EPA/OPP MICROBIOLOGY LABORATORY
                                ESC, Ft. Meade, MD

                              Standard Operating Procedure
                                           for
                              Calibration of Thermometers

                                SOP Number: EQ-02-03

                                 Date Revised: 04-25-05



Prepared By: ________________________________________ Date: ___/___/___
               Print Name:______________________________



Reviewed By: ________________________________________ Date: ___/___/___
               Print Name:______________________________
               Technical Staff

               ________________________________________Date: ___/___/___
               Print Name:______________________________
               QA Officer

               ________________________________________Date: ___/___/___
               Print Name:______________________________
               Laboratory Director

Date Issued:   ___/___/___


Withdrawn By: ______________________________________ Date: ___/___/___


Controlled Copy No.: _________________
SOP No. EQ-02-03
                                                                                            Date Revised 04-25-05
                                                                                            Page 1 of 6



                                                 TABLE OF CONTENTS


                    Contents                                                                Page Number

 1.0 SCOPE AND APPLICATION..............................................................2

 2.0 DEFINITIONS......................................................................................2

 3.0 HEALTH AND SAFETY.....................................................................2

 4.0 CAUTIONS...........................................................................................2

 5.0 INTERFERENCES...............................................................................2

 6.0 PERSONNEL QUALIFICATIONS......................................................2

 7.0 SPECIAL APPARATUS AND MATERIALS.....................................2

 8.0 INSTRUMENT OR METHOD CALIBRATION.................................3

 9.0 SAMPLE HANDLING AND STORAGE............................................3

10.0 PROCEDURE AND ANALYSIS.........................................................3

11.0 DATA ANALYSIS/CALCULATIONS................................................4

12.0 DATA MANAGEMENT/RECORDS MANAGEMENT.....................5

13.0 QUALITY CONTROL..........................................................................5

14.0 NONCONFORMANCE AND CORRECTIVE ACTION....................5

15.0 REFERENCES......................................................................................5

16.0 FORMS AND DATA SHEETS............................................................5
SOP No. EQ-02-03
                                                                Date Revised 04-25-05
                                                                Page 2 of 6
1.0   SCOPE AND APPLICATION:

      1.1   This protocol describes the procedure used to point-check the accuracy of
            laboratory thermometers.

2.0   DEFINITIONS:

      2.1   NIST = National Institute of Standards and Technology.

      2.2   Correction Factor = the difference in the NIST traceable thermometer and the
            Observed Temperature measured by the NIST traceable thermometer. This is
            calculated and reported during the annual certification of the thermometers.

3.0   HEALTH AND SAFETY:

      3.1   In the event that a mercury thermometer breaks, evacuate the laboratory and call
            the SHEM manager (or contact security at extension 2600 if SHEM manager
            cannot be reached) immediately for assistance. Analysts are not to attempt to
            clean up the spill.

4.0   CAUTIONS: None

5.0   INTERFERENCES:

      5.1   Allow the thermometer to equilibrate with the solutions before taking the
            temperature reading.

6.0   PERSONNEL QUALIFICATIONS:

      6.1   Personnel are required to be knowledgeable of the procedures in this SOP.

7.0   SPECIAL APPARATUS AND MATERIALS:

      7.1   NIST traceable thermometer (ERTCO, serial number H99-106) certified annually
            by Barnstead International.

      7.2   NIST traceable thermometer (ERTCO, serial number 3321) certified annually by
            Barnstead International.

      7.3   NIST traceable Kessler maximum registration thermometers (ERTCO, serial
            numbers 3236, 3279, 3302, 3306, 3400, 4265, 4315, 4330 and 5936) certified
            annually by Barnstead International at 121EC. Thermometers are used to verify
            autoclave temperatures.
SOP No. EQ-02-03
                                                                   Date Revised 04-25-05
                                                                   Page 3 of 6

       7.4    Thermometers used to measure the temperature of water baths, incubators,
              refrigerators, and freezers, including FRIO-Temp Precision Thermometers (H-B
              Instrument Company). To determine which thermometers are currently in use in
              the laboratory, refer to the Point Check Calibration of Laboratory Thermometers
              Record Book.

