SlideShare une entreprise Scribd logo
1  sur  41
1
Quality by design approach for
establishment of stability indicating
method for determination of cefditoren
pivoxil
Dr.Sawsan M.Amer
Professor of Analytical Chemistry
2
This study was done as part of master
degree for one of my students ,Mohamed
Gad .
With two other colleges : Assistant Prof.Dr
Halla Zazaa from my department &
Prof.Dr.Mohamed Korany from Faculty of
Pharmacy , Alex,University
3
4
* Introduction to Cefditoren bivoxil & its degradation
* Introduction to Analytical QbD method .
* Developing & validation of stability indicating HPLC
method for determination of CTP in presence
of its degradation products .
Outlines
 It is a semi-synthetic third generation, broad-spectrum
cephalosporin orally administered for treatment of respiratory
tract infections

Cefditoren pivoxil (CTP)
6
Chemical structure
7
Chemical stability of CTP
Also, It hydrolysed either spontaneously in aqueous
medium or after oral administration, in gastrointestinal tract
in the presence of a β- lactamase .
 literature review revealed various methods for determination of CTP &
different applications of QbD in analytical method development.
Traditional approach to HPLC, method development
depends on trial and error or by changing one factor at
time (OFAT) while holding the rest constant . Although
it require a very large number of experiments to
identify the optimal conditions, they do not account for
interaction between factors.
. Computer-assisted QbD approach provides better
understanding of method parameters influencing
chromatographic process. Design Of Experiment
( DOE) ensures method application with
predictable performance during routine work
9
Development and validation of a robust and
ragged stability indicating HPLC method for
determination of CTP in the presence of its
degradation products.
Aim of work
Degradation solution preparation
10
0.001M
NaOH
0.1M
HCl
1%
H
2
O
2
H
2
O
(thermal)
H
2
O
(hƲ)
Reflux for 30 min at 60 °C
Development
solution
CTP Stock
standard
solution
Highly
complicated
matrix
Stress degradation studies performed
using acid, base, peroxide, thermal and
photolytic methods
Quality by design (QbD)
 Quality by design principles when
applied to the development of
analytical methods, it termed
“Analytical QbD” (AQbD)
11
Robust
method
Analytical
method
development
QbD
principles
 It is a systematic approach for process development that
begins with predefined objectives and emphasizes
product, process understanding and process control,
based on sound science and quality risk management
Analytical Target Profile (ATP)
Identification of Parameters &
Critical quality attributes (CQA)
Risk Assessments
Design of experment
Identification of Design space
Quality by design methodology
12
13
ATP identification includes selection of method requirements such as
target analytes, type of analytical technique, and product specifications.
1.Analytical target profile (ATP)
Critical quality attributes are defined as a property that should be
within an appropriate limit or range to ensure the desired product
quality.
2.Critical quality attributes (CQAs)
Critical quality Attributes
and Method Parameters
14
Critical quality attribute Predefined Limit
Peak asymmetry 0.9 to 1.1
Theoretical plate number > 4000 (maximize)
Pre-resolution > 4 (maximize)
Post-resolution > 4 (maximize)
Run time [min] < 10 min. (minimize)
Method Parameters
• Flow rate
• Wavelength
• Chromatographic Column type
• Mobile phase Buffer pH
• Temperature
• Methanol%
For analytical chromatographic methods, performance criteria as
resolution, asymmetry and theoretical plate number can be called critical
quality attributes (CQAs).
15 3.Risk assessment
Parameters that directly affect the quality of the method are
first sorted out, and its possible effects on method
development are studied in risk-based approach
Risk assessment aims to find out the risk of method parameters
on different aspects of response. Various tools for risk
assessment are available as Failure mode effect analysis
(FMEA) & Pareto rules .
FMEA is used to rank the factors based on risk (i.e. a product
of probability, severity, and detectability) and in combination
with Pareto analysis it is possible to select the process
parameters
6
Analytical Method and Risk Management
• Severity = Effect on efficacy of (CQAs)
• Occurrence = Chance of Failure
Related to process knowledge , changes and controls
• Detectability = Ability to Detect a Failure
Risk Factor = Severity x Occurrence x Detectability
Low High
Severity (S) 1 10
Occurrence (O) 1 10
Detectability (D) 10 1
0
200
400
600
800
1000
Wavelength
Column
Buffer pH
Temperat…
% Methanol
Risk assessment
17
Failure mode and effect analysis(FMEA)
Pareto rule: 80 % of risk is caused by 20% of causes.
Risk factor (RF) = Severity (S)
×occurrence (O) ×detectability (D)
18
