*An ultrasound artifact is a structure in an image which
does not directly similiar with actual tissue being
scanned.
*Artifact assumes different forms including :
• Structures in the image that are not actually present
• Objects that should be represented but are missing from
the image.
• Structures which are misregistered on the image.
1. REVERBERATION
2. ACOUSTIC SHADOWING
3. ACOUSTIC ENHANCEMENT
4. EDGE SHADOWING
5. BEAM WIDTH ARTIFACT
6. SLICE THICKNESS ARTIFACT
7. SIDE LOBE ARTIFACT
8. MIROR IMAGE
9. DOUBLE IMAGE
10.EQUIPMENT-GENERATED ARTIFACT
11.REFRACTION ARTIFACT
• This is the production of false echoes due to repeated
reflections between two interfaces with a high acoustic
impedance mismatch.
• The echo from the interface is received by the transducer
and displayed on the image.
• Some of the energy in the returned echo is reflected at the
transducer face, and return to the reflecting interface as if
it was a weak transmitted pulse, returning as a second
echo.
• As the time taken for the second echo to arrive is twice
that taken by the depth.
*
• This sequence of reflection and transmission can occur
many times, with the third echo taking three times as long
to return to the transducer and being displayed at three
times the depth, and so on.
• The reverberation echoes will be equally spaced because
the time for each additional echo is multiple of the time of
return of the first echo.
• These reverberation echoes will be strong because of the
high acoustic mismatch.
This artifact will be seen at the skin-transducer interface
and behind bowel gas.
Rectification:
• Increase the amount of gel used.
• Use a stand-off pad.
• Reduce the gain.
• Move the position of the transducer.
• This appears as an area of low amplitude echoes behind an
area of strongly attenuating tissue.
• It is caused by severe attenuation of the beam at an
interface, resulting in very little sound being transmitted
beyond.
• The attenuation can be due to either absorption or
reflection of the sound waves, or a combination of the
two.
• Acoustic shadowing will occur at interfaces with large
acoustic mismatch such as:
Soft tissue and gas
Soft tissue and bone or calculus
*
• This artifact appears as a localized area of increased echo
amplitude behind an area of low attenuation.
• On a scan it will appears as an area of increased brightness,
and can commonly be seen distal to fluid-filled structures
such as the urinary bladder, GB or a cyst.
• The artifact arises due to the application of the time-gain
compensation(TGC) to areas of low attenuating structures
such as fluid.
• It is caused by the low level of attenuation of the beam as it
passes through fluid relative to the greater attenuation of
the beam in the adjacent more solid tissue.
*
• This artifact can often be an useful diagnostic aid,
particularly when scanning a soft-tissue mass or cyst
containing low level echoes.
• These echoes may often cause the structure to disappear
in the image as it blend into the surrounding echo pattern.
• A combination of refraction and reflection occurring at the
edges of rounded structures will result in edge shadowing
artifact.
• It arises due to refraction of the beam caused by both the
curvature of the rounded edges and difference in speed of
two materials.
• When the ultrasound beam reaches the rounded edge of a
structure, reflection will occur, with an angle of incidence
equal to the angle of reflection.
• The outer part of the beam will be totally reflected, but
the reminder of the beam passes through the rounded
structure and is refracted.
*
• This combination of reflection and refraction of the
beam at the edges of a rounded structure results in a
thin strip of tissue behind the edge not being insonated
and causes a shadow.
• This artifact can be demonstrated by scanning a point
reflector in a phantom, where the display will clearly
portray this as a line.
• During routine scanning, the artifact can be seen when
spurious echoes are displayed in an echo-free area.
• Correct positioning of the focal zone will help to reduce
this artifact.
• The focal zone is controlled by electronically narrowing
the beam
*
• These occurs due to the thickness of the beam, and are
similar to beam width artifacts.
• These artifacts will typically be seen in transverse views of
the urinary bladder when structures adjacent to the slice
through the bladder being scanned will be incorporated
into the image.
• These echoes are then displayed as if they were arising
from within the bladder.
• Although the appearance of this artifact is similar to the
beam width artifact, the differentiating factors is that the
reflector causing the slice thickness artifact will not be
seen on the display.
*
• This artifact is a result of inherent characteristics of the
transducer, and apart from trying a different transducer,
cannot be eliminated.
• The energy within the ultrasound beam exists as several
side lobes radiating at a number of angles from a central
lobe.
• Echoes are generated by these side lobes in addition to
the main lobe, but all the returning echoes are assumed
by the transducer to have arisen from the central axis of
the main lobe.
• Side lobe echoes will therefore be misregistered in the
display.
• This artifact can often be seen in area such as the urinary
bladder and may also arise within a cyst.
*
• These artifacts results in a mirror image of a structure
occurring in an ultrasound display.
• They arise due to specular reflection of the beam at a
large smooth interface.
• An area close to a specular reflector will be imaged twice,
once by the original ultrasound beam and once by the
beam after it has reflected off the specular reflector.
• Mirror image artifacts are most commonly seen where
there is a large acoustic mismatch, such as a fluid-air
interface.
*
• Typically this artifact can occur during the scanning of a
full bladder, when air in the rectum behind the bladder
act as specular reflector and mirror image of the
bladder is displayed posteriorly.
• It will then have the appearance of a large cyst behind
the bladder.
• It can also be seen when scanning the liver, and the
diaphragm act as a specular reflector.
• This image is caused by refraction of the beam and may
occur in areas such as the rectus abdominis muscle on the
anterior abdominal wall.
• In the transverse plane the edges of the muscle act as a
lens and the ultrasound beam to be refracted and this
causes the single structure to be interrogated by two
separate refracted beams.
• Two sets of echoes will therefore be returned and these
will cause display of two structures in the image.
*
This results in, for example, two images of the transverse
aorta side by side in the abdomen.
• Incorrect use of the equipment controls can lead to artifact
appearing.
• Misuse of controls such as the gain or TGC can result in
echoes being recorded as too bright or too dark.
• Care must be taken when setting these controls, to ensure
an even brightness throughout the image.
• If the dynamic range control is incorrectly set, this can lead
to an image which has too much contrast, and result in the
loss of subtle echo information.
*
• Gain must be in medium level.
• Blurring of a moving image can occur if the frame rate is
too low or if the persistence is too high.
• It is important to ensure that the frame rate is capable of
recording a moving structure at the speed.
• Use of multiple focal zones can give rise to a prominent
banding effect within the image.
*
• The refraction is the change of the sound direction on
passing from one medium to another.
• In ultrasound, refraction is due to sound velocity
mismatches combines with oblique angles of incidence,
most commonly with convex scanheads.
• When the ultrasound wave crosses at an oblique angle the
interface of two materials, through which the waves
propagate at different velocities, refraction occurs, caused
by bending of the wave beam.
• Refraction artifact cause spatial distortion and loss of
resolution in the image.
Ultrasound artifacts

