HEMODYNAMIC
MONITORING
By
Yasmeen Rahim
OBJECTIVES
By the end of this session, learners will be
able to:
 Define what is hemodynamic monitoring.
 List down the importance of hemodynamic
monitoring in hospital.
 Discuss about pressure monitoring system.
 Identify troubleshooting in the pressure
monitoring system.
CONT.
 Explain in detail about arterial pressure
monitoring, central venous pressure
monitoring and pulmonary artery pressure
monitoring.
 Evaluate oxygen delivery and demand.
HEMODYNAMIC MONOTORING
 It is the study of interrelationship of blood
pressure, blood flow, vascular volumes,
heart rate, ventricular function and the
physical properties of blood at bedside
 Is an integral part of critical care nursing
Cont..
 CCN should have knowledge of how to
obtain accurate data, analyze waveforms
and interpret and integrate the data.
 The information provided by invasive
catheters can give us accurate and timely
information to clinicians so that appropriate
interventions can be taken.
IMPORTANCE OF HDM
 Early detection, identification, and treatment
of life-threatening conditions such as heart
failure and cardiac temponade.
 Evaluate the patient’s immediate response to
treatment such as drugs and mechanical
support.
 Evaluate the effectiveness of cardiovascular
function such as cardiac output and index.
PURPOSE OF HDM:
 The purpose of HDM is to:
 Early detection, identification, and treatment of
life-threatening conditions
 Aid in the diagnosis of various cardiovascular
disorders
 Guide therapies to optimize cardiac functions
 Minimize cardiovascular dysfunctions
 Evaluate the patient’s immediate response to
treatment modalities
INDICATION OF HDM
 It is essential in conditions when cardiac output is
insufficient to deliver oxygen to the cells (altered
preload, afterload and contractility).
 Any deficit or loss of cardiac function: such as AMI,
CHF, Cardiomyopathy.
 All types of shock; cardiogenic, septic, neurogenic or
anaphylactic.
 Dehydration, hemorrhage, G.I. bleed, severe burn,
ARDS or any major surgery.
 Severe sepsis or multiple organ failure.
 It aids in assessing body oxygen supply and demand.
HDM PARAMETERS:
 Cardiac output.
 Heart rate
 Rhythm
 Cardiac index.
 Stroke volume.
 Preload
 Afterload
 Contractility
PRESSURE MONITORING
SYSTEM
 Catheter (hollow tube).
 Central venous catheter for CVP
 Arterial catheter for Arterial pressure monitoring
 Pulmonary catheter / SWAN Ganz for PAP monitoring.
 Pressure tubing.
 Flush solution (N/S or D5W, heparinized)
 IV tubing with drip chamber
 Pressure bag (pressure of 300 mmHg) and stopcock
 Pressure transducer.
 Pressure amplifier.
 Cardiac monitor
SQUARE WAVE TEST:
 Square wave test is performed to quickly assess
the response of the system.
 Flush device is released rapidly which will
increase fluid flow through the system.
 Record the square wave formed on the monitor.
LEVELING & ZEROING:
 A fundamental step in obtaining accurate
hemodynamic values is to zero the transducer
amplifier system.
 This step is performed at least once before
obtaining the first hemodynamic reading after
catheter insertion.
 The pressure monitoring system is leveled to an
external landmark and then zeroed to
atmospheric pressure.
 Zeroing is done to eliminate the effect of
atmospheric and hydrostatic pressure.
Cont..
 The phlebostatic axis is used as the reference
point for leveling and zeroing.
 The site of the phlebostatic axis is at the
intersection of the fourth intercostal space and
mid axillary line (approx level of right atrium and
pulmonary artery.
 Leveling (referencing) and zeroing ensures that
hemodynamic values obtained with the catheter
are accurate.
STEPS FOR ZEROING AND
LEVELING:
 Position patient on their back
 Patient may be positioned with the head of the bed elevated
between 0-60°
 Flush the system
 Level transducer to phlebostatic axis (may mark this with an x
on patient)
 Turn stop-cock on transducer so that it is off to the patient.
 Remove cap
 Press zero on the module
 Ensure that zero appears on screen replace cap and turn
stop-cock so that it is open to monitoring and patient.
ARTERIAL PRESSURE
MONITORING:
 Arterial blood pressure is a basic
hemodynamic index often utilized to guide
therapeutic interventions
 Continuous monitoring of blood pressure is
indicated for patients with hemodynamic
instability that requires inotropic or
vasopressor medication
Cont.
 An arterial line is a cannula usually
positioned in a peripheral artery such as
 Radial artery (most commonly used site)
 Brachial artery (rarely used)
 Dorsalis pedis artery (rarely used)
 Femoral artery (second option other than radial
artery, more chances of getting contaminated)
Cont..
 An arterial line allows for consistent and
continuous monitoring of blood pressure to
facilitate the reliable titration of supportive
medications
 In addition, arterial lines allow for reliable
access to the arterial circulation for the
measurement of arterial oxygenation and
for frequent blood sampling
ARTERIAL LINE INSERTION:
 http://content.nejm.org/cgi/video/354/15/e13/
ARTERIAL PRESSURE
 Arterial pressure waveform has three parts:
 Rapid upstroke wave (systolic pressure)
 Dicrotic notch
 Diastolic pressure waveform
 Normal systolic BP is 90 – 140 mm Hg.
 Normal diastolic BP is 60 – 90 mm Hg.
 Dicrotic notch is small downward deflection
following the closure of semilunar valve, indicate
the end of systole and beginning of diastole.
MEAN ARTERIAL PRESSURE:
 Mean arterial pressure (MAP) is used to
evaluate perfusion of vital body organs.
