SlideShare une entreprise Scribd logo
1  sur  56
Télécharger pour lire hors ligne
2014
Spring

Ultimate Measurement
Instruments

General Catalog

HVAC Testing
Indoor Air Quality
Industrial Hygiene
Cleanroom Contamination Control
Industrial Flow Testing
Aerosol Research
Fluid Mechanics Research

Anemometers / Indoor Air Quality Monitors / Gas Monitors / Dust Monitors / Sound Meters / Vibration Meters / Particle Counters
Cleanroom Monitoring System / Aerosol Research Instruments / Mass Spectrometer / Automotive Testing Instruments / Fluid Mechanics Research Instruments

www.kanomax-usa.com
Kanomax Group
Since our inception 60 years ago, Kanomax has been the most promising manufacturer of a broad range of precision
measuring instruments for fluid mechanics research, environmental, aerosol research, particle measurement, and
customized system applications. As a company that prides itself in technology, product quality, and service, we have
been enjoying an unsurpassed reputation in the industrial and academic fields.

Global Network
Our direct subsidiaries and Kanomax’s affiliates and well-trained distributors worldwide are there to provide the most
efficient support and service for you. Our global network is always listening to the voice of customers, like you, in order to
keep providing the best measurement solutions possible.
Kanomax Holdings Inc. (New York, NY)

Kanomax USA Inc. (Andover, NJ)

Kanomax Corporation (Osaka, Japan)

Kanomax Japan Inc. (Osaka-Tokyo-Nagoya, Japan)

Kanomax Instrument Shenyang Inc. (China)

MSI. TOKYO (Tokyo, Japan)

Shenyang Kano Scientific Instrument Co., Ltd (China)

ISO Certification
Kanomax is an ISO 9001/ISO14001 certified company. Kanomax management and
production procedures adhere to these international quality standards.

Kanomax Provides Outstanding Solutions
Environmental Measurements
HVAC Testing
Indoor Air Quality
Industrial Testing

1

Aerosol/Particle Measurements
Cleanroom Contamination Control
Aerosol Research

Fluid Measurements
Fluid Mechanics
Calibration facility ensures accuracy and repeatability
Kanomax fully understands service is an essential part of the total solution provided to our valued customers.
Having already established a worldwide service network, we continuously strive to improve our support services.

For reliable measurements it is recommended that all instruments be calibrated on an
annual basis. This ensures ongoing credibility and accuracy.  ur calibration laboratory
O
in New Jersey maintains the most accurate wind tunnel of its kind. Kanomax provides
the highest quality of service available with a quick turnaround time. Our service
specialists are well trained and will calibrate your instruments to the highest standards.
Our lab certifies Kanomax product to NIST standards.

High Velocity Wind Tunnel

Open Jet Wind Tunnel

Particle Generator

Featured Products

1CFM Flow Rate
P
 ortable Particle Counter

Portable Aerosol Mobility Spectrometer

(PAMS)

Handheld Anemometer

Capture Hood

Aerosol Monitor

2
Handheld Anemometers / Air Velocity

Applications

HVAC testing and balancing

Testing airflow rate of ventilation

Handheld
Anemometers

Specifications
Model
Sensor Type
Air Velocity Ranges
ft./min
m/s
Resolution
Accuracy
Air Flow

CFM (ft3/min)

Temperature Ranges
Accuracy
Relative Humidity Ranges
Accuracy
Differential Pressure Ranges
Accuracy
Dimensions
Main Unit
Probe
Weight
3
AnemomasterTM

AnemomasterTM

Professional  Standard

ClimomasterTM Series

Palm-size and feather-weight
standard hot-wire Anemometer

Telescopic, articulating probe is
designed for HVAC testing and
balancing applications

Multi-function hot-wire
Anemometer with detachable
compatible probes

Rotating Vane Anemometer with
High Accuracy from 40 to 7800 fpm

●●Compact and lightweight

●●Simultaneous display of
temperature and airflow or air
velocity

●●Simultaneously measures and
displays air velocity, flow rate,
humidity, temperature, and
differential pressure

Handheld Anemometers / Air Velocity

AnemomasterTM Lite

●●2 sizes of vane heads are available for
users’ applications

●●Easy replacement of probe without
recalibration
●●Data Hold function
●●Includes probe with 59 in (150
cm) cable, extension rod, 4 pcs.
AA batteries, carrying case, and
NIST Certificate

●●Telescopic, articulating probe
●●Data HOLD function, record and recall
MAX/MIN/AVG
●●Store up to 1,500 measurements

●●6 interchangeable probes are
available for various applications
●●Smart probe technology: easy probe
replacement without recalibration
of the main unit

●●Data processing software
allows real-time measuring and
downloading data to PC

●●Data HOLD function, record and
recall MAX/MIN/AVG

●●Complete with telescopic probe with
79 in (200 cm) cable, Data processing
software, USB cable, AC adapter, 6
pcs. AA batteries, carrying case, and
NIST Certificate

Model 6810 Series

●●Industrial grade enclosure and metal
vane probe
●●Data HOLD, record and recall MAX/
MIN
●●Complete with metal vane sensor
(choice of 2.75” or 1.00” diameter)
with 5 ft long cable, extension rod
with handle grip, flexible extension
rod, 3 pcs. AA batteries, carrying
case, and NIST Certificate

●●Data processing software
allows real-time measuring and
downloading of data to PC

●●Store up to 20,000 measurements

●●Includes probe, 79 in (200 cm)
probe cable, 6 pcs. AA batteries,
carrying case, and NIST Certificate

LITE

Professional  Standard

Climomaster Series

6810 Series

Hot-wire

Hot-wire

Hot-wire

Rotating Vane

2 to 3940

2 to 6000

2 to 9480

40 to 7800

0.01 to 20.0

0.01 to 30.0

0.01 to 50.0

0.2 to 40

0.01 m/s

0.01 m/s

0.01 m/s

0.01 m/s

+/- 5% of reading or 0.015 m/s
whichever is greater

+/- 3% of reading or 0.015 m/s
whichever is greater

+/- 2% of reading or 0.015 m/s
whichever is greater

+/- 1% of reading

n/a

0 to 2,709,360

0 to 2,709,360

0 to 9,999

-4 to 158°F (-20 to 70°C)

-4 to 158°F (-20 to 70°C)

-4 to 158°F (-20 to 70°C)

-4 to 212°F (-20 to 100°C)

+/- 1°F (0.5°C)

+/- 1°F (0.5°C)

+/- 1°F (0.5°C)

+/- 0.5°F (0.3°C)

n/a

n/a

0.2 to 98.0% RH

0.5 to 95.0% RH

n/a

n/a

+/- 2% of reading

+/- 2% of reading

n/a

+/- 5.00 kPa *Option

+/- 5.00 kPa *Option

n/a

n/a

+/- (3% of reading +0.01) kPa

+/- (3% of reading +0.01) kPa

n/a

W2.4” x H4.7” x D1.2”

W3.4” x H7.4” x D1.6”

W3.4” x H7.4” x D1.6”

W3.3” x H6.3” x D1.5”

0.24” (6.1 mm) in diameter

0.24” (6.1 mm) in diameter

1.0 to 10 mm in diameter

Vane: 2.75” or 1.00” in diameter

0.4 lbs (180 g)

0.9 lbs (400 g)

0.9 lbs (400 g)

0.95 lbs (430 g)
4
Anemometers / Air Velocity Meters

Climomaster TMModel 6501 Series
Multi-function hot-wire anemometer with detachable
compatible probes
Features:
●●Simultaneously measures and displays air velocity, flow rate, humidity,

temperature, and differential pressure

●●Airflow rate calculation based on registered duct size
●●Now equipped with an automatic atmospheric pressure compensation

function for precise airflow measurement

●●Store up to 20,000 measurements
●●8 interchangeable probes are available for various applications
●●Smart probe technology: easy probe replacement without recalibration

of the main unit

●●Includes probe,

79 in (200cm) probe connection cable, 6 pcs. AA
batteries, carrying case, and NIST Certificate

Main Unit Specifications
Air Velocity Ranges

2 to 9840 fpm (0.01 to 50.0 m/s) *Varies by probe
Accuracy

+/- 2% of reading or 0.015 m/s whichever is greater

Temperature Ranges

-4 to 158°F (-20 to 70°C)
Accuracy

Relative Humidity Ranges

+/- 1.0°F (0.5°C)
2.0 to 98.0%RH *Varies by probe

Accuracy

+/- 2.0%RH

Differential Pressure Ranges
(Option)

+/- 5.00 kPa

Accuracy
Interface

+/- (3% of reading + 0.01) kPa
USB / RS232C (for print-out)

Datalogging

Up to 20,000 records

Analog Output (Option)

0 to 1 V

Power Supply

6 x AA Batteries or AC Adapter

Dimensions

Optional data processing software
allows real-time measuring and
downloading data to your PC

W3.4” x H7.4” x D1.6” (88 x 188 x 41 mm)

Weight

0.9 lbs (400 g)

Probe Specifications
Model

6531-2G

6541-2G

6561-2G

6542-2G

6533-2G

6543-2G

6551-2G

6552-2G

Probe Type

Uni-Directional

Uni-Directional

Uni-Directional

Omni-Directional

Omni-Directional

Omni-Directional

Mini-Spherical

Mini-Spherical

Air Velocity

2 to 6000 fpm

2 to 6000 fpm

2 to 9840 fpm

2 to 6000 fpm

2 to 1000 fpm

2 to 1000 fpm

2 to 6000 fpm

2 to 6000 fpm

Temperature

-4 to 158°F

-4 to 158°F

-4 to 158°F

-4 to 158°F

-4 to 158°F

-4 to 158°F

n/a

n/a

2.0 to 98.0 %RH

n/a

n/a

n/a

2.0 to 98.0 %RH

n/a

n/a

n/a

Relative Humidity

Accessories
6501-AE
6501-BE
6531-04:
6531-05:
6531-06:
6000-41:
6000-31:
6000-61:
6113-02:
DPU-S245:
TP-202L:

5

Main Unit with Analog Output
Main Unit with Pressure Sensor
Telescopic Extension Rod (Flex-Neck)
Telescopic Extension Rod (Straight)
2m Probe Cable (also available in 5, 10, 20m)
Data Processing Software (for Windows)
Printer Cable for DPU-S245
Hands Free Case
A
 C Adapter
Portable Thermal Printer 
Rolled Printer Paper (10 rolls)

6531-04

6531-05

6000-61
Anemomaster TMProfessional  Standard
Multi-function hot-wire Anemometer with
telescopic, articulating probe
Features:
●●Simultaneous display of temperature and airflow or air velocity
●●Telescopic probe measures are velocity and temperature in air

ducts, vents, and small openings

●●Data HOLD function, Record and recall MAX/MIN/AVG
●●Store up to 1,500 measurements (Professional only)
●●Data

processing software allows real-time measuring and
downloading data to your PC (professional only)

●●Includes

telescopic probe with 79 in (200 cm) cable, Data
processing software (Professional only), USB cable (Professional
only), AC adapter, 6 pcs. AA batteries, carrying case and NIST
Certificate

Telescopic probe is designed for HVAC
testing and balancing applications

Applications:

Articulating Tip

●●HVAC Testing

(Professional only)

●●Facility Maintenace
●●Critical Environment Certification
●●IAQ Investigation

Specifications
Model

AnemomasterTM Professional Model 6036
Telescopic and Articulating tip

Probe Type

AnemomasterTM Standard Model 6035
Telescopic

Air Velocity Ranges
Accuracy

2 to 6000 fpm (0.01 to 30.0 m/s)
+/- 3% of reading or 0.015 m/s whichever is greater

Temperature Ranges

-4 to 158°F (-20 to 70°C)

Accuracy

+/- 1.0°F (0.5°)

Differential Pressure Ranges

+/- 5.00 kPa *Option

Accuracy
Interface - Digital
Analog (Option)
Datalogging

+/- 3% of reading +0.01 kPa
USB / (RS232C for print-out)

(RS232C for print-out)

DC 0 to 3 V (only for air velocity measurements)
Up to 1500 records

Power Supply

None

Windows 7 compatible software included
(professional only)

6 x AA batteris or AC Adapter

Dimensions

W3.4” x H7.4” x D1.6”

Weight

0.9 lbs (400 g)

Theory of thermal (hot-wire) type AnemomasterTM

Accessories
6036-AE:
6036-BE:
6035-AE:
6035-BE:
6000-31:
6
 000-61:
DPU-S245:
T
 P-202L:

Professional with Analog Output
Professional with Pressure Sensor
S
 tandard with Analog Output
Standard with Pressure Sensor
Printer Cable for DPU-S245
H
 ands Free Case
Portable Thermal Printer
Rolled Printer Paper (10 rolls)

The air velocity sensor is heated and temperature elevated (relative to the surrounding
air) by means of control electronics. The temperature compensation sensor senses the
ambient, or surrounding air temperature, and forces the velocity sensor to stay at a
constant overheat above the ambient. The circuit forces the voltage to be equal by means
of an operational amplifier. Air flowing past
Air velocity sensor
the sensor tends to cool the sensor, thus
driving down its resistance. The amplifier
responds by immediately delivering more
power to the circuit to maintain voltage
equilibrium. Delivered power is converted
into electrical signals to display.
Temperature compensation sensor

6
Anemometers / Air Velocity Meters

Anemomaster TMLITE

Palm-size and feather-weight standard hot-wire Anemometer
Features:	
●●Compact and lightweight
●●Display switchable in m/s or ft/min (FPM) for air velocity and °F and °C for air temperature
●●Easy replacement of probe without recalibration
●●Average measurements over 1 or 5 seconds for air velocity
●●Data Hold function
●●4 pcs. AA batteries, extension rod, carrying case and NIST Certificate

Specifications
Model

6006

Air Velocity Ranges
Accuracy

2 to 3940 fpm (0.01 to 20.0 m/s)
+/- 5% of reading or 0.015 m/s whichever is greater

Temperature Ranges

-4 to 158°F (-20 to 70°C)

Accuracy
Power Supply
Dimensions

+/- 1.0°F (0.5°C)

Accessories

4 x AA Batteries or AC Adapter

6006-2G: Replacement Probe
6112-03: Spare Extension Rod
6006-02: Spare Hard Carrying Case

W2.4” x H7.4” x D1.2”

Weight

0.4 lbs (180 g)

Anemomaster TMModel 6113/6114
High velocity (up to 50 m/s) multi-function hot-wire Anemometer
Features:	
●●Simultaneous measurements of air velocity, air temperature and differential pressure
●●Large, easy to read LCD display
●●Durable body
●●Model 6113 includes built-in printer
●●Complete

with probe with 79 in (200 cm) cable, extension rod, shoulder strap, and 6
pcs. C cell batteries, and NIST Certificate

Specifications
Model
Air Velocity Ranges
Accuracy
Temperature Ranges
Accuracy
Differential Pressure Ranges
Accuracy
Interface
Datalogging
Analog Output
Built-in Printer
Power Supply
Dimensions
Weight

6113
6114
20 to 9840 fpm (0.10 to 50.0 m/s)
+/- 3% of reading +0.1 m/s
32 to 212°F (0 to 100°C)
+/- 2.0°F (1.0°C)
+/- 5.00 kPa *Option
+/- 3% of reading +0.01 kPa
RS232C
100 measurements
0 to 1 V *Option
⃝

―
6 x C cell Batteries
W7.9” x H5.9” x D3.9”
2.2 lbs (1 kg)

Accessories
6113-01:
6113-02:
6113-03:
6113-04:
6113-07:
6113-08:

7

Spare Probe
AC Adapter
Extension Rod
Shoulder Strap
Pressure Sensor
Analog Output

6000-41: Data Processing Software
6000-02: Communication Cable to PC
TP-202L: Rolled Printer Paper (10 rolls)
Anemometers / Air Velocity Meters

Anemomaster TMVane Model 6810 Series
Rotating Vane Digital Anemometer
Features:	
●●High Accuracy from 40 to 7800 feet per minute
●●2 sizes of vane heads are available for users’ applications
●●Industrial grade enclosure and metal vane probe
●●Large display with backlight
●●Data HOLD, Record and recall MAX/MIN
●●Includes metal vane sensor (choice of 2.75” or 1.00”

diameter) with 5 ft long cable, extension rod with handle
grip, flexible extension rod, 3 pcs. AA batteries, carrying
case, and NIST-traceable calibration certificate

1 inch Air Velocity Probe

Specifications
Model

6812

Air Velocity Ranges

6815 comes with HTP202
6815

6813

2.75” Head

1.00” Head

2.75” Head

1.00” Head

2.75” Head

1.00” Head

ft/min

40 to 7800

300 to 6800

40 to 7800

300 to 6800

40 to 7800

300 to 6800

m/s

0.20 to 40.0

1.5 to 35.0

0.20 to 40.0

1.5 to 35.0

0.20 to 40.0

1.5 to 35.0

Resolution

1 FPM or 0.01 m/s

Air Flow CFM (ft3/min)

1 FPM or 0.01 m/s

1 FPM or 0.01 m/s

0 to 9999

n/a

n/a

n/a

Temperature Ranges

-4 to 212°F (-20 to 100°C)

-4 to 176°F (-20 to 80°C) *with HTP202

Accuracy

n/a

+/- (0.3°C +0.2% of reading in °C)

+/- (0.3°C +0.2% of reading in °C)

Relative Humidity Ranges

n/a

n/a

5.0 to 95.0% RH

Accuracy

n/a

n/a

+/- 2.0% RH

3 x AA Batteries

3 x AA Batteries

3 x AA Batteries

W3.2” x H6.5” x D1.5”

W3.2” x H6.5” x D1.5”

W3.2” x H6.5” x D1.5”

0.95 lbs (430 g)

0.95 lbs (430 g)

1.1 lbs (500 g)

Power Supply
Main Unit Dimensions
Weight

Humidity  Temp. Probe

Air Velocity Probes
Model
Air Velocity
ft/min
m/s
Accuracy
Temperature
°F
°C
Accuracy

AP275

n/a
n/a
n/a

APT275
2.75” Head
40 to 7800
0.20 to 40.0
+/- (1.0% reading + 1 digit)
-4 to 212
-20 to 100
+/- (0.3°C +0.2% of reading in °C)

AP100

APT100
1.00” Head
300 to 6890
1.5 to 35.0
+/- 0.50% FS + 0.75% reading + 1 digit

n/a
n/a
n/a

-4 to 212
-20 to 100
+/- (0.3°C +0.2% of reading in °C)

Model
Relative Humidity
Range
Resolution
Accuracy
Temperature
°F
°C
Accuracy

HTP202
5.0 to 95.0% RH
0.1% RH
+/- 2.0% RH
-4 to 176
-20 to 80
+/- (0.3°C +0.2% of reading in °C)

Vane Air Velocity Transmitters
Features:	
●●Transmitters specially designed for monitoring the Indoor Air Quality in the air conditioning,

heating, and ventilation industries.
●●Three versions available:
AT400: Air Velocity Transmitter
TAT420: Air Velocity  Temperature Transmitter
THT500: Temperature-Humidity Transmitter

●●Three separate power supplies 110 VAC 50/60 Hz, 220 VAC 50/60 Hz, and 10-30 VDC
●●Three concurrent outputs for each measurement: 0-1 VDC, 0-5 VDC, and 4-20 mA DC
●●Temperature Range -4 to 212°F (-20 to 100°C)

8
TABmaster TMCapture Hood
A solution for airflow testing and balancing

Airflow Capture Hood

The new Kanomax TABmasterTM is the perfect tool for accurate supply and return
airflow measurements. Interchangeable hoods make it easy to sample the air for
any duct size. The unit is lightweight and easy to handle. The full color screen can
be tilted so it’s always at the optimal viewing angle regardless of height.

Features:	
●●23 to 2500 cfm (40 to 4250 m3/h) range
●●Simultaneously measures and displays air flow, temperature and humidity
●●Displays the direction of the airflow
●●Store up to 8,000 measurements
●●Advanced storage feature allows you to store multiple measurements under a single ID#
●●Built-in back pressure compensation ensures accuracy for large volumetric flow

measurements

●●Removable handheld micromanometer with Bluetooth® wireless capability (Model 6715)
●●Includes: standard hood, carrying case, AA batteries, PC communication cable, data

processing software, user manual, and calibration certificate

Applications:
●●HVAC testing, adjusting and balancing
●●Air volumetric flow measurements through registers, diffusers and grilles
●●Direct readout at supply and return airflow
●●Air velocity measurement in the duct
●●Check filter fouling by measuring differential pressure

Specifications
6710

Model
Airflow Range

6715
23 to 2500 CFM (40 to 4250 m3/h)

Accuracy

+/- 3% of reading +/- 10 m3/h

Resolution
Air Velocity Range
Accuracy
Resolution

1m3/h
n/a

0.15 to 40 m/s (Pitot), 0.15 to 15 m/s (Velocity Matrix)

n/a

±3%±0.05m/s of readings

n/a

0.1 m/s (10.0m/s)
0.01m/s (9.99m/s)

Temperature Range

Portable stand extends up to 6.5’ from top to base

32 to 122°F (0 to 50°C)

Accuracy

+/- 1.0°F (0.5°C)

Resolution

0.1°C

Humidity Range

0 to 100% RH

Accuracy

+/- 3.0% RH

Resolution

0.1% RH

Interface

USB

Datalogging

Up to 3000 measurements

Power Supply
Hood Dimensions

Up to 8000 measurements
AA batteries or AC adapter

2x2 ft (610x610 mm), 1x4 ft (305x1220 mm), 2x4 ft (610 x 1220 mm)
3x2 ft (915 x 610 mm), 3x3 ft (915x915 mm), 500x500mm

Weight

7.9 lbs (3.6 kg)

Accessories
6
 710-01:
6710-02:
6710-03:
6710-04:

9

Spare Hood 2x2 ft (610x610mm)
Spare Hood 2x4 ft (610x1220mm)
Spare Hood 1x4 ft (305x1220mm)
Spare Hood 3x2 ft (915x610mm)

6
 710-05:
6710-06:
6710-07:
6710-08:

Spare Hood 3x3 ft (915x915mm)
Spare Hood 500x500mm
Spare Hood Support Poles
Capture Hood Stand

The Bluetooth® feature can send data to any
Android-based device
Micromanometer DP500
Smart micrometer for TAB professionals

Pressure Measurement

Handheld micromanometer with Bluetooth® wireless capability
takes airflow and pressure readings with a pitot tube or the
optional velocity matrix. The matrix is a cross shaped grid, similar
to the one in the base of the capture hood, that makes it possible
to take face velocity readings and average velocity readings at
multiple points simultaneously. The Bluetooth® feature can send
data to your smartphone or any Android-based device.
Pitot tubes and Velocity Matrix are
available from Kanomax

Specifications
Model

DP500
0.15 to 40 m/s (with Pitot tube)
0.15 to 15 m/s (with Velocity Matrix)

Air Velocity Ranges
Accuracy

±3%±0.05m/s of readings

Accuracy

±0.25% ±1Pa of readings

Pressure Ranges

-2500 to 2500 Pa

Temperature Ranges

0 to 50°C

Accuracy

+/- 1% of reading +1°C

Relative Humidity Ranges

0 to 100%RH

Accuracy

±3%RH(10~90%RH)

Interface

USB and RS232C (for printer)

Datalogging

8000 measurements

Power Supply

4 x AA Batteries or AC Adapter

Pitot Static Tube
Stainless Steel Construction Pitot Tubes
The unique ellipsoidal nose form results in exceptional accuracy,
insensitivity to errors in alignment and elimination of the need for
calibration curves. Constructed from stainless steel, tubes can be
as small as 2.3 mm in diameter yet scaled up to 4 m in length.

Features:	
●●Totally

compatible with manometers, pressure gauges and
pressure transmitters

●●Ideal for measurement of air velocity in hostile environments
●●Integral

sensing

Thermo pitot tube

thermocouple for combined pressure and temperature

●●Can be used in permanent position when fitted with gland
●●S-Type for limited access and adverse conditions

S-Type pitot tube

Specifications
Easy Fit Ellipsoidal Pitot Tube
Overall Length (m)

Classic Ellipsoidal Pitot Tube

0.2

0.3

0.3

0.5

0.5

Tube Diameter (mm)

4

4

4

8

8

8

Head Diameter (mm)

2.3

4

4

8

8

8

Gland (inch)

1/4”

3/8”

0.75 0.75

1.0

1.0

8

8

8

8

3/8”

Pitot with K-Type Thermocouple

Overall Length (m)

Overall Length (m)

1.83

2.13

2.44

2.74

1.0

1.22

1.52

1.83

8

9.5

9.5

Tube Diameter (mm)

12.7

19

19

19

8

9.5

9.5

12.7

8

9.5

9.5

Head Diameter (mm)

9.5

9.5

9.5

9.5

8

9.5

9.5

9.5

1/2”

Gland (inch)

3/8”

1/2”

1/2”

1/2”

3/8”

S-Type Pitot Tube

Overall Length (m)

1.25 1.25

0.3

1.0

Tube Diameter (mm)

4

8

Head Diameter (mm)

4

0.7

1.2

2.2

3.2

8

10
Flow Grids
For Accurate In-Duct Volume Flow Measurements

Airflow Measurement

Kanomax flow grids produce averaged total and sub-static pressures signals to generate
an enhanced velocity pressure signal, outside the duct wall, that directly relates to the
volume flow, providing accurate and reliable outputs where other flow measuring
devices are found to be unsatisfactory. Works directly with Micro-manometers, pressure
transducers, Magnahelic gauges, and pressure activated alarm switches. These are ideal
for ventilation monitoring for LEEDS and many other facility management monitoring
applications.

Features:	
●●Sizes from 4.0” (100mm) diameter to 10 ft.(3.0 meters) square
●●Uncertainty of flow measurement is within +/- 1.0%
●●High temperature systems up to 850°C are available
●●Stainless steel grade 321 tube supplied as standard, other grades of steel or alloys

are available upon request

Standard Sizes
Rectangular Grids

Radial Grids

Type A
6.4mm (1/4”)
d
 iameter tube

Manifold length (mm)

100 to 450

Pressure Tube length (mm)

150 to 450

Type B
12.7mm (1/2”)
diameter tube

Manifold length (mm)

250 to 650

Pressure Tube length (mm)

350 to 1200

Type C
25.4mm (1”)
d
 iameter tube

Manifold length (mm)

750 to 1000

Pressure Tube length (mm)

800 to 2000

Type A
6.4mm (1/4”)
d
 iameter tube

Diameter (mm)

100 to 500

Type B
12.7mm (1/2”)
diameter tube

Diameter (mm)

500 to 1100

Type C
25.4mm (1”)
d
 iameter tube

Diameter (mm)

1100 to 2500

X Flow Grids
For Accurate In-Duct Volume Flow Measurements
X flowgrids will give useful and reliable readings in a wide variety of in-duct locations
often where other flow rate measuring devices are found to be unsatisfactory or difficult
to install. The low cost design consists of two tubes mounted diagonally through the
cross section of the duct. They generate an averaged signal for both the total and substatic (X) or static (UX) pressure along the length of each tube to generate a differential
pressure signal that directly relates to volume flow.

