Optometrist
“....far more useful and far
more used” than any other
lenses in clinical refraction.
-Dr. Edward Jackson
JCC(Sharma IP)
Objective
To understand the optics and proper use of
Jackson Crossed-Cylinder in clinical practise.
JCC(Sharma IP)
Contents
1. Introduction to JCC
2. Historical Perspective
3. Optics – How is it made?
4. The choice of JCC
5. Detection of astigmatism
6. Refinement of axis
7. Refinement of power
8. Sources of Error
9. Points to ponder-Summary
10. Conclusion
11. Reference
JCC(Sharma IP)
Introduction to JCC
Jackson Crossed-Cylinder is a
combination of two cylinders of
equal strength but of opposite signs
placed with their axis at 90 degrees
to each other and mounted in a
handle.
Jackson Crossed-Cylinder (JCC)
technique is also called the flip-
cross technique.
JCC(Sharma IP)
Historical Perspective
1849-The original concept of crossed cylinders was
described by Stokes.
1855-The Stokes lens was used in a variation of the
present technique by Dennet
Crisp brought it to worldwide attention, and it has
become known as the JCC technique.
JCC(Sharma IP)
Contd.....
The present technique
-designed and described by
Dr. Edward Jackson for the
determination of cylinder
power in 1887 and for axis in
1907.
JCC(Sharma IP)
Optics- How is it made?
A typical JCC lens is a spherocylindrical lens having
a spherical power component combined with a cylinder
power component of twice the power of the sphere, and
of opposite sign.
Eg: +0.50 DS combined with -1.00 DC. This results in
a net power of +0.50 DC in one axis and -0.50 DC in
the other axis(+-50 DC).
+0.50/-1.00 @90
JCC(Sharma IP)
Axis for + cyl
(+0.25@90)
Power: +0.25DC
+ =
+0.25DS/-0.50DCx180
Axis for – cyl
(-0.25@180)
Power: -0.25DC
Contd...
JCC(Sharma IP)
Contd...
Crossed cylinders of +0.25 DS combined with -0.50 DC
(+-25 DC) or +0.37 DS combined with -0.75 DC (to.37
DC), etc.,are available.
Thus, the two principal axes of a crossed-cylinder
lens exhibit equal cylinder power of opposite signs.
JCC(Sharma IP)
Marking of Principle meridian
The principal
meridians are marked
in the periphery of
JCC lens
In the UK, it is the
opposite.
JCC(Sharma IP)
White dots = axis of the plus
cylinder
Red dots = axis of the minus
cylinder
Handle of JCC
A handle is attached between the two marked axes, which
enables the lens to be "twirled" before the eye by rotation of
the handle.
In this manner, the positions of the minus and plus axes are
interchanged rapidly and alternately.
The common term used for the rotating the handle is known
as flipping.
Hence, the JCC is often termed as flip-cross cylinder.
JCC(Sharma IP)
Contd....
0.5 D Jackson cross-cylinder in
primary orientation. The handle
is down to the right and at 45º to
the horizontal. The label +.50 is
in the usual orientation for
reading; -.50 reads upward. The
markings are in red (as here) and
white (shown black here).
JCC(Sharma IP)
JCC(Sharma IP)
In almost all modern phoropters, the rotation of the JCC lens is automated
Choice of JCC
Vision 6/9 or better: use 0.25DC x-cyl
If results unreliable, then change up to 0.50 x-cyl. and see if
more reliable
Vision 6/12 or worse: use 0.50DC x-cyl
If results reliable and vision improves, change down to 0.25
Use a larger target until the vision improves!
Vision 6/24 or worse: try 0.75DC x-cyl
If results unreliable, use alternative method of astigmatic
correction (Astigmtic Fan, keratometry etc)
JCC(Sharma IP)
BE
Today’s
topic
JCC(Sharma IP)
Procedure
JCC(Sharma IP)
Starting point for JCC
After retinoscopy.....adjust sphere.
End point of spherical adjustment is the starting point
of JCC refinement.
Circle of least confusion must be on retina (ILM), so
check sphere first.
Remember Strums Conoid
JCC(Sharma IP)
Eg: Simple myopic astigmatism
Interval of Sturm
Circle of Least Confusion
Blur is due to combination of…
CLC in front of the retina
Focal lines being separated
“It’s very blurred”
With best spherical correction
Circle of Least Confusion
Has moved, is now on the retina
Interval of Sturm
Length unchanged
Reason the vision is still blurred
“That’s better but it still isn’t clear”
JCC(Sharma IP)
1.Discovery of Astigmatism
JCC(Sharma IP)
If astigmatism is present
But WHY
- because the correct axis can be found in the presence
of an incorrect power but the full cylindrical power
will not be found in the presence of an incorrect axis.
JCC(Sharma IP)
2. Refinement of Axis
JCC(Sharma IP)
Procedure
JCC(Sharma IP)
Contd......
JCC(Sharma IP)
Example
You have performed retinoscopy on a patient
Retinoscopy value
-1.00DS/-1.00DCx180, 6/9
You have checked the sphere power
Now lets refine axis of the cyl.
JCC(Sharma IP)
Position 1
Resultant axis
-1.00
JCC(Sharma IP)
“ Clear now”
Resultant axis
Position 1
JCC(Sharma IP)
Refining axis
Patient response
“Lens position 1 was clearer”
So, we rotate the cylinder lens towards the position of
the red markings
In this case, position 1
Initially, move by steps of about 15 deg, then use
smaller steps as we get closer
JCC(Sharma IP)
