Toxoplasma
Quantitative IgG
For the Quantitative Measurement of
IgG Antibodies to Toxoplasma gondii
in Human Serum
For use on the IMMULITE
®
and IMMULITE
®
1000 systems
IMMULITE
®
/IMMULITE 1000
®
Toxoplasma Quantitative IgG
2 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
English
Intended Use: For in vitro diagnostic use
with the IMMULITE and IMMULITE 1000
Analyzers — for the quantitative
measurement of IgG antibodies to
Toxoplasma gondii (toxoplasma) in human
serum, as an aid in the determination of
serological status to Toxoplasma gondii.
This kit is not FDA-cleared for use in
the testing (i.e. screening) of blood or
plasma donors.
Catalog Number: LKTXP1 (100 tests),
LKTXP5 (500 tests)
Test Code: TXP Color: Light Gray
CDC Analyte Identifier Code: 6113
CDC Test System Identifier Code: 10159
CLIA Complexity Category: Moderate
Summary and Explanation
Toxoplasma gondii is an obligate
intracellular parasite capable of infecting
most mammals, including humans. The
organism is transmitted through ingestion
of insufficiently cooked meat. Throughout
the world, 1 to 90 percent of populations
may be infected,
7
with 25 to 30 percent of
the adult population infected in the United
States.
3
While toxoplasmosis may be
manifested in several forms, infections are
usually clinically inapparent, and latent
infections usually persist for life.
2
Overt
clinical symptoms are similar to infectious
mononucleosis, with lymphadenopathy,
fever, headache, malaise and sometimes
pneumonia and myocarditis.
6
As with other latent infections, acute
toxoplasma infection can pose a serious
threat to immunocompromised individuals
and newborns who acquire the infection in
utero. Immunosuppressed patients may
develop encephalitis, myocarditis or
pneumonitis.
3
Congenital infections
usually result as a consequence of
asymptomatic acute maternal infection.
This infection can cause premature
delivery, spontaneous abortion or
stillbirth.
4,6
Neonates may manifest
chorioretinitis, hydrocephaly,
microcephaly, cerebral calcification and
psychomotor retardation.
1
The majority of
congenitally infected children will not
exhibit any symptoms until later in life.
4
Management of toxoplasmosis requires
serological monitoring of infected
individuals,
3
as the organism is not readily
available for culture. Quantitative testing
for the presence of toxoplasma IgG can be
useful to determine prior infection and
indicate reactivation of the infection.
Accurate diagnostic information is
important, particularly during pregnancy,
as treatment with spiramycin can reduce
the risk to the fetus.
5
Principle of the Procedure
IMMULITE/IMMULITE 1000 Toxoplasma
Quantitative IgG is a solid-phase,
sequential chemiluminescent
immunoassay.
Incubation Cycles: 2 × 30 minutes.
Specimen Collection
The use of an ultracentrifuge is
recommended to clear lipemic samples.
Samples which are cloudy or have
particulate material should be clarified by
low-speed centrifugation.
Centrifuging serum samples before a
complete clot forms may result in the
presence of fibrin. To prevent erroneous
results due to the presence of fibrin,
ensure that complete clot formation has
taken place prior to centrifugation of
samples. Some samples, particularly
those from patients receiving
anticoagulant therapy, may require
increased clotting time.
Blood collection tubes from different
manufacturers may yield differing values,
depending on materials and additives,
including gel or physical barriers, clot
activators and/or anticoagulants.
IMMULITE/IMMULITE 1000 Toxoplasma
Quantitative IgG has not been tested with
all possible variations of tube types.
Consult the section on Alternate Sample
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 3
Types for details on tubes that have been
tested.
Volume Required: 10 µL of the prediluted
patient sample (1-in-21 dilution). (Sample
cup must contain at least 100 µL more
than the total volume required.)
Storage: 3 days at 2–8°C or
6 months at –20°C.
13
Patient Sample Dilution: Serum samples
must be prediluted 1-in-21 in IgG/IgM
Sample Diluent (LIGZ2). This dilution must
be performed manually e.g. by adding
20 µL pf patient sample to 400 µL of
Sample Diluent (LIGZ2, supplied in the
kit).
Alternatively, the IMMULITE 1000
Windows
®
system performs the dilution on
board, automatically, using LIGW
(provided in the IgG/IgM Sample Diluent
Module, L1KIGW1,5).
Printed results have been corrected for the
1-in-21 dilution.
Warnings and Precautions
For in vitro diagnostic use.
Reagents: Store at 2–8°C. Dispose of in
accordance with applicable laws.
Follow universal precautions, and handle
all components as if capable of
transmitting infectious agents. Source
materials derived from human blood were
tested and found nonreactive for syphilis;
for antibodies to HIV 1 and 2; for hepatitis
B surface antigen; and for antibodies to
hepatitis C.
Sodium azide, at concentrations less than
0.1 g/dL, has been added as a
preservative. On disposal, flush with large
volumes of water to prevent the buildup of
potentially explosive metal azides in lead
and copper plumbing.
Chemiluminescent Substrate: Avoid
contamination and exposure to direct
sunlight. (See insert.)
Water: Use distilled or deionized water.
The bead is coated with inactivated
toxoplasma antigen. However, caution is
advised because of the possible presence
of residual organism when working with, or
disposing of, the materials supplied.
The anti-toxoplasma IgG results
determined for a given specimen with
assays from different manufacturers can
vary due to differences in assay methods
and reagent specificity. Therefore, the
results reported by the laboratory to the
physician should include: "The following
results were obtained with the
IMMULITE/IMMULITE 1000 Toxoplasma
Quantitative IgG EIA. Results obtained
from other manufacturers' assay methods
may not be used interchangeably."
Materials Supplied
Components are a matched set. The
barcode labels are needed for the assay.
Toxoplasma Quantitative IgG Test
Units (LTXP1)
Each barcode-labeled unit contains one
bead coated with inactivated, partially
purified Toxoplasma gondii antigen (RH
strain tachyzoites from mouse
peritoneum). Stable at 2–8°C until
expiration date.
LKTXP1: 100 units. LKTXP5: 500 units.
Allow the Test Unit bags to come to room
temperature before opening. Open by
cutting along the top edge, leaving the
ziplock ridge intact. Reseal the bags to
protect from moisture.
Toxoplasma Quantitative IgG Reagent
Wedges (LTXPA, LTXPB)
With barcodes. LTXPA: 7.5 mL of a
protein-based buffer, with preservative.
LTXPB: 7.5 mL alkaline phosphatase
(bovine calf intestine) conjugated to
monoclonal murine anti-human IgG
antibody in buffer, with preservative. Store
capped and refrigerated: stable at 2–8°C
until expiration date. Recommended
usage is within 30 days after opening
when stored as indicated.
LKTXP1: 1 set. LKTXP5: 5 sets.
Toxoplasma Quantitative IgG Adjustors
(LTXPL, LTXPH)
Two vials (Low and High), 4 mL each, of
human serum with IgG reactive to
toxoplasma, in buffer, with preservative.
(No predilution is required.) Stable at
2–8°C for 14 days after opening, or for
6 months (aliquotted) at –20°C.
LKTXP1: 1 set. LKTXP5: 2 sets.
4 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
Toxoplasma Quantitative IgG Controls
(LTXPC1, LTXPC2, LTXPC3)
Three vials (Negative, Low Positive and
Positive), 2 mL each. LTXPC1 (Negative
Control): One vial containing human
serum with IgG nonreactive to toxoplasma,
with preservative. LTXPC2, LTXPC3 (Low
Positive Control, Positive Control): Two
vials containing human serum with IgG
reactive to toxoplasma, with preservative.
Stable at 2–8°C for 14 days after opening,
or for 6 months (aliquotted) at –20°C.
LKTXP1: 1 set. LKTXP5: 2 sets.
IMMULITE: Toxoplasma IgG controls
require the same manual 1-in-21
predilution as patient samples, with
IgG/IgM Sample Diluent (LIGZ2).
IMMULITE 1000: The IMMULITE 1000
software performs automatic on-board
dilution of patient and control samples.
Enter controls as controls.
Refer to the control insert for
concentration levels.
IgG/IgM Sample Diluent (LIGZ2)
For the manual dilution of patient samples
and controls. 100 mL nonhuman
protein/buffer matrix, with preservative.
Stable at 2–8°C for 30 days after opening,
or 6 months (aliquotted) at –20°C.
LKTXP1: 1 vial.
LKTXP5: 3 vials.
For on-board dilution of patient samples
using IMMULITE 1000 Windows
®
, refer to
the Kit Components Supplied Separately
section.
Kit Components
Supplied Separately
IgG/IgM Sample Diluent Module
(L1KIGW1,5)
Required for the on-board dilution of
patient samples and controls using
IMMULITE 1000 Windows
®
. Sample
Diluent wedges (LIGW) containing
concentrated (2.5X) non-human
protein/buffer matrix, with preservative.
Stable at 2–8°C until expiration date.
L1KIGW1: 2 Wedges.
L1KIGW5: 10 Wedges.
LSUBX: Chemiluminescent Substrate
LPWS2: Probe Wash Module
LKPM: Probe Cleaning Kit
LCHx-y: Sample Cup Holders (barcoded)
LSCP: Sample Cups (disposable)
LSCC: Sample Cup Caps (optional)
DCHS: Dilution Cup Holders (barcoded)
for IMMULITE 1000 Windows
®
on-board
dilutions.
Also Required
Sample transfer pipets, distilled or
deionized water.
Assay Procedure
Note that for optimal performance, it is
important to perform all routine
maintenance procedures as defined in the
IMMULITE or IMMULITE 1000 Operator's
Manual.
See the IMMULITE or IMMULITE 1000
Operator's Manual for: preparation, setup,
dilutions, adjustment, assay and quality
control procedures.
Visually inspect each Test Unit for the
presence of a bead before loading it onto
the system.
Note that both Reagent Wedges A and B
must be loaded on the carousel to run this
assay. For IMMULITE 1000 Windows
®
,
the Sample Diluent Wedge must also be
loaded on the carousel.
Recommended Adjustment Interval:
4 weeks.
Quality Control Samples: The controls
supplied with the kit should be used as
quality control material to monitor assay
performance.
The negative and positive controls are
intended to monitor for substantial reagent
failure. The positive control provided will
not ensure precision of the cut-off.
In addition to the controls provided, users
may wish to run additional controls to
cover the upper range of the assay.
Additional controls may be tested in
accordance with guidelines or
requirements of local, state and/or federal
regulations or accrediting organizations.
12
It is also recommended that known
reactive and nonreactive specimens be
run periodically to assure pipetting
accuracy for the dilution step.
Interpretation of Results
IMMULITE/IMMULITE 1000 Toxoplasma
Quantitative IgG assay is standardized in
terms of the World Health Organization's
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 5
Third International Standard for Anti-
Toxoplasma Serum. The assay's cutoff is
determined from representative positive
and negative samples (confirmed by an
ELISA method) using the ROC (Receiver
Operating Characteristic) technique.
Positive: A result greater than or equal to
8 IU/mL is indicative of a past infection.
Negative: A result less than 6.5 IU/mL is
considered to be negative. Results less
than 6.5 IU/mL should be reported as
Negative only, and no numerical
concentration should be reported.
Indeterminate: A result greater than or
equal to 6.5 IU/mL and less than 8 IU/mL
is considered indeterminate.
Samples greater than 250 IU/mL should
be reported as "greater than 250 IU/mL".
Specimens reporting a result of greater
than 250 IU/mL, when tested at the
required 1-in-21 dilution, may be further
diluted with the IgG/IgM Sample Diluent.
The presence of IgG antibodies to
Toxoplasma gondii is an indication of
previous exposure to the organism. A
single specimen can be used to estimate
the serological status of the individual.
A result of "indeterminate" indicates an
intermediate level of Toxoplasma gondii
antibodies in the sample: the sample
should therefore be retested. Samples
which still test as "indeterminate" should
be examined by an alternate method, or a
second sample should be collected, if
possible, within a reasonable period of
time (e.g. one week).
Any interpretation of the results is not
recommended if the positive or negative
control falls outside the range specified in
the Quality Control section.
The magnitude of the reported IgG level
cannot be correlated to an endpoint titer.
Reports by the laboratory to the physician
should include: "The following results were
obtained with the IMMULITE/IMMULITE
1000 Toxoplasma Quantitative IgG EIA.
Values obtained from other manufacturers'
assay methods may not be used
interchangeably."
Expected Values
Individuals with the toxoplasma organism
may not exhibit detectable levels of IgG
antibody in the early stages of infection.
Levels of IgG antibody to Toxoplasma
gondii begin to rise one or two weeks after
infection. Peak levels are reached in 6 to 8
weeks, then gradually decline over a
period of months or even years. Low titers
are generally detectable for life. The
antibody titer does not correlate with
severity of illness.
2,4,6
The prevalence of IgG antibody to
Toxoplasma gondii will vary with age and
geographical location. Studies were
conducted with specimens from presumed
healthy, asymptomatic subjects from the
northwestern US. The subjects consisted
of 69 individuals with ages ranging from
15 to 64 years. (Samples had been
previously collected and stored frozen at
–20°C). The prevalence of toxoplasma
reactivity, based on the results from the
IMMULITE Toxoplasma Quantitative IgG
assay, was found to be 56% at this
geographical location.
Additional studies were conducted with
specimens from 42 pregnant women from
the same geographical location, with ages
ranging from 16 to 41 years. (Samples
had been previously collected and stored
frozen at –20°C). The prevalence of
toxoplasma reactivity for this specific
population was 5%, based on the results
from the IMMULITE Toxoplasma
Quantitative IgG assay. (See “Expected
Values” graph.)
Consider these limits as guidelines only.
Each laboratory should establish its own
reference ranges.
Limitations
The use of IMMULITE/IMMULITE 1000
Toxoplasma Quantitative IgG to diagnose
recent infection by testing paired sera has
not been validated.
The results of the test must be taken
within the context of the patient's clinical
history, symptomology and other
laboratory findings.
The presence of IgG antibodies in a single
specimen is not sufficient to distinguish
between active and past infection.
Patients suspected of having a primary or
active toxoplasma infection should be
tested for the presence of IgM antibodies
to Toxoplasma gondii.
For the determination of seroconversion
from nonreactive to reactive, a second
6 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
serum sample should be drawn three to
four weeks after the acute, nonreactive
sample, during the convalescent stage of
the infection. The acute phase sample
should be stored and tested in parallel
with the convalescent sample.
Individuals with an acute toxoplasma
infection may not exhibit any detectable
IgG antibodies at the early stage of
infection.
The results in HIV patients, in patients
undergoing immunosuppressive therapy,
or in patients with other disorders leading
to immunosuppression, should be
interpreted with caution.
The performance characteristics of this
assay have not been established for use
with specimens from neonates, cord
blood, pretransplant patients, or body
fluids other than serum, such as urine,
saliva or amniotic fluid.
Heterophilic antibodies in human serum
can react with the immunoglobulins
included in the assay components causing
interference with in vitro immunoassays.
[See Boscato LM, Stuart MC. Heterophilic
antibodies: a problem for all
immunoassays. Clin Chem 1988:34:27-
33.] Samples from patients routinely
exposed to animals or animal serum
products can demonstrate this type of
interference potentially causing an
anomalous result. These reagents have
been formulated to minimize the risk of
interference; however, potential
interactions between rare sera and test
components can occur. For diagnostic
purposes, the results obtained from this
assay should always be used in
combination with the clinical examination,
patient medical history, and other findings.
Performance Data
See Tables and Graphs for data
representative of the assay's performance.
Results are expressed in IU/mL. (Unless
otherwise noted, all were generated on
serum samples collected in tubes without
gel barriers or clot-promoting additives.)
Calibration Range: 5.0 to 250 IU/mL.
Standardized in terms of WHO 3rd IS for
Anti-Toxoplasma Serum, Human.
Analytical Sensitivity: ≤ 5.0 IU/mL.
Precision: Statistics were calculated for
controls and samples processed in 5
replicates each on five separate days.
(See “Intraassay Precision” table and
“Interassay Precision” table.)
Note that the negative control and Sample
1 were consistently below 5 IU/mL (the
lower limit of the assay's calibration range)
for all replicates in every run.
Specificity: The specificity of IMMULITE
Toxoplasma Quantitative IgG assay was
evaluated by testing 50 specimens
positive for cytomegalovirus (CMV),
varicella-zoster virus (VZV), Epstein-Barr
virus (EBV), Mycoplasma pneumoniae
and parvovirus B19. With these
specimens, IMMULITE Toxoplasma
Quantitative IgG and a commercially
available enzyme immunoassay for
Toxoplasma IgG (Kit D) showed 96%
agreement (48/50).
Alternate Sample Type: To assess the
effect of alternate sample types, blood
was collected from 16 volunteers into
plain, heparinized and Becton Dickinson
SST
®
vacutainer tubes. All samples were
assayed by the IMMULITE Toxoplasma
Quantitative IgG procedure, with the
following results.
(Heparin) = 1.02 (Serum) + 0.13 IU/mL
r = 1.00
(SST) = 0.98 (Plain Tubes) – 0.06 IU/mL
r = 1.00
Means:
4.4 IU/mL (Serum)
4.6 IU/mL (Heparin)
4.2 IU/mL (SST)
In another experiment, blood was
collected from 16 laboratory volunteers
into plain and EDTA and vacutainer tubes.
All samples were assayed by the
IMMULITE Toxoplasma Quantitative IgG
procedure, with the following results.
(EDTA) = 0.93 (Serum) – 0.003 IU/mL
r = 1.00
Means:
5.3 IU/mL (EDTA)
5.6 IU/mL (Serum)
Clinical Performance
The clinical performance of the IMMULITE
Toxoplasma Quantitative IgG assay was
studied using a total of 199 serum
specimens, assayed at each of three sites,
two of which were outside the US. Of the
199 specimens (previously collected and
stored frozen at –20°C), 61 were obtained
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 7
from women who presented for obstetric
and gynecological examination, 42 were
obtained from pregnant women, and the
remainder were obtained from patients
with miscellaneous diseases and
conditions (AIDS, anemia, arthritis,
fatigue, headache, lymphoma, meningitis,
pneumonia and others). There were 37
male and 162 female subjects, with ages
ranging from 15 to 68 years.
One of the retrospective studies was
conducted at a university medical center in
the northwestern US. The 199 specimens
were evaluated nonconcurrently with the
IMMULITE Toxoplasma Quantitative IgG
procedure and Kit A, a commercially
available Toxoplasma IgG enzyme-linked
fluorescent immunoassay (ELFA).
IMMULITE Toxoplasma
Quantitative IgG
Kit A Pos Neg
Relative
Sensitivity
Relative
Specificity
Positive 52 2
Negative 0 145
96.3% 100%
Agreement: 99.0%
95% Confidence Limits for Relative Sensitivity
and Specificity, respectively: 87.3% – 99.5% and
97.5% – 100%. (Confidence limits by exact test.)
The two discordant specimens, negative
by IMMULITE but positive by Kit A, were
further evaluated by Kit B, a commercially
available Toxoplasma IgG microparticle
enzyme immunoassay (MEIA), and found
to be positive.
In the second retrospective study,
conducted outside the US, 194 of the 199
specimens (insufficient sample volume for
5 specimens) were evaluated with the
IMMULITE Toxoplasma Quantitative IgG
assay and Kit B.
IMMULITE Toxoplasma
Quantitative IgG
Kit B Pos Neg
Relative
Sensitivity
Relative
Specificity
Positive 46 5
Negative 2 141
90.2% 98.6%
Agreement: 96.4%
95% Confidence Limits for Relative Sensitivity
and Specificity, respectively: 78.6% – 96.7% and
95.0% – 99.8%. (Confidence limits by exact
test.)
Of the seven discordant specimens, three
of the five negative by IMMULITE but
positive by Kit B were negative by Kit A.
The two specimens positive by IMMULITE
but negative by Kit B were positive by
Kit A.
In the third retrospective study, also
conducted outside the US, 160 of the 199
specimens (insufficient sample volume for
39 specimens) were evaluated with the
IMMULITE Toxoplasma Quantitative IgG
assay and Kit C, a commercially available
enzyme immunoassay.
IMMULITE Toxoplasma
Quantitative IgG
Kit C Pos Neg
Relative
Sensitivity
Relative
Specificity
Positive 32 2
Equivocal 0 4
Negative 2 120
94.1% 98.4%
Agreement: 97.4%
95% Confidence Limits for Relative Sensitivity
and Specificity, respectively: 80.3% – 99.3% and
94.2% – 99.8%. (Confidence limits by exact
test).
The two cases positive by the IMMULITE
but negative by Kit C, were positive by
Kit A. Of the two cases negative by
IMMULITE but positive by Kit C, one was
positive by Kit A and the other negative by
Kit A. The four cases negative by
IMMULITE but equivocal by Kit C were all
negative by Kit A. Thus, the IMMULITE
results agree with Kit A in seven of the
eight cases of discordance with Kit C.
References
1) Daffos F. Prenatal management of 746
pregnancies at risk for congenital toxoplasmosis.
N Engl J Med 1988;318: 271-5. 2) Krick JA,
Remington JS. Toxoplasmosis in the adult: an
overview. N Engl J Med 1978;298: 550-3. 3)
Krogstad DJ, et al. Blood and tissue protozoa.
In: Lennette EH, et al, editors. Manual of clinical
microbiology. 4th ed. Washington, D.C.:
American Society for Microbiology, 1985: 612-
30. 4) McCabe RE, Remington JS. Toxoplasma
gondii. In: Mandell, Douglas, Bennett, editors.
Principles and practice of infectious diseases.
2nd ed. New York: John Wiley, 1985. 5) Jeannel
D, et al. What is known about the prevention of
congenital toxoplasmosis? Lancet
1990;336:359-61. 6) Remington JS. McLeod R.
Toxoplasmosis. In: Braude, editor. Microbiology
and infectious diseases. Philadelphia: Saunders,
1981. 7) Walls KW. Toxoplasmosis. In: Balows
A, et al, editors. Laboratory diagnosis of
infectious diseases. New York: Springer-Verlag,
1988: 998-1017. 8) National Committee for
8 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
Clinical Laboratory Standards. Procedures for
the collection of diagnostic blood specimens by
venipuncture; approved standard. 4th ed.
NCCLS Document H3-A4, Wayne, PA: NCCLS,
1998. 9) Baltz ML, Searcy RL. Clinical
significance and advanced serologic diagnosis
of ToRCH infections. Am Clin Lab 1994
(Mar/Apr): 18-23. 10) Wilson M, et al.
Seroepidemiology of toxoplasmosis in the
United States. Abstract presented at the
American Society for Microbiology, May 1994.
11) Approved standard procedures for the
handling and processing of blood specimens,
NCCLS Document H18-A, 1991. 12)
International Quality Control testing: principles
and definitions, NCCLS Document C24-A, 1991.
13) Tietz NW, editor. Clinical guide to laboratory
tests. 3rd ed. Philadelphia: WB Saunders,
1995:358.
Technical Assistance
In the United States, contact DPC's
Technical Services department.
Tel: 800.372.1782 or 973.927.2828
Fax: 973.927.4101. Outside the United
States, contact your National Distributor.
The Quality System of Diagnostic Products
Corporation is registered to ISO 13485:2003.
Tables and Graphs
Intraassay Precision (IU/mL)
Mean1
SD2
CV3
Pos. Control4
18 0.95 5.3%
Neg. Control5
<5 — —
1 <5 — —
2 8.3 0.81 9.8%
3 45 3.9 8.7%
4 83 5.1 6.1%
5 122 14 11%
Interassay Precision (IU/mL)
Mean1
SD2
CV3
Pos. Control4
18 1.3 7.2%
Neg. Control5
<5 — —
1 <5 — —
2 8.3 0.85 10.2%
3 45 6.9 15%
4 83 7.5 9.0%
5 122 16 13%
Expected Values
Reactivity: IML Toxoplasma Quantitative IgG
Asymptomatic Pregnant
0%
20%
40%
60%
80%
100%
Reactive
Non-Reactive
n = 69 n = 42
Deutsch. Intraassay Precision: 1
Mittelwert, 2
S
(Standardabweichung), 3
CV
(Variationskoeffizient), 4
Positiv-Kontrolle,
5
Negativ-Kontrolle. Interassay Precision:
1
Mittelwert, 2
S (Standardabweichung), 3
CV
(Variationskoeffizient), 4
Positiv-Kontrolle
5
Negativ-Kontrolle. Reactivity: IML Toxoplasma
Quantitative IgG: IML Toxoplasma Quantitativ
IgG, Reactive: reaktiv, Non-Reactive: nicht-
reaktiv, Asymptomatic: Asymptomatic, Pregnant:
Schwanger.
Español. Intraassay Precision: 1
Media, 2
DS,
3
CV, 4
Pos. Control, 5
Neg. Control. Interassay
Precision: 1
Media, 2
DS, 3
CV, 4
Pos. Control,
5
Neg. Control. Reactivity: IML Toxoplasma
Quantitative IgG: IML Toxoplasma Quantitative