Features:	
●●Economical method of measuring volume flow within a duct
●●Uncertainty of volume or mass flow measurement within +/- 5.0%
●●Stainless steel grade 321 tube supplied as standard

Specifications
Model
Air Velocity Ranges

UX8

Maximum Duct Diagonal
Maximum Dust Diameter

UX version: 1.0 correlation ;
X version: 2.0 amplified correlation

Diameter of Tubes

Weight

UX16

+/- 5%

Maximum Temperature

●●Differential pressure correlation/correction factor:

MX16

295.2 to 5904 ft/min (1.5 to 30 m/s)

Accuracy

●●High temperature systems up to 850°C (continuous) available

11

MX8

176°F (80°C)
5/16” (8mm)

5/8” (16mm)

28.1“ (715mm)
26.7” (680mm)

60.4 “(1534mm)
59.4” (1510mm)

1 lbs (454 g)

3 lbs (1361 g)
High Temperature Velocity Meter

Anemomaster TMModel 6162
High Temperature Anemometer
Features:	
●●Air

velocity and temperature measurements in 932°F (500°C)
environments

●●Record and recall MAX/MIN/AVG, Timing graph display
●●Store up to 999 measurements
●●RS232C

interface, analog output, and remote control terminal
equipped

●●Includes

shoulder strap, AC adapter, 2 pcs. analog output cable,
and 6 pcs. C cell batteries

High temperature measuring solution requires Model 6162 with
one of the high temperature probes (0203, 0204 or 0205)
●●The

Model 0203 includes probe with 4.9 ft (1.5 m) cable, 5m
probe connection cable, probe case, and NIST Certificate

●●The

Model 0204 includes probe with 7.6 ft (2.3 m) cable, 10m
probe connection cable, probe case, and NIST Certificate

●●The

Model 0205 includes probe with 7.6 ft (2.3 m) cable, 10m
probe connection cable, probe case, and NIST Certificate

Main Unit Specifications
Model

6162

Air Velocity Ranges
Accuracy
Temperature Ranges
Accuracy

Varies by Probe, see below for Probe Specifications
+/- 3% of Full Scale
Varies by Probe, see below for Probe Specifications
+/- 1% of reading +1°C

Interface

RS232C

Datalogging

999 measurements

Analog Output
Remote Terminal

Probe Specifications
Model

0203

0204/0205

Temp. Range
Air Velocity
Measuring Range

32 to 212°F

40 to 9840 fpm (0.2 to 50.0 m/s)

212 to 392°F
n/a

138 to 9840 fpm (0.7 to 50.0 m/s)

572 to 752°F

n/a

197 to 9840 fpm (1.0 to 50.0 m/s)

Accuracy

+/- 3% Full Scale
32 to 392°F (0 to 200°C)

Accuracy

Dimensions / Weight

6 x C cell Batteries or AC Adapter

Dimensions

W8.7” x H3.3” x D5.9”

Weight

4.0 lbs (1.8 kg)

80 to 9840 fpm (0.4 to 50.0 m/s)

392 to 572°F

Temperate Measuring Range

Power Supply

0 to 1 V
START/STOP Key

32 to 932°F (0 to 500°C)
+/- 1% of reading

ø 11 x 200 mm (ø 0.4” x 7.8”)
0.4 lbs (200 g)

Probe Cables

0204

ø 14 x 1000 mm (ø 0.6” x 39.4”)
1.1 lbs (500 g)

0205

ø 14 x 500 mm (ø 0.6” x 19.7”)
0.4 lbs (200 g)

Teflon Coating
Heat-resistance

392°F (200°C)

Extension Cable

Vinyl Coating

Heat-resistance

176°F (80°C)

Accessories
0203:
0204:
0205:
6162-03:
6162-04:
6162-05:
6162-06:
6162-07:
6000-41:
6511-09:
DPU-S245:
TP-202L:

Probe for Middle Temperature
Probe for High Temperature (Long)
Probe for High Temperature (Short)
Extension Rod for 0203
Extension Rod for 0204/0205
Probe Compression fitting for 0203
Probe Compression fitting for 0204/0205
Communication Cable to PC
Data processing Software
Printer Cable for DPU-S245
Portable Thermal Printer
Rolled Printer paper (10 rolls)
*Optional probe cable lengths up to 40m are available

Probe Dimensions

*Length in mm

MODEL 0203

MODEL 0204

Extension Rod for 0203 * Option

Extension Rod for 0204 * Option

Entire Length

Entire Length

Probe Cable

MODEL 0205

12
Multi-Channel Anemometers

Multi-Channel
Measuring
Instruments

Main Unit Specifications
Model
Probe Compatible
Air Velocity Ranges
Temperature Ranges
Relative Humidity Ranges
PC Communication

Variety of probes for many applications

Interface
Analog Output
Power Supply
Dimensions
Weight

13
Multi-Channel Anemometers

Airflow Transducer
Model 6332 / 6332D

4-Channel AnemomasterTM
Model 1570

Multi-Channel AnemomasterTM
Model 1550 / 1560

Compact design Airflow Transducer, fits
into small spaces

Compact design 4-channel unit with compatible
probes

Up to 320 points multi-measurements system with
a variety of probes

●●10 interchangeable probes are available
for various applications

●●10 interchangeable probes are available for various
multi-channel applications

●●System can be scaled up with modules and probes

●●Easy probe replacement without
recalibration of the main unit

●●Simultaneously measurements of 4 channels of air
velocity

●●Selectable output options: 0 to 5 V or 4
to 20 mA

●●Software allows real-time measurements of air
velocity and airflow in 4 channels

●●The chassis may be cascaded up to 5 units via RS232C

●●The Model 1570 includes data processing software,
RS232C cable, power cable, and 2 pcs. fuse

●●Multi-channel Anemomaster includes RS232C cable,
printer cable, power cable, and 2 pcs. fuse

●●3 types of probes are available
●●4 types of modules are available

6332/6332D

1570

1550 / 1560

V Probe

V Probe

V Probe / VT Probe / VTH Probe

20 to 9840 fpm (0.10 to 50.0 m/s)

20 to 9840 fpm (0.10 to 50.0 m/s)

20 to 9840 fpm (0.10 to 50.0 m/s)

n/a

n/a

32 to 212°F (0 to 100°C)

n/a

n/a

5.0 to 95.0% RH

n/a

Data Processing Software

Data Processing Software

n/a

RS232C

RS232C, Centronics, GP-IB

DC 4 to 20mA or DC 0 to 5V

0 to 5 V

0 to 5 V

DC or AC

AC

AC

W3.1” x H5.0” x D1.2”

W10.2” x H2.8” x D7.9”

Model 1550: W19.6” x H5.5” x D16.9”
Model 1560: W8.9” x H5.5” x D12.8”

0.7 lbs (320 g)

5.7 lbs (2.6 kg)

Model 1550: 22 lbs (10 kg)
Model 1560: 11 lbs (5 kg)
14
Airflow Transducer Model 6332 / 6332D
Features:	

Multi-Channel Anemometers

●●Smart

probe technology: easy probe replacement without
recalibration of the main unit

●●Selectable output options: 0 to 5 V or 4 to 20 mA

Main Unit Specifications
Model
Display
Air Velocity Ranges
Accuracy
Analog Output

6332
6332D
―
⃝
Varies by probe, See below for Probe Specifications
+/- 3% of reading
DC 4 to 20mA or DC 0 to 5V

Power Consumption

Approx. 2.0 W

Power Supply
Dimensions
Weight

DC 12 to 24V or AC 80 to 240V
W3.1” x H5.0” x D1.2”
0.7 lbs (320 g)

6332D

6332

4-Channel Anemomaster TMModel 1570
Features:	
●●Simultaneous measurements of 4 channels of air velocity; easy to switch over each

channel display

●●Software allows real-time measurements of air velocity and airflow in 4 channels
●●The

Model 1570 includes data processing software, RS232C cable, power cable,
and 2 pc. fuse

Main Unit Specifications
Model
Air Velocity Ranges
Resolution
Interface
Analog Output
Power Supply
Dimensions
Weight

1570
Varies by probe, See below for Probe Specifications
0.01 m/s
RS232C
0 to 5V
AC 85 to 265V
W10.2” x H2.8” x D7.9”
5.7 lbs (2.6 kg)

Compatible Probes
Air Velocity Proves (V Probes)

0965-00 is with Horn

Probe Specifications
Model
Probe Type
Air Velocity

0962-00
Uni-Directional
20 to 9840 fpm (0.10 to 50.0 m/s)

0963-00
Uni-Directional
20 to 9840 fpm (0.10 to 50.0 m/s)

0965-00/-01
Omni-Directional
20 to 4920 fpm (0.10 to 25.0 m/s)

Miniature Probe
with built-in temperature compensation

0965-09 is 80mm long
0965-10 is 400mm long

0965-09/10
Omni-Directional
20 to 9840 fpm (0.10 to 50.0 m/s)

Miniature Probe
with independent temperature compensation

Air Velocity Sensor
Air Velocity Sensor

Probe Specifications
Model
Probe Type
Air Velocity

15

Temperature Compensation Sensor
0965-03
Omni-Directional
20 to 4920 fpm (0.10 to 25.0 m/s)

Temperature Compensation Sensor
0965-04
Omni-Directional
20 to 4920 fpm (0.10 to 25.0 m/s)

0965-07
Omni-Directional
20 to 4920 fpm (0.10 to 25.0 m/s)

0965-08
Omni-Directional
20 to 4920 fpm (0.10 to 25.0 m/s)
Multi-Channel Anemomaster TMModel 1550/1560
●●The

multi-channel Anemomaster is composed of chassis, module,
and probe. The Model 1550 (chassis) holds 16 modules and the
Model 1560 holds 6 modules

●●The

chassis may be cascaded up to 5 units (5 units of Model 1550
can hold 320 channels of air velocity sensors)
1550

●●3 types of probes are available

Air velocity probe (V probe)
Air velocity / Temperature probe (VT probe)
And Air Velocity / Temperature / Humidity probe (VTH probe)

●●4 types of modules are available

1560

4 channel air velocity module
2 channel air velocity / temperature module
1 channel air velocity / temperature / humidity module
Analog output module
Main Unit Specifications

●●Multi-channel Anemomaster includes RS232C cable, printer cable,

Model

power cable, and 2 pcs. fuse

1550

Air Velocity Ranges

1560

Varies by Probe, See below for Probe Specifications

Resolution

0.01 m/s

Temperature Ranges

Varies by Probe, See below for Probe Specifications

Resolution

Accessories
1500-01:
1500-02:
1500-03:
1504-04:
1511-01:
1512-01:
S620-00:

Multi-Channel Anemometers

Features:	

0.1°C

Relative Humidity Ranges

RS232C for Cascade Connection
GP-IB Output
Ring Guards for Probe Protection
Cable for V module (10m)*
Cable for VT Module (10m)*
Cable for VTH Module (10m)*
Data Processing Software (for Windows)

Varies by Probe, See below for Probe Specifications

Resolution

0.1% RH
RS232C for PC Connection
RS232C for Cascade *Option

Interface

Centronics for Printer Output
GP-IB *Option

Analog Output

0 to 5 V *Option with D/A Module

Power Supply

*Additional cable lengths available

AC Adapter

Dimensions

W19.6” x H5.5” x D16.9”

W8.9” x H5.5” x D12.8”

22 lbs (10 kg)

11 lbs (5 kg)

Weight

Air Velocity Temperature Humidity Proves (VT / VTH Probes)

Probe Specifications
Model
Probe Type
Air Velocity
Temperature Ranges
Relative Humidity Ranges

0962-21

0963-21

0965-21

0963-31

0965-31

Uni-Directional

Uni-Directional

Omni-Directional

Uni-Directional

Omni-Directional

20 to 9840 fpm (0.10 to 50.0 m/s)

20 to 9840 fpm (0.10 to 50.0 m/s)

20 to 9840 fpm (0.10 to 50.0 m/s)

20 to 9840 fpm (0.10 to 50.0 m/s)

20 to 9840 fpm (0.10 to 50.0 m/s)

32 to 212°F (0 to 100°C)
n/a

32 to 212°F (0 to 100°C)
n/a

32 to 212°F (0 to 100°C)
n/a

32 to 212°F (0 to 100°C)
5.0 to 95.0% RH

32 to 212°F (0 to 100°C)
5.0 to 95.0% RH

Modules

Specifications
Model
Module Type
# of Channels

1504 V Module

1511 VT Module

1512 VTH Module

1510 D/A Module

Air Velocity

Air Velocity, Temp.

Air Velocity, Temp., Humidity

Analog Output

4

2

1

Model 1560 has 6 slots for modules

1

16
Applications

Indoor Ait Quality Monitors

Hospitals and Elderly Care Facilities Monitor

Green building rating system
IEQ performance testing

Indoor Air Quality
Monitors

Ventilation Testing
Thermal Comfort Control
Occupational Health Control
Monitoring Toxic Gas

17

Indoor Air Quality Investigation
Handheld IAQ Monitor Model 2211
Multi-Function Indoor Air Quality Monitor

Indoor Ait Quality Monitors

Features:	
●●Simultaneous measurements of CO, CO2, Temperature, and Relative humidity
●●Calculates Dew point, Wet bulb temperature, Absolute humidity, Humidity Ratio, and

% Outside Air

●●Store up to 1500 measurements
●●Easy user self calibration
●●Easy replacement of probe
●●PC

interface with RS232C or USB and software for real-time measurements and
downloading data to your PC

●●Complete with probe with 79 in (2m) cable, probe stand, calibration cap  connection

tube, data processing software, RS232C cable, USB to serial adapter, 6 pcs. AA
batteries, carrying case, and NIST-Traceable calibration certificate

Specifications
Model

2211

Carbon Monoxide (CO)
Accuracy
Carbon Dioxide (CO2)
Accuracy
Temperature Range
Accuracy
Relative Humidity Range
Accuracy
Interface
Datalogging

0 to 500 ppm
+/- 3% of reading
0 to 5000 ppm
+/- 3% of reading
-4 to 140 F (-20 to 60 C)
+/- 1.0°F (0.5°C)
2.0 to 98.0 %RH
+/- 2% of reading
1500 measurements

Analog Output

0 to 1 V *Option

Power Supply

6 x AA Batteries or AC Adapter

Dimensions
Weight

Software Included

RS232C

W3.4” x H7.4” x D1.6”
0.9 lbs (400 g)

Accessories
6113-02:
2211-09:
DPU-S245:
6
 000-03:
T
 P-202L:

AC Adapter
Analog Output
Portable Thermal Printer
P
 rinter Cable for DPU-S245
Rolled Printer Paper (10 rolls)

Indoor Comfort
Comfort is very subjective and will vary from one individual to the next, but
ASHRAE Standard 55 provides guidelines for a “comfort zone” that most people
will feel comfortable in. This comfort zone is derived from both temperature
and humidity and varies depending on the season. See the chart to the right
for details. The 2211 IAQ Monitor from Kanomax can measure both parameters
simultaneously indicating if changes are needed to make a building’s occupants
comfortable.
The monitor can also check CO2 levels and determine the rate of exchange, or the
percentage of outside air, being introduced to a building per ASHRAE Standard
62. Controlling the rate of exchange with the HVAC system is critical to ensuring
occupant well-being. Too much CO2 build-up can cause lethargy and make it
difficult for individuals to concentrate. Too much outside air may also be harmful
if it’s introducing external pollutants to the environment.
The 2211 is a great tool for IAQ investigation and spot checks to ensure an indoor
environment remains within desired levels to maintain occupant health and
comfort. The ability to measure all of these parameters in a single instrument
simplifies the labor involved and eliminates the need to purchase and learn
multiple instruments.

18
Handheld Gas Monitors

CO, CO2, Ammonia, Ozone, VOC etc. More than 30 gases
Features:	
●●Simultaneous measurements of gas concentration, temperature, and

humidity

●●The S200 is simple, easy-to-use, low-cost monitor
●●The S300 has on-board alarm and analog output

Gas Monitors

●●The S500 has data logging function and USB interface for PC

communication

●●The Remote Sensor Head Adaptor Kit is available; it is designed to allow

the sensor head to be located up to 46 ft (14 m) away from the monitor

Specifications
Model
Measurement Units
T/H* Sensor
On-board alarm
Remote Sensor
Datalogging
Interface
Analog Output
Power supply
Dimensions
Weight

S200

S300

S500

ppm or mg/m3

ppm or mg/m3

ppm or mg/m3

Option *

Option *

Option *

―
⃝
―
―
―

⃝
⃝
―
―

⃝
⃝
4,300 measurements
RS232C

0 to 5 V

0 to 5 V

9.6V Battery / 12 VDC

9.6V Battery / 12 VDC

9.6V Battery / 12 VDC

W7.4” x H4.8” x D2.1”

W7.4” x H4.8” x D2.1”

W7.4” x H4.8” x D2.1”

1.0 lbs (460 g)

1.0 lbs (460 g)

1.0 lbs (460 g)

* Temperature and Humidity

Features:	

Gas Sensor Heads

●●The gas monitor is fully compatible with all gas sensors

Gas Sensor Specifications
Gas Sensor Heads

Range (ppm)

Accuracy

Resolution

Ammonia

0 - 100

+/- 5 ppm

0.1 ppm

Ammonia (leak)

0 - 1000

+/- 15%

1 ppm

Carbon monoxide

0 - 100

+/- 5 ppm

0.1 ppm

Carbon monoxide

0 - 1000

+/- 10%

1 ppm

Carbon dioxide

0 - 2000

+/- 40 ppm + 3%

10 ppm

Carbon dioxide

0 - 5000

+/- 150 ppm + 5%

10 ppm

Hydrogen

0 - 5000

+/- 10 %

1 ppm

0 - 10

+/- 0.5 ppm

0.01 ppm

Hydrogen sulphide
Methane

0 - 9999

+/- 15 %

0 - 0.150

+/- 0.005 ppm

0 - 0.5

+/- 0.008 ppm

0 - 1.0

+/- 0.001 ppm

0.001 ppm

NMHC 3,4

0 - 25

+/- 10 %

0.1 ppm

Perchloroethylene

0 - 200

+/- 5 ppm

1 ppm

0 -10

+/- 0.5 ppm

0.01 ppm

Formaldehyde

0 -10

+/- 0.05 ppm

0.01 ppm

VOC

0 - 25

+/- 10 %

0.1 ppm

VOC

0 - 500

+/- 10 %

1 ppm

VOC PID

0 - 20

+/- 10 %

0.01 ppm

VOC PID

0 - 1000

+/- 10 %

0.1 ppm

Sulphur dioxide

Multi-Gas Sensor Heads

Sensor

Range (ppm)

Accuracy

Resolution

MS1

CO2
CO

0 - 2000
0 - 100

+/- 40 ppm +3%
+/- 10%

1 ppm
0.1 ppm

MS2

CO2
CO
VOC (PID)

0 - 2000
0 - 100
0 - 25

+/- 40 ppm +3%
+/- 10%
+/- 10%

1 ppm
0.1 ppm
0.01 ppm

* Other specific concentrations available on request

19

without recalibration of main unit

0.001 ppm

Nitrogen dioxide

●●Sensor heads for handheld units are interchangeable

0.001 ppm

Ozone

application

1 ppm

Ozone

●●Multi-gas sensor heads are available for IAQ survey

Accessories
R01:
R10:
R13:
R
 31:
R32:
R
 33:
R
 40:
R
 52:

Temperature/Relative Humidity option
Remote Sensor Adapter with Cable
Remote Sensor Adapter with Cable, Enclosure
Replacement Battery
C
 igarette Lighter Adapter Battery Charger
Wall Mount Bracket
Carrying Case
R
 S232 to USB Converter
Networked Gas Monitoring Systems
Fixed Transmitters  Controllers

Features:	
●●The

gas sensors are able to measure concentrations of Ozone,
Volatile Organic Compounds, Ammonia, Carbon Monoxide
sensor head has its own identification (ID), which allows a
“Daisy-Chained” network to be created utilizing either a computer
system or Programmable Logic Controller (PLC)

●●Networks

created

Gas Monitors

●●Each

of up to 256 individual monitors/transmitters can be

PLC Specifications
Specifications

Model
Measuring Ranges
T/H* Sensor
Alarm Output
Analog Output
Interface
Power Supply
Enclosure Size
Enclosure Casing
Weight

S905

S935

S945

See Gas Sensor Specifications

See Gas Sensor Specifications

See Gas Sensor Specifications

Option *

Option *

Option *

⃝

⃝

⃝

4 - 20 mA

4 - 20 mA and 12 to 24 V

4 - 20 mA and 12 to 24 V

RS485

RS485

RS485

24 VDC

24 VDC

24 VDC

2.5” x Ø5.1”

W7.1” X H4.3” X D3.5”

W9.1” X H5.5” X D3.7”

Fibre reiforced plastic

Fibre reiforced polycarbonate

Fibre reiforced polycarbonate

850 g

850 g

1.1 kg

Model
Sensor Channels
Digital Alarm Inputs
Alarm  Control Relays
Analog Inputs
Analog  Control Outputs
Expansion I/O
Data-storage
Communication
Power Supply
Dimensions

R1000 Series
Up to 12
Up to 10
Up to 22
Up to 13
Up to 6
Up to a maximum of 128 I/O
13,000 data points
Serial: RS232/ RS485 (Selectable)
ModBus: Supports MODBUS protocol, Master/Slave
DC 24V
W3.78” x H3.78” x D2.52” (96 x 96 x 64 mm)

Monitors Networked to a computer system

Features:	
●●Centralized control from the PC or decentralized control from individual monitors
●●Specialized application specific software is available on request

Monitors Networked to the R1000 PLC

Features:	
●●Centralized control from the PLC or decentralized control from individual monitors
●●Specialized application specific software is available on request

20
Handheld Odor Monitor
Highly Sensitive, 3 Models for Various Applications
Handheld Odor Meter is the most popular simplified tool for odor  nalysis,
a
which indicates the relative strength and odor  lassification numerically by
c
comparing the odor gas and purified air.

Features:	

Odor Monitor

●●Numerical value for relative strength of smell
●●Numerical value for classification of smell
●●Handheld Odor meter is ideal for Before and After applications, such as air purifica-

tion and cleaning service.

●●Real-time sampling mode displays odor change continuously
●●Memory sampling mode saves data based on the selected sampling rate. (up to

32732 data, 511 files)

●●Battery operated with 7 hrs continuous usage

Applications:	
OMX-SRM
●●Finding odor source at factories, incinerator,  lants, or effluent treatment facility
p

OMX-ADM
●●Evaluation for deodorizing at hospitals or nursing homes
●●Suitable for putrid odor such as ammonia

OMX-TDM
●●TVOC measurement for monitoring IAQ condition

Specifications
Model

OMX-SRM

OMX-ADM

Detection Method
Sampling Method

Continuous Sampling with Built-in Air Pump

Object Gas
Odor Strength Level
Classification
Power Supply
Memory Capacity
Storage Temperature
Operating Temperature
Dimensions
Weight

21

OMX-TDM

Two Semiconductor Gas Sensors

Ethanol, Acetone, Hydrogen, etc

Hydrogen, Sulfide,
Methyl Mercaptan,
Ammonia, etc

TVOC (Toluene), etc

0.0 to 999

0 to 999

0.0 to 9999 (µg/m3)

Class 0 to 90

Odor Intensity 2.5 to 5.0

n/a

4 x AA batteries or AC adaptor *Battery life for continuous measurement is approx. 7 hours
Up to 32732 data
0 to 40°C (32 to 104 ºF) * No condensation
-10 to 50°C (14 to 122 ºF) * No condensation
74 (W) x 167.5 (L) x 35 (H) mm
250g (without batteries)
Dif-Kit Tracer Gas Hardware
Professional Fume Hood Diagnostic Tools

Fume Hood Diagnostic Tools

Features:	
●●The

ANSI/ASHRAE 110 test is a method of testing the performance of laboratory fume hoods.
Kanomax Dif-Kit tracer gas hardware is ideal for use in performing the Tracer Gas test in
accordance with ANSI/ASHRAE Standard 110-1995.

●●Diffuser is made to the design and specifications of Standard drawing #110-83M
●●The internal critical orifice ensures a flow rate of 4 liters per minute. Other orifice sizes can be

inserted

●●Test-Mannequin and the vapor analyzer for the tracer gas test are available
●●Face velocity measurement data can be stored and downloaded with Anemometer
●●Dif-Kit

includes tracer gas diffuser, nozzle assembly, in-line flow meter, tank regulator, and 20
ft of tubing

●●Test-mannequin includes lab coat and stand

ASHRAE 110 - 1995
Performance Testing

Dif-Kit
Flow Meter

Regulator

AS 110-1995 Ejector
Specifications
Model

Dif-Kit

Critical Orifice
Regulator
Flow Meter

4 liter per minute
Dual Stage, Specialty Gas
Calibrated 150 mm/200 psi, Glass Tube

Pressure Gauge

0 to 60 psi

Vapor Analyzer
The Vapor Analyzer is factory calibrated for Sulfur Hexafluoride (SF6).
Response time is less than 10 seconds to 90% response with a stable
maximum concentration in less than 30 seconds.
Specifications
Type
Sample Flow
Sample Cell Volume
Gas Purge
Concentration Alarm
Interface
Power Supply
Dimensions

Vapor Analyzer

Weight

Single Beam Infrared Spectrophotometer
10 liter per min (21.2 ft3/hr)
0.45 liters (0.0159 ft3)
5 Cell Volumes for 99% purge of cell
High  Low (selectable by user)
USB, RS232
NiMH battery or 120-220V AC
15 x 7.3 x 7.5 inches (381 x 185 x 191 mm)
18 lbs (8.2 kg)

22
Applications

Sound  Vibration Meters
23

Environmental noise measurement

Noise exposure measurement

Industrial vibration testing

Sound and
Vibration Meters
Sound Level Meter Model 4431
Precision Sound Level Meter with 0-dB Function

Sound  Vibration Meters

The new Kanomax sound level meter is designed to be compact,
lightweight and easy to use. It complies with the type 2 ANSI
S1.4 1983 standard , and has an innovative, unique 0-dB feature
that eliminates the self-noise of the microphone. This technology
extends the lower limit of the measurement range to lower than 0
dB-SPL.

Features:	
●●Lightweight and compact design
●●Equipped with highly sensitive electret condenser microphone
●●Large 4 digit display with 0.1 dB resolution with backlighting

and analog bar graph

●●Add additional functions with the program cards
●●Includes SD card for data storage, windshield, carrying case, AA

Optional program cards give you the flexibility
to add additional functions as needed

batteries, hand strap, and calibration certificate

Specifications
Model

4431

Parameters

Lp, LA, LAeq, LAE, LAmax, LAmin, LAN, Lpeak, LAtm5

Ranges

A: 28 - 130 dB, Z 39 - 130 dB

Weighting

A, C and Z

Time Response

Fast or Slow

Microphone

TYPE 7146nl (-28dB, Stand-alone -26dB)
Type 1 (4432) or Type 2 (4431)
standards for ANSI S1.4 1983, IEC 61672-1

Standards
Power Supply

4 AA Batteries or optional AC Adapter (Battery life: approx. 9 hours)

Size  Weight

W3.4” x H12.9” x D1.9”, Approx 1 lb.

Accessories
AC-1026:
A
 CBC-0046-3:
ACBC-0046-5:
A
 CBC-0071:
ACNA-0038W:
ACNA-0038:
A
 CNA-0038F:
A
 CNA-0038R:
ACNA-0333:

AC adapter
Microphone Cable (3m)*
Extension cable (5m)*
BNC-Pin Cable
Data processing software
Program card (1/1, 1/3 octave analyzer)
Program card (FFT analyzer)
Program card (Real Sound Recording)
Tripod
*Additional cable lengths available

Sound Calibrator
Sound Calibrator Model AC2127 implements stable and highprecision calibration by its feedback control function with
temperature compensation.
Calibrator Specifications
Model
Microphone sizes

AC2127
1 inch and 1/2 inch (1/2 inch microphone adapter is included)

Sound Pressure Level
Frequency Level
Standards

94dB (+/-0.3dB)
1000Hz (+/-0.1%)
IEC 60942: 2003 Class 1
JIS C1515: 2004 Class 1

Power Supply

2 x AA Batteries

Size  Weight

W4.3” x H2.9” x D1.8”, Approx 0.3 lb. (150g)

Pistonphone
Model AC2124A is a standard sound source, which emits
124dB-SPL pure sine tone at 250Hz for calibration of sound
level meter.This is used for precise calibration of sound level
meter provided with 1-inch, 1/2-inch and 1/4-inch condenser
microphones.

Pistonphone Specifications
Model
Sound Pressure Level
Accuracy

AC2124A
124dB
+/- 0.2dB

Frequency

250Hz (+/- 2%)

Power Supply

4 x AA Batteries

Size  Weight

1.5” x 8.8”, Approx 1.7 lb. (800g)

24
Sound Measurement Microphones
Sound  Vibration Meters

Features:	
●●Wide

frequency range: wide range of measurement from low-frequency
to ultrasonic frequency (1Hz to 200 kHz) is possible.

●●Wide

dynamic range: measurement of high sound pressure level up to
170dB is possible.

●●Preamplifier integrated microphone: it is also possible to connect directly

to an analyzer by using a BNC cable.

●●High

sensitivity and certain measurement: it is possible to connect
directly to analyzer such as an FFT and is easy to measure with low sound
pressure levels.

Preamplifier integrated microphone
The preamplifier integrated microphone is compatible with IEPE (Integrated
Electronic Piezoelectric Microphone) and can be connected directly to an FFT
analyzer or other analyzers, enabling measurements with a highly accurate
resolution at low cost. Long-term stability is accomplished by an accelerated
aging process under 360°C.

