SlideShare a Scribd company logo
1 of 47
equipment AND QUALITY CONTROL
BY MUMBA CHILIMBOYI
EQUIPMENT USED IN NUCLEAR MEDICINE
AND QUALITY CONTROL
IONIZATION CHAMBER
DOSE CALIBRATOR
GAMMA SCINTILLATION CAMERA
COMPUTERS
SINGLE-PHOTON EMISSION COMPUTED TOMOGRAPHY (SPECT)
RADIATION MONITOR GEIGER-MUELLER COUNTER
DETECTORS
IONIZATION CHAMBER
• Ionization chambers are handheld survey instruments used to measure low or
high exposure rates.
• They have an air or gas-filled chamber but a low efficiency for detection of
gamma rays.
• These instruments have a relatively low applied voltage from anode to cathode;
as a result, there is no avalanche effect and no dead time problem.
• Ionization chambers typically are useful at exposure rates ranging from 0.1 mR
(2.5 x 10-8 C/kg)/hour to 100 R (2.5 x 10-2 C/kg)/hour.
• A dose calibrator is a special form of an ionization chamber.
DOSE CALIBRATOR AND GAMMA CAMERA
• DEFINITION
• COMPONENTS
• PRINCIPLES OF OPERATION
• QUALITY CONTROL
DOSE CALIBRATOR
• The dose calibrator is an ionization chamber used to assay the
amount of activity in vials and syringes
• This includes the assay of individual doses before
administration to patients, as required by regulation.
• The dose calibrator operates over a very wide range of
activities, from hundreds of kilobecquerels (10s of μCi) to tens
of gigabecquerels (up to a curie).
• Displayed units are mCI millicurie or Megabecquerel
DOSE CALIBRATOR COMPONENTS1
3
2
OPERATION OF DOSE CALIBRATOR
• The chamber is cylindrical and holds a
defined volume of pressurized inert gas
(usually argon).
• Within the chamber is a collecting
electrode. As radiation emanates from the
radiopharmaceutical in the syringe, it enters
the chamber and interacts with the gas,
causing ionization.
• If no voltage is applied to the electrodes, the
ion pairs recombine.
• If an electrical differential applied between
the chamber and the collecting electrode is
applied it causes the ions to be captured and
measured. This measurement is used to
calculate the dose contained in the syringe.
QUALITY CONTROL FOR DOSE CALIBRATORS
• The dose calibrator is used to assay the activity administered
to the patient, and thus a comprehensive quality control
program is necessary.
• This program comprises four basic quality control tests:
-Geometry
-Accuracy
-Linearity
-Constancy
THE GEOMETRY PROTOCOL TESTS
• This test of the dose calibrator provides the same reading for the same
amount of activity irrespective of the volume or orientation of the
sample.
• A reading of a certain amount of activity in a 0.5-mL volume is
obtained. The volume is then increased by augmenting the sample with
amounts of nonradioactive water or saline and taking additional
readings. The subsequent readings should not vary from the original
readings by more than 10%.
• The geometry test is performed during acceptance testing and after a
major repair or movement of the equipment to another location.
ACCURACY
• For accuracy, calibrated sources (typically cobalt-57
and 137Cs) are analysed.
• The resultant reading cannot vary by more than 10%
from the calibrated activity decay corrected to the day
of the test.
• The accuracy test should be performed during
acceptance testing, annually thereafter, and after a
major repair or move.
LINEARITY PROTOCOL TESTS
•This test of the dose calibrator operates appropriately over
the wide activity range to which it is applied. The device
is tested from 10 μCi (370 kBq) to a level higher than that
routinely used in the clinic and perhaps as high as 1 Ci
(37 GBq).
•The activity readings are varied by starting with a sample
of radioactivity of Tc-99m at the highest value to be tested
(e.g., tens of gigabequerels).
LINEARITY PROTOCOL TESTS Cont.…
• The activity readings are then varied by either allowing the
source to radioactively decay over several days or using a set
of lead shields of varying thicknesses until a reading close to
370 kBq is obtained. Each reading should not vary by more
than 10% from the line drawn through the calculated activity
values.
• The linearity test should be performed during acceptance
testing, quarterly thereafter, and after a major repair or move.
THE CONSTANCY PROTOCOL TESTS
• This tests is the reproducibility of the readings as compared to
a decay-corrected estimate for a reference reading obtained
from the dose calibrator on a particular day.
• Today’s constancy reading cannot vary from the decay-
corrected reference reading by more than 10%.
• The constancy test varies from accuracy in that it evaluates the
precision of the readings from day to day rather than accuracy.
• The constancy test should be performed on every day that the
device is used to assay a dose to be administered to a patient.
DOSE CALIBRATOR RADIATION PROTECTION
• Lead shielding around the ionization chamber
1. Protects the operator
2. Reduces the response background radiation
• Sample holder can be cleaned in the event of
radioactive contamination of the chamber well
GAMMA CAMERA
The most widely used cameras in nuclear medicine are:
1. Simple gamma scintillation (Anger) camera
2. Single-photon emission computed tomography (SPECT) capable
gamma camera.
FUNCTION OF GAMMA CAMERA
• A gamma camera converts photons emitted by the radionuclide in the
patient into a light pulse and subsequently into a voltage signal.
GAMMA CAMERA SYSTEM
COMPONENTS
• The collimator
• The scintillation crystal
• An array of photomultiplier tubes (PMTs)
• Preamplifiers
• A pulse height analyzer (PHA)
• Digital correction circuitry,
• A cathode ray tube (CRT)
• The control console.
• A computer and
• Picture archiving systems(PACs)
GAMMA CAMERAASSEMBLY
COLLIMATORS
• The collimator is made of perforated or folded lead and is
interposed between the patient and the scintillation crystal. It
allows the gamma camera to localize accurately the
radionuclide in the patient’s body.
• Collimators perform this function by absorbing and stopping
most radiation except that arriving almost perpendicular to the
