SlideShare une entreprise Scribd logo
1  sur  31
Hygiena SUPERSNAP
ATP and PRO-Clean
Protein Swab Testing
Hygiena SUPERSNAP ATP Swabs and
Luminometer
• . The SuperSnap swabs allows for detection of concentrations of ATP
from 1 -1000 ppm and emits light directly proportional to the
concentrateion swabbed by virtue of the Luciferase/ Luciferin
reagent.
• The ATP luminometer (Figure 1) measures light generated by the
Luciferase/ Luciferin system and reports in Relative Light Units (RLU);
the higher the concentration of ATP the higher the RLU.
Hygiena SUPERSNAP ATP Swab and
Luminometer
Figure1: Hygenia SuerpSnap swab
Hygeian PRO-Clean Protein Swab
• The PRO-Clean Rapid Protein Residue test (Figure 2) is a colorimetric
test utilizing the BCA assay which turns from green to purple if
protein is present in the sample; green indicating a negative protein
sample and purple for a positive protein sample.
Hygiena PRO-Clean Protein Swab
Figure 2: Hygenia Pro-Clean Protein swab Figure 3: The colorimetric indicator on the PRO-Clean swab
METHODS
Methods: Stainless Steel Strips
• To test the efficacy of the Hygiena swabs using the stainless steel
strips as a surface to inoculate, four sample sets were prepared: a
control, a 5.0% Horse blood, 0.5% Horse blood and Bovine serum
albumin set. Each set consisted of 25 stainless steel strips which were
sonicated with house deionized water for 20 minutes.
• 10 µl were used to inoculate each strip with the exception of the
control group. The blood solutions and BSA were allowed to dry for
20 minutes and the first readings were taken denoted as day 0.
• Readings were taken every 7 days for 14 days.
Methods: Lumbar RAPIDCLEAN Ronguers
• Using Lumbar RAPDICLEAN Ronguers as a substrate to inoculate, five
samples sets were prepared: the four sets originally prepared in the
stainless steel strips model with the addition of a modified version of
nelson soil (15 ml egg yolk, 15ml Horse blood, and 300 mg of Hog
musin).
• Eight areas, as shown in Figure 4, were selected to inoculate with the
four protein samples (1 control group).
• The same protocol was followed for the Lumbar rongeurs as for the
stainless steel strips with the exception of swab readings be obtained
at day 0 only.
Lumbar RAPIDCLEAN Ronguers
Figure 4: Two hard to clean (HTC) and two Easy to clean (ETC)
areas were selected to be inoculated for swab analysis. The two
hard to clean areas are indentations in the device.
Methods: micro BCA assay Standard curve
• In order to determine the protein absorbance and concentration of
the four samples the micro BCA assay kit was used to obtain a
standard curve using Ultraviolet visible spectrophotometry.
• The protocol provided by Pierce for the micro BCA assay standard
curve was followed in order to receive delivery counts of protein
concentrations of the blood samples used to inoculate the stainless
steel strips and the lumbar RAPIDCLEAN Rongeurs.
• The samples were prepared for mass spec analysis by aliquoting 10µl
in to 990µl of water and mixing with the working reagent provided by
the micro BCA assay.
RESULTS
Results: ATP and protein residue readings
from Stainless Steel Strips
DAY 0 ATP RLU PROTEIN SWAB
-control 0 -
Control 1, 2, 3 -, -, -
5.0% Horse Blood 3920, 3576, 3448 +++, +++, +++
0.5% Horse Blood 1708, 1904, 2366 +, +, -
BSA 4, 4, 3 ++, ++, ++
DAY 7 ATP RLU PROTEIN SWAB
-control 0 -
Control 2, 2, 10 -, -, -
5.0% Horse Blood 2835, 2782, 3412 +++, +++, +++
0.5% Horse Blood 751, 1155, 1683 +,+,+
