OCT MACHINESOCT MACHINES
Principle of OCT machines
TD- OCT in brief
SD-OCT
Spectralis OCT heidelberg
SS-OCT in brief
What is Optical coherence tomography (OCT) ?
“OCT represents a non invasive method for cross sectional imaging of the internal
retinal structures, through the detection of optical reflections and echo time delays
of light, using low coherence interferometry, to subsequently produce in vivo two
dimensional images of internal tissue microstructure.”
Optical Coherence Tomography: Newer Techniques, Newer Machines
• InterferometryInterferometry is a technique in which electromagnetic waves are superimposed
causing the phenomenon of interference in order to extract information
• In an interferometer, light from a single source is split into two beams that travel
different optical paths, then combined again to produce interference.
• The resulting interference fringes give information about the difference in optical
path length.
‘Michelson interferometer’
 Initially for in vivo measurements of axial length of eye by Fercher et al. (Medical University
of Vienna, 1988)
 In early 90´s David Huang, a PhD student at MIT, on his research on low-coherence
interferometry developed OCT.
 Cross-sectional tomographic imaging of retinal tissue.
 Uses backscattered light.
 Similar to ultrasound B-Scan imaging but uses infrared-light instead of ultrasound
waves.
Optical analog of ultrasound
Fujimoto, James, and Eric Swanson. "The development, commercialization, and impact of optical coherence tomography." Investigative ophthalmology
& visual science57.9 (2016): OCT1-OCT13.
Fujimoto, James, and Eric Swanson. "The development, commercialization, and impact of optical coherence tomography." Investigative ophthalmology & visual science57.9 (2016): OCT1-OCT13.
First OCT protoype at MIT
Advantages:Advantages:
 Obtains high resolution images
 “Histological images” in vivo
 In real time
 Non-invasive
 Uses harmless radiation
 Ophthalmology, cardiology, odontology and dermatology
OCT TIMELINE
Fujimoto J, Swanson E. The development, commercialization, and impact of optical coherence tomography. Invest Ophthalmol Vis Sci. 2016;57:OCT1–OCT13.
DOI:10.1167/ iovs.16-19963
Time domain ( TD-OCT) :Time domain ( TD-OCT) :
• Stratus OCT, Carl zeiss meditech
• Mobile reference arm mirror: allows backscattered tissue intensity levels
to be detected from different depths in the tissue sample.
• It is referred to as ‘time-domain’ because time-encoded signals are
obtained directly.
• 400 A scans/ sec
• Axial resolution 10 um
• lateral resolution 20um
Longer acquisition time
Motion artefacts
Fujimoto, James, and Eric Swanson. "The development, commercialization, and impact of optical coherence tomography." Investigative ophthalmology & visual science57.9
(2016): OCT1-OCT13.
Spectral domain(SD-OCT) / Fourier domain OCT(FD-OCT) :Spectral domain(SD-OCT) / Fourier domain OCT(FD-OCT) :
• Wavelength 800-850um
• Stationary reference arm
• High speed spectrometer
• A charged couple device(CCD) line scan camera
• Acquisition speed of 25k to 52k A scans/sec
• Axial resolution 5-7microns.
• lateral resolution 14-20microns.
Principle of SD-OCT
ADVANTAGES:
• Reduction of motion artefacts because of averaging. i.e multiple scans are
acquired from the same position.
• Better visualisation of individual retinal layers.
• Improved signal to noise ratio
Description of Commercially Available SD-OCT Systems
Device (Manufacturer) Description
• 3D-OCT 2000 (Topcon,Tokyo, Japan) SD-OCT and high-resolution fundus camera;
axial resolution 5um; A-scan acquisition
rate 27 kHz.
• Bioptigen SD-OCT
(Bioptigen,NC) Designed for both clinical and research use
and includes a hand-held probe and
microscope setup; axial resolution 4um;
A-scan acquisition rate 20 kHz
• Cirrus HD-OCT
(Carl Zeiss Meditec, Dublin, CA) Software includes guided progression analysis for
glaucoma progression detection; axial resolution,
5um; A-scan acquisition rate 27 kHz.
Gabriele, Michelle L., et al. "Optical coherence tomography: history, current status, and laboratory work." Investigative ophthalmology & visual science 52.5 (2011): 2425-
Device (Manufacturer) Description
• RTVue-100
(Optovue, Fremont, CA) Offers multiple scanning protocols
for glaucoma detection, including ganglion
cell complex analysis; axial resolution 5um;
A-scan acquisition rate 26 kHz.
• SOCT Copernicus
(Optopol, Zawiercie, Poland) Software includes progression analysis software that
incorporates disk damage likelihood
scale,asymmetry
between the discs, and RNFL thickness; axial
resolution, 6um; A-scan 27 kHz.
• Spectral OCT SLO
(Opko, Miami, FL) Combines SD-OCT, scanning laser
ophthalmoscopy, and microperimetry. Axial
resolution 6um, A-scan 27 kHz.
Gabriele, Michelle L., et al. "Optical coherence tomography: history, current status, and laboratory work." Investigative ophthalmology & visual science 52.5 (2011): 2425-
Device (Manufacturer) Devices
• Spectralis OCT (Heidelberg Engineering,
Heidelberg, Germany) High-speed SD-OCT device with eye-eye-
trackingtracking, fluorescein angiographyfluorescein angiography, ICGICG
angiography, and autofluorescenceautofluorescence. Axial
resolution,4um4um; A-scan acquisition rate
40 kHz40 kHz.
Gabriele, Michelle L., et al. "Optical coherence tomography: history, current status, and laboratory work." Investigative ophthalmology & visual science 52.5 (2011): 2425-
2436.
Spectralis OCTSpectralis OCT ( Heidelberg Engineering , Germany )( Heidelberg Engineering , Germany )
• Dual beam SD-OCT and cSLO system.
• Acquires volume scans of the retina and an infrared reference image
simultaneously.
• Eye tracker stablilizes the OCT scan
• 40,000 A scans/sec
• Short examination timing
• Simultaneous real time FFA/OCT and ICG/OCT.
• Spectralis HRA+OCT is a combination of the Spectralis HRA and Spectralis
OCT.
• Allows angiography, reflectance imaging, autofluorescence imaging, and
OCT cross sectional imaging, alone or in various combinations.
Confocal Laser ScanningConfocal Laser Scanning
• For confocal images, a laser beam is focused on the retina.
• SLO devices employ a confocal (pinhole) aperture, generating a single
point of laser light, at a specific wavelength.
• Can be scanned across the retina in a raster pattern.
• As only a small area of the fundus is illuminated at any one time, the
effects of light scatter are reduced and images with higher contrast are
generated.
• Spectralis HRA emit laser light with three different wavelengths:
1. For excitation of fluorescein, a blue solid state laser (wavelength 488 nm) is used.
A barrier filter at 500 nm separates excitation and fluorescence light.
The same wavelength (but without the barrier filter) is used to create blue
reflectance and autofluorescence images.
Representations of confocal laser scanning microscopy to vi-sualize fluorescence staining and reflection properties of isolated reti-nal tissue. aA laser beam is conducted through a dichroic
mirror andan objective to excite the fluorescent dye in the specimen. bTheemitted fluorescent light is conducted back through the objective anda dichroic mirror (worked as a light filter) to the
light detector (e.g.photomultiplier). The pinhole enables the exclusive detection of lightcoming from a certain depth in the focused tissue area (confocality). cIn the reflection mode of the confocal
laser scanning microscope, thelaser light for excitation is reflected from cellular structures in theretinal tissue, passes optical filters and is detected by thephotomultiplier. The pinhole is used to
allow confocality
VIDEOVIDEO
2. For excitation of indocyanine green, a diode laser at 790 nm
wavelength is used, together with a barrier filter at 830 nm to separate
excitation and fluorescence light.
3.A second diode laser at 820 nm wavelength serves to produce infrared
reflectance (IR) images.
