White Paper | CARESTREAM DRX-Revolution Mobile X-ray System
The CARESTREAM Tube and Grid Alignment System
provides better image quality and consistent
techniques for portable diagnostic radiography
Anti-Scatter Grids Improve Image Quality
Scattered radiation is a major contributor to reduced image
quality in diagnostic radiography. Anti-scatter grids, when
properly aligned, provide an effective means for reducing
scatter [1-6]. The preferential transmission of primary versus
scattered X-rays improves the contrast-to-noise ratio (CNR).
Figure 1 shows a pair of portable chest images captured of an
ICU patient with and without grids and under the same
exposure conditions. Comparing the two images, the contrast
improvement with grids is significant.
Figure 1 -- Comparison of portable chest images captured of an ICU patient without (left) and with (right) anti-scatter grid
Grid Benefits Can be Realized with Less Exposure
Increments in Digital Radiography
The Bucky factor is the reciprocal of the total X-ray
penetration through an anti-scatter grid. The analog screen-
film system has a fixed sensitometry response. When a grid is
used, a Bucky factor technique increment was necessary to
ensure sufficient exposure to produce a film image with
proper density and contrast for diagnosis.
The benefits of anti-scatter grids can be realized in digital
radiography (DR) with less exposure technique increments
than would be indicated by traditional Bucky factors [7]. This
is because DR does not have a fixed sensitomety response;
the global contrast and brightness of a DR image can be
arbitrarily modified by digital image processing. DR systems
are fundamentally noise limited – within a broad range of
exposure levels to the patient, the image quality is
determined by the anatomy CNR in the captured image.
Anti-scatter grids can improve the image CNR at a minor
expense of reduced primary X-ray transmission. This suggests
some exposure technique increase may be required to
compensate for primary X-ray attenuation, but usually at a
lower level than the Bucky factor.
White Paper | CARESTREAM DRX-Revolution Mobile X-ray System
2
Challenges in Using Grids for Portable
Radiography
Grid usage in portable radiography is often sporadic and
inconsistent. This causes greater variability in image quality,
and a greater number of radiographs of poorer quality than
those captured in the radiology department being delivered
for interpretation.
From the perspective of the radiographic technologist, using
grids for portable exams involves a variety of
time-consuming workflow implications. These include
attaching and detaching the add-on grids to the X-ray
cassettes; the stringent requirements to properly position
and align the X-ray source relative to the cassette behind the
patient to avoid grid cutoff; the increased probability that
repeated exposures will be required due to grid-cutoff
artifact; and more.
In addition, there is the misperception that grids are not
required in digital radiography because increasing the
exposure can overcome the scatter-noise level, and that
image processing adjustments, such as window and level
manipulations, can sufficiently compensate for the quality
losses that are introduced by scattered radiation. With all of
these considerations in mind, there would seem to be little
motivation for the technologists to use grids in portable
digital radiography.
DRX-Revolution Tube and Grid Alignment System
The CARESTREAM Tube and Grid Alignment System (TGA)
for the CARESTREAM DRX-Revolution Mobile X-ray System
provides guidance to the radiographic technologist for
proper alignment of the X-ray source relative to the
grid/detector. This optional feature integrates seamlessly
with the DRX-Revolution System – no additional operational
steps are required in the normal workflow (Figure 2).
Two transmitters
mounted under the
collimator
Eight receivers inside
the grid holder frame
Figure 2 – The DRX-Revolution and schematics of the optional Tube
and Alignment System. The grid-alignment transmitters are built
inside the tube-head assembly, and the receivers are in the
grid holder.
Alignment
Status
Tube Angle
Detector Angle
Actual
Distance
Alignment
Status
Bed Type
Selection
Figure 3 – The tube-head display on the DRX-Revolution System
reports real-time X-ray source and grid/detector position
information for proper alignment.
Carestream’s TGA helps the radiographic technologist to
achieve better image quality and consistent imaging
technique for portable exams. It automatically detects the
X-ray source position relative to the grid holder/detector, and
then reports the amount of X-ray source displacements from
the proper alignment (Figure 3). The displacements are
presented in real time on the tube-head display in separate
up/down, left/right and in/out directions. The alignment
status turns green when the TGA decides that the X-ray
source is within the optimal grid-operation range, specific to
the target source-to-grid distance (SGD).
White Paper | CARESTREAM DRX-Revolution Mobile X-ray System
3
Alignment Technology Overview
The TGA uses multiple electromagnetic fields for position
triangulation. Two transmitters are mounted under the X-ray