IgG, Reactive: Reactive, Non-Reactive: Non-
Reactive, Asymptomatic: Asymptomatic,
Pregnant: Pregnant.
Français. Intrraassay Precision: 1
Moyenne,
2
SD, 3
CV, 4
Pos. Control, 5
Neg. Control.
Interassay Precision: 1
Moyenne, 2
SD, 3
CV,
4
Pos. Control, 5
Neg. Control. Reactivity: IML
Toxoplasma Quantitative IgG: IML Toxoplasma
Quantitative IgG, Reactive: Reactive, Non-
Reactive: Non-Reactive, Asymptomatic:
Asymptomatic, Pregnant: Pregnant.
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 9
Italiano. intraassay Precision: 1
Media, 2
SD
(Deviazione Standard), 3
CV (Coefficiente di
Variazione), 4
Pos. Control, 5
Neg. Control.
Interassay Precision: 1
Media, 2
SD (Deviazione
Standard), 3
CV (Coefficiente di Variazione),
4
Pos. Control, 5
Neg. Control. Reactivity: IML
Toxoplasma Quantitative IgG: IML Toxoplasma
Quantitative IgG, Reactive: Reactive, Non-
Reactive: Non-Reactive, Asymptomatic:
Asymptomatic, Pregnant: Pregnant.
Português. Intraassay Precision: 1
Média,
2
Desvio padrão, 3
Coeficiente de variação, 4
Pos.
Control, 5
Neg. Control. Interassay Precision:
1
Média, 2
Desvio padrão, 3
Coeficiente de
variação, 4
Pos. Control, 5
Neg. Control.
Reactivity: IML Toxoplasma Quantitative IgG:
IML Toxoplasma Quantitative IgG, Reactive:
Reactive, Non-Reactive: Non-Reactive,
Asymptomatic: Asymptomatic, Pregnant:
Pregnant.
Deutsch
Toxoplasma IgG quantitativ
Anwendung: Für in-vitro-diagnostische
Tests mit dem Analysegerät IMMULITE
und IMMULITE 1000 zum quantitativen
Nachweis von IgG-Antikörpern gegen
Toxoplasma gondii (Toxoplasma) in
Serum als Hilfestellung zur Bestimmung
des serologischen Status gegen
Toxoplasma gondii.
Dieses Testsystem ist von der US-
amerikanischen Lebensmittel- und
Medikamentenbehörde FDA nicht für
Screening-Tests mit Blut- oder
Plasmaspendern freigegeben.
Artikelnummern: Catalog Number:
LKTXP1 (100 Tests), LKTXP5 (500 Tests)
Testcode: TXP Farbe: hellgrau
Klinische Relevanz
Toxoplasma gondii ist ein obligater
intrazellulärer Parasit, der die meisten
Säugetiere – einschließlich Menschen –
befällt und zu Infektionen führt. Der
Erreger wird durch den Verzehr von
ungenügend gekochtem Fleisch
übertragen. Weltweit sind möglicherweise
zwischen 1 und 90 Prozent aller
Populationen infiziert
7
, in den USA sind es
25–30 % der erwachsenen Bevölkerung.
3
Toxoplasmosen können sich in
verschiedenen Formen manifestieren,
normalerweise sind die Infektionen jedoch
klinisch unauffällig und in ihrer latenten
Form lebenslang präsent.
2
Zu den offenen
klinischen Symptomen zählen infektiöse
Mononukleosen mit Lymphadenopathien,
Fieber, Kopfschmerzen, Unwohlsein und
gelegentlich auch Lungenentzündungen
und Myokarditiden.
6
So wie andere latente Infektionen können
auch akute Toxoplasmosen bei
immungeschwächten Patienten und in
utero infizierten Neugeborenen eine
ernsthafte Bedrohung darstellen. Bei
immunsupprimierten Patienten können
Enzephalitis, Myokarditis oder
Pneumonitis entstehen.
3
Kongenitale
Infektionen sind normalerweise das
Resultat einer asymptomatischen akuten
Infektion der Mutter. Diese Infektionen
könen zu Frühgeburten, Spontanaborten
oder Totgeburten führen.
4,6
Bei
Neugeborenen können Chorioretinitiden,
Hydrozephalie, Mikrozephalie, zerebrale
Kalkablagerungen sowie
psychomotorische Behinderungen
auftreten.
1
Die Mehrzahl der kongenital
infizierten Kinder zeigt erst im späteren
Leben Symptome.
4
Da der Erreger von Toxoplasmosen nicht
ohne weiteres kultiviert werden kann,
müssen die infizierten Patienten
serologisch überwacht werden.
3
Quantitative Bestimmungen von
Toxoplasma-IgG können nützlich sein, um
frühere und reaktivierte Infektionen
anzuzeigen. Genaue diagnostische Daten
sind vor allem bei Schwangerschaften
wichtig, da eine Behandlung mit
Spiramycin das Risiko für den Fetus
herabsetzen kann.
5
Methodik
Der IMMULITE/IMMULITE 1000
Toxoplasma IgG quantitativ ist ein
Festphasen-, sequenzieller
Chemilumineszens-Immunoassay.
Inkubationszyklen: 2 × 30 minuten.
Probengewinnung
Der Einsatz einer Ultrazentrifuge wird zur
Klärung von lipämischen Proben
empfohlen.
10 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
Trübe oder partikelhaltige Proben sollten
bei niedriger Geschwindigkeit zentrifugiert
werden, bis sie klar sind
Die Zentrifugation der Serumproben vor
dem völligen Abschluss der Gerinnung
kann zu Fibringerinnseln führen. Um
fehlerhaften Analysenergebnissen infolge
von Gerinnseln vorzubeugen, ist
sicherzustellen, dass die Gerinnung vor
der Zentrifugation der Proben vollständig
abgeschlossen ist. Insbesondere Proben
von Patienten unter Antikoagulantien-
therapie können eine verlängerte
Gerinnungszeit aufweisen.
Blutentnahmeröhrchen von verschiedenen
Herstellern können differierende Werte
verursachen. Dies hängt von den
verwendeten Materialien und Additiven
(Gel oder physische Trennbarrieren,
Gerinnungsaktivatoren und /oder
Antikoagulantien) ab.
IMMULITE/IMMULITE 1000 Toxoplasma
IgG quantitativ sind nicht mit allen
möglichen Röhrchenvariationen
ausgetestet worden. Details der
getesteten Röhrchenarten sind dem
Kapitel "Alternative Probenarten" zu
entnehmen.
Erforderliche Menge: 10 µl der
vorverdünnten Patientenprobe (1:21
Verdünnen). (Inhalt der Probenschale
muss mindestens 100 µl über der
erforderlichen Gesamtmenge liegen.)
Lagerung: 3 Tage bei 2–8°C oder
6 Monate bei –20°C.
13
Verdünnung der Patientenproben:
Serumproben müssen 1:21 im IgG/IgM
Probenverdünnungsreagenz (LIGZ2)
vorverdünnt werden. Diese Verdünnung
ist manuell durchzuführen, z.B. durch
Zugabe von 20 µl der Patientenprobe in
400 µl Probenverdünnungsreagenz
(LIGZ2, im Kit enthalten).
Alternativ, bei Verwendung des IMMULITE
1000 Windows, erfolgt die Verdünnung
automatisch on-board. Das System
verwendet LIGW, Inhalt des IgG/IgM
Probenverdünnungsreagenz-Moduls
(L1KIGW1,5).
Die ausgedruckten Resultate
entsprechend der 1:21 Verdünnung
korrigiert sind.
Hinweise und
Vorsichtsmaßnahmen
Zur In-vitro-Diagnostik.
Reagenzien: Bei 2–8°C lagern. Unter
Einhaltung der geltenden gesetzlichen
Vorschriften entsorgen.
Die generell geltenden
Vorsichtsmaßnahmen sind einzuhalten
und alle Komponenten als potenziell
infektiös zu behandeln. Alle aus
menschlichem Blut gewonnenen
Materialien wurden auf Syphilis,
Antikörper gegen HIV-1 und HIV-2,
Hepatitis-B-Oberflächenantigen und
Hepatitis-C-Antikörper untersucht und
negativ befundet.
Bestimmten Komponenten wurde
Natriumazid (<0,1 g/dl) hinzugefügt. Um
die Bildung von explosiven Metallaziden in
Blei- und Kupferrohren zu vermeiden,
sollten die Reagenzien nur zusammen mit
großen Wassermengen in die Kanalisation
gespült werden.
Chemilumineszenz-Substrat:
Kontamination und direkte
Sonneneinstrahlung vermeiden. Siehe
Packungsbeilage.
Wasser: Destilliertes oder deionisiertes
Wasser verwenden.
Die Phase ist mit inaktiviertem
Toxoplasma-Antigen beschichtet. Wegen
des potenziellen Vorhandenseins
überlebender Mikroorganismen ist jedoch
beim Arbeiten mit und beim Entsorgen von
den im Lieferumfang enthaltenen
Materialien Vorsicht geboten.
Unterschiede in der jeweiligen Methodik
oder der Spezifität der Reagenzien
könnten jedoch dazu führen, das die mit
Testsystemen von verschiedenen
Herstellern ermittelten Anti-
Toxoplasmose-IgG-Ergebnisse für
dieselben Proben nicht einheitlich sind.
Die vom Labor an den Arzt
weitergegebenen Ergebnisse sollten
daher den folgenden Passus enthalten:
„Die folgenden Ergebnisse wurden mit
dem IMMULITE/IMMULITE 1000-
Testsystem zur quantitativen Bestimmung
von Toxoplasmose-IgG erzielt." Sie sind
nicht mit den Ergebnissen der
Testsysteme anderer Hersteller
austauschbar.
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 11
Im Lieferumfang enthalten
Die Komponenten sind aufeinander
abgestimmt. Die Barcode-Etiketten
werden für den Assay benötigt.
Toxoplasma quantitativ IgG
Testeinheiten (LTXP1)
Jede mit Barcode-Etikette versehene
Einheit enthält eine mit beschichtet mit
inaktiviertem, teilweise gereinigtem
Toxoplasma-gondii-Antigen (Tachyzoiten,
Stamm RH, aus Mausperitoneum).
beschichtete Kugel. Bei 2–8°C bis zum
Ablaufdatum haltbar.
LKTXP1: 100 Testeinheiten.
LKTXP5: 500 Testeinheiten.
Verpackte Testeinheiten vor dem Öffnen
stehen lassen, bis sie Zimmertemperatur
erreicht haben. Oben entlang der Kante
aufschneiden, ohne den Plastikverschluss
zu beschädigen. Verpackungen wieder
dicht verschließen, damit der Inhalt
trocken bleibt.
Toxoplasma quantitativ IgG –
Reagenzbehälter (LTXPA, LTXPB)
Mit Barcodes. LTXPA: 7,5 ml einer
proteinbasierten Pufferlösung (mit
Konservierungsmittel). LTXPB: 7,5 ml mit
alkalischer Phosphatase
(Rinderkalbsdarm) konjugierter
monoklonaler antihumaner IgG
Mausantikörper in Pufferlösung (mit
Konservierungsmittel). Verschlossen und
gekühlt aufbewahren: Bei 2–8°C bis zum
Ablaufdatum haltbar. Bei entsprechender
Lagerung beträgt die empfohlene
Aufbrauchsfrist nach dem Öffnen 30 Tage.
LKTXP1: 1 Set. LKTXP5: 5 Sets.
Toxoplasma quantitativ IgG -
Kalibratoren (LTXPL, LTXPH)
Zwei Fläschchen (niedrig und hoch) mit
4 ml Tosoplasma-reaktivem IgG in einer
Pufferlösung (mit Konservierungsmittel)
(Muss nicht vorverdünnt werden.) 14 Tage
nach dem Öffnen bei 2–8°C haltbar oder
6 Monate bei –20°C (aliquotiert).
LKTXP1: 1 Set. LKTXP5: 2 Sets.
Toxoplasma quantitativ IgG Kontrollen
(LTXPC1, LTXPC2, LTXPC3)
Drei Fläschchen (Negative, Schwachpositiv
und Positiv) jeweils mit 2 ml Kontrollen.
LTXPC1 (Negativkontrolle): Ein
Fläschchen Humanserum mit Toxoplasma-
nichtreaktivem IgG (mit
Konservierungsmittel). LTXPC2, LTXPC3
(Schwachpositivkontrolle, Positiv-
kontrolle): Zwei Fläschchen Humanserum
mit Toxoplasma-reaktivem IgG (mit
Konservierungsmittel). 14 Tage nach dem
Öffnen bei 2–8°C haltbar oder 6 Monate
bei –20°C (aliquotiert).
LKTXP1: 1 Set. LKTXP5: 2 Sets.
IMMULITE: Für die Toxoplasma IgG -
Kontrollen ist die gleiche manuelle 1- in-
21- Vorverdünnung mit der IgG/IgM-
Probenverdünner wie für die
Patientenproben erforderlich (LIGZ2).
IMMULITE 1000: Die IMMULITE 1000
Software führt eine automatische on-
board-Verdünnung der Patienproben und
der Kontrollen durch. Kontrollen als
Kontrollen eingeben. Kontrollen als
Kontrollen eingeben.
Die Konzentrationen entnehmen Sie bitte
der Beilage zu den Kontrollen.
IgG/IgM–Verdünnungspuffer (LIGZ2)
Zum manuellen Verdünnen der
Patientenproben und Kontrollen. 100 ml
Nichthumane Protein/Puffermatrix mit
Konservierungsmittel. 30 Tage nach dem
Öffnen bei 2–8°C oder 6 Monate
bei –20 °C haltbar (aliquotiert).
LKTXP1: 1 Fläschchen.
LKTXP5: 3 Fläschchen.
Für On-board-Verdünnungen mit
IMMULITE 1000 Windows
®
von
Patientenproben ist die Liste der separat
erhältlichen Testsystem-Komponenten
heranzuziehen.
Separat erhältliche Testsystem-
Komponenten
IgG/IgM Probenverdünnungsreagenz-
Modul (L1KIGW1,5)
Erforderlich für die on-board Verdünnung
der Patientenproben und Kontrollen bei
Verwendung des Immulite 1000 Windows.
Probenverdünnungsreagenz-Behälter
(LIGW) enthalten konzentrierte (2,5fach)
nicht-humane Protein/Puffer Matrix mit
Konservierungsmittel. Haltbar bei 2-8°C
bis zum Verfallsdatum.
L1KIGW1: 2 Behälter.
L1KIGW5: 10 Behälter.
LSUBX: Chemilumineszenz-
Substratmodul
LPWS2: Waschmodul
12 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
LKPM: Reinigungsmodul
LCHx-y: Barcodierte Probenträger
LSCP: Einmal-Probenröhrchen
LSCC: Deckel für Probenröhrchen
DCHS: Verdünnungsgefäßträger
(barcodiert) für die IMMULITE 1000
Windows on-board-Verdünnung.
Ebenfalls benötigt:
Transferpipetten für die Proben;
destilliertes bzw. deionisiertes Wasser.
Testdurchführung
Für eine optimale Funktion des Gerätes ist
unbedingt zu beachten, dass die
Wartungen, wie im IMMULITE oder
IMMULITE 1000-Handbuch beschrieben,
regelmäßig durchgeführt werden.
Das Handbuch für das IMMULITE bzw.
IMMULITE 1000 enthält die Anweisungen
für: Vorbereitung, Geräteeinstellungen,
Verdünnungen, Kalibrierung,
Testdurchführung und Qualitätskontrollen.
Überprüfen Sie jedes Testeinheit auf das
Vorhandensein der Polystyrol-Kugel vor
dem Einsetzen in das Gerät.
Beachten Sie, dass zur Durchführung
dieses Tests beide Reagenzbehälter
(A und B) auf das Karussell geladen
werden müssen. Beim IMMULITE 1000
Windows muss der
Probenverdünnungsreagenz-Behälter
ebenfalls auf dem Reagenzienkarussell
platziert sein.
Empfohlenes Kalibrationsintervall:
4 Wochen.
Proben zur Qualitätskontrolle: Die im
Lieferumfang enthaltenen Kontrolle(n)
verwenden.
Die Negativ- und Positivkontrollen dienen
zur Überwachung der Reagenzien auf
wesentliche Mängel. Die gelieferte
Positivkontrolle eignet sich nicht dazu, die
Präzision des Cutoff sicherzustellen.
Zusätzlich zu den gelieferten Kontrollen
können zur Abdeckung des höheren
Messbereiches weitere Kontrollen
eingesetzt werden. Entsprechend den
Richtlinien bzw. Vorschriften von
regionalen bzw. staatlichen Behörden
oder zertifizierenden Stellen können
weitere Kontrollen getestet werden.
12
Ferner empfiehlt es sich, regelmäßig
Testansätze mit bekannten reaktiven und
nichtreaktiven Proben durchzuführen, um
die Pipettiergenauigkeit für den
Verdünnungsschritt zu gewährleisten.
Interpretation der Ergebnisse
Das quantitative IMMULITE/IMMULITE
1000-Testsystem zur quantitativen
Bestimmung von Toxoplasma-IgG ist
entsprechend der Dritten Internationalen
WHO-Norm für Anti-Toxoplasma-Serum
standardisiert. Der Cutoff des Testsystems
wird mit Hilfe der ROC-Technik (Receiver
Operating Characteristic) aus
repräsentativen positiven und negativen
Proben (verifiziert mittels ELISA) ermittelt.
Positiv: Ein Ergebnis von ≥ 8 IU/ml weist
auf eine vergangene Infektion hin.
Negativ: Ein Ergebnis von unter 6,5 IU/ml
gilt als negativ. Ergebnisse von unter
6,5 IU/ml sollten lediglich als negativ
ausgewiesen werden, ohne dabei die
gemessene Konzentration numerisch
darzustellen.
Grenzwertig: Ein Ergebnis von
≥ 6,5 IU/ml und < 8 IU/ml gilt als
grenzwertig.
Proben mit mehr als 250 IU/ml sollten
dementsprechend (d.h. „höher als
250 IU/ml“) ausgewiesen werden. Proben,
die bei der erforderlichen Autoverdünnung
von 1:21 mehr als 250 IU/ml aufweisen,
können mit dem IgG/IgM-
Probenverdünner weiterverdünnt werden.
Sind IgG-Antikörper gegen Toxoplasma
gondii vorhanden, so deutet dies auf eine
frühere Exposition gegen den Erreger hin.
Einzelproben eignen sich ausschließlich
zur Bewertung des serologischen Status
der betreffenden Person.
Ein „indeterminate“ // „unentschiedenes“
Ergebnis weist auf einen Zwischenwert bei
den Antikörpern gegen Toxoplasma gondii
in der Probe hin: Die Probe sollte daher
erneut getestet werden. Ergebnis bringen,
sind entweder mit einer alternativen
Methode zu testen, oder, sofern dies
möglich ist, innerhalb eines vernünftigen
Zeitraums (z.B. eine Woche) mit einer neu
entnommenen zweiten Probe zu
ergänzen.
Von einer Interpretation der Ergebnisse
wird abgeraten, wenn die Positiv- oder
Negativkontrollen außerhalb des im
Abschnitt „Quality Control“ //
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 13
„Qualitätskontrolle“ angegebenen
Bereichs liegen.
Die Größenordnung der ausgewiesenen
IgG-Konzentration kann nicht mit einem
Endtiter korreliert werden.
Die vom Labor an den Arzt
weitergegebenen Ergebnisse sollten
daher den folgenden Passus enthalten:
„Die folgenden Ergebnisse wurden mit
dem IMMULITE/IMMULITE 1000-
Testsystem zur quantitativen Bestimmung
von Toxoplasmose-IgG erzielt." Sie sind
nicht mit den Ergebnissen der
Testsysteme anderer Hersteller
austauschbar.
Referenzwerte
Im Frühstadium einer Infektion mit dem
Toxoplasma-Erreger sind möglicherweise
keine nachweisbaren Konzentrationen an
IgG-Antikörpern vorhanden. Die
Konzentrationen an IgG-Antikörpern
gegen Toxoplasma gondii steigen
erstmals ein bis zwei Wochen nach der
Primärinfektion an. Die Spitzenwerte
werden innerhalb von 6–8 Wochen
erreicht und fallen dann über Monate oder
sogar Jahre hinaus allmählich wieder ab.
Niedrige Titer bleiben generell für den
Rest des Lebens nachweisbar. Es besteht
keine Korrelation zwischen Antikörpertiter
und dem Schweregrad der Erkrankung.
2,4,6
Die Prävalenz von IgG-Antikörpern für
Toxoplasma gondii kann abhängig vom
Alter und der geographischen Lage
variieren. Es wurden mit gesunden,
asymptomatischen Patienten aus dem
Nordwesten der USA Studien
durchgeführt. Die Patienten setzten sich
aus 69 Personen im Alter zwischen 15
und 64 Jahren zusammen. (Die Proben
wurden vorher gesammelt und bei –20°C
eingefroren gelagert). In dieser
geographischen Gegend wurde eine
Prävalenz der Toxoplasmosereaktivität
von 56% gefunden.
Weiterhin wurden Studien mit
Patientenseren von schwangeren Frauen
im Alter zwischen 16 und 41 Jahren aus
der gleichen geographischen Region
durchgeführt. (Die Proben wurden vorher
gesammelt und bei –20°C eingefroren
gelagert). Die Prävalenz der
Toxoplasmosereaktivität dieser speziellen
Bevölkerungsgruppe lag bei 5%,
basierend auf den Resultaten des
IMMULITE Toxoplasma quantitativem IgG
Test. (siehe grafik „Grenzen der
Methode“)
Betrachten Sie diese Grenzwerte nur als
Richtlinien. Jedes Labor sollte eigene
Referenzbereiche ermitteln.
Grenzen der Methode
Der Einsatz des IMMULITE/IMMULITE
1000 Testsystems zur quantitativen
Bestimmung von Toxoplasma-IgG ist nicht
validiert, was die Diagnose kürzlicher
Infektionen mittels gepaarten Seren
betrifft.
Die Testergebnisse sind vor dem
Hintergrund der klinischen Anamnese, der
Beschwerden des Patienten sowie
weiterer Laborbefunde zu bewerten.
Das Vorhandensein von IgG-Antikörpern
in Einzelproben reicht nicht aus, um
zwischen einer aktiven und einer
vergangenen Infektion zu unterscheiden.
Patienten mit Verdacht auf eine Primär-
bzw. aktive Infektion sollten auf das
Vorhandensein von IgM-Antikörpern
gegen das Toxoplasma gondii untersucht
werden.
Zur Bestimmung einer Serokonversion
von nichtreaktiv zu reaktiv sollte während
der Rekonvaleszenzphase drei bis vier
Wochen nach der akuten nichtreaktiven
Probe eine zweite Serumprobe
genommen werden. Die Probe aus der
Akutphase sollte gelagert und parallel zur
Rekonvaleszenzprobe getestet werden.
Die Ergebnisse bei HIV-Patienten sowie
behandlungs- oder krankheitsbedingt
immunsupprimierten Patienten sollten
zurückhaltend interpretiert werden.
Im Frühstadium einer akuten
Toxoplasmose sind möglicherweise keine
nachweisbaren IgG-Antikörper vorhanden.
Die Eckdaten dieses Testsystems wurden
nicht für den Gebrauch mit Proben von
Neugeborenen, Nabelschnurblut,
Patienten mit bevorstehender
Organverpflanzung und nicht für andere
Körperflüssigkeiten als Serum wie etwa
Harn, Speichel oder Fruchtwasser
etabliert.
Heterophile Antikörper in Humanseren
können mit Immunglobulinen aus den
Assaykomponenten reagieren und
Interferenzerscheinungen innerhalb des in
14 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
vitro Immunoassays verursachen. (Clin.
Chem. 1988:34:27-33) Proben von
Patienten, die häufig mit Tier- bzw.
Tierserumprodukten zu tun haben, können
die erwähnten Interferenzen verursachen
und zu anomalen Resultaten führen. Die
verwendeten Reagenzien sind so
konzipiert, dass das Risiko einer
Interferenz mit den zu messenden Proben
minimiert ist. Dennoch können potentiell
Interaktionen zwischen seltenen Seren
und den Testkomponenten auftreten. Zu
diagnostischen Zwecken sollten die mit
dem Assay erhaltenen Ergebnisse immer
in Kombination mit der klinischen
Untersuchung, der Patientenanamnese
und anderen Befunden gesehen werden
Leistungsdaten
Repräsentative Daten zum Test
entnehmen Sie bitte den Tabellen und
graphischen Darstellungen im englischen
Teil dieser Anleitung. Die Ergebnisse sind
in IU/ml angegeben. (Sofern nicht anders
angegeben, wurden hierfür Serumproben
in Röhrchen ohne Geltrennung und
Gerinnungshilfen eingesetzt.)
Messbereich: 5,0 – 250 IU/ml.
Standardisiert nach der Dritten
Internationalen WHO-Norm für humane
Anti-Toxoplasma-Seren.
Analytische Sensitivität: ≤ 5,0 IU/ml
Präzision: Statisik aus einem einzelnen
Testansatz mit 5 Proben verschiedener
IgG-Konzentrationen in 5-facher
Bestimmung am gleichen Tag. (Siehe
Tabelle „Intraassay / Interassay -
Precision“).
Die Negativ-Kontrolle sowie Probe 1 zeigt
durchgehend Werte unter 5 IU/ml.
Spezifität: Die Spezifität des IMMULITE
Toxoplasma Quantitativ IgG Tests wurde
mit 50 Proben, die positiv für Cytomegalie
(CMV), Varizelle zoster (VZV), Epstein-
Barr-Virus (EBV), Mycoplasma
pneumoniae und Parvovirus B19 waren,
evaluiert. Diese Proben zeigten mit dem
IMMULITE Toxoplasma Quantitativ IgG
Test eine 96%ige Übereinstimmung
(48/50) mit einem kommerziell erhältlichen
Toxoplasma IgG Enzymimmunotest.
Alternativer Probentyp: Um die
Auswirkungen von verschiedenen
Probenarten zu untersuchen, wurde Blut
von 16 Freiwilligen in Röhrchen ohne
Additiva, in Heparin- und Becton
Dickinson SST
®
Vacutainer-Rörchen
gesammelt. Alle Proben wurden mit dem
Toxoplasma IgG Verfahren für den
IMMULITE getestet, mit der folgenden
Ergebnisse.
(Heparin) = 1,02 (Serum) + 0,13 IU/ml
r = 1,00
(SST) = 0,98 (einfachen Röhrchen) – 0,06 IU/ml
r = 1,00
Mittelwerte:
4,4 IU/ml (Serum)
4,6 IU/ml (Heparin)
4,2 IU/ml (SST)
In einer weiteren Studie wurde 16
Freiwilligen Blut entnommen. Als
Abnahmegefäße wurden Serumröhrchen
ohne Zusätze, Heparin-, und EDTA
Vakutainer röhrchen verwendet. Alle
Proben wurden mit dem Toxoplasma IgG
Verfahren für den IMMULITE getestet, mit
der folgenden Ergebnisse.
(EDTA) = 0,93 (Serum) – 0,003 IU/ml
r = 1,00
Mittelwerte:
5,3 IU/ml (EDTA)
5,6 IU/ml (Serum)
Klinische Studien
Die klinischen Leistungsdaten des
IMMULITE Toxoplasma Quantitativ IgG
Test wurdem mit 199 Serumproben aus
drei verschiedenen Regionen, von denen
zwei außerhalb der USA stammten,
geprüft. Die 199 Patientenproben (die
Proben wurden vorher gesammelt und bei
–20°C eingefroren gelagert) wurden von
Frauen erhalten, die aus derzeitigen
Entbindungs- und gynäkologischen
Untersuchungen stammten. 42 Proben
wurden von schwangeren Frauen
erhalten, und die restlichen Proben
stammten von Patienten mit
unterschiedlichen Krankheiten und
Gegebenheiten (AIDS, Anämie, Arthritis,
Müdigkeitssyndrom, Kopfschmerzen,
Lymphom, Meningitis, Lungenentzündung
und andere Krankheiten). Es handelte
sich um 37 männliche und 162 weibliche
Patienten im Alter von 15 bis 68 Jahren.
Eine der retrospektiven Studien wurde an
einer Medizinischen Universität im
Nordwesten der USA durchgeführt. Die
199 Blutproben wurden mit dem
IMMULITE Toxoplasma Quantitativ IgG
Test und einem kommerziell erhältlichen
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 15
Toxoplasma IgG Enzymfluoresz
Enzimmunoassay (ELFA) evaluiert.
IMMULITE Toxoplasma
Quantitativ IgG
Kit A Positiv Negativ
Relative
Sensitivität
Relative
Spezifität
Positiv 52 2
Negativ 0 145
96,3% 100%
Übereinstimmung: 99,0%
95% Vertrauensbereich für die relative
Sensitivität und Spezifität (resp.: 87,3% - 99,5%
und 97,5% - 100%).
Die zwei diskrepanten Blutproben, negativ
im IMMULITE, aber positiv im Kit A, wurde
desweiteren mit einem kommerziell
erhältlichen Toxoplasma IgG
Mikropartikelenzymimmunotest (MEIA)
evaluiert und als positiv gemessen.
In der zweiten retrospektiven Studie, die
außerhalb der USA durchgeführt wurde,
wurden von 199 Proben 194 (5 Proben mit
nicht ausreichendem Probenvolumen) mit
dem IMMULITE Toxoplasma Quantitativ
IgG-Test und Kit B evaluiert.
IMMULITE Toxoplasma
Quantitativ IgG
Kit B Positiv Negativ
Relative
Sensitivität
Relative
Spezifität
Positiv 46 5
Negativ 2 141
90,2% 98,6%
Übereinstimmung: 96,4%
95% Vertrauensbereich für die relative
Sensititvät und Spezifität, resp.: 78,6% - 96,7%
und 95% - 99,8%.
Von den 7 diskrepanten Blutproben waren
3 der 5 negativ gefundenen Proben im
IMMULITE und im Kit B positiv, negativ im
Kit A. Zwei der positiv gemessenen
Blutproben im IMMULITE, aber negativ
gemessenen im Kit B, waren positiv im
Kit A.
In der dritten retrospektiven Studie, die
ebenfalls außerhalb der USA durchgeführt
wurde, wurden von 199 Proben 160 (39
Blutproben mit nicht ausreichendem
Probenvolumen) mit dem IMMULITE
Toxoplasma IgG-Test und einem
kommerziell erhältlichen
Enzymimmunoassay Kit C, evaluiert.
IMMULITE Toxoplasma
Quantitativ IgG
Kit C Positiv Negativ
Relative
Sensitivität
Relative
Spezifität
Positiv 32 2
grenzwertig 0 4
Negativ 2 120
94,1% 98,4%
Übereinstimmung: 97,4%
95% Vertrauensbereich für die relative
Sensitivität und Spezifität, resp.: 80,3% – 99,3%
und 94,2% – 99,8%.
In zwei Fällen mit positivem IMMULITE
und negativem Kit C Ergebnis, waren
ebenfalls positiv in Kit A. In den zwei
Fällen, die negativ im IMMULITE und
positiv in Kit C waren, war eine Probe
positiv und eine negativ im Kit A. In vier
Fällen mit negativem IMMULITE und
grenzwertigen Ergebnis in Kit C, waren
alle Proben negativ in Kit A. Demnach
stimmen die IMMULITE Ergebnisse in
sieben von acht Fällen mit Kit A überein,
die diskrepant im Kit C liefen.
Anwendungsberatung
Bei Rückfragen wenden Sie sich bitte an
Ihre DPC Niederlassung.
Das Qualitätssystem der Diagnostic Products