• SPECTRALIS®
MultiColor Module
incorporates a “multicolour” mode,
where separate fundus images are
obtained with blue, green, and near-
infrared filters, and which can then be
examined individually or in
combination.
• Shows distinct structures at different
depths within the retina.
• MultiColor images can highlight
structures and pathologies not visible
on ophthalmoscopy and fundus
photography.
CSCR
MultiColor images may even be acquired in
patients with cataract or nystagmus.
•Longer wavelengths of light penetrate more deeply – the use of near infrared
filters thus greatly improves visualization of subretinal and choroidal structures
(e.g. drusen subtypes in age-related macular degeneration (AMD)).
•Shorter wavelengths of light (blue light), allow improved imaging of superficial
retinal structures (e.g. retinal nerve fibre layer defects in glaucoma and epiretinal membranes).
•Intermediate wavelengths of light (green light (“red-free”)), provide optimal
fundus images for general purposes.
•Green wavelength is reflected by RPE, but absorbed by retinal haemorrhage
and blood vessels.
• In the simultaneous imaging modes, each image line is scanned twice
subsequently with alternating lasers.
• So, during simultaneous FFA and ICGA, only the 488 nm laser is switched
on for the first scan of a line, while during the second scan only the 790
nm laser is switched on.
Enhanced depth imaging ( EDI)Enhanced depth imaging ( EDI)
• Spaide was the first to describe.
• In standard OCT, the zero-delay line is typically at the vitreous
inner retina junction, providing excellent resolution.
• The “zero-delay” line is a reference point set by the software where
the image capture is optimal.
• EDI is achieved by moving the zero delay line deeper into the eye,
focusing the OCT scanner on the choroid instead of the retina.
• Moving away from the zero-delay line decreases the resolution from
structures within the choroid resulting in diminished choroidal detail.
Filter wheel position ‘R’ (Reflectance)
How To Capture various modesHow To Capture various modes
Filter wheel position ‘A’ (Angiography)
Filter wheel position ‘P’ (Polarisation)
xyz-Micromanipulator
For the adjustment of the camera position:
• Inner knob: Z adjustment (forward/backward)
• Middle knob: Y adjustment (height)
• outer knob: x adjustment (left/right)
X
Y
Z
Scanning techniques:Scanning techniques:
• The commonly used scan protocols for macular scanning are
three dimensional (3D)scan, radial scan, and raster scan.‑
• A 3D scan consists of a number of horizontal line scans
composing a 6 mm × 6 mm/ 7 mm × 7 mm / 12 mm× 9 mm
rectangular box.
• A minimum pupil diameter of 3 mm is required to obtain a
good OCT image.
Types of Radial ScansTypes of Radial Scans
30°30°
4 Lines, 45°4 Lines, 45° 6 Lines, 30°6 Lines, 30°
12 Lines, 15°12 Lines, 15° 24 Lines, 7.5°24 Lines, 7.5° 48 Lines, 3.8°48 Lines, 3.8°
2 Lines, 90°2 Lines, 90°
Types of Raster ScansTypes of Raster Scans
Single line scanSingle line scan
Detail Scan, 30umDetail Scan, 30um
7 line scan, 240um7 line scan, 240um Fast scan , 240umFast scan , 240um
Dense Scan, 120umDense Scan, 120um
• Thickness map displays the average retinal thickness values in 1 mm, 3
mm, and 6 mm diameter circles.
• Divided into sectors with color map.
• Warm colors indicate thicker retinal areas and cool color indicate thinner
retinal areas
• The SPECTRALIS®
Widefield Imaging Module provides a 55-degree field of view for all fundus
imaging modalities including MultiColor, BluePeak, IR, angiography, and OCT.
• The Widefield Imaging Module facilitates comprehensive diagnostics.
• The high-resolution, high-contrast scanning laser images captured with the Widefield Imaging
Module offer diagnostic capabilities beyond those of conventional fundus photography.
The optional Ultra-Widefield Angiography Module (102 °) delivers scanning laser
angiography images from the macula to the periphery for more comprehensive
retinal examinations.
Changing lenses for ImagingChanging lenses for Imaging
• To remove the objective lens , it is rotated by 50° to the left (counter
clockwise).
30°Objective lens
• The optional 55°wide angle lens
• For IR, RF and angiography wide-field imaging
• Has to be inserted with the red dot orientated upwards (12 o’clock)
• Then rotated by about 50°clockwise until it snaps in.
55°wide angle lens
Step 1 Step 2 Step 3
102°Ultra wide field lens
FOCUS KNOB:
• By turning this knob, refractive errors of the patient’s eye can be
compensated.
• For examination of most patients, an initial focus setting of ‘0’ (zero
diopters) is suitable.
• Also called Optical Frequency
Domain OCT
• Obtains time-encoded information
by sweeping a narrow bandwidth
laser through a broad optical
spectrum
• Backscattered intensity is detected with a
photodetector.
• This process is in contrast to SD-OCT, which uses
a broad bandwidth light source and detects the
interference spectra with a CCD camera and
spectrometer.
First high speed OFDI system: S.H. Yun et al, Opt. Express 11, 2953-2963 (2003)
Swept source oct
• Speeds of up to 249,000A-scans/s have been attained
• Eye motion artifacts are greatly reduced compared with TD-OCT.
• A drawback to camera-based SD-OCT detection is a drop-off in signal with
depth of scanning because of the finite pixel size of the CCD camera.
Taken from “Swept-Source Optical Coherence Tomography” presentation by Dr Muna Bhende
• Wide 12 mm scans – evaluation of disc- macula relationships in one frame
Taken from “Swept-Source Optical Coherence Tomography” presentation by Dr Muna Bhende
12 mm 12 mm 12 mm
Taken from “Swept-Source Optical Coherence Tomography” presentation by Dr Muna Bhende
36 mm !!
Comparison of TD-, SD-, and SS-OCT DevicesComparison of TD-, SD-, and SS-OCT Devices
•Employs motion contrast imaging to high-resolution volumetric blood flow
information  generating angiographic images.
•Compares the ‘decorrelation signal’‘decorrelation signal’ between sequential OCT B-Scans taken at
precisely the same cross section in order to construct a map of blood flow.
Other OCT systems:Other OCT systems:
Optical coherence tomography angiography (OCTA)Optical coherence tomography angiography (OCTA)
En face OCT:En face OCT:
•Derived from SD OCT.
Produces frontal sections of retinal layers, also called ‘C-scan OCT’.
Adaptive optics OCTAdaptive optics OCT
• Corrects for higher-order ocular aberrations
Doppler OCTDoppler OCT
• measuring the Doppler shift and relative angle between the OCT beam
and a blood vessel, assess the blood flow velocity.
Polarization sensitive OCTPolarization sensitive OCT
• used for functional tissue assessment by means of evaluation of light
polarization which allows individual retinal layer identification
Image taken from “Swept-Source Optical Coherence Tomography” presentation by Dr Muna Bhende
OCT is a quick, non invasive and harmless examination technique
Can be used to rapidly access the pathologies of retina and choroid
Better and improved machines with better resolution.
Future of OCT with newer advances is even more promising.
OCT is here to stay !!OCT is here to stay !!
To summarizeTo summarize
THANK YOU
• Automated segmentation into the “superficial” and “deep” inner retinal vascular
plexuses, outer retina, and choriocapillaris
• The OCT angiogram segmentation of the superficial inner retina contains a
projection of the vasculature in the retinal nerve fiber layer (RNFL) and ganglion
cell layer (GCL)
Superficial capillary plexusSuperficial capillary plexus Deep capillary plexusDeep capillary plexus
• The deep inner retina OCT angiogram shows a composite of the vascular plexuses
at the border of the IPL/INL and the border of the INL/ OPL.
Deep capillary plexusDeep capillary plexus choriocapillarieschoriocapillaries
Sattler’s layerSattler’s layer Haller’s layerHaller’s layer

