Technical Information
TOCII CA72TOC
Analyzer for online-determination of TOC in aqueous media using
thermic catalytic combustion
Application
• Industrial wastewater monitoring
• Control of process wastewater
• Monitoring of industrial surface water
• Municipal wastewater monitoring
• Recording of the carbon load to control the nutrient dosage
Your benefits
• Optimized for industrial applications with varying pH values
and high salt loads
• Fast and easy maintenance thanks to good accessibility of all
components
• Heatable salt trap significantly increases the service life
• Exchangeable furnace concept significantly decreases service
time thanks to prepared furnace
• pH-controlled acid dosage for TIC stripping minimizes acid
consumption
• Adjustable measuring range: 0.25 mg/l to 12,000 mg/l TOC
expandable to 240 000 mg/l by predilution of factor 20
• Measurement proceeds as "Double Batch"
• One and two-channel measurement available
• External triggered self testing with TOC standard (e.g. for
limit violation)
TOCII CA72TOC
2 Endress+Hauser
Function and system design
Measuring principle The analyzer determines the TOC (total organic carbon) content of an aqueous sample. It uses thermic catalytic
combustion with subsequent NDIR (non-dispersive infrared) detection of the produced CO2. TOC indicates
the total organic load of water. It is used to assess the water quality and can serve as a basis for calculation of
wastewater charges.
The analyzer works with two interconnected circuits, a liquid and a gas circuit, to determine TOC by the direct
method. In the liquid circuit, the sample is pumped into the analyzer where it can be diluted in case of high
salt loads or TOC values. The sample is acidified and the inorganic carbon is eliminated (Stripping). Then the
sample is transferred through a rotation slit filter in which the particles are separated according to DIN
standards. After that, it is dosed into the furnace.
The sample is combusted at a temperature of approx. 850 °C (1550 °F). Then, the combustion gas is cooled
down and the CO2 content is determined by the NDIR detector.
The double-batch operation makes sure that the reactor and gas circuit are flushed with CO2-free carrier gas
and the basic trace is determined while the next sample is prepared in the liquid circuit.
Measurement; calibration and
adjustment of the system
The double-batch operation links and matches two different batch procedures. On the one hand, the sample is
batch-wise fed into the furnace. On the other hand, the reactor is batch-wise flushed with CO2-free carrier gas
after each measurement. To monitor the measurement, a 1-point calibration can be triggered externally or by
a timer. The 2-point adjustment allows to adapt the analyzer to system changes.
TOC as sum parameter The total carbon of an aqueous sample is divided into total inorganic carbon (TIC) and total organic carbon
(TOC). TOC can again be divided into three groups:
• Dissolved organic carbon (DOC)
• Non-purgeable organic carbon (NPOC))
• Purgeable organic carbon (POC).
For the classification of organic carbon, it is important to note the difference between POC and VOC (volatile
organic carbon). POC is actively eliminated during online analysis of TOC (e.g. during stripping). VOC is a
scientific term that describes the properties of volatile organic carbons. Volatilization of substances is a passive
process strongly influenced by temperature and pressure.
a0011787-en
Classification of carbon parameters
Important features Heatable salt trap
The heatable salt trap simplifies maintenance in applications with high salt loads. When samples with high salt
loads are combusted, deposits form on the catalyst and the furnace which leads to inaccurate measurements
and might result in congestion of the system. With the heatable salt trap, salts can be removed from the furnace
quickly and efficiently. This increases the availability of the measuring point dramatically. It also reduces
operating costs thanks to longer maintenance intervals and easy handling of the salt trap.
The salt trap can be removed from the furnace without having to switch the furnace off. The salts in the trap
can easily be rinsed off and the trap can be screwed back into position.
Inorganic C
Carbonates,
cyanide compounds,
carbonic acids... Sum of and
carbon in and
(i.e. particulate) compounds
inorganic organic
dissolved undissolved
Organic C
Albumen, sugar, organic
acids, humic acids
aromatic compounds,
fats...
TC
Total carbon
TIC
Total
Inorganic
Carbon
TOC
Total
Organic
Carbon
NPOC
Non purgeable
Organic
Carbon
POC
Purgeable
Organic
Carbon
DOC
Dissolved
Organic
Carbon
TOCII CA72TOC
Endress+Hauser 3
Exchangeable furnace concept
The exchangeable furnace increases the availability of the measuring point by considerably shortening the
maintenance process. A second furnace can be prepared in an optional separate preparation unit (burn out of
catalyst, rinsing and preheating of furnace). This means that burn out and conditioning of the catalyst do not
need to be done in the analyzer. The used furnace can simply be exchanged with a prepared fresh furnace and
the system is back to operation within a short period of time.
pH-controlled acid dosing
pH-controlled acid dosing offers the advantage that the exact amount of acid is added to the sample that is
necessary to reach a pH value of 2.5. Excess dosing is avoided which results in reduced acid consumption. At
