Corneal pachymetry
INTRODUCTION :
 Pachymetry (Greek words: Pachos= thick+metry)
is term used for measurement of corneal thickness.
 It is important indicator of health status of the
cornea especially of corneal endothelial function.
 The thickness of cornea was first reported in
ancient textbooks on physiological optics (Helmholtz
and Gullstrand).
CORNEAL THICKNESS IN
NORMAL EYES
 It ranges from 0.7-0.9 mm at the limbus and varies
between 0.49-0.56 mm at the center.
 The CCT reading of 0.7 mm or more is indicative
of endothelial decompensation.
 Peripheral corneal thickness is asymmetric so that
temporal cornea is thinnest followed by the inferior
cornea.
CORNEAL THICKNESS IN NEWBORNS AND
CHILDREN :
 Is similar to that of adult cornea.
 It has been found that cornea on day one
is significantly thicker and decreases in
thickness as child grows older.
 The average corneal thickness in infants is
585±52 microns.
 The superior peripheral cornea is thinnest
in newborn.
 Normal Values
 Normal corneal thickness – 500-575 microns.
 Abnormal results
 Abnormally thick or thin measurements may
indicate –
 corneal thinning.
 Corneal edema
 Lower or higher than suspected IOP.
FACTORS AFFECTING CCT
 The mean CCT in black children is thinner
than that of white children.
 CCT does not correlate with refraction or
systemic hypertension.
TECHNIQUES
ULTRASONIC OPTICAL ALTERNATIVE
MEASUREMENTS
• Conventional ultrasonic
pachymetry
Manual Pentacam
• Ultrasound
biomicroscpy (UBM).
Specular microscopy • Pachycam
Optical coherence
Tomography(OCT)
1.ULTRASONIC PACHYMETRY
 Broadly accepted as method of choice, regarded as the gold
standard.
 PRINCIPLE :
 Instruments functions by measuring the amount of time
(transmit time) needed for ultrasound pulse pass from one end
of transducer to descemet’s membrane and back to transducer.
 Corneal thickness=transmit time × propagation velocity)/2
 Speed of sound in cornea : current standard is 1640 m/sec
COMPONENTS :
 PROBE HANDLE - It has piezoelectric crystal
that emits an ultrasonic beam of 20 MHz
 TRANSDUCER – It sends ultrasound rays
through the probe to the cornea and receives
echoes from the cornea.
 TIP - The diameter of the tip should not be
more than 2mm.
ADVANTAGES DISADVANTAGES
• Faster ,simpler and easy to use. • Accuracy is dependent on the
perpendicularity of the probe’s
application to the cornea.
Portable Contact method.
• Dry(no coupling medium
required)
• Can be used intraoperatively. Not accurate in edematous cornea.
2. ULTRASOUND
BIOMICROSCOPY:
 High resolution ultrasound
machine which images anterior
segment of the eye.
 Using a 50-MHz imaging probe
 images of the cross-sectional
anterior chamber anatomy with
high resolution are achieved.
 12.5–50 MHz probe , the depth
of penetration is lesser (4mm)than
conventional ultrasound.
ADVANTAGES DISADVANTAGES
• Useful in cases where cornea is
opaque
The main limitation of UBM is that it
requires a water-bath coupling
media
and a very experienced examiner.
• Various layers of cornea can be
identified.
• Contact method and patient to lie
supine
• Cannot be used intraoperatively.
3.MANUAL OPTICAL PACHYMETER:
 This was the original method to measure
corneal thickness.
 The instrument contains two Plano glass
plates that splits the image of the corneal
parallelepiped.
 There are two methods to measure
corneal thickness.
 Just touch method.
 Overlap method.
 The corneal thickness is then directly read
from the scale.
 Range is 0-1.2mm, with a least gradation
of 0.02 mm.
ADVANTAGE DISADVANTAGE
Non contact method. • Lack of accuracy in
measurements, the usual range of
error with an optical pachymeter is
±2%
• Lack of repeatability.
• Requires slit lamp and therefore
has poor portability and cannot be
used in operating room.
4. SPECULAR PACHYMETRY :
 This is the oldest method to measure corneal
thickness.
 PRINCIPLE : This measures the distance
between the anterior and the posterior surfaces of
cornea and depends on the focusing of light rays
through front back cornea unlike sound waves in
ultrasound pachymeter.
 2 types 1.contact 2.Non contact
ADVANTAGES DISADVANTAGES
Operator independent. The exact point where reading is
taken is
not known.
Non invasive. • Time consuming.
Simultaneous measurement of
cell count.
• Impractical to use in operation
room.
6. ANTERIOR SEGMENT OPTICAL
COHERANCE TOMOGRAPHY :
 High resolution, non-contact OCT
customized anterior segment.
 Gives color coded map of corneal
thickness.
ADVANTAGES
 Non contact
 Rapid acquisition during the pachymetry scan ensures an
accurate and repeatable pachymetry map.
 High resolution
 It measures and documents both corneal flap thickness and
residual stromal thickness immediately following LASIK
surgery.
 Measures through corneal opacity.
10. PENTACAM
It is3-dimensional (3D) rotating
scheimpflug.
It can perform five functions in 2
sec.
1. Scheimpflug image of anterior
segment.
2. Pachymetry
3. 3-D anterior chamber
analyzer
4. Corneal topography.
5. Cataract analyzer
 Pachymetry by pentacam is displayed as a
color image over its entire area from limbus
to limbus..
ADVANTAGE DISADVANTAGE
Non invasive, non contact
gives precise representation and
repeatability.
Even minute eye movements are
captured and corrected simultaneously.
▪ It underestimates the corneal thickness in
comparison to ultrasonic pachymetry.
PACHYCAM
 compact and portable
non- contact pachymeter
with built-in keratometer.
It can be mounted on slit
lamp.
 Image acquisition is
done with the help of a
3D alignment screen.
 Advantages
1. Noncontact
2. Compact,portable,light weight
Corneal pachymetry.ppt

