WFHSS world forum for hospital sterile supply
education group
Guideline No.04 June 2012
Reprocessing of Medical Devices
in/for
Healthcare Establishments
wfhss Guideline on Reprocessing Medical Devices
Status:June 2012 Page 2 of 15
TABLE OF CONTENTS
1 General requirements 3
2 Competencies 3
3 Risk analysis, evaluation, selection and definition of suitable reprocessing methods 4
4 Organisational measures, quality management (QM), validation, monitoring 4
5 Personnel qualification 5
6 Documentation 6
7 External reprocessing (outsourcing) 6
8 Authors 7
ANNEX 1: STANDARD OPERATING PROCEDURES (SOPS) 8
ANNEX 2: CLASSIFICATION OF MEDICAL DEVICES INTO RISK GROUPS 10
ANNEX 3: REQUIREMENTS FOR RUMEDS IN HEALTHCARE ESTABLISHMENTS 14
wfhss Guideline on Reprocessing Medical Devices
Status:June 2012 Page 3 of 15
WFHSS Guideline on Reprocessing Medical
Devices in / for Healthcare Establishments
This Guideline has been compiled pursuant to the Medical Devices Directive of the EU.
1 General requirements
Medical device (MD) reprocessing in or for healthcare establishments must be con-
ducted with validated processes that demonstrate that a process continually meets
the given specifications.
This does not apply for MDs belonging to the non-critical group, as defined by the
Robert Koch Institute (RKI) (see Annex 2.)1
Note: “Validation serves to furnish documented proof of the ongoing effectiveness of the reprocessing
process under the operating conditions prevailing at the installation site, using the medical devices
encountered in routine operation in their respective packaging and with the reference loads used (i.e. to
ensure clean, disinfected or sterile devices).”
The functions of a MD, as dictated by its intended purpose, or compliance with the
relevant safety requirements must not be adversely affected when the device is re-
processed.
Medical device reprocessing must be conducted in accordance with the state of the
art in science and technology.
Note: The state of the art in science and technology is dictated by pertinent harmonised standards and by
relevant national or international standards and guidelines, such as those published by specialist
associations.
2 Competencies
Overall responsibility for medical device reprocessing in prescribed form is borne by
the proprietor or authorities in charge of the respective establishment.
Before MDs are procured, those parties responsible for their reprocessing must be
aware of the attendant reprocessing conditions and the requisite equipment must be
available in the establishment (see wfhss guideline No 02: “Check List for Procure-
ment of Medical Devices pursuant to EN ISO 17664”).
The manufacturer of a reusable MD must supply instructions on how to reprocess the
1
The RKI scheme rather than the Spaulding scheme takes into account the consequences for reprocessing of assigning
MDs to risk groups.
wfhss Guideline on Reprocessing Medical Devices
Status:June 2012 Page 4 of 15
device, using (pre-) validated2
processes as per EN ISO 17664.
Competencies and working practices related to all steps of the reprocessing chain
must be set out in writing in the form of standard operating
3 Risk analysis, evaluation, selection and definition of
suitable reprocessing methods
To ensure that they are reprocessed in prescribed form, medical devices must be
classified into risk groups in accordance with Annex 2.
Based on this classification, it must be set out in writing
• whether,
• how often,
• with which process
• and under what conditions
medical devices are to be reprocessed and stored.
The following aspects, in particular, must be borne in mind:
• the manufacturer’s instructions, if relevant,
• the resistance profiles of the microorganisms and potentially infectious agents
(e.g. prions) expected on the MDs,
• expected intervals between the time the MDs are used and the time repro-
cessed, with correspondingly more challenging reprocessing demands as well
as
• special risks posed by the MD when using and/or reprocessing it.
4 Organisational measures, quality management (QM),
validation, monitoring
Reprocessing Units for Medical Devices (RUMEDs) entrusted with medical device re-
processing are divided into three categories on the basis of the highest risk group to
which the medical devices they reprocess are assigned (see Annex 3).
Competencies and responsibilities within the RUMED must be defined (e.g. in the
form of an organogram). RUMED managers must have corresponding deputy manag-
ers.
2
The term “prevalidation” is understood to mean tests conducted for the purpose of identifying a suitable
reprocessing procedure; these are carried out on behalf of the manufacturer and should preferably be
performed by an independent test body
wfhss Guideline on Reprocessing Medical Devices
Status:June 2012 Page 5 of 15
The structural conditions must permit high quality reprocessing in line with the state
of the art in science and technology.
The equipment, fittings and adjuncts needed for MD reprocessing must be available
and comply with the state of the art (see Annex 3).
For medical device reprocessing all precautions and measures must be taken to as-
sure a demonstrable and continually high quality of reprocessing, in line with the vali-
dation specifications.
Reprocessing of medical devices subject to ultra stringent reprocessing demands
(“Critical C” devices as per Annex 2) must be conducted in line with a QM system
pursuant to EN ISO 13485 (see Annex 3).
The quality of automated reprocessing (cleaning, disinfection and sterilisation) must
be assured by the following measures:
• installation qualification
• validation of reprocessing procedures
• periodic (at least annual) revalidation
• routine daily tests on each day the RUMED is in operation
• batch-related routine tests
• monitoring and testing of process parameters using measurement technology.
5 Personnel qualification
Reprocessing in general, or any reprocessing task, may only be conducted by persons
who by virtue of their training, knowledge and practical experiences are able to as-
sure proper execution of such tasks (see Annex 3):
• All persons entrusted with MD reprocessing must have successfully attended
Specialist Training Course 1 (series of training courses specifically designed for
sterilisation assistants), or be in possession of an equivalent qualification.
• Managers and deputy managers of Category II RUMEDs must additionally have
successfully attended Specialist Training Course 2
• Managers and deputy managers of Category III RUMEDs must additionally
have successfully attended Specialist Training Course 3
Knowledge and skills must be acquired by attendance at appropriate continuing pro-
fessional development seminars based on the state of the art in science and technol-
ogy.
wfhss Guideline on Reprocessing Medical Devices
Status:June 2012 Page 6 of 15
6 Documentation
When reprocessing MDs belonging to Semi-Critical B as well as to Critical B and C risk
groups, the measured values recorded for the process parameters and underlying the
decision-making process for MD release must be documented, while indicating the
batch name and name of the person responsible for release. These values must fur-
nish proof that the reprocessing procedure was carried out in compliance with the
standard operating procedures (SOPs), using the parameters set out in the validation
protocol. This batch documentation must comprise at least:
• Date
• Equipment serial number
• Programme
• Serial batch numbers
• Designation of reprocessed load or similar code
• Operator’s name
• Release by responsible person
• Printout of measured values in analog or digital form (actual values)
• If applicable, results of batch control (e.g. cleaning indictors, chemical indicators)
Documentation of reprocessing MDs belonging to the non-critical risk group is not re-
quired, while that of Semi-Critical A and Critical A risk groups must be adapted to the
intended use on the basis of risk assessment and the mode of reprocessing inferred
from that.
Category I RUMED documentation, as per Annex 3, must be archived in a well-
assured readable form for at least five years, and Category II and III RUMED docu-
mentation for at least 10 years. Electronic documentation, including release docu-
mentation, is permitted provided that readability and data safety are assured for the
specified period.
7 External reprocessing (outsourcing)
The term “external reprocessing” is used to describe a situation where healthcare es-
tablishments outsource reprocessing in general, or parts thereof, to an authorised
external service provider. The same requirements refer to insourcing.
Outsourcing of reprocessing to a third party must not in any way negatively impact
on the healthcare establishment’s ability to meet the terms of its own service con-
tract.
Reprocessing service providers or healthcare establishments that conduct MD repro-
wfhss Guideline on Reprocessing Medical Devices
Status:June 2012 Page 7 of 15
cessing for other healthcare institutions must use validated reprocessing procedures.
Commercial service providers conducting MD reprocessing for healthcare establish-
ments with Category II and III RUMEDs as per Annex 3 must have in place a certified
QM system as per ÖNORM EN ISO 13485.
The following aspects, in particular, must be contractually regulated between the ex-
ternal service providers and the person (s) responsible for the respective healthcare
establishment (Category II and III RUMEDs as per Annex 3):
• The conditions to be met to assure running operations (e.g. emergency measures as well
as regulations on coping with bottle necks, breakdowns, standby);
• Duties, responsibilities and competencies of contract awarder and contractor (interfaces);
• Ownership rights as regards the MDs;
• Condition and quality of MDs to be reprocessed;
• Liability for (transport) damage;
• Handover conditions;
• Feedback system;
• Quality assurance measures (if necessary, certification as per ÖNORM EN ISO 13485, see
above);
• Transport logistics, including protected transport of contaminated instruments;
• Maximum time interval before soiled MDs are reprocessed;
• Whether and how any pre-treatment measures are to be conducted for MDs already within
the respective establishment, while taking account of specific types of soils;
• (Special) release regulations;
• Loaned instruments;
• Validation obligations of contract awarder and contractor;
• Qualifications of managers and staff in both premises (as per Annex 3);
• Nature, frequency and scope of checks to be conducted by contract awarder;
• Nature and scope of documentation to be provided by the contract awarder (validation, re-