Patient response
We rotated the axis of the trial lens AND the handle of the x-
cyl by 20 deg
“Both lenses are equally blurred”
This means that the cyl axis of the trial frame now matches
the patient’s cyl axis
The true axis is 160 deg
In real life, you would continue until the patient sends us in
the other direction (reversal)
There is usually a range where the images appear equal, and we
need to find the limits
Choose the axis mid-way between the two reversals
JCC(Sharma IP)
“ They are same”
May be on axis, therefore move cyl axis by about 20deg and check to
see if it returns
May be within range of uncertainty (next slide)
0.25DC JCC may give insufficient difference
Try 0.50DC
0.50DC JCC may give too much distortion
Move down to 0.25DC
If none of the above help, use alternative technique.
JCC(Sharma IP)
Range of uncertainty
In real life, most patients will report that both lenses are equally
blurred over a range of axes
This is more common with low cyl power
You need to identify the range
Find where the patient tells you to rotate in the
opposite direction at each end
Select the axis in the middle of the range
JCC(Sharma IP)
How much axis to move?
Del Priore and Guyton
gave the guidelines
suggested in Table 20-2
for the initial change in
correcting axis position
relative to power of the
correcting cylinder while
checking the axis.
JCC(Sharma IP)
Source:William J Benjamin,2006, Borish’s Clinical Refraction, Butterworth Heineman Elsevier. 20: 818
2. Refinement of Power
JCC(Sharma IP)
Optical principle of power
refinement
When determining the power, JCC will either increase
or decrease residual cyl, either expanding or collapsing
the astigmatic interval and circle of least confusion
Thereby making the target less or more clear
JCC(Sharma IP)
Option 1 Circle of Least Confusion
Increases in size
Does not change position!
Interval of Sturm
Longer
“That looks awful”
JCC(Sharma IP)
Option 2 Circle of Least Confusion
Decreases in size
Does not change position!
Interval of Sturm
Shorter
“That is much better”
JCC(Sharma IP)
Procedure
JCC(Sharma IP)
3. JCC is placed with its axis parallel to the axis of the cylinder in the trial frame.
Contd...
JCC(Sharma IP)
Example
Using ± 0.50 JCC
Retioscopy : -1.00DS/-0.75 DC X 120
If the patient prefers the lens
When red marks are aligned with trial cyl axis (120
deg),
add -0.50 DC
When white marks are aligned with trial cyl axis (120),
reduce -0.50DC
Equally clear: you have the right power
JCC(Sharma IP)
Contd.....
JCC(Sharma IP)
 For each -0.50DC change, you need to add +0.25DS, to keep the circle of
least confusion on the retina
 Add -0.25DS for each +0.50DC change
So in the example : -1.00DS/-0.75 DC X 120
If the patient prefers red marks
Final power is:
-0.75DS/ -1.25 DC X 120
Refining power
JCC(Sharma IP)
Add – 0.50
Deduct – 0.50
Sources of Error
JCC(Sharma IP)
Common errors
1.Not keeping the circle of least confusion on the retina
Starting with the wrong sphere power
Forgetting to change sphere power if cyl is changed by
0.50DC or more
2.Assuming the axis is correct if the patient says “they
look the same” without checking
Could be no astigmatism at all
Could be 90deg off
3.Incorrect presentation time – esp too quick
4.Poor alignment of JCC and trial frame axis
JCC(Sharma IP)
Points to ponder- Summary
JCC is always a sphero-cylindrical lens such that one meridian is plus
power and the other meridian is of equal minus power.
 The red dots identify the axis of the minus power and the white dot
is of plus power.
While refining axis: JCC handle is parallel to trail lens cylinder axis
While refining power: JCC lens axis is parallel to trail lens cylinder
axis.
When patient says “ both sides are same” exclude all posibilities.
JCC(Sharma IP)
Conclusion
The most accepted procedure used by the
overwhelming majority of examiners is the JCC
technique.
Freeman and Purdum and Goar considered the JCC to
be the most delicate test for astigmatism.
The use of JCC is very important in clinical parctise of
refraction and an it serves as an important instrument
for optometrist.
JCC(Sharma IP)
Reference
1. William J Benjamin,2006, Borish’s Clinical Refraction, Butterworth Heineman
Elsevier. 20: 816-829
2. Wunsch SE. 1971. The cross cylinder. Int Ophthalmol Clin 11:131-153.
3.Brookman KE. 1993. The Jackson cross-cylinder: Historical perspective. JAm Optom
Assoc 64:329-331.
4. Crisp WHo 1943. Photographing cross cylinder tests. Am J Opht 26:758-760.
5. Duke Elder, 1998. Practise of Clinical Refraction, Butterworth Heineman Elsevier. 4:
181-183
6. Khurana AK 2013. Optics and refraction, 6: 133-134
7. Perlstein SII. 1982. Mounted cross-cylinder: A new mounted Jackson cross-
cylinder. Ann Opht/Ullmol 14:992.
8..Del Priore LV, Guyton DL. 1986. The Jackson cross cylinder, a reappraisal.
Ophthalmology 93: 1461-1465.
9. Sims CN, Durham DG. 1986. The Jackson cross-cylinder disproved. 'hans Am
Ophthailltol Soc 84:355-386.
JCC(Sharma IP)
Orthoptics as a career (ZG)
Blog: optomindra@blogspot.com