BSA 35, 6, 10 ++, ++, ++
DAY 14 ATP RLU PROTEIN SWAB
-control 0 -
control 5, 12, 7 -, -, -
5.0% Horse Blood 2838, 3321, 1422 +++, +++, +++
Table 1: Chart showing the
data obtained from the
stainless steel strips when
inoculated with 5.0%, 0.5%
Horse blood and Bovine
Serum Albumin. The
negatives and positives
correlate with the color
indicator on the PRO-Clean
swab. A (-) is green, (+) is
grey, (++) is light purple and
(+++) is purple and the color
change indicates protein
concentration from no
protein concentration to a
high protein concentration.
Results: ATP and protein residue readings
from Lumbar RAPIDCLEAN Rongeurs
CONTROL ATP RLU PROTEIN SWAB
Easy to Clean 0, 0 -, -
Hard to Clean 0, 0 -, -
5.0% Horse Blood ATP RLU PROTEIN SWAB
Easy to Clean 3097, 2396 +++, +++
Hard to Clean 1180, 1410 +++, +++
0.5% Horse Blood ATP RLU PROTEIN SWAB
Easy to Clean 2026, 2499 +++,+++
Hard to Clean 900, 27 +,+
BSA ATP RLU PROTEIN SWAB
Easy to Clean 6, 14 -, -
Hard to Clean 7, 10 -,-
Nelson Soil ATP RLU PROTEIN SWAB
Easy to Clean 6008, 5887 +++, +++
Hard to clean 2467, 3014 +++, +++
Table 2: Results
from the ATP and
protein swabbing of
the Lumbar
Rongeurs.
The negatives and
positives correlate
with the color
indicator on the
PRO-Clean swab.
A (-) is green, (+)
is grey, (++) is light
purple and (+++)
is purple and the
color change
indicates protein
concentration
from no protein
concentration to a
high protein
concentration.
Results: Protein absorbance and
concentrations from stainless steel strips
Sample (10µl) A @562nm Protein conc. µg/ml
5.0% Horse Blood 2.041
2.207
2.240
110.581
125.100
128.087
0.5% Horse Blood 0.181
0.151
0.170
5.280
4.445
4.971
BSA 0.885
0.930
0.942
32.741
35.010
35.625
Table 3: Protein absorbance and concentration of the horse
blood samples and the BSA sample on day 0 of inoculation of
the stainless steel strips.
Results: Protein absorbance and concentrations
from Lumbar RAPIDLCLEAN Ronguers
Sample (10µl) A @ 562nm Protein conc. µg/ml
5.0% Horse Blood 2.813
2.691
2.412
88.883
84.589
74.890
0.5% Horse Blood 0.308
0.348
0.264
7.231
8.427
5.919
BSA 0.825
0.948
0.924
22.572
26.787
26.037
Nelson Soil 3.806
3.845
3.825
125.049
114.986
125.762
Table 4: Protein absorbance and concentration of horse blood
samples, BSA, and Nelson soil after inoculation of the Lumbar
Rongeurs.
Conclusion
• Both the Hygiena SuperSnap ATP swabs and PRO-Clean Rapid Protein
Test swabs are both effective at detecting ATP and protein
concentrations after exposure to air for a period of time.
• The results obtained from the two tests indicate that the Hygiena
SuperSnap ATP swabs seem to be more sensitive, and the sensitivity
for both swabs may be determined by the surface area swabbed.
• This study allowed the efficacy of both swabs to be tested indicating
that both can be used to determine whether surface has been
cleaned thoroughly and can be applicable to a hospital or food
industry setting as well as testing loaner medical kits for any microbial
contamination after use.
CRUDE PROBIOTIC
BACTERIAL LYSATE
ACTIVITY TESTING
USING THE BacT/ALERT
SYSTEM
Nutraceutix Crude Bacterial Probiotic lysates
and Powder Lysates from Biocare
• The crude bacterial probiotic lysates prepared by JJSA from cultrues
provided by Nutraceutix were exposed to 10kgy of electron beam
radiation.
• The prepared lysates were transferred from sterile bags to sterile
45ml Falcon tubes and stored at -70°C.
• The Powder Lysates were sent from Biocare Coppenhagen.