Contenu connexe

Tendances

Retinoscopy/ Objective Refraction / Retinoscopy of eye (Principle & Techniqu...
Retinoscopy/ Objective Refraction / Retinoscopy of eye  (Principle & Techniqu...Retinoscopy/ Objective Refraction / Retinoscopy of eye  (Principle & Techniqu...
Retinoscopy/ Objective Refraction / Retinoscopy of eye (Principle & Techniqu...
Bikash Sapkota
 
anomalous retinal correspondence
anomalous retinal correspondenceanomalous retinal correspondence
anomalous retinal correspondence
Rajeshwori
 
Contrast sensitivity
Contrast sensitivityContrast sensitivity
Contrast sensitivity
Jagdish Dukre
 

Tendances (20)

OPTICS OF CONTACT LENSES
OPTICS OF CONTACT LENSESOPTICS OF CONTACT LENSES
OPTICS OF CONTACT LENSES
 
Orthoptic evaluation 1
Orthoptic evaluation 1Orthoptic evaluation 1
Orthoptic evaluation 1
 
Lacrimal function test kapil
Lacrimal function test kapilLacrimal function test kapil
Lacrimal function test kapil
 
Retinoscopy/ Objective Refraction / Retinoscopy of eye (Principle & Techniqu...
Retinoscopy/ Objective Refraction / Retinoscopy of eye  (Principle & Techniqu...Retinoscopy/ Objective Refraction / Retinoscopy of eye  (Principle & Techniqu...
Retinoscopy/ Objective Refraction / Retinoscopy of eye (Principle & Techniqu...
 
Soft toric Contact Lens
Soft toric Contact LensSoft toric Contact Lens
Soft toric Contact Lens
 
Freshnel prism final
Freshnel prism finalFreshnel prism final
Freshnel prism final
 
A scan ultrasonography
A scan ultrasonographyA scan ultrasonography
A scan ultrasonography
 
Optical aberrations
Optical aberrationsOptical aberrations
Optical aberrations
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Examination protocol for binocular vision
Examination protocol for binocular visionExamination protocol for binocular vision
Examination protocol for binocular vision
 
Orthoptic examination
Orthoptic examinationOrthoptic examination
Orthoptic examination
 
Real subjective refraction in astigmatism
Real subjective refraction in astigmatismReal subjective refraction in astigmatism
Real subjective refraction in astigmatism
 
Testing for npa
Testing for npaTesting for npa
Testing for npa
 
Measurements of the optical constants of the eye
Measurements  of  the optical constants of the eyeMeasurements  of  the optical constants of the eye
Measurements of the optical constants of the eye
 
anomalous retinal correspondence
anomalous retinal correspondenceanomalous retinal correspondence
anomalous retinal correspondence
 