Contenu connexe

Tendances

Ophthalmic Prisms: Prismatic Effects and Decentration
Ophthalmic Prisms: Prismatic Effects and DecentrationOphthalmic Prisms: Prismatic Effects and Decentration
Ophthalmic Prisms: Prismatic Effects and DecentrationRabindraAdhikary
 
Dynamic retinoscopy
Dynamic retinoscopy Dynamic retinoscopy
Dynamic retinoscopy Rajeshwori
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refractionAzizul Islam
 
Simple & Toric Transposition
Simple & Toric TranspositionSimple & Toric Transposition
Simple & Toric TranspositionAzizul Islam
 
Maddox rod and double maddox rod
Maddox rod and double maddox rodMaddox rod and double maddox rod
Maddox rod and double maddox rodAnuMusyakhwo7
 
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...Bikash Sapkota
 
Special purpose frames
Special purpose framesSpecial purpose frames
Special purpose framesEyenirvaan
 
ARC: abnormal retinal correspondence, eccentric fixation
ARC: abnormal retinal correspondence, eccentric fixationARC: abnormal retinal correspondence, eccentric fixation
ARC: abnormal retinal correspondence, eccentric fixationaditi Jain
 
Hess chart, diplopia chart, cover tests
Hess chart, diplopia chart, cover testsHess chart, diplopia chart, cover tests
Hess chart, diplopia chart, cover testsNikhil Rp
 
Corneal topography
Corneal topographyCorneal topography
Corneal topographySatish Jeria
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopyHira Dahal
 
Binocular refraction techniques, binocular balancing
Binocular refraction techniques, binocular balancing Binocular refraction techniques, binocular balancing
Binocular refraction techniques, binocular balancing Mohammad Arman Bin Aziz
 

Tendances (20)

Ophthalmic Prisms: Prismatic Effects and Decentration
Ophthalmic Prisms: Prismatic Effects and DecentrationOphthalmic Prisms: Prismatic Effects and Decentration
Ophthalmic Prisms: Prismatic Effects and Decentration
 
Dynamic retinoscopy
Dynamic retinoscopy Dynamic retinoscopy
Dynamic retinoscopy
 
Worth 4 dot test
Worth 4 dot testWorth 4 dot test
Worth 4 dot test
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Maddox rod n wing
Maddox rod n wingMaddox rod n wing
Maddox rod n wing
 
