nice NeÖtech Medical Systems Pvt. Ltd
Presentation Flow
COMPANY OVERVIEW
CLINICAL ASPECTS OF nice 8000
series HUMIDIFIER
PRODUCT OVERVIEW
TROUBLESHOOTING
Company INFO
 Established in the year 1997
 Group of highly skilled professionals
 4000+ highly satisfied customers
 Deep penetration in NICU and respiratory domain
 User friendly range of products
 With deep skills and high end technology , we always strive
for growth and delighted customers
Timeline-nice Neotech
2005
•First New
Factory Unit 1
Opened
•Increase in
production Line
2001
•Registered as
Pvt ltd
•Production
started for
different models
2013
•Second New
Factory Unit 2
Opened
1999
•Neotech Logo
Designed
•Designed first
incubator & open
care prototype
2003
•Export to
Vietnam
2009-
2011
•CE certification
for products
2015
•125+skilled
employee strong
firm
•Multiple range
of NICU products
1997
•Started Neotech
Medical Systems
with 2 employees
•First Infant
Radiant Warmer
Why nice Neotech ?
 User friendly range of products
 Customized solutions
 Ethical Business practices
 Stringent quality control management
 Complete customer satisfaction
Humidity
Humidity is the presence of water vapour in gas.
Absolute humidity is the amount of water vapour per litre of gas,
expressed in mg/l.
Relative humidity is a measure of how much water vapour is in a
gas compared to its capacity to hold water vapour. It is measured as
a percentage.
Principle
•When dry gas passes over a moist surface evaporation occurs and
the gas becomes more humid.
•When moist gas cools, water droplets condense and the gas
becomes less Humid.
•The warmer a gas, the more moisture (absolute humidity) it can
carry.
•If a gas already has 100% humidity no more moisture will
evaporate into it unless it is heated further.
What is Respiratory Humidifier?
During mechanical ventilation ,When dry air/gas is administered to the
patient there occurs airway cooling, airway drying ,inflammation which can
lead to many complications. Optimum humidity is required for ventilation -
33mg/L in non invasive mode and 44 mg/L in invasive mode. Exposure to
humidity levels below 25mg H2O /l or 100% saturated gas can lead to
respiratory complications. Less humidity can cause airway mucosal
dysfunction and more humidity can cause reduced mucus viscosity and
increased pericellular depth fluid.
To avoid the above said conditions and to administer optimal humidity.
respiratory humidifiers are used during ventilation. Respiratory Humidifiers
heats the water contained in the humidification chamber, humidifying the air
passing through it.
Why Humidifiers are Used?
Prevent Airway
Cooling
Prevent Airway Drying
and Water Loss
Prevent Airway
Inflammation and
Constriction
Ensure patient is
stabilized
Infant
Humidifiers
Less Humidity = More Evaporation
Introducing nice NeÖtech 8000 series
Respiratory Humidifiers
Designed to perfection
nice 8000 series Respiratory Humidifiers
•Invasive or non invasive control mode with flow detection.
•Flexibility in Operation
•Invasive ventilation, Infant Bubble CPAP, Resuscitation, and Oxygen therapy
for Neonatal Care
•for invasive ventilation, non invasive ventilation and oxygen therapy for Adult
Care
•Dual servo control feedback system
•Comprehensive audible & visual alarm
•Water alarm if water alarm falls below required level
•High and low flow ventilation therapies
•Heated and non heated wire breathing system
•Precise, reliable performance
•Compact and Handy
•Mounting Options
Features
How to operate?
Control Panel Operation
Alarms
Heater Wire Operation
Non-Heater Wire Operation
 Automatic Humidity Compensation
Display & Controls
Indicators
Control Panel Operation
Temperature Probe : if the temperature probe is damaged or disconnected
Heater Wire : if the heater wire is defected or disconnected
Heater : if the heater is defected or disconnected
Maximum System
Operating Pressure : 20 kPa,
Gas leakage at max.
Pressure : <100 mL/minute.
Alarms
•Humidified gas from the chamber pass through the inspiratory limb, where
its temperature must be maintained in order to prevent the generated
humidity from condensing.
•This is achieved with a heater wire encapsulated within the inspiratory
limb. The humidifier maintains the temperature along the inspiratory limb
by monitoring the temperature at the airway probe and controlling the
voltage delivered to the heater wire.
• Under normal conditions the gas is heated to 40 °C in the invasive mode,
34 °C for the non-invasive mode.
Heater Wire Operation
Respiratory Humidifier maintains the airway temperature at the desired set
point (invasive 37 °C or non-invasive 31 °C) by heating the chamber of
water through the heater plate. As the gas cools considerably down the
unheated circuit, a water trap circuit must be used to collect the resulting
condensate.
Non-Heater Wire Operation
Automatic Humidity Compensation
nice 8050 Respiratory humidifier calculates the power required to
adequately humidify the gas flow through the chamber. If the minimum
power level is not met then the chamber set point will automatically be
increased in 0.5 °C steps until the minimum power is achieved. The
maximum amount of compensation applied is either 3 to 5 °C
depending on the mode (Invasive/Non-invasive)
If humidification is improved and too much power is being applied, then
the humidifier will automatically reduce the chamber set temperature.
Display & Controls
Power Switch
The humidifier will power ON and OFF by the rocker switch is provided in
the right side of the humidifier. After power-on the humidifier starts self test
which checks for possible problems in the humidifier setup. If everything is
working correctly, normal control is initiated.
Mode Key
By press mode key select humid ifier to toggle Invasive & Non-Invasive. The
Mode indicator LED shows the user which mode is selected.
Invasive mode
Invasive mode is for use with patients whose upper airways have been
bypassed by either a tracheostomy or endotracheal tube. In this mode of
operation the humidifier attempts to deliver optimal humidity to the patient
(37 °C, 100 % RH).
