510(k) Summary per 21 CFR 807.92
Page I of4

510(k) U K120398

AUG 162012
5 10(k) SUBMITTER:

ESTABLISHMENT
REGISTRATION:
CONTACT:
DATE PREPARED:
PROPRIETARY NAME:
PANEL:
REGULATION NUMBER:
CLASSIFICATION:
PRODUCT CODES:

Hollywog, LLC
2830 Amnicola Highway
Chattanooga, TN 37406
3008585473
Michael W. Treas,
Chief Compliance Officer
August 15, 2012
WiTouch'Pro
Neurology
CFR Title 21, 882.5890
Class 11
GZJ (Stimulator, Nerve,

Transcutaneous, For Pain RelieQ;

NYN (Stimulator, Electrical, Transcutaneous, For Arthritis)

COMMON NAME:

Transcutaneous electrical nerve stimulator for pain
relief

Description:
M
The WiTouch T Pro medical device is a Transcutaneous Electrical Nerve Stimulator (TENS) used to
reduce the perception of pain by electrically stimulating peripheral nerves across the skin
(transcutaneously). The design consists of a battery powered current generator with integral electrodes
and replaceable electrical dispersive hydrogel pads (gel-pads). One side of the adhesive gel-pad adhere
to the integral electrode, and the other side adhere the device to the healthy intact skin of the patient's
back to provide a transcutaneous analgesic electrical stimulus to the painfulI area.

Indications for Use:

The WiTouh m Pro Transcutaneous Electrical Nerve Stimulator Device is used* for the symptomatic
relief and management of chronic intractable back pain and relief of pain of the upper and lower back
associated with arthritis. It is also used for adjunctive treatnment for post-surgical and post-trauma acute
back pain.
Intended Use:

The intended use is to provide analgesic electrical stimulus to reduce the perception of back pain by
electrically stimulating peripheral nerves across the skin (transcutaneously).
510(k) Summary per 21 CFR 807.92

510(k) # K120398

Page 2of 4

Accessories:
The device utilizes hydrogel pads (gel-pads) for achieving the indications for use and intended use. The
composition of the gel-pads is common materials found in the electrode industry. The uniqueness of the
gel-pads is in the shape. The maximum average power density of the electrodes with the gel-pads
applied i less than 0.25 watts per square centimeter of electrode conductive surface area to reduce the
risk of thermal burns which is consistent with the referenced predicate devices.
510(k) #

Substantially Equivalent Predicate Devices
PredicateRegulation, Classification, and Product Code(s)
Predicate Proprietary
Trade Name

K061.650
K061516

Empi SELECT"
Staodyn' Max Preset TENS

_______________________________

CFR Title 21, Sec. 882.5890, Class 11, GZJ, NYN
CER Title 2 1, Sec. 882.5 890, Class 11, GZJ, NYN

The predicate devices utilize flexible wires between the electrodes and the electrical stimulus generator;
thus, increases the indications for use to the lower back and to body surfaces with greater ranges of
motion (e.g., knee, shoulder, elbow, and hip).
The intended design of the WiTouch TMPro device limits the application for use to the anatomical sites of
the upper and lower back. The design includes carbon rubber electrodes that are intended for reuse and
are permanently-affixed to a rigid surface of the electrical stimulus generator. The unique connection
makes the electrodes integral to the generator. The connection and shape of the electrodes limit
application of the device to the contours of the back. These unique characteristics of the integral
electrodes are insignificant as it relates to safety and effectiveness, and is not critical to the intended use
between the device and the referenced predicate devices.
The referenced predicate devices utilize affixed buttons as the sole method to control the electrical
stimulus generator on/off and intensity up/down. The WiTouchTm Pro utilizes an additional method of a
wireless remote control radio frequency transceiver to control the electrical stimulus generator on/off
and intensity up/down. The transceiver operates in the ISM radio frequency band for wireless medical
technology. This uniqueness of controlling the electrical stimulus generator by utilizing a radio
frequency transceiver is insignificant as it relates to safety and effectiveness, and is not critical to the
intended use between the device and the referenced predicate devices
The characteristics of the analgesic electrical stimulus output between the device and the referenced
predicate devices are substantially equivalent as it relates to safety and effectiveness.
510(k) Summary per 21 CPU 807.92
Page 3 of 4

510(k) # K120398

Technological characteristics and physical properties of the WiTouch Pro device and the referenced predicate
devices:
510(k) Number
Device Name and Model

