MASS SPECTROMETRY
(Theory, Principle and
Instrumentation)
Prepared by
Prof. (Dr.) Dinesh Kumar Mehta
MMCP, (MMDU), Mullana
Mass Spectrometry (MS) is an analytic technique used to determine
the relative masses of molecular ions and fragments by calculating the
degree of deflection of charged particles in a magnetic field.
It provides a great deal of information with very small amount of samples.
Mass spectrometry is used to:
 Determine molecular mass
 Determine molecular formula of a compound
 Determine structural features of the compound
 Find out the structure of an unknown substance
 Provide data on isotopic abundance
 Verify the identity and purity of a known substance
Mass spectrometry is an instrumental technique that does not involve
electromagnetic radiation. Thus, it is called spectrometry, whereas the
others such as UV, 1
HNMR, IR are called spectroscopy.
 It involves the production and separation of ionised molecules and
their ionic decompositon product and finally the measurement of the
relative abundance of different ions produced. It is, thus a destructive
technique in that the sample is consumed during analysis.
 In most cases, the nascent molecular ion of the analyte produced
fragment ions by cleavage of the bond and the resulting
fragmentation pattern constitutes the mass spectrum.
 Thus, the mass spectrum of each compound is unique and can be
used as a chemical fingerprint to characterize the sample.
The principle of mass spectrometry involves the generation of ions
from either inorganic or organic compounds, to separate the ions by their
mass-to-charge ratio (m/z) and to detect them qualitatively and
quantitatively by their m/z value and relative abundance.
In the first step a beam of energetic electron produce gas phase ions of the
compound. Removal of one electron from the molecule (M) results into
generation of parent ion (M.+
or molecular ion). The molecular mass of the
compound is equal to the m/z value of the parent ion.
This parent ion or molecular ion normally undergoes fragmentations
(fragment ions or daughter ions).
M + e-
M.+
+ 2e-
The parent ion (M.+
) is a radical cation with an odd number of electrons, it
can fragment to give either a radical (R.
) and an ion with an even number
of electrons (EE+
), or a molecule (N) and a new radical cation (OE.+
).
These two types of ions derived from the molecular ion can, further
undergo fragmentation, and so on. All these ions formed are then
separated by the mass spectrometer according to their mass-to-charge
ratio. Only positively charged species reach the collector. The charge (z)
on all the fragments is +1, therefore m/z is equal to the molecular mass
(m) of the fragment.
A graph of the relative abundance of each fragment plotted against its m/z
value is called a mass spectrum.
For example, in the case of methane (CH4), the impact of high energy
electrons causes the molecule to lose an electron and form a radical cation
with m/z 16. A species, with a positive charge and one unpaired electron is
called radical cation.
The peak with the highest m/z 72 value in the spectrum of pentane is due
to the fragment that results when an electron is knocked out of a molecule
(figure 4). It is called molecular ion peak and it is equal to molecular
mass of pentane.
Peaks with m/z values less than 72 are called fragment ion peaks. The
peak at m/z 43 with the greatest intensity is called base peak.
The main components of a mass spectrometer are:
Inlet system (LC, GC, Direct probe etc...)
Ion source (EI, CI, ESI, MALDI, etc...)
Mass analyzer (Quadrupole, TOF, Ion Trap, Magnetic Sector)
Detector (Electron Multiplier, Micro Channel Plates MCPs)
There are four key stages in the process for Mass Spectrometry.
 Ionization
 Acceleration
 Deflection
 Detection
Ionization
The atom is ionised by knocking one or more electrons off to give a
positive ion. (Mass spectrometers always work with positive ions).
 The particles in the sample (atoms or molecules) are bombarded with
a stream of electrons to knock one or more electrons out of the
sample particles to make positive ions.
 Most of the positive ions formed will carry a charge of +1
 These positive ions are persuaded out into the rest of the machine by
the ion repeller which is another metal plate carrying a slight positive
