The Complementary Roles of
Computer-Aided Diagnosis and
Quantitative Image Analysis:
Similarities and Differences in
Assessment, Quality Assurance
and Training
Berkman Sahiner1, PhD
Samuel G. Armato III2, PhD
Zhimin Huo3, PhD
Heang-Ping Chan4, PhD
Ronald M. Summers5, MD, PhD
Nicholas Petrick1, PhD
1: US Food and Drug Administration, Center for Devices and Radiological Health
2: University of Chicago, Department of Radiology
3: Carestream Inc.
4: University of Michigan, Department of Radiology
5: National Institutes of Health, Clinical Center
Introduction
ī‚§ Quantitative image analysis (QIA) and computer-aided
diagnosis (CAD) are closely linked
ī‚§ This exhibit reviews the methods developed for assessment,
quality assurance, and user training for CAD, and highlights
parallels and distinctions between CAD and QIA
ī‚§ Comparison between CAD and QIA in this computer exhibit
is performed through several example applications
developed by the authors or that are publicly available
What is CAD?
ī‚§ CAD systems incorporate pattern recognition and data
analysis capabilities* and are intended to
ī‚§ Mark regions of an image that may reveal specific abnormalities
and alert the clinician to these regions during image interpretation
(computer-aided detection (CADe) systems)**
ī‚§ Provide to the clinician an assessment of disease, disease type,
severity, stage, progression (computer-aided diagnosis (CADx)
systems)**
*Guidance for industry and FDA staff: “Clinical performance assessment: Considerations for computer-assisted
detection devices applied to radiology images and radiology device data - premarket approval (PMA) and
premarket notification [510(k)] submissions," (FDA, 2012).
**N. Petrick et al., “Evaluation of computer-aided detection and diagnosis systems,” Med. Phys. 2013, 40:087001.
What is QIA?
ī‚§ QIA is the process of extraction of quantitative imaging
biomarkers from medical images, typically involving
computerized tools
ī‚§ A quantitative imaging biomarker is an objectively measured
characteristic derived from an in vivo image as an indicator of
normal biological processes, pathogenic processes, or response to a
therapeutic intervention*
*Sullivan DC et al., “Metrology standards for quantitative Imaging biomarkers,” Radiology 2015 Aug 12:142202.
(Epub ahead of print).
What is Quantitative Imaging?
ī‚§ Both QIA and quantitative imaging biomarkers are within
the larger context of quantitative imaging, defined as
ī‚§ the extraction of quantifiable features from medical images for the
assessment of normal [findings] or the severity, degree of change,
or status of a disease, injury, or chronic condition relative to normal
[findings]
ī‚§ Quantitative imaging includes the development,
standardization, and optimization of anatomical, functional,
and molecular imaging acquisition protocols, data analyses,
display methods, and reporting structures
*Quantitative Imaging Biomarkers Alliance. http://rsna.org/QIBA.aspx. Accessed Oct. 15, 2015.
How are CAD and QIA Similar?
ī‚§ Both aim at aiding clinicians through advanced image
analysis techniques
ī‚§ Both commonly use computer methods to extract features
ī‚§ CAD may use other patient-related information, e.g., age, gender,
risk factors, or non-image biomarkers
ī‚§ CAD features may include quantitative imaging biomarkers, but can
also be relative, ordinal features
ī‚§ User interaction may be somewhat more prevalent in QIA (e.g.,
semi-automated segmentation)
ī‚§ Both emphasize appropriate image acquisition protocols,
display methods, training, and reporting
How are CAD and QIA Different?
ī‚§ In CAD, the emphasis is on how the CAD output can provide
decision support to clinicians
ī‚§ In full assessment, it is required to demonstrate that a CAD system
aids the clinician
ī‚§ May integrate imaging and non-imaging information or biomarkers
ī‚§ Not necessary to establish a direct link between individual CAD
features and a specific disease condition
ī‚§ Standalone testing (performance of the CAD system alone) is
nonetheless important
ī‚§ In QIA, the emphasis is on the extraction of biomarkers
ī‚§ Bias and variability across devices, patients and time are major
considerations
Assessment, Quality Assurance,
and Training
ī‚§ The differences and similarities between CAD and QIA have
a number of implications for assessment, quality assurance
(QA), and training
ī‚§ We briefly summarize our work in CAD in these areas and
draw parallels with QIA
Assessment - CAD
ī‚§ CAD systems are typically assessed by evaluating the effect
of the system on clinicians
ī‚§ Reader performance assessment: Evaluate performance of a
clinician using CAD as part of the decision making process
ī‚§ Retrospective multiple-reader multiple case (MRMC) studies
ī‚§ Prospective field trials
ī‚§ Standalone performance of a CAD system is also useful as a
performance indicator, both in
ī‚§ System design phase and
ī‚§ Final system assessment
Assessment - CAD
ī‚§ A recent paper by AAPM outlined important considerations in CAD
assessment including*:
ī‚§ Data set selection
ī‚§ Representativeness of the test set for the targeted population
ī‚§ Reference standard and mark labeling
ī‚§ Disease status of a case (sometimes includes location information)
ī‚§ Standalone assessment metrics
ī‚§ True-positive fraction, false-positive fraction pairs
ī‚§ Receiver operating characteristic (ROC) curves
ī‚§ Location-specific ROC
ī‚§ Reader study design and analysis
ī‚§ Prospective versus retrospective studies
ī‚§ MRMC study design
ī‚§ Reader training for reader studies
*N. Petrick et al., “Evaluation of computer-aided detection and diagnosis systems,” Med. Phys. 2013, 40:087001.
Assessment - Quantitative Imaging
ī‚§ Most groups working in quantitative imaging are interested
in technical performance assessment
ī‚§ How a test performs in reference objects or subjects under
controlled conditions
ī‚§ Other quantitative imaging assessment
ī‚§ Clinical assessment
ī‚§ Clinical impact
Technical Assessment - Quantitative Imaging
ī‚§ Technical assessment can be performed using
ī‚§ Anthropomorphic phantoms
ī‚§ Large number of replicate measurements easily obtained
ī‚§ The same phantoms can be used repeatedly
ī‚§ Reference standard easier to obtain
ī‚§ Phantom realism may be limited
ī‚§ Generalizability to human data may be limited because of limited
abnormality variation and complexity
ī‚§ Patient data
ī‚§ More realistic
ī‚§ Number of replicate measurements may be limited
ī‚§ Reference standard more difficult to obtain
ī‚§ Number of available patients usually limited
Technical Assessment - Quantitative Imaging
ī‚§ Important considerations include*
ī‚§ Measurand/Reference
ī‚§ True value of the quantity intended to be measured
ī‚§ Bias
ī‚§ Difference between the expected value of the biomarker and the measurand
ī‚§ Linearity
ī‚§ Is a change in the measurand reflected as a proportional change in the biomarker?
ī‚§ Repeatability
ī‚§ Ability to repeatedly measure the same feature under identical or near-identical
conditions
ī‚§ Reproducibility
ī‚§ Ability to measure the same feature under different conditions expected in
ī‚§ A preclinical study
ī‚§ Clinical trial
ī‚§ Clinical practice
D.L. Raunig et al., “Quantitative imaging biomarkers: a review of statistical methods for technical performance
assessment,” Stat Methods Med Res. 2015, 24:27-67.
Clinical Assessment - Quantitative Imaging
ī‚§ Clinical assessment of a QIA method
ī‚§ How does the method help clinicians in diagnosis or treatment?
ī‚§ Recommendations developed for clinical assessment of CAD
may be useful for the clinical assessment of quantitative
imaging biomarkers after proper modifications
ī‚§ Methodology for prospective and retrospective studies
ī‚§ MRMC methods
ī‚§ Assessment metrics
Quality Assurance - CAD
ī‚§ CAD systems are medical devices, and can benefit from QA
procedures like all medical devices
ī‚§ Clinical image acquisition:
ī‚§ Follow manufacturer’s QA procedure for imaging device
ī‚§ Additional QA for CAD:
ī‚§ Assure functionality and performance of CAD device according to
vendor’s specifications
ī‚§ Assure consistency of CAD device performance over time
ī‚§ A recent paper by AAPM outlined important considerations
in CAD quality assurance*
*Z. Huo, et al., “Quality assurance and training procedures for computer-aided detection and diagnosis systems,”
Med. Phys. 2013, 40:077001.
Quality Assurance - Quantitative Imaging
ī‚§ Clinical image acquisition:
ī‚§ Follow manufacturer’s QA procedure for imaging device
ī‚§ Image acquisition for quantitative imaging:
ī‚§ QA procedures to reduce variability across devices, patients and
time
ī‚§ QA procedures to ensure that acquired images are of quantitative
quality
ī‚§ Calibration methods and phantom tests in addition to those
recommended by manufacturers for clinical image quality
Clinician Training - CAD
ī‚§ Inform clinicians about
ī‚§ Intended use of the CAD system
ī‚§ CAD system’s limitations
ī‚§ Particular strengths and weaknesses of different CAD systems
ī‚§ E.g., with respect to lesion type, acquisition parameters
ī‚§ Although many clinicians and researchers have underlined
the need for clinician training for CAD
ī‚§ Additional awareness needs to be raised for proper use and
potential adverse effects of improper use
ī‚§ Best practices for CAD training are yet to be established
Clinician Training - Quantitative Imaging
ī‚§ Clinicians using the quantitative imaging tool in the clinic
should be qualified in accordance with local and national
requirements and standards
ī‚§ Training of operators in image acquisition and analysis can
influence the value of the extracted imaging biomarker
ī‚§ e.g., training of an operator in a semi-automated segmentation task
for a lesion volume biomarker
ī‚§ There is a small number of studies on clinician training and
its effect on the extracted quantitative imaging biomarker
â€ĸ The rest of the computer exhibit consists of a summary
and example applications developed by the authors or
that are publicly available
â€ĸ Click to go to the slide of your choice
â€ĸ Summary
â€ĸ Example applications
â€ĸ Click on the E on lower right from any slide to go to the list of
example applications
E
Summary
ī‚§ CAD and QIA fields have a number of shared attributes as
well as some differences
ī‚§ Methodologies developed in one field are likely to be
applicable in the other after careful consideration of the
differences
ī‚§ We expect that the development of standardization,
assessment and clinical use recommendations in these two
fields will strengthen each other
E
Example Applications
ī‚§ Example 1
ī‚§ CADe assessment: A
commercial CAD system
ī‚§ Example 2
ī‚§ Quantitative imaging
technical assessment: A
phantom study
ī‚§ Example 3
ī‚§ Quantitative volume change
analysis for treatment
response assessment of
head and neck neoplasms
on CT scans
ī‚§ Example 4
ī‚§ Computer-aided analysis of treatment
response of bladder cancers on CT scans
ī‚§ Example 5
ī‚§ Radiomic biomarkers and decision
support system
ī‚§ Example 6
ī‚§ Measurement of pleural effusion in
thoracic CT volumes
ī‚§ Example 7
ī‚§ Radiomics-based assessment of normal
lung tissue damage in radiation therapy
Example 1
CADe Assessment:
A Commercial CADe System
E
A Commercial CADe System
ī‚§ M-Vu Algorithm Engine
ī‚§ CADe device intended to aid radiologists in reading
screening mammograms
ī‚§ Mark areas for review by a radiologist, who
ī‚§ First reviews each case in the conventional manner,
ī‚§ then re-examines regions marked by the M-Vu system before
making a BI-RADS assessment
ī‚§ Pre-market approval (PMA) by the FDA in 2012
E
Test Data Set Selection
ī‚§ 140 positive and 140 negative cases
ī‚§ 11 sites in the United States
ī‚§ Academic, specialty, and community clinics.
ī‚§ Pre-defined inclusion and exclusion criteria
ī‚§ Sequential, eligible positive mammograms
ī‚§ A mammography exam for which a biopsy-proven breast cancer was
found within 15 months following the exam date
ī‚§ Eligible negative mammograms
ī‚§ A mammography exam for which
ī‚§ Breast cancer was not found within 15 months prior or 15 months after the exam
date
ī‚§ At least one associated subsequent negative mammogram was acquired at least 11
months after the exam date.
E
Reference Standard
ī‚§ Positive cases:
ī‚§ Pathology reports from biopsies or surgeries
ī‚§ Mammograms and radiology reports for lesion location
ī‚§ Each biopsy-proven lesion is outlined by site investigators
ī‚§ Negative cases
ī‚§ Radiology reports
E
Standalone Assessment: Sensitivity
ī‚§ A malignant region was considered to be detected if
ī‚§ A CAD mark centroid was inside the region, or
ī‚§ if the region centroid was inside a CAD mark
ī‚§ Having the cancer within the CAD mark, as above, can lead to an
optimistic estimate of sensitivity
ī‚§ A large mark may cover the lesion, but may not attract the attention of
the radiologist to the abnormality
ī‚§ When this is used as a detection criterion, other additional data are
evaluated to check if sensitivity is overestimated