Specifications
Model

4156N

4158N

Nominal Diameter
Release Voltage
Pressure Sensitivity

4160N

4152N

4153N

Maximum sound pressure level
Self-noise level
Temperature coefficient

7147A

7312

7313

1/2 inch

-56dB (1.6mV/Pa)

-48dB (4.0mV/Pa)

-58dB +/- 3dB

-50dB +/- 3dB

-32dB (25.1mV/Pa)

-26dB (50.0mV/Pa)

-36dB (15.8mV/Pa)

-38dB (12.5mV/Pa)

-33dB +/- 3dB

-40dB +/- 3dB

-27dB +/- 2dB

-37dB +/- 2dB

-39dB +/- 2dB

Polarization Voltage
Frequency Characteristics

7146

1/4 inch

0V
20Hz to 40kHz

20Hz to 100kHz

20Hz to 10kHz

168dB

150dB

130dB

140dB

135dB

18dB

17dB

0.7dB/°C

0.01dB/°C

45dB

20Hz to 20kHz

20Hz to 10kHz

23dB

0.01dB/°C

0.009dB/°C

Power supply and voltage

20Hz to 20kHz

20Hz to 10kHz

20Hz to 40kHz

20Hz to 20kHz
140dB
20dB

0.009dB/°C

DC15 to 28V

Constant current

0.5 to 4mA

Connector

SMB connector

Dimensions

Ø 0.27 x 1.9

BNC connector
Ø 0.5” x 2.9”

Ø 0.5 x 3.1

Ø 0.5” x 2.9”

Measurement condenser microphone
The condenser microphone can be used for measurements of sound
pressure levels with high resolution over wide frequency ranges. A variety of
microphone diaphragm diameters 1, 1/2, 1/4, 1/6, 1/8 inch and corresponding
frequency characteristics support various applications.

Specifications
Model

7012

Nominal Diameter
Release Voltage

7013

7016

1/2 inch

7017

7116

7118

1/6 inch

1/4 inch

1/8 inch
-72dB
(
 0.25mV/Pa)

-20dB
(
 100.0mV/Pa)

-26dB
(
 50.0mV/Pa)

-28dB
(39.8mV/Pa)

-26dB
(
 50.0mV/Pa)

20Hz to 20kHz

20Hz to 10kHz

20Hz to 10kHz

20Hz to 40kHz

20Hz to 20kHz

-36dB
(1.58mV/Pa)

-38dB
(
 12.5mV/Pa)

-49dB
(3.5mV/Pa)

-58dB
(1.3mV/Pa

-60dB
(
 1.0mV/Pa)

10Hz to 40kHz

10Hz to 20kHz

20Hz to 100kHz

20Hz to 20kHz

20Hz to 10kHz

Polarization Voltage
Frequency Characteristics

160dB
19dB

164dB
21dB

34dB

Dimensions

25

7023

7047A

7146NL
1/2 inch

0V

170dB
46dB

45dB

Temperature coefficient
Preamplifier Type

7022
1 inch

200V

Maximum sound pressure level
Self-noise level

7020

140dB
65dB

2dB

146dB
11dB

160dB
13dB

134dB

20dB

15dB

0.007dB/°C
Type 4212
Ø 0.5 x 0.5

Type 4116
Ø 0.3 x 0.4

Ø 0.2” x 0.4”

Type 4022
Ø 0.1” x 0.2”

Ø 0.9” x 1.1”

Ø 0.9” x 0.7”

Type 4212

Type 4011

Ø 0.5” x 0.6”
Vibration Meter Model 4200
Compact, Easy-to-Use Vibration Meter

Sound  Vibration Meters

The new Kanomax vibration meter is designed to be compact, lightweight and
easy to use. The magnetic accelerometer attaches easily to machinery for
increased accuracy and precise operation. It’s the perfect tool to diagnosis
problematic vibrations with your industrial machinery or manufactured
products.

Features:	
●●Compact size maximizes technician mobility
●●Magnetic accelerometer increases measuring accuracy and ease-of-use
●●Includes

meter, accelerometer with cable and magnet, contact pin, 2 x
AAA batteries, carrying case, and Calibration Certificate

Specifications
Measurement Range
Acceleration

0.02 to 200 m/s2 RMS

Velocity

0.02 to 200mm/s RMS

Displacement

2 to 2000 µm EQp-p

Frequency Range
Acceleration
Velocity

3Hz to 10kHz
10Hz to 1kHz (Compliant with JIS B0907-1989)

Displacement
Readings

10 to 400Hz
RMS, Peak, EQ Peak, EQ Peak-to-Peak

Output
AC Output
Headphone Output
Interface

1Vrms (Full Scale)
Portable Headphones w/Volume function
RS232C

Operating Environment
Temperature
Humidity
Power Supply

-10 to 50°C
30% to 90% (no condensing)

Accessories
AC-1046:
AC7812B:
ACPV-0148:
A
 CPV-5050:
ACNA-0134:
ACSS-22M:
ACBC-0071:
ACBC-0116-3:
A
 CNA-0116:
A
 CBC-0026:

AC adapter
Pickup
Spare Magnet
Strong Magnet
Auscultation Rod
Stud
BNC-Pin Cable
E
 xtension Cable 3m*
Data processing software
Communication Cable (RS232)
*Additional cable lengths available

2 x AAA batteries or AC adapter

Dimensions  Weight

5.7”(H)x1.9”(W)x0.9”(D) Approx. 4.6 oz.

Vibration Monitoring System
Model ACCM-393 is a system to monitor vibration of machinery and other
applications. It can quickly detect abnormal vibration condition and provide
alarm signal. CM-393 is ideal for continuous vibration monitoring at power
plants or production facilities. Either of 2 measurement modes, acceleration
or displacement, can be easily selected on the front panel.
Includes storage case, 1 x acceleration pickup with built-in amplifier (Model
7828), pickup cable, and pickup cover.
Specifications
Model
Input Impedance

ACCM-393
Approx. 100kΩ

Measurement Sensitivity

20mV/m/s2 (200mV/G)

Measurement
R
 ange

Acceleration

1 to 100mm/s2 (0.1 to 10G)

Displacement

1 to 1000µp-p

Frequency
Response

Acceleration

5Hz to 1kHz

Displacement

10Hz to 200Hz

Accuracy
Filter

Output
Power Supply
Dimensions

Within +/- 5% of full scale
High-pass
Low-pass
Signal for
level recorder
Waveform  ignal
s

Cut-off Frequency 5Hz to 12dB/oct
Cut-off Frequency 200Hz to 12dB/oct
Cut-off Frequency 1kHz to 12dB/oct
4 to 20mA/full-scale
Terminal block on the back panel (M4)
0 to 2 Vp-p/full-scale
BNC connector on the front panel
Terminal block on the back panel (M4)
AC 100V +/- 10%, 60Hz
7.9”(H) x 7.5”(W) x 18.5”(D)

*Variety of acceleration pickups are available.

26
Applications

Dust Monitors
27

Monitoring worker exposure to airborne contaminants

Indoor Air Quality Assessments

Dust Monitors
●●Measures

PM 10, Respirable, or PM 2.5
particle matters

●●Real-time

measurements
concentration

of

dust

●●Data logging up to 500 measurements
●●Simple

cleaning mechanism for easy
maintenance

●●Complete

with data processing software,
RS232C cable, cleaning kit, Ni-MH battery
pack, AC adapter, carrying case, and
calibration certificate

Digital Dust Monitor
Model 3443

Dust Monitors

Piezobalance Dust Monitor
Model 3521 / 3522

●●Measures PM 10 particle matters
●●Compact and Lightweight unit
●●Analog output controls other devices
●●Data logging up to 100,000 measurements
●●PC

interface with USB and software for
downloading data to your PC

●●Complete with rubber protector, shoulder

strap, data processing software, USB
cable, AC adapter, 2 pcs. LCD protective
sheet, rubber cap, 2 pcs. filter, and
calibration certificate

Specifications

Model

3521 / 3522

3443

Measuring Method

Piezobalance

Light Scattering

Particle Size Range

0.1 to 10 µm (Model 3521)
0.1 to 2.5 µm (Model 3522)

0.1 to 10 µm

Measuring Range

0.01 to 10.00 mg/m3

0.001 to 10.000 mg/m3

1.0 L/min

1.0 L/min

500 measurements

100,000 measurements

RS232C

USB

Analog Output

n/a

0 to 1 V / Pulse / Alarm

Power Supply

Ni-MH Battery or AC 100 - 240 V

Li-ion Battery or AC 100 - 240 V

W2.6” x H7.1” x D5.9”

W6.7” x H2.7” x D4.3”

3.9 lbs (1.75 kg)

2.9 lbs (1.3 kg)

Flow Rate
Datalogging
Interface

Dimensions
Weight

28
Piezobalance Dust Monitor Model 3521/3522
Optimal Tool for Monitoring Oil Mist
Features:	
●●Measures

PM 10, Respirable, or PM 2.5 particle matters, such as dust, oil
mist, fumes, and soot

●●Real-time measurements of dust concentration

Dust Monitors

●●Data

logging up to 500 measurements and data may be reviewed on screen
or printed

●●PC interface with RS232C and software for downloading data to your PC
●●Easy operation and requires no special training
●●Simple cleaning mechanism for easy maintenance
●●Includes data processing software, RS232C cable,

cleaning kit, Ni-MH battery
pack, AC adapter, carrying case, and calibration certificate

Specifications
Model

3521

3522

Measuring Method
Particle Size Range
Measuring Range

Piezobalance
0.1 to 10 µm

1.0 L/min
500 measurements

Interface
Power Supply
Dimensions
Weight

Software Included

0.01 to 10.00 mg/m

3

Flow Rate
Datalogging

Accessories

0.1 to 2.5 µm

RS232C
Ni-MH Battery or AC 100 - 240 V
W2.6” x H7.1” x D5.9”
3.9 lbs (1.75 kg)

3521-01:
3521-02:
3521-03: 
3521-04:
3
 521-05:
3
 521-06:
3521-07:
3521-08:
3521-20:
D
 PU-S245:
T
 P-202L:

Rechargeable Battery Pack
Carrying Case
10 µm Impactor Nozzle (for 3521)
4 µm Impactor Nozzle (for 3521)
AC Adapter
Cleaning Sponges (3 pieces)
Cleaning Fluid
Communication Cable to PC
P
 rinter Cable
Portable Thermal Printer
Rolled Printer Paper (10 rolls)

Dust Measuring Methods
Piezobalance Method
An air sample enters the system, it travels through
the impactor, which captures and removes larger
particulates away from the sample. Smaller
particulates become electrically charged and
deposited on the piezo-crystal. The total mass
of the deposited particulates affects the piezocrystal’s frequency. Since the change in frequency
is proportional to the mass of the particulates, the
actual weight of the particulates is obtained.
Since some particle matters such as oil mist absorb
lasers, the Piezobalance dust monitor would be ideal
(the light scattering method would not give correct
measurements).

Applications:
• Monitoring milling operation 
• Monitoring honing
• Monitoring boring operation

Monitoring Milling Operation

Light Scattering Method
When a laser hits particle matter, light scattering occurs. A dust monitor collects the amount
of scattering light and calculates the mass concentration in proportion to the scattering light.
Mass concentration is based on the density of particle matter, thus gravimetric sampling is
required if the density is unknown.
Applications for light scattering dust monitor include Indoor air quality investigations, Point
source monitoring, and Personal exposure monitoring.

29

Monitoring Welding Operation
Digital Dust Monitor Model 3443
Features:	
●●Measures PM 10 particle matters, such as dust, fumes, and smoke
●●Real-time and long term measurements of dust concentration
●●Compact and Lightweight unit
●●PC

Dust Monitors

interface with USB cable and software for downloading data to
your PC

●●Analog output controls other devices
●●Data logging up to 100,000 measurements and displays MIN / MAX /

AVG and Timing graph for review

●●Complete

with rubber protector, shoulder strap, data processing
software, USB cable, AC adapter, 2 pcs. LCD protective sheet, rubber
cap, 2 pcs. filter, and calibration certificate

Specifications
Model
Measuring Method
Particle Size Range
Measuring Range
Flow Rate
Datalogging
Interface
Analog Output
Power Supply
Dimensions
Weight

3443
Light Scattering
0.1 to 10 µm
0.001 to 10.000 mg/m3
1.0 L/min
100,000 measurements
USB
0 to 1 V / Pulse / Alarm
AA Battery or AC 100 - 240 V
W6.7” x H2.7” x D4.3”
2.9 lbs (1.3 kg)

Software Included

With Rubber Protector

Accessories
3442-01:
3442-02:
3442-04:
3442-05:
3442-10:
3431-03:
6113-02:
C
 X-440:

Analog Output Cable
Rubber Protector
L
 CD Protective Sheet
C
 arrying Case
Rubber Cap
Filter (10 pc)
A
 C Adapter
T
 ripod

CEGRIT Automatic Flyash Sampler
Isokinetic Sampling for Particle Emission Testing

With no moving parts, each CEGRIT sampler collects a sample from
one point in the boiler duct. Operating on duct vacuum to drive its
atmospheric-air ejector, the CEGRIT maintains near-isokinetic sampling
to keep collecting unbiased sample as boiler load and duct vacuum vary.

Features:	
●●Operates continuously with no moving parts
●●High efficiency sampling rate for fine pulverized dust
●●Industrial construction
●●Cyclone removes to facilitate periodic cleaning
●●Two manometer tap points to monitor draft pressure

Accessories

Applications:
●●All

combustion or other processes where airborne particle emission
occur

●●Smoke and particle emissions or grit burdens from boiler stacks
●●Carbon determination in Fly ash
●●Incinerator emission compliance

C8406:
C8408:
C8407:
C3101:
C3102:
C3103:
C3104:
C5019:
C
 5020:

Probe (2 m)
P
 robe (3 m)
P
 robe Extension (10 ft)
Inlet Nozzle 1/2” (12.7 mm) 
Inlet Nozzle, 5/8” (15.9mm)
Inlet Nozzle, 3/4” (19.0 mm)
Inlet Nozzle, 3/8” (9.5 mm) 
Heater Jacket, 100W, 240 Volt 
Heater Jacket, 100W, 110 Volt

30
Applications

Particle Counters
31

Cleanroom Certification

Indoor Air Quality Investigation

Filter Testing

Particle Counters
Particle Counters

Handheld Particle Counter
Model 3887

Portable Particle Counter
Model 3910  3905

Handheld Particle Counter
Model 3886 Geo-α

Simple and easy-to-use, affordable
handheld particle counter

Lightweight
handheld
particle
counter with environmental sensors

Small
footprint,
lightweight
portable particle counter available
with two different flow rates

●●Simultaneously measures and

●●Simultaneous measurements of

●●Simultaneous measurements of

●●Built-in flow sensor ensures high

●●Air velocity, temperature, and

displays 3 particle sizes

5 particle sizes

accuracy measurement

●●Store

up
measurements

to

10,000

●●Data

logging
measurements

interface with RS232C
and software for real-time
measurements and downloading
data to your PC

Light Source
Counting Efficiency

to

enclosure

touch
displays
all
simultaneously

●●Store

up
measurements

Stainless

screen LCD
measurements
to

10,000

●●PC

software allows Remote
control, Real-time measuring,
and Registering map

●●Includes

zero filter, metal
handle, 4 pcs. AA batteries,
battery charger, AC adapter, and
calibration certificate

3887

footprint,

●●Large

500

optional software for real-time
monitoring with timing graph

zero filter, meter stand, data
processing software, RS232C
cable, 4 pcs. AA batteries,
battery charger, AC adapter, and
calibration certificate

Flow Rate

up

●●Small

●●PC interface with RS232C and

●●Complete with Isokinetic probe,

Particle Sizes

●●50.0 or 28.3 L/min flow rate

humidity measurements with
optional probes

●●PC

Specifications
Model

6 particle sizes

●●Complete

with
Quick-start
guide, AC adapter, zero filter,
Isokinetic probe with 79 inch
(2 m) tubing, inlet nozzle, data
processing software, memory
card, 2 rolls of cleanroom-grade
printer paper, Li-ion battery, and
calibration certificate

3886

3910 and 3905

0.3 / 0.5 / 5.0 µm

0.3 / 0.5 / 1.0 / 3.0 / 5.0 µm

0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0 µm

0.1 cfm (2.83 L/min)

0.1 cfm (2.83 L/min)

50.0 L/min (3910) or 28.3L/min (3905)

Laser Diode

Laser Diode

Laser Diode

50+/-20% @ 0.3 µm

50+/-20% @ 0.3 µm

50+/-20% @ 0.3 µm

Coincidence Loss

Less than 5% at 2,000,000 particles/ft

Less than 5% at 2,000,000 particles/ft

Less than 10% at 500,000 particles/ft3

Zero Count Level

Less than 1 count per 5 minutes

Less than 1 count per 5 minutes

Less than 1 count per 5 minutes

10,000 measurements

500 measurements

10,000 measurements

RS232C or USB

RS232C or USB

Ethernet, USB, Memory card slot (MMC)

Datalogging
Interface
Option Sensor
Enclosure
Power Supply
Dimensions
Weight

3

3

n/a

Air Velocity / Temperature, Humidity

Air Velocity, Temperature, Humidity

Molded Plastic

Molded Plastic

Stainless Steel

4 x AA Batteries or AC 100 - 240 V

4 x AA Batteries or AC 100 - 240 V

Li-ion Battery or AC 100 - 240 V

W4.4” x H7.8” x D2.8”

W4.5” x H8.3” x D2.8”

W7.9” x H8.1” x D7.9”

1.5 lbs (680 g)

2.2 lbs (980 g)

14.2 lbs (6.44 kg)
32
Handheld Particle Counter Model 3887
Features:	
●●Simultaneously measures and displays 3 particle sizes
●●Built-in flow sensor ensures high accuracy measurements
●●Store up to 10,000 measurements

Particle Counters

●●PC interface with RS232C and software for real-time measurements and

downloading data to your PC
●●Complete with isokinetic probe, zero filter, meter stand, data processing
software, RS232C cable, USB to serial adapter, 4 pc. AA batteries, battery
charger, AC adapter, carrying case, and calibration certificate
Specifications
Model

3887

Particle Sizes

0.3 / 0.5 / 5.0 µm

Flow Rate
Light Source
Counting Efficiency

0.1 cfm (2.83 L/min)
Laser Diode
50+/-20% @ 0.3 µm

Coincidence Loss

Less than 5% at 2,000,000 particles/ft3

Zero Count Level

Less than 1 count per 5 minutes

Datalogging
Interface
Enclosure
Power Supply
Dimensions
Weight

10,000 measurements
RS232C or USB
Molded Plastic
4 x AA Batteries or AC 100 - 240 V
W4.4” x H7.8” x D2.8”
1.5 lbs (680 g)

Software Included

Accessories
3887-02:
3
 887-07:
DPU-S245:
T
 P-202L:
CX-440:
301APK:

Carrying Case
Printer Cable
Portable Thermal Printer
R
 olled Printer Paper (10 rolls)
Tripod
Intl. Plug Adapter Kit

Handheld Particle Counter Model 3886 Geo- α
Features:	
●●Simultaneous measurements of 5 particle sizes
●●Air

velocity, temperature, and humidity measurements with optional
probes
●●Data logging up to 500 measurements
●●PC interface with RS232C and optional software for real-time monitoring
with timing graph
●●Compact and lightweight, easy handling
●●Includes AC adapter, zero filter, metal handle, 4 pc. AA batteries, battery
charger and calibration certificate
Specifications
Model
Particle Sizes
Flow Rate
Light Source
Counting Efficiency

3886
0.3 / 0.5 / 1.0 / 3.0 / 5.0 µm
0.1 cfm (2.83 L/min)
Laser Diode
50+/-20% @ 0.3 µm

Coincidence Loss

Less than 5% at 2,000,000 particles/ft3

Zero Count Level

Less than 1 count per 5 minutes

Datalogging
Interface
Optional Sensor
Enclosure
Power Supply
Dimensions
Weight

33

500 measurements
RS232C or USB
Air Velocity / Temperature, Humidity
Molded Plastic
4 x AA Batteries or AC 100 - 240 V
W4.5” x H8.3” x D2.8”
2.2 lbs (980 g)

Software for Monitoring
and Remote Control

Accessories
0842:
0843:
0843-01:
3886-02:
3886-04:
S388-61:
3886-07:
3886-08:
DPU-S245:
TP-202L:
CX-440:

Temperature and RH Probe
Air Velocity Probe
E
 xtension Rod for 0843
Carrying Case
I
sokinetic Probe
D
 ata Processing Software for Windows
Printer Cable
Communication Cable to PC
Portable Thermal Printer 
Rolled Printer Paper (10 rolls)
T
 ripod

Carrying Case
Portable Particle Counter Models 3910/3905
Smallest and lightest portable particle counters
Features:	
●●Simultaneous measurements of 6 particle sizes

Particle Counters

●●Small footprint, Stainless enclosure
●●Availabe in 50.0 or 28.3 L/min flow rates
●●Complies with all requirements of ISO 21501-4
●●Large touch screen LCD displays all measurements simultaneously
●●Store up to 10,000 measurements
●●21 CFR Part 11 compliance
●●Complete

with quick-start guide, AC adapter, zero filter, isokinetic probe
with 79 inch (2 m) tubing, inlet nozzle, data processing software, memory
card, 2 rolls of cleanroom-grade printer paper, 1 x Li-ion battery, and
calibration certificate

Specifications
Model
Particle Sizes
Flow Rate
Light Source
Counting Efficiency
Coincidence Loss
Zero Count Level
Datalogging
Interface
Optional Sensor
Enclosure
Power Supply
Dimensions
Weight

3910
3905
0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0 µm
50.0 L/min
28.3L/min
Laser Diode
50+/-20% @ 0.3 µm
Less than 10% at 500,000 particles/ft3
Less than 1 count per 5 minutes
10,000 measurements
Ethernet, USB, Memory card slot (MMC)
Air Velocity, Temperature, Humidity
Stainless Steel
Li-ion Battery or AC 100 - 240 V
W7.9” x H8.1” x D7.9”
14.2 lbs (6.44 kg)

Model 3910/05

Model 3900

Small, Lightweight Unit

Optional environmental sensor

Optional Climomaster Environmental Sensor measures airflow,
temperature and humidity.
Probe Specifications
Model

6531-2G-P

Probe Type

Uni-Directional

Air Velocity

2 to 6000 fpm

Temperature
Relative Humidity

-4 to 158°F

Climomaster probe for
environmental measurements

2.0 to 98.0% RH

Accessories
6531-2G-P:
3910-01:
3910-02:
3910-03:
3910-04:
3910-05:

Air velocity, Temp, RH probe with 2 m cable
Carry case
Battery charger
Spare Li-ion battery
Alarm-output cable
Pressure-sensor (w/ connection cable)

3910-08:
3910-09:
CRVAL:

Zero filter
Printer paper
V
 alidation IQ/OQ Document

34
Handheld Condensation Particle Counter Model 3800
Optimal Screening Tool for Nano Size Particles

Features:	

Particle Counters

●●0.015 micron (15nm) minimum sensitivity
●●Handheld and Lightweight Aerosol Research Instrument
●●Store up to 10,000 measurements
●●PC

interface with USB and software for real-time
measurements and display time fluctuation graph

●●Complete with

zero filter, data processing software, USB
cable, alcohol bottle, 6 pc. AA batteries, carrying case,
and calibration certificate

Measuring Priciple
Software Included
Specifications
Model

3800

Measuring Object

Airborne Particle Matter

Particle Sizes

0.015 to 1.0 µm

Flow Rate

0.7 L/min

Light Source

Laser Diode
100% @ 0.05 µm
More than 50% @ 0.015 µm

Counting Efficiency
Coincidence Loss

Less than 5% at 100,000 particles/cm3

Zero Count Level

 1 count per 5 minutes

Alcohol Supply
Datalogging
Interface

Isopropyl Alcohol
10,000 measurements
USB

Power Supply

6 x AA Batteries or AC 100 - 240 V

Dimensions
Weight

W4.7” x H11.0” x D5.1”
3.3 lbs (1.5 kg)

Applications

Checking Worker Exposure
to Airborne Contaminants

Leakage Testing

Engine Exhaust Testing

Accessories
3800-01:
3800-02:
3800-03:
3800-04:
3800-05:
3800-06:

35

AC Adapter
Zero Filter
Alcohol Bottle
Storage Cap
Alcohol Cartridge
Spare Felt and Wire Mesh

3800-07:
3800-08:
3800-10:
3800-11:
DPU-S245:
TP-202L:

Communication Cable to PC
Carrying Case
Sampling Probe with 1m Tubing
Printer Cable
Portable Thermal Printer
Rolled Printer Paper (10 rolls)
Sampling Probe
Remote Particle Sensor with Analog Output
2-channel remote particle sensors for facility monitoring applications

Analog Particle Sensors

Kanomax particle sensors with analog output are designed to fit into your existing
monitoring system, or they can be used as a stand-alone unit to monitor a critical area
when connected to an alarm or controller. The Kanomax particle sensor is available
with a 1.0 CFM flowrate and also comes in two different ranges: 0.3 and 0.5 microns or
0.5 and 5.0 microns.

Features:	
●●Light scattering particle sensor
●●1.0 cfm flow rate (28.3 lpm)
●●Analog output makes it easy to integrate into existing systems
●●0.3 or 0.5 micron sensitivity
●●Compact body with stainless enclosure
●●Fully compatible with the Kanomax Cleanroom Monitoring System as

well as third party existing systems

Applications:	
●●Cleanroom Facility Monitoring
●●Food Industry
●●Aerospace
●●Hospital Surgical Rooms

Specifications
Model
Measuring Object
Particle Sizes
Flow Rate
Light Source
Counting Efficiency
Coincidence Loss
Zero Count Level
Interface
Enclosure
Power Supply
Dimensions
Weight

3716A

3717A
Airborne Particle Matter
0.3 / 0.5 µm
0.5 / 5.0 µm
1.0 cfm (28.3 L/min)
*External vacuum source is required
Laser Diode
50% @ 0.3 µm
50% @ 0.5 µm
Less than 5% at 500,000 particles/ft3
 1 count per 5 minutes
RS485  4-20mA
Stainless Steel
9 - 28V (AC adpter supplies 12V)
W6.3” x H5.1” x D3.9”
3 lbs (1.5 kg)

Example Configuration: Shows 6 sensors connected to datalogger via the 4-20 mA output. It’s easy to connect them to
a datalogger or any other device that will accept an analog signal. Note that a vaccuum source is required for the sensors.
Vacuum Pump
Connect up to 6 particle sensors

4-20mA connection to
datalogger or acquisition system

36
Particle Sensors and OEM Sensors

Particle Sensors

Large Particle, PM and Pollen Sensors
The pollen sensor detects the particle with light scattering method, using one light emitter and
two light receptors, and discriminates pollen from other particles with two factors “scattered
light intensity” and “degree of polarization”.
The large particle sensor is the sensor for industrial applications to detect particles sized over 10
micron. This sensor is used in production lines that dislike large particles.
“PM2.5” is known as the fine particle which is one of the cause of asthma or respiratory disease, etc. Especially, diesel soot commonly referred to as PM2.5 is suspected to raise the risk of
lung cancer. Our PM sensor has a high correlation with Federal Equivalent Method (FEM) and
achieves low-cost, simple installation for multipoint time-series monitoring.
Specifications
Model
Detectable particle size

PS2 Pollen Sensor

Large Particle Sensor

PM Sensor

approx. 15µm and larger

approx. 15µm and larger

approx. 0.3µm and larger
Concentration Range: equiv. PM2.5 (0-200 µg/m3)

Supply Voltage

12VDC

Operating Temperature

-10 to 45°C

Operating Humidity

20 to 80% RH (no condensation)

Dimensions

71.4 x 76.4 x 36.7 mm

Weight

120g (approx)

120g (approx)

130g (approx)

Particle and Aerosol OEM Sensors
Designed for easy OEM Integration
Particle Sensor Units - Based on the light scattering method, it always detects airborne particles.
Pulse output that corresponds to concentration per unit volume of particles can be obtained, with
using an original detection method based on light scattered principle similar to the particle counter.

Particle Sensor Units

Aerosol Sensor Unit - Particulates of 0.5 microns or larger can be monitored. The aerosol sensor unit
can be used to keep optimum driving of filtering equipment related to a clean room depending on
actual condition of cleanliness level.

Features:	
●●Unique air sampling method using “heater” to generate updraft, preventing choked flow. Suc-

tion pump is not necessary.

PPD42NJ

●●Stable and sensitive detection for particulates of 0.5 microns and larger can be monitored.
●●Long life laser. 20,000 hours
●●Compact size, low cost

and easy to integrate into your original equipment.

Particle Sensor Applications:	

Aerosol Sensor Applications:	

PPD60PV

●●Air Purifier

●●Clean booth/small

●●Air Quality Monitor

●●Fan Filter Unit

●●Air Conditioner

●●Clean Unit

●●Ventilation

●●Clean Room related filtration system

Aerosol Sensor Unit

●●Vaccuum Cleaner

Specifications
Model No.
Detectable Particle Size

PPD42NJ
Approx. 1 µm  larger

Supply Voltage

Output Method
Dimensions
Weight

37

AES1-05N

Approx. 0.5 µm  larger
5VDC

12VDC

0 to 45°C

0 to 40°C

95% RH or less (no condensation)

30 to 80% RH
(no condensation)

Operating Temperature
Operating Humidity

PPD60PV

Digital

Digital or Analog

Digital

59x45x22 mm

88x60x20 mm

90x90x23 mm

24 g

36 g

130 g

AES1-05N
Particle Monitor AES-1000

Continuous Particle Monitoring for Modular Clean Room Applications
The AES-1000 particle monitoring system provides near real time
stable particle measurements at low cost. Measurement covers the
equivalent range of ISO class 5 through ISO class 8.

