Defibrillators
Use & How
Dr.Imran
Definition
™Defibrillator is a device that deliver a
therapeutic dose of electrical energy (electric
shock) to the affected heart (fibrillated heart or
other shockable rhythm) to force the heart to
produce more normal cardiac rhythm.
First demonstrated on dogs in 1899 by Jean-Louis Prévost & Frederic
Batelli, two physiologists from University of Geneva, Switzerland.
The first use on a human was in 1947 by Claude Beck,
professor of surgery at Case Western Reserve University.
Transthoracic defibrillation was first used in humans using
alternating current (AC)
Bernard Lown and his coworkers introduced direct current
(DC) defibrillators into clinical practice.
History
Defibrillator
Does not re-start the heart.
Stops all electrical activity.
If heart is still viable its natural pace maker will take over.
It is an emergency life saving procedure.
Probability decreases with time.
Mechanism
Current depolarizes the myocardium.
Induces asystole temporarily.
Allows SA node to take over coz it‟s the first to repolarize.
Indications
- Pulse-less polymorphic VT/VF.
- When unsure weather monomorphic or
polymorphic VT in a un-stable patient.
Non Shockable rhythm’s
- Asystole
- Pulse less electrical activity (PEA)
Loss of synchronization in heart is called as
fibrillation.
 Types of fibrillation
 Atrial fibrillation
 Ventricular fibrillation
Defibrillator……
™Defibrillation is the definitive treatment for the life threatening
cardiac arrhythmias ventricular fibrillation and pulseless
ventricular tachycardia
™
Ventricular fibrillation results from:
-Coronary occlusion
-Electrical shock
-Abnormalities of body chemistry
™
This irregular contraction of the muscle fibers causes non
effectively blood pumping and that results in a steep fall of cardiac
output.
Need for a Defibrillator
™Ventricular fibrillation can be converted into a more efficient rhythm
by applying high energy shock to the heart.
™
This causes all muscle fibers to contract simultaneously, which may
then respond to normal physiological pacemaking pulses
Restoration of normal rhythm in fibrillating heart as achieved by direct current shock (arrow)
across the chest wall. The horizontal line after the shock shows that the cardiograph was blocked
or disconnected for its protection during the period of shock4
Ventricular fibrillation
This recommendation regarding CPR prior
to attempted defibrillation is supported by
2 clinical studies
JAMA. 2003;289:1389 –1395
JAMA. 1999;281:1182–1188
 Applying a brief(.25 to 1 sec) burst of 60 HZ ac at an
intensity of around 6 A.
 This application of an electrical shock to resynchronize
the heart is sometimes called counter shock.
 If the patient does not respond, the burst is repeated
until defibrillation occurs. this method is known as ac
defibrillation.
AC DEFIBRILLATION
• It is cannot be successfully used to correct atrial
fibrillation.
• Successive attempts to correct ventricular fibrillation are
often required.
• Attempts to correct atrial fibrillation by this method often
result more serious ventricular fibrillation.
Disadvantage of using ac Defibrillator
 In this method a capacitor is charged to a high dc voltage and
then rapidly discharged.
 The amount of energy discharged by the capacitor may range
between 2 to 400joules with peak value of current 20A.
 A corrective shock of 750-800 volts is applied within a tenth of a
second .
DC Defibrillation
CIRCUIT OF DC DEFIBRILLATOR
• Energy storage capacitor is charged at relatively slow
rate from AC line.
• Energy stored in capacitor is then delivered at a
relatively rapid rate to chest of the patient.
• Simple arrangement involve the discharge of capacitor
energy through the patient’s own resistance.
PRINCIPLE OF DEFIBRILLATOR
Cont…..
The discharge resistance which the patient represents is
roughly a ohmic resistance of 50 – 100 ohms for a typical
electrode size of 80cm2.
The particular wave form is called “Lown” wave form.
The pulse width of this waveform is 10ms.
1. Manual external defibrillator
2. Manual internal defibrillator
3. Semi-Automated External Defibrillator
4. Automated external defibrillator (AED)
5. Implantable cardioverter-defibrillator (ICD)
{automatic internal cardiac defibrillator (AICD)}
6. Wearable cardiac defibrillator
Types of Defibrillators
Manual external defibrillator
Electrodes placed directly around the heart area of chest.
Higher Voltage required than internal defibrillator.
