SlideShare une entreprise Scribd logo
1  sur  58
Télécharger pour lire hors ligne
1
Dr.V.Ramana reddy
Dr.V.Ramana reddy
2
Fluoroscopy
Purpose
Purpose
To visualize, in real time:
To visualize, in real time:
 organ motion
organ motion
 ingested or injected contrast agents
ingested or injected contrast agents
 insert stents
insert stents
 cathetarize small blood vessels
cathetarize small blood vessels
REAL TIME IMAGING
3
X-rays were discovered because of
there ability to cause fluorescence .
The first x-ray image of human part was
observed fluoroscopically-Dr. Glasser.
4
THE FLUOROSCOPE
 First generation fluoroscopes consisted of an x-ray
tube, an x-ray table and a fluoroscopic screen.
 The fluorescent material used in screen was copper
activated zinc cadmium sulfide that emitted light in
yellow-green spectrum.
 A sheet of lead glass covered the screen, so that
radiologist could stare directly into the screen with out
having the x-ray beam strike his eyes.
 Screen fluoroscence was very faint so, the
examination was carried out in a dark room by the
radiologist who had to adapt his eyes by wearing red
goggles for 20-30 mins prior to the examination 
technique is now obsolete & gone.
5
FLUOROSCOPE
FLUOROSCOPE
6
Photograph
Photograph
shows an early
shows an early
(1933)
(1933)
fluoroscopic
fluoroscopic
system in use
system in use
before the
before the
development
development
of image
of image
intensification.
intensification.
An actual
An actual
fluoroscopic
fluoroscopic
examination
examination
with this
with this
device would
device would
have occurred
have occurred
in a darkened
in a darkened
room.
room.
7
IMAGE INTENSIFIER DESIGN
 Image intensifier was discovered in 1950s-to
produce an image bright enough to allow cone
vision without giving the pt an excess radiation
exposure.
 The components of an x-ray image intensifier
 The tube itself is an evacuated glass
envelope ,a vacuum tube containing-
1.input phosphor and photocathode .
2.electrostatic focusing lens.
3.accelerating anode.
4.out put phosphor.
8
 After an x-ray beam passes the pt it enters the image
intensifier tube the input fluorescent screen absorbs
x-ray photons and converts their energy into light
photons.
 The light photons strike the photo cathode, causing it
to emit photoelectrons  these electrons are
immediately drawn away from the photocathode by
the high potential difference betn it &the accelerating
anode.
 As the electrons flow from the cathode towards the
anode, they are focused by an electrostatic lens which
guides them to the output fluorescent screen without
distorting their geometric configuration.
9
 The electrons strike the output screen, which
emits the light photons that carry the
fluoroscopic images to the eye of the observer.
 In intensifier tube, the image is first carried by
the x-ray photons, then by the light photons,
next by the electrons &finally by the light
photons.
10
IMAGE INTENSIFIER
11
12
INPUT PHOSPHOR & PHOTO CATHODE
INPUT PHOSPHOR & PHOTO CATHODE:
:
 The input fluorescent screen in image intensifiers is
cesium iodide (CsI). (older intensifier- silver activated
zinc cadmium sulfide).
 CsI is deposited on a thin aluminum substrate by a
process called “vapor deposition”.  an interesting &
useful characteristic of CsI is that during the
deposition process the crystals of CsI grow in tiny
needles perpendicular to the substrate.  There by
reducing scattering.
13
INPUT PHOSPHOR & PHOTO CATHODE:
 Image quality is dramatically better with CSI input
screen than it was with zinc cadmium sulfide screens.
 Three physical characteristics of CsI make it superior.
1. vertical orientation of the crystals.
2. A greater packing density &
3. A more favorable effective atomic number.
Phosphor thickness have been reduced comparably
from app. 0.3mm with Zn-Cd su to 0.1mm with CsI. The
principal advantage of thinner phosphor layer combined
with needle shaped crystals is improved resolution.
14
PHOTO CATHODE
PHOTO CATHODE:
:
 The photo cathode is a photoemissive metal
(commonly a combination of antimony & cesium
compounds).
 When the light from the fluorescent screen strikes the
photo cathode, photo electrons are emitted in numbers
proportional to the brightness of the screen.
 The photo cathode is applied directly to the CsI input
phosphor.
 The photoelectrons thus produced has to be moved to
the other end of the image intensifier. This can be
done using an electrostatic focusing lens and an
accelerating anode.
anode.
15
ELECTROSTATIC FOCUSING CUP:
 The lens is made up of a series positively charged
electrodes that are usually plated on to the inside surface
of the glass envelope.
 These electrodes focus the electron beam as it flows from
the photocathode toward the output phosphor.
 Electron focusing inverts & reverses the image which is
called “point inversion” because all the electrons pass
through a common focal point on their way to output
phosphor.
 The input phosphor is curved to ensure that electrons
emitted at the peripheral regions of the photocathode
travel the same distance as those emitted from the
central region.
 The image on the output phosphor is reduced in size
,which is one of the principle reasons why it is brighter.
16
ACCELERATING ANODE
ACCELERATING ANODE :
:
 The anode is located in the neck of the image tube.
 Its function is to accelerate electrons emitted from the
photocathode towards the output screen. the anode has a
+ve potential of 25 to 35 kv relative to the photocathode,
so it accelerates electrons to a tremendous velocity.
OUTPUT PHOSPHOR:
OUTPUT PHOSPHOR:
 The output fluorescent screen of image intensifiers is
The output fluorescent screen of image intensifiers is
silver activated zn-cd sulfide, the same used in Ist
silver activated zn-cd sulfide, the same used in Ist
generation input phosphor.
generation input phosphor.
 Crystal size and layer thickness are reduced to maintain
Crystal size and layer thickness are reduced to maintain
resolution in the minified image.
resolution in the minified image.
 A thin layer of aluminum is plated onto the fluorescent
A thin layer of aluminum is plated onto the fluorescent
screen
screen
 prevent light from moving retrograde through
prevent light from moving retrograde through
the tube & activating the photocathode.
the tube & activating the photocathode.
17
 The glass tube of the image intensifier is abt 2 to 4mm
thick &is enclosed in a lead lined metal container
protects the operator from stray radiation.
 The output phosphor image is viewed either directly
through a series of lenses and mirrors or indirectly
through closed circuit TV.
BRIGHTNESS GAIN:
BRIGHTNESS GAIN:
 Two methods are used to evaluate the brightness gain
of image intensifiers. the first compares the luminance
of an intensifier output screen to that of a Patterson
type B2 fluoroscopy screen when both are exposed to
same quantity of radiation.
 The brightness gain is the ratio of the two
illuminations.
 Brightness gain=intensifier luminance/Patterson b-2
lumin.
18
 Another way of evaluation of brightness gain is called as
“conversion factor”
 Conversion factor =cd/m2 by mR/sec
 Output screen luminance is measured in candelas.
 Radiation quality & output luminance are explicitly
defined, so the method is accurate & reproducible.
 The brightness gain tends to deteriorate as the image
intensifier ages. ie the pt dose with an older image
intensifier tends to be higher than with a new intensifier
of the same type.
 The brightness gain of an imag inten comes from 2
completely unrelated sources called minification gain
&flux gain.
19
MINIFICATION GAIN:
MINIFICATION GAIN:
 The brightness gain from minification is produced by a
reduction in image size.
 The quantity of gain depends on the relative areas of
input &output screens. coz the size of the intensifier is
usually indicated by its diameter, so minif gain is
expressed as MG=(d1/d0)2.d1=diameter of input
screen,d0=diameter of output screen.
 Most img inten have an input screen from 5 to9 in.& an
output screen of app 1in in diameter.
 Theoretically minification can be increased indefinitely
as we can see from above formula, but as the
minification is increased the image becomes smaller.-
disadvantage.
 Hence image has to be magnified and viewed which will
result in reduce brightness & precipitous drop in
resolution.
20
FLUX GAIN:
FLUX GAIN:
 FLUX gain increases the brightness of the fluoroscopic
image by a factor of app 50.
 The total brightness of an imag intes is product of
minification & flux gain: ie
