SlideShare une entreprise Scribd logo
1  sur  6
Télécharger pour lire hors ligne
Application
Note:
AP-MIB-VARIO12-0108
Comparison of Flash and Glow ATP
Assays with Thermo Scientific Varioskan
Flash Luminometry
Jorma Lampinen, Marika Raitio, Arto Perälä, Vuokko Kytöniemi and Reija-Riitta Harinen
Thermo Fisher Scientific Oy, Vantaa, Finland
Introduction
One of the most common lumi-
nescent reactions used in research
is an ATP reaction catalysed by
firefly luciferases. In this reaction
ATP is degraded to AMP and the
energy is released as visible light.
ATP reaction can be used for the
measurement of ATP (1-3), its deg-
radation products ADP and AMP
(4) or any enzyme activity that
can be linked to ATP production
or consumption (5). It can also be
used for the measurement of firefly
enzyme activity in applications
where firefly luciferase gene is used
as a reporter gene (6-9).
Light production reaction cata-
lyzed by firefly is an oxidation
reaction where a special substrate,
D-luciferin, is oxidized and ATP is
reduced to AMP:
	
Luciferase
ATP + D-luciferin + O2 ------------>
Mg2+
Oxyluciferin + AMP + PPi + CO2
+ light (λmax 562 nm (10-11))
	
When this reaction is used to mea-
sure ATP, all other substrates are
kept in saturated concentrations,
therefore ATP is the limiting com-
ponent of the reaction and light
production is directly proportion-
al to the concentration of ATP.
Firefly-ATP reaction can be
controlled two different ways
in common applications. Light
production can be either fast flash
type reaction or a stable glow
type reaction, depending on the
assay environment. Natural wild
type firefly-ATP reaction repre-
sents this fast flash reaction and
that kind of light production is
obtained when just substrates and
luciferase are added in the reac-
tion mixture (1-2). However, this
reaction type is not very conve-
nient for applications and there-
fore the reaction is quite often
chemically stabilized to produce
stable long term light emission
(12-13). Most of the commercial
ATP assay kits use this stabilized
ATP glow reaction but lately
certain kits using ATP flash reac-
tion have been introduced. This
flash reaction offers improved
sensitivity in ATP measurement
but has a disadvantage of requir-
ing automatic dispensers for the
substrate addition. In this applica-
tion note, the basic performance
of Varioskan Flash microplate
luminometer was tested with both
glow and flash type ATP assay.
Materials and Experimental
protocols
Kits and reagents
Two luminometric ATP kits were
chosen for the evaluation: ATP
Biomass Kit HS (BioThema AB,
Sweden) representing glow type
ATP reaction, and CheckLiteTM
HS Set (Kikkoman Corp. Japan)
that is based on flash type ATP
reaction. Both kits were used
according to manufacturer’s
instructions. All tests were per-
formed with commercial
10 umol/L ATP standard
(BioThema AB, Sweden).
Instrumentation
All measurements were performed
with Thermo Scientific Varioskan
Flash multitechnology microplate
readers that were equipped with
Varioskan LumiSens high sensitivi-
ty luminometric module and auto-
matic dispenser. Instruments were
controlled using Thermo Scientific
SkanIt software.
Plastics
All measurements were per-
formed on white 96- or 384
well microplates and ATP glow
reaction was measured also
with white 1536-well plates.
The 96-well plates used were
Microlite1 96-well 12-strip plates,
384-well plates were Microlite1+
384 square well plates (both
Thermo Scientific.) and 1536-well
plates were Lumitrac 600 HiBase
1536-well plates from Greiner
Bio-One.
ATP Reaction Kinetics
ATP reaction kinetics of both flash
and glow ATP light emission reac-
tions were measured with white
384 well plates. A 30 ul aliquot of
0.01 µM ATP standard was added
into the well and plate was placed
into the microplate luminometer.
An equal volume of either flash
or glow ATP reagent was added
with the automatic dispenser and
luminescence signal was imme-
diately measured with 1 second
integration time for 10 min (flash
type ATP reaction) or with 500 ms
integration time for 60 min (glow
type ATP reaction). For the kinetic
measurements with ATP flash
chemistry, Varioskan Flash was
programmed as follows:
•	 STEP 1: Well loop step,
Execution type “by well”, Well
count 1.
Sets the instrument to perform the
following actions well by well. All
following steps are executed for
one well, then next well and so on
until all wells have been measured.
•	 STEP 2 (inside well loop):
Kinetic loop step “KineticLoop1”,
Readings 600, Interval 0s.
This sets the instrument to per-
form kinetic assay with 600
kinetic points. Kinetic reading is
performed as fast as possible with-
out any interval time between the
readings.
•	 STEP 3 (inside kinetic loop
1): Dispensing step, Dispenser
1, Dispensing volume 30 µl (for
384-well plates), Dispensing speed
“High”, Dispensing position
“Luminometric 1”, Tip priming
“No”, Dispense at reading 1.
Dispenses 30 µl of luciferin-
luciferase reagent into the well
with high speed. Dispensing is
performed simultaneously with the
first kinetic reading point.
•	 STEP 4 (inside kinetic loop 1):
Measurement step, Luminometric
measurement, Luminometric
optics “Normal”,  Dynamic range
“AutoRange”, Measurement time
1000 ms.
Performs kinetic measurement with
1000 ms signal integration time,
sampling speed 60 points/ minute.
Total kinetic measurement time is
10 min.
Step structure of the protocol is
shown in Figure 1.
When glow type ATP chemistry
was used, a different measure-
ment protocol was used, the pro-
tocol included following actions:
•	 STEP 1: Kinetic loop step
“KineticLoop1”, Readings 60,
Interval 1 min.
This sets the instrument to per-
form kinetic assay with 60 kinetic
points with the interval time of
one minute between sequential
readings of the same well.
•	 STEP 2: (inside kinetic loop
1): Dispensing step, Dispenser
1, Dispensing volume 30 µl (for
384-well plates), Dispensing speed
“High”, Dispensing position
“Luminometric 1”, Tip priming
“No”, Dispense at reading 1.
Dispenses 30 µl of luciferin-