ī‚§ A case was considered a true positive if
ī‚§ At least one malignant region was detected by CAD
E
Standalone Assessment: Sensitivity
Cases Sensitivity 95% Conf. Interval
Overall 140 79.3% (72.6% 86.0%)
Microcalcification 69 79.7% (70.2%, 89.2%)
Mass 86 81.4% (73.2%, 89.6%)
ī‚§ Sensitivity reported for two important sub-groups: Masses and
microcalcifications
ī‚§ Additional data on sensitivity with respect to
ī‚§ Lesion size
ī‚§ Lesion pathology (invasive, DCIS, other)
ī‚§ Breast density E
Standalone Performance: False
Positives
ī‚§ Additional data on
ī‚§ False-positives by breast density
ī‚§ Specificity (percentage of mammograms with no FP marks)
False-positives per mammo. 95% Conf. Interval
Total 0.418 (0.346, 0.490)
Microcalcification 0.088 (0.042,0.133)
Mass 0.330 (0.275, 0.385)
E
Pivotal Study
ī‚§ Multi-reader multi-case (MRMC) study to investigate the
effect of the CADe device on radiologists’ performance
ī‚§ 21 radiologists from a variety of academic, specialty, and
community clinics located across the US
ī‚§ For each case:
ī‚§ Evaluate the case without CAD and record an assessment
ī‚§ View CAD marks
ī‚§ Record a "with CAD" assessment
E
Pivotal Study
ī‚§ With and without CAD assessments
ī‚§ Recall/do not recall patient
ī‚§ Screening BI-RADS (0, 1, 2, 3, 4a, 4b, 4c, 5)
ī‚§ Forced BI-RADS (1, 2, 3, 4a, 4b, 4c, 5)
ī‚§ Lesion findings
ī‚§ Lesion findings:
ī‚§ Laterality
ī‚§ Type (Mass, Architectural Distortion, Asymmetry,
microcalcification)
ī‚§ BI-RADS (1, 2, 3, 4a, 4b, 4c, 5)
ī‚§ Probability of Malignancy (POM) (0-100%) E
Pivotal Study - ROC Analysis
ī‚§ POM data analyzed using ROC
methodology
ī‚§ Area under the ROC curve (AUC)
ī‚§ AUC w/0 CAD = 0.885
ī‚§ AUC with CAD = 0.902
ī‚§ Difference statistically significant
with p=0.013 (DBM MRMC method)
E
Pivotal Study – Sens., Spec.
ī‚§ Radiologist per-case sensitivity and specificity based on
decision to recall
W/o CAD With CAD Difference P-value
Sensitivity 0.865 0.901 0.036 (0.014, 0.058) 0.002
Specificity 0.649 0.623 -0.026 (-0.039, -0.013) < 0.001
ī‚§ Sensitivity increased and specificity decreased with CAD
ī‚§ Typical in reader studies for most second-read CADe systems
ī‚§ Lack of CADe mark for a lesion does not dissuade a reader from recall
ī‚§ Marks for missed lesions and some false-positive marks add to the
number of recalled cases E
Summary
ī‚§ Components of CAD assessment
ī‚§ Test data set selection
ī‚§ Representative of the targeted population
ī‚§ Reference standard
ī‚§ Clear and accurate definitions of actual positives and negatives
ī‚§ Standalone performance
ī‚§ How does the CAD system alone perform in the task for which it is intended to help
the radiologist ?
ī‚§ Performance metrics, confidence intervals
ī‚§ Reader performance
ī‚§ What is the effect of the CAD system on readers?
ī‚§ Performance metrics
ī‚§ Performance with and without CAD
ī‚§ Difference, confidence intervals E
Example 2
Quantitative Imaging
Technical Assessment:
A Phantom Study
E
Designing an Analysis Plan*
Step 1: Define the quantitative imaging biomarker and its relationship to quantity to
be measured (measurand)
Step 2: Define question to be addressed
ī‚§ Hypothesis or bounds on performance
Step 3: Define the experimental unit
ī‚§ Lesion-level, patient-level
Step 4: Define statistical measures of performance
ī‚§ What are the metrics for bias, repeatability, reproducibility?
Step 5: Specify elements of statistical design
ī‚§ Reference, reproducibility conditions, etc.
Step 6: Determine the data requirements
ī‚§ Patient population, type of images, sample size, etc.
Step 7: Collect data and perform statistical analysis
*Raunig et al., Stat Methods
Med Res, 2014 E
Main Considerations in
Technical Assessment
E
E
Main Considerations in
Technical Assessment
Purpose of the Study
ī‚§ Evaluate technical performance of a nodule volume
estimation tool
E
Study Design
ī‚§ Phantom
ī‚§ Thorax phantom with
vascular insert
ī‚§ Synthetic nodules
ī‚§ 4 spherical nodules
ī‚§ 5, 8, 9, 10 mm
â€ĸ +100 HU
E
Study Design
ī‚§ Image collection protocol
ī‚§ 10 repeat acquisitions
Phantom
16×0.75 mm
100 mAs
0.75 mm 1.5 mm
Reconstruction
Acquisition
Detailed Detailed
E
Bias/Linearity
ī‚§ Visually assess data to define limits of quantitation
ī‚§ Evaluate means/variances
ī‚§ Is data transformation necessary/appropriate to stabilize the variance?
-60
-40
-20
0
20
40
60
5 mm 8 mm 9 mm 10 mm
Nodule Size
50
150
250
350
450
550
0.75 1.5 0.75 1.5 0.75 1.5 0.75 1.5
Slice Thickness (mm)
MeasuredValue(mm3
)
5 mm 8 mm 9 mm 10 mm
Nodule Size
MeasuredValue-Median(mm3) E
Bias/Linearity
0 100 200 300 400 500 600
0
100
200
300
400
500
600
Reference Standard (mm3)
MeasuredValue(mm3
)
E
Repeatability/Reproducibility
ī‚§ RC
ī‚§ Estimated over 10 repeat acquisitions, 4 nodules
ī‚§ RDC
ī‚§ Estimated over 10 repeat acquisitions, 4 nodules, and 2 slice
thicknesses
Slice Thickness
0.75 mm 1.5 mm
RC 14.4 mm3 27.6 mm3
RDC 29.0 mm3
E
Repeatability Analysis
1.5 mm Slices
0 100 200 300 400 500 600
-60
-40
-20
0
20
40
60
Bland-Altman Plot (1.5 mm Slices)
RC=27.62 mm3
0.75 mm Slices
0 100 200 300 400 500 600
-60
-40
-20
0
20
40
60
Bland-Altman Plot (0.75 mm Slices)
RC=14.44 mm3
Mean(Esti,Estj) (mm3)Mean(Esti,Estj) (mm3)
Diff(Esti,Estj)(mm3)
Diff(Esti,Estj)(mm3) E
Reproducibility Analysis
0 100 200 300 400 500 600
-60
-40
-20
0
20
40
60
Bland-Altman Plot (Reproducibility)
Mean(Esti ) (mm,Estj
3
)
RDC=28.95 mm3
Diff(Esti,Estj)(mm3)
E
Summary
ī‚§ Main components of quantitative imaging biomarker technical
assessment
ī‚§ Bias/linearity analysis
ī‚§ Repeatability analysis
ī‚§ Reproducibility analysis
ī‚§ Others
ī‚§ Identification of significant factors/subgroups
ī‚§ â€Ļ.
ī‚§ Challenging to maintain consistency across studies
ī‚§ Phantom/clinical data
ī‚§ Transformation of data
ī‚§ Reference standard
ī‚§ Is test-retest data available?
ī‚§ Reproducibility conditions E
Example 3
Quantitative Volume Change Analysis
for Treatment Response Assessment of
Head and Neck Neoplasms on CT Scans
Ref. Hadjiiski L, Mukherji SK, Gujar SK, Sahiner B, Ibrahim M, Street E, Moyer J, Worden FP, Chan HP.
Treatment response assessment of head and neck cancers on CT using computerized volume analysis.
American Journal of Neuroradiology (AJNR) 2010;31(9):1744-1751.
E
Lesion Segmentation
Region of Interest
Preprocessing
Level Set
Automatic
Segmentation
E
Example: Tongue Base Tumor
E
Example: Tongue Base Tumor
E
Pre-treatment [cm3]
0 10 20 30 40 50
Post-treatement[cm3]
0
10
20
30
40
50
Pre-treatment [cm3]
0 10 20 30 40 50
Post-treatement[cm3]
0
10
20
30
40
50
Volume Segmentation
â€ĸ Pre-to-Post-treatment change in segmented primary
tumor volumes – 23 pairs
Auto Manual
E
Results - % Volume Change Comparison
â€ĸ Automatic vs. Manual % volume change
comparison – 23 tumor pairs
- Automatic vs. Manual:
r = 0.89
Pearson’s correlation r
E% Volume Change Automatic
-20 0 20 40 60 80 100
%VolumeChangeManual
-20
0
20
40
60
80
100
Results – Summary
WHO (longest diameter):
Primary tumors
RECIST (longest diameter and
perpendicular):
0.73
0.58
Pearson’s correlation coeff.
â€ĸ % Pre-to-Post-treatment change in volume estimates by 3
methods, relative to radiologist’s segmentation:
Quantitative image analysis: 0.89
E
Example: Tonsillar Tumor
Pre-treatment Post-treatment E
Example: Tonsillar Tumor
Pre-treatment Post-treatment E
Example: Tumor
Pre-treatment Post-treatment E
Example: Tumor
Pre-treatment Post-treatment E
Example 4
Computer-aided Analysis of Treatment
Response of Bladder Cancers on CT Scans
Ref. Hadjiiski L, Weizer AZ, Alva A, Caoili EM, Cohan RH, Cha K, Chan HP. Bladder cancers on CT: preliminary study
of treatment response assessment based on computerized volume analysis, WHO and RECIST Criteria.
American Journal of Roentgenology 2015, 205(2) pp 348-352.
E
Bladder Lesion Segmentation
Cascaded
Level Set
Region of Interest
Automatic
Segmentation
Auto-Initialized Cascaded Level Set (AI-CALS)
Preprocessing
E
Bladder Tumor: Pre- & Post-Treatment
Pre-
treatment
Post-
treatment
E
Feature Descriptors
ī‚§ Descriptors automatically extracted from the segmented
lesions:
- 15 radiomic features (RF) based on pre- and post-treatment
changes in:
- volume (V)
- 5 gray level descriptors (GL)
- 9 shape descriptors (S)
- Selected features merged into a Combined Response Index (CRI)
E
Combined Response Index (CRI)
â€ĸ AUC pT0 stage (complete response) vs. others
– 35 primary site tumors
Volume (3D): 0.68 Âą 0.09
Reference (3D): 0.66 Âą 0.11
CRI (Volume,
Shape):
0.75 Âą 0.10
CRI (Volume,
Shape, Gray
Level):
0.76 Âą 0.09
EFalse Positive Fraction
0.0 0.2 0.4 0.6 0.8 1.0
TruePositiveFraction
0.0
0.2
0.4
0.6
0.8
1.0
V
Reference (3D)
CRI (V, S)
CRI (V, S, GL)
Example 5
Radiomic Biomarkers and
Decision Support System
for Multiple Myeloma
Ref. Zhou C, Chan HP, Dong Q, Couriel DR, Pawarode A, Hadjiiski LM, Wei J. Quantitative analysis of MR
imaging to assess trestment response for patients with multiple myeloma by using dynamic intensity
entropy transformation 2015, Ahead of Print, 10.1148/radiol.2015142804
E
ī€ž Multiple Myeloma (MM) – a cancer formed by
malignant plasma cells in the bone marrow
o MR radiomic biomarkers (+ clinical biomarkers)
- Staging
- Treatment response assessment
- Prognosis prediction
- Recurrence – early detection
E
Radiomic Biomarkers for MM Treatment Response
T1W sagittal views
VOI of vertebral bodies and intervertebral discs
3D Dynamic Intensity Energy Transformation (DIET)
Quantitative Energy Enhancement Value:
Response Index DIET response map E
p-qEEVvbr and m-DqEEVvbr
T1-weighted MR DIET response map
Pre-Bone Marrow
Transplant
55 days post-BMT Pre-BMT 55 days post-BMT
BMT Treatment Response Assessment
Responder
E
56 days post-BMT Pre-BMT 56 days post-BMT
T1-weighted MR DIET response map
Non-responder
BMT Treatment Response Assessment
Pre-Bone Marrow
Transplant
E
Kaplan-Meier Survival Curve
predicted by p-qEEV biomarker
0 5 10 15 20 25 30 35
0.2
0.4
0.6
0.8
1
Time Since BMT (Months)
Survivalprobability
p-qEEV >= 10%
p-qEEV < 10%
E
Prediction of treatment response by
qEEV-based response index
Computer-Aided Decision Support
for Multiple Myeloma
Example 6
Measurement of Pleural Effusion
in Thoracic CT Volumes
Ref. Yao J, Bliton J, Summers RM. Automatic segmentation and measurement of pleural effusions on CT. IEEE
Transactions on Biomedical Engineering 2013, 60(7) pp 1834-1840.
E
Significance and Workflow
ī‚§ Pleural effusion:
ī‚§ Size, location and temporal change
can be significant for diagnosis and
patient care
ī‚§ Very time consuming to manually
measure in 3D chest CT
E
Segmentation and Surface Modeling
Segmentation
before
deformable
surface
modeling
Segmentation
after
deformable
surface
modeling
Segmented PE
surface before
deformable
surface modeling
Segmented PE
surface after
deformable
surface modeling
E
Automated vs. Manual Segmentations
E
Comparison to Manual Segmentation
E
Example 7
Radiomics-Based Assessment of
Normal Lung Tissue Damage in
Radiation Therapy
E
Ref. Cunliffe AR, et al.: Lung texture in serial thoracic CT scans: Correlation with radiologist-defined severity of
acute changes following radiation therapy. Physics in Medicine and Biology 59: 5387–5398, 2014.
Ref. Cunliffe AR, et al.: Lung texture in serial thoracic computed tomography scans: Correlation of radiomics-based
features with radiation therapy dose and radiation pneumonitis development. International Journal of
Radiation Oncology â€ĸ Biology â€ĸ Physics 91: 1048–1056, 2015.
Purpose
ī‚§ Develop quantitative methods to measure dose-
dependent normal lung tissue damage in patients
who receive thoracic radiation therapy
ī‚§ Characterize CT scan appearance based on pixel values and
spatial relationship between pixels
ī‚§ Use these quantitative techniques to distinguish
between patients with and without radiation
pneumonitis E
Radiation-Induced Lung Damage
E
Pre-RT ~3 months post-RT
Texture Analysis
E
Mean -785 -385
Median -820 -348
St. dev. 148 234
IQR 90 296
Fractal dimension 3.0 2.7
Entropy of ripple-filtered region 4.0 7.1
Follow-upBaseline
Identified using
deformable registration
Results
ī‚§ Extent of texture change between scans increases with
severity of RT-induced damage E
Texture Analysis
ī‚§ Image texture change related to dose through the dose map
of the treatment planning scan
ETreatment planning dose mapBaseline
deformable registration
Results
E
ī‚§ Extent of texture change between scans increases with
increased dose
Patient Classification
E
ī‚§ Area under the ROC curve for the ability of texture features
to differentiate patients with and without radiation
pneumonitis
Mean AUC Across 20 Features
Low Dose
Medium
Dose
High Dose Fitted Slope
0.64 0.68 0.71 0.71
Summary
ī‚§ Identified a set of reliable texture features for use in lung
texture analysis applications
ī‚§ Fully automated method for quantitative analysis of
changes in lung parenchyma between serial CT scans
ī‚§ Demonstrated relationship between texture change and
ī‚§ Radiation dose
ī‚§ Severity of radiation therapy-induced damage
ī‚§ Radiation pneumonitis status
E