Nutraceutix Crude Probioitc Lysates
Table 1: Only lysates
1,2,4,6,9,10,13,14, and
16 were tested using
the BacT/ALERT
system.
Biocare Powder Lysates
BacT/ALERT system
BacT/ALERT system
METHODS and
METHODOLOGY
Methodology
• With traditional detection methodslooking for ZOI, 10µl of 10^8 of
S.aureus was added into a top agar 7ml (TSB with 0.7% agar) then
poured onto a TSA plate.
• 50µl of the tested lysates were deposited onto the S.aureus
inoculated plates.
• With the Biocare lysates there was a milky deposit on the plate,
making it diffuclt to obtain ZOI. This was corrected using the
BacT/ALERT system which can detect bacterial growth in turbidity.
Methods: Inoculation of BacT/ALERT bottles
and treatment with the Lysates
• A Quanti-Cult of Staphylococcus aureus was grown in A BacT/ALERT
bottle overnight in the BacT/ALERT system at 35°C.
• The overnight culture was then serially diluted to 10⁻⁴, where the
10⁻² dilution was used to inoculate the BacT/AELRT bottles treated
with the bacterial lysates.
• The Biocare powder lysates were weighed out to 0.4 grams and then
asepcitcally transferred from sterile microcentrifuge tubes to the
BacT/ALERT bottles.
• For the Crude Probiotic Bacterial Lysates from Nutraceutix, 0.4ml was
used to treat the BacT/ALERT bottles.
Methods: Inoculation of BacT/ALERT bottles
and treatment with the Lysates
• Five samples sets were prepared.
• Two samples sets with 0.4ml and 0.4g treated with the Crude Probiotic
Bacterial Lysate and the Powder Lysate respectively, Sample sets 1A and 1B.
• Four sample sample sets with 2ml and 2.0g treated with the Crude Porbioitc
Bacterial Lysate and the Powder Lysate respectively, Sample sets 2A, 2B, 3A
and 3B.
• One sample set of 2.0g of Powder lysate without any inoculum, sample set
4A.
• The sample sets were incubated in the BacT/Alert system for 7 days in order to
asses the absence of bacterial growth or time of detection of bacterial growth,
with the exception of sample set3A. Sample set 3A, was left to be agitated by
the BacT/ALERT system in order to effectively dissolve the powder lysate.
Sample Set 1A: Biocare Lysates 0.4g
Sample (Set 1A) Time to detection (hours)
BL Negative control No Growth
S. aureus 10⁻¹ 0.17
S. aureus 10⁻² 0.29
S. aureus 10⁻³ 0.44
S. aureus 10⁻⁴ 0.67
BL #1 0.29
BL #2 0.28
BL #3 0.29
BL #4 0.29
BL #5 0.27
BL #6 0.27
BL #7 0.28
BL #8 0.27
BL #9 0.29
Sample set 1B: Nutraceutix Lysates 0.4ml
Sample (Set 1B) Time to detection (hours)
BL Negative control No Growth
S. aureus 10⁻¹ 0.17
S. aureus 10⁻² 0.29
S. aureus 10⁻³ 0.44
S. aureus 10⁻⁴ 0.67
CPBL #1 0.28
CPBL #2 0.27
CPBL #4 0.27
CPBL #6 0.28
CPBL #9 0.28
CPBL #10 0.28
CPBL #13 0.29
CPBL #14 0.27
CPBL #16 0.33
Sample set 2A: Biocare Lysates 2.0g
Sample Time to detection (hours, Set 2A)
BL Negative control No Growth
S. aureus 10⁻¹ 0.18
S. aureus 10⁻² 0.28
S. aureus 10⁻³ 0.41
S. aureus 10⁻⁴ 0.68
BL #1 0.52
BL #2 0.48
BL #3 0.47
BL #4 0.44
BL #5 0.47
BL #6 0.44
BL #7 0.47
BL #8 0.45
BL #9 0.44
Time to detection (hours, Set 3A)
No Growth
0.19
0.29
0.43
0.61
0.46
0.41
0.46
0.43
0.39
0.39
0.46
0.36
0.40
Sample Set 3B: Nutraceutix Lysates 2ml
Sample
BL Negative control
S. aureus 10⁻¹
S. aureus 10⁻²
S. aureus 10⁻³
S. aureus 10⁻⁴
CPBL #1
CPBL #2
CPBL #4
CPBL #6
CPBL #9
CPBL #10
CPBL #13
CPBL #14
CPBL #16
Time to detection (hours, 2B)
No Growth
0.18
0.28
0.41
0.68
0.51
0.40
0.43
0.52
0.40
0.55
0.69
0.37
1.00
Time to detection (hours, Set 3B)
No Growth
0.20
0.33
0.46
0.59
1.27
0.74
0.72
0.75
0.83
0.97
1.42
0.39
0.82

