, a compound with a high intensity signal but with a significant 
interference within the same 20 Da SWATHTM window. Despite using background 
subtraction to obtain the MS/MS, there are significant interferences, resulting in a poor 
library match purity score of 2.2%. 
• B) Amlodipine, a compound with low intensity signal and numerous interferences 
within the 20 Da SWATHTM window. The MS/MS is clearly a mixture of several 
compounds. Purity score of 38.8% 
In both examples, the overlap SWATHTM acquisition (after demultiplexing was 
performed) resulted in MS/MS that were significantly reduced in interferences. The 
library match purity scores were improved to 97.5% and 92.7% respectively. The 
spectral quality of the demultiplexed spectra were equivalent to spectra acquired from 
SWATHTM windows that were actually set to 10 Da wide. 
9 © 2014 AB SCIEX RUO-MKT-10-1442-A
Improved cycle times 
Extractions of berberine from the MS1 experiment of either a 
20 Da window SWATHTM acquisition, an overlap SWATHTM 
acquisition (30 Da windows with a 10 Da overlap between 
cycles) or a 10 Da window SWATHTM acquisition. 
The accumulation time for MS/MS experiments was held 
constant at 25 msec. Having narrower isolation windows 
improves the specificity of MS/MS data, but at a cost. Either 
accumulation times must be decreased (which would make 
signal to noise worse) or cycle times will get longer (reducing 
the number of points across a peak). 
The demultiplexed MS/MS from overlap SWATHTM acquisition 
can approach the quality of a true 10 Da SWATHTM 
acquisition MS/MS, while having an improved cycle time. 
10 © 2014 AB SCIEX
11 © 2014 AB SCIEX RUO-MKT-10-1442-A 
SWATH 20 Da Window 
626 msec cycle time 
Overlap SWATH 
10 Da Effective Window 
475 msec cycle time 
SWATH 10 Da Window 
1100 msec cycle time
Screening using various acquisition modes 
• Various methods were used to screen compounds spiked into either solvent or urine. 
Identifications were made using mass accuracy, isotope patterns, retention time accuracy, 
and library scores. For this comparison, only library scores were plotted. Fit: A measure of 
how well a library spectrum matches the unknown spectrum. Purity: A measure of how 
well the unknown spectrum matches the library spectrum. Lower values indicate that either 
the match is less certain or that additional fragment ion peaks from another compound are 
present in the unknown spectrum. 
• In a spiked clean solvent sample, using IDA acquisition, both fit and purity scores were 
good for all screened compounds. 
• Spiked urine resulted in some ion suppression and interferences. A few compounds did 
not trigger MS/MS, resulting on no score. For the majority of compounds, an MS/MS was 
triggered, and the unit resolution isolation resulted in good purity scores. 
• With SWATHTM, since all possible MS/MS are acquired, no compounds were missed, 
resulting in acceptable fit scores for all compounds. However, purity scores were lower 
due to the wider 20 Da isolation windows. 
• With overlap SWATHTM, the effective 10 Da windows gave slightly improved purity scores 
compared to 20 Da windows. 
12 © 2014 AB SCIEX
140 
120 
100 
80 
60 
40 
20 
0 
Fit Purity Fit Purity Fit Purity Fit Purity 
Spiked Solvent 
Spiked Urine 
IDA 
IDA 
13 © 2014 AB SCIEX RUO-MKT-10-1442-A 
Spiked Urine 
SWATH 20 Da 
Spiked Urine 
Overlap SWATH 
No match, or score < 40% 
Requires review, score > 40% 
Confident ID, score > 70% 
# of compounds
• Deconvolution of SWATHTM MS/MS improves library 
match purity scores 
The wider windows of SWATHTM acquisition often lead to convolved spectra, resulting 
in lower purity scores. Compounds were still confidently identified using mass 
accuracy, isotope patterns, retention time accuracy, and library fit scores. However, 
achieving consistently high purity scores is important, particularly when an unknown 
compound is not in a library and must be identified by other methods, such as formula 
finding and ChemSpider. These methods work best with reasonably pure MS/MS. 
Fortunately, there are many techniques for deconvolving MS/MS that result from a 
mixture of co-eluting compounds. Unprocessed SWATHTM MS/MS had significantly 
lower purity scores for many compounds. Simple background subtraction resulted in 
MS/MS of much better quality. Two other deconvolution techniques were tried. 
Method A was similar to techniques used for deconvolving GC-MS signals, and was 
implemented to run on an NVIDIA 660 graphics card. Method B is a novel technique 
making use of principal components variable grouping (PCVG) to obtain a SWATHTM 
MS/MS. When the techniques were combined, results were equivalent to those 
achieved using unit resolution IDA. For a few compounds, IDA was not triggered, 
resulting in no identification, while the SWATHTM acquisition was able to confidently 
identify these compounds with good purity scores. 
14 © 2014 AB SCIEX RUO-MKT-10-1442-A
140 
120 
100 
80 
60 
40 
20 
0 
IDA (98.3) Unprocessed (70.0) Background Subtract (82.5) Deconvolve A (94.8) Deconvolve B (95.3) 
15 © 2014 AB SCIEX RUO-MKT-10-1442-A 
No match, or score < 40% 
Requires review, score > 40% 
Confident ID, score > 70% 
Solvent 
IDA 
Urine 
IDA 
Urine 
SWATH 
Unprocessed 
Urine 
SWATH 
Background 
Subtract 
Urine 
SWATH 
Deconvolve A 
Urine 
SWATH 
Deconvolve B 
Urine 
SWATH 
Combined
CONCLUSIONS 
SWATHTM acquisition methods: 
• Acquire MS/MS for all compounds, at every 
time point 
• Achieve identification results comparable to 
unit resolution 
IDA methods 
• Overlap SWATHTM acquisition can improve 
cycle times and improve identification results 
16 © 2014 AB SCIEX RUO-MKT-10-1442-A
REFERENCES 
1.R. G. Dromey, Mark J. Stefik, Thomas C. Rindfleisch, and Alan M. 
Duffield. (1976). Extraction of Mass Spectra Free of Background and 
Neighboring Component Contributions from Gas 
Chromatography/Mass Spectrometry Data. Analytical Chemistry, 48(9), 
1368-1375. 
2.Vladimir A Likić. (2009). Extraction of pure components from 
overlapped signals in gas chromatography-mass spectrometry (GC-MS). 
BioData Mining 2(6). 
3.Gordana Ivosev, Lyle Burton and Ron Bonner. (2008). Dimensionality 
Reduction and Visualization in Principal Component Analysis. 
Analytical Chemistry, 80(13), 4933-4944. 
17 © 2014 AB SCIEX RUO-MKT-10-1442-A
TRADEMARKS/LICENSING 
• For Research Use Only. Not for use in diagnostic 
procedures. 
• Document #: RUO-MKT-10-1442-A 
• © 2014 AB SCIEX. SCIEX is part of AB SCIEX. The 
trademarks mentioned herein are the property of AB Sciex 
Pte. Ltd. or their respective owners. AB SCIEX™ is being 
used under license. 
• www.absciex.com 
18 © 2014 AB SCIEX RUO-MKT-10-1442-A
19 © 2014 AB SCIEX

