CT-to-TPS-to-verification
summary of simulation process
Vibha Chaswal, Ph.D.
Know the co-ordinate system TPS
uses

Patient co-ordinate system (not
used in Pinnacle)

CT co-ordinate system
Image data on acquisition on CT
• Image request form: Patient positioned using
lap lasers (ct bb’s on lap lasers)
• Immobilization devices marked
• Orientation and technique (protocol) selected
in CT
• Topogram acquisition
• CT scan acquisition
• Dicom images send to TPS and archive
TPS processes
• Image data imported
• Scanning information verified ( patient position,
orientation)
• CT bb’s are localized w.r.t. CT room lasers co-ordinate
system
• ROIs are contoured and isocenter is defined w.r.t machine
co-ordinate system
• Shifts from ct-bb’s to isocenter are given w.r.t treatment
table (up/down, L/R, in/out)
• Segmentation, TP (beams, blocks, dose) and DRRs
generated (set-up and field)
• Transfer TP to record and verify (LEX), DRRs to machine
(coherence)
Verification processes
•
•
•
•

Set patient to CT-bb’s
Get shifts to isocenter from DRR’s, make shifts
Check SSD
Take portal image using field DRRs or Set-up
DRRs (orthogonal and oblique DRRs)
• Compare DRR with portal image and verify shift
(orthogonal gives sup-inf and left-right info
oblique gives A-P)
• Mark lasers on patient if everything matches,
these are treatment isocenter localization marks
• Bite-block – physicist verifies the physical
positioning of bite-block before CT scan,
radiocam localization no ct bb’s
• Cone-beam – Cone beam field with CT image
and segmented structures sent to machine,
MVCT image compared with the help of bony
land-marks and contours
Thank You!
Thank You!

Contenu connexe

Tendances

Treatment verification and set up errors
Treatment verification and set up errorsTreatment verification and set up errors
Treatment verification and set up errorssailakshmi pullookkara
 
Beam Modification in Radiotherapy
Beam Modification in RadiotherapyBeam Modification in Radiotherapy
Beam Modification in RadiotherapySantam Chakraborty
 
Role of immobilisation and devices in radiotherapy
Role of immobilisation and devices in radiotherapyRole of immobilisation and devices in radiotherapy
Role of immobilisation and devices in radiotherapySwarnita Sahu
 
Beam modification-in-radiotherapy-
Beam modification-in-radiotherapy- Beam modification-in-radiotherapy-
Beam modification-in-radiotherapy- Sathish Kumar
 
Evaluation of radiotherapy treatment planning
Evaluation of radiotherapy treatment planningEvaluation of radiotherapy treatment planning
Evaluation of radiotherapy treatment planningAmin Amin
 
Isodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYIsodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYPaul George
 
Treatment Planning Ii Patient Data, Corrections, And Set Up
Treatment Planning Ii Patient Data, Corrections, And Set UpTreatment Planning Ii Patient Data, Corrections, And Set Up
Treatment Planning Ii Patient Data, Corrections, And Set Upfondas vakalis
 
Principles of beam direction and use of simulators
Principles of beam direction and use of simulators Principles of beam direction and use of simulators
Principles of beam direction and use of simulators Anil Gupta
 
planning systems in radiotherapy
 planning systems in radiotherapy planning systems in radiotherapy
planning systems in radiotherapyfondas vakalis
 
PARTICLE BEAM RADIOTHERAPY
PARTICLE BEAM RADIOTHERAPYPARTICLE BEAM RADIOTHERAPY
PARTICLE BEAM RADIOTHERAPYKanhu Charan
 
Beam modification devices
Beam modification devicesBeam modification devices
Beam modification devicesVibhay Pareek
 

Tendances (20)

Treatment verification and set up errors
Treatment verification and set up errorsTreatment verification and set up errors
Treatment verification and set up errors
 