validation reports, if applicable batch release documentation).
• Compatibility with the quality management system of the hospital
• Compatibility with the documentation system of the hospital
8 Authors
A. Blacky, V. Buchrieser, T. Freundlinger, M. Gehrer, H. Getreuer, A. Gruber, M. Hell, W. Koller,
P. Lachner, T. Miorini, A. Percht, G. Palmisano, U. Prüfert-Freese, M. Suchomel, A. Steinhardt, B.
Weinmayr
The guideline has been proof read and authorized by the wfhss education group
wfhss Guideline on Reprocessing Medical Devices
Status:June 2012 Page 8 of 15
ANNEX 1: Standard operating procedures (SOPs)
The RUMED management must ensure that standard operating procedures are availalble and
clearly displayed for the persons entrusted with the respective tasks. The RUMED manage-
ment, or a person explicitly designated in writing to that effect by the latter, must ensure that
RUMED employees are also able to understand the content of such SOPs.
This applies in particular for the following:
1. All partial steps, such as:
• Measures conducted immediately after using an MD, e.g. pretreatment, dis-
mantling, collection and intermediate storage
• Transport
• Cleaning
• Rinsing
• Disinfection
• Drying
• Verification of steps conducted hitherto (cleaning results, etc.)
• Maintenance and care
• Functional testing
• Packaging
• Arrangement of supplies for sterilisation
• Sterilisation
• Labelling
• Release
• Transport of sterile supplies
• Storage of sterile supplies
• Measures conducted immediately before use (e.g. checking that sterile sup-
plies are undamaged)
2. Special processes
2.1 Reprocessing MDs that have not been used
These medical devices include:
- MDs which were supplied in an unsterile condition but must be
sterile when put to use,
- Sterilised MDs whose packaging was damaged or opened without the
MD having been used and which, as specified by the manufacturer’s in-
structions, can be reprocessed.
- MDs whose sterile storage period has expired before the expiry date
wfhss Guideline on Reprocessing Medical Devices
Status:June 2012 Page 9 of 15
and which, in terms of their composition, are amenable to reprocessing.
2.2 Reprocessing of MDs which may be used or reprocessed only for a limited number
of times
2.3 Reprocessing of textile materials
2.4 Patient-related specific forms of reprocessing.
3. Critical processes
3.1. Reprocessing of heat-sensitive MDs (e.g. endoscopes)
3.2. Reprocessing of MDs with poorly accessible sites (e.g. minimally invasive surgical
(MIS) instruments)
3.3. Reprocessing of MDs posing a risk of injury to users, personnel or patients (e.g.
faulty electrical currents, pointed/sharp implements)
3.4. Reprocessing of MDs subject to more stringent reprocessing requirements (e.g.
residues that are difficult to remove)
3.5. Reprocessing of MDs made of special materials or technical configuration (e.g.
kinks, liable to scratches, antimagnetic, cannot be immersed in liquid, alumini-
um)
3.6. Manual reprocessing steps
3.7. Reprocessing of MDs subject to special reprocessing requirements because of
specific infectious agents (e.g. prions)
wfhss Guideline on Reprocessing Medical Devices
Status: June 2012 Page 10 of 15
ANNEX 2: Classification of medical devices into risk groups
(modified as per RKI)
Classification of MDs into risk groups is done on the basis of previous and subsequent
use, transport and storage conditions, design and materials. If there is any doubt
about the risk group to which the MD belongs, assign it to the higher group.
Preference must be given to automated reprocessing in a washer-disinfector because
it assures better standardisation and occupational protection. Disinfection should
preferably be performed with a thermal process.
Classification Description
Medical
device
Example
Pretreat-
treat-
ment
a
Cleaning /
Disinfec-
tion
Spec.
label
Sterili-
sation
Critical process
steps, special re-
quirements
NON-CRITICAL Contact with
intact skin
ECGelectrode,
Kidney dish
X ( X
1
) Inspection:visual cleanli-
ness
SEMI-
CRITICAL Contact with mucosa or pathologically altered skin
A
No special repro-
cessing require-
ments
b
Specula ,
Oral mirror,
dental scoop,
endoscopes
without lumens
(X
2
) X (X
3
)
Preferablyautomated
cleaning (C)and disinfec-
tion (D),atleastdisinfec-
tion withsuitableagents/
process
B
Subjecttomore
stringentrepro-
cessing require-
ments
Flexible endo-
scopeswith
lumens
c
,
Hand and
angled
pieces
c,d
,,
X
4 X (X
3
) Automated C/D
CRITICAL Skin or mucosa penetration, contact with open wounds, drugs, blood or blood products
that must be administered in sterile condition
A
No special repro-
cessing require-
ments
Woundretrac-
tors,
Needleholders
(X
2
) X X
Automated C/Dfor
Cat.IIIRUMED,otherwise
preferablyautomated C/D
Steamsterilisation
B
Subjecttomore
stringentrepro-
cessing require-
ments
Heat--sensitive
MISinstru-
ments,
surgical drapes
X
4, e
X (X
5
) X
Forall partscominginto
directcontactwith tissue,
onlyautomated thermal
C/Din suitableWDand
steam sterilisation
C
Subjecttoultra
stringentrepro-
cessing require-
ments
Angioscopes
Epiduroscopes
X
4
X X
5
X
Fornon-thermalsterilisa-
tionprocesses,noevi-
dence ofprion inactivation
hasbeenfurnished to
date.This mustbe borne
in mindforMDs belonging
to this group
a
Pretreatment: this is understood to mean removal of course soils or a measure aimed at prevention of
incrustations, but need not be limited to these. Pretreatment is done at the site of MD use.
b
e.g. can cleanliness be checked through visual inspection
c
if it cannot be classified as “critical”;
d
preferably, automated reprocessing, if this is not possible, then steam sterilisation with suitable pro-
wfhss Guideline on Reprocessing Medical Devices
Status: June 2012 Page 11 of 15
cesses following manual reprocessing
e
does not apply to surgical drapes
(X) optional working step
X
1
steam sterilisation may be needed, bearing in mind specific requirements.
X
2
pretreatment immediately after use with non-protein fixing agents/process
X
3
for MDs that are not used in physiologically sterile body cavities, final disinfection suffices apart from
cases of intraoperative use.
X
4
these MDs (semi-critical B, critical B, C) must be treated immediately after use to pave the way for
subsequent cleaning and disinfection, so that final cleaning is conducted as quickly as possible us-
ing automated processes, while preventing corrosion or other form of damage to the instruments. All
factors of influence here must be defined and taken account of during validation..
X
5
A special label is used for identification, information and tracking (as required, information must be
displayed on the following: batch no., serial no., manufacturer’s name, type and number of repro-
cessing cycles., max. number of reprocessing cycles permitted, safety and warning instructions,
etc.).
wfhssGuidelineonReprocessingMedicalDevices
Status:June2012Page12of15
wfhss Guideline on Reprocessing Medical Devices
Status: June 2012 Page 13 of 15
wfhssGuidelineonReprocessingMedicalDevices
Status:June2012Page14of15
ANNEX3:RequirementsforRUMEDsinhealthcareestablishments
ReprocessingunitcategoryIIIIII
Riskgroupofmedicalde-
vice(MD)tobereprocessed
Non-critical,semi-criticalA,
criticalA(1)
Non-critical,semi-criticalA,B,
criticalA
Allgroups
Healthcareestablishments
e.g.(homesfortheelderly/nursinghomes,
outpatientdepartments,medicalpractices
(2)
e.g.healthcareestablishmentswithoutsur-
gicalunits
e.g.healthcareestablishmentswithsurgical
units
QualityassuranceAppropriatequalityassurancemeasures
Appropriatequalityassurancemeasures
(minimumrequirementsasper
RobertKochInstitute-RKI)
QMsystempreferablybasedon,forcriticalC
devicesbasedonENISO13485:2003
Structuralrequirements
-Dedicatedarea
-Preferablywithseparateunclean/
clean/sterilezones(temporalsepara-
tionpossible)
-Dedicatedreprocessingroom
-Separateunclean/clean/sterilezones(3)
-Dedicatedpremises
-Separateunclean/clean/sterilepremises(4)
Staffqualifications
ManagementandstaffatleastSpecialist
CourseI(5)
Management:atleastSpecialistCourseII
Staff:atleastSpecialistCourseI
Managementanddeputies:SpecialistCourse
III
Staff:atleastSpecialistCourseI
Technicalequipment
-Ifnecessarysmallsteamsterilizer
(ClassBrecommended)asperEN
13060
-If,necessaryultrasoniccleaningequip-
ment
-Ifnecessary,washer-disinfector(WD)
asperENISO15883
-Ifnecessary,heatsealingmachine
-Steamsterilizers(bigorsmall)asper
EN285orEN13060
-EndoscopeWDasperENISO15883
-Ifnecessary,rotarysealingmachine
-Ifnecessary,ultrasoniccleaningequip-
ment
-Steamsterilizers(big)asperEN285
-Ifnecessary,equipmentforspecialsterili-
zationprocesses(e.g.low-temperature
sterilizers)
-EndoscopeWDasperENISO15883
-Rotarysealingmachine
-Ultrasoniccleaningequipment
(1)Callforspecialconsideration;(2)withtheexceptionoflumenedendoscopes(Cat.II)andsurgicalprocedures(Cat.III),exceptformanagementofsmallwounds;(3)For
newendoscopyunits–centralisationandseparateunclean/cleanpremisesarerecommended;(4)Fornew,extendedorconvertedpremises,cleanareashouldmeetre-
quirementsofISOclass8asperISO14644-1,asfaraspossiblealsoforexistingestablishments;(5)deemedtohavebeenmetfordegree-levelnursingstafforformedi-
calpersonnel
wfhssGuidelineonReprocessingMedicalDevices
Status:June2012Page15of15
Conceptforcontinuingprofessionaldevelopmentcoursesfor"Reprocessingmedicaldevicesin/forhealthcareestablishments"
COURSESPECIALISTCOURSEISPECIALISTCOURSEIISPECIALISTCOURSEIII
TARGETGROUP
Semi-skilledstaffandassistants;
Dentalstaff**ordentalassistants
Semi-skilledstaffwithspecialduties(e.g.
abletoreleaseMDs)andgraduateswithre-
sponsibilityforasection
Managersanddeputymanagers
EXTENT
80hours:
theoryandpracticalexercises
80hours:
40hourstheory
16hourspracticalexercises
24hourspracticalwork
80hours:
40hourstheory
40hoursconcludingassignment
ADMISSIONREQUIRE-
MENTS
None
SuccessfulcompletionofSpecialistCourseI.
andatleast1year’sexperienceinaMDre-
processingunit
or
SuitablyrecognisedQualification(inaccord-
ancewithnationalrequirements)andwith
successfulexaminationoncontentofSpe-
cialistCourseI(withoutattendingcourse)
SuitablyrecognisedQualification(inaccord-
ancewithnationalrequirements)andsuc-
cessfulcompletionofSpecialistCourseII
REQUIREMENTFOR
ManagementofRUMEDCat.I**,
StaffofRUMDCat.I**,II,III
ManagersanddeputymanagersofRUMED
Cat.II
ResponsibilityforsectionofRUMEDCat.III.
ManagersanddeputymanagersofRUMED
Cat.III
DEADLINEFORCOM-
PLETION
Within2yearsoftakinguppositionWithin2yearsoftakinguppositionWithin2yearsoftakingupposition
EXAMINATIONExaminationofknowledgeExaminationandassignmentExaminationandfinalassignment
*ifnotcoveredinspecialistORcourse,**exceptfordegree-levelnursingstafforformedicalpersonnel
Source:AustrianSocietyforSterileSupply(ÖGSV),www.oegsv.com