detector face.
• Most radiation striking the collimator at oblique angles is not
included in the final image.
TYPES COLLIMATORS
As the energy of the radionuclide increases, the
best collimator usually has thicker and longer
septa. For a given septal thickness, spatial
resolution of a collimator increases with septal
length but sensitivity decreases.
The two basic types of collimators are pinhole and multihole
SEPTAL PENETRATION AND PHOTON SCATTERING
EFFECT OF SEPTAL LENGTH ON COLLIMATOR SENSITIVITY
AND RESOLUTION
EFFECT OF DIFFERENT SOURCE-TO-CAMERA DISTANCES
CRYSTAL AND OTHER PHOTON DETECTOR DEVICES
• Radiation emerging from the patient and passing through the
collimator interacts with a thallium activated sodium iodide
crystal.
• Crystals are made with thallium or sodium activated cesium
iodide or even lanthanum bromide are used. They convert
Gamma rays to light.
• PMTs situated along the posterior crystal face detect this light
and amplify it to an electrical pulse.
PHOTONMULTIPLIER TUBEs (PMTs)
• A photomultiplier tube (PMT) converts a light pulse into an electrical
signal of measurable magnitude.
• Localization of the event in the final image depends on the amount of
light sensed by each PMT and thus on the pattern of PMT voltage
output.
• VOLTAGE AMPLIFIER / VOLTAGE SUPPLY
 A High voltage supply for the PMT
 An amplifier increases the size of the pulse.
PULSE HEIGHT ANALYZER
• The basic principle of the PHA is to discard signals from background
and scattered radiation and/or radiation from interfering isotopes so
that only primary photons known to come from the photopeak of the
isotope being imaged are recorded
• The PHA discriminates between events occurring in the crystal that
will be displayed or stored in the computer and events that will be
rejected.
CONSOLE CONTROLS
• Image exposure time is selected by console control
• Its usually a preset count, a preset time, or preset information
density for the image accumulation.
Other console controls are present for orientation and allow
the image to be reversed on the x- and y-axes.
In addition, the CRT image may be manipulated by an
intensity control
Hard copy images on film may be obtained directly from the
computer, display digital images on monitors and store the
images in a picture archiving system.
OPERATION OF GAMMA SYSTEM
 Gamma rays emitted from within the patient pass
through the holes of an absorptive collimator to
reach the NaI crystal.
 On interaction of the gamma ray with the NaI
scintillating crystal, thousands of light photons are
emitted, a portion of which are collected by an
array of PMTs. By taking weighted sums of the
PMT signals within the associated computer
 The two-dimensional (2D) x and y location and
the total energy of the detection event deposited is
estimated. If the energy deposited is within a
prespecified energy window (e.g., within 10% of
the photopeak energy), the event is accepted and
the location of the event recorded.
 In this manner, the gamma camera image is
constructed on an event-by-event basis, and a
single nuclear medicine image may consist of
hundreds of thousands of such events.
GAMMA CAMERA QUALITY CONTROL
•This involves acceptance testing of the device before its
initial use and a program of routine tests and evaluations
applied on a regular basis.
•It is essential that the performance be evaluated regularly to
ensure that the images adequately demonstrate the in vivo
distribution of the administered radiopharmaceutical and
that any quantitation performed with the camera yields
values that are accurate and precise
GAMMA CAMERA QUALITY CONTROL SUMMARY
PARAMETER COMMENT
DAILY
• UNIFORMITY
• WINDOW SETTING
• FLOOD FIELD; INTRINSIC (WITHOUT COLLIMATOR) OR EXTRINSIC (WITH COLLIMATOR)
• CONFIRM ENERGY WINDOW SETTING RELATIVE TO PHOTOPEAK FOR EACH
RADIONUCLIDE USED WITH EACH PATIENT
WEEKLY OR MONTHLY
• SPATIAL RESOLUTION
• LINEARITY CHECK
• REQUIRES A “RESOLUTION” PHANTOM SUCH AS THE FOUR-QUADRANT BAR
• QUALITATIVE ASSESSMENT OF BAR PATTERN LINEARITY
The purpose of quality control is to detect changes in the performance of a gamma camera and below summary:
GAMMA CAMERA QUALITY CONTROL SUMMARY Cont....
PARAMETER COMMENT
ANNUALLY
• SYSTEM UNIFORMITY
• MULTI-WINDOW REGISTRATION
• COUNT RATE PERFORMANCE
• ENERGY RESOLUTION
• SYSTEM SENSITIVITY
• HIGH COUNT FLOOD WITH EACH COLLIMATOR
• FOR CAMERAS WITH CAPABILITY OF IMAGING MULTIPLE ENERGY
WINDOWS SIMULTANEOUSLY
• VARY COUNTS USING DECAY OR ABSORBER METHOD
• EASIEST IN CAMERAS WITH BUILT IN MULTICHANNEL ANALYZERS
• COUNT RATE PERFORMANCE PER UNIT OF ACTIVITY FOR EACH
COLLIMATOR
EFFECT OF INCREASING THE PATIENT-TO-DETECTOR FACE DISTANCE
ON CLINICAL IMAGES
SINGLE PHOTON EMISSION COMPUTED
TOMOGRAPHY (SPECT)
•SPECT is a nuclear medicine tomographic
imaging technique that uses gamma rays.
•It is very similar to conventional nuclear
medicine planar imaging using a gamma camera.
•It is able to provide true 3D information..
SPECT
SPECT CONT.…
•SPECT imaging is performed by using a gamma
camera to acquire multiple 2-D images from multiple
angles.
• A computer is then used to apply a tomographic
reconstruction algorithm to the multiple projections,
yielding a 3-D dataset.
•The dataset may then be manipulated to show thin
slices along any chosen axis of the body.
IMAGE RECONSTRUCTION
• SPECT acquire raw data in the form of projection data at a
variety of angles about the patient.
• Image reconstruction involves the processing of these data to
generate a series of cross-sectional images through the object
of interest.
PROJECTION DATA
APPLICATION OF SPECT
• Cerebral Blood flow imaging
• Myocardial perfusion imaging with thalium-201 or technetium-99
perfusion agents
• Imaging of tumors or infections with agents such as gallium-67 or 111
WBCs
• Certain Cases of Bone Imaging
• Brain studies
• Liver/ Spleen Imaging
• Renal Imaging
FOURIER TRANSFORM
• Image data may be best represented in either spatial (real) or frequency
space.
• The mathematician Joseph Fourier noted in 1807 that any arbitrary