*The previous figure revealed high risk rank for methanol
percentage in mobile phase (MeOH%) and elution
temperature (T), relatively lower risk rank for buffer pH on
CQAs.
* While factors like detection wavelength and column type
show minor risk rank. These factors were easily controlled &
Buffer pH was studied in univariate mode.
* MeOH% and T as the major risky factors were subjected
to extensive study using multivariate design of experiment
(DOE) to model them with CQAs.
Risk assessment
Parameters optimization
(flow rate & wavelength)
19
 Flow rate:
Stability indicating method are lengthy methods. So, reduction of run
time would be advantageous as long as we maintain acceptable
CQAs Flow rate 1.5 mL min-1
 Wavelength:
For chromatographic detection wavelength , CTP was scanned
between 200-400 nm where λ=225 nm was the best in terms of
sensitivity and precision that was selected as optimum wavelength
m
i
n
0 1 2 3 4 5 6 7 8
m
A
U
0
5
0
1
0
0
1
5
0
2
0
0
2
5
0
H
y
p
e
r
s
i
lB
D
S
C
8
S
B
1
5
0
x
4
.
6
m
m
5
m
i
c
1.
5
63
1.
6
86
2.
5
82
3.
0
43
3.
7
11
5.
6
36
m
i
n
0 1 2 3 4 5 6 7 8 9
m
A
U
0
2
0
4
0
6
0
8
0
1.
2
43
1.
4
15
2.
4
74
2.
9
05
3.
8
30
5.
2
23
5.
5
05
6.
1
20
9.
1
38
E
c
l
i
p
s
p
l
u
sC
1
8
1
0
0
x
4
.
6
m
m
3
.
5
m
i
c
Reversed phase C18 stationery phase was superior to RP-C8 in terms of
number of eluted peaks and resolution.
Using short RP-C18 column with smaller particle size, superior results were
obtained.
Parameters optimization (Column)
20
pH
2.5 3.5 4.5 5.5 6.5
Resolution
4.5
5.0
5.5
6.0
6.5
7.0
CTP pre-resolution
CTP post-resolution
Effect of mobile phase buffer pH on Resolution
between CTP and previous or next eluted peaks
pH
2.5 3.5 4.5 5.5 6.5
CTP
Last
peak
retention
time
8.5
8.6
8.7
8.8
8.9
9.0
9.1
9.2
Effect of mobile phase buffer pH on Retention
time of last eluted peak.
pH
2.5 3.5 4.5 5.5 6.5
CTP
Asymmetry
0.90
0.92
0.94
0.96
0.98
1.00
Effect of mobile phase buffer pH on CTP peak
asymmetry.
pH
2.5 3.5 4.5 5.5 6.5
CTP
Theoretical
plate
number
(N)
5900
6000
6100
6200
6300
6400
6500
6600
6700
Effect of mobile phase buffer pH on CTP peak
theoretical plate number.
Parameters optimization (pH)
21
22 4-Design of experiments (DOE)
It is a useful tool for studying effect of different factors in addition
to interaction between factors on given response.
The Outcomes of DOE are models relate CQAs to input
method parameters.
Mathematical and statistical manipulations involved in QbD
approach were performed using the Design expert software
package Version 7.0.0 (Stat-Ease Inc.).
23 Optimization of chromatographic method was performed
using Central composite design ( CCD ) evaluating theoretical
plates , peak asymmetry & resolution as the CAAs.
The selected experimental design is face-centered Central
composite design .
* It is one of the response surface design.
* It can detect curvature in response surface.
• Factorial points= 4
• Center points= 1
• Axial points= 4
- Replicates= 1
- Replicates = 6
- Replicates = 2
24
It is used to investigate the response surfaces resulted from combined
effect of elution temperature and mobile phase methanol percentage
namely; asymmetry, theoretical plate, resolutions, retention time of
last eluted peak as indication on run time.
Replications of the center point & axial point was done to
enhance the performance of the Design .
Face-centered Central composite design
50.00
52.50
55.00
57.50
60.00 25.00
28.75
32.50
36.25
40.00
3.3
4.775
6.25
7.725
9.2
Post-Res
Methanol% Temperature
50.00
52.50
55.00
57.50
60.00 25.00
28.75
32.50
36.25
40.00
3
6.75
10.5
14.25
18
Pre-Res
Methanol%
Temperature
50.00
52.50
55.00
57.50
60.00 25.00
28.75
32.50
36.25
40.00
0.92
0.95
0.98
1.01
1.04
Asymmetry
Methanol%
Temperature
50.00
52.50
55.00
57.50
60.00 25.00
28.75
32.50
36.25
40.00
5
12.75
20.5
28.25
36
Last
peak
retention
time
Methanol%
Temperature
50.00
52.50
55.00
57.50
60.00
25.00
28.75
32.50
36.25
40.00
0.00012
0.0001625
0.000205
0.0002475
0.00029
1.0/(Plates
No.)
Methanol%
Temperature
CQAs models of Temperature and Methanol%
25
3D-Response surface of CTP , (a) asymmetry; (b) Theoretical
plate number ; (c) CTP pre- resolution,(d) CTP post-resolution;
(e) last peak retention time as function of mobile phase
methanol% and elution temperature.
26
According to predefined limits of CQAs , each CQA response surface
has two distinct spaces: first, failure space where the CQA limits are
not satisfied. Second, design space (DS) within which CQA limits are
satisfied.
All response surfaces were overlaid in order to define the common
design space of MeOH% and T that satisfies all CQAs predefined
limits as shown in the next Figure .
ICH pharmaceutical development guideline, defines DS as the
multidimensional combination and interaction of input variables
(process parameters) that have been demonstrated to provide
assurance of quality.
27
50. 00 52. 50 55. 00 57. 50 60. 00
25. 00
28. 75
32. 50