Classified as -
Monophasic
Biphasic
Monophasic waveform Defibrillators
• Deliver current of one polarity.
• Current travels in one direction through the patients heart from one
paddle to another.
2 types :-
• The monophasic damped sinusoidal waveform (MDS) returns to
zero gradually
• Monophasic truncated exponential waveform (MTE) current is
abruptly returned to baseline (truncated) to zero current flow
MDS v/s MTE wave form
Biphasic waveform Defibrillators
Current travels towards the +ve paddle & then reverses back.
Reversing of polarity, depolarizes all cells – called “burping”
response.
Classified into –
Biphasic truncated exponential waveform (BTE)
Rectilinear biphasic waveform (RLB)
RLB is better than BTE.
Biphasic truncated exponential waveform (BTE)
v/s
Rectilinear biphasic waveform (RLB)
RBL BTE
Advantages of Biphasic over Monophasic
Less power – Less trauma – Less battery.
Defibrillation more effective at low energy.
Fewer burns.
Less myocardial damage.
1st shock success rate in cardiac arrest due to shockable
rhythm –
Monophasic 60%
Biphasic increases to 90%
Manual internal defibrillator
Just like normal defilbrillator.
Electric charge is usually pumped in by internal
paddles in close or direct contact with the heart.
These devices are mainly used in operating theatres,
where the chest can be opened or being treated by a
surgeon
Semi-Automated External Defibrillator
Carries features of both full-manuated as well as
automated units of defibrillator.
Has an ECG display and manual override in it.
Generally used by paramedics and emergency medicine
technicians.
Implantable cardioverter-defibrillator (ICD)
{automatic internal cardiac defibrillator (AICD)}
A implanted device that detacts and terminats life threatning
episodes of VF / VT in high risk patients.
Wearable cardiac defibrillator
- Life vest
• AED is a portable type of external defibrillator that
automatically diagnoses the ventricular fibrillation in a patient.
• Automatic refers to the ability to autonomously analyze the patient's
condition.
• AED is provided with self-adhesive electrodes instead of hand held
paddles.
AUTOMATIC EXTERNAL
DEFIBRILLATOR
Automated external defibrillator (AED) cont…
Use does not require special medical training.
Found in public places – offices, airport, shopping mall.
The electrodes transmit information about the person's heart rhythm to a
controller in the AED.
The controller examines the electrical output from the heart and determine if
the patient is in a shockable rhythm or not and whether shock is needed.
The AED uses voice / visual prompts to tell user when to give the shock,
and the electrodes deliver it.
Paddle placement
4 Positions ----
Anterior-lateral > most convenient
Anterior-posterior
Anterior-left infrascapular
Anterior-right infrascapular
Anterior-lateral position ----
one – right of sternum below clavicle (2nd & 3rd ICS)
other – left 4th or 5th ICS mid axillary line
Reversing paddle markings “sternum & apex” does not affect defibrillation
Paddles placed along axis of heart.
AP- placement used in children with adult paddles.
Position of electrodes in pregnant patients :-
-One on right of sternum below the right clivcle
-Other left mid-axillary line avoiding breast
Paddle Size
- Adult – large paddles
10-13 cm diameter
- Pediatric – small paddles <1yr
Infant <10kgs – 4.5cm
Children >10kgs – 8cm
Steps of Defibrillation
Give 2min CPR before analysis.
Power on ADE
Attach electrode pad.
Check rhythm.
CLEAR.
Give shock(120-200J)
Resume CPR for 2mins before analysis cardiac rhythm again.
Consider giving – vasopressors / anti-arrhythmic‟s during subsequent
shock.
Defibrillation Success
Termination of VF into an organized rhythm or
asystole for atleast 5secs after shock.
Regardless of haemodynamic response
Start with –
150-200J Bisphasic truncated waveform
120J Rectilinear biphasic waveform
Monitoring
Monitor that patient stay in converted rhythm.
Keep patient well oxygenated.(SPO2 >98%)
Check serum K+ & Mg+ levels.
Maintain acid base balance.
Get 12 lead ECG after procedure.
Check for chest pain & access.
Get CPK & Troponin done.
Access patients skin.
Efficacy of Defibrillation
Trans – thoracic resistance is the major factor :
Which depends upon
Chest width & configuration
Ventilatory phase
Electrode skin interface
Electrode position
Electrode size
Force of electrode application
Energy level
Time between & no. of shocks
