SlideShare une entreprise Scribd logo
1  sur  25
Presented by: Facilitated to:
Mr. L. Sanathoiba Singha
M. Pharm 1st Semester
Department of Pharmaceutical Analysis.
Dr. C. Sreedhar
Head of Department
Department of Pharmaceutical Analysis.
Karnataka College of Pharmacy
Bengaluru-64, Karnataka.
1
1. ELECTRONIC BALANCE
DESIGN QUALIFICATION:
The supplier has ISO 9001:2000 certification.
INSTALLATION QUALIFICATION:
• Checking of all requirements set during the selection of instrument such as electricity, humidity, temperature, etc.
• Allowing sufficient shelf space for the equipment, SOPs, operating manuals, etc.
• Comparing equipment, as received, with purchase order, including accessories, spare parts, etc.
• Checking documentation for completeness like operating manuals, maintenance instructions, health and safety instructions, etc.
• Checking equipment for any damage.
• Reading the supplier’s safety instructions, if there are any.
• Installing hardware following the manufacturer’s recommendation.
• Switching on the electronic balance and ensuring that all the modules power up.
• Preparing an installation report
2
Test procedure Acceptance limits Test frequency Remarks
• Measurement of reference
weight by using 10mg, 50mg,
100mg, 500mg, 1g, 5g, 10g and
20g.
• Comparing the actual results
with reference weights.
0.1% • Daily or when used, whatever is
longer with internal reference
weights.
• Yearly with traceable external
weights through instrument
vendor.
External standard should be
traceable to national standard.
OPERATIONAL QUALIFICATION:
PERFORMANCE QUALIFICATION:
• Defining weights and weight classes to be used.
• Defining acceptance limits of results.
• Defining test intervals.
• Defining corrective actions on what to do if the electronic balance does not meet the criteria, in other words if the results are out of
specification.
Operational qualification for electronic balance.
3
2. pH METER
DESIGN QUALIFICATION:
• Selection of a manufacturer that can satisfy the needs of the laboratory.
• Should have ISO 9001 certification.
INSTALLATION QUALIFICATION:
• Checking of all requirements set during the selection of instrument such as electricity, humidity, temperature, etc.
• Allowing sufficient shelf space for the equipment, SOPs, operating manuals, etc.
• Comparing equipment, as received, with purchase order, including accessories, spare parts, etc.
• Checking documentation for completeness like operating manuals, maintenance instructions, health and safety instructions, etc.
• Checking equipment for any damage.
• Reading the supplier’s safety instructions, if there are any.
• Installing hardware following the manufacturer’s recommendation.
• Switching on the pH meter and ensuring that all the modules power up.
• Preparing an installation report.
4
PERFORMANCE QUALIFICATION:
A log book is prepared for operators and service technicians to record all equipment related activities in chronological order.
Parameter to be checked Typical frequency Typical tolerance limit
Slope Each day of use or before each series of
measurements
95.0 to 105.0%
Offset (pH asymmetry) +/-30mV (offset of manufacturer +/-0.5 pH
units at defined temperature)
Reference buffer solution +/-0.05 pH units
Test procedure Acceptance limits Test frequency Remarks
• Calibration with buffer solution
at pH 7 and pH 4.
• Verification with buffer solution
at pH 5 or pH 6.
<0.005 pH units Before each use • Buffer solutions are
commercially available in either
tablet or solution form.
• Some pH meters also have a
temperature sensor, which
should be calibrated every six
months.
• The electrodes should be
washed between each
measurement and the buffers
should be visually checked for
cleanliness and absence of
microbial growth before each
use.
OPERATIONAL QUALIFICATION:
5
3. UV-VISIBLE SPECTROPHOTOMETER
DESIGN QUALIFICATIONS:
General considerations
Feature Consideration
Instrument set-up and control • Pc based or integrated system.
• Software control of operating conditions and parameters.
• Data acquisition, processing and presentation needs.
• In-built diagnostic facilities.
• Self-testing diagnostics.
• Detector options.
• Source options.
Sample introduction and throughput • Sample throughput, presentation and introduction needs.
• Sample thermostating requirements.
• Sample volume requirements.
Materials of construction Resistance to corrosion, contaminated by solvents and sample.
6
Contd…
Feature Consideration
Installation requirements • Size and weight in shipped form.
• Access restrictions to permanent site.
Operational requirements • Limitations on, requirements for and expected consumption of services,
utilities, and consumables (e.g. lamps).
• Ventilation requirements.
• Controlling software embedded or separate package.
Environmental conditions • Environmental conditions within which, or range over which, the
instrument is required to work within specification.
• Recyclability of instrument.
Maintenance and support • Ease of user maintenance and cleaning.
• Cost, longevity and availability of spares and parts.
• Cost and availability of service contacts and technical support.
• Suggested intervals between and procedures for maintenance and
calibration of the instrument.
• The period for which support (qualification, maintenance, parts, etc.)
for the instrument can be guaranteed.
Training requirements The level of skill required to operate the instrument and details of any
training necessary and courses provided by the supplier.
Accessories Desired adaptation of the system
Documentation • Clarity and ease of use of documentation (e.g. operating manuals,
qualification protocols, model SOPs).
• Manuals available as separate hardcopies or computer files (e.g. on CD
ROM), or held as digitally embedded copies within the instrument.
• Unique document identification by version number and date of issue.
Health and safety • Health and safety and environmental issues and /or requirements.
• Significant generation of ozone.
7
Design Qualification of UV-Visible Spectrophotometer
Feature Consideration
Photometric drift Stability of the measurement over time, making comparisons meaningful.
Stray light Affects accuracy and linearity
Data collection • Expression of data on % absorption or % transmission scales.
• Single or repeat scan facility.
• Single or multiple wavelength monitoring.
• Full spectra data (e.g. diode array).
Data manipulation • Spectra subtraction.
• Spectral derivatives.
• Audit trail facility.
• Is raw data uncorruptable ?
• Back-up and restore capabilities.
Data storage • No storage.
• Embedded memory.
• File transmission to external device.
• Stored output via optical or magnetic media.
• System security.
Data output • Inbuilt hardcopy output-data only, spectrum only, data+ spectrum.
• PC based data processing.
• Custom report generation.
8
INSTALLATION QUALIFICATION:
• The instrument has been delivered as ordered, e.g. according to the DQ or purchase order.
• The instrument has been checked and verified as undamaged.
• The appropriate documentation has been supplied and it is of correct issue and uniquely identified by part number, version number and date.
• Details of recommended service and calibration intervals (carried out by the supplier) have been provided.
• Intervals, methods and instructions for user-maintenance and calibration have been provided along with contact points for service and spare parts.
• The correct hardware, firmware and software have been supplied and it is of correct tissue and uniquely identified by part number.
• Information has been provided on consumables required during the normal operation of the instrument.
• The selected environment for the instrument system is suitable, with adequate room for unpacking, installation, operation and servicing, and have appropriate
services and utilities have been provided.
• Health and safety and environmental information relating to the operation of the instrument have been provided and the proposed working environment is consistent
with these requirements.
• The response of the instrument to the initial application of power is as expected and checked for any deviations.
OPERATIONAL QUALIFICATION:
Test procedure Acceptance limits Test frequency Remarks
Wavelength maxima of holmium
oxide solution is measured at 241,
287, 361 and 536 nm.