Contenu connexe

Tendances

Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...SCIEX
 
Analysis of Environmental Contaminants using High Performance Quantitative LC...
Analysis of Environmental Contaminants using High Performance Quantitative LC...Analysis of Environmental Contaminants using High Performance Quantitative LC...
Analysis of Environmental Contaminants using High Performance Quantitative LC...Waters Corporation - Environmental Analysis
 
Biopharmaceutical Attribute Monitoring with the Waters ACQUITY QDa Mass Detector
Biopharmaceutical Attribute Monitoring with the Waters ACQUITY QDa Mass DetectorBiopharmaceutical Attribute Monitoring with the Waters ACQUITY QDa Mass Detector
Biopharmaceutical Attribute Monitoring with the Waters ACQUITY QDa Mass DetectorWaters Corporation
 
Monitoring Released N-Glycans in Biopharma Development/QC with Fluorescence &...
Monitoring Released N-Glycans in Biopharma Development/QC with Fluorescence &...Monitoring Released N-Glycans in Biopharma Development/QC with Fluorescence &...
Monitoring Released N-Glycans in Biopharma Development/QC with Fluorescence &...Waters Corporation
 
Adding Mass Detection to Monitor Peptides in Biopharmaceutical Development & QC
Adding Mass Detection to Monitor Peptides in Biopharmaceutical Development & QCAdding Mass Detection to Monitor Peptides in Biopharmaceutical Development & QC
Adding Mass Detection to Monitor Peptides in Biopharmaceutical Development & QCWaters Corporation
 