Contenu connexe

Tendances

B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING Ganesan Yogananthem
 
CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis Upakar Paudel
 
Chest x ray positioning
Chest x ray  positioningChest x ray  positioning
Chest x ray positioningairwave12
 
Contrast enhanced ultrasound
Contrast enhanced ultrasoundContrast enhanced ultrasound
Contrast enhanced ultrasoundNeeraj Patange
 
Mri coils
Mri coilsMri coils
Mri coilsSelf
 
Interventional radiology part 1
Interventional radiology part 1Interventional radiology part 1
Interventional radiology part 1Chandni Wadhwani
 
ULTRASOUND PHYSICS
ULTRASOUND PHYSICSULTRASOUND PHYSICS
ULTRASOUND PHYSICSNavni Garg
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkDr pradeep Kumar
 
Radiographic cassettes
Radiographic cassettesRadiographic cassettes
Radiographic cassettesSudil Paudyal
 
Doppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USGDoppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USGSudil Paudyal
 
Interactions of ultrasound with matter
Interactions of ultrasound with matterInteractions of ultrasound with matter
Interactions of ultrasound with matterSwapnil Shetty
 
Procedure of ercp and t tube cholangiography
Procedure of ercp and t tube cholangiographyProcedure of ercp and t tube cholangiography
Procedure of ercp and t tube cholangiographyYashawant Yadav
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiationNitish Rajput
 
CT Physics
CT PhysicsCT Physics
CT PhysicsRMLIMS
 

Tendances (20)