K120398
WiTouch'M Pro

K061650
EmpiO Select

Manufacturer
TENS Device Power Source (DC battery)

Hollywog
Battery 1.SVDC

EMPI
Battery i.SVDC

K(061516
StaodynO
Max preset TENS
Compex Technologies
Battery 1.5VDC

(2-Alkaline AAA)

(3-Alkaline AAA)

(3-Alkaline AAA)

4

4

4

1 Channel
Asynchronous

2 Channels
Asynchronous

2 Channels
Synchronous

Yes
Yes

Yes
Yes

Yes
Yes

Yes

Yes

Yes

______________________________

Number of Output Modes
Number of
Synchronous or
Output
Asynchronous?
Channels:
Software/Firmware/MIicroprocessor
Control?
Automatic Shut Off?
User

Override Control?

{

On/Off St atus?
Low Batte ry?
Voltage/Current Level?
Timer Range (minutes)
Weight (lbs., oz,)

Indicator
Display:

Dimensions (in.) [W x H x DJ
Housing Material and Construction
Waveform (e.g., pulsed monophasic,
biphasic)

Yes
Yes
No
30
4.8 oz.
w/ batteries included
7.5"(W) x 3.5(H)" x
D.7"(D)
Silicone & ABS
Pulsed biphasic
Square

sinusoidal)

Symmetrical phases?
Step-i and Step-3
Phase Duration (include
units), (state range, if
applicable), (both
phases, if
asymmetrical)
Step-2
Phase Duration (include
units), (state range, if
applicable), (both
phases, if
asymmetrical)
Maximum Current Density, (mA/cm 2,
r.m.s.)
Maximum Average Current (average
absolute value), MA
Maximum Average Power Density,
For multiphasic
waveforms only:

(W/CM ),

Yes
No
No
i to 30
5.15 oz.
w/ batteries included
2.5"(W) x 5.25"(H) x
1.0"(D)
Plastic
Pulsed biphasic
____________

____________

Shape (e.g., rectangular, spike, rectified

2

Yes
Yes
No
i to 30
4.9 oz.
w/ batteries included
2.38"(W) x 1.38"(H) x
4.31"(D)
Plastic
Pulsed biphasic
Square
____________

Square
____________

No
12011S

No
0 to 40011s at 50% peak
amplitude

Yes
0 to 300pis

480p

0 to 400115 at 50% peak
amplitude

0 to 30li

0.l2mA@500 0

<l0mA @500 0

<l0mA @500 0

1.GmA@500 0

<l0mA @500 0

<l0mA @5000Q

0.00069W/cm2@ 500 a

<0.25 W/cm2 at 500 0)

<0.25 W/cm2 at 500 0

(using smallest electrode

conductive surface area)

___________________________________
510(k) Summary per 21 CFR 807.92

510(k) # K120398

Page 4 of4

Declarations of Conformity
The device complies with the following FDA recognized standards:
FDA Recognized Number 5-4, IEC 60601-1, Medical Electrical Equipment - Part 1: General
Requirements for Safety, 1988 Amendment 1, 1991 -1 l,,Amendment 2, 1995. (General)
FDA Recognized Number 5-60, IEC 60601-1-2 Int. I Third Edition/I-SR 01:2007, Medical electrical
equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral
standard: Electromagnetic compatibility - Requirements and tests, Interpretation Sheet. (General)
FDA Recognized Number 5-41, Medical electrical equipment - Part 1-4: General requirements for
safety- Collateral standard: Programmable electrical medical systems, edition 1.1I. (General)
FDA Recognized Number 17-5, lEG 60601-2-10 1987/Amendment 1 2001, Medical electrical
equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators.
(Neurology)
FDA Recognition Number 2-156: AAMI/AMSI/ ISO 10993-1:2009, Biological evaluation of medical
devices - Part 1: Evaluation and testing within a risk management process. (Biocompatibility)
FDA Recognition Number 2-153 (Electrodes) ISO 10993-5:2009, Biological evaluation of medical
devices - PartS5: Tests for In Vitro cytotoxicity. (Biocompatibility)
FDA Recognized Standard 2-173 (Electrodes) Recognition Number 2-173: AAMI / ANSI / ISO 1099310:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
(Biocompatibility)
Truthful and Accurate Statement
A statement was included in the Premarket Notification attesting to the truthfulness and accuracy of the
information provided.
Further Information
In the event that additional information is required, please contact:
Michael W. Treas
Chief Compliance Officer
Hollywog, LLC
2830 Amnicola Highway
Chattanooga, TN 37406
Telephone: (423) 305-7778 ext. 108
Fax: (423) 305-7867
E-mail: mike.treas~hollywog.com
DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration
10903 Newv Hampshire Avenue
Document Control Room -W066-G609
Silver Spring, MD 20993-0002