signal can be generated by adding a large number of sine and cosine
signals of varying frequencies and amplitudes.
• The plot of amplitude as a function of frequency is referred to as the
Fourier transform, and it defines the components of the image at each
frequency.
• The low frequencies provide the overall shape of the object, whereas the
high frequencies help define the sharp edges and fine detail within the
image.
APPLICATIONS OF FOURIER TRANSFORM
• The Fourier method is often used on images from astronomy,
microbiology, images of repetitive structures such as crystals and so on.
• Fourier transform is good for identifying a periodic component or lattice
in an image.
• Identifying regular patterns on an image has other advantages like
removing regular dirty spots or noise from an image
• Components of higher frequency can be removed to achieve anti-aliasing
effect, i.e. removing ugly zaggy edges.
APPLICATIONS OF FOURIER TRANSFORM Cont.…..
• There are other techniques associated with Fourier
transform:
convolution theory
correlation, sampling
Reconstruction
image compression, and more.
RADIATION MONITOR-GEIGER-MUELLER
COUNTER
• Geiger-Mueller (GM) counters are handheld,
very sensitive, inexpensive survey instruments
used primarily to detect small amounts of
radioactive contamination.
• The detector is usually pancake shaped, it may
also be cylindrical.
• The detector is gas-filled and has a high applied
voltage from the anode to the cathode. This
causes one ionization to result in an “avalanche”
of other electrons, allowing high efficiency for
detection of even a single gamma ray.
TYPES OF RADIATION DETECTORS
Basically three types of radiation detectors are
used in nuclear medicine
Gas detectors
Scintillators
Semiconductors
GAS DETECTORS
• A gas radiation detector is filled with a volume of gas that acts as the sensitive
material of the detector.
• In some cases, it is air and in others it is an inert gas such as argon or xenon,
depending on the particular detector.
• Electrodes are located at either end of the sensitive volume.
• The detector circuit also contains a variable voltage supply and a current detector.
• As radiation passes through the sensitive volume, it causes ionization in the gas.
• If a voltage is applied across the volume, the resulting ions (electrons and positive
ions) will start to drift, causing a measureable current in the circuit. The current will
last until all of the charge that was liberated in the event is collected at the
electrodes.
• The resulting current entity is referred to as a pulse and is associated with a
particular detection event.
SCINTILLATION AND SEMICONDUCTOR DETECTORS
• Some crystalline materials emit a large number of light
photons upon the absorption of ionizing radiation.This
process is referred to as scintillation, and materials are
referred to as scintillators.
• As radiation interacts within the scintillator, a large number
of excitations and ionizations occur. On deexcitation, the
number of light photons emitted is directly proportional to the
amount of energy deposited within the scintillator.
SCINTILLATION AND SEMICONDUCTOR DETECTORS Cont.…
• Thermal energy can lead to a measureable current in some
semiconductor detectors such as GeLi, even in the absence of
radiation, and thus these semiconductor detectors must be operated at
cryogenic temperatures.
• On the other hand, semiconductor detectors such as cadmium telluride
(CdTe) or cadmium zinc telluride (CZT) can operate at room
temperature. CdTe and CZT do not have the excellent energy
resolution of GeLi, but at approximately 5%, it is still significantly
better than that of sodium iodide.
CONCLUSION
• Radiation detection and counting is the corner stone of nuclear
medicine.
• Detectors of all types—gas detectors, scintillators, and
semiconductors—are used every day in the nuclear medicine clinic.
• Some are used for ancillary purposes that support the clinic, such as
those used in the context of radiation protection.
• Others are used to specifically acquire biological data for a particular
clinical purpose. Most notably, the gamma camera is used to obtain
images of the in vivo distribution of the administered
radiopharmaceutical from which the patient’s physiology or function
can be inferred to further define the patient’s medical picture.
Conclusion Cont….
• A rigorous quality control program must be maintained for all equipment
used in the nuclear medicine clinic to ensure the integrity of the data
obtained from the patient. The quality control program for the gamma
camera includes acceptance testing and tests that need to be performed on a
routine basis. The nuclear medicine image acquired with the gamma camera
provides a snapshot of the patient’s in vivo radiopharmaceutical distribution
from a certain view and at a particular point in time. These images also can
be acquired as a dynamic (time-sequence) study or in conjunction with a
physiological gate such as the ECG.
• Regions of interest can be drawn about specific features to provide regional
quantitation or TACs of dynamic processes.
• Finally, nuclear medicine instrumentation continues to evolve, including the
development of devices designed for a specific clinical task such as breast
imaging. It is expected that this development will continue in the years
ahead.
REFERENCES
• Essentials of nuclear medicine imaging; Fred A.Mettler 2012
• The Requisites Ziessman - Nuclear Medicine, 4th ed. FRED A. METTLER
• Camera Systems. Publication 1141. Vienna, Austria: International Atomic Energy
Agency; 2003.
• IAEA Comprehensive Clinical Audits Of Diagnostic Radiology Practices A Tool For
Quality Improvement
• Nuclear Medicine Instrumentation Lecture Notes:2005; Ghoorun S
• Operational Levels In Radiopharmacy For Realigning Our Profession With International
Guidelines Richard E. 2015
• Chandra R. Nuclear Medicine Physics: The Basics. 7th ed. Philadelphia: Williams &
Wilkins; 2011.
THANK YOU VERY MUCH AND GOD
BLESS YOU