Non-Invasive mode
The humidifier normally controls the chamber outlet temperature to 31°C,
and the airway temperature to 34°C, maintaining a +3°C temperature
gradient along
the inspiratory limb.
Standby Mode
If the humidifier detects a problem with its setup or operation it will alarm.
Depending on the severity of the alarm condition, the humidifier will either
remove all power from the heating systems, or enter stand-by. The humidifier
will also enter stand-by if the gas flow through the breathing circuit has
stopped.
Heater wire power is set at 15%.
Control of chamber temperature is attempted, within the following limits:
Heater plate temperature is limited to 50°C.
Mute Key
The mute key silences the humidifier's audible alarm.
Heater Wire
By press heater wire key select humidifier to toggle heater wire ON or OFF
CH/AW Temp Key
By press CH/AW Temp key to shows the Chamber Actual Temperature and
Airway Actual Temperature.
Temperature Display
The front panel display window shows the saturated gas temperature (the
lower of the airway and chamber temperatures in °C) delivered to the patient.
Display & Controls(Cont’d…..)
Indicators
Heater wire Indicator
Heater wire indicator LED glow if the heater wire in the breathing circuit has
not been connected correctly, or if the heater wire or heater wire adaptor is
faulty. An intermittent connection or excessive current (total current in all
limbs > 3.5 A) in the heater wires will also produce this alarm.
Temperature / Flow Probe Indicator
Temperature Probe indicator will glow if the temperature probe is not
correctly plugged in, or the probe used is
faulty. The humidifier tests for the following probe fault conditions:
Temperature probe disconnected
Chamber thermistor open or short circuit
Airway thermistor open or short circuit
Flow thermistor open or short circuit
One thermistor shorted to another
Chamber Probe & Airway Probe Indicator
These indicators are used to show that either the chamber probe or airway
probe is not inserted into the breathing circuit correctly.
Add Water Indicator
If there is insufficient water in the humidification chamber, Add Water LED
will glow
Comparison of nice 8000 series Respiratory
Humidifiers
nice 8050 nice 8010
Invasive Mode Y N
Non Invasive Mode Y Y
Low temp setting N Y
Flow detection Y N
Airway Probe Y N
Heater wire adaptor Y Y
Chamber sensor Y Y
How nice Respiratory Humidifiers fare over
competitors?
nice
Respiratory
Humidifiers
User-friendly:
easy to access all
parameters
Water
Condensation is
minimal
Precise Delivery
Compact and
handy
Fully Automatic
Other company
Respiratory
Humidifiers
Not user friendly:
needs extensive
training to operate
Maximum in other
machines
Not so precise
Bulky
Parameters
should be
adjusted manually
The addition of Air oxygen blender and resuscitator along with humidifier allows
delivery of warmed and humidified flow of air- oxygen mixture to the infant
Air Oxygen Blender
Resuscitator
Humidifier
Cont’d…………….
TROUBLESHOOTING
AND
MAINTENANCE
Cleaning Procedure
Cleaning the Humidifiers
It is recommended that only the following cleaners be used on the nice 8050.
Clean the humidifier with one of the following recommended disinfectants
using a damp cloth:
Isopropyl Alcohol
Soft dishwashing detergent
Note: Follow the manufacturer’s instructions carefully.
Use the correct dilution of the disinfectant.
DO NOT immerse the humidifier in any liquid.
Wipe the humidifier clear of any cleaning residues before use.
Cleaning of Temperature & Flow Probe
1.Physically clean the probes, removing all visible contaminants.
2.Disinfect the probes with one of the following solutions:
Sporicidin; Sporox; Cidex & OPA OR Sterilize the probe using Ethylene Oxide
sterilization at 55 °C (131 °F), 80kPa allow at least 15 hours for residual
ETO to disperse before use.
3. Wipe the airway temperature probes clear of any cleaning residues before use
4. Store in clean conditions.
Preventive Maintenance
Checking the Humidifier
Annually
Check for physical damage:
Check the mains cable for damage, replace if necessary.
Check the heater plate for deep scratching etc., replace if necessary.
Check the heater wire adaptor for kinks, abrasions and damaged connectors.
Check that the plugs couple with the sockets on the humidifier.
Carry out a full performance test.
Checking the Temperature Probe
Every Six months
a. Visually check the humidifier probes for physical damage:
Check that the chamber probe's glass thermistor has not been damaged.
Replace probe if required.
Inspect the chamber probe's glass thermistor for deposits or foreign material.
Clean probe as required.
Check the probe cable for kinks and abrasions etc.
Check that the probe connectors couple with the humidifier sockets.
b. A temperature accuracy check and flow accuracy check should be performed
on the nice 8050 temperature probe.
Cautions
 Use cleaning solution sparingly on a cloth when cleaning
the unit.
 Do not saturate the unit - excessive solution causes
damage to internal components.
 Do not autoclave or gas sterilize the unit.
 Only competent individuals trained in the repair of this
equipment should attempt to service it as detailed in the
service manual. The Service Manual provides detailed
information solely for use by individuals having proper
knowledge, tools and test equipment, and for service
representatives trained by nice Neotech .
 Use of nonstandard components: Consult the
manufacturer for repair and replacement of components.
Use of incorrect component can adversely affect Safety,
performance and/or damage the equipment performance
Certifications and Approvals
Electrical Safety : IEC 60601-1
EMC Safety : IEC 60601-1-2
Particular Standard : ISO 8185
Contact Us
nice Neotech Medical Systems Private Limited
No 85, Krishna Industrial Estate
Mettukuppam, Vannagaram,
Chennai – 600095, Tamil Nadu,
INDIA
Tel: +91-44-24762594,24764608.
Fax: +91-44-24766920
Web: www.niceneotech.com.
E-mail: info@niceneotech.com , marketing@niceneotech.com
Overseas Enquiry: overseas@niceneotech.com
Toll Free : +91-1800-425-2594.