From WCBP 2015: GlycoWorks RapiFluor-MS for Glycan Profiling
From WCBP 2015: GlycoWorks RapiFluor-MS for Glycan ProfilingFrom WCBP 2015: GlycoWorks RapiFluor-MS for Glycan Profiling
From WCBP 2015: GlycoWorks RapiFluor-MS for Glycan ProfilingWaters Corporation
 
Native and IM-MS characterization of ADCs
Native and IM-MS characterization of ADCsNative and IM-MS characterization of ADCs
Native and IM-MS characterization of ADCsWaters Corporation
 
It's All About Peak Performance - Altus™ UPLC
It's All About Peak Performance - Altus™ UPLCIt's All About Peak Performance - Altus™ UPLC
It's All About Peak Performance - Altus™ UPLCPerkinElmer, Inc.
 
Semtech PPMD Portable Particulate Measuring Device
Semtech PPMD Portable Particulate Measuring DeviceSemtech PPMD Portable Particulate Measuring Device
Semtech PPMD Portable Particulate Measuring Devicemarkpartridge
 
Olympus AU480 – An Ideal Choice for Low- to Mid-volume Laboratories
Olympus AU480 – An Ideal Choice for Low- to Mid-volume LaboratoriesOlympus AU480 – An Ideal Choice for Low- to Mid-volume Laboratories
Olympus AU480 – An Ideal Choice for Low- to Mid-volume Laboratoriesblockscientific
 
Comprehensive Investigation of the Utilization of SFC/ESI Positive Mode MS fo...
Comprehensive Investigation of the Utilization of SFC/ESI Positive Mode MS fo...Comprehensive Investigation of the Utilization of SFC/ESI Positive Mode MS fo...
Comprehensive Investigation of the Utilization of SFC/ESI Positive Mode MS fo...Waters Corporation
 
AT Precision presentation on Environmental Testing
AT Precision presentation on Environmental TestingAT Precision presentation on Environmental Testing
AT Precision presentation on Environmental TestingActive Technologies Limited
 

Tendances (20)

Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
 
Analysis of Environmental Contaminants using High Performance Quantitative LC...
Analysis of Environmental Contaminants using High Performance Quantitative LC...Analysis of Environmental Contaminants using High Performance Quantitative LC...
Analysis of Environmental Contaminants using High Performance Quantitative LC...
 
Biopharmaceutical Attribute Monitoring with the Waters ACQUITY QDa Mass Detector
Biopharmaceutical Attribute Monitoring with the Waters ACQUITY QDa Mass DetectorBiopharmaceutical Attribute Monitoring with the Waters ACQUITY QDa Mass Detector
Biopharmaceutical Attribute Monitoring with the Waters ACQUITY QDa Mass Detector
 
Monitoring Released N-Glycans in Biopharma Development/QC with Fluorescence &...
Monitoring Released N-Glycans in Biopharma Development/QC with Fluorescence &...Monitoring Released N-Glycans in Biopharma Development/QC with Fluorescence &...
Monitoring Released N-Glycans in Biopharma Development/QC with Fluorescence &...
 
Adding Mass Detection to Monitor Peptides in Biopharmaceutical Development & QC
Adding Mass Detection to Monitor Peptides in Biopharmaceutical Development & QCAdding Mass Detection to Monitor Peptides in Biopharmaceutical Development & QC
Adding Mass Detection to Monitor Peptides in Biopharmaceutical Development & QC
 
Altus™ HPLC
Altus™ HPLCAltus™ HPLC
Altus™ HPLC
 
The Advantages of Mass Detection for the Food Testing Laboratory - Waters Cor...
The Advantages of Mass Detection for the Food Testing Laboratory - Waters Cor...The Advantages of Mass Detection for the Food Testing Laboratory - Waters Cor...
The Advantages of Mass Detection for the Food Testing Laboratory - Waters Cor...
 
From WCBP 2015: GlycoWorks RapiFluor-MS for Glycan Profiling
From WCBP 2015: GlycoWorks RapiFluor-MS for Glycan ProfilingFrom WCBP 2015: GlycoWorks RapiFluor-MS for Glycan Profiling
From WCBP 2015: GlycoWorks RapiFluor-MS for Glycan Profiling
 
Native and IM-MS characterization of ADCs
Native and IM-MS characterization of ADCsNative and IM-MS characterization of ADCs
Native and IM-MS characterization of ADCs
 
RGT- 01-Vacuum-Packaging-Analyzer
RGT- 01-Vacuum-Packaging-AnalyzerRGT- 01-Vacuum-Packaging-Analyzer
RGT- 01-Vacuum-Packaging-Analyzer
 