Factors to consider during defibrillation
1) Duration of VF
- the longer VF lasts, the harder it is to cure
- the quicker the batter
- shock early, shock often
- likelihood of resuscitation decrease by 7-10% with every
passing minute (Ann Emerg Med. 1993;22:1652–1658 )
2) Myocardial environment / condition
Hypoxia, acidosis, hypothermia, electrolyte imbalance,
drug toxicity – impede conversion.
DO NOT DELAY SHOCK trying to correct these problems.
3) Heart size / Body type
Pediatric requirement lower than adult
2J /kg initial shock
4J /kg repeat shock
Higher dose (upto 10J/kg)
Or adult maximum dose
Direct size / energy relationship in adults unknown
4) Use largest size paddles
- completely contact chest without paddles touching each
other
- In pediatric minimum of 3cm distance between pads.
NOTE :-
- Small paddles : concentrate current, burn heart.
- Large paddles : reduces current density
5) Previous counter shock
- repeated shocks lower resistance
- give one shock at a time & then continue CPR
- subesquent shock either equal or higher energy
6) Paddle size
( as discussed before)
7) Paddle placement
- In pacemaker / ICD
atleast 12cm from generator
90 degree to AICD electrode
avoid placing pads directly over
no delay in defibrillation
- for other as described before…….
8) Paddles – Skin interface
- only gel should be used (ECG gelly)
- cream, paste, saline pads etc.- not recommended
- gel decreases resistance to the flow of current
- never use alcohol
9) Paddle contact pressure
- firm pressure of 25 pounds
- in child <10kgs --- 3kg pressure
- in large children >10kgs --- 5kg pressure
- deflate lung, shortens the path of current
- do not lean on paddles : they slip
CARDIOVERSION
Cardioversion is the delivery of energy that is
synchronized to the large R waves or QRS complex.
It uses energy less than that used for defibrillation in
shockable SCA
Avoids delivering shock during repolarization period (T
wave on ECG)
Indications
- Atrial flutter
- Atrial fibrillation
- Re entry SVT
- Mono morphic VT
- Poly morphic VT
- Wide complex tachycardia of uncertain type
Contraindications
- Digitalis induced dysrythmias
- refractory to cardioversion
- may precipitate to more serious ventricular dysarythmias
- Junctional tachycardia or ectopics / multifocal atrial
tachycardia
- automatic foci not reverted by cardioversion
CARDIOVERSION
Cardioversion can be
Elective or Emergency
Elective Cardioversion
> Pre procedure conciderations
- History & physical examination (neurological)
- Concurrent illness
- Current medication including anticoagulation
- Fasting 6hrs
- Correct electrolyte imbalance
- Obtain 12 lead ECG
- Shaving at site if nessacery
- Digoxin to be withheld for 48hrs
- Continue other medications
- Transthoracic Echo or TEE
Cardioversion Anaesthetic Technique
Cardioversion is almost always performed under induction or sedation
(short-acting agent such as midazolam)
exceptions : if patient is hemodynamically unstable or if
cardiovascular collapse is imminent
- Amnesia / Sedation / GA – may be required
- Premedication – Midazolam 1 to 2 mg
- Pre oxygenation
- Airway maintained & supported
Cardioversion Anaesthetic Drugs
-- IV sedation – Propofol, Etomidate, Thiopentone,
Benzodiazepines
-- Etomidate
haemodynamically more stable
myoclonus 40%- interferes with ECG interpretation
-- Propofol
hypotension (boluses)
slow induction can attenuate this drop
Steps
- Check environment at procedure site
- Turn on defibrillator
- Anaesthetic technique as required
- Apply electrodes
- Press SYNC control
- Select applicationenergy level
- „3 clear‟ shout & look
- Deliver shock
Post Procedure Monitoring
- Record delivery energy & result
- Continuous ECG monitoring
- 12 lead ECG
- If successful response
check for peripheral pulses, BP, Airway patency & LOC
- Inspect skin under the padds
- If not successful, check & reassess
Complications of Cardioversion
- Systemic embolization
- Post shock cardiac arythmias
asystole
heart block
atrial / ventricular ectopics
ventricular tachyarrythmias
- Trasient ST & T wave changes
Cardioversion Defibrillation
Elective planned procedure Emergency life saving procedure
Synchronized shock Un-synchronized shock
Low energy shock High energy shock
There can be some delay No delay, immediate
Anti-coagulation needed No anti-coagulation needed
Less damage to myocardium More damage to myocardium
Used in most of the arrhythmias
except VT /VF
Used in VT / VF