Corporation ist nach ISO 13485:2003 registriert.
Español
Toxoplasma Cuantitativo IgG
Utilidad del análisis: Para su uso en el
diagnóstico in vitro con los analizadores
IMMULITE e IMMULITE 1000 — para la
medición cuantitativa de anticuerpos IgG
contra Toxoplasma gondii (toxoplasma),
como una ayuda en la determinación del
estado serológico para Toxoplasma
gondii.
Este kit no tiene autorización de la
FDA para su uso en pruebas
(detección precoz) de donantes de
sangre o plasma.
Números de Catálogo:
LKTXP1 (100 tests), LKTXP5 (500 tests)
Código del Test: TXP Color: Gris claro
16 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
Resumen y Explicación
del Test
Toxoplasma gondii es un parásito
intracelular obligado, capaz de infectar a
la mayoría de los mamíferos, incluyendo
al humano. El organismo se transmite por
ingestión de carne poco cocinada. Por
todo el mundo, del 1 al 90% de la
población puede ser infectada,
7
con un 25
– 30 por ciento de la población adulta de
Estados Unidos infectada.
3
Mientras que
la toxoplasmosis puede manifestarse de
diversas formas, las infecciones son
normalmente clínicamente inaparentes, y
las infecciones latentes persisten
normalmente durante toda la vida.
2
Los
síntomas clínicos son similares a los de
una infección por mononucleosis, con
linfadenopatía, fiebre, dolor de cabeza,
malestar general y algunas veces
neumomía y miocarditis.
6
Al igual que otras infecciones latentes, la
infección aguda por toxoplasma puede
presentar una serie amenaza para los
individuos inmunocomprometidos y los
recién nacidos, quienes adquieren la
infección en el utero. Los individuos
inmunosuprimidos pueden desarrollar
encefalitis, miocarditis o neumonitis.
3
Las
infecciones congénitas normalmente
resultan como consecuencia de una
infección aguda asintomática en la madre.
Esta infección puede causar: parto
prematuro, aborto espontáneo o el
nacimiento sin vida del feto.
4,6
Los
neonatos pueden presentar coriorretinitis,
hidrocefalia, microcefalía, calcificación
cerebral y retraso psicomotor.
1
La mayoría
de los niños infectados congénitamente
no mostrarán ningún síntoma hasta más
tarde.
4
El seguimiento de la toxoplasmosis
requiere el control serológico de los
individuos infectados,
3
ya que el cultivo de
este organismo no es posible. Un análisis
cuantitativo para estudiar la presencia de
IgG contra toxoplasma puede ser útil para
determinar una infección anterior e indicar
la reactivación de la infección. Una
información precisa del diagnóstico es
importante, especialmente durante el
embarazo, puesto que el tratamiento con
espiramicina puede reducir los riesgos
para el feto.
5
Principio del análisis
IMMULITE/IMMULITE 1000 Toxoplasma
Cuantitativa IgG es un inmunoanálisis
quimioluminiscente en fase sólida
secuencial.
Ciclos de incubación: 2 × 30 minutos.
Recogida de la muestra
Se recomienda el uso de una
ultracentrífuga para aclarar las muestras
lipémicas.
Las muestras que estén turbias o tengan
un material particular deben aclararse
mediante una centrifugación a baja
velocidad.
La centrifugacion de las muestras de
suero antes de que se forme el coagulo
puede ocasionar la presencia de fibrina.
Para evitar resultados erroneos debidos a
la presencia de fibrina, asegurarse que se
ha formado el coagulo completamente
antes de centrifugar las muestras.
Algunas muestras, particularmente
aquellas de pacientes sometidos a terapia
anticoagulante, pueden requerir mayor
tiempo de coagulacion.
Los tubos para recoger sangre de
distintos fabricantes pueden producir
valores diferentes, dependiendo del
material del tubo y de los aditivos,
incluyendo barreras de gel o barreras
físicas, activadores de la coagulación y/o
anticoagulantes. El IgG Cuantitativo para
Toxoplasma IMMULITE/IMMULITE 1000
no ha sido analizado con todos los
distintos tipos de tubos. Para obtener
detalles sobre los tipos tubos que se han
analizado, consulte la sección de Tipos de
Muestras Alternativos.
Volumen requerido: 10 µl de la muestra
de paciente prediluida (1 por 21 dilución).
(El recipiente de la muestre debe
contener, como mínimo, 100 µl más que
el volumen total requerido).
Conservación: 3 días a 2–8°C, o
6 meses a –20°C.
13
Dilución de Muestras de Pacientes: Las
muestras de suero deben ser prediluídas
1-en-21 con Diluyente de Muestra IgG/IgM
(LIGZ2). Esta dilución debe ser realizada
manualmente, p.e. añadiendo 20 µl de
muestra del paciente con 400 µl de
Diluyente de Muestra (LIGZ2,
suministrados en el kit).
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 17
Como alternativa, utilizando IMMULITE
1000 Windows
®
, la dilución la realiza el
sistema automaticamente en el equipo
usando LIGW, incluido en el Modulo de
Diluyente de Muestra IgG/IgM
(L1KIGW 1,5).
Los resultados imprimidos han sido
corregidos para una dilución de 1-en-21.
Advertencias y Precauciones
Para uso diagnóstico in vitro.
Reactivos: Mantener a 2–8°C. Desechar
de acuerdo con las normas aplicables.
Siga las precauciones universales y
manipule todos los componentes como si
fueran capaces de transmitir agentes
infecciosos. Los materiales derivados de
sangre humana han sido analizados y son
negativos para sífilis; para anticuerpos
frente al HIV 1 y 2; para el antígeno de
superficie de hepatitis B y para los
anticuerpos de hepatitis C.
Se ha usado Azida sodica, en
concentraciones menores de 0,1 g/dl,
como conservante. Para su eliminacion,
lavar con grandes cantidades de agua
para evitar la constitucion de residuos de
azidas metalicas, potencialmente
explosivas, en las cañerías de cobre y
plomo.
Sustrato quimioluminiscente: evite la
contaminación y exposición a la luz
directa del sol. (Ver el prospecto.)
Agua: Usar agua destilada o desionizada.
La bola está recubierta con antígeno
inactivado de toxoplasma. Sin embargo,
se recomienda precaución ante la posible
presencia de organismo residual cuando
trabajamos o deshechamos los materiales
suministrados.
La concentración de IgG contra
Toxoplasma en una muestra dada,
determinada mediante ensayos de
distintos fabricantes, puede variar debido
a diferencias en los métodos de los
ensayos y a la especificidad del reactivo.
Por lo tanto, en el informe de resultados
del laboratorio para el médico debe
incluirse: "Los siguientes resultados se
obtuvieron con EIA IgG Cuantitativo
contra Toxoplasma IMMULITE/IMMULITE
1000. No es posible intercambiar los
resultados obtenidos por métodos de
ensayo de otros fabricantes".
Materiales Suministrados
Los componentes representan un juego
completo. Las etiquetas incluidas en la
caja son necesarias para el ensayo.
Unidades de análisis de Toxoplasma
Cuantitativo IgG (LTXP1)
Cada unidad etiquetada con código de
barras contiene una bola recubierta de
antígeno inactivado, parcialmente
purificado de Toxoplasma gondii
(taquizoitos de cepa RH de peritoneo de
ratón). Estable a 2–8°C hasta la fecha de
caducidad.
LKTXP1: 100 unidades.
LKTXP5: 500 unidades.
Espere a que las bolsas de las unidades
de análisis alcancen la temperatura
ambiente antes de abrirlas. Ábralas
cortando por el extremo superior, dejando
el borde del cierre de cremallera intacto.
Vuelva a cerrar las bolsas
herméticamente para protegerlas de la
humedad.
Viales de reactivo de Toxoplasma
Cuantitativo IgG (LTXPA, LTXPB)
Con códigos de barras. LTXPA: 7,5 ml de
una matriz de proteína en solución
tampón, con conservante. LTXPB: 7,5 ml
de fosfatasa alcalina (de intestino de
ternera) conjugada a anticuerpo
monoclonal murino anti-IgG humana, en
una solución tampón, con conservante.
Guardar tapado y refrigerado: estable a
2–8°C hasta la fecha de caducidad. Se
recomienda utilizarlo antes de que pasen
30 días después de abrirlo cuando se
guarda según lo indicado.
LKTXP1: 1 juego. LKTXP5: 5 juegos.
Ajustadores de Toxoplasma
Cuantitativo IgG (LTXPL, LTXPH)
Dos viales (bajo y alto) de cada uno con
4 ml de suero humano con IgG reactivo
para Toxoplasma, en solución tampón,
con conservante. (No es necesaria
ninguna dilución previa.) Estable a 2–8°C
durante 14 días después de abrise, o
hasta 6 meses (alicuotados) a –20°C.
LKTXP1: 1 juego. LKTXP5: 2 juegos.
Controles de Toxoplasma Cuantitativo
IgG (LTXPC1, LTXPC2, LTXPC3)
Tres viales (Negativo, Bajo Positivo y
Positivo) de cada uno con 2 ml. LTXPC1
(Control negativo): Un vial que contiene
18 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
suero humano con IgG no reactivo para
toxoplasma, con conservante. LTXPC2,
LTXPC3 (Control bajo positivo, Control
Positivo): contienen suero humano con
IgG reactivo para toxoplasma, con
conservante. Estable a 2–8°C durante
14 días después de abrise, o hasta
6 meses (alicuotados) a –20°C.
LKTXP1: 1 juego. LKTXP5: 2 juegos.
IMMULITE: Los controles de Toxoplasma
IgG requieren la misma predilución
manual 1-en-21 que las muestras de
pacientes, con la diluyente de muestra
para IgG e IgM (LIGZ2).
IMMULITE 1000: El software del
IMMULITE 1000 realiza una dilución
automática en el instrumento de las
muestras de pacientes y de los controles.
Introduzca los controles como controles.
Consulte los niveles de concentración en
el prospecto del control.
Diluyente de IgG/IgM (LIGZ2)
Para la dilución manual de muestras de
pacientes y controles. 100 ml matriz
proteica no humana en solución tampón,
con conservante. Estable a 2–8°C durante
30 días después de abrise o estable a
–20°C durante 6 meses (alicuotados).
LKTXP1: 1 vial.
LKTXP5: 3 viales.
Para diluciones de muestras de pacientes
dentro del equipo usando IMMULITE 1000
Windows
®
, remitirse a la sección del Kit
de Componentes Suministrados Aparte.
Componentes del kit que se
suministran por separado
Módulo de Diluyente de Muestra
IgG/IgM (L1KIGW1,5)
Requerido para la dilución de muestras de
pacientes y controles en el equipo
utilizando IMMULITE 1000 Windows
®
. Los
Viales (LIGW) de Diluyente de Muestras
contienen una matriz concentrada de
proteína/buffer (2.5X) no-humana, con
conservante. Estable a 2-8°C hasta la
fecha de caducidad.
L1KIGW1: 2 Viales. L1KIGW5: 10 Viales.
LSUBX: Substrato quimioluminiscente
LPWS2: Lavado de sonda
LKPM: Kit de limpieza de sonda
LCHx-y: Soportes de recipientes de
muestras (con códigos de barras)
LSCP: Recipientes de muestras
(desechables)
LSCC: Tapas para los recipientes de
muestras (opcionales)
DCHS: Cubetas de dilución (con código
de barras) para diluciones en el equipo
(IMMULITE 1000 Windows
®
).
También necesario
Pipetas de transferencia de muestras;
agua destilada o desionizada.
Ensayo
Aviso: para obtener el funcionamiento
óptimo, es importante realizar todos los
procedimientos del mantenimiento general
según lo definido en el manual del
operador de IMMULITE o IMMULITE
1000.
Ver el Manual del Operador del
IMMULITE o IMMULITE 1000 para:
preparación, procesamiento, diluciones,
ajuste, procedimientos de ensayo y
control de calidad.
Inspeccionar visulamente cada unidad de
rección para asegurarse de que hay una
bola antes de introducirla en el Sistema.
Tenga en cuenta que hay que cargar los
viales de reactivos A y B en el carrusel
para ejecutar este ensayo. Para el
IMMULITE 1000 Windows
®
, el Vial del
Diluyente de la Muestra también debe
estar cargado en el carrusel.
Intervalo de ajuste recomendado:
4 semanas.
Muestras de Control de Calidad: Los
controles provistos con el kit deberán
usarse como control de calidad para
monitorear el rendimiento del ensayo.
Los controles positivo y negativo están
diseñados para controlar un fallo
importante del reactivo. El control positivo
proporcionado no asegurará la precísión
del valor de corte.
Además de los controles provistos, los
usuarios pueden desear correr controles
adicionales para cubrir el intervalo
superior del ensayo. Pueden analizarse
otros controles según las directrices o
requisitos de las normas locales, estatales
o federales o de las organizaciones
acreditadas.
12
También se recomienda analizar
periódicamente muestras conocidas,
reactivas o no reactivas, para asegurar la
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 19
precisión del pipeteado para el paso de
dilución.
Interpretación de los
Resultados
El ensayo IgG Cuantitativo para
Toxoplasma IMMULITE/IMMULITE 1000
está estandarizado en términos del Tercer
Estándar Internacional para Suero de
Anti-Toxoplasma de la Organización
Mundial de la Salud. El valor de corte del
ensayo se determinó utilizando muestras
positivas y negativas representativas
(confirmadas por el método ELISA)
utilizando la técnica ROC (Curva de
rendimiento diagnóstico).
Positivo: Un resultado superior o igual a
8 IU/ml es indicativo de una infección
pasada.
Negativo: Un resultado inferior a
6,5 IU/ml es considerado negativo.
Resultados inferiores a 6,5 IU/ml deben
comunicarse solamente como Negativos,
y no debe comunicarse la concentración.
Indeterminado: Un resultado superior o
igual a 6,5 IU/ml e inferior a 8 IU/ml se
considera indeterminado.
Las muestras con valores superiores a
250 IU/ml deben comunicarse como
“superiores a 250 IU/ml". Las muestras
que dieron resultados superiores a
250 IU/ml, cuando se examinan con la
autodilución necesaria 1-en-21, pueden
necesitar diluciones adicionales con el
diluyente para muestras IgG/IgM.
La presencia de anticuerpos IgG contra
Toxoplasma gondii es indicativo de una
exposición previa al organismo. Puede
utilizarse sólo una única muestra para
estimar el estado serológico del individuo.
Un resultado “indeterminado” indica un
nivel intermedio de anticuerpos contra
Toxoplasma gondii en la muestra: la
muestra debe volverse a analizarse. Las
muestras que vuelvan a dar un resultado
“indeterminado” deben examinarse por un
método alternativo, o debe recogerse una
segunda muestra, si es posible, en un
plazo de tiempo razonable (p.e. una
semana).
No se recomienda interpretar los
resultados si el control positivo o negativo
cae fuera del margen especificado en la
Sección Control de Calidad.
La magnitud del nivel de IgG comunicado
no puede relacionarse con un título final.
Por lo tanto, en el informe de resultados
del laboratorio para el médico debe
incluirse: "Los siguientes resultados se
obtuvieron con EIA IgG Cuantitativo
contra Toxoplasma IMMULITE/IMMULITE
1000." No es posible intercambiar los
resultados obtenidos por métodos de
ensayo de otros fabricantes.
Valores Esperados
Individuos con el organismo toxoplasma
pueden no mostrar niveles de anticuerpo
IgG detectables en las fases tempranas
de la infección. Los niveles de IgG para
Toxoplasma gondii empiezan a elevarse 1
o 2 semanas después de la infección. Los
niveles máximos se alcanzan en 6 – 8
semanas, después descienden
gradualmente en los meses y años
posteriores. Se detectan títulos bajos
generalmente durante toda la vida. El
título de anticuerpo no guarda relación
con la severidad de la enfermedad.
2,4,6
La preponderancia de anticuerpos IgG
anti-Toxoplasma gondii variará con la
edad y la zona geográfica. Se condujeron
estudios con muestras tomadas de
sujetos presumiblemente sanos y
asintomáticos del noroeste de los EE.UU.
Este grupo de personas se compuso de
69 individuos de 15 a 64 años de edad.
(Las muestras fueron previamente
recolectadas y guardadas congeladas a –
20ºC). Se encontró que la preponderancia
de reactividad al toxoplasma, basada en
los resultados del ensayo cuantitativo IgG
anti -Toxoplasma IMMULITE, fue de 56%
en esta zona geográfica.
Se realizaron estudios adicionales con
muestras provenientes de 42 mujeres
embarazadas de 16 a 41 años de edad de
la misma zona geográfica. (Las muestras
fueron previamente recolectadas y
guardadas congeladas a –20º C). La
preponderancia de reactividad al
toxoplasma para esta población
específica fue del 5%, basándose en los
resultados del ensayo cuantitativo IgG anti
-Toxoplasma IMMULITE. (Ver el gráfico
de “Expected Values”),
Estos límites han de considerarse sólo
como una guía. Cada Laboratorio deberá
establecer sus propios rangos de
referencia.
20 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
Limitaciones
El uso de IgG Cuantitativo Toxoplasma
IMMULITE/IMMULITE 1000 para
diagnosticar una infección reciente
analizando parejas de sueros no se ha
validado.
Los resultados del análisis deben
contemplarse en el contexto del historial
clínico de los pacientes, de su
sintomatología y de los demás hallazgos
del laboratorio.
La presencia de anticuerpos IgG en una
sola muestra no es suficiente para
distinguir entre una infección activa y una
antigua. Los pacientes de los que se
sospeche que padecen una infección
primaria o activa deben analizarse para
detectar la presencia de anticuerpos IgM
para Toxoplasma gondii.
Para determinar la seroconversión de no
reactivo a reactivo, debe recogerse una
segunda muestra de suero tres o cuatro
semanas después de la infección,
muestra no reactiva, durante la etapa de
convalecencia de la infección. La muestra
de la fase aguda debe almacenarse y
analizarse paralelamente a la muestra de
la fase de convalecencia.
Los individuos con infección aguda por
toxoplasma puede que no presenten
niveles detectables de anticuerpos IgG en
la fase temprana de la infección.
Los resultados en pacientes infectados
con el virus VIH, pacientes bajo una
terapia inmunosupresiva, o en pacientes
con otros desórdenes que causen
inmunosupresión, deben interpretarse con
precaución.
No se han establecido las características
de funcionamiento del ensayo para
utilizarlo con muestras de neonatos,
sangre del cordón umbilical, pacientes
pretrasplante, u otros fluidos distintos del
suero, tales como orina, saliva o líquido
amniótico.
Los anticuerpos heterofílicos en el suero
humano pueden reaccionar con las
inmunoglobulinas de los componentes del
ensayo provocando interferencias con los
inmunoanálisis in vitro. [Ver Boscato LM,
Stuart MC. Heterophilic antibodies: a
problem for all immunoassays. Clin Chem
1988:34:27-33.] Las muestras de los
pacientes que frecuentemente están
expuestos a animales o a productos
séricos animales pueden presentar este
tipo de interferencia que potencialmente
ocasione un resultado anómalo. Estos
reactivos han sido formulados para
minimizar el riesgo de interferencia, no
obstante, pueden darse interacciones
anómalas entre sueros conflictivos y los
componentes del ensayo. Con fines de
diagnóstico, los resultados obtenidos con
este ensayo siempre deben ser usados en
combinación con el examen clínico, la
historia médica del paciente y cualquier
otro dato clínico relevante.
Características Analíticas
Para ver resultados representativos de las
cualidades del ensayo ver las tablas y los
gráficos. Los resultados se expresan
en IU/ml. (A no ser que se indique lo
contrario, todos los resultados fueron
generados en muestras de suero
recogidas en tubos sin geles o
activadores de la coagulación).
Intervalo de calibración: 5,0 – 250 IU/ml.
Normalizado en términos del tercer
estándar WHO IS para Suero Humano
Anti-Toxoplasma.
Sensibilidad: ≤ 5,0 IU/ml
Precisión: Se han calculado datos
estadísticos para las muestras a partir de
los resultados de 5 replicados, cada uno
en cinco días separados. (Ver las tablas
de "Intraassay Precision" y “Interassay
Precision”).
Note que los controles negativos de la
Muestra 1 siempre se encontraron por
debajo de 5 IU/ml (el límite inferior del
intervalo de calibración del ensayo) para
todos los replicados de cada corrida.
Especificidad: Se evaluó la especificidad
del ensayo cuantitativo IgG anti-
Toxoplasma IMMULITE analizando 50
muestras positivas para citomegalovirus
(CMV), virus de varicella-zoster (VZV),
virus Epstein-Barr (EBV), Mycoplasma
penumoniae y parvovirus B19. En estas
muestras, el ensayo cuantitativo IgG anti-
toxoplasma IMMULITE y un
inmunoensayo enzimático para IgG anti-
Toxoplasma comercialmente disponible
(Kit D) mostraron 96% de concordancia
(48/50).
Tipo de Muestra Alternativa: para
evaluar el efecto de los diferentes tipos de
muestras alternativos, se recogió sangre
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 21
de 16 voluntarios en tubos normales,
tubos con Heparina y tubos vacutainer
SST
®
de Becton Dickinson. Todas las
muestras se ensayo de IgG Cuantitativo
para Toxoplasma de IMMULITE,
obteniendo los siguientes resultados.
(Heparina) = 1,02 (Suero) + 0,13 IU/ml
r = 1,00
(SST) = 0,98 (tubos simples) – 0,06 IU/ml
r = 1,00
Medias:
4,4 IU/ml (Suero)
4,6 IU/ml (Heparina)
4,2 IU/ml (SST)
En otro estudio, la sangre de 16
voluntarios se recogió en tubos normales,
y EDTA vacutainer. Todas las muestras
se ensayo de IgG Cuantitativo para
Toxoplasma de IMMULITE, obteniendo
los siguientes resultados.
(EDTA) = 0,93 (Suero) – 0,003 IU/ml
r = 1,00
Medias:
5,3 IU/ml (EDTA)
5,6 IU/ml (Suero)
Rendimiento Clínico
Se estudió el rendimiento clínico del
ensayo cuantitativo IgG anti-Toxoplasma
IMMULITE, usando un total de 199
muestras de suero. Cada una de estas
muestras fue analizada en tres sitios, dos
de los cuales se encuentran fuera de los
EE.UU. De las 199 muestras (previamente
recolectadas y guardadas congeladas a
–20°C ), 61 se obtuvieron de mujeres que
se presentaron para un examen obstétrico
y ginecológico, 42 se obtuvieron de
mujeres embarazadas y el resto se obtuvo
de pacientes con distintas enfermedades
y afecciones (SIDA, anemia, arteritis,
fatiga, dolor de cabeza, linfoma,
meningitis, neumonía y otros). El grupo
consistió de 37 hombres y 162 mujeres de
15 a 68 años de edad.
Uno de los estudios retrospectivos fue
conducido en el centro médico de una
universidad en el noroeste de los EE.UU.
Las 199 muestras fueron evaluadas en
forma no simultánea con el procedimiento
cuantitativo IgG anti-Toxoplasma
IMMULITE y el Kit A, un ensayo de
inmunofluorescencia ligado a una enzima
para IgG anti -Toxoplasma (ELFA)
comercialmente disponible.
Ensayo Cuantitativo para IgG anti-
Toxoplasma IMMULITE
Kit A Pos Neg
Sensibilidad
relativa
Especificidad
relativaa
Positivo 52 2
Negativo 0 145
96,3% 100%
Concordancia: 99,0%
95% de límites de confianza para la sensibilidad
y la especificidad relativas: 87,3% – 99,5% y
97,5% – 100%. (Límites de confianza mediante
análisis exacto.)
Las dos muestras discordantes, negativas
para IMMULITE pero positivas para el Kit
A, fueron evaluadas con el Kit B, un
inmunoensayo de micropartículas para
IgG anti-Toxoplasma (MEIA)
comercialmente disponible, el cual dio un
resultado positivo.
En el segundo estudio retrospectivo que
se realizó fuera de los EE.UU., se
evaluaron194 de las 199 muestras (5
muestras tuvieron insuficiente volumen de
muestra) con el ensayo cuantitativo IgG
anti -Toxoplasma IMMULITE y el Kit B.
Ensayo Cuantitativo IgG anti-
Toxoplasma IMMULITE
Kit B Pos Neg
Sensibilidad
relativa
Especificidad
relativaa
Positivo 46 5
Negativo 2 141
90,2% 98,6%
Concordancia: 96,4%
95% de límites de confianza para la sensibilidad
y la especificidad relativas: 78,6% – 96,7% y
95,0% – 99,8%. (Límites de confianza mediante
análisis exacto.)
De las siete muestras discordantes, tres
de las cinco que dieron negativas con
IMMULITE pero positivas con el Kit B
dieron negativas con el Kit A. Las dos
muestras que dieron positivas con
IMMULITE pero negativas con el Kit B
dieron positivas con el Kit A.
En el tercer estudio retrospectivo, también
realizado fuera de los EE.UU., 160 de las
199 muestras (39 muestras tuvieron
insuficiente volumen de muestra) fueron
analizadas con el ensayo cuantitativo IgG
anti-Toxoplasma IMMULITE y el Kit C, un
inmunoensayo enzimático comercialmente
disponible.
22 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
Ensayo Cuantitativo IgG anti-
Toxoplasma IMMULITE
Kit C Pos Neg
Relative
Sensitivity
Relative
Specificity
Positivo 32 2
Ambiguo 0 4
Negativo 2 120
94,1% 98,4%
Concordancia: 97,4%
95% de límites de confianza para la sensibilidad
y la especificidad relativas: 80,3% – 99,3% y
94,2% – 99,8%. (Límites de confianza mediante
análisis exacto).
Los dos casos positivos para el ensayo
IMMULITE pero negativos para el Kit C
dieron positivos para el Kit A. De los dos
casos negativos para el ensayo
IMMULITE pero positivos para el Kit C,
uno dio positivo y el otro dio negativo con
el Kit A. Los cuatro casos negativos para
el ensayo IMMULITE, pero ambiguos para
el Kit C, dieron todos negativos con el Kit
A. Por lo tanto, los resultados de
IMMULITE concuerdan con los del Kit A
en siete de los ocho casos en que no hay
concordancia con el Kit C.
Asistencia técnica
Contáctese con su Distribuidor Nacional.
El Sistema de Calidad de Diagnostic Products
Corporation está registrado para la
ISO 13485:2003.
Français
IMMULITE/IMMULITE 1000
Toxoplasmose quantitative IgG
Domaine d'utilisation : Dosage
quantitatif des IgG spécifiques dirigées
contre Toxoplasma gondii dans le sérum.
Réservé à un usage diagnostique in vitro
avec l'Analyseur IMMULITE et de
IMMULITE 1000, ce test constitue une
aide à la détermination du statut
sérologique vis-à-vis de Toxoplasma
gondii.
Ce coffret n'est pas habilité auprès de