AUG 162012
Hollywog, LLC
Mr. Michael Treas
Chief Compliance Officer
2830 Amnicola Highway
Chattanooga, TN 37406
Re: K120398
TradeIDevice Name: WiTouch Pro
Regulation Number: 21 CFR 882.5 890
Regulation Name: Transcutaneous nerve stimulator, for pain relief
Regulatory Class: II
Product Code: GZJ, NYN
Dated: July 10, 2012
Received: July 12, 2012
Dear Mr. Treas:
We have reviewed your Section 5 10(k) premarket notification of intent to market the device
referenced above and have determined the device is substantially equivalent (for the indications
for use stated in the enclosure) to legally marketed predicate devices marketed in interstate
commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to
devices that have been reclassified in accordance with the provisions of the Federal Food, Drug,
and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA).
You may, therefore, market the device, subject to the general controls provisions of the Act. The
general controls provisions of the Act include requirements for annual registration, listing of
devices, good manufacturing practice, labeling, and prohibitions against misbranding and
adulteration. Please note: CDRH does not evaluate information related to contract liability
warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class 11 (Special Controls) or class Ill (PMA), it
may be subject to additional controls. Existing major regulations affecting your device can be
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may
publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean
that FDA has made a determination that your device complies with other requirements of the Act
or any Federal statutes and regulations administered by other Federal agencies. You must
Page 2 - Mr. Michael Treas
comply with all the Act's requirements, including, but not limited to: registration and listing (21
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic
product radiation control provisions (Sections 53 1-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
go to http://www.fda.gov/AboutFDA/CenterSOffices/CDRH/CDRHOffices/ucm 15809.htm for
the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please
note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21
CFR Part 803), please go to
http://www.fda.pov/MedicalDevices/SafetLReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the
Division of Small Manufacturers, International and Consumer Assistance at its toll-free number
(800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.pov/MedicalDevices/ResourcesforYou/Industr/default.htm.
Sincerely yours,

Malvina B. Eydelman, M.D.
Director
Division of Ophthalmic, Neurological,
and Ear, Nose and Throat Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
Indications for Use
510(k) Number:

K120398

Device Trade Name: WiTouch® Pro
Indications for Use:
The WiTouch® Pro Transcutaneous Electrical Nerve Stimulator Device is used
for the symptomatic relief and management of chronic intractable back pain and
relief of pain of the upper and lower back associated with arthritis. It is also used
for adjunctive treatment for post-surgical and post-trauma acute back pain.

FPrescription Use
(Part 21 CER 801

____AND/OR]

-atD

Over-The-Counter Use

F(21

_

_

CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)'

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division VnOf
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices

5l0(k)Nunber.

More Related Content

Similar to ZMPCHW070000.13.04 US FDA approval letter K 120398

Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture DeviceNeuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture Devicemcalipo
 
Contactors and Thermal Overload Relays - FJ Series - Fuji Electric
Contactors and Thermal Overload Relays - FJ Series - Fuji ElectricContactors and Thermal Overload Relays - FJ Series - Fuji Electric
Contactors and Thermal Overload Relays - FJ Series - Fuji ElectricCTY TNHH HẠO PHƯƠNG
 
Contactors and Thermal Overload Relays - FJ Series - Fuji Electric
Contactors and Thermal Overload Relays - FJ Series - Fuji ElectricContactors and Thermal Overload Relays - FJ Series - Fuji Electric
Contactors and Thermal Overload Relays - FJ Series - Fuji ElectricCTY TNHH HẠO PHƯƠNG
 
Original NPN Transistor TIP31C TIP31 TO-220F New STMicroelectronics
Original NPN Transistor TIP31C TIP31 TO-220F New STMicroelectronicsOriginal NPN Transistor TIP31C TIP31 TO-220F New STMicroelectronics
Original NPN Transistor TIP31C TIP31 TO-220F New STMicroelectronicsAUTHELECTRONIC
 