       7.5    NIST traceable digital thermometers for freezers: 1) Fisher Scientific, model No.
              06-664-23, serial number 240106521, 2) Fisher Scientific, model No. 15-077-14,
              serial number 230317798, and 3) Fisher Scientific, model No. 15-077-14, serial
              number 41299983.

8.0    INSTRUMENT OR METHOD CALIBRATION:

       8.1    All thermometers are calibrated using a NIST traceable thermometer. The NIST
              traceable thermometers are certified annually by a contractor outside of the OPP
              Microbiology Laboratory.

9.0    SAMPLE HANDLING AND STORAGE: Not applicable

10.0   PROCEDURE AND ANALYSIS:

       10.1   Annual Thermometer Calibration Reports for NIST traceable thermometers
              (sections 7.1, 7.2, and 7.3) are stored in the Thermometers Calibration Certificates
              notebook.

       10.2   Once a year, all of the thermometers falling under section 7.4 of this SOP are
              checked at operating temperatures against one of the NIST traceable
              thermometers (sections 7.1 and 7.2).

       10.3   To point-check the accuracy of FRIO-Temp Precision Thermometers, place a
              flask of de-ionized water or ethanol next to the thermometer in the incubator,
              refrigerator, or freezer. Place the NIST traceable thermometer into the flask of
              de-ionized water or ethanol. Once the de-ionized water or ethanol has reached the
              temperature of interest, conduct the accuracy check. The NIST thermometer must
              be suspended in the liquid in the flask, not resting on the bottom of the flask.

       10.4   A water bath is used, where applicable, to check the temperature of the
              thermometers (excluding the FRIO-Temp Precision Thermometers) by
              simultaneous immersion of the NIST traceable thermometer and the thermometer
              to be calibrated into the water bath at the temperature of interest. Otherwise,
              point checks are taken by simultaneous immersion of the thermometers in a flask
              of de-ionized water or ethanol and then placing the flask containing both
SOP No. EQ-02-03
                                                                Date Revised 04-25-05
                                                                Page 4 of 6
              thermometers in the instrument being monitored until the de-ionized water or
              ethanol reaches the temperature of interest.

       10.5   Any difference in temperature readings between the NIST traceable thermometer
              and the laboratory thermometer is recorded on the Point-Check Calibration of
              Laboratory Thermometers Record Form (see 16.0).

              A label displaying the date of calibration and correction factor (even if it is zero),
              is placed around the top of the corresponding thermometer.

       10.6   Digital thermometers are calibrated against NIST traceable instrumentation and
              are indicated to be within tolerance. No correction factor is indicated (see
              certificates of NIST traceability).

11.0   DATA ANALYSIS/CALCULATIONS:

       11.1   The Check Point for both the NIST thermometer and the thermometer being
              calibrated is the reading on the NIST thermometer during the point check process.
              The temperature reading on the NIST thermometer must be within the range at
              which the thermometer to be calibrated will be used or is being used (e.g., from 2-
              5°C for thermometers in refrigerators, from -10°C to -20°C for thermometers in
              freezers, from 36°C to 38°C for thermometers in incubators).

       11.2   The Observed Temperature is the temperature reading of the thermometer during
              the check point process.

       11.3   The correction factor for the NIST thermometer is determined by the certified
              company that performs the calibration verification for this thermometer. The
              True Temperature measured by the NIST traceable thermometer = the Observed
              Temperature of the NIST traceable thermometer plus its correction factor.

       11.4   The correction factor for the laboratory thermometer that is being calibrated
              against the NIST traceable thermometer is calculated by subtracting the Observed
              Temperature of the laboratory thermometer from the True temperature of the
              NIST traceable thermometer.

       11.5   The True Temperature measured by the laboratory thermometer = the Observed
              Temperature plus the correction factor for that thermometer.

12.0   DATA MANAGEMENT/RECORDS MANAGEMENT:

       12.1   Data will be recorded promptly, legibly and in indelible ink on the Point Check
              Calibration of Laboratory Thermometers Record Form. Completed forms are
SOP No. EQ-02-03
                                                                    Date Revised 04-25-05
                                                                    Page 5 of 6
              archived in the Point Check Calibration of Laboratory Thermometers Record
              Book. Annual Thermometer Calibration Reports for NIST traceable
              thermometers (sections 7.1, 7.2, and 7.3) are stored in the Thermometers and
              Hygrometer Calibration Certificates notebook. The books are kept in a secured
              file cabinet in the file room D217. Only authorized personnel have access to the
              secured files. Archived data are subject to OPP=s official retention schedule
              contained in SOP ADM-03, Records and Archives.