Simple & Toric Transposition
Simple & Toric TranspositionSimple & Toric Transposition
Simple & Toric Transposition
 
Stereopsis
Stereopsis  Stereopsis
Stereopsis
 
Maddox rod and double maddox rod
Maddox rod and double maddox rodMaddox rod and double maddox rod
Maddox rod and double maddox rod
 
Hvid
Hvid Hvid
Hvid
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
 
Lensometers
LensometersLensometers
Lensometers
 
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
 
Special purpose frames
Special purpose framesSpecial purpose frames
Special purpose frames
 
ARC: abnormal retinal correspondence, eccentric fixation
ARC: abnormal retinal correspondence, eccentric fixationARC: abnormal retinal correspondence, eccentric fixation
ARC: abnormal retinal correspondence, eccentric fixation
 
Hess chart, diplopia chart, cover tests
Hess chart, diplopia chart, cover testsHess chart, diplopia chart, cover tests
Hess chart, diplopia chart, cover tests
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Binocular refraction techniques, binocular balancing
Binocular refraction techniques, binocular balancing Binocular refraction techniques, binocular balancing
Binocular refraction techniques, binocular balancing
 
Aniseikonia
AniseikoniaAniseikonia
Aniseikonia
 
Testing for npa
Testing for npaTesting for npa
Testing for npa
 

Similaire à JCC -Jackson Cross Cylinder

Similaire à JCC -Jackson Cross Cylinder (20)

Jackson Cross Cylinder
Jackson Cross CylinderJackson Cross Cylinder
Jackson Cross Cylinder
 
JACKSON CROSS CYLINDER
JACKSON CROSS CYLINDERJACKSON CROSS CYLINDER
JACKSON CROSS CYLINDER
 
Technique of Jackson cross cylinder
Technique of Jackson cross cylinderTechnique of Jackson cross cylinder
Technique of Jackson cross cylinder
 
Cross Cylinder 4th Yr.pptx
Cross Cylinder 4th Yr.pptxCross Cylinder 4th Yr.pptx
Cross Cylinder 4th Yr.pptx
 
Retinoscopy.pptx
Retinoscopy.pptxRetinoscopy.pptx
Retinoscopy.pptx
 
Subjective refraction by optom praveen
Subjective refraction by optom praveenSubjective refraction by optom praveen
Subjective refraction by optom praveen
 
subjective verification of refraction
subjective verification of refractionsubjective verification of refraction
subjective verification of refraction
 
Methods of subjective refraction
Methods of subjective refractionMethods of subjective refraction
Methods of subjective refraction
 
FLACS ie Femtosecond Laser Assisted Cataract Surgery - Unbiased Review..
FLACS ie Femtosecond Laser Assisted Cataract Surgery - Unbiased Review..FLACS ie Femtosecond Laser Assisted Cataract Surgery - Unbiased Review..
FLACS ie Femtosecond Laser Assisted Cataract Surgery - Unbiased Review..
 
Iol
IolIol
Iol
 
lecture(3) phoropter.ppt
lecture(3) phoropter.pptlecture(3) phoropter.ppt
lecture(3) phoropter.ppt
 
Biometry
Biometry Biometry
Biometry
 
Static retinoscope presentation by Tamalika
Static retinoscope presentation by TamalikaStatic retinoscope presentation by Tamalika
Static retinoscope presentation by Tamalika
 
PRESCRIPTION & TRIAL SET.pptx
PRESCRIPTION & TRIAL SET.pptxPRESCRIPTION & TRIAL SET.pptx
PRESCRIPTION & TRIAL SET.pptx
 
Biometry made easy
Biometry made easy Biometry made easy
Biometry made easy
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Jcc .pptx
Jcc .pptxJcc .pptx
Jcc .pptx
 
Errors 2
Errors 2Errors 2
Errors 2
 
Online femto ppt (no video)
Online femto ppt (no video)Online femto ppt (no video)
Online femto ppt (no video)
 
Astigmatism
AstigmatismAstigmatism
Astigmatism
 

Plus de Indra Prasad Sharma

Plus de Indra Prasad Sharma (14)

Impact of myopia and high myopia
Impact of myopia and high myopiaImpact of myopia and high myopia
Impact of myopia and high myopia
 
Eye health Promotion using technology
Eye health Promotion using technologyEye health Promotion using technology
Eye health Promotion using technology
 
Research Hypothesis
Research HypothesisResearch Hypothesis
Research Hypothesis
 
OCT in Ophthalmology
OCT in OphthalmologyOCT in Ophthalmology
OCT in Ophthalmology
 