• ±1 nm in UV range
• ±3 nm in visible range
Six months Standard is traceable to national standard.
Wavelength accuracy- holmium oxide solution
Wavelength accuracy- holmium oxide filter
Test procedure Acceptance limits Test frequency Remarks
Wavelength maxima of holmium
oxide filter is measured at 361, 418
and 536 nm.
• ±1 nm in UV range
• ±3 nm in visible range
Six months Filter is traceable to national standard. 9
Test procedure Acceptance limits Test frequency Remarks
wavelength A 1%/1cm Limits
Absorption of potassium chromate
solution is measured.
• 235 (min)
• 257 (max)
• 313 (min)
• 350 (max)
• 124.5
• 144.0
• 48.6
• 106.6
• 122.9-126.2
• 142.4-145.7
• 47.0-50.3
• 104.9-108.2
Six months Standard is traceable to national standard.
Absorption intensity
Test procedure Acceptance limits Test frequency Remarks
Absorption of 1.2% potassium
chloride solution at 200 nm is
measured against water.
Absorption at 200 nm >2.0 AU Six months Alternative standards:
• Sodium nitrite
• Sodium iodide.
Stray light
Test procedure Acceptance limits Test frequency Remarks
Transmission of cuvette (water)
against air.
a) Quartz cuvettes:
• 85% at 220 nm
• 88% at 240 nm
b) Glass cuvettes:
• 85% at 356 nm
• 88% at 650 nm
Six months Standard is traceable to national standard.
Control of cuvette
10
Test procedure Acceptance limits Test frequency
Spectrum of toluene is measured from 260 to 275 nm.
Ratio of peak height to valley is calculated.
Absorbance ratio peak to valley at 266/269 nm > 1.5 Six months
Wavelength resolution
PERFORMANCE QUALIFICATION:
Parameter Reason Procedure
Wavelength calibration Critical to accuracy of results. Can be determined using only two calibration
wavelengths, preferably bracketing the analytical
wavelength.
Photometric accuracy Critical to accuracy of results Can be determined for particular absorbance at
particular wavelengths using calibrated filters or
cuvettes filled with standard solutions.
Linearity of photometric response Critical to accuracy of results Can be determined by checking the accuracy of a
number of normal absorbances across the desired
absorbance range. Extrapolation beyond highest
standard is inadvisable.
Signal to noise ratio Important for sensitivity and limit detection Can be determined from the response of a detector
to a dilute standard solution and/or a blank.
Stray light Important measure of ‘health’ of whole system Record uncorrected spectrum in single beam mode
for double beam instruments or with baseline
correction switched off for single beam instruments.
11
Parameter Reason Procedure
Software accuracy and precision, derivatives Importance for precise and accurate measurement
of peaks including partially resolved, broad and
asymmetric peaks- does it do its sums properly? Is
the software fully and correctly loaded? Does
manipulation and storage corrupt the raw data?
Can be determined and verified using software
packages or using reference materials.
Where instruments are designed to accept checking
using datasets, these should be used.
The extent of checks may need to satisfy regulatory
requirements.
Absorption Important for accuracy and precision of results,
checks that the optical faces of the cell are parallel.
Can be determined by filling the cell with water,
measuring at required wavelength(s), using air as
the reference and repeating the process after
rotating the cell through 180°.
Contd…
12
3. FT-IR SPECTROMETER
DESIGN QUALIFICATION:
• Supplier must provide documented evidence that the product has been designed, developed and manufactured in a quality environment e.g. ISO 9001:2000
certification.
• Supplier must provide phone and on-site support in case of defects.
• Supplier must provide information through the internet on availability of new firmware upgrades.
The below is a sample of technical attributes that may be considered in design qualification of an FT-IR spectrometer and may be adapted if necessary.
Attribute Specifications
Detector range The optical bench shall include a DTGS detector with a frequency range of
7400 to 350 cm-1
.
It shall include a compressed air interferometer -
The instrument shall come with an air-cooled standard infrared source -
Special resolution The instrument shall have a spectral resolution not exceeding 1.0 cm-1. 13
Attribute Specifications
Wave number accuracy Wave number accuracy shall be better than ±0.01 cm-1.
The interferometer shall have at least four basic velocity levels; software shall
permit the selection of a greater number of velocities between the basic levels.
-
The mirror’s greatest velocity shall allow a speed of at least five sweeps per
second.
-
The laser and infrared beams must be coaxial to enable rapid, easy alignment
of the system, depending on the samples or accessories.
-
The optical bench shall have a main experimentation module with a device to
allow purging with nitrogen.
-
At purchase, the main experimentation module shall be designed so that it can
receive 13 and 5 mm potassium bromide pellets.
-
Contd…
INSTALLATION QUALIFICATION:
• Equipment is compared as received with purchase order, including software, accessories, spare parts and consumables.
• Documentation checked for completeness of operating, maintenance instructions, standard operating procedures for testing and safety, validation certificates and
health and safety instruments.
• Equipment is checked for any damage.
• The supplier’s instruction for installation is read.
• The supplier’s safety instructions are read, if there are any.
• Hardware (computer, equipment, fittings and tubings for fluid connections, power cables, data flow and instrument control cables) is installed following the
manufacturer’s recommendation.
• The instruments are switched on and ensured that all modules power up and an electronic self test is performed. Any deviations are recorded.
• Software is installed on computer following the manufacturer’s recommendation.
• Correct software installation is verified.
• A backup copy of software is made.
• Peripherals e.g. printers and equipment modules are configured.
14
Contd…
• A list with a description of all hardware are identified and made, including drawings where appropriate.
• A list with a description of all software installed on the computer is made.
• Equipment manuals and SOPs are listed.
• An installation report is prepared.
OPERATIONAL QUALIFICATION:
Test procedure Acceptance limits Test frequency Remarks
Measurement of polystyrene
spectrum at 1144, 1680, 2167 and
2307 nm. The results are
compared with reference values.
±2 nm Daily Standard should be traceable to
national standard.
Wavelength accuracy
Control of 0 and 100%
Test procedure Acceptance limits Test frequency
Control at 0 and 100%. 1% transmission (measured at 4000 cm-1) Daily and after changing measurement
parameters
Wavelength resolution
Test procedure Acceptance limits Test frequency
Resolution of polystyrene at 2870 / 2851 and at
1589 / 1883 nm.
• T (band 2870 cm-1 – band 2851 cm-1) 18
• T (band 1589 cm-1 – band 1583 cm-1) 12
Daily
15
Test procedure Tansmission minima (cm-1) Acceptable tolerance (cm-1)
Recording the spectrum of a polystyrene film which
has transmission minima (absorption maxima) at
the wave numbers given in the next column.
3060.0 ±1.0
2849.5 ±1.0
1942.9 ±1.0
1601.2 ±1.0
1583.0 ±1.0
1154.5 ±1.0
1028.3 ±1.0
Wave number scale
Test procedure Acceptance limits
The minimum energy ratio value for at least one of the following is measured
and it is compared to the vendor’s specifications:
-Energy at 3990 cm-1/ energy at 2000 cm-1
-Energy at 3400 cm-1/ energy at 1300 cm-1
Energy ratio test specifications vary for each spectrometer configuration and
must be referred to the manufacturer’s specifications.
Detector Energy ratio
16
Test procedure Peak-to-peak noise betweeen RMS (root mean square)
noise between
Limits (%T)
The maximum noise level is recorded
for each of the regions given in the
next two columns.
• 4050 cm-1 and 3950 cm-1
• 2050 cm-1 and 1950 cm-1
• 1050 cm-1 and 950 cm-1
• 550 cm-1 and 450 cm-1
• 4050 cm-1 and 3950 cm-1
• 2050 cm-1 and 1950 cm-1
• 1050 cm-1 and 950 cm-1
• 550 cm-1 and 450 cm-1
Noise level test specifications vary for each
spectrometer configuration and must be
referred to the manufacturer’s specifications.
Signal/ noise ratio
Test procedure Limits (%T)