Simple & Toric Transposition
Simple & Toric TranspositionSimple & Toric Transposition
Simple & Toric Transposition
 
Pachymetry
PachymetryPachymetry
Pachymetry
 
Keratometry
KeratometryKeratometry
Keratometry
 
Contrast sensitivity
Contrast sensitivityContrast sensitivity
Contrast sensitivity
 
Orthoptics Introduction test
Orthoptics  Introduction testOrthoptics  Introduction test
Orthoptics Introduction test
 

Similaire à Corneal pachymetry.ppt

pachymetry confocal microscopy cornea ophthalmology diagnostics
pachymetry confocal microscopy cornea ophthalmology diagnosticspachymetry confocal microscopy cornea ophthalmology diagnostics
pachymetry confocal microscopy cornea ophthalmology diagnostics
Paavan Kalra
 
OCULAR BIOMETRY AND IOL.pptx
OCULAR BIOMETRY AND IOL.pptxOCULAR BIOMETRY AND IOL.pptx
OCULAR BIOMETRY AND IOL.pptx
Ayienwi Wilson
 

Similaire à Corneal pachymetry.ppt (20)

Pachymetry sivateja
Pachymetry sivatejaPachymetry sivateja
Pachymetry sivateja
 
pachymetry confocal microscopy cornea ophthalmology diagnostics
pachymetry confocal microscopy cornea ophthalmology diagnosticspachymetry confocal microscopy cornea ophthalmology diagnostics
pachymetry confocal microscopy cornea ophthalmology diagnostics
 
Biometry instruments & equipment
Biometry instruments & equipmentBiometry instruments & equipment
Biometry instruments & equipment
 
pachy.pptx
pachy.pptxpachy.pptx
pachy.pptx
 
biometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfbiometry-160212135415 (1).pdf
biometry-160212135415 (1).pdf
 
Biometry & Iol calculations
Biometry & Iol calculationsBiometry & Iol calculations
Biometry & Iol calculations
 
A SCAN, B SCAN & BIOMETRY---19.pptx
A SCAN, B SCAN & BIOMETRY---19.pptxA SCAN, B SCAN & BIOMETRY---19.pptx
A SCAN, B SCAN & BIOMETRY---19.pptx
 
AS OCT & UBM - Dr Shylesh B Dabke
AS OCT & UBM - Dr Shylesh B DabkeAS OCT & UBM - Dr Shylesh B Dabke
AS OCT & UBM - Dr Shylesh B Dabke
 
Optical Coherence Tomography
Optical Coherence TomographyOptical Coherence Tomography
Optical Coherence Tomography
 
Oct introduction
Oct  introductionOct  introduction
Oct introduction
 
Slit lamp biomicroscope
Slit lamp biomicroscopeSlit lamp biomicroscope
Slit lamp biomicroscope
 
biometry for ON.ppt
biometry for ON.pptbiometry for ON.ppt
biometry for ON.ppt
 
oct-ujjval solanki
oct-ujjval solankioct-ujjval solanki
oct-ujjval solanki
 
OCULAR BIOMETRY AND IOL.pptx
OCULAR BIOMETRY AND IOL.pptxOCULAR BIOMETRY AND IOL.pptx
OCULAR BIOMETRY AND IOL.pptx
 
OCT
OCTOCT
OCT
 
ultrasound biomicroscopy
ultrasound biomicroscopyultrasound biomicroscopy
ultrasound biomicroscopy
 
Oct and visual field by dr abdul basir safi
Oct and visual field by dr abdul basir safiOct and visual field by dr abdul basir safi
Oct and visual field by dr abdul basir safi
 
Refractive surgeries.pptx arjun
Refractive surgeries.pptx arjunRefractive surgeries.pptx arjun
Refractive surgeries.pptx arjun
 
OCT , Laser therapy for DR , Vitrectomy
OCT , Laser therapy for DR , VitrectomyOCT , Laser therapy for DR , Vitrectomy
OCT , Laser therapy for DR , Vitrectomy
 