It's All About Peak Performance - Altus™ UPLC
It's All About Peak Performance - Altus™ UPLCIt's All About Peak Performance - Altus™ UPLC
It's All About Peak Performance - Altus™ UPLC
 
Print products CIGRE-2014
Print products CIGRE-2014Print products CIGRE-2014
Print products CIGRE-2014
 
Semtech PPMD Portable Particulate Measuring Device
Semtech PPMD Portable Particulate Measuring DeviceSemtech PPMD Portable Particulate Measuring Device
Semtech PPMD Portable Particulate Measuring Device
 
Olympus AU480 – An Ideal Choice for Low- to Mid-volume Laboratories
Olympus AU480 – An Ideal Choice for Low- to Mid-volume LaboratoriesOlympus AU480 – An Ideal Choice for Low- to Mid-volume Laboratories
Olympus AU480 – An Ideal Choice for Low- to Mid-volume Laboratories
 
Comprehensive Investigation of the Utilization of SFC/ESI Positive Mode MS fo...
Comprehensive Investigation of the Utilization of SFC/ESI Positive Mode MS fo...Comprehensive Investigation of the Utilization of SFC/ESI Positive Mode MS fo...
Comprehensive Investigation of the Utilization of SFC/ESI Positive Mode MS fo...
 
Measuring Sulfur in Tier III Gasoline
Measuring Sulfur in Tier III GasolineMeasuring Sulfur in Tier III Gasoline
Measuring Sulfur in Tier III Gasoline
 
Rapid Detection of Pesticides in Fruit Juice without Sample Preparation - Wat...
Rapid Detection of Pesticides in Fruit Juice without Sample Preparation - Wat...Rapid Detection of Pesticides in Fruit Juice without Sample Preparation - Wat...
Rapid Detection of Pesticides in Fruit Juice without Sample Preparation - Wat...
 
OR2-411-Organic-Gas-Permeability-Analyzer
OR2-411-Organic-Gas-Permeability-AnalyzerOR2-411-Organic-Gas-Permeability-Analyzer
OR2-411-Organic-Gas-Permeability-Analyzer
 
AT Precision presentation on Environmental Testing
AT Precision presentation on Environmental TestingAT Precision presentation on Environmental Testing
AT Precision presentation on Environmental Testing
 
Tier 3 Webinar
Tier 3 Webinar Tier 3 Webinar
Tier 3 Webinar
 

Similaire à Impact of novel ms ms all acquisition and processing techniques on forensic toxicological screening

Impact of novel ms ms all acquisition and processing techniques on forensic t...
Impact of novel ms ms all acquisition and processing techniques on forensic t...Impact of novel ms ms all acquisition and processing techniques on forensic t...
Impact of novel ms ms all acquisition and processing techniques on forensic t...Sara Feltesse
 
Online Detection of Shutdown Periods in Chemical Plants: A Case Study
Online Detection of Shutdown Periods in Chemical Plants: A Case StudyOnline Detection of Shutdown Periods in Chemical Plants: A Case Study
Online Detection of Shutdown Periods in Chemical Plants: A Case StudyManuel Martín
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingPerkinElmer, Inc.
 
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...KBI Biopharma
 
LVTS - Image Resolution Monitor for Litho-Metrology
LVTS - Image Resolution Monitor for Litho-MetrologyLVTS - Image Resolution Monitor for Litho-Metrology
LVTS - Image Resolution Monitor for Litho-MetrologyVladislav Kaplan
 
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Shimadzu Scientific Instruments
 
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...PerkinElmer, Inc.
 
Jordi Labs Agilent Extractables & Leachables Webinar Presentation (Part 1)
Jordi Labs Agilent Extractables & Leachables Webinar Presentation (Part 1)Jordi Labs Agilent Extractables & Leachables Webinar Presentation (Part 1)
Jordi Labs Agilent Extractables & Leachables Webinar Presentation (Part 1)Jordi Labs
 
REST SUMMARY
REST SUMMARYREST SUMMARY
REST SUMMARYShan Guan
 
FL52308_E 0212M_GCMS_crops_REVISED FINAL
FL52308_E 0212M_GCMS_crops_REVISED FINALFL52308_E 0212M_GCMS_crops_REVISED FINAL
FL52308_E 0212M_GCMS_crops_REVISED FINALJennifer Christianson
 