Aerosol Monitor

It uses a long life semiconductor laser to detect optical scatter
by particles and an integral heater to provide continuous flow by
thermal convection to provide a long maintenance free life.

Features:	
●●Stable

and sensitive detection for particulates of 0.5 microns
and larger.

●●Easy-to-read

air cleanliness level, equivalent of ISO5/ Class 100
-ISO8/Class 100,000.

●●Unique air sampling method using “heater” to generate updraft,

preventing choked flow. Suction pump is not necessary.

●●Alarm

comes on when the setting level goes beyond the
required cleanliness, while monitor gives alarm signal output.

LED Light continuously shows current air cleanliness
(equiv. of Class 100 - Class 100,000)

●●Link

to Ethernet system with purpose of monitoring at multiple
locations.

●●20,000 hour long life laser.
●●Compact and light, easy installation and Low cost.

Applications:	
●●Monitor cleanrooms
●●Pharmaceutical / Semiconductor / Food Industry

Alarm comes on when the setting
level goes beyond the required
cleanliness level while monitor gives
alarm signal ouput.

Set and Reset for alarm

●●Buffer Zone Monitoring
●●Indoor Air Quality

Wall-mount Monitor

Specifications
Model
Monitoring Particle Size
Display
Sampling Method
Optics

AES-1000
0.5 microns
10 x LED lights shows equivalent of Class;
100(ISO 5), 1,000(ISO 6), 2,000, 3,000,5,000,10,000(ISO 7),
20,000, 30,000, 10,000, 100,000(ISO 8)
Updraft airflow generated by built-in heater enables air sampling
Semiconductor Laser, 20,000 hour life

Alarm Display

Alarm LED light comes on when reaching to setting level

Alarm Output

TTL Open collector output in sync with alarm display

Interface
Supply Voltage
Operating Temp/Humidity

Ethernet X 1 port
12VDC+/-10% 500mA
0 to 40 deg. C, 30 to 80%rh

PC Software is Included

38
Cleanroom Monitoring System
Cleanroom Monitoring System

The Kanomax Cleanroom Monitoring System (CRMS)
provides an automated means to monitor and gather
airborne particle counts and other parameter levels
in controlled environments. The CRMS allows users
to perform a variety of functions from a PC, including
changing alarm settings and viewing particle count
concentrations.

Features:	
●●Compact stainless enclosure with sensors

●●1 PC system controls up to 128 sensors

●●Multi-parameter

●●Alarm outputs: warning light, on-screen, or pager notification

measurements: Particle count, Air velocity,
Temperature, Humidity, and Differential pressure
●●Multi-function, user-friendly monitoring software

Particle Sensors
Features:	
●●Light scattering particle sensor
●●Durable

stainless enclosure is easy to
sanitize during whole facility cleaning

●●No

system down-time: each sensor is replaceable for repair and
recalibration

Specifications
Model
Measuring Object
Particle Sizes
Flow Rate
Light Source
Counting Efficiency
Coincidence Loss
Zero Count Level
Interface
Enclosure
Power Supply
Dimensions
Weight

3714

3715
Airborne Particle Matter
0.3 / 0.5 µm
0.5 / 5.0 µm
0.1 cfm (2.83 L/min)
*External vacuum source is required
Laser Diode
50% @ 0.3 µm
50% @ 0.5 µm
Less than 5% at 1,000,000 particles/ft3
 1 count per 5 minutes
RS485
Stainless Steel
DC24V (Supplied from the 3770)
W5.0” x H2.8” x D1.6”
1.1 lbs (500 g)

Specifications

Features:	

Model
Measuring Object
Particle Sizes
Flow Rate

●●Light scattering particle sensor
●●1.0 cfm flow rate (28.3 lpm)
●●Analog

output makes it easy to integrate
into existing systems

Light Source
Counting Efficiency
Coincidence Loss
Zero Count Level
Interface
Enclosure
Power Supply
Dimensions
Weight

3716A

3717A
Airborne Particle Matter
0.3 / 0.5 µm
0.5 / 5.0 µm
1.0 cfm (28.3 L/min)
*External vacuum source is required
Laser Diode
50% @ 0.3 µm
50% @ 0.5 µm
Less than 5% at 500,000 particles/ft3
 1 count per 5 minutes
RS485  4-20mA
Stainless Steel
9 - 28V
W5.9” x H4.7” x D3.9”
3 lbs (1.5 kg)

Specifications

Features:	
●●0.2 µm sensitivity particle counter
●●Digital and analog outputs are available
●●Analog output for Multiplexer

39

Model
Measuring Object
Particle Sizes
Flow Rate
Light Source
Counting Efficiency
Coincidence Loss
Zero Count Level
Interface
Power Supply
Dimensions
Weight

3792-01
Airborne Particle Matter
0.2 / 0.3 µm
0.1 cfm (2.83 L/min)
Laser Diode
50% @ 0.2 µm
Less than 5% at 1,000,000 particles/ft3
 1 count per 5 minutes
RS485 / 4 to 20 mA for Multiplexer
AC 100 to 240 V
W4.6” x H5.3” x D6.7”
6.6 lbs (3 kg)
Interface Box
●●Converts analog input to digital
●●Supplies power to the sensor

Specifications
Model
Input
Output
Power Supply
Dimensions
Weight

3772-02
4 to 20 mA / 0 to 1 V / 1 to 5 V
RS485
DC24V (Supplied from the 3770)
W5.5” x H3.1” x D1.6”
1.1 lbs (500 g)

with Differential Pressure Sensor

Distributors
Features:	

Cleanroom Monitoring System

Features:	

Specifications

●●Supplies

data communication and power
to sensors via RS485
●●1 unit connects up to 8 sensors

Model
# of Channels
Interface
Power Supply
Dimensions
Weight

3770
8
RS485
AC 85 to 132 V or AC 170 to 267 V
W11.8” x H3.9” x D7.9”
6.6 lbs (3 kg)

Cleanroom Monitoring Software
Features:	
●●Continuous monitoring and data processing software
●●Remote monitoring via LAN
●●Multi-function:

Map display at a glance, Real-time graph,
Maintenance Indication, and Data table

Map

Trend Graph

System Examples
Industrial Cleanroom

Bio-medical Cleanroom

Semiconductor, HDD, Flat Panel Display, Electronics

Food, Pharmaceutical, Hospital surgical rooms

40
Aerosol Research Instruments

Our line of aerosol research
instruments covers a variety
of applications.

Aerosol Research
Instruments

Industrial hygiene

Nanosampler

Health effect research
Nano-particle research

Portable Aerosol Mobility Spectrometer
(PAMS)

Condensation
Particle Counter

Air pollution
Ambient air quality research

Ambient Air Quality Monitor

Earth sciences

Black Carbon Monitor

Environmental research
Climate change research
41

Aerosol Particle Mass Analyzer
Small, lightweight mobility size spectrometer
Features:	
●●Electrical

mobility size spectrometer designed for portable, mobile, or
handheld aerosol sampling applications

●●Provides number-weighted diameter distribution of aerosols over the entire

submicrometer range (10 to 863nm) in one scan

●●Uses

a non-radioactive bipolar aerosol charger to allow easy access to
sampling sites with tight safety regulations

●●Bipolar

charger significantly reduces measurement uncertainty of larger
particles in the submicrometer range

●●Can be used in two modes: Single diameter count mode: to get a total count

within a narrow size range or Size distribution mode: to get an automated
size distribution measurement over desired size range or size resolution

Aerosol Research Instruments

Portable Aerosol Mobility Spectrometer (PAMS)

●●Stand-alone, battery-operated instrument; no external computer needed
●●Weighs only 4.5 kg and measures 23 x 23 x 15 cm

Operation:
Sampled aerosol is charge-conditioned using a
non-radioactive, bipolar charger which brings
the particles to steady-state charge distribution.
Particles are subsequently sorted according to their
electrical mobility in a differential mobility classifier
(DMA). Classified particles are detected and counted
downstream using a condensation particle counter
(CPC).

Specifications
Model

3300

Measuring Modes

Single size mode (single size concentration)
Scan mode (size distributions)
0.2LPM: 14.5 to 863 nm
0.4LPM: 10 to 433 nm

Size Range

0.2LPM: 14 channels
0.4LPM: 27 channels

Scan Resolution

0.2LPM: 42 sec to minutes for one scan
0.4LPM: 81 sec to minutes for one scan

Scan Time
Concentration Range

0 to 100,000 particles /cm3

Flow Rate

Inlet: 0.7 LPM
Sample: 0.05 LPM

Bipolar Charger

Inlet: 0.7 LPM
Sample: 0.05 LPM

Condensing Fluid

Isopropyl alcohol

Display

Color touchscreen

Dimensions

9.1 x 9.1 x 5.9 inch (23 cm x 23 cm x 15 cm)

Weight

9.9 lbs (4.5 kg)

Accessories
3300-05:
3300-08:
3300-09:
3300-10:

Spare Alcohol Cartridge
Spare Lithium-ion Battery
Battery Charger
Carrying Case

42
Aerosol Research Instruments

Aerosol Particle Mass Analyzer Model APM-3601
Lightweight, Desktop Aerosol Particle Mass Analyzer
Aerosol Particle Mass Analyzer classifies particles by mass based on the
balance between centrifugal force and electrostatic force. Particle size
distribution measurement is normally used in order to measure nanosized
particle distribution. While DMA (Differential Mobility Analyzer) classifies
particles by particle size utilizing electrostatic force, APM classifies particles
by mass based on entirely new classification principles. APM classifies
aerosol particles of 0.01 ~ 100 femtograms.
Particle density distribution can be attained by combining the APM and DMA.

Applications:
●●Mass distribution measurements
●●Particle density research
●●Monodisperse aerosol generation

Features:	
●●Lightweight, Desktop Unit
●●Measurement software, communication cable, extension board, extension

adapter and box, and plastic case are available as options

Specifications
Model
Particle Mass Range
Maximum Rotation Speed
Maximum Voltage
Rotating Cylinder Dimensions
Sampling Flow Rate
Power Supply
Dimensions
Weight

APM-3601

Rotating Cylinder Assembly

0.01 to 1000 femtogram
Equivalent to approx. 14 nm to 1.3 µm for particle density of 1 g/cm3)
1,000 to 14,000 rpm
Up to 2000V
Inner Cylinder Diameter: 48 mm
Gap between Inner and Outer Cylinders: 1 mm
Cylinder Length: 100 mm
0.3 to 1.0 L/min (0.3 L/min is recommended)
Single-phase AC100 ~ 240V
Main Unit: W 16.9” x H 5.5” x D 7.9”
Control Unit: W 16.9” x H 7.1” x D 13.8”
Main Unit: 23 lbs (10.5 kg)
Control Unit: 15 lbs (7 kg)

Control Unit

APM experimental data comparison with
theoretical curve at 100nm (PSL)

43
Aerosol Research Instruments

Black Carbon Monitor Model 3130
Continuous Soot Monitoring System

The Model 3130 monitors changes in transmittance across an automatically advancing
quartz fiber filter tape using an LED at a 565 nm wavelength. To achieve measurements
with high sensitivity and a lower detectable light absorption coefficient, the Model
3130 uses a double-convex lens and optical bundle pipes to maintain high light
intensity and signal data, obtained at 1000 Hz.

Features:	
●●Measures black carbon concentration in the air in real time
●●Preprocessing of sample air improves measurement accuracy
●●Automatic filter feed enables continuous measurement for extended periods
●●Advanced detection sensitivity enables measurement in low concentrated areas
●●Complete with inlet heater, pump, pump controller, 2.5 µm-cut impactor, software,

and power cable

Specifications
Model
Measurement Method
Light Source
Lowest Detection

3130
Light Absorption Method
LED Wavelength 565 nm
0.05 µg/m3 @1 min @0.8 L/min

Collection Flow Rate
Collection Filter

0.8 L/min
Fiberglass Filter (Length: 25 m)

PC Interface

USB

Power Supply
Dimensions

100 VAC, 6A
W 17” x H 8.7” x D 13.9”

Weight

37.5 lbs (17kg)

System Diagram

Nanosampler Model 3180
Inertial Filter for Aerosol Collection
Fiber filters utilize several mechanisms for aerosol collection such as inertial,
gravitational settling, interception, and diffusion. Each collection efficiency depends
on particle size and filtration velocity. Large particles are collected in a filter by inertial
impaction at a high filtration velocity while small particles are removed from the air
by Brownian diffusion.The Model 3180 controls its diffusion collection increasing
filtration velocity, and collecting particles only by inertia.

Features:	
●●Particle classification as small as 0.1 µm
●●5 stages of PM10, 2.5, 1.0, 0.5, 0.1 and backup filter
●●Large sample flow rate of 40 L/min
●●Includes 1 box of quartz fiber filter (55mm x 100 pc.), and 5 sets

of PM0.1 cartridge

●●Pump unit (Vacuum pump + Rota meter with valve), PM0.1

cartridge case (1 case for 5 cartridges), Quartz fiber filter case
(1 set of 20 cases), PM0.1 absorption materials (10 sets) are
available

Specifications
Model
Classification Method

3180
Inertial Collection

Classification Range

PM 10, PM 2.5, PM 1.0, PM 0.5, PM 0.1 and Backup filter

Absorption Material

Quartz Fiber Filter: PM 10, PM 2.5, PM 1.0, PM 0.5, and Backup fiter
SUS fiber: PM0.1

Sampling Flow Rate

40 L/min

Dimensions

Ø 3.5” x 8.3”

Weight

6.6 lbs (3 kg)

44
Dust Sentry
Aerosol Research Instruments

Accurate and affordable real-time outdoor particulate monitoring
The Dust Sentry provides real time, continuous particle measurement of TSP,
PM10, PM2.5 or PM1 with datalogging and alarm relay outputs. Numerous
integrated options including SMS and/or email notifications, weather sensors,
noise and mounting equipment make the Dust Sentry a flexible base system
suitable for a wide range of applications.

Features:	
●●Real time particle monitoring, Measures and reports data in 1 minute intervals
●●Accurately measures TSP, PM10, PM2.5 or PM1
●●On board data storage for up to 15 years of data
●●Web-based and GSM data retrieval and alerts
●●Gravimetric filter
●●Option to integrate weather, wind, temp/RH sensors, and sirens
●●Tough weatherproof enclosure with solar shielding for hot climates
●●Quick installation and relocatable
●●Datalogging and configuration software included

Specifications
Model

DS 10

Measurement Technology

DS 25

DS 1

Near forward light scattering nephelometry

Range

0-2000 µg/m3

Accuracy

+/- (2 µg/m3 + 5% of reading)

Cut Points Available

PM 10

PM 2.5

Sample Flowrate

PM 1.0

2.0 LPM

Inlet

Heated

Alarm

Relay and SMS (optional)

Datalogging

2GB SD card (15 years data storage)

Reference Filter

25mm GF filter

Outputs

RS232  cellular IP Gateway

Wireless Communication Options

Low power RF modem

Environmental Operating Range

-10°C to +45°C; 10 to 95% RH (NC)

Power Requirements

Mains (100 to 260VAC) or regulated 12VDC (10W)
Lockable  weatherproof GRP cabinet IP65; 685 x 350 x 300mm (including
solar shield armour  mounting brackets)

Enclosure
Weight

Compared to reference-equivalent monitors like TEOMs and
BAMs which lack sensitivity at short sampling periods, the
Dust Sentry reports accurately at 1 minute intervals and has a
much lower total cost of ownership (up to 5x less). In addition,
the Dust Sentry is compact, re-locatable and can integrate
wind and weather sensors for maximum utility.

12kg

Accessories
OPTIONAL SHARP CUT CYCLONES

REMOTE COMMUNICATION OPTIONS

DS SC10:
DS SC25:
DS SC1:

DS R39:
DS R40:
DS O12:

PM10 Cyclone
PM2.5 Cyclone
PM1.0 Cyclone

Gatetel GT-HE910 Modem
Orbit Serial Port Extender
12 Month Orbit Data Access

OTHER SENSORS AND MODULES

45

CONSUMABLES, PARTS AND TOOLS

DS AUX:
DS 420:
DS R47:
DS R1:
DS R25:

DS CM:
DS R10:
DS R31:
DS R20B:
DS R34:
DS R56:
CAL R4:

Auxillary Module
4-20mA Module
47mm AnalysisFilter
Gill WindSonic Sensor
Vaisala WTX520 Weather Transmitter

Control Module
Sampling Pump
PM Filters
Flowmeters (x2)
PM Purge Pump
Zero Filter  Flow Assembly
Factory Calibration PM
AQM60 Ambient Air Quality Monitor
●●Multiple gas sensors (up to six)

●●Wired or Wireless communication

●●Temperature and humidity sensors

●●Large data storage capacity (15 years)

●●Particulate monitor (PM2.5 or PM10)

●●Thermal management system

●●Wind speed and direction anemometer

●●Weatherproof and compact enclosures

●●Noise and meteorological sensors

●●Insulated

roadside boxes, Metal enclosures,
Mounting accessories are available

●●Rapid real-time data sampling (2-minute)

Specifications
Model

AQM60

Aerosol Research Instruments

Features:	

Standard: 50 to 77°F (10°C to 25°C)
With Heater/Cooler: -4 to 113°F (-20°C to 45°C)
10 to 90 %RH (non condensing)

Environment Operation Range
Communications Options

GSM modem / RF modem / Ethernet LAN

Other Sensor Options

Noise sensors / Particle counters / Meterological sensors

Power Supply

12VDC / AC power module 100 - 250 VAC

Enclosure

Standard: Fiber reinforced polycarbonate

Dimensions

H 23.6” x W 19.7” x D 9.1” (600 x 500 x 230 mm)

Weight

33 to 55 lb (15 to 25 kg), varies depending on configuration

Gas Sensor Specifications
Gas Sensor Heads

Sensor

Range (ppm)

Accuracy

Resolution

Ozone

GSS

0 - 0.150

+/- 0.005 ppm

0.001 ppm

Power
2.9 W

Nitrogen dioxide

GSS

0 - 0.200

+/- 15%

0.001 ppm

12.3 W

Carbon monoxide

GSS

0 - 100

+/- 5 ppm

0.1 ppm

2.5 W

Sulphur dioxide

GSS

0 - 10

+/- 0.5 ppm

0.01 ppm

2.5 W

Non-methane hydrocarbon

GSS

0 - 25

+/- 0.5 ppm

0.1 ppm

2.5 W

VOC isobutylene

GSS

0 - 25

+/- 10%

0.1 ppm

2.5 W

VOC isobutylene

PID

0 - 20

10 %

0.01 ppm

1.5 W

Hydrogen sulphide

GSS

0 - 10

+/- 0.5 ppm

0.01 ppm

2.5 W

Carbon dioxide

NDIR

0 - 2,000

40 ppm + 3%

1 ppm

1.0 W

Fluidized Particle Generator
Kanomax Fluidized Particle Generators generate standard particles
continuously for extended periods. Applications include pharmaceutical and
chemical compound exposure research, and filtration performance testing.

Features:	
●●Available for variety of sample particles
●●Long time operation
●●Stably generated particle concentration
●●Easy control on the concentration level

Specifications
Model
Particle Sizes
Density
Flow Rate
Power Supply
Dimensions

Model 3211

3211

3216

Less than 20 µm

0.2 to 10 µm

100 mg/m3 to 1 g/m3

20 g/m3

30 L/min
AC100V, 50/60Hz
W 6.7” x H 17.5” x D17.7”

W 20.1” x H 61.4” x D15.9”

Model 3216
46
Applications

Mass Spectrometry

Food analysis

Forensic Science

Medical

Environmental analysis

Mass Spectrometry

47
Time of Flight Mass Spectrometer

High performance and small foot print time of flight mass
spectrometer by using multi-turn technology

Mass Spectrometry

Features:	
●●High resolution and compact time of flight mass spectrometer
●●Impure substance analysis in gases with the detection limit of 10ppb
●●Onsite detailed accurate analysis

Applications:
●●Food analysis
●●Particle analysis
●●Environmental analysis
●●Forensic: detection of forged paintings
●●Medical: blood test, hair analysis etc

Specifications
Resolution

More than 30,000

Dynamic Range

10 bit

Sampling Flow Rate

2GS/s

Dimensions
Weight

H9.2” x W17.9” x D25.2”
79 lbs (36kg)

with Gas chromatography (optional)

Key Technology “Multi-Turn Time of Flight Mass Spectrometry”
There are four orbiting electrodes and two injection/
ejection electrodes on a palm-sized optics bench.
The ion source gives kinetic-energy for orbiting
motion in the infinite loop. Injection and ejection
electrodes are synchronized with ion source pulsing
triggering. Injection electrode has to be in the on
state while ions enter the analyzer, and then has to
be turned off before first ion (smallest ion) returns
to it. Orbiting electrodes are constant, so orbiting
ion can be held until ejection electrode is ON.

48
Automotive Testing Instruments

Automotive
Testing
Instruments
Amenity Manikin System
Cabin Comfort Test Rig

Kanomax Amenity Manikin System is a solution for precise interior
cabin comfort evaluation. The system measures parameters; air velocity,
temperature, humidity, and radiant heat; which relate to human comfort
level. One mannequin equips more than 120 sensors all over its surface and
provides sophisticated measurement.

Features:	
●●4 mannequins measure simultaneously; understanding the entire cabin
●●Wireless connection brings easy operation
●●Graphical software for both real-time measurement and data retrieval
●●Excel compatible data output as well as saved graphical data for review

Amenity Manikin

Sensor Allocations by Part
Head
Torso
Lower Body
Total

Air Velocity
4
12
20
36

Temperature
12
33
33
78

Humidity
1
0
1
2

Radiant Heat
3
5
4
12

Mannequin Specifications
Height
Weight
Material

Approx. 5’6” (170 cm)
90 lbs (40 kg)
Plastic (FRP)

Sensor Specifications
Air Velocity Ranges
Accuracy
Temperature Ranges
Accuracy
Humidity Ranges
Accuracy
Radiant Heat Ranges
Accuracy
Wavelength

49

0.10 to 5.00 m/s
0.05 m/s
30 to 100°C
+/- 3.0°C
3 to 95% RH
+/- 3% RH
0 to kw/m2
+/- 7%
0.3 to 40 µm

Measurement Software
Automotive Testing Instruments

Vehicle Cabin Leakage Testers
Cabin Comfort Test Rig
Leakage testing is performed by pressurizing or depressurizing the vehicle cabin.
The tester measures the changes in cabin pressure. The control unit consists of a
manometer and pressure transducers. The manometer detects the leakage flow,
which is calculated by measuring the nozzle pressure relative to static pressure. The
tester controls an adjustable fan to maintain static test pressure. Five selectable inlet
nozzles provide reasonable leakage flow rate measurements of 2.5 to 800 cfm.

CLT-APC-NPP

1. The cab is pressurized (either positively or negatively)
2. Pressure will be stable, normally stable conditions are achieved quickly
3. The leakage rate is measured

Features:	
●●Automatic Pressure Control

●●System includes inlet nozzles, certificate of conformance

●●Negative Pressure Testing
●●Features High Accuracy Manometer

Specifications
Model

CLT-MPC

to ISO 5801: 1997, and pressure measurement
calibration certificate

Nozzle Specifications
CLT-MPC

Fan Speed
Pressure Measurement Accuracy
Leakage Flow Measurement
Accuracy
Cabin Pressure Preset Ranges
Automatic Pressure Control
Negative Pressure Plenum
Small Transport Cart
Power Supply
Duct Size
Dimensions

CLT-MPC-NPP

CLT-APC

CLT-APC-NPP

847 cfm (400 L/sec)
+/- 2 % of reading +/- 1 digit
+/- 3 % of reading +/- 1 digit
Providing flow rate is varied by the nozzle * See nozzle specs
0.2 to 2 in. H2O
―
(5 to 499 Pa)
―
⃝
―
⃝
―
⃝
⃝
―
120/240 V, 1 phase
16.4 ft x 8 inches (5 m x 203 mm)
Small Transport Cart: W 22.5” x H 23.5” x D 54”
Large Transport Cart: W 27.5” x H 42” x D 58.5”

Nozzle
F
G
H
D
E

Inlet Diameter
0.590” (15 mm)
1.122” (28.5 mm)
2.204” (56 mm)
3.543” (90 mm)
6.023” (153mm)

Flow Ranges
2.5 to 10 cfm (1.2 to 4.8 l/s)
8.5 to 36 cfm (4 to 17 l/s)
32 to 142 cfm (15 to 67 l/s)
53 to 296 cfm (25 to 140 l/s)
190 to 847 cfm (90 to 400 l/s)

Smoke Generator Model 8304
Smoke Generator for Air Flow Visualization
The Model 8304 is a smoke generator for wind tunnel testing
and aerodynamics visualization. It produces vaporizing
white smoke of 15 to 80L/min. Compressed air moves
smoke to the comb-shaped nozzle and makes streamlines.

Features:	
●●Versatile and Portable unit
●●High output, non-toxic, water-soluble smoke
●●System

cart, air compressor, and comb-shaped nozzle are
available as options

Nozzle
Specifications
Model
Smoke Output
Smoke Particle Size
Warm Up Time
Power Supply
Dimensions
Weight

8304
15 to 80 L/min
0.3 to 1.0 µm
4 minutes
AC100V, 550W
W11.8” x H13.8” x D25.6”
61.7 lbs (28 kg)

50
Katalog produk-kanomax-2014-tridinamika
Katalog produk-kanomax-2014-tridinamika
Katalog produk-kanomax-2014-tridinamika
Katalog produk-kanomax-2014-tridinamika
Katalog produk-kanomax-2014-tridinamika

Contenu connexe

En vedette

catalogo
catalogocatalogo
catalogolitus13
 
Evaluation question 3
Evaluation question 3Evaluation question 3
Evaluation question 3SeyiDaramola
 
Effects of Wind Direction on Trace Metal Concentration in Southeast Kansas
Effects of Wind Direction on  Trace Metal Concentration in Southeast KansasEffects of Wind Direction on  Trace Metal Concentration in Southeast Kansas
Effects of Wind Direction on Trace Metal Concentration in Southeast KansasSergio A. Guerra
 
Trials and Tribulations of Founding a Private Equity Business
Trials and Tribulations of Founding a Private Equity BusinessTrials and Tribulations of Founding a Private Equity Business
Trials and Tribulations of Founding a Private Equity Businessyinizhang13
 
JongwingsDesignPortfolio2016
JongwingsDesignPortfolio2016JongwingsDesignPortfolio2016
JongwingsDesignPortfolio2016Jongwings
 
Explicacion abp cursoedmodomatematicasgrupo5
Explicacion abp cursoedmodomatematicasgrupo5Explicacion abp cursoedmodomatematicasgrupo5
Explicacion abp cursoedmodomatematicasgrupo5Zully Moreno Celis
 
Pete Fraser - Top Tips (2)
Pete Fraser - Top Tips (2)Pete Fraser - Top Tips (2)
Pete Fraser - Top Tips (2)anouska1234
 
Pairing aermod concentrations with the 50th percentile monitored value
Pairing aermod concentrations with the 50th percentile monitored valuePairing aermod concentrations with the 50th percentile monitored value
Pairing aermod concentrations with the 50th percentile monitored valueSergio A. Guerra
 
Universal design for learning
Universal design for learningUniversal design for learning
Universal design for learningsusimg
 

En vedette (12)

catalogo
catalogocatalogo
catalogo
 
Evaluation question 3
Evaluation question 3Evaluation question 3
Evaluation question 3
 
Effects of Wind Direction on Trace Metal Concentration in Southeast Kansas
Effects of Wind Direction on  Trace Metal Concentration in Southeast KansasEffects of Wind Direction on  Trace Metal Concentration in Southeast Kansas
Effects of Wind Direction on Trace Metal Concentration in Southeast Kansas
 
Trials and Tribulations of Founding a Private Equity Business
Trials and Tribulations of Founding a Private Equity BusinessTrials and Tribulations of Founding a Private Equity Business
Trials and Tribulations of Founding a Private Equity Business
 
Almer grup
Almer grupAlmer grup
Almer grup
 
JongwingsDesignPortfolio2016
JongwingsDesignPortfolio2016JongwingsDesignPortfolio2016
JongwingsDesignPortfolio2016
 
Good, Fast, or Cheap.
Good, Fast, or Cheap. Good, Fast, or Cheap.
Good, Fast, or Cheap.
 