Oct
OctOct
Oct
 

Dernier

FULL NIGHT — 9999894380 Call Girls In New Ashok Nagar | Delhi
FULL NIGHT — 9999894380 Call Girls In New Ashok Nagar | DelhiFULL NIGHT — 9999894380 Call Girls In New Ashok Nagar | Delhi
FULL NIGHT — 9999894380 Call Girls In New Ashok Nagar | Delhi
SaketCallGirlsCallUs
 
FULL NIGHT — 9999894380 Call Girls In Badarpur | Delhi
FULL NIGHT — 9999894380 Call Girls In Badarpur | DelhiFULL NIGHT — 9999894380 Call Girls In Badarpur | Delhi
FULL NIGHT — 9999894380 Call Girls In Badarpur | Delhi
SaketCallGirlsCallUs
 
FULL NIGHT — 9999894380 Call Girls In Dwarka Mor | Delhi
FULL NIGHT — 9999894380 Call Girls In Dwarka Mor | DelhiFULL NIGHT — 9999894380 Call Girls In Dwarka Mor | Delhi
FULL NIGHT — 9999894380 Call Girls In Dwarka Mor | Delhi
SaketCallGirlsCallUs
 
Authentic # 00971556872006 # Hot Call Girls Service in Dubai By International...
Authentic # 00971556872006 # Hot Call Girls Service in Dubai By International...Authentic # 00971556872006 # Hot Call Girls Service in Dubai By International...
Authentic # 00971556872006 # Hot Call Girls Service in Dubai By International...
home
 
DELHI NCR —@9711106444 Call Girls In Majnu Ka Tilla (MT)| Delhi
DELHI NCR —@9711106444 Call Girls In Majnu Ka Tilla (MT)| DelhiDELHI NCR —@9711106444 Call Girls In Majnu Ka Tilla (MT)| Delhi
DELHI NCR —@9711106444 Call Girls In Majnu Ka Tilla (MT)| Delhi
delhimunirka444
 
FULL NIGHT — 9999894380 Call Girls In Uttam Nagar | Delhi
FULL NIGHT — 9999894380 Call Girls In Uttam Nagar | DelhiFULL NIGHT — 9999894380 Call Girls In Uttam Nagar | Delhi
FULL NIGHT — 9999894380 Call Girls In Uttam Nagar | Delhi
SaketCallGirlsCallUs
 
Dubai Call Girls # 00971547881831 # Indian Call Girls In Dubai # (UAE)
Dubai Call Girls # 00971547881831 # Indian Call Girls In Dubai # (UAE)Dubai Call Girls # 00971547881831 # Indian Call Girls In Dubai # (UAE)
Dubai Call Girls # 00971547881831 # Indian Call Girls In Dubai # (UAE)
Business Bay Call Girls || 0529877582 || Call Girls Service in Business Bay Dubai
 
(9711106444 )🫦#Sexy Desi Call Girls Noida Sector 4 Escorts Service Delhi 🫶
(9711106444 )🫦#Sexy Desi Call Girls Noida Sector 4 Escorts Service Delhi 🫶(9711106444 )🫦#Sexy Desi Call Girls Noida Sector 4 Escorts Service Delhi 🫶
(9711106444 )🫦#Sexy Desi Call Girls Noida Sector 4 Escorts Service Delhi 🫶
delhimunirka444
 
Top # 971526963005 # Call Girls Near Burjuman By philippines call girls in du...
Top # 971526963005 # Call Girls Near Burjuman By philippines call girls in du...Top # 971526963005 # Call Girls Near Burjuman By philippines call girls in du...
Top # 971526963005 # Call Girls Near Burjuman By philippines call girls in du...
home
 
FULL NIGHT — 9999894380 Call Girls In Shivaji Enclave | Delhi
FULL NIGHT — 9999894380 Call Girls In Shivaji Enclave | DelhiFULL NIGHT — 9999894380 Call Girls In Shivaji Enclave | Delhi
FULL NIGHT — 9999894380 Call Girls In Shivaji Enclave | Delhi
SaketCallGirlsCallUs
 
Bobbie goods coloring book 81 pag_240127_163802.pdf
Bobbie goods coloring book 81 pag_240127_163802.pdfBobbie goods coloring book 81 pag_240127_163802.pdf
Bobbie goods coloring book 81 pag_240127_163802.pdf
MARIBEL442158
 