More Related Content

What's hot

Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyRamzee Small
 
Radiation emergencies and preparedness in radiotherapy
Radiation emergencies and preparedness in radiotherapyRadiation emergencies and preparedness in radiotherapy
Radiation emergencies and preparedness in radiotherapyDeepjyoti saha
 
Nuclear medicine 2
Nuclear medicine 2Nuclear medicine 2
Nuclear medicine 2Rad Tech
 
Radiation protection
Radiation protection   Radiation protection
Radiation protection Anil Gupta
 
Difference Between Single Slice and Multi Slice CT Scanner
Difference Between Single Slice and Multi Slice CT ScannerDifference Between Single Slice and Multi Slice CT Scanner
Difference Between Single Slice and Multi Slice CT ScannerAbdul Rehman (R)(CT)(MR)
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das AkashDas169
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose IndexAnjan Dangal
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectorssandip suman
 
Basic principle of ct and ct generations
Basic principle of ct and ct generationsBasic principle of ct and ct generations
Basic principle of ct and ct generationsTarun Goyal
 
Nuclear Medicine - PET/CT
Nuclear Medicine - PET/CTNuclear Medicine - PET/CT
Nuclear Medicine - PET/CT@Saudi_nmc
 

What's hot (20)

Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed Tomography
 
Nuclear imaging
Nuclear imagingNuclear imaging
Nuclear imaging
 
Radiation emergencies and preparedness in radiotherapy
Radiation emergencies and preparedness in radiotherapyRadiation emergencies and preparedness in radiotherapy
Radiation emergencies and preparedness in radiotherapy
 
Nuclear medicine 2
Nuclear medicine 2Nuclear medicine 2
Nuclear medicine 2
 
Radiation protection
Radiation protection   Radiation protection
Radiation protection
 
Radiation protection in nuclear medicine shafiee
Radiation protection in nuclear medicine shafieeRadiation protection in nuclear medicine shafiee
Radiation protection in nuclear medicine shafiee
 
Mri contrast
Mri contrastMri contrast
Mri contrast
 
Difference Between Single Slice and Multi Slice CT Scanner
Difference Between Single Slice and Multi Slice CT ScannerDifference Between Single Slice and Multi Slice CT Scanner
Difference Between Single Slice and Multi Slice CT Scanner
 
Nuclear Medicine - Rivin
Nuclear Medicine - RivinNuclear Medicine - Rivin
Nuclear Medicine - Rivin
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose Index
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Nuclear medicine scan.ppt
Nuclear medicine scan.pptNuclear medicine scan.ppt
Nuclear medicine scan.ppt
 
Epid
EpidEpid
Epid
 
Basic principles of CT scanning
Basic principles of CT scanningBasic principles of CT scanning
Basic principles of CT scanning
 
Basic principle of ct and ct generations
Basic principle of ct and ct generationsBasic principle of ct and ct generations
Basic principle of ct and ct generations
 
Nuclear Medicine - PET/CT
Nuclear Medicine - PET/CTNuclear Medicine - PET/CT
Nuclear Medicine - PET/CT
 
Automatic exposure control
Automatic exposure controlAutomatic exposure control
Automatic exposure control
 
Linear Accelerator
Linear AcceleratorLinear Accelerator
Linear Accelerator
 
Ct Generations
Ct  GenerationsCt  Generations
Ct Generations
 

Similar to Nuclear medicine equipment quality control and calibration

Gamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfGamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfNehaMishra787239
 
Gamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfGamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfNehaMishra787239
 
Gamma Camera.pptx
Gamma Camera.pptxGamma Camera.pptx
Gamma Camera.pptxMISSCOM1
 
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxAkashSharma640477
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfAngelina Protik
 
QC Gamma Camera
QC Gamma CameraQC Gamma Camera
QC Gamma CameraHelbert
 
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxZubairUlIslam5
 
Task group report 135 cyberknie
Task group report 135 cyberknie Task group report 135 cyberknie
Task group report 135 cyberknie Slidevikram
 
Neutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopitalNeutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopitalmedicalHunter2
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.pptBBAdhikari
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCAnupriyaNR
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptJgdishrathi
 
25 -radiation_detection_&_measurement_i
25  -radiation_detection_&_measurement_i25  -radiation_detection_&_measurement_i
25 -radiation_detection_&_measurement_imurty61
 
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptxGAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptxTaushifulHoque
 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapyfondas vakalis
 

Similar to Nuclear medicine equipment quality control and calibration (20)

Gamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfGamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdf
 
Gamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfGamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdf
 
Gamma Camera.pptx
Gamma Camera.pptxGamma Camera.pptx
Gamma Camera.pptx
 
QUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptxQUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptx
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 
QC Gamma Camera
QC Gamma CameraQC Gamma Camera
QC Gamma Camera
 