36. 25
40. 00
Overlay Plot
Methanol%
Temperature
Asymmetry: 0.93
Plates No.: 4000
Plates No.: 7000
Pre-Res: 4
Pre-Res: 8
Post-Res: 4
Post-Res: 7
Last peak retention time: 6
Last peak retention time: 12
6
6
6
6
6
6
2
2 2
2
2
2
2
2
Control
space
Design space
Overlay plot of all response surfaces showing the failure space
“gray area”, design space “white area”, control space “green area” and
cross point normal operation parameter
Control space (CS) is subdivision of DS that is defined according
to desirability function.
50.00 52.50 55.00 57.50 60.00
25.00
28.75
32.50
36.25
40.00
Desirability
Methanol%
Temperature
0.159 0.159
0.318
0.477
0.637
0.796
6
6
6
6
6
6
2
2 2
2
2
2
2
2
Design space Desirability
28
Number
Theoretical plate
(N)
Desirability function enabled finding the most
desirable space within the DS to be identified
as control space (CS) “54-56.5 methanol%
and 34-40 °C”; then
A maximum desirable point was identified as
normal operating parameters (NOP) “55
methanol% and 40 °C” within the control space.
29
Benefits of Application of QbD
Approach to Analytical Methods
• Development of a robust method
• Applicable throughout the life cycle of the procedure
• Regulatory flexibility
The Movements within “Design Space” are not a change in method
C18
column(100×4.6
mm 3.5µm)
methanol: acetate buffer [pH
4.5]:
(55 : 45 , v/v )
Flow rate : 1.5 ml/min
Injection volume: 10 µL
Detection : 225 nm
Temperature: 40 C°
RP-HPLC
30
HPLC chromatogram of a CTP & its stress
degradation products mixture
Parameters Value
Mobile phase Acetate buffer pH 4.5 : Methanol = 45:55
Flow rate [mL/min] 1.5
Temperature [°C] 40
injection volume [µL] 10
Wavelength [nm] 225
Method parameters
31
Oxidation
DP
Basic
DP
Basic
DP
Photolytic
DP
Acidic
DP
Thermal
DP
Acidic
DP
Thermal
DP
Acidic
DP
Typical development solution chromatogram
32
Chromatograms showing CTP spiked with forced
degradation products and placebo chromatogram,
indicating absence of interference
Parameters DP
1
DP2
DP
3
DP
4
DP
5
CTP
DP
6
DP
7
DP
8
DP
9
Reference
value
Resolution (R) 1.95 5.3 2.7 3.16 5.56 6.21 1.1 2.25 8.44 R >1.5
Selectivity
factor (α)
1.14 1.07 1.01 1.19 1.32 1.36 1.05 1.11 1.49 >1
CTP
Asymmetry
factor (As)
0.95 0.9-1.2
CTP Capacity
factor (K΄)
2.82 1–10
CTP Column
efficiency (N)
6700 >2000
Method validation
33
The chromatographic selectivity factor and chromatographic
resolution between CTP and degradation products results
indicate method specificity.
34
700
600
500
400
300
200
100
9000
8000
7000
6000
5000
4000
3000
2000
1000
0
Conc (µg/ml)
Peak
area
Peak area = 30.19 + 13.12 Concentration; r=0.9999
A calibration curve was constructed by plotting the peak area
versus concentration of CTP in range of 90-675µg mL-1,
Results of assay validation parameters of the proposed HPLC method for
determination of CTP
Method parameter HPLC method
Linearity range 90-675 µgmL-1
Regression equation (A = bC + a)*
 Intercept (a) 30.2
 Slope (b) 13.1
 Correlation coefficient
(r)
0.9999
Accuracy
Mean ± St.dev 100.4 ± 0.28
50% 100.50 %
100% 100.64 %
150% 100.06 %
Precision
(Intraday %RSD)b 0.11 %
(Interday %RSD)c 0.44 %
t-test (2.228)d 0.62
Robustness 100.1 – 100.7 %
LOD [µg mL-1] 5.31
LOQ [µg mL-1] 16.1
*A is the peak area and C is the concentration.
b Intraday precision (6different determinations at 100% concentrations of / 2 replicate each (n=6))
cInterday precision (6 different determinations at 100% concentrations of / 2 replicate each (n=6)).
dt-tabulated for degree of freedom=10, two sided test at α=0.05.
35 Method validation
36
Dosage form Claimed [mg mL-1] Found[mg mL-1] Recovery%
Giasion®400 Tablet
Lot No. EE0279
0.450
0.457 101.56
0.448 99.56
0.451 100.22
0.447 99.33
0.455 101.11
Mean ± SD 100.36±0.96
Determination of Cefditoren pivoxil in pharmaceutical
formulation (Giasion®400 Tablet ) by the proposed HPLC method
Giasion® film coated tablets by Zambon claimed to contain 490.2 mg of
cefditoren pivoxil / tablet = 400 mg of cefditoren.
Item
Cefditoren pivoxil
Proposed method Reported method
a
Mean ± St.dev 99.89±0.69 99.92±0.60
n 7 7
Variance 0.48 0.36
F- value ( 4.28 )b 0.75
Student's t-test
(2.45)b
0.09
a Reported method: HPLC method, C18, water- Acetonitrile (50: 50, v/v) as a mobile phase, flow
rate of 1.2 ml min-1 and UV detection at 218 nm.
b The figures in parenthesis are the corresponding tabulated values at α= 0.05
Statistical comparison between proposed and
reported method for the determination of CTP in
pure powder form.
37
38 CONCLUSION
39
• Validation results demonstrated highly specific,
accurate, linear, precise and robust method
performance.
• AQbD development approach introduced good
separation, high robustness and confidence in
method ability to deliver intended performance.
Design space created during method development,
enabled flexibility of method transfer, Provide
guidance for troubleshooting method performance.
CONCLUSION
40
41