Contenu connexe

Tendances

Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Swapnil Fernandes
 
CMC, post approval regulatory affairs, etc
CMC, post approval regulatory affairs, etcCMC, post approval regulatory affairs, etc
CMC, post approval regulatory affairs, etcJayeshRajput7
 
GTPs vs. cGMPs
GTPs vs. cGMPsGTPs vs. cGMPs
GTPs vs. cGMPswrtolbert
 
Cmc and post regulatory
Cmc and post regulatoryCmc and post regulatory
Cmc and post regulatoryDev Jain
 
Process validation-and-critical-regulatory-requirements-in-manufacturing-of-i...
Process validation-and-critical-regulatory-requirements-in-manufacturing-of-i...Process validation-and-critical-regulatory-requirements-in-manufacturing-of-i...
Process validation-and-critical-regulatory-requirements-in-manufacturing-of-i...Ahmed Hasham
 
Computerized systems-validation-csv-in-biopharmaceutical-industries
Computerized systems-validation-csv-in-biopharmaceutical-industriesComputerized systems-validation-csv-in-biopharmaceutical-industries
Computerized systems-validation-csv-in-biopharmaceutical-industriesAhmed Hasham
 
Computerized system validation (CSV) as a requirement for good manufacturing ...
Computerized system validation (CSV) as a requirement for good manufacturing ...Computerized system validation (CSV) as a requirement for good manufacturing ...
Computerized system validation (CSV) as a requirement for good manufacturing ...Ahmed Hasham
 
Process validation and critical regulatory requirements in manufacturing of i...
Process validation and critical regulatory requirements in manufacturing of i...Process validation and critical regulatory requirements in manufacturing of i...
Process validation and critical regulatory requirements in manufacturing of i...Ahmed Hasham
 
Pharmaceutical Equipment
Pharmaceutical EquipmentPharmaceutical Equipment
Pharmaceutical EquipmentGomtesh447
 
Scale up process and post marketing survilence
Scale up process and post marketing survilenceScale up process and post marketing survilence
Scale up process and post marketing survilenceKAVITAAGRE
 
CTD and eCTD Format
CTD and eCTD Format CTD and eCTD Format
CTD and eCTD Format NikhilThorane
 
Presentation on regulatory affairs 30032013
Presentation on regulatory affairs 30032013Presentation on regulatory affairs 30032013
Presentation on regulatory affairs 30032013Nishodh Saxena Ph. D.
 
Documentation In Pharmaceutical Industry(Master Formula Record,DMF,Distributi...
Documentation In Pharmaceutical Industry(Master Formula Record,DMF,Distributi...Documentation In Pharmaceutical Industry(Master Formula Record,DMF,Distributi...
Documentation In Pharmaceutical Industry(Master Formula Record,DMF,Distributi...RUSHIKESHSHINDE80
 
Dmf & distribution records seminor
Dmf & distribution records seminorDmf & distribution records seminor
Dmf & distribution records seminorOmkar Sasane
 

Tendances (20)

Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
 
DMF -Drug Master File
DMF -Drug Master File DMF -Drug Master File
DMF -Drug Master File
 
Docmentation complete
Docmentation completeDocmentation complete
Docmentation complete
 
CMC, post approval regulatory affairs, etc
CMC, post approval regulatory affairs, etcCMC, post approval regulatory affairs, etc
CMC, post approval regulatory affairs, etc
 
GTPs vs. cGMPs
GTPs vs. cGMPsGTPs vs. cGMPs
GTPs vs. cGMPs
 
Phụ lục 10. Thủ tục sơ kiểm các sản phẩm dược
Phụ lục 10. Thủ tục sơ kiểm các sản phẩm dượcPhụ lục 10. Thủ tục sơ kiểm các sản phẩm dược
Phụ lục 10. Thủ tục sơ kiểm các sản phẩm dược
 
clinical trials
clinical trialsclinical trials
clinical trials
 
CMC
CMCCMC
CMC
 
Cmc and post regulatory
Cmc and post regulatoryCmc and post regulatory
Cmc and post regulatory
 
Process validation-and-critical-regulatory-requirements-in-manufacturing-of-i...
Process validation-and-critical-regulatory-requirements-in-manufacturing-of-i...Process validation-and-critical-regulatory-requirements-in-manufacturing-of-i...
Process validation-and-critical-regulatory-requirements-in-manufacturing-of-i...
 
Computerized systems-validation-csv-in-biopharmaceutical-industries
Computerized systems-validation-csv-in-biopharmaceutical-industriesComputerized systems-validation-csv-in-biopharmaceutical-industries
Computerized systems-validation-csv-in-biopharmaceutical-industries
 
Medical devices
Medical devicesMedical devices
Medical devices
 
Computerized system validation (CSV) as a requirement for good manufacturing ...
Computerized system validation (CSV) as a requirement for good manufacturing ...Computerized system validation (CSV) as a requirement for good manufacturing ...
Computerized system validation (CSV) as a requirement for good manufacturing ...
 