luciferase reagent into all sample
wells in the plate with high speed.
The first kinetic measurement is
taken simultaneously with the
dispensing.
•	 STEP 3: (inside kinetic loop 1):
Measurement step, Luminometric
measurement, Luminometric
optics “Normal”,  Dynamic range
“AutoRange”, Measurement time
500 ms.
Performs kinetic measurement
with 500 ms integration time for
all sample wells, then waits until
defined one minute interval is
filled and then repeats the mea-
surement for 60 rounds.
Step structure of the protocol is
shown in Figure 2.
ATP Sensitivity Tests
ATP sensitivity was determined
for a set of Varioskan Flash units.
Number of Varioskan Flash units
used in each test is shown in
Table I. An ATP standard dilu-
tion series was prepared between
0.1 pM to 1 µM with 10-fold
dilutions. With 384-well plates
30 µl of each ATP dilution or
blank solution was pipetted into
the microplate wells and 30 µl of
luciferin-luciferase reagent from
either ATP flash or glow kit was
added with the instrument’s dis-
penser. For 96-well plates 50 µl
of ATP dilution or blank solution
and 50 µl of luciferin-luciferase
reagent and for 1536-well plates
5 µl of both ATP dilution and
reagent was used. Test included
with all plate types 16 replicates
of the blanks and eight replicates
of the ATP dilutions. With flash
type ATP reaction luminescence
of each well was measured after
two second delay time with
four second signal integration.
Measurement protocol used in
this assay had following proce-
dural steps:
•	 STEP 1: Well loop step,
Execution type “by well”, Well
count 1.
Sets the instrument to perform
the following actions well by
well. All following steps are
executed for one well, then next
well and so on until all wells have
been measured.
•	 STEP 2: Dispensing step,
Dispenser 1, Dispensing vol-
ume 30 µl (for 384-well
plates), Dispensing speed
“High”, Dispensing position
“Luminometric 1”, Tip priming
“No”.
Dispenses 30 µl of luciferin-
luciferase reagent into one sample
well with high speed.
•	 STEP 3: Measurement step,
Luminometric measurement,
Luminometric optics “Normal”,  
Dynamic range “AutoRange”,
Measurement time 4000 ms, Lag
time 2 s.
This step first waits for 2 s delay
time until the signal has been
generated and then it measures
the luminescent signal for 4000
ms. Resulting luminescence result
is calculated and given for 1 sec-
ond integration time.
Step structure of the protocol is
shown in Figure 3.
Part of Thermo Fisher Scientific
Figure 1. Protocol steps for the kinetic measure-
ment of flash ATP reaction.
Figure 2. Protocol steps for the kinetic meas-
urement of glow ATP reaction.
Part of Thermo Fisher Scientific
When glow type ATP reaction
was measured, all wells were
first dispensed with the luciferin-
luciferase reagent and signal was
measured with one second inte-
gration time from the wells after
two minute waiting period.
•	 STEP 1: Dispensing step,
Dispenser 1, Dispensing vol-
ume 30 µl (for 384-well
plates), Dispensing speed
“High”, Dispensing position
“Luminometric 1”, Tip priming
“No”.
Dispenses 30 µl of luciferin-
luciferase reagent into all sample
wells in the plate with high
speed.
•	 STEP 2: Measurement step,
Luminometric measurement,
Luminometric optics “Normal”,  
Dynamic range “AutoRange”,
Measurement time 1000 ms, Lag
time 2 min.
Step will wait two minutes at the
beginning of the step, then all
wells are measured with 1000 ms
integration time.
Step structure of the protocol is
shown in Figure 4.
Theoretical assay sensitivity was
calculated as a crossing point
between the blank subtracted
ATP calibration curve and 2*SD
level of assay blank wells and
Z-prime values were calculated
according to standard formulas
(14).
Results and Discussion
ATP Reaction Kinetics
As expected, flash and glow type
luminescence from the ATP reac-
tion show totally different kinetic
behaviour. Typical kinetic lumines-
cence curves are shown in Figure 5.
Flash type ATP luminescence reac-
tion shown in Figure 5A has very
fast signal generation where peak
maximum is reached in about four
seconds after addition of luciferin-
luciferase reagent and the signal
decays quite rapidly, showing little
below two minutes half-life time.
Therefore, this assay was measured
by integrating a signal from two
to six seconds, where maximal
signal is emitted by the reaction.
Because of this fast signal increase
and decrease, this type of ATP reac-
tion can be measured only with the
instrument equipped with auto-
matic dispenser module. If luciferin-
luciferase reagent would be added
into the ATP sample manually
outside the instrument, the signal
maximum would be totally lost
before any measurements could be
done. In addition, time difference
between dispensing the luciferin-
luciferase reagent and measuring
the signal has to be constant with
rather high precision for each well,
otherwise incorrect results would
be obtained because of signal
instability.
Glow type ATP luminescence reac-
tion in Figure 5B on the other
hand has totally different reaction
kinetics. Luminescence maximum
is reached within the minute after
ATP and luciferin-luciferase are
mixed but reaction stays stable for
much longer time period. After an
hour, there is still more than 70%
of the signal present. Therefore,
when this glow reaction is used to
measure ATP, it is possible to add
all components manually and read
the plate afterwards. This will not
cause remarkable variations in the
result values.
Figure 3. Protocol steps used to measure ATP
sensitivity assay with flash chemistry.
Figure 4. Protocol steps used to measure ATP
sensitivity assay with glow chemistry.
Figure 5. Kinetic curves of two luminescent ATP reactions. A) ATP flash reaction. B) ATP glow reaction.
Part of Thermo Fisher Scientific
ATP Sensitivity Tests
Sensitivity of these two ATP assays
was determined with different plate
formats, glow type ATP sensitiv-
ity with all three common plate
formats: 96-, 384- and 1536-well
plates and flash type assay only
with 96- and 384-well plates.
	
