OCT provides high-resolution, cross-sectional imaging of the retina and anterior segment of the eye in a non-invasive manner. It works on the principles of interferometry and low coherence reflectometry to obtain micrometer-level resolution images. Time domain OCT uses a moving reference mirror while Fourier domain OCT obtains entire scans simultaneously using a spectrometer. OCT is useful for diagnosing and monitoring various retinal diseases like macular edema, glaucoma, age-related macular degeneration and corneal pathologies by visualizing intraretinal layers and thickness maps. It has become the gold standard for evaluation and management of diseases affecting the retina.
2. DR DINESH MITTAL DR SONALEE MITTAL
DRISHTI EYE HOSP VIJAYNAGAR INDORE
3. non invasive,
non contact imaging modality
generates micrometer resolution
cross sectional images
OCT
4. Scattering is a fundamental property of a
heterogeneous medium, and occurs because of
variations in the refractive index within tissue.
4
5. Principle of OCT
• Interferometry is the
technique of superimposing
(interfering ) two or more
waves, to detect differences
between them.
• Interferometry works
because two waves with the
same frequency that have the
same phase will add each
other while two waves that
have opposite phase will
subtract.
6. Principles of OCT Technology
An A-scan is the intensity of reflected light at various
retinal depths at a single retinal location
Combining many A-scans produces a B-scan
A-scan A-scan
+ + . . . =
B-scanA-scans
RetinalDepth
Reflectance Intensity
7.
8. • Light from a source is
directed onto a
partially reflecting
mirror and is split into
a reference and a
measurement beam.
• The measurement
beam reflected from
the specimen with
different time delays
according to its internal
microstructure.
WORKING OF OCT
9. • The light in the reference
beam is reflected from a
reference mirror at a
variable distance which
produces a variable time
delay.
• The light from the
specimen, consisting of
multiple echoes, and the
light from the reference
mirror, consisting of a
single echo at a known
delay are combined and
detected.
WORKING OF OCT
12. Time Domain OCT
SLD
Lens
Detector
Data Acquisition
Processing
Combines light
from reference
with reflected
light from
retina
Distance determines
depth in A scan
Reference mirror
moves back and forth
Scanning mirror
directs SLD
beam on retina
Interferometer
Broadband
Light Source
Creates
A-scan
1 pixel
at a
time
Final A-scan
Process
repeated many
times to create
B-scan
13. Time Domain OCT
• The Michelson interferometer splits the light from the
broadband source into two paths, the reference and sample
arms.
• The interference signal between the reflected reference
wave and the backscattered sample wave is then recorded.
• The axial optical sectioning ability of the technique is
inversely proportional to its optical bandwidth.
14. Fourier Domain OCT
SLD
Spectrometer
analyzes
signal by
wavelength FFT
Grating splits
signal by
wavelength
Broadband
Light Source
Reference mirror
stationary
Combines light
from reference
with reflected
light from
retina
Interferometer
Spectral
interferogram
Fourier transform
converts signal to
typical A-scan
Entire A-scan
created at a
single time
Process
repeated many
times to create
B-scan
15. Fourier domain OCT
• In FD-OCT ,the detector arm of the Michelson
interferometer uses a spectrometer instead a single
detector.
• The spectrometer measures spectral modulations
produced by interference between the sample and
reference reflections.
16. SPECTRAL/FOURIER DOMAIN OCT
• No need for mobile reference mirror
• assessment of interference patterns as a function of
frequency rather than of time.
• use spectral interferometry and a mathematical
function (Fourier transformation)
• images can be acquired 50 to 100 times more quickly
than in TD systems ( 20,000 A-scans /s )
18. Time vs Fourier domain OCT
Time domain OCT
• A scan generated sequentially,
one pixel at a time of 1.6
seconds
• Moving reference mirror
• 400 scans/sec
• Resolution – 10 micron
• Slower than eye movement
Fourier domain OCT
• Entire A scan is generated at
once based on Fourier
transformation of
spectrometer analysis
• Stationary reference mirror
• 26,000 scans/sec
• Resolution – 5 micron
• Faster than eye movement
18
19. Spectral OCT/SLO
• Limitation of OCT technology was difficulty in accurately localizing the
cross-sectional images and correlating them with a conventional en face
view of the fundus.
• To localize and visually interpret the images, integrating a scanning
laser ophthalmoscopy (SLO) into the OCT was needed.
• This rationale was used by OTI technologies (Toronto, Canada) to
develop the Spectral OCT/SLO.
20. • The Spectral OCT/SLO is a computerized optical scanner
device providing high-resolution, high-definition images
of the fundus anatomy.
• It integrats SLO’s confocal imaging principles with OCT’s
high resolution tomographic images.
• The system simultaneously produces SLO and OCT
images that are created through the same optical path,
and therefore correspond pixel to pixel.
• It produces a new image format called as C scan
21.
22.
23. OCT Fundus Images
• .