B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
 
CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis
 
the lower limb positioning
the lower limb positioningthe lower limb positioning
the lower limb positioning
 
Chest x ray positioning
Chest x ray  positioningChest x ray  positioning
Chest x ray positioning
 
Contrast enhanced ultrasound
Contrast enhanced ultrasoundContrast enhanced ultrasound
Contrast enhanced ultrasound
 
Mri coils
Mri coilsMri coils
Mri coils
 
Interventional radiology part 1
Interventional radiology part 1Interventional radiology part 1
Interventional radiology part 1
 
ULTRASOUND PHYSICS
ULTRASOUND PHYSICSULTRASOUND PHYSICS
ULTRASOUND PHYSICS
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pk
 
Radiographic cassettes
Radiographic cassettesRadiographic cassettes
Radiographic cassettes
 
Doppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USGDoppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USG
 
Interactions of ultrasound with matter
Interactions of ultrasound with matterInteractions of ultrasound with matter
Interactions of ultrasound with matter
 
Usg Basics
Usg BasicsUsg Basics
Usg Basics
 
Procedure of ercp and t tube cholangiography
Procedure of ercp and t tube cholangiographyProcedure of ercp and t tube cholangiography
Procedure of ercp and t tube cholangiography
 
CR & DR
CR & DRCR & DR
CR & DR
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiation
 
CT Physics
CT PhysicsCT Physics
CT Physics
 
soft tissue radiography
soft tissue radiographysoft tissue radiography
soft tissue radiography
 
CT Artifacts
CT ArtifactsCT Artifacts
CT Artifacts
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
 

Similaire à Ultrasound artifacts

Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound ArtifactAmer Sohail
 
Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pkDr pradeep Kumar
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundArjun Reddy
 
ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxAnayaAnaya14
 
Ultrasound of eye - B scan
Ultrasound of eye - B scan Ultrasound of eye - B scan
Ultrasound of eye - B scan Shruti Laddha
 
Artifacts in echocardiography
Artifacts  in echocardiographyArtifacts  in echocardiography
Artifacts in echocardiographyDilzith Katepogu
 
Physics of ultrasound imaging
Physics of ultrasound imagingPhysics of ultrasound imaging
Physics of ultrasound imagingmtuzer
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxShubham661884
 
ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT  ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT PAWAN JARWAL
 
Probe construction, types and usg artefacts
Probe construction, types and usg artefactsProbe construction, types and usg artefacts
Probe construction, types and usg artefactsVishakha Mittal
 
Artifacts and Bioeffects
Artifacts and Bioeffects Artifacts and Bioeffects
Artifacts and Bioeffects Nusirat Adedewe
 
Basics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallBasics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallGamal Agmy
 
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...Gamal Agmy
 

Similaire à Ultrasound artifacts (20)

Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound Artifact
 
Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pk
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasound
 
ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptx
 
Us artefacts
Us artefactsUs artefacts
Us artefacts
 
Ultrasound of eye - B scan
Ultrasound of eye - B scan Ultrasound of eye - B scan
Ultrasound of eye - B scan
 
Us physics 3
Us physics 3Us physics 3
Us physics 3
 
Ct artiact
Ct artiactCt artiact
Ct artiact
 
Usg
UsgUsg
Usg
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
Artifacts in echocardiography
Artifacts  in echocardiographyArtifacts  in echocardiography
Artifacts in echocardiography
 
Physics of ultrasound imaging
Physics of ultrasound imagingPhysics of ultrasound imaging
Physics of ultrasound imaging
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
ULTRA.pptx
ULTRA.pptxULTRA.pptx
ULTRA.pptx
 
ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT  ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT
 
Probe construction, types and usg artefacts
Probe construction, types and usg artefactsProbe construction, types and usg artefacts
Probe construction, types and usg artefacts
 
Artifacts and Bioeffects
Artifacts and Bioeffects Artifacts and Bioeffects
Artifacts and Bioeffects
 
Basics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallBasics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wall
 
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
 

Dernier

Virtual-Orientation-on-the-Administration-of-NATG12-NATG6-and-ELLNA.pdf
Virtual-Orientation-on-the-Administration-of-NATG12-NATG6-and-ELLNA.pdfVirtual-Orientation-on-the-Administration-of-NATG12-NATG6-and-ELLNA.pdf
Virtual-Orientation-on-the-Administration-of-NATG12-NATG6-and-ELLNA.pdfErwinPantujan2
 