Results of these tests are collected
in Table I and examples of typical
ATP standard curves are shown in
Figure 6. Varioskan Flash instru-
ment is designed to have optimal
performance with 384-well plates
and that is seen from these results.
Both ATP assays using either flash
or glow type ATP chemistry show
the best sensitivity with 384-well
plates. The detection limit of ATP
with flash type chemistry and with
this plate format is about 1.5 amol/
well and with glow type chemistry
about 80 amol/well. This around
50 times sensitivity difference is
very understandable because of the
difference between the two chemis-
tries. With flash type ATP reaction,
the light from the chemical reaction
is released within the short period
of time, therefore average signal
over time is much higher. This is
seen also from the Figure 6, the
same amount of ATP in the well
produces about 30 times more light
in flash reaction. For example ATP
sample of 300 000 amol/well gives
about 5 000 000 RLU in flash
reaction when glow reaction gives
only 150 000 RLU with the same
amount of ATP.
When the plate format is changed
from 384-wells to 96-wells, assay
sensitivity is decreased. This effect
is not remarkable with flash type ATP
assay, only about 20% decrease in the
detection limit, but with glow type
ATP assay the sensitivity is decreased
almost 10 times. Therefore the dif-
ference between flash and glow assay
sensitivity with 96-well plates is much
bigger with 96-well plates than with
384-well plates, being about 400
times.
The very high density 1536-well
plate was tested only with
glow type ATP chemistry and
it was shown to produce about
the same ATP sensitivity than
96-well plate, having the detec-
tion limit value of 200 amol/well
ATP, both slightly lower than
with 384-well plate.
Plate
format
ATP Flash chemistry ATP Glow Chemistry
Glow/
Flash
ratio
Average detection
limit (amol/well))
Range, (Min-Max) Number of Varioskan
Flash units
Average detection
limit (amol/well))
Range, (Min-
Max)
Number of Varioskan
Flash units
96 1.8 0.8 – 2.6 7 130 600 - 900 4 409
384 1.5 0.5 – 2.9 30 80 60 - 130 6 52
1536 -- -- -- 200 180 - 210 2 --
Figure 6. Typical ATP calibration curves with different plate formats. A) Flash type ATP assay. B)
Glow type ATP assay.
Table I. Sensitivity test results of flash and glow ATP assays with Varioskan Flash.
Part of Thermo Fisher Scientific
All sensitivity tests were performed
with several Varioskan Flash units.
As expected, there are differences
between the performances of the
individual instruments in these
ATP assays.  Every instrument is
always individual in its perfor-
mance because there are always
some differences in electronics,
noise levels etc. The total perfor-
mance range in the detection limits
between the instruments is any-
how amazingly narrow. When the
average ATP detection limit with
the flash chemistry and 384-well
plates is for example 1.8 amol/
well, all tested 30 instrument units
had their detection limit below 3
amol/well ATP. Figure 7. shows
the cumulative distribution of
these detection limit values of the
individual Varioskan Flash instru-
ments. This distribution is clearly
following expected cumulative
Gaussian distribution.
Another very important assay
quality parameter is Z-prime
value. It is a numerical calculation
that reflects the assay capability
to separate positive results from
the negative ones (14). Theoretical
maximum of the Z-prime value is
1.0 that suggests that reliability
of the separation of two signals
is 100.0%. This is of course
never possible to be met in real
life because there is always some
standard deviation in the signal
values. If the Z-prime value is over
0.5, that assay can be normally
considered good and reliable (14)
but sometimes values around
0.6 or higher might be required.
Calculated Z-prime values against
the assay blanks over the whole
ATP calibration series are shown
for both flash and glow ATP
chemistries and all plate formats in
Figure 8.
Figure 8. clearly shows the excel-
lent reliability of these ATP assays.
With the flash type ATP assay,
that 3 amol/well ATP sample
that is very close to the theoreti-
cal detection limit shows already
almost sufficient Z-prime values.
Theoretical detection limit is
calculated based on blank SD
only when Z-prime calculation
formula includes both blank and
standard SD values. Therefore, cal-
culated theoretical detection limit
is always below the concentration
range where good Z-prime values
are obtained. With the ATP sample
of 30 amol/well Z-prime values
are clearly over required 0.5
level, with both 96- and 384-well
formats.
When the Z-prime values of
glow type ATP assay are ana-
lyzed, it is immediately noticed
that higher ATP amounts than
with flash chemistry are required
for the good z-prime level. About
300 amol/well ATP is required
before calculated Z-prime val-
ues are constantly over required
level.
Figure 7. Cumulative distribution of the Varioskan Flash detection limits in flash type ATP assay
with 384-well plates.
Figure 8. The Z-prime values of both flash and glow type ATP assays measured with 96-, 384- and
1536-well plate formats.
Part of Thermo Fisher Scientific
Conclusions
These results clearly show the high
sensitivity and reliability of ATP
assays with Varioskan Flash lumi-
nometry. ATP assay using flash type
chemistry gives better assay sensitivi-
ty than assays using glow type chem-
istry and 384-well plate format gives
the best sensitivity with both types
of chemistries. For the statistically
reliable results, around 10-30 amol
of ATP is required in the sample
with flash type ATP assay and about
300 -500 amol/well is required for
the glow type assay.
References
1.	 Lundin, A., et al. 1976. Continuous
monitoring of ATP-converting reactions by
purified firefly luciferase. Anal. Biochem.
75, pp. 611 - 620.
2.	 Webster, J.J. and Leach, F. 1980.
Optimization of the firefly luciferase assay
for ATP. J. Applied Biochem. 2, pp. 469 -
79.
3.	 Karl, D.M. 1980. Cellular nucleotide
measurements and applications in micro-
bial ecology. Microbiol. Rev. 44, pp. 739
- 96.
4.	 Lundin, A. et al. 1986. Estimation
of biomass in growing cell lines by
adenosine triphosphate assay. Methods in
Enzymology 133, pp. 27 - 42.
5.	 Fan, F. and Wood K.V. 2007.
Bioluminescent Assays for High-
Throughput Screening. Assay and Drug
Development Technologies. 5. pp.
127-136.
6.	 Ow, D.W. et al. 1986. Transient and
stable expression of the firefly luciferase
gene in plant cells and transgenic plants.
Science 234, pp. 856 - 859.
7.	 De Wet, J.R. et al. 1987. Firefly
luciferase gene: Structure and expression in
mammalian cells. Mol. Cell. Biol. 7, 725 -
737.
8.	 Brasier, A.R. et al. 1989. Optimized use
of firefly luciferase assay as a reporter gene
in mammalian cell lines. BioTechniques. 7,
pp. 1116 - 1122.
9.	 Koncz, C. et al. 1990. Bacterial and
Firefly Luciferase Genes in Transgenic
Plants: Advantages and Disadvantages of
a Reporter Gene. Dev. Genet. 11, pp. 224
- 232.
10.	 Seliger, H.H. et al. 1964. The Spectral
Distribution of Firefly Light. J. Gen. Physiol.
48, pp. 95-104.
11.	 Ugarova, N.N. and L.Y. Brovko. 2002.
Protein structure and bioluminescent spectra
for firefly bioluminescence. Luminescence
17, pp. 321-330.
12.	 Lundin, et al. 1982. Optimized biolu-
minescence assay of creatine kinase and cre-
atine kinase B-subunit activity. Clin. Chem.
28, pp. 609-614.
13.	 Wood, K V. et al. 1991. Recent
Advances and Prospects for Use of Beetle
Luciferases as Genetic Reporters. In:
Bioluminescence & Chemiluminescence:
Current Status, Proceedings of the
VIth International Symposium on
Bioluminescence and Chemiluminescence,
Stanley, P. E., et al., Editors, John Wiley &
Sons, Chichester, United Kingdom. Pp. 543-
546.
14.	 Carter, J.M. 2003. A Guide to Assay
Development. D&MD Publications. Pp.
12/10 – 12/11.
North America:
USA / Canada
+1 800 522 7763
Europe:
Austria
+43 1 801 40 0
Belgium
+32 2 482 30 30
France
+33 2 2803 2000
Germany national toll free
08001-536 376
Germany international
+49 6184 90 6940
Italy
+39 02 02 95059 434-254
Netherlands
+31 76 571 4440
Nordic countries
+358 9 329 100
Russia/CIS
+7 (812) 703 42 15
Spain/Portugal
+34 93 223 09 18
Switzerland
+41 44 454 12 12
UK/Ireland
+44 870 609 9203
Asia:
China
+86 21 6865 4588 or
+86 10 8419 3588
India
+91 22 6716 2200
Japan
+81 45 453 9220
Other Asian countries
+852 2885 4613
Countries not listed:
+49 6184 90 6940 or
+33 2 2803 2000
www.thermo.com/
readingroom
www.thermo.com/mpi
In addition to these offices,
Thermo Fisher Scientific
maintains a network of
representative organizations
throughout the world.
© 2008 Thermo Fisher Scientific
Inc. All rights reserved. All
trademarks are the property of
Thermo Fisher Scientific Inc. and
its subsidiaries. Specifications,
terms and pricing are subject to
change. Not all products are
available in all countries. Please
consult your local sales repre-
sentative for details.