Brightness gain =minification gain x flux gain.
21
MULTIPLE-FIELD IMAGE INTENSIFIERS
MULTIPLE-FIELD IMAGE INTENSIFIERS
 Dual field or triple field imsg intes attempt to
resolve the conflicts btn image size & quality.
 They can be operated in several modes, including
the 4.5in, a 6in, or a 9in mode. the 9in mode is
used to view large anatomic areas. When size is
unimportant the 4.5in or 6in mode is used coz of
better resultant image quality.
 Field size is changed by a simple electronic
principle: the higher the voltage on the
electrostatic focusing lens, the more the electron
beam is focused.
22
 This figure shows this principle
applied to a dual field imag
intes.
 In the 9in mode, the
electrostatic focusing voltage is
decreased. the electrons focus to
a point or cross, close to the
output phosphor & the final
image is actually smaller than
the phosphor.
 In 6in mode the electrostatic
focusing voltage is increased &
the electrons focus farther away
from the output phosphor. after
the electrons cross they diverge,
so the image on the output
phosphor is larger than in the
23
Optical coupling
Optical coupling
 Optical system transmits the output of the image
Optical system transmits the output of the image
intensifier to the light sensitive area of the video
intensifier to the light sensitive area of the video
camera
camera
 The optical distributor include beam-splitting
The optical distributor include beam-splitting
mirror, which directs a portion of the light from
mirror, which directs a portion of the light from
the image intensifier output window to an
the image intensifier output window to an
accessory device for image recording and passes
accessory device for image recording and passes
the remainder to the video camera.
the remainder to the video camera.
 Two lenses are mounted in tandem
Two lenses are mounted in tandem
 The II and the vidicon are placed at the focal
The II and the vidicon are placed at the focal
planes of the two lenses
planes of the two lenses
24
25
Closed-circuit Television System
 used to view the image intensifier output image
used to view the image intensifier output image
 Consists of
Consists of
1)Television camera
1)Television camera
2) Camera control unit
2) Camera control unit
3) Monitor
3) Monitor
 The television system allows for real-time
The television system allows for real-time
viewing of the fluoroscopic image by several
viewing of the fluoroscopic image by several
people at once from one monitor or multiple
people at once from one monitor or multiple
monitors
monitors
26
Television Camera Tube
Television Camera Tube
 Output phosphor is directly coupled to a TV
camera tube
 Plumbicon
 Vidicon
 CCD
27
 The basic video camera consists of
1) vacuum tube cylinder (approximately 2.5 cm in
diameter) surrounded by electromagnetic focusing
coils ,2 pairs of electrostatic deflecting coils
2) photoconductive target
3) a scanning electron beam
Target assembly:
a) Glass plate assembly
b) signal plate
c) Target
28
PICK UP TUBE
29
Target
 Functionally most important i n tube
 Thin film of photoconductive material, antimony sulfide
suspended as globules in mica matrix.
 The optical coupling lens focuses the image intensifier output
image onto the target, forming a charge image within the
photoconductive layer
 This latent image is read out by the electron beam, which
scans across the target in a series of horizontal raster lines.
 As the scanning electron beam moves across the target, a
current signal is produced that represents the two-dimensional
image as a continuous series of raster lines with varying
voltage levels.
30
Video signal
 When globules absorbs light ,photoelectrons are
When globules absorbs light ,photoelectrons are
emitted
emitted
 The globule becomes positively charged
The globule becomes positively charged
 The electron beam scans the electrical image stored
The electron beam scans the electrical image stored
on the target & fills in the holes left by the emitted
on the target & fills in the holes left by the emitted
photoelectrons, discharging the tiny globular
photoelectrons, discharging the tiny globular
capacitors
capacitors
 When the electrons in beam neutralize the positive
When the electrons in beam neutralize the positive
charge in the globules , the electrons on the signal
charge in the globules , the electrons on the signal
plate leave the plate via resistor
plate leave the plate via resistor
 These moving electrons form a current flowing
These moving electrons form a current flowing
through a resistor and voltage across the resistor
through a resistor and voltage across the resistor
 This voltage ,when collected for each neutralized
This voltage ,when collected for each neutralized
globule, constitutes the video signal
globule, constitutes the video signal
31
32
Television monitor
Television monitor
 The video signal produced by the video camera is converted into a visible
image by the monitor
 Contains picture tube & controls for regulating brightness & contrast
 Picture tube contains-
Electron gun
control grid
anode
focusing coil, deflecting coil-control the electron beam
synchrony with the camera tube
 Control grid-receive video signal from Camera control unit ,uses this signal
to regulate the no. of electrons in electron beam
 Anode –carries higher potential (10,000V) accelerates the electron beam to
much higher velocity
 The electron Strike the fluorescent screen ,emit large number of light
photons.
33
34
Image Recording
Image Recording
 Two modes of recording th flouroscopic
image -
1) light image from output phosphor of II
recorded on film with a photospot camera
or cine camera
2) electrical signal generated by TV camera-
includes magnetic tape, magnetic discs &
optical discs
35
DIRECT FILM RECORDING
SPOT FILM DEVICES:
 Fluoroscopic systems designed for gastrointestinal imaging
Fluoroscopic systems designed for gastrointestinal imaging
are generally equipped with a spot film device.
are generally equipped with a spot film device.
 The spot film device allows exposure of a conventional
The spot film device allows exposure of a conventional
screen-film cassette in conjunction with fluoroscopic
screen-film cassette in conjunction with fluoroscopic
viewing. This rather familiar system, located in front of the
viewing. This rather familiar system, located in front of the
image intensifier, accepts the screen-film cassette and
image intensifier, accepts the screen-film cassette and
"parks" it out of the way during fluoroscopy
"parks" it out of the way during fluoroscopy
 Cassettes may be loaded from the front or rear depending on
Cassettes may be loaded from the front or rear depending on
the design of the system.
the design of the system.
36
Standard spot film
Standard spot film
imaging configuration
imaging configuration
typical of
typical of
gastrointestinal
gastrointestinal
fluoroscopy equipment.
fluoroscopy equipment.
The screen-film cassette
The screen-film cassette
is parked out of the x-
is parked out of the x-
ray field until the spot
ray field until the spot
film trigger is pressed,
film trigger is pressed,
causing both the
causing both the
cassette and the
cassette and the
formatting mask to
formatting mask to
move into position.
move into position.
37
 The x-ray field size is also reduced automatically by the
collimators at the time of exposure to minimize scattered
radiation and patient radiation dose.
 The fluoroscopist can override this automatic collimation to
further reduce the x-ray field.
 Spot film imaging uses essentially the same technology as
conventional screen-film radiography.
Differences and limitations of spot film imaging
compared with general radiography.
 One major limitation is the range of film sizes available for
spot film imaging. Although some older fluoroscopy
equipment is limited to a single size, usually 24 x 24 cm,
current equipment allows a range of film sizes to be used,
from 20 x 25 cm to 24 x 35 cm.
 Spot film devices usually allow more than one image to be
obtained on a single film.
 Formats typically include one, two, three, four, or six images
on a film.
38
 Moving the spot film device closer to the patient
reduces the amount of magnification and decreases
the patient radiation dose.
 A number of factors affect patient doses in spot film
imaging.
 The source-to-skin distance is shorter in spot film imaging
than in general radiography.
 Although the automatic exposure control system fixes the
exposure to the screen, the shorter source-to-skin distance
