17. MTTT: Mean transit time, DST: Down slope time
Bendjelid K et al. Crit Care. 2012;16(3):R98.
Reuter DA et al. Anesth Analg. 2010;110(3):799-811.
S2
S1
EDTT : Exponential decay time,
S1: maximum ascending slope, S2: maximum descending slope
MTT
DST
Bendjelid K et al. Crit Care. 2010;14(6):R209.
MTTはAUC1=AUC2となるようにtを設定
DSTはEDTTより算出
17
18. GEDV: global end-diastolic volume
PiCCOTM
VolumeViewTM
Reuter DA et al. Anesth Analg. 2010;110(3):799-811.
Bendjelid K et al. Crit Care. 2012 May 30;16(3):R98.
FT= Cardiac Output
ITTV: intrathoracic thermal volume
PTV: pulmonary thermal volume
EVLW: extravascular lung water
18
35. 近年の循環管理モニター機器研究で
頻用される統計について
Critchley LA, Lee A, Ho AM.
A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac
output. Anesth Analg. 2010;111(5):1180-92.
Critchley LA, Yang XX, Lee A.
Assessment of trending ability of cardiac output monitors by polar plot methodology.
J Cardiothorac Vasc Anesth. 2011 Jun;25(3):536-46. PMID: 21419654
35
40. Bland – Altman Plot
比較したい計測機器の値の差と平均を用いて
BiasとPrecisionを評価
比較したい2つの機器で同時に得られた値 A,
B
縦軸:difference(差); A-B
横軸:average (平均); (A+B)/2
+2SD or +1.96SD
Dr. Martin Bland, Dr. Douglas. Altman
Bias: 測定値差の平均
Limits of Agreement (LOA): ±2 SD or ±1.96 SD
LOA
Percentage error (% error): 2SD/Mean Bias
Biasの理想は0
-2SD or -1.96SD
% error
値は正規分布
%error <30%であれば許容される
Critchley LA, Critchley JA. J Clin Monit Comput. 1999;15(2):85-91.
40
41. Four quadrant plot
Bland-Altman Plotの欠点
比較している計測値はある時点のものであるため“変化”を反映させていな
い
値の変化を考慮した評価方法が近年用いられるようになってきた
→ four quadrant plot, polar plot
Four qudrant plotの作成例
横軸: 基準となる機器の値(reference), 縦軸: 評価したい機器の値(test)
Plot A
各測定においてBefore (1,1), Treatment A (3,2), B (4,5), C (7,7)の測定結果を得る
Plot B
Before (1,1)→Treatment A (3,2)の
変化は(3-1,2-1)→(2,1)となる
同様にA→B, B→Cは(1,3), (3,2)と
なり,左図のようなプロットに
なる。
45度の直線を描く
45度の直線と原点-プロット直
線との角度がバラつき,プ
41
ロットと原点との距離が変化
42. Four quadrant plot
2
1
Exclusion zone
3
4
Concordance ratio
第1,3象限にあるプロットの割合
Good: >95%
Marginal: 90-95%
Poor: <90%
Exclusion Zone
変化率が小さいサンプルは過大評価
→検証から除外
通常ΔCOが0.5-1.0 L/min または10-15%
ROC曲線よりAUC 50%となるような値を設
第1象限,第3象限のプロット
↓
2つの機器で同じような値の変化
変化が完全に一致していれば
45度の直線上にある
定
Ex) ΔCO 12%
→右のようなROC曲線
Exclusion Zone: 12%
Monnet X, Anguel N, Jozwiak M et al.
Br J Anaesth. 2012;108(4): 615-22.
42
43. シミュレーションプロット
After excluding central zone data, good trending
ability was associated with concordance rates above 95%, marginal
with 90% to 95%, and poor below 90%.
43
44. Polar plot
Four quadrant plotのベクトルと距離を変換したも
45° の
Plot C
原点からの距離を(reference ΔCO+test ΔCO)/2に短
縮(●→△)
45°の直線を基準にプロットを時計回りに45°回転
第2,3象限を180°回転させる場合もある
30° Plot D
水平線から30°の直線を描く
95%信頼区間が±30°
330° 原点-プロット直線と水平直線との角度(θ)によって
ばらつきを評価できる
44
原点とプロットとの距離を縮小することでばらつきを
50. 1) Kiefer N, Hofer CK, Marx G et al.