Contenu connexe

Tendances

Can LCMSMS be used in horse meat detection?
Can LCMSMS be used in horse meat detection?Can LCMSMS be used in horse meat detection?
Can LCMSMS be used in horse meat detection?SCIEX
 
Detection of Melamine in Milk
Detection of Melamine in MilkDetection of Melamine in Milk
Detection of Melamine in MilkArto Per
 
Life science products of GL Sciences
Life science products of GL SciencesLife science products of GL Sciences
Life science products of GL SciencesGeert Alkema
 
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...
"Productivity and Simplicity" -  Streamlining Cumbersome Sample Preparation i..."Productivity and Simplicity" -  Streamlining Cumbersome Sample Preparation i...
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...Oscar Cabrices PhD
 
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDATocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDAShimadzu Scientific Instruments
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsSCIEX
 
Bmb 496 spring '10
Bmb 496  spring '10Bmb 496  spring '10
Bmb 496 spring '10buk11
 
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Shimadzu Scientific Instruments
 
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...Shimadzu Scientific Instruments
 
Y01062155162
Y01062155162Y01062155162
Y01062155162iosrjce
 
Manual of quality analysis - Soya products
Manual of quality analysis - Soya productsManual of quality analysis - Soya products
Manual of quality analysis - Soya productsVasu Chithiravelu
 
Proteins and amino acids; Brief introduction of methods for quantification of...
Proteins and amino acids; Brief introduction of methods for quantification of...Proteins and amino acids; Brief introduction of methods for quantification of...
Proteins and amino acids; Brief introduction of methods for quantification of...umair7894
 
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Shimadzu Scientific Instruments
 
Proteins and amino acids
Proteins and amino acidsProteins and amino acids
Proteins and amino acidsShruti Tyagi
 
The Analysis of Baby Foods and Juices for Metals to Protect a Sensitive Popul...
The Analysis of Baby Foods and Juices for Metals to Protect a Sensitive Popul...The Analysis of Baby Foods and Juices for Metals to Protect a Sensitive Popul...
The Analysis of Baby Foods and Juices for Metals to Protect a Sensitive Popul...PerkinElmer, Inc.
 
Automated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterAutomated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterRick Youngblood
 

Tendances (20)

Can LCMSMS be used in horse meat detection?
Can LCMSMS be used in horse meat detection?Can LCMSMS be used in horse meat detection?
Can LCMSMS be used in horse meat detection?
 
SITI AZIAH
SITI AZIAHSITI AZIAH
SITI AZIAH
 
Detection of Melamine in Milk
Detection of Melamine in MilkDetection of Melamine in Milk
Detection of Melamine in Milk
 
Life science products of GL Sciences
Life science products of GL SciencesLife science products of GL Sciences
Life science products of GL Sciences
 
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...
"Productivity and Simplicity" -  Streamlining Cumbersome Sample Preparation i..."Productivity and Simplicity" -  Streamlining Cumbersome Sample Preparation i...
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...
 
HPLC Combinatorial Libraries
HPLC Combinatorial LibrariesHPLC Combinatorial Libraries
HPLC Combinatorial Libraries
 
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDATocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminants
 
Bmb 496 spring '10
Bmb 496  spring '10Bmb 496  spring '10
Bmb 496 spring '10
 
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
 
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
 
Y01062155162
Y01062155162Y01062155162
Y01062155162
 
Manual of quality analysis - Soya products
Manual of quality analysis - Soya productsManual of quality analysis - Soya products
Manual of quality analysis - Soya products
 
Proteins and amino acids; Brief introduction of methods for quantification of...
Proteins and amino acids; Brief introduction of methods for quantification of...Proteins and amino acids; Brief introduction of methods for quantification of...
Proteins and amino acids; Brief introduction of methods for quantification of...
 
Lal presentation
Lal presentation Lal presentation
Lal presentation
 
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
 
Proteins and amino acids
Proteins and amino acidsProteins and amino acids
Proteins and amino acids
 
Chromatography: Feedstock Characterization and Fermentation Monitoring of Bio...
Chromatography: Feedstock Characterization and Fermentation Monitoring of Bio...Chromatography: Feedstock Characterization and Fermentation Monitoring of Bio...
Chromatography: Feedstock Characterization and Fermentation Monitoring of Bio...
 
The Analysis of Baby Foods and Juices for Metals to Protect a Sensitive Popul...
The Analysis of Baby Foods and Juices for Metals to Protect a Sensitive Popul...The Analysis of Baby Foods and Juices for Metals to Protect a Sensitive Popul...
The Analysis of Baby Foods and Juices for Metals to Protect a Sensitive Popul...
 
Automated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterAutomated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling Poster
 

Similaire à JJSA

Athira v c
Athira v cAthira v c
Athira v cAthiravc
 
Determination of Cholesterol Level in Human Serum
Determination of Cholesterol Level in Human SerumDetermination of Cholesterol Level in Human Serum
Determination of Cholesterol Level in Human SerumPerkinElmer, Inc.
 