Contenu connexe

Tendances

PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...PerkinElmer, Inc.
 
gass chromatography(dr.islam).docx
gass chromatography(dr.islam).docxgass chromatography(dr.islam).docx
gass chromatography(dr.islam).docxZORAIZ HAIDER
 
Almac: Development and Scale-up of a Photochemical Route to a Steroidal API
Almac: Development and Scale-up of a Photochemical Route to a Steroidal APIAlmac: Development and Scale-up of a Photochemical Route to a Steroidal API
Almac: Development and Scale-up of a Photochemical Route to a Steroidal APISusan1Beattie
 
C13 NUCLEAR MAGNETIC RESONANCE
C13 NUCLEAR MAGNETIC RESONANCE C13 NUCLEAR MAGNETIC RESONANCE
C13 NUCLEAR MAGNETIC RESONANCE Sujitlal Bhakta
 
ResearchStatement2016_v2
ResearchStatement2016_v2ResearchStatement2016_v2
ResearchStatement2016_v2Feng-Kuo Hsu
 

Tendances (10)

PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
 
ICP-MS
ICP-MSICP-MS
ICP-MS
 
gass chromatography(dr.islam).docx
gass chromatography(dr.islam).docxgass chromatography(dr.islam).docx
gass chromatography(dr.islam).docx
 
Water Analysis using LAMBDA
Water Analysis using LAMBDAWater Analysis using LAMBDA
Water Analysis using LAMBDA
 
LC-MS
LC-MSLC-MS
LC-MS
 
qHNMR for purity determination
qHNMR for purity determinationqHNMR for purity determination
qHNMR for purity determination
 
Almac: Development and Scale-up of a Photochemical Route to a Steroidal API
Almac: Development and Scale-up of a Photochemical Route to a Steroidal APIAlmac: Development and Scale-up of a Photochemical Route to a Steroidal API
Almac: Development and Scale-up of a Photochemical Route to a Steroidal API
 
C13 NUCLEAR MAGNETIC RESONANCE
C13 NUCLEAR MAGNETIC RESONANCE C13 NUCLEAR MAGNETIC RESONANCE
C13 NUCLEAR MAGNETIC RESONANCE
 
ResearchStatement2016_v2
ResearchStatement2016_v2ResearchStatement2016_v2
ResearchStatement2016_v2
 
No ions
No ionsNo ions
No ions
 

En vedette

програма обслуговування дисків
програма обслуговування дисківпрограма обслуговування дисків
програма обслуговування дисківAnya Vasilkovskaya
 
Основи алгоритмізації
Основи алгоритмізаціїОснови алгоритмізації
Основи алгоритмізаціїvetochka16
 
Dreissenid Mussel Monitoring for the Great Lakes Nutrient Initiative (GLNI) 2...
Dreissenid Mussel Monitoring for the Great Lakes Nutrient Initiative (GLNI) 2...Dreissenid Mussel Monitoring for the Great Lakes Nutrient Initiative (GLNI) 2...
Dreissenid Mussel Monitoring for the Great Lakes Nutrient Initiative (GLNI) 2...Zach Leslie
 
Refinement seminaari 13_11_15_grigori_joffe
Refinement seminaari 13_11_15_grigori_joffeRefinement seminaari 13_11_15_grigori_joffe
Refinement seminaari 13_11_15_grigori_joffeJohanna Cresswell-Smith
 
Brett_Kobold_Dont_Be_Miserable_Final
Brett_Kobold_Dont_Be_Miserable_FinalBrett_Kobold_Dont_Be_Miserable_Final
Brett_Kobold_Dont_Be_Miserable_FinalBrett S. Kobold
 
Virtual symposium c christophersen
Virtual symposium c christophersenVirtual symposium c christophersen
Virtual symposium c christophersenCeciRabbit
 
Основи алгоритмізації
Основи алгоритмізаціїОснови алгоритмізації
Основи алгоритмізаціїvetochka16
 
New features in java se 8
New features in java se 8New features in java se 8
New features in java se 8melaniamitisor
 
Ms2 coursework
Ms2 courseworkMs2 coursework
Ms2 courseworkchloekeogh
 
TDCP Jeep Rally 2016
TDCP Jeep Rally 2016TDCP Jeep Rally 2016
TDCP Jeep Rally 2016MusaTDCP
 

En vedette (20)

програма обслуговування дисків
програма обслуговування дисківпрограма обслуговування дисків
програма обслуговування дисків
 
Kaptewn
KaptewnKaptewn
Kaptewn
 
Derecho de los Negocios
Derecho de los NegociosDerecho de los Negocios
Derecho de los Negocios
 
Основи алгоритмізації
Основи алгоритмізаціїОснови алгоритмізації
Основи алгоритмізації
 
Diapositivas leyton
Diapositivas leytonDiapositivas leyton
Diapositivas leyton
 
Derecho de los Negocios
Derecho de los NegociosDerecho de los Negocios
Derecho de los Negocios
 
Dreissenid Mussel Monitoring for the Great Lakes Nutrient Initiative (GLNI) 2...
Dreissenid Mussel Monitoring for the Great Lakes Nutrient Initiative (GLNI) 2...Dreissenid Mussel Monitoring for the Great Lakes Nutrient Initiative (GLNI) 2...
Dreissenid Mussel Monitoring for the Great Lakes Nutrient Initiative (GLNI) 2...
 