More Related Content

What's hot

Study on Physicians Request for Computed Tomography Examinations for Patients...
Study on Physicians Request for Computed Tomography Examinations for Patients...Study on Physicians Request for Computed Tomography Examinations for Patients...
Study on Physicians Request for Computed Tomography Examinations for Patients...IRJESJOURNAL
 
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...Carestream
 
IHE / RSNA Image Sharing Project - IHE Colombia Workshop (12/2014) Module 5c
IHE / RSNA Image Sharing Project - IHE Colombia Workshop (12/2014) Module 5cIHE / RSNA Image Sharing Project - IHE Colombia Workshop (12/2014) Module 5c
IHE / RSNA Image Sharing Project - IHE Colombia Workshop (12/2014) Module 5cIHE Brasil
 
COMPLETE OVERVIEW ON ADAPTIVE RADIOTHERAPY OVER DAILY IMAGE GUIDED RADIOTHERAPY
COMPLETE OVERVIEW ON ADAPTIVE RADIOTHERAPY OVER DAILY IMAGE GUIDED RADIOTHERAPYCOMPLETE OVERVIEW ON ADAPTIVE RADIOTHERAPY OVER DAILY IMAGE GUIDED RADIOTHERAPY
COMPLETE OVERVIEW ON ADAPTIVE RADIOTHERAPY OVER DAILY IMAGE GUIDED RADIOTHERAPYSubrata Roy
 
3DP France
3DP France3DP France
3DP FranceJenny Chen
 
Emerson et al-2013-journal_of_ultrasound_in_medicine
Emerson et al-2013-journal_of_ultrasound_in_medicineEmerson et al-2013-journal_of_ultrasound_in_medicine
Emerson et al-2013-journal_of_ultrasound_in_medicinelinhnguyen1927
 
High Dimensional Fused-Informatics
High Dimensional Fused-InformaticsHigh Dimensional Fused-Informatics
High Dimensional Fused-InformaticsJoel Saltz
 
The Patient Journey with AI
The Patient Journey with AIThe Patient Journey with AI
The Patient Journey with AIImagia Cybernetics
 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Carestream
 
Challenges in quality management of diagnostic medical imaging
Challenges in quality management of diagnostic medical imagingChallenges in quality management of diagnostic medical imaging
Challenges in quality management of diagnostic medical imagingMohammad Fathi
 
Artificial Intelligence in Radiation Oncology
Artificial Intelligence in Radiation OncologyArtificial Intelligence in Radiation Oncology
Artificial Intelligence in Radiation OncologyWookjin Choi
 
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaPatient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaRamzee Small
 
IRJET - Automatic Detection of Diabetic Retinopathy in Retinal Image
IRJET -  	  Automatic Detection of Diabetic Retinopathy in Retinal ImageIRJET -  	  Automatic Detection of Diabetic Retinopathy in Retinal Image
IRJET - Automatic Detection of Diabetic Retinopathy in Retinal ImageIRJET Journal
 
Igrt prostate
Igrt prostateIgrt prostate
Igrt prostateJyoti Sharma
 
Quantitative Cancer Image Analysis
Quantitative Cancer Image AnalysisQuantitative Cancer Image Analysis
Quantitative Cancer Image AnalysisWookjin Choi
 
PERFORMANCE OF DATA MINING TECHNIQUES TO PREDICT IN HEALTHCARE CASE STUDY: CH...
PERFORMANCE OF DATA MINING TECHNIQUES TO PREDICT IN HEALTHCARE CASE STUDY: CH...PERFORMANCE OF DATA MINING TECHNIQUES TO PREDICT IN HEALTHCARE CASE STUDY: CH...
PERFORMANCE OF DATA MINING TECHNIQUES TO PREDICT IN HEALTHCARE CASE STUDY: CH...ijdms
 

What's hot (20)

Study on Physicians Request for Computed Tomography Examinations for Patients...
Study on Physicians Request for Computed Tomography Examinations for Patients...Study on Physicians Request for Computed Tomography Examinations for Patients...
Study on Physicians Request for Computed Tomography Examinations for Patients...
 