ICSA17 - Bula Dosagem Proteínas
 ICSA17 - Bula Dosagem Proteínas ICSA17 - Bula Dosagem Proteínas
ICSA17 - Bula Dosagem ProteínasRicardo Portela
 
Advances in size exclusion chromatography for the analysis of small proteins ...
Advances in size exclusion chromatography for the analysis of small proteins ...Advances in size exclusion chromatography for the analysis of small proteins ...
Advances in size exclusion chromatography for the analysis of small proteins ...Eduardo Castro
 
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MSQuantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MSCovance
 
A visual chip-based coimmunoprecipitation technique for analysis of protein–p...
A visual chip-based coimmunoprecipitation technique for analysis of protein–p...A visual chip-based coimmunoprecipitation technique for analysis of protein–p...
A visual chip-based coimmunoprecipitation technique for analysis of protein–p...Qing Chen
 
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Kate?ina Svobodov
 
Bioanalytical method development and validation .
Bioanalytical method development and validation .Bioanalytical method development and validation .
Bioanalytical method development and validation .Shubham Bora
 
WESTERN BLOTTING ADYA.pptx
WESTERN BLOTTING ADYA.pptxWESTERN BLOTTING ADYA.pptx
WESTERN BLOTTING ADYA.pptxAADYARAJPANDEY1
 
2 collinspres
2 collinspres2 collinspres
2 collinspresdream10f
 
Automated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPAutomated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPMark Hayward
 
Method Development Poster-MicroLiter
Method Development Poster-MicroLiterMethod Development Poster-MicroLiter
Method Development Poster-MicroLiterRick Youngblood
 
Conceição et al, 2006. orpotrin a novel vasoconstrictor peptide from the veno...
Conceição et al, 2006. orpotrin a novel vasoconstrictor peptide from the veno...Conceição et al, 2006. orpotrin a novel vasoconstrictor peptide from the veno...
Conceição et al, 2006. orpotrin a novel vasoconstrictor peptide from the veno...pryloock
 
Method-of-analysis-of-milk-contaminanta.pptx
Method-of-analysis-of-milk-contaminanta.pptxMethod-of-analysis-of-milk-contaminanta.pptx
Method-of-analysis-of-milk-contaminanta.pptxSurendra664913
 
EVALUATION OF AQUAFEED INGREDIENTS AND FEED QUALITY
EVALUATION OF AQUAFEED INGREDIENTS AND FEED QUALITYEVALUATION OF AQUAFEED INGREDIENTS AND FEED QUALITY
EVALUATION OF AQUAFEED INGREDIENTS AND FEED QUALITYsouravfnftmb306
 
Determination of Structural Carbohydrates & Lignin in Biomass
Determination of Structural Carbohydrates & Lignin in BiomassDetermination of Structural Carbohydrates & Lignin in Biomass
Determination of Structural Carbohydrates & Lignin in BiomassBiorefineryEPC™
 

Similaire à JJSA (20)

Athira v c
Athira v cAthira v c
Athira v c
 
Chromatography: Trends and Developments in MAb Screening and Characterization
Chromatography: Trends and Developments in MAb Screening and CharacterizationChromatography: Trends and Developments in MAb Screening and Characterization
Chromatography: Trends and Developments in MAb Screening and Characterization
 
Determination of Cholesterol Level in Human Serum
Determination of Cholesterol Level in Human SerumDetermination of Cholesterol Level in Human Serum
Determination of Cholesterol Level in Human Serum
 
Khushbu_Simoa_06-16
Khushbu_Simoa_06-16Khushbu_Simoa_06-16
Khushbu_Simoa_06-16
 
Stages of Production of Escozine For Pets™
Stages of Production of Escozine For Pets™ Stages of Production of Escozine For Pets™
Stages of Production of Escozine For Pets™
 
ICSA17 - Bula Dosagem Proteínas
 ICSA17 - Bula Dosagem Proteínas ICSA17 - Bula Dosagem Proteínas
ICSA17 - Bula Dosagem Proteínas
 
Advances in size exclusion chromatography for the analysis of small proteins ...
Advances in size exclusion chromatography for the analysis of small proteins ...Advances in size exclusion chromatography for the analysis of small proteins ...
Advances in size exclusion chromatography for the analysis of small proteins ...
 
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MSQuantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
 
A visual chip-based coimmunoprecipitation technique for analysis of protein–p...
A visual chip-based coimmunoprecipitation technique for analysis of protein–p...A visual chip-based coimmunoprecipitation technique for analysis of protein–p...
A visual chip-based coimmunoprecipitation technique for analysis of protein–p...
 
E5 protein
E5 proteinE5 protein
E5 protein
 
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
 
Bioanalytical method development and validation .
Bioanalytical method development and validation .Bioanalytical method development and validation .
Bioanalytical method development and validation .
 