la Food and Drug Administration
américaine pour le dépistage des
donneurs de sang ou de plasma.
Ce réactif est enregistré auprès de
l'AFSSAPS.
Référence catalogue :
LKTXP1 (100 tests), LKTXP5 (500 tests)
Code produit : TXP
Code couleur : gris clair
Introduction
Toxoplasma gondii est un parasite
intracellulaire capable d'infecter la plupart
des mammifères, y compris l'homme. Le
parasite se transmet par l'ingestion de
viande insuffisamment cuite. Dans le
monde, 1 à 90 % des populations est
susceptible d'être infecté,
7
25 à 30 % de la
population adulte étant infecté au Etats-
Unis.
3
Pour la France, 30 à 60 % de la
population est infectée. Même si la
toxoplasmose peut se manifester sous
plusieurs formes, les individus infectés
restent généralement asymptomatiques et
une infection latente persiste
habituellement toute la vie.
2
Les
symptômes cliniques visibles sont
similaires à ceux de la mononucléose
infectieuse, avec une lymphadénopathie,
de la fièvre, des céphalées, une sensation
de malaise et, parfois, une pneumonie et
une myocardite.
6
Comme toute infection latente, une
infection aiguë par Toxoplasma gondii
peut représenter un risque sérieux pour
un individu immunodéprimé ou pour un
nouveau-né infecté in utero. Les patients
immunodéprimés infectés sont
susceptibles de développer une
encéphalite, une myocardite ou une
pneumopathie inflammatoire.
3
Les
infections congénitales proviennent
habituellement d'une infection maternelle
aiguë asymptomatique. Cette infection
peut se traduire par une naissance
prématurée, un avortement spontané ou
la naissance d'un enfant mort-né.
4,6
Le
nouveau-né peut présenter une
choriorétinite, une hydrocéphalie, une
microcéphalie, des calcifications intra-
cérébrales et un retard psychomoteur.
1
La
plupart des enfants infectés in utero sont
asymptomatiques dans les premiers
temps de vie.
4
Dans la mesure où le parasite n'est pas
facile à mettre en culture, la prise en
charge de la toxoplasmose demande un
contrôle sérologique régulier des individus
infectés.
3
Un dosage quantitatif d'IgG
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 23
spécifiques dirigées contre Toxoplasma
gondii peut être utile pour mettre en
évidence une infection antérieure et
révéler une réactivation de l'infection. Un
diagnostic précis est important, en
particulier au cours de la grossesse,
puisqu'un traitement par la spiramycine
pourra réduire les risques encourus par le
fœtus.
5
Principe du test
IMMULITE/IMMULITE 1000
Toxoplasmose IgG quantitatif est un
immunodosage chimiluminescent
séquentiel, en phase solide.
Cycles d'incubation : 2 × 30 minutes.
Recueil des échantillons
Il est recommandé de clarifier les
échantillons hyperlipémiques par
ultracentrifugation.
Les échantillons troubles ou présentant
des particules en suspension devront être
clarifiés par centrifugation à vitesse
réduite.
La centrifugation des échantillons
sériques avant la formation complète du
caillot peut entraîner la présence de
fibrine. Pour éviter les résultats erronés
dûs à la présence de fibrine, s'assurer de
la formation complète du caillot avant de
centrifuger les échantillons. Certains
échantillons, en particulier ceux provenant
de patients sous anti-coagulants, peuvent
nécessiter un temps plus long pour la
formation du caillot.
Des tubes pour prélèvements sanguins
provenant de fabricants différents peuvent
donner des résultats différents, selon les
matériaux et additifs utilisés, y compris
gels ou barrières physiques, activateurs
de la coagulation et/ou anticoagulants. Le
coffret Toxoplasmose IgG quantitatif
IMMULITE/IMMULITE 1000 n'a pas été
testé sur tous les types de tubes
possibles. Veuillez consulter le chapitre
intitulé Autres Types d'Échantillons pour
plus de renseignements sur les tubes qui
ont été évalués.
Volume nécessaire : 10 µl d 'échantillon
pré-dilué (dilution au 1:21). (L'unité-
échantillon doit contenir au moins 100 µl
de plus que le volume total nécessaire.)
Conservation : 3 jours à +2°C/+8°C ou
6 mois à –20°C.
13
Dilution des échantillons de patients :
Les échantillons doivent être dilués au
1/21
ème
avec du diluant échantillon
IgG/IgM (LIGZ2). Cette dilution doit être
faite manuellement. Exemple : ajouter
20 µl d'échantillon à 400 µl de diluant
échantillon (LIGZ2) fournis avec le coffret.
Le système IMMULITE 1000 Windows
®
réalise également les dilutions
automatiquement, en utilisant le diluant
échantillon IgG/IgM (LIGW) fourni dans le
coffret (L1KIGW1,5).
Les résultats imprimés ont été corrigés en
tenant compte d'une dilution au 1/21
ème
.
Précautions d'emploi
Réservé à un usage diagnostique in vitro.
Réactifs : conserver les réactifs à +2/
+8 °C. Eliminer les déchets conformément
à la réglementation en vigueur.
Respecter les précautions d'emploi et
manipuler tous les composants du coffret
comme des produits potentiellement
infectieux. Les réactifs dérivés de produits
humains et utilisés dans ce coffret ont subi
un test sérologique pour la Syphilis et des
tests de dépistage pour les anticorps anti-
VIH1 et 2, anti-HCV et pour l'antigène de
surface de l'hépatite B, qui se sont tous
avérés négatifs.
De l'azide de sodium à des concentrations
inférieures à 0,1 g/dl a été ajouté comme
conservateur ; lors de l'élimination,
l'évacuer avec de grandes quantités d'eau
pour éviter une accumulation d'azides
métalliques explosifs dans les
canalisations.
Substrat chimiluminescent : éviter la
contamination et l'exposition directe au
soleil. (Voir notice).
Eau : utiliser de l'eau distillée ou
désionisée.
La bille est revêtue d'antigène de
Toxoplasma gondii inactivé. Cependant,
des parasites ayant pu subsister, la
prudence est recommandée lors de
l'utilisation ou de l'élimination des produits
fournis.
Les résultats d'IgG dirigées contre
Toxoplasma gondii obtenus sur un
échantillon donné, avec les tests de
24 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
différents fabricants, pourront présenter
une disparité en raison de différences de
méthodologie et de spécificité des réactifs.
Par conséquent, les résultats
communiqués au médecin par le
laboratoire devront comporter la mention
suivante : « Les résultats suivant ont été
obtenus avec le dosage
IMMULITE/IMMULITE 1000
Toxoplasmose IgG quantitatif. Les
résultats obtenus avec des procédés de
dosage d'autres fabricants ne doivent pas
y être substitués. »
Matériel fourni
Les composants de la trousse ne peuvent
être utilisés que conjointement. Les
étiquettes à l'intérieur du coffret sont
nécessaires au dosage.
Tests unitaires Toxoplasmose
quantitative IgG (LTXP1)
Avec code-barre. Chaque unité-test
contient une bille revêtue d'un antigène de
Toxoplasma gondii inactivé, partiellement
purifié (tachyzoïtes de souche RH, issus
de péritoine de souris). Stable à
+2°C/+8°C jusqu'à la date de péremption.
LKTXP1 : 100 unités.
LKTXP5 : 500 unités.
Porter les sachets à température ambiante
avant d'ouvrir. Ouvrir le sachet avec des
ciseaux en préservant le dispositif de
fermeture. Refermer les sachets pour les
protéger de l'humidité.
Cartouches de réactif Toxoplasmose
quantitative IgG (LTXPA, LTXPB)
Avec code-barres. LTXPA : 7,5 ml de
tampon à base de protéines, avec
conservateur. LTXPB : 7,5 ml d'anticorps
monoclonal murin anti-IgG humaines
marqué à la phosphatase alcaline dans un
tampon avec conservateur. Conserver
bouchés et réfrigérés : stables à
+2°C/+8°C jusqu'à la date de péremption.
A utiliser de préférence dans les 30 jours
qui suivent l'ouverture, si les
recommandations de stockage sont
respectées.
LKTXP1 : 1 jeu. LKTXP5 : 5 jeux.
Ajusteurs Toxoplasmose quantitative
IgG (LTXPL, LTXPH)
2 flacons (haut et bas), 4 ml chacun,
contenant des IgG anti-Toxoplasma gondii
dans du sérum humain avec tampon avec
conservateur. (Il n’est pas nécessaire de
prédiluer.) Stable à +2°C/+8 °C pendant
14 jours après ouverture ou 6 mois
(aliquoté) à –20 °C.
LKTXP1 : 1 jeu. LKTXP5 : 2 jeux.
Contrôles Toxoplasmose quantitative
IgG (LTXPC1, LTXPC2, LTXPC3)
Trois flacons (Négatif, Positif Faible et
Positif), 2 ml chacun. LTXPC1 (Contrôle
négatif) : un flacon de sérum humain ne
contenant pas d'IgG dirigées contre
Toxoplasma gondii, avec conservateur.
LTXPC2, LTXPC3 (Contrôle positif
faible, Contrôle positif) : deux flacons
contenant des IgG dirigées contre
Toxoplasma gondii dans du sérum humain
avec conservateur. Stable à +2°C/+8°C
pendant 14 jours après ouverture ou
6 mois (aliquoté) à –20°C.
LKTXP1 : 1 jeu. LKTXP5 : 2 jeux.
IMMULITE : Les contrôles Toxoplasmose
IgG nécessite la même prédilution
manuelle 1-dans-21 que les échantillons
de patients, avec diluant échantillon IgG/
IgM (LIGZ2).
IMMULITE 1000 : Le logiciel de
l'IMMULITE 1000 réalise
automatiquement les dilutions à bord des
échantillons de patients et de contrôles.
Saisir les contrôles comme des contrôles.
Se reporter à la fiche technique du
contrôle pour les concentrations.
Diluant échantillon IgG/IgM (LIGZ2)
Pour la dilution manuelle des échantillons
de patients et des contrôles. 100 ml
matrice tampon/protéines non-humaines
avec conservateur. Stable à +2°C/+8°C
pendant 30 jours après ouverture ou à –
20°C pendant 6 mois (aliquoté).
LKTXP1 : 1 flacon.
LKTXP5 : 3 flacons.
Pour la dilution à bord automatique des
échantillons de patients, avec IMMULITE
1000 Windows
®
, se référer au châpitre
« Composants du Coffret fournis
séparément ».
Composants du coffret fournis
séparément
Diluant échantillon IgG/IgM
(L1KIGW1,5)
Requis pour la dilution à bord des
échantillons de patients et des contrôles
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 25
avec IMMULITE 1000 Windows
®
. Les
flacons de diluant échantillon (LIGW)
contiennent une matrice protéïque non-
humaine dans un tampon concentré
(2.5X) avec un conservateur. Stable à 2–
8°C jusqu'à la date d'expiration.
L1KIGW1 : 2 flacons.
L1KIGW5 : 10 flacons.
LSUBX : Substrat chimiluminescent
LPWSM : Solution de lavage
LKPM : Coffret de décontamination de
l'aiguille de prélèvement
LCHx-y : portoirs à cupules réactionnelles
(avec code-barre)
LSCP : cupules réactionnelles (à usage
unique)
LSCC : bouchons pour cupules
réactionnelles (optionnel)
DCHS : Support pour unités de dilution
automatisée par l'IMMULITE 1000
Windows
®
.
Egalement requis
Pipettes pour le transfert des échantillons ;
eau distillée ou désionisée.
Protocole de dosage
Noter que pour des performances
optimales, il est important de réaliser
toutes les procédures de maintenance de
routine selon les instructions du Manuel
d'Utilisation de l'IMMULITE ou de
l'IMMULITE 1000.
Voir le manuel d'utilisation de l'IMMULITE
ou de l'IMMULITE 1000 pour la
préparation, le démarrage du système, les
ajustements, le dosage et les procédures
de contrôle de qualité.
Vérifier visuellement que chaque Unité-
Test contient bien une bille avant de la
charger dans l'automate.
Noter que les deux flacons de réactif A et
B doivent être placés sur le carrousel pour
réaliser ce dosage. Pour l'IMMULITE 1000
Windows
®
, le flacon contenant le diluant
échantillon doit également être chargé sur
le carrousel.
Intervalle d'ajustement recommandé :
4 semaines.
Echantillons pour le contrôle de qualité :
Les contrôles fournis dans le coffret
doivent utilisés pour le contrôle de qualité
des performances du dosage.
Les témoins positifs et négatif sont conçus
pour détecter un défaut majeur du réactif.
Le témoin positif fourni ne pourra pas
garantir la précision du seuil.
En plus des contrôles fournis, les
utilisateurs peuvent souhaiter utiliser des
contrôles supplémentaires pour les
valeurs élevées du domaine de mesure du
dosages. Ces contrôle supplémentaires
pourront être testés, conformément aux
recommandations ou aux directives des
autorités locales ou nationales.
12
Il est également recommandé de tester
périodiquement des échantillons positifs
ou négatifs connus, pour contrôler
l'exactitude du pipetage lors de l'étape de
dilution.
Interprétations des résultats
Le dosage IMMULITE/IMMULITE 1000
Toxoplasmose IgG quantitatif est
standardisé par rapport au Troisième
Standard International de l'OMS pour les
anticorps sériques anti-toxoplasmose. Le
seuil du dosage est déterminé par courbe
ROC à partir d'échantillons positifs et
négatifs représentatifs (confirmé par
dosage ELISA).
Positif : un résultat supérieur ou égal à
8 UI/ml indique une infection passée.
Négatif : un résultat inférieur à 6,5 UI/ml
est considéré comme négatif. (Les
résultats inférieurs à 6,5 UI/ml doivent être
rendus comme Négatif sans mentionner
de concentration numérique).
Indéterminé : un résultat supérieur ou
égal à 6,5 UI/ml et inférieur à 8 UI/ml est
considéré comme indéterminé.
Une concentration supérieure à 250 UI/ml
devra être rapportée comme « supérieure
à 250 UI/ml ». Un échantillon dont la
concentration, obtenue sur une dilution au
1:21 par l'appareil, est supérieure à
250 UI/ml, devra être dilué d'un facteur
supplémentaire avec le diluant échantillon
IgG/IgM.
La présence d'anticorps IgG anti-
Toxoplasma gondii est révélatrice d'un
contact antérieur avec le parasite. Un seul
échantillon peut uniquement servir à
évaluer le statut sérologique de l'individu.
Un résultat « indéterminé » indique des
taux d'anticorps anti-Toxoplasma gondii
intermédiaires : l'échantillon devra donc
être dosé de nouveau. Les échantillons
qui restent alors « indéterminés » devront
26 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
être testés par une autre méthode, ou un
autre prélèvement devra être obtenu – si
possible – dans un délai raisonnable (une
semaine, par exemple).
Toute interprétation de résultats est
déconseillée si les contrôles positif ou
négatif donne un résultat situé en dehors
de l'intervalle de référence défini à la
section « Contrôle de qualité ».
L'importance du taux d'IgG rapporté n'est
pas corrélée à un titrage en point final.
Les résultats communiqués au médecin
par le laboratoire devront comporter la
mention suivante : « Les résultats suivant
ont été obtenus avec le dosage
IMMULITE/IMMULITE 1000
Toxoplasmose IgG quantitatif. Les
résultats obtenus avec des procédés de
dosage d'autres fabricants ne doivent pas
y être substitués. »
Valeurs de référence
Il est possible que les IgG spécifiques ne
soient pas détectables dans les phases
précoces de l'infection par Toxoplasma
gondii. En cas de suspicion d'infection au
stade précoce, il faudra impérativement
effectuer un dosage IgM anti-Toxoplasma
gondii. Les taux d'IgG anti-Toxoplasma
gondii commencent à augmenter une à
deux semaines après la contamination. La
réponse anticorps atteint son maximum en
6 à 8 semaines, puis diminue
progressivement sur plusieurs mois, voire
plusieurs années. Le titre d'anticorps n'est
pas corrélé à la gravité de la maladie.
2,4,6
La prévalence des anticorps anti-
Toxoplasma gondii varie avec l'âge et la
localisation géographique. Des études ont
été menées sur des échantillons
provenant de sujets asymptomatiques en
bonne santé apparente, du nord-ouest
des Etats-Unis. Les sujets, au nombre de
69, étaient âgés de 15 à 64 ans. (Les
échantillons avaient été recueillis et
congelés à –20 °C avant d'être dosés). La
prévalence de la séropositivité pour
Toxoplasma gondii, sur la base des
résultats obtenus par le dosage
IMMULITE Toxoplasmose IgG quantitatif,
est de 56 % dans cette population.
Des études complémentaires ont été
menées sur des échantillons prélevés
chez 42 femmes enceintes, âgées de 16 à
41 ans, habitant la même région. (Les
échantillons avaient été recueillis et
congelés à –20 °C avant d'être dosés). La
prévalence de la séropositivité pour
Toxoplasma gondii, sur la base des
résultats obtenus par le dosage
IMMULITE Toxoplasmose IgG quantitatif,
est de 5 % pour cette population. (Voir le
graphique “Expected Values”.)
Utiliser ces valeurs à titre indicatif
uniquement. Chaque laboratoire devrait
établir ses propres valeurs de référence.
Limites
L'utilisation du dosage
IMMULITE/IMMULITE 1000
Toxoplasmose IgG quantitatif pour le
diagnostic d'une infection récente en
dosant des paires de sérums n'a pas été
validée.
Les résultats doivent impérativement être
interprétés dans le contexte clinique du
patient, de la symptomatologie et d'autres
résultats de laboratoire.
La mise en évidence d'IgG spécifiques
dans un échantillon unique ne permet pas
de distinguer une infection active d'une
infection passée. Les patients suspects de
primo-infection ou d'infection active
doivent faire l'objet d'une recherche d'IgM
anti-Toxoplasma gondii.
Pour mettre en évidence une
séroconversion, un second échantillon
sérique devra être prélevé trois ou quatre
semaines après le prélèvement négatif de
la phase aiguë, au cours de la phase de
convalescence de l'infection. L'échantillon
de phase aiguë doit être conservé et testé
en parallèle avec l'échantillon de la phase
de convalescence.
Il est possible que les IgG spécifiques ne
soient pas détectables dans la phase
précoce d'une infection aiguë par
Toxoplasma gondii.
Les résultats obtenus chez des patients
séropositifs pour le VIH, des patients
suivant un traitement immunosuppresseur
ou des patients ayant d'autres troubles
conduisant à une immunosuppression,
devront être interprétés avec prudence.
Les performances de ce dosage n'ont pas
été évaluées pour l'utilisation
d'échantillons provenant de sang de
cordon, de nouveau-nés, de patients
allant recevoir une transplantation ou pour
des liquides biologiques autres que le
IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 27
sérum, comme l'urine, la salive ou le
liquide amniotique.
Les anticorps hétérophiles du sérum
humain peuvent réagir avec les
immunoglobulines faisant partie des
composants du coffret et interférer avec
les immunodosages in vitro. [Voir Boscato
LM, Stuart MC. Heterophilic antibodies: a
problem for all immunoassays. Clin Chem
1988:34:27-33.] Les échantillons
provenant de patients fréquemment
exposés aux animaux ou aux produits
sériques d'origine animale peuvent
présenter ce type d'interférence pouvant
potentiellement donner un résultat
anormal. Ces réactifs ont été mis au point
afin de minimiser le risque d'interférence,
cependant des interactions potentielles
entre des sérums rares et les composants
du test peuvent se produire. Dans un but
diagnostique, les résultats obtenus avec
ce dosage doivent toujours être utilisés en
association avec un examen clinique,
l'histoire médicale du patient et d'autres
résultats.
Performances du test
Consulter les tableaux et graphiques pour
obtenir les données représentatives des
performances du test. Les résultats sont
donnés en UI/ml. (En l'absence de
précision supplémentaire, tous les
résultats ont été obtenus sur des
échantillons sériques prélevés sur tubes
sans gel, ni activateur de la coagulation).
Domaine de mesure : 5,0 – 250 UI/ml.
Standardisé par rapport au Troisième
Standard International de l'OMS pour les
anticorps sériques anti-toxoplasmose chez
l'homme.
Sensibilité analytique : ≤ 5,0 UI/ml.
Précision : les statistiques ont été
réalisées sur les résultats de 5 doublets
testés chacun pendant 5 jours. (Voir le
tableau « Intraassay Precision » et
«Interassay Precision».)
N.B. : le contrôle négatif et l'échantillon
n° 1 étaient toujours inférieurs à 5 UI/ml
(la limite inférieure du domaine de mesure
pour tous les doublets de chaque série).
Spécificité : La spécificité du test
IMMULITE Toxoplasmose IgG quantitatif a
été évaluée sur 50 échantillons positifs
pour le cytomégalovirus (CMV), le virus de
la varicelle et du zona (VZV), le virus
Epstein-Barr (EBV), pour Mycoplasma
pneumoniae et le parvovirus B19. Avec
ces échantillons, le dosage IMMULITE
Toxoplasmose IgG quantitatif et un
immunodosage pour le dosage des IgG
anti-Toxoplasmose (coffret D) avaint une
concordance de 96% (48/50).
Autres types d'échantillons : pour
estimer l'effet de l'utilisation de différents
type d'échantillons, 16 volontaires ont été
prélevés sur tubes secs, héparinés et sur
tubes vacutainer SST
®
Becton Dickinson.
Tous les échantillons puis dosés selon le
protocole Toxoplasmose IgG quantitatif
IMMULITE, avec les résultats suivants.
(Hépariné) = 1,02 (Sérum) + 0,13 UI/ml
r = 1,00
(SST) = 0,98 (tubes ordinaires) – 0,06 UI/ml
r = 1,00
Moyennes :
4,4 UI/ml (Sérum)
4,6 UI/ml (Hépariné)
4,2 UI/ml (SST)
Dans une autre étude, le sang de 16
volontaires a été prélevé sur tubes
vacutainer en ordinaire et EDTA.
Tous les échantillons puis dosés selon le
protocole Toxoplasmose IgG quantitatif
IMMULITE, avec les résultats suivants.
(EDTA) = 0,93 (Sérum) – 0,003 UI/ml
r = 1,00
Moyennes :
5,3 UI/ml (EDTA)
5,6 UI/ml (Sérum)
Performance clinique
La performance clinique du dosage
IMMULITE Toxoplasmose IgG quantitatif a
été étudiée sur un total de 199
échantillons sériques, dosés chacun par
trois centres investigateurs, dont deux
situés en dehors des Etats-Unis. Sur les
199 échantillons (recueillis et congelés à
-20 °C avant d'être dosés), 61 avaient été
prélevés chez des femmes venues
consulter pour un examen obstétrique ou
gynécologique, 42 avaient été prélevés
chez des femmes enceintes et le reste
provenait de patients présentant divers
états cliniques (SIDA, anémie, artérite,
fatigue, céphalées, lymphome, méningite,
pneumonie, etc.). On comptait 37 hommes
et 162 femmes, âgés de 15 à 68 ans.
Une des études rétrospectives a été
menée dans un centre hospitalo-
universitaire du Nord-ouest des Etats-
28 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08)
Unis. Les 199 échantillons ont été
analysés avec le dosage IMMULITE
Toxoplasmose IgG quantitatif et avec le
coffret A, un dosage immunofluorescent
(ELFA) du commerce.
IMMULITE Toxoplasmose
IgG Quantitatif
Coffret A Pos Nég
Sensibilité
relative
Spécificité
relative
Positif 52 2
Négatif 0 145
96,3% 100%
Concordance : 99,0%
Limites de confiance à 95% pour la sensibilité
relative et la spécificité relative, respectivement:
87,3% – 99,5% et 97,5% – 100%. (Limites de
confiance par test exact.)
Les deux échantillons discordants,
négatifs selon IMMULITE, mais positifs
selon le coffret A, ont retestés avec le
coffret B, un dosage immuno-enzymatique
à microparticules (MEIA) disponible dans
le commerce, et trouvés positifs.
Dans une seconde étude rétrospective,
menée en dehors des Etats-Unis, 194
échantillons sur 199 (5 échantillons
avaient un volume insuffisant) ont été
dosé avec le test IMMULITE
Toxoplasmose IgG quantitatif et avec le
coffret B.
IMMULITE Toxoplasmose
IgG quantitatif
Coffret B Pos Nég
Sensibilité
relative
Spécificité
relative
Positif 46 5
Négatif 2 141
90,2% 98,6%
Concordance : 96,4%
Limites de confiance à 95% pour la sensibilité
relative et la spécificité relative, respectivement:
78,6% – 96,7% et 95,0% – 99,8%. (Limites de
confiance par test exact.)
Parmi les sept échantillons discordants,
trois sur cinq, qui étaient négatifs selon
IMMULITE mais positifs selon le coffret B,
étaient négatifs avec le coffret A. Les deux
échantillons positifs selon IMMULITE,
mais négatifs selon le coffret B étaient
positifs avec le coffret A.
Dans la troisième étude rétrospective,
également menée en dehors des Etats-
Unis, 160 échantillons sur 199 (39
échantillons étaient d'un volume
insuffisant) ont été analysés avec le
dosage IMMULITE Toxoplasmose IgG
quantitatif et avec le coffret C, un dosage
immuno-enzymatique du commerce.
IMMULITE Toxoplasmose
IgG quantitatif
Coffret C Pos Nég
Sensibilité
relative
Spécificité
relative
Positif 32 2
Douteux 0 4
Négatif 2 120
94,1% 98,4%
Concordance : 97,4%
Limites de confiance à 95% pour la sensibilité
relative et la spécificité relative, respectivement:
80,3% – 99,3% et 94,2% – 99,8%. (Limites de
confiance par test exact).
Les deux échantillons positifs selon
IMMULITE, mais négatifs selon le coffret
C, étaient positifs avec le coffret A. Parmi
les échantillons négatifs selon IMMULITE,
mais positifs selon le coffret C, l'un était
positif avec le coffret A et l'autre négatif
avec le coffret A. Les quatre échantillons
négatifs selon IMMULITE, mais douteux
selon le coffret C, étaient tous négatifs
avec le coffret A. Ainsi, les mesures
IMMULITE sont concordantes avec celles
du coffret A pour sept des huit cas de
divergence avec le coffret C.
Assistance technique
En France distribué par DPC France 90
bd National 92257 La Garenne-Colombes.
Le système d’assurance qualité de DPC est
certifié ISO 13485 (2003).
Italiano
IMMULITE/IMMULITE 1000
Toxoplasma Quantitativo IgG
Uso: Ad uso diagnostico in vitro con gli
Analizzatori IMMULITE ed IMMULITE
1000 — per la misurazione quantitativa
della anticorpi Anti-Toxoplasma gondii nel
siero, quale ausilio nella determinazione
dello stato sierologico vs.Toxoplasma
gondii.
Il kit non è stato approvato dall'FDA
per lo screening dei donatori di
sangue o di plasma.
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)
IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)