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
 
IHE / RSNA Image Sharing Project - IHE Colombia Workshop (12/2014) Module 5c
IHE / RSNA Image Sharing Project - IHE Colombia Workshop (12/2014) Module 5cIHE / RSNA Image Sharing Project - IHE Colombia Workshop (12/2014) Module 5c
IHE / RSNA Image Sharing Project - IHE Colombia Workshop (12/2014) Module 5c
 
COMPLETE OVERVIEW ON ADAPTIVE RADIOTHERAPY OVER DAILY IMAGE GUIDED RADIOTHERAPY
COMPLETE OVERVIEW ON ADAPTIVE RADIOTHERAPY OVER DAILY IMAGE GUIDED RADIOTHERAPYCOMPLETE OVERVIEW ON ADAPTIVE RADIOTHERAPY OVER DAILY IMAGE GUIDED RADIOTHERAPY
COMPLETE OVERVIEW ON ADAPTIVE RADIOTHERAPY OVER DAILY IMAGE GUIDED RADIOTHERAPY
 
3DP France
3DP France3DP France
3DP France
 
Emerson et al-2013-journal_of_ultrasound_in_medicine
Emerson et al-2013-journal_of_ultrasound_in_medicineEmerson et al-2013-journal_of_ultrasound_in_medicine
Emerson et al-2013-journal_of_ultrasound_in_medicine
 
High Dimensional Fused-Informatics
High Dimensional Fused-InformaticsHigh Dimensional Fused-Informatics
High Dimensional Fused-Informatics
 
IMAGE GUIDED SURGERY
IMAGE GUIDED SURGERYIMAGE GUIDED SURGERY
IMAGE GUIDED SURGERY
 
The Patient Journey with AI
The Patient Journey with AIThe Patient Journey with AI
The Patient Journey with AI
 
Random and systematic errors 25.10.12
Random and systematic errors 25.10.12Random and systematic errors 25.10.12
Random and systematic errors 25.10.12
 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
 
Challenges in quality management of diagnostic medical imaging
Challenges in quality management of diagnostic medical imagingChallenges in quality management of diagnostic medical imaging
Challenges in quality management of diagnostic medical imaging
 
Artificial Intelligence in Radiation Oncology
Artificial Intelligence in Radiation OncologyArtificial Intelligence in Radiation Oncology
Artificial Intelligence in Radiation Oncology
 
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaPatient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
 
IRJET - Automatic Detection of Diabetic Retinopathy in Retinal Image
IRJET -  	  Automatic Detection of Diabetic Retinopathy in Retinal ImageIRJET -  	  Automatic Detection of Diabetic Retinopathy in Retinal Image
IRJET - Automatic Detection of Diabetic Retinopathy in Retinal Image
 
Assessment
AssessmentAssessment
Assessment
 
Igrt prostate
Igrt prostateIgrt prostate
Igrt prostate
 
Quantitative Cancer Image Analysis
Quantitative Cancer Image AnalysisQuantitative Cancer Image Analysis
Quantitative Cancer Image Analysis
 
PERFORMANCE OF DATA MINING TECHNIQUES TO PREDICT IN HEALTHCARE CASE STUDY: CH...
PERFORMANCE OF DATA MINING TECHNIQUES TO PREDICT IN HEALTHCARE CASE STUDY: CH...PERFORMANCE OF DATA MINING TECHNIQUES TO PREDICT IN HEALTHCARE CASE STUDY: CH...
PERFORMANCE OF DATA MINING TECHNIQUES TO PREDICT IN HEALTHCARE CASE STUDY: CH...
 
Cyber Knife
Cyber KnifeCyber Knife
Cyber Knife
 

Viewers also liked

Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging StrategyWhitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging StrategyCarestream
 
Clinical Data Collaboration Across the Enterprise
Clinical Data Collaboration Across the EnterpriseClinical Data Collaboration Across the Enterprise
Clinical Data Collaboration Across the EnterpriseCarestream
 
Impacto assistencial dos laudos estruturados em radiologia: revisÃŖo integrativa
Impacto assistencial dos laudos estruturados em radiologia: revisÃŖo integrativaImpacto assistencial dos laudos estruturados em radiologia: revisÃŖo integrativa
Impacto assistencial dos laudos estruturados em radiologia: revisÃŖo integrativaThiago Julio, MD
 
Intel next-generation-medical-imaging-data-and-analytics
Intel next-generation-medical-imaging-data-and-analyticsIntel next-generation-medical-imaging-data-and-analytics
Intel next-generation-medical-imaging-data-and-analyticsCarestream
 
Carestream VNA, an Oxymoron?
Carestream VNA, an Oxymoron?Carestream VNA, an Oxymoron?
Carestream VNA, an Oxymoron?Carestream
 
useR2011 - Whitcher
useR2011 - WhitcheruseR2011 - Whitcher
useR2011 - Whitcherrusersla
 
Radiology nurse kpi
Radiology nurse kpiRadiology nurse kpi
Radiology nurse kpiasfoskenfur
 
The Case for Imaging Analytics
The Case for Imaging AnalyticsThe Case for Imaging Analytics
The Case for Imaging AnalyticsJoachim Surich
 
Radiomics Data Management, Computation, and Analysis for QIN F2F 2016
Radiomics Data Management, Computation, and Analysis for QIN F2F 2016Radiomics Data Management, Computation, and Analysis for QIN F2F 2016
Radiomics Data Management, Computation, and Analysis for QIN F2F 2016Ashish Sharma
 
Conquest for dummies: how to setup a pacs server
Conquest for dummies: how to setup a pacs serverConquest for dummies: how to setup a pacs server
Conquest for dummies: how to setup a pacs serverBogazici University
 
Radiomics: Novel Paradigm of Deep Learning for Clinical Decision Support towa...
Radiomics: Novel Paradigm of Deep Learning for Clinical Decision Support towa...Radiomics: Novel Paradigm of Deep Learning for Clinical Decision Support towa...
Radiomics: Novel Paradigm of Deep Learning for Clinical Decision Support towa...Wookjin Choi
 
InteligÃĒncia de NegÃŗcios em Radiologia / Business Intelligence in Radiology
InteligÃĒncia de NegÃŗcios em Radiologia / Business Intelligence in Radiology InteligÃĒncia de NegÃŗcios em Radiologia / Business Intelligence in Radiology
InteligÃĒncia de NegÃŗcios em Radiologia / Business Intelligence in Radiology Thiago Julio, MD
 
CT computer aided diagnosis system
CT computer aided diagnosis systemCT computer aided diagnosis system
CT computer aided diagnosis systemAboul Ella Hassanien
 
computer aided detection of pulmonary nodules in ct scans
computer aided detection of pulmonary nodules in ct scanscomputer aided detection of pulmonary nodules in ct scans
computer aided detection of pulmonary nodules in ct scansWookjin Choi
 
Thesis seminar
Thesis seminarThesis seminar
Thesis seminargvesom
 
kelainan kongenital muskuloskeletalblog
kelainan kongenital muskuloskeletalblogkelainan kongenital muskuloskeletalblog
kelainan kongenital muskuloskeletalblogyudhasetya01
 
Image processing in lung cancer screening and treatment
Image processing in lung cancer screening and treatmentImage processing in lung cancer screening and treatment
Image processing in lung cancer screening and treatmentWookjin Choi
 
DICOM Structure Basics
DICOM Structure BasicsDICOM Structure Basics
DICOM Structure BasicsGunjan Patel
 
Breast Cancer - A Study (Early Detection)
Breast Cancer - A Study  (Early Detection)Breast Cancer - A Study  (Early Detection)
Breast Cancer - A Study (Early Detection)Mathankumar S
 

Viewers also liked (19)

Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging StrategyWhitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
 
Clinical Data Collaboration Across the Enterprise
Clinical Data Collaboration Across the EnterpriseClinical Data Collaboration Across the Enterprise
Clinical Data Collaboration Across the Enterprise
 
Impacto assistencial dos laudos estruturados em radiologia: revisÃŖo integrativa
Impacto assistencial dos laudos estruturados em radiologia: revisÃŖo integrativaImpacto assistencial dos laudos estruturados em radiologia: revisÃŖo integrativa
Impacto assistencial dos laudos estruturados em radiologia: revisÃŖo integrativa
 
Intel next-generation-medical-imaging-data-and-analytics
Intel next-generation-medical-imaging-data-and-analyticsIntel next-generation-medical-imaging-data-and-analytics
Intel next-generation-medical-imaging-data-and-analytics
 
Carestream VNA, an Oxymoron?
Carestream VNA, an Oxymoron?Carestream VNA, an Oxymoron?
Carestream VNA, an Oxymoron?
 
useR2011 - Whitcher
useR2011 - WhitcheruseR2011 - Whitcher
useR2011 - Whitcher
 
Radiology nurse kpi
Radiology nurse kpiRadiology nurse kpi
Radiology nurse kpi
 
The Case for Imaging Analytics
The Case for Imaging AnalyticsThe Case for Imaging Analytics
The Case for Imaging Analytics
 
Radiomics Data Management, Computation, and Analysis for QIN F2F 2016
Radiomics Data Management, Computation, and Analysis for QIN F2F 2016Radiomics Data Management, Computation, and Analysis for QIN F2F 2016
Radiomics Data Management, Computation, and Analysis for QIN F2F 2016
 
Conquest for dummies: how to setup a pacs server
Conquest for dummies: how to setup a pacs serverConquest for dummies: how to setup a pacs server
Conquest for dummies: how to setup a pacs server
 
Radiomics: Novel Paradigm of Deep Learning for Clinical Decision Support towa...
Radiomics: Novel Paradigm of Deep Learning for Clinical Decision Support towa...Radiomics: Novel Paradigm of Deep Learning for Clinical Decision Support towa...
Radiomics: Novel Paradigm of Deep Learning for Clinical Decision Support towa...
 