charge.
MALDI (Matrix Assisted Laser Desorption)
MALDI is a method of ionization in which the sample is bombarded with
a laser. The sample is typically mixed with a matrix that absorbs the laser
radiation and transfer a proton to the sample. Some small mass samples
can be ionized without matrix, but this is typically called laser
desorption. This technique was able to analyze other type of
biomolecules, such as oligosaccharides, glycolipids, nucleotides, and
synthetic polymers.
 Analyte mixed with organic solvent containing some other organic
compound
 Solvent is evaporated creating “solid solution” called matrix
 Molecules are ionized by gas- phase photoionization or excited state
proton transfer.
For example for proteins, the matrix of choice is often sinapinic
acid. A 337 nm radiation from nitrogen laser is most commonly
used. The laser helps introducing energy into the molecular
system in such a way preventing thermal decomposition.
MALDI is often used with time-of-flights mass spectrometers (TOF) due
to the pulsing nature of the technique, and the mass range capability.
Molecular weights up to few hundreds of daltons could be measured.
Acceleration
The ions are accelerated so that they all have the same kinetic energy.
 The positive ions are repelled away from the positive ionisation chamber
and pass through three slits with voltage in the decreasing order.
 The middle slit carries some intermediate voltage and the final at 0
volts.
 All the ions are accelerated into a finely focused beam.
Deflection
 The ions are then deflected by a magnetic field according to their
masses. The lighter they are, the more they are deflected.
 The amount of deflection also depends on the number of positive
charges on the ion.
 The more the ion is charged, the more it gets deflected.
 Different ions are deflected by the magnetic field by different
amounts.
The amount of deflection depends on:
The mass of the ion: Lighter ions are deflected more than heavier ones.
The charge on the ion: Ions with 2 (or more) positive charges are
deflected more than ones with only1 positive charge.
Mass analyzer
The mass analyzer is the component of mass spectrometry instrumentation
which we used to resolve ions with different mass to charge ratios. There
are various types of mass analyzers that we used in the mass spectrometer.
 Single focusing mass analyzer with magnetic deflection
 Double focusing analyzer
 Time to flight mass analyzer
 Quadrupole mass analyzer
Time to flight mass analyzer (TOF-MS)
The principle of TOF mass analyzer involves the separation of ions based
on the time it takes for the ions to travel through a flight tube with known
length and reach the detector
In time to flight mass analyzer, all the ions leave from an accelerating field
with the same kinetic energies but different velocities depending upon
their masses. In magnetic focusing, ions are separated by changing their
directions. When they float in a straight line through a field-free region,
they required different times to travel a given distance. Based on classical
physics, ions with lower m/z will travel the fastest and arrive at the
detector first while ions with larger m/z will travel the slowest and arrive
at the detector last. A ToF layout is shown in Figure
Quadrupole mass analyzer
A quadrupole mass analyzer contains short parallel metal rods
arranged around the beam. A quadrupole mass spectrometer
consists of an ionizer (bombardment by electrons from a
hot filament), an ion accelerator, and a mass filter consisting of
four parallel metal rods arranged as in the given figure. The
combined field causes the particle to oscillate about their central
axis where they travel. The oppositely placed electrodes are
connected together and a DC voltage is applied.
Detection
 The beam of ions passing through the machine is detected
electrically.
 When an ion hits the metal box, its charge is neutralized by an
electron jumping from the metal on to the ion.
 That leaves a space amongst the electrons in the metal, and the
electrons in the wire shuffle along to fill it.
 A flow of electrons in the wire is detected as an electric current
which can be amplified and recorded. The more ions arriving, the
greater the current.

