©2009 Mourne Training Services Dear Lab Tech Guy, Sometimes while using Agilent HPLC, I am facing problems of more back pressure while purging. Someone told me that it is due to the frit. What is the exact function of this frit and how can I increase the life of the frit? Also, is this frit present in Waters HPLC as well? Yours, Sanjeev (Haryana, India)
Dear Sanjeev, The frit you refer to is located in the purge valve on an Agilent HPLC system. Its purpose is to protect the HPLC system and the column by removing particulates from the mobile phase. This PTFE frit (shown on the right) is intended as a disposable part which is changed on a regular basis, hence the low cost at approximately $30 for a pack of 5. The frit should be changed when you experience high back pressure during purging, the Agilent recommendation is as follows: “The most common way to determine if the Purge Valve frit needs to be replaced is by opening the Purge Valve and establishing a flow of 1 mL/min of Methanol. If the pressure is more than 7 bar, the frit should be replaced”. Exchanging the frit is simple (there is a useful demonstration video on the Agilent website). Make sure to inspect the gold seal when you are changing the frit and replace it too if there is any sign of damage. ©2009 Mourne Training Services Expanded view of the purge valve PTFE frit Gold seal
©2009 Mourne Training Services When you remove an old frit, inspect it carefully. Does it shown any signs of discolouration? Black stains are indicative of microbial growth and this may be one reason why your frits need to be changed very often. To maximise the lifetime of the purge valve frit you need to pay particular attention to your mobile phase. Make sure that your mobile phase is very clean and if possible filtered. Mobile phase reservoirs which contain water or solutions of buffers are susceptible to growth of microorganisms and this is the main source of the discoloration on the purge valve frit. Be very careful when applying expiry dates to these types of solutions, it is preferable to prepare these freshly each time they are required.
The Waters Alliance HPLC systems do not have a frit in the purge valve. However, there is an in-line filter which is located between the pump and the injector (shown on the left). The purpose of this component is also to capture particulate contamination.  When the filter becomes blocked (pressure increase will be observed), it may be removed from the casing and replaced, or it is also possible to extend its life by cleaning in methanol using sonication.  In summary then, filtering the previous for the main advice, the frit is there to help protect your HPLC system and changing it is a normal maintenance procedure.  Make sure your mobile phase is as clean as possible (no particulates, no microbial growth) to extend the lifetime of your frits. Yours preventatively,  ©2009 Mourne Training Services
Brought to you by... www.mournetrainingservices.co.uk ©2009 Mourne Training Services

Contenu connexe

En vedette (7)

Quality control measures in pharmaceutical industry
Quality control measures in pharmaceutical industryQuality control measures in pharmaceutical industry
Quality control measures in pharmaceutical industry
 
Hplc presentation for class
Hplc presentation for classHplc presentation for class
Hplc presentation for class
 
HPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid ChromatographyHPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid Chromatography
 
Integration of chromatographic peaks
Integration of chromatographic peaksIntegration of chromatographic peaks
Integration of chromatographic peaks
 
Quality control of pharmaceutical products
Quality control of pharmaceutical productsQuality control of pharmaceutical products
Quality control of pharmaceutical products
 
Quality Control Tests For Tablets and Capsules(QC)
Quality Control Tests For Tablets and Capsules(QC)Quality Control Tests For Tablets and Capsules(QC)
Quality Control Tests For Tablets and Capsules(QC)
 
HPLC - Peak integration for chromatography
HPLC - Peak integration for chromatographyHPLC - Peak integration for chromatography
HPLC - Peak integration for chromatography
 

Dernier

Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 

Dernier (20)

GenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdfGenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdf
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWEREMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
 
Artificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : UncertaintyArtificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : Uncertainty
 
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
 
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot TakeoffStrategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
 
2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
 
A Beginners Guide to Building a RAG App Using Open Source Milvus
A Beginners Guide to Building a RAG App Using Open Source MilvusA Beginners Guide to Building a RAG App Using Open Source Milvus
A Beginners Guide to Building a RAG App Using Open Source Milvus
 
TrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
TrustArc Webinar - Unlock the Power of AI-Driven Data DiscoveryTrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
TrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
 
ICT role in 21st century education and its challenges
ICT role in 21st century education and its challengesICT role in 21st century education and its challenges
ICT role in 21st century education and its challenges
 
MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a Fresher
 
Corporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptxCorporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptx
 
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
 
AWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of TerraformAWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of Terraform
 
Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024
 
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
 
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
 

Lab Tech Guy on Agilent HPLC purge valve frits

  • 2. ©2009 Mourne Training Services Dear Lab Tech Guy, Sometimes while using Agilent HPLC, I am facing problems of more back pressure while purging. Someone told me that it is due to the frit. What is the exact function of this frit and how can I increase the life of the frit? Also, is this frit present in Waters HPLC as well? Yours, Sanjeev (Haryana, India)
  • 3. Dear Sanjeev, The frit you refer to is located in the purge valve on an Agilent HPLC system. Its purpose is to protect the HPLC system and the column by removing particulates from the mobile phase. This PTFE frit (shown on the right) is intended as a disposable part which is changed on a regular basis, hence the low cost at approximately $30 for a pack of 5. The frit should be changed when you experience high back pressure during purging, the Agilent recommendation is as follows: “The most common way to determine if the Purge Valve frit needs to be replaced is by opening the Purge Valve and establishing a flow of 1 mL/min of Methanol. If the pressure is more than 7 bar, the frit should be replaced”. Exchanging the frit is simple (there is a useful demonstration video on the Agilent website). Make sure to inspect the gold seal when you are changing the frit and replace it too if there is any sign of damage. ©2009 Mourne Training Services Expanded view of the purge valve PTFE frit Gold seal
  • 4. ©2009 Mourne Training Services When you remove an old frit, inspect it carefully. Does it shown any signs of discolouration? Black stains are indicative of microbial growth and this may be one reason why your frits need to be changed very often. To maximise the lifetime of the purge valve frit you need to pay particular attention to your mobile phase. Make sure that your mobile phase is very clean and if possible filtered. Mobile phase reservoirs which contain water or solutions of buffers are susceptible to growth of microorganisms and this is the main source of the discoloration on the purge valve frit. Be very careful when applying expiry dates to these types of solutions, it is preferable to prepare these freshly each time they are required.
  • 5. The Waters Alliance HPLC systems do not have a frit in the purge valve. However, there is an in-line filter which is located between the pump and the injector (shown on the left). The purpose of this component is also to capture particulate contamination. When the filter becomes blocked (pressure increase will be observed), it may be removed from the casing and replaced, or it is also possible to extend its life by cleaning in methanol using sonication. In summary then, filtering the previous for the main advice, the frit is there to help protect your HPLC system and changing it is a normal maintenance procedure. Make sure your mobile phase is as clean as possible (no particulates, no microbial growth) to extend the lifetime of your frits. Yours preventatively, ©2009 Mourne Training Services
  • 6. Brought to you by... www.mournetrainingservices.co.uk ©2009 Mourne Training Services