WESTERN BLOTTING ADYA.pptx
WESTERN BLOTTING ADYA.pptxWESTERN BLOTTING ADYA.pptx
WESTERN BLOTTING ADYA.pptx
 
2 collinspres
2 collinspres2 collinspres
2 collinspres
 
Automated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPAutomated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSP
 
Method Development Poster-MicroLiter
Method Development Poster-MicroLiterMethod Development Poster-MicroLiter
Method Development Poster-MicroLiter
 
Conceição et al, 2006. orpotrin a novel vasoconstrictor peptide from the veno...
Conceição et al, 2006. orpotrin a novel vasoconstrictor peptide from the veno...Conceição et al, 2006. orpotrin a novel vasoconstrictor peptide from the veno...
Conceição et al, 2006. orpotrin a novel vasoconstrictor peptide from the veno...
 
Method-of-analysis-of-milk-contaminanta.pptx
Method-of-analysis-of-milk-contaminanta.pptxMethod-of-analysis-of-milk-contaminanta.pptx
Method-of-analysis-of-milk-contaminanta.pptx
 
EVALUATION OF AQUAFEED INGREDIENTS AND FEED QUALITY
EVALUATION OF AQUAFEED INGREDIENTS AND FEED QUALITYEVALUATION OF AQUAFEED INGREDIENTS AND FEED QUALITY
EVALUATION OF AQUAFEED INGREDIENTS AND FEED QUALITY
 
Determination of Structural Carbohydrates & Lignin in Biomass
Determination of Structural Carbohydrates & Lignin in BiomassDetermination of Structural Carbohydrates & Lignin in Biomass
Determination of Structural Carbohydrates & Lignin in Biomass
 