Process validation and critical regulatory requirements in manufacturing of i...
Process validation and critical regulatory requirements in manufacturing of i...Process validation and critical regulatory requirements in manufacturing of i...
Process validation and critical regulatory requirements in manufacturing of i...
 
Pharmaceutical Equipment
Pharmaceutical EquipmentPharmaceutical Equipment
Pharmaceutical Equipment
 
Scale up process and post marketing survilence
Scale up process and post marketing survilenceScale up process and post marketing survilence
Scale up process and post marketing survilence
 
CTD and eCTD Format
CTD and eCTD Format CTD and eCTD Format
CTD and eCTD Format
 
Presentation on regulatory affairs 30032013
Presentation on regulatory affairs 30032013Presentation on regulatory affairs 30032013
Presentation on regulatory affairs 30032013
 
Documentation In Pharmaceutical Industry(Master Formula Record,DMF,Distributi...
Documentation In Pharmaceutical Industry(Master Formula Record,DMF,Distributi...Documentation In Pharmaceutical Industry(Master Formula Record,DMF,Distributi...
Documentation In Pharmaceutical Industry(Master Formula Record,DMF,Distributi...
 
Dmf & distribution records seminor
Dmf & distribution records seminorDmf & distribution records seminor
Dmf & distribution records seminor
 

En vedette

Innovaciones en investigacion y docencia enfermera
Innovaciones en investigacion y docencia enfermeraInnovaciones en investigacion y docencia enfermera
Innovaciones en investigacion y docencia enfermeraAzucena Santillan García
 
Enfermeria Basada en la Evidencia: ¿Realidad o ficcion?
Enfermeria Basada en la Evidencia: ¿Realidad o ficcion?Enfermeria Basada en la Evidencia: ¿Realidad o ficcion?
Enfermeria Basada en la Evidencia: ¿Realidad o ficcion?Azucena Santillan García
 
Estrategia socialmedia para la divulgacion de la enfermeria basada en evidenc...
Estrategia socialmedia para la divulgacion de la enfermeria basada en evidenc...Estrategia socialmedia para la divulgacion de la enfermeria basada en evidenc...
Estrategia socialmedia para la divulgacion de la enfermeria basada en evidenc...Azucena Santillan García
 
Manual basico de supervivencia para enfermeras innovadoras
Manual basico de supervivencia para enfermeras innovadorasManual basico de supervivencia para enfermeras innovadoras
Manual basico de supervivencia para enfermeras innovadorasAzucena Santillan García
 
Como elaborar una estrategia socialmedia para favorecer la visibilidad de los...
Como elaborar una estrategia socialmedia para favorecer la visibilidad de los...Como elaborar una estrategia socialmedia para favorecer la visibilidad de los...
Como elaborar una estrategia socialmedia para favorecer la visibilidad de los...Azucena Santillan García
 
Nuevas formas de hacer investigacion y docencia enfermera
Nuevas formas de hacer investigacion y docencia enfermeraNuevas formas de hacer investigacion y docencia enfermera
Nuevas formas de hacer investigacion y docencia enfermeraAzucena Santillan García
 
Enfermeria y Redes Sociales: Internet como escenario
Enfermeria y Redes Sociales: Internet como escenarioEnfermeria y Redes Sociales: Internet como escenario
Enfermeria y Redes Sociales: Internet como escenarioAzucena Santillan García
 
Como escribir un blog de exito y no morir en el intento
Como escribir un blog de exito y no morir en el intento Como escribir un blog de exito y no morir en el intento
Como escribir un blog de exito y no morir en el intento Azucena Santillan García
 
TIC en investigacion de cuidados y salud comunitaria
TIC en investigacion de cuidados y salud comunitariaTIC en investigacion de cuidados y salud comunitaria
TIC en investigacion de cuidados y salud comunitariaAzucena Santillan García
 
Practica basada en la evidencia
Practica basada en la evidenciaPractica basada en la evidencia
Practica basada en la evidencialuisasalamanca
 
VII encuentro metis avila 2014. LAS NUEVAS TECNOLOGÍAS DE LA INFORMACIÓN AL S...
VII encuentro metis avila 2014. LAS NUEVAS TECNOLOGÍAS DE LA INFORMACIÓN AL S...VII encuentro metis avila 2014. LAS NUEVAS TECNOLOGÍAS DE LA INFORMACIÓN AL S...
VII encuentro metis avila 2014. LAS NUEVAS TECNOLOGÍAS DE LA INFORMACIÓN AL S...Carmen Villar Bustos
 
Azucena Santillán García
Azucena Santillán GarcíaAzucena Santillán García
Azucena Santillán Garcíainvestenisciii
 
Conferencia Inaugural: Carlos Segovia Pérez
Conferencia Inaugural: Carlos Segovia PérezConferencia Inaugural: Carlos Segovia Pérez
Conferencia Inaugural: Carlos Segovia Pérezinvestenisciii
 
Gestion clinica de los cuidados a través de Gacela Care: experiencias de impl...
Gestion clinica de los cuidados a través de Gacela Care: experiencias de impl...Gestion clinica de los cuidados a través de Gacela Care: experiencias de impl...
Gestion clinica de los cuidados a través de Gacela Care: experiencias de impl...Azucena Santillan García
 
Como generar evidencia practica y no morir en el intento
Como generar evidencia practica y no morir en el intentoComo generar evidencia practica y no morir en el intento
Como generar evidencia practica y no morir en el intentoAzucena Santillan García
 
Programa del VIII Premio de Investigación Enfermería Raquel Recuero
Programa del VIII Premio de Investigación Enfermería Raquel RecueroPrograma del VIII Premio de Investigación Enfermería Raquel Recuero
Programa del VIII Premio de Investigación Enfermería Raquel RecueroCarmen Villar Bustos
 
Enfermería basada en la evidencia
Enfermería basada en la evidenciaEnfermería basada en la evidencia
Enfermería basada en la evidenciaCamolaNavarra
 
Contexto historico de las practicas basadas en evidencias
Contexto historico de las practicas basadas en evidenciasContexto historico de las practicas basadas en evidencias
Contexto historico de las practicas basadas en evidenciasAzucena Santillan García
 

En vedette (20)

Innovaciones en investigacion y docencia enfermera
Innovaciones en investigacion y docencia enfermeraInnovaciones en investigacion y docencia enfermera
Innovaciones en investigacion y docencia enfermera
 
Enfermeria Basada en la Evidencia: ¿Realidad o ficcion?
Enfermeria Basada en la Evidencia: ¿Realidad o ficcion?Enfermeria Basada en la Evidencia: ¿Realidad o ficcion?
Enfermeria Basada en la Evidencia: ¿Realidad o ficcion?
 
Estrategia socialmedia para la divulgacion de la enfermeria basada en evidenc...
Estrategia socialmedia para la divulgacion de la enfermeria basada en evidenc...Estrategia socialmedia para la divulgacion de la enfermeria basada en evidenc...
Estrategia socialmedia para la divulgacion de la enfermeria basada en evidenc...
 
Manual basico de supervivencia para enfermeras innovadoras
Manual basico de supervivencia para enfermeras innovadorasManual basico de supervivencia para enfermeras innovadoras
Manual basico de supervivencia para enfermeras innovadoras
 
Como elaborar una estrategia socialmedia para favorecer la visibilidad de los...
Como elaborar una estrategia socialmedia para favorecer la visibilidad de los...Como elaborar una estrategia socialmedia para favorecer la visibilidad de los...
Como elaborar una estrategia socialmedia para favorecer la visibilidad de los...
 
Nuevas formas de hacer investigacion y docencia enfermera
Nuevas formas de hacer investigacion y docencia enfermeraNuevas formas de hacer investigacion y docencia enfermera
Nuevas formas de hacer investigacion y docencia enfermera
 
Enfermeria y Redes Sociales: Internet como escenario
Enfermeria y Redes Sociales: Internet como escenarioEnfermeria y Redes Sociales: Internet como escenario
Enfermeria y Redes Sociales: Internet como escenario
 
Como escribir un blog de exito y no morir en el intento
Como escribir un blog de exito y no morir en el intento Como escribir un blog de exito y no morir en el intento
Como escribir un blog de exito y no morir en el intento
 
TIC en investigacion de cuidados y salud comunitaria
TIC en investigacion de cuidados y salud comunitariaTIC en investigacion de cuidados y salud comunitaria
TIC en investigacion de cuidados y salud comunitaria
 
Enfermeria 2.0
Enfermeria 2.0Enfermeria 2.0
Enfermeria 2.0
 
Practica basada en la evidencia
Practica basada en la evidenciaPractica basada en la evidencia
Practica basada en la evidencia
 
VII encuentro metis avila 2014. LAS NUEVAS TECNOLOGÍAS DE LA INFORMACIÓN AL S...
VII encuentro metis avila 2014. LAS NUEVAS TECNOLOGÍAS DE LA INFORMACIÓN AL S...VII encuentro metis avila 2014. LAS NUEVAS TECNOLOGÍAS DE LA INFORMACIÓN AL S...
VII encuentro metis avila 2014. LAS NUEVAS TECNOLOGÍAS DE LA INFORMACIÓN AL S...
 