More Related Content

Viewers also liked

Viewers also liked (12)

DNA vaccine of toxoplsma gondii
DNA vaccine of toxoplsma gondiiDNA vaccine of toxoplsma gondii
DNA vaccine of toxoplsma gondii
 
Dysregulation of macrophage signal transduction by Toxoplasma
Dysregulation of macrophage signal transduction by ToxoplasmaDysregulation of macrophage signal transduction by Toxoplasma
Dysregulation of macrophage signal transduction by Toxoplasma
 
1302208 - Lit. Rev
1302208 - Lit. Rev1302208 - Lit. Rev
1302208 - Lit. Rev
 
Ifmt p13 cancers induits par les parasites
Ifmt p13 cancers induits par les parasitesIfmt p13 cancers induits par les parasites
Ifmt p13 cancers induits par les parasites
 
INMUNIDAD
INMUNIDADINMUNIDAD
INMUNIDAD
 
Toxoplasmosis Epidemiology
Toxoplasmosis EpidemiologyToxoplasmosis Epidemiology
Toxoplasmosis Epidemiology
 
Toxoplasmosis
ToxoplasmosisToxoplasmosis
Toxoplasmosis
 
Plasmodium & Toxoplasma
Plasmodium & ToxoplasmaPlasmodium & Toxoplasma
Plasmodium & Toxoplasma
 
Toxoplasma
ToxoplasmaToxoplasma
Toxoplasma
 
Innate immunity
Innate immunityInnate immunity
Innate immunity
 
Toxoplasma
ToxoplasmaToxoplasma
Toxoplasma
 
Toxoplasmosis
ToxoplasmosisToxoplasmosis
Toxoplasmosis
 

IMMULITE/IMMULITE 1000. Toxoplasma Quantitative IgG (2006)