Flat Plans
Flat PlansFlat Plans
Flat Plans
 
Explicacion abp cursoedmodomatematicasgrupo5
Explicacion abp cursoedmodomatematicasgrupo5Explicacion abp cursoedmodomatematicasgrupo5
Explicacion abp cursoedmodomatematicasgrupo5
 
Pete Fraser - Top Tips (2)
Pete Fraser - Top Tips (2)Pete Fraser - Top Tips (2)
Pete Fraser - Top Tips (2)
 
Pairing aermod concentrations with the 50th percentile monitored value
Pairing aermod concentrations with the 50th percentile monitored valuePairing aermod concentrations with the 50th percentile monitored value
Pairing aermod concentrations with the 50th percentile monitored value
 
Universal design for learning
Universal design for learningUniversal design for learning
Universal design for learning
 

Plus de PT. Tridinamika Jaya Instrument

Katalog-Katronic-ultrasonic-flow-measurement-tridinamika
Katalog-Katronic-ultrasonic-flow-measurement-tridinamikaKatalog-Katronic-ultrasonic-flow-measurement-tridinamika
Katalog-Katronic-ultrasonic-flow-measurement-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamikaKatalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-digital-multimeter-L4411 a-system-tridinamika
Katalog agilent-digital-multimeter-L4411 a-system-tridinamikaKatalog agilent-digital-multimeter-L4411 a-system-tridinamika
Katalog agilent-digital-multimeter-L4411 a-system-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaKatalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaKatalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamikaKatalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...PT. Tridinamika Jaya Instrument
 
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamikaKatalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamikaKatalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamikaPT. Tridinamika Jaya Instrument
 
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamika
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamikaKatalog siglent-digital-oscilloscope-sds1000cml-tridinamika
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamikaPT. Tridinamika Jaya Instrument
 

Plus de PT. Tridinamika Jaya Instrument (20)

Company profile Tridinamika 2017
Company profile Tridinamika 2017Company profile Tridinamika 2017
Company profile Tridinamika 2017
 
Katalog-Katronic-ultrasonic-flow-measurement-tridinamika
Katalog-Katronic-ultrasonic-flow-measurement-tridinamikaKatalog-Katronic-ultrasonic-flow-measurement-tridinamika
Katalog-Katronic-ultrasonic-flow-measurement-tridinamika
 
Katalog-Katronic technologies brochure 2014
Katalog-Katronic technologies brochure 2014Katalog-Katronic technologies brochure 2014
Katalog-Katronic technologies brochure 2014
 
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
 
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
 
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
 
Katalog agilent-34420 a-micro-ohm-meter-tridinamika
Katalog agilent-34420 a-micro-ohm-meter-tridinamikaKatalog agilent-34420 a-micro-ohm-meter-tridinamika
Katalog agilent-34420 a-micro-ohm-meter-tridinamika
 
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamikaKatalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
 
Katalog agilent-digital-multimeter-L4411 a-system-tridinamika
Katalog agilent-digital-multimeter-L4411 a-system-tridinamikaKatalog agilent-digital-multimeter-L4411 a-system-tridinamika
Katalog agilent-digital-multimeter-L4411 a-system-tridinamika
 
Katalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamikaKatalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamika
 
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaKatalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
 
Katalog agilent-digital-multimeter-34401 a-tangerang
Katalog agilent-digital-multimeter-34401 a-tangerangKatalog agilent-digital-multimeter-34401 a-tangerang
Katalog agilent-digital-multimeter-34401 a-tangerang
 
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaKatalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
 
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamikaKatalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
 
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
 
Katalog agilent-dc-power-analyzer-n6705 b-tridinamika
Katalog agilent-dc-power-analyzer-n6705 b-tridinamikaKatalog agilent-dc-power-analyzer-n6705 b-tridinamika
Katalog agilent-dc-power-analyzer-n6705 b-tridinamika
 
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamikaKatalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
 
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamikaKatalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
 
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamika
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamikaKatalog siglent-digital-oscilloscope-sds1000cml-tridinamika
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamika
 
Katalog kanomax ih-solutions-tridinamika
Katalog kanomax ih-solutions-tridinamikaKatalog kanomax ih-solutions-tridinamika
Katalog kanomax ih-solutions-tridinamika
 

Dernier

How AI, OpenAI, and ChatGPT impact business and software.
How AI, OpenAI, and ChatGPT impact business and software.How AI, OpenAI, and ChatGPT impact business and software.
How AI, OpenAI, and ChatGPT impact business and software.Curtis Poe
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
unit 4 immunoblotting technique complete.pptx
unit 4 immunoblotting technique complete.pptxunit 4 immunoblotting technique complete.pptx
unit 4 immunoblotting technique complete.pptxBkGupta21
 
DevoxxFR 2024 Reproducible Builds with Apache Maven
DevoxxFR 2024 Reproducible Builds with Apache MavenDevoxxFR 2024 Reproducible Builds with Apache Maven
DevoxxFR 2024 Reproducible Builds with Apache MavenHervé Boutemy
 
WordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your BrandWordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your Brandgvaughan
 
From Family Reminiscence to Scholarly Archive .
From Family Reminiscence to Scholarly Archive .From Family Reminiscence to Scholarly Archive .
From Family Reminiscence to Scholarly Archive .Alan Dix
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity PlanDatabarracks
 
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek SchlawackFwdays
 
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
The Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsThe Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsPixlogix Infotech
 
Gen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfGen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfAddepto
 
Scale your database traffic with Read & Write split using MySQL Router
Scale your database traffic with Read & Write split using MySQL RouterScale your database traffic with Read & Write split using MySQL Router
Scale your database traffic with Read & Write split using MySQL RouterMydbops
 
Anypoint Exchange: It’s Not Just a Repo!
Anypoint Exchange: It’s Not Just a Repo!Anypoint Exchange: It’s Not Just a Repo!
Anypoint Exchange: It’s Not Just a Repo!Manik S Magar
 
DevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsDevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsSergiu Bodiu
 
DSPy a system for AI to Write Prompts and Do Fine Tuning
DSPy a system for AI to Write Prompts and Do Fine TuningDSPy a system for AI to Write Prompts and Do Fine Tuning
DSPy a system for AI to Write Prompts and Do Fine TuningLars Bell
 
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Mark Simos
 
Sample pptx for embedding into website for demo
Sample pptx for embedding into website for demoSample pptx for embedding into website for demo
Sample pptx for embedding into website for demoHarshalMandlekar2
 
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxLoriGlavin3
 
Unraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfUnraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfAlex Barbosa Coqueiro
 
Generative AI for Technical Writer or Information Developers
Generative AI for Technical Writer or Information DevelopersGenerative AI for Technical Writer or Information Developers
Generative AI for Technical Writer or Information DevelopersRaghuram Pandurangan
 

Dernier (20)

How AI, OpenAI, and ChatGPT impact business and software.
How AI, OpenAI, and ChatGPT impact business and software.How AI, OpenAI, and ChatGPT impact business and software.
How AI, OpenAI, and ChatGPT impact business and software.
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
unit 4 immunoblotting technique complete.pptx
unit 4 immunoblotting technique complete.pptxunit 4 immunoblotting technique complete.pptx
unit 4 immunoblotting technique complete.pptx
 
DevoxxFR 2024 Reproducible Builds with Apache Maven
DevoxxFR 2024 Reproducible Builds with Apache MavenDevoxxFR 2024 Reproducible Builds with Apache Maven
DevoxxFR 2024 Reproducible Builds with Apache Maven
 
WordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your BrandWordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your Brand
 
From Family Reminiscence to Scholarly Archive .
From Family Reminiscence to Scholarly Archive .From Family Reminiscence to Scholarly Archive .
From Family Reminiscence to Scholarly Archive .
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity Plan
 
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
 
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
The Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsThe Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and Cons
 
Gen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfGen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdf
 
Scale your database traffic with Read & Write split using MySQL Router
Scale your database traffic with Read & Write split using MySQL RouterScale your database traffic with Read & Write split using MySQL Router
Scale your database traffic with Read & Write split using MySQL Router
 
Anypoint Exchange: It’s Not Just a Repo!
Anypoint Exchange: It’s Not Just a Repo!Anypoint Exchange: It’s Not Just a Repo!
Anypoint Exchange: It’s Not Just a Repo!
 
DevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsDevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platforms
 
DSPy a system for AI to Write Prompts and Do Fine Tuning
DSPy a system for AI to Write Prompts and Do Fine TuningDSPy a system for AI to Write Prompts and Do Fine Tuning
DSPy a system for AI to Write Prompts and Do Fine Tuning
 
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
 
Sample pptx for embedding into website for demo
Sample pptx for embedding into website for demoSample pptx for embedding into website for demo
Sample pptx for embedding into website for demo
 
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
 
Unraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfUnraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdf
 
Generative AI for Technical Writer or Information Developers
Generative AI for Technical Writer or Information DevelopersGenerative AI for Technical Writer or Information Developers
Generative AI for Technical Writer or Information Developers
 