Vcu Chemistry Reasearch Facilities
Vcu Chemistry Reasearch FacilitiesVcu Chemistry Reasearch Facilities
Vcu Chemistry Reasearch FacilitiesJoseph Turner 'Jody'
 
Turbidity poster presentation final edit
Turbidity poster presentation final editTurbidity poster presentation final edit
Turbidity poster presentation final editKala Drake
 
Veracel andritz - i io-t - online wirelless vibration project implementa...
Veracel    andritz - i io-t -  online wirelless vibration project  implementa...Veracel    andritz - i io-t -  online wirelless vibration project  implementa...
Veracel andritz - i io-t - online wirelless vibration project implementa...Jaques Valle
 
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_finalLawrence Hwang
 

Similaire à Impact of novel ms ms all acquisition and processing techniques on forensic toxicological screening (20)

Impact of novel ms ms all acquisition and processing techniques on forensic t...
Impact of novel ms ms all acquisition and processing techniques on forensic t...Impact of novel ms ms all acquisition and processing techniques on forensic t...
Impact of novel ms ms all acquisition and processing techniques on forensic t...
 
Lcms 2020
Lcms 2020Lcms 2020
Lcms 2020
 
Online Detection of Shutdown Periods in Chemical Plants: A Case Study
Online Detection of Shutdown Periods in Chemical Plants: A Case StudyOnline Detection of Shutdown Periods in Chemical Plants: A Case Study
Online Detection of Shutdown Periods in Chemical Plants: A Case Study
 
EMSDVici
EMSDViciEMSDVici
EMSDVici
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
 
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
 
Routine impurity analysis with UPLC/MS and UPC2/MS
Routine impurity analysis with UPLC/MS and UPC2/MSRoutine impurity analysis with UPLC/MS and UPC2/MS
Routine impurity analysis with UPLC/MS and UPC2/MS
 
LVTS - Image Resolution Monitor for Litho-Metrology
LVTS - Image Resolution Monitor for Litho-MetrologyLVTS - Image Resolution Monitor for Litho-Metrology
LVTS - Image Resolution Monitor for Litho-Metrology
 
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
 
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
 
Jordi Labs Agilent Extractables & Leachables Webinar Presentation (Part 1)
Jordi Labs Agilent Extractables & Leachables Webinar Presentation (Part 1)Jordi Labs Agilent Extractables & Leachables Webinar Presentation (Part 1)
Jordi Labs Agilent Extractables & Leachables Webinar Presentation (Part 1)
 
REST SUMMARY
REST SUMMARYREST SUMMARY
REST SUMMARY
 
FL52308_E 0212M_GCMS_crops_REVISED FINAL
FL52308_E 0212M_GCMS_crops_REVISED FINALFL52308_E 0212M_GCMS_crops_REVISED FINAL
FL52308_E 0212M_GCMS_crops_REVISED FINAL
 
Vcu Chemistry Reasearch Facilities
Vcu Chemistry Reasearch FacilitiesVcu Chemistry Reasearch Facilities
Vcu Chemistry Reasearch Facilities
 
LabOnAChip 13_4864
LabOnAChip 13_4864LabOnAChip 13_4864
LabOnAChip 13_4864
 
Turbidity poster presentation final edit
Turbidity poster presentation final editTurbidity poster presentation final edit
Turbidity poster presentation final edit
 
Innovation in size based polymer separations
Innovation in size based polymer separationsInnovation in size based polymer separations
Innovation in size based polymer separations
 
03 broderick qsts_sand2016-4697 c
03 broderick qsts_sand2016-4697 c03 broderick qsts_sand2016-4697 c
03 broderick qsts_sand2016-4697 c
 
Veracel andritz - i io-t - online wirelless vibration project implementa...
Veracel    andritz - i io-t -  online wirelless vibration project  implementa...Veracel    andritz - i io-t -  online wirelless vibration project  implementa...
Veracel andritz - i io-t - online wirelless vibration project implementa...
 
