12. IO Insertion
Attach the provided EZ-connect extension set to the
catheter hub’s standard luer lock
Prepare to give a
10 ml syringe flush
13. IO Insertion
Do not attach a syringe directly to the EZ-IO catheter
Attaching a syringe
may lead to an
increase in the
diameter of the
insertion hole
Extravasation
15. IO Confirmation
• Catheter is firmly seated and does not move
• You note blood at catheter hub
• You are able to aspirate blood or marrow from
catheter
– We recommend aspiration of only a small amount of
blood due to its extremely viscous nature
16. IO Confirmation
• Drugs or fluids flow without difficulty; no signs of
leakage in or around tissue
– CAUTION: concious patients will experience pain with
infusion prior to lidocaine!
• Flow rates may be slow or non existent prior to the 10 ml
bolus
• Check stylet tip for blood prior to placing it in the
shuttle or the bio hazard container