Azucena Santillán García
Azucena Santillán GarcíaAzucena Santillán García
Azucena Santillán García
 
Conferencia Inaugural: Carlos Segovia Pérez
Conferencia Inaugural: Carlos Segovia PérezConferencia Inaugural: Carlos Segovia Pérez
Conferencia Inaugural: Carlos Segovia Pérez
 
Gestion clinica de los cuidados a través de Gacela Care: experiencias de impl...
Gestion clinica de los cuidados a través de Gacela Care: experiencias de impl...Gestion clinica de los cuidados a través de Gacela Care: experiencias de impl...
Gestion clinica de los cuidados a través de Gacela Care: experiencias de impl...
 
Como generar evidencia practica y no morir en el intento
Como generar evidencia practica y no morir en el intentoComo generar evidencia practica y no morir en el intento
Como generar evidencia practica y no morir en el intento
 
Programa del VIII Premio de Investigación Enfermería Raquel Recuero
Programa del VIII Premio de Investigación Enfermería Raquel RecueroPrograma del VIII Premio de Investigación Enfermería Raquel Recuero
Programa del VIII Premio de Investigación Enfermería Raquel Recuero
 
Enfermería basada en la evidencia
Enfermería basada en la evidenciaEnfermería basada en la evidencia
Enfermería basada en la evidencia
 
Contexto historico de las practicas basadas en evidencias
Contexto historico de las practicas basadas en evidenciasContexto historico de las practicas basadas en evidencias
Contexto historico de las practicas basadas en evidencias
 
APPs en salud: nuevos retos para el cuidado
APPs en salud: nuevos retos para el cuidadoAPPs en salud: nuevos retos para el cuidado
APPs en salud: nuevos retos para el cuidado
 

Similaire à WFHSS Guideline on Reprocessing Medical Devices in / for Healthcare Establishments

Final_Schedule M2.pdf
Final_Schedule M2.pdfFinal_Schedule M2.pdf
Final_Schedule M2.pdfZeelshah2258
 
Clinical Evaluation Report.pdf
Clinical Evaluation Report.pdfClinical Evaluation Report.pdf
Clinical Evaluation Report.pdfJonesmiller3
 
Medical Device Registration in India_ A Comprehensive Guide.pdf
Medical Device Registration in India_ A Comprehensive Guide.pdfMedical Device Registration in India_ A Comprehensive Guide.pdf
Medical Device Registration in India_ A Comprehensive Guide.pdfPranshuCorpseed
 
Post-Market Clinical Follow Up Studies Under EU MDR and IVDR
Post-Market Clinical Follow Up Studies Under EU MDR and IVDRPost-Market Clinical Follow Up Studies Under EU MDR and IVDR
Post-Market Clinical Follow Up Studies Under EU MDR and IVDREMMAIntl
 
Presentation: Conformity Assessment Evidence
Presentation: Conformity Assessment EvidencePresentation: Conformity Assessment Evidence
Presentation: Conformity Assessment EvidenceTGA Australia
 
Clinical evaluation report cer in a more stringent regulatory- Pepgra Healthcare
Clinical evaluation report cer in a more stringent regulatory- Pepgra HealthcareClinical evaluation report cer in a more stringent regulatory- Pepgra Healthcare
Clinical evaluation report cer in a more stringent regulatory- Pepgra HealthcarePEPGRA Healthcare
 
Clinical Evaluation Report
Clinical Evaluation ReportClinical Evaluation Report
Clinical Evaluation ReportAdamsBrown4
 
Clinical Evaluation Report.pptx
Clinical Evaluation Report.pptxClinical Evaluation Report.pptx
Clinical Evaluation Report.pptxAllensmith572606
 
2010 Medical Equipment Management Plan Medical Equipment Management Plan
2010 Medical Equipment Management Plan Medical Equipment Management Plan2010 Medical Equipment Management Plan Medical Equipment Management Plan
2010 Medical Equipment Management Plan Medical Equipment Management PlanNat Rice
 
WHO guideline on Validation .pdf
WHO guideline on Validation .pdfWHO guideline on Validation .pdf
WHO guideline on Validation .pdfHassanHani5
 
American radiology accreditation reqmts 2013
American radiology accreditation reqmts 2013American radiology accreditation reqmts 2013
American radiology accreditation reqmts 2013Bhanu Sharadha
 
Clinical Evaluation Report for Medical Devices
Clinical Evaluation Report for Medical DevicesClinical Evaluation Report for Medical Devices
Clinical Evaluation Report for Medical DevicesI 3 Consulting
 
Malaysia Medical Devices Regulations
Malaysia Medical Devices RegulationsMalaysia Medical Devices Regulations
Malaysia Medical Devices RegulationsMourad Kholti
 
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...Brajesh Kumar
 
TGA Presentation: Potential reforms for the regulation of system and procedur...
TGA Presentation: Potential reforms for the regulation of system and procedur...TGA Presentation: Potential reforms for the regulation of system and procedur...
TGA Presentation: Potential reforms for the regulation of system and procedur...TGA Australia
 
Understanding Post-market Surveillance under EU MDR: Being Proactive, not Rea...
Understanding Post-market Surveillance under EU MDR: Being Proactive, not Rea...Understanding Post-market Surveillance under EU MDR: Being Proactive, not Rea...
Understanding Post-market Surveillance under EU MDR: Being Proactive, not Rea...Greenlight Guru
 
Future of EU In Vitro Diagnostics Regulation
Future of EU In Vitro Diagnostics RegulationFuture of EU In Vitro Diagnostics Regulation
Future of EU In Vitro Diagnostics RegulationErik Vollebregt
 
clinical-evaluation-report (CER)
clinical-evaluation-report (CER)clinical-evaluation-report (CER)
clinical-evaluation-report (CER)PEPGRA Healthcare
 

Similaire à WFHSS Guideline on Reprocessing Medical Devices in / for Healthcare Establishments (20)

Final_Schedule M2.pdf
Final_Schedule M2.pdfFinal_Schedule M2.pdf
Final_Schedule M2.pdf
 
Clinical Evaluation Report.pdf
Clinical Evaluation Report.pdfClinical Evaluation Report.pdf
Clinical Evaluation Report.pdf
 
Medical Device Registration in India_ A Comprehensive Guide.pdf
Medical Device Registration in India_ A Comprehensive Guide.pdfMedical Device Registration in India_ A Comprehensive Guide.pdf
Medical Device Registration in India_ A Comprehensive Guide.pdf
 
Post-Market Clinical Follow Up Studies Under EU MDR and IVDR
Post-Market Clinical Follow Up Studies Under EU MDR and IVDRPost-Market Clinical Follow Up Studies Under EU MDR and IVDR
Post-Market Clinical Follow Up Studies Under EU MDR and IVDR
 
Presentation: Conformity Assessment Evidence
Presentation: Conformity Assessment EvidencePresentation: Conformity Assessment Evidence
Presentation: Conformity Assessment Evidence
 
Clinical evaluation report cer in a more stringent regulatory- Pepgra Healthcare
Clinical evaluation report cer in a more stringent regulatory- Pepgra HealthcareClinical evaluation report cer in a more stringent regulatory- Pepgra Healthcare
Clinical evaluation report cer in a more stringent regulatory- Pepgra Healthcare
 
Euipment q a
Euipment q aEuipment q a
Euipment q a
 
Clinical Evaluation Report
Clinical Evaluation ReportClinical Evaluation Report
Clinical Evaluation Report
 
Clinical Evaluation Report.pptx
Clinical Evaluation Report.pptxClinical Evaluation Report.pptx
Clinical Evaluation Report.pptx
 
Advamed MDR IVDR update
Advamed MDR IVDR updateAdvamed MDR IVDR update
Advamed MDR IVDR update
 
2010 Medical Equipment Management Plan Medical Equipment Management Plan
2010 Medical Equipment Management Plan Medical Equipment Management Plan2010 Medical Equipment Management Plan Medical Equipment Management Plan
2010 Medical Equipment Management Plan Medical Equipment Management Plan
 
WHO guideline on Validation .pdf
WHO guideline on Validation .pdfWHO guideline on Validation .pdf
WHO guideline on Validation .pdf
 
American radiology accreditation reqmts 2013
American radiology accreditation reqmts 2013American radiology accreditation reqmts 2013
American radiology accreditation reqmts 2013
 
Clinical Evaluation Report for Medical Devices
Clinical Evaluation Report for Medical DevicesClinical Evaluation Report for Medical Devices
Clinical Evaluation Report for Medical Devices
 
Malaysia Medical Devices Regulations
Malaysia Medical Devices RegulationsMalaysia Medical Devices Regulations
Malaysia Medical Devices Regulations
 
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
 
TGA Presentation: Potential reforms for the regulation of system and procedur...
TGA Presentation: Potential reforms for the regulation of system and procedur...TGA Presentation: Potential reforms for the regulation of system and procedur...
TGA Presentation: Potential reforms for the regulation of system and procedur...
 