Qc
QcQc
Qc
 
Radioactivity
RadioactivityRadioactivity
Radioactivity
 
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
 
Task group report 135 cyberknie
Task group report 135 cyberknie Task group report 135 cyberknie
Task group report 135 cyberknie
 
Neutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopitalNeutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopital
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.ppt
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.ppt
 
25 -radiation_detection_&_measurement_i
25  -radiation_detection_&_measurement_i25  -radiation_detection_&_measurement_i
25 -radiation_detection_&_measurement_i
 
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptxGAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapy
 
linac QA.pptx
linac QA.pptxlinac QA.pptx
linac QA.pptx
 

Recently uploaded

call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowNehru place Escorts
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformKweku Zurek
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 

Recently uploaded (20)

call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy Platform
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 

Nuclear medicine equipment quality control and calibration

  • 1. equipment AND QUALITY CONTROL BY MUMBA CHILIMBOYI
  • 2. EQUIPMENT USED IN NUCLEAR MEDICINE AND QUALITY CONTROL IONIZATION CHAMBER DOSE CALIBRATOR GAMMA SCINTILLATION CAMERA COMPUTERS SINGLE-PHOTON EMISSION COMPUTED TOMOGRAPHY (SPECT) RADIATION MONITOR GEIGER-MUELLER COUNTER DETECTORS
  • 3. IONIZATION CHAMBER • Ionization chambers are handheld survey instruments used to measure low or high exposure rates. • They have an air or gas-filled chamber but a low efficiency for detection of gamma rays. • These instruments have a relatively low applied voltage from anode to cathode; as a result, there is no avalanche effect and no dead time problem. • Ionization chambers typically are useful at exposure rates ranging from 0.1 mR (2.5 x 10-8 C/kg)/hour to 100 R (2.5 x 10-2 C/kg)/hour. • A dose calibrator is a special form of an ionization chamber.
  • 4. DOSE CALIBRATOR AND GAMMA CAMERA • DEFINITION • COMPONENTS • PRINCIPLES OF OPERATION • QUALITY CONTROL
  • 5. DOSE CALIBRATOR • The dose calibrator is an ionization chamber used to assay the amount of activity in vials and syringes • This includes the assay of individual doses before administration to patients, as required by regulation. • The dose calibrator operates over a very wide range of activities, from hundreds of kilobecquerels (10s of μCi) to tens of gigabecquerels (up to a curie). • Displayed units are mCI millicurie or Megabecquerel
  • 7. OPERATION OF DOSE CALIBRATOR • The chamber is cylindrical and holds a defined volume of pressurized inert gas (usually argon). • Within the chamber is a collecting electrode. As radiation emanates from the radiopharmaceutical in the syringe, it enters the chamber and interacts with the gas, causing ionization. • If no voltage is applied to the electrodes, the ion pairs recombine. • If an electrical differential applied between the chamber and the collecting electrode is applied it causes the ions to be captured and measured. This measurement is used to calculate the dose contained in the syringe.
  • 8. QUALITY CONTROL FOR DOSE CALIBRATORS • The dose calibrator is used to assay the activity administered to the patient, and thus a comprehensive quality control program is necessary. • This program comprises four basic quality control tests: -Geometry -Accuracy -Linearity -Constancy
  • 9. THE GEOMETRY PROTOCOL TESTS • This test of the dose calibrator provides the same reading for the same amount of activity irrespective of the volume or orientation of the sample. • A reading of a certain amount of activity in a 0.5-mL volume is obtained. The volume is then increased by augmenting the sample with amounts of nonradioactive water or saline and taking additional readings. The subsequent readings should not vary from the original readings by more than 10%. • The geometry test is performed during acceptance testing and after a major repair or movement of the equipment to another location.
  • 10. ACCURACY • For accuracy, calibrated sources (typically cobalt-57 and 137Cs) are analysed. • The resultant reading cannot vary by more than 10% from the calibrated activity decay corrected to the day of the test. • The accuracy test should be performed during acceptance testing, annually thereafter, and after a major repair or move.
  • 11. LINEARITY PROTOCOL TESTS •This test of the dose calibrator operates appropriately over the wide activity range to which it is applied. The device is tested from 10 μCi (370 kBq) to a level higher than that routinely used in the clinic and perhaps as high as 1 Ci (37 GBq). •The activity readings are varied by starting with a sample of radioactivity of Tc-99m at the highest value to be tested (e.g., tens of gigabequerels).
  • 12. LINEARITY PROTOCOL TESTS Cont.… • The activity readings are then varied by either allowing the source to radioactively decay over several days or using a set of lead shields of varying thicknesses until a reading close to 370 kBq is obtained. Each reading should not vary by more than 10% from the line drawn through the calculated activity values. • The linearity test should be performed during acceptance testing, quarterly thereafter, and after a major repair or move.
  • 13. THE CONSTANCY PROTOCOL TESTS • This tests is the reproducibility of the readings as compared to a decay-corrected estimate for a reference reading obtained from the dose calibrator on a particular day. • Today’s constancy reading cannot vary from the decay- corrected reference reading by more than 10%. • The constancy test varies from accuracy in that it evaluates the precision of the readings from day to day rather than accuracy. • The constancy test should be performed on every day that the device is used to assay a dose to be administered to a patient.
  • 14. DOSE CALIBRATOR RADIATION PROTECTION • Lead shielding around the ionization chamber 1. Protects the operator 2. Reduces the response background radiation • Sample holder can be cleaned in the event of radioactive contamination of the chamber well
  • 15. GAMMA CAMERA The most widely used cameras in nuclear medicine are: 1. Simple gamma scintillation (Anger) camera 2. Single-photon emission computed tomography (SPECT) capable gamma camera. FUNCTION OF GAMMA CAMERA • A gamma camera converts photons emitted by the radionuclide in the patient into a light pulse and subsequently into a voltage signal.
