2. LCSU 3
● Battery powered
portable suction unit, up
to 45 mins constant use.
● 300ml canister, single
use.
● Variable vacuum 50 –
500mmHg.
● Rechargeable and
removable battery.
3. LCSU 3 top
Battery low indicator
Suction power
mmHg
Power on /off
Charging (out when External
fully charged) power
4. LCSU 3 Side
Filter assembly
Graduated 300mL
canister. Disposable
single use only
Tubing attachment
Suction
control
5. USE
1. Attach tubing and Yankauer to canister
inlet.
2. Turn on and check suction by occluding
outlet – look for green light to 550mmHg.
3. Leave off until suction actually needed –
(conserve battery and lessen noise)
4. Ensure unit is upright to avoid contact of
fluid with filter.
6. Stowage
Prehospital Trauma Pack 2 (Red).
The unit is self contained, including one canister,
Yankauer Sucker and tubing length.
7. Cannister
● Canister snaps
into place.
● Note, if filter
becomes wet,
the suction
will cease to
protect the
internal motor.
● Always use in
upright
position to
maximise
available
volume
8. ● Use coin or similar to open Batteries
housing.
● Remove Battery and
disconnect cable.
● Replace with fully charged
battery.
● Battery can be charged within
unit.
● Device can run while being
charged
● Charge duration 5 hours max
● Battery life – 45 mins
maximum continuous use.
9. Important Notes
● The suction will cease if the filter becomes wet, this
protects the motor.
● Avoid canister content from coming into contact with
filter, eg by careless handling.
● The low battery warning indicates < 5 mins power.
● If suction fails consider moving to alternative, eg
Vehicle, Aircraft, Bridge.
● The catheter tip 60mL syringe is available in the
intubation kit as a last resort.
10. Usage
● This device is to be utilised on all RSI if vehicle
suction is not practical or available.
● The 300mL canister may fill rapidly, source
alternate suction if likely, by moving patient to
vehicle.
● The canister cannot be emptied and re-used.
● The device will be checked on daily level 3 checks.
● The entire unit should be rotated and placed on
charge, for level 1 checks and after use.
11. Specifications
● Weight: 1.53 kg
● Size: 18cmH x 26.7cmW x 7.4cmD
● Disposable Canister Capacity: 300ml
● Integrated Filter: Internal Bacterial Filter/ Fluid Shut-off
● Patient Suction Tubing (non-sterile)
● Length: 0.9m
● Vacuum - Range: 50-550mmHg (6.7 - 73kPa)
● (+/-27.5 mmHg)
● Battery Operation Time: Up to 45 minutes
● Noise Level: < 70 dB(A)
● *Conditions may vary according to environmental factors, system configuration, power source,
age/maintenance of battery, and/or state of charge.
● Electrical Requirements: 100-240 V AC 47-63 Hz, 0.75A max; 12 V DC, 33 W max
● Battery: High Capacity Rechargeable NiMH
● Charging Time: Up to 5 hours
● Operating
● Temperature Range: 0 degrees C - 40 degrees C
● Relative Humidity: 0-95%
● Atmospheric Pressure: 10.2 Psi (70 kPA) - 15.4 Psi (106 kPA)
● Storage and Transport
● Temperature Range: -40 degrees C - 70 degrees C
● Relative Humidity: 0-95%
● Atmospheric Pressure: 7.3 Psi (50 kPA) - 15.4 Psi (106 kPA)