Blepharitis : Overview and management
Blepharitis : Overview and managementBlepharitis : Overview and management
Blepharitis : Overview and management
 
Slit lamp in Ophthalmology
Slit lamp in OphthalmologySlit lamp in Ophthalmology
Slit lamp in Ophthalmology
 
Challenging Refraction
Challenging RefractionChallenging Refraction
Challenging Refraction
 
Xanthalesma
Xanthalesma Xanthalesma
Xanthalesma
 
Basics of binocular vision
Basics of binocular visionBasics of binocular vision
Basics of binocular vision
 
Fluorescein Stain
Fluorescein Stain Fluorescein Stain
Fluorescein Stain
 
Pinguecula - An overview
Pinguecula - An overviewPinguecula - An overview
Pinguecula - An overview
 
Measuring interpupillary distance
Measuring interpupillary distanceMeasuring interpupillary distance
Measuring interpupillary distance
 
Prescription for ametropias
Prescription for ametropiasPrescription for ametropias
Prescription for ametropias
 
Ectopia lentis and its low vision management
Ectopia lentis and its low vision managementEctopia lentis and its low vision management
Ectopia lentis and its low vision management
 

Dernier

COVID-19 (NOVEL CORONA VIRUS DISEASE PANDEMIC ).pptx
COVID-19  (NOVEL CORONA  VIRUS DISEASE PANDEMIC ).pptxCOVID-19  (NOVEL CORONA  VIRUS DISEASE PANDEMIC ).pptx
COVID-19 (NOVEL CORONA VIRUS DISEASE PANDEMIC ).pptxBibekananda shah
 
Presentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPresentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPrerana Jadhav
 
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfLippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfSreeja Cherukuru
 
The next social challenge to public health: the information environment.pptx
The next social challenge to public health:  the information environment.pptxThe next social challenge to public health:  the information environment.pptx
The next social challenge to public health: the information environment.pptxTina Purnat
 
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptxPERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptxdrashraf369
 
Basic principles involved in the traditional systems of medicine PDF.pdf
Basic principles involved in the traditional systems of medicine PDF.pdfBasic principles involved in the traditional systems of medicine PDF.pdf
Basic principles involved in the traditional systems of medicine PDF.pdfDivya Kanojiya
 
Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxDr. Dheeraj Kumar
 
Case Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxCase Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxNiranjan Chavan
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptxDr.Nusrat Tariq
 
LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS  CLASSIFICATIONS.pdfLUNG TUMORS AND ITS  CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS CLASSIFICATIONS.pdfDolisha Warbi
 
Presentation for Bella Mahl 2024-03-28-24-MW-Overview-Bella.pptx
Presentation for Bella Mahl 2024-03-28-24-MW-Overview-Bella.pptxPresentation for Bella Mahl 2024-03-28-24-MW-Overview-Bella.pptx
Presentation for Bella Mahl 2024-03-28-24-MW-Overview-Bella.pptxpdamico1
 
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...Badalona Serveis Assistencials
 
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisVarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisGolden Helix
 
Apiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptApiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptkedirjemalharun
 
Culture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxCulture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxDr. Dheeraj Kumar
 
Measurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxMeasurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxDr. Dheeraj Kumar
 
systemic bacteriology (7)............pptx
systemic bacteriology (7)............pptxsystemic bacteriology (7)............pptx
systemic bacteriology (7)............pptxEyobAlemu11
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAAjennyeacort
 
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdfPULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdfDolisha Warbi
 
maternal mortality and its causes and how to reduce maternal mortality
maternal mortality and its causes and how to reduce maternal mortalitymaternal mortality and its causes and how to reduce maternal mortality
maternal mortality and its causes and how to reduce maternal mortalityhardikdabas3
 

Dernier (20)

COVID-19 (NOVEL CORONA VIRUS DISEASE PANDEMIC ).pptx
COVID-19  (NOVEL CORONA  VIRUS DISEASE PANDEMIC ).pptxCOVID-19  (NOVEL CORONA  VIRUS DISEASE PANDEMIC ).pptx
COVID-19 (NOVEL CORONA VIRUS DISEASE PANDEMIC ).pptx
 
Presentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPresentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous System
 
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfLippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
 
The next social challenge to public health: the information environment.pptx
The next social challenge to public health:  the information environment.pptxThe next social challenge to public health:  the information environment.pptx
The next social challenge to public health: the information environment.pptx
 