 Normal MAP is 70 to 105 mm Hg.
 MAP can be calculated by:
Systolic BP + (Diastolic *2) / 3
EJECTION FRACTION
 Ejection Fraction (EF) is the fraction of
blood ejected by the ventricle relative to its
end-diastolic volume.
 EF is calculated from:
EF = (SV / EDV) * 100
 For example if SV is 75 and EDV is 120
then EF would be 63%.
POTENTIAL COMPLICATIONS
 HAEMORRHAGE:
PREVENTION:
 Keep limb visible at all times
 Ensure alarm is on so that any accidental disconnection
can be dealt quickly
 Ensure that arm is immobile with arm board
 Ensure all connections are tight
SOLUTION
 Apply pressure to limb
 Assess leak
 If hemorrhage persists notify medical officer
CONT..
 INFECTION
PREVENTION
 Assess area regularly for redness or swelling
 Avoid interrupting circuit as much as possible
 Use gloves when touching arterial line
SOLUTION
 Remove arterial Line
 Ensure proper hand washing when handling
arterial line or transducer
CONT..
 BLOCKAGE, CLOTTING & AIR EMBOLI
PREVENTION
 Keep pressure bag inflated to 300 mmHg
 ensure 3-5ml auto flush is continuous.
 Attempt to aspirate blood
 Use fast flush device to clear line to prevent clot
formation
SOLUTION
 Attempt to aspirate blood to remove clot
 Ensure all connections are secure
CONT..
 INTERUPTION TO PERIPHERAL
CIRCULATION
PREVENTION
 Regularly check distal pulses and capillary refill.
SOLUTION
 Notify MO and consider removing arterial line
NURSING CONSIDERATIONS
 Nursing care mainly directed to preventing
complications
 Ensure that the insertion site is visible at all
times for early detection of disconnection
 All connections must be secured
 Ensure that the cannula site is covered with
an appropriate dressing to maintain asepsis
Cont.
 Never inject anything into an arterial cannula
or arterial line
 Ensure that the flush bag has adequate fluid.
 Use only heparinized 0.9% sodium chloride
 Ensure that the pressure in the pressure bag
is maintained at 300mmHg.
 Always set and keep the alarms on.
Cont.
 Do not allow the flush bag to empty
 To maintain patency of arterial cannula.
 To prevent air embolism
 To maintain accuracy of blood pressure reading
 To maintain accuracy of fluid balance chart
 To prevent backflow of blood
Cont.
 Monitor color & temperature of limb distal to
arterial line & compare to other limb
 To confirm that circulation to the limb is
adequate.
 To ensure the early detection of impaired
circulation
Cont.
 Monitor and display the arterial waveform at
all times
 To detect cannula disconnection.
 Level and zero transducer once per shift
 To ensure accuracy in measuring blood pressure
 Maintain the transducer level with the
patient’s phlebostatic axis (fourth intercostal
space midaxillary line)
Cont.
 On removal of arterial cannula maintain
pressure over puncture site for at least 5
minutes until bleeding has stopped
 To prevent bleeding and haematoma formation
 Send cannula tip to microbiology
 Only if suspected infection
 To detect infection
CENTRAL VENOUS PRESSURE
 Is the pressure within the superior vena cava or
the right atrium
 It serve as a guide to fluid balance in critically ill
patients
 It give estimation of the circulating blood volume
 It also assist in monitoring of rt ventricular function
 Route for delivery of medications
Cont.
 CVP is a helpful tool in the assessment of
cardiac function, circulating blood volume, and
patient’s response to treatment
 CVP should not be interpreted solely but in
conjunction with other systemic measurements,
as isolated CVP measurements can be
misleading
 Normal CVP is less than 8 mm Hg
 CVP is raised in mechanical ventilation.
CVP INSERTION:
 http://content.nejm.org/cgi/video/356/21/e21/
METHODS OF CVP
MONITORING
 There are two methods of CVP monitoring
 manometer system: enables intermittent
readings and is less accurate than the
transducer system
 transducer system: enables continuous readings
which are displayed on a monitor.
MONITORING WITH
TRANSDUCERS
 Transducers enable the pressure readings
from invasive monitoring to be displayed on
a monitor
 To maintain patency of the cannula a bag
of normal saline or heparinized saline
should be connected to the transducer
tubing and kept under continuous pressure
of 300mmHg. (autoflush 3ml/hr)
PROCEDURE FOR CVP MEASUREMENT
USING A TRANSDUCER
 Explain the procedure to the patient
 Ensure that central line is patent
 Position the patient supine (if possible) and
align the transducer with the phlebostatic axis
 Zero the monitor
 Observe the CVP tracing
 Document the reading and report any changes
or abnormalities
COMPLICATIONS
 Infection
 Arterial puncture
 Hematoma
 Pneumothorax
 Hemothorax
 Arrhythmias
 Thrombosis
 Air embolism
MANAGEMENT OF A PATIENT
WITH A CVP LINE
 Monitor patient for signs of complications
 Label CVP lines with drugs/fluids etc. being infused
in order to minimise the risk of accidental bolus
injection
 Ensure all connections are secure to prevent
infection and introduction of air emboli
 Observe the insertion site frequently for signs of
infection.
 CVP lines should be removed when clinically
indicated
REMOVAL OF CENTRAL LINES
 This is an aseptic procedure
 The patient should be supine with head tilted down
 Ensure no drugs are attached and running via the
central line
 Remove dressing
 Cut the stitches
 Slowly remove the catheter
 If there is resistance then call for assistance
 Apply digital pressure with gauze until bleeding stops
 Dress with gauze and do clear dressing eg tegaderm
PA PRESSURE MONITORING
 Pulmonary artery pressure monitoring is
measuring the pressure in the pulmonary