the same time, salt drag-in into the furnace due to excess acid is minimized.
In municipal wastewater treatment plants, acidification of the sample can lead to precipitation of humic acids
which can mask the carbonate. This inorganic carbon ends up in the furnace and causes too high measured
values.
Double-batch operation
The patented double-batch operation links the water and the gas circuits. The aqueous sample is continuously
prepared in the analyzer and fed batch-wise into the furnace. During measurement, the CO2-containing gas is
circulated and accumulated in the gas circuit. This allows to record large sample volumes (1200 μl) which leads
to a high sensitivity. After measurement, the gas circuit is flushed with CO2-free carrier gas and the basic line
for the next measurement is determined.
Two-channel measurement
The two-channel measurement allows to measure two different sample streams, e.g. two inlets of a wastewater
treatment plant, with one device. This option is suitable for sample streams with similar TOC values.
Sample preparation The following models of sample preparation are available:
Measuring system
a0011809-en
Schematics of measuring point
Model Flow Material Recommendation
PA-2 1 to 8 m3
/h (260 to 2000 gal/hr) PVC Recommended for multiple analyzer systems in
series (e.g. CA72TOC an CA72TP); for probe
temperatures <55 °C (<130 °F) and diluted
solutions
PA-3 0,1 to 1 m3
/h (26 to 260 gal/hr) PVC Recommended for probe temperatures <55 °C
(<130 °F) and diluted solutions
PA-9 0,1 to 1 m3
/h (26 to 260 gal/hr) PP For problematic wastewater due to high chemical
resistance (except oxidizing acids and halogens)
for probe temperatures <55 °C (<130 °F)
Sample
Sample
processor Analyzer CO -scrubber2
Gas
supply
Water
supply
TOCII CA72TOC
4 Endress+Hauser
Input
Measured variables TOC (TC possible without stripper)
Measuring range
With the optional pre-dilution the measuring range will expand by the factor of 20.
Signal inputs 8 signal inputs 24 V DC, active, load max. 500 Ω
Output
Output signals Channel 1: 0/4 to 20 mA, galvanically separated
Channel 2 (two channel version only): 0/4 to 20 mA, galvanically separated
Load: max. 500 Ω
Alarm signals Limit alarm, malfunction message, standby message and operation control, potential free, normally closed
(max. 0.25 A / 50 V)
Interface RS 232 C, proprietary, for data output and remote control (optional)
Version Measuring range
A 0.25 to 600 mg TOC/l
B 1 to 2400 mg TOC/l
C 2.5 to 6000 mg TOC/l
D 5 to 12000 mg TOC/l
Input #1 Activation service calibration
Input #2 Activation service adjustment
Input #3 Activation service flush screen
Input #4 Activation service power-flush
Input #5
Input #6
Input #7 Activation standby
Input #8 Activation channel switch over (optional)
TOCII CA72TOC
Endress+Hauser 5
Power supply
Signal connection
a0011911
Signal connection
Note!
TOCII CA72TOC is suitable for connection to industrial mains support according to EN 61326-1, class A.
Electrical connection
Fuses
binaryoutbinaryout
1
2
3
4
5
6
7
8
binaryinbinaryin
0/4-20
mA
0/4-20
mA
RS 232RS 232
IIIIIVII
+
-
+
-
40
41
1
2
1
2
1
2
1
2
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
I Alert 1 External activation of calibration
II Alarm for limits 2 External activation of adjustment
III Standby 3 External activation of flush screen
IV Operating control 4 External activation of power flush
40 Signal output channel 1 5
41 Signal output channel 2 (optional) 6
7 External activation of standby
8 External activation of channel switch over (optional)
Message I to IV Relay contact (max. 0.2 A and 50 V), normally closed
Signal output 40 to 41 0 to 20 mA or 4 to 20 mA switchable, galvanically separated, load max.
500 Ω
Signal input 1 to 8 24 V DC active, load max. 500 Ω
Version Average power consumption Maximum power consumption
115 V AC 50 Hz 490 VA 650 VA
115 V AC 60 Hz 490 VA 650 VA
230 V AC 50 Hz 490 VA 650 VA
230 V AC 60 Hz 490 VA 650 VA
Modules Fuses
Power distribution 16 A, slow blow
Relays per relay 4 A, slow blow, type: TR5
Power supply 2 A, slow blow, type: microfuse 5 x 20
TOCII CA72TOC
6 Endress+Hauser
Performance characteristic
According to ISO 15839 According to ISO 15839
The performance characteristics are determined according to ISO 15839 appendix B. 300 μl sample per
measurement are dosed at TOCII (CA72TOC-A1A0B1) resulting in a measuring range of 4 to 800 mg/l. The
following data correspond to this system.
It is assumed that the performance characteristics can be transferred to additional measuring ranges with minor
deviations.
Inorganic carbon removal > 95 % using a pH controlled TIC scrubber
Installation conditions
Water connection
Drain Drain underneath the analyzer. Use drain 6/8 mm made of PTFE. Do not install the drain with any restrictions
or back pressure.
Compressed air The compressed air must be dry and must meet the following specifications:
• < 3 ppm CO2
• < 3 ppm hydrocarbon
• Constant pressure 2 bar (29 psi)
• Pressure tolerance ±5 %
• Consumption 650 to 750 cm3/min (1.37 to 1.6 ft3/hr)
• The in-house compressed air supply must be equipped with a CO2 scrubber (input pressure 4.0 to 10 bar
(58 to 145 psi)) and a pressure reducing valve.
Connection: 4/6 mm
Exhaust Inside of buildings an exhaust is required. No accumulation of halogens or other vapors is allowed in this area.
In the exhaust hose (4/6 mm) no back pressure is allowed.
Performance characteristics according to ISO 15839 chapter 5 Value