INTRODUCTION TO CATHOLIC CHRISTOLOGY.pptx
INTRODUCTION TO CATHOLIC CHRISTOLOGY.pptxINTRODUCTION TO CATHOLIC CHRISTOLOGY.pptx
INTRODUCTION TO CATHOLIC CHRISTOLOGY.pptxHumphrey A Beña
 
ROLES IN A STAGE PRODUCTION in arts.pptx
ROLES IN A STAGE PRODUCTION in arts.pptxROLES IN A STAGE PRODUCTION in arts.pptx
ROLES IN A STAGE PRODUCTION in arts.pptxVanesaIglesias10
 
4.16.24 Poverty and Precarity--Desmond.pptx
4.16.24 Poverty and Precarity--Desmond.pptx4.16.24 Poverty and Precarity--Desmond.pptx
4.16.24 Poverty and Precarity--Desmond.pptxmary850239
 
Field Attribute Index Feature in Odoo 17
Field Attribute Index Feature in Odoo 17Field Attribute Index Feature in Odoo 17
Field Attribute Index Feature in Odoo 17Celine George
 
ICS2208 Lecture6 Notes for SL spaces.pdf
ICS2208 Lecture6 Notes for SL spaces.pdfICS2208 Lecture6 Notes for SL spaces.pdf
ICS2208 Lecture6 Notes for SL spaces.pdfVanessa Camilleri
 
TEACHER REFLECTION FORM (NEW SET........).docx
TEACHER REFLECTION FORM (NEW SET........).docxTEACHER REFLECTION FORM (NEW SET........).docx
TEACHER REFLECTION FORM (NEW SET........).docxruthvilladarez
 
How to do quick user assign in kanban in Odoo 17 ERP
How to do quick user assign in kanban in Odoo 17 ERPHow to do quick user assign in kanban in Odoo 17 ERP
How to do quick user assign in kanban in Odoo 17 ERPCeline George
 
The Contemporary World: The Globalization of World Politics
The Contemporary World: The Globalization of World PoliticsThe Contemporary World: The Globalization of World Politics
The Contemporary World: The Globalization of World PoliticsRommel Regala
 
EMBODO Lesson Plan Grade 9 Law of Sines.docx
EMBODO Lesson Plan Grade 9 Law of Sines.docxEMBODO Lesson Plan Grade 9 Law of Sines.docx
EMBODO Lesson Plan Grade 9 Law of Sines.docxElton John Embodo
 
Active Learning Strategies (in short ALS).pdf
Active Learning Strategies (in short ALS).pdfActive Learning Strategies (in short ALS).pdf
Active Learning Strategies (in short ALS).pdfPatidar M
 
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...Postal Advocate Inc.
 
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptxQ4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptxlancelewisportillo
 
Millenials and Fillennials (Ethical Challenge and Responses).pptx
Millenials and Fillennials (Ethical Challenge and Responses).pptxMillenials and Fillennials (Ethical Challenge and Responses).pptx
Millenials and Fillennials (Ethical Challenge and Responses).pptxJanEmmanBrigoli
 
Presentation Activity 2. Unit 3 transv.pptx
Presentation Activity 2. Unit 3 transv.pptxPresentation Activity 2. Unit 3 transv.pptx
Presentation Activity 2. Unit 3 transv.pptxRosabel UA
 
Grade 9 Quarter 4 Dll Grade 9 Quarter 4 DLL.pdf
Grade 9 Quarter 4 Dll Grade 9 Quarter 4 DLL.pdfGrade 9 Quarter 4 Dll Grade 9 Quarter 4 DLL.pdf
Grade 9 Quarter 4 Dll Grade 9 Quarter 4 DLL.pdfJemuel Francisco
 
MULTIDISCIPLINRY NATURE OF THE ENVIRONMENTAL STUDIES.pptx
MULTIDISCIPLINRY NATURE OF THE ENVIRONMENTAL STUDIES.pptxMULTIDISCIPLINRY NATURE OF THE ENVIRONMENTAL STUDIES.pptx
MULTIDISCIPLINRY NATURE OF THE ENVIRONMENTAL STUDIES.pptxAnupkumar Sharma
 