increases the inverse square reduction in radiation intensity
as it passes through the patient.
 This increase tends to make the skin entrance exposure
higher.
 The field size in spot film imaging is generally smaller than
that used in general radiography.
 This smaller field size reduces scatter and therefore tends to
reduce dose. For the same reason, grids used in fluoroscopy
generally have a lower grid ratio and therefore a smaller
Bucky factor, which also leads to lower dose.
 These effects tend to offset each other to a large extent.
39
 One of the major shortcomings of conventional spot
film devices is the delay involved in moving the
cassette into position for exposure.
 In gastrointestinal imaging, this delay can be
overcome by using photofluorography.
 In vascular imaging, more rapid film movement is
achieved with automatic film changers.
40
AUTOMATIC FILM CHANGERS
 The automatic film changers used in vascular imaging are
also screen-film systems.
 They can be found in several varieties. Some are large,
floor-mounted boxes, but systems more commonly used
today mount on the image intensifier
 The system consists of a supply magazine for holding
unexposed film, a receiving magazine, a pair of
radiographic screens, and a mechanism for transferring the
film.
 When an exposure is required, the screens are mechanically
separated, the film is pulled into place between them, and
they are closed.
 After the film is exposed, the screens separate again.
 The film is moved to the receiver, and another film is pulled
into place for the next exposure.
 The number of films and filming rates must be
preprogrammed for proper operation.
41
Photospot camera
 Records the the image output of an II on a film
 Film – role film/cut film(10 cm)
 Advantage –1)reduction in pt exposure
2)film does not have to be changed
b/w exposures
3)exposure times are shorter-motion is less
likely problem
4)films can be taken more rapidly
5)possible to record & view image at same
time
42
Framing with spot film cameras:
Framing with spot film cameras:
 Framing –utilisation of available area on film
Framing –utilisation of available area on film
 The output phosphor of II tube is round,
The output phosphor of II tube is round,
shape of film is square
shape of film is square
 4 framing patterns
4 framing patterns
1) Exact framing
1) Exact framing
2) Equal area framing
2) Equal area framing
3) mean diameter framing
3) mean diameter framing
4) total area framing
4) total area framing
 Equal area framing or mean diameter framing
Equal area framing or mean diameter framing
is recommended for most clinical situations.
is recommended for most clinical situations.
43
44
Cineflourography
Cineflourography
 Process of recording fluoroscopic images on
Process of recording fluoroscopic images on
movie (cine) film
movie (cine) film
 Two film sizes- 16 mm, 35 mm
Two film sizes- 16 mm, 35 mm
 Cine camera-components are lens, iris
Cine camera-components are lens, iris
diaphragm, shutter, aperture, pressure plate, pull
diaphragm, shutter, aperture, pressure plate, pull
down arm & film transport mechanism
down arm & film transport mechanism
45
46
TV image recorders :
3) TAPE RECORDER : used for both recoding& playback
• as a recorder receives video signal from camera
control unit
• for playback transmits the signal to one or more
several TV monitors
Components:1)magnetic tube
2)writing head
3) tape transport system
Writing head converts an electrical signal in to magnetic
field for recoding & converts magnetic signal to electric
signal for replay
47
48
49
DIGITAL FLUOROGRAPHY
Digital charge-coupled device (CCD) TV cameras are
rapidly replacing conventional TV cameras in fluoroscopic
systems.
An analog, high-resolution (1,023-line) TV camera has
a vertical resolution of about 358 line pairs.
A high-resolution CCD camera with a 1,024 x 1,024
matrix will provide equivalent vertical resolution. However,
the digital camera will have the same vertical and
horizontal resolution, whereas the horizontal resolution of
the analog camera is defined by its electronic band-pass.
50
For a 15-cm-diameter intensifier input, the
limiting resolution of the CCD camera would be
358 lp/150 mm or 2.4 lp/mm. This result is about
half the resolution of a photospot film. This
resolution loss is made up for by the ability to
digitally increase display contrast, reduce noise,
and enhance the edges of digital images.
There are several other advantages to digital
photospot images. Mechanical devices are not
needed for film transport. Film processing is not
required. Images can be viewed immediately. The
linear response of the digital system makes it very
forgiving of under- or overexposure.
51
Fluoroscopic Equipment Configurations
The basic configurations include radiography/fluoroscopy (R/F) tables with
either an under-table or over-table x-ray tube and fixed C-arm, mobile C-arm,
and mini C-arm
R/F Units with Under-Table X-ray Tube:
most common fluoroscopic equipment configuration
The x-ray tube and collimator are mounted below the tabletop with the image
intensifier tower mounted above the table on a carriage that can be panned
over the patient.
In addition to the standard fluoroscopic imaging chain, R/F systems include
an overhead x-ray tube that can be used for regular radiography with a Bucky
incorporated into the table.
Other common features include a tilting table and image recording devices.
52
Under-table x-ray tube R/F system. Photograph shows an example of an
Under-table x-ray tube R/F system. Photograph shows an example of an
R/F table that includes a spot film device and side-mounted video camera.
R/F table that includes a spot film device and side-mounted video camera.
53
R/F Units with Over-Table X-ray Tube
R/F Units with Over-Table X-ray Tube
 x-ray tube mounted over the table with the
x-ray tube mounted over the table with the
image intensifier below.
image intensifier below.
 this configuration results in increased patient
this configuration results in increased patient
access, which is helpful for interventional
access, which is helpful for interventional
procedures.
procedures.
 Radiography can be performed with the same x-
Radiography can be performed with the same x-
ray tube and a Bucky incorporated into the table.
ray tube and a Bucky incorporated into the table.
 The x-ray tube can be angled to acquire
The x-ray tube can be angled to acquire
angulated projections or tomograms.
angulated projections or tomograms.
54
Over-table x-ray tube R/F system. Photograph shows a sample system
Over-table x-ray tube R/F system. Photograph shows a sample system
that can be controlled from within the procedure room with the pedestal
that can be controlled from within the procedure room with the pedestal
control panel (left) or from outside the room from the remote desk controls
control panel (left) or from outside the room from the remote desk controls
(right).
(right).
55
SUMMARY:
 Early fluroscopy was accomplished by
radiologists looking directly at a fluoroscopic
screen.
 The image on the screen was only 0.0001 as bright
as the image of a routinely viewed radiograph, so
dark adaptation of eyes was required.
 In 1950s the image intensifier alleviated this
situation by producing an image bright enough to
be viewed with cone vision.
 The input phosphor of modern image intensifier
is CsI; the output phosphor is zinc cadmium
56
 Brightness gain is the product of minification gain and
flux gain.
 Imaging characteristics important in the evaluation of
image intensifier fluoroscopy include contrast, lag and
distortion.
 Large field of view image intensifier tubes are available
to fill special needs, such as digital and spot film
angiography.
 Most image intensifiers allow dual field or triple field
imaging.
57
 Output phospher image is processed by television
Output phospher image is processed by television
camera tube (vidicon, plumbicon, CCD)
camera tube (vidicon, plumbicon, CCD)
 The image is displayed on TV monitor
The image is displayed on TV monitor
 Standard x-ray closed-circuit television uses
Standard x-ray closed-circuit television uses
525x525 format with 5-MHZ band pass
525x525 format with 5-MHZ band pass
 Vertical resolution limited by scan line format,
Vertical resolution limited by scan line format,
horizontal resolution is a function of band pass
horizontal resolution is a function of band pass
 Light image is recorded by photo spot camera or
Light image is recorded by photo spot camera or
cine camera
cine camera
58