FULL NIGHT — 9999894380 Call Girls In Delhi Cantt | Delhi
FULL NIGHT — 9999894380 Call Girls In Delhi Cantt | DelhiFULL NIGHT — 9999894380 Call Girls In Delhi Cantt | Delhi
FULL NIGHT — 9999894380 Call Girls In Delhi Cantt | Delhi
SaketCallGirlsCallUs
 

Dernier (20)

Barasat call girls 📞 8617697112 At Low Cost Cash Payment Booking
Barasat call girls 📞 8617697112 At Low Cost Cash Payment BookingBarasat call girls 📞 8617697112 At Low Cost Cash Payment Booking
Barasat call girls 📞 8617697112 At Low Cost Cash Payment Booking
 
FULL NIGHT — 9999894380 Call Girls In New Ashok Nagar | Delhi
FULL NIGHT — 9999894380 Call Girls In New Ashok Nagar | DelhiFULL NIGHT — 9999894380 Call Girls In New Ashok Nagar | Delhi
FULL NIGHT — 9999894380 Call Girls In New Ashok Nagar | Delhi
 
(NEHA) Call Girls Mumbai Call Now 8250077686 Mumbai Escorts 24x7
(NEHA) Call Girls Mumbai Call Now 8250077686 Mumbai Escorts 24x7(NEHA) Call Girls Mumbai Call Now 8250077686 Mumbai Escorts 24x7
(NEHA) Call Girls Mumbai Call Now 8250077686 Mumbai Escorts 24x7
 
FULL NIGHT — 9999894380 Call Girls In Badarpur | Delhi
FULL NIGHT — 9999894380 Call Girls In Badarpur | DelhiFULL NIGHT — 9999894380 Call Girls In Badarpur | Delhi
FULL NIGHT — 9999894380 Call Girls In Badarpur | Delhi
 
Completed Event Presentation for Huma 1305
Completed Event Presentation for Huma 1305Completed Event Presentation for Huma 1305
Completed Event Presentation for Huma 1305
 
FULL NIGHT — 9999894380 Call Girls In Dwarka Mor | Delhi
FULL NIGHT — 9999894380 Call Girls In Dwarka Mor | DelhiFULL NIGHT — 9999894380 Call Girls In Dwarka Mor | Delhi
FULL NIGHT — 9999894380 Call Girls In Dwarka Mor | Delhi
 
Authentic # 00971556872006 # Hot Call Girls Service in Dubai By International...
Authentic # 00971556872006 # Hot Call Girls Service in Dubai By International...Authentic # 00971556872006 # Hot Call Girls Service in Dubai By International...
Authentic # 00971556872006 # Hot Call Girls Service in Dubai By International...
 
Storyboard short: Ferrarius Tries to Sing
Storyboard short: Ferrarius Tries to SingStoryboard short: Ferrarius Tries to Sing
Storyboard short: Ferrarius Tries to Sing
 
DELHI NCR —@9711106444 Call Girls In Majnu Ka Tilla (MT)| Delhi
DELHI NCR —@9711106444 Call Girls In Majnu Ka Tilla (MT)| DelhiDELHI NCR —@9711106444 Call Girls In Majnu Ka Tilla (MT)| Delhi
DELHI NCR —@9711106444 Call Girls In Majnu Ka Tilla (MT)| Delhi
 
FULL NIGHT — 9999894380 Call Girls In Uttam Nagar | Delhi
FULL NIGHT — 9999894380 Call Girls In Uttam Nagar | DelhiFULL NIGHT — 9999894380 Call Girls In Uttam Nagar | Delhi
FULL NIGHT — 9999894380 Call Girls In Uttam Nagar | Delhi
 
Dubai Call Girls # 00971547881831 # Indian Call Girls In Dubai # (UAE)
Dubai Call Girls # 00971547881831 # Indian Call Girls In Dubai # (UAE)Dubai Call Girls # 00971547881831 # Indian Call Girls In Dubai # (UAE)
Dubai Call Girls # 00971547881831 # Indian Call Girls In Dubai # (UAE)
 
(INDIRA) Call Girl Jammu Call Now 8617697112 Jammu Escorts 24x7
(INDIRA) Call Girl Jammu Call Now 8617697112 Jammu Escorts 24x7(INDIRA) Call Girl Jammu Call Now 8617697112 Jammu Escorts 24x7
(INDIRA) Call Girl Jammu Call Now 8617697112 Jammu Escorts 24x7
 