collimator to generate the electromagnetic fields, and a total
of eight receivers are installed inside the grid holder around
the grid/detector to detect the fields from the transmitters
(Figure 2). The transmitters use magnetic coils to synthesize
magnetic fields that are functionally the same as those of
spinning, permanent bar magnets (Figure 4). The receivers
sense the spinning magnetic-field intensity and phase, and
then each generates a unique sine-wave signal (Figure 5).
N S
Receiver
Transmitter (Spinning Magnet)
Figure 4 – The Tube and Grid Alignment System works by detecting the intensity and phase of synthesized spinning magnetic fields. These
diagrams show three examples of different relative positions between a transmitter and a receiver.
Figure 5 – A snapshot of the detected magnetic-field signals from eight receivers
Sophisticated mathematical models are built inside the
software to decode the receiver sine-wave information. The
magnitude and phase of the sine wave from each receiver
are extracted from the electronic signals, and then compared
with their expected theoretical values, which are calculated
based on an initial assumption of the geometry between the
X-ray source and the grid holder/detector. A total error is
summed from the comparison differences for all the
receivers. A non-zero total error value indicates that the
initial geometry assumption is incorrect, to which fine
adjustments are made and the calculation process is
reiterated. The X-ray source position is best estimated when
White Paper | CARESTREAM DRX-Revolution Mobile X-ray System
www.carestream.com
Carestream Health, Inc., 2012. CARESTREAM is a trademark of
Carestream Health. CAT 200 0016 8.5x11 9/12
the total error is minimized. Finally, the estimated position
value is compared to the optimal grid-operation range, and
the X-ray source displacements from the proper alignment
are presented on the tube-head display.
The grid-alignment transmitters use amplitude-modulated
radio-frequency (RF) signals for extended operation range
and improved signal-to-noise ratio. Eddy current can be
induced by the transmitter RF signals in the metallic material
that is in close proximity to the receivers. The Eddy current
introduces secondary electromagnetic waves that can
negatively influence the accuracy of the grid-alignment
results. To overcome this issue, different bed types can be
selected on the software’s GUI to better compensate for the
interference. Specifically, the bed-type selection (intensive
care units vs. emergency departments) is primarily based on
the thickness of the mattress on top of the bed
metal frames.
Definition of Optimal Grid-Operation Range
The grid built inside the grid holder has a center focal
distance of 112 cm, a grid ratio of 8:1 and a frequency
resolution of 80 lines per centimeter.
The optimal grid-operation range built inside the TGA
software is determined by the minimum performance
requirement for image-quality improvement. The grid
performance is characterized in terms of the signal-to-noise
ratio improvement factor (SIF)[8]. The higher the SIF value,
the better the performance. The benefit of using the grid
diminishes when the SIF value equals 1.0. The optimal
grid-operation range at a particular SGD is defined as the
maximum allowable lateral (across the grid-line direction)
misalignment in the X-ray source position, at which the
minimum SIF value across the detector’s effective imaging
area would be greater than 1.0.
Summary
Anti-scatter grids improve radiography image quality, and
the benefit of grid usage can be realized in digital
radiography with less exposure technique increments. The
CARESTREAM Tube and Grid Alignment System for the
DRX-Revolution is fully integrated with the radiographic
technologist’s normal workflow, and provides easy and
intuitive guidance for X-ray-source alignment to achieve
consistent and optimal image quality.
References
1. H.-P. Chan, K. L. Lam, and Y. Wu, “Studies of Performance of Anti-Scatter Grids in Digital Radiography: Effect on Signal-
to-Noise Ratio,” Medical Physics, 17(4), 655-664 (1990).
2. L. N. Rill, L. Brateman and M. Arreola, “Evaluating Radiographic Parameters for Mobile Chest Computed Radiography:
Phantoms, Image Quality and Effective Dose,” Medical Physics, 30(10), 2727-2735 (2003).
3. D. W. Anderson, “Introduction of Grids to Mobile ICU Radiography in a Teaching Hospital,” British Journal of Radiology,
79, 315-318 (2006).
4. J. C. Wandtke, “Bedside Chest Radiography,” Radiology, 190:1-10 (1994).
5. M. D. Carlin, R. M. Nishikawa, H. MacMahon and K. Doi, “The Effect of X-ray Beam Alignment on the Performance of
Anti-Scatter Grids,” Medical Physics, 23(8), 1347-1350 (1996).
6. H. MacMahon, “Digital Chest Radiography: Practical Issues,” Journal of Thoracic Imaging, 18, 138-147 (2003).
7. D. H. Foos, D. F. Yankelevitz, X. Wang, W. J. Sehnert, , J. Yorkston, C. I. Henschke, “The Bucky Stops Here: Redefining the
Bucky Factor for Digital Portable Chest Radiography,” United Kingdom Radiological Congress 2012 (http://profile-
eposters.co.uk/eposter/action/view/layout/2/id/284).
8. “Diagnostic X-ray Imaging Equipment – Characteristics of General Purpose and Mammographic Anti-Scatter grids,”
IEC-60627, 2001.