Contenu connexe

Similaire à flouroscopicimging-120617043323-phpapp01.pdf

Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxRohit Bansal
 
Fluoroscopy presentation1
Fluoroscopy presentation1Fluoroscopy presentation1
Fluoroscopy presentation1Huzaifa Oxford
 
basic of angiography physics and equipement.pdf
basic of angiography physics and equipement.pdfbasic of angiography physics and equipement.pdf
basic of angiography physics and equipement.pdfnaima SENHOU
 
Different types of microscopes
Different types of microscopesDifferent types of microscopes
Different types of microscopesUTTAM KUMAR DAS
 
RDT-112-PRELIM-LESSON-1.docx
RDT-112-PRELIM-LESSON-1.docxRDT-112-PRELIM-LESSON-1.docx
RDT-112-PRELIM-LESSON-1.docxJianSoliman2
 
fluoroscopy imaging
 fluoroscopy imaging  fluoroscopy imaging
fluoroscopy imaging Estam Udith
 
Lectures Notes Types of microscopes.pptx
Lectures Notes Types of microscopes.pptxLectures Notes Types of microscopes.pptx
Lectures Notes Types of microscopes.pptxPeninaTabitha
 
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)mr_koky
 
Computed radiography
Computed radiographyComputed radiography
Computed radiographyAnjan Dangal
 
2. Microscopy (Biochemistry)
2. Microscopy (Biochemistry)2. Microscopy (Biochemistry)
2. Microscopy (Biochemistry)Jay Khaniya
 
Transmission electron microscope
Transmission electron microscopeTransmission electron microscope
Transmission electron microscopeSenthil Arasan
 

Similaire à flouroscopicimging-120617043323-phpapp01.pdf (20)

Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptx
 
Fluoroscopic Imaging.pptx
Fluoroscopic Imaging.pptxFluoroscopic Imaging.pptx
Fluoroscopic Imaging.pptx
 
Flouroscopy JP.pptx
Flouroscopy JP.pptxFlouroscopy JP.pptx
Flouroscopy JP.pptx
 
Fluroscopy.pptx
Fluroscopy.pptxFluroscopy.pptx
Fluroscopy.pptx
 
Fluroscopy
FluroscopyFluroscopy
Fluroscopy
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
Fluoroscopy presentation1
Fluoroscopy presentation1Fluoroscopy presentation1
Fluoroscopy presentation1
 
basic of angiography physics and equipement.pdf
basic of angiography physics and equipement.pdfbasic of angiography physics and equipement.pdf
basic of angiography physics and equipement.pdf
 
flouroscopy.pptx
flouroscopy.pptxflouroscopy.pptx
flouroscopy.pptx
 
Night vision glasses
Night vision glassesNight vision glasses
Night vision glasses
 
Different types of microscopes
Different types of microscopesDifferent types of microscopes
Different types of microscopes
 
FLUOROSCOPIC IMAGING.ppt
FLUOROSCOPIC IMAGING.pptFLUOROSCOPIC IMAGING.ppt
FLUOROSCOPIC IMAGING.ppt
 
RDT-112-PRELIM-LESSON-1.docx
RDT-112-PRELIM-LESSON-1.docxRDT-112-PRELIM-LESSON-1.docx
RDT-112-PRELIM-LESSON-1.docx
 
Xri fin
Xri finXri fin
Xri fin
 
fluoroscopy imaging
 fluoroscopy imaging  fluoroscopy imaging
fluoroscopy imaging
 
Lectures Notes Types of microscopes.pptx
Lectures Notes Types of microscopes.pptxLectures Notes Types of microscopes.pptx
Lectures Notes Types of microscopes.pptx
 
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
2. Microscopy (Biochemistry)
2. Microscopy (Biochemistry)2. Microscopy (Biochemistry)
2. Microscopy (Biochemistry)
 
Transmission electron microscope
Transmission electron microscopeTransmission electron microscope
Transmission electron microscope
 

Plus de VishnuDutt40

Pediatric obesity and its related complications.pptx
Pediatric obesity and its related complications.pptxPediatric obesity and its related complications.pptx
Pediatric obesity and its related complications.pptxVishnuDutt40
 
scattered radiation.pptx
scattered radiation.pptxscattered radiation.pptx
scattered radiation.pptxVishnuDutt40
 
Abdominal x ray- views.pptx
Abdominal x ray- views.pptxAbdominal x ray- views.pptx
Abdominal x ray- views.pptxVishnuDutt40
 
PULSE OXIMETRY.pptx
PULSE OXIMETRY.pptxPULSE OXIMETRY.pptx
PULSE OXIMETRY.pptxVishnuDutt40
 