Influencing policy (training slides from Fast Track Impact)
Influencing policy (training slides from Fast Track Impact)Influencing policy (training slides from Fast Track Impact)
Influencing policy (training slides from Fast Track Impact)Mark Reed
 

Dernier (20)

Virtual-Orientation-on-the-Administration-of-NATG12-NATG6-and-ELLNA.pdf
Virtual-Orientation-on-the-Administration-of-NATG12-NATG6-and-ELLNA.pdfVirtual-Orientation-on-the-Administration-of-NATG12-NATG6-and-ELLNA.pdf
Virtual-Orientation-on-the-Administration-of-NATG12-NATG6-and-ELLNA.pdf
 
INTRODUCTION TO CATHOLIC CHRISTOLOGY.pptx
INTRODUCTION TO CATHOLIC CHRISTOLOGY.pptxINTRODUCTION TO CATHOLIC CHRISTOLOGY.pptx
INTRODUCTION TO CATHOLIC CHRISTOLOGY.pptx
 
ROLES IN A STAGE PRODUCTION in arts.pptx
ROLES IN A STAGE PRODUCTION in arts.pptxROLES IN A STAGE PRODUCTION in arts.pptx
ROLES IN A STAGE PRODUCTION in arts.pptx
 
4.16.24 Poverty and Precarity--Desmond.pptx
4.16.24 Poverty and Precarity--Desmond.pptx4.16.24 Poverty and Precarity--Desmond.pptx
4.16.24 Poverty and Precarity--Desmond.pptx
 
Field Attribute Index Feature in Odoo 17
Field Attribute Index Feature in Odoo 17Field Attribute Index Feature in Odoo 17
Field Attribute Index Feature in Odoo 17
 
YOUVE_GOT_EMAIL_PRELIMS_EL_DORADO_2024.pptx
YOUVE_GOT_EMAIL_PRELIMS_EL_DORADO_2024.pptxYOUVE_GOT_EMAIL_PRELIMS_EL_DORADO_2024.pptx
YOUVE_GOT_EMAIL_PRELIMS_EL_DORADO_2024.pptx
 
ICS2208 Lecture6 Notes for SL spaces.pdf
ICS2208 Lecture6 Notes for SL spaces.pdfICS2208 Lecture6 Notes for SL spaces.pdf
ICS2208 Lecture6 Notes for SL spaces.pdf
 
TEACHER REFLECTION FORM (NEW SET........).docx
TEACHER REFLECTION FORM (NEW SET........).docxTEACHER REFLECTION FORM (NEW SET........).docx
TEACHER REFLECTION FORM (NEW SET........).docx
 
How to do quick user assign in kanban in Odoo 17 ERP
How to do quick user assign in kanban in Odoo 17 ERPHow to do quick user assign in kanban in Odoo 17 ERP
How to do quick user assign in kanban in Odoo 17 ERP
 
The Contemporary World: The Globalization of World Politics
The Contemporary World: The Globalization of World PoliticsThe Contemporary World: The Globalization of World Politics
The Contemporary World: The Globalization of World Politics
 
EMBODO Lesson Plan Grade 9 Law of Sines.docx
EMBODO Lesson Plan Grade 9 Law of Sines.docxEMBODO Lesson Plan Grade 9 Law of Sines.docx
EMBODO Lesson Plan Grade 9 Law of Sines.docx
 
Active Learning Strategies (in short ALS).pdf
Active Learning Strategies (in short ALS).pdfActive Learning Strategies (in short ALS).pdf
Active Learning Strategies (in short ALS).pdf
 
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
 
INCLUSIVE EDUCATION PRACTICES FOR TEACHERS AND TRAINERS.pptx
INCLUSIVE EDUCATION PRACTICES FOR TEACHERS AND TRAINERS.pptxINCLUSIVE EDUCATION PRACTICES FOR TEACHERS AND TRAINERS.pptx
INCLUSIVE EDUCATION PRACTICES FOR TEACHERS AND TRAINERS.pptx
 
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptxQ4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
 