Refinement seminaari 13_11_15_grigori_joffe
Refinement seminaari 13_11_15_grigori_joffeRefinement seminaari 13_11_15_grigori_joffe
Refinement seminaari 13_11_15_grigori_joffe
 
Brett_Kobold_Dont_Be_Miserable_Final
Brett_Kobold_Dont_Be_Miserable_FinalBrett_Kobold_Dont_Be_Miserable_Final
Brett_Kobold_Dont_Be_Miserable_Final
 
Abraham Sanni CV
Abraham Sanni CVAbraham Sanni CV
Abraham Sanni CV
 
Virtual symposium c christophersen
Virtual symposium c christophersenVirtual symposium c christophersen
Virtual symposium c christophersen
 
Основи алгоритмізації
Основи алгоритмізаціїОснови алгоритмізації
Основи алгоритмізації
 
SIP_Rajasekar
SIP_RajasekarSIP_Rajasekar
SIP_Rajasekar
 
New features in java se 8
New features in java se 8New features in java se 8
New features in java se 8
 
Ms2 coursework
Ms2 courseworkMs2 coursework
Ms2 coursework
 
Presentación1
Presentación1Presentación1
Presentación1
 
My 5 minutes bell
My 5 minutes bellMy 5 minutes bell
My 5 minutes bell
 
Cmm ingles imagenes
Cmm ingles imagenesCmm ingles imagenes
Cmm ingles imagenes
 
TDCP Jeep Rally 2016
TDCP Jeep Rally 2016TDCP Jeep Rally 2016
TDCP Jeep Rally 2016
 
Purva pravesh
Purva praveshPurva pravesh
Purva pravesh
 

Similaire à Comparison of Flash & Glow ATP

Versatile Luciferases for Reporter Assays
Versatile Luciferases for Reporter AssaysVersatile Luciferases for Reporter Assays
Versatile Luciferases for Reporter AssaysArto Per
 
methods_of_enzyme_assay_4.pdf
methods_of_enzyme_assay_4.pdfmethods_of_enzyme_assay_4.pdf
methods_of_enzyme_assay_4.pdfsaraso888
 
Pharmacheutical impurities analysis
Pharmacheutical impurities analysisPharmacheutical impurities analysis
Pharmacheutical impurities analysisBejoUtomo4
 
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Kate?ina Svobodov
 
A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Co...
A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Co...A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Co...
A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Co...PerkinElmer, Inc.
 
Amino acid analysis and peptide mapping Likhith K
Amino acid analysis  and peptide mapping Likhith KAmino acid analysis  and peptide mapping Likhith K
Amino acid analysis and peptide mapping Likhith KLIKHITHK1
 
electro_enzyme_assay.pptx
electro_enzyme_assay.pptxelectro_enzyme_assay.pptx
electro_enzyme_assay.pptxCobraPrsj
 
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...Sandy Simmons
 
Anachem-trans-Introduction-to-Analytical-Techniques.docx
Anachem-trans-Introduction-to-Analytical-Techniques.docxAnachem-trans-Introduction-to-Analytical-Techniques.docx
Anachem-trans-Introduction-to-Analytical-Techniques.docxJamaicaSalvador
 
Novel Approaches to Omega-3 Stability Testing
Novel Approaches to Omega-3 Stability Testing Novel Approaches to Omega-3 Stability Testing
Novel Approaches to Omega-3 Stability Testing Nutrasource
 
SCU LC-MS Agilent 6100 training_060710
SCU LC-MS Agilent 6100 training_060710SCU LC-MS Agilent 6100 training_060710
SCU LC-MS Agilent 6100 training_060710pead
 
Combining Reichert’s SPR Systems With Other Informative Techniques
Combining Reichert’s SPR Systems With Other Informative TechniquesCombining Reichert’s SPR Systems With Other Informative Techniques
Combining Reichert’s SPR Systems With Other Informative TechniquesReichertSPR
 
Automated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPAutomated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPMark Hayward
 
Method Development Poster-MicroLiter
Method Development Poster-MicroLiterMethod Development Poster-MicroLiter
Method Development Poster-MicroLiterRick Youngblood
 
FORMULATION AND EVALUATION OF ATORVASTATIN CALCIUM NIOSOMES.pptx
FORMULATION AND EVALUATION OF ATORVASTATIN CALCIUM NIOSOMES.pptxFORMULATION AND EVALUATION OF ATORVASTATIN CALCIUM NIOSOMES.pptx
FORMULATION AND EVALUATION OF ATORVASTATIN CALCIUM NIOSOMES.pptxshivalisingla5
 

Similaire à Comparison of Flash & Glow ATP (20)

Versatile Luciferases for Reporter Assays
Versatile Luciferases for Reporter AssaysVersatile Luciferases for Reporter Assays
Versatile Luciferases for Reporter Assays
 
methods_of_enzyme_assay_4.pdf
methods_of_enzyme_assay_4.pdfmethods_of_enzyme_assay_4.pdf
methods_of_enzyme_assay_4.pdf
 
Pharmacheutical impurities analysis
Pharmacheutical impurities analysisPharmacheutical impurities analysis
Pharmacheutical impurities analysis
 
Uplc
UplcUplc
Uplc
 
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
 
A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Co...
A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Co...A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Co...
A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Co...
 
Amino acid analysis and peptide mapping Likhith K
Amino acid analysis  and peptide mapping Likhith KAmino acid analysis  and peptide mapping Likhith K
Amino acid analysis and peptide mapping Likhith K
 
electro_enzyme_assay.pptx
electro_enzyme_assay.pptxelectro_enzyme_assay.pptx
electro_enzyme_assay.pptx
 
أيات
أياتأيات
أيات
 
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
 
Anachem-trans-Introduction-to-Analytical-Techniques.docx
Anachem-trans-Introduction-to-Analytical-Techniques.docxAnachem-trans-Introduction-to-Analytical-Techniques.docx
Anachem-trans-Introduction-to-Analytical-Techniques.docx
 
Novel Approaches to Omega-3 Stability Testing
Novel Approaches to Omega-3 Stability Testing Novel Approaches to Omega-3 Stability Testing
Novel Approaches to Omega-3 Stability Testing
 
Ir spectroscopy
Ir spectroscopyIr spectroscopy
Ir spectroscopy
 
SCU LC-MS Agilent 6100 training_060710
SCU LC-MS Agilent 6100 training_060710SCU LC-MS Agilent 6100 training_060710
SCU LC-MS Agilent 6100 training_060710
 
Combining Reichert’s SPR Systems With Other Informative Techniques
Combining Reichert’s SPR Systems With Other Informative TechniquesCombining Reichert’s SPR Systems With Other Informative Techniques
Combining Reichert’s SPR Systems With Other Informative Techniques
 
Automated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPAutomated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSP
 