בחינת הצללה ברמת המודול לאלקטרוניקת אנרגיה
בחינת הצללה ברמת המודול לאלקטרוניקת אנרגיהבחינת הצללה ברמת המודול לאלקטרוניקת אנרגיה
בחינת הצללה ברמת המודול לאלקטרוניקת אנרגיהTashtiot media
 
Ultrasonic welding questions
Ultrasonic welding questionsUltrasonic welding questions
Ultrasonic welding questionsRutvik
 
Original Transistor NPN D882 2SD882 3A 40V TO-126 New
Original Transistor NPN D882 2SD882 3A 40V TO-126 NewOriginal Transistor NPN D882 2SD882 3A 40V TO-126 New
Original Transistor NPN D882 2SD882 3A 40V TO-126 NewAUTHELECTRONIC
 
SOI RF Switch for Wireless Sensor Network
SOI RF Switch for Wireless Sensor NetworkSOI RF Switch for Wireless Sensor Network
SOI RF Switch for Wireless Sensor NetworkADEIJ Journal
 
Power generation from shoes & utilize it to charge the mobile's, laptop's or ...
Power generation from shoes & utilize it to charge the mobile's, laptop's or ...Power generation from shoes & utilize it to charge the mobile's, laptop's or ...
Power generation from shoes & utilize it to charge the mobile's, laptop's or ...Shantesh Singh
 
Original NPN SILICON POWER DARLINGTONS Transistor TIP142 10A 100V TO-220F New...
Original NPN SILICON POWER DARLINGTONS Transistor TIP142 10A 100V TO-220F New...Original NPN SILICON POWER DARLINGTONS Transistor TIP142 10A 100V TO-220F New...
Original NPN SILICON POWER DARLINGTONS Transistor TIP142 10A 100V TO-220F New...AUTHELECTRONIC
 
Catalog ELCB Mitsubishi-Beeteco.com
Catalog ELCB Mitsubishi-Beeteco.comCatalog ELCB Mitsubishi-Beeteco.com
Catalog ELCB Mitsubishi-Beeteco.comBeeteco
 
Trans Block Show
Trans Block ShowTrans Block Show
Trans Block Showmjsmith9
 
Original Transistor NPN TIP41C TIP41 6A 100V TO-220 New
Original Transistor NPN TIP41C TIP41 6A 100V TO-220 NewOriginal Transistor NPN TIP41C TIP41 6A 100V TO-220 New
Original Transistor NPN TIP41C TIP41 6A 100V TO-220 NewAUTHELECTRONIC
 
Original PNP Transistor A1020-Y A1020 1020 2SA1020 TO-92 New ON semiconductor
Original PNP Transistor A1020-Y A1020 1020 2SA1020 TO-92 New ON semiconductorOriginal PNP Transistor A1020-Y A1020 1020 2SA1020 TO-92 New ON semiconductor
Original PNP Transistor A1020-Y A1020 1020 2SA1020 TO-92 New ON semiconductorAUTHELECTRONIC
 
CPRI-Power quality report
CPRI-Power quality reportCPRI-Power quality report
CPRI-Power quality reportNageswar Rao
 
Save energy and protect your facility with voltage optimisation
Save energy and protect your facility with voltage optimisationSave energy and protect your facility with voltage optimisation
Save energy and protect your facility with voltage optimisationNick Lee
 
Megger Power And Industrial Test Equipment - Cable Fault Locators, High Volta...
Megger Power And Industrial Test Equipment - Cable Fault Locators, High Volta...Megger Power And Industrial Test Equipment - Cable Fault Locators, High Volta...
Megger Power And Industrial Test Equipment - Cable Fault Locators, High Volta...Thorne & Derrick International
 
Introduction final a
Introduction final   aIntroduction final   a
Introduction final aAmala Putrevu
 

Similar to ZMPCHW070000.13.04 US FDA approval letter K 120398 (20)

Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture DeviceNeuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
 
Contactors and Thermal Overload Relays - FJ Series - Fuji Electric
Contactors and Thermal Overload Relays - FJ Series - Fuji ElectricContactors and Thermal Overload Relays - FJ Series - Fuji Electric
Contactors and Thermal Overload Relays - FJ Series - Fuji Electric
 
Contactors and Thermal Overload Relays - FJ Series - Fuji Electric
Contactors and Thermal Overload Relays - FJ Series - Fuji ElectricContactors and Thermal Overload Relays - FJ Series - Fuji Electric
Contactors and Thermal Overload Relays - FJ Series - Fuji Electric
 