(9711106444 )🫦#Sexy Desi Call Girls Noida Sector 4 Escorts Service Delhi 🫶
(9711106444 )🫦#Sexy Desi Call Girls Noida Sector 4 Escorts Service Delhi 🫶(9711106444 )🫦#Sexy Desi Call Girls Noida Sector 4 Escorts Service Delhi 🫶
(9711106444 )🫦#Sexy Desi Call Girls Noida Sector 4 Escorts Service Delhi 🫶
 
VIP Ramnagar Call Girls, Ramnagar escorts Girls 📞 8617697112
VIP Ramnagar Call Girls, Ramnagar escorts Girls 📞 8617697112VIP Ramnagar Call Girls, Ramnagar escorts Girls 📞 8617697112
VIP Ramnagar Call Girls, Ramnagar escorts Girls 📞 8617697112
 
Call Girls Ludhiana Just Call 98765-12871 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 98765-12871 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 98765-12871 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 98765-12871 Top Class Call Girl Service Available
 
Top # 971526963005 # Call Girls Near Burjuman By philippines call girls in du...
Top # 971526963005 # Call Girls Near Burjuman By philippines call girls in du...Top # 971526963005 # Call Girls Near Burjuman By philippines call girls in du...
Top # 971526963005 # Call Girls Near Burjuman By philippines call girls in du...
 
FULL NIGHT — 9999894380 Call Girls In Shivaji Enclave | Delhi
FULL NIGHT — 9999894380 Call Girls In Shivaji Enclave | DelhiFULL NIGHT — 9999894380 Call Girls In Shivaji Enclave | Delhi
FULL NIGHT — 9999894380 Call Girls In Shivaji Enclave | Delhi
 
Bobbie goods coloring book 81 pag_240127_163802.pdf
Bobbie goods coloring book 81 pag_240127_163802.pdfBobbie goods coloring book 81 pag_240127_163802.pdf
Bobbie goods coloring book 81 pag_240127_163802.pdf
 
FULL NIGHT — 9999894380 Call Girls In Delhi Cantt | Delhi
FULL NIGHT — 9999894380 Call Girls In Delhi Cantt | DelhiFULL NIGHT — 9999894380 Call Girls In Delhi Cantt | Delhi
FULL NIGHT — 9999894380 Call Girls In Delhi Cantt | Delhi
 
Deconstructing Gendered Language; Feminist World-Making 2024
Deconstructing Gendered Language; Feminist World-Making 2024Deconstructing Gendered Language; Feminist World-Making 2024
Deconstructing Gendered Language; Feminist World-Making 2024
 