13.0   QUALITY CONTROL:

       13.1   The OPP Microbiology laboratory conforms to 40 CFR Part 160, Good
              Laboratory Practices. Appropriate quality control measures are integrated into
              each SOP.

       13.2   Thermometers are point-checked once a year either in the laboratory or by a
              company certified to perform calibration verification of thermometers.

       13.3   For quality control purposes, the required information is documented on the
              appropriate form(s) (see 16.0).

14.0   NONCONFORMANCE AND CORRECTIVE ACTION:

       14.1   When routinely recording temperatures for the laboratory, laboratory equipment,
              media etc., the observed temperature reading of the thermometer read plus the
              correction factor for that specific thermometer must be recorded.

       14.2   On occasion, thermometers must be discarded (e.g., broken columns,
              unreasonably large correction factor). Contact the facility Safety, Health, and
              Environmental Management Program Manager for proper disposal procedures.

15.0   REFERENCES: None

16.0   FORMS AND DATA SHEETS:

       16.1   Point-Check Calibration of Laboratory Thermometers Record Form




Point Check Calibration of Laboratory Thermometers Record
OPP Microbiology Laboratory
SOP No. EQ-02-03
                                                                                       Date Revised 04-25-05
                                                                                       Page 6 of 6
                                                                                                                     True
                              Thermometer                         Check-         Observed         Correction        Temp.
   Date        Initials           Type            Serial #        point†         Temp.††           Factor *           **

                           NIST Traceable



 Description of Calibration




                            Thermometer                           Check-        Observed        Correction      True Temp.
   Date       Initials          Type              Serial #        point†        Temp.††          Factor *           **

                          NIST Traceable



 Description of Calibration




† Reading on the NIST thermometer during the point-check process. The temperature reading on the NIST thermometer must be
within the range at which the thermometer to be calibrated will be or is being used (e.g., from 2°C to 5°C for thermometers in
refrigerators, from -10°C to -20°C for thermometers in freezers, from 36°C to 38°C for thermometers in incubators).
†† Temperature reading of thermometer during the point-check process.
* Correction factor for NIST Thermometer determined by the Certification company.
    Correction factor for the thermometer being calibrated is calculated by subtracting its observed temperature from true
     temperature of NIST Thermometer
** True Temperature of NIST Thermometer = Observed Temp.+ Correction factor :
   True temperature of Thermometer being Calibrated = Observed Temp. + correction factor for that thermometer.

More Related Content

What's hot

Temperature Measuring and Control
Temperature Measuring and ControlTemperature Measuring and Control
Temperature Measuring and Control
JIMMY MANDLA KABINI
 

What's hot (20)

Using a Portable Bath to Improve Calibration of Sanitary RTD's and Transmitte...
Using a Portable Bath to Improve Calibration of Sanitary RTD's and Transmitte...Using a Portable Bath to Improve Calibration of Sanitary RTD's and Transmitte...
Using a Portable Bath to Improve Calibration of Sanitary RTD's and Transmitte...
 
Contact temperature sensor calibration
Contact temperature sensor calibrationContact temperature sensor calibration
Contact temperature sensor calibration
 
Field Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke CalibrationField Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke Calibration
 
Testing and Troubleshooting 4-20 mA Control Loops Presented by Fluke and Tra...
Testing and Troubleshooting 4-20 mA Control Loops  Presented by Fluke and Tra...Testing and Troubleshooting 4-20 mA Control Loops  Presented by Fluke and Tra...
Testing and Troubleshooting 4-20 mA Control Loops Presented by Fluke and Tra...
 
How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2
 
Temperature Screening Solutions and Returning to Work Safely Webinar with Ame...
Temperature Screening Solutions and Returning to Work Safely Webinar with Ame...Temperature Screening Solutions and Returning to Work Safely Webinar with Ame...
Temperature Screening Solutions and Returning to Work Safely Webinar with Ame...
 