More Related Content

What's hot

High voltage transformer and high frequency generator
High voltage transformer and high frequency generatorHigh voltage transformer and high frequency generator
High voltage transformer and high frequency generatorYashawant Yadav
 
Mammography physics and technique
Mammography  physics and techniqueMammography  physics and technique
Mammography physics and techniqueArchana Koshy
 
Computed radiography and digital radiography- CR/DR
Computed radiography and digital radiography- CR/DRComputed radiography and digital radiography- CR/DR
Computed radiography and digital radiography- CR/DRAshim Budhathoki
 
X ray tubes types and advancements
X  ray tubes types and advancementsX  ray tubes types and advancements
X ray tubes types and advancementsAnwesaSardar1
 
Presentation2.pptx technique chest 2
Presentation2.pptx technique chest 2Presentation2.pptx technique chest 2
Presentation2.pptx technique chest 2Yashawant Yadav
 
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING Ganesan Yogananthem
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Sarin Vidya
 
OT theatre radiography.ppt
OT theatre radiography.pptOT theatre radiography.ppt
OT theatre radiography.pptdrparamesh93
 
Emegency drugs in radiology department
Emegency drugs in radiology department Emegency drugs in radiology department
Emegency drugs in radiology department AbubakarMustaphaAman
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x raySushilPattar
 
Dose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyDose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyTarun Goyal
 
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...Prem Murti
 
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxZubairUlIslam5
 

What's hot (20)

Automatic exposure control
Automatic exposure controlAutomatic exposure control
Automatic exposure control
 
High voltage transformer and high frequency generator
High voltage transformer and high frequency generatorHigh voltage transformer and high frequency generator
High voltage transformer and high frequency generator
 
Mammography physics and technique
Mammography  physics and techniqueMammography  physics and technique
Mammography physics and technique
 
CR & DR
CR & DRCR & DR
CR & DR
 
Computed radiography and digital radiography- CR/DR
Computed radiography and digital radiography- CR/DRComputed radiography and digital radiography- CR/DR
Computed radiography and digital radiography- CR/DR
 
Emergency radiology
Emergency  radiologyEmergency  radiology
Emergency radiology
 
Ct protocols of thorax
Ct protocols of thoraxCt protocols of thorax
Ct protocols of thorax
 
Screen Film Radiography
Screen Film RadiographyScreen Film Radiography
Screen Film Radiography
 
Fluroscopy
FluroscopyFluroscopy
Fluroscopy
 
X ray tubes types and advancements
X  ray tubes types and advancementsX  ray tubes types and advancements
X ray tubes types and advancements
 
Presentation2.pptx technique chest 2
Presentation2.pptx technique chest 2Presentation2.pptx technique chest 2
Presentation2.pptx technique chest 2
 
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968
 
OT theatre radiography.ppt
OT theatre radiography.pptOT theatre radiography.ppt
OT theatre radiography.ppt
 
Emegency drugs in radiology department
Emegency drugs in radiology department Emegency drugs in radiology department
Emegency drugs in radiology department
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
Dose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyDose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and Fluoroscopy
 
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
 
Mammography
MammographyMammography
Mammography
 
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
 

Similar to Tube and Grid Alignment System

whitepaper-advanced-ultrasound-imaging-201507
whitepaper-advanced-ultrasound-imaging-201507whitepaper-advanced-ultrasound-imaging-201507
whitepaper-advanced-ultrasound-imaging-201507Ajay Anand
 