When using a polystyrene film of approximately 35 µm in thickness as standard
at the wavelength of 2925 cm-1 and 700 cm-1, almost complete absorption of the
irradiated energy can be observed. With this test, the remaining transmission is
measured. As the maximum absorption can be observed at 700cm-1 negative
values may be observed. The objective of the test is to evaluate if, despite the
fact that there is almost complete absorption, energy is still detectable.
Non-valid results are an indication of a non-linear behavior of the detector and
the electronic system .
Results vary for each spectrometer configuration and must be referred to the
manufacturer’s specifications.
Zero test
Test procedure Wavenumber (cm-1) Upper limit (A)
The automated function of the instrument is used to perform this test (If
automation is not available then a background spectrum is recorded).
This test checks the presence of peaks that signal a contamination problem.
3100.0-2800.0 0.1
1800.0-1600.0 0.1
1400.0-1100.0 0.2
Contamination test
17
PERFORMANCE QUALIFICATION:
Description Wave number (cm-1) Standard Value (FTIR-8400S) Remarks
Power spectrum gives the plot of
portion of signal’s power (energy per
unit time) falling within the given
frequency bins.
This test estimates the intensity of
power spectrum at a specified wave
number.
4600 10% or more of max When the measured intensity is equal to
or larger than the criterion value, the
test is passed.
4000 25% or more of max
3000 50% or more of max
Maximum value 50.0
700 10% or more of max
500 2% or more of max
403 0.5% or more of max
351 0.01% or more of max
Power spectrum
Wave number accuracy
The wave number scale is usually calibrated by the use of several characteristics wave numbers of a polystyrene film:
• 3060.0 (±/-1.5) cm-1
• 2849.5 (±/-1.5) cm-1
• 1942.9 (±/-1.5) cm-1
• 1601.2 (±/-1.0) cm-1
• 1583.0 (±/-1.0) cm-1
• 1154.5 (±/-1.0) cm-1
• 1028.3 (±/-1.0) cm-1
The software then judges whether the values are within the allowable range and the program labels the results ‘PASS’ if all the peak numbers are within the range.
18
Wave number reproducibility:
• This program specifies three points to measure the peak wave numbers.
• Then it obtains the actual peak wave numbers at each point by measuring the polystyrene film twice.
• It should satisfy 5 cm-1 around 3000 cm-1 of polystyrene absorption wave number, 1 cm-1 around 1000 cm-1.
• The software determines whether the differences between each of two measurements are within the allowable range and it labels the result ‘PASS’ if they are within the range.
Transmittance reproducibility:
• This program specifies peak wave number at three points and the transmittance at each point is measured twice.
• The transmittance reproducibility should satisfy 0.5% T when the several points of polystyrene absorption from 3000 cm-1 to 1000 cm-1 are measured twice.
• Then it is determined whether the differences between the two data are within the range.
• The FTIR abnormalities or large changes over short term and long term is assessed by these tests.
• The two parameters checked by this program are:
a) Energy spectrum test- Power spectra obtained in the inspection are compared with reference data and the spectra are checked for changes over long periods. 100% T line
spectra are calculated for power spectra and are measured continuously in inspection and the spectra are checked for changes over short periods.
b) Polystyrene test- Evaluation is performed using differences between spectra obtained for polystyrene film in inspection and the stored reference data. When the differences
are within the standard, it passes the test.
19
5. GAS CHROMATOGRAPHY (GC)
DESIGN QUALIFICATION:
• Supplier must provide documented evidence that the product has been designed, developed and manufactured in a quality environment e.g. ISO 9001:2000 certification.
• Supplier must provide phone and on-site support in case of defects.
• Supplier must provide information through the internet on availability of new firmware upgrades.
INSTALLATION QUALIFICATION:
• Equipment is compared as received with purchase order, including software, accessories, spare parts and consumables.
• Documentation checked for completeness of operating, maintenance instructions, standard operating procedures for testing and safety, validation certificates and health and
safety instruments.
• Equipment is checked for any damage.
• The supplier’s instruction for installation is read.
• The supplier’s safety instructions are read, if there are any.
• Hardware (computer, equipment, fittings and tubings for fluid connections, power cables, data flow and instrument control cables) is installed following the manufacturer’s
recommendation.
• The instruments are switched on and ensured that all modules power up and an electronic self test is performed. Any deviations are recorded.
• Software is installed on computer following the manufacturer’s recommendation.
• Correct software installation is verified.
• A backup copy of software is made.
• Peripherals e.g. printers and equipment modules are configured.
20
OPERATIONAL QUALIFICATION:
Test procedure Acceptance limits Test frequency
• Five injections of a standard.
• Calculation of relative standard deviation.
<1% RSD (Relative Standard Deviation) Yearly
Precision of peak retention times
Precision of peak areas
Test procedure Acceptance limits Test frequency
• Five injections of a standard.
• Calculation of relative standard deviation.
<2% RSD Yearly
Accuracy of the temperature of the column oven
Test procedure Acceptance limits Test frequency Remarks
Temperature in column oven is
measured and compared with setpoint.
±1°C Yearly The temperature measurement device
should be calibrated and traceable to a
national standard.
Precision of heated zone temperature
Test procedure Acceptance limits Test frequency Remarks
Actual temperature is measured and
compared with setpoints at 50,70 and
90°C
2°C Yearly The temperature measurement device
should be calibrated and traceable to a
national standard.
21
Precision of injection
Test procedure Acceptance limits Test frequency
• Five injections of a standard.
• Calculation of relative standard deviation.
<2% RSD Yearly
Carryover of injection
Test procedure Acceptance limits Test frequency
Blank solvent is injected after standard. Peak ratio is
measured between blank and standard injection.
<5% Yearly
PERFORMANCE QUALIFICATION:
Detector linearity
22
Procedure Acceptance criteria
• Blank is injected followed by ‘detector linearity solutions’ and the peak
responses are recorded.
• A standard plot is drawn between the concentrations vs the peak responses.
The plot should be linear and regression coefficient (R2) should not be less than
0.99.
Flow rate accuracy:
• The digital flow meter is connected to the detector outlet port.
• The carrier gas flow is set and wait till it reaches the set flow.
• The observed flow in replicate is noted.
• The procedure is repeated for other carrier gases such as hydrogen and air.
• The flow rate of carrier gas should be ±10% of set flow
System precision:
• 20ml of methanol, ethanol and acetone is transferred into 100ml volumetric flask and made up with ethyl acetate.
• Blank is injected followed by standard preparation in six replicates.
• The areas and retention times are noted down.
• The %RSD (Relative standard deviation) of retention time should not be more than 1.0% and peak area NMT (not more than) 5.0%
Detector noise and Drift test:
• After GC is ready, the system is run up to 15 minutes through single run.
• When the run is completed noise and drift is calculated through software.
• The acceptance values are:
a) Noise not more than 100µV.
b) Drift not more than 2500µV/hr.
23
REFERENCES:
 Huber L. Validation and Qualification in Analytical Laboratories. Informa Healthcare USA, New York. 2007 (2).11-19.
 Chan C. C, Lee Y. C, Lam H and Zhang X. M. Analytical Method Validation and Instrument Performance Verification. John Wiley & Sons, Inc., Hoboken, New
Jersey. 2004.137-148.
 Krull I. S and Swartz M. E. Operational and Performance Qualification. LCGC North America, April, 2002, 20 (4), 356-362.
 Gowrisankar D, Abbulu K, Souri O. B and Sujana K. Validation and Calibration of Analytical Instruments. Journal of Biomedical Sciences and Research, 2010, 2
(2), 89-99.
24
End of presentation
Thank you !
25