InteligÃĒncia de NegÃŗcios em Radiologia / Business Intelligence in Radiology
InteligÃĒncia de NegÃŗcios em Radiologia / Business Intelligence in Radiology InteligÃĒncia de NegÃŗcios em Radiologia / Business Intelligence in Radiology
InteligÃĒncia de NegÃŗcios em Radiologia / Business Intelligence in Radiology
 
CT computer aided diagnosis system
CT computer aided diagnosis systemCT computer aided diagnosis system
CT computer aided diagnosis system
 
computer aided detection of pulmonary nodules in ct scans
computer aided detection of pulmonary nodules in ct scanscomputer aided detection of pulmonary nodules in ct scans
computer aided detection of pulmonary nodules in ct scans
 
Thesis seminar
Thesis seminarThesis seminar
Thesis seminar
 
kelainan kongenital muskuloskeletalblog
kelainan kongenital muskuloskeletalblogkelainan kongenital muskuloskeletalblog
kelainan kongenital muskuloskeletalblog
 
Image processing in lung cancer screening and treatment
Image processing in lung cancer screening and treatmentImage processing in lung cancer screening and treatment
Image processing in lung cancer screening and treatment
 
DICOM Structure Basics
DICOM Structure BasicsDICOM Structure Basics
DICOM Structure Basics
 
Breast Cancer - A Study (Early Detection)
Breast Cancer - A Study  (Early Detection)Breast Cancer - A Study  (Early Detection)
Breast Cancer - A Study (Early Detection)
 

Similar to The Complementary Roles of Computer-Aided Diagnosis and Quantitative Image Analysis

Comparison of the CDE guidance and the FDA guidance for antitumor clinical tr...
Comparison of the CDE guidance and the FDA guidance for antitumor clinical tr...Comparison of the CDE guidance and the FDA guidance for antitumor clinical tr...
Comparison of the CDE guidance and the FDA guidance for antitumor clinical tr...KanakoIshiguro
 
How to establish and evaluate clinical prediction models - Statswork
How to establish and evaluate clinical prediction models - StatsworkHow to establish and evaluate clinical prediction models - Statswork
How to establish and evaluate clinical prediction models - StatsworkStats Statswork
 
How to establish and evaluate clinical prediction models - Statswork
How to establish and evaluate clinical prediction models - StatsworkHow to establish and evaluate clinical prediction models - Statswork
How to establish and evaluate clinical prediction models - StatsworkStats Statswork
 
KCVI_analytics_NGWeiskopf research week v4
KCVI_analytics_NGWeiskopf research week v4KCVI_analytics_NGWeiskopf research week v4
KCVI_analytics_NGWeiskopf research week v4Deborah Woodcock
 
IRJET - A Conceptual Method for Breast Tumor ClassiīŦcation using SHAP Values ...
IRJET - A Conceptual Method for Breast Tumor ClassiīŦcation using SHAP Values ...IRJET - A Conceptual Method for Breast Tumor ClassiīŦcation using SHAP Values ...
IRJET - A Conceptual Method for Breast Tumor ClassiīŦcation using SHAP Values ...IRJET Journal
 
Ror editorial advance magazine (v3)
Ror editorial   advance magazine (v3)Ror editorial   advance magazine (v3)
Ror editorial advance magazine (v3)wtroyer
 
Pharma Quality Metrics
Pharma Quality MetricsPharma Quality Metrics
Pharma Quality MetricsMarcep Inc.
 
PACS AND IMAGING INFORMATICS -2018
PACS AND IMAGING INFORMATICS -2018PACS AND IMAGING INFORMATICS -2018
PACS AND IMAGING INFORMATICS -2018Arizona ACS-Healthcare
 
[Typ]Poster[Sbj]1593Synoptics[Dte]20150906
[Typ]Poster[Sbj]1593Synoptics[Dte]20150906[Typ]Poster[Sbj]1593Synoptics[Dte]20150906
[Typ]Poster[Sbj]1593Synoptics[Dte]20150906Mark Gusack
 
Care expert assistant for Medicare system using Machine learning
Care expert assistant for Medicare system using Machine learningCare expert assistant for Medicare system using Machine learning
Care expert assistant for Medicare system using Machine learningIRJET Journal
 
Evaluation methods in heathcare systems
Evaluation methods in heathcare systemsEvaluation methods in heathcare systems
Evaluation methods in heathcare systemsMarsa Gholamzadeh
 
Sepsis Prediction Using Machine Learning
Sepsis Prediction Using Machine LearningSepsis Prediction Using Machine Learning
Sepsis Prediction Using Machine LearningIRJET Journal
 
Predicting Life Expectancy of Hepatitis B Patients
Predicting Life Expectancy of Hepatitis B PatientsPredicting Life Expectancy of Hepatitis B Patients
Predicting Life Expectancy of Hepatitis B Patientsnabeelali11101999
 
lecture 10a
lecture 10alecture 10a
lecture 10aCMDLMS
 
Copy of Copy of Research_krishna_part.pdf
Copy of Copy of Research_krishna_part.pdfCopy of Copy of Research_krishna_part.pdf
Copy of Copy of Research_krishna_part.pdfKrishna840372
 
IRJET - Breast Cancer Prediction using Supervised Machine Learning Algorithms...
IRJET - Breast Cancer Prediction using Supervised Machine Learning Algorithms...IRJET - Breast Cancer Prediction using Supervised Machine Learning Algorithms...
IRJET - Breast Cancer Prediction using Supervised Machine Learning Algorithms...IRJET Journal
 
08 Am09 Presentations Kliger
08 Am09 Presentations   Kliger08 Am09 Presentations   Kliger
08 Am09 Presentations KligerSimon Prince
 
Accelerating Medical Labeling for Diagnostic AI: Add Non-Clinicians to Labeli...
Accelerating Medical Labeling for Diagnostic AI: Add Non-Clinicians to Labeli...Accelerating Medical Labeling for Diagnostic AI: Add Non-Clinicians to Labeli...
Accelerating Medical Labeling for Diagnostic AI: Add Non-Clinicians to Labeli...Cognizant
 
Big data, RWE and AI in Clinical Trials made simple
Big data, RWE and AI in Clinical Trials made simpleBig data, RWE and AI in Clinical Trials made simple
Big data, RWE and AI in Clinical Trials made simpleHadas Jacoby
 
Breast Cancer Prediction using Machine Learning
Breast Cancer Prediction using Machine LearningBreast Cancer Prediction using Machine Learning
Breast Cancer Prediction using Machine LearningIRJET Journal
 

Similar to The Complementary Roles of Computer-Aided Diagnosis and Quantitative Image Analysis (20)

Comparison of the CDE guidance and the FDA guidance for antitumor clinical tr...
Comparison of the CDE guidance and the FDA guidance for antitumor clinical tr...Comparison of the CDE guidance and the FDA guidance for antitumor clinical tr...
Comparison of the CDE guidance and the FDA guidance for antitumor clinical tr...
 
How to establish and evaluate clinical prediction models - Statswork
How to establish and evaluate clinical prediction models - StatsworkHow to establish and evaluate clinical prediction models - Statswork
How to establish and evaluate clinical prediction models - Statswork
 
How to establish and evaluate clinical prediction models - Statswork
How to establish and evaluate clinical prediction models - StatsworkHow to establish and evaluate clinical prediction models - Statswork
How to establish and evaluate clinical prediction models - Statswork
 
KCVI_analytics_NGWeiskopf research week v4
KCVI_analytics_NGWeiskopf research week v4KCVI_analytics_NGWeiskopf research week v4
KCVI_analytics_NGWeiskopf research week v4
 
IRJET - A Conceptual Method for Breast Tumor ClassiīŦcation using SHAP Values ...
IRJET - A Conceptual Method for Breast Tumor ClassiīŦcation using SHAP Values ...IRJET - A Conceptual Method for Breast Tumor ClassiīŦcation using SHAP Values ...
IRJET - A Conceptual Method for Breast Tumor ClassiīŦcation using SHAP Values ...
 
Ror editorial advance magazine (v3)
Ror editorial   advance magazine (v3)Ror editorial   advance magazine (v3)
Ror editorial advance magazine (v3)
 
Pharma Quality Metrics
Pharma Quality MetricsPharma Quality Metrics
Pharma Quality Metrics
 
PACS AND IMAGING INFORMATICS -2018
PACS AND IMAGING INFORMATICS -2018PACS AND IMAGING INFORMATICS -2018
PACS AND IMAGING INFORMATICS -2018
 
[Typ]Poster[Sbj]1593Synoptics[Dte]20150906
[Typ]Poster[Sbj]1593Synoptics[Dte]20150906[Typ]Poster[Sbj]1593Synoptics[Dte]20150906
[Typ]Poster[Sbj]1593Synoptics[Dte]20150906
 
Care expert assistant for Medicare system using Machine learning
Care expert assistant for Medicare system using Machine learningCare expert assistant for Medicare system using Machine learning
Care expert assistant for Medicare system using Machine learning
 
Evaluation methods in heathcare systems
Evaluation methods in heathcare systemsEvaluation methods in heathcare systems
Evaluation methods in heathcare systems
 
Sepsis Prediction Using Machine Learning
Sepsis Prediction Using Machine LearningSepsis Prediction Using Machine Learning
Sepsis Prediction Using Machine Learning
 
Predicting Life Expectancy of Hepatitis B Patients
Predicting Life Expectancy of Hepatitis B PatientsPredicting Life Expectancy of Hepatitis B Patients
Predicting Life Expectancy of Hepatitis B Patients
 
lecture 10a
lecture 10alecture 10a
lecture 10a
 
Copy of Copy of Research_krishna_part.pdf
Copy of Copy of Research_krishna_part.pdfCopy of Copy of Research_krishna_part.pdf
Copy of Copy of Research_krishna_part.pdf
 
IRJET - Breast Cancer Prediction using Supervised Machine Learning Algorithms...
IRJET - Breast Cancer Prediction using Supervised Machine Learning Algorithms...IRJET - Breast Cancer Prediction using Supervised Machine Learning Algorithms...
IRJET - Breast Cancer Prediction using Supervised Machine Learning Algorithms...
 
08 Am09 Presentations Kliger
08 Am09 Presentations   Kliger08 Am09 Presentations   Kliger
08 Am09 Presentations Kliger
 
Accelerating Medical Labeling for Diagnostic AI: Add Non-Clinicians to Labeli...
Accelerating Medical Labeling for Diagnostic AI: Add Non-Clinicians to Labeli...Accelerating Medical Labeling for Diagnostic AI: Add Non-Clinicians to Labeli...
Accelerating Medical Labeling for Diagnostic AI: Add Non-Clinicians to Labeli...
 