Katalog produk-kanomax-2014-tridinamika

  • 1. 2014 Spring Ultimate Measurement Instruments General Catalog HVAC Testing Indoor Air Quality Industrial Hygiene Cleanroom Contamination Control Industrial Flow Testing Aerosol Research Fluid Mechanics Research Anemometers / Indoor Air Quality Monitors / Gas Monitors / Dust Monitors / Sound Meters / Vibration Meters / Particle Counters Cleanroom Monitoring System / Aerosol Research Instruments / Mass Spectrometer / Automotive Testing Instruments / Fluid Mechanics Research Instruments www.kanomax-usa.com
  • 2. Kanomax Group Since our inception 60 years ago, Kanomax has been the most promising manufacturer of a broad range of precision measuring instruments for fluid mechanics research, environmental, aerosol research, particle measurement, and customized system applications. As a company that prides itself in technology, product quality, and service, we have been enjoying an unsurpassed reputation in the industrial and academic fields. Global Network Our direct subsidiaries and Kanomax’s affiliates and well-trained distributors worldwide are there to provide the most efficient support and service for you. Our global network is always listening to the voice of customers, like you, in order to keep providing the best measurement solutions possible. Kanomax Holdings Inc. (New York, NY) Kanomax USA Inc. (Andover, NJ) Kanomax Corporation (Osaka, Japan) Kanomax Japan Inc. (Osaka-Tokyo-Nagoya, Japan) Kanomax Instrument Shenyang Inc. (China) MSI. TOKYO (Tokyo, Japan) Shenyang Kano Scientific Instrument Co., Ltd (China) ISO Certification Kanomax is an ISO 9001/ISO14001 certified company. Kanomax management and production procedures adhere to these international quality standards. Kanomax Provides Outstanding Solutions Environmental Measurements HVAC Testing Indoor Air Quality Industrial Testing 1 Aerosol/Particle Measurements Cleanroom Contamination Control Aerosol Research Fluid Measurements Fluid Mechanics
  • 3. Calibration facility ensures accuracy and repeatability Kanomax fully understands service is an essential part of the total solution provided to our valued customers. Having already established a worldwide service network, we continuously strive to improve our support services. For reliable measurements it is recommended that all instruments be calibrated on an annual basis. This ensures ongoing credibility and accuracy. ur calibration laboratory O in New Jersey maintains the most accurate wind tunnel of its kind. Kanomax provides the highest quality of service available with a quick turnaround time. Our service specialists are well trained and will calibrate your instruments to the highest standards. Our lab certifies Kanomax product to NIST standards. High Velocity Wind Tunnel Open Jet Wind Tunnel Particle Generator Featured Products 1CFM Flow Rate P ortable Particle Counter Portable Aerosol Mobility Spectrometer (PAMS) Handheld Anemometer Capture Hood Aerosol Monitor 2
  • 4. Handheld Anemometers / Air Velocity Applications HVAC testing and balancing Testing airflow rate of ventilation Handheld Anemometers Specifications Model Sensor Type Air Velocity Ranges ft./min m/s Resolution Accuracy Air Flow CFM (ft3/min) Temperature Ranges Accuracy Relative Humidity Ranges Accuracy Differential Pressure Ranges Accuracy Dimensions Main Unit Probe Weight 3
  • 5. AnemomasterTM AnemomasterTM Professional Standard ClimomasterTM Series Palm-size and feather-weight standard hot-wire Anemometer Telescopic, articulating probe is designed for HVAC testing and balancing applications Multi-function hot-wire Anemometer with detachable compatible probes Rotating Vane Anemometer with High Accuracy from 40 to 7800 fpm ●●Compact and lightweight ●●Simultaneous display of temperature and airflow or air velocity ●●Simultaneously measures and displays air velocity, flow rate, humidity, temperature, and differential pressure Handheld Anemometers / Air Velocity AnemomasterTM Lite ●●2 sizes of vane heads are available for users’ applications ●●Easy replacement of probe without recalibration ●●Data Hold function ●●Includes probe with 59 in (150 cm) cable, extension rod, 4 pcs. AA batteries, carrying case, and NIST Certificate ●●Telescopic, articulating probe ●●Data HOLD function, record and recall MAX/MIN/AVG ●●Store up to 1,500 measurements ●●6 interchangeable probes are available for various applications ●●Smart probe technology: easy probe replacement without recalibration of the main unit ●●Data processing software allows real-time measuring and downloading data to PC ●●Data HOLD function, record and recall MAX/MIN/AVG ●●Complete with telescopic probe with 79 in (200 cm) cable, Data processing software, USB cable, AC adapter, 6 pcs. AA batteries, carrying case, and NIST Certificate Model 6810 Series ●●Industrial grade enclosure and metal vane probe ●●Data HOLD, record and recall MAX/ MIN ●●Complete with metal vane sensor (choice of 2.75” or 1.00” diameter) with 5 ft long cable, extension rod with handle grip, flexible extension rod, 3 pcs. AA batteries, carrying case, and NIST Certificate ●●Data processing software allows real-time measuring and downloading of data to PC ●●Store up to 20,000 measurements ●●Includes probe, 79 in (200 cm) probe cable, 6 pcs. AA batteries, carrying case, and NIST Certificate LITE Professional Standard Climomaster Series 6810 Series Hot-wire Hot-wire Hot-wire Rotating Vane 2 to 3940 2 to 6000 2 to 9480 40 to 7800 0.01 to 20.0 0.01 to 30.0 0.01 to 50.0 0.2 to 40 0.01 m/s 0.01 m/s 0.01 m/s 0.01 m/s +/- 5% of reading or 0.015 m/s whichever is greater +/- 3% of reading or 0.015 m/s whichever is greater +/- 2% of reading or 0.015 m/s whichever is greater +/- 1% of reading n/a 0 to 2,709,360 0 to 2,709,360 0 to 9,999 -4 to 158°F (-20 to 70°C) -4 to 158°F (-20 to 70°C) -4 to 158°F (-20 to 70°C) -4 to 212°F (-20 to 100°C) +/- 1°F (0.5°C) +/- 1°F (0.5°C) +/- 1°F (0.5°C) +/- 0.5°F (0.3°C) n/a n/a 0.2 to 98.0% RH 0.5 to 95.0% RH n/a n/a +/- 2% of reading +/- 2% of reading n/a +/- 5.00 kPa *Option +/- 5.00 kPa *Option n/a n/a +/- (3% of reading +0.01) kPa +/- (3% of reading +0.01) kPa n/a W2.4” x H4.7” x D1.2” W3.4” x H7.4” x D1.6” W3.4” x H7.4” x D1.6” W3.3” x H6.3” x D1.5” 0.24” (6.1 mm) in diameter 0.24” (6.1 mm) in diameter 1.0 to 10 mm in diameter Vane: 2.75” or 1.00” in diameter 0.4 lbs (180 g) 0.9 lbs (400 g) 0.9 lbs (400 g) 0.95 lbs (430 g) 4
  • 6. Anemometers / Air Velocity Meters Climomaster TMModel 6501 Series Multi-function hot-wire anemometer with detachable compatible probes Features: ●●Simultaneously measures and displays air velocity, flow rate, humidity, temperature, and differential pressure ●●Airflow rate calculation based on registered duct size ●●Now equipped with an automatic atmospheric pressure compensation function for precise airflow measurement ●●Store up to 20,000 measurements ●●8 interchangeable probes are available for various applications ●●Smart probe technology: easy probe replacement without recalibration of the main unit ●●Includes probe, 79 in (200cm) probe connection cable, 6 pcs. AA batteries, carrying case, and NIST Certificate Main Unit Specifications Air Velocity Ranges 2 to 9840 fpm (0.01 to 50.0 m/s) *Varies by probe Accuracy +/- 2% of reading or 0.015 m/s whichever is greater Temperature Ranges -4 to 158°F (-20 to 70°C) Accuracy Relative Humidity Ranges +/- 1.0°F (0.5°C) 2.0 to 98.0%RH *Varies by probe Accuracy +/- 2.0%RH Differential Pressure Ranges (Option) +/- 5.00 kPa Accuracy Interface +/- (3% of reading + 0.01) kPa USB / RS232C (for print-out) Datalogging Up to 20,000 records Analog Output (Option) 0 to 1 V Power Supply 6 x AA Batteries or AC Adapter Dimensions Optional data processing software allows real-time measuring and downloading data to your PC W3.4” x H7.4” x D1.6” (88 x 188 x 41 mm) Weight 0.9 lbs (400 g) Probe Specifications Model 6531-2G 6541-2G 6561-2G 6542-2G 6533-2G 6543-2G 6551-2G 6552-2G Probe Type Uni-Directional Uni-Directional Uni-Directional Omni-Directional Omni-Directional Omni-Directional Mini-Spherical Mini-Spherical Air Velocity 2 to 6000 fpm 2 to 6000 fpm 2 to 9840 fpm 2 to 6000 fpm 2 to 1000 fpm 2 to 1000 fpm 2 to 6000 fpm 2 to 6000 fpm Temperature -4 to 158°F -4 to 158°F -4 to 158°F -4 to 158°F -4 to 158°F -4 to 158°F n/a n/a 2.0 to 98.0 %RH n/a n/a n/a 2.0 to 98.0 %RH n/a n/a n/a Relative Humidity Accessories 6501-AE 6501-BE 6531-04: 6531-05: 6531-06: 6000-41: 6000-31: 6000-61: 6113-02: DPU-S245: TP-202L: 5 Main Unit with Analog Output Main Unit with Pressure Sensor Telescopic Extension Rod (Flex-Neck) Telescopic Extension Rod (Straight) 2m Probe Cable (also available in 5, 10, 20m) Data Processing Software (for Windows) Printer Cable for DPU-S245 Hands Free Case A C Adapter Portable Thermal Printer Rolled Printer Paper (10 rolls) 6531-04 6531-05 6000-61
  • 7. Anemomaster TMProfessional Standard Multi-function hot-wire Anemometer with telescopic, articulating probe Features: ●●Simultaneous display of temperature and airflow or air velocity ●●Telescopic probe measures are velocity and temperature in air ducts, vents, and small openings ●●Data HOLD function, Record and recall MAX/MIN/AVG ●●Store up to 1,500 measurements (Professional only) ●●Data processing software allows real-time measuring and downloading data to your PC (professional only) ●●Includes telescopic probe with 79 in (200 cm) cable, Data processing software (Professional only), USB cable (Professional only), AC adapter, 6 pcs. AA batteries, carrying case and NIST Certificate Telescopic probe is designed for HVAC testing and balancing applications Applications: Articulating Tip ●●HVAC Testing (Professional only) ●●Facility Maintenace ●●Critical Environment Certification ●●IAQ Investigation Specifications Model AnemomasterTM Professional Model 6036 Telescopic and Articulating tip Probe Type AnemomasterTM Standard Model 6035 Telescopic Air Velocity Ranges Accuracy 2 to 6000 fpm (0.01 to 30.0 m/s) +/- 3% of reading or 0.015 m/s whichever is greater Temperature Ranges -4 to 158°F (-20 to 70°C) Accuracy +/- 1.0°F (0.5°) Differential Pressure Ranges +/- 5.00 kPa *Option Accuracy Interface - Digital Analog (Option) Datalogging +/- 3% of reading +0.01 kPa USB / (RS232C for print-out) (RS232C for print-out) DC 0 to 3 V (only for air velocity measurements) Up to 1500 records Power Supply None Windows 7 compatible software included (professional only) 6 x AA batteris or AC Adapter Dimensions W3.4” x H7.4” x D1.6” Weight 0.9 lbs (400 g) Theory of thermal (hot-wire) type AnemomasterTM Accessories 6036-AE: 6036-BE: 6035-AE: 6035-BE: 6000-31: 6 000-61: DPU-S245: T P-202L: Professional with Analog Output Professional with Pressure Sensor S tandard with Analog Output Standard with Pressure Sensor Printer Cable for DPU-S245 H ands Free Case Portable Thermal Printer Rolled Printer Paper (10 rolls) The air velocity sensor is heated and temperature elevated (relative to the surrounding air) by means of control electronics. The temperature compensation sensor senses the ambient, or surrounding air temperature, and forces the velocity sensor to stay at a constant overheat above the ambient. The circuit forces the voltage to be equal by means of an operational amplifier. Air flowing past Air velocity sensor the sensor tends to cool the sensor, thus driving down its resistance. The amplifier responds by immediately delivering more power to the circuit to maintain voltage equilibrium. Delivered power is converted into electrical signals to display. Temperature compensation sensor 6
  • 8. Anemometers / Air Velocity Meters Anemomaster TMLITE Palm-size and feather-weight standard hot-wire Anemometer Features: ●●Compact and lightweight ●●Display switchable in m/s or ft/min (FPM) for air velocity and °F and °C for air temperature ●●Easy replacement of probe without recalibration ●●Average measurements over 1 or 5 seconds for air velocity ●●Data Hold function ●●4 pcs. AA batteries, extension rod, carrying case and NIST Certificate Specifications Model 6006 Air Velocity Ranges Accuracy 2 to 3940 fpm (0.01 to 20.0 m/s) +/- 5% of reading or 0.015 m/s whichever is greater Temperature Ranges -4 to 158°F (-20 to 70°C) Accuracy Power Supply Dimensions +/- 1.0°F (0.5°C) Accessories 4 x AA Batteries or AC Adapter 6006-2G: Replacement Probe 6112-03: Spare Extension Rod 6006-02: Spare Hard Carrying Case W2.4” x H7.4” x D1.2” Weight 0.4 lbs (180 g) Anemomaster TMModel 6113/6114 High velocity (up to 50 m/s) multi-function hot-wire Anemometer Features: ●●Simultaneous measurements of air velocity, air temperature and differential pressure ●●Large, easy to read LCD display ●●Durable body ●●Model 6113 includes built-in printer ●●Complete with probe with 79 in (200 cm) cable, extension rod, shoulder strap, and 6 pcs. C cell batteries, and NIST Certificate Specifications Model Air Velocity Ranges Accuracy Temperature Ranges Accuracy Differential Pressure Ranges Accuracy Interface Datalogging Analog Output Built-in Printer Power Supply Dimensions Weight 6113 6114 20 to 9840 fpm (0.10 to 50.0 m/s) +/- 3% of reading +0.1 m/s 32 to 212°F (0 to 100°C) +/- 2.0°F (1.0°C) +/- 5.00 kPa *Option +/- 3% of reading +0.01 kPa RS232C 100 measurements 0 to 1 V *Option ⃝ ― 6 x C cell Batteries W7.9” x H5.9” x D3.9” 2.2 lbs (1 kg) Accessories 6113-01: 6113-02: 6113-03: 6113-04: 6113-07: 6113-08: 7 Spare Probe AC Adapter Extension Rod Shoulder Strap Pressure Sensor Analog Output 6000-41: Data Processing Software 6000-02: Communication Cable to PC TP-202L: Rolled Printer Paper (10 rolls)
  • 9. Anemometers / Air Velocity Meters Anemomaster TMVane Model 6810 Series Rotating Vane Digital Anemometer Features: ●●High Accuracy from 40 to 7800 feet per minute ●●2 sizes of vane heads are available for users’ applications ●●Industrial grade enclosure and metal vane probe ●●Large display with backlight ●●Data HOLD, Record and recall MAX/MIN ●●Includes metal vane sensor (choice of 2.75” or 1.00” diameter) with 5 ft long cable, extension rod with handle grip, flexible extension rod, 3 pcs. AA batteries, carrying case, and NIST-traceable calibration certificate 1 inch Air Velocity Probe Specifications Model 6812 Air Velocity Ranges 6815 comes with HTP202 6815 6813 2.75” Head 1.00” Head 2.75” Head 1.00” Head 2.75” Head 1.00” Head ft/min 40 to 7800 300 to 6800 40 to 7800 300 to 6800 40 to 7800 300 to 6800 m/s 0.20 to 40.0 1.5 to 35.0 0.20 to 40.0 1.5 to 35.0 0.20 to 40.0 1.5 to 35.0 Resolution 1 FPM or 0.01 m/s Air Flow CFM (ft3/min) 1 FPM or 0.01 m/s 1 FPM or 0.01 m/s 0 to 9999 n/a n/a n/a Temperature Ranges -4 to 212°F (-20 to 100°C) -4 to 176°F (-20 to 80°C) *with HTP202 Accuracy n/a +/- (0.3°C +0.2% of reading in °C) +/- (0.3°C +0.2% of reading in °C) Relative Humidity Ranges n/a n/a 5.0 to 95.0% RH Accuracy n/a n/a +/- 2.0% RH 3 x AA Batteries 3 x AA Batteries 3 x AA Batteries W3.2” x H6.5” x D1.5” W3.2” x H6.5” x D1.5” W3.2” x H6.5” x D1.5” 0.95 lbs (430 g) 0.95 lbs (430 g) 1.1 lbs (500 g) Power Supply Main Unit Dimensions Weight Humidity Temp. Probe Air Velocity Probes Model Air Velocity ft/min m/s Accuracy Temperature °F °C Accuracy AP275 n/a n/a n/a APT275 2.75” Head 40 to 7800 0.20 to 40.0 +/- (1.0% reading + 1 digit) -4 to 212 -20 to 100 +/- (0.3°C +0.2% of reading in °C) AP100 APT100 1.00” Head 300 to 6890 1.5 to 35.0 +/- 0.50% FS + 0.75% reading + 1 digit n/a n/a n/a -4 to 212 -20 to 100 +/- (0.3°C +0.2% of reading in °C) Model Relative Humidity Range Resolution Accuracy Temperature °F °C Accuracy HTP202 5.0 to 95.0% RH 0.1% RH +/- 2.0% RH -4 to 176 -20 to 80 +/- (0.3°C +0.2% of reading in °C) Vane Air Velocity Transmitters Features: ●●Transmitters specially designed for monitoring the Indoor Air Quality in the air conditioning, heating, and ventilation industries. ●●Three versions available: AT400: Air Velocity Transmitter TAT420: Air Velocity Temperature Transmitter THT500: Temperature-Humidity Transmitter ●●Three separate power supplies 110 VAC 50/60 Hz, 220 VAC 50/60 Hz, and 10-30 VDC ●●Three concurrent outputs for each measurement: 0-1 VDC, 0-5 VDC, and 4-20 mA DC ●●Temperature Range -4 to 212°F (-20 to 100°C) 8
  • 10. TABmaster TMCapture Hood A solution for airflow testing and balancing Airflow Capture Hood The new Kanomax TABmasterTM is the perfect tool for accurate supply and return airflow measurements. Interchangeable hoods make it easy to sample the air for any duct size. The unit is lightweight and easy to handle. The full color screen can be tilted so it’s always at the optimal viewing angle regardless of height. Features: ●●23 to 2500 cfm (40 to 4250 m3/h) range ●●Simultaneously measures and displays air flow, temperature and humidity ●●Displays the direction of the airflow ●●Store up to 8,000 measurements ●●Advanced storage feature allows you to store multiple measurements under a single ID# ●●Built-in back pressure compensation ensures accuracy for large volumetric flow measurements ●●Removable handheld micromanometer with Bluetooth® wireless capability (Model 6715) ●●Includes: standard hood, carrying case, AA batteries, PC communication cable, data processing software, user manual, and calibration certificate Applications: ●●HVAC testing, adjusting and balancing ●●Air volumetric flow measurements through registers, diffusers and grilles ●●Direct readout at supply and return airflow ●●Air velocity measurement in the duct ●●Check filter fouling by measuring differential pressure Specifications 6710 Model Airflow Range 6715 23 to 2500 CFM (40 to 4250 m3/h) Accuracy +/- 3% of reading +/- 10 m3/h Resolution Air Velocity Range Accuracy Resolution 1m3/h n/a 0.15 to 40 m/s (Pitot), 0.15 to 15 m/s (Velocity Matrix) n/a ±3%±0.05m/s of readings n/a 0.1 m/s (10.0m/s) 0.01m/s (9.99m/s) Temperature Range Portable stand extends up to 6.5’ from top to base 32 to 122°F (0 to 50°C) Accuracy +/- 1.0°F (0.5°C) Resolution 0.1°C Humidity Range 0 to 100% RH Accuracy +/- 3.0% RH Resolution 0.1% RH Interface USB Datalogging Up to 3000 measurements Power Supply Hood Dimensions Up to 8000 measurements AA batteries or AC adapter 2x2 ft (610x610 mm), 1x4 ft (305x1220 mm), 2x4 ft (610 x 1220 mm) 3x2 ft (915 x 610 mm), 3x3 ft (915x915 mm), 500x500mm Weight 7.9 lbs (3.6 kg) Accessories 6 710-01: 6710-02: 6710-03: 6710-04: 9 Spare Hood 2x2 ft (610x610mm) Spare Hood 2x4 ft (610x1220mm) Spare Hood 1x4 ft (305x1220mm) Spare Hood 3x2 ft (915x610mm) 6 710-05: 6710-06: 6710-07: 6710-08: Spare Hood 3x3 ft (915x915mm) Spare Hood 500x500mm Spare Hood Support Poles Capture Hood Stand The Bluetooth® feature can send data to any Android-based device
  • 11. Micromanometer DP500 Smart micrometer for TAB professionals Pressure Measurement Handheld micromanometer with Bluetooth® wireless capability takes airflow and pressure readings with a pitot tube or the optional velocity matrix. The matrix is a cross shaped grid, similar to the one in the base of the capture hood, that makes it possible to take face velocity readings and average velocity readings at multiple points simultaneously. The Bluetooth® feature can send data to your smartphone or any Android-based device. Pitot tubes and Velocity Matrix are available from Kanomax Specifications Model DP500 0.15 to 40 m/s (with Pitot tube) 0.15 to 15 m/s (with Velocity Matrix) Air Velocity Ranges Accuracy ±3%±0.05m/s of readings Accuracy ±0.25% ±1Pa of readings Pressure Ranges -2500 to 2500 Pa Temperature Ranges 0 to 50°C Accuracy +/- 1% of reading +1°C Relative Humidity Ranges 0 to 100%RH Accuracy ±3%RH(10~90%RH) Interface USB and RS232C (for printer) Datalogging 8000 measurements Power Supply 4 x AA Batteries or AC Adapter Pitot Static Tube Stainless Steel Construction Pitot Tubes The unique ellipsoidal nose form results in exceptional accuracy, insensitivity to errors in alignment and elimination of the need for calibration curves. Constructed from stainless steel, tubes can be as small as 2.3 mm in diameter yet scaled up to 4 m in length. Features: ●●Totally compatible with manometers, pressure gauges and pressure transmitters ●●Ideal for measurement of air velocity in hostile environments ●●Integral sensing Thermo pitot tube thermocouple for combined pressure and temperature ●●Can be used in permanent position when fitted with gland ●●S-Type for limited access and adverse conditions S-Type pitot tube Specifications Easy Fit Ellipsoidal Pitot Tube Overall Length (m) Classic Ellipsoidal Pitot Tube 0.2 0.3 0.3 0.5 0.5 Tube Diameter (mm) 4 4 4 8 8 8 Head Diameter (mm) 2.3 4 4 8 8 8 Gland (inch) 1/4” 3/8” 0.75 0.75 1.0 1.0 8 8 8 8 3/8” Pitot with K-Type Thermocouple Overall Length (m) Overall Length (m) 1.83 2.13 2.44 2.74 1.0 1.22 1.52 1.83 8 9.5 9.5 Tube Diameter (mm) 12.7 19 19 19 8 9.5 9.5 12.7 8 9.5 9.5 Head Diameter (mm) 9.5 9.5 9.5 9.5 8 9.5 9.5 9.5 1/2” Gland (inch) 3/8” 1/2” 1/2” 1/2” 3/8” S-Type Pitot Tube Overall Length (m) 1.25 1.25 0.3 1.0 Tube Diameter (mm) 4 8 Head Diameter (mm) 4 0.7 1.2 2.2 3.2 8 10
  • 12. Flow Grids For Accurate In-Duct Volume Flow Measurements Airflow Measurement Kanomax flow grids produce averaged total and sub-static pressures signals to generate an enhanced velocity pressure signal, outside the duct wall, that directly relates to the volume flow, providing accurate and reliable outputs where other flow measuring devices are found to be unsatisfactory. Works directly with Micro-manometers, pressure transducers, Magnahelic gauges, and pressure activated alarm switches. These are ideal for ventilation monitoring for LEEDS and many other facility management monitoring applications. Features: ●●Sizes from 4.0” (100mm) diameter to 10 ft.(3.0 meters) square ●●Uncertainty of flow measurement is within +/- 1.0% ●●High temperature systems up to 850°C are available ●●Stainless steel grade 321 tube supplied as standard, other grades of steel or alloys are available upon request Standard Sizes Rectangular Grids Radial Grids Type A 6.4mm (1/4”) d iameter tube Manifold length (mm) 100 to 450 Pressure Tube length (mm) 150 to 450 Type B 12.7mm (1/2”) diameter tube Manifold length (mm) 250 to 650 Pressure Tube length (mm) 350 to 1200 Type C 25.4mm (1”) d iameter tube Manifold length (mm) 750 to 1000 Pressure Tube length (mm) 800 to 2000 Type A 6.4mm (1/4”) d iameter tube Diameter (mm) 100 to 500 Type B 12.7mm (1/2”) diameter tube Diameter (mm) 500 to 1100 Type C 25.4mm (1”) d iameter tube Diameter (mm) 1100 to 2500 X Flow Grids For Accurate In-Duct Volume Flow Measurements X flowgrids will give useful and reliable readings in a wide variety of in-duct locations often where other flow rate measuring devices are found to be unsatisfactory or difficult to install. The low cost design consists of two tubes mounted diagonally through the cross section of the duct. They generate an averaged signal for both the total and substatic (X) or static (UX) pressure along the length of each tube to generate a differential pressure signal that directly relates to volume flow. Features: ●●Economical method of measuring volume flow within a duct ●●Uncertainty of volume or mass flow measurement within +/- 5.0% ●●Stainless steel grade 321 tube supplied as standard Specifications Model Air Velocity Ranges UX8 Maximum Duct Diagonal Maximum Dust Diameter UX version: 1.0 correlation ; X version: 2.0 amplified correlation Diameter of Tubes Weight UX16 +/- 5% Maximum Temperature ●●Differential pressure correlation/correction factor: MX16 295.2 to 5904 ft/min (1.5 to 30 m/s) Accuracy ●●High temperature systems up to 850°C (continuous) available 11 MX8 176°F (80°C) 5/16” (8mm) 5/8” (16mm) 28.1“ (715mm) 26.7” (680mm) 60.4 “(1534mm) 59.4” (1510mm) 1 lbs (454 g) 3 lbs (1361 g)
  • 13. High Temperature Velocity Meter Anemomaster TMModel 6162 High Temperature Anemometer Features: ●●Air velocity and temperature measurements in 932°F (500°C) environments ●●Record and recall MAX/MIN/AVG, Timing graph display ●●Store up to 999 measurements ●●RS232C interface, analog output, and remote control terminal equipped ●●Includes shoulder strap, AC adapter, 2 pcs. analog output cable, and 6 pcs. C cell batteries High temperature measuring solution requires Model 6162 with one of the high temperature probes (0203, 0204 or 0205) ●●The Model 0203 includes probe with 4.9 ft (1.5 m) cable, 5m probe connection cable, probe case, and NIST Certificate ●●The Model 0204 includes probe with 7.6 ft (2.3 m) cable, 10m probe connection cable, probe case, and NIST Certificate ●●The Model 0205 includes probe with 7.6 ft (2.3 m) cable, 10m probe connection cable, probe case, and NIST Certificate Main Unit Specifications Model 6162 Air Velocity Ranges Accuracy Temperature Ranges Accuracy Varies by Probe, see below for Probe Specifications +/- 3% of Full Scale Varies by Probe, see below for Probe Specifications +/- 1% of reading +1°C Interface RS232C Datalogging 999 measurements Analog Output Remote Terminal Probe Specifications Model 0203 0204/0205 Temp. Range Air Velocity Measuring Range 32 to 212°F 40 to 9840 fpm (0.2 to 50.0 m/s) 212 to 392°F n/a 138 to 9840 fpm (0.7 to 50.0 m/s) 572 to 752°F n/a 197 to 9840 fpm (1.0 to 50.0 m/s) Accuracy +/- 3% Full Scale 32 to 392°F (0 to 200°C) Accuracy Dimensions / Weight 6 x C cell Batteries or AC Adapter Dimensions W8.7” x H3.3” x D5.9” Weight 4.0 lbs (1.8 kg) 80 to 9840 fpm (0.4 to 50.0 m/s) 392 to 572°F Temperate Measuring Range Power Supply 0 to 1 V START/STOP Key 32 to 932°F (0 to 500°C) +/- 1% of reading ø 11 x 200 mm (ø 0.4” x 7.8”) 0.4 lbs (200 g) Probe Cables 0204 ø 14 x 1000 mm (ø 0.6” x 39.4”) 1.1 lbs (500 g) 0205 ø 14 x 500 mm (ø 0.6” x 19.7”) 0.4 lbs (200 g) Teflon Coating Heat-resistance 392°F (200°C) Extension Cable Vinyl Coating Heat-resistance 176°F (80°C) Accessories 0203: 0204: 0205: 6162-03: 6162-04: 6162-05: 6162-06: 6162-07: 6000-41: 6511-09: DPU-S245: TP-202L: Probe for Middle Temperature Probe for High Temperature (Long) Probe for High Temperature (Short) Extension Rod for 0203 Extension Rod for 0204/0205 Probe Compression fitting for 0203 Probe Compression fitting for 0204/0205 Communication Cable to PC Data processing Software Printer Cable for DPU-S245 Portable Thermal Printer Rolled Printer paper (10 rolls) *Optional probe cable lengths up to 40m are available Probe Dimensions *Length in mm MODEL 0203 MODEL 0204 Extension Rod for 0203 * Option Extension Rod for 0204 * Option Entire Length Entire Length Probe Cable MODEL 0205 12
  • 14. Multi-Channel Anemometers Multi-Channel Measuring Instruments Main Unit Specifications Model Probe Compatible Air Velocity Ranges Temperature Ranges Relative Humidity Ranges PC Communication Variety of probes for many applications Interface Analog Output Power Supply Dimensions Weight 13
  • 15. Multi-Channel Anemometers Airflow Transducer Model 6332 / 6332D 4-Channel AnemomasterTM Model 1570 Multi-Channel AnemomasterTM Model 1550 / 1560 Compact design Airflow Transducer, fits into small spaces Compact design 4-channel unit with compatible probes Up to 320 points multi-measurements system with a variety of probes ●●10 interchangeable probes are available for various applications ●●10 interchangeable probes are available for various multi-channel applications ●●System can be scaled up with modules and probes ●●Easy probe replacement without recalibration of the main unit ●●Simultaneously measurements of 4 channels of air velocity ●●Selectable output options: 0 to 5 V or 4 to 20 mA ●●Software allows real-time measurements of air velocity and airflow in 4 channels ●●The chassis may be cascaded up to 5 units via RS232C ●●The Model 1570 includes data processing software, RS232C cable, power cable, and 2 pcs. fuse ●●Multi-channel Anemomaster includes RS232C cable, printer cable, power cable, and 2 pcs. fuse ●●3 types of probes are available ●●4 types of modules are available 6332/6332D 1570 1550 / 1560 V Probe V Probe V Probe / VT Probe / VTH Probe 20 to 9840 fpm (0.10 to 50.0 m/s) 20 to 9840 fpm (0.10 to 50.0 m/s) 20 to 9840 fpm (0.10 to 50.0 m/s) n/a n/a 32 to 212°F (0 to 100°C) n/a n/a 5.0 to 95.0% RH n/a Data Processing Software Data Processing Software n/a RS232C RS232C, Centronics, GP-IB DC 4 to 20mA or DC 0 to 5V 0 to 5 V 0 to 5 V DC or AC AC AC W3.1” x H5.0” x D1.2” W10.2” x H2.8” x D7.9” Model 1550: W19.6” x H5.5” x D16.9” Model 1560: W8.9” x H5.5” x D12.8” 0.7 lbs (320 g) 5.7 lbs (2.6 kg) Model 1550: 22 lbs (10 kg) Model 1560: 11 lbs (5 kg) 14
  • 16. Airflow Transducer Model 6332 / 6332D Features: Multi-Channel Anemometers ●●Smart probe technology: easy probe replacement without recalibration of the main unit ●●Selectable output options: 0 to 5 V or 4 to 20 mA Main Unit Specifications Model Display Air Velocity Ranges Accuracy Analog Output 6332 6332D ― ⃝ Varies by probe, See below for Probe Specifications +/- 3% of reading DC 4 to 20mA or DC 0 to 5V Power Consumption Approx. 2.0 W Power Supply Dimensions Weight DC 12 to 24V or AC 80 to 240V W3.1” x H5.0” x D1.2” 0.7 lbs (320 g) 6332D 6332 4-Channel Anemomaster TMModel 1570 Features: ●●Simultaneous measurements of 4 channels of air velocity; easy to switch over each channel display ●●Software allows real-time measurements of air velocity and airflow in 4 channels ●●The Model 1570 includes data processing software, RS232C cable, power cable, and 2 pc. fuse Main Unit Specifications Model Air Velocity Ranges Resolution Interface Analog Output Power Supply Dimensions Weight 1570 Varies by probe, See below for Probe Specifications 0.01 m/s RS232C 0 to 5V AC 85 to 265V W10.2” x H2.8” x D7.9” 5.7 lbs (2.6 kg) Compatible Probes Air Velocity Proves (V Probes) 0965-00 is with Horn Probe Specifications Model Probe Type Air Velocity 0962-00 Uni-Directional 20 to 9840 fpm (0.10 to 50.0 m/s) 0963-00 Uni-Directional 20 to 9840 fpm (0.10 to 50.0 m/s) 0965-00/-01 Omni-Directional 20 to 4920 fpm (0.10 to 25.0 m/s) Miniature Probe with built-in temperature compensation 0965-09 is 80mm long 0965-10 is 400mm long 0965-09/10 Omni-Directional 20 to 9840 fpm (0.10 to 50.0 m/s) Miniature Probe with independent temperature compensation Air Velocity Sensor Air Velocity Sensor Probe Specifications Model Probe Type Air Velocity 15 Temperature Compensation Sensor 0965-03 Omni-Directional 20 to 4920 fpm (0.10 to 25.0 m/s) Temperature Compensation Sensor 0965-04 Omni-Directional 20 to 4920 fpm (0.10 to 25.0 m/s) 0965-07 Omni-Directional 20 to 4920 fpm (0.10 to 25.0 m/s) 0965-08 Omni-Directional 20 to 4920 fpm (0.10 to 25.0 m/s)