Limit of detection LOD 0.75 % of full scale
Limit of quantitation LOQ 2.5 % of full scale
Short term drift 0.5 %/day
Repeatability for 20 % of full scale 0.4 %
Limit of resolution for 20 % of full scale(LDC) 1.1 %
Systematic measured error for 20 % of full scale (BIAS) 0.4 %
Repeatability for 80 % of full scale 1.6 %
Limit of resolution for 80 % of full scale(LDC) 4.6 %
Systematic measured error for 80 % of full scale (BIAS) 2.4 %
Pressure 2 to 4 bar (29 to 58 psi)
Connection 6/8 mm DN or G3/8 threaded connection
TOCII CA72TOC
Endress+Hauser 7
Environment
Ambient temperature 5 to 35 °C (41 to 95 °F)
Humidity 10 to 90 %, non-condensing
Ingress protection IP 54
Process
Sample temperature 0 to 40 °C (32 to 104 °F)
Sample flow rate 20 ml/min (0.32 US gal/hr)
Sample volume 90 ml (3 fl.oz)
Sample preparations PA-2, PA-3 and PA-9 are available.
Sample consistency Aqueous
Flammable substances are not allowed in ignitable concentrations - in such cases a sample dilution is required.
Sample inlet Unpressurized outlet of the sample processor to the analyzer
TOCII CA72TOC
8 Endress+Hauser
Mechanical construction
Dimensions
a0012435
Dimensions
Weight approx. 75 kg (165 lbs)
Materials
400 (15.8)
600 (23.6)
785 (30.9)
1010(39.8)
1558(61.3)
mm (inch)
50
(1.97)
Housing Aluminium, powder coated (RAL7035)
Window Glass, conductive coated
Valve seals EPDM, FPM, FFKM
Pump hoses Ismapren
Pumps and pump seals PTFE, FFKM
Reagents- und sample hoses PTFE, PE
Exhaust hoses PTFE, PE
Drain hoses PTFE
TOCII CA72TOC
Endress+Hauser 9
Human interface
Display and operating
elements
a0007996
Display and operating elements
1 Diaplay, 16 rows a 40 characters
2 Keyboard
3 Arrow keys (cursors)
4 Function key"Measuring"
5 Function key "Service"
6 Function key "Programming"
7 Function key "Help"
8 Key "Enter"
9 Key "Clear"
Certificates and approvals
4 symbol Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus complies with the legal
requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.
1 2 3
4
5
6
7
89
TOCII CA72TOC
10 Endress+Hauser
Ordering information
Product structure
Note!
To complete your order code, simply add the optional features to the end of order code. If you have any
questions, please contact your local sales office.
Measuring range
A 0.25 to 600 mg/l TOC
B 1 to 2400 mg/l TOC
C 2.5 to 6000 mg/l TOC
D 5 to 12000 mg/l TOC
Y Special version according to customer specification
Sample transfer
1 1 measuring point
2 2 measuring points; time controlled
3 2 measuring points; external controlled
Operation mode
A Batch mode
Power supply
0 230 VAC, 50 Hz
1 115 VAC, 60 Hz
2 115 VAC, 50 Hz
3 230 VAC, 60 Hz
9 Special version according to customer specification
Sample preparation
A 1 x PA-2 PVC, 1 to 8 m3
/h (4.4 to 35 gpm) wastewater (not with CA72TOC-*2*********)
B 1 x PA-3 PVC, 0.1 to 1.0 m3
/h (0.4 to 4.4 gpm) wastewater (not with CA72TOC-*2*********)
C 1 x PA-9 PP, 0.1 to 1.0 m3
/h (0.4 to 4.4 gpm) wastewater (not with CA72TOC-*2*********)
D 2 x PA-2 PVC, 1 to 8 m3
/h (4.4 to 35 gpm) wastewater (not with CA72TOC-*1*********)
E 2 x PA-3 PVC, 0.1 to 1.0 m3
/h (0.4 to 4.4 gpm) wastewater (not with CA72TOC-*1*********)
F 2 x PA-9 PVC, 0.1 to 1.0 m3/h (0.4 to 4.4 gpm) wastewater (not with CA72TOC-*1*********)
Y Special version according to customer specification
Language instrument, documentation
1 German
2 English
CA72TOC- Order code
Carrying gas (optinal, select one option only)
E1 CO2-scrubber
E2 CO2-scrubber + compressor
Analyzer mounting (optinal, select one option only)
F1 Wall mounted
F2 Stand
Accessory mounted (optinal, multiple selection possible)
G1 Sample predilution
G4 Heated salt trap
G5 Pipe back flushing
G6 Standby mode PA-2/-3 failure
G7 sample valve for aggressive media
Accessory enclosed (optinal, multiple selection possible)
H1 230 VAC furnace conditioning station (not with CA72TOC-***1******* and not with
CA72TOC-***2*******)
H2 115 VAC furnace conditioning station (not with CA72TOC-***0******* and not with
CA72TOC-***3*******)
H3 Standard furnace (not with CA72TOC-********G4**)
H4 Furnace for heated salt trap
TOCII CA72TOC
Endress+Hauser 11
Scope of delivery The scope of delivery comprises:
• 1 analyzer CA72TOC
• Tool kit for leak check
• Tool kit to remove the glass ball and the media
• Acid filter kit
• Kit for commisioning of stripper and separation chamber
• Maintenance kit for furnace
• Hose kit for inlet and outlet
• 1 canister 5 liter
• 2 canisters 2 liter
• 3 covers for canisters
• Key for door lock
• Connection box EMV
• Measuring glas 10 ml
• Sponge
• Safety glasses
• Gloves, acid-proof and base proof
• Silicone
• Operating Instructions
Accessories
Reagents
• CAY450-V10AAE, 1000 ml stripper reagent für CA72TOC
• CAY451-V10C01AAH, 1000 ml parent solution 10 000 mg/l TOC
• CAY451-V10C10AAH, 1000 ml parent solution100 000 mg/l TOC
Furnace conditioning station
• 230 V AC 50/60 Hz order number: 71103492
• 115 V AC 50/60 Hz order number: 71103493
Standard furnace
• As additional furnace or als replacement
• Not for heated salt trap
• Order number: 71091188
Furnace for heated salt trap
• As additional furnace or as replacement
• Order number: 71103888
Kit CA72TOC heated salt trap
• As replacement for maintenance (decreases the maintenance time) or as replacement
• Order number: 71101532
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
Tel.+41 61 715 81 00
Fax+41 61 715 25 00
www.endress.com
info@ii.endress.com
TI448C/07/en/10.09
71098868
Printed in Germany / FM+SGML 6.0 / DT