More Related Content

What's hot

3. ITIL Categorization for Remedyforce- An Evolution into Service Desk Self S...
3.	ITIL Categorization for Remedyforce- An Evolution into Service Desk Self S...3.	ITIL Categorization for Remedyforce- An Evolution into Service Desk Self S...
3. ITIL Categorization for Remedyforce- An Evolution into Service Desk Self S...BMC Software
 
ITIL Service Value System.pdf
ITIL Service Value System.pdfITIL Service Value System.pdf
ITIL Service Value System.pdfSSGC
 
Itil service desk business case
Itil service desk business caseItil service desk business case
Itil service desk business caseDavid Lutes
 
IT Service Desk
IT Service DeskIT Service Desk
IT Service DeskSkyCentral
 
mechanical ventilation in children
mechanical ventilation in children mechanical ventilation in children
mechanical ventilation in children mariem ahmed
 
Mechanical ventilation, understanding modes.
Mechanical ventilation, understanding modes.Mechanical ventilation, understanding modes.
Mechanical ventilation, understanding modes.ceswyn
 
Physiology of mechanical ventilation upload
Physiology of mechanical ventilation   uploadPhysiology of mechanical ventilation   upload
Physiology of mechanical ventilation uploadDr Deepa Chandramohan
 
Trouble shooting of mechanical ventilator
Trouble shooting of mechanical ventilatorTrouble shooting of mechanical ventilator
Trouble shooting of mechanical ventilatorAnkit Gajjar
 
Endotracheal intubation &amp; its complications
Endotracheal intubation &amp; its complicationsEndotracheal intubation &amp; its complications
Endotracheal intubation &amp; its complicationsDr Shrikant Phatak
 
NON INVASIVE VENTILATION IN NEONATES-PART 2
NON INVASIVE VENTILATION IN NEONATES-PART 2NON INVASIVE VENTILATION IN NEONATES-PART 2
NON INVASIVE VENTILATION IN NEONATES-PART 2Adhi Arya
 
Anesthetic equipment
Anesthetic equipmentAnesthetic equipment
Anesthetic equipmentSUNY Ulster
 
Mechanical ventilation
Mechanical ventilationMechanical ventilation
Mechanical ventilationBakti Setiadi
 
التنفس الصناعى فى حديثى الولادة د اسامه حسين
التنفس الصناعى فى حديثى الولادة د اسامه حسينالتنفس الصناعى فى حديثى الولادة د اسامه حسين
التنفس الصناعى فى حديثى الولادة د اسامه حسينmohamed osama hussein
 
Fio2 Calculation -Fraction of Inspired Oxygen
Fio2 Calculation -Fraction of Inspired OxygenFio2 Calculation -Fraction of Inspired Oxygen
Fio2 Calculation -Fraction of Inspired OxygenTansika Malar
 
The four phases of intravenous fluid therapy: Manu Malbrain
The four phases of intravenous fluid therapy: Manu MalbrainThe four phases of intravenous fluid therapy: Manu Malbrain
The four phases of intravenous fluid therapy: Manu MalbrainSMACC Conference
 
Physiologically difficult airway
Physiologically difficult airwayPhysiologically difficult airway
Physiologically difficult airwayShreyas Kate
 

What's hot (20)

3. ITIL Categorization for Remedyforce- An Evolution into Service Desk Self S...
3.	ITIL Categorization for Remedyforce- An Evolution into Service Desk Self S...3.	ITIL Categorization for Remedyforce- An Evolution into Service Desk Self S...
3. ITIL Categorization for Remedyforce- An Evolution into Service Desk Self S...
 
ITIL Service Value System.pdf
ITIL Service Value System.pdfITIL Service Value System.pdf
ITIL Service Value System.pdf
 
Itil service desk business case
Itil service desk business caseItil service desk business case
Itil service desk business case
 
ARTERIAL CANNULATION.pptx
ARTERIAL CANNULATION.pptxARTERIAL CANNULATION.pptx
ARTERIAL CANNULATION.pptx
 
IT Service Desk
IT Service DeskIT Service Desk
IT Service Desk
 
mechanical ventilation in children
mechanical ventilation in children mechanical ventilation in children
mechanical ventilation in children
 
Mechanical ventilation, understanding modes.
Mechanical ventilation, understanding modes.Mechanical ventilation, understanding modes.
Mechanical ventilation, understanding modes.
 