artery leading to the lungs
 It also allows for indirect measurement of
left heart pressures since the pulmonary
veins have no valves in them and collects
the information needed to calculate cardiac
output and resistance
Cont.
 The PA catheter assesses right ventricular
function, pulmonary vascular status,
indirectly left ventricular function and all 3
components of stroke volume
 PAC aids in diagnosis of cardiovascular
and cardiopulmonary dysfunction, therapy
needed, and evaluate effectiveness of
interventions.
PULMONARY ARTERY CATHETER
 A Pulmonary artery catheter is a multi
lumen catheter inserted into pulmonary
artery.
 Each lumen or port has specific functions.
 Veins used for PA catheter insertion include
the internal jugular, subclavian, femoral and
very less commonly brachial.
Components of Swan-Ganz [con’t]
 Proximal port – [Blue] used to measure
central venous pressure/RAP and
injectate port for measurement of cardiac
output
 Distal port – [Yellow] used to measure
pulmonary artery pressure and for
withdrawal of mixed venous saturation.
Medication administration from this port
is not recommended.
Cont..
 Balloon port – [Red] used to determine
pulmonary wedge pressure;1.5 ml air is
injected via special syringe already
attached with the catheter.
 Thermister port- [White] it measures
patients temperature in the pulmonary
artery and reflect the temperature change
when fluid is injected for cardiac output.
PA CATHETER INSERTION
 http://www.edwards.com/products/pacatheters/c
 http://www.edwards.com/Products/PACathe
ters/HDMTroubleshooting.htm
INDICATIONS:
 To assess volume status and myocardial function.
 To assist in making a differential diagnosis
 To guide the management of the patient with heart/lung
disease/shock of all types
 To monitor hemodynamic pressures during fluid resuscitation
 Inotropic/vasoconstrictor/vasodilator drug infusion therapy
 To assess complications of MI and heart failure
 To monitor hemodyanamics with complicated surgical
procedures
 Sepsis and multi system organ failure
 Complex surgery, complicated myocardial infarction
CONTRAINDICATION:
 Severe coagulopathy.
 Patient receiving thrombolytics (e.g-TPA).
 Prosthetic right heart valve - catheter may cause the valve to
malfunction.
 Endocardial pacemaker - catheter may dislodge or knot around
the electrode.
 Severe pulmonary hypertension - increased risk of PA rupture.
 Severe vascular disease - catheter may puncture an abnormal
vessel.
 Significant immunodeficiency.
 If there are no skilled physicians/staff.
 If the patient's disease or injury can't be modified or corrected by
therapy.
COMPLICATIONS DURING
INSERTION:
 Pneumothorax
 Venous air embolism
 Dysrhythmias
 Dislodgement of the catheter guide wire
 Excessive bleeding
COMPLICATIONS AFTER
INSERTION:
 Dysrhythmias
 Infection
 Catheter dislodgement
 Thrombophlebitis
 Pulmonary artery rupture
 Tension Pnuemothorax
Cont..
 Catheter wedges permanently—considered an
emergency, notify MD immediately, can occur
when balloon is left inflated or catheter migrates
too far into pulmonary artery (flat PA waveform)…
can cause pulmonary infarct after only a few
minutes!
 Ventricular irritation – occurs when catheter
migrates back into RV or is looped through the
ventricle, notify MD immediately…can cause VT
• Check coagulation labs (pt, ptt, INR, platelets)
• Transfuse if Platelets < than 70 and INR > 1.5
• Ensure Packed Red Blood Cells in cooler at bedside
(Remember two RN check for PRBCs. Instructions
for blood in cooler, taped to cooler)
• Ensure good vascular access
• Evaluate need for sedation. (if too active
↑ BP may → bleeding)
NURSING CARE BEFORE REMOVAL OF PAC
• Keep PRBCs for a minimum of 1 hour
• Continuous hemodynamic monitoring for minimum 1hour
• Assess for signs of tamponade-dampening arterial wave
form, narrowing pulse pressure and bleeding- blood in
chest tubes, decrease blood pressure, pallor altered LOC)
• Document vital signs every 15 minutes
• Check HCT if bleeding suspected
• Ensure patency of chest tubes
• Do not transfer patient for at least 2 hours
NURSING CARE AFTER REMOVAL OF PAC
TROUBLESHOOTING:
 Remove multiple stopcocks, multiple injection
ports, and long lengths of tubing whenever
possible
 The optimal length of pressure tubing is 4 feet
 Overly compliant tubing leads to over damping
 Avoid large diameter tubing
 Remove all air bubbles from the system
 Ensure that all connections are tight and
periodically flush all tubing and stopcocks to
remove air bubbles
Cont…
 Whenever you are evaluating a patient’s changing
hemodyanamics check all transducers, stopcocks,
tubing, and injection ports for air
 Gently tap the tubing and stopcocks as the
continuous flush valve is opened to dislodge any
bubbles.
 This will usually clear the system and restore
measurement accuracy
 Flushing a few small bubbles through the catheter
is OK; if more air is present, aspirate it from the
tubing
Cont…
 Changes in bed positioning generally
require re zeroing the pressure transducer
 If the transducer is below the phlebostatic
axis, the resulting arterial pressure will be
erroneously high
 If the transducer is above the phlebostatic
axis, the resulting arterial pressure will be
erroneously low
Documentation
 Document PAS, PAD, and PCWP on nursing flowsheet
under Hemodynamic Parameters
 PCWP will rarely be > PAD (if so, means blood is flowing
backwards) If PCWP = PAD, look for tamponade
 Under circumstances where the catheter will not wedge (or
should not be), do not document any values in the PCWP
column on the flowsheet.