More Related Content

More from Edress Hauser: Flow Meter, Level, Pressure, Temperature

More from Edress Hauser: Flow Meter, Level, Pressure, Temperature (20)

Level Limit Switch-Soliswitch FTE31
Level Limit Switch-Soliswitch FTE31Level Limit Switch-Soliswitch FTE31
Level Limit Switch-Soliswitch FTE31
 
Level limit switch-Soliswitch FTE30
Level limit switch-Soliswitch FTE30Level limit switch-Soliswitch FTE30
Level limit switch-Soliswitch FTE30
 
Flow indicator for bulk solids-Solimotion FTR20
Flow indicator for bulk solids-Solimotion FTR20Flow indicator for bulk solids-Solimotion FTR20
Flow indicator for bulk solids-Solimotion FTR20
 
Level measurement with Prothermo NMT532
Level measurement with Prothermo NMT532 Level measurement with Prothermo NMT532
Level measurement with Prothermo NMT532
 
High accuracy liquid level measurement-Proservo NMS5
High accuracy liquid level measurement-Proservo NMS5High accuracy liquid level measurement-Proservo NMS5
High accuracy liquid level measurement-Proservo NMS5
 
Conductive Limit Detection-Rod Probe 11263
Conductive Limit Detection-Rod Probe 11263Conductive Limit Detection-Rod Probe 11263
Conductive Limit Detection-Rod Probe 11263
 
FTW360 compact pump protector
FTW360 compact pump protectorFTW360 compact pump protector
FTW360 compact pump protector
 
Conductive-Point level switch for multiple point
Conductive-Point level switch for multiple pointConductive-Point level switch for multiple point
Conductive-Point level switch for multiple point
 
Level limit switch compact design-Liquiphant T FTL20H
Level limit switch compact design-Liquiphant T FTL20HLevel limit switch compact design-Liquiphant T FTL20H
Level limit switch compact design-Liquiphant T FTL20H
 
Vibronic-Point level switch for all kinds of liquids
Vibronic-Point level switch for all kinds of liquidsVibronic-Point level switch for all kinds of liquids
Vibronic-Point level switch for all kinds of liquids
 
Vibronic-Point level switch-Liquiphant M FTL51C
Vibronic-Point level switch-Liquiphant M FTL51CVibronic-Point level switch-Liquiphant M FTL51C
Vibronic-Point level switch-Liquiphant M FTL51C
 
Level limit switch for liquids-Liquiphant T FTL20
Level limit switch for liquids-Liquiphant T FTL20Level limit switch for liquids-Liquiphant T FTL20
Level limit switch for liquids-Liquiphant T FTL20
 
Radiometric measuring technology-Gammapilot FTG20
Radiometric measuring technology-Gammapilot FTG20Radiometric measuring technology-Gammapilot FTG20
Radiometric measuring technology-Gammapilot FTG20
 
Differential pressure measurement-Level transmiter
Differential pressure measurement-Level transmiterDifferential pressure measurement-Level transmiter
Differential pressure measurement-Level transmiter
 
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
 
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
 
Capacitance level measurement-Liquicap M FMI51, FMI52
Capacitance level measurement-Liquicap M FMI51, FMI52Capacitance level measurement-Liquicap M FMI51, FMI52
Capacitance level measurement-Liquicap M FMI51, FMI52
 
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
 
Ultrasonic Level Measurement-Prosonic T FMU30
Ultrasonic Level Measurement-Prosonic T FMU30Ultrasonic Level Measurement-Prosonic T FMU30
Ultrasonic Level Measurement-Prosonic T FMU30
 
Electromechanical Level Measuring System-Silopilot M FMM50
Electromechanical Level Measuring System-Silopilot M FMM50Electromechanical Level Measuring System-Silopilot M FMM50
Electromechanical Level Measuring System-Silopilot M FMM50
 

Recently uploaded

Partners Life - Insurer Innovation Award 2024
Partners Life - Insurer Innovation Award 2024Partners Life - Insurer Innovation Award 2024
Partners Life - Insurer Innovation Award 2024The Digital Insurer
 