Contenu connexe

Similaire à sawsan-mohamed-amer-cairo-university-egypt.pptx

D BHARATH KUMAR MPA III SEM Proposal Presentation.ppt
D BHARATH KUMAR MPA III SEM Proposal Presentation.pptD BHARATH KUMAR MPA III SEM Proposal Presentation.ppt
D BHARATH KUMAR MPA III SEM Proposal Presentation.pptVenkatesh Mantha
 
Development and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC methodDevelopment and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC methodUshaKhanal3
 
Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptxBholakant raut
 
The Analytical Method Transfer Process SK-Sep 2013
The Analytical Method Transfer Process  SK-Sep 2013The Analytical Method Transfer Process  SK-Sep 2013
The Analytical Method Transfer Process SK-Sep 2013Stephan O. Krause, PhD
 
Sagar kanade m pharm indrustial project
Sagar kanade m pharm indrustial projectSagar kanade m pharm indrustial project
Sagar kanade m pharm indrustial projectSagar Kanade
 
Valganciclovir Tablet
Valganciclovir Tablet Valganciclovir Tablet
Valganciclovir Tablet google
 
Particles in the Biotech Product Life Cycle: Analysis, Identification and Con...
Particles in the Biotech Product Life Cycle: Analysis, Identification and Con...Particles in the Biotech Product Life Cycle: Analysis, Identification and Con...
Particles in the Biotech Product Life Cycle: Analysis, Identification and Con...SGS
 
Practical Implementation of the New Elemental Impurities Guidelines May 2015
Practical Implementation of the New Elemental Impurities Guidelines May 2015Practical Implementation of the New Elemental Impurities Guidelines May 2015
Practical Implementation of the New Elemental Impurities Guidelines May 2015SGS
 
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...PerkinElmer, Inc.
 
Analytical quality by design
Analytical quality by designAnalytical quality by design
Analytical quality by designAnnalisa Forlenza
 
Method Development and Method Validation for the estimation of Valganciclovir...
Method Development and Method Validation for the estimation of Valganciclovir...Method Development and Method Validation for the estimation of Valganciclovir...
Method Development and Method Validation for the estimation of Valganciclovir...google
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationSai Praveen Reddy
 
Development and validation of rp hplc method for simultaneous
Development and validation of rp hplc method for simultaneousDevelopment and validation of rp hplc method for simultaneous
Development and validation of rp hplc method for simultaneouschandu chatla
 
Process Scale Chromatography
Process Scale ChromatographyProcess Scale Chromatography
Process Scale ChromatographyMilliporeSigma
 
Analytical method development for new
Analytical method development for newAnalytical method development for new
Analytical method development for newVeeprho Laboratories
 

Similaire à sawsan-mohamed-amer-cairo-university-egypt.pptx (20)

D BHARATH KUMAR MPA III SEM Proposal Presentation.ppt
D BHARATH KUMAR MPA III SEM Proposal Presentation.pptD BHARATH KUMAR MPA III SEM Proposal Presentation.ppt
D BHARATH KUMAR MPA III SEM Proposal Presentation.ppt
 