Helical Tomotherapy
Helical TomotherapyHelical Tomotherapy
Helical Tomotherapy
 
Flattening filter Free
Flattening filter FreeFlattening filter Free
Flattening filter Free
 
Beam Modification in Radiotherapy
Beam Modification in RadiotherapyBeam Modification in Radiotherapy
Beam Modification in Radiotherapy
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
Role of immobilisation and devices in radiotherapy
Role of immobilisation and devices in radiotherapyRole of immobilisation and devices in radiotherapy
Role of immobilisation and devices in radiotherapy
 
Cobalt & linac
Cobalt & linacCobalt & linac
Cobalt & linac
 
Beam modification-in-radiotherapy-
Beam modification-in-radiotherapy- Beam modification-in-radiotherapy-
Beam modification-in-radiotherapy-
 
Icru 58.
Icru 58.Icru 58.
Icru 58.
 
Evaluation of radiotherapy treatment planning
Evaluation of radiotherapy treatment planningEvaluation of radiotherapy treatment planning
Evaluation of radiotherapy treatment planning
 
Isodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYIsodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGY
 
Treatment Planning Ii Patient Data, Corrections, And Set Up
Treatment Planning Ii Patient Data, Corrections, And Set UpTreatment Planning Ii Patient Data, Corrections, And Set Up
Treatment Planning Ii Patient Data, Corrections, And Set Up
 
Principles of beam direction and use of simulators
Principles of beam direction and use of simulators Principles of beam direction and use of simulators
Principles of beam direction and use of simulators
 
planning systems in radiotherapy
 planning systems in radiotherapy planning systems in radiotherapy
planning systems in radiotherapy
 
Treatment plannings i kiran
Treatment plannings i   kiranTreatment plannings i   kiran
Treatment plannings i kiran
 
PARTICLE BEAM RADIOTHERAPY
PARTICLE BEAM RADIOTHERAPYPARTICLE BEAM RADIOTHERAPY
PARTICLE BEAM RADIOTHERAPY
 
Beam modification devices
Beam modification devicesBeam modification devices
Beam modification devices
 
Srs and srt
Srs and srtSrs and srt
Srs and srt
 
Total body irradiation
Total body irradiationTotal body irradiation
Total body irradiation
 
5 rs radiotherapy
5 rs radiotherapy5 rs radiotherapy
5 rs radiotherapy
 

En vedette

importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapymakhhi
 
Immobilization in radiotherapy
Immobilization in radiotherapy Immobilization in radiotherapy
Immobilization in radiotherapy prabhurt
 
Immobilization in Radiotherapy-Quality assurance
Immobilization in Radiotherapy-Quality assuranceImmobilization in Radiotherapy-Quality assurance
Immobilization in Radiotherapy-Quality assuranceVishvendra Gaur
 
Radiation therapy
Radiation therapyRadiation therapy
Radiation therapyRad Tech
 
beam modifying devises
beam modifying devisesbeam modifying devises
beam modifying devisessaikishore15
 
Intensity Modulated Radiation Therapy (IMRT)
Intensity Modulated Radiation Therapy (IMRT)Intensity Modulated Radiation Therapy (IMRT)
Intensity Modulated Radiation Therapy (IMRT)Dilshad KP
 
Sad calculations
Sad calculationsSad calculations
Sad calculationsCSULB
 
Linear accelator
Linear accelatorLinear accelator
Linear accelatorSUST\RICK
 
Dosimetric calculations
Dosimetric calculationsDosimetric calculations
Dosimetric calculationsCSULB
 
Radiation Therapy
Radiation TherapyRadiation Therapy
Radiation Therapykathrnrt
 
IMRT: Intensity Modulated Radiotherapy
IMRT: Intensity Modulated RadiotherapyIMRT: Intensity Modulated Radiotherapy
IMRT: Intensity Modulated RadiotherapyShatha M
 
Radiation therapy patient treatment planning & post treatment care new/ Labia...
Radiation therapy patient treatment planning & post treatment care new/ Labia...Radiation therapy patient treatment planning & post treatment care new/ Labia...
Radiation therapy patient treatment planning & post treatment care new/ Labia...Indian dental academy
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapyjyotimannath
 

