1. Medical Equipment III
X-ray Imaging
Shereen M. El-Metwally
Associate Professor,
Systems and Biomedical Engineering Department,
Faculty of Engineering - Cairo University
sh.elmetwally@eng1.cu.edu.eg
4. Reference Books:
Russell K. Hobbie, Bradley J. Roth,
"Intermediate Physics and Biology (Chapter
14, 15), Springer, 2007."
N. Smith et. Al., “Introduction to Medical
Imaging Physics, Engineering and Clinical
Applications”, 2011.
P. Mayles et. Al., “Handbook of
Radiotherapy Physics: Theory and
Practice”, 2007.
6. Light travels in vacuum with a velocity c = 3
×108 m/s
When light travels through matter, its speed is
less than this and is given by
n is index of refraction of substance
depends on both the composition of substance and
color of light
n
c
cn
Nature of light
Light as an electromagnetic wave:
7. A traveling wave of light can be described by:
f(x − cnt)
represents a disturbance traveling along the x-axis in the
positive direction
As a sinusoidal wave, the period T, frequency v , and
wavelength λ, are related by
As light moves from one medium to another,
wavelength changes as the speed changes, while
frequency remains the same.
nc
T
,
1
Nature of light
Light as an electromagnetic wave:
9. Nature of light
Particle concept
Each particle of light or “photon” has an energy E
given by:
where h is the Planck’s constant
h= 6.63 x 10-34 Js = 4.14x10-15 eVs
10. Atomic Energy Levels and Atomic
Spectra Energy levels:
Isolated atoms have specific discrete internal energies.
The internal energy of the atom is the sum of the energies
of each electron.
Photon emission or absorption:
Let the energy of the ith state be labeled Ei.
If Ei is greater than the lowest possible energy,
then the atom can lose energy by emitting a photon of energy (Ei
−Ef ) and exist in a lower energy
state Ef
At the lowest possible internal energy: in this state no
further energy loss can take place.
ph i fE E E
11. Atomic Energy Levels and Atomic
Spectra
Ionization energy is the smallest amount of energy
required to remove an electron from the atom when
the atom is in its ground state.
Radiatio
n
Ionizing
Non-
Ionizing
Direct
Indirec
t
Microwaves
Radiofrequen
cy
Ultrasound
Infrared
13. Photon Interaction with
matter
Photons in a vacuum travel in a straight line.
When they travel through matter
Passing through (Transmission)
Scattering
Absorption
Loss of photons due to scattering and absorption is
called “attenuation”.
14. Photon Interaction with matter
Beer-Lambert law:
The number of photons transmitted decreases with
distance:
( )
s
a
( )
where : total linear attenuation coefficient
:linear scattering coefficient
:linear absorption coefficient
z:distance
s a zz
o oN z N e N e
15. Imaging Tool: Full Definition
Effect produced
Interaction with the body
Effect to be detected
Image construction
17. X-ray Imaging
The tube – X-rays are produced
The body – X-rays interact with the body
The image – X-rays interact with film, detectors
Film processing, Signal analysis
X-ray Tube
Body
Detectors
Film
processing,
Signal
analysis
18. X-rays
Made of photons
Travel at speed of light
Travel in a straight line
Has no mass nor charge (cannot be focused by
magnets)
X-ray beam has a mix of energies
Diagnostic X-ray range 20-150 keV
19. Diagnostic and Therapeutic X-
Ray
Diagnostic radiology
Uses low energy X-rays for imaging
No tissue damage.
Radiotherapy
Uses high energy X-rays to treat tumor.
Ionizes the water in tumor cells and induces the
formation of free radicals which may cause damage of
genetic material (DNA).
Normal cells are also affected adversely by radiation but
have the capability to repair.
20. The set-up for planar
radiography
The basis of X-ray imaging is the differential absorption of X-rays
by various tissues
20
21. X-ray Imaging
The tube – X-rays are produced
The body – X-rays interact with the body
The image – X-rays interact with film, detectors
Film processing, Signal analysis
X-ray Tube
Body
Detectors
Film
processing,
Signal
analysis
24. X-ray production
A current of a few amperes heats the filament. Electrons
are liberated at a rate that increases with the filament
current. This is called “thermionic emission”.
