Plethysmographic Waveform



                             Plethysmographic Waveform Shapes
                             Displayed by Pulse Oximeters


                             Introduction

                                 Pulse oximeters utilize two wavelengths of light, red and infrared, to measure arterial oxygen
                                 saturation and pulse rate. Since the infrared (IR) light signal is relatively unaffected by changes
                                 in arterial oxygen saturation, it is often displayed on the front of the pulse oximeter as a
                                 plethysmographic waveform. Plethysmographic pulse waveform morphology varies from site to
                                 site depending on a variety of physiologic factors. For example, the waveform morphology is
                                 different if measured from the fingertip or from the ear. These differences have important physiologic
                                 causes including but not limited to variations in sympathetic innervation of the arteriolar bed at
                                 the monitoring site. Normal plethysmographic waveforms may or may not contain an incisura, or
                                 notch, resembling the classic dicrotic notch seen in arterial blood pressure tracings. This variation
                                 in appearance depends on many conditions. Clinicians have attempted to utilize this waveform and
                                 its variation for assessing signal integrity, physiological and artifactual changes such as perfusion
                                 changes, dysrhythmia, motion artifact, and electrical interference. For this reason accurate and
                                 reliable presentation of the plethysmographic waveform is of vital importance.

                                 Manufacturers normally filter the infrared signal to remove environmental effects and other
                                 disturbances, and to scale it for display. There are a number of filtering techniques and the method
                                 used may cause little or no change to the morphology of the patient signal or it may result in
                                 substantial distortion. It is desirable to preserve as much of the actual patient waveform shape as
                                 possible.

                                 We have examined the effects of two types of filters for various IR signals in an effort to understand
                                 the consequence of each on waveform integrity. Two common methods of filtering are finite impulse
                                 response (FIR) and infinite impulse response (IIR) filtering. These two methods vary in execution, as
                                 they require substantially different computational power for processing. IIR filters are typically found
w w w. m a s i m o . c o m




                                 either in signal processing units consisting of analog circuitry or low-end digital processors. While
                                 highly desirable from a cost standpoint, IIR filters are also known for potentially unstable designs
                                 and nonlinear phase characteristics, which result in waveform distortion. FIR filters overcome both
                                 stability and waveform distortion, but at the expense of greater processing load. To facilitate the
                                 use of FIR filters, and thus preserve the integrity of the waveform, Masimo SET devices utilize a
                                 high end, digital signal processing (DSP) chip. While a DSP chip comes at a much greater expense,
                                 relative to low-end processors, waveform integrity is maintained.
92618




                                 We tested the theoretical effects of FIR and IIR filtering using computer models. We then tested
                                 various input waveform shapes on three pulse oximeters, the Masimo Radical, Nellcor N-395, and
ca




                                 the Datex-Ohmeda 3800, to examine the effects of each of their processing on waveform integrity
irvine




                                 - i.e. did the waveform displayed on the pulse oximeters replicate the input waveform?
40 parker




                             Methods
                                 Characteristic digit waveform shapes were obtained from both a Bio-Tek Index II and a Clinical
                                 Dynamics simulator for various pulse rates to assess the effects of the filters on changing
masimo corporation




                                 physiological conditions. Plethysmographic waveforms seen in the clinic may vary greatly but
                                 include sharp ascending and descending limbs, and may contain, a characteristic notch on the
                                 descending limb. These characteristic waveforms were then subjected to filtering by FIR and IIR
                                 filters to examine their effect on the shape of the original input waveform when reproduced on the
                                 pulse oximeter display.

                                 In a separate experiment, these waveforms were applied to three pulse oximeters, the Masimo
                                 Radical, the Nellcor N-395, and the Datex-Ohmeda 3800, to examine the ability of each instrument
                                 to maintain waveform fidelity. The question being asked: is the output waveform displayed by the
                                 pulse oximeter identical to the input waveform?



                                                                                                                                            1
Plethysmographic Waveform



                             Computer model
                                Several waveform shapes were obtained from the Bio-Tek Index II and the Clinical Dynamics Smart
                                Sat simulators. These shapes are typically seen in the clinical environment. The waveform shapes
                                were then subjected to both FIR and IIR filtering on a personal computer. The results of these tests
                                are depicted in Figures 1-4.