Contenu connexe

Tendances

pentacam
pentacampentacam
pentacamnrvdad
 
Glare testing and dark adaptation
Glare testing and dark adaptationGlare testing and dark adaptation
Glare testing and dark adaptationHira Dahal
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopyHira Dahal
 
Slitlamp bimicroscopy
Slitlamp bimicroscopy Slitlamp bimicroscopy
Slitlamp bimicroscopy Obehi Osoata
 
anomalous retinal correspondence
anomalous retinal correspondenceanomalous retinal correspondence
anomalous retinal correspondenceRajeshwori
 
The Slit lamp Biomicroscope
The Slit lamp BiomicroscopeThe Slit lamp Biomicroscope
The Slit lamp BiomicroscopeAkshay Nayak
 
Retinoscopy @adi
Retinoscopy @adiRetinoscopy @adi
Retinoscopy @adiFarhana Adi
 
Indirect ophthalmoscopy and fundus drawing
Indirect ophthalmoscopy and fundus drawingIndirect ophthalmoscopy and fundus drawing
Indirect ophthalmoscopy and fundus drawingSonali Singh
 
Optical Coherence Tomography
Optical Coherence TomographyOptical Coherence Tomography
Optical Coherence TomographyManoj Aryal
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refractionsanju_95
 
Optical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --maculaOptical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --maculaAkshay Nayak
 

Tendances (20)

pentacam
pentacampentacam
pentacam
 
Oct
Oct Oct
Oct
 
B scan
B scanB scan
B scan
 
Glare testing and dark adaptation
Glare testing and dark adaptationGlare testing and dark adaptation
Glare testing and dark adaptation
 