Contenu connexe

Tendances

Care of patient on ventilator
Care of patient on ventilatorCare of patient on ventilator
Care of patient on ventilator
Nursing Path
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slides
maltiaziz24
 

Tendances (20)

Defibrilation
DefibrilationDefibrilation
Defibrilation
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Central venous pressure
Central venous pressureCentral venous pressure
Central venous pressure
 
DEFIBRILLATOR
DEFIBRILLATORDEFIBRILLATOR
DEFIBRILLATOR
 
Care of patient on ventilator
Care of patient on ventilatorCare of patient on ventilator
Care of patient on ventilator
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slides
 
Cardioversion
Cardioversion Cardioversion
Cardioversion
 
Cardiac monitoring & ECG
Cardiac monitoring & ECGCardiac monitoring & ECG
Cardiac monitoring & ECG
 
Pacemaker powerpoint presentation med surg
Pacemaker powerpoint presentation med surgPacemaker powerpoint presentation med surg
Pacemaker powerpoint presentation med surg
 
Defibrillation & Cardioversion
Defibrillation & CardioversionDefibrillation & Cardioversion
Defibrillation & Cardioversion
 
Defibrillator
Defibrillator Defibrillator
Defibrillator
 
Nursing Care of Ventilated Patient
Nursing Care of Ventilated PatientNursing Care of Ventilated Patient
Nursing Care of Ventilated Patient
 
Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemaker
 
Pulse oxymetry
Pulse oxymetryPulse oxymetry
Pulse oxymetry
 
Defibrillator power point presentation for medical students
Defibrillator power point presentation for medical studentsDefibrillator power point presentation for medical students
Defibrillator power point presentation for medical students
 
Mechanical Ventilator by AJ
Mechanical Ventilator by AJMechanical Ventilator by AJ
Mechanical Ventilator by AJ
 
Syringe Pump
Syringe PumpSyringe Pump
Syringe Pump
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
 
Mechanical ventilation
Mechanical ventilationMechanical ventilation
Mechanical ventilation
 
Central venous pressure monitoring
Central venous pressure monitoring Central venous pressure monitoring
Central venous pressure monitoring
 

En vedette

defibrillation and cardioversion
 defibrillation and cardioversion defibrillation and cardioversion
defibrillation and cardioversion
Mohammed Awwad
 
Market Research Report : External defibrillator market in india 2014 - Sample
Market Research Report : External defibrillator market in india 2014 - SampleMarket Research Report : External defibrillator market in india 2014 - Sample
Market Research Report : External defibrillator market in india 2014 - Sample
Netscribes, Inc.
 
Defibrillater
DefibrillaterDefibrillater
Defibrillater
mans4ani
 

En vedette (11)

defibrillation and cardioversion
 defibrillation and cardioversion defibrillation and cardioversion
defibrillation and cardioversion
 
Defibrillation
DefibrillationDefibrillation
Defibrillation
 
Nursing case-study-burns-philnursingstudent
Nursing case-study-burns-philnursingstudentNursing case-study-burns-philnursingstudent
Nursing case-study-burns-philnursingstudent
 
Nursing case-study-dhf-dengue-hemorrhagic-fever-philnursingstudent
Nursing case-study-dhf-dengue-hemorrhagic-fever-philnursingstudentNursing case-study-dhf-dengue-hemorrhagic-fever-philnursingstudent
Nursing case-study-dhf-dengue-hemorrhagic-fever-philnursingstudent
 
Cardiac arrest
Cardiac arrest Cardiac arrest
Cardiac arrest
 
Defibrillation presentation
Defibrillation presentationDefibrillation presentation
Defibrillation presentation
 
Global Pharmaceuticals &amp; Biotechnology Industry
Global Pharmaceuticals &amp; Biotechnology IndustryGlobal Pharmaceuticals &amp; Biotechnology Industry
Global Pharmaceuticals &amp; Biotechnology Industry
 
Market Research Report : External defibrillator market in india 2014 - Sample
Market Research Report : External defibrillator market in india 2014 - SampleMarket Research Report : External defibrillator market in india 2014 - Sample
Market Research Report : External defibrillator market in india 2014 - Sample
 