ROLE OF INTRAVENOUS THIAMINE IN NEONATES WITH MECONIUM.pptx
ROLE OF INTRAVENOUS THIAMINE IN NEONATES  WITH MECONIUM.pptxROLE OF INTRAVENOUS THIAMINE IN NEONATES  WITH MECONIUM.pptx
ROLE OF INTRAVENOUS THIAMINE IN NEONATES WITH MECONIUM.pptxVishnuDutt40
 
AORTIC ARCH ANOMALIES 2.pptx
AORTIC ARCH ANOMALIES 2.pptxAORTIC ARCH ANOMALIES 2.pptx
AORTIC ARCH ANOMALIES 2.pptxVishnuDutt40
 
benign focal lesions in liver.pptx
benign focal lesions in liver.pptxbenign focal lesions in liver.pptx
benign focal lesions in liver.pptxVishnuDutt40
 
classification.pptx
classification.pptxclassification.pptx
classification.pptxVishnuDutt40
 
protocol ppt final.pptx
protocol ppt final.pptxprotocol ppt final.pptx
protocol ppt final.pptxVishnuDutt40
 
THESIS PPT Dr vishnu.pptx
THESIS PPT Dr vishnu.pptxTHESIS PPT Dr vishnu.pptx
THESIS PPT Dr vishnu.pptxVishnuDutt40
 
THESIS PPT Dr Hafsal.pptx
THESIS PPT Dr Hafsal.pptxTHESIS PPT Dr Hafsal.pptx
THESIS PPT Dr Hafsal.pptxVishnuDutt40
 
Bag and Mask Ventilation.pptx
Bag and Mask Ventilation.pptxBag and Mask Ventilation.pptx
Bag and Mask Ventilation.pptxVishnuDutt40
 
ctanatomy-130625014916-phpapp02.pdf
ctanatomy-130625014916-phpapp02.pdfctanatomy-130625014916-phpapp02.pdf
ctanatomy-130625014916-phpapp02.pdfVishnuDutt40
 
seldingertechnique-171213172044.pdf
seldingertechnique-171213172044.pdfseldingertechnique-171213172044.pdf
seldingertechnique-171213172044.pdfVishnuDutt40
 
bariumprocedures-180530182835.pptx
bariumprocedures-180530182835.pptxbariumprocedures-180530182835.pptx
bariumprocedures-180530182835.pptxVishnuDutt40
 
cvjanomalies-181113135852.pdf
cvjanomalies-181113135852.pdfcvjanomalies-181113135852.pdf
cvjanomalies-181113135852.pdfVishnuDutt40
 

Plus de VishnuDutt40 (20)

Pediatric obesity and its related complications.pptx
Pediatric obesity and its related complications.pptxPediatric obesity and its related complications.pptx
Pediatric obesity and its related complications.pptx
 
scattered radiation.pptx
scattered radiation.pptxscattered radiation.pptx
scattered radiation.pptx
 
Abdominal x ray- views.pptx
Abdominal x ray- views.pptxAbdominal x ray- views.pptx
Abdominal x ray- views.pptx
 
PULSE OXIMETRY.pptx
PULSE OXIMETRY.pptxPULSE OXIMETRY.pptx
PULSE OXIMETRY.pptx
 
ROLE OF INTRAVENOUS THIAMINE IN NEONATES WITH MECONIUM.pptx
ROLE OF INTRAVENOUS THIAMINE IN NEONATES  WITH MECONIUM.pptxROLE OF INTRAVENOUS THIAMINE IN NEONATES  WITH MECONIUM.pptx
ROLE OF INTRAVENOUS THIAMINE IN NEONATES WITH MECONIUM.pptx
 
MST (1).pptx
MST (1).pptxMST (1).pptx
MST (1).pptx
 
spotter.pptx
spotter.pptxspotter.pptx
spotter.pptx
 
AORTIC ARCH ANOMALIES 2.pptx
AORTIC ARCH ANOMALIES 2.pptxAORTIC ARCH ANOMALIES 2.pptx
AORTIC ARCH ANOMALIES 2.pptx
 
benign focal lesions in liver.pptx
benign focal lesions in liver.pptxbenign focal lesions in liver.pptx
benign focal lesions in liver.pptx
 
vish ankle.pptx
vish ankle.pptxvish ankle.pptx
vish ankle.pptx
 
classification.pptx
classification.pptxclassification.pptx
classification.pptx
 
mri physics.pptx
mri physics.pptxmri physics.pptx
mri physics.pptx
 
protocol ppt final.pptx
protocol ppt final.pptxprotocol ppt final.pptx
protocol ppt final.pptx
 
THESIS PPT Dr vishnu.pptx
THESIS PPT Dr vishnu.pptxTHESIS PPT Dr vishnu.pptx
THESIS PPT Dr vishnu.pptx
 
THESIS PPT Dr Hafsal.pptx
THESIS PPT Dr Hafsal.pptxTHESIS PPT Dr Hafsal.pptx
THESIS PPT Dr Hafsal.pptx
 
Bag and Mask Ventilation.pptx
Bag and Mask Ventilation.pptxBag and Mask Ventilation.pptx
Bag and Mask Ventilation.pptx
 
ctanatomy-130625014916-phpapp02.pdf
ctanatomy-130625014916-phpapp02.pdfctanatomy-130625014916-phpapp02.pdf
ctanatomy-130625014916-phpapp02.pdf
 
seldingertechnique-171213172044.pdf
seldingertechnique-171213172044.pdfseldingertechnique-171213172044.pdf
seldingertechnique-171213172044.pdf
 
bariumprocedures-180530182835.pptx
bariumprocedures-180530182835.pptxbariumprocedures-180530182835.pptx
bariumprocedures-180530182835.pptx
 
cvjanomalies-181113135852.pdf
cvjanomalies-181113135852.pdfcvjanomalies-181113135852.pdf
cvjanomalies-181113135852.pdf
 

Dernier

Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsMedicoseAcademics
 
tongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacytongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacyDrMohamed Assadawy
 
VIP Hyderabad Call Girls KPHB 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls KPHB 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls KPHB 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls KPHB 7877925207 ₹5000 To 25K With AC Room 💚😋mahima pandey
 
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...Sheetaleventcompany
 
Nagpur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Nagpur No💰...
Nagpur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Nagpur No💰...Nagpur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Nagpur No💰...
Nagpur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Nagpur No💰...Sheetaleventcompany
 
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...GENUINE ESCORT AGENCY
 
Control of Local Blood Flow: acute and chronic
Control of Local Blood Flow: acute and chronicControl of Local Blood Flow: acute and chronic
Control of Local Blood Flow: acute and chronicMedicoseAcademics
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...gragneelam30
 