Clinical validation of a new thermodilution system for the assessment of cardiac output and
volumetric parameters. Crit Care. 2012 May 30;16(3):R98.
2) Bendjelid K, Marx G, Kiefer N et al.
Performance of a new pulse contour method for continuous cardiac output monitoring: validation in
critically ill patients. Br J Anaesth. 2013 Oct;111(4):573-9. PMID: 23625132
PiCCOTM vs. VolumeViewTM
Setting:
Prospective, multicenter observational study
2010.4 - 2010.12, 4 Hospital, surgical and interdisciplinary ICU
Germany and Switzerland
Patients:
TPTDによる循環管理を必要とする患者72人
1) 370 paired changes / 443 paired boluses
2) 338 matched data / 72hr collected
50
51. 1) Kiefer N, Hofer CK, Marx G et al.
Clinical validation of a new thermodilution system for the assessment of cardiac output and
volumetric parameters. Crit Care. 2012 May 30;16(3):R98.
2) Bendjelid K, Marx G, Kiefer N et al.
Performance of a new pulse contour method for continuous cardiac output monitoring: validation in
critically ill patients. Br J Anaesth. 2013 Oct;111(4):573-9. PMID: 23625132
PiCCOTM vs. VolumeViewTM
Methods:
全患者にVolumeViewカテーテルを挿入
EV1000TMにより波形データを用いた持続COを測定(CCOVolumeView)
5分後に冷水20ml x3注入, EV1000TMにてTPTDを測定
VolumeViewTMの圧波形データを同時にPiCCO2TMへ転送, 持続CO測定(CCOPiCCO)
15%以上の誤差は除外
1) CCOVolumeView とCCOPiCCOの誤差を検証
2) TPTDをreferenceとしてCCOVolumeView とCCOPiCCOとの誤差を検証
51
52. Kiefer N, Hofer CK, Marx G et al.
Crit Care. 2012 May 30;16(3):R98.
Bland Altman Plot CO
0.82
0.2
-0.45
r2=0.98
% error: 9.7%
CO average COVolumeView & COPiCCO
ΔCOVolumeView (%)
4 quadrant plot % Change CO
ΔCOPiCCO (%)
COVolumeView (ml/min)
Results:
Continuous CO
VolumeViewTM
vs. PiCCOTM
CO difference COVolumeView & COPiCCO
COVolumeView vs. COPiCCO
COPiCCO (ml/min)
Polar Plot % Change CO
150°
30°
VolumeViewTM は
PiCCOTMと比べバラつき
が尐なく,同程度の計測結
果が得られる
Concordance: 98.5%
Exclusion zone: 10%
210°
330°
Concordance: 100%
52
53. Bendjelid K, Marx G, Kiefer N et al.
Br J Anaesth. 2013 Oct;111(4):573-9. PMID: 23625132
Results:
CO, CCOVolumeView vs. TPTD (VolumeViewTMのContinuous CO vs. Intermitted CO)
N=338
Bias = - 0.07 L/min
Limits of agreement : ±2.0 L/min
N=338
Concordance: 81%
Exclusion zone: 15%
% error: 29%
53
54. Bendjelid K, Marx G, Kiefer N et al.
Br J Anaesth. 2013 Oct;111(4):573-9. PMID: 23625132
Results:
CO, CCOPiCCO vs. TPTD
(PiCCOTMのContinuous CO vs. VolumeViewTMのIntermitted CO)