Contenu connexe

Tendances

Echocardiography to assess Fluid Responsiveness in icu
Echocardiography to assess Fluid Responsiveness in icuEchocardiography to assess Fluid Responsiveness in icu
Echocardiography to assess Fluid Responsiveness in icuDr.Mahmoud Abbas
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slidesmaltiaziz24
 
cardiac output monitoring
cardiac output monitoringcardiac output monitoring
cardiac output monitoringmadhu chaitanya
 
Advanced hemodynamic monitoring
Advanced hemodynamic monitoringAdvanced hemodynamic monitoring
Advanced hemodynamic monitoringGhaleb Almekhlafi
 
Invasive blood pressure_monitoring
Invasive blood pressure_monitoringInvasive blood pressure_monitoring
Invasive blood pressure_monitoringUbaidur Rahaman
 
Intra Aortic Balloon Pump by Rubina Shehzadi RN
Intra Aortic Balloon Pump by Rubina Shehzadi RNIntra Aortic Balloon Pump by Rubina Shehzadi RN
Intra Aortic Balloon Pump by Rubina Shehzadi RNRubina Shehzadi
 
Hemodynamic Pressure Monitoring
Hemodynamic Pressure MonitoringHemodynamic Pressure Monitoring
Hemodynamic Pressure MonitoringKhalid
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoringAlbert Blesson
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoringIrvan Setiawan
 
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING vikramnaidu2311
 
Central venous catheterization
Central venous catheterizationCentral venous catheterization
Central venous catheterizationMominul Haider
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheterArun Aru
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheterrajkumarsrihari
 
Use of echocardiography in icu
Use of echocardiography in icuUse of echocardiography in icu
Use of echocardiography in icuNisheeth Patel
 

Tendances (20)

Central venous cannulation
Central venous cannulation Central venous cannulation
Central venous cannulation
 
Echocardiography to assess Fluid Responsiveness in icu
Echocardiography to assess Fluid Responsiveness in icuEchocardiography to assess Fluid Responsiveness in icu
Echocardiography to assess Fluid Responsiveness in icu
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slides
 
Hemodynamics
HemodynamicsHemodynamics
Hemodynamics
 
cardiac output monitoring
cardiac output monitoringcardiac output monitoring
cardiac output monitoring
 
Advanced hemodynamic monitoring
Advanced hemodynamic monitoringAdvanced hemodynamic monitoring
Advanced hemodynamic monitoring
 
Invasive blood pressure_monitoring
Invasive blood pressure_monitoringInvasive blood pressure_monitoring
Invasive blood pressure_monitoring
 
Intra Aortic Balloon Pump by Rubina Shehzadi RN
Intra Aortic Balloon Pump by Rubina Shehzadi RNIntra Aortic Balloon Pump by Rubina Shehzadi RN
Intra Aortic Balloon Pump by Rubina Shehzadi RN
 
Hemodynamic Pressure Monitoring
Hemodynamic Pressure MonitoringHemodynamic Pressure Monitoring
Hemodynamic Pressure Monitoring
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
Pulmonary artery Catheter
Pulmonary artery CatheterPulmonary artery Catheter
Pulmonary artery Catheter
 
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
 
Central venous catheterization
Central venous catheterizationCentral venous catheterization
Central venous catheterization
 
Icp monitoring seminar
Icp monitoring seminarIcp monitoring seminar
Icp monitoring seminar
 
CVP Monitoring_Dr. Subrata Kumar_BSMMU_2014
CVP Monitoring_Dr. Subrata Kumar_BSMMU_2014CVP Monitoring_Dr. Subrata Kumar_BSMMU_2014
CVP Monitoring_Dr. Subrata Kumar_BSMMU_2014
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheter
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheter
 
Use of echocardiography in icu
Use of echocardiography in icuUse of echocardiography in icu
Use of echocardiography in icu
 
Cardioversion
Cardioversion Cardioversion
Cardioversion
 

En vedette

Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoringNurseKim
 
REVIEW OF HEMODYNAMIC MONITORING
REVIEW OF HEMODYNAMIC MONITORINGREVIEW OF HEMODYNAMIC MONITORING
REVIEW OF HEMODYNAMIC MONITORINGGhaleb Almekhlafi
 
Hemodynamic monitoring- Dr Sandeep Gampa
Hemodynamic monitoring- Dr Sandeep GampaHemodynamic monitoring- Dr Sandeep Gampa
Hemodynamic monitoring- Dr Sandeep GampaSiddharth Pandey
 
Hemodynamic monitoring r weinzierl
Hemodynamic monitoring r weinzierlHemodynamic monitoring r weinzierl
Hemodynamic monitoring r weinzierlLaurie Crane
 
Minimaly invasive hemodynamic monitoring for hepatic patients
Minimaly invasive hemodynamic monitoring for hepatic patients Minimaly invasive hemodynamic monitoring for hepatic patients
Minimaly invasive hemodynamic monitoring for hepatic patients Dr.Mahmoud Abbas
 
Manu Malbrain - Nursing thisisit final monitoring - IFAD 2012
Manu Malbrain - Nursing thisisit final monitoring - IFAD 2012Manu Malbrain - Nursing thisisit final monitoring - IFAD 2012
Manu Malbrain - Nursing thisisit final monitoring - IFAD 2012International Fluid Academy
 
Static and dynamic indices of hemodynamic monitoring
Static and dynamic indices of hemodynamic monitoringStatic and dynamic indices of hemodynamic monitoring
Static and dynamic indices of hemodynamic monitoringBhargav Mundlapudi
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoringPratik Tantia
 
Hemodynamic monitoring ppt
Hemodynamic monitoring pptHemodynamic monitoring ppt
Hemodynamic monitoring pptUma Binoy
 

