a p p l i c at i o n N o t e

Atomic Absorption
Authors:
Lee Davidowski, Ph.D.
Lorraine Foglio
PerkinElmer, Inc.
Shelton, CT 06484 USA

The Determination of
Minerals and Metals
in Multi-Mineral/MultiVitamin Tablets by Flame
Atomic Absorption
Spectroscopy

Introduction

There are many mineral dietary supplements
available in today’s marketplace to ensure
that mineral deficiencies do not occur in
one’s diet. The mineral content of these
products must be verified for quality
control (QC) purposes. Furthermore, the
Nutritional Labeling and Education Act of
1990 mandates accurate labeling of all
food supplements sold in the U.S. which
means accurate testing of the products
is mandatory. In many labs, this task is
accomplished by the technique of flame
atomic absorption spectroscopy (FAAS). FAAS has the advantages of lower initial
cost, low cost per analysis, and requires less operator training than many other
trace elemental techniques. The objective of this work is to demonstrate the
applicability of FAAS using the PerkinElmer® PinAAcle™ 900T to accomplish this
task. Seven elements are determined in two commercially available multi-mineral
tablets, a NIST® Standard Reference Material, and a commercial reference material
which simulates a mixed food diet.
Experimental
Instrumentation
The PinAAcle 900T flame and longitudinal Zeeman furnace
atomic absorption spectrometer controlled by WinLab32™
for AA software, running under Microsoft® Windows® 7,
was used for all analyses (PerkinElmer, Inc., Shelton, CT). A
high-sensitivity nebulizer (Part No. N3160112) with a spacer
was employed. Single-element Lumina™ hollow cathode
lamps (HCLs) were used and the instrumental operating
conditions for each element in this application are shown in
Table 1. A four-second integration time and three replicates
were used for all elements.

approximately one hour and warmed, not to boiling, on a
hot plate. Deionized water was then added to the volumetric
mark after removal of the magnetic stir bar. The sample
digestion solution was then filtered through Whatman®
46 paper (passes < 8 micron) to remove waxy undissolved
material, presumably from the tablet’s coating. The filtrate
was then analyzed, after appropriate dilution, against singleelement aqueous standards (PerkinElmer Pure) for elemental
quantitation. A Certified Reference Material Mixed Food
Diet (CRM-MFD) (High Purity Standards, Charleston, SC)
was prepared by dissolution, following the manufacturer’s
instructions, to an appropriate concentration for individual
elements.
All tablets also contained about 2 mg/tablet of silicon, much
of which may have remained undissolved. For an accurate
measurement of the silicon, HF (trace metal grade) and plastic
volumetric flasks should be employed for the dissolution.
For calcium and magnesium determinations, La(NO3)2
(reagent grade) was added to all standards, blank and
samples to a concentration level of 0.2% to remove chemical
inferences from other matrix elements. For potassium
determinations, CsCl (reagent grade) was added to all test
solutions to a concentration level of 0.2% as an ionization
buffer.

Figure 1. PinAAcle 900T atomic absorption spectrometer.

Results
Sample and Standard Preparation
One NIST® SRM 3280 Multi-vitamin/Multi-mineral Tablet and
two over the counter multi-vitamin/multi-mineral tablets were
coarsely ground. Nominally, 0.75 g were weighed out and
rinsed into a 1000 mL volumetric flask containing about
20 mL of ASTM® Type I deionized water (>16MΩ • cm).
Four mL of HCl (trace metal grade), two mL of HNO3 (trace
metal grade), and a magnetic stirring bar were then added
to the flask. This sample digestion solution was stirred for

All elements showed acceptable calibration criteria. The
results obtained for the FAAS analyses of the multi-vitamin/
multi-mineral tablets are shown in Table 2. The values are
corrected for laboratory sample dissolution steps and dilutions.
The commercial tablets were calculated and reported in mg/
tablet as this was the unit shown and guaranteed on each
label. All elements were found to be within the certified
range for the NIST® 3280 tablet. Both commercial tablets
showed good correlation to labeled quantity guaranteed by
the manufacturer.

Table 1. PinAAcle 900T Instrumental Parameters.
	