  • 16. GAMMA CAMERA SYSTEM COMPONENTS • The collimator • The scintillation crystal • An array of photomultiplier tubes (PMTs) • Preamplifiers • A pulse height analyzer (PHA) • Digital correction circuitry, • A cathode ray tube (CRT) • The control console. • A computer and • Picture archiving systems(PACs)
  • 18. COLLIMATORS • The collimator is made of perforated or folded lead and is interposed between the patient and the scintillation crystal. It allows the gamma camera to localize accurately the radionuclide in the patient’s body. • Collimators perform this function by absorbing and stopping most radiation except that arriving almost perpendicular to the detector face. • Most radiation striking the collimator at oblique angles is not included in the final image.
  • 19. TYPES COLLIMATORS As the energy of the radionuclide increases, the best collimator usually has thicker and longer septa. For a given septal thickness, spatial resolution of a collimator increases with septal length but sensitivity decreases. The two basic types of collimators are pinhole and multihole
  • 20. SEPTAL PENETRATION AND PHOTON SCATTERING EFFECT OF SEPTAL LENGTH ON COLLIMATOR SENSITIVITY AND RESOLUTION EFFECT OF DIFFERENT SOURCE-TO-CAMERA DISTANCES
  • 21. CRYSTAL AND OTHER PHOTON DETECTOR DEVICES • Radiation emerging from the patient and passing through the collimator interacts with a thallium activated sodium iodide crystal. • Crystals are made with thallium or sodium activated cesium iodide or even lanthanum bromide are used. They convert Gamma rays to light. • PMTs situated along the posterior crystal face detect this light and amplify it to an electrical pulse.
  • 22. PHOTONMULTIPLIER TUBEs (PMTs) • A photomultiplier tube (PMT) converts a light pulse into an electrical signal of measurable magnitude. • Localization of the event in the final image depends on the amount of light sensed by each PMT and thus on the pattern of PMT voltage output. • VOLTAGE AMPLIFIER / VOLTAGE SUPPLY  A High voltage supply for the PMT  An amplifier increases the size of the pulse.
  • 23. PULSE HEIGHT ANALYZER • The basic principle of the PHA is to discard signals from background and scattered radiation and/or radiation from interfering isotopes so that only primary photons known to come from the photopeak of the isotope being imaged are recorded • The PHA discriminates between events occurring in the crystal that will be displayed or stored in the computer and events that will be rejected.
  • 24. CONSOLE CONTROLS • Image exposure time is selected by console control • Its usually a preset count, a preset time, or preset information density for the image accumulation. Other console controls are present for orientation and allow the image to be reversed on the x- and y-axes. In addition, the CRT image may be manipulated by an intensity control Hard copy images on film may be obtained directly from the computer, display digital images on monitors and store the images in a picture archiving system.
  • 25. OPERATION OF GAMMA SYSTEM  Gamma rays emitted from within the patient pass through the holes of an absorptive collimator to reach the NaI crystal.  On interaction of the gamma ray with the NaI scintillating crystal, thousands of light photons are emitted, a portion of which are collected by an array of PMTs. By taking weighted sums of the PMT signals within the associated computer  The two-dimensional (2D) x and y location and the total energy of the detection event deposited is estimated. If the energy deposited is within a prespecified energy window (e.g., within 10% of the photopeak energy), the event is accepted and the location of the event recorded.  In this manner, the gamma camera image is constructed on an event-by-event basis, and a single nuclear medicine image may consist of hundreds of thousands of such events.
  • 26. GAMMA CAMERA QUALITY CONTROL •This involves acceptance testing of the device before its initial use and a program of routine tests and evaluations applied on a regular basis. •It is essential that the performance be evaluated regularly to ensure that the images adequately demonstrate the in vivo distribution of the administered radiopharmaceutical and that any quantitation performed with the camera yields values that are accurate and precise
  • 27. GAMMA CAMERA QUALITY CONTROL SUMMARY PARAMETER COMMENT DAILY • UNIFORMITY • WINDOW SETTING • FLOOD FIELD; INTRINSIC (WITHOUT COLLIMATOR) OR EXTRINSIC (WITH COLLIMATOR) • CONFIRM ENERGY WINDOW SETTING RELATIVE TO PHOTOPEAK FOR EACH RADIONUCLIDE USED WITH EACH PATIENT WEEKLY OR MONTHLY • SPATIAL RESOLUTION • LINEARITY CHECK • REQUIRES A “RESOLUTION” PHANTOM SUCH AS THE FOUR-QUADRANT BAR • QUALITATIVE ASSESSMENT OF BAR PATTERN LINEARITY The purpose of quality control is to detect changes in the performance of a gamma camera and below summary:
  • 28. GAMMA CAMERA QUALITY CONTROL SUMMARY Cont.... PARAMETER COMMENT ANNUALLY • SYSTEM UNIFORMITY • MULTI-WINDOW REGISTRATION • COUNT RATE PERFORMANCE • ENERGY RESOLUTION • SYSTEM SENSITIVITY • HIGH COUNT FLOOD WITH EACH COLLIMATOR • FOR CAMERAS WITH CAPABILITY OF IMAGING MULTIPLE ENERGY WINDOWS SIMULTANEOUSLY • VARY COUNTS USING DECAY OR ABSORBER METHOD • EASIEST IN CAMERAS WITH BUILT IN MULTICHANNEL ANALYZERS • COUNT RATE PERFORMANCE PER UNIT OF ACTIVITY FOR EACH COLLIMATOR
  • 29. EFFECT OF INCREASING THE PATIENT-TO-DETECTOR FACE DISTANCE ON CLINICAL IMAGES
  • 30. SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY (SPECT) •SPECT is a nuclear medicine tomographic imaging technique that uses gamma rays. •It is very similar to conventional nuclear medicine planar imaging using a gamma camera. •It is able to provide true 3D information..
  • 31. SPECT
  • 32. SPECT CONT.… •SPECT imaging is performed by using a gamma camera to acquire multiple 2-D images from multiple angles. • A computer is then used to apply a tomographic reconstruction algorithm to the multiple projections, yielding a 3-D dataset. •The dataset may then be manipulated to show thin slices along any chosen axis of the body.
  • 33. IMAGE RECONSTRUCTION • SPECT acquire raw data in the form of projection data at a variety of angles about the patient. • Image reconstruction involves the processing of these data to generate a series of cross-sectional images through the object of interest.