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
 

Dernier

Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPirithiRaju
 
Speech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxSpeech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxpriyankatabhane
 
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptxTHE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptxNandakishor Bhaurao Deshmukh
 
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxLIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxmalonesandreagweneth
 
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)riyaescorts54
 
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...lizamodels9
 
Topic 9- General Principles of International Law.pptx
Topic 9- General Principles of International Law.pptxTopic 9- General Principles of International Law.pptx
Topic 9- General Principles of International Law.pptxJorenAcuavera1
 
Microteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringMicroteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringPrajakta Shinde
 
Functional group interconversions(oxidation reduction)
Functional group interconversions(oxidation reduction)Functional group interconversions(oxidation reduction)
Functional group interconversions(oxidation reduction)itwameryclare
 
Davis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologyDavis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologycaarthichand2003
 
User Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationUser Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationColumbia Weather Systems
 
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfBehavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfSELF-EXPLANATORY
 
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...Universidade Federal de Sergipe - UFS
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentationtahreemzahra82
 
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)Columbia Weather Systems
 
Dubai Calls Girl Lisa O525547819 Lexi Call Girls In Dubai
Dubai Calls Girl Lisa O525547819 Lexi Call Girls In DubaiDubai Calls Girl Lisa O525547819 Lexi Call Girls In Dubai
Dubai Calls Girl Lisa O525547819 Lexi Call Girls In Dubaikojalkojal131
 
Environmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial BiosensorEnvironmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial Biosensorsonawaneprad
 
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPirithiRaju
 

Dernier (20)

Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
 
Speech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxSpeech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptx
 
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptxTHE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
 
Hot Sexy call girls in Moti Nagar,🔝 9953056974 🔝 escort Service
Hot Sexy call girls in  Moti Nagar,🔝 9953056974 🔝 escort ServiceHot Sexy call girls in  Moti Nagar,🔝 9953056974 🔝 escort Service
Hot Sexy call girls in Moti Nagar,🔝 9953056974 🔝 escort Service
 
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxLIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
 
Volatile Oils Pharmacognosy And Phytochemistry -I
Volatile Oils Pharmacognosy And Phytochemistry -IVolatile Oils Pharmacognosy And Phytochemistry -I
Volatile Oils Pharmacognosy And Phytochemistry -I
 
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
 
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
 
Topic 9- General Principles of International Law.pptx
Topic 9- General Principles of International Law.pptxTopic 9- General Principles of International Law.pptx
Topic 9- General Principles of International Law.pptx
 
Microteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringMicroteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical Engineering
 
Functional group interconversions(oxidation reduction)
Functional group interconversions(oxidation reduction)Functional group interconversions(oxidation reduction)
Functional group interconversions(oxidation reduction)
 
Davis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologyDavis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technology
 
User Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationUser Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather Station
 
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfBehavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
 
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentation
 
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
 
Dubai Calls Girl Lisa O525547819 Lexi Call Girls In Dubai
Dubai Calls Girl Lisa O525547819 Lexi Call Girls In DubaiDubai Calls Girl Lisa O525547819 Lexi Call Girls In Dubai
Dubai Calls Girl Lisa O525547819 Lexi Call Girls In Dubai
 
Environmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial BiosensorEnvironmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial Biosensor
 
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
 

Impact of novel ms ms all acquisition and processing techniques on forensic toxicological screening