  • 1. Toxoplasma Quantitative IgG For the Quantitative Measurement of IgG Antibodies to Toxoplasma gondii in Human Serum For use on the IMMULITE ® and IMMULITE ® 1000 systems
  • 2. IMMULITE ® /IMMULITE 1000 ® Toxoplasma Quantitative IgG 2 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) English Intended Use: For in vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers — for the quantitative measurement of IgG antibodies to Toxoplasma gondii (toxoplasma) in human serum, as an aid in the determination of serological status to Toxoplasma gondii. This kit is not FDA-cleared for use in the testing (i.e. screening) of blood or plasma donors. Catalog Number: LKTXP1 (100 tests), LKTXP5 (500 tests) Test Code: TXP Color: Light Gray CDC Analyte Identifier Code: 6113 CDC Test System Identifier Code: 10159 CLIA Complexity Category: Moderate Summary and Explanation Toxoplasma gondii is an obligate intracellular parasite capable of infecting most mammals, including humans. The organism is transmitted through ingestion of insufficiently cooked meat. Throughout the world, 1 to 90 percent of populations may be infected, 7 with 25 to 30 percent of the adult population infected in the United States. 3 While toxoplasmosis may be manifested in several forms, infections are usually clinically inapparent, and latent infections usually persist for life. 2 Overt clinical symptoms are similar to infectious mononucleosis, with lymphadenopathy, fever, headache, malaise and sometimes pneumonia and myocarditis. 6 As with other latent infections, acute toxoplasma infection can pose a serious threat to immunocompromised individuals and newborns who acquire the infection in utero. Immunosuppressed patients may develop encephalitis, myocarditis or pneumonitis. 3 Congenital infections usually result as a consequence of asymptomatic acute maternal infection. This infection can cause premature delivery, spontaneous abortion or stillbirth. 4,6 Neonates may manifest chorioretinitis, hydrocephaly, microcephaly, cerebral calcification and psychomotor retardation. 1 The majority of congenitally infected children will not exhibit any symptoms until later in life. 4 Management of toxoplasmosis requires serological monitoring of infected individuals, 3 as the organism is not readily available for culture. Quantitative testing for the presence of toxoplasma IgG can be useful to determine prior infection and indicate reactivation of the infection. Accurate diagnostic information is important, particularly during pregnancy, as treatment with spiramycin can reduce the risk to the fetus. 5 Principle of the Procedure IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG is a solid-phase, sequential chemiluminescent immunoassay. Incubation Cycles: 2 × 30 minutes. Specimen Collection The use of an ultracentrifuge is recommended to clear lipemic samples. Samples which are cloudy or have particulate material should be clarified by low-speed centrifugation. Centrifuging serum samples before a complete clot forms may result in the presence of fibrin. To prevent erroneous results due to the presence of fibrin, ensure that complete clot formation has taken place prior to centrifugation of samples. Some samples, particularly those from patients receiving anticoagulant therapy, may require increased clotting time. Blood collection tubes from different manufacturers may yield differing values, depending on materials and additives, including gel or physical barriers, clot activators and/or anticoagulants. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG has not been tested with all possible variations of tube types. Consult the section on Alternate Sample
  • 3. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 3 Types for details on tubes that have been tested. Volume Required: 10 µL of the prediluted patient sample (1-in-21 dilution). (Sample cup must contain at least 100 µL more than the total volume required.) Storage: 3 days at 2–8°C or 6 months at –20°C. 13 Patient Sample Dilution: Serum samples must be prediluted 1-in-21 in IgG/IgM Sample Diluent (LIGZ2). This dilution must be performed manually e.g. by adding 20 µL pf patient sample to 400 µL of Sample Diluent (LIGZ2, supplied in the kit). Alternatively, the IMMULITE 1000 Windows ® system performs the dilution on board, automatically, using LIGW (provided in the IgG/IgM Sample Diluent Module, L1KIGW1,5). Printed results have been corrected for the 1-in-21 dilution. Warnings and Precautions For in vitro diagnostic use. Reagents: Store at 2–8°C. Dispose of in accordance with applicable laws. Follow universal precautions, and handle all components as if capable of transmitting infectious agents. Source materials derived from human blood were tested and found nonreactive for syphilis; for antibodies to HIV 1 and 2; for hepatitis B surface antigen; and for antibodies to hepatitis C. Sodium azide, at concentrations less than 0.1 g/dL, has been added as a preservative. On disposal, flush with large volumes of water to prevent the buildup of potentially explosive metal azides in lead and copper plumbing. Chemiluminescent Substrate: Avoid contamination and exposure to direct sunlight. (See insert.) Water: Use distilled or deionized water. The bead is coated with inactivated toxoplasma antigen. However, caution is advised because of the possible presence of residual organism when working with, or disposing of, the materials supplied. The anti-toxoplasma IgG results determined for a given specimen with assays from different manufacturers can vary due to differences in assay methods and reagent specificity. Therefore, the results reported by the laboratory to the physician should include: "The following results were obtained with the IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG EIA. Results obtained from other manufacturers' assay methods may not be used interchangeably." Materials Supplied Components are a matched set. The barcode labels are needed for the assay. Toxoplasma Quantitative IgG Test Units (LTXP1) Each barcode-labeled unit contains one bead coated with inactivated, partially purified Toxoplasma gondii antigen (RH strain tachyzoites from mouse peritoneum). Stable at 2–8°C until expiration date. LKTXP1: 100 units. LKTXP5: 500 units. Allow the Test Unit bags to come to room temperature before opening. Open by cutting along the top edge, leaving the ziplock ridge intact. Reseal the bags to protect from moisture. Toxoplasma Quantitative IgG Reagent Wedges (LTXPA, LTXPB) With barcodes. LTXPA: 7.5 mL of a protein-based buffer, with preservative. LTXPB: 7.5 mL alkaline phosphatase (bovine calf intestine) conjugated to monoclonal murine anti-human IgG antibody in buffer, with preservative. Store capped and refrigerated: stable at 2–8°C until expiration date. Recommended usage is within 30 days after opening when stored as indicated. LKTXP1: 1 set. LKTXP5: 5 sets. Toxoplasma Quantitative IgG Adjustors (LTXPL, LTXPH) Two vials (Low and High), 4 mL each, of human serum with IgG reactive to toxoplasma, in buffer, with preservative. (No predilution is required.) Stable at 2–8°C for 14 days after opening, or for 6 months (aliquotted) at –20°C. LKTXP1: 1 set. LKTXP5: 2 sets.
  • 4. 4 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) Toxoplasma Quantitative IgG Controls (LTXPC1, LTXPC2, LTXPC3) Three vials (Negative, Low Positive and Positive), 2 mL each. LTXPC1 (Negative Control): One vial containing human serum with IgG nonreactive to toxoplasma, with preservative. LTXPC2, LTXPC3 (Low Positive Control, Positive Control): Two vials containing human serum with IgG reactive to toxoplasma, with preservative. Stable at 2–8°C for 14 days after opening, or for 6 months (aliquotted) at –20°C. LKTXP1: 1 set. LKTXP5: 2 sets. IMMULITE: Toxoplasma IgG controls require the same manual 1-in-21 predilution as patient samples, with IgG/IgM Sample Diluent (LIGZ2). IMMULITE 1000: The IMMULITE 1000 software performs automatic on-board dilution of patient and control samples. Enter controls as controls. Refer to the control insert for concentration levels. IgG/IgM Sample Diluent (LIGZ2) For the manual dilution of patient samples and controls. 100 mL nonhuman protein/buffer matrix, with preservative. Stable at 2–8°C for 30 days after opening, or 6 months (aliquotted) at –20°C. LKTXP1: 1 vial. LKTXP5: 3 vials. For on-board dilution of patient samples using IMMULITE 1000 Windows ® , refer to the Kit Components Supplied Separately section. Kit Components Supplied Separately IgG/IgM Sample Diluent Module (L1KIGW1,5) Required for the on-board dilution of patient samples and controls using IMMULITE 1000 Windows ® . Sample Diluent wedges (LIGW) containing concentrated (2.5X) non-human protein/buffer matrix, with preservative. Stable at 2–8°C until expiration date. L1KIGW1: 2 Wedges. L1KIGW5: 10 Wedges. LSUBX: Chemiluminescent Substrate LPWS2: Probe Wash Module LKPM: Probe Cleaning Kit LCHx-y: Sample Cup Holders (barcoded) LSCP: Sample Cups (disposable) LSCC: Sample Cup Caps (optional) DCHS: Dilution Cup Holders (barcoded) for IMMULITE 1000 Windows ® on-board dilutions. Also Required Sample transfer pipets, distilled or deionized water. Assay Procedure Note that for optimal performance, it is important to perform all routine maintenance procedures as defined in the IMMULITE or IMMULITE 1000 Operator's Manual. See the IMMULITE or IMMULITE 1000 Operator's Manual for: preparation, setup, dilutions, adjustment, assay and quality control procedures. Visually inspect each Test Unit for the presence of a bead before loading it onto the system. Note that both Reagent Wedges A and B must be loaded on the carousel to run this assay. For IMMULITE 1000 Windows ® , the Sample Diluent Wedge must also be loaded on the carousel. Recommended Adjustment Interval: 4 weeks. Quality Control Samples: The controls supplied with the kit should be used as quality control material to monitor assay performance. The negative and positive controls are intended to monitor for substantial reagent failure. The positive control provided will not ensure precision of the cut-off. In addition to the controls provided, users may wish to run additional controls to cover the upper range of the assay. Additional controls may be tested in accordance with guidelines or requirements of local, state and/or federal regulations or accrediting organizations. 12 It is also recommended that known reactive and nonreactive specimens be run periodically to assure pipetting accuracy for the dilution step. Interpretation of Results IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG assay is standardized in terms of the World Health Organization's
  • 5. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 5 Third International Standard for Anti- Toxoplasma Serum. The assay's cutoff is determined from representative positive and negative samples (confirmed by an ELISA method) using the ROC (Receiver Operating Characteristic) technique. Positive: A result greater than or equal to 8 IU/mL is indicative of a past infection. Negative: A result less than 6.5 IU/mL is considered to be negative. Results less than 6.5 IU/mL should be reported as Negative only, and no numerical concentration should be reported. Indeterminate: A result greater than or equal to 6.5 IU/mL and less than 8 IU/mL is considered indeterminate. Samples greater than 250 IU/mL should be reported as "greater than 250 IU/mL". Specimens reporting a result of greater than 250 IU/mL, when tested at the required 1-in-21 dilution, may be further diluted with the IgG/IgM Sample Diluent. The presence of IgG antibodies to Toxoplasma gondii is an indication of previous exposure to the organism. A single specimen can be used to estimate the serological status of the individual. A result of "indeterminate" indicates an intermediate level of Toxoplasma gondii antibodies in the sample: the sample should therefore be retested. Samples which still test as "indeterminate" should be examined by an alternate method, or a second sample should be collected, if possible, within a reasonable period of time (e.g. one week). Any interpretation of the results is not recommended if the positive or negative control falls outside the range specified in the Quality Control section. The magnitude of the reported IgG level cannot be correlated to an endpoint titer. Reports by the laboratory to the physician should include: "The following results were obtained with the IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG EIA. Values obtained from other manufacturers' assay methods may not be used interchangeably." Expected Values Individuals with the toxoplasma organism may not exhibit detectable levels of IgG antibody in the early stages of infection. Levels of IgG antibody to Toxoplasma gondii begin to rise one or two weeks after infection. Peak levels are reached in 6 to 8 weeks, then gradually decline over a period of months or even years. Low titers are generally detectable for life. The antibody titer does not correlate with severity of illness. 2,4,6 The prevalence of IgG antibody to Toxoplasma gondii will vary with age and geographical location. Studies were conducted with specimens from presumed healthy, asymptomatic subjects from the northwestern US. The subjects consisted of 69 individuals with ages ranging from 15 to 64 years. (Samples had been previously collected and stored frozen at –20°C). The prevalence of toxoplasma reactivity, based on the results from the IMMULITE Toxoplasma Quantitative IgG assay, was found to be 56% at this geographical location. Additional studies were conducted with specimens from 42 pregnant women from the same geographical location, with ages ranging from 16 to 41 years. (Samples had been previously collected and stored frozen at –20°C). The prevalence of toxoplasma reactivity for this specific population was 5%, based on the results from the IMMULITE Toxoplasma Quantitative IgG assay. (See “Expected Values” graph.) Consider these limits as guidelines only. Each laboratory should establish its own reference ranges. Limitations The use of IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG to diagnose recent infection by testing paired sera has not been validated. The results of the test must be taken within the context of the patient's clinical history, symptomology and other laboratory findings. The presence of IgG antibodies in a single specimen is not sufficient to distinguish between active and past infection. Patients suspected of having a primary or active toxoplasma infection should be tested for the presence of IgM antibodies to Toxoplasma gondii. For the determination of seroconversion from nonreactive to reactive, a second
  • 6. 6 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) serum sample should be drawn three to four weeks after the acute, nonreactive sample, during the convalescent stage of the infection. The acute phase sample should be stored and tested in parallel with the convalescent sample. Individuals with an acute toxoplasma infection may not exhibit any detectable IgG antibodies at the early stage of infection. The results in HIV patients, in patients undergoing immunosuppressive therapy, or in patients with other disorders leading to immunosuppression, should be interpreted with caution. The performance characteristics of this assay have not been established for use with specimens from neonates, cord blood, pretransplant patients, or body fluids other than serum, such as urine, saliva or amniotic fluid. Heterophilic antibodies in human serum can react with the immunoglobulins included in the assay components causing interference with in vitro immunoassays. [See Boscato LM, Stuart MC. Heterophilic antibodies: a problem for all immunoassays. Clin Chem 1988:34:27- 33.] Samples from patients routinely exposed to animals or animal serum products can demonstrate this type of interference potentially causing an anomalous result. These reagents have been formulated to minimize the risk of interference; however, potential interactions between rare sera and test components can occur. For diagnostic purposes, the results obtained from this assay should always be used in combination with the clinical examination, patient medical history, and other findings. Performance Data See Tables and Graphs for data representative of the assay's performance. Results are expressed in IU/mL. (Unless otherwise noted, all were generated on serum samples collected in tubes without gel barriers or clot-promoting additives.) Calibration Range: 5.0 to 250 IU/mL. Standardized in terms of WHO 3rd IS for Anti-Toxoplasma Serum, Human. Analytical Sensitivity: ≤ 5.0 IU/mL. Precision: Statistics were calculated for controls and samples processed in 5 replicates each on five separate days. (See “Intraassay Precision” table and “Interassay Precision” table.) Note that the negative control and Sample 1 were consistently below 5 IU/mL (the lower limit of the assay's calibration range) for all replicates in every run. Specificity: The specificity of IMMULITE Toxoplasma Quantitative IgG assay was evaluated by testing 50 specimens positive for cytomegalovirus (CMV), varicella-zoster virus (VZV), Epstein-Barr virus (EBV), Mycoplasma pneumoniae and parvovirus B19. With these specimens, IMMULITE Toxoplasma Quantitative IgG and a commercially available enzyme immunoassay for Toxoplasma IgG (Kit D) showed 96% agreement (48/50). Alternate Sample Type: To assess the effect of alternate sample types, blood was collected from 16 volunteers into plain, heparinized and Becton Dickinson SST ® vacutainer tubes. All samples were assayed by the IMMULITE Toxoplasma Quantitative IgG procedure, with the following results. (Heparin) = 1.02 (Serum) + 0.13 IU/mL r = 1.00 (SST) = 0.98 (Plain Tubes) – 0.06 IU/mL r = 1.00 Means: 4.4 IU/mL (Serum) 4.6 IU/mL (Heparin) 4.2 IU/mL (SST) In another experiment, blood was collected from 16 laboratory volunteers into plain and EDTA and vacutainer tubes. All samples were assayed by the IMMULITE Toxoplasma Quantitative IgG procedure, with the following results. (EDTA) = 0.93 (Serum) – 0.003 IU/mL r = 1.00 Means: 5.3 IU/mL (EDTA) 5.6 IU/mL (Serum) Clinical Performance The clinical performance of the IMMULITE Toxoplasma Quantitative IgG assay was studied using a total of 199 serum specimens, assayed at each of three sites, two of which were outside the US. Of the 199 specimens (previously collected and stored frozen at –20°C), 61 were obtained
  • 7. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 7 from women who presented for obstetric and gynecological examination, 42 were obtained from pregnant women, and the remainder were obtained from patients with miscellaneous diseases and conditions (AIDS, anemia, arthritis, fatigue, headache, lymphoma, meningitis, pneumonia and others). There were 37 male and 162 female subjects, with ages ranging from 15 to 68 years. One of the retrospective studies was conducted at a university medical center in the northwestern US. The 199 specimens were evaluated nonconcurrently with the IMMULITE Toxoplasma Quantitative IgG procedure and Kit A, a commercially available Toxoplasma IgG enzyme-linked fluorescent immunoassay (ELFA). IMMULITE Toxoplasma Quantitative IgG Kit A Pos Neg Relative Sensitivity Relative Specificity Positive 52 2 Negative 0 145 96.3% 100% Agreement: 99.0% 95% Confidence Limits for Relative Sensitivity and Specificity, respectively: 87.3% – 99.5% and 97.5% – 100%. (Confidence limits by exact test.) The two discordant specimens, negative by IMMULITE but positive by Kit A, were further evaluated by Kit B, a commercially available Toxoplasma IgG microparticle enzyme immunoassay (MEIA), and found to be positive. In the second retrospective study, conducted outside the US, 194 of the 199 specimens (insufficient sample volume for 5 specimens) were evaluated with the IMMULITE Toxoplasma Quantitative IgG assay and Kit B. IMMULITE Toxoplasma Quantitative IgG Kit B Pos Neg Relative Sensitivity Relative Specificity Positive 46 5 Negative 2 141 90.2% 98.6% Agreement: 96.4% 95% Confidence Limits for Relative Sensitivity and Specificity, respectively: 78.6% – 96.7% and 95.0% – 99.8%. (Confidence limits by exact test.) Of the seven discordant specimens, three of the five negative by IMMULITE but positive by Kit B were negative by Kit A. The two specimens positive by IMMULITE but negative by Kit B were positive by Kit A. In the third retrospective study, also conducted outside the US, 160 of the 199 specimens (insufficient sample volume for 39 specimens) were evaluated with the IMMULITE Toxoplasma Quantitative IgG assay and Kit C, a commercially available enzyme immunoassay. IMMULITE Toxoplasma Quantitative IgG Kit C Pos Neg Relative Sensitivity Relative Specificity Positive 32 2 Equivocal 0 4 Negative 2 120 94.1% 98.4% Agreement: 97.4% 95% Confidence Limits for Relative Sensitivity and Specificity, respectively: 80.3% – 99.3% and 94.2% – 99.8%. (Confidence limits by exact test). The two cases positive by the IMMULITE but negative by Kit C, were positive by Kit A. Of the two cases negative by IMMULITE but positive by Kit C, one was positive by Kit A and the other negative by Kit A. The four cases negative by IMMULITE but equivocal by Kit C were all negative by Kit A. Thus, the IMMULITE results agree with Kit A in seven of the eight cases of discordance with Kit C. References 1) Daffos F. Prenatal management of 746 pregnancies at risk for congenital toxoplasmosis. N Engl J Med 1988;318: 271-5. 2) Krick JA, Remington JS. Toxoplasmosis in the adult: an overview. N Engl J Med 1978;298: 550-3. 3) Krogstad DJ, et al. Blood and tissue protozoa. In: Lennette EH, et al, editors. Manual of clinical microbiology. 4th ed. Washington, D.C.: American Society for Microbiology, 1985: 612- 30. 4) McCabe RE, Remington JS. Toxoplasma gondii. In: Mandell, Douglas, Bennett, editors. Principles and practice of infectious diseases. 2nd ed. New York: John Wiley, 1985. 5) Jeannel D, et al. What is known about the prevention of congenital toxoplasmosis? Lancet 1990;336:359-61. 6) Remington JS. McLeod R. Toxoplasmosis. In: Braude, editor. Microbiology and infectious diseases. Philadelphia: Saunders, 1981. 7) Walls KW. Toxoplasmosis. In: Balows A, et al, editors. Laboratory diagnosis of infectious diseases. New York: Springer-Verlag, 1988: 998-1017. 8) National Committee for
  • 8. 8 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) Clinical Laboratory Standards. Procedures for the collection of diagnostic blood specimens by venipuncture; approved standard. 4th ed. NCCLS Document H3-A4, Wayne, PA: NCCLS, 1998. 9) Baltz ML, Searcy RL. Clinical significance and advanced serologic diagnosis of ToRCH infections. Am Clin Lab 1994 (Mar/Apr): 18-23. 10) Wilson M, et al. Seroepidemiology of toxoplasmosis in the United States. Abstract presented at the American Society for Microbiology, May 1994. 11) Approved standard procedures for the handling and processing of blood specimens, NCCLS Document H18-A, 1991. 12) International Quality Control testing: principles and definitions, NCCLS Document C24-A, 1991. 13) Tietz NW, editor. Clinical guide to laboratory tests. 3rd ed. Philadelphia: WB Saunders, 1995:358. Technical Assistance In the United States, contact DPC's Technical Services department. Tel: 800.372.1782 or 973.927.2828 Fax: 973.927.4101. Outside the United States, contact your National Distributor. The Quality System of Diagnostic Products Corporation is registered to ISO 13485:2003. Tables and Graphs Intraassay Precision (IU/mL) Mean1 SD2 CV3 Pos. Control4 18 0.95 5.3% Neg. Control5 <5 — — 1 <5 — — 2 8.3 0.81 9.8% 3 45 3.9 8.7% 4 83 5.1 6.1% 5 122 14 11% Interassay Precision (IU/mL) Mean1 SD2 CV3 Pos. Control4 18 1.3 7.2% Neg. Control5 <5 — — 1 <5 — — 2 8.3 0.85 10.2% 3 45 6.9 15% 4 83 7.5 9.0% 5 122 16 13% Expected Values Reactivity: IML Toxoplasma Quantitative IgG Asymptomatic Pregnant 0% 20% 40% 60% 80% 100% Reactive Non-Reactive n = 69 n = 42 Deutsch. Intraassay Precision: 1 Mittelwert, 2 S (Standardabweichung), 3 CV (Variationskoeffizient), 4 Positiv-Kontrolle, 5 Negativ-Kontrolle. Interassay Precision: 1 Mittelwert, 2 S (Standardabweichung), 3 CV (Variationskoeffizient), 4 Positiv-Kontrolle 5 Negativ-Kontrolle. Reactivity: IML Toxoplasma Quantitative IgG: IML Toxoplasma Quantitativ IgG, Reactive: reaktiv, Non-Reactive: nicht- reaktiv, Asymptomatic: Asymptomatic, Pregnant: Schwanger. Español. Intraassay Precision: 1 Media, 2 DS, 3 CV, 4 Pos. Control, 5 Neg. Control. Interassay Precision: 1 Media, 2 DS, 3 CV, 4 Pos. Control, 5 Neg. Control. Reactivity: IML Toxoplasma Quantitative IgG: IML Toxoplasma Quantitative IgG, Reactive: Reactive, Non-Reactive: Non- Reactive, Asymptomatic: Asymptomatic, Pregnant: Pregnant. Français. Intrraassay Precision: 1 Moyenne, 2 SD, 3 CV, 4 Pos. Control, 5 Neg. Control. Interassay Precision: 1 Moyenne, 2 SD, 3 CV, 4 Pos. Control, 5 Neg. Control. Reactivity: IML Toxoplasma Quantitative IgG: IML Toxoplasma Quantitative IgG, Reactive: Reactive, Non- Reactive: Non-Reactive, Asymptomatic: Asymptomatic, Pregnant: Pregnant.
  • 9. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 9 Italiano. intraassay Precision: 1 Media, 2 SD (Deviazione Standard), 3 CV (Coefficiente di Variazione), 4 Pos. Control, 5 Neg. Control. Interassay Precision: 1 Media, 2 SD (Deviazione Standard), 3 CV (Coefficiente di Variazione), 4 Pos. Control, 5 Neg. Control. Reactivity: IML Toxoplasma Quantitative IgG: IML Toxoplasma Quantitative IgG, Reactive: Reactive, Non- Reactive: Non-Reactive, Asymptomatic: Asymptomatic, Pregnant: Pregnant. Português. Intraassay Precision: 1 Média, 2 Desvio padrão, 3 Coeficiente de variação, 4 Pos. Control, 5 Neg. Control. Interassay Precision: 1 Média, 2 Desvio padrão, 3 Coeficiente de variação, 4 Pos. Control, 5 Neg. Control. Reactivity: IML Toxoplasma Quantitative IgG: IML Toxoplasma Quantitative IgG, Reactive: Reactive, Non-Reactive: Non-Reactive, Asymptomatic: Asymptomatic, Pregnant: Pregnant. Deutsch Toxoplasma IgG quantitativ Anwendung: Für in-vitro-diagnostische Tests mit dem Analysegerät IMMULITE und IMMULITE 1000 zum quantitativen Nachweis von IgG-Antikörpern gegen Toxoplasma gondii (Toxoplasma) in Serum als Hilfestellung zur Bestimmung des serologischen Status gegen Toxoplasma gondii. Dieses Testsystem ist von der US- amerikanischen Lebensmittel- und Medikamentenbehörde FDA nicht für Screening-Tests mit Blut- oder Plasmaspendern freigegeben. Artikelnummern: Catalog Number: LKTXP1 (100 Tests), LKTXP5 (500 Tests) Testcode: TXP Farbe: hellgrau Klinische Relevanz Toxoplasma gondii ist ein obligater intrazellulärer Parasit, der die meisten Säugetiere – einschließlich Menschen – befällt und zu Infektionen führt. Der Erreger wird durch den Verzehr von ungenügend gekochtem Fleisch übertragen. Weltweit sind möglicherweise zwischen 1 und 90 Prozent aller Populationen infiziert 7 , in den USA sind es 25–30 % der erwachsenen Bevölkerung. 3 Toxoplasmosen können sich in verschiedenen Formen manifestieren, normalerweise sind die Infektionen jedoch klinisch unauffällig und in ihrer latenten Form lebenslang präsent. 2 Zu den offenen klinischen Symptomen zählen infektiöse Mononukleosen mit Lymphadenopathien, Fieber, Kopfschmerzen, Unwohlsein und gelegentlich auch Lungenentzündungen und Myokarditiden. 6 So wie andere latente Infektionen können auch akute Toxoplasmosen bei immungeschwächten Patienten und in utero infizierten Neugeborenen eine ernsthafte Bedrohung darstellen. Bei immunsupprimierten Patienten können Enzephalitis, Myokarditis oder Pneumonitis entstehen. 3 Kongenitale Infektionen sind normalerweise das Resultat einer asymptomatischen akuten Infektion der Mutter. Diese Infektionen könen zu Frühgeburten, Spontanaborten oder Totgeburten führen. 4,6 Bei Neugeborenen können Chorioretinitiden, Hydrozephalie, Mikrozephalie, zerebrale Kalkablagerungen sowie psychomotorische Behinderungen auftreten. 1 Die Mehrzahl der kongenital infizierten Kinder zeigt erst im späteren Leben Symptome. 4 Da der Erreger von Toxoplasmosen nicht ohne weiteres kultiviert werden kann, müssen die infizierten Patienten serologisch überwacht werden. 3 Quantitative Bestimmungen von Toxoplasma-IgG können nützlich sein, um frühere und reaktivierte Infektionen anzuzeigen. Genaue diagnostische Daten sind vor allem bei Schwangerschaften wichtig, da eine Behandlung mit Spiramycin das Risiko für den Fetus herabsetzen kann. 5 Methodik Der IMMULITE/IMMULITE 1000 Toxoplasma IgG quantitativ ist ein Festphasen-, sequenzieller Chemilumineszens-Immunoassay. Inkubationszyklen: 2 × 30 minuten. Probengewinnung Der Einsatz einer Ultrazentrifuge wird zur Klärung von lipämischen Proben empfohlen.
  • 10. 10 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) Trübe oder partikelhaltige Proben sollten bei niedriger Geschwindigkeit zentrifugiert werden, bis sie klar sind Die Zentrifugation der Serumproben vor dem völligen Abschluss der Gerinnung kann zu Fibringerinnseln führen. Um fehlerhaften Analysenergebnissen infolge von Gerinnseln vorzubeugen, ist sicherzustellen, dass die Gerinnung vor der Zentrifugation der Proben vollständig abgeschlossen ist. Insbesondere Proben von Patienten unter Antikoagulantien- therapie können eine verlängerte Gerinnungszeit aufweisen. Blutentnahmeröhrchen von verschiedenen Herstellern können differierende Werte verursachen. Dies hängt von den verwendeten Materialien und Additiven (Gel oder physische Trennbarrieren, Gerinnungsaktivatoren und /oder Antikoagulantien) ab. IMMULITE/IMMULITE 1000 Toxoplasma IgG quantitativ sind nicht mit allen möglichen Röhrchenvariationen ausgetestet worden. Details der getesteten Röhrchenarten sind dem Kapitel "Alternative Probenarten" zu entnehmen. Erforderliche Menge: 10 µl der vorverdünnten Patientenprobe (1:21 Verdünnen). (Inhalt der Probenschale muss mindestens 100 µl über der erforderlichen Gesamtmenge liegen.) Lagerung: 3 Tage bei 2–8°C oder 6 Monate bei –20°C. 13 Verdünnung der Patientenproben: Serumproben müssen 1:21 im IgG/IgM Probenverdünnungsreagenz (LIGZ2) vorverdünnt werden. Diese Verdünnung ist manuell durchzuführen, z.B. durch Zugabe von 20 µl der Patientenprobe in 400 µl Probenverdünnungsreagenz (LIGZ2, im Kit enthalten). Alternativ, bei Verwendung des IMMULITE 1000 Windows, erfolgt die Verdünnung automatisch on-board. Das System verwendet LIGW, Inhalt des IgG/IgM Probenverdünnungsreagenz-Moduls (L1KIGW1,5). Die ausgedruckten Resultate entsprechend der 1:21 Verdünnung korrigiert sind. Hinweise und Vorsichtsmaßnahmen Zur In-vitro-Diagnostik. Reagenzien: Bei 2–8°C lagern. Unter Einhaltung der geltenden gesetzlichen Vorschriften entsorgen. Die generell geltenden Vorsichtsmaßnahmen sind einzuhalten und alle Komponenten als potenziell infektiös zu behandeln. Alle aus menschlichem Blut gewonnenen Materialien wurden auf Syphilis, Antikörper gegen HIV-1 und HIV-2, Hepatitis-B-Oberflächenantigen und Hepatitis-C-Antikörper untersucht und negativ befundet. Bestimmten Komponenten wurde Natriumazid (<0,1 g/dl) hinzugefügt. Um die Bildung von explosiven Metallaziden in Blei- und Kupferrohren zu vermeiden, sollten die Reagenzien nur zusammen mit großen Wassermengen in die Kanalisation gespült werden. Chemilumineszenz-Substrat: Kontamination und direkte Sonneneinstrahlung vermeiden. Siehe Packungsbeilage. Wasser: Destilliertes oder deionisiertes Wasser verwenden. Die Phase ist mit inaktiviertem Toxoplasma-Antigen beschichtet. Wegen des potenziellen Vorhandenseins überlebender Mikroorganismen ist jedoch beim Arbeiten mit und beim Entsorgen von den im Lieferumfang enthaltenen Materialien Vorsicht geboten. Unterschiede in der jeweiligen Methodik oder der Spezifität der Reagenzien könnten jedoch dazu führen, das die mit Testsystemen von verschiedenen Herstellern ermittelten Anti- Toxoplasmose-IgG-Ergebnisse für dieselben Proben nicht einheitlich sind. Die vom Labor an den Arzt weitergegebenen Ergebnisse sollten daher den folgenden Passus enthalten: „Die folgenden Ergebnisse wurden mit dem IMMULITE/IMMULITE 1000- Testsystem zur quantitativen Bestimmung von Toxoplasmose-IgG erzielt." Sie sind nicht mit den Ergebnissen der Testsysteme anderer Hersteller austauschbar.