En vedette (20)

importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapy
 
Immobilization in radiotherapy
Immobilization in radiotherapy Immobilization in radiotherapy
Immobilization in radiotherapy
 
Immobilization in Radiotherapy-Quality assurance
Immobilization in Radiotherapy-Quality assuranceImmobilization in Radiotherapy-Quality assurance
Immobilization in Radiotherapy-Quality assurance
 
Imobilazation Device
Imobilazation DeviceImobilazation Device
Imobilazation Device
 
Immobilizations devices
Immobilizations devicesImmobilizations devices
Immobilizations devices
 
Beam modification in radiotherapy
Beam modification in radiotherapyBeam modification in radiotherapy
Beam modification in radiotherapy
 
Radiation therapy
Radiation therapyRadiation therapy
Radiation therapy
 
beam modifying devises
beam modifying devisesbeam modifying devises
beam modifying devises
 
Intensity Modulated Radiation Therapy (IMRT)
Intensity Modulated Radiation Therapy (IMRT)Intensity Modulated Radiation Therapy (IMRT)
Intensity Modulated Radiation Therapy (IMRT)
 
CT
CTCT
CT
 
Sad calculations
Sad calculationsSad calculations
Sad calculations
 
Linear accelator
Linear accelatorLinear accelator
Linear accelator
 
Dosimetric calculations
Dosimetric calculationsDosimetric calculations
Dosimetric calculations
 
Linear Accelerator
Linear AcceleratorLinear Accelerator
Linear Accelerator
 
Radiation Therapy
Radiation TherapyRadiation Therapy
Radiation Therapy
 
IMRT: Intensity Modulated Radiotherapy
IMRT: Intensity Modulated RadiotherapyIMRT: Intensity Modulated Radiotherapy
IMRT: Intensity Modulated Radiotherapy
 
Radiation therapy patient treatment planning & post treatment care new/ Labia...
Radiation therapy patient treatment planning & post treatment care new/ Labia...Radiation therapy patient treatment planning & post treatment care new/ Labia...
Radiation therapy patient treatment planning & post treatment care new/ Labia...
 
Beam Direction
Beam DirectionBeam Direction
Beam Direction
 
Beam modifications
Beam modificationsBeam modifications
Beam modifications
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
 

Similaire à CT-TPS-Verification Simulation Process

Immobilization techniques in SRS and SBRT
Immobilization techniques in SRS and SBRTImmobilization techniques in SRS and SBRT
Immobilization techniques in SRS and SBRTShreya Singh
 
Commissionin of medical linear accelerator
Commissionin of medical linear acceleratorCommissionin of medical linear accelerator
Commissionin of medical linear acceleratorRam Pukar Bharat
 
Robotics and navigation in Orthopaedic surgery - Dr. Sachin M
Robotics and navigation in Orthopaedic surgery - Dr. Sachin MRobotics and navigation in Orthopaedic surgery - Dr. Sachin M
Robotics and navigation in Orthopaedic surgery - Dr. Sachin MSachinMalayaiah1
 
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...Abhishek Soni
 
Image registraiton in Radiation Therapy
Image registraiton in Radiation TherapyImage registraiton in Radiation Therapy
Image registraiton in Radiation TherapyTalluri Anil Kumar
 
How high frequency ultrasound imaging is supporting preclinical research appl...
How high frequency ultrasound imaging is supporting preclinical research appl...How high frequency ultrasound imaging is supporting preclinical research appl...
How high frequency ultrasound imaging is supporting preclinical research appl...Scintica Instrumentation
 
CT image acquisition
CT image acquisitionCT image acquisition
CT image acquisitiondypradio
 
Intensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
Intensity Modulated Radiotherapy (IMRT) - Dr. S. SachinIntensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
Intensity Modulated Radiotherapy (IMRT) - Dr. S. SachinSACHINS700327
 
importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapymakhhi
 
CT perfusion physics and its application in Neuroimaging
CT perfusion physics and its application in NeuroimagingCT perfusion physics and its application in Neuroimaging
CT perfusion physics and its application in NeuroimagingDr.Suhas Basavaiah
 
Components of CT Scan Machine
Components of CT Scan MachineComponents of CT Scan Machine
Components of CT Scan MachineSyed Jamal
 
Motion Management in Lung Cancer Radiotherapy
Motion Management in Lung Cancer RadiotherapyMotion Management in Lung Cancer Radiotherapy
Motion Management in Lung Cancer RadiotherapyJyotirup Goswami
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdfmitians
 
Cyberknife Treatment Delivery Radiation Therapist Perspective
Cyberknife Treatment Delivery  Radiation Therapist PerspectiveCyberknife Treatment Delivery  Radiation Therapist Perspective
Cyberknife Treatment Delivery Radiation Therapist PerspectiveSubrata Roy
 
Computed tomography - Physics and Clinical
Computed tomography - Physics and ClinicalComputed tomography - Physics and Clinical
Computed tomography - Physics and ClinicalSambasivaselli R
 
Image guided radiation therapy
Image guided radiation therapyImage guided radiation therapy
Image guided radiation therapySwarnita Sahu
 
Optimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detectionOptimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detectionWookjin Choi
 

Similaire à CT-TPS-Verification Simulation Process (20)

Immobilization techniques in SRS and SBRT
Immobilization techniques in SRS and SBRTImmobilization techniques in SRS and SBRT
Immobilization techniques in SRS and SBRT
 
Commissionin of medical linear accelerator
Commissionin of medical linear acceleratorCommissionin of medical linear accelerator
Commissionin of medical linear accelerator
 
Robotics and navigation in Orthopaedic surgery - Dr. Sachin M
Robotics and navigation in Orthopaedic surgery - Dr. Sachin MRobotics and navigation in Orthopaedic surgery - Dr. Sachin M
Robotics and navigation in Orthopaedic surgery - Dr. Sachin M
 
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
 
Rt
RtRt
Rt
 
Image registraiton in Radiation Therapy
Image registraiton in Radiation TherapyImage registraiton in Radiation Therapy
Image registraiton in Radiation Therapy
 
How high frequency ultrasound imaging is supporting preclinical research appl...
How high frequency ultrasound imaging is supporting preclinical research appl...How high frequency ultrasound imaging is supporting preclinical research appl...
How high frequency ultrasound imaging is supporting preclinical research appl...
 
CT image acquisition
CT image acquisitionCT image acquisition
CT image acquisition
 
Intensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
Intensity Modulated Radiotherapy (IMRT) - Dr. S. SachinIntensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
Intensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
 
4.pptx
4.pptx4.pptx
4.pptx
 
importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapy
 
CT perfusion physics and its application in Neuroimaging
CT perfusion physics and its application in NeuroimagingCT perfusion physics and its application in Neuroimaging
CT perfusion physics and its application in Neuroimaging
 
Components of CT Scan Machine
Components of CT Scan MachineComponents of CT Scan Machine
Components of CT Scan Machine
 
Motion Management in Lung Cancer Radiotherapy
Motion Management in Lung Cancer RadiotherapyMotion Management in Lung Cancer Radiotherapy
Motion Management in Lung Cancer Radiotherapy
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdf
 
CT COMPONENTS
CT COMPONENTSCT COMPONENTS
CT COMPONENTS
 
Cyberknife Treatment Delivery Radiation Therapist Perspective
Cyberknife Treatment Delivery  Radiation Therapist PerspectiveCyberknife Treatment Delivery  Radiation Therapist Perspective
Cyberknife Treatment Delivery Radiation Therapist Perspective
 
Computed tomography - Physics and Clinical
Computed tomography - Physics and ClinicalComputed tomography - Physics and Clinical
Computed tomography - Physics and Clinical
 