Method Development Poster-MicroLiter
Method Development Poster-MicroLiterMethod Development Poster-MicroLiter
Method Development Poster-MicroLiter
 
Creatinine estimation
Creatinine estimationCreatinine estimation
Creatinine estimation
 
journal journals
journal journalsjournal journals
journal journals
 
FORMULATION AND EVALUATION OF ATORVASTATIN CALCIUM NIOSOMES.pptx
FORMULATION AND EVALUATION OF ATORVASTATIN CALCIUM NIOSOMES.pptxFORMULATION AND EVALUATION OF ATORVASTATIN CALCIUM NIOSOMES.pptx
FORMULATION AND EVALUATION OF ATORVASTATIN CALCIUM NIOSOMES.pptx
 

Comparison of Flash & Glow ATP

  • 1. Application Note: AP-MIB-VARIO12-0108 Comparison of Flash and Glow ATP Assays with Thermo Scientific Varioskan Flash Luminometry Jorma Lampinen, Marika Raitio, Arto Perälä, Vuokko Kytöniemi and Reija-Riitta Harinen Thermo Fisher Scientific Oy, Vantaa, Finland Introduction One of the most common lumi- nescent reactions used in research is an ATP reaction catalysed by firefly luciferases. In this reaction ATP is degraded to AMP and the energy is released as visible light. ATP reaction can be used for the measurement of ATP (1-3), its deg- radation products ADP and AMP (4) or any enzyme activity that can be linked to ATP production or consumption (5). It can also be used for the measurement of firefly enzyme activity in applications where firefly luciferase gene is used as a reporter gene (6-9). Light production reaction cata- lyzed by firefly is an oxidation reaction where a special substrate, D-luciferin, is oxidized and ATP is reduced to AMP: Luciferase ATP + D-luciferin + O2 ------------> Mg2+ Oxyluciferin + AMP + PPi + CO2 + light (λmax 562 nm (10-11)) When this reaction is used to mea- sure ATP, all other substrates are kept in saturated concentrations, therefore ATP is the limiting com- ponent of the reaction and light production is directly proportion- al to the concentration of ATP. Firefly-ATP reaction can be controlled two different ways in common applications. Light production can be either fast flash type reaction or a stable glow type reaction, depending on the assay environment. Natural wild type firefly-ATP reaction repre- sents this fast flash reaction and that kind of light production is obtained when just substrates and luciferase are added in the reac- tion mixture (1-2). However, this reaction type is not very conve- nient for applications and there- fore the reaction is quite often chemically stabilized to produce stable long term light emission (12-13). Most of the commercial ATP assay kits use this stabilized ATP glow reaction but lately certain kits using ATP flash reac- tion have been introduced. This flash reaction offers improved sensitivity in ATP measurement but has a disadvantage of requir- ing automatic dispensers for the substrate addition. In this applica- tion note, the basic performance of Varioskan Flash microplate luminometer was tested with both glow and flash type ATP assay. Materials and Experimental protocols Kits and reagents Two luminometric ATP kits were chosen for the evaluation: ATP Biomass Kit HS (BioThema AB, Sweden) representing glow type ATP reaction, and CheckLiteTM HS Set (Kikkoman Corp. Japan) that is based on flash type ATP reaction. Both kits were used according to manufacturer’s instructions. All tests were per- formed with commercial 10 umol/L ATP standard (BioThema AB, Sweden). Instrumentation All measurements were performed with Thermo Scientific Varioskan Flash multitechnology microplate readers that were equipped with Varioskan LumiSens high sensitivi- ty luminometric module and auto- matic dispenser. Instruments were controlled using Thermo Scientific SkanIt software. Plastics All measurements were per- formed on white 96- or 384 well microplates and ATP glow reaction was measured also with white 1536-well plates. The 96-well plates used were Microlite1 96-well 12-strip plates, 384-well plates were Microlite1+ 384 square well plates (both Thermo Scientific.) and 1536-well plates were Lumitrac 600 HiBase 1536-well plates from Greiner Bio-One. ATP Reaction Kinetics ATP reaction kinetics of both flash and glow ATP light emission reac- tions were measured with white 384 well plates. A 30 ul aliquot of 0.01 µM ATP standard was added into the well and plate was placed into the microplate luminometer. An equal volume of either flash or glow ATP reagent was added with the automatic dispenser and luminescence signal was imme- diately measured with 1 second integration time for 10 min (flash
  • 2. type ATP reaction) or with 500 ms integration time for 60 min (glow type ATP reaction). For the kinetic measurements with ATP flash chemistry, Varioskan Flash was programmed as follows: • STEP 1: Well loop step, Execution type “by well”, Well count 1. Sets the instrument to perform the following actions well by well. All following steps are executed for one well, then next well and so on until all wells have been measured. • STEP 2 (inside well loop): Kinetic loop step “KineticLoop1”, Readings 600, Interval 0s. This sets the instrument to per- form kinetic assay with 600 kinetic points. Kinetic reading is performed as fast as possible with- out any interval time between the readings. • STEP 3 (inside kinetic loop 1): Dispensing step, Dispenser 1, Dispensing volume 30 µl (for 384-well plates), Dispensing speed “High”, Dispensing position “Luminometric 1”, Tip priming “No”, Dispense at reading 1. Dispenses 30 µl of luciferin- luciferase reagent into the well with high speed. Dispensing is performed simultaneously with the first kinetic reading point. • STEP 4 (inside kinetic loop 1): Measurement step, Luminometric measurement, Luminometric optics “Normal”, Dynamic range “AutoRange”, Measurement time 1000 ms. Performs kinetic measurement with 1000 ms signal integration time, sampling speed 60 points/ minute. Total kinetic measurement time is 10 min. Step structure of the protocol is shown in Figure 1. When glow type ATP chemistry was used, a different measure- ment protocol was used, the pro- tocol included following actions: • STEP 1: Kinetic loop step “KineticLoop1”, Readings 60, Interval 1 min. This sets the instrument to per- form kinetic assay with 60 kinetic points with the interval time of one minute between sequential readings of the same well. • STEP 2: (inside kinetic loop 1): Dispensing step, Dispenser 1, Dispensing volume 30 µl (for 384-well plates), Dispensing speed “High”, Dispensing position “Luminometric 1”, Tip priming “No”, Dispense at reading 1. Dispenses 30 µl of luciferin- luciferase reagent into all sample wells in the plate with high speed. The first kinetic measurement is taken simultaneously with the dispensing. • STEP 3: (inside kinetic loop 1): Measurement step, Luminometric measurement, Luminometric optics “Normal”, Dynamic range “AutoRange”, Measurement time 500 ms. Performs kinetic measurement with 500 ms integration time for all sample wells, then waits until defined one minute interval is filled and then repeats the mea- surement for 60 rounds. Step structure of