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptxPERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
 
Basic principles involved in the traditional systems of medicine PDF.pdf
Basic principles involved in the traditional systems of medicine PDF.pdfBasic principles involved in the traditional systems of medicine PDF.pdf
Basic principles involved in the traditional systems of medicine PDF.pdf
 
Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptx
 
Case Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxCase Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptx
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptx
 
LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS  CLASSIFICATIONS.pdfLUNG TUMORS AND ITS  CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
 
Presentation for Bella Mahl 2024-03-28-24-MW-Overview-Bella.pptx
Presentation for Bella Mahl 2024-03-28-24-MW-Overview-Bella.pptxPresentation for Bella Mahl 2024-03-28-24-MW-Overview-Bella.pptx
Presentation for Bella Mahl 2024-03-28-24-MW-Overview-Bella.pptx
 
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
 
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisVarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
 
Apiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptApiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.ppt
 
Culture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxCulture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptx
 
Measurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxMeasurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptx
 
systemic bacteriology (7)............pptx
systemic bacteriology (7)............pptxsystemic bacteriology (7)............pptx
systemic bacteriology (7)............pptx
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA
 
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdfPULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
PULMONARY EMBOLISM AND ITS MANAGEMENTS.pdf
 
maternal mortality and its causes and how to reduce maternal mortality
maternal mortality and its causes and how to reduce maternal mortalitymaternal mortality and its causes and how to reduce maternal mortality
maternal mortality and its causes and how to reduce maternal mortality
 