                                                                                                      Notch Test




                                                   0           1               2              3   4          5          6   7   8   9   10
                                                                                                         time
                                                                                                              i
                                                                                                      Time (Second s)

                                                       In pu t W av efo rm
                                                       FIR Filter
                                                       IIR Filter
                                                       Zero Level fo r In p u t W av efo rm
                                                       Zero Level fo r FIR Filter Ou tp u t
                                                       Zero Level fo r IIR Ou tp u t


                                Figure 1. FIR and IIR Filter Performance for a Waveform with Notch Input
w w w. m a s i m o . c o m




                                Figure 1 shows that the IIR filter causes distortion in the waveform. The notch is below midline,
                                the descending limb is elongated (above the notch), and the descending limb (below the notch) is
                                curved.
92618
ca
irvine
40 parker
masimo corporation
technical bulletin



                                                                                                     R amp Test




                                    0             1               2              3        4                 5              6       7       8       9        10
                                                                                                        time
                                                                                                             i
                                                                                                     Time (Second s)

                                          In pu t W av efo rm
                                          FIR Filter
                                          IIR Filter
                                          Zero Level fo r In p u t W av efo rm
                                          Zero Level fo r FIR Filter Ou tp u t
                                          Zero Level fo r IIR Ou tp u t


Figure 2. FIR and IIR Filter Performance for a Ramp Input


Figure 2 shows that the IIR filter causes distortion to the input waveform. The sharp, straight descending limb on
the input signal is preserved with the FIR filter. In contrast, the IIR filter causes a curvilinear descending limb, which
deviates greatly from the actual input waveform.

                                                                                         Lo w Heart R ate Test




                                0           1               2              3         4               5                 6       7       8       9       10
                                                                                                 time
                                                                                                      i
                                                                                              Time (Second s)

                                    In pu t W av efo rm
                                    FIR Filter
                                    IIR Filter
                                    Zero Level fo r In p u t W av efo rm
                                    Zero Level fo r FIR Filter Ou tp u t
                                    Zero Level fo r IIR Ou tp u t



Figure 3. FIR and IIR Filter on a Low Heart Rate Input


Figure 3 shows that the IIR filter causes distortion in the waveform. The notch is below midline, the
descending limb is elongated (above the notch), and the descending limb (below the notch) is shortened and
curved. In marked contrast, the FIR filter preserves the waveform integrity.




                                                                                                                                                                            3
Plethysmographic Waveform


                                                                                                         Asysto le Test




                                                      0           1               2              3   4           5          6   7   8   9   10
                                                                                                             time
                                                                                                                  i
                                                                                                          Time (Second s)

                                                          In pu t W av efo rm
                                                          FIR Filter
                                                          IIR Filter
                                                          Zero Level fo r In p u t W av efo rm
                                                          Zero Level fo r FIR Filter Ou tp u t
                                                          Zero Level fo r IIR Ou tp u t


                             Figure 4. FIR and IIR Filter Performance for an Asystole Input


                                     Figure 4 demonstrates the delays associated with the different types of filtering. While the IIR filter
                                     distorts the waveform, it appears almost in sync with the input waveform. In contrast, the increased
                                     integrity of the waveform with the FIR filter comes at the cost of a slight delay in waveform
                                     appearance.

                                     The four figures show that the processing performed by the FIR filter maintains the integrity of
w w w. m a s i m o . c o m




                                     the input waveform. However, due to the increased computational requirements of the FIR filter
                                     the resultant waveform is delayed slightly in time. The IIR filtering causes marked distortion of the
                                     waveform with little delay in waveform appearance.



                             Conclusions
92618




                                     The processing used by pulse oximeters in their plethysmographic waveform generation may have
                                     an effect on the integrity of the waveform and/or its temporal display. As shown above, the IIR filter
ca




                                     clearly causes marked distortion of the shape of the waveform with a minor delay in the displayed
irvine




                                     waveform appearing on the pulse oximeter screen. In contrast, the FIR filter with only a slight time
                                     increase to display the waveform, maintains the true shape of the waveform.
40 parker
masimo corporation
technical bulletin




Waveforms Displayed by Pulse Oximeter




   Figure 5. This depicts an input waveform at 70 bpm. Notice sharp ascending and descending limbs.