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
 
Transcat and Fluke Webinar: Streamlining Instrument Calibration
Transcat and Fluke Webinar: Streamlining Instrument CalibrationTranscat and Fluke Webinar: Streamlining Instrument Calibration
Transcat and Fluke Webinar: Streamlining Instrument Calibration
 
Temperature Calibration equipment for Thermcouple, RTDs, and Pyrometers
Temperature Calibration equipment for Thermcouple, RTDs, and PyrometersTemperature Calibration equipment for Thermcouple, RTDs, and Pyrometers
Temperature Calibration equipment for Thermcouple, RTDs, and Pyrometers
 
Industrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke CalibrationIndustrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke Calibration
 
Transcat and Fluke Calibration Webinar: Automating Pressure Calibration
Transcat and Fluke Calibration Webinar: Automating Pressure CalibrationTranscat and Fluke Calibration Webinar: Automating Pressure Calibration
Transcat and Fluke Calibration Webinar: Automating Pressure Calibration
 
Temperature calibration
Temperature calibrationTemperature calibration
Temperature calibration
 
The beginners guide to heat treatment control v2
The beginners guide to heat treatment control v2The beginners guide to heat treatment control v2
The beginners guide to heat treatment control v2
 
Fluke Calibration Tips for High Pressure Calibration
Fluke Calibration Tips for High Pressure CalibrationFluke Calibration Tips for High Pressure Calibration
Fluke Calibration Tips for High Pressure Calibration
 
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
 
Calibration
CalibrationCalibration
Calibration
 
Understanding cqi9 iss3 (intermediate)
Understanding cqi9 iss3 (intermediate)Understanding cqi9 iss3 (intermediate)
Understanding cqi9 iss3 (intermediate)
 
Overcoming Challenges in Pressure Calibration Webinar
Overcoming Challenges in Pressure Calibration WebinarOvercoming Challenges in Pressure Calibration Webinar
Overcoming Challenges in Pressure Calibration Webinar
 
Evaluating Out Of Tolerance Instruments Webinar
Evaluating Out Of Tolerance Instruments WebinarEvaluating Out Of Tolerance Instruments Webinar
Evaluating Out Of Tolerance Instruments Webinar
 
Temperature Measuring and Control
Temperature Measuring and ControlTemperature Measuring and Control
Temperature Measuring and Control
 

Viewers also liked

Relief and safety valves for thermal power plants
Relief and safety valves for thermal power plantsRelief and safety valves for thermal power plants
Relief and safety valves for thermal power plants
SHIVAJI CHOUDHURY
 
control valve functional testing - Swetha
control valve functional testing - Swethacontrol valve functional testing - Swetha
control valve functional testing - Swetha
Swetha Sundaram
 
Control valve positioners
Control valve positionersControl valve positioners
Control valve positioners
Sagar Jathan
 

Viewers also liked (16)

Calibration dasibi-ozone
Calibration dasibi-ozoneCalibration dasibi-ozone
Calibration dasibi-ozone
 
Calibration
CalibrationCalibration
Calibration
 
Preassure Safety Valve
Preassure Safety ValvePreassure Safety Valve
Preassure Safety Valve
 
Instrumentation positioner presentation
Instrumentation positioner presentationInstrumentation positioner presentation
Instrumentation positioner presentation
 
Digital thermometer ppt
Digital thermometer pptDigital thermometer ppt
Digital thermometer ppt
 
Equipment calibration PPT by Shravan Kumar
Equipment calibration PPT by Shravan KumarEquipment calibration PPT by Shravan Kumar
Equipment calibration PPT by Shravan Kumar
 
104980410 control-valve-calibration-procedure-fisher-hc6010
104980410 control-valve-calibration-procedure-fisher-hc6010104980410 control-valve-calibration-procedure-fisher-hc6010
104980410 control-valve-calibration-procedure-fisher-hc6010
 
Testing of Safety Valves
Testing of Safety ValvesTesting of Safety Valves
Testing of Safety Valves
 
Relief and safety valves for thermal power plants
Relief and safety valves for thermal power plantsRelief and safety valves for thermal power plants
Relief and safety valves for thermal power plants
 
Thermometers
ThermometersThermometers
Thermometers
 
Instrument Calibration
Instrument Calibration Instrument Calibration
Instrument Calibration
 
01 General Control Valves Training.
01 General Control Valves Training.01 General Control Valves Training.
01 General Control Valves Training.
 