  • 35. APPLICATION OF SPECT • Cerebral Blood flow imaging • Myocardial perfusion imaging with thalium-201 or technetium-99 perfusion agents • Imaging of tumors or infections with agents such as gallium-67 or 111 WBCs • Certain Cases of Bone Imaging • Brain studies • Liver/ Spleen Imaging • Renal Imaging
  • 36. FOURIER TRANSFORM • Image data may be best represented in either spatial (real) or frequency space. • The mathematician Joseph Fourier noted in 1807 that any arbitrary signal can be generated by adding a large number of sine and cosine signals of varying frequencies and amplitudes. • The plot of amplitude as a function of frequency is referred to as the Fourier transform, and it defines the components of the image at each frequency. • The low frequencies provide the overall shape of the object, whereas the high frequencies help define the sharp edges and fine detail within the image.
  • 37. APPLICATIONS OF FOURIER TRANSFORM • The Fourier method is often used on images from astronomy, microbiology, images of repetitive structures such as crystals and so on. • Fourier transform is good for identifying a periodic component or lattice in an image. • Identifying regular patterns on an image has other advantages like removing regular dirty spots or noise from an image • Components of higher frequency can be removed to achieve anti-aliasing effect, i.e. removing ugly zaggy edges.
  • 38. APPLICATIONS OF FOURIER TRANSFORM Cont.….. • There are other techniques associated with Fourier transform: convolution theory correlation, sampling Reconstruction image compression, and more.
  • 39. RADIATION MONITOR-GEIGER-MUELLER COUNTER • Geiger-Mueller (GM) counters are handheld, very sensitive, inexpensive survey instruments used primarily to detect small amounts of radioactive contamination. • The detector is usually pancake shaped, it may also be cylindrical. • The detector is gas-filled and has a high applied voltage from the anode to the cathode. This causes one ionization to result in an “avalanche” of other electrons, allowing high efficiency for detection of even a single gamma ray.
  • 40. TYPES OF RADIATION DETECTORS Basically three types of radiation detectors are used in nuclear medicine Gas detectors Scintillators Semiconductors
  • 41. GAS DETECTORS • A gas radiation detector is filled with a volume of gas that acts as the sensitive material of the detector. • In some cases, it is air and in others it is an inert gas such as argon or xenon, depending on the particular detector. • Electrodes are located at either end of the sensitive volume. • The detector circuit also contains a variable voltage supply and a current detector. • As radiation passes through the sensitive volume, it causes ionization in the gas. • If a voltage is applied across the volume, the resulting ions (electrons and positive ions) will start to drift, causing a measureable current in the circuit. The current will last until all of the charge that was liberated in the event is collected at the electrodes. • The resulting current entity is referred to as a pulse and is associated with a particular detection event.
  • 42. SCINTILLATION AND SEMICONDUCTOR DETECTORS • Some crystalline materials emit a large number of light photons upon the absorption of ionizing radiation.This process is referred to as scintillation, and materials are referred to as scintillators. • As radiation interacts within the scintillator, a large number of excitations and ionizations occur. On deexcitation, the number of light photons emitted is directly proportional to the amount of energy deposited within the scintillator.
  • 43. SCINTILLATION AND SEMICONDUCTOR DETECTORS Cont.… • Thermal energy can lead to a measureable current in some semiconductor detectors such as GeLi, even in the absence of radiation, and thus these semiconductor detectors must be operated at cryogenic temperatures. • On the other hand, semiconductor detectors such as cadmium telluride (CdTe) or cadmium zinc telluride (CZT) can operate at room temperature. CdTe and CZT do not have the excellent energy resolution of GeLi, but at approximately 5%, it is still significantly better than that of sodium iodide.
  • 44. CONCLUSION • Radiation detection and counting is the corner stone of nuclear medicine. • Detectors of all types—gas detectors, scintillators, and semiconductors—are used every day in the nuclear medicine clinic. • Some are used for ancillary purposes that support the clinic, such as those used in the context of radiation protection. • Others are used to specifically acquire biological data for a particular clinical purpose. Most notably, the gamma camera is used to obtain images of the in vivo distribution of the administered radiopharmaceutical from which the patient’s physiology or function can be inferred to further define the patient’s medical picture.
  • 45. Conclusion Cont…. • A rigorous quality control program must be maintained for all equipment used in the nuclear medicine clinic to ensure the integrity of the data obtained from the patient. The quality control program for the gamma camera includes acceptance testing and tests that need to be performed on a routine basis. The nuclear medicine image acquired with the gamma camera provides a snapshot of the patient’s in vivo radiopharmaceutical distribution from a certain view and at a particular point in time. These images also can be acquired as a dynamic (time-sequence) study or in conjunction with a physiological gate such as the ECG. • Regions of interest can be drawn about specific features to provide regional quantitation or TACs of dynamic processes. • Finally, nuclear medicine instrumentation continues to evolve, including the development of devices designed for a specific clinical task such as breast imaging. It is expected that this development will continue in the years ahead.
  • 46. REFERENCES • Essentials of nuclear medicine imaging; Fred A.Mettler 2012 • The Requisites Ziessman - Nuclear Medicine, 4th ed. FRED A. METTLER • Camera Systems. Publication 1141. Vienna, Austria: International Atomic Energy Agency; 2003. • IAEA Comprehensive Clinical Audits Of Diagnostic Radiology Practices A Tool For Quality Improvement • Nuclear Medicine Instrumentation Lecture Notes:2005; Ghoorun S • Operational Levels In Radiopharmacy For Realigning Our Profession With International Guidelines Richard E. 2015 • Chandra R. Nuclear Medicine Physics: The Basics. 7th ed. Philadelphia: Williams & Wilkins; 2011.
  • 47. THANK YOU VERY MUCH AND GOD BLESS YOU