   Figure 6. Photo of waveform at 70 bpm displayed by Masimo Radical in response to input signals shown in figure 5. Notice that the
   waveform integrity is maintained.




   Figure 7. Photo of waveform at 70 bpm displayed by Nellcor N-395 in response to input signals shown in figure 5. Notice that the
   waveform is distorted and shows a curvature, which is not present in the input signal.




   Figure 8. Photo of waveform at 70 bpm displayed by Datex-Ohmeda 3800 in response to input signals shown in figure 5. Notice that
   the waveform is distorted and shows a curvature, which is not present in the input signal.




                                                                                                                                       5
Plethysmographic Waveform



                             Waveforms Displayed by Pulse Oximeter




                                Figure 9. This depicts an input waveform at 72 bpm with notch at mid level of descending limb.




                                Figure 10. Photo of notched waveform at 72 bpm displayed by Masimo Radical in response to input signals shown
                                in figure 9. Notice the notch appears at mid level and waveform integrity is excellent.
w w w. m a s i m o . c o m
92618




                                Figure 11. Photo of notched waveform at 72 bpm displayed by Nellcor N-395 in response to input signals shown in
ca




                                figure 9. Notice the distortion in waveform and the notch well below midline.
irvine
40 parker
masimo corporation




                                Figure 12. Photo of notched waveform at 72 bpm as displayed by Datex-Ohmeda 3800 in response to input shown
                                in figure 9. Notice the distortion in waveform and the notch well below midline.
technical bulletin




Waveforms Displayed by Pulse Oximeter




   Figure 13. This input waveform at 45 bpm with notch at mid level of descending limb.




   Figure 14. Photo of notched waveforms at 45 bpm as displayed by Masimo Radical in response to input signal shown in figure 13.
   Notice that the waveform integrity is preserved.




   Figure 15. Photo of notched waveform at 45 bpm as displayed by Nellcor N-395 in response to input signal shown in figure 13. Notice
   that the waveform is distorted and the notch is well below midline.




   Figure 16. Photo of notched waveform at 45 bpm as displayed by Datex-Ohmeda 3800 in response to input signal shown in figure 13.
   Notice that the waveform is distorted and the notch is well below midline.




                                                                                                                                         7
7396-3618B-0108




Masimo Corporation 40 Parker Irvine, California 92618 Tel 949-297-7000 Fax 949-297-7001 www.masimo.com

©2008 Masimo Corporation. All rights reserved. Masimo, Radical, SET, ,and Signal Extraction Technology are registered
  trademarks of Masimo Corporation. All rights reserved. All other products are trademarks of their respective owners.

More Related Content

Similar to Plethysmographic Waveform Shapes

IRJET-A Comparative Study of Digital FIR and IIR Band- Pass Filter
IRJET-A Comparative Study of Digital FIR and IIR Band- Pass FilterIRJET-A Comparative Study of Digital FIR and IIR Band- Pass Filter
IRJET-A Comparative Study of Digital FIR and IIR Band- Pass FilterIRJET Journal
 
IRJET - Essential Features Extraction from Aaroh and Avroh of Indian Clas...
IRJET -  	  Essential Features Extraction from Aaroh and Avroh of Indian Clas...IRJET -  	  Essential Features Extraction from Aaroh and Avroh of Indian Clas...
IRJET - Essential Features Extraction from Aaroh and Avroh of Indian Clas...IRJET Journal
 
Design of Photosynthetically Active Radiation Sensor_Crimson Publishers
Design of Photosynthetically Active Radiation Sensor_Crimson PublishersDesign of Photosynthetically Active Radiation Sensor_Crimson Publishers
Design of Photosynthetically Active Radiation Sensor_Crimson PublishersCrimsonpublishers-Rehabilitation
 
International Refereed Journal of Engineering and Science (IRJES)
International Refereed Journal of Engineering and Science (IRJES)International Refereed Journal of Engineering and Science (IRJES)
International Refereed Journal of Engineering and Science (IRJES)irjes
 
Performance Analysis of FIR Filter using FDATool
Performance Analysis of FIR Filter using FDAToolPerformance Analysis of FIR Filter using FDATool
Performance Analysis of FIR Filter using FDAToolijtsrd
 