Element

Parameter	

Ca	

Mg	

K	

Fe	

Zn	

Cu	

Mn

Wavelength (nm)	

422.7	

285.2	

766.5	

248.3	

213.9	

324.8	

279.5

Slit Width (nm)	

0.7	

0.7	

0.7	

0.2	

0.7	

0.7	

0.2

Oxidant Flow (L/min)	

10	

10	

10	

10	

10	

8.7	

10

C2H2 Flow (L/min)	

2.5	

2.5	

2.2	

2.5	

2.0	

2.5

5	

1	

4	

4	

1	

4	

2

HCL Part Nos. 	

N3050114	

N3050144	

N3050139	

N3050126	

N3050191	

N3050121	

N3050145

Standard Part Nos.	

2

2.5	

Working Range (mg/L)	

N9303763	

N9300179	

N9303779	

N9303771	

N9300178	

N9300183	

N9303783
Table 2. Corrected FAAS results for one certified reference material and two commercial multi-vitamin/multi-mineral samples.
	
	
Element	

NIST® 3280	
(mg/g)	
Certified	

Found	

Commercial Tablet 1	
(mg/Tab)	
SD	

Label	

Found	

SD	

Commercial Tablet 2
(mg/Tab)
Label	

Found	

SD

Ca	

110.7	

107	

0.8	

162	

163	

0.6	

200	

211	

0.6

Mg	

67.8	

69.1	

0.2	

100	

107	

0.4	

100	

109	

0.4

K	

53.1	

53.8	

0.3	

80	

83.8	

0.4	

80	

90.7	

0.4

Fe	

12.35	

12.9	

0.05	

18	

19.1	

0.08	

n/a	

0.23	

0.006

Zn	

10.15	

10.1	

0.02	

15	

15.5	

0.04	

15	

15.2	

0.03

Cu	

1.4	

1.42	

0.01	

2	

2.19	

0.01	

2	

2.04	

0.01

Mn	

1.44	

1.48	

0.01	

2	

1.91	

0.01	

2	

2.04	

0.01

The results for the FAAS analysis of the Certified Reference
Material – Mixed Food Diet from High Purity Standards – are
given in Table 3. All elements were found to be within 5%
of the certified value, showing excellent agreement with the
reference material.
Table 3. Mixed Food Diet CRM results using aqueous
standards and FAAS on a PinAAcle 900T.
	
	

CRM-MFD
(mg/L)

Element	

Certified	

Found	

SD

Ca	

40	

39.8	

0.2

Mg	

12	

11.8	

160	

159	

0.005

Fe	

0.8	

0.802	

0.005

Zn	

0.3	

0.290	

0.001

Cu	

0.06	

0.061	

0.001

Mn	

0.2	

0.191	

The mineral content of multi-vitamin/multi-mineral tablets
must be determined in order to ensure the quality of
commercial dietary supplements. This method demonstrates
the ability of the PinAAcle 900T flame atomic absorption
system to accurately measure minerals in both commercial
tablets and food simulation diets. It is an efficacious method
that allows for cost effectiveness and ease of use, with
less operator training than other analytical methods. The
PinAAcle 900H (Flame and Deuterium Furnace) and PinAAcle
900F (Flame only) spectrometers can also be used for this
application.

0.03

K	

Conclusions

0.003

PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA	
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com

For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
009636A_01

Contenu connexe

Tendances

Get Proactive With Your Stability Program
Get Proactive With Your Stability ProgramGet Proactive With Your Stability Program
Get Proactive With Your Stability Program
Merck Life Sciences
 
Identification of Bioactive Compounds in the acetone extract of Daedalea eleg...
Identification of Bioactive Compounds in the acetone extract of Daedalea eleg...Identification of Bioactive Compounds in the acetone extract of Daedalea eleg...
Identification of Bioactive Compounds in the acetone extract of Daedalea eleg...
AI Publications
 
TetraQ Services Presentation
TetraQ Services PresentationTetraQ Services Presentation
TetraQ Services Presentation
simonsend
 
003 BI Verification Change Control
003 BI Verification Change Control003 BI Verification Change Control
003 BI Verification Change Control
Mark Thompson
 
Presentation1rohitha reddy
Presentation1rohitha reddyPresentation1rohitha reddy
Presentation1rohitha reddy
ceutics1315
 
Ds cto optimizeantibodymanuf
Ds cto optimizeantibodymanufDs cto optimizeantibodymanuf
Ds cto optimizeantibodymanuf
mganguly123
 

Tendances (18)