Development and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC methodDevelopment and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC method
 
Tools of the Trade
Tools of the TradeTools of the Trade
Tools of the Trade
 
Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptx
 
The Analytical Method Transfer Process SK-Sep 2013
The Analytical Method Transfer Process  SK-Sep 2013The Analytical Method Transfer Process  SK-Sep 2013
The Analytical Method Transfer Process SK-Sep 2013
 
Sagar kanade m pharm indrustial project
Sagar kanade m pharm indrustial projectSagar kanade m pharm indrustial project
Sagar kanade m pharm indrustial project
 
Valganciclovir Tablet
Valganciclovir Tablet Valganciclovir Tablet
Valganciclovir Tablet
 
Particles in the Biotech Product Life Cycle: Analysis, Identification and Con...
Particles in the Biotech Product Life Cycle: Analysis, Identification and Con...Particles in the Biotech Product Life Cycle: Analysis, Identification and Con...
Particles in the Biotech Product Life Cycle: Analysis, Identification and Con...
 
Practical Implementation of the New Elemental Impurities Guidelines May 2015
Practical Implementation of the New Elemental Impurities Guidelines May 2015Practical Implementation of the New Elemental Impurities Guidelines May 2015
Practical Implementation of the New Elemental Impurities Guidelines May 2015
 
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
 
Analytical quality by design
Analytical quality by designAnalytical quality by design
Analytical quality by design
 
Method Development and Method Validation for the estimation of Valganciclovir...
Method Development and Method Validation for the estimation of Valganciclovir...Method Development and Method Validation for the estimation of Valganciclovir...
Method Development and Method Validation for the estimation of Valganciclovir...
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Development and validation of rp hplc method for simultaneous
Development and validation of rp hplc method for simultaneousDevelopment and validation of rp hplc method for simultaneous
Development and validation of rp hplc method for simultaneous
 
Process Scale Chromatography
Process Scale ChromatographyProcess Scale Chromatography
Process Scale Chromatography
 
Process Scale Chromatography
Process Scale ChromatographyProcess Scale Chromatography
Process Scale Chromatography
 
A Comparative Analysis of Semiconductor Electroplating Bath Additives by Cali...
A Comparative Analysis of Semiconductor Electroplating Bath Additives by Cali...A Comparative Analysis of Semiconductor Electroplating Bath Additives by Cali...
A Comparative Analysis of Semiconductor Electroplating Bath Additives by Cali...
 
Analytical method development for new
Analytical method development for newAnalytical method development for new
Analytical method development for new
 
Project ppt
Project pptProject ppt
Project ppt
 
Innovation in size based polymer separations
Innovation in size based polymer separationsInnovation in size based polymer separations
Innovation in size based polymer separations
 

Dernier

Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Glomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxGlomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxDr.Nusrat Tariq
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...rajnisinghkjn
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...narwatsonia7
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000aliya bhat
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 

Dernier (20)

Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
Glomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxGlomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptx
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 