Contenu connexe

Tendances

Flame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptFlame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Sachin G
 

Tendances (20)

Mass analyser
Mass analyserMass analyser
Mass analyser
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Mass Spectroscopy
Mass SpectroscopyMass Spectroscopy
Mass Spectroscopy
 
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptFlame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy ppt
 
SlideShare Presentation on Mass spectrophotometer
SlideShare Presentation on Mass spectrophotometerSlideShare Presentation on Mass spectrophotometer
SlideShare Presentation on Mass spectrophotometer
 
Maldi (matrix attested laser desorption ionization technique
Maldi (matrix attested laser desorption ionization techniqueMaldi (matrix attested laser desorption ionization technique
Maldi (matrix attested laser desorption ionization technique
 
Ionizaion Techniques - Mass Spectroscopy
Ionizaion Techniques - Mass SpectroscopyIonizaion Techniques - Mass Spectroscopy
Ionizaion Techniques - Mass Spectroscopy
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
NMR SPECTROSCOPY
NMR SPECTROSCOPYNMR SPECTROSCOPY
NMR SPECTROSCOPY
 
X-ray Crystallography
X-ray CrystallographyX-ray Crystallography
X-ray Crystallography
 
Applications of mass spectrometry
Applications of mass spectrometryApplications of mass spectrometry
Applications of mass spectrometry
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Infrared instrumentation
Infrared instrumentationInfrared instrumentation
Infrared instrumentation
 
Atmospheric Pressure Chemical Ionization
Atmospheric Pressure Chemical IonizationAtmospheric Pressure Chemical Ionization
Atmospheric Pressure Chemical Ionization
 
Flame emission spectroscopy
Flame emission spectroscopyFlame emission spectroscopy
Flame emission spectroscopy
 
Nmr spectroscopy
Nmr spectroscopyNmr spectroscopy
Nmr spectroscopy
 
X ray crystallography for mpharm
X ray crystallography for mpharm X ray crystallography for mpharm
X ray crystallography for mpharm
 
AFFINITY CHROMATOGRAPHY
AFFINITY CHROMATOGRAPHYAFFINITY CHROMATOGRAPHY
AFFINITY CHROMATOGRAPHY
 
Lc ms
Lc msLc ms
Lc ms
 
FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY
 

Similaire à MASS SPECTROMETRY=.pdf

2 amrutamass-130210233002-phpapp02
2 amrutamass-130210233002-phpapp022 amrutamass-130210233002-phpapp02
2 amrutamass-130210233002-phpapp02
illyas_77
 

Similaire à MASS SPECTROMETRY=.pdf (20)

mass spectrOSCOPY
 mass spectrOSCOPY mass spectrOSCOPY
mass spectrOSCOPY
 
Presentation on mass spectroscopy
Presentation on mass spectroscopyPresentation on mass spectroscopy
Presentation on mass spectroscopy
 
mass spectroscopy presentation.pptx
mass spectroscopy presentation.pptxmass spectroscopy presentation.pptx
mass spectroscopy presentation.pptx
 
Mass spectrometry- full lecture
Mass spectrometry- full lecture Mass spectrometry- full lecture
Mass spectrometry- full lecture
 
Mass spectroscopy -> by Mohit kumar
Mass spectroscopy -> by Mohit kumarMass spectroscopy -> by Mohit kumar
Mass spectroscopy -> by Mohit kumar
 
Mass spectroscopy(1)
Mass spectroscopy(1)Mass spectroscopy(1)
Mass spectroscopy(1)
 
mass spectro pres.pptx
mass spectro pres.pptxmass spectro pres.pptx
mass spectro pres.pptx
 
Mass Spectrometry
Mass SpectrometryMass Spectrometry
Mass Spectrometry
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Bilal Ibrahim Dan-Iya_GS57595.pptx
Bilal Ibrahim Dan-Iya_GS57595.pptxBilal Ibrahim Dan-Iya_GS57595.pptx
Bilal Ibrahim Dan-Iya_GS57595.pptx
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
2 amrutamass-130210233002-phpapp02
2 amrutamass-130210233002-phpapp022 amrutamass-130210233002-phpapp02
2 amrutamass-130210233002-phpapp02
 
Mass spectrometry.pptx
Mass spectrometry.pptxMass spectrometry.pptx
Mass spectrometry.pptx
 
4. mass jntu pharmacy
4. mass jntu pharmacy4. mass jntu pharmacy
4. mass jntu pharmacy
 
Spectroscopy 1.pdf, atomic absorption spectroscopy
Spectroscopy 1.pdf, atomic absorption spectroscopySpectroscopy 1.pdf, atomic absorption spectroscopy
Spectroscopy 1.pdf, atomic absorption spectroscopy
 