Original NPN Transistor TIP31C TIP31 TO-220F New STMicroelectronics
Original NPN Transistor TIP31C TIP31 TO-220F New STMicroelectronicsOriginal NPN Transistor TIP31C TIP31 TO-220F New STMicroelectronics
Original NPN Transistor TIP31C TIP31 TO-220F New STMicroelectronics
 
בחינת הצללה ברמת המודול לאלקטרוניקת אנרגיה
בחינת הצללה ברמת המודול לאלקטרוניקת אנרגיהבחינת הצללה ברמת המודול לאלקטרוניקת אנרגיה
בחינת הצללה ברמת המודול לאלקטרוניקת אנרגיה
 
Ultrasonic welding questions
Ultrasonic welding questionsUltrasonic welding questions
Ultrasonic welding questions
 
Original Transistor NPN D882 2SD882 3A 40V TO-126 New
Original Transistor NPN D882 2SD882 3A 40V TO-126 NewOriginal Transistor NPN D882 2SD882 3A 40V TO-126 New
Original Transistor NPN D882 2SD882 3A 40V TO-126 New
 
SOI RF Switch for Wireless Sensor Network
SOI RF Switch for Wireless Sensor NetworkSOI RF Switch for Wireless Sensor Network
SOI RF Switch for Wireless Sensor Network
 
Power generation from shoes & utilize it to charge the mobile's, laptop's or ...
Power generation from shoes & utilize it to charge the mobile's, laptop's or ...Power generation from shoes & utilize it to charge the mobile's, laptop's or ...
Power generation from shoes & utilize it to charge the mobile's, laptop's or ...
 
Original NPN SILICON POWER DARLINGTONS Transistor TIP142 10A 100V TO-220F New...
Original NPN SILICON POWER DARLINGTONS Transistor TIP142 10A 100V TO-220F New...Original NPN SILICON POWER DARLINGTONS Transistor TIP142 10A 100V TO-220F New...
Original NPN SILICON POWER DARLINGTONS Transistor TIP142 10A 100V TO-220F New...
 
Catalog ELCB Mitsubishi-Beeteco.com
Catalog ELCB Mitsubishi-Beeteco.comCatalog ELCB Mitsubishi-Beeteco.com
Catalog ELCB Mitsubishi-Beeteco.com
 
Trans Block Show
Trans Block ShowTrans Block Show
Trans Block Show
 
Original Transistor NPN TIP41C TIP41 6A 100V TO-220 New
Original Transistor NPN TIP41C TIP41 6A 100V TO-220 NewOriginal Transistor NPN TIP41C TIP41 6A 100V TO-220 New
Original Transistor NPN TIP41C TIP41 6A 100V TO-220 New
 
Original PNP Transistor A1020-Y A1020 1020 2SA1020 TO-92 New ON semiconductor
Original PNP Transistor A1020-Y A1020 1020 2SA1020 TO-92 New ON semiconductorOriginal PNP Transistor A1020-Y A1020 1020 2SA1020 TO-92 New ON semiconductor
Original PNP Transistor A1020-Y A1020 1020 2SA1020 TO-92 New ON semiconductor
 
CPRI-Power quality report
CPRI-Power quality reportCPRI-Power quality report
CPRI-Power quality report
 
Save energy and protect your facility with voltage optimisation
Save energy and protect your facility with voltage optimisationSave energy and protect your facility with voltage optimisation
Save energy and protect your facility with voltage optimisation
 
Megger Power And Industrial Test Equipment - Cable Fault Locators, High Volta...
Megger Power And Industrial Test Equipment - Cable Fault Locators, High Volta...Megger Power And Industrial Test Equipment - Cable Fault Locators, High Volta...
Megger Power And Industrial Test Equipment - Cable Fault Locators, High Volta...
 
The 7 types of Power Problems
The 7 types of Power ProblemsThe 7 types of Power Problems
The 7 types of Power Problems
 
MY PAPER
MY PAPERMY PAPER
MY PAPER
 
Introduction final a
Introduction final   aIntroduction final   a
Introduction final a
 

More from Painezee Specialist

ZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsPainezee Specialist
 
ZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsPainezee Specialist
 
Neuromove – pathway to recovery
Neuromove – pathway to recoveryNeuromove – pathway to recovery
Neuromove – pathway to recoveryPainezee Specialist
 
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Painezee Specialist
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeePainezee Specialist
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeePainezee Specialist
 
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWaveZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWavePainezee Specialist
 
ZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartPainezee Specialist
 
ZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualPainezee Specialist
 
ZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualPainezee Specialist
 
ZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualPainezee Specialist
 
ZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualPainezee Specialist
 
ZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guideZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guidePainezee Specialist
 
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesPainezee Specialist
 
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementPainezee Specialist
 

More from Painezee Specialist (20)

ZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specifications
 
ZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructions
 
Neuromove – pathway to recovery
Neuromove – pathway to recoveryNeuromove – pathway to recovery
Neuromove – pathway to recovery
 
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
 
ZMPCCM111000.03
ZMPCCM111000.03 ZMPCCM111000.03
ZMPCCM111000.03
 
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWaveZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
 
ZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chart
 
Zywie
ZywieZywie
Zywie
 
ZMPCZM019000.10.01
ZMPCZM019000.10.01ZMPCZM019000.10.01
ZMPCZM019000.10.01
 
ZMPCZM016000.12.07
ZMPCZM016000.12.07ZMPCZM016000.12.07
ZMPCZM016000.12.07
 
ZMPCZM016000.12.06
ZMPCZM016000.12.06ZMPCZM016000.12.06
ZMPCZM016000.12.06
 
ZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manual
 
ZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manual
 
ZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manual
 
ZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's Manual
 
ZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guideZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guide
 
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
 
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
 

Recently uploaded

Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...Nehru place Escorts
 
Glomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxGlomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxDr.Nusrat Tariq
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersBook Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersnarwatsonia7
 
9873777170 Full Enjoy @24/7 Call Girls In North Avenue Delhi Ncr
9873777170 Full Enjoy @24/7 Call Girls In North Avenue Delhi Ncr9873777170 Full Enjoy @24/7 Call Girls In North Avenue Delhi Ncr
9873777170 Full Enjoy @24/7 Call Girls In North Avenue Delhi NcrDelhi Call Girls
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfMedicoseAcademics
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any TimeCall Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Timevijaych2041
 
High Profile Call Girls Kodigehalli - 7001305949 Escorts Service with Real Ph...
High Profile Call Girls Kodigehalli - 7001305949 Escorts Service with Real Ph...High Profile Call Girls Kodigehalli - 7001305949 Escorts Service with Real Ph...
High Profile Call Girls Kodigehalli - 7001305949 Escorts Service with Real Ph...narwatsonia7
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingCall Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingNehru place Escorts
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...rajnisinghkjn
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.ANJALI
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 

Recently uploaded (20)

Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
 
Glomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxGlomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptx
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
 
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersBook Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
 
9873777170 Full Enjoy @24/7 Call Girls In North Avenue Delhi Ncr
9873777170 Full Enjoy @24/7 Call Girls In North Avenue Delhi Ncr9873777170 Full Enjoy @24/7 Call Girls In North Avenue Delhi Ncr
9873777170 Full Enjoy @24/7 Call Girls In North Avenue Delhi Ncr
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any TimeCall Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
 
High Profile Call Girls Kodigehalli - 7001305949 Escorts Service with Real Ph...
High Profile Call Girls Kodigehalli - 7001305949 Escorts Service with Real Ph...High Profile Call Girls Kodigehalli - 7001305949 Escorts Service with Real Ph...
High Profile Call Girls Kodigehalli - 7001305949 Escorts Service with Real Ph...
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingCall Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 