Contenu connexe

Tendances

Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.begalavenu
 
Analytical methods validation as per ich & usp
Analytical methods validation as per ich & uspAnalytical methods validation as per ich & usp
Analytical methods validation as per ich & uspGANESH NIGADE
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaningKUNDLAJAYALAKSHMI
 
Impurities in residual solvents
Impurities in residual solventsImpurities in residual solvents
Impurities in residual solventsMehulJain143
 
Rationale for the reporting control of degradation products
Rationale for the reporting control of degradation products Rationale for the reporting control of degradation products
Rationale for the reporting control of degradation products ManiKandan1405
 
Analytical method validation as per ich and usp
Analytical method validation as per ich and usp Analytical method validation as per ich and usp
Analytical method validation as per ich and usp shreyas B R
 
Stability testing protocols
Stability testing protocolsStability testing protocols
Stability testing protocolsMehulJain143
 
Qualification of GC & FTIR
Qualification of GC & FTIRQualification of GC & FTIR
Qualification of GC & FTIRMehulJain143
 
Concept &amp; evolution of qa &amp; qc
Concept &amp; evolution of qa &amp; qcConcept &amp; evolution of qa &amp; qc
Concept &amp; evolution of qa &amp; qcChowdaryPavani
 
Documentation In Pharmaceutical Industry.pptx
Documentation In Pharmaceutical Industry.pptxDocumentation In Pharmaceutical Industry.pptx
Documentation In Pharmaceutical Industry.pptxsaurabh11102000
 

Tendances (20)

CLEANING METHOD DEVELOPMENT
CLEANING METHOD DEVELOPMENT CLEANING METHOD DEVELOPMENT
CLEANING METHOD DEVELOPMENT
 
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
 
QUALIFICATION OF FT-IR
QUALIFICATION OF FT-IRQUALIFICATION OF FT-IR
QUALIFICATION OF FT-IR
 
Requalification
RequalificationRequalification
Requalification
 
Analytical methods validation as per ich & usp
Analytical methods validation as per ich & uspAnalytical methods validation as per ich & usp
Analytical methods validation as per ich & usp
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaning
 
Qualification of Gas Chromatography
Qualification of Gas Chromatography Qualification of Gas Chromatography
Qualification of Gas Chromatography
 
Impurities in residual solvents
Impurities in residual solventsImpurities in residual solvents
Impurities in residual solvents
 
Rationale for the reporting control of degradation products
Rationale for the reporting control of degradation products Rationale for the reporting control of degradation products
Rationale for the reporting control of degradation products
 
Validation qualification
Validation qualificationValidation qualification
Validation qualification
 
Analytical method validation as per ich and usp
Analytical method validation as per ich and usp Analytical method validation as per ich and usp
Analytical method validation as per ich and usp
 
Stability testing protocols
Stability testing protocolsStability testing protocols
Stability testing protocols
 
Qualification of Weights and Measures
Qualification of Weights and MeasuresQualification of Weights and Measures
Qualification of Weights and Measures
 
Qualification of GC & FTIR
Qualification of GC & FTIRQualification of GC & FTIR
Qualification of GC & FTIR
 
Computer system validations
Computer system validationsComputer system validations
Computer system validations
 
Computerized system validation
Computerized system validationComputerized system validation
Computerized system validation
 
Concept &amp; evolution of qa &amp; qc
Concept &amp; evolution of qa &amp; qcConcept &amp; evolution of qa &amp; qc
Concept &amp; evolution of qa &amp; qc
 
QUALIFICATION OF HPTLC,HPTLC
QUALIFICATION  OF HPTLC,HPTLCQUALIFICATION  OF HPTLC,HPTLC
QUALIFICATION OF HPTLC,HPTLC
 
Documentation In Pharmaceutical Industry.pptx
Documentation In Pharmaceutical Industry.pptxDocumentation In Pharmaceutical Industry.pptx
Documentation In Pharmaceutical Industry.pptx
 
Qualification of HPLC
Qualification of HPLCQualification of HPLC
Qualification of HPLC
 

Similaire à Qualification of analytical instruments

Module_4_GCLP_LABORATORY_EQUIPMENT.ppt
Module_4_GCLP_LABORATORY_EQUIPMENT.pptModule_4_GCLP_LABORATORY_EQUIPMENT.ppt
Module_4_GCLP_LABORATORY_EQUIPMENT.pptssuserbf5fe71
 
Laboratory Equipment (from CLSI)
Laboratory Equipment (from CLSI)Laboratory Equipment (from CLSI)
Laboratory Equipment (from CLSI)Mohammed Alzhrani
 
Krishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instrumentsKrishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instrumentseslamgouda
 
ICH guidelines for validation Of Equipments by Nikita Sahu[1].pptx
ICH guidelines for validation Of Equipments by Nikita Sahu[1].pptxICH guidelines for validation Of Equipments by Nikita Sahu[1].pptx
ICH guidelines for validation Of Equipments by Nikita Sahu[1].pptxNikitaSahu39
 
Validation of equipments
Validation of equipmentsValidation of equipments
Validation of equipmentssuhasini
 
validation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptxvalidation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptxDevaPundkar
 
validation-1.pptx
validation-1.pptxvalidation-1.pptx
validation-1.pptxSridharA50
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments PUNEET NIRMAL
 
Different types of Lab control record
Different types of Lab control recordDifferent types of Lab control record
Different types of Lab control recordVinit Gohel
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemAkshay Trivedi , Maliba Pharmacy College
 
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptxICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptxABG
 
Zablocki, Shawn, Olsson Associates, Mechanical Integrity Programs for MRP and...
Zablocki, Shawn, Olsson Associates, Mechanical Integrity Programs for MRP and...Zablocki, Shawn, Olsson Associates, Mechanical Integrity Programs for MRP and...
Zablocki, Shawn, Olsson Associates, Mechanical Integrity Programs for MRP and...Kevin Perry
 
Equipment qualification & validation of autoclave, dry heat sterilization and...
Equipment qualification & validation of autoclave, dry heat sterilization and...Equipment qualification & validation of autoclave, dry heat sterilization and...
Equipment qualification & validation of autoclave, dry heat sterilization and...Malla Reddy College of Pharmacy
 
Support and utilities validation
Support and utilities validationSupport and utilities validation
Support and utilities validationVaishali Dudhabale
 