N=338
Bias = 0.03 L/min
Limits of agreement : ±2.5 L/min
N=338
Concordance: 77%
Exclusion zone: 15%
% error: 37%
54
59. Hadian M, Kim HK, Severyn DA, Pinsky MR.
Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery
catheters. Crit Care. 2010;14(6):R212. PMCID: PMC3220011
2nd FloTracTM vs. PiCCOTM, 2nd FloTracTM vs. pulmonary artery catheter
Patients/Methods:
心臓血管外科術後でPAカテーテル挿入している患者17人,
容量負荷した後,血圧,HRが安定したところでCO測定
COTD:10ml冷水注入の後PACからVigilanceTMにて測定
CCO: PACからVigilanceTMにてContinuous COを測定
FloTrac: FloTrac/VigilanceTM ver 1.26
PiCCO: pulse contourによるcontinuous COを測定
59
60. Hadian M, Kim HK, Severyn DA, Pinsky MR.
Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery
catheters. Crit Care. 2010;14(6):R212. PMCID: PMC3220011
2nd FloTracTM vs. PiCCOTM, 2nd FloTracTM vs. pulmonary artery catheter
Results:
% error: 59%
% error: 53%
Limits of Agreement: 1.96SD
60
61. Monnet X, Anguel N, Naudin B, et al.
Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and
uncalibrated pressure waveform devices. Crit Care. 2010;14(3):R109. PMID: 20537159
2nd Flotrac/Vigileo™ vs. PiCCO™, Cardiac Index (CI)を評価
Setting:
single center, medical ICU, France
Patients / Methods:
敗血症性ショックで循環管理された患者80人
Group 1 (n=40): 輸液負荷に反応すると予測された患者
(CI上昇: Passive leg raising CI≥10%,
end-expiratory occlusionでCI≥5%)
Group 2 (n=40): MAP 65-75 mmHgになるようNAD調節
計測方法:
CIpcとCIpwは常にモニター上に表示
CItdでは 冷水15ml注入, 3回測定した平均値を記録
CIpc, CIpwでは 冷水注入直後の値を記録
CIpw: arterial pressure waveform-derived CI ( Flotrac/Vigileo™ ver1.10),
CIpc: pulse contour CI (PiCCO plus™) , CItd: transpulmonary thermodilution CI (PiCCO plus™)
61
83. Mahjoub Y, Lakhdari M, Lorne E, et al.
Assessment of an uncalibrated pressure waveform device's ability to track cardiac output changes due to
norepinephrine dose adjustments in patients with septic shock: a comparison with Doppler
echocardiography. Ann Fr Anesth Reanim. 2012;31(9):677-81.PMID: 22776771.
3rd Vigileo/FloTrac™ vs. 経食道Doppler, Septic shock患者に対するCOを検証
Setting:
Single center, Prospective observational study, medical and surgical ICU, France
Patients:
人工呼吸管理された患者20人を含む昇圧薬(NAD or dopamine)を必要とする成人
septic shock患者24人
Sample: 107 paired measurement CO; 90 paired measurement ΔCO
Methods:
PAカテーテルを抜去するまでの期間COを同時に測定, 3回施行し平均値を記録
COED: Philips Envisor HDTMにてCOFTVを隠した状態で測定
COFTV: FloTrac/VigileoTM(ver3.02) にてCO測定
83
84. Mahjoub Y, Lakhdari M, Lorne E, et al.
Assessment of an uncalibrated pressure waveform device's ability to track cardiac output changes due to
norepinephrine dose adjustments in patients with septic shock: a comparison with Doppler
echocardiography. Ann Fr Anesth Reanim. 2012;31(9):677-81.PMID: 22776771.
Results: COFTV vs. COED (3rd Vigileo/FloTrac™ vs. 経食道Doppler)
絶対値CO
ΔCO
Bias: 0.1 L/min
LOA: - 2.6 to 2.7
Bias: - 1.7 L/min
LOA: - 6.2 to 2.8
% error: 81%
経食道DopplerとFloTracTMではCOのバラつきが大きい
ΔCOでのBiasはほとんどない
Four qudrant plot, polar plotによる検証はない
84
85. Meng L, Tran NP, Alexander BS, et al.
Theimpact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FloTrac and
esophageal doppler cardiac output measurements. Anesth Analg.2011 Oct;113(4):751-7.
PMID: 21821516.