JCC -Jackson Cross Cylinder

  • 2. “....far more useful and far more used” than any other lenses in clinical refraction. -Dr. Edward Jackson JCC(Sharma IP)
  • 3. Objective To understand the optics and proper use of Jackson Crossed-Cylinder in clinical practise. JCC(Sharma IP)
  • 4. Contents 1. Introduction to JCC 2. Historical Perspective 3. Optics – How is it made? 4. The choice of JCC 5. Detection of astigmatism 6. Refinement of axis 7. Refinement of power 8. Sources of Error 9. Points to ponder-Summary 10. Conclusion 11. Reference JCC(Sharma IP)
  • 5. Introduction to JCC Jackson Crossed-Cylinder is a combination of two cylinders of equal strength but of opposite signs placed with their axis at 90 degrees to each other and mounted in a handle. Jackson Crossed-Cylinder (JCC) technique is also called the flip- cross technique. JCC(Sharma IP)
  • 6. Historical Perspective 1849-The original concept of crossed cylinders was described by Stokes. 1855-The Stokes lens was used in a variation of the present technique by Dennet Crisp brought it to worldwide attention, and it has become known as the JCC technique. JCC(Sharma IP)
  • 7. Contd..... The present technique -designed and described by Dr. Edward Jackson for the determination of cylinder power in 1887 and for axis in 1907. JCC(Sharma IP)
  • 8. Optics- How is it made? A typical JCC lens is a spherocylindrical lens having a spherical power component combined with a cylinder power component of twice the power of the sphere, and of opposite sign. Eg: +0.50 DS combined with -1.00 DC. This results in a net power of +0.50 DC in one axis and -0.50 DC in the other axis(+-50 DC). +0.50/-1.00 @90 JCC(Sharma IP)
  • 9. Axis for + cyl (+0.25@90) Power: +0.25DC + = +0.25DS/-0.50DCx180 Axis for – cyl (-0.25@180) Power: -0.25DC Contd... JCC(Sharma IP)
  • 10. Contd... Crossed cylinders of +0.25 DS combined with -0.50 DC (+-25 DC) or +0.37 DS combined with -0.75 DC (to.37 DC), etc.,are available. Thus, the two principal axes of a crossed-cylinder lens exhibit equal cylinder power of opposite signs. JCC(Sharma IP)
  • 11. Marking of Principle meridian The principal meridians are marked in the periphery of JCC lens In the UK, it is the opposite. JCC(Sharma IP) White dots = axis of the plus cylinder Red dots = axis of the minus cylinder
  • 12. Handle of JCC A handle is attached between the two marked axes, which enables the lens to be "twirled" before the eye by rotation of the handle. In this manner, the positions of the minus and plus axes are interchanged rapidly and alternately. The common term used for the rotating the handle is known as flipping. Hence, the JCC is often termed as flip-cross cylinder. JCC(Sharma IP)
  • 13. Contd.... 0.5 D Jackson cross-cylinder in primary orientation. The handle is down to the right and at 45º to the horizontal. The label +.50 is in the usual orientation for reading; -.50 reads upward. The markings are in red (as here) and white (shown black here). JCC(Sharma IP)
  • 14. JCC(Sharma IP) In almost all modern phoropters, the rotation of the JCC lens is automated
  • 15. Choice of JCC Vision 6/9 or better: use 0.25DC x-cyl If results unreliable, then change up to 0.50 x-cyl. and see if more reliable Vision 6/12 or worse: use 0.50DC x-cyl If results reliable and vision improves, change down to 0.25 Use a larger target until the vision improves! Vision 6/24 or worse: try 0.75DC x-cyl If results unreliable, use alternative method of astigmatic correction (Astigmtic Fan, keratometry etc) JCC(Sharma IP)
  • 18. Starting point for JCC After retinoscopy.....adjust sphere. End point of spherical adjustment is the starting point of JCC refinement. Circle of least confusion must be on retina (ILM), so check sphere first. Remember Strums Conoid JCC(Sharma IP)
  • 19. Eg: Simple myopic astigmatism Interval of Sturm Circle of Least Confusion Blur is due to combination of… CLC in front of the retina Focal lines being separated “It’s very blurred”
  • 20. With best spherical correction Circle of Least Confusion Has moved, is now on the retina Interval of Sturm Length unchanged Reason the vision is still blurred “That’s better but it still isn’t clear” JCC(Sharma IP)
  • 22. If astigmatism is present But WHY - because the correct axis can be found in the presence of an incorrect power but the full cylindrical power will not be found in the presence of an incorrect axis. JCC(Sharma IP)
  • 23. 2. Refinement of Axis JCC(Sharma IP)
  • 26. Example You have performed retinoscopy on a patient Retinoscopy value -1.00DS/-1.00DCx180, 6/9 You have checked the sphere power Now lets refine axis of the cyl. JCC(Sharma IP)
  • 29. Refining axis Patient response “Lens position 1 was clearer” So, we rotate the cylinder lens towards the position of the red markings In this case, position 1 Initially, move by steps of about 15 deg, then use smaller steps as we get closer JCC(Sharma IP)
  • 30. Patient response We rotated the axis of the trial lens AND the handle of the x- cyl by 20 deg “Both lenses are equally blurred” This means that the cyl axis of the trial frame now matches the patient’s cyl axis The true axis is 160 deg In real life, you would continue until the patient sends us in the other direction (reversal) There is usually a range where the images appear equal, and we need to find the limits Choose the axis mid-way between the two reversals JCC(Sharma IP)
  • 31. “ They are same” May be on axis, therefore move cyl axis by about 20deg and check to see if it returns May be within range of uncertainty (next slide) 0.25DC JCC may give insufficient difference Try 0.50DC 0.50DC JCC may give too much distortion Move down to 0.25DC If none of the above help, use alternative technique. JCC(Sharma IP)
  • 32. Range of uncertainty In real life, most patients will report that both lenses are equally blurred over a range of axes This is more common with low cyl power You need to identify the range Find where the patient tells you to rotate in the opposite direction at each end Select the axis in the middle of the range JCC(Sharma IP)
  • 33. How much axis to move? Del Priore and Guyton gave the guidelines suggested in Table 20-2 for the initial change in correcting axis position relative to power of the correcting cylinder while checking the axis. JCC(Sharma IP) Source:William J Benjamin,2006, Borish’s Clinical Refraction, Butterworth Heineman Elsevier. 20: 818
  • 34. 2. Refinement of Power JCC(Sharma IP)
  • 35. Optical principle of power refinement When determining the power, JCC will either increase or decrease residual cyl, either expanding or collapsing the astigmatic interval and circle of least confusion Thereby making the target less or more clear JCC(Sharma IP)
  • 36. Option 1 Circle of Least Confusion Increases in size Does not change position! Interval of Sturm Longer “That looks awful” JCC(Sharma IP)
  • 37. Option 2 Circle of Least Confusion Decreases in size Does not change position! Interval of Sturm Shorter “That is much better” JCC(Sharma IP)
  • 38. Procedure JCC(Sharma IP) 3. JCC is placed with its axis parallel to the axis of the cylinder in the trial frame.
  • 40. Example Using ± 0.50 JCC Retioscopy : -1.00DS/-0.75 DC X 120 If the patient prefers the lens When red marks are aligned with trial cyl axis (120 deg), add -0.50 DC When white marks are aligned with trial cyl axis (120), reduce -0.50DC Equally clear: you have the right power JCC(Sharma IP)
  • 41. Contd..... JCC(Sharma IP)  For each -0.50DC change, you need to add +0.25DS, to keep the circle of least confusion on the retina  Add -0.25DS for each +0.50DC change So in the example : -1.00DS/-0.75 DC X 120 If the patient prefers red marks Final power is: -0.75DS/ -1.25 DC X 120
  • 42. Refining power JCC(Sharma IP) Add – 0.50 Deduct – 0.50
  • 44. Common errors 1.Not keeping the circle of least confusion on the retina Starting with the wrong sphere power Forgetting to change sphere power if cyl is changed by 0.50DC or more 2.Assuming the axis is correct if the patient says “they look the same” without checking Could be no astigmatism at all Could be 90deg off 3.Incorrect presentation time – esp too quick 4.Poor alignment of JCC and trial frame axis JCC(Sharma IP)
  • 45. Points to ponder- Summary JCC is always a sphero-cylindrical lens such that one meridian is plus power and the other meridian is of equal minus power.  The red dots identify the axis of the minus power and the white dot is of plus power. While refining axis: JCC handle is parallel to trail lens cylinder axis While refining power: JCC lens axis is parallel to trail lens cylinder axis. When patient says “ both sides are same” exclude all posibilities. JCC(Sharma IP)
  • 46. Conclusion The most accepted procedure used by the overwhelming majority of examiners is the JCC technique. Freeman and Purdum and Goar considered the JCC to be the most delicate test for astigmatism. The use of JCC is very important in clinical parctise of refraction and an it serves as an important instrument for optometrist. JCC(Sharma IP)
  • 47. Reference 1. William J Benjamin,2006, Borish’s Clinical Refraction, Butterworth Heineman Elsevier. 20: 816-829 2. Wunsch SE. 1971. The cross cylinder. Int Ophthalmol Clin 11:131-153. 3.Brookman KE. 1993. The Jackson cross-cylinder: Historical perspective. JAm Optom Assoc 64:329-331. 4. Crisp WHo 1943. Photographing cross cylinder tests. Am J Opht 26:758-760. 5. Duke Elder, 1998. Practise of Clinical Refraction, Butterworth Heineman Elsevier. 4: 181-183 6. Khurana AK 2013. Optics and refraction, 6: 133-134 7. Perlstein SII. 1982. Mounted cross-cylinder: A new mounted Jackson cross- cylinder. Ann Opht/Ullmol 14:992. 8..Del Priore LV, Guyton DL. 1986. The Jackson cross cylinder, a reappraisal. Ophthalmology 93: 1461-1465. 9. Sims CN, Durham DG. 1986. The Jackson cross-cylinder disproved. 'hans Am Ophthailltol Soc 84:355-386. JCC(Sharma IP)
  • 48. Orthoptics as a career (ZG) Blog: optomindra@blogspot.com