  • 17. Multi-Channel Anemomaster TMModel 1550/1560 ●●The multi-channel Anemomaster is composed of chassis, module, and probe. The Model 1550 (chassis) holds 16 modules and the Model 1560 holds 6 modules ●●The chassis may be cascaded up to 5 units (5 units of Model 1550 can hold 320 channels of air velocity sensors) 1550 ●●3 types of probes are available Air velocity probe (V probe) Air velocity / Temperature probe (VT probe) And Air Velocity / Temperature / Humidity probe (VTH probe) ●●4 types of modules are available 1560 4 channel air velocity module 2 channel air velocity / temperature module 1 channel air velocity / temperature / humidity module Analog output module Main Unit Specifications ●●Multi-channel Anemomaster includes RS232C cable, printer cable, Model power cable, and 2 pcs. fuse 1550 Air Velocity Ranges 1560 Varies by Probe, See below for Probe Specifications Resolution 0.01 m/s Temperature Ranges Varies by Probe, See below for Probe Specifications Resolution Accessories 1500-01: 1500-02: 1500-03: 1504-04: 1511-01: 1512-01: S620-00: Multi-Channel Anemometers Features: 0.1°C Relative Humidity Ranges RS232C for Cascade Connection GP-IB Output Ring Guards for Probe Protection Cable for V module (10m)* Cable for VT Module (10m)* Cable for VTH Module (10m)* Data Processing Software (for Windows) Varies by Probe, See below for Probe Specifications Resolution 0.1% RH RS232C for PC Connection RS232C for Cascade *Option Interface Centronics for Printer Output GP-IB *Option Analog Output 0 to 5 V *Option with D/A Module Power Supply *Additional cable lengths available AC Adapter Dimensions W19.6” x H5.5” x D16.9” W8.9” x H5.5” x D12.8” 22 lbs (10 kg) 11 lbs (5 kg) Weight Air Velocity Temperature Humidity Proves (VT / VTH Probes) Probe Specifications Model Probe Type Air Velocity Temperature Ranges Relative Humidity Ranges 0962-21 0963-21 0965-21 0963-31 0965-31 Uni-Directional Uni-Directional Omni-Directional Uni-Directional Omni-Directional 20 to 9840 fpm (0.10 to 50.0 m/s) 20 to 9840 fpm (0.10 to 50.0 m/s) 20 to 9840 fpm (0.10 to 50.0 m/s) 20 to 9840 fpm (0.10 to 50.0 m/s) 20 to 9840 fpm (0.10 to 50.0 m/s) 32 to 212°F (0 to 100°C) n/a 32 to 212°F (0 to 100°C) n/a 32 to 212°F (0 to 100°C) n/a 32 to 212°F (0 to 100°C) 5.0 to 95.0% RH 32 to 212°F (0 to 100°C) 5.0 to 95.0% RH Modules Specifications Model Module Type # of Channels 1504 V Module 1511 VT Module 1512 VTH Module 1510 D/A Module Air Velocity Air Velocity, Temp. Air Velocity, Temp., Humidity Analog Output 4 2 1 Model 1560 has 6 slots for modules 1 16
  • 18. Applications Indoor Ait Quality Monitors Hospitals and Elderly Care Facilities Monitor Green building rating system IEQ performance testing Indoor Air Quality Monitors Ventilation Testing Thermal Comfort Control Occupational Health Control Monitoring Toxic Gas 17 Indoor Air Quality Investigation
  • 19. Handheld IAQ Monitor Model 2211 Multi-Function Indoor Air Quality Monitor Indoor Ait Quality Monitors Features: ●●Simultaneous measurements of CO, CO2, Temperature, and Relative humidity ●●Calculates Dew point, Wet bulb temperature, Absolute humidity, Humidity Ratio, and % Outside Air ●●Store up to 1500 measurements ●●Easy user self calibration ●●Easy replacement of probe ●●PC interface with RS232C or USB and software for real-time measurements and downloading data to your PC ●●Complete with probe with 79 in (2m) cable, probe stand, calibration cap connection tube, data processing software, RS232C cable, USB to serial adapter, 6 pcs. AA batteries, carrying case, and NIST-Traceable calibration certificate Specifications Model 2211 Carbon Monoxide (CO) Accuracy Carbon Dioxide (CO2) Accuracy Temperature Range Accuracy Relative Humidity Range Accuracy Interface Datalogging 0 to 500 ppm +/- 3% of reading 0 to 5000 ppm +/- 3% of reading -4 to 140 F (-20 to 60 C) +/- 1.0°F (0.5°C) 2.0 to 98.0 %RH +/- 2% of reading 1500 measurements Analog Output 0 to 1 V *Option Power Supply 6 x AA Batteries or AC Adapter Dimensions Weight Software Included RS232C W3.4” x H7.4” x D1.6” 0.9 lbs (400 g) Accessories 6113-02: 2211-09: DPU-S245: 6 000-03: T P-202L: AC Adapter Analog Output Portable Thermal Printer P rinter Cable for DPU-S245 Rolled Printer Paper (10 rolls) Indoor Comfort Comfort is very subjective and will vary from one individual to the next, but ASHRAE Standard 55 provides guidelines for a “comfort zone” that most people will feel comfortable in. This comfort zone is derived from both temperature and humidity and varies depending on the season. See the chart to the right for details. The 2211 IAQ Monitor from Kanomax can measure both parameters simultaneously indicating if changes are needed to make a building’s occupants comfortable. The monitor can also check CO2 levels and determine the rate of exchange, or the percentage of outside air, being introduced to a building per ASHRAE Standard 62. Controlling the rate of exchange with the HVAC system is critical to ensuring occupant well-being. Too much CO2 build-up can cause lethargy and make it difficult for individuals to concentrate. Too much outside air may also be harmful if it’s introducing external pollutants to the environment. The 2211 is a great tool for IAQ investigation and spot checks to ensure an indoor environment remains within desired levels to maintain occupant health and comfort. The ability to measure all of these parameters in a single instrument simplifies the labor involved and eliminates the need to purchase and learn multiple instruments. 18
  • 20. Handheld Gas Monitors CO, CO2, Ammonia, Ozone, VOC etc. More than 30 gases Features: ●●Simultaneous measurements of gas concentration, temperature, and humidity ●●The S200 is simple, easy-to-use, low-cost monitor ●●The S300 has on-board alarm and analog output Gas Monitors ●●The S500 has data logging function and USB interface for PC communication ●●The Remote Sensor Head Adaptor Kit is available; it is designed to allow the sensor head to be located up to 46 ft (14 m) away from the monitor Specifications Model Measurement Units T/H* Sensor On-board alarm Remote Sensor Datalogging Interface Analog Output Power supply Dimensions Weight S200 S300 S500 ppm or mg/m3 ppm or mg/m3 ppm or mg/m3 Option * Option * Option * ― ⃝ ― ― ― ⃝ ⃝ ― ― ⃝ ⃝ 4,300 measurements RS232C 0 to 5 V 0 to 5 V 9.6V Battery / 12 VDC 9.6V Battery / 12 VDC 9.6V Battery / 12 VDC W7.4” x H4.8” x D2.1” W7.4” x H4.8” x D2.1” W7.4” x H4.8” x D2.1” 1.0 lbs (460 g) 1.0 lbs (460 g) 1.0 lbs (460 g) * Temperature and Humidity Features: Gas Sensor Heads ●●The gas monitor is fully compatible with all gas sensors Gas Sensor Specifications Gas Sensor Heads Range (ppm) Accuracy Resolution Ammonia 0 - 100 +/- 5 ppm 0.1 ppm Ammonia (leak) 0 - 1000 +/- 15% 1 ppm Carbon monoxide 0 - 100 +/- 5 ppm 0.1 ppm Carbon monoxide 0 - 1000 +/- 10% 1 ppm Carbon dioxide 0 - 2000 +/- 40 ppm + 3% 10 ppm Carbon dioxide 0 - 5000 +/- 150 ppm + 5% 10 ppm Hydrogen 0 - 5000 +/- 10 % 1 ppm 0 - 10 +/- 0.5 ppm 0.01 ppm Hydrogen sulphide Methane 0 - 9999 +/- 15 % 0 - 0.150 +/- 0.005 ppm 0 - 0.5 +/- 0.008 ppm 0 - 1.0 +/- 0.001 ppm 0.001 ppm NMHC 3,4 0 - 25 +/- 10 % 0.1 ppm Perchloroethylene 0 - 200 +/- 5 ppm 1 ppm 0 -10 +/- 0.5 ppm 0.01 ppm Formaldehyde 0 -10 +/- 0.05 ppm 0.01 ppm VOC 0 - 25 +/- 10 % 0.1 ppm VOC 0 - 500 +/- 10 % 1 ppm VOC PID 0 - 20 +/- 10 % 0.01 ppm VOC PID 0 - 1000 +/- 10 % 0.1 ppm Sulphur dioxide Multi-Gas Sensor Heads Sensor Range (ppm) Accuracy Resolution MS1 CO2 CO 0 - 2000 0 - 100 +/- 40 ppm +3% +/- 10% 1 ppm 0.1 ppm MS2 CO2 CO VOC (PID) 0 - 2000 0 - 100 0 - 25 +/- 40 ppm +3% +/- 10% +/- 10% 1 ppm 0.1 ppm 0.01 ppm * Other specific concentrations available on request 19 without recalibration of main unit 0.001 ppm Nitrogen dioxide ●●Sensor heads for handheld units are interchangeable 0.001 ppm Ozone application 1 ppm Ozone ●●Multi-gas sensor heads are available for IAQ survey Accessories R01: R10: R13: R 31: R32: R 33: R 40: R 52: Temperature/Relative Humidity option Remote Sensor Adapter with Cable Remote Sensor Adapter with Cable, Enclosure Replacement Battery C igarette Lighter Adapter Battery Charger Wall Mount Bracket Carrying Case R S232 to USB Converter
  • 21. Networked Gas Monitoring Systems Fixed Transmitters Controllers Features: ●●The gas sensors are able to measure concentrations of Ozone, Volatile Organic Compounds, Ammonia, Carbon Monoxide sensor head has its own identification (ID), which allows a “Daisy-Chained” network to be created utilizing either a computer system or Programmable Logic Controller (PLC) ●●Networks created Gas Monitors ●●Each of up to 256 individual monitors/transmitters can be PLC Specifications Specifications Model Measuring Ranges T/H* Sensor Alarm Output Analog Output Interface Power Supply Enclosure Size Enclosure Casing Weight S905 S935 S945 See Gas Sensor Specifications See Gas Sensor Specifications See Gas Sensor Specifications Option * Option * Option * ⃝ ⃝ ⃝ 4 - 20 mA 4 - 20 mA and 12 to 24 V 4 - 20 mA and 12 to 24 V RS485 RS485 RS485 24 VDC 24 VDC 24 VDC 2.5” x Ø5.1” W7.1” X H4.3” X D3.5” W9.1” X H5.5” X D3.7” Fibre reiforced plastic Fibre reiforced polycarbonate Fibre reiforced polycarbonate 850 g 850 g 1.1 kg Model Sensor Channels Digital Alarm Inputs Alarm Control Relays Analog Inputs Analog Control Outputs Expansion I/O Data-storage Communication Power Supply Dimensions R1000 Series Up to 12 Up to 10 Up to 22 Up to 13 Up to 6 Up to a maximum of 128 I/O 13,000 data points Serial: RS232/ RS485 (Selectable) ModBus: Supports MODBUS protocol, Master/Slave DC 24V W3.78” x H3.78” x D2.52” (96 x 96 x 64 mm) Monitors Networked to a computer system Features: ●●Centralized control from the PC or decentralized control from individual monitors ●●Specialized application specific software is available on request Monitors Networked to the R1000 PLC Features: ●●Centralized control from the PLC or decentralized control from individual monitors ●●Specialized application specific software is available on request 20
  • 22. Handheld Odor Monitor Highly Sensitive, 3 Models for Various Applications Handheld Odor Meter is the most popular simplified tool for odor nalysis, a which indicates the relative strength and odor lassification numerically by c comparing the odor gas and purified air. Features: Odor Monitor ●●Numerical value for relative strength of smell ●●Numerical value for classification of smell ●●Handheld Odor meter is ideal for Before and After applications, such as air purifica- tion and cleaning service. ●●Real-time sampling mode displays odor change continuously ●●Memory sampling mode saves data based on the selected sampling rate. (up to 32732 data, 511 files) ●●Battery operated with 7 hrs continuous usage Applications: OMX-SRM ●●Finding odor source at factories, incinerator, lants, or effluent treatment facility p OMX-ADM ●●Evaluation for deodorizing at hospitals or nursing homes ●●Suitable for putrid odor such as ammonia OMX-TDM ●●TVOC measurement for monitoring IAQ condition Specifications Model OMX-SRM OMX-ADM Detection Method Sampling Method Continuous Sampling with Built-in Air Pump Object Gas Odor Strength Level Classification Power Supply Memory Capacity Storage Temperature Operating Temperature Dimensions Weight 21 OMX-TDM Two Semiconductor Gas Sensors Ethanol, Acetone, Hydrogen, etc Hydrogen, Sulfide, Methyl Mercaptan, Ammonia, etc TVOC (Toluene), etc 0.0 to 999 0 to 999 0.0 to 9999 (µg/m3) Class 0 to 90 Odor Intensity 2.5 to 5.0 n/a 4 x AA batteries or AC adaptor *Battery life for continuous measurement is approx. 7 hours Up to 32732 data 0 to 40°C (32 to 104 ºF) * No condensation -10 to 50°C (14 to 122 ºF) * No condensation 74 (W) x 167.5 (L) x 35 (H) mm 250g (without batteries)
  • 23. Dif-Kit Tracer Gas Hardware Professional Fume Hood Diagnostic Tools Fume Hood Diagnostic Tools Features: ●●The ANSI/ASHRAE 110 test is a method of testing the performance of laboratory fume hoods. Kanomax Dif-Kit tracer gas hardware is ideal for use in performing the Tracer Gas test in accordance with ANSI/ASHRAE Standard 110-1995. ●●Diffuser is made to the design and specifications of Standard drawing #110-83M ●●The internal critical orifice ensures a flow rate of 4 liters per minute. Other orifice sizes can be inserted ●●Test-Mannequin and the vapor analyzer for the tracer gas test are available ●●Face velocity measurement data can be stored and downloaded with Anemometer ●●Dif-Kit includes tracer gas diffuser, nozzle assembly, in-line flow meter, tank regulator, and 20 ft of tubing ●●Test-mannequin includes lab coat and stand ASHRAE 110 - 1995 Performance Testing Dif-Kit Flow Meter Regulator AS 110-1995 Ejector Specifications Model Dif-Kit Critical Orifice Regulator Flow Meter 4 liter per minute Dual Stage, Specialty Gas Calibrated 150 mm/200 psi, Glass Tube Pressure Gauge 0 to 60 psi Vapor Analyzer The Vapor Analyzer is factory calibrated for Sulfur Hexafluoride (SF6). Response time is less than 10 seconds to 90% response with a stable maximum concentration in less than 30 seconds. Specifications Type Sample Flow Sample Cell Volume Gas Purge Concentration Alarm Interface Power Supply Dimensions Vapor Analyzer Weight Single Beam Infrared Spectrophotometer 10 liter per min (21.2 ft3/hr) 0.45 liters (0.0159 ft3) 5 Cell Volumes for 99% purge of cell High Low (selectable by user) USB, RS232 NiMH battery or 120-220V AC 15 x 7.3 x 7.5 inches (381 x 185 x 191 mm) 18 lbs (8.2 kg) 22
  • 24. Applications Sound Vibration Meters 23 Environmental noise measurement Noise exposure measurement Industrial vibration testing Sound and Vibration Meters
  • 25. Sound Level Meter Model 4431 Precision Sound Level Meter with 0-dB Function Sound Vibration Meters The new Kanomax sound level meter is designed to be compact, lightweight and easy to use. It complies with the type 2 ANSI S1.4 1983 standard , and has an innovative, unique 0-dB feature that eliminates the self-noise of the microphone. This technology extends the lower limit of the measurement range to lower than 0 dB-SPL. Features: ●●Lightweight and compact design ●●Equipped with highly sensitive electret condenser microphone ●●Large 4 digit display with 0.1 dB resolution with backlighting and analog bar graph ●●Add additional functions with the program cards ●●Includes SD card for data storage, windshield, carrying case, AA Optional program cards give you the flexibility to add additional functions as needed batteries, hand strap, and calibration certificate Specifications Model 4431 Parameters Lp, LA, LAeq, LAE, LAmax, LAmin, LAN, Lpeak, LAtm5 Ranges A: 28 - 130 dB, Z 39 - 130 dB Weighting A, C and Z Time Response Fast or Slow Microphone TYPE 7146nl (-28dB, Stand-alone -26dB) Type 1 (4432) or Type 2 (4431) standards for ANSI S1.4 1983, IEC 61672-1 Standards Power Supply 4 AA Batteries or optional AC Adapter (Battery life: approx. 9 hours) Size Weight W3.4” x H12.9” x D1.9”, Approx 1 lb. Accessories AC-1026: A CBC-0046-3: ACBC-0046-5: A CBC-0071: ACNA-0038W: ACNA-0038: A CNA-0038F: A CNA-0038R: ACNA-0333: AC adapter Microphone Cable (3m)* Extension cable (5m)* BNC-Pin Cable Data processing software Program card (1/1, 1/3 octave analyzer) Program card (FFT analyzer) Program card (Real Sound Recording) Tripod *Additional cable lengths available Sound Calibrator Sound Calibrator Model AC2127 implements stable and highprecision calibration by its feedback control function with temperature compensation. Calibrator Specifications Model Microphone sizes AC2127 1 inch and 1/2 inch (1/2 inch microphone adapter is included) Sound Pressure Level Frequency Level Standards 94dB (+/-0.3dB) 1000Hz (+/-0.1%) IEC 60942: 2003 Class 1 JIS C1515: 2004 Class 1 Power Supply 2 x AA Batteries Size Weight W4.3” x H2.9” x D1.8”, Approx 0.3 lb. (150g) Pistonphone Model AC2124A is a standard sound source, which emits 124dB-SPL pure sine tone at 250Hz for calibration of sound level meter.This is used for precise calibration of sound level meter provided with 1-inch, 1/2-inch and 1/4-inch condenser microphones. Pistonphone Specifications Model Sound Pressure Level Accuracy AC2124A 124dB +/- 0.2dB Frequency 250Hz (+/- 2%) Power Supply 4 x AA Batteries Size Weight 1.5” x 8.8”, Approx 1.7 lb. (800g) 24
  • 26. Sound Measurement Microphones Sound Vibration Meters Features: ●●Wide frequency range: wide range of measurement from low-frequency to ultrasonic frequency (1Hz to 200 kHz) is possible. ●●Wide dynamic range: measurement of high sound pressure level up to 170dB is possible. ●●Preamplifier integrated microphone: it is also possible to connect directly to an analyzer by using a BNC cable. ●●High sensitivity and certain measurement: it is possible to connect directly to analyzer such as an FFT and is easy to measure with low sound pressure levels. Preamplifier integrated microphone The preamplifier integrated microphone is compatible with IEPE (Integrated Electronic Piezoelectric Microphone) and can be connected directly to an FFT analyzer or other analyzers, enabling measurements with a highly accurate resolution at low cost. Long-term stability is accomplished by an accelerated aging process under 360°C. Specifications Model 4156N 4158N Nominal Diameter Release Voltage Pressure Sensitivity 4160N 4152N 4153N Maximum sound pressure level Self-noise level Temperature coefficient 7147A 7312 7313 1/2 inch -56dB (1.6mV/Pa) -48dB (4.0mV/Pa) -58dB +/- 3dB -50dB +/- 3dB -32dB (25.1mV/Pa) -26dB (50.0mV/Pa) -36dB (15.8mV/Pa) -38dB (12.5mV/Pa) -33dB +/- 3dB -40dB +/- 3dB -27dB +/- 2dB -37dB +/- 2dB -39dB +/- 2dB Polarization Voltage Frequency Characteristics 7146 1/4 inch 0V 20Hz to 40kHz 20Hz to 100kHz 20Hz to 10kHz 168dB 150dB 130dB 140dB 135dB 18dB 17dB 0.7dB/°C 0.01dB/°C 45dB 20Hz to 20kHz 20Hz to 10kHz 23dB 0.01dB/°C 0.009dB/°C Power supply and voltage 20Hz to 20kHz 20Hz to 10kHz 20Hz to 40kHz 20Hz to 20kHz 140dB 20dB 0.009dB/°C DC15 to 28V Constant current 0.5 to 4mA Connector SMB connector Dimensions Ø 0.27 x 1.9 BNC connector Ø 0.5” x 2.9” Ø 0.5 x 3.1 Ø 0.5” x 2.9” Measurement condenser microphone The condenser microphone can be used for measurements of sound pressure levels with high resolution over wide frequency ranges. A variety of microphone diaphragm diameters 1, 1/2, 1/4, 1/6, 1/8 inch and corresponding frequency characteristics support various applications. Specifications Model 7012 Nominal Diameter Release Voltage 7013 7016 1/2 inch 7017 7116 7118 1/6 inch 1/4 inch 1/8 inch -72dB ( 0.25mV/Pa) -20dB ( 100.0mV/Pa) -26dB ( 50.0mV/Pa) -28dB (39.8mV/Pa) -26dB ( 50.0mV/Pa) 20Hz to 20kHz 20Hz to 10kHz 20Hz to 10kHz 20Hz to 40kHz 20Hz to 20kHz -36dB (1.58mV/Pa) -38dB ( 12.5mV/Pa) -49dB (3.5mV/Pa) -58dB (1.3mV/Pa -60dB ( 1.0mV/Pa) 10Hz to 40kHz 10Hz to 20kHz 20Hz to 100kHz 20Hz to 20kHz 20Hz to 10kHz Polarization Voltage Frequency Characteristics 160dB 19dB 164dB 21dB 34dB Dimensions 25 7023 7047A 7146NL 1/2 inch 0V 170dB 46dB 45dB Temperature coefficient Preamplifier Type 7022 1 inch 200V Maximum sound pressure level Self-noise level 7020 140dB 65dB 2dB 146dB 11dB 160dB 13dB 134dB 20dB 15dB 0.007dB/°C Type 4212 Ø 0.5 x 0.5 Type 4116 Ø 0.3 x 0.4 Ø 0.2” x 0.4” Type 4022 Ø 0.1” x 0.2” Ø 0.9” x 1.1” Ø 0.9” x 0.7” Type 4212 Type 4011 Ø 0.5” x 0.6”
  • 27. Vibration Meter Model 4200 Compact, Easy-to-Use Vibration Meter Sound Vibration Meters The new Kanomax vibration meter is designed to be compact, lightweight and easy to use. The magnetic accelerometer attaches easily to machinery for increased accuracy and precise operation. It’s the perfect tool to diagnosis problematic vibrations with your industrial machinery or manufactured products. Features: ●●Compact size maximizes technician mobility ●●Magnetic accelerometer increases measuring accuracy and ease-of-use ●●Includes meter, accelerometer with cable and magnet, contact pin, 2 x AAA batteries, carrying case, and Calibration Certificate Specifications Measurement Range Acceleration 0.02 to 200 m/s2 RMS Velocity 0.02 to 200mm/s RMS Displacement 2 to 2000 µm EQp-p Frequency Range Acceleration Velocity 3Hz to 10kHz 10Hz to 1kHz (Compliant with JIS B0907-1989) Displacement Readings 10 to 400Hz RMS, Peak, EQ Peak, EQ Peak-to-Peak Output AC Output Headphone Output Interface 1Vrms (Full Scale) Portable Headphones w/Volume function RS232C Operating Environment Temperature Humidity Power Supply -10 to 50°C 30% to 90% (no condensing) Accessories AC-1046: AC7812B: ACPV-0148: A CPV-5050: ACNA-0134: ACSS-22M: ACBC-0071: ACBC-0116-3: A CNA-0116: A CBC-0026: AC adapter Pickup Spare Magnet Strong Magnet Auscultation Rod Stud BNC-Pin Cable E xtension Cable 3m* Data processing software Communication Cable (RS232) *Additional cable lengths available 2 x AAA batteries or AC adapter Dimensions Weight 5.7”(H)x1.9”(W)x0.9”(D) Approx. 4.6 oz. Vibration Monitoring System Model ACCM-393 is a system to monitor vibration of machinery and other applications. It can quickly detect abnormal vibration condition and provide alarm signal. CM-393 is ideal for continuous vibration monitoring at power plants or production facilities. Either of 2 measurement modes, acceleration or displacement, can be easily selected on the front panel. Includes storage case, 1 x acceleration pickup with built-in amplifier (Model 7828), pickup cable, and pickup cover. Specifications Model Input Impedance ACCM-393 Approx. 100kΩ Measurement Sensitivity 20mV/m/s2 (200mV/G) Measurement R ange Acceleration 1 to 100mm/s2 (0.1 to 10G) Displacement 1 to 1000µp-p Frequency Response Acceleration 5Hz to 1kHz Displacement 10Hz to 200Hz Accuracy Filter Output Power Supply Dimensions Within +/- 5% of full scale High-pass Low-pass Signal for level recorder Waveform ignal s Cut-off Frequency 5Hz to 12dB/oct Cut-off Frequency 200Hz to 12dB/oct Cut-off Frequency 1kHz to 12dB/oct 4 to 20mA/full-scale Terminal block on the back panel (M4) 0 to 2 Vp-p/full-scale BNC connector on the front panel Terminal block on the back panel (M4) AC 100V +/- 10%, 60Hz 7.9”(H) x 7.5”(W) x 18.5”(D) *Variety of acceleration pickups are available. 26
  • 28. Applications Dust Monitors 27 Monitoring worker exposure to airborne contaminants Indoor Air Quality Assessments Dust Monitors
  • 29. ●●Measures PM 10, Respirable, or PM 2.5 particle matters ●●Real-time measurements concentration of dust ●●Data logging up to 500 measurements ●●Simple cleaning mechanism for easy maintenance ●●Complete with data processing software, RS232C cable, cleaning kit, Ni-MH battery pack, AC adapter, carrying case, and calibration certificate Digital Dust Monitor Model 3443 Dust Monitors Piezobalance Dust Monitor Model 3521 / 3522 ●●Measures PM 10 particle matters ●●Compact and Lightweight unit ●●Analog output controls other devices ●●Data logging up to 100,000 measurements ●●PC interface with USB and software for downloading data to your PC ●●Complete with rubber protector, shoulder strap, data processing software, USB cable, AC adapter, 2 pcs. LCD protective sheet, rubber cap, 2 pcs. filter, and calibration certificate Specifications Model 3521 / 3522 3443 Measuring Method Piezobalance Light Scattering Particle Size Range 0.1 to 10 µm (Model 3521) 0.1 to 2.5 µm (Model 3522) 0.1 to 10 µm Measuring Range 0.01 to 10.00 mg/m3 0.001 to 10.000 mg/m3 1.0 L/min 1.0 L/min 500 measurements 100,000 measurements RS232C USB Analog Output n/a 0 to 1 V / Pulse / Alarm Power Supply Ni-MH Battery or AC 100 - 240 V Li-ion Battery or AC 100 - 240 V W2.6” x H7.1” x D5.9” W6.7” x H2.7” x D4.3” 3.9 lbs (1.75 kg) 2.9 lbs (1.3 kg) Flow Rate Datalogging Interface Dimensions Weight 28
  • 30. Piezobalance Dust Monitor Model 3521/3522 Optimal Tool for Monitoring Oil Mist Features: ●●Measures PM 10, Respirable, or PM 2.5 particle matters, such as dust, oil mist, fumes, and soot ●●Real-time measurements of dust concentration Dust Monitors ●●Data logging up to 500 measurements and data may be reviewed on screen or printed ●●PC interface with RS232C and software for downloading data to your PC ●●Easy operation and requires no special training ●●Simple cleaning mechanism for easy maintenance ●●Includes data processing software, RS232C cable, cleaning kit, Ni-MH battery pack, AC adapter, carrying case, and calibration certificate Specifications Model 3521 3522 Measuring Method Particle Size Range Measuring Range Piezobalance 0.1 to 10 µm 1.0 L/min 500 measurements Interface Power Supply Dimensions Weight Software Included 0.01 to 10.00 mg/m 3 Flow Rate Datalogging Accessories 0.1 to 2.5 µm RS232C Ni-MH Battery or AC 100 - 240 V W2.6” x H7.1” x D5.9” 3.9 lbs (1.75 kg) 3521-01: 3521-02: 3521-03: 3521-04: 3 521-05: 3 521-06: 3521-07: 3521-08: 3521-20: D PU-S245: T P-202L: Rechargeable Battery Pack Carrying Case 10 µm Impactor Nozzle (for 3521) 4 µm Impactor Nozzle (for 3521) AC Adapter Cleaning Sponges (3 pieces) Cleaning Fluid Communication Cable to PC P rinter Cable Portable Thermal Printer Rolled Printer Paper (10 rolls) Dust Measuring Methods Piezobalance Method An air sample enters the system, it travels through the impactor, which captures and removes larger particulates away from the sample. Smaller particulates become electrically charged and deposited on the piezo-crystal. The total mass of the deposited particulates affects the piezocrystal’s frequency. Since the change in frequency is proportional to the mass of the particulates, the actual weight of the particulates is obtained. Since some particle matters such as oil mist absorb lasers, the Piezobalance dust monitor would be ideal (the light scattering method would not give correct measurements). Applications: • Monitoring milling operation • Monitoring honing • Monitoring boring operation Monitoring Milling Operation Light Scattering Method When a laser hits particle matter, light scattering occurs. A dust monitor collects the amount of scattering light and calculates the mass concentration in proportion to the scattering light. Mass concentration is based on the density of particle matter, thus gravimetric sampling is required if the density is unknown. Applications for light scattering dust monitor include Indoor air quality investigations, Point source monitoring, and Personal exposure monitoring. 29 Monitoring Welding Operation
  • 31. Digital Dust Monitor Model 3443 Features: ●●Measures PM 10 particle matters, such as dust, fumes, and smoke ●●Real-time and long term measurements of dust concentration ●●Compact and Lightweight unit ●●PC Dust Monitors interface with USB cable and software for downloading data to your PC ●●Analog output controls other devices ●●Data logging up to 100,000 measurements and displays MIN / MAX / AVG and Timing graph for review ●●Complete with rubber protector, shoulder strap, data processing software, USB cable, AC adapter, 2 pcs. LCD protective sheet, rubber cap, 2 pcs. filter, and calibration certificate Specifications Model Measuring Method Particle Size Range Measuring Range Flow Rate Datalogging Interface Analog Output Power Supply Dimensions Weight 3443 Light Scattering 0.1 to 10 µm 0.001 to 10.000 mg/m3 1.0 L/min 100,000 measurements USB 0 to 1 V / Pulse / Alarm AA Battery or AC 100 - 240 V W6.7” x H2.7” x D4.3” 2.9 lbs (1.3 kg) Software Included With Rubber Protector Accessories 3442-01: 3442-02: 3442-04: 3442-05: 3442-10: 3431-03: 6113-02: C X-440: Analog Output Cable Rubber Protector L CD Protective Sheet C arrying Case Rubber Cap Filter (10 pc) A C Adapter T ripod CEGRIT Automatic Flyash Sampler Isokinetic Sampling for Particle Emission Testing With no moving parts, each CEGRIT sampler collects a sample from one point in the boiler duct. Operating on duct vacuum to drive its atmospheric-air ejector, the CEGRIT maintains near-isokinetic sampling to keep collecting unbiased sample as boiler load and duct vacuum vary. Features: ●●Operates continuously with no moving parts ●●High efficiency sampling rate for fine pulverized dust ●●Industrial construction ●●Cyclone removes to facilitate periodic cleaning ●●Two manometer tap points to monitor draft pressure Accessories Applications: ●●All combustion or other processes where airborne particle emission occur ●●Smoke and particle emissions or grit burdens from boiler stacks ●●Carbon determination in Fly ash ●●Incinerator emission compliance C8406: C8408: C8407: C3101: C3102: C3103: C3104: C5019: C 5020: Probe (2 m) P robe (3 m) P robe Extension (10 ft) Inlet Nozzle 1/2” (12.7 mm) Inlet Nozzle, 5/8” (15.9mm) Inlet Nozzle, 3/4” (19.0 mm) Inlet Nozzle, 3/8” (9.5 mm) Heater Jacket, 100W, 240 Volt Heater Jacket, 100W, 110 Volt 30
  • 32. Applications Particle Counters 31 Cleanroom Certification Indoor Air Quality Investigation Filter Testing Particle Counters
  • 33. Particle Counters Handheld Particle Counter Model 3887 Portable Particle Counter Model 3910 3905 Handheld Particle Counter Model 3886 Geo-α Simple and easy-to-use, affordable handheld particle counter Lightweight handheld particle counter with environmental sensors Small footprint, lightweight portable particle counter available with two different flow rates ●●Simultaneously measures and ●●Simultaneous measurements of ●●Simultaneous measurements of ●●Built-in flow sensor ensures high ●●Air velocity, temperature, and displays 3 particle sizes 5 particle sizes accuracy measurement ●●Store up measurements to 10,000 ●●Data logging measurements interface with RS232C and software for real-time measurements and downloading data to your PC Light Source Counting Efficiency to enclosure touch displays all simultaneously ●●Store up measurements Stainless screen LCD measurements to 10,000 ●●PC software allows Remote control, Real-time measuring, and Registering map ●●Includes zero filter, metal handle, 4 pcs. AA batteries, battery charger, AC adapter, and calibration certificate 3887 footprint, ●●Large 500 optional software for real-time monitoring with timing graph zero filter, meter stand, data processing software, RS232C cable, 4 pcs. AA batteries, battery charger, AC adapter, and calibration certificate Flow Rate up ●●Small ●●PC interface with RS232C and ●●Complete with Isokinetic probe, Particle Sizes ●●50.0 or 28.3 L/min flow rate humidity measurements with optional probes ●●PC Specifications Model 6 particle sizes ●●Complete with Quick-start guide, AC adapter, zero filter, Isokinetic probe with 79 inch (2 m) tubing, inlet nozzle, data processing software, memory card, 2 rolls of cleanroom-grade printer paper, Li-ion battery, and calibration certificate 3886 3910 and 3905 0.3 / 0.5 / 5.0 µm 0.3 / 0.5 / 1.0 / 3.0 / 5.0 µm 0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0 µm 0.1 cfm (2.83 L/min) 0.1 cfm (2.83 L/min) 50.0 L/min (3910) or 28.3L/min (3905) Laser Diode Laser Diode Laser Diode 50+/-20% @ 0.3 µm 50+/-20% @ 0.3 µm 50+/-20% @ 0.3 µm Coincidence Loss Less than 5% at 2,000,000 particles/ft Less than 5% at 2,000,000 particles/ft Less than 10% at 500,000 particles/ft3 Zero Count Level Less than 1 count per 5 minutes Less than 1 count per 5 minutes Less than 1 count per 5 minutes 10,000 measurements 500 measurements 10,000 measurements RS232C or USB RS232C or USB Ethernet, USB, Memory card slot (MMC) Datalogging Interface Option Sensor Enclosure Power Supply Dimensions Weight 3 3 n/a Air Velocity / Temperature, Humidity Air Velocity, Temperature, Humidity Molded Plastic Molded Plastic Stainless Steel 4 x AA Batteries or AC 100 - 240 V 4 x AA Batteries or AC 100 - 240 V Li-ion Battery or AC 100 - 240 V W4.4” x H7.8” x D2.8” W4.5” x H8.3” x D2.8” W7.9” x H8.1” x D7.9” 1.5 lbs (680 g) 2.2 lbs (980 g) 14.2 lbs (6.44 kg) 32
  • 34. Handheld Particle Counter Model 3887 Features: ●●Simultaneously measures and displays 3 particle sizes ●●Built-in flow sensor ensures high accuracy measurements ●●Store up to 10,000 measurements Particle Counters ●●PC interface with RS232C and software for real-time measurements and downloading data to your PC ●●Complete with isokinetic probe, zero filter, meter stand, data processing software, RS232C cable, USB to serial adapter, 4 pc. AA batteries, battery charger, AC adapter, carrying case, and calibration certificate Specifications Model 3887 Particle Sizes 0.3 / 0.5 / 5.0 µm Flow Rate Light Source Counting Efficiency 0.1 cfm (2.83 L/min) Laser Diode 50+/-20% @ 0.3 µm Coincidence Loss Less than 5% at 2,000,000 particles/ft3 Zero Count Level Less than 1 count per 5 minutes Datalogging Interface Enclosure Power Supply Dimensions Weight 10,000 measurements RS232C or USB Molded Plastic 4 x AA Batteries or AC 100 - 240 V W4.4” x H7.8” x D2.8” 1.5 lbs (680 g) Software Included Accessories 3887-02: 3 887-07: DPU-S245: T P-202L: CX-440: 301APK: Carrying Case Printer Cable Portable Thermal Printer R olled Printer Paper (10 rolls) Tripod Intl. Plug Adapter Kit Handheld Particle Counter Model 3886 Geo- α Features: ●●Simultaneous measurements of 5 particle sizes ●●Air velocity, temperature, and humidity measurements with optional probes ●●Data logging up to 500 measurements ●●PC interface with RS232C and optional software for real-time monitoring with timing graph ●●Compact and lightweight, easy handling ●●Includes AC adapter, zero filter, metal handle, 4 pc. AA batteries, battery charger and calibration certificate Specifications Model Particle Sizes Flow Rate Light Source Counting Efficiency 3886 0.3 / 0.5 / 1.0 / 3.0 / 5.0 µm 0.1 cfm (2.83 L/min) Laser Diode 50+/-20% @ 0.3 µm Coincidence Loss Less than 5% at 2,000,000 particles/ft3 Zero Count Level Less than 1 count per 5 minutes Datalogging Interface Optional Sensor Enclosure Power Supply Dimensions Weight 33 500 measurements RS232C or USB Air Velocity / Temperature, Humidity Molded Plastic 4 x AA Batteries or AC 100 - 240 V W4.5” x H8.3” x D2.8” 2.2 lbs (980 g) Software for Monitoring and Remote Control Accessories 0842: 0843: 0843-01: 3886-02: 3886-04: S388-61: 3886-07: 3886-08: DPU-S245: TP-202L: CX-440: Temperature and RH Probe Air Velocity Probe E xtension Rod for 0843 Carrying Case I sokinetic Probe D ata Processing Software for Windows Printer Cable Communication Cable to PC Portable Thermal Printer Rolled Printer Paper (10 rolls) T ripod Carrying Case