  • 11. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 11 Im Lieferumfang enthalten Die Komponenten sind aufeinander abgestimmt. Die Barcode-Etiketten werden für den Assay benötigt. Toxoplasma quantitativ IgG Testeinheiten (LTXP1) Jede mit Barcode-Etikette versehene Einheit enthält eine mit beschichtet mit inaktiviertem, teilweise gereinigtem Toxoplasma-gondii-Antigen (Tachyzoiten, Stamm RH, aus Mausperitoneum). beschichtete Kugel. Bei 2–8°C bis zum Ablaufdatum haltbar. LKTXP1: 100 Testeinheiten. LKTXP5: 500 Testeinheiten. Verpackte Testeinheiten vor dem Öffnen stehen lassen, bis sie Zimmertemperatur erreicht haben. Oben entlang der Kante aufschneiden, ohne den Plastikverschluss zu beschädigen. Verpackungen wieder dicht verschließen, damit der Inhalt trocken bleibt. Toxoplasma quantitativ IgG – Reagenzbehälter (LTXPA, LTXPB) Mit Barcodes. LTXPA: 7,5 ml einer proteinbasierten Pufferlösung (mit Konservierungsmittel). LTXPB: 7,5 ml mit alkalischer Phosphatase (Rinderkalbsdarm) konjugierter monoklonaler antihumaner IgG Mausantikörper in Pufferlösung (mit Konservierungsmittel). Verschlossen und gekühlt aufbewahren: Bei 2–8°C bis zum Ablaufdatum haltbar. Bei entsprechender Lagerung beträgt die empfohlene Aufbrauchsfrist nach dem Öffnen 30 Tage. LKTXP1: 1 Set. LKTXP5: 5 Sets. Toxoplasma quantitativ IgG - Kalibratoren (LTXPL, LTXPH) Zwei Fläschchen (niedrig und hoch) mit 4 ml Tosoplasma-reaktivem IgG in einer Pufferlösung (mit Konservierungsmittel) (Muss nicht vorverdünnt werden.) 14 Tage nach dem Öffnen bei 2–8°C haltbar oder 6 Monate bei –20°C (aliquotiert). LKTXP1: 1 Set. LKTXP5: 2 Sets. Toxoplasma quantitativ IgG Kontrollen (LTXPC1, LTXPC2, LTXPC3) Drei Fläschchen (Negative, Schwachpositiv und Positiv) jeweils mit 2 ml Kontrollen. LTXPC1 (Negativkontrolle): Ein Fläschchen Humanserum mit Toxoplasma- nichtreaktivem IgG (mit Konservierungsmittel). LTXPC2, LTXPC3 (Schwachpositivkontrolle, Positiv- kontrolle): Zwei Fläschchen Humanserum mit Toxoplasma-reaktivem IgG (mit Konservierungsmittel). 14 Tage nach dem Öffnen bei 2–8°C haltbar oder 6 Monate bei –20°C (aliquotiert). LKTXP1: 1 Set. LKTXP5: 2 Sets. IMMULITE: Für die Toxoplasma IgG - Kontrollen ist die gleiche manuelle 1- in- 21- Vorverdünnung mit der IgG/IgM- Probenverdünner wie für die Patientenproben erforderlich (LIGZ2). IMMULITE 1000: Die IMMULITE 1000 Software führt eine automatische on- board-Verdünnung der Patienproben und der Kontrollen durch. Kontrollen als Kontrollen eingeben. Kontrollen als Kontrollen eingeben. Die Konzentrationen entnehmen Sie bitte der Beilage zu den Kontrollen. IgG/IgM–Verdünnungspuffer (LIGZ2) Zum manuellen Verdünnen der Patientenproben und Kontrollen. 100 ml Nichthumane Protein/Puffermatrix mit Konservierungsmittel. 30 Tage nach dem Öffnen bei 2–8°C oder 6 Monate bei –20 °C haltbar (aliquotiert). LKTXP1: 1 Fläschchen. LKTXP5: 3 Fläschchen. Für On-board-Verdünnungen mit IMMULITE 1000 Windows ® von Patientenproben ist die Liste der separat erhältlichen Testsystem-Komponenten heranzuziehen. Separat erhältliche Testsystem- Komponenten IgG/IgM Probenverdünnungsreagenz- Modul (L1KIGW1,5) Erforderlich für die on-board Verdünnung der Patientenproben und Kontrollen bei Verwendung des Immulite 1000 Windows. Probenverdünnungsreagenz-Behälter (LIGW) enthalten konzentrierte (2,5fach) nicht-humane Protein/Puffer Matrix mit Konservierungsmittel. Haltbar bei 2-8°C bis zum Verfallsdatum. L1KIGW1: 2 Behälter. L1KIGW5: 10 Behälter. LSUBX: Chemilumineszenz- Substratmodul LPWS2: Waschmodul
  • 12. 12 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) LKPM: Reinigungsmodul LCHx-y: Barcodierte Probenträger LSCP: Einmal-Probenröhrchen LSCC: Deckel für Probenröhrchen DCHS: Verdünnungsgefäßträger (barcodiert) für die IMMULITE 1000 Windows on-board-Verdünnung. Ebenfalls benötigt: Transferpipetten für die Proben; destilliertes bzw. deionisiertes Wasser. Testdurchführung Für eine optimale Funktion des Gerätes ist unbedingt zu beachten, dass die Wartungen, wie im IMMULITE oder IMMULITE 1000-Handbuch beschrieben, regelmäßig durchgeführt werden. Das Handbuch für das IMMULITE bzw. IMMULITE 1000 enthält die Anweisungen für: Vorbereitung, Geräteeinstellungen, Verdünnungen, Kalibrierung, Testdurchführung und Qualitätskontrollen. Überprüfen Sie jedes Testeinheit auf das Vorhandensein der Polystyrol-Kugel vor dem Einsetzen in das Gerät. Beachten Sie, dass zur Durchführung dieses Tests beide Reagenzbehälter (A und B) auf das Karussell geladen werden müssen. Beim IMMULITE 1000 Windows muss der Probenverdünnungsreagenz-Behälter ebenfalls auf dem Reagenzienkarussell platziert sein. Empfohlenes Kalibrationsintervall: 4 Wochen. Proben zur Qualitätskontrolle: Die im Lieferumfang enthaltenen Kontrolle(n) verwenden. Die Negativ- und Positivkontrollen dienen zur Überwachung der Reagenzien auf wesentliche Mängel. Die gelieferte Positivkontrolle eignet sich nicht dazu, die Präzision des Cutoff sicherzustellen. Zusätzlich zu den gelieferten Kontrollen können zur Abdeckung des höheren Messbereiches weitere Kontrollen eingesetzt werden. Entsprechend den Richtlinien bzw. Vorschriften von regionalen bzw. staatlichen Behörden oder zertifizierenden Stellen können weitere Kontrollen getestet werden. 12 Ferner empfiehlt es sich, regelmäßig Testansätze mit bekannten reaktiven und nichtreaktiven Proben durchzuführen, um die Pipettiergenauigkeit für den Verdünnungsschritt zu gewährleisten. Interpretation der Ergebnisse Das quantitative IMMULITE/IMMULITE 1000-Testsystem zur quantitativen Bestimmung von Toxoplasma-IgG ist entsprechend der Dritten Internationalen WHO-Norm für Anti-Toxoplasma-Serum standardisiert. Der Cutoff des Testsystems wird mit Hilfe der ROC-Technik (Receiver Operating Characteristic) aus repräsentativen positiven und negativen Proben (verifiziert mittels ELISA) ermittelt. Positiv: Ein Ergebnis von ≥ 8 IU/ml weist auf eine vergangene Infektion hin. Negativ: Ein Ergebnis von unter 6,5 IU/ml gilt als negativ. Ergebnisse von unter 6,5 IU/ml sollten lediglich als negativ ausgewiesen werden, ohne dabei die gemessene Konzentration numerisch darzustellen. Grenzwertig: Ein Ergebnis von ≥ 6,5 IU/ml und < 8 IU/ml gilt als grenzwertig. Proben mit mehr als 250 IU/ml sollten dementsprechend (d.h. „höher als 250 IU/ml“) ausgewiesen werden. Proben, die bei der erforderlichen Autoverdünnung von 1:21 mehr als 250 IU/ml aufweisen, können mit dem IgG/IgM- Probenverdünner weiterverdünnt werden. Sind IgG-Antikörper gegen Toxoplasma gondii vorhanden, so deutet dies auf eine frühere Exposition gegen den Erreger hin. Einzelproben eignen sich ausschließlich zur Bewertung des serologischen Status der betreffenden Person. Ein „indeterminate“ // „unentschiedenes“ Ergebnis weist auf einen Zwischenwert bei den Antikörpern gegen Toxoplasma gondii in der Probe hin: Die Probe sollte daher erneut getestet werden. Ergebnis bringen, sind entweder mit einer alternativen Methode zu testen, oder, sofern dies möglich ist, innerhalb eines vernünftigen Zeitraums (z.B. eine Woche) mit einer neu entnommenen zweiten Probe zu ergänzen. Von einer Interpretation der Ergebnisse wird abgeraten, wenn die Positiv- oder Negativkontrollen außerhalb des im Abschnitt „Quality Control“ //
  • 13. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 13 „Qualitätskontrolle“ angegebenen Bereichs liegen. Die Größenordnung der ausgewiesenen IgG-Konzentration kann nicht mit einem Endtiter korreliert werden. Die vom Labor an den Arzt weitergegebenen Ergebnisse sollten daher den folgenden Passus enthalten: „Die folgenden Ergebnisse wurden mit dem IMMULITE/IMMULITE 1000- Testsystem zur quantitativen Bestimmung von Toxoplasmose-IgG erzielt." Sie sind nicht mit den Ergebnissen der Testsysteme anderer Hersteller austauschbar. Referenzwerte Im Frühstadium einer Infektion mit dem Toxoplasma-Erreger sind möglicherweise keine nachweisbaren Konzentrationen an IgG-Antikörpern vorhanden. Die Konzentrationen an IgG-Antikörpern gegen Toxoplasma gondii steigen erstmals ein bis zwei Wochen nach der Primärinfektion an. Die Spitzenwerte werden innerhalb von 6–8 Wochen erreicht und fallen dann über Monate oder sogar Jahre hinaus allmählich wieder ab. Niedrige Titer bleiben generell für den Rest des Lebens nachweisbar. Es besteht keine Korrelation zwischen Antikörpertiter und dem Schweregrad der Erkrankung. 2,4,6 Die Prävalenz von IgG-Antikörpern für Toxoplasma gondii kann abhängig vom Alter und der geographischen Lage variieren. Es wurden mit gesunden, asymptomatischen Patienten aus dem Nordwesten der USA Studien durchgeführt. Die Patienten setzten sich aus 69 Personen im Alter zwischen 15 und 64 Jahren zusammen. (Die Proben wurden vorher gesammelt und bei –20°C eingefroren gelagert). In dieser geographischen Gegend wurde eine Prävalenz der Toxoplasmosereaktivität von 56% gefunden. Weiterhin wurden Studien mit Patientenseren von schwangeren Frauen im Alter zwischen 16 und 41 Jahren aus der gleichen geographischen Region durchgeführt. (Die Proben wurden vorher gesammelt und bei –20°C eingefroren gelagert). Die Prävalenz der Toxoplasmosereaktivität dieser speziellen Bevölkerungsgruppe lag bei 5%, basierend auf den Resultaten des IMMULITE Toxoplasma quantitativem IgG Test. (siehe grafik „Grenzen der Methode“) Betrachten Sie diese Grenzwerte nur als Richtlinien. Jedes Labor sollte eigene Referenzbereiche ermitteln. Grenzen der Methode Der Einsatz des IMMULITE/IMMULITE 1000 Testsystems zur quantitativen Bestimmung von Toxoplasma-IgG ist nicht validiert, was die Diagnose kürzlicher Infektionen mittels gepaarten Seren betrifft. Die Testergebnisse sind vor dem Hintergrund der klinischen Anamnese, der Beschwerden des Patienten sowie weiterer Laborbefunde zu bewerten. Das Vorhandensein von IgG-Antikörpern in Einzelproben reicht nicht aus, um zwischen einer aktiven und einer vergangenen Infektion zu unterscheiden. Patienten mit Verdacht auf eine Primär- bzw. aktive Infektion sollten auf das Vorhandensein von IgM-Antikörpern gegen das Toxoplasma gondii untersucht werden. Zur Bestimmung einer Serokonversion von nichtreaktiv zu reaktiv sollte während der Rekonvaleszenzphase drei bis vier Wochen nach der akuten nichtreaktiven Probe eine zweite Serumprobe genommen werden. Die Probe aus der Akutphase sollte gelagert und parallel zur Rekonvaleszenzprobe getestet werden. Die Ergebnisse bei HIV-Patienten sowie behandlungs- oder krankheitsbedingt immunsupprimierten Patienten sollten zurückhaltend interpretiert werden. Im Frühstadium einer akuten Toxoplasmose sind möglicherweise keine nachweisbaren IgG-Antikörper vorhanden. Die Eckdaten dieses Testsystems wurden nicht für den Gebrauch mit Proben von Neugeborenen, Nabelschnurblut, Patienten mit bevorstehender Organverpflanzung und nicht für andere Körperflüssigkeiten als Serum wie etwa Harn, Speichel oder Fruchtwasser etabliert. Heterophile Antikörper in Humanseren können mit Immunglobulinen aus den Assaykomponenten reagieren und Interferenzerscheinungen innerhalb des in
  • 14. 14 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) vitro Immunoassays verursachen. (Clin. Chem. 1988:34:27-33) Proben von Patienten, die häufig mit Tier- bzw. Tierserumprodukten zu tun haben, können die erwähnten Interferenzen verursachen und zu anomalen Resultaten führen. Die verwendeten Reagenzien sind so konzipiert, dass das Risiko einer Interferenz mit den zu messenden Proben minimiert ist. Dennoch können potentiell Interaktionen zwischen seltenen Seren und den Testkomponenten auftreten. Zu diagnostischen Zwecken sollten die mit dem Assay erhaltenen Ergebnisse immer in Kombination mit der klinischen Untersuchung, der Patientenanamnese und anderen Befunden gesehen werden Leistungsdaten Repräsentative Daten zum Test entnehmen Sie bitte den Tabellen und graphischen Darstellungen im englischen Teil dieser Anleitung. Die Ergebnisse sind in IU/ml angegeben. (Sofern nicht anders angegeben, wurden hierfür Serumproben in Röhrchen ohne Geltrennung und Gerinnungshilfen eingesetzt.) Messbereich: 5,0 – 250 IU/ml. Standardisiert nach der Dritten Internationalen WHO-Norm für humane Anti-Toxoplasma-Seren. Analytische Sensitivität: ≤ 5,0 IU/ml Präzision: Statisik aus einem einzelnen Testansatz mit 5 Proben verschiedener IgG-Konzentrationen in 5-facher Bestimmung am gleichen Tag. (Siehe Tabelle „Intraassay / Interassay - Precision“). Die Negativ-Kontrolle sowie Probe 1 zeigt durchgehend Werte unter 5 IU/ml. Spezifität: Die Spezifität des IMMULITE Toxoplasma Quantitativ IgG Tests wurde mit 50 Proben, die positiv für Cytomegalie (CMV), Varizelle zoster (VZV), Epstein- Barr-Virus (EBV), Mycoplasma pneumoniae und Parvovirus B19 waren, evaluiert. Diese Proben zeigten mit dem IMMULITE Toxoplasma Quantitativ IgG Test eine 96%ige Übereinstimmung (48/50) mit einem kommerziell erhältlichen Toxoplasma IgG Enzymimmunotest. Alternativer Probentyp: Um die Auswirkungen von verschiedenen Probenarten zu untersuchen, wurde Blut von 16 Freiwilligen in Röhrchen ohne Additiva, in Heparin- und Becton Dickinson SST ® Vacutainer-Rörchen gesammelt. Alle Proben wurden mit dem Toxoplasma IgG Verfahren für den IMMULITE getestet, mit der folgenden Ergebnisse. (Heparin) = 1,02 (Serum) + 0,13 IU/ml r = 1,00 (SST) = 0,98 (einfachen Röhrchen) – 0,06 IU/ml r = 1,00 Mittelwerte: 4,4 IU/ml (Serum) 4,6 IU/ml (Heparin) 4,2 IU/ml (SST) In einer weiteren Studie wurde 16 Freiwilligen Blut entnommen. Als Abnahmegefäße wurden Serumröhrchen ohne Zusätze, Heparin-, und EDTA Vakutainer röhrchen verwendet. Alle Proben wurden mit dem Toxoplasma IgG Verfahren für den IMMULITE getestet, mit der folgenden Ergebnisse. (EDTA) = 0,93 (Serum) – 0,003 IU/ml r = 1,00 Mittelwerte: 5,3 IU/ml (EDTA) 5,6 IU/ml (Serum) Klinische Studien Die klinischen Leistungsdaten des IMMULITE Toxoplasma Quantitativ IgG Test wurdem mit 199 Serumproben aus drei verschiedenen Regionen, von denen zwei außerhalb der USA stammten, geprüft. Die 199 Patientenproben (die Proben wurden vorher gesammelt und bei –20°C eingefroren gelagert) wurden von Frauen erhalten, die aus derzeitigen Entbindungs- und gynäkologischen Untersuchungen stammten. 42 Proben wurden von schwangeren Frauen erhalten, und die restlichen Proben stammten von Patienten mit unterschiedlichen Krankheiten und Gegebenheiten (AIDS, Anämie, Arthritis, Müdigkeitssyndrom, Kopfschmerzen, Lymphom, Meningitis, Lungenentzündung und andere Krankheiten). Es handelte sich um 37 männliche und 162 weibliche Patienten im Alter von 15 bis 68 Jahren. Eine der retrospektiven Studien wurde an einer Medizinischen Universität im Nordwesten der USA durchgeführt. Die 199 Blutproben wurden mit dem IMMULITE Toxoplasma Quantitativ IgG Test und einem kommerziell erhältlichen
  • 15. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 15 Toxoplasma IgG Enzymfluoresz Enzimmunoassay (ELFA) evaluiert. IMMULITE Toxoplasma Quantitativ IgG Kit A Positiv Negativ Relative Sensitivität Relative Spezifität Positiv 52 2 Negativ 0 145 96,3% 100% Übereinstimmung: 99,0% 95% Vertrauensbereich für die relative Sensitivität und Spezifität (resp.: 87,3% - 99,5% und 97,5% - 100%). Die zwei diskrepanten Blutproben, negativ im IMMULITE, aber positiv im Kit A, wurde desweiteren mit einem kommerziell erhältlichen Toxoplasma IgG Mikropartikelenzymimmunotest (MEIA) evaluiert und als positiv gemessen. In der zweiten retrospektiven Studie, die außerhalb der USA durchgeführt wurde, wurden von 199 Proben 194 (5 Proben mit nicht ausreichendem Probenvolumen) mit dem IMMULITE Toxoplasma Quantitativ IgG-Test und Kit B evaluiert. IMMULITE Toxoplasma Quantitativ IgG Kit B Positiv Negativ Relative Sensitivität Relative Spezifität Positiv 46 5 Negativ 2 141 90,2% 98,6% Übereinstimmung: 96,4% 95% Vertrauensbereich für die relative Sensititvät und Spezifität, resp.: 78,6% - 96,7% und 95% - 99,8%. Von den 7 diskrepanten Blutproben waren 3 der 5 negativ gefundenen Proben im IMMULITE und im Kit B positiv, negativ im Kit A. Zwei der positiv gemessenen Blutproben im IMMULITE, aber negativ gemessenen im Kit B, waren positiv im Kit A. In der dritten retrospektiven Studie, die ebenfalls außerhalb der USA durchgeführt wurde, wurden von 199 Proben 160 (39 Blutproben mit nicht ausreichendem Probenvolumen) mit dem IMMULITE Toxoplasma IgG-Test und einem kommerziell erhältlichen Enzymimmunoassay Kit C, evaluiert. IMMULITE Toxoplasma Quantitativ IgG Kit C Positiv Negativ Relative Sensitivität Relative Spezifität Positiv 32 2 grenzwertig 0 4 Negativ 2 120 94,1% 98,4% Übereinstimmung: 97,4% 95% Vertrauensbereich für die relative Sensitivität und Spezifität, resp.: 80,3% – 99,3% und 94,2% – 99,8%. In zwei Fällen mit positivem IMMULITE und negativem Kit C Ergebnis, waren ebenfalls positiv in Kit A. In den zwei Fällen, die negativ im IMMULITE und positiv in Kit C waren, war eine Probe positiv und eine negativ im Kit A. In vier Fällen mit negativem IMMULITE und grenzwertigen Ergebnis in Kit C, waren alle Proben negativ in Kit A. Demnach stimmen die IMMULITE Ergebnisse in sieben von acht Fällen mit Kit A überein, die diskrepant im Kit C liefen. Anwendungsberatung Bei Rückfragen wenden Sie sich bitte an Ihre DPC Niederlassung. Das Qualitätssystem der Diagnostic Products Corporation ist nach ISO 13485:2003 registriert. Español Toxoplasma Cuantitativo IgG Utilidad del análisis: Para su uso en el diagnóstico in vitro con los analizadores IMMULITE e IMMULITE 1000 — para la medición cuantitativa de anticuerpos IgG contra Toxoplasma gondii (toxoplasma), como una ayuda en la determinación del estado serológico para Toxoplasma gondii. Este kit no tiene autorización de la FDA para su uso en pruebas (detección precoz) de donantes de sangre o plasma. Números de Catálogo: LKTXP1 (100 tests), LKTXP5 (500 tests) Código del Test: TXP Color: Gris claro
  • 16. 16 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) Resumen y Explicación del Test Toxoplasma gondii es un parásito intracelular obligado, capaz de infectar a la mayoría de los mamíferos, incluyendo al humano. El organismo se transmite por ingestión de carne poco cocinada. Por todo el mundo, del 1 al 90% de la población puede ser infectada, 7 con un 25 – 30 por ciento de la población adulta de Estados Unidos infectada. 3 Mientras que la toxoplasmosis puede manifestarse de diversas formas, las infecciones son normalmente clínicamente inaparentes, y las infecciones latentes persisten normalmente durante toda la vida. 2 Los síntomas clínicos son similares a los de una infección por mononucleosis, con linfadenopatía, fiebre, dolor de cabeza, malestar general y algunas veces neumomía y miocarditis. 6 Al igual que otras infecciones latentes, la infección aguda por toxoplasma puede presentar una serie amenaza para los individuos inmunocomprometidos y los recién nacidos, quienes adquieren la infección en el utero. Los individuos inmunosuprimidos pueden desarrollar encefalitis, miocarditis o neumonitis. 3 Las infecciones congénitas normalmente resultan como consecuencia de una infección aguda asintomática en la madre. Esta infección puede causar: parto prematuro, aborto espontáneo o el nacimiento sin vida del feto. 4,6 Los neonatos pueden presentar coriorretinitis, hidrocefalia, microcefalía, calcificación cerebral y retraso psicomotor. 1 La mayoría de los niños infectados congénitamente no mostrarán ningún síntoma hasta más tarde. 4 El seguimiento de la toxoplasmosis requiere el control serológico de los individuos infectados, 3 ya que el cultivo de este organismo no es posible. Un análisis cuantitativo para estudiar la presencia de IgG contra toxoplasma puede ser útil para determinar una infección anterior e indicar la reactivación de la infección. Una información precisa del diagnóstico es importante, especialmente durante el embarazo, puesto que el tratamiento con espiramicina puede reducir los riesgos para el feto. 5 Principio del análisis IMMULITE/IMMULITE 1000 Toxoplasma Cuantitativa IgG es un inmunoanálisis quimioluminiscente en fase sólida secuencial. Ciclos de incubación: 2 × 30 minutos. Recogida de la muestra Se recomienda el uso de una ultracentrífuga para aclarar las muestras lipémicas. Las muestras que estén turbias o tengan un material particular deben aclararse mediante una centrifugación a baja velocidad. La centrifugacion de las muestras de suero antes de que se forme el coagulo puede ocasionar la presencia de fibrina. Para evitar resultados erroneos debidos a la presencia de fibrina, asegurarse que se ha formado el coagulo completamente antes de centrifugar las muestras. Algunas muestras, particularmente aquellas de pacientes sometidos a terapia anticoagulante, pueden requerir mayor tiempo de coagulacion. Los tubos para recoger sangre de distintos fabricantes pueden producir valores diferentes, dependiendo del material del tubo y de los aditivos, incluyendo barreras de gel o barreras físicas, activadores de la coagulación y/o anticoagulantes. El IgG Cuantitativo para Toxoplasma IMMULITE/IMMULITE 1000 no ha sido analizado con todos los distintos tipos de tubos. Para obtener detalles sobre los tipos tubos que se han analizado, consulte la sección de Tipos de Muestras Alternativos. Volumen requerido: 10 µl de la muestra de paciente prediluida (1 por 21 dilución). (El recipiente de la muestre debe contener, como mínimo, 100 µl más que el volumen total requerido). Conservación: 3 días a 2–8°C, o 6 meses a –20°C. 13 Dilución de Muestras de Pacientes: Las muestras de suero deben ser prediluídas 1-en-21 con Diluyente de Muestra IgG/IgM (LIGZ2). Esta dilución debe ser realizada manualmente, p.e. añadiendo 20 µl de muestra del paciente con 400 µl de Diluyente de Muestra (LIGZ2, suministrados en el kit).
  • 17. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 17 Como alternativa, utilizando IMMULITE 1000 Windows ® , la dilución la realiza el sistema automaticamente en el equipo usando LIGW, incluido en el Modulo de Diluyente de Muestra IgG/IgM (L1KIGW 1,5). Los resultados imprimidos han sido corregidos para una dilución de 1-en-21. Advertencias y Precauciones Para uso diagnóstico in vitro. Reactivos: Mantener a 2–8°C. Desechar de acuerdo con las normas aplicables. Siga las precauciones universales y manipule todos los componentes como si fueran capaces de transmitir agentes infecciosos. Los materiales derivados de sangre humana han sido analizados y son negativos para sífilis; para anticuerpos frente al HIV 1 y 2; para el antígeno de superficie de hepatitis B y para los anticuerpos de hepatitis C. Se ha usado Azida sodica, en concentraciones menores de 0,1 g/dl, como conservante. Para su eliminacion, lavar con grandes cantidades de agua para evitar la constitucion de residuos de azidas metalicas, potencialmente explosivas, en las cañerías de cobre y plomo. Sustrato quimioluminiscente: evite la contaminación y exposición a la luz directa del sol. (Ver el prospecto.) Agua: Usar agua destilada o desionizada. La bola está recubierta con antígeno inactivado de toxoplasma. Sin embargo, se recomienda precaución ante la posible presencia de organismo residual cuando trabajamos o deshechamos los materiales suministrados. La concentración de IgG contra Toxoplasma en una muestra dada, determinada mediante ensayos de distintos fabricantes, puede variar debido a diferencias en los métodos de los ensayos y a la especificidad del reactivo. Por lo tanto, en el informe de resultados del laboratorio para el médico debe incluirse: "Los siguientes resultados se obtuvieron con EIA IgG Cuantitativo contra Toxoplasma IMMULITE/IMMULITE 1000. No es posible intercambiar los resultados obtenidos por métodos de ensayo de otros fabricantes". Materiales Suministrados Los componentes representan un juego completo. Las etiquetas incluidas en la caja son necesarias para el ensayo. Unidades de análisis de Toxoplasma Cuantitativo IgG (LTXP1) Cada unidad etiquetada con código de barras contiene una bola recubierta de antígeno inactivado, parcialmente purificado de Toxoplasma gondii (taquizoitos de cepa RH de peritoneo de ratón). Estable a 2–8°C hasta la fecha de caducidad. LKTXP1: 100 unidades. LKTXP5: 500 unidades. Espere a que las bolsas de las unidades de análisis alcancen la temperatura ambiente antes de abrirlas. Ábralas cortando por el extremo superior, dejando el borde del cierre de cremallera intacto. Vuelva a cerrar las bolsas herméticamente para protegerlas de la humedad. Viales de reactivo de Toxoplasma Cuantitativo IgG (LTXPA, LTXPB) Con códigos de barras. LTXPA: 7,5 ml de una matriz de proteína en solución tampón, con conservante. LTXPB: 7,5 ml de fosfatasa alcalina (de intestino de ternera) conjugada a anticuerpo monoclonal murino anti-IgG humana, en una solución tampón, con conservante. Guardar tapado y refrigerado: estable a 2–8°C hasta la fecha de caducidad. Se recomienda utilizarlo antes de que pasen 30 días después de abrirlo cuando se guarda según lo indicado. LKTXP1: 1 juego. LKTXP5: 5 juegos. Ajustadores de Toxoplasma Cuantitativo IgG (LTXPL, LTXPH) Dos viales (bajo y alto) de cada uno con 4 ml de suero humano con IgG reactivo para Toxoplasma, en solución tampón, con conservante. (No es necesaria ninguna dilución previa.) Estable a 2–8°C durante 14 días después de abrise, o hasta 6 meses (alicuotados) a –20°C. LKTXP1: 1 juego. LKTXP5: 2 juegos. Controles de Toxoplasma Cuantitativo IgG (LTXPC1, LTXPC2, LTXPC3) Tres viales (Negativo, Bajo Positivo y Positivo) de cada uno con 2 ml. LTXPC1 (Control negativo): Un vial que contiene
  • 18. 18 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) suero humano con IgG no reactivo para toxoplasma, con conservante. LTXPC2, LTXPC3 (Control bajo positivo, Control Positivo): contienen suero humano con IgG reactivo para toxoplasma, con conservante. Estable a 2–8°C durante 14 días después de abrise, o hasta 6 meses (alicuotados) a –20°C. LKTXP1: 1 juego. LKTXP5: 2 juegos. IMMULITE: Los controles de Toxoplasma IgG requieren la misma predilución manual 1-en-21 que las muestras de pacientes, con la diluyente de muestra para IgG e IgM (LIGZ2). IMMULITE 1000: El software del IMMULITE 1000 realiza una dilución automática en el instrumento de las muestras de pacientes y de los controles. Introduzca los controles como controles. Consulte los niveles de concentración en el prospecto del control. Diluyente de IgG/IgM (LIGZ2) Para la dilución manual de muestras de pacientes y controles. 100 ml matriz proteica no humana en solución tampón, con conservante. Estable a 2–8°C durante 30 días después de abrise o estable a –20°C durante 6 meses (alicuotados). LKTXP1: 1 vial. LKTXP5: 3 viales. Para diluciones de muestras de pacientes dentro del equipo usando IMMULITE 1000 Windows ® , remitirse a la sección del Kit de Componentes Suministrados Aparte. Componentes del kit que se suministran por separado Módulo de Diluyente de Muestra IgG/IgM (L1KIGW1,5) Requerido para la dilución de muestras de pacientes y controles en el equipo utilizando IMMULITE 1000 Windows ® . Los Viales (LIGW) de Diluyente de Muestras contienen una matriz concentrada de proteína/buffer (2.5X) no-humana, con conservante. Estable a 2-8°C hasta la fecha de caducidad. L1KIGW1: 2 Viales. L1KIGW5: 10 Viales. LSUBX: Substrato quimioluminiscente LPWS2: Lavado de sonda LKPM: Kit de limpieza de sonda LCHx-y: Soportes de recipientes de muestras (con códigos de barras) LSCP: Recipientes de muestras (desechables) LSCC: Tapas para los recipientes de muestras (opcionales) DCHS: Cubetas de dilución (con código de barras) para diluciones en el equipo (IMMULITE 1000 Windows ® ). También necesario Pipetas de transferencia de muestras; agua destilada o desionizada. Ensayo Aviso: para obtener el funcionamiento óptimo, es importante realizar todos los procedimientos del mantenimiento general según lo definido en el manual del operador de IMMULITE o IMMULITE 1000. Ver el Manual del Operador del IMMULITE o IMMULITE 1000 para: preparación, procesamiento, diluciones, ajuste, procedimientos de ensayo y control de calidad. Inspeccionar visulamente cada unidad de rección para asegurarse de que hay una bola antes de introducirla en el Sistema. Tenga en cuenta que hay que cargar los viales de reactivos A y B en el carrusel para ejecutar este ensayo. Para el IMMULITE 1000 Windows ® , el Vial del Diluyente de la Muestra también debe estar cargado en el carrusel. Intervalo de ajuste recomendado: 4 semanas. Muestras de Control de Calidad: Los controles provistos con el kit deberán usarse como control de calidad para monitorear el rendimiento del ensayo. Los controles positivo y negativo están diseñados para controlar un fallo importante del reactivo. El control positivo proporcionado no asegurará la precísión del valor de corte. Además de los controles provistos, los usuarios pueden desear correr controles adicionales para cubrir el intervalo superior del ensayo. Pueden analizarse otros controles según las directrices o requisitos de las normas locales, estatales o federales o de las organizaciones acreditadas. 12 También se recomienda analizar periódicamente muestras conocidas, reactivas o no reactivas, para asegurar la
  • 19. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 19 precisión del pipeteado para el paso de dilución. Interpretación de los Resultados El ensayo IgG Cuantitativo para Toxoplasma IMMULITE/IMMULITE 1000 está estandarizado en términos del Tercer Estándar Internacional para Suero de Anti-Toxoplasma de la Organización Mundial de la Salud. El valor de corte del ensayo se determinó utilizando muestras positivas y negativas representativas (confirmadas por el método ELISA) utilizando la técnica ROC (Curva de rendimiento diagnóstico). Positivo: Un resultado superior o igual a 8 IU/ml es indicativo de una infección pasada. Negativo: Un resultado inferior a 6,5 IU/ml es considerado negativo. Resultados inferiores a 6,5 IU/ml deben comunicarse solamente como Negativos, y no debe comunicarse la concentración. Indeterminado: Un resultado superior o igual a 6,5 IU/ml e inferior a 8 IU/ml se considera indeterminado. Las muestras con valores superiores a 250 IU/ml deben comunicarse como “superiores a 250 IU/ml". Las muestras que dieron resultados superiores a 250 IU/ml, cuando se examinan con la autodilución necesaria 1-en-21, pueden necesitar diluciones adicionales con el diluyente para muestras IgG/IgM. La presencia de anticuerpos IgG contra Toxoplasma gondii es indicativo de una exposición previa al organismo. Puede utilizarse sólo una única muestra para estimar el estado serológico del individuo. Un resultado “indeterminado” indica un nivel intermedio de anticuerpos contra Toxoplasma gondii en la muestra: la muestra debe volverse a analizarse. Las muestras que vuelvan a dar un resultado “indeterminado” deben examinarse por un método alternativo, o debe recogerse una segunda muestra, si es posible, en un plazo de tiempo razonable (p.e. una semana). No se recomienda interpretar los resultados si el control positivo o negativo cae fuera del margen especificado en la Sección Control de Calidad. La magnitud del nivel de IgG comunicado no puede relacionarse con un título final. Por lo tanto, en el informe de resultados del laboratorio para el médico debe incluirse: "Los siguientes resultados se obtuvieron con EIA IgG Cuantitativo contra Toxoplasma IMMULITE/IMMULITE 1000." No es posible intercambiar los resultados obtenidos por métodos de ensayo de otros fabricantes. Valores Esperados Individuos con el organismo toxoplasma pueden no mostrar niveles de anticuerpo IgG detectables en las fases tempranas de la infección. Los niveles de IgG para Toxoplasma gondii empiezan a elevarse 1 o 2 semanas después de la infección. Los niveles máximos se alcanzan en 6 – 8 semanas, después descienden gradualmente en los meses y años posteriores. Se detectan títulos bajos generalmente durante toda la vida. El título de anticuerpo no guarda relación con la severidad de la enfermedad. 2,4,6 La preponderancia de anticuerpos IgG anti-Toxoplasma gondii variará con la edad y la zona geográfica. Se condujeron estudios con muestras tomadas de sujetos presumiblemente sanos y asintomáticos del noroeste de los EE.UU. Este grupo de personas se compuso de 69 individuos de 15 a 64 años de edad. (Las muestras fueron previamente recolectadas y guardadas congeladas a – 20ºC). Se encontró que la preponderancia de reactividad al toxoplasma, basada en los resultados del ensayo cuantitativo IgG anti -Toxoplasma IMMULITE, fue de 56% en esta zona geográfica. Se realizaron estudios adicionales con muestras provenientes de 42 mujeres embarazadas de 16 a 41 años de edad de la misma zona geográfica. (Las muestras fueron previamente recolectadas y guardadas congeladas a –20º C). La preponderancia de reactividad al toxoplasma para esta población específica fue del 5%, basándose en los resultados del ensayo cuantitativo IgG anti -Toxoplasma IMMULITE. (Ver el gráfico de “Expected Values”), Estos límites han de considerarse sólo como una guía. Cada Laboratorio deberá establecer sus propios rangos de referencia.