Physiology of mechanical ventilation upload
Physiology of mechanical ventilation   uploadPhysiology of mechanical ventilation   upload
Physiology of mechanical ventilation upload
 
Trouble shooting of mechanical ventilator
Trouble shooting of mechanical ventilatorTrouble shooting of mechanical ventilator
Trouble shooting of mechanical ventilator
 
Endotracheal intubation &amp; its complications
Endotracheal intubation &amp; its complicationsEndotracheal intubation &amp; its complications
Endotracheal intubation &amp; its complications
 
NON INVASIVE VENTILATION IN NEONATES-PART 2
NON INVASIVE VENTILATION IN NEONATES-PART 2NON INVASIVE VENTILATION IN NEONATES-PART 2
NON INVASIVE VENTILATION IN NEONATES-PART 2
 
Anesthetic equipment
Anesthetic equipmentAnesthetic equipment
Anesthetic equipment
 
Mechanical ventilation
Mechanical ventilationMechanical ventilation
Mechanical ventilation
 
التنفس الصناعى فى حديثى الولادة د اسامه حسين
التنفس الصناعى فى حديثى الولادة د اسامه حسينالتنفس الصناعى فى حديثى الولادة د اسامه حسين
التنفس الصناعى فى حديثى الولادة د اسامه حسين
 
Fio2 Calculation -Fraction of Inspired Oxygen
Fio2 Calculation -Fraction of Inspired OxygenFio2 Calculation -Fraction of Inspired Oxygen
Fio2 Calculation -Fraction of Inspired Oxygen
 
Global service desk
Global service deskGlobal service desk
Global service desk
 
The four phases of intravenous fluid therapy: Manu Malbrain
The four phases of intravenous fluid therapy: Manu MalbrainThe four phases of intravenous fluid therapy: Manu Malbrain
The four phases of intravenous fluid therapy: Manu Malbrain
 
Physiologically difficult airway
Physiologically difficult airwayPhysiologically difficult airway
Physiologically difficult airway
 
How Ticketing System Helps Businesses Improve Customer Support
How Ticketing System Helps Businesses Improve Customer SupportHow Ticketing System Helps Businesses Improve Customer Support
How Ticketing System Helps Businesses Improve Customer Support
 
Service Support Process PPT
Service Support Process PPTService Support Process PPT
Service Support Process PPT
 

Viewers also liked

Entrevista jp molinero
Entrevista jp molineroEntrevista jp molinero
Entrevista jp molineroDesme Mou
 
Busqueda de una empresa y como rastrear la dirección ip.
Busqueda de una empresa y como rastrear la dirección ip.Busqueda de una empresa y como rastrear la dirección ip.
Busqueda de una empresa y como rastrear la dirección ip.Claudia Rodriguez
 
NIRPOA submission on dealing with the past to Haass
NIRPOA submission on dealing with the past to HaassNIRPOA submission on dealing with the past to Haass
NIRPOA submission on dealing with the past to HaassSlugger Consults
 
ICAWC 2011 Sara Turetta & Jenny Vestlund
ICAWC 2011 Sara Turetta & Jenny VestlundICAWC 2011 Sara Turetta & Jenny Vestlund
ICAWC 2011 Sara Turetta & Jenny VestlundJenny Vestlund
 
Topologías inalámbricas gran tarea
Topologías inalámbricas gran tareaTopologías inalámbricas gran tarea
Topologías inalámbricas gran tareaLarry Ruiz Barcayola
 
MongoDB: Built for Speed
MongoDB: Built for SpeedMongoDB: Built for Speed
MongoDB: Built for SpeedMitch Pirtle
 
Demanda Contencioso Plan General Transitorio Denia Asociacion
Demanda Contencioso Plan General Transitorio Denia AsociacionDemanda Contencioso Plan General Transitorio Denia Asociacion
Demanda Contencioso Plan General Transitorio Denia Asociacionalmelini
 
Mekke yi tanıyalım
Mekke yi tanıyalımMekke yi tanıyalım
Mekke yi tanıyalımguestb8ea82
 
Factura electrónica con SAP
Factura electrónica con SAPFactura electrónica con SAP
Factura electrónica con SAPXavier Terré
 
Reelicon felicia 3 bhk apartments at pashan
Reelicon felicia   3 bhk apartments  at pashanReelicon felicia   3 bhk apartments  at pashan
Reelicon felicia 3 bhk apartments at pashanMilind Jadhav
 
QuimicaDAÑOS A LA SALUD Y EL AMBIENTE CAUSADOS POR EL USO INDISCRIMINADO DE D...
QuimicaDAÑOS A LA SALUD Y EL AMBIENTE CAUSADOS POR EL USO INDISCRIMINADO DE D...QuimicaDAÑOS A LA SALUD Y EL AMBIENTE CAUSADOS POR EL USO INDISCRIMINADO DE D...
QuimicaDAÑOS A LA SALUD Y EL AMBIENTE CAUSADOS POR EL USO INDISCRIMINADO DE D...Yuleni Glez
 
Mi Vida Y Los Videojuegos Video Games Life Time
Mi Vida Y Los Videojuegos   Video Games Life TimeMi Vida Y Los Videojuegos   Video Games Life Time
Mi Vida Y Los Videojuegos Video Games Life TimeOliver Barreto Rodríguez
 
Economia y medicina en la transicion demografica
Economia y medicina en la transicion demograficaEconomia y medicina en la transicion demografica
Economia y medicina en la transicion demograficaConstanza Contreras
 

Viewers also liked (20)

Entrevista jp molinero
Entrevista jp molineroEntrevista jp molinero
Entrevista jp molinero
 
Vuelo de gansos
Vuelo de gansosVuelo de gansos
Vuelo de gansos
 
Nordpeis Taipei
Nordpeis TaipeiNordpeis Taipei
Nordpeis Taipei
 
Busqueda de una empresa y como rastrear la dirección ip.
Busqueda de una empresa y como rastrear la dirección ip.Busqueda de una empresa y como rastrear la dirección ip.
Busqueda de una empresa y como rastrear la dirección ip.
 