Pentacam and topography
Pentacam and topographyPentacam and topography
Pentacam and topography
 
Keratometry
KeratometryKeratometry
Keratometry
 
Contrast sensitivity
Contrast sensitivityContrast sensitivity
Contrast sensitivity
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Slitlamp bimicroscopy
Slitlamp bimicroscopy Slitlamp bimicroscopy
Slitlamp bimicroscopy
 
Ffa
FfaFfa
Ffa
 
anomalous retinal correspondence
anomalous retinal correspondenceanomalous retinal correspondence
anomalous retinal correspondence
 
The Slit lamp Biomicroscope
The Slit lamp BiomicroscopeThe Slit lamp Biomicroscope
The Slit lamp Biomicroscope
 
Retinoscopy @adi
Retinoscopy @adiRetinoscopy @adi
Retinoscopy @adi
 
Indirect ophthalmoscopy and fundus drawing
Indirect ophthalmoscopy and fundus drawingIndirect ophthalmoscopy and fundus drawing
Indirect ophthalmoscopy and fundus drawing
 
B SCAN
B SCAN B SCAN
B SCAN
 
OCT in glaucoma ppt;1
OCT in glaucoma ppt;1OCT in glaucoma ppt;1
OCT in glaucoma ppt;1
 
Optical Coherence Tomography
Optical Coherence TomographyOptical Coherence Tomography
Optical Coherence Tomography
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Optical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --maculaOptical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --macula
 
Dynamic retinoscopy srijana
Dynamic retinoscopy srijanaDynamic retinoscopy srijana
Dynamic retinoscopy srijana
 

Similaire à OCT Machines

Similaire à OCT Machines (20)

OCT MACULA INTERPRETATION.
OCT MACULA  INTERPRETATION.OCT MACULA  INTERPRETATION.
OCT MACULA INTERPRETATION.
 
Oct(optical cohorence tomography)
Oct(optical cohorence tomography)Oct(optical cohorence tomography)
Oct(optical cohorence tomography)
 
Oct introduction
Oct  introductionOct  introduction
Oct introduction
 
Principles of optical coherence tomography
Principles of optical coherence tomographyPrinciples of optical coherence tomography
Principles of optical coherence tomography
 
Oct
OctOct
Oct
 
Final oct
Final octFinal oct
Final oct
 
OPTICAL COHERENCE DEMYSTIFIED
OPTICAL COHERENCE DEMYSTIFIED OPTICAL COHERENCE DEMYSTIFIED
OPTICAL COHERENCE DEMYSTIFIED
 
ASOCT NEW SLIDES.pptx
ASOCT NEW SLIDES.pptxASOCT NEW SLIDES.pptx
ASOCT NEW SLIDES.pptx
 
Optical Coherence Tomography(OCT) in posterior segment diseases
Optical Coherence Tomography(OCT) in posterior segment diseasesOptical Coherence Tomography(OCT) in posterior segment diseases
Optical Coherence Tomography(OCT) in posterior segment diseases
 
OCT
OCTOCT
OCT
 
Lasers in ophthalmology
Lasers in ophthalmologyLasers in ophthalmology
Lasers in ophthalmology
 
Role of oct in ophthalmology
Role of oct in ophthalmologyRole of oct in ophthalmology
Role of oct in ophthalmology
 
OCT
OCTOCT
OCT
 
Retinal imaging
Retinal imagingRetinal imaging
Retinal imaging
 
Microscopy
MicroscopyMicroscopy
Microscopy
 
Optical coherence tomography
Optical coherence tomographyOptical coherence tomography
Optical coherence tomography
 
LASER IN OPHTHALMOLOGY
LASER IN OPHTHALMOLOGYLASER IN OPHTHALMOLOGY
LASER IN OPHTHALMOLOGY
 
OCT
OCTOCT
OCT
 
11
1111
11
 
Oct in post seg disorders
Oct in post seg disordersOct in post seg disorders
Oct in post seg disorders
 

Plus de Dr Samarth Mishra (20)

Cover tests
Cover testsCover tests
Cover tests
 
Retina quiz
Retina quizRetina quiz
Retina quiz
 
Cone and Rod Dystrophy
Cone and Rod DystrophyCone and Rod Dystrophy
Cone and Rod Dystrophy
 
History of Indirect Ophthalmoscope
History of Indirect OphthalmoscopeHistory of Indirect Ophthalmoscope
History of Indirect Ophthalmoscope
 
Vitrectomy: Development And Steps
Vitrectomy: Development And StepsVitrectomy: Development And Steps
Vitrectomy: Development And Steps
 
Evolution of retinal detachment surgery
Evolution of retinal detachment surgery Evolution of retinal detachment surgery
Evolution of retinal detachment surgery
 
Secondary open angle glaucoma
Secondary open angle glaucomaSecondary open angle glaucoma
Secondary open angle glaucoma
 
Normal tension glaucoma
Normal tension glaucomaNormal tension glaucoma
Normal tension glaucoma
 
Glaucoma risk factors
Glaucoma risk factorsGlaucoma risk factors
Glaucoma risk factors
 
Choroiditis
ChoroiditisChoroiditis
Choroiditis
 
Target IOP
Target IOPTarget IOP
Target IOP
 
Ocular hypertension
Ocular hypertensionOcular hypertension
Ocular hypertension
 
Importance of diurnal variation
Importance of diurnal variationImportance of diurnal variation
Importance of diurnal variation
 
Aqueous humour
Aqueous humourAqueous humour
Aqueous humour
 
Role of oct in glaucoma
Role of oct in glaucomaRole of oct in glaucoma
Role of oct in glaucoma
 
Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.
 