3rd Vigileo/FloTrac™ vs. 経食道Doppler
昇圧薬(Phenylephrine, Ephedrine)の影響を評価
Setting:
Single center
Prospective cohort study,
Univ. California, U.S.A
Patients:
定時手術患者33人
176 paired measurement
Methods:
左図
各昇圧薬投与前後で値を記
録
85
86. Meng L, Tran NP, Alexander BS, et al.
Theimpact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FloTrac and
esophageal Doppler cardiac output measurements. Anesth Analg.2011 Oct;113(4):751-7.
PMID: 21821516.
concordance: 23%
Exclusion zone 3%
No zone: 22%
concordance: 69%
Exclusion zone 3%
No zone: 67%
concordance: 96%
Exclusion zone 1.2%
No zone: 92%
Results:
A: Phenylephrine
経食道DopplerとFloTracTMのCOが逆相関→末梢血管収縮がFloTracTMのCOに影
響
B: Ephedrine
逆相関してはいないがバラつきは大きい
C:Preload increase
輸液負荷によるCO上昇はEs DopplerとFloTracTMで遜色がない
86
89. Hamzaoui O, Monnet X, Richard C, et al.
Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary
thermodilution cardiac output measurements within an up to 6-hour calibration-free period.
Crit Care Med. 2008 Feb;36(2):434-40. PMID:18091547.
PiCCOTMplusのキャリブレーション間隔を検証
Setting/Patients:
Prospective observational study
侵襲的モニタリングを必要とした非心臓手術患者73人, 330 paired measurement
Methods:
PiCCOTMplusにてcontinuous CO(PCCO)とIntermitted CO (COTCP)を測定
キャリブレーションの時間ごとのバラつきをBland-Altman plotを用いて評価
(8時間おきの投与を勧めているが,投与時刻は臨床医の判断で施行)
Results:
初回から1時間以降の投与
→ %error > 30%
⇓
キャリブレーションか
ら
1時間以降ではPCCOの
バラつきが大きくなる
89
90. Dufour N, Delville M, Teboul JL, Monnet X al.
Transpulmonary thermodilution measurements are not affected by continuous veno-venous
hemofiltration at high blood pump flow. Intensive Care Med. 2012 Jul;38(7):1162-8. PMID: 22543424.
PiCCOTMによるintermitted CI vs. continuous CI, CVVHDによる検証
Patients:
CVVHDを受けている患者69人
Blood pump flow: 250 ml/min n=31, 350 ml/min n=38 , filtration flow
Methods:
PiCCO2TMにてTPTD でのintermitted CIとpulse contourでのcontinuous CIを測定
TDon: CVVHD中のintermitted CI, PCoff: CVVH停止時点のcontinuous CI
TDoff: CVVH停止時点のintermitted CI
TDon-last:
CVVHD再開直後のintermitted CI
冷水15ml注入しCI計測
90
91. Dufour N, Delville M, Teboul JL, Monnet X al
. Transpulmonary thermodilution measurements are not affected by continuous veno-venous
hemofiltration at high blood pump flow. Intensive Care Med. 2012 Jul;38(7):1162-8. PMID: 22543424.
250 ml/min
Results:
CVVHDの使用の有無でTPTDによるCIに
有意差はない
CVVHD中のpulse contourによるCIは
検証されていない
350 ml/min
91
95. De Backer D, Donadello K, Sakr Y, Vincent JL et al.
Microcirculatory alterations in patients with severe sepsis: impact of time of assessment and relationship
with outcome. Crit Care Med. 2013;41(3):791-9. PMID: 23318492.
Microvideoscopy を用いて微小血管循環の評価を行いseptic shock を管理
Setting:
Analysis f prospectively collected
data
Results:
微小循環モニターをした場合の
微小循環モニ
ター
Survivor, non survivorの感度,特異度
↓
MAP, SvO2, lactateを指標にした場
合よりも高い
Area under Curve:
微小循環モニター 0.818
MAP 0.576
SvO2 0.616
lactate 0.612
95
96. … we must try and choose devices that have a
maximum of these attributes,
bearing in mind that there is no ‘one size fits all’
type of system.
And one should, therefore, select the system most
appropriate for each patient and, perhaps even more
importantly, for each type of problem.
Jean-Louis Vincent
Critical Care 2011, 15:229
96