Mass Spectrometry
Mass SpectrometryMass Spectrometry
Mass Spectrometry
 
mass shruti
mass shrutimass shruti
mass shruti
 

Plus de Dr. Dinesh Mehta

Plus de Dr. Dinesh Mehta (20)

MS Fragmentation Process and Application of MS.pdf
MS Fragmentation Process and Application of MS.pdfMS Fragmentation Process and Application of MS.pdf
MS Fragmentation Process and Application of MS.pdf
 
MASS SPECTROMETRY INTRODUCTION.pdf
MASS SPECTROMETRY INTRODUCTION.pdfMASS SPECTROMETRY INTRODUCTION.pdf
MASS SPECTROMETRY INTRODUCTION.pdf
 
MS Fragmentation Process and Application of MS.pdf
MS Fragmentation Process and Application of MS.pdfMS Fragmentation Process and Application of MS.pdf
MS Fragmentation Process and Application of MS.pdf
 
UV-VIS.pdf
UV-VIS.pdfUV-VIS.pdf
UV-VIS.pdf
 
INSTRUMENTATIONS UV-VIS.pdf
INSTRUMENTATIONS UV-VIS.pdfINSTRUMENTATIONS UV-VIS.pdf
INSTRUMENTATIONS UV-VIS.pdf
 
Quality System and Audit.pptx
Quality System and Audit.pptxQuality System and Audit.pptx
Quality System and Audit.pptx
 
QMS.pptx
QMS.pptxQMS.pptx
QMS.pptx
 
AUDIT Ph Manufacturing.pptx
AUDIT Ph Manufacturing.pptxAUDIT Ph Manufacturing.pptx
AUDIT Ph Manufacturing.pptx
 
Auditing Manufacturing Process and Product and Process Information.pdf
Auditing Manufacturing Process and Product and Process Information.pdfAuditing Manufacturing Process and Product and Process Information.pdf
Auditing Manufacturing Process and Product and Process Information.pdf
 
Building of Raw Material.pdf
Building of Raw Material.pdfBuilding of Raw Material.pdf
Building of Raw Material.pdf
 
CHAPTER-3 Product and Process Information.pdf
CHAPTER-3 Product and Process Information.pdfCHAPTER-3 Product and Process Information.pdf
CHAPTER-3 Product and Process Information.pdf
 
CHAPTER-1 DEFICIENCIES.pdf
CHAPTER-1 DEFICIENCIES.pdfCHAPTER-1 DEFICIENCIES.pdf
CHAPTER-1 DEFICIENCIES.pdf
 
CHAPTER-1 Information Gathering and Administration.pdf
CHAPTER-1 Information Gathering and Administration.pdfCHAPTER-1 Information Gathering and Administration.pdf
CHAPTER-1 Information Gathering and Administration.pdf
 
CHAPTER-1 Management Audit and Planning procedure.pdf
CHAPTER-1 Management Audit and Planning procedure.pdfCHAPTER-1 Management Audit and Planning procedure.pdf
CHAPTER-1 Management Audit and Planning procedure.pdf
 
HPTLC MS=
HPTLC MS=HPTLC MS=
HPTLC MS=
 
GC -MS/MS and LC MS/MS
GC -MS/MS and LC MS/MSGC -MS/MS and LC MS/MS
GC -MS/MS and LC MS/MS
 
Benchmarking
BenchmarkingBenchmarking
Benchmarking
 
Application of hyphenated techniques
Application of hyphenated techniques Application of hyphenated techniques
Application of hyphenated techniques
 
Application of hyphenated techniques(GC-MS)
Application of hyphenated techniques(GC-MS)Application of hyphenated techniques(GC-MS)
Application of hyphenated techniques(GC-MS)
 
Hyphenated techniques(GC-MS/MS, LC-MS/MS, HPTLC-MS)
Hyphenated techniques(GC-MS/MS, LC-MS/MS,  HPTLC-MS)Hyphenated techniques(GC-MS/MS, LC-MS/MS,  HPTLC-MS)
Hyphenated techniques(GC-MS/MS, LC-MS/MS, HPTLC-MS)
 