the protocol is shown in Figure 2. ATP Sensitivity Tests ATP sensitivity was determined for a set of Varioskan Flash units. Number of Varioskan Flash units used in each test is shown in Table I. An ATP standard dilu- tion series was prepared between 0.1 pM to 1 µM with 10-fold dilutions. With 384-well plates 30 µl of each ATP dilution or blank solution was pipetted into the microplate wells and 30 µl of luciferin-luciferase reagent from either ATP flash or glow kit was added with the instrument’s dis- penser. For 96-well plates 50 µl of ATP dilution or blank solution and 50 µl of luciferin-luciferase reagent and for 1536-well plates 5 µl of both ATP dilution and reagent was used. Test included with all plate types 16 replicates of the blanks and eight replicates of the ATP dilutions. With flash type ATP reaction luminescence of each well was measured after two second delay time with four second signal integration. Measurement protocol used in this assay had following proce- dural steps: • STEP 1: Well loop step, Execution type “by well”, Well count 1. Sets the instrument to perform the following actions well by well. All following steps are executed for one well, then next well and so on until all wells have been measured. • STEP 2: Dispensing step, Dispenser 1, Dispensing vol- ume 30 µl (for 384-well plates), Dispensing speed “High”, Dispensing position “Luminometric 1”, Tip priming “No”. Dispenses 30 µl of luciferin- luciferase reagent into one sample well with high speed. • STEP 3: Measurement step, Luminometric measurement, Luminometric optics “Normal”, Dynamic range “AutoRange”, Measurement time 4000 ms, Lag time 2 s. This step first waits for 2 s delay time until the signal has been generated and then it measures the luminescent signal for 4000 ms. Resulting luminescence result is calculated and given for 1 sec- ond integration time. Step structure of the protocol is shown in Figure 3. Part of Thermo Fisher Scientific Figure 1. Protocol steps for the kinetic measure- ment of flash ATP reaction. Figure 2. Protocol steps for the kinetic meas- urement of glow ATP reaction.
  • 3. Part of Thermo Fisher Scientific When glow type ATP reaction was measured, all wells were first dispensed with the luciferin- luciferase reagent and signal was measured with one second inte- gration time from the wells after two minute waiting period. • STEP 1: Dispensing step, Dispenser 1, Dispensing vol- ume 30 µl (for 384-well plates), Dispensing speed “High”, Dispensing position “Luminometric 1”, Tip priming “No”. Dispenses 30 µl of luciferin- luciferase reagent into all sample wells in the plate with high speed. • STEP 2: Measurement step, Luminometric measurement, Luminometric optics “Normal”, Dynamic range “AutoRange”, Measurement time 1000 ms, Lag time 2 min. Step will wait two minutes at the beginning of the step, then all wells are measured with 1000 ms integration time. Step structure of the protocol is shown in Figure 4. Theoretical assay sensitivity was calculated as a crossing point between the blank subtracted ATP calibration curve and 2*SD level of assay blank wells and Z-prime values were calculated according to standard formulas (14). Results and Discussion ATP Reaction Kinetics As expected, flash and glow type luminescence from the ATP reac- tion show totally different kinetic behaviour. Typical kinetic lumines- cence curves are shown in Figure 5. Flash type ATP luminescence reac- tion shown in Figure 5A has very fast signal generation where peak maximum is reached in about four seconds after addition of luciferin- luciferase reagent and the signal decays quite rapidly, showing little below two minutes half-life time. Therefore, this assay was measured by integrating a signal from two to six seconds, where maximal signal is emitted by the reaction. Because of this fast signal increase and decrease, this type of ATP reac- tion can be measured only with the instrument equipped with auto- matic dispenser module. If luciferin- luciferase reagent would be added into the ATP sample manually outside the instrument, the signal maximum would be totally lost before any measurements could be done. In addition, time difference between dispensing the luciferin- luciferase reagent and measuring the signal has to be constant with rather high precision for each well, otherwise incorrect results would be obtained because of signal instability. Glow type ATP luminescence reac- tion in Figure 5B on the other hand has totally different reaction kinetics. Luminescence maximum is reached within the minute after ATP and luciferin-luciferase are mixed but reaction stays stable for much longer time period. After an hour, there is still more than 70% of the signal present. Therefore, when this glow reaction is used to measure ATP, it is possible to add all components manually and read the plate afterwards. This will not cause remarkable variations in the result values. Figure 3. Protocol steps used to measure ATP sensitivity assay with flash chemistry. Figure 4. Protocol steps used to measure ATP sensitivity assay with glow chemistry. Figure 5. Kinetic curves of two luminescent ATP reactions. A) ATP flash reaction. B) ATP glow reaction.
  • 4. Part of Thermo Fisher Scientific ATP Sensitivity Tests Sensitivity of these two ATP assays was determined with different plate formats, glow type ATP sensitiv- ity with all three common plate formats: 96-, 384- and 1536-well plates and flash type assay only with 96- and 384-well plates. Results of these tests are collected in Table I and examples of typical ATP standard curves are shown in Figure 6. Varioskan Flash instru- ment is designed to have optimal performance with 384-well plates and that is seen from these results. Both ATP assays using either flash or glow type ATP chemistry show the best sensitivity with 384-well plates. The detection limit of ATP with flash type chemistry and with this plate format is about 1.5 amol/ well and with glow type chemistry about 80 amol/well. This around 50 times sensitivity difference is very understandable because of the difference between the two chemis- tries. With flash type ATP reaction, the light from the chemical reaction is released within the short period of time, therefore average signal over time is much higher. This is seen also from the Figure 6, the same amount of ATP in the well produces about 30 times more light in flash reaction. For example ATP sample of 300 000 amol/well gives about 5 000 000 RLU in flash reaction when glow reaction gives only 150 000 RLU with the same amount of ATP. When the plate format is changed from 384-wells to 96-wells, assay sensitivity is decreased. This effect is not remarkable with flash type ATP assay, only about 20% decrease in the detection limit, but with glow type ATP assay the sensitivity is decreased almost 10 times. Therefore the dif- ference between flash and glow assay sensitivity with 96-well plates is much bigger with 96-well plates than with 384-well plates, being about 400 times. The very high density 1536-well plate was tested only with glow type ATP chemistry and it was shown to produce about the same ATP sensitivity than 96-well plate, having the detec- tion limit value of 200 amol/well ATP, both slightly lower than with 384-well plate. Plate format ATP Flash chemistry ATP Glow Chemistry Glow/ Flash ratio Average detection limit (amol/well)) Range, (Min-Max) Number of Varioskan Flash units Average detection limit (amol/well)) Range, (Min- Max) Number of Varioskan Flash units 96 1.8 0.8 – 2.6 7 130 600 - 900 4 409 384 1.5 0.5 – 2.9 30 80 60 - 130 6 52 1536 -- -- -- 200 180 - 210 2 -- Figure 6. Typical ATP calibration curves with different plate formats. A) Flash type ATP assay. B) Glow type ATP assay. Table I. Sensitivity test results of flash and glow ATP assays with Varioskan Flash.