Similaire à Qualification of analytical instruments (20)

Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
ISO 17025 – technical requirements
ISO 17025 – technical requirementsISO 17025 – technical requirements
ISO 17025 – technical requirements
 
Module_4_GCLP_LABORATORY_EQUIPMENT.ppt
Module_4_GCLP_LABORATORY_EQUIPMENT.pptModule_4_GCLP_LABORATORY_EQUIPMENT.ppt
Module_4_GCLP_LABORATORY_EQUIPMENT.ppt
 
Laboratory Equipment (from CLSI)
Laboratory Equipment (from CLSI)Laboratory Equipment (from CLSI)
Laboratory Equipment (from CLSI)
 
Krishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instrumentsKrishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instruments
 
ICH guidelines for validation Of Equipments by Nikita Sahu[1].pptx
ICH guidelines for validation Of Equipments by Nikita Sahu[1].pptxICH guidelines for validation Of Equipments by Nikita Sahu[1].pptx
ICH guidelines for validation Of Equipments by Nikita Sahu[1].pptx
 
Validation of equipments
Validation of equipmentsValidation of equipments
Validation of equipments
 
validation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptxvalidation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptx
 
validation-1.pptx
validation-1.pptxvalidation-1.pptx
validation-1.pptx
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Different types of Lab control record
Different types of Lab control recordDifferent types of Lab control record
Different types of Lab control record
 
Dr. ajmal nasir
Dr. ajmal nasirDr. ajmal nasir
Dr. ajmal nasir
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility System
 
2. Ppt on GLP
2. Ppt on GLP2. Ppt on GLP
2. Ppt on GLP
 
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptxICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
 
Validation of gc instrument
Validation of gc instrumentValidation of gc instrument
Validation of gc instrument
 
Zablocki, Shawn, Olsson Associates, Mechanical Integrity Programs for MRP and...
Zablocki, Shawn, Olsson Associates, Mechanical Integrity Programs for MRP and...Zablocki, Shawn, Olsson Associates, Mechanical Integrity Programs for MRP and...
Zablocki, Shawn, Olsson Associates, Mechanical Integrity Programs for MRP and...
 
Equipment qualification & validation of autoclave, dry heat sterilization and...
Equipment qualification & validation of autoclave, dry heat sterilization and...Equipment qualification & validation of autoclave, dry heat sterilization and...
Equipment qualification & validation of autoclave, dry heat sterilization and...
 
Support and utilities validation
Support and utilities validationSupport and utilities validation
Support and utilities validation
 
Validation
ValidationValidation
Validation
 

Plus de Sanathoiba Singha

Problems with analysis of biological matrices
Problems with analysis of biological matricesProblems with analysis of biological matrices
Problems with analysis of biological matricesSanathoiba Singha
 
Notes* for the subject 'Advanced Pharmaceutical Analysis'
Notes* for the subject 'Advanced Pharmaceutical Analysis'Notes* for the subject 'Advanced Pharmaceutical Analysis'
Notes* for the subject 'Advanced Pharmaceutical Analysis'Sanathoiba Singha
 
Notes* for the subject 'Food Analysis'
Notes* for the subject 'Food Analysis'Notes* for the subject 'Food Analysis'
Notes* for the subject 'Food Analysis'Sanathoiba Singha
 
Notes for the subject 'Pharmaceutical Validation'
Notes for the subject 'Pharmaceutical Validation' Notes for the subject 'Pharmaceutical Validation'
Notes for the subject 'Pharmaceutical Validation' Sanathoiba Singha
 
Analysis of drugs in biological matrix
Analysis of drugs in biological matrixAnalysis of drugs in biological matrix
Analysis of drugs in biological matrixSanathoiba Singha
 
Quality Control tests for pharmaceutical containers, closures and secondary p...
Quality Control tests for pharmaceutical containers, closures and secondary p...Quality Control tests for pharmaceutical containers, closures and secondary p...
Quality Control tests for pharmaceutical containers, closures and secondary p...Sanathoiba Singha
 
A collection of notes for the subject 'Modern Pharmaceutical Analysis (MPA)'
A collection of notes for the subject 'Modern Pharmaceutical Analysis (MPA)' A collection of notes for the subject 'Modern Pharmaceutical Analysis (MPA)'
A collection of notes for the subject 'Modern Pharmaceutical Analysis (MPA)' Sanathoiba Singha
 

Plus de Sanathoiba Singha (15)

Problems with analysis of biological matrices
Problems with analysis of biological matricesProblems with analysis of biological matrices
Problems with analysis of biological matrices
 
Notes* for the subject 'Advanced Pharmaceutical Analysis'
Notes* for the subject 'Advanced Pharmaceutical Analysis'Notes* for the subject 'Advanced Pharmaceutical Analysis'
Notes* for the subject 'Advanced Pharmaceutical Analysis'
 
Notes* for the subject 'Food Analysis'
Notes* for the subject 'Food Analysis'Notes* for the subject 'Food Analysis'
Notes* for the subject 'Food Analysis'
 
Notes for the subject 'Pharmaceutical Validation'
Notes for the subject 'Pharmaceutical Validation' Notes for the subject 'Pharmaceutical Validation'
Notes for the subject 'Pharmaceutical Validation'
 
Analysis of drugs in biological matrix
Analysis of drugs in biological matrixAnalysis of drugs in biological matrix
Analysis of drugs in biological matrix
 
Quality Control tests for pharmaceutical containers, closures and secondary p...
Quality Control tests for pharmaceutical containers, closures and secondary p...Quality Control tests for pharmaceutical containers, closures and secondary p...
Quality Control tests for pharmaceutical containers, closures and secondary p...
 
Sebumeter
SebumeterSebumeter
Sebumeter
 
A collection of notes for the subject 'Modern Pharmaceutical Analysis (MPA)'
A collection of notes for the subject 'Modern Pharmaceutical Analysis (MPA)' A collection of notes for the subject 'Modern Pharmaceutical Analysis (MPA)'
A collection of notes for the subject 'Modern Pharmaceutical Analysis (MPA)'
 
Bioluminescence Assay
Bioluminescence AssayBioluminescence Assay
Bioluminescence Assay
 
Pesticide analysis
Pesticide analysisPesticide analysis
Pesticide analysis
 
Vaccines
VaccinesVaccines
Vaccines
 
Ketone bodies
Ketone bodiesKetone bodies
Ketone bodies
 
Theory of IR spectroscopy
Theory of IR spectroscopy Theory of IR spectroscopy
Theory of IR spectroscopy
 
pKa and log p determination
pKa and log p determinationpKa and log p determination
pKa and log p determination
 
Protein analysis
Protein analysisProtein analysis
Protein analysis
 

Dernier

Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...parulsinha
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...khalifaescort01
 
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...chennailover
 
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...mahaiklolahd
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls ServiceGENUINE ESCORT AGENCY
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Dipal Arora
 
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...GENUINE ESCORT AGENCY
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...chandars293
 
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...khalifaescort01
 
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableTrichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Sheetaleventcompany
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappInaaya Sharma
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426jennyeacort
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Sheetaleventcompany
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...parulsinha
 

Dernier (20)

Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
 
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
 
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
 
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableTrichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
 