Retinopathy of prematurity
Retinopathy of prematurityRetinopathy of prematurity
Retinopathy of prematurity
 
Normal fundus
Normal fundusNormal fundus
Normal fundus
 
MANAGEMENT OF RETINOBLASTOMA & CURRENT TRENDS
MANAGEMENT OF RETINOBLASTOMA & CURRENT TRENDSMANAGEMENT OF RETINOBLASTOMA & CURRENT TRENDS
MANAGEMENT OF RETINOBLASTOMA & CURRENT TRENDS
 
Macular hole
Macular holeMacular hole
Macular hole
 

Dernier

Gurgaon Sushant Lok Phase 2 Call Girls Service ( 9873940964 ) unlimited hard...
Gurgaon Sushant Lok Phase 2 Call Girls Service ( 9873940964 )  unlimited hard...Gurgaon Sushant Lok Phase 2 Call Girls Service ( 9873940964 )  unlimited hard...
Gurgaon Sushant Lok Phase 2 Call Girls Service ( 9873940964 ) unlimited hard...ggsonu500
 
Rohini Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Rohini Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Rohini Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Rohini Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...ddev2574
 
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near MeBook Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Call Girls Hsr Layout Whatsapp 7001305949 Independent Escort Service
Call Girls Hsr Layout Whatsapp 7001305949 Independent Escort ServiceCall Girls Hsr Layout Whatsapp 7001305949 Independent Escort Service
Call Girls Hsr Layout Whatsapp 7001305949 Independent Escort Servicenarwatsonia7
 
EMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareEMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareRommie Duckworth
 
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...ggsonu500
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed RuleShelby Lewis
 
Globalny raport: „Prawdziwe piękno 2024" od Dove
Globalny raport: „Prawdziwe piękno 2024" od DoveGlobalny raport: „Prawdziwe piękno 2024" od Dove
Globalny raport: „Prawdziwe piękno 2024" od Doveagatadrynko
 
Call Girls Laxmi Nagar 9999965857 Cheap and Best with original Photos
Call Girls Laxmi Nagar 9999965857 Cheap and Best with original PhotosCall Girls Laxmi Nagar 9999965857 Cheap and Best with original Photos
Call Girls Laxmi Nagar 9999965857 Cheap and Best with original Photosparshadkalavatidevi7
 
Call Girls Delhi 9873940964 Elite Escort Service Available 24/7 Hire
Call Girls Delhi 9873940964 Elite Escort Service Available 24/7 HireCall Girls Delhi 9873940964 Elite Escort Service Available 24/7 Hire
Call Girls Delhi 9873940964 Elite Escort Service Available 24/7 HireCall Girls Delhi
 
Russian Call Girls Sadashivanagar | 7001305949 At Low Cost Cash Payment Booking
Russian Call Girls Sadashivanagar | 7001305949 At Low Cost Cash Payment BookingRussian Call Girls Sadashivanagar | 7001305949 At Low Cost Cash Payment Booking
Russian Call Girls Sadashivanagar | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...narwatsonia7
 
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...ggsonu500
 
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...scanFOAM
 
Call Girls South Delhi 9999965857 Cheap and Best with original Photos
Call Girls South Delhi 9999965857 Cheap and Best with original PhotosCall Girls South Delhi 9999965857 Cheap and Best with original Photos
Call Girls South Delhi 9999965857 Cheap and Best with original Photosparshadkalavatidevi7
 
Russian Call Girls South Delhi | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls South Delhi | 9711199171 | High Profile -New Model -Availa...Russian Call Girls South Delhi | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls South Delhi | 9711199171 | High Profile -New Model -Availa...satishsharma69855
 
MVP Health Care City of Schenectady Presentation
MVP Health Care City of Schenectady PresentationMVP Health Care City of Schenectady Presentation
MVP Health Care City of Schenectady PresentationMVP Health Care
 
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service MumbaiCollege Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...sandeepkumar69420
 

Dernier (20)

Gurgaon Sushant Lok Phase 2 Call Girls Service ( 9873940964 ) unlimited hard...
Gurgaon Sushant Lok Phase 2 Call Girls Service ( 9873940964 )  unlimited hard...Gurgaon Sushant Lok Phase 2 Call Girls Service ( 9873940964 )  unlimited hard...
Gurgaon Sushant Lok Phase 2 Call Girls Service ( 9873940964 ) unlimited hard...
 
Rohini Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Rohini Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Rohini Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Rohini Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
 
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near MeBook Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
 
Call Girls Hsr Layout Whatsapp 7001305949 Independent Escort Service
Call Girls Hsr Layout Whatsapp 7001305949 Independent Escort ServiceCall Girls Hsr Layout Whatsapp 7001305949 Independent Escort Service
Call Girls Hsr Layout Whatsapp 7001305949 Independent Escort Service
 
EMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareEMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical Care
 
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
 
Globalny raport: „Prawdziwe piękno 2024" od Dove
Globalny raport: „Prawdziwe piękno 2024" od DoveGlobalny raport: „Prawdziwe piękno 2024" od Dove
Globalny raport: „Prawdziwe piękno 2024" od Dove
 
Call Girls Laxmi Nagar 9999965857 Cheap and Best with original Photos
Call Girls Laxmi Nagar 9999965857 Cheap and Best with original PhotosCall Girls Laxmi Nagar 9999965857 Cheap and Best with original Photos
Call Girls Laxmi Nagar 9999965857 Cheap and Best with original Photos
 
Call Girls Delhi 9873940964 Elite Escort Service Available 24/7 Hire
Call Girls Delhi 9873940964 Elite Escort Service Available 24/7 HireCall Girls Delhi 9873940964 Elite Escort Service Available 24/7 Hire
Call Girls Delhi 9873940964 Elite Escort Service Available 24/7 Hire
 