Get Proactive With Your Stability Program
Get Proactive With Your Stability ProgramGet Proactive With Your Stability Program
Get Proactive With Your Stability Program
 
Upstream Viral Safety: A Holistic Approach to Mitigating Contamination Risks
Upstream Viral Safety: A Holistic Approach to Mitigating Contamination RisksUpstream Viral Safety: A Holistic Approach to Mitigating Contamination Risks
Upstream Viral Safety: A Holistic Approach to Mitigating Contamination Risks
 
Physical, Thermal & Spectral Properties of p-Chloro-m-cresol
Physical, Thermal & Spectral Properties of p-Chloro-m-cresolPhysical, Thermal & Spectral Properties of p-Chloro-m-cresol
Physical, Thermal & Spectral Properties of p-Chloro-m-cresol
 
Identification of Bioactive Compounds in the acetone extract of Daedalea eleg...
Identification of Bioactive Compounds in the acetone extract of Daedalea eleg...Identification of Bioactive Compounds in the acetone extract of Daedalea eleg...
Identification of Bioactive Compounds in the acetone extract of Daedalea eleg...
 
Excipients selection for high risk formulations Smita Rajput
Excipients selection for high risk formulations Smita RajputExcipients selection for high risk formulations Smita Rajput
Excipients selection for high risk formulations Smita Rajput
 
Formulation and evaluation of effect of different stabilizer at nanosuspensio...
Formulation and evaluation of effect of different stabilizer at nanosuspensio...Formulation and evaluation of effect of different stabilizer at nanosuspensio...
Formulation and evaluation of effect of different stabilizer at nanosuspensio...
 
Platform Techniques for Preformulation Development for Non-Antibody Products
Platform Techniques for Preformulation Development for Non-Antibody ProductsPlatform Techniques for Preformulation Development for Non-Antibody Products
Platform Techniques for Preformulation Development for Non-Antibody Products
 
A Turn-Key Flow-Through-Mode Purification Process to improve Quality and Safe...
A Turn-Key Flow-Through-Mode Purification Process to improve Quality and Safe...A Turn-Key Flow-Through-Mode Purification Process to improve Quality and Safe...
A Turn-Key Flow-Through-Mode Purification Process to improve Quality and Safe...
 
TetraQ Services Presentation
TetraQ Services PresentationTetraQ Services Presentation
TetraQ Services Presentation
 
Q3D - Elemental Impurities: What implications for APIs & excipients suppliers?
Q3D - Elemental Impurities: What implications for APIs & excipients suppliers?Q3D - Elemental Impurities: What implications for APIs & excipients suppliers?
Q3D - Elemental Impurities: What implications for APIs & excipients suppliers?
 
Stability testing protocols
Stability testing protocolsStability testing protocols
Stability testing protocols
 
The Butterfly Effect: How to see the impact of small changes to your ADC
The Butterfly Effect: How to see the impact of small changes to your ADCThe Butterfly Effect: How to see the impact of small changes to your ADC
The Butterfly Effect: How to see the impact of small changes to your ADC
 
Stability study for formulation
Stability study for formulation Stability study for formulation
Stability study for formulation
 
A Platform Approach to Preformulation Development for Antibody Products
A Platform Approach to Preformulation Development for Antibody ProductsA Platform Approach to Preformulation Development for Antibody Products
A Platform Approach to Preformulation Development for Antibody Products
 
003 BI Verification Change Control
003 BI Verification Change Control003 BI Verification Change Control
003 BI Verification Change Control
 
Stability indicating assay
Stability indicating assayStability indicating assay
Stability indicating assay
 
Presentation1rohitha reddy
Presentation1rohitha reddyPresentation1rohitha reddy
Presentation1rohitha reddy
 
Ds cto optimizeantibodymanuf
Ds cto optimizeantibodymanufDs cto optimizeantibodymanuf
Ds cto optimizeantibodymanuf
 

Similaire à The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tablets by Flame Atomic Absorption Spectroscopy

Nitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdfNitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdf
NguyenMinhChau22
 
Determination of 3-chloropropanediol in soy sauce samples by liquid phase ext...
Determination of 3-chloropropanediol in soy sauce samples by liquid phase ext...Determination of 3-chloropropanediol in soy sauce samples by liquid phase ext...
Determination of 3-chloropropanediol in soy sauce samples by liquid phase ext...
International Multispeciality Journal of Health
 