Rigol RF basics_knowledge_applications
Rigol RF basics_knowledge_applicationsRigol RF basics_knowledge_applications
Rigol RF basics_knowledge_applicationsNIHON DENKEI SINGAPORE
 
IRJET- Design Simulation and Analysis of Efficient De-Noising of ECG Signals ...
IRJET- Design Simulation and Analysis of Efficient De-Noising of ECG Signals ...IRJET- Design Simulation and Analysis of Efficient De-Noising of ECG Signals ...
IRJET- Design Simulation and Analysis of Efficient De-Noising of ECG Signals ...IRJET Journal
 
Soliton comb generation using add-drop ring resonators
Soliton comb generation using add-drop ring resonatorsSoliton comb generation using add-drop ring resonators
Soliton comb generation using add-drop ring resonatorsPremier Publishers
 
Characterization of transients and fault diagnosis in transformer by discrete
Characterization of transients and fault diagnosis in transformer by discreteCharacterization of transients and fault diagnosis in transformer by discrete
Characterization of transients and fault diagnosis in transformer by discreteIAEME Publication
 
New Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet TransformNew Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet TransformCSCJournals
 
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...IRJET Journal
 
Removal of Clutter by Using Wavelet Transform For Wind Profiler
Removal of Clutter by Using Wavelet Transform For Wind ProfilerRemoval of Clutter by Using Wavelet Transform For Wind Profiler
Removal of Clutter by Using Wavelet Transform For Wind ProfilerIJMER
 
Empirical Mode Decomposition and Data Hiding in ECG Signal
Empirical Mode Decomposition and Data Hiding in ECG SignalEmpirical Mode Decomposition and Data Hiding in ECG Signal
Empirical Mode Decomposition and Data Hiding in ECG SignalIRJET Journal
 
Iisrt anjana francis (ec)
Iisrt anjana francis (ec)Iisrt anjana francis (ec)
Iisrt anjana francis (ec)IISRT
 

Similar to Plethysmographic Waveform Shapes (20)

IRJET-A Comparative Study of Digital FIR and IIR Band- Pass Filter
IRJET-A Comparative Study of Digital FIR and IIR Band- Pass FilterIRJET-A Comparative Study of Digital FIR and IIR Band- Pass Filter
IRJET-A Comparative Study of Digital FIR and IIR Band- Pass Filter
 
Ftir principle
Ftir principleFtir principle
Ftir principle
 
IRJET - Essential Features Extraction from Aaroh and Avroh of Indian Clas...
IRJET -  	  Essential Features Extraction from Aaroh and Avroh of Indian Clas...IRJET -  	  Essential Features Extraction from Aaroh and Avroh of Indian Clas...
IRJET - Essential Features Extraction from Aaroh and Avroh of Indian Clas...
 
Validation of IR instrument
Validation of IR instrumentValidation of IR instrument
Validation of IR instrument
 
Design of Photosynthetically Active Radiation Sensor_Crimson Publishers
Design of Photosynthetically Active Radiation Sensor_Crimson PublishersDesign of Photosynthetically Active Radiation Sensor_Crimson Publishers
Design of Photosynthetically Active Radiation Sensor_Crimson Publishers
 
International Refereed Journal of Engineering and Science (IRJES)
International Refereed Journal of Engineering and Science (IRJES)International Refereed Journal of Engineering and Science (IRJES)
International Refereed Journal of Engineering and Science (IRJES)
 
Performance Analysis of FIR Filter using FDATool
Performance Analysis of FIR Filter using FDAToolPerformance Analysis of FIR Filter using FDATool
Performance Analysis of FIR Filter using FDATool
 
UNIT I - ST.pptx
UNIT I - ST.pptxUNIT I - ST.pptx
UNIT I - ST.pptx
 
Rigol RF basics_knowledge_applications
Rigol RF basics_knowledge_applicationsRigol RF basics_knowledge_applications
Rigol RF basics_knowledge_applications
 
IRJET- Design Simulation and Analysis of Efficient De-Noising of ECG Signals ...
IRJET- Design Simulation and Analysis of Efficient De-Noising of ECG Signals ...IRJET- Design Simulation and Analysis of Efficient De-Noising of ECG Signals ...
IRJET- Design Simulation and Analysis of Efficient De-Noising of ECG Signals ...
 