JJSA

  • 1. Hygiena SUPERSNAP ATP and PRO-Clean Protein Swab Testing
  • 2. Hygiena SUPERSNAP ATP Swabs and Luminometer • . The SuperSnap swabs allows for detection of concentrations of ATP from 1 -1000 ppm and emits light directly proportional to the concentrateion swabbed by virtue of the Luciferase/ Luciferin reagent. • The ATP luminometer (Figure 1) measures light generated by the Luciferase/ Luciferin system and reports in Relative Light Units (RLU); the higher the concentration of ATP the higher the RLU.
  • 3. Hygiena SUPERSNAP ATP Swab and Luminometer Figure1: Hygenia SuerpSnap swab
  • 4. Hygeian PRO-Clean Protein Swab • The PRO-Clean Rapid Protein Residue test (Figure 2) is a colorimetric test utilizing the BCA assay which turns from green to purple if protein is present in the sample; green indicating a negative protein sample and purple for a positive protein sample.
  • 5. Hygiena PRO-Clean Protein Swab Figure 2: Hygenia Pro-Clean Protein swab Figure 3: The colorimetric indicator on the PRO-Clean swab
  • 7. Methods: Stainless Steel Strips • To test the efficacy of the Hygiena swabs using the stainless steel strips as a surface to inoculate, four sample sets were prepared: a control, a 5.0% Horse blood, 0.5% Horse blood and Bovine serum albumin set. Each set consisted of 25 stainless steel strips which were sonicated with house deionized water for 20 minutes. • 10 µl were used to inoculate each strip with the exception of the control group. The blood solutions and BSA were allowed to dry for 20 minutes and the first readings were taken denoted as day 0. • Readings were taken every 7 days for 14 days.
  • 8. Methods: Lumbar RAPIDCLEAN Ronguers • Using Lumbar RAPDICLEAN Ronguers as a substrate to inoculate, five samples sets were prepared: the four sets originally prepared in the stainless steel strips model with the addition of a modified version of nelson soil (15 ml egg yolk, 15ml Horse blood, and 300 mg of Hog musin). • Eight areas, as shown in Figure 4, were selected to inoculate with the four protein samples (1 control group). • The same protocol was followed for the Lumbar rongeurs as for the stainless steel strips with the exception of swab readings be obtained at day 0 only.
  • 9. Lumbar RAPIDCLEAN Ronguers Figure 4: Two hard to clean (HTC) and two Easy to clean (ETC) areas were selected to be inoculated for swab analysis. The two hard to clean areas are indentations in the device.
  • 10. Methods: micro BCA assay Standard curve • In order to determine the protein absorbance and concentration of the four samples the micro BCA assay kit was used to obtain a standard curve using Ultraviolet visible spectrophotometry. • The protocol provided by Pierce for the micro BCA assay standard curve was followed in order to receive delivery counts of protein concentrations of the blood samples used to inoculate the stainless steel strips and the lumbar RAPIDCLEAN Rongeurs. • The samples were prepared for mass spec analysis by aliquoting 10µl in to 990µl of water and mixing with the working reagent provided by the micro BCA assay.
  • 12. Results: ATP and protein residue readings from Stainless Steel Strips DAY 0 ATP RLU PROTEIN SWAB -control 0 - Control 1, 2, 3 -, -, - 5.0% Horse Blood 3920, 3576, 3448 +++, +++, +++ 0.5% Horse Blood 1708, 1904, 2366 +, +, - BSA 4, 4, 3 ++, ++, ++ DAY 7 ATP RLU PROTEIN SWAB -control 0 - Control 2, 2, 10 -, -, - 5.0% Horse Blood 2835, 2782, 3412 +++, +++, +++ 0.5% Horse Blood 751, 1155, 1683 +,+,+ BSA 35, 6, 10 ++, ++, ++ DAY 14 ATP RLU PROTEIN SWAB -control 0 - control 5, 12, 7 -, -, - 5.0% Horse Blood 2838, 3321, 1422 +++, +++, +++ Table 1: Chart showing the data obtained from the stainless steel strips when inoculated with 5.0%, 0.5% Horse blood and Bovine Serum Albumin. The negatives and positives correlate with the color indicator on the PRO-Clean swab. A (-) is green, (+) is grey, (++) is light purple and (+++) is purple and the color change indicates protein concentration from no protein concentration to a high protein concentration.
  • 13. Results: ATP and protein residue readings from Lumbar RAPIDCLEAN Rongeurs CONTROL ATP RLU PROTEIN SWAB Easy to Clean 0, 0 -, - Hard to Clean 0, 0 -, - 5.0% Horse Blood ATP RLU PROTEIN SWAB Easy to Clean 3097, 2396 +++, +++ Hard to Clean 1180, 1410 +++, +++ 0.5% Horse Blood ATP RLU PROTEIN SWAB Easy to Clean 2026, 2499 +++,+++ Hard to Clean 900, 27 +,+ BSA ATP RLU PROTEIN SWAB Easy to Clean 6, 14 -, - Hard to Clean 7, 10 -,- Nelson Soil ATP RLU PROTEIN SWAB Easy to Clean 6008, 5887 +++, +++ Hard to clean 2467, 3014 +++, +++ Table 2: Results from the ATP and protein swabbing of the Lumbar Rongeurs. The negatives and positives correlate with the color indicator on the PRO-Clean swab. A (-) is green, (+) is grey, (++) is light purple and (+++) is purple and the color change indicates protein concentration from no protein concentration to a high protein concentration.
  • 14. Results: Protein absorbance and concentrations from stainless steel strips Sample (10µl) A @562nm Protein conc. µg/ml 5.0% Horse Blood 2.041 2.207 2.240 110.581 125.100 128.087 0.5% Horse Blood 0.181 0.151 0.170 5.280 4.445 4.971 BSA 0.885 0.930 0.942 32.741 35.010 35.625 Table 3: Protein absorbance and concentration of the horse blood samples and the BSA sample on day 0 of inoculation of the stainless steel strips.