Control valve ppt
Control valve pptControl valve ppt
Control valve ppt
 
control valve functional testing - Swetha
control valve functional testing - Swethacontrol valve functional testing - Swetha
control valve functional testing - Swetha
 
Control valve positioners
Control valve positionersControl valve positioners
Control valve positioners
 
Valve Testing Guidelines
Valve Testing GuidelinesValve Testing Guidelines
Valve Testing Guidelines
 

Similar to Thermometers calibration

E+H-Transmitter sensor matching improves RTD accuracy
E+H-Transmitter sensor matching improves RTD accuracyE+H-Transmitter sensor matching improves RTD accuracy
E+H-Transmitter sensor matching improves RTD accuracy
Carotek
 
Krishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instrumentsKrishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instruments
eslamgouda
 
temperature measurement
temperature measurementtemperature measurement
temperature measurement
DimaJawhar
 
Quality DocumentationDocumentation OverviewThe Q.docx
Quality DocumentationDocumentation OverviewThe Q.docxQuality DocumentationDocumentation OverviewThe Q.docx
Quality DocumentationDocumentation OverviewThe Q.docx
makdul
 

Similar to Thermometers calibration (20)

Fluke Temperature Calibration Catalog
Fluke Temperature Calibration CatalogFluke Temperature Calibration Catalog
Fluke Temperature Calibration Catalog
 
TEMPERATURE TRANSMITTER TEMPE TEMPERATURE TRANSMITTER
TEMPERATURE TRANSMITTER TEMPE TEMPERATURE TRANSMITTERTEMPERATURE TRANSMITTER TEMPE TEMPERATURE TRANSMITTER
TEMPERATURE TRANSMITTER TEMPE TEMPERATURE TRANSMITTER
 
TEMPERATURE TRANSMITTER_2023 TEMPERATURE TRANSMITTER_2023
TEMPERATURE TRANSMITTER_2023 TEMPERATURE TRANSMITTER_2023TEMPERATURE TRANSMITTER_2023 TEMPERATURE TRANSMITTER_2023
TEMPERATURE TRANSMITTER_2023 TEMPERATURE TRANSMITTER_2023
 
Understanding AMS2750E (intermediate) v2
Understanding AMS2750E (intermediate) v2Understanding AMS2750E (intermediate) v2
Understanding AMS2750E (intermediate) v2
 
Hướng dẫn sử dụng testo 550
Hướng dẫn sử dụng testo 550Hướng dẫn sử dụng testo 550
Hướng dẫn sử dụng testo 550
 
Hướng dẫn sử dụng máy đo đa năng Testo 550 KIT
Hướng dẫn sử dụng máy đo đa năng Testo 550 KIT Hướng dẫn sử dụng máy đo đa năng Testo 550 KIT
Hướng dẫn sử dụng máy đo đa năng Testo 550 KIT
 
Oep
OepOep
Oep
 
E+H-Transmitter sensor matching improves RTD accuracy
E+H-Transmitter sensor matching improves RTD accuracyE+H-Transmitter sensor matching improves RTD accuracy
E+H-Transmitter sensor matching improves RTD accuracy
 
Krishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instrumentsKrishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instruments
 
Thermal mass flow meter English manual
Thermal mass flow meter English manualThermal mass flow meter English manual
Thermal mass flow meter English manual
 
Hướng dẫn sử dụng Testo 310
Hướng dẫn sử dụng Testo 310Hướng dẫn sử dụng Testo 310
Hướng dẫn sử dụng Testo 310
 
Hướng dẫn sử dụng Testo 310
Hướng dẫn sử dụng Testo 310Hướng dẫn sử dụng Testo 310
Hướng dẫn sử dụng Testo 310
 
Calibration times march 2011
Calibration times march 2011Calibration times march 2011
Calibration times march 2011
 
Calibration times march 2011
Calibration times march 2011Calibration times march 2011
Calibration times march 2011
 
custody-transfer-temperature-measurement-api-mpms.pdf
custody-transfer-temperature-measurement-api-mpms.pdfcustody-transfer-temperature-measurement-api-mpms.pdf
custody-transfer-temperature-measurement-api-mpms.pdf
 