Soliton comb generation using add-drop ring resonators
Soliton comb generation using add-drop ring resonatorsSoliton comb generation using add-drop ring resonators
Soliton comb generation using add-drop ring resonators
 
FTIR spectroscopy
FTIR spectroscopyFTIR spectroscopy
FTIR spectroscopy
 
Characterization of transients and fault diagnosis in transformer by discrete
Characterization of transients and fault diagnosis in transformer by discreteCharacterization of transients and fault diagnosis in transformer by discrete
Characterization of transients and fault diagnosis in transformer by discrete
 
New Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet TransformNew Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet Transform
 
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...
 
Removal of Clutter by Using Wavelet Transform For Wind Profiler
Removal of Clutter by Using Wavelet Transform For Wind ProfilerRemoval of Clutter by Using Wavelet Transform For Wind Profiler
Removal of Clutter by Using Wavelet Transform For Wind Profiler
 
Ecg
EcgEcg
Ecg
 
Empirical Mode Decomposition and Data Hiding in ECG Signal
Empirical Mode Decomposition and Data Hiding in ECG SignalEmpirical Mode Decomposition and Data Hiding in ECG Signal
Empirical Mode Decomposition and Data Hiding in ECG Signal
 
Iisrt anjana francis (ec)
Iisrt anjana francis (ec)Iisrt anjana francis (ec)
Iisrt anjana francis (ec)
 
Ih3514441454
Ih3514441454Ih3514441454
Ih3514441454
 

Recently uploaded

Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxUse of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxLoriGlavin3
 
Take control of your SAP testing with UiPath Test Suite
Take control of your SAP testing with UiPath Test SuiteTake control of your SAP testing with UiPath Test Suite
Take control of your SAP testing with UiPath Test SuiteDianaGray10
 
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptx
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptxPasskey Providers and Enabling Portability: FIDO Paris Seminar.pptx
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptxLoriGlavin3
 
"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr BaganFwdays
 
DevoxxFR 2024 Reproducible Builds with Apache Maven
DevoxxFR 2024 Reproducible Builds with Apache MavenDevoxxFR 2024 Reproducible Builds with Apache Maven
DevoxxFR 2024 Reproducible Builds with Apache MavenHervé Boutemy
 
A Deep Dive on Passkeys: FIDO Paris Seminar.pptx
A Deep Dive on Passkeys: FIDO Paris Seminar.pptxA Deep Dive on Passkeys: FIDO Paris Seminar.pptx
A Deep Dive on Passkeys: FIDO Paris Seminar.pptxLoriGlavin3
 
Nell’iperspazio con Rocket: il Framework Web di Rust!
Nell’iperspazio con Rocket: il Framework Web di Rust!Nell’iperspazio con Rocket: il Framework Web di Rust!
Nell’iperspazio con Rocket: il Framework Web di Rust!Commit University
 
Advanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionAdvanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionDilum Bandara
 
Training state-of-the-art general text embedding
Training state-of-the-art general text embeddingTraining state-of-the-art general text embedding
Training state-of-the-art general text embeddingZilliz
 
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxThe Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxLoriGlavin3
 
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxLoriGlavin3
 
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
Gen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfGen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfAddepto
 
What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024Stephanie Beckett
 
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024BookNet Canada
 
What is DBT - The Ultimate Data Build Tool.pdf
What is DBT - The Ultimate Data Build Tool.pdfWhat is DBT - The Ultimate Data Build Tool.pdf
What is DBT - The Ultimate Data Build Tool.pdfMounikaPolabathina
 
Artificial intelligence in cctv survelliance.pptx
Artificial intelligence in cctv survelliance.pptxArtificial intelligence in cctv survelliance.pptx
Artificial intelligence in cctv survelliance.pptxhariprasad279825
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity PlanDatabarracks
 
SIP trunking in Janus @ Kamailio World 2024
SIP trunking in Janus @ Kamailio World 2024SIP trunking in Janus @ Kamailio World 2024
SIP trunking in Janus @ Kamailio World 2024Lorenzo Miniero
 