PDQAthenaPoster - Jan2015
PDQAthenaPoster - Jan2015PDQAthenaPoster - Jan2015
PDQAthenaPoster - Jan2015
Rosalind Ramsey
 

Similaire à The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tablets by Flame Atomic Absorption Spectroscopy (20)

Nitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdfNitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdf
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
 
Analysis of Pb, Cd and As in Spice Mixtures using Graphite Furnace Atomic Abs...
Analysis of Pb, Cd and As in Spice Mixtures using Graphite Furnace Atomic Abs...Analysis of Pb, Cd and As in Spice Mixtures using Graphite Furnace Atomic Abs...
Analysis of Pb, Cd and As in Spice Mixtures using Graphite Furnace Atomic Abs...
 
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
 
Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...
Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...
Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...
 
Trace Elemental Characterization of Edible Oils with Graphite Furnace Atomic ...
Trace Elemental Characterization of Edible Oils with Graphite Furnace Atomic ...Trace Elemental Characterization of Edible Oils with Graphite Furnace Atomic ...
Trace Elemental Characterization of Edible Oils with Graphite Furnace Atomic ...
 
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
 
Subvisible Particle Characterization: Why Simply Counting Shadows Leaves You ...
Subvisible Particle Characterization: Why Simply Counting Shadows Leaves You ...Subvisible Particle Characterization: Why Simply Counting Shadows Leaves You ...
Subvisible Particle Characterization: Why Simply Counting Shadows Leaves You ...
 
ICH guidelines
ICH guidelines ICH guidelines
ICH guidelines
 
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
 
Brochure
BrochureBrochure
Brochure
 
ICH Guidelines Q1 - Q10
ICH Guidelines Q1 - Q10ICH Guidelines Q1 - Q10
ICH Guidelines Q1 - Q10
 
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...
 
Determination of 3-chloropropanediol in soy sauce samples by liquid phase ext...
Determination of 3-chloropropanediol in soy sauce samples by liquid phase ext...Determination of 3-chloropropanediol in soy sauce samples by liquid phase ext...
Determination of 3-chloropropanediol in soy sauce samples by liquid phase ext...
 
Sumita
SumitaSumita
Sumita
 
PDQAthenaPoster - Jan2015
PDQAthenaPoster - Jan2015PDQAthenaPoster - Jan2015
PDQAthenaPoster - Jan2015
 
Analysis of heavy_metal_content_in_conventional_an
Analysis of heavy_metal_content_in_conventional_anAnalysis of heavy_metal_content_in_conventional_an
Analysis of heavy_metal_content_in_conventional_an
 
STUDY OF ICH Q8
STUDY OF ICH Q8STUDY OF ICH Q8
STUDY OF ICH Q8
 
optimization of microcrystalline cellulose production from groundnuts husk
optimization of microcrystalline cellulose production from groundnuts huskoptimization of microcrystalline cellulose production from groundnuts husk
optimization of microcrystalline cellulose production from groundnuts husk
 
Determination of Residual Solvents in Flexible Packaging
Determination of Residual Solvents in Flexible PackagingDetermination of Residual Solvents in Flexible Packaging
Determination of Residual Solvents in Flexible Packaging
 

Plus de PerkinElmer, Inc.

Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
PerkinElmer, Inc.
 

Plus de PerkinElmer, Inc. (20)

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º Trimestre
 
É A VEZ DO PRIME
É A VEZ DO PRIMEÉ A VEZ DO PRIME
É A VEZ DO PRIME
 
TREC PARA A VIDA
TREC PARA A VIDATREC PARA A VIDA
TREC PARA A VIDA
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o Desempenho
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no Brasil
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGS
 
O que é X Frágil?
O que é X Frágil?O que é X Frágil?
O que é X Frágil?
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedor
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
 
CG Clarus® All In One
CG Clarus® All In OneCG Clarus® All In One
CG Clarus® All In One
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
 
XXI Encontro de Genética
XXI Encontro de GenéticaXXI Encontro de Genética
XXI Encontro de Genética
 

Dernier

Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 
Architecting Cloud Native Applications
Architecting Cloud Native ApplicationsArchitecting Cloud Native Applications
Architecting Cloud Native Applications
WSO2
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
panagenda
 
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Victor Rentea
 

Dernier (20)

Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
Architecting Cloud Native Applications
Architecting Cloud Native ApplicationsArchitecting Cloud Native Applications
Architecting Cloud Native Applications
 
DEV meet-up UiPath Document Understanding May 7 2024 Amsterdam
DEV meet-up UiPath Document Understanding May 7 2024 AmsterdamDEV meet-up UiPath Document Understanding May 7 2024 Amsterdam
DEV meet-up UiPath Document Understanding May 7 2024 Amsterdam
 
Artificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : UncertaintyArtificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : Uncertainty
 
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ..."I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
 
2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...
 