Big data, RWE and AI in Clinical Trials made simple
Big data, RWE and AI in Clinical Trials made simpleBig data, RWE and AI in Clinical Trials made simple
Big data, RWE and AI in Clinical Trials made simple
 
Breast Cancer Prediction using Machine Learning
Breast Cancer Prediction using Machine LearningBreast Cancer Prediction using Machine Learning
Breast Cancer Prediction using Machine Learning
 

More from Carestream

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Carestream
 
Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationCarestream
 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareCarestream
 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...Carestream
 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsCarestream
 
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographySmart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographyCarestream
 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-rayCarestream
 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentCarestream
 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityCarestream
 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Carestream
 
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Carestream
 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesCarestream
 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seramCarestream
 
Cloud Security
Cloud Security Cloud Security
Cloud Security Carestream
 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareCarestream
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundCarestream
 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Carestream
 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Carestream
 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Carestream
 
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA163 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA16Carestream
 

More from Carestream (20)

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
 
Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of Operation
 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization Software
 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
 
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographySmart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-ray
 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray Equipment
 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image Quality
 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions
 
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seram
 
Cloud Security
Cloud Security Cloud Security
Cloud Security
 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
 
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA163 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
 

Recently uploaded

Bangalore call girl đŸ‘¯â€â™€ī¸@ Simran Independent Call Girls in Bangalore GIUXUZ...
Bangalore call girl  đŸ‘¯â€â™€ī¸@ Simran Independent Call Girls in Bangalore  GIUXUZ...Bangalore call girl  đŸ‘¯â€â™€ī¸@ Simran Independent Call Girls in Bangalore  GIUXUZ...
Bangalore call girl đŸ‘¯â€â™€ī¸@ Simran Independent Call Girls in Bangalore GIUXUZ...Gfnyt
 
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012Call Girls Service Gurgaon
 
Hot Call Girl In Chandigarh 👅đŸĨĩ 9053'900678 Call Girls Service In Chandigarh
Hot  Call Girl In Chandigarh 👅đŸĨĩ 9053'900678 Call Girls Service In ChandigarhHot  Call Girl In Chandigarh 👅đŸĨĩ 9053'900678 Call Girls Service In Chandigarh
Hot Call Girl In Chandigarh 👅đŸĨĩ 9053'900678 Call Girls Service In ChandigarhVip call girls In Chandigarh
 
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Niamh verma
 
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Call Girls Service Chandigarh Ayushi
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknowgragteena
 
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591adityaroy0215
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...indiancallgirl4rent
 
Udaipur Call Girls 📲 9999965857 Call Girl in Udaipur
Udaipur Call Girls 📲 9999965857 Call Girl in UdaipurUdaipur Call Girls 📲 9999965857 Call Girl in Udaipur
Udaipur Call Girls 📲 9999965857 Call Girl in Udaipurseemahedar019
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★indiancallgirl4rent
 
Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Vipesco
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
Vip Kolkata Call Girls Cossipore 👉 8250192130 âŖī¸đŸ’¯ Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 âŖī¸đŸ’¯ Available With Room 24×7Vip Kolkata Call Girls Cossipore 👉 8250192130 âŖī¸đŸ’¯ Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 âŖī¸đŸ’¯ Available With Room 24×7Miss joya
 
Dehradun Call Girls Service 08854095900 Real Russian Girls Looking Models
Dehradun Call Girls Service 08854095900 Real Russian Girls Looking ModelsDehradun Call Girls Service 08854095900 Real Russian Girls Looking Models
Dehradun Call Girls Service 08854095900 Real Russian Girls Looking Modelsindiancallgirl4rent
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591adityaroy0215
 
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...gurkirankumar98700
 
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 

Recently uploaded (20)

Bangalore call girl đŸ‘¯â€â™€ī¸@ Simran Independent Call Girls in Bangalore GIUXUZ...
Bangalore call girl  đŸ‘¯â€â™€ī¸@ Simran Independent Call Girls in Bangalore  GIUXUZ...Bangalore call girl  đŸ‘¯â€â™€ī¸@ Simran Independent Call Girls in Bangalore  GIUXUZ...
Bangalore call girl đŸ‘¯â€â™€ī¸@ Simran Independent Call Girls in Bangalore GIUXUZ...
 
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
 
Hot Call Girl In Chandigarh 👅đŸĨĩ 9053'900678 Call Girls Service In Chandigarh
Hot  Call Girl In Chandigarh 👅đŸĨĩ 9053'900678 Call Girls Service In ChandigarhHot  Call Girl In Chandigarh 👅đŸĨĩ 9053'900678 Call Girls Service In Chandigarh
Hot Call Girl In Chandigarh 👅đŸĨĩ 9053'900678 Call Girls Service In Chandigarh
 
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
 
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
 
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
 
Udaipur Call Girls 📲 9999965857 Call Girl in Udaipur
Udaipur Call Girls 📲 9999965857 Call Girl in UdaipurUdaipur Call Girls 📲 9999965857 Call Girl in Udaipur
Udaipur Call Girls 📲 9999965857 Call Girl in Udaipur
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
 
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
 
Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
Vip Kolkata Call Girls Cossipore 👉 8250192130 âŖī¸đŸ’¯ Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 âŖī¸đŸ’¯ Available With Room 24×7Vip Kolkata Call Girls Cossipore 👉 8250192130 âŖī¸đŸ’¯ Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 âŖī¸đŸ’¯ Available With Room 24×7
 
Dehradun Call Girls Service 08854095900 Real Russian Girls Looking Models
Dehradun Call Girls Service 08854095900 Real Russian Girls Looking ModelsDehradun Call Girls Service 08854095900 Real Russian Girls Looking Models
Dehradun Call Girls Service 08854095900 Real Russian Girls Looking Models
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
 
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
 
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 

The Complementary Roles of Computer-Aided Diagnosis and Quantitative Image Analysis