sawsan-mohamed-amer-cairo-university-egypt.pptx

  • 1. 1 Quality by design approach for establishment of stability indicating method for determination of cefditoren pivoxil Dr.Sawsan M.Amer Professor of Analytical Chemistry
  • 2. 2 This study was done as part of master degree for one of my students ,Mohamed Gad . With two other colleges : Assistant Prof.Dr Halla Zazaa from my department & Prof.Dr.Mohamed Korany from Faculty of Pharmacy , Alex,University
  • 3. 3
  • 4. 4
  • 5. * Introduction to Cefditoren bivoxil & its degradation * Introduction to Analytical QbD method . * Developing & validation of stability indicating HPLC method for determination of CTP in presence of its degradation products . Outlines
  • 6.  It is a semi-synthetic third generation, broad-spectrum cephalosporin orally administered for treatment of respiratory tract infections  Cefditoren pivoxil (CTP) 6 Chemical structure
  • 7. 7 Chemical stability of CTP Also, It hydrolysed either spontaneously in aqueous medium or after oral administration, in gastrointestinal tract in the presence of a β- lactamase .  literature review revealed various methods for determination of CTP & different applications of QbD in analytical method development.
  • 8. Traditional approach to HPLC, method development depends on trial and error or by changing one factor at time (OFAT) while holding the rest constant . Although it require a very large number of experiments to identify the optimal conditions, they do not account for interaction between factors. . Computer-assisted QbD approach provides better understanding of method parameters influencing chromatographic process. Design Of Experiment ( DOE) ensures method application with predictable performance during routine work
  • 9. 9 Development and validation of a robust and ragged stability indicating HPLC method for determination of CTP in the presence of its degradation products. Aim of work
  • 10. Degradation solution preparation 10 0.001M NaOH 0.1M HCl 1% H 2 O 2 H 2 O (thermal) H 2 O (hƲ) Reflux for 30 min at 60 °C Development solution CTP Stock standard solution Highly complicated matrix Stress degradation studies performed using acid, base, peroxide, thermal and photolytic methods
  • 11. Quality by design (QbD)  Quality by design principles when applied to the development of analytical methods, it termed “Analytical QbD” (AQbD) 11 Robust method Analytical method development QbD principles  It is a systematic approach for process development that begins with predefined objectives and emphasizes product, process understanding and process control, based on sound science and quality risk management
  • 12. Analytical Target Profile (ATP) Identification of Parameters & Critical quality attributes (CQA) Risk Assessments Design of experment Identification of Design space Quality by design methodology 12
  • 13. 13 ATP identification includes selection of method requirements such as target analytes, type of analytical technique, and product specifications. 1.Analytical target profile (ATP) Critical quality attributes are defined as a property that should be within an appropriate limit or range to ensure the desired product quality. 2.Critical quality attributes (CQAs)
  • 14. Critical quality Attributes and Method Parameters 14 Critical quality attribute Predefined Limit Peak asymmetry 0.9 to 1.1 Theoretical plate number > 4000 (maximize) Pre-resolution > 4 (maximize) Post-resolution > 4 (maximize) Run time [min] < 10 min. (minimize) Method Parameters • Flow rate • Wavelength • Chromatographic Column type • Mobile phase Buffer pH • Temperature • Methanol% For analytical chromatographic methods, performance criteria as resolution, asymmetry and theoretical plate number can be called critical quality attributes (CQAs).
  • 15. 15 3.Risk assessment Parameters that directly affect the quality of the method are first sorted out, and its possible effects on method development are studied in risk-based approach Risk assessment aims to find out the risk of method parameters on different aspects of response. Various tools for risk assessment are available as Failure mode effect analysis (FMEA) & Pareto rules . FMEA is used to rank the factors based on risk (i.e. a product of probability, severity, and detectability) and in combination with Pareto analysis it is possible to select the process parameters
  • 16. 6 Analytical Method and Risk Management • Severity = Effect on efficacy of (CQAs) • Occurrence = Chance of Failure Related to process knowledge , changes and controls • Detectability = Ability to Detect a Failure Risk Factor = Severity x Occurrence x Detectability Low High Severity (S) 1 10 Occurrence (O) 1 10 Detectability (D) 10 1
  • 17. 0 200 400 600 800 1000 Wavelength Column Buffer pH Temperat… % Methanol Risk assessment 17 Failure mode and effect analysis(FMEA) Pareto rule: 80 % of risk is caused by 20% of causes. Risk factor (RF) = Severity (S) ×occurrence (O) ×detectability (D)
