3. Asystole is the end
60 year old man ?previously well
Presented to regional hospital ED with
syncope
Asystolic arrest soon after arrival
Intubated
CPR and adrenaline ~ 20 mins
Called for senior advice (no ED SS)
7. Clinical examination
Patient looks surprisingly pink for at least 20 mins
asystole
Confirm no peripheral pulses but palpable
“pulses” with CPR
“Gut feeling” is hmmmm…….
What’s that rule about asystole?
8. Is it really asystole?
Check other leads (I,III, V) – no obvious
difference
Closer look at lead II – are those small bumps?
Rhythm strip with increased amplitude:
12. Transcutaneous pacing
“Older style” TCPM (Lifepak)
Rate set at 70bpm
Amplitude increased by 20 mA increments
Capture at 120 mA
Good pulses and SBP 130mmHg!
13. Rembrandt van Rijn “The Raising of Lazarus” Los
Angeles County Museum of Art, Los Angeles
15. The Raising of Lazarus…..
Patient started to move!.....
…opened eyes!!......
..responded to simple instructions!!!
Transcutaneous pacing did not appear to be
causing him to much discomfort
What now?
16. Temporary pacing wire
Insertion of temporary transvenous pacing wire
in ED (RIJV; balloon flotation)
17. Temporary transvenous
pacing
How do I know it’s in the correct position?
Length of insertion
ECG electrode
Ectopics
CXR / bedside imaging
20. Settings
Capture threshold - the minimum electricity output
(mA) required to pace the heart (capture)
Asynchronous – paces at a fixed rate, ignoring any
of the patient’s intrinsic activity
Demand - paces only if the intrinsic heart rate gets
lower than the set rate. This is the usual setting.
Sensitivity – the sensitivity of the pacemaker to the
patient’s intrinsic activity. When an intrinsic beat is
“sensed’ (sensing light), this “inhibits” the
pacemaker from firing
21. Settings
Setting sensitivity – turn the output to 0 (prevent
discharge); set rate below intrinsic rate; reduce
sensitivity (increase amplitude) until no sensing
of intrinsic activity (sensing threshold); set at ~
half the amplitude of this threshold.
Setting output – turn output to 0 and set rate at
10-15 bpm higher than intrinsic rate; increase
output until capture occurs (capture
threshold); set at 2-3 times higher than this
threshold.
22. Eventual outcome
Extubated inED
Admitted to CICU
History of syncope over several months –
neurological work-up was normal!
Probable sick sinus syndrome
Referred to Metropolitan hospital for permanent
pacemaker
Full neurological recovery and return to usual
independent activities
23.
24. Summary
Don’t ignore those gut instincts
The old rule of thumb: a rhythm disturbance with
haemodynamic compromise needs electricity
Temporary pacing can save lives
Transcutaneous pacing is getting better and
easier
Temporary transvenous pacemaker insertion is
getting better and easier