NIRPOA submission on dealing with the past to Haass
NIRPOA submission on dealing with the past to HaassNIRPOA submission on dealing with the past to Haass
NIRPOA submission on dealing with the past to Haass
 
ICAWC 2011 Sara Turetta & Jenny Vestlund
ICAWC 2011 Sara Turetta & Jenny VestlundICAWC 2011 Sara Turetta & Jenny Vestlund
ICAWC 2011 Sara Turetta & Jenny Vestlund
 
Topologías inalámbricas gran tarea
Topologías inalámbricas gran tareaTopologías inalámbricas gran tarea
Topologías inalámbricas gran tarea
 
MongoDB: Built for Speed
MongoDB: Built for SpeedMongoDB: Built for Speed
MongoDB: Built for Speed
 
Web 2.0
Web 2.0Web 2.0
Web 2.0
 
Demanda Contencioso Plan General Transitorio Denia Asociacion
Demanda Contencioso Plan General Transitorio Denia AsociacionDemanda Contencioso Plan General Transitorio Denia Asociacion
Demanda Contencioso Plan General Transitorio Denia Asociacion
 
Mekke yi tanıyalım
Mekke yi tanıyalımMekke yi tanıyalım
Mekke yi tanıyalım
 
Factura electrónica con SAP
Factura electrónica con SAPFactura electrónica con SAP
Factura electrónica con SAP
 
Reelicon felicia 3 bhk apartments at pashan
Reelicon felicia   3 bhk apartments  at pashanReelicon felicia   3 bhk apartments  at pashan
Reelicon felicia 3 bhk apartments at pashan
 
RESUME NYOE 2015
RESUME NYOE 2015RESUME NYOE 2015
RESUME NYOE 2015
 
QuimicaDAÑOS A LA SALUD Y EL AMBIENTE CAUSADOS POR EL USO INDISCRIMINADO DE D...
QuimicaDAÑOS A LA SALUD Y EL AMBIENTE CAUSADOS POR EL USO INDISCRIMINADO DE D...QuimicaDAÑOS A LA SALUD Y EL AMBIENTE CAUSADOS POR EL USO INDISCRIMINADO DE D...
QuimicaDAÑOS A LA SALUD Y EL AMBIENTE CAUSADOS POR EL USO INDISCRIMINADO DE D...
 
Mi Vida Y Los Videojuegos Video Games Life Time
Mi Vida Y Los Videojuegos   Video Games Life TimeMi Vida Y Los Videojuegos   Video Games Life Time
Mi Vida Y Los Videojuegos Video Games Life Time
 
Gestor de proyectos G9 Sub-grupoC
Gestor de proyectos G9 Sub-grupoCGestor de proyectos G9 Sub-grupoC
Gestor de proyectos G9 Sub-grupoC
 
Economia y medicina en la transicion demografica
Economia y medicina en la transicion demograficaEconomia y medicina en la transicion demografica
Economia y medicina en la transicion demografica
 
Tartamudez
TartamudezTartamudez
Tartamudez
 
Medio pan y un libro. Lorca
Medio pan y un libro. LorcaMedio pan y un libro. Lorca
Medio pan y un libro. Lorca
 

Similar to 8000 series humidifier

3010 incubators
3010 incubators3010 incubators
3010 incubatorsJerry John
 
3000 transport incubators
3000 transport  incubators3000 transport  incubators
3000 transport incubatorsJerry John
 
5000 series air oxygen blenders
5000 series air oxygen blenders5000 series air oxygen blenders
5000 series air oxygen blendersJerry John
 
Controlling Relative Humidity and Condensation in a CO2 Incubator
Controlling Relative Humidity and Condensation in a CO2 IncubatorControlling Relative Humidity and Condensation in a CO2 Incubator
Controlling Relative Humidity and Condensation in a CO2 Incubatorthebakerco
 
HVAC SYSTEM IN PHARMACEUTICAL INDUSTRY
HVAC SYSTEM IN PHARMACEUTICAL INDUSTRYHVAC SYSTEM IN PHARMACEUTICAL INDUSTRY
HVAC SYSTEM IN PHARMACEUTICAL INDUSTRYNamdeo Shinde
 
2000 series infant radiant warmers
2000 series infant radiant  warmers2000 series infant radiant  warmers
2000 series infant radiant warmersJerry John
 
Anesthesia Vaporizors In Brief.pptx
Anesthesia Vaporizors In Brief.pptxAnesthesia Vaporizors In Brief.pptx
Anesthesia Vaporizors In Brief.pptxMahmood Hasan Taha
 
Air Humidity is crucial!
 Control and monitoring solution for mission-critica...
Air Humidity is crucial!
 Control and monitoring solution for mission-critica...Air Humidity is crucial!
 Control and monitoring solution for mission-critica...
Air Humidity is crucial!
 Control and monitoring solution for mission-critica...CAREL Industries S.p.A
 
2007 series warmers
2007 series warmers2007 series warmers
2007 series warmersJerry John
 
Allied autovent-4000-with-cpap-instruction-manual
Allied autovent-4000-with-cpap-instruction-manualAllied autovent-4000-with-cpap-instruction-manual
Allied autovent-4000-with-cpap-instruction-manualMario O
 

Similar to 8000 series humidifier (20)

3010 incubators
3010 incubators3010 incubators
3010 incubators
 
3000 transport incubators
3000 transport  incubators3000 transport  incubators
3000 transport incubators
 
5000 series air oxygen blenders
5000 series air oxygen blenders5000 series air oxygen blenders
5000 series air oxygen blenders
 
Controlling Relative Humidity and Condensation in a CO2 Incubator
Controlling Relative Humidity and Condensation in a CO2 IncubatorControlling Relative Humidity and Condensation in a CO2 Incubator
Controlling Relative Humidity and Condensation in a CO2 Incubator
 
Environmental Chamber
Environmental Chamber Environmental Chamber
Environmental Chamber
 