Corneal pachymetry.ppt

  • 2. INTRODUCTION :  Pachymetry (Greek words: Pachos= thick+metry) is term used for measurement of corneal thickness.  It is important indicator of health status of the cornea especially of corneal endothelial function.  The thickness of cornea was first reported in ancient textbooks on physiological optics (Helmholtz and Gullstrand).
  • 3.
  • 4. CORNEAL THICKNESS IN NORMAL EYES  It ranges from 0.7-0.9 mm at the limbus and varies between 0.49-0.56 mm at the center.  The CCT reading of 0.7 mm or more is indicative of endothelial decompensation.  Peripheral corneal thickness is asymmetric so that temporal cornea is thinnest followed by the inferior cornea.
  • 5. CORNEAL THICKNESS IN NEWBORNS AND CHILDREN :  Is similar to that of adult cornea.  It has been found that cornea on day one is significantly thicker and decreases in thickness as child grows older.  The average corneal thickness in infants is 585±52 microns.  The superior peripheral cornea is thinnest in newborn.
  • 6.  Normal Values  Normal corneal thickness – 500-575 microns.  Abnormal results  Abnormally thick or thin measurements may indicate –  corneal thinning.  Corneal edema  Lower or higher than suspected IOP.
  • 7. FACTORS AFFECTING CCT  The mean CCT in black children is thinner than that of white children.  CCT does not correlate with refraction or systemic hypertension.
  • 8. TECHNIQUES ULTRASONIC OPTICAL ALTERNATIVE MEASUREMENTS • Conventional ultrasonic pachymetry Manual Pentacam • Ultrasound biomicroscpy (UBM). Specular microscopy • Pachycam Optical coherence Tomography(OCT)
  • 9. 1.ULTRASONIC PACHYMETRY  Broadly accepted as method of choice, regarded as the gold standard.  PRINCIPLE :  Instruments functions by measuring the amount of time (transmit time) needed for ultrasound pulse pass from one end of transducer to descemet’s membrane and back to transducer.  Corneal thickness=transmit time × propagation velocity)/2  Speed of sound in cornea : current standard is 1640 m/sec
  • 10. COMPONENTS :  PROBE HANDLE - It has piezoelectric crystal that emits an ultrasonic beam of 20 MHz  TRANSDUCER – It sends ultrasound rays through the probe to the cornea and receives echoes from the cornea.  TIP - The diameter of the tip should not be more than 2mm.
  • 11.
  • 12. ADVANTAGES DISADVANTAGES • Faster ,simpler and easy to use. • Accuracy is dependent on the perpendicularity of the probe’s application to the cornea. Portable Contact method. • Dry(no coupling medium required) • Can be used intraoperatively. Not accurate in edematous cornea.
  • 13. 2. ULTRASOUND BIOMICROSCOPY:  High resolution ultrasound machine which images anterior segment of the eye.  Using a 50-MHz imaging probe  images of the cross-sectional anterior chamber anatomy with high resolution are achieved.  12.5–50 MHz probe , the depth of penetration is lesser (4mm)than conventional ultrasound.
  • 14. ADVANTAGES DISADVANTAGES • Useful in cases where cornea is opaque The main limitation of UBM is that it requires a water-bath coupling media and a very experienced examiner. • Various layers of cornea can be identified. • Contact method and patient to lie supine • Cannot be used intraoperatively.
  • 15. 3.MANUAL OPTICAL PACHYMETER:  This was the original method to measure corneal thickness.  The instrument contains two Plano glass plates that splits the image of the corneal parallelepiped.  There are two methods to measure corneal thickness.  Just touch method.  Overlap method.  The corneal thickness is then directly read from the scale.  Range is 0-1.2mm, with a least gradation of 0.02 mm.
  • 16. ADVANTAGE DISADVANTAGE Non contact method. • Lack of accuracy in measurements, the usual range of error with an optical pachymeter is ±2% • Lack of repeatability. • Requires slit lamp and therefore has poor portability and cannot be used in operating room.
  • 17. 4. SPECULAR PACHYMETRY :  This is the oldest method to measure corneal thickness.  PRINCIPLE : This measures the distance between the anterior and the posterior surfaces of cornea and depends on the focusing of light rays through front back cornea unlike sound waves in ultrasound pachymeter.  2 types 1.contact 2.Non contact
  • 18.
  • 19.
  • 20. ADVANTAGES DISADVANTAGES Operator independent. The exact point where reading is taken is not known. Non invasive. • Time consuming. Simultaneous measurement of cell count. • Impractical to use in operation room.
  • 21. 6. ANTERIOR SEGMENT OPTICAL COHERANCE TOMOGRAPHY :  High resolution, non-contact OCT customized anterior segment.  Gives color coded map of corneal thickness.
  • 22.
  • 23. ADVANTAGES  Non contact  Rapid acquisition during the pachymetry scan ensures an accurate and repeatable pachymetry map.  High resolution  It measures and documents both corneal flap thickness and residual stromal thickness immediately following LASIK surgery.  Measures through corneal opacity.
  • 24. 10. PENTACAM It is3-dimensional (3D) rotating scheimpflug. It can perform five functions in 2 sec. 1. Scheimpflug image of anterior segment. 2. Pachymetry 3. 3-D anterior chamber analyzer 4. Corneal topography. 5. Cataract analyzer  Pachymetry by pentacam is displayed as a color image over its entire area from limbus to limbus..
  • 25. ADVANTAGE DISADVANTAGE Non invasive, non contact gives precise representation and repeatability. Even minute eye movements are captured and corrected simultaneously. ▪ It underestimates the corneal thickness in comparison to ultrasonic pachymetry.
  • 26. PACHYCAM  compact and portable non- contact pachymeter with built-in keratometer. It can be mounted on slit lamp.  Image acquisition is done with the help of a 3D alignment screen.
  • 27.  Advantages 1. Noncontact 2. Compact,portable,light weight