CNIC Information System with Pakdata Cf In Pakistan
CNIC Information System with Pakdata Cf In PakistanCNIC Information System with Pakdata Cf In Pakistan
CNIC Information System with Pakdata Cf In Pakistan
 
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdfRising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a Fresher
 
[BuildWithAI] Introduction to Gemini.pdf
[BuildWithAI] Introduction to Gemini.pdf[BuildWithAI] Introduction to Gemini.pdf
[BuildWithAI] Introduction to Gemini.pdf
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
Ransomware_Q4_2023. The report. [EN].pdf
Ransomware_Q4_2023. The report. [EN].pdfRansomware_Q4_2023. The report. [EN].pdf
Ransomware_Q4_2023. The report. [EN].pdf
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
 
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
 
TrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
TrustArc Webinar - Unlock the Power of AI-Driven Data DiscoveryTrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
TrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
 
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot TakeoffStrategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
 
Exploring the Future Potential of AI-Enabled Smartphone Processors
Exploring the Future Potential of AI-Enabled Smartphone ProcessorsExploring the Future Potential of AI-Enabled Smartphone Processors
Exploring the Future Potential of AI-Enabled Smartphone Processors
 
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemkeProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
 
ICT role in 21st century education and its challenges
ICT role in 21st century education and its challengesICT role in 21st century education and its challenges
ICT role in 21st century education and its challenges
 

The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tablets by Flame Atomic Absorption Spectroscopy