Whitefield { Call Girl in Bangalore ₹7.5k Pick Up & Drop With Cash Payment 63...
Whitefield { Call Girl in Bangalore ₹7.5k Pick Up & Drop With Cash Payment 63...Whitefield { Call Girl in Bangalore ₹7.5k Pick Up & Drop With Cash Payment 63...
Whitefield { Call Girl in Bangalore ₹7.5k Pick Up & Drop With Cash Payment 63...dishamehta3332
 
Electrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdfElectrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdfMedicoseAcademics
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan 087776558899
 
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...Oleg Kshivets
 
Jaipur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Jaipur No💰...
Jaipur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Jaipur No💰...Jaipur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Jaipur No💰...
Jaipur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Jaipur No💰...Sheetaleventcompany
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Sheetaleventcompany
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...soniya pandit
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...Sheetaleventcompany
 
Intramuscular & Intravenous Injection.pptx
Intramuscular & Intravenous Injection.pptxIntramuscular & Intravenous Injection.pptx
Intramuscular & Intravenous Injection.pptxsaranpratha12
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana GuptaLifecare Centre
 
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Sheetaleventcompany
 

Dernier (20)

Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
tongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacytongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacy
 
VIP Hyderabad Call Girls KPHB 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls KPHB 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls KPHB 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls KPHB 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
 
Nagpur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Nagpur No💰...
Nagpur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Nagpur No💰...Nagpur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Nagpur No💰...
Nagpur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Nagpur No💰...
 
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
 
Control of Local Blood Flow: acute and chronic
Control of Local Blood Flow: acute and chronicControl of Local Blood Flow: acute and chronic
Control of Local Blood Flow: acute and chronic
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
Whitefield { Call Girl in Bangalore ₹7.5k Pick Up & Drop With Cash Payment 63...
Whitefield { Call Girl in Bangalore ₹7.5k Pick Up & Drop With Cash Payment 63...Whitefield { Call Girl in Bangalore ₹7.5k Pick Up & Drop With Cash Payment 63...
Whitefield { Call Girl in Bangalore ₹7.5k Pick Up & Drop With Cash Payment 63...
 
Electrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdfElectrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdf
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
 
Jaipur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Jaipur No💰...
Jaipur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Jaipur No💰...Jaipur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Jaipur No💰...
Jaipur Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Jaipur No💰...
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
 
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
 
Intramuscular & Intravenous Injection.pptx
Intramuscular & Intravenous Injection.pptxIntramuscular & Intravenous Injection.pptx
Intramuscular & Intravenous Injection.pptx
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
 