The released electrons are accelerated across a high
voltage onto a target.
X-rays are produced as the electrons interact in the
target.
Vacuum is maintained inside the glass envelope of the
x-ray tube to prevent the electrons from interacting with
gas molecules.
24
25. X-ray production
Push the “rotor” or
“prep” button
Charges the filament –
causes thermionic emission
(e- cloud)
Begins rotating the anode.
Push the “exposure” or
“x-ray” button
Accelerating voltage is
applied, so e-’s move
towards anode target to
produce x-rays
26. Focal spot size and Coverage
calculation The “focal spot” is the volume of target within which electrons
are absorbed and x-rays are produced.
Most diagnostic x-ray tubes use a target angle between 6 and 17
degrees wrt the e- beam direction. The smaller the angle, the
smaller the effective focal spot size (f):
where θ is the target angle and
F is the width of the electron beam.
Coverage:
27. Focal spot size
A finite effective spot size
as well as the tube–
patient (S0) and tube–
detector (S1) distances
determine the spatial
resolution of the image.
To improve the image
spatial resolution, S0 should
be as large, and the value of
f as small, as possible, with
the patient placed directly
on top of the detector.
where P is size of the
‘penumbra’ region, or
geometric unsharpness. 27
28. Dual-focus tubes
Many X-ray tubes contain two
cathode filaments of different
lengths, called “dual-focus”
tubes.
A “small” or “fine” filament
produces a narrower e-beam, so
achieves a small focal spot for
radiographs of greater detail.
A “larger” or “coarse” filament
produces a wider e-beam. Used
for x-ray exposures of high
intensity and short duration in
order to limit the blurring effectsCathode assembly of a dual-focus x-ray tu
29. Energy Spectrum of X-ray
Many of the very low energy X-rays
are absorbed by the housing of the X-
ray tube itself.
The distribution of photon energies
produced by a typical x-ray tube is
referred to as an emission
spectrum. It is composed of:
A wide spectrum of X-ray
energies, termed ‘general
radiation’ or ‘bremsstrahlung’
(braking radiation in German).
Sharp peaks whose energy is
characteristic of the particular
target material, hence the name
‘characteristic radiation’.
30. Anode Heel Effect
The X-ray beam has a higher
intensity at the ‘cathode-end’ than at
the ‘anode-end’, a phenomenon
known as the “Anode Heel” effect.
This effect is due to X-rays on the
anode side have to travel further
through the anode itself before
leaving the tube, and are therefore
more highly attenuated.
31. Exposure Factors:
Accelerating voltage (kVp)
Tube current (mA)
Exposure time (s)
mAs – product of mA and s
Exposure factors are set by
radiographer
Notes de l'éditeur
X rays and gamma rays have high frequencies.
The internal energy of the atom is the sum of the energies of each electron.
The energy of each electron depends on the values of five quantum numbers: n, l, s, nl, ns
An atom can change from one energy level to another by emitting or absorbing a photon with an energy equal to the energy difference between the two levels
Ionizing radiation as radiation from high-energy photons (X and gamma rays), and charged particles.
Or, Ionization energy: amount of energy required to remove the least-bound electron from the atom.
n is the shell no. or principal quantum no.
Units of μ is m-1
Bones attenuate X-rays to a much greater degree than the soft tissue of the lungs, and appear brighter on the image.
(a) The set-up for planar radiography.
(b) An anti-scatter grid is placed directly in front of the detector to reduce the contribution from scattered X-rays.
(c) An example of a chest radiograph: there is far greater attenuation in bone than in tissue, and excellent tissue contrast.
The x-rays emerge from the target in all directions but are restricted by collimators to form a useful beam of x rays.
The size of the ‘penumbra’ region, or geometric unsharpness, is denoted P.
The two cathode filaments are of different lengths, each with a focusing cup, to produce a wider or narrower e-beam depending upon the particular application.
This means that the signal intensity on one side of a planar radiograph is different from that on the other.