Carestream Whitepaper Advanced Ultrasound Imaging
Carestream Whitepaper Advanced Ultrasound ImagingCarestream Whitepaper Advanced Ultrasound Imaging
Carestream Whitepaper Advanced Ultrasound ImagingCarestream
 
Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationCarestream
 
Beam steering in smart antennas by using low complex
Beam steering in smart antennas by using low complexBeam steering in smart antennas by using low complex
Beam steering in smart antennas by using low complexeSAT Publishing House
 
Beam steering in smart antennas by using low complex adaptive algorithms
Beam steering in smart antennas by using low complex adaptive algorithmsBeam steering in smart antennas by using low complex adaptive algorithms
Beam steering in smart antennas by using low complex adaptive algorithmseSAT Journals
 
Relative Dosimetry_02.04.2019_ Final.pptx
Relative Dosimetry_02.04.2019_ Final.pptxRelative Dosimetry_02.04.2019_ Final.pptx
Relative Dosimetry_02.04.2019_ Final.pptxAshikurRahman204838
 
CT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QACT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QASambasivaselli R
 
Reduction of Azimuth Uncertainties in SAR Images Using Selective Restoration
Reduction of Azimuth Uncertainties in SAR Images Using Selective RestorationReduction of Azimuth Uncertainties in SAR Images Using Selective Restoration
Reduction of Azimuth Uncertainties in SAR Images Using Selective RestorationIJTET Journal
 
Adaptive 3D ray tracing approach for indoor radio signal prediction at 3.5 GHz
Adaptive 3D ray tracing approach for indoor radio signal prediction at 3.5 GHzAdaptive 3D ray tracing approach for indoor radio signal prediction at 3.5 GHz
Adaptive 3D ray tracing approach for indoor radio signal prediction at 3.5 GHzIJECEIAES
 
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdfNASERALHAQ
 
A REVIEW ON SYNTHETIC APERTURE RADAR
A REVIEW ON SYNTHETIC APERTURE RADARA REVIEW ON SYNTHETIC APERTURE RADAR
A REVIEW ON SYNTHETIC APERTURE RADARAM Publications
 
IRJET- Simulation Measurement for Detection of the Breast Tumors by using Ult...
IRJET- Simulation Measurement for Detection of the Breast Tumors by using Ult...IRJET- Simulation Measurement for Detection of the Breast Tumors by using Ult...
IRJET- Simulation Measurement for Detection of the Breast Tumors by using Ult...IRJET Journal
 
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy Indian dental academy
 
DRX Plus Detectors: Going from Good to Great
DRX Plus Detectors: Going from Good to GreatDRX Plus Detectors: Going from Good to Great
DRX Plus Detectors: Going from Good to GreatCarestream
 
V5_I1_2016_Paper19.doc
V5_I1_2016_Paper19.docV5_I1_2016_Paper19.doc
V5_I1_2016_Paper19.docIIRindia
 
An overview of adaptive antenna technologies for wireless communication
An overview of adaptive antenna technologies for  wireless communication An overview of adaptive antenna technologies for  wireless communication
An overview of adaptive antenna technologies for wireless communication marwaeng
 

Similar to Tube and Grid Alignment System (20)

gr 8.pdf
gr 8.pdfgr 8.pdf
gr 8.pdf
 
whitepaper-advanced-ultrasound-imaging-201507
whitepaper-advanced-ultrasound-imaging-201507whitepaper-advanced-ultrasound-imaging-201507
whitepaper-advanced-ultrasound-imaging-201507
 
Carestream Whitepaper Advanced Ultrasound Imaging
Carestream Whitepaper Advanced Ultrasound ImagingCarestream Whitepaper Advanced Ultrasound Imaging
Carestream Whitepaper Advanced Ultrasound Imaging
 
Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of Operation
 
Beam steering in smart antennas by using low complex
Beam steering in smart antennas by using low complexBeam steering in smart antennas by using low complex
Beam steering in smart antennas by using low complex
 
Beam steering in smart antennas by using low complex adaptive algorithms
Beam steering in smart antennas by using low complex adaptive algorithmsBeam steering in smart antennas by using low complex adaptive algorithms
Beam steering in smart antennas by using low complex adaptive algorithms
 