En vedette (9)

Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
REVIEW OF HEMODYNAMIC MONITORING
REVIEW OF HEMODYNAMIC MONITORINGREVIEW OF HEMODYNAMIC MONITORING
REVIEW OF HEMODYNAMIC MONITORING
 
Hemodynamic monitoring- Dr Sandeep Gampa
Hemodynamic monitoring- Dr Sandeep GampaHemodynamic monitoring- Dr Sandeep Gampa
Hemodynamic monitoring- Dr Sandeep Gampa
 
Hemodynamic monitoring r weinzierl
Hemodynamic monitoring r weinzierlHemodynamic monitoring r weinzierl
Hemodynamic monitoring r weinzierl
 
Minimaly invasive hemodynamic monitoring for hepatic patients
Minimaly invasive hemodynamic monitoring for hepatic patients Minimaly invasive hemodynamic monitoring for hepatic patients
Minimaly invasive hemodynamic monitoring for hepatic patients
 
Manu Malbrain - Nursing thisisit final monitoring - IFAD 2012
Manu Malbrain - Nursing thisisit final monitoring - IFAD 2012Manu Malbrain - Nursing thisisit final monitoring - IFAD 2012
Manu Malbrain - Nursing thisisit final monitoring - IFAD 2012
 
Static and dynamic indices of hemodynamic monitoring
Static and dynamic indices of hemodynamic monitoringStatic and dynamic indices of hemodynamic monitoring
Static and dynamic indices of hemodynamic monitoring
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
Hemodynamic monitoring ppt
Hemodynamic monitoring pptHemodynamic monitoring ppt
Hemodynamic monitoring ppt
 

Similaire à Lec # 6 hemodynamic monitoring

arterial and CVP monitoring in perioperative period.pptx
arterial and CVP monitoring in perioperative period.pptxarterial and CVP monitoring in perioperative period.pptx
arterial and CVP monitoring in perioperative period.pptxkhelifakolea
 
Hemodynamic monitoring.pptx
Hemodynamic monitoring.pptxHemodynamic monitoring.pptx
Hemodynamic monitoring.pptxGETAYE5
 
Clinicalteaching [autosaved]
Clinicalteaching [autosaved]Clinicalteaching [autosaved]
Clinicalteaching [autosaved]Anvin Thomas
 
bpmonitoring.pdf
bpmonitoring.pdfbpmonitoring.pdf
bpmonitoring.pdfshafina27
 
Non Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTNon Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTRanjith Thampi
 
Hemodynamic monitoring of critically ill patients
Hemodynamic monitoring of critically ill patientsHemodynamic monitoring of critically ill patients
Hemodynamic monitoring of critically ill patientsV4Veeru25
 
Arterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptArterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptssuser35745f
 
Cardiovascular monitoring Part I
Cardiovascular monitoring Part ICardiovascular monitoring Part I
Cardiovascular monitoring Part ISiddhanta Choudhury
 
HAEMODYNAMIC MONITORING PART 1&2.pptx
HAEMODYNAMIC MONITORING PART 1&2.pptxHAEMODYNAMIC MONITORING PART 1&2.pptx
HAEMODYNAMIC MONITORING PART 1&2.pptxashishnair22
 
Basic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxBasic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxamitkalirawana07
 
Arterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptArterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptssuser35745f
 
Rush Exam with Ultrasound Cases.pdf
Rush Exam with Ultrasound Cases.pdfRush Exam with Ultrasound Cases.pdf
Rush Exam with Ultrasound Cases.pdfAlireza Bahmani
 
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...prateek gupta
 
anaesthesia.Monitoring.(dr.amr)
anaesthesia.Monitoring.(dr.amr)anaesthesia.Monitoring.(dr.amr)
anaesthesia.Monitoring.(dr.amr)student
 

Similaire à Lec # 6 hemodynamic monitoring (20)

arterial and CVP monitoring in perioperative period.pptx
arterial and CVP monitoring in perioperative period.pptxarterial and CVP monitoring in perioperative period.pptx
arterial and CVP monitoring in perioperative period.pptx
 
Hemodynamic monitoring.pptx
Hemodynamic monitoring.pptxHemodynamic monitoring.pptx
Hemodynamic monitoring.pptx
 
Perioprative monitoring
Perioprative monitoringPerioprative monitoring
Perioprative monitoring
 
Clinicalteaching [autosaved]
Clinicalteaching [autosaved]Clinicalteaching [autosaved]
Clinicalteaching [autosaved]
 
bpmonitoring.pdf
bpmonitoring.pdfbpmonitoring.pdf
bpmonitoring.pdf
 
Non Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTNon Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRT
 
Hemodynamic monitoring of critically ill patients
Hemodynamic monitoring of critically ill patientsHemodynamic monitoring of critically ill patients
Hemodynamic monitoring of critically ill patients
 
Monitoring in ICU
Monitoring in ICUMonitoring in ICU
Monitoring in ICU
 
Cvp
CvpCvp
Cvp
 
Arterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptArterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.ppt
 
Cardiovascular monitoring Part I
Cardiovascular monitoring Part ICardiovascular monitoring Part I
Cardiovascular monitoring Part I
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
HAEMODYNAMIC MONITORING PART 1&2.pptx
HAEMODYNAMIC MONITORING PART 1&2.pptxHAEMODYNAMIC MONITORING PART 1&2.pptx
HAEMODYNAMIC MONITORING PART 1&2.pptx
 
Advanced Hemodynamics
Advanced HemodynamicsAdvanced Hemodynamics
Advanced Hemodynamics
 
Basic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxBasic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptx
 
Arterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptArterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.ppt
 
