Ce diaporama a bien été signalé.
Le téléchargement de votre SlideShare est en cours. ×

ABO-DISCREPANCIES.8615883.powerpoint.pptx

Publicité
Publicité
Publicité
Publicité
Publicité
Publicité
Publicité
Publicité
Publicité
Publicité
Publicité
Publicité
Prochain SlideShare
Chapter 2.ppt
Chapter 2.ppt
Chargement dans…3
×

Consultez-les par la suite

1 sur 22 Publicité

Plus De Contenu Connexe

Similaire à ABO-DISCREPANCIES.8615883.powerpoint.pptx (20)

Plus récents (20)

Publicité

ABO-DISCREPANCIES.8615883.powerpoint.pptx

  1. 1. ABO DISCREPENCIES Dr. BIPIN B PALLATH 06/09/2016
  2. 2. FORWARD BLOOD GROUPING • Can be done by Slide method, tile method, or tube method • Slide or tile method is usually used in case of emergencies or in camps • The most accepted method for routine grouping is tube method
  3. 3. • Steps of tube method 1. Get the sample, see for any blood lysis 2. Separate Cell and Serum 3. Wash the Cell 3 times with normal saline and clear the supernatant away. 4. Make 5% cell suspension by taking 19 drops of normal saline and 1 drop of red cell.
  4. 4. 5. Take 4 test tubes and mark as Anti A, Anti B , Anti AB. And Control 6 Add one drop of corresponding antisera to the test tubes and Normal saline to control 7.Add one drop of 5% Red cell suspension to the above test tubes
  5. 5. 8. Mix it gently and wait for 10 min, and centrifuge at 1000 rpm for 1 min 9. Then look for agglutination. Free cell suspension indicates negative results. 10. Negative cell results should be conformed with microscope.
  6. 6. SERUM GROUPING Reverse Grouping • A, B, O cell reagents are prepared by pooling three sample of each group • Wash each with normal saline for 3 times and make 5% solution of each by adding 19 drops of saline and 1 drop of red cell
  7. 7. • Procedure 1. Take 3 test tubes and mark A, B, O 2. Place 2 drops of test serum to each test tubes 3. Put one drop of A, B, O cells into corresponding test tubes 4. Mix the test tubes and keep it in room temperature for 10min
  8. 8. 5. Centrifuge at 1000 rpm for 1 min Examine the tubes for haemolysis or agglutination which indicates positive results
  9. 9. Grading of results • One solid aggregate - +4 • Several large aggregate of cells - +3 • Medium sized aggregates of cells - +2 • Small aggregates in reddish back ground - +1 • Microscopic aggregates only - +W • No aggregate - 0
  10. 10. DISCREPANCY • A discrepancy occurs when the red cell testing does NOT match the serum testing results • In other words, the forward does NOT match the reverse
  11. 11. MAY BE BECAUSE • Reaction strengths could be weaker than expected • Some reactions may be missing in the reverse or forward typing • Extra reactions may occur
  12. 12. THINGS TO DO • Identify the problem Most of the time, the problem is technical – Mislabeled tube – Failure to add reagent Either repeat test on same sample, request a new sample, or wash cells • Some times, there is a real discrepancy due to problems with the patient’s red cells or serum
  13. 13. ERRORS
  14. 14. Technical Errors • Clerical errors –Mislabeled tubes –Patient misidentification –Inaccurate interpretations recorded –Transcription error –Computer entry error
  15. 15. • Reagent or equipment problems –Using expired reagents –Using an uncalibrated centrifuge –Contaminated or hemolysed reagents –Incorrect storage temperatures
  16. 16. • Procedural errors –Reagents not added –Manufacturer’s directions not followed –RBC suspensions incorrect concentration –Cell buttons not resuspended before grading agglutination
  17. 17. CLOTTING DEFICIENCIES • Serum that does not clot may be due to: –Low platelet counts –Anticoagulant therapy (Heparin, Aspirin, etc) –Factor deficiencies
  18. 18. • Serum that does not clot completely before testing is prone to developing fibrin clots that may mimic agglutination • Thrombin can be added to serum to activate clot formation • Tubes containing EDTA can be used
  19. 19. Contaminated samples or reagents • Sample contamination – Microbial growth in tube • Reagent contamination – Bacterial growth causes cloudy or discolored appearance, so do not use if you see this. – Reagents contaminated with other reagents (don’t touch side of tube when dispensing)
  20. 20. Equipment problems • Routine maintenance should be performed on a regular basis (daily, weekly, etc) • Keep instruments like centrifuges, thermometers, and timers calibrated –Uncalibrated centrifuges can cause false results
  21. 21. Hemolysis • Detected in serum after centrifugation (red) • Important if not documented • Can result from: –Complement binding • Anti-A, anti-B, anti-H, and anti-Lea –Bacterial contamination
  22. 22. Interesting, right? This is just a sneak preview of the full presentation. We hope you like it! To see the rest of it, just click here to view it in full on PowerShow.com. Then, if you’d like, you can also log in to PowerShow.com to download the entire presentation for free.

×