Notes de l'éditeur

  1. The original concept of crossed cylinders was described by Stokes in 1849, who combined cylinders of +4.00 DC and -4.00 DC so that they could be rotated in opposite directions, giving a variety of powers from plano to a +4.00 OS sphere combined with -8.00 DC cylinder. The Stokes lens was used in a variation of the present technique by Dennet in 1855. However, the present technique was first promulgated and described by Jackson for the determination of cylinder power in 1887 and for axis in 1907.
  2. The Jackson Crossed-Cylinder (JCC) lens in position before the phoropter's lens aperture, as would occur during assessment of cylinder power of the right eye. The correcting minus cylinder in the phoropter is at x180. A, The ICC axis of minus cylinder (red dots, Ilrrows) is aligned with that of the correcting cylinder lens. B, The JCC lens has been flipped so as to reverse the positions of the JCC axes, and the JCC axis of plus cylinder (white dots) is aligned with the axis of the correcting cylinder.
  3. The original concept of crossed cylinders was described by Stokes in 1849, who combined cylinders of +4.00 DC and -4.00 DC so that they could be rotated in opposite directions, giving a variety of powers from plano to a +4.00 OS sphere combined with -8.00 DC cylinder. The Stokes lens was used in a variation of the present technique by Dennet in 1855. However, the present technique was first promulgated and described by Jackson for the determination of cylinder power in 1887 and for axis in 1907.