  • 20. 20 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) Limitaciones El uso de IgG Cuantitativo Toxoplasma IMMULITE/IMMULITE 1000 para diagnosticar una infección reciente analizando parejas de sueros no se ha validado. Los resultados del análisis deben contemplarse en el contexto del historial clínico de los pacientes, de su sintomatología y de los demás hallazgos del laboratorio. La presencia de anticuerpos IgG en una sola muestra no es suficiente para distinguir entre una infección activa y una antigua. Los pacientes de los que se sospeche que padecen una infección primaria o activa deben analizarse para detectar la presencia de anticuerpos IgM para Toxoplasma gondii. Para determinar la seroconversión de no reactivo a reactivo, debe recogerse una segunda muestra de suero tres o cuatro semanas después de la infección, muestra no reactiva, durante la etapa de convalecencia de la infección. La muestra de la fase aguda debe almacenarse y analizarse paralelamente a la muestra de la fase de convalecencia. Los individuos con infección aguda por toxoplasma puede que no presenten niveles detectables de anticuerpos IgG en la fase temprana de la infección. Los resultados en pacientes infectados con el virus VIH, pacientes bajo una terapia inmunosupresiva, o en pacientes con otros desórdenes que causen inmunosupresión, deben interpretarse con precaución. No se han establecido las características de funcionamiento del ensayo para utilizarlo con muestras de neonatos, sangre del cordón umbilical, pacientes pretrasplante, u otros fluidos distintos del suero, tales como orina, saliva o líquido amniótico. Los anticuerpos heterofílicos en el suero humano pueden reaccionar con las inmunoglobulinas de los componentes del ensayo provocando interferencias con los inmunoanálisis in vitro. [Ver Boscato LM, Stuart MC. Heterophilic antibodies: a problem for all immunoassays. Clin Chem 1988:34:27-33.] Las muestras de los pacientes que frecuentemente están expuestos a animales o a productos séricos animales pueden presentar este tipo de interferencia que potencialmente ocasione un resultado anómalo. Estos reactivos han sido formulados para minimizar el riesgo de interferencia, no obstante, pueden darse interacciones anómalas entre sueros conflictivos y los componentes del ensayo. Con fines de diagnóstico, los resultados obtenidos con este ensayo siempre deben ser usados en combinación con el examen clínico, la historia médica del paciente y cualquier otro dato clínico relevante. Características Analíticas Para ver resultados representativos de las cualidades del ensayo ver las tablas y los gráficos. Los resultados se expresan en IU/ml. (A no ser que se indique lo contrario, todos los resultados fueron generados en muestras de suero recogidas en tubos sin geles o activadores de la coagulación). Intervalo de calibración: 5,0 – 250 IU/ml. Normalizado en términos del tercer estándar WHO IS para Suero Humano Anti-Toxoplasma. Sensibilidad: ≤ 5,0 IU/ml Precisión: Se han calculado datos estadísticos para las muestras a partir de los resultados de 5 replicados, cada uno en cinco días separados. (Ver las tablas de "Intraassay Precision" y “Interassay Precision”). Note que los controles negativos de la Muestra 1 siempre se encontraron por debajo de 5 IU/ml (el límite inferior del intervalo de calibración del ensayo) para todos los replicados de cada corrida. Especificidad: Se evaluó la especificidad del ensayo cuantitativo IgG anti- Toxoplasma IMMULITE analizando 50 muestras positivas para citomegalovirus (CMV), virus de varicella-zoster (VZV), virus Epstein-Barr (EBV), Mycoplasma penumoniae y parvovirus B19. En estas muestras, el ensayo cuantitativo IgG anti- toxoplasma IMMULITE y un inmunoensayo enzimático para IgG anti- Toxoplasma comercialmente disponible (Kit D) mostraron 96% de concordancia (48/50). Tipo de Muestra Alternativa: para evaluar el efecto de los diferentes tipos de muestras alternativos, se recogió sangre
  • 21. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 21 de 16 voluntarios en tubos normales, tubos con Heparina y tubos vacutainer SST ® de Becton Dickinson. Todas las muestras se ensayo de IgG Cuantitativo para Toxoplasma de IMMULITE, obteniendo los siguientes resultados. (Heparina) = 1,02 (Suero) + 0,13 IU/ml r = 1,00 (SST) = 0,98 (tubos simples) – 0,06 IU/ml r = 1,00 Medias: 4,4 IU/ml (Suero) 4,6 IU/ml (Heparina) 4,2 IU/ml (SST) En otro estudio, la sangre de 16 voluntarios se recogió en tubos normales, y EDTA vacutainer. Todas las muestras se ensayo de IgG Cuantitativo para Toxoplasma de IMMULITE, obteniendo los siguientes resultados. (EDTA) = 0,93 (Suero) – 0,003 IU/ml r = 1,00 Medias: 5,3 IU/ml (EDTA) 5,6 IU/ml (Suero) Rendimiento Clínico Se estudió el rendimiento clínico del ensayo cuantitativo IgG anti-Toxoplasma IMMULITE, usando un total de 199 muestras de suero. Cada una de estas muestras fue analizada en tres sitios, dos de los cuales se encuentran fuera de los EE.UU. De las 199 muestras (previamente recolectadas y guardadas congeladas a –20°C ), 61 se obtuvieron de mujeres que se presentaron para un examen obstétrico y ginecológico, 42 se obtuvieron de mujeres embarazadas y el resto se obtuvo de pacientes con distintas enfermedades y afecciones (SIDA, anemia, arteritis, fatiga, dolor de cabeza, linfoma, meningitis, neumonía y otros). El grupo consistió de 37 hombres y 162 mujeres de 15 a 68 años de edad. Uno de los estudios retrospectivos fue conducido en el centro médico de una universidad en el noroeste de los EE.UU. Las 199 muestras fueron evaluadas en forma no simultánea con el procedimiento cuantitativo IgG anti-Toxoplasma IMMULITE y el Kit A, un ensayo de inmunofluorescencia ligado a una enzima para IgG anti -Toxoplasma (ELFA) comercialmente disponible. Ensayo Cuantitativo para IgG anti- Toxoplasma IMMULITE Kit A Pos Neg Sensibilidad relativa Especificidad relativaa Positivo 52 2 Negativo 0 145 96,3% 100% Concordancia: 99,0% 95% de límites de confianza para la sensibilidad y la especificidad relativas: 87,3% – 99,5% y 97,5% – 100%. (Límites de confianza mediante análisis exacto.) Las dos muestras discordantes, negativas para IMMULITE pero positivas para el Kit A, fueron evaluadas con el Kit B, un inmunoensayo de micropartículas para IgG anti-Toxoplasma (MEIA) comercialmente disponible, el cual dio un resultado positivo. En el segundo estudio retrospectivo que se realizó fuera de los EE.UU., se evaluaron194 de las 199 muestras (5 muestras tuvieron insuficiente volumen de muestra) con el ensayo cuantitativo IgG anti -Toxoplasma IMMULITE y el Kit B. Ensayo Cuantitativo IgG anti- Toxoplasma IMMULITE Kit B Pos Neg Sensibilidad relativa Especificidad relativaa Positivo 46 5 Negativo 2 141 90,2% 98,6% Concordancia: 96,4% 95% de límites de confianza para la sensibilidad y la especificidad relativas: 78,6% – 96,7% y 95,0% – 99,8%. (Límites de confianza mediante análisis exacto.) De las siete muestras discordantes, tres de las cinco que dieron negativas con IMMULITE pero positivas con el Kit B dieron negativas con el Kit A. Las dos muestras que dieron positivas con IMMULITE pero negativas con el Kit B dieron positivas con el Kit A. En el tercer estudio retrospectivo, también realizado fuera de los EE.UU., 160 de las 199 muestras (39 muestras tuvieron insuficiente volumen de muestra) fueron analizadas con el ensayo cuantitativo IgG anti-Toxoplasma IMMULITE y el Kit C, un inmunoensayo enzimático comercialmente disponible.
  • 22. 22 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) Ensayo Cuantitativo IgG anti- Toxoplasma IMMULITE Kit C Pos Neg Relative Sensitivity Relative Specificity Positivo 32 2 Ambiguo 0 4 Negativo 2 120 94,1% 98,4% Concordancia: 97,4% 95% de límites de confianza para la sensibilidad y la especificidad relativas: 80,3% – 99,3% y 94,2% – 99,8%. (Límites de confianza mediante análisis exacto). Los dos casos positivos para el ensayo IMMULITE pero negativos para el Kit C dieron positivos para el Kit A. De los dos casos negativos para el ensayo IMMULITE pero positivos para el Kit C, uno dio positivo y el otro dio negativo con el Kit A. Los cuatro casos negativos para el ensayo IMMULITE, pero ambiguos para el Kit C, dieron todos negativos con el Kit A. Por lo tanto, los resultados de IMMULITE concuerdan con los del Kit A en siete de los ocho casos en que no hay concordancia con el Kit C. Asistencia técnica Contáctese con su Distribuidor Nacional. El Sistema de Calidad de Diagnostic Products Corporation está registrado para la ISO 13485:2003. Français IMMULITE/IMMULITE 1000 Toxoplasmose quantitative IgG Domaine d'utilisation : Dosage quantitatif des IgG spécifiques dirigées contre Toxoplasma gondii dans le sérum. Réservé à un usage diagnostique in vitro avec l'Analyseur IMMULITE et de IMMULITE 1000, ce test constitue une aide à la détermination du statut sérologique vis-à-vis de Toxoplasma gondii. Ce coffret n'est pas habilité auprès de la Food and Drug Administration américaine pour le dépistage des donneurs de sang ou de plasma. Ce réactif est enregistré auprès de l'AFSSAPS. Référence catalogue : LKTXP1 (100 tests), LKTXP5 (500 tests) Code produit : TXP Code couleur : gris clair Introduction Toxoplasma gondii est un parasite intracellulaire capable d'infecter la plupart des mammifères, y compris l'homme. Le parasite se transmet par l'ingestion de viande insuffisamment cuite. Dans le monde, 1 à 90 % des populations est susceptible d'être infecté, 7 25 à 30 % de la population adulte étant infecté au Etats- Unis. 3 Pour la France, 30 à 60 % de la population est infectée. Même si la toxoplasmose peut se manifester sous plusieurs formes, les individus infectés restent généralement asymptomatiques et une infection latente persiste habituellement toute la vie. 2 Les symptômes cliniques visibles sont similaires à ceux de la mononucléose infectieuse, avec une lymphadénopathie, de la fièvre, des céphalées, une sensation de malaise et, parfois, une pneumonie et une myocardite. 6 Comme toute infection latente, une infection aiguë par Toxoplasma gondii peut représenter un risque sérieux pour un individu immunodéprimé ou pour un nouveau-né infecté in utero. Les patients immunodéprimés infectés sont susceptibles de développer une encéphalite, une myocardite ou une pneumopathie inflammatoire. 3 Les infections congénitales proviennent habituellement d'une infection maternelle aiguë asymptomatique. Cette infection peut se traduire par une naissance prématurée, un avortement spontané ou la naissance d'un enfant mort-né. 4,6 Le nouveau-né peut présenter une choriorétinite, une hydrocéphalie, une microcéphalie, des calcifications intra- cérébrales et un retard psychomoteur. 1 La plupart des enfants infectés in utero sont asymptomatiques dans les premiers temps de vie. 4 Dans la mesure où le parasite n'est pas facile à mettre en culture, la prise en charge de la toxoplasmose demande un contrôle sérologique régulier des individus infectés. 3 Un dosage quantitatif d'IgG
  • 23. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 23 spécifiques dirigées contre Toxoplasma gondii peut être utile pour mettre en évidence une infection antérieure et révéler une réactivation de l'infection. Un diagnostic précis est important, en particulier au cours de la grossesse, puisqu'un traitement par la spiramycine pourra réduire les risques encourus par le fœtus. 5 Principe du test IMMULITE/IMMULITE 1000 Toxoplasmose IgG quantitatif est un immunodosage chimiluminescent séquentiel, en phase solide. Cycles d'incubation : 2 × 30 minutes. Recueil des échantillons Il est recommandé de clarifier les échantillons hyperlipémiques par ultracentrifugation. Les échantillons troubles ou présentant des particules en suspension devront être clarifiés par centrifugation à vitesse réduite. La centrifugation des échantillons sériques avant la formation complète du caillot peut entraîner la présence de fibrine. Pour éviter les résultats erronés dûs à la présence de fibrine, s'assurer de la formation complète du caillot avant de centrifuger les échantillons. Certains échantillons, en particulier ceux provenant de patients sous anti-coagulants, peuvent nécessiter un temps plus long pour la formation du caillot. Des tubes pour prélèvements sanguins provenant de fabricants différents peuvent donner des résultats différents, selon les matériaux et additifs utilisés, y compris gels ou barrières physiques, activateurs de la coagulation et/ou anticoagulants. Le coffret Toxoplasmose IgG quantitatif IMMULITE/IMMULITE 1000 n'a pas été testé sur tous les types de tubes possibles. Veuillez consulter le chapitre intitulé Autres Types d'Échantillons pour plus de renseignements sur les tubes qui ont été évalués. Volume nécessaire : 10 µl d 'échantillon pré-dilué (dilution au 1:21). (L'unité- échantillon doit contenir au moins 100 µl de plus que le volume total nécessaire.) Conservation : 3 jours à +2°C/+8°C ou 6 mois à –20°C. 13 Dilution des échantillons de patients : Les échantillons doivent être dilués au 1/21 ème avec du diluant échantillon IgG/IgM (LIGZ2). Cette dilution doit être faite manuellement. Exemple : ajouter 20 µl d'échantillon à 400 µl de diluant échantillon (LIGZ2) fournis avec le coffret. Le système IMMULITE 1000 Windows ® réalise également les dilutions automatiquement, en utilisant le diluant échantillon IgG/IgM (LIGW) fourni dans le coffret (L1KIGW1,5). Les résultats imprimés ont été corrigés en tenant compte d'une dilution au 1/21 ème . Précautions d'emploi Réservé à un usage diagnostique in vitro. Réactifs : conserver les réactifs à +2/ +8 °C. Eliminer les déchets conformément à la réglementation en vigueur. Respecter les précautions d'emploi et manipuler tous les composants du coffret comme des produits potentiellement infectieux. Les réactifs dérivés de produits humains et utilisés dans ce coffret ont subi un test sérologique pour la Syphilis et des tests de dépistage pour les anticorps anti- VIH1 et 2, anti-HCV et pour l'antigène de surface de l'hépatite B, qui se sont tous avérés négatifs. De l'azide de sodium à des concentrations inférieures à 0,1 g/dl a été ajouté comme conservateur ; lors de l'élimination, l'évacuer avec de grandes quantités d'eau pour éviter une accumulation d'azides métalliques explosifs dans les canalisations. Substrat chimiluminescent : éviter la contamination et l'exposition directe au soleil. (Voir notice). Eau : utiliser de l'eau distillée ou désionisée. La bille est revêtue d'antigène de Toxoplasma gondii inactivé. Cependant, des parasites ayant pu subsister, la prudence est recommandée lors de l'utilisation ou de l'élimination des produits fournis. Les résultats d'IgG dirigées contre Toxoplasma gondii obtenus sur un échantillon donné, avec les tests de
  • 24. 24 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) différents fabricants, pourront présenter une disparité en raison de différences de méthodologie et de spécificité des réactifs. Par conséquent, les résultats communiqués au médecin par le laboratoire devront comporter la mention suivante : « Les résultats suivant ont été obtenus avec le dosage IMMULITE/IMMULITE 1000 Toxoplasmose IgG quantitatif. Les résultats obtenus avec des procédés de dosage d'autres fabricants ne doivent pas y être substitués. » Matériel fourni Les composants de la trousse ne peuvent être utilisés que conjointement. Les étiquettes à l'intérieur du coffret sont nécessaires au dosage. Tests unitaires Toxoplasmose quantitative IgG (LTXP1) Avec code-barre. Chaque unité-test contient une bille revêtue d'un antigène de Toxoplasma gondii inactivé, partiellement purifié (tachyzoïtes de souche RH, issus de péritoine de souris). Stable à +2°C/+8°C jusqu'à la date de péremption. LKTXP1 : 100 unités. LKTXP5 : 500 unités. Porter les sachets à température ambiante avant d'ouvrir. Ouvrir le sachet avec des ciseaux en préservant le dispositif de fermeture. Refermer les sachets pour les protéger de l'humidité. Cartouches de réactif Toxoplasmose quantitative IgG (LTXPA, LTXPB) Avec code-barres. LTXPA : 7,5 ml de tampon à base de protéines, avec conservateur. LTXPB : 7,5 ml d'anticorps monoclonal murin anti-IgG humaines marqué à la phosphatase alcaline dans un tampon avec conservateur. Conserver bouchés et réfrigérés : stables à +2°C/+8°C jusqu'à la date de péremption. A utiliser de préférence dans les 30 jours qui suivent l'ouverture, si les recommandations de stockage sont respectées. LKTXP1 : 1 jeu. LKTXP5 : 5 jeux. Ajusteurs Toxoplasmose quantitative IgG (LTXPL, LTXPH) 2 flacons (haut et bas), 4 ml chacun, contenant des IgG anti-Toxoplasma gondii dans du sérum humain avec tampon avec conservateur. (Il n’est pas nécessaire de prédiluer.) Stable à +2°C/+8 °C pendant 14 jours après ouverture ou 6 mois (aliquoté) à –20 °C. LKTXP1 : 1 jeu. LKTXP5 : 2 jeux. Contrôles Toxoplasmose quantitative IgG (LTXPC1, LTXPC2, LTXPC3) Trois flacons (Négatif, Positif Faible et Positif), 2 ml chacun. LTXPC1 (Contrôle négatif) : un flacon de sérum humain ne contenant pas d'IgG dirigées contre Toxoplasma gondii, avec conservateur. LTXPC2, LTXPC3 (Contrôle positif faible, Contrôle positif) : deux flacons contenant des IgG dirigées contre Toxoplasma gondii dans du sérum humain avec conservateur. Stable à +2°C/+8°C pendant 14 jours après ouverture ou 6 mois (aliquoté) à –20°C. LKTXP1 : 1 jeu. LKTXP5 : 2 jeux. IMMULITE : Les contrôles Toxoplasmose IgG nécessite la même prédilution manuelle 1-dans-21 que les échantillons de patients, avec diluant échantillon IgG/ IgM (LIGZ2). IMMULITE 1000 : Le logiciel de l'IMMULITE 1000 réalise automatiquement les dilutions à bord des échantillons de patients et de contrôles. Saisir les contrôles comme des contrôles. Se reporter à la fiche technique du contrôle pour les concentrations. Diluant échantillon IgG/IgM (LIGZ2) Pour la dilution manuelle des échantillons de patients et des contrôles. 100 ml matrice tampon/protéines non-humaines avec conservateur. Stable à +2°C/+8°C pendant 30 jours après ouverture ou à – 20°C pendant 6 mois (aliquoté). LKTXP1 : 1 flacon. LKTXP5 : 3 flacons. Pour la dilution à bord automatique des échantillons de patients, avec IMMULITE 1000 Windows ® , se référer au châpitre « Composants du Coffret fournis séparément ». Composants du coffret fournis séparément Diluant échantillon IgG/IgM (L1KIGW1,5) Requis pour la dilution à bord des échantillons de patients et des contrôles
  • 25. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 25 avec IMMULITE 1000 Windows ® . Les flacons de diluant échantillon (LIGW) contiennent une matrice protéïque non- humaine dans un tampon concentré (2.5X) avec un conservateur. Stable à 2– 8°C jusqu'à la date d'expiration. L1KIGW1 : 2 flacons. L1KIGW5 : 10 flacons. LSUBX : Substrat chimiluminescent LPWSM : Solution de lavage LKPM : Coffret de décontamination de l'aiguille de prélèvement LCHx-y : portoirs à cupules réactionnelles (avec code-barre) LSCP : cupules réactionnelles (à usage unique) LSCC : bouchons pour cupules réactionnelles (optionnel) DCHS : Support pour unités de dilution automatisée par l'IMMULITE 1000 Windows ® . Egalement requis Pipettes pour le transfert des échantillons ; eau distillée ou désionisée. Protocole de dosage Noter que pour des performances optimales, il est important de réaliser toutes les procédures de maintenance de routine selon les instructions du Manuel d'Utilisation de l'IMMULITE ou de l'IMMULITE 1000. Voir le manuel d'utilisation de l'IMMULITE ou de l'IMMULITE 1000 pour la préparation, le démarrage du système, les ajustements, le dosage et les procédures de contrôle de qualité. Vérifier visuellement que chaque Unité- Test contient bien une bille avant de la charger dans l'automate. Noter que les deux flacons de réactif A et B doivent être placés sur le carrousel pour réaliser ce dosage. Pour l'IMMULITE 1000 Windows ® , le flacon contenant le diluant échantillon doit également être chargé sur le carrousel. Intervalle d'ajustement recommandé : 4 semaines. Echantillons pour le contrôle de qualité : Les contrôles fournis dans le coffret doivent utilisés pour le contrôle de qualité des performances du dosage. Les témoins positifs et négatif sont conçus pour détecter un défaut majeur du réactif. Le témoin positif fourni ne pourra pas garantir la précision du seuil. En plus des contrôles fournis, les utilisateurs peuvent souhaiter utiliser des contrôles supplémentaires pour les valeurs élevées du domaine de mesure du dosages. Ces contrôle supplémentaires pourront être testés, conformément aux recommandations ou aux directives des autorités locales ou nationales. 12 Il est également recommandé de tester périodiquement des échantillons positifs ou négatifs connus, pour contrôler l'exactitude du pipetage lors de l'étape de dilution. Interprétations des résultats Le dosage IMMULITE/IMMULITE 1000 Toxoplasmose IgG quantitatif est standardisé par rapport au Troisième Standard International de l'OMS pour les anticorps sériques anti-toxoplasmose. Le seuil du dosage est déterminé par courbe ROC à partir d'échantillons positifs et négatifs représentatifs (confirmé par dosage ELISA). Positif : un résultat supérieur ou égal à 8 UI/ml indique une infection passée. Négatif : un résultat inférieur à 6,5 UI/ml est considéré comme négatif. (Les résultats inférieurs à 6,5 UI/ml doivent être rendus comme Négatif sans mentionner de concentration numérique). Indéterminé : un résultat supérieur ou égal à 6,5 UI/ml et inférieur à 8 UI/ml est considéré comme indéterminé. Une concentration supérieure à 250 UI/ml devra être rapportée comme « supérieure à 250 UI/ml ». Un échantillon dont la concentration, obtenue sur une dilution au 1:21 par l'appareil, est supérieure à 250 UI/ml, devra être dilué d'un facteur supplémentaire avec le diluant échantillon IgG/IgM. La présence d'anticorps IgG anti- Toxoplasma gondii est révélatrice d'un contact antérieur avec le parasite. Un seul échantillon peut uniquement servir à évaluer le statut sérologique de l'individu. Un résultat « indéterminé » indique des taux d'anticorps anti-Toxoplasma gondii intermédiaires : l'échantillon devra donc être dosé de nouveau. Les échantillons qui restent alors « indéterminés » devront
  • 26. 26 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) être testés par une autre méthode, ou un autre prélèvement devra être obtenu – si possible – dans un délai raisonnable (une semaine, par exemple). Toute interprétation de résultats est déconseillée si les contrôles positif ou négatif donne un résultat situé en dehors de l'intervalle de référence défini à la section « Contrôle de qualité ». L'importance du taux d'IgG rapporté n'est pas corrélée à un titrage en point final. Les résultats communiqués au médecin par le laboratoire devront comporter la mention suivante : « Les résultats suivant ont été obtenus avec le dosage IMMULITE/IMMULITE 1000 Toxoplasmose IgG quantitatif. Les résultats obtenus avec des procédés de dosage d'autres fabricants ne doivent pas y être substitués. » Valeurs de référence Il est possible que les IgG spécifiques ne soient pas détectables dans les phases précoces de l'infection par Toxoplasma gondii. En cas de suspicion d'infection au stade précoce, il faudra impérativement effectuer un dosage IgM anti-Toxoplasma gondii. Les taux d'IgG anti-Toxoplasma gondii commencent à augmenter une à deux semaines après la contamination. La réponse anticorps atteint son maximum en 6 à 8 semaines, puis diminue progressivement sur plusieurs mois, voire plusieurs années. Le titre d'anticorps n'est pas corrélé à la gravité de la maladie. 2,4,6 La prévalence des anticorps anti- Toxoplasma gondii varie avec l'âge et la localisation géographique. Des études ont été menées sur des échantillons provenant de sujets asymptomatiques en bonne santé apparente, du nord-ouest des Etats-Unis. Les sujets, au nombre de 69, étaient âgés de 15 à 64 ans. (Les échantillons avaient été recueillis et congelés à –20 °C avant d'être dosés). La prévalence de la séropositivité pour Toxoplasma gondii, sur la base des résultats obtenus par le dosage IMMULITE Toxoplasmose IgG quantitatif, est de 56 % dans cette population. Des études complémentaires ont été menées sur des échantillons prélevés chez 42 femmes enceintes, âgées de 16 à 41 ans, habitant la même région. (Les échantillons avaient été recueillis et congelés à –20 °C avant d'être dosés). La prévalence de la séropositivité pour Toxoplasma gondii, sur la base des résultats obtenus par le dosage IMMULITE Toxoplasmose IgG quantitatif, est de 5 % pour cette population. (Voir le graphique “Expected Values”.) Utiliser ces valeurs à titre indicatif uniquement. Chaque laboratoire devrait établir ses propres valeurs de référence. Limites L'utilisation du dosage IMMULITE/IMMULITE 1000 Toxoplasmose IgG quantitatif pour le diagnostic d'une infection récente en dosant des paires de sérums n'a pas été validée. Les résultats doivent impérativement être interprétés dans le contexte clinique du patient, de la symptomatologie et d'autres résultats de laboratoire. La mise en évidence d'IgG spécifiques dans un échantillon unique ne permet pas de distinguer une infection active d'une infection passée. Les patients suspects de primo-infection ou d'infection active doivent faire l'objet d'une recherche d'IgM anti-Toxoplasma gondii. Pour mettre en évidence une séroconversion, un second échantillon sérique devra être prélevé trois ou quatre semaines après le prélèvement négatif de la phase aiguë, au cours de la phase de convalescence de l'infection. L'échantillon de phase aiguë doit être conservé et testé en parallèle avec l'échantillon de la phase de convalescence. Il est possible que les IgG spécifiques ne soient pas détectables dans la phase précoce d'une infection aiguë par Toxoplasma gondii. Les résultats obtenus chez des patients séropositifs pour le VIH, des patients suivant un traitement immunosuppresseur ou des patients ayant d'autres troubles conduisant à une immunosuppression, devront être interprétés avec prudence. Les performances de ce dosage n'ont pas été évaluées pour l'utilisation d'échantillons provenant de sang de cordon, de nouveau-nés, de patients allant recevoir une transplantation ou pour des liquides biologiques autres que le
  • 27. IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) 27 sérum, comme l'urine, la salive ou le liquide amniotique. Les anticorps hétérophiles du sérum humain peuvent réagir avec les immunoglobulines faisant partie des composants du coffret et interférer avec les immunodosages in vitro. [Voir Boscato LM, Stuart MC. Heterophilic antibodies: a problem for all immunoassays. Clin Chem 1988:34:27-33.] Les échantillons provenant de patients fréquemment exposés aux animaux ou aux produits sériques d'origine animale peuvent présenter ce type d'interférence pouvant potentiellement donner un résultat anormal. Ces réactifs ont été mis au point afin de minimiser le risque d'interférence, cependant des interactions potentielles entre des sérums rares et les composants du test peuvent se produire. Dans un but diagnostique, les résultats obtenus avec ce dosage doivent toujours être utilisés en association avec un examen clinique, l'histoire médicale du patient et d'autres résultats. Performances du test Consulter les tableaux et graphiques pour obtenir les données représentatives des performances du test. Les résultats sont donnés en UI/ml. (En l'absence de précision supplémentaire, tous les résultats ont été obtenus sur des échantillons sériques prélevés sur tubes sans gel, ni activateur de la coagulation). Domaine de mesure : 5,0 – 250 UI/ml. Standardisé par rapport au Troisième Standard International de l'OMS pour les anticorps sériques anti-toxoplasmose chez l'homme. Sensibilité analytique : ≤ 5,0 UI/ml. Précision : les statistiques ont été réalisées sur les résultats de 5 doublets testés chacun pendant 5 jours. (Voir le tableau « Intraassay Precision » et «Interassay Precision».) N.B. : le contrôle négatif et l'échantillon n° 1 étaient toujours inférieurs à 5 UI/ml (la limite inférieure du domaine de mesure pour tous les doublets de chaque série). Spécificité : La spécificité du test IMMULITE Toxoplasmose IgG quantitatif a été évaluée sur 50 échantillons positifs pour le cytomégalovirus (CMV), le virus de la varicelle et du zona (VZV), le virus Epstein-Barr (EBV), pour Mycoplasma pneumoniae et le parvovirus B19. Avec ces échantillons, le dosage IMMULITE Toxoplasmose IgG quantitatif et un immunodosage pour le dosage des IgG anti-Toxoplasmose (coffret D) avaint une concordance de 96% (48/50). Autres types d'échantillons : pour estimer l'effet de l'utilisation de différents type d'échantillons, 16 volontaires ont été prélevés sur tubes secs, héparinés et sur tubes vacutainer SST ® Becton Dickinson. Tous les échantillons puis dosés selon le protocole Toxoplasmose IgG quantitatif IMMULITE, avec les résultats suivants. (Hépariné) = 1,02 (Sérum) + 0,13 UI/ml r = 1,00 (SST) = 0,98 (tubes ordinaires) – 0,06 UI/ml r = 1,00 Moyennes : 4,4 UI/ml (Sérum) 4,6 UI/ml (Hépariné) 4,2 UI/ml (SST) Dans une autre étude, le sang de 16 volontaires a été prélevé sur tubes vacutainer en ordinaire et EDTA. Tous les échantillons puis dosés selon le protocole Toxoplasmose IgG quantitatif IMMULITE, avec les résultats suivants. (EDTA) = 0,93 (Sérum) – 0,003 UI/ml r = 1,00 Moyennes : 5,3 UI/ml (EDTA) 5,6 UI/ml (Sérum) Performance clinique La performance clinique du dosage IMMULITE Toxoplasmose IgG quantitatif a été étudiée sur un total de 199 échantillons sériques, dosés chacun par trois centres investigateurs, dont deux situés en dehors des Etats-Unis. Sur les 199 échantillons (recueillis et congelés à -20 °C avant d'être dosés), 61 avaient été prélevés chez des femmes venues consulter pour un examen obstétrique ou gynécologique, 42 avaient été prélevés chez des femmes enceintes et le reste provenait de patients présentant divers états cliniques (SIDA, anémie, artérite, fatigue, céphalées, lymphome, méningite, pneumonie, etc.). On comptait 37 hommes et 162 femmes, âgés de 15 à 68 ans. Une des études rétrospectives a été menée dans un centre hospitalo- universitaire du Nord-ouest des Etats-
  • 28. 28 IMMULITE/IMMULITE 1000 Toxoplasma Quantitative IgG (PILKTXP-12, 2006-08-08) Unis. Les 199 échantillons ont été analysés avec le dosage IMMULITE Toxoplasmose IgG quantitatif et avec le coffret A, un dosage immunofluorescent (ELFA) du commerce. IMMULITE Toxoplasmose IgG Quantitatif Coffret A Pos Nég Sensibilité relative Spécificité relative Positif 52 2 Négatif 0 145 96,3% 100% Concordance : 99,0% Limites de confiance à 95% pour la sensibilité relative et la spécificité relative, respectivement: 87,3% – 99,5% et 97,5% – 100%. (Limites de confiance par test exact.) Les deux échantillons discordants, négatifs selon IMMULITE, mais positifs selon le coffret A, ont retestés avec le coffret B, un dosage immuno-enzymatique à microparticules (MEIA) disponible dans le commerce, et trouvés positifs. Dans une seconde étude rétrospective, menée en dehors des Etats-Unis, 194 échantillons sur 199 (5 échantillons avaient un volume insuffisant) ont été dosé avec le test IMMULITE Toxoplasmose IgG quantitatif et avec le coffret B. IMMULITE Toxoplasmose IgG quantitatif Coffret B Pos Nég Sensibilité relative Spécificité relative Positif 46 5 Négatif 2 141 90,2% 98,6% Concordance : 96,4% Limites de confiance à 95% pour la sensibilité relative et la spécificité relative, respectivement: 78,6% – 96,7% et 95,0% – 99,8%. (Limites de confiance par test exact.) Parmi les sept échantillons discordants, trois sur cinq, qui étaient négatifs selon IMMULITE mais positifs selon le coffret B, étaient négatifs avec le coffret A. Les deux échantillons positifs selon IMMULITE, mais négatifs selon le coffret B étaient positifs avec le coffret A. Dans la troisième étude rétrospective, également menée en dehors des Etats- Unis, 160 échantillons sur 199 (39 échantillons étaient d'un volume insuffisant) ont été analysés avec le dosage IMMULITE Toxoplasmose IgG quantitatif et avec le coffret C, un dosage immuno-enzymatique du commerce. IMMULITE Toxoplasmose IgG quantitatif Coffret C Pos Nég Sensibilité relative Spécificité relative Positif 32 2 Douteux 0 4 Négatif 2 120 94,1% 98,4% Concordance : 97,4% Limites de confiance à 95% pour la sensibilité relative et la spécificité relative, respectivement: 80,3% – 99,3% et 94,2% – 99,8%. (Limites de confiance par test exact). Les deux échantillons positifs selon IMMULITE, mais négatifs selon le coffret C, étaient positifs avec le coffret A. Parmi les échantillons négatifs selon IMMULITE, mais positifs selon le coffret C, l'un était positif avec le coffret A et l'autre négatif avec le coffret A. Les quatre échantillons négatifs selon IMMULITE, mais douteux selon le coffret C, étaient tous négatifs avec le coffret A. Ainsi, les mesures IMMULITE sont concordantes avec celles du coffret A pour sept des huit cas de divergence avec le coffret C. Assistance technique En France distribué par DPC France 90 bd National 92257 La Garenne-Colombes. Le système d’assurance qualité de DPC est certifié ISO 13485 (2003). Italiano IMMULITE/IMMULITE 1000 Toxoplasma Quantitativo IgG Uso: Ad uso diagnostico in vitro con gli Analizzatori IMMULITE ed IMMULITE 1000 — per la misurazione quantitativa della anticorpi Anti-Toxoplasma gondii nel siero, quale ausilio nella determinazione dello stato sierologico vs.Toxoplasma gondii. Il kit non è stato approvato dall'FDA per lo screening dei donatori di sangue o di plasma.