Powerful Google developer tools for immediate impact! (2023-24 C)
Powerful Google developer tools for immediate impact! (2023-24 C)Powerful Google developer tools for immediate impact! (2023-24 C)
Powerful Google developer tools for immediate impact! (2023-24 C)wesley chun
 
Real Time Object Detection Using Open CV
Real Time Object Detection Using Open CVReal Time Object Detection Using Open CV
Real Time Object Detection Using Open CVKhem
 
Artificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and MythsArtificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and MythsJoaquim Jorge
 
2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...Martijn de Jong
 
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, AdobeApidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobeapidays
 
Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024The Digital Insurer
 
Driving Behavioral Change for Information Management through Data-Driven Gree...
Driving Behavioral Change for Information Management through Data-Driven Gree...Driving Behavioral Change for Information Management through Data-Driven Gree...
Driving Behavioral Change for Information Management through Data-Driven Gree...Enterprise Knowledge
 
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Miguel Araújo
 
GenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationGenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationMichael W. Hawkins
 
Finology Group – Insurtech Innovation Award 2024
Finology Group – Insurtech Innovation Award 2024Finology Group – Insurtech Innovation Award 2024
Finology Group – Insurtech Innovation Award 2024The Digital Insurer
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherRemote DBA Services
 
Handwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed textsHandwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed textsMaria Levchenko
 
Tech Trends Report 2024 Future Today Institute.pdf
Tech Trends Report 2024 Future Today Institute.pdfTech Trends Report 2024 Future Today Institute.pdf
Tech Trends Report 2024 Future Today Institute.pdfhans926745
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptxHampshireHUG
 
🐬 The future of MySQL is Postgres 🐘
🐬  The future of MySQL is Postgres   🐘🐬  The future of MySQL is Postgres   🐘
🐬 The future of MySQL is Postgres 🐘RTylerCroy
 
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...DianaGray10
 
Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...apidays
 
Boost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityBoost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityPrincipled Technologies
 
Advantages of Hiring UIUX Design Service Providers for Your Business
Advantages of Hiring UIUX Design Service Providers for Your BusinessAdvantages of Hiring UIUX Design Service Providers for Your Business
Advantages of Hiring UIUX Design Service Providers for Your BusinessPixlogix Infotech
 

Recently uploaded (20)

Partners Life - Insurer Innovation Award 2024
Partners Life - Insurer Innovation Award 2024Partners Life - Insurer Innovation Award 2024
Partners Life - Insurer Innovation Award 2024
 
Powerful Google developer tools for immediate impact! (2023-24 C)
Powerful Google developer tools for immediate impact! (2023-24 C)Powerful Google developer tools for immediate impact! (2023-24 C)
Powerful Google developer tools for immediate impact! (2023-24 C)
 
Real Time Object Detection Using Open CV
Real Time Object Detection Using Open CVReal Time Object Detection Using Open CV
Real Time Object Detection Using Open CV
 
Artificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and MythsArtificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and Myths
 
2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...
 
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, AdobeApidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
 
Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024
 
Driving Behavioral Change for Information Management through Data-Driven Gree...
Driving Behavioral Change for Information Management through Data-Driven Gree...Driving Behavioral Change for Information Management through Data-Driven Gree...
Driving Behavioral Change for Information Management through Data-Driven Gree...
 
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
 
GenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationGenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day Presentation
 
Finology Group – Insurtech Innovation Award 2024
Finology Group – Insurtech Innovation Award 2024Finology Group – Insurtech Innovation Award 2024
Finology Group – Insurtech Innovation Award 2024
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a Fresher
 
Handwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed textsHandwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed texts
 
Tech Trends Report 2024 Future Today Institute.pdf
Tech Trends Report 2024 Future Today Institute.pdfTech Trends Report 2024 Future Today Institute.pdf
Tech Trends Report 2024 Future Today Institute.pdf
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
 
🐬 The future of MySQL is Postgres 🐘
🐬  The future of MySQL is Postgres   🐘🐬  The future of MySQL is Postgres   🐘
🐬 The future of MySQL is Postgres 🐘
 
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
 
Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...
 
Boost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityBoost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivity
 
Advantages of Hiring UIUX Design Service Providers for Your Business
Advantages of Hiring UIUX Design Service Providers for Your BusinessAdvantages of Hiring UIUX Design Service Providers for Your Business
Advantages of Hiring UIUX Design Service Providers for Your Business
 