ZMPCHW070000.13.04 US FDA approval letter K 120398

  • 1. 510(k) Summary per 21 CFR 807.92 Page I of4 510(k) U K120398 AUG 162012 5 10(k) SUBMITTER: ESTABLISHMENT REGISTRATION: CONTACT: DATE PREPARED: PROPRIETARY NAME: PANEL: REGULATION NUMBER: CLASSIFICATION: PRODUCT CODES: Hollywog, LLC 2830 Amnicola Highway Chattanooga, TN 37406 3008585473 Michael W. Treas, Chief Compliance Officer August 15, 2012 WiTouch'Pro Neurology CFR Title 21, 882.5890 Class 11 GZJ (Stimulator, Nerve, Transcutaneous, For Pain RelieQ; NYN (Stimulator, Electrical, Transcutaneous, For Arthritis) COMMON NAME: Transcutaneous electrical nerve stimulator for pain relief Description: M The WiTouch T Pro medical device is a Transcutaneous Electrical Nerve Stimulator (TENS) used to reduce the perception of pain by electrically stimulating peripheral nerves across the skin (transcutaneously). The design consists of a battery powered current generator with integral electrodes and replaceable electrical dispersive hydrogel pads (gel-pads). One side of the adhesive gel-pad adhere to the integral electrode, and the other side adhere the device to the healthy intact skin of the patient's back to provide a transcutaneous analgesic electrical stimulus to the painfulI area. Indications for Use: The WiTouh m Pro Transcutaneous Electrical Nerve Stimulator Device is used* for the symptomatic relief and management of chronic intractable back pain and relief of pain of the upper and lower back associated with arthritis. It is also used for adjunctive treatnment for post-surgical and post-trauma acute back pain. Intended Use: The intended use is to provide analgesic electrical stimulus to reduce the perception of back pain by electrically stimulating peripheral nerves across the skin (transcutaneously).
  • 2. 510(k) Summary per 21 CFR 807.92 510(k) # K120398 Page 2of 4 Accessories: The device utilizes hydrogel pads (gel-pads) for achieving the indications for use and intended use. The composition of the gel-pads is common materials found in the electrode industry. The uniqueness of the gel-pads is in the shape. The maximum average power density of the electrodes with the gel-pads applied i less than 0.25 watts per square centimeter of electrode conductive surface area to reduce the risk of thermal burns which is consistent with the referenced predicate devices. 510(k) # Substantially Equivalent Predicate Devices PredicateRegulation, Classification, and Product Code(s) Predicate Proprietary Trade Name K061.650 K061516 Empi SELECT" Staodyn' Max Preset TENS _______________________________ CFR Title 21, Sec. 882.5890, Class 11, GZJ, NYN CER Title 2 1, Sec. 882.5 890, Class 11, GZJ, NYN The predicate devices utilize flexible wires between the electrodes and the electrical stimulus generator; thus, increases the indications for use to the lower back and to body surfaces with greater ranges of motion (e.g., knee, shoulder, elbow, and hip). The intended design of the WiTouch TMPro device limits the application for use to the anatomical sites of the upper and lower back. The design includes carbon rubber electrodes that are intended for reuse and are permanently-affixed to a rigid surface of the electrical stimulus generator. The unique connection makes the electrodes integral to the generator. The connection and shape of the electrodes limit application of the device to the contours of the back. These unique characteristics of the integral electrodes are insignificant as it relates to safety and effectiveness, and is not critical to the intended use between the device and the referenced predicate devices. The referenced predicate devices utilize affixed buttons as the sole method to control the electrical stimulus generator on/off and intensity up/down. The WiTouchTm Pro utilizes an additional method of a wireless remote control radio frequency transceiver to control the electrical stimulus generator on/off and intensity up/down. The transceiver operates in the ISM radio frequency band for wireless medical technology. This uniqueness of controlling the electrical stimulus generator by utilizing a radio frequency transceiver is insignificant as it relates to safety and effectiveness, and is not critical to the intended use between the device and the referenced predicate devices The characteristics of the analgesic electrical stimulus output between the device and the referenced predicate devices are substantially equivalent as it relates to safety and effectiveness.