  • 15. Results: Protein absorbance and concentrations from Lumbar RAPIDLCLEAN Ronguers Sample (10µl) A @ 562nm Protein conc. µg/ml 5.0% Horse Blood 2.813 2.691 2.412 88.883 84.589 74.890 0.5% Horse Blood 0.308 0.348 0.264 7.231 8.427 5.919 BSA 0.825 0.948 0.924 22.572 26.787 26.037 Nelson Soil 3.806 3.845 3.825 125.049 114.986 125.762 Table 4: Protein absorbance and concentration of horse blood samples, BSA, and Nelson soil after inoculation of the Lumbar Rongeurs.
  • 16. Conclusion • Both the Hygiena SuperSnap ATP swabs and PRO-Clean Rapid Protein Test swabs are both effective at detecting ATP and protein concentrations after exposure to air for a period of time. • The results obtained from the two tests indicate that the Hygiena SuperSnap ATP swabs seem to be more sensitive, and the sensitivity for both swabs may be determined by the surface area swabbed. • This study allowed the efficacy of both swabs to be tested indicating that both can be used to determine whether surface has been cleaned thoroughly and can be applicable to a hospital or food industry setting as well as testing loaner medical kits for any microbial contamination after use.
  • 17. CRUDE PROBIOTIC BACTERIAL LYSATE ACTIVITY TESTING USING THE BacT/ALERT SYSTEM
  • 18. Nutraceutix Crude Bacterial Probiotic lysates and Powder Lysates from Biocare • The crude bacterial probiotic lysates prepared by JJSA from cultrues provided by Nutraceutix were exposed to 10kgy of electron beam radiation. • The prepared lysates were transferred from sterile bags to sterile 45ml Falcon tubes and stored at -70°C. • The Powder Lysates were sent from Biocare Coppenhagen.
  • 19. Nutraceutix Crude Probioitc Lysates Table 1: Only lysates 1,2,4,6,9,10,13,14, and 16 were tested using the BacT/ALERT system.
  • 24. Methodology • With traditional detection methodslooking for ZOI, 10µl of 10^8 of S.aureus was added into a top agar 7ml (TSB with 0.7% agar) then poured onto a TSA plate. • 50µl of the tested lysates were deposited onto the S.aureus inoculated plates. • With the Biocare lysates there was a milky deposit on the plate, making it diffuclt to obtain ZOI. This was corrected using the BacT/ALERT system which can detect bacterial growth in turbidity.
  • 25.
  • 26. Methods: Inoculation of BacT/ALERT bottles and treatment with the Lysates • A Quanti-Cult of Staphylococcus aureus was grown in A BacT/ALERT bottle overnight in the BacT/ALERT system at 35°C. • The overnight culture was then serially diluted to 10⁻⁴, where the 10⁻² dilution was used to inoculate the BacT/AELRT bottles treated with the bacterial lysates. • The Biocare powder lysates were weighed out to 0.4 grams and then asepcitcally transferred from sterile microcentrifuge tubes to the BacT/ALERT bottles. • For the Crude Probiotic Bacterial Lysates from Nutraceutix, 0.4ml was used to treat the BacT/ALERT bottles.
  • 27. Methods: Inoculation of BacT/ALERT bottles and treatment with the Lysates • Five samples sets were prepared. • Two samples sets with 0.4ml and 0.4g treated with the Crude Probiotic Bacterial Lysate and the Powder Lysate respectively, Sample sets 1A and 1B. • Four sample sample sets with 2ml and 2.0g treated with the Crude Porbioitc Bacterial Lysate and the Powder Lysate respectively, Sample sets 2A, 2B, 3A and 3B. • One sample set of 2.0g of Powder lysate without any inoculum, sample set 4A. • The sample sets were incubated in the BacT/Alert system for 7 days in order to asses the absence of bacterial growth or time of detection of bacterial growth, with the exception of sample set3A. Sample set 3A, was left to be agitated by the BacT/ALERT system in order to effectively dissolve the powder lysate.
  • 28. Sample Set 1A: Biocare Lysates 0.4g Sample (Set 1A) Time to detection (hours) BL Negative control No Growth S. aureus 10⁻¹ 0.17 S. aureus 10⁻² 0.29 S. aureus 10⁻³ 0.44 S. aureus 10⁻⁴ 0.67 BL #1 0.29 BL #2 0.28 BL #3 0.29 BL #4 0.29 BL #5 0.27 BL #6 0.27 BL #7 0.28 BL #8 0.27 BL #9 0.29
  • 29. Sample set 1B: Nutraceutix Lysates 0.4ml Sample (Set 1B) Time to detection (hours) BL Negative control No Growth S. aureus 10⁻¹ 0.17 S. aureus 10⁻² 0.29 S. aureus 10⁻³ 0.44 S. aureus 10⁻⁴ 0.67 CPBL #1 0.28 CPBL #2 0.27 CPBL #4 0.27 CPBL #6 0.28 CPBL #9 0.28 CPBL #10 0.28 CPBL #13 0.29 CPBL #14 0.27 CPBL #16 0.33
  • 30. Sample set 2A: Biocare Lysates 2.0g Sample Time to detection (hours, Set 2A) BL Negative control No Growth S. aureus 10⁻¹ 0.18 S. aureus 10⁻² 0.28 S. aureus 10⁻³ 0.41 S. aureus 10⁻⁴ 0.68 BL #1 0.52 BL #2 0.48 BL #3 0.47 BL #4 0.44 BL #5 0.47 BL #6 0.44 BL #7 0.47 BL #8 0.45 BL #9 0.44 Time to detection (hours, Set 3A) No Growth 0.19 0.29 0.43 0.61 0.46 0.41 0.46 0.43 0.39 0.39 0.46 0.36 0.40
  • 31. Sample Set 3B: Nutraceutix Lysates 2ml Sample BL Negative control S. aureus 10⁻¹ S. aureus 10⁻² S. aureus 10⁻³ S. aureus 10⁻⁴ CPBL #1 CPBL #2 CPBL #4 CPBL #6 CPBL #9 CPBL #10 CPBL #13 CPBL #14 CPBL #16 Time to detection (hours, 2B) No Growth 0.18 0.28 0.41 0.68 0.51 0.40 0.43 0.52 0.40 0.55 0.69 0.37 1.00 Time to detection (hours, Set 3B) No Growth 0.20 0.33 0.46 0.59 1.27 0.74 0.72 0.75 0.83 0.97 1.42 0.39 0.82