Dernier

Introduction,importance and scope of horticulture.pptx
Introduction,importance and scope of horticulture.pptxIntroduction,importance and scope of horticulture.pptx
Introduction,importance and scope of horticulture.pptx
Bhagirath Gogikar
 
Formation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disksFormation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disks
Sérgio Sacani
 
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
PirithiRaju
 
Bacterial Identification and Classifications
Bacterial Identification and ClassificationsBacterial Identification and Classifications
Bacterial Identification and Classifications
Areesha Ahmad
 
Seismic Method Estimate velocity from seismic data.pptx
Seismic Method Estimate velocity from seismic  data.pptxSeismic Method Estimate velocity from seismic  data.pptx
Seismic Method Estimate velocity from seismic data.pptx
AlMamun560346
 
dkNET Webinar "Texera: A Scalable Cloud Computing Platform for Sharing Data a...
dkNET Webinar "Texera: A Scalable Cloud Computing Platform for Sharing Data a...dkNET Webinar "Texera: A Scalable Cloud Computing Platform for Sharing Data a...
dkNET Webinar "Texera: A Scalable Cloud Computing Platform for Sharing Data a...
dkNET
 

Dernier (20)

Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
 
IDENTIFICATION OF THE LIVING- forensic medicine
IDENTIFICATION OF THE LIVING- forensic medicineIDENTIFICATION OF THE LIVING- forensic medicine
IDENTIFICATION OF THE LIVING- forensic medicine
 
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 60009654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
 
Clean In Place(CIP).pptx .
Clean In Place(CIP).pptx                 .Clean In Place(CIP).pptx                 .
Clean In Place(CIP).pptx .
 
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRLKochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
 
Introduction,importance and scope of horticulture.pptx
Introduction,importance and scope of horticulture.pptxIntroduction,importance and scope of horticulture.pptx
Introduction,importance and scope of horticulture.pptx
 
Formation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disksFormation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disks
 
Unit5-Cloud.pptx for lpu course cse121 o
Unit5-Cloud.pptx for lpu course cse121 oUnit5-Cloud.pptx for lpu course cse121 o
Unit5-Cloud.pptx for lpu course cse121 o
 
GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)
 
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
 
COST ESTIMATION FOR A RESEARCH PROJECT.pptx
COST ESTIMATION FOR A RESEARCH PROJECT.pptxCOST ESTIMATION FOR A RESEARCH PROJECT.pptx
COST ESTIMATION FOR A RESEARCH PROJECT.pptx
 
pumpkin fruit fly, water melon fruit fly, cucumber fruit fly
pumpkin fruit fly, water melon fruit fly, cucumber fruit flypumpkin fruit fly, water melon fruit fly, cucumber fruit fly
pumpkin fruit fly, water melon fruit fly, cucumber fruit fly
 
Bacterial Identification and Classifications
Bacterial Identification and ClassificationsBacterial Identification and Classifications
Bacterial Identification and Classifications
 
Seismic Method Estimate velocity from seismic data.pptx
Seismic Method Estimate velocity from seismic  data.pptxSeismic Method Estimate velocity from seismic  data.pptx
Seismic Method Estimate velocity from seismic data.pptx
 
Vip profile Call Girls In Lonavala 9748763073 For Genuine Sex Service At Just...
Vip profile Call Girls In Lonavala 9748763073 For Genuine Sex Service At Just...Vip profile Call Girls In Lonavala 9748763073 For Genuine Sex Service At Just...
Vip profile Call Girls In Lonavala 9748763073 For Genuine Sex Service At Just...
 
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...
 
Connaught Place, Delhi Call girls :8448380779 Model Escorts | 100% verified
Connaught Place, Delhi Call girls :8448380779 Model Escorts | 100% verifiedConnaught Place, Delhi Call girls :8448380779 Model Escorts | 100% verified
Connaught Place, Delhi Call girls :8448380779 Model Escorts | 100% verified
 
dkNET Webinar "Texera: A Scalable Cloud Computing Platform for Sharing Data a...
dkNET Webinar "Texera: A Scalable Cloud Computing Platform for Sharing Data a...dkNET Webinar "Texera: A Scalable Cloud Computing Platform for Sharing Data a...
dkNET Webinar "Texera: A Scalable Cloud Computing Platform for Sharing Data a...
 