Defibrillater
DefibrillaterDefibrillater
Defibrillater
 
Defibrillation, cardioversion and pacing
Defibrillation, cardioversion and pacingDefibrillation, cardioversion and pacing
Defibrillation, cardioversion and pacing
 
Pre operative care
Pre operative carePre operative care
Pre operative care
 

Similaire à Defibrillators

Similaire à Defibrillators (20)

Defibrillators
DefibrillatorsDefibrillators
Defibrillators
 
Defibrillatorppt 131028115457-phpapp01
Defibrillatorppt 131028115457-phpapp01Defibrillatorppt 131028115457-phpapp01
Defibrillatorppt 131028115457-phpapp01
 
defibrillator.pptx
defibrillator.pptxdefibrillator.pptx
defibrillator.pptx
 
DEFIBRILLATION.pptx
DEFIBRILLATION.pptxDEFIBRILLATION.pptx
DEFIBRILLATION.pptx
 
Defibrillation and cardioversion
Defibrillation and cardioversionDefibrillation and cardioversion
Defibrillation and cardioversion
 
Defibrillator prakash
Defibrillator prakashDefibrillator prakash
Defibrillator prakash
 
Dc shock
Dc shock  Dc shock
Dc shock
 
D & C 2012
D & C 2012D & C 2012
D & C 2012
 
Medical Electronics - Hints for Slow Learner
Medical Electronics - Hints for Slow Learner Medical Electronics - Hints for Slow Learner
Medical Electronics - Hints for Slow Learner
 
Cardiovertor
CardiovertorCardiovertor
Cardiovertor
 
Defibrillator ppt.pptx
Defibrillator ppt.pptxDefibrillator ppt.pptx
Defibrillator ppt.pptx
 
Bio-Medical Therapeutic of in Pacemaker& Respiratory
Bio-Medical Therapeutic of in  Pacemaker& RespiratoryBio-Medical Therapeutic of in  Pacemaker& Respiratory
Bio-Medical Therapeutic of in Pacemaker& Respiratory
 
Electrical therapies in cpr
Electrical therapies in cprElectrical therapies in cpr
Electrical therapies in cpr
 
DEFIBRILLATOR.pptx
DEFIBRILLATOR.pptxDEFIBRILLATOR.pptx
DEFIBRILLATOR.pptx
 
Defibrilator
DefibrilatorDefibrilator
Defibrilator
 
Defibrillation & cardioversion by DJ
Defibrillation & cardioversion by DJDefibrillation & cardioversion by DJ
Defibrillation & cardioversion by DJ
 
Defibrillators
DefibrillatorsDefibrillators
Defibrillators
 
Advanced Cardiovascular Life Support (ACLS).pptx
Advanced Cardiovascular Life Support (ACLS).pptxAdvanced Cardiovascular Life Support (ACLS).pptx
Advanced Cardiovascular Life Support (ACLS).pptx
 
Electrical therapy in bls 2018
Electrical therapy  in bls 2018Electrical therapy  in bls 2018
Electrical therapy in bls 2018
 
2. defibrillators
2. defibrillators2. defibrillators
2. defibrillators
 

Plus de Imran Sheikh (8)

Double Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptDouble Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes ppt
 
Stages of labour &amp; labour analgesia
Stages of labour &amp; labour analgesiaStages of labour &amp; labour analgesia
Stages of labour &amp; labour analgesia
 
Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit's
 
ANATOMY OF TRACHEA & TRACHEOSTOMY
ANATOMY OF TRACHEA & TRACHEOSTOMYANATOMY OF TRACHEA & TRACHEOSTOMY
ANATOMY OF TRACHEA & TRACHEOSTOMY
 
Coronary circuln 07 02-2012
Coronary  circuln 07 02-2012Coronary  circuln 07 02-2012
Coronary circuln 07 02-2012
 
Hepatic physiology & liver function tests
Hepatic physiology & liver function testsHepatic physiology & liver function tests
Hepatic physiology & liver function tests
 
Cardiac reflex
Cardiac reflexCardiac reflex
Cardiac reflex
 
Respiratory changes during anesthesia and ippv
Respiratory changes during anesthesia and ippvRespiratory changes during anesthesia and ippv
Respiratory changes during anesthesia and ippv
 

Dernier

Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Dipal Arora
 
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Dipal Arora
 

Dernier (20)

Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 

Defibrillators

  • 2. Definition ™Defibrillator is a device that deliver a therapeutic dose of electrical energy (electric shock) to the affected heart (fibrillated heart or other shockable rhythm) to force the heart to produce more normal cardiac rhythm.
  • 3. First demonstrated on dogs in 1899 by Jean-Louis Prévost & Frederic Batelli, two physiologists from University of Geneva, Switzerland. The first use on a human was in 1947 by Claude Beck, professor of surgery at Case Western Reserve University. Transthoracic defibrillation was first used in humans using alternating current (AC) Bernard Lown and his coworkers introduced direct current (DC) defibrillators into clinical practice. History
  • 4. Defibrillator Does not re-start the heart. Stops all electrical activity. If heart is still viable its natural pace maker will take over. It is an emergency life saving procedure. Probability decreases with time. Mechanism Current depolarizes the myocardium. Induces asystole temporarily. Allows SA node to take over coz it‟s the first to repolarize.
  • 5. Indications - Pulse-less polymorphic VT/VF. - When unsure weather monomorphic or polymorphic VT in a un-stable patient.
  • 6. Non Shockable rhythm’s - Asystole - Pulse less electrical activity (PEA)
  • 7. Loss of synchronization in heart is called as fibrillation.  Types of fibrillation  Atrial fibrillation  Ventricular fibrillation Defibrillator……
  • 8. ™Defibrillation is the definitive treatment for the life threatening cardiac arrhythmias ventricular fibrillation and pulseless ventricular tachycardia ™ Ventricular fibrillation results from: -Coronary occlusion -Electrical shock -Abnormalities of body chemistry ™ This irregular contraction of the muscle fibers causes non effectively blood pumping and that results in a steep fall of cardiac output. Need for a Defibrillator
  • 9. ™Ventricular fibrillation can be converted into a more efficient rhythm by applying high energy shock to the heart. ™ This causes all muscle fibers to contract simultaneously, which may then respond to normal physiological pacemaking pulses Restoration of normal rhythm in fibrillating heart as achieved by direct current shock (arrow) across the chest wall. The horizontal line after the shock shows that the cardiograph was blocked or disconnected for its protection during the period of shock4 Ventricular fibrillation
  • 10. This recommendation regarding CPR prior to attempted defibrillation is supported by 2 clinical studies JAMA. 2003;289:1389 –1395 JAMA. 1999;281:1182–1188
  • 11.  Applying a brief(.25 to 1 sec) burst of 60 HZ ac at an intensity of around 6 A.  This application of an electrical shock to resynchronize the heart is sometimes called counter shock.  If the patient does not respond, the burst is repeated until defibrillation occurs. this method is known as ac defibrillation. AC DEFIBRILLATION
  • 12. • It is cannot be successfully used to correct atrial fibrillation. • Successive attempts to correct ventricular fibrillation are often required. • Attempts to correct atrial fibrillation by this method often result more serious ventricular fibrillation. Disadvantage of using ac Defibrillator
  • 13.  In this method a capacitor is charged to a high dc voltage and then rapidly discharged.  The amount of energy discharged by the capacitor may range between 2 to 400joules with peak value of current 20A.  A corrective shock of 750-800 volts is applied within a tenth of a second . DC Defibrillation
  • 14. CIRCUIT OF DC DEFIBRILLATOR
  • 15. • Energy storage capacitor is charged at relatively slow rate from AC line. • Energy stored in capacitor is then delivered at a relatively rapid rate to chest of the patient. • Simple arrangement involve the discharge of capacitor energy through the patient’s own resistance. PRINCIPLE OF DEFIBRILLATOR
  • 16. Cont….. The discharge resistance which the patient represents is roughly a ohmic resistance of 50 – 100 ohms for a typical electrode size of 80cm2. The particular wave form is called “Lown” wave form. The pulse width of this waveform is 10ms.
  • 17. 1. Manual external defibrillator 2. Manual internal defibrillator 3. Semi-Automated External Defibrillator 4. Automated external defibrillator (AED) 5. Implantable cardioverter-defibrillator (ICD) {automatic internal cardiac defibrillator (AICD)} 6. Wearable cardiac defibrillator Types of Defibrillators