W6P3622650776P65
W6P3622650776P65W6P3622650776P65
W6P3622650776P65
 
Relative Dosimetry_02.04.2019_ Final.pptx
Relative Dosimetry_02.04.2019_ Final.pptxRelative Dosimetry_02.04.2019_ Final.pptx
Relative Dosimetry_02.04.2019_ Final.pptx
 
CT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QACT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QA
 
Reduction of Azimuth Uncertainties in SAR Images Using Selective Restoration
Reduction of Azimuth Uncertainties in SAR Images Using Selective RestorationReduction of Azimuth Uncertainties in SAR Images Using Selective Restoration
Reduction of Azimuth Uncertainties in SAR Images Using Selective Restoration
 
Adaptive 3D ray tracing approach for indoor radio signal prediction at 3.5 GHz
Adaptive 3D ray tracing approach for indoor radio signal prediction at 3.5 GHzAdaptive 3D ray tracing approach for indoor radio signal prediction at 3.5 GHz
Adaptive 3D ray tracing approach for indoor radio signal prediction at 3.5 GHz
 
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
 
A REVIEW ON SYNTHETIC APERTURE RADAR
A REVIEW ON SYNTHETIC APERTURE RADARA REVIEW ON SYNTHETIC APERTURE RADAR
A REVIEW ON SYNTHETIC APERTURE RADAR
 
Vinci-Medphs
Vinci-MedphsVinci-Medphs
Vinci-Medphs
 
IRJET- Simulation Measurement for Detection of the Breast Tumors by using Ult...
IRJET- Simulation Measurement for Detection of the Breast Tumors by using Ult...IRJET- Simulation Measurement for Detection of the Breast Tumors by using Ult...
IRJET- Simulation Measurement for Detection of the Breast Tumors by using Ult...
 
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
 
CT dose reduction
CT dose reductionCT dose reduction
CT dose reduction
 
DRX Plus Detectors: Going from Good to Great
DRX Plus Detectors: Going from Good to GreatDRX Plus Detectors: Going from Good to Great
DRX Plus Detectors: Going from Good to Great
 
V5_I1_2016_Paper19.doc
V5_I1_2016_Paper19.docV5_I1_2016_Paper19.doc
V5_I1_2016_Paper19.doc
 
An overview of adaptive antenna technologies for wireless communication
An overview of adaptive antenna technologies for  wireless communication An overview of adaptive antenna technologies for  wireless communication
An overview of adaptive antenna technologies for wireless communication
 

More from Carestream

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Carestream
 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareCarestream
 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...Carestream
 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsCarestream
 
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographySmart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographyCarestream
 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-rayCarestream
 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentCarestream
 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityCarestream
 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Carestream
 
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Carestream
 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesCarestream
 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seramCarestream
 
Cloud Security
Cloud Security Cloud Security
Cloud Security Carestream
 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Carestream
 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareCarestream
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundCarestream
 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Carestream
 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Carestream
 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Carestream
 
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA163 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA16Carestream
 

More from Carestream (20)

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization Software
 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
 
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographySmart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-ray
 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray Equipment
 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image Quality
 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions
 
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seram
 
Cloud Security
Cloud Security Cloud Security
Cloud Security
 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
 
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA163 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
 

Recently uploaded

The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...chandars293
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...narwatsonia7
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Dipal Arora
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableDipal Arora
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...parulsinha
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Ishani Gupta
 
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Sheetaleventcompany
 
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...parulsinha
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...narwatsonia7
 
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In AhmedabadO963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In AhmedabadGenuine Call Girls
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 

Recently uploaded (20)

The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
 
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In AhmedabadO963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 