Russian Call Girls Sadashivanagar | 7001305949 At Low Cost Cash Payment Booking
Russian Call Girls Sadashivanagar | 7001305949 At Low Cost Cash Payment BookingRussian Call Girls Sadashivanagar | 7001305949 At Low Cost Cash Payment Booking
Russian Call Girls Sadashivanagar | 7001305949 At Low Cost Cash Payment Booking
 
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
 
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
 
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
 
Russian Call Girls Jor Bagh 9711199171 discount on your booking
Russian Call Girls Jor Bagh 9711199171 discount on your bookingRussian Call Girls Jor Bagh 9711199171 discount on your booking
Russian Call Girls Jor Bagh 9711199171 discount on your booking
 
Call Girls South Delhi 9999965857 Cheap and Best with original Photos
Call Girls South Delhi 9999965857 Cheap and Best with original PhotosCall Girls South Delhi 9999965857 Cheap and Best with original Photos
Call Girls South Delhi 9999965857 Cheap and Best with original Photos
 
Russian Call Girls South Delhi | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls South Delhi | 9711199171 | High Profile -New Model -Availa...Russian Call Girls South Delhi | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls South Delhi | 9711199171 | High Profile -New Model -Availa...
 
MVP Health Care City of Schenectady Presentation
MVP Health Care City of Schenectady PresentationMVP Health Care City of Schenectady Presentation
MVP Health Care City of Schenectady Presentation
 
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service MumbaiCollege Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
 