Hướng dẫn sử dụng Teso 570
Hướng dẫn sử dụng Teso 570Hướng dẫn sử dụng Teso 570
Hướng dẫn sử dụng Teso 570
 
calibration.pptx
calibration.pptxcalibration.pptx
calibration.pptx
 
temperature measurement
temperature measurementtemperature measurement
temperature measurement
 
350619 634044039582135000
350619 634044039582135000350619 634044039582135000
350619 634044039582135000
 
Quality DocumentationDocumentation OverviewThe Q.docx
Quality DocumentationDocumentation OverviewThe Q.docxQuality DocumentationDocumentation OverviewThe Q.docx
Quality DocumentationDocumentation OverviewThe Q.docx
 

Recently uploaded

Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Victor Rentea
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
panagenda
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 

Recently uploaded (20)

Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
 
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, AdobeApidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
 
Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
 
Introduction to Multilingual Retrieval Augmented Generation (RAG)
Introduction to Multilingual Retrieval Augmented Generation (RAG)Introduction to Multilingual Retrieval Augmented Generation (RAG)
Introduction to Multilingual Retrieval Augmented Generation (RAG)
 
Elevate Developer Efficiency & build GenAI Application with Amazon Q​
Elevate Developer Efficiency & build GenAI Application with Amazon Q​Elevate Developer Efficiency & build GenAI Application with Amazon Q​
Elevate Developer Efficiency & build GenAI Application with Amazon Q​
 
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ..."I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a Fresher
 
MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024
 
Boost Fertility New Invention Ups Success Rates.pdf
Boost Fertility New Invention Ups Success Rates.pdfBoost Fertility New Invention Ups Success Rates.pdf
Boost Fertility New Invention Ups Success Rates.pdf
 
CNIC Information System with Pakdata Cf In Pakistan
CNIC Information System with Pakdata Cf In PakistanCNIC Information System with Pakdata Cf In Pakistan
CNIC Information System with Pakdata Cf In Pakistan
 
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
Corporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptxCorporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptx
 
DBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor PresentationDBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor Presentation
 
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot TakeoffStrategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
 