Tube and Grid Alignment System

  • 1. White Paper | CARESTREAM DRX-Revolution Mobile X-ray System The CARESTREAM Tube and Grid Alignment System provides better image quality and consistent techniques for portable diagnostic radiography Anti-Scatter Grids Improve Image Quality Scattered radiation is a major contributor to reduced image quality in diagnostic radiography. Anti-scatter grids, when properly aligned, provide an effective means for reducing scatter [1-6]. The preferential transmission of primary versus scattered X-rays improves the contrast-to-noise ratio (CNR). Figure 1 shows a pair of portable chest images captured of an ICU patient with and without grids and under the same exposure conditions. Comparing the two images, the contrast improvement with grids is significant. Figure 1 -- Comparison of portable chest images captured of an ICU patient without (left) and with (right) anti-scatter grid Grid Benefits Can be Realized with Less Exposure Increments in Digital Radiography The Bucky factor is the reciprocal of the total X-ray penetration through an anti-scatter grid. The analog screen- film system has a fixed sensitometry response. When a grid is used, a Bucky factor technique increment was necessary to ensure sufficient exposure to produce a film image with proper density and contrast for diagnosis. The benefits of anti-scatter grids can be realized in digital radiography (DR) with less exposure technique increments than would be indicated by traditional Bucky factors [7]. This is because DR does not have a fixed sensitomety response; the global contrast and brightness of a DR image can be arbitrarily modified by digital image processing. DR systems are fundamentally noise limited – within a broad range of exposure levels to the patient, the image quality is determined by the anatomy CNR in the captured image. Anti-scatter grids can improve the image CNR at a minor expense of reduced primary X-ray transmission. This suggests some exposure technique increase may be required to compensate for primary X-ray attenuation, but usually at a lower level than the Bucky factor.
  • 2. White Paper | CARESTREAM DRX-Revolution Mobile X-ray System 2 Challenges in Using Grids for Portable Radiography Grid usage in portable radiography is often sporadic and inconsistent. This causes greater variability in image quality, and a greater number of radiographs of poorer quality than those captured in the radiology department being delivered for interpretation. From the perspective of the radiographic technologist, using grids for portable exams involves a variety of time-consuming workflow implications. These include attaching and detaching the add-on grids to the X-ray cassettes; the stringent requirements to properly position and align the X-ray source relative to the cassette behind the patient to avoid grid cutoff; the increased probability that repeated exposures will be required due to grid-cutoff artifact; and more. In addition, there is the misperception that grids are not required in digital radiography because increasing the exposure can overcome the scatter-noise level, and that image processing adjustments, such as window and level manipulations, can sufficiently compensate for the quality losses that are introduced by scattered radiation. With all of these considerations in mind, there would seem to be little motivation for the technologists to use grids in portable digital radiography. DRX-Revolution Tube and Grid Alignment System The CARESTREAM Tube and Grid Alignment System (TGA) for the CARESTREAM DRX-Revolution Mobile X-ray System provides guidance to the radiographic technologist for proper alignment of the X-ray source relative to the grid/detector. This optional feature integrates seamlessly with the DRX-Revolution System – no additional operational steps are required in the normal workflow (Figure 2). Two transmitters mounted under the collimator Eight receivers inside the grid holder frame Figure 2 – The DRX-Revolution and schematics of the optional Tube and Alignment System. The grid-alignment transmitters are built inside the tube-head assembly, and the receivers are in the grid holder. Alignment Status Tube Angle Detector Angle Actual Distance Alignment Status Bed Type Selection Figure 3 – The tube-head display on the DRX-Revolution System reports real-time X-ray source and grid/detector position information for proper alignment. Carestream’s TGA helps the radiographic technologist to achieve better image quality and consistent imaging technique for portable exams. It automatically detects the X-ray source position relative to the grid holder/detector, and then reports the amount of X-ray source displacements from the proper alignment (Figure 3). The displacements are presented in real time on the tube-head display in separate up/down, left/right and in/out directions. The alignment status turns green when the TGA decides that the X-ray source is within the optimal grid-operation range, specific to the target source-to-grid distance (SGD).
