1. Technical Information
BOD Measurement BIOX
Continuous on-line analyzer for BOD measurement in industrial
and municipal wastewater treatment plants
Application
Measurement in wastewater liquids:
Detection of product loss
Detection of BOD load peaks
Reduction of electricity and operating costs
Increase in operational safety
Your benefits
True continuous measurement
Response time of 3 to 15 minutes
Measurement range of 5 to 100,000 mg/l BOD
Sample preparation with self-cleaning coarse filter
Programmable concentration alarms
Limit value monitoring
Up to 14 days historical data can be recalled and displayed
graphically
BOD load curve and current value on LCD screen
•
•
•
•
•
•
•
•
•
•
•
•
TI806C/24/ae
2. Technical data, BIOX-1010
Detection limit: 5 mg BOD/l
Range: 20 to 1000 mg BOD/l
20 to 100,000 mg/l BOD with high concentration option
BIOX-1015 range: 20 to 100,000 mg/l BOD
Detection limit: 5 mg/l BOD
Reproducibility: 3%
T90
time: 3 to 15 minutes
Dimensions: 25.4” x 39.8” x 15.7” (645 x 1010 x 400 mm)
Weight: 143 lbs (65 kg)
Power supply: 230 VAC, 50/60 Hz
Current usage: Average, 3.6 A; maximum 8.15 A
Enclosure: IP 54 (similar to NEMA 3)
Protection class: 1 per DIN 40050, EN 50022)
EMC emission: EN 50081-1, 12/96
EMC noise immunity: EN 50082-2, 12/96
Certification: CE Mark (declaration of conformity)
Test method: Per DIN/VDE 0701
Display: LCD grapics display, 16 lines, 40 characters, backlit
Keypad: 21 keys
Data presentation: 6-hour graphic (14-day scrollable); current value with 5-digit resolution
Data storage: 14 days in RAM; 90 days with diskette drive (data interval 2 minutes)
Monitoring: Logging of warnings, malfuncitons, limit value alarms and calibrations for the last 4 weeks (90 days with
diskette drive), alerts for leakage, deficient dilution water, deficient wastewater, incorrect reactor
temperature, incorrect oxygen reading, broken membrane on oxygen sensor, decreasing electrode slop
Analog outputs: 0/4 to 20 mA selectable, electrically isolated load max.; 500 ohm, 10 V
Limit value alarms: Dry contact relay (normally closed, max. 0.2A and 50 V) for high limit, low limit and slope
Fault signal: Dry contact relay (normally closed, max. 0.2 A and 50 V)
Diskette drive: PC-compatible 3.5” diskete drive for data storage
Computer interface: RS 232C for data output and remote maintenance
3. Dilution water
Wastewater
Wastewater is continuously pumped through a sample bypass. The built-in
pump continuously feeds a small stream of wastewater from the bypass to
the bioreactor. Before reaching the bioreactor, the wastewater stream is
diluted with oxygen-saturated dilution water. Inside the bioreactor, micro-
organisms grow inside small plastic cylinders where they are protected
against mechanical abrasion caused by turbulent mixing. The respiration
rate of the microbial population is automatically maintained at a constant
level by a feedback loop that varies the dilution ratio. Increasing BOD
concentration in the wastewater increases the respiration rate, which in
turn increases the dilution rate. Similarly, if the BOD concentration level
of the wastewater decreases, so does the dilution rate.
As the oxygen usage varies, the volume of wastewater pumped into the
chamber is controlled so as to maintain the target rate of 3 mg/liter O2
consumption.
The mixing ratio of wastewater and dilution water is used to calculate the
BOD parameter.
Ordering Information
The BIOX analyzer use a turbulent-bed bioreactor to continuously measure BOD or toxicity.
Please use the Water Quality Analyzer Application Questionnaire to assess water characteristics and analyzer requirements.
To receive a complete, operational STIP analyzer system, the order must include:
The correct analyzer model and measurement range.
A sample prep/bypass system (see Sample Prep/accessories section)
Either a free-standing rack, or a hinged bracket for wall mounting
Any other options needed for the specific site
NOTE: Startup is REQUIRED and must be provided by Endress+Hauser service group or an authorized service provider.
SAP Material # Order code Description
71019633 10008002 BIOX-1010 BOD analyzer (20 to 1000 mg/L BOD)
71019816 C0008003 BIOX-1015 BOD analyzer (5 to 100,000 ppm BOD) includes nutrient and substrate
dosing, second oxygen probe, high concentration range capability.
Sample Preparation
71013798 20807601 Filtered sample bypass system PA-2 (5 to 35 GPM sample flow stream). Inlet and
outlet have 45° elbow, for cart-mounted STIP analyzer.
71013804 20807603 Filtered sample bypass system PA-2 (5 to 35 GPM sample flow stream, filtered to
solids < 0.5 cm). Straight inlet and outlet, for wall mounted STIP analyzer.
71013806 20807605 Filtered sample bypass system PA-3 (0.1 to 5 GPM sample flow stream, filtered to
solids < 0.5 cm). Inlet and outlet have 45° elbow, for cart-mounted STIP analyzer.
71013805 20807604 Filtered smaple bypass system PA-3 (0.1 to 5 GPM sample flow stream, filtered to
solids < 0.5 cm). Straight inlet and outlet, for wall mounted STIP analyzer.
71013812 20807609 Filtered sample bypass system PA-8 (5 to 35 GPM sample flow stream, filtered to
solids < 0.5 cm). Stainless steel for hi-temperature/hi-pressure sample, for STIP
analyzer.
71013646 20487704 Separation system for PA’s, prevents sample from flowing back into tap water system.
71013702 20687601 Sample reservoir chamber 1010 and EZ-TOC
71013604 20037601 Sample cooler BIOX
Mounting Options
71013631 20407601 Analyzer stand on wheels (movable cart)
71013640 20407605 Analyzer stand without wheels, height adjustable
71013637 20407602 Hinged bracket for wall mounting STIP analyzer
•
•
•
•