  • 3. 510(k) Summary per 21 CPU 807.92 Page 3 of 4 510(k) # K120398 Technological characteristics and physical properties of the WiTouch Pro device and the referenced predicate devices: 510(k) Number Device Name and Model K120398 WiTouch'M Pro K061650 EmpiO Select Manufacturer TENS Device Power Source (DC battery) Hollywog Battery 1.SVDC EMPI Battery i.SVDC K(061516 StaodynO Max preset TENS Compex Technologies Battery 1.5VDC (2-Alkaline AAA) (3-Alkaline AAA) (3-Alkaline AAA) 4 4 4 1 Channel Asynchronous 2 Channels Asynchronous 2 Channels Synchronous Yes Yes Yes Yes Yes Yes Yes Yes Yes ______________________________ Number of Output Modes Number of Synchronous or Output Asynchronous? Channels: Software/Firmware/MIicroprocessor Control? Automatic Shut Off? User Override Control? { On/Off St atus? Low Batte ry? Voltage/Current Level? Timer Range (minutes) Weight (lbs., oz,) Indicator Display: Dimensions (in.) [W x H x DJ Housing Material and Construction Waveform (e.g., pulsed monophasic, biphasic) Yes Yes No 30 4.8 oz. w/ batteries included 7.5"(W) x 3.5(H)" x D.7"(D) Silicone & ABS Pulsed biphasic Square sinusoidal) Symmetrical phases? Step-i and Step-3 Phase Duration (include units), (state range, if applicable), (both phases, if asymmetrical) Step-2 Phase Duration (include units), (state range, if applicable), (both phases, if asymmetrical) Maximum Current Density, (mA/cm 2, r.m.s.) Maximum Average Current (average absolute value), MA Maximum Average Power Density, For multiphasic waveforms only: (W/CM ), Yes No No i to 30 5.15 oz. w/ batteries included 2.5"(W) x 5.25"(H) x 1.0"(D) Plastic Pulsed biphasic ____________ ____________ Shape (e.g., rectangular, spike, rectified 2 Yes Yes No i to 30 4.9 oz. w/ batteries included 2.38"(W) x 1.38"(H) x 4.31"(D) Plastic Pulsed biphasic Square ____________ Square ____________ No 12011S No 0 to 40011s at 50% peak amplitude Yes 0 to 300pis 480p 0 to 400115 at 50% peak amplitude 0 to 30li 0.l2mA@500 0 <l0mA @500 0 <l0mA @500 0 1.GmA@500 0 <l0mA @500 0 <l0mA @5000Q 0.00069W/cm2@ 500 a <0.25 W/cm2 at 500 0) <0.25 W/cm2 at 500 0 (using smallest electrode conductive surface area) ___________________________________
  • 4. 510(k) Summary per 21 CFR 807.92 510(k) # K120398 Page 4 of4 Declarations of Conformity The device complies with the following FDA recognized standards: FDA Recognized Number 5-4, IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988 Amendment 1, 1991 -1 l,,Amendment 2, 1995. (General) FDA Recognized Number 5-60, IEC 60601-1-2 Int. I Third Edition/I-SR 01:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, Interpretation Sheet. (General) FDA Recognized Number 5-41, Medical electrical equipment - Part 1-4: General requirements for safety- Collateral standard: Programmable electrical medical systems, edition 1.1I. (General) FDA Recognized Number 17-5, lEG 60601-2-10 1987/Amendment 1 2001, Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators. (Neurology) FDA Recognition Number 2-156: AAMI/AMSI/ ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. (Biocompatibility) FDA Recognition Number 2-153 (Electrodes) ISO 10993-5:2009, Biological evaluation of medical devices - PartS5: Tests for In Vitro cytotoxicity. (Biocompatibility) FDA Recognized Standard 2-173 (Electrodes) Recognition Number 2-173: AAMI / ANSI / ISO 1099310:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. (Biocompatibility) Truthful and Accurate Statement A statement was included in the Premarket Notification attesting to the truthfulness and accuracy of the information provided. Further Information In the event that additional information is required, please contact: Michael W. Treas Chief Compliance Officer Hollywog, LLC 2830 Amnicola Highway Chattanooga, TN 37406 Telephone: (423) 305-7778 ext. 108 Fax: (423) 305-7867 E-mail: mike.treas~hollywog.com
  • 5. DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 Newv Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002 AUG 162012 Hollywog, LLC Mr. Michael Treas Chief Compliance Officer 2830 Amnicola Highway Chattanooga, TN 37406 Re: K120398 TradeIDevice Name: WiTouch Pro Regulation Number: 21 CFR 882.5 890 Regulation Name: Transcutaneous nerve stimulator, for pain relief Regulatory Class: II Product Code: GZJ, NYN Dated: July 10, 2012 Received: July 12, 2012 Dear Mr. Treas: We have reviewed your Section 5 10(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class 11 (Special Controls) or class Ill (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
  • 6. Page 2 - Mr. Michael Treas comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 53 1-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CenterSOffices/CDRH/CDRHOffices/ucm 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.pov/MedicalDevices/SafetLReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.pov/MedicalDevices/ResourcesforYou/Industr/default.htm. Sincerely yours, Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure
  • 7. Indications for Use 510(k) Number: K120398 Device Trade Name: WiTouch® Pro Indications for Use: The WiTouch® Pro Transcutaneous Electrical Nerve Stimulator Device is used for the symptomatic relief and management of chronic intractable back pain and relief of pain of the upper and lower back associated with arthritis. It is also used for adjunctive treatment for post-surgical and post-trauma acute back pain. FPrescription Use (Part 21 CER 801 ____AND/OR] -atD Over-The-Counter Use F(21 _ _ CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)' Concurrence of CDRH, Office of Device Evaluation (ODE) (Division VnOf Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 5l0(k)Nunber.