• SD OCT can generate
“OCT fundus images”
that mimic the images
obtained from fundus
photography
30. OCT IN DIFFERENT RETINAL DISEASES
• differentiate lamellar / pseudo / full-thickness
macular holes
• diagnosing vitreomacular traction syndrome
• differentiating various presentations of diabetic
macular edema
• monitoring the course of CSR
• making treatment decisions in ARMD
31. OCT role in DME
•Confirm presence of macular edema
•Know type of macular edema
•Assess macular thickness
•Vitero macular interface abnormalities
•Intra retinal exudates
32. OCT role in DME
•Sub retinal fluid
•Photoreceptor IS / OS junction
abnormalities
•Know response to laser , IV
pharmacotherapy & surgery
•For follow up & documentation.
33. OCT can also produce a retinal thickness map.
The OCT software automatically determines
the inner and outer retinal boundaries and
produces a false-color topographic map
showing areas of increased thickening in
brighter colors and areas of lesser thicken -ing
in darker colors
RETINAL THICKNESS MAP
34. OCT can also produce a
retinal thickness map.
The OCT software
automatically determines
the inner and outer retinal
boundaries and produces
a false-color topographic
map showing areas of
increased thickening in
brighter colors and areas
of lesser thicken -ing in
darker colors
35. An assessment of macular volume can also be
obtained from the retinal thickness map. By
evaluating differences in retinal volume over
time, the clinician can judge the efficacy of
therapy
RETINAL THICKNESS MAP
36. OCT gold standard in monitoring the progression and
treatment response in DME patients .
Retinal thickness is the most commonly used
quantitative parameter.
cirrhus measures the retinal thickness between ILM
& anterior edge of rpe layer .
normal subjects central retinal thickness is 265 µm
with cirrhus oct .
37. DME can be classified as
Diffuse retinal thickness Sponge like generalized mild hypo reflective swelling of retina
Cystoid macular odema presence of intra retinal cystoid areas of low reflectivity & separated by
higher reflectivity septa
Serous retinal detachment focal elevation of neurosensory retina overlying a hyporeflective dome
shaped space .
Viteromacular interface abnormalilities may involve epiretinal membrane or vitreo macular traction or both
DME
44. OCT Software analyses-
Peri papilary
Nerve fiber layer
Macular thickness
Optic nerve
head
1. Macula
Thickness
2. RNFL
3. Optic Nerve
Head
(ONH)
45. Ganglion cells- 30-35% of total retinal thickness at macula
upto 50%of ganglion cells in macula
glaucoma preferentially involves the ganglion cell complex (GCC).
Macula Thickness Analysis
Glaucoma with thinner GCC
Normal
46.
47. RNFL Analysis
• Analysis of RNFL aids in identification of early glaucomatous loss
• Circular scans of 3.4 mm diameter in the peripapillary region (cylindrical
retinal cross-section)
• RNFL thickness measurement is graphed in a TSNIT orientation
• Compared to age-matched normative data
48. RNFL analysis
• Circular scanning
around ONH at a radius
of 1. 73mm
• Three scans are
acquired and data is
averaged and compared
with normative data
base of age matched
subjects
• Scan begins temporally
49. RNFL thickness average analysis
printout -7 zones
• Zone -1-Pt. I.D
• Zone -2-TSNIT with age matched normative data
base
• Zone-3-TSNIT overlap of two eyes
• Zone -4-circular scan-quadrant/clockwise
• Zone-5-DATA TABLE-ratio/average
• Zone-6-RED FREE PHOTOGRAPH-position
• Zone-7-PERCENTILE COLOR CODING
53. Optic Nerve Head Analysis
• Radial line scans through optic disc provide crosssectional information on
cupping and neuroretinal rim area
• Disc margins are objectively identified using signal from
end of RPE
• Parameters:
• Disc
• cup and rim area
• horizontal and vertical cup-to-disc ratio
• vertical integrated rim area
• horizontal integrated rim width
54. Optic nerve head analysis
Optic nerve head scans
are composed of six
linear scans in a spoke
pattern separated by
30-degree intervals
centered on the ONH
55. Cup/disk ratios and cup Volumes
Disc size:
by measuring the distance
between the terminal ends
of the choriod at the level of
the pigment epithelium
(green line)
Cup:
determined by drawing a line
b/w both sides of the cup at
a point 140um above the
green line.
Area below the line is cup
and above is neuroretinal
rim
57. OCT ANTERIOR EYE
• OCT image of a
normal anterior
chamber
• cornea, sclera, iris
and lens anterior
capsule identified
58. • It gives optical
images with higher
resolution than
Scheimpflug-based
devices
•
T Tear film
Ep Epithelium
B Bowman layer
S Stroma
En & D Endothelium & Descemets layer
59. • • It measures corneal thickness and gives a
that is less affected by corneal opacities
• • It can be used in diagnosis and treatment of refractive
complications and some corneal pathologies
• • It is taking an important role in detecting early KC and
other ectatic corneal disorders