  • 1. The Complementary Roles of Computer-Aided Diagnosis and Quantitative Image Analysis: Similarities and Differences in Assessment, Quality Assurance and Training
  • 2. Berkman Sahiner1, PhD Samuel G. Armato III2, PhD Zhimin Huo3, PhD Heang-Ping Chan4, PhD Ronald M. Summers5, MD, PhD Nicholas Petrick1, PhD 1: US Food and Drug Administration, Center for Devices and Radiological Health 2: University of Chicago, Department of Radiology 3: Carestream Inc. 4: University of Michigan, Department of Radiology 5: National Institutes of Health, Clinical Center
  • 3. Introduction ī‚§ Quantitative image analysis (QIA) and computer-aided diagnosis (CAD) are closely linked ī‚§ This exhibit reviews the methods developed for assessment, quality assurance, and user training for CAD, and highlights parallels and distinctions between CAD and QIA ī‚§ Comparison between CAD and QIA in this computer exhibit is performed through several example applications developed by the authors or that are publicly available
  • 4. What is CAD? ī‚§ CAD systems incorporate pattern recognition and data analysis capabilities* and are intended to ī‚§ Mark regions of an image that may reveal specific abnormalities and alert the clinician to these regions during image interpretation (computer-aided detection (CADe) systems)** ī‚§ Provide to the clinician an assessment of disease, disease type, severity, stage, progression (computer-aided diagnosis (CADx) systems)** *Guidance for industry and FDA staff: “Clinical performance assessment: Considerations for computer-assisted detection devices applied to radiology images and radiology device data - premarket approval (PMA) and premarket notification [510(k)] submissions," (FDA, 2012). **N. Petrick et al., “Evaluation of computer-aided detection and diagnosis systems,” Med. Phys. 2013, 40:087001.
  • 5. What is QIA? ī‚§ QIA is the process of extraction of quantitative imaging biomarkers from medical images, typically involving computerized tools ī‚§ A quantitative imaging biomarker is an objectively measured characteristic derived from an in vivo image as an indicator of normal biological processes, pathogenic processes, or response to a therapeutic intervention* *Sullivan DC et al., “Metrology standards for quantitative Imaging biomarkers,” Radiology 2015 Aug 12:142202. (Epub ahead of print).
  • 6. What is Quantitative Imaging? ī‚§ Both QIA and quantitative imaging biomarkers are within the larger context of quantitative imaging, defined as ī‚§ the extraction of quantifiable features from medical images for the assessment of normal [findings] or the severity, degree of change, or status of a disease, injury, or chronic condition relative to normal [findings] ī‚§ Quantitative imaging includes the development, standardization, and optimization of anatomical, functional, and molecular imaging acquisition protocols, data analyses, display methods, and reporting structures *Quantitative Imaging Biomarkers Alliance. http://rsna.org/QIBA.aspx. Accessed Oct. 15, 2015.
  • 7. How are CAD and QIA Similar? ī‚§ Both aim at aiding clinicians through advanced image analysis techniques ī‚§ Both commonly use computer methods to extract features ī‚§ CAD may use other patient-related information, e.g., age, gender, risk factors, or non-image biomarkers ī‚§ CAD features may include quantitative imaging biomarkers, but can also be relative, ordinal features ī‚§ User interaction may be somewhat more prevalent in QIA (e.g., semi-automated segmentation) ī‚§ Both emphasize appropriate image acquisition protocols, display methods, training, and reporting
  • 8. How are CAD and QIA Different? ī‚§ In CAD, the emphasis is on how the CAD output can provide decision support to clinicians ī‚§ In full assessment, it is required to demonstrate that a CAD system aids the clinician ī‚§ May integrate imaging and non-imaging information or biomarkers ī‚§ Not necessary to establish a direct link between individual CAD features and a specific disease condition ī‚§ Standalone testing (performance of the CAD system alone) is nonetheless important ī‚§ In QIA, the emphasis is on the extraction of biomarkers ī‚§ Bias and variability across devices, patients and time are major considerations
  • 9. Assessment, Quality Assurance, and Training ī‚§ The differences and similarities between CAD and QIA have a number of implications for assessment, quality assurance (QA), and training ī‚§ We briefly summarize our work in CAD in these areas and draw parallels with QIA
  • 10. Assessment - CAD ī‚§ CAD systems are typically assessed by evaluating the effect of the system on clinicians ī‚§ Reader performance assessment: Evaluate performance of a clinician using CAD as part of the decision making process ī‚§ Retrospective multiple-reader multiple case (MRMC) studies ī‚§ Prospective field trials ī‚§ Standalone performance of a CAD system is also useful as a performance indicator, both in ī‚§ System design phase and ī‚§ Final system assessment
  • 11. Assessment - CAD ī‚§ A recent paper by AAPM outlined important considerations in CAD assessment including*: ī‚§ Data set selection ī‚§ Representativeness of the test set for the targeted population ī‚§ Reference standard and mark labeling ī‚§ Disease status of a case (sometimes includes location information) ī‚§ Standalone assessment metrics ī‚§ True-positive fraction, false-positive fraction pairs ī‚§ Receiver operating characteristic (ROC) curves ī‚§ Location-specific ROC ī‚§ Reader study design and analysis ī‚§ Prospective versus retrospective studies ī‚§ MRMC study design ī‚§ Reader training for reader studies *N. Petrick et al., “Evaluation of computer-aided detection and diagnosis systems,” Med. Phys. 2013, 40:087001.
  • 12. Assessment - Quantitative Imaging ī‚§ Most groups working in quantitative imaging are interested in technical performance assessment ī‚§ How a test performs in reference objects or subjects under controlled conditions ī‚§ Other quantitative imaging assessment ī‚§ Clinical assessment ī‚§ Clinical impact
  • 13. Technical Assessment - Quantitative Imaging ī‚§ Technical assessment can be performed using ī‚§ Anthropomorphic phantoms ī‚§ Large number of replicate measurements easily obtained ī‚§ The same phantoms can be used repeatedly ī‚§ Reference standard easier to obtain ī‚§ Phantom realism may be limited ī‚§ Generalizability to human data may be limited because of limited abnormality variation and complexity ī‚§ Patient data ī‚§ More realistic ī‚§ Number of replicate measurements may be limited ī‚§ Reference standard more difficult to obtain ī‚§ Number of available patients usually limited
  • 14. Technical Assessment - Quantitative Imaging ī‚§ Important considerations include* ī‚§ Measurand/Reference ī‚§ True value of the quantity intended to be measured ī‚§ Bias ī‚§ Difference between the expected value of the biomarker and the measurand ī‚§ Linearity ī‚§ Is a change in the measurand reflected as a proportional change in the biomarker? ī‚§ Repeatability ī‚§ Ability to repeatedly measure the same feature under identical or near-identical conditions ī‚§ Reproducibility ī‚§ Ability to measure the same feature under different conditions expected in ī‚§ A preclinical study ī‚§ Clinical trial ī‚§ Clinical practice D.L. Raunig et al., “Quantitative imaging biomarkers: a review of statistical methods for technical performance assessment,” Stat Methods Med Res. 2015, 24:27-67.
  • 15. Clinical Assessment - Quantitative Imaging ī‚§ Clinical assessment of a QIA method ī‚§ How does the method help clinicians in diagnosis or treatment? ī‚§ Recommendations developed for clinical assessment of CAD may be useful for the clinical assessment of quantitative imaging biomarkers after proper modifications ī‚§ Methodology for prospective and retrospective studies ī‚§ MRMC methods ī‚§ Assessment metrics
  • 16. Quality Assurance - CAD ī‚§ CAD systems are medical devices, and can benefit from QA procedures like all medical devices ī‚§ Clinical image acquisition: ī‚§ Follow manufacturer’s QA procedure for imaging device ī‚§ Additional QA for CAD: ī‚§ Assure functionality and performance of CAD device according to vendor’s specifications ī‚§ Assure consistency of CAD device performance over time ī‚§ A recent paper by AAPM outlined important considerations in CAD quality assurance* *Z. Huo, et al., “Quality assurance and training procedures for computer-aided detection and diagnosis systems,” Med. Phys. 2013, 40:077001.
  • 17. Quality Assurance - Quantitative Imaging ī‚§ Clinical image acquisition: ī‚§ Follow manufacturer’s QA procedure for imaging device ī‚§ Image acquisition for quantitative imaging: ī‚§ QA procedures to reduce variability across devices, patients and time ī‚§ QA procedures to ensure that acquired images are of quantitative quality ī‚§ Calibration methods and phantom tests in addition to those recommended by manufacturers for clinical image quality
  • 18. Clinician Training - CAD ī‚§ Inform clinicians about ī‚§ Intended use of the CAD system ī‚§ CAD system’s limitations ī‚§ Particular strengths and weaknesses of different CAD systems ī‚§ E.g., with respect to lesion type, acquisition parameters ī‚§ Although many clinicians and researchers have underlined the need for clinician training for CAD ī‚§ Additional awareness needs to be raised for proper use and potential adverse effects of improper use ī‚§ Best practices for CAD training are yet to be established
  • 19. Clinician Training - Quantitative Imaging ī‚§ Clinicians using the quantitative imaging tool in the clinic should be qualified in accordance with local and national requirements and standards ī‚§ Training of operators in image acquisition and analysis can influence the value of the extracted imaging biomarker ī‚§ e.g., training of an operator in a semi-automated segmentation task for a lesion volume biomarker ī‚§ There is a small number of studies on clinician training and its effect on the extracted quantitative imaging biomarker
  • 20. â€ĸ The rest of the computer exhibit consists of a summary and example applications developed by the authors or that are publicly available â€ĸ Click to go to the slide of your choice â€ĸ Summary â€ĸ Example applications â€ĸ Click on the E on lower right from any slide to go to the list of example applications E
  • 21. Summary ī‚§ CAD and QIA fields have a number of shared attributes as well as some differences ī‚§ Methodologies developed in one field are likely to be applicable in the other after careful consideration of the differences ī‚§ We expect that the development of standardization, assessment and clinical use recommendations in these two fields will strengthen each other E
  • 22. Example Applications ī‚§ Example 1 ī‚§ CADe assessment: A commercial CAD system ī‚§ Example 2 ī‚§ Quantitative imaging technical assessment: A phantom study ī‚§ Example 3 ī‚§ Quantitative volume change analysis for treatment response assessment of head and neck neoplasms on CT scans ī‚§ Example 4 ī‚§ Computer-aided analysis of treatment response of bladder cancers on CT scans ī‚§ Example 5 ī‚§ Radiomic biomarkers and decision support system ī‚§ Example 6 ī‚§ Measurement of pleural effusion in thoracic CT volumes ī‚§ Example 7 ī‚§ Radiomics-based assessment of normal lung tissue damage in radiation therapy
  • 23. Example 1 CADe Assessment: A Commercial CADe System E
  • 24. A Commercial CADe System ī‚§ M-Vu Algorithm Engine ī‚§ CADe device intended to aid radiologists in reading screening mammograms ī‚§ Mark areas for review by a radiologist, who ī‚§ First reviews each case in the conventional manner, ī‚§ then re-examines regions marked by the M-Vu system before making a BI-RADS assessment ī‚§ Pre-market approval (PMA) by the FDA in 2012 E
  • 25. Test Data Set Selection ī‚§ 140 positive and 140 negative cases ī‚§ 11 sites in the United States ī‚§ Academic, specialty, and community clinics. ī‚§ Pre-defined inclusion and exclusion criteria ī‚§ Sequential, eligible positive mammograms ī‚§ A mammography exam for which a biopsy-proven breast cancer was found within 15 months following the exam date ī‚§ Eligible negative mammograms ī‚§ A mammography exam for which ī‚§ Breast cancer was not found within 15 months prior or 15 months after the exam date ī‚§ At least one associated subsequent negative mammogram was acquired at least 11 months after the exam date. E
  • 26. Reference Standard ī‚§ Positive cases: ī‚§ Pathology reports from biopsies or surgeries ī‚§ Mammograms and radiology reports for lesion location ī‚§ Each biopsy-proven lesion is outlined by site investigators ī‚§ Negative cases ī‚§ Radiology reports E
  • 27. Standalone Assessment: Sensitivity ī‚§ A malignant region was considered to be detected if ī‚§ A CAD mark centroid was inside the region, or ī‚§ if the region centroid was inside a CAD mark ī‚§ Having the cancer within the CAD mark, as above, can lead to an optimistic estimate of sensitivity ī‚§ A large mark may cover the lesion, but may not attract the attention of the radiologist to the abnormality ī‚§ When this is used as a detection criterion, other additional data are evaluated to check if sensitivity is overestimated ī‚§ A case was considered a true positive if ī‚§ At least one malignant region was detected by CAD E