Editor's Notes

  1.  After measuring the dose rate its possible to calculate the total dose which would be received at a particular place for a particular time. Radiation Dose= Dose Rate x time of Exposure.
  2. Dose calibrator Lead shielding, Outer electrode, Collector Electrode, Well( Sample Holder), Voltage power supply Electrometer(Isotope and Range Selectors) and Display. A. The sample is placed in the shielded ionization chamber (arrow) which is behind the technologist’s protective shielding. B. The selector buttons on the control panel and display require the user to select the appropriate radionuclide in order to display the correct activity. C. Schematic diagram.
  3. Processing of the electrical signal For a given a radionuclide the amount of ionisation (number of ion pairs) produced and the amount of electrical current is directly proportional to the activity of the radionuclide. The very small current(10-9 to 10-8 amps ) produced is converted to a voltage and amplified by the electrometer. The processed voltage is finally displayed in in digital form, in units of activity (e.g. mCi or MBq). The processing involves the application of a calibration coefficient kQ that corresponds to the ionization current produced by unit activity of the radionuclide being assayed Most dose calibrators have such conversion circuits.The value of the kQ depends primarily on the types, energies and abundance of the radiations emitted by the radionuclide It also depends on the attenuation of the radiations in their passage from the point of disintegration in the sample into the gas volume
  4. Modification of the regulations since 2003 specify that the quality control program must meet the manufacturer’s recommendations or national standards.
  5. Lead shielding around the ionization chamber provides protection to personnel against radiation hazards;
  6. This signal is used to form an image of the distribution of the radionuclide.
  7. Of all the photons emitted by an administered radiopharmaceutical, more than 99% are “wasted” and not recorded by the gamma camera; less than 1% are used to generate the desired image. Thus the collimator is the “rate limiting” step in the imaging chain of gamma camera technology. Collimation is required to determine the directionality of the detected event. This is because gamma rays cannot be easily focused.
  8. Interaction of the gamma ray with the crystal may result in ejection of an orbital electron (photoelectric absorption), producing a pulse of fluorescent light (scintillation event) proportional in intensity to the energy of the gamma ray.
  9. An array of these tubes is situated behind the sodium iodide crystal and may be placed directly on the crystal, connected to the crystal by light pipes, or optically coupled to the crystal with a silicone-like material. A scintillation event occurring in the crystal is recorded by one or more PMTs.
  10. Most gamma cameras allow for a fine adjustment known as automatic peaking of the isotope. This essentially divides the photopeak window into halves and calculates the number of counts in each half. If the machine is correctly peaked, each half of the window has the same number of counts from the upper and lower portions of the photopeak. Occasionally, an asymmetric window is used to improve resolution by eliminating some of the Compton scatter
  11. In the earliest days of nuclear medicine, counting devices similar to the thyroid probe described in the previous section were used to evaluate the amount of activity in a particular tissue For example, probes could be used to evaluate the iodine uptake of the thyroid gland. However, it was not long before clinicians realized that it would be helpful to not only know the total uptake of the radiopharmaceutical within the tissue of interest but also to be able to discern the spatial distribution of the uptake within the tissue.
  12. To ensure proper operation of the gamma camera, it is essential that a comprehensive quality control program be applied. Operational Checks and Acceptance and Reference testing should be done
  13. This information is typically presented as cross-sectional slices through the patient, but can be freely reformatted or manipulated as required
  14. The dataset may then be manipulated to show thin slices along any chosen axis of the body , similar to those obtained from other tomographic techniques, such as MRI, CT, and PET
  15. SPECT and PET acquisition geometries. For SPECT (left), the gamma camera rotates about the patient, acquiring a projection image at each angle. Each projection image represents the projections of many slices acquired at that angle. For PET (right), the patient is located within a ring of detectors. A positron annihilation event leads to two photons emitted in opposite directions. The detection of two events within a small timing acceptance window (5-12 ns) are considered to be from the same event and assumed to have originated along the line of response that connects the two detectors.
  16. There are many situations in Graphics and Vision, specially in image processing and filtering, where Fourier transform is useful.
  17. Geiger-Mueller survey meter. A, This instrument is used for low levels of radiation or activity. On the instrument, the pancake detector is located at the end of the handle and the face is covered with a red plastic cap. The selector knob has various multipliers to use with the displayed reading. Note the radiation check source affixed to the side, which is used to make sure the instrument is functional. Also there is a calibration sticker. B, The dial reads in either counts per minute (CPM) or milliroentgens per hour (mR/hr). There is also a battery test range that is used when the battery check button is pushed or the selector knob is switched to battery check.
  18. These three operate on different principles and are typically used for different purposes
  19. Evolve- Develop gradually