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
 

OCT Machines

  • 1. DR. SAMARTH MISHRA Moderator: DR. DEBMALYA DAS Nil financial interest OCT MACHINESOCT MACHINES
  • 2. Principle of OCT machines TD- OCT in brief SD-OCT Spectralis OCT heidelberg SS-OCT in brief
  • 3. What is Optical coherence tomography (OCT) ? “OCT represents a non invasive method for cross sectional imaging of the internal retinal structures, through the detection of optical reflections and echo time delays of light, using low coherence interferometry, to subsequently produce in vivo two dimensional images of internal tissue microstructure.” Optical Coherence Tomography: Newer Techniques, Newer Machines
  • 4. • InterferometryInterferometry is a technique in which electromagnetic waves are superimposed causing the phenomenon of interference in order to extract information • In an interferometer, light from a single source is split into two beams that travel different optical paths, then combined again to produce interference. • The resulting interference fringes give information about the difference in optical path length. ‘Michelson interferometer’
  • 5.  Initially for in vivo measurements of axial length of eye by Fercher et al. (Medical University of Vienna, 1988)  In early 90´s David Huang, a PhD student at MIT, on his research on low-coherence interferometry developed OCT.  Cross-sectional tomographic imaging of retinal tissue.  Uses backscattered light.  Similar to ultrasound B-Scan imaging but uses infrared-light instead of ultrasound waves. Optical analog of ultrasound Fujimoto, James, and Eric Swanson. "The development, commercialization, and impact of optical coherence tomography." Investigative ophthalmology & visual science57.9 (2016): OCT1-OCT13.
  • 6. Fujimoto, James, and Eric Swanson. "The development, commercialization, and impact of optical coherence tomography." Investigative ophthalmology & visual science57.9 (2016): OCT1-OCT13. First OCT protoype at MIT
  • 7. Advantages:Advantages:  Obtains high resolution images  “Histological images” in vivo  In real time  Non-invasive  Uses harmless radiation  Ophthalmology, cardiology, odontology and dermatology
  • 9. Fujimoto J, Swanson E. The development, commercialization, and impact of optical coherence tomography. Invest Ophthalmol Vis Sci. 2016;57:OCT1–OCT13. DOI:10.1167/ iovs.16-19963
  • 10. Time domain ( TD-OCT) :Time domain ( TD-OCT) : • Stratus OCT, Carl zeiss meditech • Mobile reference arm mirror: allows backscattered tissue intensity levels to be detected from different depths in the tissue sample. • It is referred to as ‘time-domain’ because time-encoded signals are obtained directly. • 400 A scans/ sec • Axial resolution 10 um • lateral resolution 20um Longer acquisition time Motion artefacts Fujimoto, James, and Eric Swanson. "The development, commercialization, and impact of optical coherence tomography." Investigative ophthalmology & visual science57.9 (2016): OCT1-OCT13.
  • 11.
  • 12. Spectral domain(SD-OCT) / Fourier domain OCT(FD-OCT) :Spectral domain(SD-OCT) / Fourier domain OCT(FD-OCT) : • Wavelength 800-850um • Stationary reference arm • High speed spectrometer • A charged couple device(CCD) line scan camera • Acquisition speed of 25k to 52k A scans/sec • Axial resolution 5-7microns. • lateral resolution 14-20microns.
  • 14. ADVANTAGES: • Reduction of motion artefacts because of averaging. i.e multiple scans are acquired from the same position. • Better visualisation of individual retinal layers. • Improved signal to noise ratio
  • 15. Description of Commercially Available SD-OCT Systems Device (Manufacturer) Description • 3D-OCT 2000 (Topcon,Tokyo, Japan) SD-OCT and high-resolution fundus camera; axial resolution 5um; A-scan acquisition rate 27 kHz. • Bioptigen SD-OCT (Bioptigen,NC) Designed for both clinical and research use and includes a hand-held probe and microscope setup; axial resolution 4um; A-scan acquisition rate 20 kHz • Cirrus HD-OCT (Carl Zeiss Meditec, Dublin, CA) Software includes guided progression analysis for glaucoma progression detection; axial resolution, 5um; A-scan acquisition rate 27 kHz. Gabriele, Michelle L., et al. "Optical coherence tomography: history, current status, and laboratory work." Investigative ophthalmology & visual science 52.5 (2011): 2425-
  • 16. Device (Manufacturer) Description • RTVue-100 (Optovue, Fremont, CA) Offers multiple scanning protocols for glaucoma detection, including ganglion cell complex analysis; axial resolution 5um; A-scan acquisition rate 26 kHz. • SOCT Copernicus (Optopol, Zawiercie, Poland) Software includes progression analysis software that incorporates disk damage likelihood scale,asymmetry between the discs, and RNFL thickness; axial resolution, 6um; A-scan 27 kHz. • Spectral OCT SLO (Opko, Miami, FL) Combines SD-OCT, scanning laser ophthalmoscopy, and microperimetry. Axial resolution 6um, A-scan 27 kHz. Gabriele, Michelle L., et al. "Optical coherence tomography: history, current status, and laboratory work." Investigative ophthalmology & visual science 52.5 (2011): 2425-
  • 17. Device (Manufacturer) Devices • Spectralis OCT (Heidelberg Engineering, Heidelberg, Germany) High-speed SD-OCT device with eye-eye- trackingtracking, fluorescein angiographyfluorescein angiography, ICGICG angiography, and autofluorescenceautofluorescence. Axial resolution,4um4um; A-scan acquisition rate 40 kHz40 kHz. Gabriele, Michelle L., et al. "Optical coherence tomography: history, current status, and laboratory work." Investigative ophthalmology & visual science 52.5 (2011): 2425- 2436.
  • 18. Spectralis OCTSpectralis OCT ( Heidelberg Engineering , Germany )( Heidelberg Engineering , Germany ) • Dual beam SD-OCT and cSLO system. • Acquires volume scans of the retina and an infrared reference image simultaneously. • Eye tracker stablilizes the OCT scan • 40,000 A scans/sec
  • 19. • Short examination timing • Simultaneous real time FFA/OCT and ICG/OCT. • Spectralis HRA+OCT is a combination of the Spectralis HRA and Spectralis OCT. • Allows angiography, reflectance imaging, autofluorescence imaging, and OCT cross sectional imaging, alone or in various combinations.
  • 20.
  • 21.
  • 22. Confocal Laser ScanningConfocal Laser Scanning • For confocal images, a laser beam is focused on the retina. • SLO devices employ a confocal (pinhole) aperture, generating a single point of laser light, at a specific wavelength. • Can be scanned across the retina in a raster pattern. • As only a small area of the fundus is illuminated at any one time, the effects of light scatter are reduced and images with higher contrast are generated.
  • 23. • Spectralis HRA emit laser light with three different wavelengths: 1. For excitation of fluorescein, a blue solid state laser (wavelength 488 nm) is used. A barrier filter at 500 nm separates excitation and fluorescence light. The same wavelength (but without the barrier filter) is used to create blue reflectance and autofluorescence images. Representations of confocal laser scanning microscopy to vi-sualize fluorescence staining and reflection properties of isolated reti-nal tissue. aA laser beam is conducted through a dichroic mirror andan objective to excite the fluorescent dye in the specimen. bTheemitted fluorescent light is conducted back through the objective anda dichroic mirror (worked as a light filter) to the light detector (e.g.photomultiplier). The pinhole enables the exclusive detection of lightcoming from a certain depth in the focused tissue area (confocality). cIn the reflection mode of the confocal laser scanning microscope, thelaser light for excitation is reflected from cellular structures in theretinal tissue, passes optical filters and is detected by thephotomultiplier. The pinhole is used to allow confocality
  • 25. 2. For excitation of indocyanine green, a diode laser at 790 nm wavelength is used, together with a barrier filter at 830 nm to separate excitation and fluorescence light. 3.A second diode laser at 820 nm wavelength serves to produce infrared reflectance (IR) images.