Font
FontFont
Font
 
Growth Chamber
Growth Chamber Growth Chamber
Growth Chamber
 
HVAC validation.pdf
HVAC validation.pdfHVAC validation.pdf
HVAC validation.pdf
 
Bubble cpap
Bubble cpapBubble cpap
Bubble cpap
 
Dehumidifier System
Dehumidifier SystemDehumidifier System
Dehumidifier System
 
What is hvac
What is hvacWhat is hvac
What is hvac
 
HVAC SYSTEM IN PHARMACEUTICAL INDUSTRY
HVAC SYSTEM IN PHARMACEUTICAL INDUSTRYHVAC SYSTEM IN PHARMACEUTICAL INDUSTRY
HVAC SYSTEM IN PHARMACEUTICAL INDUSTRY
 
Presentation on HVAC
Presentation on HVAC Presentation on HVAC
Presentation on HVAC
 
DH PRESENTATION
DH PRESENTATIONDH PRESENTATION
DH PRESENTATION
 
DH PRESENTATION
DH PRESENTATIONDH PRESENTATION
DH PRESENTATION
 
2000 series infant radiant warmers
2000 series infant radiant  warmers2000 series infant radiant  warmers
2000 series infant radiant warmers
 
Anesthesia Vaporizors In Brief.pptx
Anesthesia Vaporizors In Brief.pptxAnesthesia Vaporizors In Brief.pptx
Anesthesia Vaporizors In Brief.pptx
 
Air Humidity is crucial!
 Control and monitoring solution for mission-critica...
Air Humidity is crucial!
 Control and monitoring solution for mission-critica...Air Humidity is crucial!
 Control and monitoring solution for mission-critica...
Air Humidity is crucial!
 Control and monitoring solution for mission-critica...
 
2007 series warmers
2007 series warmers2007 series warmers
2007 series warmers
 
Allied autovent-4000-with-cpap-instruction-manual
Allied autovent-4000-with-cpap-instruction-manualAllied autovent-4000-with-cpap-instruction-manual
Allied autovent-4000-with-cpap-instruction-manual
 

More from Jerry John

Phototherapy led&cfl
Phototherapy led&cflPhototherapy led&cfl
Phototherapy led&cflJerry John
 
Nice neotech profile
Nice neotech profileNice neotech profile
Nice neotech profileJerry John
 
Oxygen analyser
Oxygen analyserOxygen analyser
Oxygen analyserJerry John
 
5020 hornpuff resuscitator
5020 hornpuff resuscitator5020 hornpuff resuscitator
5020 hornpuff resuscitatorJerry John
 
2010 series warmers
2010 series warmers2010 series warmers
2010 series warmersJerry John
 
5000 series warmers
5000 series warmers5000 series warmers
5000 series warmersJerry John
 

More from Jerry John (7)

Phototherapy led&cfl
Phototherapy led&cflPhototherapy led&cfl
Phototherapy led&cfl
 
Nice neotech profile
Nice neotech profileNice neotech profile
Nice neotech profile
 
Oxygen analyser
Oxygen analyserOxygen analyser
Oxygen analyser
 
5020 hornpuff resuscitator
5020 hornpuff resuscitator5020 hornpuff resuscitator
5020 hornpuff resuscitator
 
Apnea monitor
Apnea monitorApnea monitor
Apnea monitor
 
2010 series warmers
2010 series warmers2010 series warmers
2010 series warmers
 
5000 series warmers
5000 series warmers5000 series warmers
5000 series warmers
 

Recently uploaded

Pregnancy and Breastfeeding Dental Considerations.pptx
Pregnancy and Breastfeeding Dental Considerations.pptxPregnancy and Breastfeeding Dental Considerations.pptx
Pregnancy and Breastfeeding Dental Considerations.pptxcrosalofton
 
2024 HCAT Healthcare Technology Insights
2024 HCAT Healthcare Technology Insights2024 HCAT Healthcare Technology Insights
2024 HCAT Healthcare Technology InsightsHealth Catalyst
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...narwatsonia7
 
EMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareEMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareRommie Duckworth
 
Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Timedelhimodelshub1
 
Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949ps5894268
 
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...ggsonu500
 
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...High Profile Call Girls Chandigarh Aarushi
 
Single Assessment Framework - What We Know So Far
Single Assessment Framework - What We Know So FarSingle Assessment Framework - What We Know So Far
Single Assessment Framework - What We Know So FarCareLineLive
 
Call Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any TimeCall Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any Timedelhimodelshub1
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed RuleShelby Lewis
 
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...scanFOAM
 
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call GirlsBook Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call GirlsCall Girls Noida
 
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...narwatsonia7
 

Recently uploaded (20)

Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
 
Pregnancy and Breastfeeding Dental Considerations.pptx
Pregnancy and Breastfeeding Dental Considerations.pptxPregnancy and Breastfeeding Dental Considerations.pptx
Pregnancy and Breastfeeding Dental Considerations.pptx
 
2024 HCAT Healthcare Technology Insights
2024 HCAT Healthcare Technology Insights2024 HCAT Healthcare Technology Insights
2024 HCAT Healthcare Technology Insights
 
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service GuwahatiCall Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
 
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
 
EMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareEMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical Care
 
Russian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your bookingRussian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your booking
 
Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Time
 
Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949
 
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
 
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
 
Single Assessment Framework - What We Know So Far
Single Assessment Framework - What We Know So FarSingle Assessment Framework - What We Know So Far
Single Assessment Framework - What We Know So Far
 
Call Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any TimeCall Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any Time
 
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service GuwahatiCall Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
 
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
 
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
 
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call GirlsBook Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
 