Qualification of analytical instruments

  • 1. Presented by: Facilitated to: Mr. L. Sanathoiba Singha M. Pharm 1st Semester Department of Pharmaceutical Analysis. Dr. C. Sreedhar Head of Department Department of Pharmaceutical Analysis. Karnataka College of Pharmacy Bengaluru-64, Karnataka. 1
  • 2. 1. ELECTRONIC BALANCE DESIGN QUALIFICATION: The supplier has ISO 9001:2000 certification. INSTALLATION QUALIFICATION: • Checking of all requirements set during the selection of instrument such as electricity, humidity, temperature, etc. • Allowing sufficient shelf space for the equipment, SOPs, operating manuals, etc. • Comparing equipment, as received, with purchase order, including accessories, spare parts, etc. • Checking documentation for completeness like operating manuals, maintenance instructions, health and safety instructions, etc. • Checking equipment for any damage. • Reading the supplier’s safety instructions, if there are any. • Installing hardware following the manufacturer’s recommendation. • Switching on the electronic balance and ensuring that all the modules power up. • Preparing an installation report 2
  • 3. Test procedure Acceptance limits Test frequency Remarks • Measurement of reference weight by using 10mg, 50mg, 100mg, 500mg, 1g, 5g, 10g and 20g. • Comparing the actual results with reference weights. 0.1% • Daily or when used, whatever is longer with internal reference weights. • Yearly with traceable external weights through instrument vendor. External standard should be traceable to national standard. OPERATIONAL QUALIFICATION: PERFORMANCE QUALIFICATION: • Defining weights and weight classes to be used. • Defining acceptance limits of results. • Defining test intervals. • Defining corrective actions on what to do if the electronic balance does not meet the criteria, in other words if the results are out of specification. Operational qualification for electronic balance. 3
  • 4. 2. pH METER DESIGN QUALIFICATION: • Selection of a manufacturer that can satisfy the needs of the laboratory. • Should have ISO 9001 certification. INSTALLATION QUALIFICATION: • Checking of all requirements set during the selection of instrument such as electricity, humidity, temperature, etc. • Allowing sufficient shelf space for the equipment, SOPs, operating manuals, etc. • Comparing equipment, as received, with purchase order, including accessories, spare parts, etc. • Checking documentation for completeness like operating manuals, maintenance instructions, health and safety instructions, etc. • Checking equipment for any damage. • Reading the supplier’s safety instructions, if there are any. • Installing hardware following the manufacturer’s recommendation. • Switching on the pH meter and ensuring that all the modules power up. • Preparing an installation report. 4
  • 5. PERFORMANCE QUALIFICATION: A log book is prepared for operators and service technicians to record all equipment related activities in chronological order. Parameter to be checked Typical frequency Typical tolerance limit Slope Each day of use or before each series of measurements 95.0 to 105.0% Offset (pH asymmetry) +/-30mV (offset of manufacturer +/-0.5 pH units at defined temperature) Reference buffer solution +/-0.05 pH units Test procedure Acceptance limits Test frequency Remarks • Calibration with buffer solution at pH 7 and pH 4. • Verification with buffer solution at pH 5 or pH 6. <0.005 pH units Before each use • Buffer solutions are commercially available in either tablet or solution form. • Some pH meters also have a temperature sensor, which should be calibrated every six months. • The electrodes should be washed between each measurement and the buffers should be visually checked for cleanliness and absence of microbial growth before each use. OPERATIONAL QUALIFICATION: 5
  • 6. 3. UV-VISIBLE SPECTROPHOTOMETER DESIGN QUALIFICATIONS: General considerations Feature Consideration Instrument set-up and control • Pc based or integrated system. • Software control of operating conditions and parameters. • Data acquisition, processing and presentation needs. • In-built diagnostic facilities. • Self-testing diagnostics. • Detector options. • Source options. Sample introduction and throughput • Sample throughput, presentation and introduction needs. • Sample thermostating requirements. • Sample volume requirements. Materials of construction Resistance to corrosion, contaminated by solvents and sample. 6
  • 7. Contd… Feature Consideration Installation requirements • Size and weight in shipped form. • Access restrictions to permanent site. Operational requirements • Limitations on, requirements for and expected consumption of services, utilities, and consumables (e.g. lamps). • Ventilation requirements. • Controlling software embedded or separate package. Environmental conditions • Environmental conditions within which, or range over which, the instrument is required to work within specification. • Recyclability of instrument. Maintenance and support • Ease of user maintenance and cleaning. • Cost, longevity and availability of spares and parts. • Cost and availability of service contacts and technical support. • Suggested intervals between and procedures for maintenance and calibration of the instrument. • The period for which support (qualification, maintenance, parts, etc.) for the instrument can be guaranteed. Training requirements The level of skill required to operate the instrument and details of any training necessary and courses provided by the supplier. Accessories Desired adaptation of the system Documentation • Clarity and ease of use of documentation (e.g. operating manuals, qualification protocols, model SOPs). • Manuals available as separate hardcopies or computer files (e.g. on CD ROM), or held as digitally embedded copies within the instrument. • Unique document identification by version number and date of issue. Health and safety • Health and safety and environmental issues and /or requirements. • Significant generation of ozone. 7
  • 8. Design Qualification of UV-Visible Spectrophotometer Feature Consideration Photometric drift Stability of the measurement over time, making comparisons meaningful. Stray light Affects accuracy and linearity Data collection • Expression of data on % absorption or % transmission scales. • Single or repeat scan facility. • Single or multiple wavelength monitoring. • Full spectra data (e.g. diode array). Data manipulation • Spectra subtraction. • Spectral derivatives. • Audit trail facility. • Is raw data uncorruptable ? • Back-up and restore capabilities. Data storage • No storage. • Embedded memory. • File transmission to external device. • Stored output via optical or magnetic media. • System security. Data output • Inbuilt hardcopy output-data only, spectrum only, data+ spectrum. • PC based data processing. • Custom report generation. 8
  • 9. INSTALLATION QUALIFICATION: • The instrument has been delivered as ordered, e.g. according to the DQ or purchase order. • The instrument has been checked and verified as undamaged. • The appropriate documentation has been supplied and it is of correct issue and uniquely identified by part number, version number and date. • Details of recommended service and calibration intervals (carried out by the supplier) have been provided. • Intervals, methods and instructions for user-maintenance and calibration have been provided along with contact points for service and spare parts. • The correct hardware, firmware and software have been supplied and it is of correct tissue and uniquely identified by part number. • Information has been provided on consumables required during the normal operation of the instrument. • The selected environment for the instrument system is suitable, with adequate room for unpacking, installation, operation and servicing, and have appropriate services and utilities have been provided. • Health and safety and environmental information relating to the operation of the instrument have been provided and the proposed working environment is consistent with these requirements. • The response of the instrument to the initial application of power is as expected and checked for any deviations. OPERATIONAL QUALIFICATION: Test procedure Acceptance limits Test frequency Remarks Wavelength maxima of holmium oxide solution is measured at 241, 287, 361 and 536 nm. • ±1 nm in UV range • ±3 nm in visible range Six months Standard is traceable to national standard. Wavelength accuracy- holmium oxide solution Wavelength accuracy- holmium oxide filter Test procedure Acceptance limits Test frequency Remarks Wavelength maxima of holmium oxide filter is measured at 361, 418 and 536 nm. • ±1 nm in UV range • ±3 nm in visible range Six months Filter is traceable to national standard. 9