TrustArc Webinar - How to Build Consumer Trust Through Data Privacy
TrustArc Webinar - How to Build Consumer Trust Through Data PrivacyTrustArc Webinar - How to Build Consumer Trust Through Data Privacy
TrustArc Webinar - How to Build Consumer Trust Through Data PrivacyTrustArc
 

Recently uploaded (20)

Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxUse of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
 
Take control of your SAP testing with UiPath Test Suite
Take control of your SAP testing with UiPath Test SuiteTake control of your SAP testing with UiPath Test Suite
Take control of your SAP testing with UiPath Test Suite
 
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptx
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptxPasskey Providers and Enabling Portability: FIDO Paris Seminar.pptx
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptx
 
"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan
 
DevoxxFR 2024 Reproducible Builds with Apache Maven
DevoxxFR 2024 Reproducible Builds with Apache MavenDevoxxFR 2024 Reproducible Builds with Apache Maven
DevoxxFR 2024 Reproducible Builds with Apache Maven
 
A Deep Dive on Passkeys: FIDO Paris Seminar.pptx
A Deep Dive on Passkeys: FIDO Paris Seminar.pptxA Deep Dive on Passkeys: FIDO Paris Seminar.pptx
A Deep Dive on Passkeys: FIDO Paris Seminar.pptx
 
Nell’iperspazio con Rocket: il Framework Web di Rust!
Nell’iperspazio con Rocket: il Framework Web di Rust!Nell’iperspazio con Rocket: il Framework Web di Rust!
Nell’iperspazio con Rocket: il Framework Web di Rust!
 
Advanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionAdvanced Computer Architecture – An Introduction
Advanced Computer Architecture – An Introduction
 
Training state-of-the-art general text embedding
Training state-of-the-art general text embeddingTraining state-of-the-art general text embedding
Training state-of-the-art general text embedding
 
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxThe Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
 
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
 
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
Gen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfGen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdf
 
What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024
 
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
 
What is DBT - The Ultimate Data Build Tool.pdf
What is DBT - The Ultimate Data Build Tool.pdfWhat is DBT - The Ultimate Data Build Tool.pdf
What is DBT - The Ultimate Data Build Tool.pdf
 
Artificial intelligence in cctv survelliance.pptx
Artificial intelligence in cctv survelliance.pptxArtificial intelligence in cctv survelliance.pptx
Artificial intelligence in cctv survelliance.pptx
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity Plan
 
SIP trunking in Janus @ Kamailio World 2024
SIP trunking in Janus @ Kamailio World 2024SIP trunking in Janus @ Kamailio World 2024
SIP trunking in Janus @ Kamailio World 2024
 
TrustArc Webinar - How to Build Consumer Trust Through Data Privacy
TrustArc Webinar - How to Build Consumer Trust Through Data PrivacyTrustArc Webinar - How to Build Consumer Trust Through Data Privacy
TrustArc Webinar - How to Build Consumer Trust Through Data Privacy
 