STS-UNIT 4 CLIMATE CHANGE POWERPOINT PRESENTATION
STS-UNIT 4 CLIMATE CHANGE POWERPOINT PRESENTATIONSTS-UNIT 4 CLIMATE CHANGE POWERPOINT PRESENTATION
STS-UNIT 4 CLIMATE CHANGE POWERPOINT PRESENTATION
 
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
 

MASS SPECTROMETRY=.pdf

  • 1. MASS SPECTROMETRY (Theory, Principle and Instrumentation) Prepared by Prof. (Dr.) Dinesh Kumar Mehta MMCP, (MMDU), Mullana
  • 2. Mass Spectrometry (MS) is an analytic technique used to determine the relative masses of molecular ions and fragments by calculating the degree of deflection of charged particles in a magnetic field. It provides a great deal of information with very small amount of samples. Mass spectrometry is used to:  Determine molecular mass  Determine molecular formula of a compound  Determine structural features of the compound  Find out the structure of an unknown substance  Provide data on isotopic abundance  Verify the identity and purity of a known substance Mass spectrometry is an instrumental technique that does not involve electromagnetic radiation. Thus, it is called spectrometry, whereas the others such as UV, 1 HNMR, IR are called spectroscopy.  It involves the production and separation of ionised molecules and their ionic decompositon product and finally the measurement of the relative abundance of different ions produced. It is, thus a destructive technique in that the sample is consumed during analysis.
  • 3.  In most cases, the nascent molecular ion of the analyte produced fragment ions by cleavage of the bond and the resulting fragmentation pattern constitutes the mass spectrum.  Thus, the mass spectrum of each compound is unique and can be used as a chemical fingerprint to characterize the sample. The principle of mass spectrometry involves the generation of ions from either inorganic or organic compounds, to separate the ions by their mass-to-charge ratio (m/z) and to detect them qualitatively and quantitatively by their m/z value and relative abundance. In the first step a beam of energetic electron produce gas phase ions of the compound. Removal of one electron from the molecule (M) results into generation of parent ion (M.+ or molecular ion). The molecular mass of the compound is equal to the m/z value of the parent ion. This parent ion or molecular ion normally undergoes fragmentations (fragment ions or daughter ions). M + e- M.+ + 2e- The parent ion (M.+ ) is a radical cation with an odd number of electrons, it can fragment to give either a radical (R. ) and an ion with an even number of electrons (EE+ ), or a molecule (N) and a new radical cation (OE.+ ).
  • 4. These two types of ions derived from the molecular ion can, further undergo fragmentation, and so on. All these ions formed are then separated by the mass spectrometer according to their mass-to-charge ratio. Only positively charged species reach the collector. The charge (z) on all the fragments is +1, therefore m/z is equal to the molecular mass (m) of the fragment. A graph of the relative abundance of each fragment plotted against its m/z value is called a mass spectrum. For example, in the case of methane (CH4), the impact of high energy electrons causes the molecule to lose an electron and form a radical cation with m/z 16. A species, with a positive charge and one unpaired electron is called radical cation.
  • 5. The peak with the highest m/z 72 value in the spectrum of pentane is due to the fragment that results when an electron is knocked out of a molecule (figure 4). It is called molecular ion peak and it is equal to molecular mass of pentane. Peaks with m/z values less than 72 are called fragment ion peaks. The peak at m/z 43 with the greatest intensity is called base peak. The main components of a mass spectrometer are: Inlet system (LC, GC, Direct probe etc...) Ion source (EI, CI, ESI, MALDI, etc...) Mass analyzer (Quadrupole, TOF, Ion Trap, Magnetic Sector) Detector (Electron Multiplier, Micro Channel Plates MCPs)
  • 6. There are four key stages in the process for Mass Spectrometry.  Ionization  Acceleration  Deflection  Detection
  • 7. Ionization The atom is ionised by knocking one or more electrons off to give a positive ion. (Mass spectrometers always work with positive ions).  The particles in the sample (atoms or molecules) are bombarded with a stream of electrons to knock one or more electrons out of the sample particles to make positive ions.  Most of the positive ions formed will carry a charge of +1  These positive ions are persuaded out into the rest of the machine by the ion repeller which is another metal plate carrying a slight positive charge.