  • 28. Standalone Assessment: Sensitivity Cases Sensitivity 95% Conf. Interval Overall 140 79.3% (72.6% 86.0%) Microcalcification 69 79.7% (70.2%, 89.2%) Mass 86 81.4% (73.2%, 89.6%) ī‚§ Sensitivity reported for two important sub-groups: Masses and microcalcifications ī‚§ Additional data on sensitivity with respect to ī‚§ Lesion size ī‚§ Lesion pathology (invasive, DCIS, other) ī‚§ Breast density E
  • 29. Standalone Performance: False Positives ī‚§ Additional data on ī‚§ False-positives by breast density ī‚§ Specificity (percentage of mammograms with no FP marks) False-positives per mammo. 95% Conf. Interval Total 0.418 (0.346, 0.490) Microcalcification 0.088 (0.042,0.133) Mass 0.330 (0.275, 0.385) E
  • 30. Pivotal Study ī‚§ Multi-reader multi-case (MRMC) study to investigate the effect of the CADe device on radiologists’ performance ī‚§ 21 radiologists from a variety of academic, specialty, and community clinics located across the US ī‚§ For each case: ī‚§ Evaluate the case without CAD and record an assessment ī‚§ View CAD marks ī‚§ Record a "with CAD" assessment E
  • 31. Pivotal Study ī‚§ With and without CAD assessments ī‚§ Recall/do not recall patient ī‚§ Screening BI-RADS (0, 1, 2, 3, 4a, 4b, 4c, 5) ī‚§ Forced BI-RADS (1, 2, 3, 4a, 4b, 4c, 5) ī‚§ Lesion findings ī‚§ Lesion findings: ī‚§ Laterality ī‚§ Type (Mass, Architectural Distortion, Asymmetry, microcalcification) ī‚§ BI-RADS (1, 2, 3, 4a, 4b, 4c, 5) ī‚§ Probability of Malignancy (POM) (0-100%) E
  • 32. Pivotal Study - ROC Analysis ī‚§ POM data analyzed using ROC methodology ī‚§ Area under the ROC curve (AUC) ī‚§ AUC w/0 CAD = 0.885 ī‚§ AUC with CAD = 0.902 ī‚§ Difference statistically significant with p=0.013 (DBM MRMC method) E
  • 33. Pivotal Study – Sens., Spec. ī‚§ Radiologist per-case sensitivity and specificity based on decision to recall W/o CAD With CAD Difference P-value Sensitivity 0.865 0.901 0.036 (0.014, 0.058) 0.002 Specificity 0.649 0.623 -0.026 (-0.039, -0.013) < 0.001 ī‚§ Sensitivity increased and specificity decreased with CAD ī‚§ Typical in reader studies for most second-read CADe systems ī‚§ Lack of CADe mark for a lesion does not dissuade a reader from recall ī‚§ Marks for missed lesions and some false-positive marks add to the number of recalled cases E
  • 34. Summary ī‚§ Components of CAD assessment ī‚§ Test data set selection ī‚§ Representative of the targeted population ī‚§ Reference standard ī‚§ Clear and accurate definitions of actual positives and negatives ī‚§ Standalone performance ī‚§ How does the CAD system alone perform in the task for which it is intended to help the radiologist ? ī‚§ Performance metrics, confidence intervals ī‚§ Reader performance ī‚§ What is the effect of the CAD system on readers? ī‚§ Performance metrics ī‚§ Performance with and without CAD ī‚§ Difference, confidence intervals E
  • 35. Example 2 Quantitative Imaging Technical Assessment: A Phantom Study E
  • 36. Designing an Analysis Plan* Step 1: Define the quantitative imaging biomarker and its relationship to quantity to be measured (measurand) Step 2: Define question to be addressed ī‚§ Hypothesis or bounds on performance Step 3: Define the experimental unit ī‚§ Lesion-level, patient-level Step 4: Define statistical measures of performance ī‚§ What are the metrics for bias, repeatability, reproducibility? Step 5: Specify elements of statistical design ī‚§ Reference, reproducibility conditions, etc. Step 6: Determine the data requirements ī‚§ Patient population, type of images, sample size, etc. Step 7: Collect data and perform statistical analysis *Raunig et al., Stat Methods Med Res, 2014 E
  • 39. Purpose of the Study ī‚§ Evaluate technical performance of a nodule volume estimation tool E
  • 40. Study Design ī‚§ Phantom ī‚§ Thorax phantom with vascular insert ī‚§ Synthetic nodules ī‚§ 4 spherical nodules ī‚§ 5, 8, 9, 10 mm â€ĸ +100 HU E
  • 41. Study Design ī‚§ Image collection protocol ī‚§ 10 repeat acquisitions Phantom 16×0.75 mm 100 mAs 0.75 mm 1.5 mm Reconstruction Acquisition Detailed Detailed E
  • 42. Bias/Linearity ī‚§ Visually assess data to define limits of quantitation ī‚§ Evaluate means/variances ī‚§ Is data transformation necessary/appropriate to stabilize the variance? -60 -40 -20 0 20 40 60 5 mm 8 mm 9 mm 10 mm Nodule Size 50 150 250 350 450 550 0.75 1.5 0.75 1.5 0.75 1.5 0.75 1.5 Slice Thickness (mm) MeasuredValue(mm3 ) 5 mm 8 mm 9 mm 10 mm Nodule Size MeasuredValue-Median(mm3) E
  • 43. Bias/Linearity 0 100 200 300 400 500 600 0 100 200 300 400 500 600 Reference Standard (mm3) MeasuredValue(mm3 ) E
  • 44. Repeatability/Reproducibility ī‚§ RC ī‚§ Estimated over 10 repeat acquisitions, 4 nodules ī‚§ RDC ī‚§ Estimated over 10 repeat acquisitions, 4 nodules, and 2 slice thicknesses Slice Thickness 0.75 mm 1.5 mm RC 14.4 mm3 27.6 mm3 RDC 29.0 mm3 E
  • 45. Repeatability Analysis 1.5 mm Slices 0 100 200 300 400 500 600 -60 -40 -20 0 20 40 60 Bland-Altman Plot (1.5 mm Slices) RC=27.62 mm3 0.75 mm Slices 0 100 200 300 400 500 600 -60 -40 -20 0 20 40 60 Bland-Altman Plot (0.75 mm Slices) RC=14.44 mm3 Mean(Esti,Estj) (mm3)Mean(Esti,Estj) (mm3) Diff(Esti,Estj)(mm3) Diff(Esti,Estj)(mm3) E
  • 46. Reproducibility Analysis 0 100 200 300 400 500 600 -60 -40 -20 0 20 40 60 Bland-Altman Plot (Reproducibility) Mean(Esti ) (mm,Estj 3 ) RDC=28.95 mm3 Diff(Esti,Estj)(mm3) E
  • 47. Summary ī‚§ Main components of quantitative imaging biomarker technical assessment ī‚§ Bias/linearity analysis ī‚§ Repeatability analysis ī‚§ Reproducibility analysis ī‚§ Others ī‚§ Identification of significant factors/subgroups ī‚§ â€Ļ. ī‚§ Challenging to maintain consistency across studies ī‚§ Phantom/clinical data ī‚§ Transformation of data ī‚§ Reference standard ī‚§ Is test-retest data available? ī‚§ Reproducibility conditions E
  • 48. Example 3 Quantitative Volume Change Analysis for Treatment Response Assessment of Head and Neck Neoplasms on CT Scans Ref. Hadjiiski L, Mukherji SK, Gujar SK, Sahiner B, Ibrahim M, Street E, Moyer J, Worden FP, Chan HP. Treatment response assessment of head and neck cancers on CT using computerized volume analysis. American Journal of Neuroradiology (AJNR) 2010;31(9):1744-1751. E
  • 49. Lesion Segmentation Region of Interest Preprocessing Level Set Automatic Segmentation E
  • 52. Pre-treatment [cm3] 0 10 20 30 40 50 Post-treatement[cm3] 0 10 20 30 40 50 Pre-treatment [cm3] 0 10 20 30 40 50 Post-treatement[cm3] 0 10 20 30 40 50 Volume Segmentation â€ĸ Pre-to-Post-treatment change in segmented primary tumor volumes – 23 pairs Auto Manual E
  • 53. Results - % Volume Change Comparison â€ĸ Automatic vs. Manual % volume change comparison – 23 tumor pairs - Automatic vs. Manual: r = 0.89 Pearson’s correlation r E% Volume Change Automatic -20 0 20 40 60 80 100 %VolumeChangeManual -20 0 20 40 60 80 100
  • 54. Results – Summary WHO (longest diameter): Primary tumors RECIST (longest diameter and perpendicular): 0.73 0.58 Pearson’s correlation coeff. â€ĸ % Pre-to-Post-treatment change in volume estimates by 3 methods, relative to radiologist’s segmentation: Quantitative image analysis: 0.89 E
  • 59. Example 4 Computer-aided Analysis of Treatment Response of Bladder Cancers on CT Scans Ref. Hadjiiski L, Weizer AZ, Alva A, Caoili EM, Cohan RH, Cha K, Chan HP. Bladder cancers on CT: preliminary study of treatment response assessment based on computerized volume analysis, WHO and RECIST Criteria. American Journal of Roentgenology 2015, 205(2) pp 348-352. E
  • 60. Bladder Lesion Segmentation Cascaded Level Set Region of Interest Automatic Segmentation Auto-Initialized Cascaded Level Set (AI-CALS) Preprocessing E
  • 61. Bladder Tumor: Pre- & Post-Treatment Pre- treatment Post- treatment E
  • 62. Feature Descriptors ī‚§ Descriptors automatically extracted from the segmented lesions: - 15 radiomic features (RF) based on pre- and post-treatment changes in: - volume (V) - 5 gray level descriptors (GL) - 9 shape descriptors (S) - Selected features merged into a Combined Response Index (CRI) E
  • 63. Combined Response Index (CRI) â€ĸ AUC pT0 stage (complete response) vs. others – 35 primary site tumors Volume (3D): 0.68 Âą 0.09 Reference (3D): 0.66 Âą 0.11 CRI (Volume, Shape): 0.75 Âą 0.10 CRI (Volume, Shape, Gray Level): 0.76 Âą 0.09 EFalse Positive Fraction 0.0 0.2 0.4 0.6 0.8 1.0 TruePositiveFraction 0.0 0.2 0.4 0.6 0.8 1.0 V Reference (3D) CRI (V, S) CRI (V, S, GL)
  • 64. Example 5 Radiomic Biomarkers and Decision Support System for Multiple Myeloma Ref. Zhou C, Chan HP, Dong Q, Couriel DR, Pawarode A, Hadjiiski LM, Wei J. Quantitative analysis of MR imaging to assess trestment response for patients with multiple myeloma by using dynamic intensity entropy transformation 2015, Ahead of Print, 10.1148/radiol.2015142804 E
  • 65. ī€ž Multiple Myeloma (MM) – a cancer formed by malignant plasma cells in the bone marrow o MR radiomic biomarkers (+ clinical biomarkers) - Staging - Treatment response assessment - Prognosis prediction - Recurrence – early detection E
  • 66. Radiomic Biomarkers for MM Treatment Response T1W sagittal views VOI of vertebral bodies and intervertebral discs 3D Dynamic Intensity Energy Transformation (DIET) Quantitative Energy Enhancement Value: Response Index DIET response map E p-qEEVvbr and m-DqEEVvbr
  • 67. T1-weighted MR DIET response map Pre-Bone Marrow Transplant 55 days post-BMT Pre-BMT 55 days post-BMT BMT Treatment Response Assessment Responder E
  • 68. 56 days post-BMT Pre-BMT 56 days post-BMT T1-weighted MR DIET response map Non-responder BMT Treatment Response Assessment Pre-Bone Marrow Transplant E
  • 69. Kaplan-Meier Survival Curve predicted by p-qEEV biomarker 0 5 10 15 20 25 30 35 0.2 0.4 0.6 0.8 1 Time Since BMT (Months) Survivalprobability p-qEEV >= 10% p-qEEV < 10% E Prediction of treatment response by qEEV-based response index Computer-Aided Decision Support for Multiple Myeloma
  • 70. Example 6 Measurement of Pleural Effusion in Thoracic CT Volumes Ref. Yao J, Bliton J, Summers RM. Automatic segmentation and measurement of pleural effusions on CT. IEEE Transactions on Biomedical Engineering 2013, 60(7) pp 1834-1840. E
  • 71. Significance and Workflow ī‚§ Pleural effusion: ī‚§ Size, location and temporal change can be significant for diagnosis and patient care ī‚§ Very time consuming to manually measure in 3D chest CT E
  • 72. Segmentation and Surface Modeling Segmentation before deformable surface modeling Segmentation after deformable surface modeling Segmented PE surface before deformable surface modeling Segmented PE surface after deformable surface modeling E
  • 73. Automated vs. Manual Segmentations E
  • 74. Comparison to Manual Segmentation E
  • 75. Example 7 Radiomics-Based Assessment of Normal Lung Tissue Damage in Radiation Therapy E Ref. Cunliffe AR, et al.: Lung texture in serial thoracic CT scans: Correlation with radiologist-defined severity of acute changes following radiation therapy. Physics in Medicine and Biology 59: 5387–5398, 2014. Ref. Cunliffe AR, et al.: Lung texture in serial thoracic computed tomography scans: Correlation of radiomics-based features with radiation therapy dose and radiation pneumonitis development. International Journal of Radiation Oncology â€ĸ Biology â€ĸ Physics 91: 1048–1056, 2015.
  • 76. Purpose ī‚§ Develop quantitative methods to measure dose- dependent normal lung tissue damage in patients who receive thoracic radiation therapy ī‚§ Characterize CT scan appearance based on pixel values and spatial relationship between pixels ī‚§ Use these quantitative techniques to distinguish between patients with and without radiation pneumonitis E
  • 78. Texture Analysis E Mean -785 -385 Median -820 -348 St. dev. 148 234 IQR 90 296 Fractal dimension 3.0 2.7 Entropy of ripple-filtered region 4.0 7.1 Follow-upBaseline Identified using deformable registration
  • 79. Results ī‚§ Extent of texture change between scans increases with severity of RT-induced damage E
  • 80. Texture Analysis ī‚§ Image texture change related to dose through the dose map of the treatment planning scan ETreatment planning dose mapBaseline deformable registration
  • 81. Results E ī‚§ Extent of texture change between scans increases with increased dose
  • 82. Patient Classification E ī‚§ Area under the ROC curve for the ability of texture features to differentiate patients with and without radiation pneumonitis Mean AUC Across 20 Features Low Dose Medium Dose High Dose Fitted Slope 0.64 0.68 0.71 0.71
  • 83. Summary ī‚§ Identified a set of reliable texture features for use in lung texture analysis applications ī‚§ Fully automated method for quantitative analysis of changes in lung parenchyma between serial CT scans ī‚§ Demonstrated relationship between texture change and ī‚§ Radiation dose ī‚§ Severity of radiation therapy-induced damage ī‚§ Radiation pneumonitis status E