Plethysmographic Waveform Shapes

  • 1. Plethysmographic Waveform Plethysmographic Waveform Shapes Displayed by Pulse Oximeters Introduction Pulse oximeters utilize two wavelengths of light, red and infrared, to measure arterial oxygen saturation and pulse rate. Since the infrared (IR) light signal is relatively unaffected by changes in arterial oxygen saturation, it is often displayed on the front of the pulse oximeter as a plethysmographic waveform. Plethysmographic pulse waveform morphology varies from site to site depending on a variety of physiologic factors. For example, the waveform morphology is different if measured from the fingertip or from the ear. These differences have important physiologic causes including but not limited to variations in sympathetic innervation of the arteriolar bed at the monitoring site. Normal plethysmographic waveforms may or may not contain an incisura, or notch, resembling the classic dicrotic notch seen in arterial blood pressure tracings. This variation in appearance depends on many conditions. Clinicians have attempted to utilize this waveform and its variation for assessing signal integrity, physiological and artifactual changes such as perfusion changes, dysrhythmia, motion artifact, and electrical interference. For this reason accurate and reliable presentation of the plethysmographic waveform is of vital importance. Manufacturers normally filter the infrared signal to remove environmental effects and other disturbances, and to scale it for display. There are a number of filtering techniques and the method used may cause little or no change to the morphology of the patient signal or it may result in substantial distortion. It is desirable to preserve as much of the actual patient waveform shape as possible. We have examined the effects of two types of filters for various IR signals in an effort to understand the consequence of each on waveform integrity. Two common methods of filtering are finite impulse response (FIR) and infinite impulse response (IIR) filtering. These two methods vary in execution, as they require substantially different computational power for processing. IIR filters are typically found w w w. m a s i m o . c o m either in signal processing units consisting of analog circuitry or low-end digital processors. While highly desirable from a cost standpoint, IIR filters are also known for potentially unstable designs and nonlinear phase characteristics, which result in waveform distortion. FIR filters overcome both stability and waveform distortion, but at the expense of greater processing load. To facilitate the use of FIR filters, and thus preserve the integrity of the waveform, Masimo SET devices utilize a high end, digital signal processing (DSP) chip. While a DSP chip comes at a much greater expense, relative to low-end processors, waveform integrity is maintained. 92618 We tested the theoretical effects of FIR and IIR filtering using computer models. We then tested various input waveform shapes on three pulse oximeters, the Masimo Radical, Nellcor N-395, and ca the Datex-Ohmeda 3800, to examine the effects of each of their processing on waveform integrity irvine - i.e. did the waveform displayed on the pulse oximeters replicate the input waveform? 40 parker Methods Characteristic digit waveform shapes were obtained from both a Bio-Tek Index II and a Clinical Dynamics simulator for various pulse rates to assess the effects of the filters on changing masimo corporation physiological conditions. Plethysmographic waveforms seen in the clinic may vary greatly but include sharp ascending and descending limbs, and may contain, a characteristic notch on the descending limb. These characteristic waveforms were then subjected to filtering by FIR and IIR filters to examine their effect on the shape of the original input waveform when reproduced on the pulse oximeter display. In a separate experiment, these waveforms were applied to three pulse oximeters, the Masimo Radical, the Nellcor N-395, and the Datex-Ohmeda 3800, to examine the ability of each instrument to maintain waveform fidelity. The question being asked: is the output waveform displayed by the pulse oximeter identical to the input waveform? 1
  • 2. Plethysmographic Waveform Computer model Several waveform shapes were obtained from the Bio-Tek Index II and the Clinical Dynamics Smart Sat simulators. These shapes are typically seen in the clinical environment. The waveform shapes were then subjected to both FIR and IIR filtering on a personal computer. The results of these tests are depicted in Figures 1-4. Notch Test 0 1 2 3 4 5 6 7 8 9 10 time i Time (Second s) In pu t W av efo rm FIR Filter IIR Filter Zero Level fo r In p u t W av efo rm Zero Level fo r FIR Filter Ou tp u t Zero Level fo r IIR Ou tp u t Figure 1. FIR and IIR Filter Performance for a Waveform with Notch Input w w w. m a s i m o . c o m Figure 1 shows that the IIR filter causes distortion in the waveform. The notch is below midline, the descending limb is elongated (above the notch), and the descending limb (below the notch) is curved. 92618 ca irvine 40 parker masimo corporation