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
 

8000 series humidifier

  • 1. nice NeÖtech Medical Systems Pvt. Ltd
  • 2. Presentation Flow COMPANY OVERVIEW CLINICAL ASPECTS OF nice 8000 series HUMIDIFIER PRODUCT OVERVIEW TROUBLESHOOTING
  • 3. Company INFO  Established in the year 1997  Group of highly skilled professionals  4000+ highly satisfied customers  Deep penetration in NICU and respiratory domain  User friendly range of products  With deep skills and high end technology , we always strive for growth and delighted customers
  • 4. Timeline-nice Neotech 2005 •First New Factory Unit 1 Opened •Increase in production Line 2001 •Registered as Pvt ltd •Production started for different models 2013 •Second New Factory Unit 2 Opened 1999 •Neotech Logo Designed •Designed first incubator & open care prototype 2003 •Export to Vietnam 2009- 2011 •CE certification for products 2015 •125+skilled employee strong firm •Multiple range of NICU products 1997 •Started Neotech Medical Systems with 2 employees •First Infant Radiant Warmer
  • 5. Why nice Neotech ?  User friendly range of products  Customized solutions  Ethical Business practices  Stringent quality control management  Complete customer satisfaction
  • 6. Humidity Humidity is the presence of water vapour in gas. Absolute humidity is the amount of water vapour per litre of gas, expressed in mg/l. Relative humidity is a measure of how much water vapour is in a gas compared to its capacity to hold water vapour. It is measured as a percentage. Principle •When dry gas passes over a moist surface evaporation occurs and the gas becomes more humid. •When moist gas cools, water droplets condense and the gas becomes less Humid. •The warmer a gas, the more moisture (absolute humidity) it can carry. •If a gas already has 100% humidity no more moisture will evaporate into it unless it is heated further.
  • 7. What is Respiratory Humidifier? During mechanical ventilation ,When dry air/gas is administered to the patient there occurs airway cooling, airway drying ,inflammation which can lead to many complications. Optimum humidity is required for ventilation - 33mg/L in non invasive mode and 44 mg/L in invasive mode. Exposure to humidity levels below 25mg H2O /l or 100% saturated gas can lead to respiratory complications. Less humidity can cause airway mucosal dysfunction and more humidity can cause reduced mucus viscosity and increased pericellular depth fluid. To avoid the above said conditions and to administer optimal humidity. respiratory humidifiers are used during ventilation. Respiratory Humidifiers heats the water contained in the humidification chamber, humidifying the air passing through it.
  • 8. Why Humidifiers are Used? Prevent Airway Cooling Prevent Airway Drying and Water Loss Prevent Airway Inflammation and Constriction Ensure patient is stabilized Infant Humidifiers Less Humidity = More Evaporation
  • 9. Introducing nice NeÖtech 8000 series Respiratory Humidifiers Designed to perfection
  • 10. nice 8000 series Respiratory Humidifiers
  • 11. •Invasive or non invasive control mode with flow detection. •Flexibility in Operation •Invasive ventilation, Infant Bubble CPAP, Resuscitation, and Oxygen therapy for Neonatal Care •for invasive ventilation, non invasive ventilation and oxygen therapy for Adult Care •Dual servo control feedback system •Comprehensive audible & visual alarm •Water alarm if water alarm falls below required level •High and low flow ventilation therapies •Heated and non heated wire breathing system •Precise, reliable performance •Compact and Handy •Mounting Options Features
  • 12. How to operate? Control Panel Operation Alarms Heater Wire Operation Non-Heater Wire Operation  Automatic Humidity Compensation Display & Controls Indicators
  • 14. Temperature Probe : if the temperature probe is damaged or disconnected Heater Wire : if the heater wire is defected or disconnected Heater : if the heater is defected or disconnected Maximum System Operating Pressure : 20 kPa, Gas leakage at max. Pressure : <100 mL/minute. Alarms
  • 15. •Humidified gas from the chamber pass through the inspiratory limb, where its temperature must be maintained in order to prevent the generated humidity from condensing. •This is achieved with a heater wire encapsulated within the inspiratory limb. The humidifier maintains the temperature along the inspiratory limb by monitoring the temperature at the airway probe and controlling the voltage delivered to the heater wire. • Under normal conditions the gas is heated to 40 °C in the invasive mode, 34 °C for the non-invasive mode. Heater Wire Operation
  • 16. Respiratory Humidifier maintains the airway temperature at the desired set point (invasive 37 °C or non-invasive 31 °C) by heating the chamber of water through the heater plate. As the gas cools considerably down the unheated circuit, a water trap circuit must be used to collect the resulting condensate. Non-Heater Wire Operation
  • 17. Automatic Humidity Compensation nice 8050 Respiratory humidifier calculates the power required to adequately humidify the gas flow through the chamber. If the minimum power level is not met then the chamber set point will automatically be increased in 0.5 °C steps until the minimum power is achieved. The maximum amount of compensation applied is either 3 to 5 °C depending on the mode (Invasive/Non-invasive) If humidification is improved and too much power is being applied, then the humidifier will automatically reduce the chamber set temperature.
  • 18. Display & Controls Power Switch The humidifier will power ON and OFF by the rocker switch is provided in the right side of the humidifier. After power-on the humidifier starts self test which checks for possible problems in the humidifier setup. If everything is working correctly, normal control is initiated. Mode Key By press mode key select humid ifier to toggle Invasive & Non-Invasive. The Mode indicator LED shows the user which mode is selected. Invasive mode Invasive mode is for use with patients whose upper airways have been bypassed by either a tracheostomy or endotracheal tube. In this mode of operation the humidifier attempts to deliver optimal humidity to the patient (37 °C, 100 % RH). Non-Invasive mode The humidifier normally controls the chamber outlet temperature to 31°C, and the airway temperature to 34°C, maintaining a +3°C temperature gradient along the inspiratory limb.