  • 18. 18 *The previous figure revealed high risk rank for methanol percentage in mobile phase (MeOH%) and elution temperature (T), relatively lower risk rank for buffer pH on CQAs. * While factors like detection wavelength and column type show minor risk rank. These factors were easily controlled & Buffer pH was studied in univariate mode. * MeOH% and T as the major risky factors were subjected to extensive study using multivariate design of experiment (DOE) to model them with CQAs. Risk assessment
  • 19. Parameters optimization (flow rate & wavelength) 19  Flow rate: Stability indicating method are lengthy methods. So, reduction of run time would be advantageous as long as we maintain acceptable CQAs Flow rate 1.5 mL min-1  Wavelength: For chromatographic detection wavelength , CTP was scanned between 200-400 nm where λ=225 nm was the best in terms of sensitivity and precision that was selected as optimum wavelength
  • 20. m i n 0 1 2 3 4 5 6 7 8 m A U 0 5 0 1 0 0 1 5 0 2 0 0 2 5 0 H y p e r s i lB D S C 8 S B 1 5 0 x 4 . 6 m m 5 m i c 1. 5 63 1. 6 86 2. 5 82 3. 0 43 3. 7 11 5. 6 36 m i n 0 1 2 3 4 5 6 7 8 9 m A U 0 2 0 4 0 6 0 8 0 1. 2 43 1. 4 15 2. 4 74 2. 9 05 3. 8 30 5. 2 23 5. 5 05 6. 1 20 9. 1 38 E c l i p s p l u sC 1 8 1 0 0 x 4 . 6 m m 3 . 5 m i c Reversed phase C18 stationery phase was superior to RP-C8 in terms of number of eluted peaks and resolution. Using short RP-C18 column with smaller particle size, superior results were obtained. Parameters optimization (Column) 20
  • 21. pH 2.5 3.5 4.5 5.5 6.5 Resolution 4.5 5.0 5.5 6.0 6.5 7.0 CTP pre-resolution CTP post-resolution Effect of mobile phase buffer pH on Resolution between CTP and previous or next eluted peaks pH 2.5 3.5 4.5 5.5 6.5 CTP Last peak retention time 8.5 8.6 8.7 8.8 8.9 9.0 9.1 9.2 Effect of mobile phase buffer pH on Retention time of last eluted peak. pH 2.5 3.5 4.5 5.5 6.5 CTP Asymmetry 0.90 0.92 0.94 0.96 0.98 1.00 Effect of mobile phase buffer pH on CTP peak asymmetry. pH 2.5 3.5 4.5 5.5 6.5 CTP Theoretical plate number (N) 5900 6000 6100 6200 6300 6400 6500 6600 6700 Effect of mobile phase buffer pH on CTP peak theoretical plate number. Parameters optimization (pH) 21
  • 22. 22 4-Design of experiments (DOE) It is a useful tool for studying effect of different factors in addition to interaction between factors on given response. The Outcomes of DOE are models relate CQAs to input method parameters. Mathematical and statistical manipulations involved in QbD approach were performed using the Design expert software package Version 7.0.0 (Stat-Ease Inc.).
  • 23. 23 Optimization of chromatographic method was performed using Central composite design ( CCD ) evaluating theoretical plates , peak asymmetry & resolution as the CAAs. The selected experimental design is face-centered Central composite design . * It is one of the response surface design. * It can detect curvature in response surface.
  • 24. • Factorial points= 4 • Center points= 1 • Axial points= 4 - Replicates= 1 - Replicates = 6 - Replicates = 2 24 It is used to investigate the response surfaces resulted from combined effect of elution temperature and mobile phase methanol percentage namely; asymmetry, theoretical plate, resolutions, retention time of last eluted peak as indication on run time. Replications of the center point & axial point was done to enhance the performance of the Design . Face-centered Central composite design
  • 25. 50.00 52.50 55.00 57.50 60.00 25.00 28.75 32.50 36.25 40.00 3.3 4.775 6.25 7.725 9.2 Post-Res Methanol% Temperature 50.00 52.50 55.00 57.50 60.00 25.00 28.75 32.50 36.25 40.00 3 6.75 10.5 14.25 18 Pre-Res Methanol% Temperature 50.00 52.50 55.00 57.50 60.00 25.00 28.75 32.50 36.25 40.00 0.92 0.95 0.98 1.01 1.04 Asymmetry Methanol% Temperature 50.00 52.50 55.00 57.50 60.00 25.00 28.75 32.50 36.25 40.00 5 12.75 20.5 28.25 36 Last peak retention time Methanol% Temperature 50.00 52.50 55.00 57.50 60.00 25.00 28.75 32.50 36.25 40.00 0.00012 0.0001625 0.000205 0.0002475 0.00029 1.0/(Plates No.) Methanol% Temperature CQAs models of Temperature and Methanol% 25 3D-Response surface of CTP , (a) asymmetry; (b) Theoretical plate number ; (c) CTP pre- resolution,(d) CTP post-resolution; (e) last peak retention time as function of mobile phase methanol% and elution temperature.
  • 26. 26 According to predefined limits of CQAs , each CQA response surface has two distinct spaces: first, failure space where the CQA limits are not satisfied. Second, design space (DS) within which CQA limits are satisfied. All response surfaces were overlaid in order to define the common design space of MeOH% and T that satisfies all CQAs predefined limits as shown in the next Figure . ICH pharmaceutical development guideline, defines DS as the multidimensional combination and interaction of input variables (process parameters) that have been demonstrated to provide assurance of quality.