flouroscopicimging-120617043323-phpapp01.pdf

  • 2. 2 Fluoroscopy Purpose Purpose To visualize, in real time: To visualize, in real time:  organ motion organ motion  ingested or injected contrast agents ingested or injected contrast agents  insert stents insert stents  cathetarize small blood vessels cathetarize small blood vessels REAL TIME IMAGING
  • 3. 3 X-rays were discovered because of there ability to cause fluorescence . The first x-ray image of human part was observed fluoroscopically-Dr. Glasser.
  • 4. 4 THE FLUOROSCOPE  First generation fluoroscopes consisted of an x-ray tube, an x-ray table and a fluoroscopic screen.  The fluorescent material used in screen was copper activated zinc cadmium sulfide that emitted light in yellow-green spectrum.  A sheet of lead glass covered the screen, so that radiologist could stare directly into the screen with out having the x-ray beam strike his eyes.  Screen fluoroscence was very faint so, the examination was carried out in a dark room by the radiologist who had to adapt his eyes by wearing red goggles for 20-30 mins prior to the examination  technique is now obsolete & gone.
  • 6. 6 Photograph Photograph shows an early shows an early (1933) (1933) fluoroscopic fluoroscopic system in use system in use before the before the development development of image of image intensification. intensification. An actual An actual fluoroscopic fluoroscopic examination examination with this with this device would device would have occurred have occurred in a darkened in a darkened room. room.
  • 7. 7 IMAGE INTENSIFIER DESIGN  Image intensifier was discovered in 1950s-to produce an image bright enough to allow cone vision without giving the pt an excess radiation exposure.  The components of an x-ray image intensifier  The tube itself is an evacuated glass envelope ,a vacuum tube containing- 1.input phosphor and photocathode . 2.electrostatic focusing lens. 3.accelerating anode. 4.out put phosphor.
  • 8. 8  After an x-ray beam passes the pt it enters the image intensifier tube the input fluorescent screen absorbs x-ray photons and converts their energy into light photons.  The light photons strike the photo cathode, causing it to emit photoelectrons  these electrons are immediately drawn away from the photocathode by the high potential difference betn it &the accelerating anode.  As the electrons flow from the cathode towards the anode, they are focused by an electrostatic lens which guides them to the output fluorescent screen without distorting their geometric configuration.
  • 9. 9  The electrons strike the output screen, which emits the light photons that carry the fluoroscopic images to the eye of the observer.  In intensifier tube, the image is first carried by the x-ray photons, then by the light photons, next by the electrons &finally by the light photons.
  • 11. 11
  • 12. 12 INPUT PHOSPHOR & PHOTO CATHODE INPUT PHOSPHOR & PHOTO CATHODE: :  The input fluorescent screen in image intensifiers is cesium iodide (CsI). (older intensifier- silver activated zinc cadmium sulfide).  CsI is deposited on a thin aluminum substrate by a process called “vapor deposition”.  an interesting & useful characteristic of CsI is that during the deposition process the crystals of CsI grow in tiny needles perpendicular to the substrate.  There by reducing scattering.
  • 13. 13 INPUT PHOSPHOR & PHOTO CATHODE:  Image quality is dramatically better with CSI input screen than it was with zinc cadmium sulfide screens.  Three physical characteristics of CsI make it superior. 1. vertical orientation of the crystals. 2. A greater packing density & 3. A more favorable effective atomic number. Phosphor thickness have been reduced comparably from app. 0.3mm with Zn-Cd su to 0.1mm with CsI. The principal advantage of thinner phosphor layer combined with needle shaped crystals is improved resolution.
  • 14. 14 PHOTO CATHODE PHOTO CATHODE: :  The photo cathode is a photoemissive metal (commonly a combination of antimony & cesium compounds).  When the light from the fluorescent screen strikes the photo cathode, photo electrons are emitted in numbers proportional to the brightness of the screen.  The photo cathode is applied directly to the CsI input phosphor.  The photoelectrons thus produced has to be moved to the other end of the image intensifier. This can be done using an electrostatic focusing lens and an accelerating anode. anode.
  • 15. 15 ELECTROSTATIC FOCUSING CUP:  The lens is made up of a series positively charged electrodes that are usually plated on to the inside surface of the glass envelope.  These electrodes focus the electron beam as it flows from the photocathode toward the output phosphor.  Electron focusing inverts & reverses the image which is called “point inversion” because all the electrons pass through a common focal point on their way to output phosphor.  The input phosphor is curved to ensure that electrons emitted at the peripheral regions of the photocathode travel the same distance as those emitted from the central region.  The image on the output phosphor is reduced in size ,which is one of the principle reasons why it is brighter.
  • 16. 16 ACCELERATING ANODE ACCELERATING ANODE : :  The anode is located in the neck of the image tube.  Its function is to accelerate electrons emitted from the photocathode towards the output screen. the anode has a +ve potential of 25 to 35 kv relative to the photocathode, so it accelerates electrons to a tremendous velocity. OUTPUT PHOSPHOR: OUTPUT PHOSPHOR:  The output fluorescent screen of image intensifiers is The output fluorescent screen of image intensifiers is silver activated zn-cd sulfide, the same used in Ist silver activated zn-cd sulfide, the same used in Ist generation input phosphor. generation input phosphor.  Crystal size and layer thickness are reduced to maintain Crystal size and layer thickness are reduced to maintain resolution in the minified image. resolution in the minified image.  A thin layer of aluminum is plated onto the fluorescent A thin layer of aluminum is plated onto the fluorescent screen screen  prevent light from moving retrograde through prevent light from moving retrograde through the tube & activating the photocathode. the tube & activating the photocathode.
  • 17. 17  The glass tube of the image intensifier is abt 2 to 4mm thick &is enclosed in a lead lined metal container protects the operator from stray radiation.  The output phosphor image is viewed either directly through a series of lenses and mirrors or indirectly through closed circuit TV. BRIGHTNESS GAIN: BRIGHTNESS GAIN:  Two methods are used to evaluate the brightness gain of image intensifiers. the first compares the luminance of an intensifier output screen to that of a Patterson type B2 fluoroscopy screen when both are exposed to same quantity of radiation.  The brightness gain is the ratio of the two illuminations.  Brightness gain=intensifier luminance/Patterson b-2 lumin.
  • 18. 18  Another way of evaluation of brightness gain is called as “conversion factor”  Conversion factor =cd/m2 by mR/sec  Output screen luminance is measured in candelas.  Radiation quality & output luminance are explicitly defined, so the method is accurate & reproducible.  The brightness gain tends to deteriorate as the image intensifier ages. ie the pt dose with an older image intensifier tends to be higher than with a new intensifier of the same type.  The brightness gain of an imag inten comes from 2 completely unrelated sources called minification gain &flux gain.
  • 19. 19 MINIFICATION GAIN: MINIFICATION GAIN:  The brightness gain from minification is produced by a reduction in image size.  The quantity of gain depends on the relative areas of input &output screens. coz the size of the intensifier is usually indicated by its diameter, so minif gain is expressed as MG=(d1/d0)2.d1=diameter of input screen,d0=diameter of output screen.  Most img inten have an input screen from 5 to9 in.& an output screen of app 1in in diameter.  Theoretically minification can be increased indefinitely as we can see from above formula, but as the minification is increased the image becomes smaller.- disadvantage.  Hence image has to be magnified and viewed which will result in reduce brightness & precipitous drop in resolution.
  • 20. 20 FLUX GAIN: FLUX GAIN:  FLUX gain increases the brightness of the fluoroscopic image by a factor of app 50.  The total brightness of an imag intes is product of minification & flux gain: ie Brightness gain =minification gain x flux gain.
  • 21. 21 MULTIPLE-FIELD IMAGE INTENSIFIERS MULTIPLE-FIELD IMAGE INTENSIFIERS  Dual field or triple field imsg intes attempt to resolve the conflicts btn image size & quality.  They can be operated in several modes, including the 4.5in, a 6in, or a 9in mode. the 9in mode is used to view large anatomic areas. When size is unimportant the 4.5in or 6in mode is used coz of better resultant image quality.  Field size is changed by a simple electronic principle: the higher the voltage on the electrostatic focusing lens, the more the electron beam is focused.
  • 22. 22  This figure shows this principle applied to a dual field imag intes.  In the 9in mode, the electrostatic focusing voltage is decreased. the electrons focus to a point or cross, close to the output phosphor & the final image is actually smaller than the phosphor.  In 6in mode the electrostatic focusing voltage is increased & the electrons focus farther away from the output phosphor. after the electrons cross they diverge, so the image on the output phosphor is larger than in the
  • 23. 23 Optical coupling Optical coupling  Optical system transmits the output of the image Optical system transmits the output of the image intensifier to the light sensitive area of the video intensifier to the light sensitive area of the video camera camera  The optical distributor include beam-splitting The optical distributor include beam-splitting mirror, which directs a portion of the light from mirror, which directs a portion of the light from the image intensifier output window to an the image intensifier output window to an accessory device for image recording and passes accessory device for image recording and passes the remainder to the video camera. the remainder to the video camera.  Two lenses are mounted in tandem Two lenses are mounted in tandem  The II and the vidicon are placed at the focal The II and the vidicon are placed at the focal planes of the two lenses planes of the two lenses
  • 24. 24