Understanding Post-market Surveillance under EU MDR: Being Proactive, not Rea...
Understanding Post-market Surveillance under EU MDR: Being Proactive, not Rea...Understanding Post-market Surveillance under EU MDR: Being Proactive, not Rea...
Understanding Post-market Surveillance under EU MDR: Being Proactive, not Rea...
 
Future of EU In Vitro Diagnostics Regulation
Future of EU In Vitro Diagnostics RegulationFuture of EU In Vitro Diagnostics Regulation
Future of EU In Vitro Diagnostics Regulation
 
clinical-evaluation-report (CER)
clinical-evaluation-report (CER)clinical-evaluation-report (CER)
clinical-evaluation-report (CER)
 

Plus de Carmen Villar Bustos

Esterilización versus desinfección de alto nivel
Esterilización versus desinfección de alto nivelEsterilización versus desinfección de alto nivel
Esterilización versus desinfección de alto nivelCarmen Villar Bustos
 
Libro Resúmenes Esterilización 2011
Libro Resúmenes Esterilización 2011Libro Resúmenes Esterilización 2011
Libro Resúmenes Esterilización 2011Carmen Villar Bustos
 
Protocolo de vigilancia y control de microorganismos multirresistentes
Protocolo de vigilancia y control de microorganismos multirresistentes Protocolo de vigilancia y control de microorganismos multirresistentes
Protocolo de vigilancia y control de microorganismos multirresistentes Carmen Villar Bustos
 
Dificultades para la Investigación en Enfermería. Artículo de Cristina Quesada
Dificultades para la Investigación en Enfermería. Artículo de Cristina QuesadaDificultades para la Investigación en Enfermería. Artículo de Cristina Quesada
Dificultades para la Investigación en Enfermería. Artículo de Cristina QuesadaCarmen Villar Bustos
 
Infection control-core-competencies
Infection control-core-competenciesInfection control-core-competencies
Infection control-core-competenciesCarmen Villar Bustos
 
Control de Calidad en Esterilización en los centros de Castilla y León
Control de Calidad en Esterilización en  los centros de Castilla y LeónControl de Calidad en Esterilización en  los centros de Castilla y León
Control de Calidad en Esterilización en los centros de Castilla y LeónCarmen Villar Bustos
 

Plus de Carmen Villar Bustos (12)

Manifiesto #enfermería visible
Manifiesto #enfermería visibleManifiesto #enfermería visible
Manifiesto #enfermería visible
 
Esterilización versus desinfección de alto nivel
Esterilización versus desinfección de alto nivelEsterilización versus desinfección de alto nivel
Esterilización versus desinfección de alto nivel
 
Libro Resúmenes Esterilización 2011
Libro Resúmenes Esterilización 2011Libro Resúmenes Esterilización 2011
Libro Resúmenes Esterilización 2011
 
Protocolo de vigilancia y control de microorganismos multirresistentes
Protocolo de vigilancia y control de microorganismos multirresistentes Protocolo de vigilancia y control de microorganismos multirresistentes
Protocolo de vigilancia y control de microorganismos multirresistentes
 
Dificultades para la Investigación en Enfermería. Artículo de Cristina Quesada
Dificultades para la Investigación en Enfermería. Artículo de Cristina QuesadaDificultades para la Investigación en Enfermería. Artículo de Cristina Quesada
Dificultades para la Investigación en Enfermería. Artículo de Cristina Quesada
 
Infection control-core-competencies
Infection control-core-competenciesInfection control-core-competencies
Infection control-core-competencies
 
Control de Calidad en Esterilización en los centros de Castilla y León
Control de Calidad en Esterilización en  los centros de Castilla y LeónControl de Calidad en Esterilización en  los centros de Castilla y León
Control de Calidad en Esterilización en los centros de Castilla y León
 
Ipyou spanish hvcn-sacyl
Ipyou spanish hvcn-sacylIpyou spanish hvcn-sacyl
Ipyou spanish hvcn-sacyl
 
Enfermería quirúrgica 2011 AMFE
Enfermería quirúrgica 2011 AMFEEnfermería quirúrgica 2011 AMFE
Enfermería quirúrgica 2011 AMFE
 
Póster Trazabilidad
Póster TrazabilidadPóster Trazabilidad
Póster Trazabilidad
 
Investigación enfermería
Investigación enfermeríaInvestigación enfermería
Investigación enfermería
 
E saludcyl 2012_Avila
E saludcyl 2012_AvilaE saludcyl 2012_Avila
E saludcyl 2012_Avila
 

Dernier

call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptxDr.Nusrat Tariq
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingCall Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingNehru place Escorts
 
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service LucknowVIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 

Dernier (20)

call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptx
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingCall Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
 
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service LucknowVIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 

WFHSS Guideline on Reprocessing Medical Devices in / for Healthcare Establishments