  • 8. MALDI (Matrix Assisted Laser Desorption) MALDI is a method of ionization in which the sample is bombarded with a laser. The sample is typically mixed with a matrix that absorbs the laser radiation and transfer a proton to the sample. Some small mass samples can be ionized without matrix, but this is typically called laser desorption. This technique was able to analyze other type of biomolecules, such as oligosaccharides, glycolipids, nucleotides, and synthetic polymers.  Analyte mixed with organic solvent containing some other organic compound  Solvent is evaporated creating “solid solution” called matrix  Molecules are ionized by gas- phase photoionization or excited state proton transfer.
  • 9. For example for proteins, the matrix of choice is often sinapinic acid. A 337 nm radiation from nitrogen laser is most commonly used. The laser helps introducing energy into the molecular system in such a way preventing thermal decomposition. MALDI is often used with time-of-flights mass spectrometers (TOF) due to the pulsing nature of the technique, and the mass range capability. Molecular weights up to few hundreds of daltons could be measured. Acceleration The ions are accelerated so that they all have the same kinetic energy.
  • 10.  The positive ions are repelled away from the positive ionisation chamber and pass through three slits with voltage in the decreasing order.  The middle slit carries some intermediate voltage and the final at 0 volts.  All the ions are accelerated into a finely focused beam. Deflection  The ions are then deflected by a magnetic field according to their masses. The lighter they are, the more they are deflected.  The amount of deflection also depends on the number of positive charges on the ion.  The more the ion is charged, the more it gets deflected.  Different ions are deflected by the magnetic field by different amounts.
  • 11. The amount of deflection depends on: The mass of the ion: Lighter ions are deflected more than heavier ones. The charge on the ion: Ions with 2 (or more) positive charges are deflected more than ones with only1 positive charge. Mass analyzer The mass analyzer is the component of mass spectrometry instrumentation which we used to resolve ions with different mass to charge ratios. There are various types of mass analyzers that we used in the mass spectrometer.  Single focusing mass analyzer with magnetic deflection  Double focusing analyzer  Time to flight mass analyzer  Quadrupole mass analyzer Time to flight mass analyzer (TOF-MS) The principle of TOF mass analyzer involves the separation of ions based on the time it takes for the ions to travel through a flight tube with known length and reach the detector In time to flight mass analyzer, all the ions leave from an accelerating field with the same kinetic energies but different velocities depending upon their masses. In magnetic focusing, ions are separated by changing their
  • 12. directions. When they float in a straight line through a field-free region, they required different times to travel a given distance. Based on classical physics, ions with lower m/z will travel the fastest and arrive at the detector first while ions with larger m/z will travel the slowest and arrive at the detector last. A ToF layout is shown in Figure Quadrupole mass analyzer A quadrupole mass analyzer contains short parallel metal rods arranged around the beam. A quadrupole mass spectrometer consists of an ionizer (bombardment by electrons from a hot filament), an ion accelerator, and a mass filter consisting of four parallel metal rods arranged as in the given figure. The combined field causes the particle to oscillate about their central
  • 13. axis where they travel. The oppositely placed electrodes are connected together and a DC voltage is applied. Detection  The beam of ions passing through the machine is detected electrically.  When an ion hits the metal box, its charge is neutralized by an electron jumping from the metal on to the ion.
  • 14.  That leaves a space amongst the electrons in the metal, and the electrons in the wire shuffle along to fill it.  A flow of electrons in the wire is detected as an electric current which can be amplified and recorded. The more ions arriving, the greater the current.