TOC On-line continuous monitoring-TOCII CA72TOC

  • 1. TI448C/07/en/10.09 71098868 Technical Information TOCII CA72TOC Analyzer for online-determination of TOC in aqueous media using thermic catalytic combustion Application • Industrial wastewater monitoring • Control of process wastewater • Monitoring of industrial surface water • Municipal wastewater monitoring • Recording of the carbon load to control the nutrient dosage Your benefits • Optimized for industrial applications with varying pH values and high salt loads • Fast and easy maintenance thanks to good accessibility of all components • Heatable salt trap significantly increases the service life • Exchangeable furnace concept significantly decreases service time thanks to prepared furnace • pH-controlled acid dosage for TIC stripping minimizes acid consumption • Adjustable measuring range: 0.25 mg/l to 12,000 mg/l TOC expandable to 240 000 mg/l by predilution of factor 20 • Measurement proceeds as "Double Batch" • One and two-channel measurement available • External triggered self testing with TOC standard (e.g. for limit violation)
  • 2. TOCII CA72TOC 2 Endress+Hauser Function and system design Measuring principle The analyzer determines the TOC (total organic carbon) content of an aqueous sample. It uses thermic catalytic combustion with subsequent NDIR (non-dispersive infrared) detection of the produced CO2. TOC indicates the total organic load of water. It is used to assess the water quality and can serve as a basis for calculation of wastewater charges. The analyzer works with two interconnected circuits, a liquid and a gas circuit, to determine TOC by the direct method. In the liquid circuit, the sample is pumped into the analyzer where it can be diluted in case of high salt loads or TOC values. The sample is acidified and the inorganic carbon is eliminated (Stripping). Then the sample is transferred through a rotation slit filter in which the particles are separated according to DIN standards. After that, it is dosed into the furnace. The sample is combusted at a temperature of approx. 850 °C (1550 °F). Then, the combustion gas is cooled down and the CO2 content is determined by the NDIR detector. The double-batch operation makes sure that the reactor and gas circuit are flushed with CO2-free carrier gas and the basic trace is determined while the next sample is prepared in the liquid circuit. Measurement; calibration and adjustment of the system The double-batch operation links and matches two different batch procedures. On the one hand, the sample is batch-wise fed into the furnace. On the other hand, the reactor is batch-wise flushed with CO2-free carrier gas after each measurement. To monitor the measurement, a 1-point calibration can be triggered externally or by a timer. The 2-point adjustment allows to adapt the analyzer to system changes. TOC as sum parameter The total carbon of an aqueous sample is divided into total inorganic carbon (TIC) and total organic carbon (TOC). TOC can again be divided into three groups: • Dissolved organic carbon (DOC) • Non-purgeable organic carbon (NPOC)) • Purgeable organic carbon (POC). For the classification of organic carbon, it is important to note the difference between POC and VOC (volatile organic carbon). POC is actively eliminated during online analysis of TOC (e.g. during stripping). VOC is a scientific term that describes the properties of volatile organic carbons. Volatilization of substances is a passive process strongly influenced by temperature and pressure. a0011787-en Classification of carbon parameters Important features Heatable salt trap The heatable salt trap simplifies maintenance in applications with high salt loads. When samples with high salt loads are combusted, deposits form on the catalyst and the furnace which leads to inaccurate measurements and might result in congestion of the system. With the heatable salt trap, salts can be removed from the furnace quickly and efficiently. This increases the availability of the measuring point dramatically. It also reduces operating costs thanks to longer maintenance intervals and easy handling of the salt trap. The salt trap can be removed from the furnace without having to switch the furnace off. The salts in the trap can easily be rinsed off and the trap can be screwed back into position. Inorganic C Carbonates, cyanide compounds, carbonic acids... Sum of and carbon in and (i.e. particulate) compounds inorganic organic dissolved undissolved Organic C Albumen, sugar, organic acids, humic acids aromatic compounds, fats... TC Total carbon TIC Total Inorganic Carbon TOC Total Organic Carbon NPOC Non purgeable Organic Carbon POC Purgeable Organic Carbon DOC Dissolved Organic Carbon