  • 3. White Paper | CARESTREAM DRX-Revolution Mobile X-ray System 3 Alignment Technology Overview The TGA uses multiple electromagnetic fields for position triangulation. Two transmitters are mounted under the X-ray collimator to generate the electromagnetic fields, and a total of eight receivers are installed inside the grid holder around the grid/detector to detect the fields from the transmitters (Figure 2). The transmitters use magnetic coils to synthesize magnetic fields that are functionally the same as those of spinning, permanent bar magnets (Figure 4). The receivers sense the spinning magnetic-field intensity and phase, and then each generates a unique sine-wave signal (Figure 5). N S Receiver Transmitter (Spinning Magnet) Figure 4 – The Tube and Grid Alignment System works by detecting the intensity and phase of synthesized spinning magnetic fields. These diagrams show three examples of different relative positions between a transmitter and a receiver. Figure 5 – A snapshot of the detected magnetic-field signals from eight receivers Sophisticated mathematical models are built inside the software to decode the receiver sine-wave information. The magnitude and phase of the sine wave from each receiver are extracted from the electronic signals, and then compared with their expected theoretical values, which are calculated based on an initial assumption of the geometry between the X-ray source and the grid holder/detector. A total error is summed from the comparison differences for all the receivers. A non-zero total error value indicates that the initial geometry assumption is incorrect, to which fine adjustments are made and the calculation process is reiterated. The X-ray source position is best estimated when
  • 4. White Paper | CARESTREAM DRX-Revolution Mobile X-ray System www.carestream.com Carestream Health, Inc., 2012. CARESTREAM is a trademark of Carestream Health. CAT 200 0016 8.5x11 9/12 the total error is minimized. Finally, the estimated position value is compared to the optimal grid-operation range, and the X-ray source displacements from the proper alignment are presented on the tube-head display. The grid-alignment transmitters use amplitude-modulated radio-frequency (RF) signals for extended operation range and improved signal-to-noise ratio. Eddy current can be induced by the transmitter RF signals in the metallic material that is in close proximity to the receivers. The Eddy current introduces secondary electromagnetic waves that can negatively influence the accuracy of the grid-alignment results. To overcome this issue, different bed types can be selected on the software’s GUI to better compensate for the interference. Specifically, the bed-type selection (intensive care units vs. emergency departments) is primarily based on the thickness of the mattress on top of the bed metal frames. Definition of Optimal Grid-Operation Range The grid built inside the grid holder has a center focal distance of 112 cm, a grid ratio of 8:1 and a frequency resolution of 80 lines per centimeter. The optimal grid-operation range built inside the TGA software is determined by the minimum performance requirement for image-quality improvement. The grid performance is characterized in terms of the signal-to-noise ratio improvement factor (SIF)[8]. The higher the SIF value, the better the performance. The benefit of using the grid diminishes when the SIF value equals 1.0. The optimal grid-operation range at a particular SGD is defined as the maximum allowable lateral (across the grid-line direction) misalignment in the X-ray source position, at which the minimum SIF value across the detector’s effective imaging area would be greater than 1.0. Summary Anti-scatter grids improve radiography image quality, and the benefit of grid usage can be realized in digital radiography with less exposure technique increments. The CARESTREAM Tube and Grid Alignment System for the DRX-Revolution is fully integrated with the radiographic technologist’s normal workflow, and provides easy and intuitive guidance for X-ray-source alignment to achieve consistent and optimal image quality. References 1. H.-P. Chan, K. L. Lam, and Y. Wu, “Studies of Performance of Anti-Scatter Grids in Digital Radiography: Effect on Signal- to-Noise Ratio,” Medical Physics, 17(4), 655-664 (1990). 2. L. N. Rill, L. Brateman and M. Arreola, “Evaluating Radiographic Parameters for Mobile Chest Computed Radiography: Phantoms, Image Quality and Effective Dose,” Medical Physics, 30(10), 2727-2735 (2003). 3. D. W. Anderson, “Introduction of Grids to Mobile ICU Radiography in a Teaching Hospital,” British Journal of Radiology, 79, 315-318 (2006). 4. J. C. Wandtke, “Bedside Chest Radiography,” Radiology, 190:1-10 (1994). 5. M. D. Carlin, R. M. Nishikawa, H. MacMahon and K. Doi, “The Effect of X-ray Beam Alignment on the Performance of Anti-Scatter Grids,” Medical Physics, 23(8), 1347-1350 (1996). 6. H. MacMahon, “Digital Chest Radiography: Practical Issues,” Journal of Thoracic Imaging, 18, 138-147 (2003). 7. D. H. Foos, D. F. Yankelevitz, X. Wang, W. J. Sehnert, , J. Yorkston, C. I. Henschke, “The Bucky Stops Here: Redefining the Bucky Factor for Digital Portable Chest Radiography,” United Kingdom Radiological Congress 2012 (http://profile- eposters.co.uk/eposter/action/view/layout/2/id/284). 8. “Diagnostic X-ray Imaging Equipment – Characteristics of General Purpose and Mammographic Anti-Scatter grids,” IEC-60627, 2001.