  • 5. Part of Thermo Fisher Scientific All sensitivity tests were performed with several Varioskan Flash units. As expected, there are differences between the performances of the individual instruments in these ATP assays. Every instrument is always individual in its perfor- mance because there are always some differences in electronics, noise levels etc. The total perfor- mance range in the detection limits between the instruments is any- how amazingly narrow. When the average ATP detection limit with the flash chemistry and 384-well plates is for example 1.8 amol/ well, all tested 30 instrument units had their detection limit below 3 amol/well ATP. Figure 7. shows the cumulative distribution of these detection limit values of the individual Varioskan Flash instru- ments. This distribution is clearly following expected cumulative Gaussian distribution. Another very important assay quality parameter is Z-prime value. It is a numerical calculation that reflects the assay capability to separate positive results from the negative ones (14). Theoretical maximum of the Z-prime value is 1.0 that suggests that reliability of the separation of two signals is 100.0%. This is of course never possible to be met in real life because there is always some standard deviation in the signal values. If the Z-prime value is over 0.5, that assay can be normally considered good and reliable (14) but sometimes values around 0.6 or higher might be required. Calculated Z-prime values against the assay blanks over the whole ATP calibration series are shown for both flash and glow ATP chemistries and all plate formats in Figure 8. Figure 8. clearly shows the excel- lent reliability of these ATP assays. With the flash type ATP assay, that 3 amol/well ATP sample that is very close to the theoreti- cal detection limit shows already almost sufficient Z-prime values. Theoretical detection limit is calculated based on blank SD only when Z-prime calculation formula includes both blank and standard SD values. Therefore, cal- culated theoretical detection limit is always below the concentration range where good Z-prime values are obtained. With the ATP sample of 30 amol/well Z-prime values are clearly over required 0.5 level, with both 96- and 384-well formats. When the Z-prime values of glow type ATP assay are ana- lyzed, it is immediately noticed that higher ATP amounts than with flash chemistry are required for the good z-prime level. About 300 amol/well ATP is required before calculated Z-prime val- ues are constantly over required level. Figure 7. Cumulative distribution of the Varioskan Flash detection limits in flash type ATP assay with 384-well plates. Figure 8. The Z-prime values of both flash and glow type ATP assays measured with 96-, 384- and 1536-well plate formats.
  • 6. Part of Thermo Fisher Scientific Conclusions These results clearly show the high sensitivity and reliability of ATP assays with Varioskan Flash lumi- nometry. ATP assay using flash type chemistry gives better assay sensitivi- ty than assays using glow type chem- istry and 384-well plate format gives the best sensitivity with both types of chemistries. For the statistically reliable results, around 10-30 amol of ATP is required in the sample with flash type ATP assay and about 300 -500 amol/well is required for the glow type assay. References 1. Lundin, A., et al. 1976. Continuous monitoring of ATP-converting reactions by purified firefly luciferase. Anal. Biochem. 75, pp. 611 - 620. 2. Webster, J.J. and Leach, F. 1980. Optimization of the firefly luciferase assay for ATP. J. Applied Biochem. 2, pp. 469 - 79. 3. Karl, D.M. 1980. Cellular nucleotide measurements and applications in micro- bial ecology. Microbiol. Rev. 44, pp. 739 - 96. 4. Lundin, A. et al. 1986. Estimation of biomass in growing cell lines by adenosine triphosphate assay. Methods in Enzymology 133, pp. 27 - 42. 5. Fan, F. and Wood K.V. 2007. Bioluminescent Assays for High- Throughput Screening. Assay and Drug Development Technologies. 5. pp. 127-136. 6. Ow, D.W. et al. 1986. Transient and stable expression of the firefly luciferase gene in plant cells and transgenic plants. Science 234, pp. 856 - 859. 7. De Wet, J.R. et al. 1987. Firefly luciferase gene: Structure and expression in mammalian cells. Mol. Cell. Biol. 7, 725 - 737. 8. Brasier, A.R. et al. 1989. Optimized use of firefly luciferase assay as a reporter gene in mammalian cell lines. BioTechniques. 7, pp. 1116 - 1122. 9. Koncz, C. et al. 1990. Bacterial and Firefly Luciferase Genes in Transgenic Plants: Advantages and Disadvantages of a Reporter Gene. Dev. Genet. 11, pp. 224 - 232. 10. Seliger, H.H. et al. 1964. The Spectral Distribution of Firefly Light. J. Gen. Physiol. 48, pp. 95-104. 11. Ugarova, N.N. and L.Y. Brovko. 2002. Protein structure and bioluminescent spectra for firefly bioluminescence. Luminescence 17, pp. 321-330. 12. Lundin, et al. 1982. Optimized biolu- minescence assay of creatine kinase and cre- atine kinase B-subunit activity. Clin. Chem. 28, pp. 609-614. 13. Wood, K V. et al. 1991. Recent Advances and Prospects for Use of Beetle Luciferases as Genetic Reporters. In: Bioluminescence & Chemiluminescence: Current Status, Proceedings of the VIth International Symposium on Bioluminescence and Chemiluminescence, Stanley, P. E., et al., Editors, John Wiley & Sons, Chichester, United Kingdom. Pp. 543- 546. 14. Carter, J.M. 2003. A Guide to Assay Development. D&MD Publications. Pp. 12/10 – 12/11. North America: USA / Canada +1 800 522 7763 Europe: Austria +43 1 801 40 0 Belgium +32 2 482 30 30 France +33 2 2803 2000 Germany national toll free 08001-536 376 Germany international +49 6184 90 6940 Italy +39 02 02 95059 434-254 Netherlands +31 76 571 4440 Nordic countries +358 9 329 100 Russia/CIS +7 (812) 703 42 15 Spain/Portugal +34 93 223 09 18 Switzerland +41 44 454 12 12 UK/Ireland +44 870 609 9203 Asia: China +86 21 6865 4588 or +86 10 8419 3588 India +91 22 6716 2200 Japan +81 45 453 9220 Other Asian countries +852 2885 4613 Countries not listed: +49 6184 90 6940 or +33 2 2803 2000 www.thermo.com/ readingroom www.thermo.com/mpi In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world. © 2008 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales repre- sentative for details.