Millenials and Fillennials (Ethical Challenge and Responses).pptx
Millenials and Fillennials (Ethical Challenge and Responses).pptxMillenials and Fillennials (Ethical Challenge and Responses).pptx
Millenials and Fillennials (Ethical Challenge and Responses).pptx
 
Presentation Activity 2. Unit 3 transv.pptx
Presentation Activity 2. Unit 3 transv.pptxPresentation Activity 2. Unit 3 transv.pptx
Presentation Activity 2. Unit 3 transv.pptx
 
Grade 9 Quarter 4 Dll Grade 9 Quarter 4 DLL.pdf
Grade 9 Quarter 4 Dll Grade 9 Quarter 4 DLL.pdfGrade 9 Quarter 4 Dll Grade 9 Quarter 4 DLL.pdf
Grade 9 Quarter 4 Dll Grade 9 Quarter 4 DLL.pdf
 
MULTIDISCIPLINRY NATURE OF THE ENVIRONMENTAL STUDIES.pptx
MULTIDISCIPLINRY NATURE OF THE ENVIRONMENTAL STUDIES.pptxMULTIDISCIPLINRY NATURE OF THE ENVIRONMENTAL STUDIES.pptx
MULTIDISCIPLINRY NATURE OF THE ENVIRONMENTAL STUDIES.pptx
 
Influencing policy (training slides from Fast Track Impact)
Influencing policy (training slides from Fast Track Impact)Influencing policy (training slides from Fast Track Impact)
Influencing policy (training slides from Fast Track Impact)
 