Thermometers calibration

  • 1. EPA/OPP MICROBIOLOGY LABORATORY ESC, Ft. Meade, MD Standard Operating Procedure for Calibration of Thermometers SOP Number: EQ-02-03 Date Revised: 04-25-05 Prepared By: ________________________________________ Date: ___/___/___ Print Name:______________________________ Reviewed By: ________________________________________ Date: ___/___/___ Print Name:______________________________ Technical Staff ________________________________________Date: ___/___/___ Print Name:______________________________ QA Officer ________________________________________Date: ___/___/___ Print Name:______________________________ Laboratory Director Date Issued: ___/___/___ Withdrawn By: ______________________________________ Date: ___/___/___ Controlled Copy No.: _________________
  • 2. SOP No. EQ-02-03 Date Revised 04-25-05 Page 1 of 6 TABLE OF CONTENTS Contents Page Number 1.0 SCOPE AND APPLICATION..............................................................2 2.0 DEFINITIONS......................................................................................2 3.0 HEALTH AND SAFETY.....................................................................2 4.0 CAUTIONS...........................................................................................2 5.0 INTERFERENCES...............................................................................2 6.0 PERSONNEL QUALIFICATIONS......................................................2 7.0 SPECIAL APPARATUS AND MATERIALS.....................................2 8.0 INSTRUMENT OR METHOD CALIBRATION.................................3 9.0 SAMPLE HANDLING AND STORAGE............................................3 10.0 PROCEDURE AND ANALYSIS.........................................................3 11.0 DATA ANALYSIS/CALCULATIONS................................................4 12.0 DATA MANAGEMENT/RECORDS MANAGEMENT.....................5 13.0 QUALITY CONTROL..........................................................................5 14.0 NONCONFORMANCE AND CORRECTIVE ACTION....................5 15.0 REFERENCES......................................................................................5 16.0 FORMS AND DATA SHEETS............................................................5
  • 3. SOP No. EQ-02-03 Date Revised 04-25-05 Page 2 of 6 1.0 SCOPE AND APPLICATION: 1.1 This protocol describes the procedure used to point-check the accuracy of laboratory thermometers. 2.0 DEFINITIONS: 2.1 NIST = National Institute of Standards and Technology. 2.2 Correction Factor = the difference in the NIST traceable thermometer and the Observed Temperature measured by the NIST traceable thermometer. This is calculated and reported during the annual certification of the thermometers. 3.0 HEALTH AND SAFETY: 3.1 In the event that a mercury thermometer breaks, evacuate the laboratory and call the SHEM manager (or contact security at extension 2600 if SHEM manager cannot be reached) immediately for assistance. Analysts are not to attempt to clean up the spill. 4.0 CAUTIONS: None 5.0 INTERFERENCES: 5.1 Allow the thermometer to equilibrate with the solutions before taking the temperature reading. 6.0 PERSONNEL QUALIFICATIONS: 6.1 Personnel are required to be knowledgeable of the procedures in this SOP. 7.0 SPECIAL APPARATUS AND MATERIALS: 7.1 NIST traceable thermometer (ERTCO, serial number H99-106) certified annually by Barnstead International. 7.2 NIST traceable thermometer (ERTCO, serial number 3321) certified annually by Barnstead International. 7.3 NIST traceable Kessler maximum registration thermometers (ERTCO, serial numbers 3236, 3279, 3302, 3306, 3400, 4265, 4315, 4330 and 5936) certified annually by Barnstead International at 121EC. Thermometers are used to verify autoclave temperatures.
  • 4. SOP No. EQ-02-03 Date Revised 04-25-05 Page 3 of 6 7.4 Thermometers used to measure the temperature of water baths, incubators, refrigerators, and freezers, including FRIO-Temp Precision Thermometers (H-B Instrument Company). To determine which thermometers are currently in use in the laboratory, refer to the Point Check Calibration of Laboratory Thermometers Record Book. 7.5 NIST traceable digital thermometers for freezers: 1) Fisher Scientific, model No. 06-664-23, serial number 240106521, 2) Fisher Scientific, model No. 15-077-14, serial number 230317798, and 3) Fisher Scientific, model No. 15-077-14, serial number 41299983. 8.0 INSTRUMENT OR METHOD CALIBRATION: 8.1 All thermometers are calibrated using a NIST traceable thermometer. The NIST traceable thermometers are certified annually by a contractor outside of the OPP Microbiology Laboratory. 9.0 SAMPLE HANDLING AND STORAGE: Not applicable 10.0 PROCEDURE AND ANALYSIS: 10.1 Annual Thermometer Calibration Reports for NIST traceable thermometers (sections 7.1, 7.2, and 7.3) are stored in the Thermometers Calibration Certificates notebook. 10.2 Once a year, all of the thermometers falling under section 7.4 of this SOP are checked at operating temperatures against one of the NIST traceable thermometers (sections 7.1 and 7.2). 10.3 To point-check the accuracy of FRIO-Temp Precision Thermometers, place a flask of de-ionized water or ethanol next to the thermometer in the incubator, refrigerator, or freezer. Place the NIST traceable thermometer into the flask of de-ionized water or ethanol. Once the de-ionized water or ethanol has reached the temperature of interest, conduct the accuracy check. The NIST thermometer must be suspended in the liquid in the flask, not resting on the bottom of the flask. 10.4 A water bath is used, where applicable, to check the temperature of the thermometers (excluding the FRIO-Temp Precision Thermometers) by simultaneous immersion of the NIST traceable thermometer and the thermometer to be calibrated into the water bath at the temperature of interest. Otherwise, point checks are taken by simultaneous immersion of the thermometers in a flask of de-ionized water or ethanol and then placing the flask containing both