  • 18. Manual external defibrillator Electrodes placed directly around the heart area of chest. Higher Voltage required than internal defibrillator. Classified as - Monophasic Biphasic
  • 19. Monophasic waveform Defibrillators • Deliver current of one polarity. • Current travels in one direction through the patients heart from one paddle to another. 2 types :- • The monophasic damped sinusoidal waveform (MDS) returns to zero gradually • Monophasic truncated exponential waveform (MTE) current is abruptly returned to baseline (truncated) to zero current flow
  • 20. MDS v/s MTE wave form
  • 21. Biphasic waveform Defibrillators Current travels towards the +ve paddle & then reverses back. Reversing of polarity, depolarizes all cells – called “burping” response. Classified into – Biphasic truncated exponential waveform (BTE) Rectilinear biphasic waveform (RLB) RLB is better than BTE.
  • 22. Biphasic truncated exponential waveform (BTE) v/s Rectilinear biphasic waveform (RLB) RBL BTE
  • 23. Advantages of Biphasic over Monophasic Less power – Less trauma – Less battery. Defibrillation more effective at low energy. Fewer burns. Less myocardial damage. 1st shock success rate in cardiac arrest due to shockable rhythm – Monophasic 60% Biphasic increases to 90%
  • 24. Manual internal defibrillator Just like normal defilbrillator. Electric charge is usually pumped in by internal paddles in close or direct contact with the heart. These devices are mainly used in operating theatres, where the chest can be opened or being treated by a surgeon
  • 25. Semi-Automated External Defibrillator Carries features of both full-manuated as well as automated units of defibrillator. Has an ECG display and manual override in it. Generally used by paramedics and emergency medicine technicians.
  • 26. Implantable cardioverter-defibrillator (ICD) {automatic internal cardiac defibrillator (AICD)} A implanted device that detacts and terminats life threatning episodes of VF / VT in high risk patients.
  • 28. • AED is a portable type of external defibrillator that automatically diagnoses the ventricular fibrillation in a patient. • Automatic refers to the ability to autonomously analyze the patient's condition. • AED is provided with self-adhesive electrodes instead of hand held paddles. AUTOMATIC EXTERNAL DEFIBRILLATOR
  • 29. Automated external defibrillator (AED) cont… Use does not require special medical training. Found in public places – offices, airport, shopping mall. The electrodes transmit information about the person's heart rhythm to a controller in the AED. The controller examines the electrical output from the heart and determine if the patient is in a shockable rhythm or not and whether shock is needed. The AED uses voice / visual prompts to tell user when to give the shock, and the electrodes deliver it.
  • 30. Paddle placement 4 Positions ---- Anterior-lateral > most convenient Anterior-posterior Anterior-left infrascapular Anterior-right infrascapular Anterior-lateral position ---- one – right of sternum below clavicle (2nd & 3rd ICS) other – left 4th or 5th ICS mid axillary line Reversing paddle markings “sternum & apex” does not affect defibrillation Paddles placed along axis of heart. AP- placement used in children with adult paddles.
  • 31. Position of electrodes in pregnant patients :- -One on right of sternum below the right clivcle -Other left mid-axillary line avoiding breast Paddle Size - Adult – large paddles 10-13 cm diameter - Pediatric – small paddles <1yr Infant <10kgs – 4.5cm Children >10kgs – 8cm
  • 32. Steps of Defibrillation Give 2min CPR before analysis. Power on ADE Attach electrode pad. Check rhythm. CLEAR. Give shock(120-200J) Resume CPR for 2mins before analysis cardiac rhythm again. Consider giving – vasopressors / anti-arrhythmic‟s during subsequent shock.
  • 33. Defibrillation Success Termination of VF into an organized rhythm or asystole for atleast 5secs after shock. Regardless of haemodynamic response Start with – 150-200J Bisphasic truncated waveform 120J Rectilinear biphasic waveform
  • 34. Monitoring Monitor that patient stay in converted rhythm. Keep patient well oxygenated.(SPO2 >98%) Check serum K+ & Mg+ levels. Maintain acid base balance. Get 12 lead ECG after procedure. Check for chest pain & access. Get CPK & Troponin done. Access patients skin.
  • 35. Efficacy of Defibrillation Trans – thoracic resistance is the major factor : Which depends upon Chest width & configuration Ventilatory phase Electrode skin interface Electrode position Electrode size Force of electrode application Energy level Time between & no. of shocks