  • 10. Test procedure Acceptance limits Test frequency Remarks wavelength A 1%/1cm Limits Absorption of potassium chromate solution is measured. • 235 (min) • 257 (max) • 313 (min) • 350 (max) • 124.5 • 144.0 • 48.6 • 106.6 • 122.9-126.2 • 142.4-145.7 • 47.0-50.3 • 104.9-108.2 Six months Standard is traceable to national standard. Absorption intensity Test procedure Acceptance limits Test frequency Remarks Absorption of 1.2% potassium chloride solution at 200 nm is measured against water. Absorption at 200 nm >2.0 AU Six months Alternative standards: • Sodium nitrite • Sodium iodide. Stray light Test procedure Acceptance limits Test frequency Remarks Transmission of cuvette (water) against air. a) Quartz cuvettes: • 85% at 220 nm • 88% at 240 nm b) Glass cuvettes: • 85% at 356 nm • 88% at 650 nm Six months Standard is traceable to national standard. Control of cuvette 10
  • 11. Test procedure Acceptance limits Test frequency Spectrum of toluene is measured from 260 to 275 nm. Ratio of peak height to valley is calculated. Absorbance ratio peak to valley at 266/269 nm > 1.5 Six months Wavelength resolution PERFORMANCE QUALIFICATION: Parameter Reason Procedure Wavelength calibration Critical to accuracy of results. Can be determined using only two calibration wavelengths, preferably bracketing the analytical wavelength. Photometric accuracy Critical to accuracy of results Can be determined for particular absorbance at particular wavelengths using calibrated filters or cuvettes filled with standard solutions. Linearity of photometric response Critical to accuracy of results Can be determined by checking the accuracy of a number of normal absorbances across the desired absorbance range. Extrapolation beyond highest standard is inadvisable. Signal to noise ratio Important for sensitivity and limit detection Can be determined from the response of a detector to a dilute standard solution and/or a blank. Stray light Important measure of ‘health’ of whole system Record uncorrected spectrum in single beam mode for double beam instruments or with baseline correction switched off for single beam instruments. 11
  • 12. Parameter Reason Procedure Software accuracy and precision, derivatives Importance for precise and accurate measurement of peaks including partially resolved, broad and asymmetric peaks- does it do its sums properly? Is the software fully and correctly loaded? Does manipulation and storage corrupt the raw data? Can be determined and verified using software packages or using reference materials. Where instruments are designed to accept checking using datasets, these should be used. The extent of checks may need to satisfy regulatory requirements. Absorption Important for accuracy and precision of results, checks that the optical faces of the cell are parallel. Can be determined by filling the cell with water, measuring at required wavelength(s), using air as the reference and repeating the process after rotating the cell through 180°. Contd… 12
  • 13. 3. FT-IR SPECTROMETER DESIGN QUALIFICATION: • Supplier must provide documented evidence that the product has been designed, developed and manufactured in a quality environment e.g. ISO 9001:2000 certification. • Supplier must provide phone and on-site support in case of defects. • Supplier must provide information through the internet on availability of new firmware upgrades. The below is a sample of technical attributes that may be considered in design qualification of an FT-IR spectrometer and may be adapted if necessary. Attribute Specifications Detector range The optical bench shall include a DTGS detector with a frequency range of 7400 to 350 cm-1 . It shall include a compressed air interferometer - The instrument shall come with an air-cooled standard infrared source - Special resolution The instrument shall have a spectral resolution not exceeding 1.0 cm-1. 13
  • 14. Attribute Specifications Wave number accuracy Wave number accuracy shall be better than ±0.01 cm-1. The interferometer shall have at least four basic velocity levels; software shall permit the selection of a greater number of velocities between the basic levels. - The mirror’s greatest velocity shall allow a speed of at least five sweeps per second. - The laser and infrared beams must be coaxial to enable rapid, easy alignment of the system, depending on the samples or accessories. - The optical bench shall have a main experimentation module with a device to allow purging with nitrogen. - At purchase, the main experimentation module shall be designed so that it can receive 13 and 5 mm potassium bromide pellets. - Contd… INSTALLATION QUALIFICATION: • Equipment is compared as received with purchase order, including software, accessories, spare parts and consumables. • Documentation checked for completeness of operating, maintenance instructions, standard operating procedures for testing and safety, validation certificates and health and safety instruments. • Equipment is checked for any damage. • The supplier’s instruction for installation is read. • The supplier’s safety instructions are read, if there are any. • Hardware (computer, equipment, fittings and tubings for fluid connections, power cables, data flow and instrument control cables) is installed following the manufacturer’s recommendation. • The instruments are switched on and ensured that all modules power up and an electronic self test is performed. Any deviations are recorded. • Software is installed on computer following the manufacturer’s recommendation. • Correct software installation is verified. • A backup copy of software is made. • Peripherals e.g. printers and equipment modules are configured. 14
  • 15. Contd… • A list with a description of all hardware are identified and made, including drawings where appropriate. • A list with a description of all software installed on the computer is made. • Equipment manuals and SOPs are listed. • An installation report is prepared. OPERATIONAL QUALIFICATION: Test procedure Acceptance limits Test frequency Remarks Measurement of polystyrene spectrum at 1144, 1680, 2167 and 2307 nm. The results are compared with reference values. ±2 nm Daily Standard should be traceable to national standard. Wavelength accuracy Control of 0 and 100% Test procedure Acceptance limits Test frequency Control at 0 and 100%. 1% transmission (measured at 4000 cm-1) Daily and after changing measurement parameters Wavelength resolution Test procedure Acceptance limits Test frequency Resolution of polystyrene at 2870 / 2851 and at 1589 / 1883 nm. • T (band 2870 cm-1 – band 2851 cm-1) 18 • T (band 1589 cm-1 – band 1583 cm-1) 12 Daily 15
  • 16. Test procedure Tansmission minima (cm-1) Acceptable tolerance (cm-1) Recording the spectrum of a polystyrene film which has transmission minima (absorption maxima) at the wave numbers given in the next column. 3060.0 ±1.0 2849.5 ±1.0 1942.9 ±1.0 1601.2 ±1.0 1583.0 ±1.0 1154.5 ±1.0 1028.3 ±1.0 Wave number scale Test procedure Acceptance limits The minimum energy ratio value for at least one of the following is measured and it is compared to the vendor’s specifications: -Energy at 3990 cm-1/ energy at 2000 cm-1 -Energy at 3400 cm-1/ energy at 1300 cm-1 Energy ratio test specifications vary for each spectrometer configuration and must be referred to the manufacturer’s specifications. Detector Energy ratio 16