  • 26. • SPECTRALIS® MultiColor Module incorporates a “multicolour” mode, where separate fundus images are obtained with blue, green, and near- infrared filters, and which can then be examined individually or in combination. • Shows distinct structures at different depths within the retina. • MultiColor images can highlight structures and pathologies not visible on ophthalmoscopy and fundus photography. CSCR MultiColor images may even be acquired in patients with cataract or nystagmus.
  • 27. •Longer wavelengths of light penetrate more deeply – the use of near infrared filters thus greatly improves visualization of subretinal and choroidal structures (e.g. drusen subtypes in age-related macular degeneration (AMD)). •Shorter wavelengths of light (blue light), allow improved imaging of superficial retinal structures (e.g. retinal nerve fibre layer defects in glaucoma and epiretinal membranes).
  • 28. •Intermediate wavelengths of light (green light (“red-free”)), provide optimal fundus images for general purposes. •Green wavelength is reflected by RPE, but absorbed by retinal haemorrhage and blood vessels.
  • 29. • In the simultaneous imaging modes, each image line is scanned twice subsequently with alternating lasers. • So, during simultaneous FFA and ICGA, only the 488 nm laser is switched on for the first scan of a line, while during the second scan only the 790 nm laser is switched on.
  • 30. Enhanced depth imaging ( EDI)Enhanced depth imaging ( EDI) • Spaide was the first to describe. • In standard OCT, the zero-delay line is typically at the vitreous inner retina junction, providing excellent resolution.
  • 31. • The “zero-delay” line is a reference point set by the software where the image capture is optimal. • EDI is achieved by moving the zero delay line deeper into the eye, focusing the OCT scanner on the choroid instead of the retina. • Moving away from the zero-delay line decreases the resolution from structures within the choroid resulting in diminished choroidal detail.
  • 32. Filter wheel position ‘R’ (Reflectance) How To Capture various modesHow To Capture various modes
  • 33. Filter wheel position ‘A’ (Angiography)
  • 34. Filter wheel position ‘P’ (Polarisation)
  • 35. xyz-Micromanipulator For the adjustment of the camera position: • Inner knob: Z adjustment (forward/backward) • Middle knob: Y adjustment (height) • outer knob: x adjustment (left/right) X Y Z
  • 36. Scanning techniques:Scanning techniques: • The commonly used scan protocols for macular scanning are three dimensional (3D)scan, radial scan, and raster scan.‑ • A 3D scan consists of a number of horizontal line scans composing a 6 mm × 6 mm/ 7 mm × 7 mm / 12 mm× 9 mm rectangular box. • A minimum pupil diameter of 3 mm is required to obtain a good OCT image.
  • 37. Types of Radial ScansTypes of Radial Scans 30°30° 4 Lines, 45°4 Lines, 45° 6 Lines, 30°6 Lines, 30° 12 Lines, 15°12 Lines, 15° 24 Lines, 7.5°24 Lines, 7.5° 48 Lines, 3.8°48 Lines, 3.8° 2 Lines, 90°2 Lines, 90°
  • 38. Types of Raster ScansTypes of Raster Scans Single line scanSingle line scan Detail Scan, 30umDetail Scan, 30um 7 line scan, 240um7 line scan, 240um Fast scan , 240umFast scan , 240um Dense Scan, 120umDense Scan, 120um
  • 39. • Thickness map displays the average retinal thickness values in 1 mm, 3 mm, and 6 mm diameter circles. • Divided into sectors with color map. • Warm colors indicate thicker retinal areas and cool color indicate thinner retinal areas
  • 40. • The SPECTRALIS® Widefield Imaging Module provides a 55-degree field of view for all fundus imaging modalities including MultiColor, BluePeak, IR, angiography, and OCT. • The Widefield Imaging Module facilitates comprehensive diagnostics. • The high-resolution, high-contrast scanning laser images captured with the Widefield Imaging Module offer diagnostic capabilities beyond those of conventional fundus photography. The optional Ultra-Widefield Angiography Module (102 °) delivers scanning laser angiography images from the macula to the periphery for more comprehensive retinal examinations.
  • 41. Changing lenses for ImagingChanging lenses for Imaging • To remove the objective lens , it is rotated by 50° to the left (counter clockwise). 30°Objective lens
  • 42. • The optional 55°wide angle lens • For IR, RF and angiography wide-field imaging • Has to be inserted with the red dot orientated upwards (12 o’clock) • Then rotated by about 50°clockwise until it snaps in. 55°wide angle lens
  • 43. Step 1 Step 2 Step 3 102°Ultra wide field lens
  • 44. FOCUS KNOB: • By turning this knob, refractive errors of the patient’s eye can be compensated. • For examination of most patients, an initial focus setting of ‘0’ (zero diopters) is suitable.
  • 45.
  • 46. • Also called Optical Frequency Domain OCT • Obtains time-encoded information by sweeping a narrow bandwidth laser through a broad optical spectrum • Backscattered intensity is detected with a photodetector. • This process is in contrast to SD-OCT, which uses a broad bandwidth light source and detects the interference spectra with a CCD camera and spectrometer. First high speed OFDI system: S.H. Yun et al, Opt. Express 11, 2953-2963 (2003) Swept source oct
  • 47. • Speeds of up to 249,000A-scans/s have been attained • Eye motion artifacts are greatly reduced compared with TD-OCT. • A drawback to camera-based SD-OCT detection is a drop-off in signal with depth of scanning because of the finite pixel size of the CCD camera. Taken from “Swept-Source Optical Coherence Tomography” presentation by Dr Muna Bhende
  • 48. • Wide 12 mm scans – evaluation of disc- macula relationships in one frame Taken from “Swept-Source Optical Coherence Tomography” presentation by Dr Muna Bhende
  • 49. 12 mm 12 mm 12 mm Taken from “Swept-Source Optical Coherence Tomography” presentation by Dr Muna Bhende 36 mm !!
  • 50. Comparison of TD-, SD-, and SS-OCT DevicesComparison of TD-, SD-, and SS-OCT Devices
  • 51. •Employs motion contrast imaging to high-resolution volumetric blood flow information  generating angiographic images. •Compares the ‘decorrelation signal’‘decorrelation signal’ between sequential OCT B-Scans taken at precisely the same cross section in order to construct a map of blood flow. Other OCT systems:Other OCT systems: Optical coherence tomography angiography (OCTA)Optical coherence tomography angiography (OCTA) En face OCT:En face OCT: •Derived from SD OCT. Produces frontal sections of retinal layers, also called ‘C-scan OCT’.
  • 52. Adaptive optics OCTAdaptive optics OCT • Corrects for higher-order ocular aberrations Doppler OCTDoppler OCT • measuring the Doppler shift and relative angle between the OCT beam and a blood vessel, assess the blood flow velocity. Polarization sensitive OCTPolarization sensitive OCT • used for functional tissue assessment by means of evaluation of light polarization which allows individual retinal layer identification Image taken from “Swept-Source Optical Coherence Tomography” presentation by Dr Muna Bhende
  • 53. OCT is a quick, non invasive and harmless examination technique Can be used to rapidly access the pathologies of retina and choroid Better and improved machines with better resolution. Future of OCT with newer advances is even more promising. OCT is here to stay !!OCT is here to stay !! To summarizeTo summarize
  • 55.
  • 56. • Automated segmentation into the “superficial” and “deep” inner retinal vascular plexuses, outer retina, and choriocapillaris • The OCT angiogram segmentation of the superficial inner retina contains a projection of the vasculature in the retinal nerve fiber layer (RNFL) and ganglion cell layer (GCL) Superficial capillary plexusSuperficial capillary plexus Deep capillary plexusDeep capillary plexus
  • 57. • The deep inner retina OCT angiogram shows a composite of the vascular plexuses at the border of the IPL/INL and the border of the INL/ OPL. Deep capillary plexusDeep capillary plexus choriocapillarieschoriocapillaries
  • 58. Sattler’s layerSattler’s layer Haller’s layerHaller’s layer