  • 27. 27 50. 00 52. 50 55. 00 57. 50 60. 00 25. 00 28. 75 32. 50 36. 25 40. 00 Overlay Plot Methanol% Temperature Asymmetry: 0.93 Plates No.: 4000 Plates No.: 7000 Pre-Res: 4 Pre-Res: 8 Post-Res: 4 Post-Res: 7 Last peak retention time: 6 Last peak retention time: 12 6 6 6 6 6 6 2 2 2 2 2 2 2 2 Control space Design space Overlay plot of all response surfaces showing the failure space “gray area”, design space “white area”, control space “green area” and cross point normal operation parameter Control space (CS) is subdivision of DS that is defined according to desirability function.
  • 28. 50.00 52.50 55.00 57.50 60.00 25.00 28.75 32.50 36.25 40.00 Desirability Methanol% Temperature 0.159 0.159 0.318 0.477 0.637 0.796 6 6 6 6 6 6 2 2 2 2 2 2 2 2 Design space Desirability 28 Number Theoretical plate (N) Desirability function enabled finding the most desirable space within the DS to be identified as control space (CS) “54-56.5 methanol% and 34-40 °C”; then A maximum desirable point was identified as normal operating parameters (NOP) “55 methanol% and 40 °C” within the control space.
  • 29. 29 Benefits of Application of QbD Approach to Analytical Methods • Development of a robust method • Applicable throughout the life cycle of the procedure • Regulatory flexibility The Movements within “Design Space” are not a change in method
  • 30. C18 column(100×4.6 mm 3.5µm) methanol: acetate buffer [pH 4.5]: (55 : 45 , v/v ) Flow rate : 1.5 ml/min Injection volume: 10 µL Detection : 225 nm Temperature: 40 C° RP-HPLC 30 HPLC chromatogram of a CTP & its stress degradation products mixture
  • 31. Parameters Value Mobile phase Acetate buffer pH 4.5 : Methanol = 45:55 Flow rate [mL/min] 1.5 Temperature [°C] 40 injection volume [µL] 10 Wavelength [nm] 225 Method parameters 31 Oxidation DP Basic DP Basic DP Photolytic DP Acidic DP Thermal DP Acidic DP Thermal DP Acidic DP Typical development solution chromatogram
  • 32. 32 Chromatograms showing CTP spiked with forced degradation products and placebo chromatogram, indicating absence of interference
  • 33. Parameters DP 1 DP2 DP 3 DP 4 DP 5 CTP DP 6 DP 7 DP 8 DP 9 Reference value Resolution (R) 1.95 5.3 2.7 3.16 5.56 6.21 1.1 2.25 8.44 R >1.5 Selectivity factor (α) 1.14 1.07 1.01 1.19 1.32 1.36 1.05 1.11 1.49 >1 CTP Asymmetry factor (As) 0.95 0.9-1.2 CTP Capacity factor (K΄) 2.82 1–10 CTP Column efficiency (N) 6700 >2000 Method validation 33 The chromatographic selectivity factor and chromatographic resolution between CTP and degradation products results indicate method specificity.
  • 34. 34 700 600 500 400 300 200 100 9000 8000 7000 6000 5000 4000 3000 2000 1000 0 Conc (µg/ml) Peak area Peak area = 30.19 + 13.12 Concentration; r=0.9999 A calibration curve was constructed by plotting the peak area versus concentration of CTP in range of 90-675µg mL-1,
  • 35. Results of assay validation parameters of the proposed HPLC method for determination of CTP Method parameter HPLC method Linearity range 90-675 µgmL-1 Regression equation (A = bC + a)*  Intercept (a) 30.2  Slope (b) 13.1  Correlation coefficient (r) 0.9999 Accuracy Mean ± St.dev 100.4 ± 0.28 50% 100.50 % 100% 100.64 % 150% 100.06 % Precision (Intraday %RSD)b 0.11 % (Interday %RSD)c 0.44 % t-test (2.228)d 0.62 Robustness 100.1 – 100.7 % LOD [µg mL-1] 5.31 LOQ [µg mL-1] 16.1 *A is the peak area and C is the concentration. b Intraday precision (6different determinations at 100% concentrations of / 2 replicate each (n=6)) cInterday precision (6 different determinations at 100% concentrations of / 2 replicate each (n=6)). dt-tabulated for degree of freedom=10, two sided test at α=0.05. 35 Method validation
  • 36. 36 Dosage form Claimed [mg mL-1] Found[mg mL-1] Recovery% Giasion®400 Tablet Lot No. EE0279 0.450 0.457 101.56 0.448 99.56 0.451 100.22 0.447 99.33 0.455 101.11 Mean ± SD 100.36±0.96 Determination of Cefditoren pivoxil in pharmaceutical formulation (Giasion®400 Tablet ) by the proposed HPLC method Giasion® film coated tablets by Zambon claimed to contain 490.2 mg of cefditoren pivoxil / tablet = 400 mg of cefditoren.
  • 37. Item Cefditoren pivoxil Proposed method Reported method a Mean ± St.dev 99.89±0.69 99.92±0.60 n 7 7 Variance 0.48 0.36 F- value ( 4.28 )b 0.75 Student's t-test (2.45)b 0.09 a Reported method: HPLC method, C18, water- Acetonitrile (50: 50, v/v) as a mobile phase, flow rate of 1.2 ml min-1 and UV detection at 218 nm. b The figures in parenthesis are the corresponding tabulated values at α= 0.05 Statistical comparison between proposed and reported method for the determination of CTP in pure powder form. 37
  • 39. 39 • Validation results demonstrated highly specific, accurate, linear, precise and robust method performance. • AQbD development approach introduced good separation, high robustness and confidence in method ability to deliver intended performance. Design space created during method development, enabled flexibility of method transfer, Provide guidance for troubleshooting method performance. CONCLUSION
  • 40. 40
  • 41. 41

Notes de l'éditeur

  1. Small Stationary Phase Particles Reduce possible pore distance for analyte diffusion hence faster diffusion, Differences in diffusion times out of the pore are reduced, Diffusion time decreased
  2. 6 or 2 repeating via readjust instrument to parameters combination each Every measurement is on duplicate bases
  3. 2D-