  • 3. technical bulletin R amp Test 0 1 2 3 4 5 6 7 8 9 10 time i Time (Second s) In pu t W av efo rm FIR Filter IIR Filter Zero Level fo r In p u t W av efo rm Zero Level fo r FIR Filter Ou tp u t Zero Level fo r IIR Ou tp u t Figure 2. FIR and IIR Filter Performance for a Ramp Input Figure 2 shows that the IIR filter causes distortion to the input waveform. The sharp, straight descending limb on the input signal is preserved with the FIR filter. In contrast, the IIR filter causes a curvilinear descending limb, which deviates greatly from the actual input waveform. Lo w Heart R ate Test 0 1 2 3 4 5 6 7 8 9 10 time i Time (Second s) In pu t W av efo rm FIR Filter IIR Filter Zero Level fo r In p u t W av efo rm Zero Level fo r FIR Filter Ou tp u t Zero Level fo r IIR Ou tp u t Figure 3. FIR and IIR Filter on a Low Heart Rate Input Figure 3 shows that the IIR filter causes distortion in the waveform. The notch is below midline, the descending limb is elongated (above the notch), and the descending limb (below the notch) is shortened and curved. In marked contrast, the FIR filter preserves the waveform integrity. 3
  • 4. Plethysmographic Waveform Asysto le Test 0 1 2 3 4 5 6 7 8 9 10 time i Time (Second s) In pu t W av efo rm FIR Filter IIR Filter Zero Level fo r In p u t W av efo rm Zero Level fo r FIR Filter Ou tp u t Zero Level fo r IIR Ou tp u t Figure 4. FIR and IIR Filter Performance for an Asystole Input Figure 4 demonstrates the delays associated with the different types of filtering. While the IIR filter distorts the waveform, it appears almost in sync with the input waveform. In contrast, the increased integrity of the waveform with the FIR filter comes at the cost of a slight delay in waveform appearance. The four figures show that the processing performed by the FIR filter maintains the integrity of w w w. m a s i m o . c o m the input waveform. However, due to the increased computational requirements of the FIR filter the resultant waveform is delayed slightly in time. The IIR filtering causes marked distortion of the waveform with little delay in waveform appearance. Conclusions 92618 The processing used by pulse oximeters in their plethysmographic waveform generation may have an effect on the integrity of the waveform and/or its temporal display. As shown above, the IIR filter ca clearly causes marked distortion of the shape of the waveform with a minor delay in the displayed irvine waveform appearing on the pulse oximeter screen. In contrast, the FIR filter with only a slight time increase to display the waveform, maintains the true shape of the waveform. 40 parker masimo corporation
  • 5. technical bulletin Waveforms Displayed by Pulse Oximeter Figure 5. This depicts an input waveform at 70 bpm. Notice sharp ascending and descending limbs. Figure 6. Photo of waveform at 70 bpm displayed by Masimo Radical in response to input signals shown in figure 5. Notice that the waveform integrity is maintained. Figure 7. Photo of waveform at 70 bpm displayed by Nellcor N-395 in response to input signals shown in figure 5. Notice that the waveform is distorted and shows a curvature, which is not present in the input signal. Figure 8. Photo of waveform at 70 bpm displayed by Datex-Ohmeda 3800 in response to input signals shown in figure 5. Notice that the waveform is distorted and shows a curvature, which is not present in the input signal. 5
  • 6. Plethysmographic Waveform Waveforms Displayed by Pulse Oximeter Figure 9. This depicts an input waveform at 72 bpm with notch at mid level of descending limb. Figure 10. Photo of notched waveform at 72 bpm displayed by Masimo Radical in response to input signals shown in figure 9. Notice the notch appears at mid level and waveform integrity is excellent. w w w. m a s i m o . c o m 92618 Figure 11. Photo of notched waveform at 72 bpm displayed by Nellcor N-395 in response to input signals shown in ca figure 9. Notice the distortion in waveform and the notch well below midline. irvine 40 parker masimo corporation Figure 12. Photo of notched waveform at 72 bpm as displayed by Datex-Ohmeda 3800 in response to input shown in figure 9. Notice the distortion in waveform and the notch well below midline.
  • 7. technical bulletin Waveforms Displayed by Pulse Oximeter Figure 13. This input waveform at 45 bpm with notch at mid level of descending limb. Figure 14. Photo of notched waveforms at 45 bpm as displayed by Masimo Radical in response to input signal shown in figure 13. Notice that the waveform integrity is preserved. Figure 15. Photo of notched waveform at 45 bpm as displayed by Nellcor N-395 in response to input signal shown in figure 13. Notice that the waveform is distorted and the notch is well below midline. Figure 16. Photo of notched waveform at 45 bpm as displayed by Datex-Ohmeda 3800 in response to input signal shown in figure 13. Notice that the waveform is distorted and the notch is well below midline. 7
  • 8. 7396-3618B-0108 Masimo Corporation 40 Parker Irvine, California 92618 Tel 949-297-7000 Fax 949-297-7001 www.masimo.com ©2008 Masimo Corporation. All rights reserved. Masimo, Radical, SET, ,and Signal Extraction Technology are registered trademarks of Masimo Corporation. All rights reserved. All other products are trademarks of their respective owners.