  • 19. Standby Mode If the humidifier detects a problem with its setup or operation it will alarm. Depending on the severity of the alarm condition, the humidifier will either remove all power from the heating systems, or enter stand-by. The humidifier will also enter stand-by if the gas flow through the breathing circuit has stopped. Heater wire power is set at 15%. Control of chamber temperature is attempted, within the following limits: Heater plate temperature is limited to 50°C. Mute Key The mute key silences the humidifier's audible alarm. Heater Wire By press heater wire key select humidifier to toggle heater wire ON or OFF CH/AW Temp Key By press CH/AW Temp key to shows the Chamber Actual Temperature and Airway Actual Temperature. Temperature Display The front panel display window shows the saturated gas temperature (the lower of the airway and chamber temperatures in °C) delivered to the patient. Display & Controls(Cont’d…..)
  • 20. Indicators Heater wire Indicator Heater wire indicator LED glow if the heater wire in the breathing circuit has not been connected correctly, or if the heater wire or heater wire adaptor is faulty. An intermittent connection or excessive current (total current in all limbs > 3.5 A) in the heater wires will also produce this alarm. Temperature / Flow Probe Indicator Temperature Probe indicator will glow if the temperature probe is not correctly plugged in, or the probe used is faulty. The humidifier tests for the following probe fault conditions: Temperature probe disconnected Chamber thermistor open or short circuit Airway thermistor open or short circuit Flow thermistor open or short circuit One thermistor shorted to another Chamber Probe & Airway Probe Indicator These indicators are used to show that either the chamber probe or airway probe is not inserted into the breathing circuit correctly. Add Water Indicator If there is insufficient water in the humidification chamber, Add Water LED will glow
  • 21. Comparison of nice 8000 series Respiratory Humidifiers nice 8050 nice 8010 Invasive Mode Y N Non Invasive Mode Y Y Low temp setting N Y Flow detection Y N Airway Probe Y N Heater wire adaptor Y Y Chamber sensor Y Y
  • 22. How nice Respiratory Humidifiers fare over competitors? nice Respiratory Humidifiers User-friendly: easy to access all parameters Water Condensation is minimal Precise Delivery Compact and handy Fully Automatic Other company Respiratory Humidifiers Not user friendly: needs extensive training to operate Maximum in other machines Not so precise Bulky Parameters should be adjusted manually
  • 23. The addition of Air oxygen blender and resuscitator along with humidifier allows delivery of warmed and humidified flow of air- oxygen mixture to the infant Air Oxygen Blender Resuscitator Humidifier Cont’d…………….
  • 25. Cleaning Procedure Cleaning the Humidifiers It is recommended that only the following cleaners be used on the nice 8050. Clean the humidifier with one of the following recommended disinfectants using a damp cloth: Isopropyl Alcohol Soft dishwashing detergent Note: Follow the manufacturer’s instructions carefully. Use the correct dilution of the disinfectant. DO NOT immerse the humidifier in any liquid. Wipe the humidifier clear of any cleaning residues before use. Cleaning of Temperature & Flow Probe 1.Physically clean the probes, removing all visible contaminants. 2.Disinfect the probes with one of the following solutions: Sporicidin; Sporox; Cidex & OPA OR Sterilize the probe using Ethylene Oxide sterilization at 55 °C (131 °F), 80kPa allow at least 15 hours for residual ETO to disperse before use. 3. Wipe the airway temperature probes clear of any cleaning residues before use 4. Store in clean conditions.
  • 26. Preventive Maintenance Checking the Humidifier Annually Check for physical damage: Check the mains cable for damage, replace if necessary. Check the heater plate for deep scratching etc., replace if necessary. Check the heater wire adaptor for kinks, abrasions and damaged connectors. Check that the plugs couple with the sockets on the humidifier. Carry out a full performance test. Checking the Temperature Probe Every Six months a. Visually check the humidifier probes for physical damage: Check that the chamber probe's glass thermistor has not been damaged. Replace probe if required. Inspect the chamber probe's glass thermistor for deposits or foreign material. Clean probe as required. Check the probe cable for kinks and abrasions etc. Check that the probe connectors couple with the humidifier sockets. b. A temperature accuracy check and flow accuracy check should be performed on the nice 8050 temperature probe.
  • 27. Cautions  Use cleaning solution sparingly on a cloth when cleaning the unit.  Do not saturate the unit - excessive solution causes damage to internal components.  Do not autoclave or gas sterilize the unit.  Only competent individuals trained in the repair of this equipment should attempt to service it as detailed in the service manual. The Service Manual provides detailed information solely for use by individuals having proper knowledge, tools and test equipment, and for service representatives trained by nice Neotech .  Use of nonstandard components: Consult the manufacturer for repair and replacement of components. Use of incorrect component can adversely affect Safety, performance and/or damage the equipment performance
  • 28. Certifications and Approvals Electrical Safety : IEC 60601-1 EMC Safety : IEC 60601-1-2 Particular Standard : ISO 8185
  • 29. Contact Us nice Neotech Medical Systems Private Limited No 85, Krishna Industrial Estate Mettukuppam, Vannagaram, Chennai – 600095, Tamil Nadu, INDIA Tel: +91-44-24762594,24764608. Fax: +91-44-24766920 Web: www.niceneotech.com. E-mail: info@niceneotech.com , marketing@niceneotech.com Overseas Enquiry: overseas@niceneotech.com Toll Free : +91-1800-425-2594.