  • 36. Factors to consider during defibrillation 1) Duration of VF - the longer VF lasts, the harder it is to cure - the quicker the batter - shock early, shock often - likelihood of resuscitation decrease by 7-10% with every passing minute (Ann Emerg Med. 1993;22:1652–1658 )
  • 37. 2) Myocardial environment / condition Hypoxia, acidosis, hypothermia, electrolyte imbalance, drug toxicity – impede conversion. DO NOT DELAY SHOCK trying to correct these problems.
  • 38. 3) Heart size / Body type Pediatric requirement lower than adult 2J /kg initial shock 4J /kg repeat shock Higher dose (upto 10J/kg) Or adult maximum dose Direct size / energy relationship in adults unknown
  • 39. 4) Use largest size paddles - completely contact chest without paddles touching each other - In pediatric minimum of 3cm distance between pads. NOTE :- - Small paddles : concentrate current, burn heart. - Large paddles : reduces current density
  • 40. 5) Previous counter shock - repeated shocks lower resistance - give one shock at a time & then continue CPR - subesquent shock either equal or higher energy 6) Paddle size ( as discussed before)
  • 41. 7) Paddle placement - In pacemaker / ICD atleast 12cm from generator 90 degree to AICD electrode avoid placing pads directly over no delay in defibrillation - for other as described before…….
  • 42. 8) Paddles – Skin interface - only gel should be used (ECG gelly) - cream, paste, saline pads etc.- not recommended - gel decreases resistance to the flow of current - never use alcohol 9) Paddle contact pressure - firm pressure of 25 pounds - in child <10kgs --- 3kg pressure - in large children >10kgs --- 5kg pressure - deflate lung, shortens the path of current - do not lean on paddles : they slip
  • 43. CARDIOVERSION Cardioversion is the delivery of energy that is synchronized to the large R waves or QRS complex. It uses energy less than that used for defibrillation in shockable SCA Avoids delivering shock during repolarization period (T wave on ECG)
  • 44. Indications - Atrial flutter - Atrial fibrillation - Re entry SVT - Mono morphic VT - Poly morphic VT - Wide complex tachycardia of uncertain type
  • 45. Contraindications - Digitalis induced dysrythmias - refractory to cardioversion - may precipitate to more serious ventricular dysarythmias - Junctional tachycardia or ectopics / multifocal atrial tachycardia - automatic foci not reverted by cardioversion
  • 48. Elective Cardioversion > Pre procedure conciderations - History & physical examination (neurological) - Concurrent illness - Current medication including anticoagulation - Fasting 6hrs - Correct electrolyte imbalance - Obtain 12 lead ECG - Shaving at site if nessacery - Digoxin to be withheld for 48hrs - Continue other medications - Transthoracic Echo or TEE
  • 49. Cardioversion Anaesthetic Technique Cardioversion is almost always performed under induction or sedation (short-acting agent such as midazolam) exceptions : if patient is hemodynamically unstable or if cardiovascular collapse is imminent - Amnesia / Sedation / GA – may be required - Premedication – Midazolam 1 to 2 mg - Pre oxygenation - Airway maintained & supported
  • 50. Cardioversion Anaesthetic Drugs -- IV sedation – Propofol, Etomidate, Thiopentone, Benzodiazepines -- Etomidate haemodynamically more stable myoclonus 40%- interferes with ECG interpretation -- Propofol hypotension (boluses) slow induction can attenuate this drop
  • 51. Steps - Check environment at procedure site - Turn on defibrillator - Anaesthetic technique as required - Apply electrodes - Press SYNC control - Select applicationenergy level - „3 clear‟ shout & look - Deliver shock
  • 52. Post Procedure Monitoring - Record delivery energy & result - Continuous ECG monitoring - 12 lead ECG - If successful response check for peripheral pulses, BP, Airway patency & LOC - Inspect skin under the padds - If not successful, check & reassess
  • 53. Complications of Cardioversion - Systemic embolization - Post shock cardiac arythmias asystole heart block atrial / ventricular ectopics ventricular tachyarrythmias - Trasient ST & T wave changes
  • 54. Cardioversion Defibrillation Elective planned procedure Emergency life saving procedure Synchronized shock Un-synchronized shock Low energy shock High energy shock There can be some delay No delay, immediate Anti-coagulation needed No anti-coagulation needed Less damage to myocardium More damage to myocardium Used in most of the arrhythmias except VT /VF Used in VT / VF