  • 1. RUO-MKT-10-1442-A Impact of novel MS/MSALL Acquisition and Processing Techniques on Forensic Toxicological Screening By David M. Cox, Michael Jarvis, Evelyn McClure, Adrian M. Taylor For Research Use Only. Not for use in diagnostic procedures.
  • 2. INTRODUCTION For research use only, not for use in diagnostic procedures. Rapid forensic toxicology screening by high resolution mass spectrometry is a powerful technique. However, some compounds cannot be unambiguously identified with high resolution MS1 measurements alone. MS/MS fragmentation yields confident identifications of these compounds, but how to ensure quality MS/MS of these compounds? Data dependent techniques, although very powerful, cannot guarantee the measurement of all possible MS/MS candidates. Targeted MS/MS ensures acquisition of the target compounds, but limits the number of compounds. Data-independent techniques, such as SWATH™ acquisition (the MS/MS of all possible candidates), improve identifications significantly and enable retrospective analysis of the data. The impact of improvements to SWATHTM acquisition, including variable precursor window sizes, overlapping windows, and data processing were evaluated. 2 © 2014 AB SCIEX • Investigate variations of SWATHTM acquisition (MS/MSALL) • Compare identification results to best practice IDA methods • Deconvolution of MS/MS from multiple precursors
  • 3. MATERIALS AND METHODS Urine was spiked with over 120 drugs and compounds often found in forensics screening panels. The data was collected on a TripleTOF® 5600 system using one of the following methods: 1) using a TOF-MS survey scan with IDA-triggering of up to 20 product ion scans or 2) SWATHTM acquisition. For SWATHTM acquisition, the precursor isolation window width was varied for each MS/MS experiment, or the windows were overlapped between each cycle. Data was processed in PeakView® software 2.0, using a research prototype of MasterView™ software. 3 © 2014 AB SCIEX RUO-MKT-10-1442-A
  • 4. 4 © 2014 AB SCIEX Information Dependent Acquisition At any time point during an LC run, unit resolution MS/MS (0.7 Da wide precursor isolation window) are acquired for the top candidates. Good method settings ensure that almost all potential candidates are acquired at least once during the run. However, inevitably, there are some gaps in the coverage, and methods must be modified depending on LC peak widths and sample complexity to achieve optimal coverage.
  • 5. 5 © 2014 AB SCIEX SWATHTM Acquisition At every time point during an LC run, an MS/MS for a window of precursor ions is acquired. MS/MS for all possible candidates are acquired at every point in the LC run. Method settings are more generic. Individual MS/MS can be convolved with multiple precursors. Available since Analyst TF 1.6.
  • 6. 6 © 2014 AB SCIEX Variable Window SWATHTM Acquisition Identical to SWATHTM acquisition, except the width of the isolation window is varied. MS/MS for all possible candidates are acquired at every point in the LC run. Convolution due to multiple precursors is reduced. Available in Analyst TF 1.7.
  • 7. 7 © 2014 AB SCIEX Overlap SWATHTM Acquisition • Identical to SWATHTM acquisition, except in each cycle, the beginning and end of each window are shifted. MS/MS for all possible candidates are acquired at every point in the LC run. Convolution due to multiple precursors is different from cycle to cycle, due to the shifted windows. • The differences between shifted windows enable the calculation of an MS/MS that would have resulted from a narrower window. • The data was acquired using prototype software, and the results were demultiplexed using custom software running on an NVDIA 660 graphics card. • Research prototype software.
  • 8. RESULTS 8 © 2014 AB SCIEX MS1 MS/MS Acquired Library Acquired Library SWATH 20 Da Window Overlap SWATH 10 Da Effective Window Acquired Library SWATH 10 Da Window MS1 MS/MS Acquired Library Acquired Library SWATH 20 Da Window Overlap SWATH 10 Da Effective Window Acquired Library SWATH 10 Da Window
  • 9. Advantages of narrower isolation windows Two examples demonstrating the advantage of having narrower isolation windows. • A) Astemizole, a compound with a high intensity signal but with a significant interference within the same 20 Da SWATHTM window. Despite using background subtraction to obtain the MS/MS, there are significant interferences, resulting in a poor library match purity score of 2.2%. • B) Amlodipine, a compound with low intensity signal and numerous interferences within the 20 Da SWATHTM window. The MS/MS is clearly a mixture of several compounds. Purity score of 38.8% In both examples, the overlap SWATHTM acquisition (after demultiplexing was performed) resulted in MS/MS that were significantly reduced in interferences. The library match purity scores were improved to 97.5% and 92.7% respectively. The spectral quality of the demultiplexed spectra were equivalent to spectra acquired from SWATHTM windows that were actually set to 10 Da wide. 9 © 2014 AB SCIEX RUO-MKT-10-1442-A