  • 35. Portable Particle Counter Models 3910/3905 Smallest and lightest portable particle counters Features: ●●Simultaneous measurements of 6 particle sizes Particle Counters ●●Small footprint, Stainless enclosure ●●Availabe in 50.0 or 28.3 L/min flow rates ●●Complies with all requirements of ISO 21501-4 ●●Large touch screen LCD displays all measurements simultaneously ●●Store up to 10,000 measurements ●●21 CFR Part 11 compliance ●●Complete with quick-start guide, AC adapter, zero filter, isokinetic probe with 79 inch (2 m) tubing, inlet nozzle, data processing software, memory card, 2 rolls of cleanroom-grade printer paper, 1 x Li-ion battery, and calibration certificate Specifications Model Particle Sizes Flow Rate Light Source Counting Efficiency Coincidence Loss Zero Count Level Datalogging Interface Optional Sensor Enclosure Power Supply Dimensions Weight 3910 3905 0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0 µm 50.0 L/min 28.3L/min Laser Diode 50+/-20% @ 0.3 µm Less than 10% at 500,000 particles/ft3 Less than 1 count per 5 minutes 10,000 measurements Ethernet, USB, Memory card slot (MMC) Air Velocity, Temperature, Humidity Stainless Steel Li-ion Battery or AC 100 - 240 V W7.9” x H8.1” x D7.9” 14.2 lbs (6.44 kg) Model 3910/05 Model 3900 Small, Lightweight Unit Optional environmental sensor Optional Climomaster Environmental Sensor measures airflow, temperature and humidity. Probe Specifications Model 6531-2G-P Probe Type Uni-Directional Air Velocity 2 to 6000 fpm Temperature Relative Humidity -4 to 158°F Climomaster probe for environmental measurements 2.0 to 98.0% RH Accessories 6531-2G-P: 3910-01: 3910-02: 3910-03: 3910-04: 3910-05: Air velocity, Temp, RH probe with 2 m cable Carry case Battery charger Spare Li-ion battery Alarm-output cable Pressure-sensor (w/ connection cable) 3910-08: 3910-09: CRVAL: Zero filter Printer paper V alidation IQ/OQ Document 34
  • 36. Handheld Condensation Particle Counter Model 3800 Optimal Screening Tool for Nano Size Particles Features: Particle Counters ●●0.015 micron (15nm) minimum sensitivity ●●Handheld and Lightweight Aerosol Research Instrument ●●Store up to 10,000 measurements ●●PC interface with USB and software for real-time measurements and display time fluctuation graph ●●Complete with zero filter, data processing software, USB cable, alcohol bottle, 6 pc. AA batteries, carrying case, and calibration certificate Measuring Priciple Software Included Specifications Model 3800 Measuring Object Airborne Particle Matter Particle Sizes 0.015 to 1.0 µm Flow Rate 0.7 L/min Light Source Laser Diode 100% @ 0.05 µm More than 50% @ 0.015 µm Counting Efficiency Coincidence Loss Less than 5% at 100,000 particles/cm3 Zero Count Level 1 count per 5 minutes Alcohol Supply Datalogging Interface Isopropyl Alcohol 10,000 measurements USB Power Supply 6 x AA Batteries or AC 100 - 240 V Dimensions Weight W4.7” x H11.0” x D5.1” 3.3 lbs (1.5 kg) Applications Checking Worker Exposure to Airborne Contaminants Leakage Testing Engine Exhaust Testing Accessories 3800-01: 3800-02: 3800-03: 3800-04: 3800-05: 3800-06: 35 AC Adapter Zero Filter Alcohol Bottle Storage Cap Alcohol Cartridge Spare Felt and Wire Mesh 3800-07: 3800-08: 3800-10: 3800-11: DPU-S245: TP-202L: Communication Cable to PC Carrying Case Sampling Probe with 1m Tubing Printer Cable Portable Thermal Printer Rolled Printer Paper (10 rolls) Sampling Probe
  • 37. Remote Particle Sensor with Analog Output 2-channel remote particle sensors for facility monitoring applications Analog Particle Sensors Kanomax particle sensors with analog output are designed to fit into your existing monitoring system, or they can be used as a stand-alone unit to monitor a critical area when connected to an alarm or controller. The Kanomax particle sensor is available with a 1.0 CFM flowrate and also comes in two different ranges: 0.3 and 0.5 microns or 0.5 and 5.0 microns. Features: ●●Light scattering particle sensor ●●1.0 cfm flow rate (28.3 lpm) ●●Analog output makes it easy to integrate into existing systems ●●0.3 or 0.5 micron sensitivity ●●Compact body with stainless enclosure ●●Fully compatible with the Kanomax Cleanroom Monitoring System as well as third party existing systems Applications: ●●Cleanroom Facility Monitoring ●●Food Industry ●●Aerospace ●●Hospital Surgical Rooms Specifications Model Measuring Object Particle Sizes Flow Rate Light Source Counting Efficiency Coincidence Loss Zero Count Level Interface Enclosure Power Supply Dimensions Weight 3716A 3717A Airborne Particle Matter 0.3 / 0.5 µm 0.5 / 5.0 µm 1.0 cfm (28.3 L/min) *External vacuum source is required Laser Diode 50% @ 0.3 µm 50% @ 0.5 µm Less than 5% at 500,000 particles/ft3 1 count per 5 minutes RS485 4-20mA Stainless Steel 9 - 28V (AC adpter supplies 12V) W6.3” x H5.1” x D3.9” 3 lbs (1.5 kg) Example Configuration: Shows 6 sensors connected to datalogger via the 4-20 mA output. It’s easy to connect them to a datalogger or any other device that will accept an analog signal. Note that a vaccuum source is required for the sensors. Vacuum Pump Connect up to 6 particle sensors 4-20mA connection to datalogger or acquisition system 36
  • 38. Particle Sensors and OEM Sensors Particle Sensors Large Particle, PM and Pollen Sensors The pollen sensor detects the particle with light scattering method, using one light emitter and two light receptors, and discriminates pollen from other particles with two factors “scattered light intensity” and “degree of polarization”. The large particle sensor is the sensor for industrial applications to detect particles sized over 10 micron. This sensor is used in production lines that dislike large particles. “PM2.5” is known as the fine particle which is one of the cause of asthma or respiratory disease, etc. Especially, diesel soot commonly referred to as PM2.5 is suspected to raise the risk of lung cancer. Our PM sensor has a high correlation with Federal Equivalent Method (FEM) and achieves low-cost, simple installation for multipoint time-series monitoring. Specifications Model Detectable particle size PS2 Pollen Sensor Large Particle Sensor PM Sensor approx. 15µm and larger approx. 15µm and larger approx. 0.3µm and larger Concentration Range: equiv. PM2.5 (0-200 µg/m3) Supply Voltage 12VDC Operating Temperature -10 to 45°C Operating Humidity 20 to 80% RH (no condensation) Dimensions 71.4 x 76.4 x 36.7 mm Weight 120g (approx) 120g (approx) 130g (approx) Particle and Aerosol OEM Sensors Designed for easy OEM Integration Particle Sensor Units - Based on the light scattering method, it always detects airborne particles. Pulse output that corresponds to concentration per unit volume of particles can be obtained, with using an original detection method based on light scattered principle similar to the particle counter. Particle Sensor Units Aerosol Sensor Unit - Particulates of 0.5 microns or larger can be monitored. The aerosol sensor unit can be used to keep optimum driving of filtering equipment related to a clean room depending on actual condition of cleanliness level. Features: ●●Unique air sampling method using “heater” to generate updraft, preventing choked flow. Suc- tion pump is not necessary. PPD42NJ ●●Stable and sensitive detection for particulates of 0.5 microns and larger can be monitored. ●●Long life laser. 20,000 hours ●●Compact size, low cost and easy to integrate into your original equipment. Particle Sensor Applications: Aerosol Sensor Applications: PPD60PV ●●Air Purifier ●●Clean booth/small ●●Air Quality Monitor ●●Fan Filter Unit ●●Air Conditioner ●●Clean Unit ●●Ventilation ●●Clean Room related filtration system Aerosol Sensor Unit ●●Vaccuum Cleaner Specifications Model No. Detectable Particle Size PPD42NJ Approx. 1 µm larger Supply Voltage Output Method Dimensions Weight 37 AES1-05N Approx. 0.5 µm larger 5VDC 12VDC 0 to 45°C 0 to 40°C 95% RH or less (no condensation) 30 to 80% RH (no condensation) Operating Temperature Operating Humidity PPD60PV Digital Digital or Analog Digital 59x45x22 mm 88x60x20 mm 90x90x23 mm 24 g 36 g 130 g AES1-05N
  • 39. Particle Monitor AES-1000 Continuous Particle Monitoring for Modular Clean Room Applications The AES-1000 particle monitoring system provides near real time stable particle measurements at low cost. Measurement covers the equivalent range of ISO class 5 through ISO class 8. Aerosol Monitor It uses a long life semiconductor laser to detect optical scatter by particles and an integral heater to provide continuous flow by thermal convection to provide a long maintenance free life. Features: ●●Stable and sensitive detection for particulates of 0.5 microns and larger. ●●Easy-to-read air cleanliness level, equivalent of ISO5/ Class 100 -ISO8/Class 100,000. ●●Unique air sampling method using “heater” to generate updraft, preventing choked flow. Suction pump is not necessary. ●●Alarm comes on when the setting level goes beyond the required cleanliness, while monitor gives alarm signal output. LED Light continuously shows current air cleanliness (equiv. of Class 100 - Class 100,000) ●●Link to Ethernet system with purpose of monitoring at multiple locations. ●●20,000 hour long life laser. ●●Compact and light, easy installation and Low cost. Applications: ●●Monitor cleanrooms ●●Pharmaceutical / Semiconductor / Food Industry Alarm comes on when the setting level goes beyond the required cleanliness level while monitor gives alarm signal ouput. Set and Reset for alarm ●●Buffer Zone Monitoring ●●Indoor Air Quality Wall-mount Monitor Specifications Model Monitoring Particle Size Display Sampling Method Optics AES-1000 0.5 microns 10 x LED lights shows equivalent of Class; 100(ISO 5), 1,000(ISO 6), 2,000, 3,000,5,000,10,000(ISO 7), 20,000, 30,000, 10,000, 100,000(ISO 8) Updraft airflow generated by built-in heater enables air sampling Semiconductor Laser, 20,000 hour life Alarm Display Alarm LED light comes on when reaching to setting level Alarm Output TTL Open collector output in sync with alarm display Interface Supply Voltage Operating Temp/Humidity Ethernet X 1 port 12VDC+/-10% 500mA 0 to 40 deg. C, 30 to 80%rh PC Software is Included 38
  • 40. Cleanroom Monitoring System Cleanroom Monitoring System The Kanomax Cleanroom Monitoring System (CRMS) provides an automated means to monitor and gather airborne particle counts and other parameter levels in controlled environments. The CRMS allows users to perform a variety of functions from a PC, including changing alarm settings and viewing particle count concentrations. Features: ●●Compact stainless enclosure with sensors ●●1 PC system controls up to 128 sensors ●●Multi-parameter ●●Alarm outputs: warning light, on-screen, or pager notification measurements: Particle count, Air velocity, Temperature, Humidity, and Differential pressure ●●Multi-function, user-friendly monitoring software Particle Sensors Features: ●●Light scattering particle sensor ●●Durable stainless enclosure is easy to sanitize during whole facility cleaning ●●No system down-time: each sensor is replaceable for repair and recalibration Specifications Model Measuring Object Particle Sizes Flow Rate Light Source Counting Efficiency Coincidence Loss Zero Count Level Interface Enclosure Power Supply Dimensions Weight 3714 3715 Airborne Particle Matter 0.3 / 0.5 µm 0.5 / 5.0 µm 0.1 cfm (2.83 L/min) *External vacuum source is required Laser Diode 50% @ 0.3 µm 50% @ 0.5 µm Less than 5% at 1,000,000 particles/ft3 1 count per 5 minutes RS485 Stainless Steel DC24V (Supplied from the 3770) W5.0” x H2.8” x D1.6” 1.1 lbs (500 g) Specifications Features: Model Measuring Object Particle Sizes Flow Rate ●●Light scattering particle sensor ●●1.0 cfm flow rate (28.3 lpm) ●●Analog output makes it easy to integrate into existing systems Light Source Counting Efficiency Coincidence Loss Zero Count Level Interface Enclosure Power Supply Dimensions Weight 3716A 3717A Airborne Particle Matter 0.3 / 0.5 µm 0.5 / 5.0 µm 1.0 cfm (28.3 L/min) *External vacuum source is required Laser Diode 50% @ 0.3 µm 50% @ 0.5 µm Less than 5% at 500,000 particles/ft3 1 count per 5 minutes RS485 4-20mA Stainless Steel 9 - 28V W5.9” x H4.7” x D3.9” 3 lbs (1.5 kg) Specifications Features: ●●0.2 µm sensitivity particle counter ●●Digital and analog outputs are available ●●Analog output for Multiplexer 39 Model Measuring Object Particle Sizes Flow Rate Light Source Counting Efficiency Coincidence Loss Zero Count Level Interface Power Supply Dimensions Weight 3792-01 Airborne Particle Matter 0.2 / 0.3 µm 0.1 cfm (2.83 L/min) Laser Diode 50% @ 0.2 µm Less than 5% at 1,000,000 particles/ft3 1 count per 5 minutes RS485 / 4 to 20 mA for Multiplexer AC 100 to 240 V W4.6” x H5.3” x D6.7” 6.6 lbs (3 kg)
  • 41. Interface Box ●●Converts analog input to digital ●●Supplies power to the sensor Specifications Model Input Output Power Supply Dimensions Weight 3772-02 4 to 20 mA / 0 to 1 V / 1 to 5 V RS485 DC24V (Supplied from the 3770) W5.5” x H3.1” x D1.6” 1.1 lbs (500 g) with Differential Pressure Sensor Distributors Features: Cleanroom Monitoring System Features: Specifications ●●Supplies data communication and power to sensors via RS485 ●●1 unit connects up to 8 sensors Model # of Channels Interface Power Supply Dimensions Weight 3770 8 RS485 AC 85 to 132 V or AC 170 to 267 V W11.8” x H3.9” x D7.9” 6.6 lbs (3 kg) Cleanroom Monitoring Software Features: ●●Continuous monitoring and data processing software ●●Remote monitoring via LAN ●●Multi-function: Map display at a glance, Real-time graph, Maintenance Indication, and Data table Map Trend Graph System Examples Industrial Cleanroom Bio-medical Cleanroom Semiconductor, HDD, Flat Panel Display, Electronics Food, Pharmaceutical, Hospital surgical rooms 40
  • 42. Aerosol Research Instruments Our line of aerosol research instruments covers a variety of applications. Aerosol Research Instruments Industrial hygiene Nanosampler Health effect research Nano-particle research Portable Aerosol Mobility Spectrometer (PAMS) Condensation Particle Counter Air pollution Ambient air quality research Ambient Air Quality Monitor Earth sciences Black Carbon Monitor Environmental research Climate change research 41 Aerosol Particle Mass Analyzer
  • 43. Small, lightweight mobility size spectrometer Features: ●●Electrical mobility size spectrometer designed for portable, mobile, or handheld aerosol sampling applications ●●Provides number-weighted diameter distribution of aerosols over the entire submicrometer range (10 to 863nm) in one scan ●●Uses a non-radioactive bipolar aerosol charger to allow easy access to sampling sites with tight safety regulations ●●Bipolar charger significantly reduces measurement uncertainty of larger particles in the submicrometer range ●●Can be used in two modes: Single diameter count mode: to get a total count within a narrow size range or Size distribution mode: to get an automated size distribution measurement over desired size range or size resolution Aerosol Research Instruments Portable Aerosol Mobility Spectrometer (PAMS) ●●Stand-alone, battery-operated instrument; no external computer needed ●●Weighs only 4.5 kg and measures 23 x 23 x 15 cm Operation: Sampled aerosol is charge-conditioned using a non-radioactive, bipolar charger which brings the particles to steady-state charge distribution. Particles are subsequently sorted according to their electrical mobility in a differential mobility classifier (DMA). Classified particles are detected and counted downstream using a condensation particle counter (CPC). Specifications Model 3300 Measuring Modes Single size mode (single size concentration) Scan mode (size distributions) 0.2LPM: 14.5 to 863 nm 0.4LPM: 10 to 433 nm Size Range 0.2LPM: 14 channels 0.4LPM: 27 channels Scan Resolution 0.2LPM: 42 sec to minutes for one scan 0.4LPM: 81 sec to minutes for one scan Scan Time Concentration Range 0 to 100,000 particles /cm3 Flow Rate Inlet: 0.7 LPM Sample: 0.05 LPM Bipolar Charger Inlet: 0.7 LPM Sample: 0.05 LPM Condensing Fluid Isopropyl alcohol Display Color touchscreen Dimensions 9.1 x 9.1 x 5.9 inch (23 cm x 23 cm x 15 cm) Weight 9.9 lbs (4.5 kg) Accessories 3300-05: 3300-08: 3300-09: 3300-10: Spare Alcohol Cartridge Spare Lithium-ion Battery Battery Charger Carrying Case 42
  • 44. Aerosol Research Instruments Aerosol Particle Mass Analyzer Model APM-3601 Lightweight, Desktop Aerosol Particle Mass Analyzer Aerosol Particle Mass Analyzer classifies particles by mass based on the balance between centrifugal force and electrostatic force. Particle size distribution measurement is normally used in order to measure nanosized particle distribution. While DMA (Differential Mobility Analyzer) classifies particles by particle size utilizing electrostatic force, APM classifies particles by mass based on entirely new classification principles. APM classifies aerosol particles of 0.01 ~ 100 femtograms. Particle density distribution can be attained by combining the APM and DMA. Applications: ●●Mass distribution measurements ●●Particle density research ●●Monodisperse aerosol generation Features: ●●Lightweight, Desktop Unit ●●Measurement software, communication cable, extension board, extension adapter and box, and plastic case are available as options Specifications Model Particle Mass Range Maximum Rotation Speed Maximum Voltage Rotating Cylinder Dimensions Sampling Flow Rate Power Supply Dimensions Weight APM-3601 Rotating Cylinder Assembly 0.01 to 1000 femtogram Equivalent to approx. 14 nm to 1.3 µm for particle density of 1 g/cm3) 1,000 to 14,000 rpm Up to 2000V Inner Cylinder Diameter: 48 mm Gap between Inner and Outer Cylinders: 1 mm Cylinder Length: 100 mm 0.3 to 1.0 L/min (0.3 L/min is recommended) Single-phase AC100 ~ 240V Main Unit: W 16.9” x H 5.5” x D 7.9” Control Unit: W 16.9” x H 7.1” x D 13.8” Main Unit: 23 lbs (10.5 kg) Control Unit: 15 lbs (7 kg) Control Unit APM experimental data comparison with theoretical curve at 100nm (PSL) 43
  • 45. Aerosol Research Instruments Black Carbon Monitor Model 3130 Continuous Soot Monitoring System The Model 3130 monitors changes in transmittance across an automatically advancing quartz fiber filter tape using an LED at a 565 nm wavelength. To achieve measurements with high sensitivity and a lower detectable light absorption coefficient, the Model 3130 uses a double-convex lens and optical bundle pipes to maintain high light intensity and signal data, obtained at 1000 Hz. Features: ●●Measures black carbon concentration in the air in real time ●●Preprocessing of sample air improves measurement accuracy ●●Automatic filter feed enables continuous measurement for extended periods ●●Advanced detection sensitivity enables measurement in low concentrated areas ●●Complete with inlet heater, pump, pump controller, 2.5 µm-cut impactor, software, and power cable Specifications Model Measurement Method Light Source Lowest Detection 3130 Light Absorption Method LED Wavelength 565 nm 0.05 µg/m3 @1 min @0.8 L/min Collection Flow Rate Collection Filter 0.8 L/min Fiberglass Filter (Length: 25 m) PC Interface USB Power Supply Dimensions 100 VAC, 6A W 17” x H 8.7” x D 13.9” Weight 37.5 lbs (17kg) System Diagram Nanosampler Model 3180 Inertial Filter for Aerosol Collection Fiber filters utilize several mechanisms for aerosol collection such as inertial, gravitational settling, interception, and diffusion. Each collection efficiency depends on particle size and filtration velocity. Large particles are collected in a filter by inertial impaction at a high filtration velocity while small particles are removed from the air by Brownian diffusion.The Model 3180 controls its diffusion collection increasing filtration velocity, and collecting particles only by inertia. Features: ●●Particle classification as small as 0.1 µm ●●5 stages of PM10, 2.5, 1.0, 0.5, 0.1 and backup filter ●●Large sample flow rate of 40 L/min ●●Includes 1 box of quartz fiber filter (55mm x 100 pc.), and 5 sets of PM0.1 cartridge ●●Pump unit (Vacuum pump + Rota meter with valve), PM0.1 cartridge case (1 case for 5 cartridges), Quartz fiber filter case (1 set of 20 cases), PM0.1 absorption materials (10 sets) are available Specifications Model Classification Method 3180 Inertial Collection Classification Range PM 10, PM 2.5, PM 1.0, PM 0.5, PM 0.1 and Backup filter Absorption Material Quartz Fiber Filter: PM 10, PM 2.5, PM 1.0, PM 0.5, and Backup fiter SUS fiber: PM0.1 Sampling Flow Rate 40 L/min Dimensions Ø 3.5” x 8.3” Weight 6.6 lbs (3 kg) 44
  • 46. Dust Sentry Aerosol Research Instruments Accurate and affordable real-time outdoor particulate monitoring The Dust Sentry provides real time, continuous particle measurement of TSP, PM10, PM2.5 or PM1 with datalogging and alarm relay outputs. Numerous integrated options including SMS and/or email notifications, weather sensors, noise and mounting equipment make the Dust Sentry a flexible base system suitable for a wide range of applications. Features: ●●Real time particle monitoring, Measures and reports data in 1 minute intervals ●●Accurately measures TSP, PM10, PM2.5 or PM1 ●●On board data storage for up to 15 years of data ●●Web-based and GSM data retrieval and alerts ●●Gravimetric filter ●●Option to integrate weather, wind, temp/RH sensors, and sirens ●●Tough weatherproof enclosure with solar shielding for hot climates ●●Quick installation and relocatable ●●Datalogging and configuration software included Specifications Model DS 10 Measurement Technology DS 25 DS 1 Near forward light scattering nephelometry Range 0-2000 µg/m3 Accuracy +/- (2 µg/m3 + 5% of reading) Cut Points Available PM 10 PM 2.5 Sample Flowrate PM 1.0 2.0 LPM Inlet Heated Alarm Relay and SMS (optional) Datalogging 2GB SD card (15 years data storage) Reference Filter 25mm GF filter Outputs RS232 cellular IP Gateway Wireless Communication Options Low power RF modem Environmental Operating Range -10°C to +45°C; 10 to 95% RH (NC) Power Requirements Mains (100 to 260VAC) or regulated 12VDC (10W) Lockable weatherproof GRP cabinet IP65; 685 x 350 x 300mm (including solar shield armour mounting brackets) Enclosure Weight Compared to reference-equivalent monitors like TEOMs and BAMs which lack sensitivity at short sampling periods, the Dust Sentry reports accurately at 1 minute intervals and has a much lower total cost of ownership (up to 5x less). In addition, the Dust Sentry is compact, re-locatable and can integrate wind and weather sensors for maximum utility. 12kg Accessories OPTIONAL SHARP CUT CYCLONES REMOTE COMMUNICATION OPTIONS DS SC10: DS SC25: DS SC1: DS R39: DS R40: DS O12: PM10 Cyclone PM2.5 Cyclone PM1.0 Cyclone Gatetel GT-HE910 Modem Orbit Serial Port Extender 12 Month Orbit Data Access OTHER SENSORS AND MODULES 45 CONSUMABLES, PARTS AND TOOLS DS AUX: DS 420: DS R47: DS R1: DS R25: DS CM: DS R10: DS R31: DS R20B: DS R34: DS R56: CAL R4: Auxillary Module 4-20mA Module 47mm AnalysisFilter Gill WindSonic Sensor Vaisala WTX520 Weather Transmitter Control Module Sampling Pump PM Filters Flowmeters (x2) PM Purge Pump Zero Filter Flow Assembly Factory Calibration PM
  • 47. AQM60 Ambient Air Quality Monitor ●●Multiple gas sensors (up to six) ●●Wired or Wireless communication ●●Temperature and humidity sensors ●●Large data storage capacity (15 years) ●●Particulate monitor (PM2.5 or PM10) ●●Thermal management system ●●Wind speed and direction anemometer ●●Weatherproof and compact enclosures ●●Noise and meteorological sensors ●●Insulated roadside boxes, Metal enclosures, Mounting accessories are available ●●Rapid real-time data sampling (2-minute) Specifications Model AQM60 Aerosol Research Instruments Features: Standard: 50 to 77°F (10°C to 25°C) With Heater/Cooler: -4 to 113°F (-20°C to 45°C) 10 to 90 %RH (non condensing) Environment Operation Range Communications Options GSM modem / RF modem / Ethernet LAN Other Sensor Options Noise sensors / Particle counters / Meterological sensors Power Supply 12VDC / AC power module 100 - 250 VAC Enclosure Standard: Fiber reinforced polycarbonate Dimensions H 23.6” x W 19.7” x D 9.1” (600 x 500 x 230 mm) Weight 33 to 55 lb (15 to 25 kg), varies depending on configuration Gas Sensor Specifications Gas Sensor Heads Sensor Range (ppm) Accuracy Resolution Ozone GSS 0 - 0.150 +/- 0.005 ppm 0.001 ppm Power 2.9 W Nitrogen dioxide GSS 0 - 0.200 +/- 15% 0.001 ppm 12.3 W Carbon monoxide GSS 0 - 100 +/- 5 ppm 0.1 ppm 2.5 W Sulphur dioxide GSS 0 - 10 +/- 0.5 ppm 0.01 ppm 2.5 W Non-methane hydrocarbon GSS 0 - 25 +/- 0.5 ppm 0.1 ppm 2.5 W VOC isobutylene GSS 0 - 25 +/- 10% 0.1 ppm 2.5 W VOC isobutylene PID 0 - 20 10 % 0.01 ppm 1.5 W Hydrogen sulphide GSS 0 - 10 +/- 0.5 ppm 0.01 ppm 2.5 W Carbon dioxide NDIR 0 - 2,000 40 ppm + 3% 1 ppm 1.0 W Fluidized Particle Generator Kanomax Fluidized Particle Generators generate standard particles continuously for extended periods. Applications include pharmaceutical and chemical compound exposure research, and filtration performance testing. Features: ●●Available for variety of sample particles ●●Long time operation ●●Stably generated particle concentration ●●Easy control on the concentration level Specifications Model Particle Sizes Density Flow Rate Power Supply Dimensions Model 3211 3211 3216 Less than 20 µm 0.2 to 10 µm 100 mg/m3 to 1 g/m3 20 g/m3 30 L/min AC100V, 50/60Hz W 6.7” x H 17.5” x D17.7” W 20.1” x H 61.4” x D15.9” Model 3216 46
  • 48. Applications Mass Spectrometry Food analysis Forensic Science Medical Environmental analysis Mass Spectrometry 47
  • 49. Time of Flight Mass Spectrometer High performance and small foot print time of flight mass spectrometer by using multi-turn technology Mass Spectrometry Features: ●●High resolution and compact time of flight mass spectrometer ●●Impure substance analysis in gases with the detection limit of 10ppb ●●Onsite detailed accurate analysis Applications: ●●Food analysis ●●Particle analysis ●●Environmental analysis ●●Forensic: detection of forged paintings ●●Medical: blood test, hair analysis etc Specifications Resolution More than 30,000 Dynamic Range 10 bit Sampling Flow Rate 2GS/s Dimensions Weight H9.2” x W17.9” x D25.2” 79 lbs (36kg) with Gas chromatography (optional) Key Technology “Multi-Turn Time of Flight Mass Spectrometry” There are four orbiting electrodes and two injection/ ejection electrodes on a palm-sized optics bench. The ion source gives kinetic-energy for orbiting motion in the infinite loop. Injection and ejection electrodes are synchronized with ion source pulsing triggering. Injection electrode has to be in the on state while ions enter the analyzer, and then has to be turned off before first ion (smallest ion) returns to it. Orbiting electrodes are constant, so orbiting ion can be held until ejection electrode is ON. 48
  • 50. Automotive Testing Instruments Automotive Testing Instruments Amenity Manikin System Cabin Comfort Test Rig Kanomax Amenity Manikin System is a solution for precise interior cabin comfort evaluation. The system measures parameters; air velocity, temperature, humidity, and radiant heat; which relate to human comfort level. One mannequin equips more than 120 sensors all over its surface and provides sophisticated measurement. Features: ●●4 mannequins measure simultaneously; understanding the entire cabin ●●Wireless connection brings easy operation ●●Graphical software for both real-time measurement and data retrieval ●●Excel compatible data output as well as saved graphical data for review Amenity Manikin Sensor Allocations by Part Head Torso Lower Body Total Air Velocity 4 12 20 36 Temperature 12 33 33 78 Humidity 1 0 1 2 Radiant Heat 3 5 4 12 Mannequin Specifications Height Weight Material Approx. 5’6” (170 cm) 90 lbs (40 kg) Plastic (FRP) Sensor Specifications Air Velocity Ranges Accuracy Temperature Ranges Accuracy Humidity Ranges Accuracy Radiant Heat Ranges Accuracy Wavelength 49 0.10 to 5.00 m/s 0.05 m/s 30 to 100°C +/- 3.0°C 3 to 95% RH +/- 3% RH 0 to kw/m2 +/- 7% 0.3 to 40 µm Measurement Software
  • 51. Automotive Testing Instruments Vehicle Cabin Leakage Testers Cabin Comfort Test Rig Leakage testing is performed by pressurizing or depressurizing the vehicle cabin. The tester measures the changes in cabin pressure. The control unit consists of a manometer and pressure transducers. The manometer detects the leakage flow, which is calculated by measuring the nozzle pressure relative to static pressure. The tester controls an adjustable fan to maintain static test pressure. Five selectable inlet nozzles provide reasonable leakage flow rate measurements of 2.5 to 800 cfm. CLT-APC-NPP 1. The cab is pressurized (either positively or negatively) 2. Pressure will be stable, normally stable conditions are achieved quickly 3. The leakage rate is measured Features: ●●Automatic Pressure Control ●●System includes inlet nozzles, certificate of conformance ●●Negative Pressure Testing ●●Features High Accuracy Manometer Specifications Model CLT-MPC to ISO 5801: 1997, and pressure measurement calibration certificate Nozzle Specifications CLT-MPC Fan Speed Pressure Measurement Accuracy Leakage Flow Measurement Accuracy Cabin Pressure Preset Ranges Automatic Pressure Control Negative Pressure Plenum Small Transport Cart Power Supply Duct Size Dimensions CLT-MPC-NPP CLT-APC CLT-APC-NPP 847 cfm (400 L/sec) +/- 2 % of reading +/- 1 digit +/- 3 % of reading +/- 1 digit Providing flow rate is varied by the nozzle * See nozzle specs 0.2 to 2 in. H2O ― (5 to 499 Pa) ― ⃝ ― ⃝ ― ⃝ ⃝ ― 120/240 V, 1 phase 16.4 ft x 8 inches (5 m x 203 mm) Small Transport Cart: W 22.5” x H 23.5” x D 54” Large Transport Cart: W 27.5” x H 42” x D 58.5” Nozzle F G H D E Inlet Diameter 0.590” (15 mm) 1.122” (28.5 mm) 2.204” (56 mm) 3.543” (90 mm) 6.023” (153mm) Flow Ranges 2.5 to 10 cfm (1.2 to 4.8 l/s) 8.5 to 36 cfm (4 to 17 l/s) 32 to 142 cfm (15 to 67 l/s) 53 to 296 cfm (25 to 140 l/s) 190 to 847 cfm (90 to 400 l/s) Smoke Generator Model 8304 Smoke Generator for Air Flow Visualization The Model 8304 is a smoke generator for wind tunnel testing and aerodynamics visualization. It produces vaporizing white smoke of 15 to 80L/min. Compressed air moves smoke to the comb-shaped nozzle and makes streamlines. Features: ●●Versatile and Portable unit ●●High output, non-toxic, water-soluble smoke ●●System cart, air compressor, and comb-shaped nozzle are available as options Nozzle Specifications Model Smoke Output Smoke Particle Size Warm Up Time Power Supply Dimensions Weight 8304 15 to 80 L/min 0.3 to 1.0 µm 4 minutes AC100V, 550W W11.8” x H13.8” x D25.6” 61.7 lbs (28 kg) 50