  • 1. a p p l i c at i o n N o t e Atomic Absorption Authors: Lee Davidowski, Ph.D. Lorraine Foglio PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Minerals and Metals in Multi-Mineral/MultiVitamin Tablets by Flame Atomic Absorption Spectroscopy Introduction There are many mineral dietary supplements available in today’s marketplace to ensure that mineral deficiencies do not occur in one’s diet. The mineral content of these products must be verified for quality control (QC) purposes. Furthermore, the Nutritional Labeling and Education Act of 1990 mandates accurate labeling of all food supplements sold in the U.S. which means accurate testing of the products is mandatory. In many labs, this task is accomplished by the technique of flame atomic absorption spectroscopy (FAAS). FAAS has the advantages of lower initial cost, low cost per analysis, and requires less operator training than many other trace elemental techniques. The objective of this work is to demonstrate the applicability of FAAS using the PerkinElmer® PinAAcle™ 900T to accomplish this task. Seven elements are determined in two commercially available multi-mineral tablets, a NIST® Standard Reference Material, and a commercial reference material which simulates a mixed food diet.
  • 2. Experimental Instrumentation The PinAAcle 900T flame and longitudinal Zeeman furnace atomic absorption spectrometer controlled by WinLab32™ for AA software, running under Microsoft® Windows® 7, was used for all analyses (PerkinElmer, Inc., Shelton, CT). A high-sensitivity nebulizer (Part No. N3160112) with a spacer was employed. Single-element Lumina™ hollow cathode lamps (HCLs) were used and the instrumental operating conditions for each element in this application are shown in Table 1. A four-second integration time and three replicates were used for all elements. approximately one hour and warmed, not to boiling, on a hot plate. Deionized water was then added to the volumetric mark after removal of the magnetic stir bar. The sample digestion solution was then filtered through Whatman® 46 paper (passes < 8 micron) to remove waxy undissolved material, presumably from the tablet’s coating. The filtrate was then analyzed, after appropriate dilution, against singleelement aqueous standards (PerkinElmer Pure) for elemental quantitation. A Certified Reference Material Mixed Food Diet (CRM-MFD) (High Purity Standards, Charleston, SC) was prepared by dissolution, following the manufacturer’s instructions, to an appropriate concentration for individual elements. All tablets also contained about 2 mg/tablet of silicon, much of which may have remained undissolved. For an accurate measurement of the silicon, HF (trace metal grade) and plastic volumetric flasks should be employed for the dissolution. For calcium and magnesium determinations, La(NO3)2 (reagent grade) was added to all standards, blank and samples to a concentration level of 0.2% to remove chemical inferences from other matrix elements. For potassium determinations, CsCl (reagent grade) was added to all test solutions to a concentration level of 0.2% as an ionization buffer. Figure 1. PinAAcle 900T atomic absorption spectrometer. Results Sample and Standard Preparation One NIST® SRM 3280 Multi-vitamin/Multi-mineral Tablet and two over the counter multi-vitamin/multi-mineral tablets were coarsely ground. Nominally, 0.75 g were weighed out and rinsed into a 1000 mL volumetric flask containing about 20 mL of ASTM® Type I deionized water (>16MΩ • cm). Four mL of HCl (trace metal grade), two mL of HNO3 (trace metal grade), and a magnetic stirring bar were then added to the flask. This sample digestion solution was stirred for All elements showed acceptable calibration criteria. The results obtained for the FAAS analyses of the multi-vitamin/ multi-mineral tablets are shown in Table 2. The values are corrected for laboratory sample dissolution steps and dilutions. The commercial tablets were calculated and reported in mg/ tablet as this was the unit shown and guaranteed on each label. All elements were found to be within the certified range for the NIST® 3280 tablet. Both commercial tablets showed good correlation to labeled quantity guaranteed by the manufacturer. Table 1. PinAAcle 900T Instrumental Parameters. Element Parameter Ca Mg K Fe Zn Cu Mn Wavelength (nm) 422.7 285.2 766.5 248.3 213.9 324.8 279.5 Slit Width (nm) 0.7 0.7 0.7 0.2 0.7 0.7 0.2 Oxidant Flow (L/min) 10 10 10 10 10 8.7 10 C2H2 Flow (L/min) 2.5 2.5 2.2 2.5 2.0 2.5 5 1 4 4 1 4 2 HCL Part Nos. N3050114 N3050144 N3050139 N3050126 N3050191 N3050121 N3050145 Standard Part Nos. 2 2.5 Working Range (mg/L) N9303763 N9300179 N9303779 N9303771 N9300178 N9300183 N9303783
  • 3. Table 2. Corrected FAAS results for one certified reference material and two commercial multi-vitamin/multi-mineral samples. Element NIST® 3280 (mg/g) Certified Found Commercial Tablet 1 (mg/Tab) SD Label Found SD Commercial Tablet 2 (mg/Tab) Label Found SD Ca 110.7 107 0.8 162 163 0.6 200 211 0.6 Mg 67.8 69.1 0.2 100 107 0.4 100 109 0.4 K 53.1 53.8 0.3 80 83.8 0.4 80 90.7 0.4 Fe 12.35 12.9 0.05 18 19.1 0.08 n/a 0.23 0.006 Zn 10.15 10.1 0.02 15 15.5 0.04 15 15.2 0.03 Cu 1.4 1.42 0.01 2 2.19 0.01 2 2.04 0.01 Mn 1.44 1.48 0.01 2 1.91 0.01 2 2.04 0.01 The results for the FAAS analysis of the Certified Reference Material – Mixed Food Diet from High Purity Standards – are given in Table 3. All elements were found to be within 5% of the certified value, showing excellent agreement with the reference material. Table 3. Mixed Food Diet CRM results using aqueous standards and FAAS on a PinAAcle 900T. CRM-MFD (mg/L) Element Certified Found SD Ca 40 39.8 0.2 Mg 12 11.8 160 159 0.005 Fe 0.8 0.802 0.005 Zn 0.3 0.290 0.001 Cu 0.06 0.061 0.001 Mn 0.2 0.191 The mineral content of multi-vitamin/multi-mineral tablets must be determined in order to ensure the quality of commercial dietary supplements. This method demonstrates the ability of the PinAAcle 900T flame atomic absorption system to accurately measure minerals in both commercial tablets and food simulation diets. It is an efficacious method that allows for cost effectiveness and ease of use, with less operator training than other analytical methods. The PinAAcle 900H (Flame and Deuterium Furnace) and PinAAcle 900F (Flame only) spectrometers can also be used for this application. 0.03 K Conclusions 0.003 PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 009636A_01