Ultrasound artifacts

  • 2. * *An ultrasound artifact is a structure in an image which does not directly similiar with actual tissue being scanned. *Artifact assumes different forms including : • Structures in the image that are not actually present • Objects that should be represented but are missing from the image. • Structures which are misregistered on the image.
  • 3. 1. REVERBERATION 2. ACOUSTIC SHADOWING 3. ACOUSTIC ENHANCEMENT 4. EDGE SHADOWING 5. BEAM WIDTH ARTIFACT 6. SLICE THICKNESS ARTIFACT 7. SIDE LOBE ARTIFACT 8. MIROR IMAGE 9. DOUBLE IMAGE 10.EQUIPMENT-GENERATED ARTIFACT 11.REFRACTION ARTIFACT
  • 4. • This is the production of false echoes due to repeated reflections between two interfaces with a high acoustic impedance mismatch. • The echo from the interface is received by the transducer and displayed on the image. • Some of the energy in the returned echo is reflected at the transducer face, and return to the reflecting interface as if it was a weak transmitted pulse, returning as a second echo. • As the time taken for the second echo to arrive is twice that taken by the depth. *
  • 5. • This sequence of reflection and transmission can occur many times, with the third echo taking three times as long to return to the transducer and being displayed at three times the depth, and so on. • The reverberation echoes will be equally spaced because the time for each additional echo is multiple of the time of return of the first echo. • These reverberation echoes will be strong because of the high acoustic mismatch.
  • 6. This artifact will be seen at the skin-transducer interface and behind bowel gas. Rectification: • Increase the amount of gel used. • Use a stand-off pad. • Reduce the gain. • Move the position of the transducer.
  • 7.
  • 8. • This appears as an area of low amplitude echoes behind an area of strongly attenuating tissue. • It is caused by severe attenuation of the beam at an interface, resulting in very little sound being transmitted beyond. • The attenuation can be due to either absorption or reflection of the sound waves, or a combination of the two. • Acoustic shadowing will occur at interfaces with large acoustic mismatch such as: Soft tissue and gas Soft tissue and bone or calculus *
  • 9.
  • 10.
  • 11. • This artifact appears as a localized area of increased echo amplitude behind an area of low attenuation. • On a scan it will appears as an area of increased brightness, and can commonly be seen distal to fluid-filled structures such as the urinary bladder, GB or a cyst. • The artifact arises due to the application of the time-gain compensation(TGC) to areas of low attenuating structures such as fluid. • It is caused by the low level of attenuation of the beam as it passes through fluid relative to the greater attenuation of the beam in the adjacent more solid tissue. *
  • 12. • This artifact can often be an useful diagnostic aid, particularly when scanning a soft-tissue mass or cyst containing low level echoes. • These echoes may often cause the structure to disappear in the image as it blend into the surrounding echo pattern.
  • 13. • A combination of refraction and reflection occurring at the edges of rounded structures will result in edge shadowing artifact. • It arises due to refraction of the beam caused by both the curvature of the rounded edges and difference in speed of two materials. • When the ultrasound beam reaches the rounded edge of a structure, reflection will occur, with an angle of incidence equal to the angle of reflection. • The outer part of the beam will be totally reflected, but the reminder of the beam passes through the rounded structure and is refracted. *
  • 14. • This combination of reflection and refraction of the beam at the edges of a rounded structure results in a thin strip of tissue behind the edge not being insonated and causes a shadow.
  • 15. • This artifact can be demonstrated by scanning a point reflector in a phantom, where the display will clearly portray this as a line. • During routine scanning, the artifact can be seen when spurious echoes are displayed in an echo-free area. • Correct positioning of the focal zone will help to reduce this artifact. • The focal zone is controlled by electronically narrowing the beam *
  • 16.
  • 17. • These occurs due to the thickness of the beam, and are similar to beam width artifacts. • These artifacts will typically be seen in transverse views of the urinary bladder when structures adjacent to the slice through the bladder being scanned will be incorporated into the image. • These echoes are then displayed as if they were arising from within the bladder. • Although the appearance of this artifact is similar to the beam width artifact, the differentiating factors is that the reflector causing the slice thickness artifact will not be seen on the display. *
  • 18. • This artifact is a result of inherent characteristics of the transducer, and apart from trying a different transducer, cannot be eliminated.
  • 19. • The energy within the ultrasound beam exists as several side lobes radiating at a number of angles from a central lobe. • Echoes are generated by these side lobes in addition to the main lobe, but all the returning echoes are assumed by the transducer to have arisen from the central axis of the main lobe. • Side lobe echoes will therefore be misregistered in the display. • This artifact can often be seen in area such as the urinary bladder and may also arise within a cyst. *
  • 20.
  • 21. • These artifacts results in a mirror image of a structure occurring in an ultrasound display. • They arise due to specular reflection of the beam at a large smooth interface. • An area close to a specular reflector will be imaged twice, once by the original ultrasound beam and once by the beam after it has reflected off the specular reflector. • Mirror image artifacts are most commonly seen where there is a large acoustic mismatch, such as a fluid-air interface. *
  • 22. • Typically this artifact can occur during the scanning of a full bladder, when air in the rectum behind the bladder act as specular reflector and mirror image of the bladder is displayed posteriorly. • It will then have the appearance of a large cyst behind the bladder. • It can also be seen when scanning the liver, and the diaphragm act as a specular reflector.
  • 23.
  • 24. • This image is caused by refraction of the beam and may occur in areas such as the rectus abdominis muscle on the anterior abdominal wall. • In the transverse plane the edges of the muscle act as a lens and the ultrasound beam to be refracted and this causes the single structure to be interrogated by two separate refracted beams. • Two sets of echoes will therefore be returned and these will cause display of two structures in the image. *
  • 25. This results in, for example, two images of the transverse aorta side by side in the abdomen.
  • 26. • Incorrect use of the equipment controls can lead to artifact appearing. • Misuse of controls such as the gain or TGC can result in echoes being recorded as too bright or too dark. • Care must be taken when setting these controls, to ensure an even brightness throughout the image. • If the dynamic range control is incorrectly set, this can lead to an image which has too much contrast, and result in the loss of subtle echo information. *
  • 27. • Gain must be in medium level. • Blurring of a moving image can occur if the frame rate is too low or if the persistence is too high. • It is important to ensure that the frame rate is capable of recording a moving structure at the speed. • Use of multiple focal zones can give rise to a prominent banding effect within the image.
  • 28.
  • 29. * • The refraction is the change of the sound direction on passing from one medium to another. • In ultrasound, refraction is due to sound velocity mismatches combines with oblique angles of incidence, most commonly with convex scanheads. • When the ultrasound wave crosses at an oblique angle the interface of two materials, through which the waves propagate at different velocities, refraction occurs, caused by bending of the wave beam. • Refraction artifact cause spatial distortion and loss of resolution in the image.