  • 17. Test procedure Peak-to-peak noise betweeen RMS (root mean square) noise between Limits (%T) The maximum noise level is recorded for each of the regions given in the next two columns. • 4050 cm-1 and 3950 cm-1 • 2050 cm-1 and 1950 cm-1 • 1050 cm-1 and 950 cm-1 • 550 cm-1 and 450 cm-1 • 4050 cm-1 and 3950 cm-1 • 2050 cm-1 and 1950 cm-1 • 1050 cm-1 and 950 cm-1 • 550 cm-1 and 450 cm-1 Noise level test specifications vary for each spectrometer configuration and must be referred to the manufacturer’s specifications. Signal/ noise ratio Test procedure Limits (%T) When using a polystyrene film of approximately 35 µm in thickness as standard at the wavelength of 2925 cm-1 and 700 cm-1, almost complete absorption of the irradiated energy can be observed. With this test, the remaining transmission is measured. As the maximum absorption can be observed at 700cm-1 negative values may be observed. The objective of the test is to evaluate if, despite the fact that there is almost complete absorption, energy is still detectable. Non-valid results are an indication of a non-linear behavior of the detector and the electronic system . Results vary for each spectrometer configuration and must be referred to the manufacturer’s specifications. Zero test Test procedure Wavenumber (cm-1) Upper limit (A) The automated function of the instrument is used to perform this test (If automation is not available then a background spectrum is recorded). This test checks the presence of peaks that signal a contamination problem. 3100.0-2800.0 0.1 1800.0-1600.0 0.1 1400.0-1100.0 0.2 Contamination test 17
  • 18. PERFORMANCE QUALIFICATION: Description Wave number (cm-1) Standard Value (FTIR-8400S) Remarks Power spectrum gives the plot of portion of signal’s power (energy per unit time) falling within the given frequency bins. This test estimates the intensity of power spectrum at a specified wave number. 4600 10% or more of max When the measured intensity is equal to or larger than the criterion value, the test is passed. 4000 25% or more of max 3000 50% or more of max Maximum value 50.0 700 10% or more of max 500 2% or more of max 403 0.5% or more of max 351 0.01% or more of max Power spectrum Wave number accuracy The wave number scale is usually calibrated by the use of several characteristics wave numbers of a polystyrene film: • 3060.0 (±/-1.5) cm-1 • 2849.5 (±/-1.5) cm-1 • 1942.9 (±/-1.5) cm-1 • 1601.2 (±/-1.0) cm-1 • 1583.0 (±/-1.0) cm-1 • 1154.5 (±/-1.0) cm-1 • 1028.3 (±/-1.0) cm-1 The software then judges whether the values are within the allowable range and the program labels the results ‘PASS’ if all the peak numbers are within the range. 18
  • 19. Wave number reproducibility: • This program specifies three points to measure the peak wave numbers. • Then it obtains the actual peak wave numbers at each point by measuring the polystyrene film twice. • It should satisfy 5 cm-1 around 3000 cm-1 of polystyrene absorption wave number, 1 cm-1 around 1000 cm-1. • The software determines whether the differences between each of two measurements are within the allowable range and it labels the result ‘PASS’ if they are within the range. Transmittance reproducibility: • This program specifies peak wave number at three points and the transmittance at each point is measured twice. • The transmittance reproducibility should satisfy 0.5% T when the several points of polystyrene absorption from 3000 cm-1 to 1000 cm-1 are measured twice. • Then it is determined whether the differences between the two data are within the range. • The FTIR abnormalities or large changes over short term and long term is assessed by these tests. • The two parameters checked by this program are: a) Energy spectrum test- Power spectra obtained in the inspection are compared with reference data and the spectra are checked for changes over long periods. 100% T line spectra are calculated for power spectra and are measured continuously in inspection and the spectra are checked for changes over short periods. b) Polystyrene test- Evaluation is performed using differences between spectra obtained for polystyrene film in inspection and the stored reference data. When the differences are within the standard, it passes the test. 19
  • 20. 5. GAS CHROMATOGRAPHY (GC) DESIGN QUALIFICATION: • Supplier must provide documented evidence that the product has been designed, developed and manufactured in a quality environment e.g. ISO 9001:2000 certification. • Supplier must provide phone and on-site support in case of defects. • Supplier must provide information through the internet on availability of new firmware upgrades. INSTALLATION QUALIFICATION: • Equipment is compared as received with purchase order, including software, accessories, spare parts and consumables. • Documentation checked for completeness of operating, maintenance instructions, standard operating procedures for testing and safety, validation certificates and health and safety instruments. • Equipment is checked for any damage. • The supplier’s instruction for installation is read. • The supplier’s safety instructions are read, if there are any. • Hardware (computer, equipment, fittings and tubings for fluid connections, power cables, data flow and instrument control cables) is installed following the manufacturer’s recommendation. • The instruments are switched on and ensured that all modules power up and an electronic self test is performed. Any deviations are recorded. • Software is installed on computer following the manufacturer’s recommendation. • Correct software installation is verified. • A backup copy of software is made. • Peripherals e.g. printers and equipment modules are configured. 20
  • 21. OPERATIONAL QUALIFICATION: Test procedure Acceptance limits Test frequency • Five injections of a standard. • Calculation of relative standard deviation. <1% RSD (Relative Standard Deviation) Yearly Precision of peak retention times Precision of peak areas Test procedure Acceptance limits Test frequency • Five injections of a standard. • Calculation of relative standard deviation. <2% RSD Yearly Accuracy of the temperature of the column oven Test procedure Acceptance limits Test frequency Remarks Temperature in column oven is measured and compared with setpoint. ±1°C Yearly The temperature measurement device should be calibrated and traceable to a national standard. Precision of heated zone temperature Test procedure Acceptance limits Test frequency Remarks Actual temperature is measured and compared with setpoints at 50,70 and 90°C 2°C Yearly The temperature measurement device should be calibrated and traceable to a national standard. 21
  • 22. Precision of injection Test procedure Acceptance limits Test frequency • Five injections of a standard. • Calculation of relative standard deviation. <2% RSD Yearly Carryover of injection Test procedure Acceptance limits Test frequency Blank solvent is injected after standard. Peak ratio is measured between blank and standard injection. <5% Yearly PERFORMANCE QUALIFICATION: Detector linearity 22
  • 23. Procedure Acceptance criteria • Blank is injected followed by ‘detector linearity solutions’ and the peak responses are recorded. • A standard plot is drawn between the concentrations vs the peak responses. The plot should be linear and regression coefficient (R2) should not be less than 0.99. Flow rate accuracy: • The digital flow meter is connected to the detector outlet port. • The carrier gas flow is set and wait till it reaches the set flow. • The observed flow in replicate is noted. • The procedure is repeated for other carrier gases such as hydrogen and air. • The flow rate of carrier gas should be ±10% of set flow System precision: • 20ml of methanol, ethanol and acetone is transferred into 100ml volumetric flask and made up with ethyl acetate. • Blank is injected followed by standard preparation in six replicates. • The areas and retention times are noted down. • The %RSD (Relative standard deviation) of retention time should not be more than 1.0% and peak area NMT (not more than) 5.0% Detector noise and Drift test: • After GC is ready, the system is run up to 15 minutes through single run. • When the run is completed noise and drift is calculated through software. • The acceptance values are: a) Noise not more than 100µV. b) Drift not more than 2500µV/hr. 23
  • 24. REFERENCES:  Huber L. Validation and Qualification in Analytical Laboratories. Informa Healthcare USA, New York. 2007 (2).11-19.  Chan C. C, Lee Y. C, Lam H and Zhang X. M. Analytical Method Validation and Instrument Performance Verification. John Wiley & Sons, Inc., Hoboken, New Jersey. 2004.137-148.  Krull I. S and Swartz M. E. Operational and Performance Qualification. LCGC North America, April, 2002, 20 (4), 356-362.  Gowrisankar D, Abbulu K, Souri O. B and Sujana K. Validation and Calibration of Analytical Instruments. Journal of Biomedical Sciences and Research, 2010, 2 (2), 89-99. 24