  • 5. SOP No. EQ-02-03 Date Revised 04-25-05 Page 4 of 6 thermometers in the instrument being monitored until the de-ionized water or ethanol reaches the temperature of interest. 10.5 Any difference in temperature readings between the NIST traceable thermometer and the laboratory thermometer is recorded on the Point-Check Calibration of Laboratory Thermometers Record Form (see 16.0). A label displaying the date of calibration and correction factor (even if it is zero), is placed around the top of the corresponding thermometer. 10.6 Digital thermometers are calibrated against NIST traceable instrumentation and are indicated to be within tolerance. No correction factor is indicated (see certificates of NIST traceability). 11.0 DATA ANALYSIS/CALCULATIONS: 11.1 The Check Point for both the NIST thermometer and the thermometer being calibrated is the reading on the NIST thermometer during the point check process. The temperature reading on the NIST thermometer must be within the range at which the thermometer to be calibrated will be used or is being used (e.g., from 2- 5°C for thermometers in refrigerators, from -10°C to -20°C for thermometers in freezers, from 36°C to 38°C for thermometers in incubators). 11.2 The Observed Temperature is the temperature reading of the thermometer during the check point process. 11.3 The correction factor for the NIST thermometer is determined by the certified company that performs the calibration verification for this thermometer. The True Temperature measured by the NIST traceable thermometer = the Observed Temperature of the NIST traceable thermometer plus its correction factor. 11.4 The correction factor for the laboratory thermometer that is being calibrated against the NIST traceable thermometer is calculated by subtracting the Observed Temperature of the laboratory thermometer from the True temperature of the NIST traceable thermometer. 11.5 The True Temperature measured by the laboratory thermometer = the Observed Temperature plus the correction factor for that thermometer. 12.0 DATA MANAGEMENT/RECORDS MANAGEMENT: 12.1 Data will be recorded promptly, legibly and in indelible ink on the Point Check Calibration of Laboratory Thermometers Record Form. Completed forms are
  • 6. SOP No. EQ-02-03 Date Revised 04-25-05 Page 5 of 6 archived in the Point Check Calibration of Laboratory Thermometers Record Book. Annual Thermometer Calibration Reports for NIST traceable thermometers (sections 7.1, 7.2, and 7.3) are stored in the Thermometers and Hygrometer Calibration Certificates notebook. The books are kept in a secured file cabinet in the file room D217. Only authorized personnel have access to the secured files. Archived data are subject to OPP=s official retention schedule contained in SOP ADM-03, Records and Archives. 13.0 QUALITY CONTROL: 13.1 The OPP Microbiology laboratory conforms to 40 CFR Part 160, Good Laboratory Practices. Appropriate quality control measures are integrated into each SOP. 13.2 Thermometers are point-checked once a year either in the laboratory or by a company certified to perform calibration verification of thermometers. 13.3 For quality control purposes, the required information is documented on the appropriate form(s) (see 16.0). 14.0 NONCONFORMANCE AND CORRECTIVE ACTION: 14.1 When routinely recording temperatures for the laboratory, laboratory equipment, media etc., the observed temperature reading of the thermometer read plus the correction factor for that specific thermometer must be recorded. 14.2 On occasion, thermometers must be discarded (e.g., broken columns, unreasonably large correction factor). Contact the facility Safety, Health, and Environmental Management Program Manager for proper disposal procedures. 15.0 REFERENCES: None 16.0 FORMS AND DATA SHEETS: 16.1 Point-Check Calibration of Laboratory Thermometers Record Form Point Check Calibration of Laboratory Thermometers Record OPP Microbiology Laboratory
  • 7. SOP No. EQ-02-03 Date Revised 04-25-05 Page 6 of 6 True Thermometer Check- Observed Correction Temp. Date Initials Type Serial # point† Temp.†† Factor * ** NIST Traceable Description of Calibration Thermometer Check- Observed Correction True Temp. Date Initials Type Serial # point† Temp.†† Factor * ** NIST Traceable Description of Calibration † Reading on the NIST thermometer during the point-check process. The temperature reading on the NIST thermometer must be within the range at which the thermometer to be calibrated will be or is being used (e.g., from 2°C to 5°C for thermometers in refrigerators, from -10°C to -20°C for thermometers in freezers, from 36°C to 38°C for thermometers in incubators). †† Temperature reading of thermometer during the point-check process. * Correction factor for NIST Thermometer determined by the Certification company. Correction factor for the thermometer being calibrated is calculated by subtracting its observed temperature from true temperature of NIST Thermometer ** True Temperature of NIST Thermometer = Observed Temp.+ Correction factor : True temperature of Thermometer being Calibrated = Observed Temp. + correction factor for that thermometer.