Rush Exam with Ultrasound Cases.pdf
Rush Exam with Ultrasound Cases.pdfRush Exam with Ultrasound Cases.pdf
Rush Exam with Ultrasound Cases.pdf
 
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
 
anaesthesia.Monitoring.(dr.amr)
anaesthesia.Monitoring.(dr.amr)anaesthesia.Monitoring.(dr.amr)
anaesthesia.Monitoring.(dr.amr)
 
Dialysis machines key features
Dialysis machines key featuresDialysis machines key features
Dialysis machines key features
 

Lec # 6 hemodynamic monitoring

  • 2. OBJECTIVES By the end of this session, learners will be able to:  Define what is hemodynamic monitoring.  List down the importance of hemodynamic monitoring in hospital.  Discuss about pressure monitoring system.  Identify troubleshooting in the pressure monitoring system.
  • 3. CONT.  Explain in detail about arterial pressure monitoring, central venous pressure monitoring and pulmonary artery pressure monitoring.  Evaluate oxygen delivery and demand.
  • 4. HEMODYNAMIC MONOTORING  It is the study of interrelationship of blood pressure, blood flow, vascular volumes, heart rate, ventricular function and the physical properties of blood at bedside  Is an integral part of critical care nursing
  • 5. Cont..  CCN should have knowledge of how to obtain accurate data, analyze waveforms and interpret and integrate the data.  The information provided by invasive catheters can give us accurate and timely information to clinicians so that appropriate interventions can be taken.
  • 6. IMPORTANCE OF HDM  Early detection, identification, and treatment of life-threatening conditions such as heart failure and cardiac temponade.  Evaluate the patient’s immediate response to treatment such as drugs and mechanical support.  Evaluate the effectiveness of cardiovascular function such as cardiac output and index.
  • 7. PURPOSE OF HDM:  The purpose of HDM is to:  Early detection, identification, and treatment of life-threatening conditions  Aid in the diagnosis of various cardiovascular disorders  Guide therapies to optimize cardiac functions  Minimize cardiovascular dysfunctions  Evaluate the patient’s immediate response to treatment modalities
  • 8. INDICATION OF HDM  It is essential in conditions when cardiac output is insufficient to deliver oxygen to the cells (altered preload, afterload and contractility).  Any deficit or loss of cardiac function: such as AMI, CHF, Cardiomyopathy.  All types of shock; cardiogenic, septic, neurogenic or anaphylactic.  Dehydration, hemorrhage, G.I. bleed, severe burn, ARDS or any major surgery.  Severe sepsis or multiple organ failure.  It aids in assessing body oxygen supply and demand.
  • 9. HDM PARAMETERS:  Cardiac output.  Heart rate  Rhythm  Cardiac index.  Stroke volume.  Preload  Afterload  Contractility
  • 10. PRESSURE MONITORING SYSTEM  Catheter (hollow tube).  Central venous catheter for CVP  Arterial catheter for Arterial pressure monitoring  Pulmonary catheter / SWAN Ganz for PAP monitoring.  Pressure tubing.  Flush solution (N/S or D5W, heparinized)  IV tubing with drip chamber  Pressure bag (pressure of 300 mmHg) and stopcock  Pressure transducer.  Pressure amplifier.  Cardiac monitor
  • 11. SQUARE WAVE TEST:  Square wave test is performed to quickly assess the response of the system.  Flush device is released rapidly which will increase fluid flow through the system.  Record the square wave formed on the monitor.
  • 12. LEVELING & ZEROING:  A fundamental step in obtaining accurate hemodynamic values is to zero the transducer amplifier system.  This step is performed at least once before obtaining the first hemodynamic reading after catheter insertion.  The pressure monitoring system is leveled to an external landmark and then zeroed to atmospheric pressure.  Zeroing is done to eliminate the effect of atmospheric and hydrostatic pressure.
  • 13. Cont..  The phlebostatic axis is used as the reference point for leveling and zeroing.  The site of the phlebostatic axis is at the intersection of the fourth intercostal space and mid axillary line (approx level of right atrium and pulmonary artery.  Leveling (referencing) and zeroing ensures that hemodynamic values obtained with the catheter are accurate.
  • 14. STEPS FOR ZEROING AND LEVELING:  Position patient on their back  Patient may be positioned with the head of the bed elevated between 0-60°  Flush the system  Level transducer to phlebostatic axis (may mark this with an x on patient)  Turn stop-cock on transducer so that it is off to the patient.  Remove cap  Press zero on the module  Ensure that zero appears on screen replace cap and turn stop-cock so that it is open to monitoring and patient.
  • 15. ARTERIAL PRESSURE MONITORING:  Arterial blood pressure is a basic hemodynamic index often utilized to guide therapeutic interventions  Continuous monitoring of blood pressure is indicated for patients with hemodynamic instability that requires inotropic or vasopressor medication
  • 16. Cont.  An arterial line is a cannula usually positioned in a peripheral artery such as  Radial artery (most commonly used site)  Brachial artery (rarely used)  Dorsalis pedis artery (rarely used)  Femoral artery (second option other than radial artery, more chances of getting contaminated)
  • 17. Cont..  An arterial line allows for consistent and continuous monitoring of blood pressure to facilitate the reliable titration of supportive medications  In addition, arterial lines allow for reliable access to the arterial circulation for the measurement of arterial oxygenation and for frequent blood sampling
  • 18. ARTERIAL LINE INSERTION:  http://content.nejm.org/cgi/video/354/15/e13/