  • 3. TOCII CA72TOC Endress+Hauser 3 Exchangeable furnace concept The exchangeable furnace increases the availability of the measuring point by considerably shortening the maintenance process. A second furnace can be prepared in an optional separate preparation unit (burn out of catalyst, rinsing and preheating of furnace). This means that burn out and conditioning of the catalyst do not need to be done in the analyzer. The used furnace can simply be exchanged with a prepared fresh furnace and the system is back to operation within a short period of time. pH-controlled acid dosing pH-controlled acid dosing offers the advantage that the exact amount of acid is added to the sample that is necessary to reach a pH value of 2.5. Excess dosing is avoided which results in reduced acid consumption. At the same time, salt drag-in into the furnace due to excess acid is minimized. In municipal wastewater treatment plants, acidification of the sample can lead to precipitation of humic acids which can mask the carbonate. This inorganic carbon ends up in the furnace and causes too high measured values. Double-batch operation The patented double-batch operation links the water and the gas circuits. The aqueous sample is continuously prepared in the analyzer and fed batch-wise into the furnace. During measurement, the CO2-containing gas is circulated and accumulated in the gas circuit. This allows to record large sample volumes (1200 μl) which leads to a high sensitivity. After measurement, the gas circuit is flushed with CO2-free carrier gas and the basic line for the next measurement is determined. Two-channel measurement The two-channel measurement allows to measure two different sample streams, e.g. two inlets of a wastewater treatment plant, with one device. This option is suitable for sample streams with similar TOC values. Sample preparation The following models of sample preparation are available: Measuring system a0011809-en Schematics of measuring point Model Flow Material Recommendation PA-2 1 to 8 m3 /h (260 to 2000 gal/hr) PVC Recommended for multiple analyzer systems in series (e.g. CA72TOC an CA72TP); for probe temperatures <55 °C (<130 °F) and diluted solutions PA-3 0,1 to 1 m3 /h (26 to 260 gal/hr) PVC Recommended for probe temperatures <55 °C (<130 °F) and diluted solutions PA-9 0,1 to 1 m3 /h (26 to 260 gal/hr) PP For problematic wastewater due to high chemical resistance (except oxidizing acids and halogens) for probe temperatures <55 °C (<130 °F) Sample Sample processor Analyzer CO -scrubber2 Gas supply Water supply
  • 4. TOCII CA72TOC 4 Endress+Hauser Input Measured variables TOC (TC possible without stripper) Measuring range With the optional pre-dilution the measuring range will expand by the factor of 20. Signal inputs 8 signal inputs 24 V DC, active, load max. 500 Ω Output Output signals Channel 1: 0/4 to 20 mA, galvanically separated Channel 2 (two channel version only): 0/4 to 20 mA, galvanically separated Load: max. 500 Ω Alarm signals Limit alarm, malfunction message, standby message and operation control, potential free, normally closed (max. 0.25 A / 50 V) Interface RS 232 C, proprietary, for data output and remote control (optional) Version Measuring range A 0.25 to 600 mg TOC/l B 1 to 2400 mg TOC/l C 2.5 to 6000 mg TOC/l D 5 to 12000 mg TOC/l Input #1 Activation service calibration Input #2 Activation service adjustment Input #3 Activation service flush screen Input #4 Activation service power-flush Input #5 Input #6 Input #7 Activation standby Input #8 Activation channel switch over (optional)
  • 5. TOCII CA72TOC Endress+Hauser 5 Power supply Signal connection a0011911 Signal connection Note! TOCII CA72TOC is suitable for connection to industrial mains support according to EN 61326-1, class A. Electrical connection Fuses binaryoutbinaryout 1 2 3 4 5 6 7 8 binaryinbinaryin 0/4-20 mA 0/4-20 mA RS 232RS 232 IIIIIVII + - + - 40 41 1 2 1 2 1 2 1 2 +24VDC GND +24VDC GND +24VDC GND +24VDC GND +24VDC GND +24VDC GND +24VDC GND +24VDC GND +24VDC GND 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 I Alert 1 External activation of calibration II Alarm for limits 2 External activation of adjustment III Standby 3 External activation of flush screen IV Operating control 4 External activation of power flush 40 Signal output channel 1 5 41 Signal output channel 2 (optional) 6 7 External activation of standby 8 External activation of channel switch over (optional) Message I to IV Relay contact (max. 0.2 A and 50 V), normally closed Signal output 40 to 41 0 to 20 mA or 4 to 20 mA switchable, galvanically separated, load max. 500 Ω Signal input 1 to 8 24 V DC active, load max. 500 Ω Version Average power consumption Maximum power consumption 115 V AC 50 Hz 490 VA 650 VA 115 V AC 60 Hz 490 VA 650 VA 230 V AC 50 Hz 490 VA 650 VA 230 V AC 60 Hz 490 VA 650 VA Modules Fuses Power distribution 16 A, slow blow Relays per relay 4 A, slow blow, type: TR5 Power supply 2 A, slow blow, type: microfuse 5 x 20
  • 6. TOCII CA72TOC 6 Endress+Hauser Performance characteristic According to ISO 15839 According to ISO 15839 The performance characteristics are determined according to ISO 15839 appendix B. 300 μl sample per measurement are dosed at TOCII (CA72TOC-A1A0B1) resulting in a measuring range of 4 to 800 mg/l. The following data correspond to this system. It is assumed that the performance characteristics can be transferred to additional measuring ranges with minor deviations. Inorganic carbon removal > 95 % using a pH controlled TIC scrubber Installation conditions Water connection Drain Drain underneath the analyzer. Use drain 6/8 mm made of PTFE. Do not install the drain with any restrictions or back pressure. Compressed air The compressed air must be dry and must meet the following specifications: • < 3 ppm CO2 • < 3 ppm hydrocarbon • Constant pressure 2 bar (29 psi) • Pressure tolerance ±5 % • Consumption 650 to 750 cm3/min (1.37 to 1.6 ft3/hr) • The in-house compressed air supply must be equipped with a CO2 scrubber (input pressure 4.0 to 10 bar (58 to 145 psi)) and a pressure reducing valve. Connection: 4/6 mm Exhaust Inside of buildings an exhaust is required. No accumulation of halogens or other vapors is allowed in this area. In the exhaust hose (4/6 mm) no back pressure is allowed. Performance characteristics according to ISO 15839 chapter 5 Value Limit of detection LOD 0.75 % of full scale Limit of quantitation LOQ 2.5 % of full scale Short term drift 0.5 %/day Repeatability for 20 % of full scale 0.4 % Limit of resolution for 20 % of full scale(LDC) 1.1 % Systematic measured error for 20 % of full scale (BIAS) 0.4 % Repeatability for 80 % of full scale 1.6 % Limit of resolution for 80 % of full scale(LDC) 4.6 % Systematic measured error for 80 % of full scale (BIAS) 2.4 % Pressure 2 to 4 bar (29 to 58 psi) Connection 6/8 mm DN or G3/8 threaded connection