Image guided radiation therapy
Image guided radiation therapyImage guided radiation therapy
Image guided radiation therapy
 
Optimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detectionOptimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detection
 

Plus de Miami Cancer Institute

Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryMiami Cancer Institute
 
IGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureIGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureMiami Cancer Institute
 
Ion chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and EvaluationIon chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and EvaluationMiami Cancer Institute
 
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QAArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QAMiami Cancer Institute
 
Linear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QALinear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QAMiami Cancer Institute
 
Special Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORTSpecial Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORTMiami Cancer Institute
 
TPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyTPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyMiami Cancer Institute
 
Mutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registrationMutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registrationMiami Cancer Institute
 
Distortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correctionDistortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correctionMiami Cancer Institute
 
Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationMiami Cancer Institute
 
TG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit CalculationsTG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit CalculationsMiami Cancer Institute
 

Plus de Miami Cancer Institute (16)

Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and Delivery
 
IGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureIGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing Procedure
 
Ion chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and EvaluationIon chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and Evaluation
 
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QAArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
 
Linear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QALinear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QA
 
Shielding Design
Shielding DesignShielding Design
Shielding Design
 
Stereotactic Radiosurgery at UIHC
Stereotactic Radiosurgery at UIHCStereotactic Radiosurgery at UIHC
Stereotactic Radiosurgery at UIHC
 
Special Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORTSpecial Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORT
 
TPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyTPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation Oncology
 
Image Contrast, Noise, Resolution
Image Contrast, Noise, ResolutionImage Contrast, Noise, Resolution
Image Contrast, Noise, Resolution
 
Mutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registrationMutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registration
 
Causes of Noise in PET imaging
Causes of Noise in PET imagingCauses of Noise in PET imaging
Causes of Noise in PET imaging
 
Distortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correctionDistortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correction
 
Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verification
 
Clinical Reference Dosimetry
Clinical Reference DosimetryClinical Reference Dosimetry
Clinical Reference Dosimetry
 
TG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit CalculationsTG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit Calculations
 

Dernier

Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...narwatsonia7
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...narwatsonia7
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 

Dernier (20)

Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
 

CT-TPS-Verification Simulation Process

  • 1. CT-to-TPS-to-verification summary of simulation process Vibha Chaswal, Ph.D.
  • 2. Know the co-ordinate system TPS uses Patient co-ordinate system (not used in Pinnacle) CT co-ordinate system
  • 3.
  • 4. Image data on acquisition on CT • Image request form: Patient positioned using lap lasers (ct bb’s on lap lasers) • Immobilization devices marked • Orientation and technique (protocol) selected in CT • Topogram acquisition • CT scan acquisition • Dicom images send to TPS and archive
  • 5. TPS processes • Image data imported • Scanning information verified ( patient position, orientation) • CT bb’s are localized w.r.t. CT room lasers co-ordinate system • ROIs are contoured and isocenter is defined w.r.t machine co-ordinate system • Shifts from ct-bb’s to isocenter are given w.r.t treatment table (up/down, L/R, in/out) • Segmentation, TP (beams, blocks, dose) and DRRs generated (set-up and field) • Transfer TP to record and verify (LEX), DRRs to machine (coherence)
  • 6. Verification processes • • • • Set patient to CT-bb’s Get shifts to isocenter from DRR’s, make shifts Check SSD Take portal image using field DRRs or Set-up DRRs (orthogonal and oblique DRRs) • Compare DRR with portal image and verify shift (orthogonal gives sup-inf and left-right info oblique gives A-P) • Mark lasers on patient if everything matches, these are treatment isocenter localization marks
  • 7.
  • 8. • Bite-block – physicist verifies the physical positioning of bite-block before CT scan, radiocam localization no ct bb’s • Cone-beam – Cone beam field with CT image and segmented structures sent to machine, MVCT image compared with the help of bony land-marks and contours