  • 10. Improved cycle times Extractions of berberine from the MS1 experiment of either a 20 Da window SWATHTM acquisition, an overlap SWATHTM acquisition (30 Da windows with a 10 Da overlap between cycles) or a 10 Da window SWATHTM acquisition. The accumulation time for MS/MS experiments was held constant at 25 msec. Having narrower isolation windows improves the specificity of MS/MS data, but at a cost. Either accumulation times must be decreased (which would make signal to noise worse) or cycle times will get longer (reducing the number of points across a peak). The demultiplexed MS/MS from overlap SWATHTM acquisition can approach the quality of a true 10 Da SWATHTM acquisition MS/MS, while having an improved cycle time. 10 © 2014 AB SCIEX
  • 11. 11 © 2014 AB SCIEX RUO-MKT-10-1442-A SWATH 20 Da Window 626 msec cycle time Overlap SWATH 10 Da Effective Window 475 msec cycle time SWATH 10 Da Window 1100 msec cycle time
  • 12. Screening using various acquisition modes • Various methods were used to screen compounds spiked into either solvent or urine. Identifications were made using mass accuracy, isotope patterns, retention time accuracy, and library scores. For this comparison, only library scores were plotted. Fit: A measure of how well a library spectrum matches the unknown spectrum. Purity: A measure of how well the unknown spectrum matches the library spectrum. Lower values indicate that either the match is less certain or that additional fragment ion peaks from another compound are present in the unknown spectrum. • In a spiked clean solvent sample, using IDA acquisition, both fit and purity scores were good for all screened compounds. • Spiked urine resulted in some ion suppression and interferences. A few compounds did not trigger MS/MS, resulting on no score. For the majority of compounds, an MS/MS was triggered, and the unit resolution isolation resulted in good purity scores. • With SWATHTM, since all possible MS/MS are acquired, no compounds were missed, resulting in acceptable fit scores for all compounds. However, purity scores were lower due to the wider 20 Da isolation windows. • With overlap SWATHTM, the effective 10 Da windows gave slightly improved purity scores compared to 20 Da windows. 12 © 2014 AB SCIEX
  • 13. 140 120 100 80 60 40 20 0 Fit Purity Fit Purity Fit Purity Fit Purity Spiked Solvent Spiked Urine IDA IDA 13 © 2014 AB SCIEX RUO-MKT-10-1442-A Spiked Urine SWATH 20 Da Spiked Urine Overlap SWATH No match, or score < 40% Requires review, score > 40% Confident ID, score > 70% # of compounds
  • 14. • Deconvolution of SWATHTM MS/MS improves library match purity scores The wider windows of SWATHTM acquisition often lead to convolved spectra, resulting in lower purity scores. Compounds were still confidently identified using mass accuracy, isotope patterns, retention time accuracy, and library fit scores. However, achieving consistently high purity scores is important, particularly when an unknown compound is not in a library and must be identified by other methods, such as formula finding and ChemSpider. These methods work best with reasonably pure MS/MS. Fortunately, there are many techniques for deconvolving MS/MS that result from a mixture of co-eluting compounds. Unprocessed SWATHTM MS/MS had significantly lower purity scores for many compounds. Simple background subtraction resulted in MS/MS of much better quality. Two other deconvolution techniques were tried. Method A was similar to techniques used for deconvolving GC-MS signals, and was implemented to run on an NVIDIA 660 graphics card. Method B is a novel technique making use of principal components variable grouping (PCVG) to obtain a SWATHTM MS/MS. When the techniques were combined, results were equivalent to those achieved using unit resolution IDA. For a few compounds, IDA was not triggered, resulting in no identification, while the SWATHTM acquisition was able to confidently identify these compounds with good purity scores. 14 © 2014 AB SCIEX RUO-MKT-10-1442-A
  • 15. 140 120 100 80 60 40 20 0 IDA (98.3) Unprocessed (70.0) Background Subtract (82.5) Deconvolve A (94.8) Deconvolve B (95.3) 15 © 2014 AB SCIEX RUO-MKT-10-1442-A No match, or score < 40% Requires review, score > 40% Confident ID, score > 70% Solvent IDA Urine IDA Urine SWATH Unprocessed Urine SWATH Background Subtract Urine SWATH Deconvolve A Urine SWATH Deconvolve B Urine SWATH Combined
  • 16. CONCLUSIONS SWATHTM acquisition methods: • Acquire MS/MS for all compounds, at every time point • Achieve identification results comparable to unit resolution IDA methods • Overlap SWATHTM acquisition can improve cycle times and improve identification results 16 © 2014 AB SCIEX RUO-MKT-10-1442-A
  • 17. REFERENCES 1.R. G. Dromey, Mark J. Stefik, Thomas C. Rindfleisch, and Alan M. Duffield. (1976). Extraction of Mass Spectra Free of Background and Neighboring Component Contributions from Gas Chromatography/Mass Spectrometry Data. Analytical Chemistry, 48(9), 1368-1375. 2.Vladimir A Likić. (2009). Extraction of pure components from overlapped signals in gas chromatography-mass spectrometry (GC-MS). BioData Mining 2(6). 3.Gordana Ivosev, Lyle Burton and Ron Bonner. (2008). Dimensionality Reduction and Visualization in Principal Component Analysis. Analytical Chemistry, 80(13), 4933-4944. 17 © 2014 AB SCIEX RUO-MKT-10-1442-A
  • 18. TRADEMARKS/LICENSING • For Research Use Only. Not for use in diagnostic procedures. • Document #: RUO-MKT-10-1442-A • © 2014 AB SCIEX. SCIEX is part of AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX™ is being used under license. • www.absciex.com 18 © 2014 AB SCIEX RUO-MKT-10-1442-A
  • 19. 19 © 2014 AB SCIEX