  • 7. TOCII CA72TOC Endress+Hauser 7 Environment Ambient temperature 5 to 35 °C (41 to 95 °F) Humidity 10 to 90 %, non-condensing Ingress protection IP 54 Process Sample temperature 0 to 40 °C (32 to 104 °F) Sample flow rate 20 ml/min (0.32 US gal/hr) Sample volume 90 ml (3 fl.oz) Sample preparations PA-2, PA-3 and PA-9 are available. Sample consistency Aqueous Flammable substances are not allowed in ignitable concentrations - in such cases a sample dilution is required. Sample inlet Unpressurized outlet of the sample processor to the analyzer
  • 8. TOCII CA72TOC 8 Endress+Hauser Mechanical construction Dimensions a0012435 Dimensions Weight approx. 75 kg (165 lbs) Materials 400 (15.8) 600 (23.6) 785 (30.9) 1010(39.8) 1558(61.3) mm (inch) 50 (1.97) Housing Aluminium, powder coated (RAL7035) Window Glass, conductive coated Valve seals EPDM, FPM, FFKM Pump hoses Ismapren Pumps and pump seals PTFE, FFKM Reagents- und sample hoses PTFE, PE Exhaust hoses PTFE, PE Drain hoses PTFE
  • 9. TOCII CA72TOC Endress+Hauser 9 Human interface Display and operating elements a0007996 Display and operating elements 1 Diaplay, 16 rows a 40 characters 2 Keyboard 3 Arrow keys (cursors) 4 Function key"Measuring" 5 Function key "Service" 6 Function key "Programming" 7 Function key "Help" 8 Key "Enter" 9 Key "Clear" Certificates and approvals 4 symbol Declaration of conformity The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives. The manufacturer confirms successful testing of the product by affixing the 4 symbol. 1 2 3 4 5 6 7 89
  • 10. TOCII CA72TOC 10 Endress+Hauser Ordering information Product structure Note! To complete your order code, simply add the optional features to the end of order code. If you have any questions, please contact your local sales office. Measuring range A 0.25 to 600 mg/l TOC B 1 to 2400 mg/l TOC C 2.5 to 6000 mg/l TOC D 5 to 12000 mg/l TOC Y Special version according to customer specification Sample transfer 1 1 measuring point 2 2 measuring points; time controlled 3 2 measuring points; external controlled Operation mode A Batch mode Power supply 0 230 VAC, 50 Hz 1 115 VAC, 60 Hz 2 115 VAC, 50 Hz 3 230 VAC, 60 Hz 9 Special version according to customer specification Sample preparation A 1 x PA-2 PVC, 1 to 8 m3 /h (4.4 to 35 gpm) wastewater (not with CA72TOC-*2*********) B 1 x PA-3 PVC, 0.1 to 1.0 m3 /h (0.4 to 4.4 gpm) wastewater (not with CA72TOC-*2*********) C 1 x PA-9 PP, 0.1 to 1.0 m3 /h (0.4 to 4.4 gpm) wastewater (not with CA72TOC-*2*********) D 2 x PA-2 PVC, 1 to 8 m3 /h (4.4 to 35 gpm) wastewater (not with CA72TOC-*1*********) E 2 x PA-3 PVC, 0.1 to 1.0 m3 /h (0.4 to 4.4 gpm) wastewater (not with CA72TOC-*1*********) F 2 x PA-9 PVC, 0.1 to 1.0 m3/h (0.4 to 4.4 gpm) wastewater (not with CA72TOC-*1*********) Y Special version according to customer specification Language instrument, documentation 1 German 2 English CA72TOC- Order code Carrying gas (optinal, select one option only) E1 CO2-scrubber E2 CO2-scrubber + compressor Analyzer mounting (optinal, select one option only) F1 Wall mounted F2 Stand Accessory mounted (optinal, multiple selection possible) G1 Sample predilution G4 Heated salt trap G5 Pipe back flushing G6 Standby mode PA-2/-3 failure G7 sample valve for aggressive media Accessory enclosed (optinal, multiple selection possible) H1 230 VAC furnace conditioning station (not with CA72TOC-***1******* and not with CA72TOC-***2*******) H2 115 VAC furnace conditioning station (not with CA72TOC-***0******* and not with CA72TOC-***3*******) H3 Standard furnace (not with CA72TOC-********G4**) H4 Furnace for heated salt trap
  • 11. TOCII CA72TOC Endress+Hauser 11 Scope of delivery The scope of delivery comprises: • 1 analyzer CA72TOC • Tool kit for leak check • Tool kit to remove the glass ball and the media • Acid filter kit • Kit for commisioning of stripper and separation chamber • Maintenance kit for furnace • Hose kit for inlet and outlet • 1 canister 5 liter • 2 canisters 2 liter • 3 covers for canisters • Key for door lock • Connection box EMV • Measuring glas 10 ml • Sponge • Safety glasses • Gloves, acid-proof and base proof • Silicone • Operating Instructions Accessories Reagents • CAY450-V10AAE, 1000 ml stripper reagent für CA72TOC • CAY451-V10C01AAH, 1000 ml parent solution 10 000 mg/l TOC • CAY451-V10C10AAH, 1000 ml parent solution100 000 mg/l TOC Furnace conditioning station • 230 V AC 50/60 Hz order number: 71103492 • 115 V AC 50/60 Hz order number: 71103493 Standard furnace • As additional furnace or als replacement • Not for heated salt trap • Order number: 71091188 Furnace for heated salt trap • As additional furnace or as replacement • Order number: 71103888 Kit CA72TOC heated salt trap • As replacement for maintenance (decreases the maintenance time) or as replacement • Order number: 71101532
  • 12. Instruments International Endress+Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel.+41 61 715 81 00 Fax+41 61 715 25 00 www.endress.com info@ii.endress.com TI448C/07/en/10.09 71098868 Printed in Germany / FM+SGML 6.0 / DT