  • 1. WFHSS world forum for hospital sterile supply education group Guideline No.04 June 2012 Reprocessing of Medical Devices in/for Healthcare Establishments
  • 2. wfhss Guideline on Reprocessing Medical Devices Status:June 2012 Page 2 of 15 TABLE OF CONTENTS 1 General requirements 3 2 Competencies 3 3 Risk analysis, evaluation, selection and definition of suitable reprocessing methods 4 4 Organisational measures, quality management (QM), validation, monitoring 4 5 Personnel qualification 5 6 Documentation 6 7 External reprocessing (outsourcing) 6 8 Authors 7 ANNEX 1: STANDARD OPERATING PROCEDURES (SOPS) 8 ANNEX 2: CLASSIFICATION OF MEDICAL DEVICES INTO RISK GROUPS 10 ANNEX 3: REQUIREMENTS FOR RUMEDS IN HEALTHCARE ESTABLISHMENTS 14
  • 3. wfhss Guideline on Reprocessing Medical Devices Status:June 2012 Page 3 of 15 WFHSS Guideline on Reprocessing Medical Devices in / for Healthcare Establishments This Guideline has been compiled pursuant to the Medical Devices Directive of the EU. 1 General requirements Medical device (MD) reprocessing in or for healthcare establishments must be con- ducted with validated processes that demonstrate that a process continually meets the given specifications. This does not apply for MDs belonging to the non-critical group, as defined by the Robert Koch Institute (RKI) (see Annex 2.)1 Note: “Validation serves to furnish documented proof of the ongoing effectiveness of the reprocessing process under the operating conditions prevailing at the installation site, using the medical devices encountered in routine operation in their respective packaging and with the reference loads used (i.e. to ensure clean, disinfected or sterile devices).” The functions of a MD, as dictated by its intended purpose, or compliance with the relevant safety requirements must not be adversely affected when the device is re- processed. Medical device reprocessing must be conducted in accordance with the state of the art in science and technology. Note: The state of the art in science and technology is dictated by pertinent harmonised standards and by relevant national or international standards and guidelines, such as those published by specialist associations. 2 Competencies Overall responsibility for medical device reprocessing in prescribed form is borne by the proprietor or authorities in charge of the respective establishment. Before MDs are procured, those parties responsible for their reprocessing must be aware of the attendant reprocessing conditions and the requisite equipment must be available in the establishment (see wfhss guideline No 02: “Check List for Procure- ment of Medical Devices pursuant to EN ISO 17664”). The manufacturer of a reusable MD must supply instructions on how to reprocess the 1 The RKI scheme rather than the Spaulding scheme takes into account the consequences for reprocessing of assigning MDs to risk groups.
  • 4. wfhss Guideline on Reprocessing Medical Devices Status:June 2012 Page 4 of 15 device, using (pre-) validated2 processes as per EN ISO 17664. Competencies and working practices related to all steps of the reprocessing chain must be set out in writing in the form of standard operating 3 Risk analysis, evaluation, selection and definition of suitable reprocessing methods To ensure that they are reprocessed in prescribed form, medical devices must be classified into risk groups in accordance with Annex 2. Based on this classification, it must be set out in writing • whether, • how often, • with which process • and under what conditions medical devices are to be reprocessed and stored. The following aspects, in particular, must be borne in mind: • the manufacturer’s instructions, if relevant, • the resistance profiles of the microorganisms and potentially infectious agents (e.g. prions) expected on the MDs, • expected intervals between the time the MDs are used and the time repro- cessed, with correspondingly more challenging reprocessing demands as well as • special risks posed by the MD when using and/or reprocessing it. 4 Organisational measures, quality management (QM), validation, monitoring Reprocessing Units for Medical Devices (RUMEDs) entrusted with medical device re- processing are divided into three categories on the basis of the highest risk group to which the medical devices they reprocess are assigned (see Annex 3). Competencies and responsibilities within the RUMED must be defined (e.g. in the form of an organogram). RUMED managers must have corresponding deputy manag- ers. 2 The term “prevalidation” is understood to mean tests conducted for the purpose of identifying a suitable reprocessing procedure; these are carried out on behalf of the manufacturer and should preferably be performed by an independent test body
  • 5. wfhss Guideline on Reprocessing Medical Devices Status:June 2012 Page 5 of 15 The structural conditions must permit high quality reprocessing in line with the state of the art in science and technology. The equipment, fittings and adjuncts needed for MD reprocessing must be available and comply with the state of the art (see Annex 3). For medical device reprocessing all precautions and measures must be taken to as- sure a demonstrable and continually high quality of reprocessing, in line with the vali- dation specifications. Reprocessing of medical devices subject to ultra stringent reprocessing demands (“Critical C” devices as per Annex 2) must be conducted in line with a QM system pursuant to EN ISO 13485 (see Annex 3). The quality of automated reprocessing (cleaning, disinfection and sterilisation) must be assured by the following measures: • installation qualification • validation of reprocessing procedures • periodic (at least annual) revalidation • routine daily tests on each day the RUMED is in operation • batch-related routine tests • monitoring and testing of process parameters using measurement technology. 5 Personnel qualification Reprocessing in general, or any reprocessing task, may only be conducted by persons who by virtue of their training, knowledge and practical experiences are able to as- sure proper execution of such tasks (see Annex 3): • All persons entrusted with MD reprocessing must have successfully attended Specialist Training Course 1 (series of training courses specifically designed for sterilisation assistants), or be in possession of an equivalent qualification. • Managers and deputy managers of Category II RUMEDs must additionally have successfully attended Specialist Training Course 2 • Managers and deputy managers of Category III RUMEDs must additionally have successfully attended Specialist Training Course 3 Knowledge and skills must be acquired by attendance at appropriate continuing pro- fessional development seminars based on the state of the art in science and technol- ogy.
  • 6. wfhss Guideline on Reprocessing Medical Devices Status:June 2012 Page 6 of 15 6 Documentation When reprocessing MDs belonging to Semi-Critical B as well as to Critical B and C risk groups, the measured values recorded for the process parameters and underlying the decision-making process for MD release must be documented, while indicating the batch name and name of the person responsible for release. These values must fur- nish proof that the reprocessing procedure was carried out in compliance with the standard operating procedures (SOPs), using the parameters set out in the validation protocol. This batch documentation must comprise at least: • Date • Equipment serial number • Programme • Serial batch numbers • Designation of reprocessed load or similar code • Operator’s name • Release by responsible person • Printout of measured values in analog or digital form (actual values) • If applicable, results of batch control (e.g. cleaning indictors, chemical indicators) Documentation of reprocessing MDs belonging to the non-critical risk group is not re- quired, while that of Semi-Critical A and Critical A risk groups must be adapted to the intended use on the basis of risk assessment and the mode of reprocessing inferred from that. Category I RUMED documentation, as per Annex 3, must be archived in a well- assured readable form for at least five years, and Category II and III RUMED docu- mentation for at least 10 years. Electronic documentation, including release docu- mentation, is permitted provided that readability and data safety are assured for the specified period. 7 External reprocessing (outsourcing) The term “external reprocessing” is used to describe a situation where healthcare es- tablishments outsource reprocessing in general, or parts thereof, to an authorised external service provider. The same requirements refer to insourcing. Outsourcing of reprocessing to a third party must not in any way negatively impact on the healthcare establishment’s ability to meet the terms of its own service con- tract. Reprocessing service providers or healthcare establishments that conduct MD repro-
  • 7. wfhss Guideline on Reprocessing Medical Devices Status:June 2012 Page 7 of 15 cessing for other healthcare institutions must use validated reprocessing procedures. Commercial service providers conducting MD reprocessing for healthcare establish- ments with Category II and III RUMEDs as per Annex 3 must have in place a certified QM system as per ÖNORM EN ISO 13485. The following aspects, in particular, must be contractually regulated between the ex- ternal service providers and the person (s) responsible for the respective healthcare establishment (Category II and III RUMEDs as per Annex 3): • The conditions to be met to assure running operations (e.g. emergency measures as well as regulations on coping with bottle necks, breakdowns, standby); • Duties, responsibilities and competencies of contract awarder and contractor (interfaces); • Ownership rights as regards the MDs; • Condition and quality of MDs to be reprocessed; • Liability for (transport) damage; • Handover conditions; • Feedback system; • Quality assurance measures (if necessary, certification as per ÖNORM EN ISO 13485, see above); • Transport logistics, including protected transport of contaminated instruments; • Maximum time interval before soiled MDs are reprocessed; • Whether and how any pre-treatment measures are to be conducted for MDs already within the respective establishment, while taking account of specific types of soils; • (Special) release regulations; • Loaned instruments; • Validation obligations of contract awarder and contractor; • Qualifications of managers and staff in both premises (as per Annex 3); • Nature, frequency and scope of checks to be conducted by contract awarder; • Nature and scope of documentation to be provided by the contract awarder (validation, re- validation reports, if applicable batch release documentation). • Compatibility with the quality management system of the hospital • Compatibility with the documentation system of the hospital 8 Authors A. Blacky, V. Buchrieser, T. Freundlinger, M. Gehrer, H. Getreuer, A. Gruber, M. Hell, W. Koller, P. Lachner, T. Miorini, A. Percht, G. Palmisano, U. Prüfert-Freese, M. Suchomel, A. Steinhardt, B. Weinmayr The guideline has been proof read and authorized by the wfhss education group