  • 19. ARTERIAL PRESSURE  Arterial pressure waveform has three parts:  Rapid upstroke wave (systolic pressure)  Dicrotic notch  Diastolic pressure waveform  Normal systolic BP is 90 – 140 mm Hg.  Normal diastolic BP is 60 – 90 mm Hg.  Dicrotic notch is small downward deflection following the closure of semilunar valve, indicate the end of systole and beginning of diastole.
  • 20. MEAN ARTERIAL PRESSURE:  Mean arterial pressure (MAP) is used to evaluate perfusion of vital body organs.  Normal MAP is 70 to 105 mm Hg.  MAP can be calculated by: Systolic BP + (Diastolic *2) / 3
  • 21. EJECTION FRACTION  Ejection Fraction (EF) is the fraction of blood ejected by the ventricle relative to its end-diastolic volume.  EF is calculated from: EF = (SV / EDV) * 100  For example if SV is 75 and EDV is 120 then EF would be 63%.
  • 22. POTENTIAL COMPLICATIONS  HAEMORRHAGE: PREVENTION:  Keep limb visible at all times  Ensure alarm is on so that any accidental disconnection can be dealt quickly  Ensure that arm is immobile with arm board  Ensure all connections are tight SOLUTION  Apply pressure to limb  Assess leak  If hemorrhage persists notify medical officer
  • 23. CONT..  INFECTION PREVENTION  Assess area regularly for redness or swelling  Avoid interrupting circuit as much as possible  Use gloves when touching arterial line SOLUTION  Remove arterial Line  Ensure proper hand washing when handling arterial line or transducer
  • 24. CONT..  BLOCKAGE, CLOTTING & AIR EMBOLI PREVENTION  Keep pressure bag inflated to 300 mmHg  ensure 3-5ml auto flush is continuous.  Attempt to aspirate blood  Use fast flush device to clear line to prevent clot formation SOLUTION  Attempt to aspirate blood to remove clot  Ensure all connections are secure
  • 25. CONT..  INTERUPTION TO PERIPHERAL CIRCULATION PREVENTION  Regularly check distal pulses and capillary refill. SOLUTION  Notify MO and consider removing arterial line
  • 26. NURSING CONSIDERATIONS  Nursing care mainly directed to preventing complications  Ensure that the insertion site is visible at all times for early detection of disconnection  All connections must be secured  Ensure that the cannula site is covered with an appropriate dressing to maintain asepsis
  • 27. Cont.  Never inject anything into an arterial cannula or arterial line  Ensure that the flush bag has adequate fluid.  Use only heparinized 0.9% sodium chloride  Ensure that the pressure in the pressure bag is maintained at 300mmHg.  Always set and keep the alarms on.
  • 28. Cont.  Do not allow the flush bag to empty  To maintain patency of arterial cannula.  To prevent air embolism  To maintain accuracy of blood pressure reading  To maintain accuracy of fluid balance chart  To prevent backflow of blood
  • 29. Cont.  Monitor color & temperature of limb distal to arterial line & compare to other limb  To confirm that circulation to the limb is adequate.  To ensure the early detection of impaired circulation
  • 30. Cont.  Monitor and display the arterial waveform at all times  To detect cannula disconnection.  Level and zero transducer once per shift  To ensure accuracy in measuring blood pressure  Maintain the transducer level with the patient’s phlebostatic axis (fourth intercostal space midaxillary line)
  • 31. Cont.  On removal of arterial cannula maintain pressure over puncture site for at least 5 minutes until bleeding has stopped  To prevent bleeding and haematoma formation  Send cannula tip to microbiology  Only if suspected infection  To detect infection
  • 32. CENTRAL VENOUS PRESSURE  Is the pressure within the superior vena cava or the right atrium  It serve as a guide to fluid balance in critically ill patients  It give estimation of the circulating blood volume  It also assist in monitoring of rt ventricular function  Route for delivery of medications
  • 33. Cont.  CVP is a helpful tool in the assessment of cardiac function, circulating blood volume, and patient’s response to treatment  CVP should not be interpreted solely but in conjunction with other systemic measurements, as isolated CVP measurements can be misleading  Normal CVP is less than 8 mm Hg  CVP is raised in mechanical ventilation.
  • 35. METHODS OF CVP MONITORING  There are two methods of CVP monitoring  manometer system: enables intermittent readings and is less accurate than the transducer system  transducer system: enables continuous readings which are displayed on a monitor.
  • 36. MONITORING WITH TRANSDUCERS  Transducers enable the pressure readings from invasive monitoring to be displayed on a monitor  To maintain patency of the cannula a bag of normal saline or heparinized saline should be connected to the transducer tubing and kept under continuous pressure of 300mmHg. (autoflush 3ml/hr)
  • 37. PROCEDURE FOR CVP MEASUREMENT USING A TRANSDUCER  Explain the procedure to the patient  Ensure that central line is patent  Position the patient supine (if possible) and align the transducer with the phlebostatic axis  Zero the monitor  Observe the CVP tracing  Document the reading and report any changes or abnormalities
  • 38. COMPLICATIONS  Infection  Arterial puncture  Hematoma  Pneumothorax  Hemothorax  Arrhythmias  Thrombosis  Air embolism
  • 39. MANAGEMENT OF A PATIENT WITH A CVP LINE  Monitor patient for signs of complications  Label CVP lines with drugs/fluids etc. being infused in order to minimise the risk of accidental bolus injection  Ensure all connections are secure to prevent infection and introduction of air emboli  Observe the insertion site frequently for signs of infection.  CVP lines should be removed when clinically indicated
  • 40. REMOVAL OF CENTRAL LINES  This is an aseptic procedure  The patient should be supine with head tilted down  Ensure no drugs are attached and running via the central line  Remove dressing  Cut the stitches  Slowly remove the catheter  If there is resistance then call for assistance  Apply digital pressure with gauze until bleeding stops  Dress with gauze and do clear dressing eg tegaderm