  • 8. wfhss Guideline on Reprocessing Medical Devices Status:June 2012 Page 8 of 15 ANNEX 1: Standard operating procedures (SOPs) The RUMED management must ensure that standard operating procedures are availalble and clearly displayed for the persons entrusted with the respective tasks. The RUMED manage- ment, or a person explicitly designated in writing to that effect by the latter, must ensure that RUMED employees are also able to understand the content of such SOPs. This applies in particular for the following: 1. All partial steps, such as: • Measures conducted immediately after using an MD, e.g. pretreatment, dis- mantling, collection and intermediate storage • Transport • Cleaning • Rinsing • Disinfection • Drying • Verification of steps conducted hitherto (cleaning results, etc.) • Maintenance and care • Functional testing • Packaging • Arrangement of supplies for sterilisation • Sterilisation • Labelling • Release • Transport of sterile supplies • Storage of sterile supplies • Measures conducted immediately before use (e.g. checking that sterile sup- plies are undamaged) 2. Special processes 2.1 Reprocessing MDs that have not been used These medical devices include: - MDs which were supplied in an unsterile condition but must be sterile when put to use, - Sterilised MDs whose packaging was damaged or opened without the MD having been used and which, as specified by the manufacturer’s in- structions, can be reprocessed. - MDs whose sterile storage period has expired before the expiry date
  • 9. wfhss Guideline on Reprocessing Medical Devices Status:June 2012 Page 9 of 15 and which, in terms of their composition, are amenable to reprocessing. 2.2 Reprocessing of MDs which may be used or reprocessed only for a limited number of times 2.3 Reprocessing of textile materials 2.4 Patient-related specific forms of reprocessing. 3. Critical processes 3.1. Reprocessing of heat-sensitive MDs (e.g. endoscopes) 3.2. Reprocessing of MDs with poorly accessible sites (e.g. minimally invasive surgical (MIS) instruments) 3.3. Reprocessing of MDs posing a risk of injury to users, personnel or patients (e.g. faulty electrical currents, pointed/sharp implements) 3.4. Reprocessing of MDs subject to more stringent reprocessing requirements (e.g. residues that are difficult to remove) 3.5. Reprocessing of MDs made of special materials or technical configuration (e.g. kinks, liable to scratches, antimagnetic, cannot be immersed in liquid, alumini- um) 3.6. Manual reprocessing steps 3.7. Reprocessing of MDs subject to special reprocessing requirements because of specific infectious agents (e.g. prions)
  • 10. wfhss Guideline on Reprocessing Medical Devices Status: June 2012 Page 10 of 15 ANNEX 2: Classification of medical devices into risk groups (modified as per RKI) Classification of MDs into risk groups is done on the basis of previous and subsequent use, transport and storage conditions, design and materials. If there is any doubt about the risk group to which the MD belongs, assign it to the higher group. Preference must be given to automated reprocessing in a washer-disinfector because it assures better standardisation and occupational protection. Disinfection should preferably be performed with a thermal process. Classification Description Medical device Example Pretreat- treat- ment a Cleaning / Disinfec- tion Spec. label Sterili- sation Critical process steps, special re- quirements NON-CRITICAL Contact with intact skin ECGelectrode, Kidney dish X ( X 1 ) Inspection:visual cleanli- ness SEMI- CRITICAL Contact with mucosa or pathologically altered skin A No special repro- cessing require- ments b Specula , Oral mirror, dental scoop, endoscopes without lumens (X 2 ) X (X 3 ) Preferablyautomated cleaning (C)and disinfec- tion (D),atleastdisinfec- tion withsuitableagents/ process B Subjecttomore stringentrepro- cessing require- ments Flexible endo- scopeswith lumens c , Hand and angled pieces c,d ,, X 4 X (X 3 ) Automated C/D CRITICAL Skin or mucosa penetration, contact with open wounds, drugs, blood or blood products that must be administered in sterile condition A No special repro- cessing require- ments Woundretrac- tors, Needleholders (X 2 ) X X Automated C/Dfor Cat.IIIRUMED,otherwise preferablyautomated C/D Steamsterilisation B Subjecttomore stringentrepro- cessing require- ments Heat--sensitive MISinstru- ments, surgical drapes X 4, e X (X 5 ) X Forall partscominginto directcontactwith tissue, onlyautomated thermal C/Din suitableWDand steam sterilisation C Subjecttoultra stringentrepro- cessing require- ments Angioscopes Epiduroscopes X 4 X X 5 X Fornon-thermalsterilisa- tionprocesses,noevi- dence ofprion inactivation hasbeenfurnished to date.This mustbe borne in mindforMDs belonging to this group a Pretreatment: this is understood to mean removal of course soils or a measure aimed at prevention of incrustations, but need not be limited to these. Pretreatment is done at the site of MD use. b e.g. can cleanliness be checked through visual inspection c if it cannot be classified as “critical”; d preferably, automated reprocessing, if this is not possible, then steam sterilisation with suitable pro-
  • 11. wfhss Guideline on Reprocessing Medical Devices Status: June 2012 Page 11 of 15 cesses following manual reprocessing e does not apply to surgical drapes (X) optional working step X 1 steam sterilisation may be needed, bearing in mind specific requirements. X 2 pretreatment immediately after use with non-protein fixing agents/process X 3 for MDs that are not used in physiologically sterile body cavities, final disinfection suffices apart from cases of intraoperative use. X 4 these MDs (semi-critical B, critical B, C) must be treated immediately after use to pave the way for subsequent cleaning and disinfection, so that final cleaning is conducted as quickly as possible us- ing automated processes, while preventing corrosion or other form of damage to the instruments. All factors of influence here must be defined and taken account of during validation.. X 5 A special label is used for identification, information and tracking (as required, information must be displayed on the following: batch no., serial no., manufacturer’s name, type and number of repro- cessing cycles., max. number of reprocessing cycles permitted, safety and warning instructions, etc.).
  • 13. wfhss Guideline on Reprocessing Medical Devices Status: June 2012 Page 13 of 15
  • 14. wfhssGuidelineonReprocessingMedicalDevices Status:June2012Page14of15 ANNEX3:RequirementsforRUMEDsinhealthcareestablishments ReprocessingunitcategoryIIIIII Riskgroupofmedicalde- vice(MD)tobereprocessed Non-critical,semi-criticalA, criticalA(1) Non-critical,semi-criticalA,B, criticalA Allgroups Healthcareestablishments e.g.(homesfortheelderly/nursinghomes, outpatientdepartments,medicalpractices (2) e.g.healthcareestablishmentswithoutsur- gicalunits e.g.healthcareestablishmentswithsurgical units QualityassuranceAppropriatequalityassurancemeasures Appropriatequalityassurancemeasures (minimumrequirementsasper RobertKochInstitute-RKI) QMsystempreferablybasedon,forcriticalC devicesbasedonENISO13485:2003 Structuralrequirements -Dedicatedarea -Preferablywithseparateunclean/ clean/sterilezones(temporalsepara- tionpossible) -Dedicatedreprocessingroom -Separateunclean/clean/sterilezones(3) -Dedicatedpremises -Separateunclean/clean/sterilepremises(4) Staffqualifications ManagementandstaffatleastSpecialist CourseI(5) Management:atleastSpecialistCourseII Staff:atleastSpecialistCourseI Managementanddeputies:SpecialistCourse III Staff:atleastSpecialistCourseI Technicalequipment -Ifnecessarysmallsteamsterilizer (ClassBrecommended)asperEN 13060 -If,necessaryultrasoniccleaningequip- ment -Ifnecessary,washer-disinfector(WD) asperENISO15883 -Ifnecessary,heatsealingmachine -Steamsterilizers(bigorsmall)asper EN285orEN13060 -EndoscopeWDasperENISO15883 -Ifnecessary,rotarysealingmachine -Ifnecessary,ultrasoniccleaningequip- ment -Steamsterilizers(big)asperEN285 -Ifnecessary,equipmentforspecialsterili- zationprocesses(e.g.low-temperature sterilizers) -EndoscopeWDasperENISO15883 -Rotarysealingmachine -Ultrasoniccleaningequipment (1)Callforspecialconsideration;(2)withtheexceptionoflumenedendoscopes(Cat.II)andsurgicalprocedures(Cat.III),exceptformanagementofsmallwounds;(3)For newendoscopyunits–centralisationandseparateunclean/cleanpremisesarerecommended;(4)Fornew,extendedorconvertedpremises,cleanareashouldmeetre- quirementsofISOclass8asperISO14644-1,asfaraspossiblealsoforexistingestablishments;(5)deemedtohavebeenmetfordegree-levelnursingstafforformedi- calpersonnel
  • 15. wfhssGuidelineonReprocessingMedicalDevices Status:June2012Page15of15 Conceptforcontinuingprofessionaldevelopmentcoursesfor"Reprocessingmedicaldevicesin/forhealthcareestablishments" COURSESPECIALISTCOURSEISPECIALISTCOURSEIISPECIALISTCOURSEIII TARGETGROUP Semi-skilledstaffandassistants; Dentalstaff**ordentalassistants Semi-skilledstaffwithspecialduties(e.g. abletoreleaseMDs)andgraduateswithre- sponsibilityforasection Managersanddeputymanagers EXTENT 80hours: theoryandpracticalexercises 80hours: 40hourstheory 16hourspracticalexercises 24hourspracticalwork 80hours: 40hourstheory 40hoursconcludingassignment ADMISSIONREQUIRE- MENTS None SuccessfulcompletionofSpecialistCourseI. andatleast1year’sexperienceinaMDre- processingunit or SuitablyrecognisedQualification(inaccord- ancewithnationalrequirements)andwith successfulexaminationoncontentofSpe- cialistCourseI(withoutattendingcourse) SuitablyrecognisedQualification(inaccord- ancewithnationalrequirements)andsuc- cessfulcompletionofSpecialistCourseII REQUIREMENTFOR ManagementofRUMEDCat.I**, StaffofRUMDCat.I**,II,III ManagersanddeputymanagersofRUMED Cat.II ResponsibilityforsectionofRUMEDCat.III. ManagersanddeputymanagersofRUMED Cat.III DEADLINEFORCOM- PLETION Within2yearsoftakinguppositionWithin2yearsoftakinguppositionWithin2yearsoftakingupposition EXAMINATIONExaminationofknowledgeExaminationandassignmentExaminationandfinalassignment *ifnotcoveredinspecialistORcourse,**exceptfordegree-levelnursingstafforformedicalpersonnel Source:AustrianSocietyforSterileSupply(ÖGSV),www.oegsv.com