  • 41. PA PRESSURE MONITORING  Pulmonary artery pressure monitoring is measuring the pressure in the pulmonary artery leading to the lungs  It also allows for indirect measurement of left heart pressures since the pulmonary veins have no valves in them and collects the information needed to calculate cardiac output and resistance
  • 42. Cont.  The PA catheter assesses right ventricular function, pulmonary vascular status, indirectly left ventricular function and all 3 components of stroke volume  PAC aids in diagnosis of cardiovascular and cardiopulmonary dysfunction, therapy needed, and evaluate effectiveness of interventions.
  • 43. PULMONARY ARTERY CATHETER  A Pulmonary artery catheter is a multi lumen catheter inserted into pulmonary artery.  Each lumen or port has specific functions.  Veins used for PA catheter insertion include the internal jugular, subclavian, femoral and very less commonly brachial.
  • 44. Components of Swan-Ganz [con’t]  Proximal port – [Blue] used to measure central venous pressure/RAP and injectate port for measurement of cardiac output  Distal port – [Yellow] used to measure pulmonary artery pressure and for withdrawal of mixed venous saturation. Medication administration from this port is not recommended.
  • 45. Cont..  Balloon port – [Red] used to determine pulmonary wedge pressure;1.5 ml air is injected via special syringe already attached with the catheter.  Thermister port- [White] it measures patients temperature in the pulmonary artery and reflect the temperature change when fluid is injected for cardiac output.
  • 46. PA CATHETER INSERTION  http://www.edwards.com/products/pacatheters/c  http://www.edwards.com/Products/PACathe ters/HDMTroubleshooting.htm
  • 47. INDICATIONS:  To assess volume status and myocardial function.  To assist in making a differential diagnosis  To guide the management of the patient with heart/lung disease/shock of all types  To monitor hemodynamic pressures during fluid resuscitation  Inotropic/vasoconstrictor/vasodilator drug infusion therapy  To assess complications of MI and heart failure  To monitor hemodyanamics with complicated surgical procedures  Sepsis and multi system organ failure  Complex surgery, complicated myocardial infarction
  • 48. CONTRAINDICATION:  Severe coagulopathy.  Patient receiving thrombolytics (e.g-TPA).  Prosthetic right heart valve - catheter may cause the valve to malfunction.  Endocardial pacemaker - catheter may dislodge or knot around the electrode.  Severe pulmonary hypertension - increased risk of PA rupture.  Severe vascular disease - catheter may puncture an abnormal vessel.  Significant immunodeficiency.  If there are no skilled physicians/staff.  If the patient's disease or injury can't be modified or corrected by therapy.
  • 49. COMPLICATIONS DURING INSERTION:  Pneumothorax  Venous air embolism  Dysrhythmias  Dislodgement of the catheter guide wire  Excessive bleeding
  • 50. COMPLICATIONS AFTER INSERTION:  Dysrhythmias  Infection  Catheter dislodgement  Thrombophlebitis  Pulmonary artery rupture  Tension Pnuemothorax
  • 51. Cont..  Catheter wedges permanently—considered an emergency, notify MD immediately, can occur when balloon is left inflated or catheter migrates too far into pulmonary artery (flat PA waveform)… can cause pulmonary infarct after only a few minutes!  Ventricular irritation – occurs when catheter migrates back into RV or is looped through the ventricle, notify MD immediately…can cause VT
  • 52. • Check coagulation labs (pt, ptt, INR, platelets) • Transfuse if Platelets < than 70 and INR > 1.5 • Ensure Packed Red Blood Cells in cooler at bedside (Remember two RN check for PRBCs. Instructions for blood in cooler, taped to cooler) • Ensure good vascular access • Evaluate need for sedation. (if too active ↑ BP may → bleeding) NURSING CARE BEFORE REMOVAL OF PAC
  • 53. • Keep PRBCs for a minimum of 1 hour • Continuous hemodynamic monitoring for minimum 1hour • Assess for signs of tamponade-dampening arterial wave form, narrowing pulse pressure and bleeding- blood in chest tubes, decrease blood pressure, pallor altered LOC) • Document vital signs every 15 minutes • Check HCT if bleeding suspected • Ensure patency of chest tubes • Do not transfer patient for at least 2 hours NURSING CARE AFTER REMOVAL OF PAC
  • 54. TROUBLESHOOTING:  Remove multiple stopcocks, multiple injection ports, and long lengths of tubing whenever possible  The optimal length of pressure tubing is 4 feet  Overly compliant tubing leads to over damping  Avoid large diameter tubing  Remove all air bubbles from the system  Ensure that all connections are tight and periodically flush all tubing and stopcocks to remove air bubbles
  • 55. Cont…  Whenever you are evaluating a patient’s changing hemodyanamics check all transducers, stopcocks, tubing, and injection ports for air  Gently tap the tubing and stopcocks as the continuous flush valve is opened to dislodge any bubbles.  This will usually clear the system and restore measurement accuracy  Flushing a few small bubbles through the catheter is OK; if more air is present, aspirate it from the tubing
  • 56. Cont…  Changes in bed positioning generally require re zeroing the pressure transducer  If the transducer is below the phlebostatic axis, the resulting arterial pressure will be erroneously high  If the transducer is above the phlebostatic axis, the resulting arterial pressure will be erroneously low
  • 57. Documentation  Document PAS, PAD, and PCWP on nursing flowsheet under Hemodynamic Parameters  PCWP will rarely be > PAD (if so, means blood is flowing backwards) If PCWP = PAD, look for tamponade  Under circumstances where the catheter will not wedge (or should not be), do not document any values in the PCWP column on the flowsheet.