  • 25. 25 Closed-circuit Television System  used to view the image intensifier output image used to view the image intensifier output image  Consists of Consists of 1)Television camera 1)Television camera 2) Camera control unit 2) Camera control unit 3) Monitor 3) Monitor  The television system allows for real-time The television system allows for real-time viewing of the fluoroscopic image by several viewing of the fluoroscopic image by several people at once from one monitor or multiple people at once from one monitor or multiple monitors monitors
  • 26. 26 Television Camera Tube Television Camera Tube  Output phosphor is directly coupled to a TV camera tube  Plumbicon  Vidicon  CCD
  • 27. 27  The basic video camera consists of 1) vacuum tube cylinder (approximately 2.5 cm in diameter) surrounded by electromagnetic focusing coils ,2 pairs of electrostatic deflecting coils 2) photoconductive target 3) a scanning electron beam Target assembly: a) Glass plate assembly b) signal plate c) Target
  • 29. 29 Target  Functionally most important i n tube  Thin film of photoconductive material, antimony sulfide suspended as globules in mica matrix.  The optical coupling lens focuses the image intensifier output image onto the target, forming a charge image within the photoconductive layer  This latent image is read out by the electron beam, which scans across the target in a series of horizontal raster lines.  As the scanning electron beam moves across the target, a current signal is produced that represents the two-dimensional image as a continuous series of raster lines with varying voltage levels.
  • 30. 30 Video signal  When globules absorbs light ,photoelectrons are When globules absorbs light ,photoelectrons are emitted emitted  The globule becomes positively charged The globule becomes positively charged  The electron beam scans the electrical image stored The electron beam scans the electrical image stored on the target & fills in the holes left by the emitted on the target & fills in the holes left by the emitted photoelectrons, discharging the tiny globular photoelectrons, discharging the tiny globular capacitors capacitors  When the electrons in beam neutralize the positive When the electrons in beam neutralize the positive charge in the globules , the electrons on the signal charge in the globules , the electrons on the signal plate leave the plate via resistor plate leave the plate via resistor  These moving electrons form a current flowing These moving electrons form a current flowing through a resistor and voltage across the resistor through a resistor and voltage across the resistor  This voltage ,when collected for each neutralized This voltage ,when collected for each neutralized globule, constitutes the video signal globule, constitutes the video signal
  • 31. 31
  • 32. 32 Television monitor Television monitor  The video signal produced by the video camera is converted into a visible image by the monitor  Contains picture tube & controls for regulating brightness & contrast  Picture tube contains- Electron gun control grid anode focusing coil, deflecting coil-control the electron beam synchrony with the camera tube  Control grid-receive video signal from Camera control unit ,uses this signal to regulate the no. of electrons in electron beam  Anode –carries higher potential (10,000V) accelerates the electron beam to much higher velocity  The electron Strike the fluorescent screen ,emit large number of light photons.
  • 33. 33
  • 34. 34 Image Recording Image Recording  Two modes of recording th flouroscopic image - 1) light image from output phosphor of II recorded on film with a photospot camera or cine camera 2) electrical signal generated by TV camera- includes magnetic tape, magnetic discs & optical discs
  • 35. 35 DIRECT FILM RECORDING SPOT FILM DEVICES:  Fluoroscopic systems designed for gastrointestinal imaging Fluoroscopic systems designed for gastrointestinal imaging are generally equipped with a spot film device. are generally equipped with a spot film device.  The spot film device allows exposure of a conventional The spot film device allows exposure of a conventional screen-film cassette in conjunction with fluoroscopic screen-film cassette in conjunction with fluoroscopic viewing. This rather familiar system, located in front of the viewing. This rather familiar system, located in front of the image intensifier, accepts the screen-film cassette and image intensifier, accepts the screen-film cassette and "parks" it out of the way during fluoroscopy "parks" it out of the way during fluoroscopy  Cassettes may be loaded from the front or rear depending on Cassettes may be loaded from the front or rear depending on the design of the system. the design of the system.
  • 36. 36 Standard spot film Standard spot film imaging configuration imaging configuration typical of typical of gastrointestinal gastrointestinal fluoroscopy equipment. fluoroscopy equipment. The screen-film cassette The screen-film cassette is parked out of the x- is parked out of the x- ray field until the spot ray field until the spot film trigger is pressed, film trigger is pressed, causing both the causing both the cassette and the cassette and the formatting mask to formatting mask to move into position. move into position.
  • 37. 37  The x-ray field size is also reduced automatically by the collimators at the time of exposure to minimize scattered radiation and patient radiation dose.  The fluoroscopist can override this automatic collimation to further reduce the x-ray field.  Spot film imaging uses essentially the same technology as conventional screen-film radiography. Differences and limitations of spot film imaging compared with general radiography.  One major limitation is the range of film sizes available for spot film imaging. Although some older fluoroscopy equipment is limited to a single size, usually 24 x 24 cm, current equipment allows a range of film sizes to be used, from 20 x 25 cm to 24 x 35 cm.  Spot film devices usually allow more than one image to be obtained on a single film.  Formats typically include one, two, three, four, or six images on a film.
  • 38. 38  Moving the spot film device closer to the patient reduces the amount of magnification and decreases the patient radiation dose.  A number of factors affect patient doses in spot film imaging.  The source-to-skin distance is shorter in spot film imaging than in general radiography.  Although the automatic exposure control system fixes the exposure to the screen, the shorter source-to-skin distance increases the inverse square reduction in radiation intensity as it passes through the patient.  This increase tends to make the skin entrance exposure higher.  The field size in spot film imaging is generally smaller than that used in general radiography.  This smaller field size reduces scatter and therefore tends to reduce dose. For the same reason, grids used in fluoroscopy generally have a lower grid ratio and therefore a smaller Bucky factor, which also leads to lower dose.  These effects tend to offset each other to a large extent.
  • 39. 39  One of the major shortcomings of conventional spot film devices is the delay involved in moving the cassette into position for exposure.  In gastrointestinal imaging, this delay can be overcome by using photofluorography.  In vascular imaging, more rapid film movement is achieved with automatic film changers.
  • 40. 40 AUTOMATIC FILM CHANGERS  The automatic film changers used in vascular imaging are also screen-film systems.  They can be found in several varieties. Some are large, floor-mounted boxes, but systems more commonly used today mount on the image intensifier  The system consists of a supply magazine for holding unexposed film, a receiving magazine, a pair of radiographic screens, and a mechanism for transferring the film.  When an exposure is required, the screens are mechanically separated, the film is pulled into place between them, and they are closed.  After the film is exposed, the screens separate again.  The film is moved to the receiver, and another film is pulled into place for the next exposure.  The number of films and filming rates must be preprogrammed for proper operation.
  • 41. 41 Photospot camera  Records the the image output of an II on a film  Film – role film/cut film(10 cm)  Advantage –1)reduction in pt exposure 2)film does not have to be changed b/w exposures 3)exposure times are shorter-motion is less likely problem 4)films can be taken more rapidly 5)possible to record & view image at same time
  • 42. 42 Framing with spot film cameras: Framing with spot film cameras:  Framing –utilisation of available area on film Framing –utilisation of available area on film  The output phosphor of II tube is round, The output phosphor of II tube is round, shape of film is square shape of film is square  4 framing patterns 4 framing patterns 1) Exact framing 1) Exact framing 2) Equal area framing 2) Equal area framing 3) mean diameter framing 3) mean diameter framing 4) total area framing 4) total area framing  Equal area framing or mean diameter framing Equal area framing or mean diameter framing is recommended for most clinical situations. is recommended for most clinical situations.
  • 43. 43
  • 44. 44 Cineflourography Cineflourography  Process of recording fluoroscopic images on Process of recording fluoroscopic images on movie (cine) film movie (cine) film  Two film sizes- 16 mm, 35 mm Two film sizes- 16 mm, 35 mm  Cine camera-components are lens, iris Cine camera-components are lens, iris diaphragm, shutter, aperture, pressure plate, pull diaphragm, shutter, aperture, pressure plate, pull down arm & film transport mechanism down arm & film transport mechanism
  • 45. 45
  • 46. 46 TV image recorders : 3) TAPE RECORDER : used for both recoding& playback • as a recorder receives video signal from camera control unit • for playback transmits the signal to one or more several TV monitors Components:1)magnetic tube 2)writing head 3) tape transport system Writing head converts an electrical signal in to magnetic field for recoding & converts magnetic signal to electric signal for replay
  • 47. 47
  • 48. 48
  • 49. 49 DIGITAL FLUOROGRAPHY Digital charge-coupled device (CCD) TV cameras are rapidly replacing conventional TV cameras in fluoroscopic systems. An analog, high-resolution (1,023-line) TV camera has a vertical resolution of about 358 line pairs. A high-resolution CCD camera with a 1,024 x 1,024 matrix will provide equivalent vertical resolution. However, the digital camera will have the same vertical and horizontal resolution, whereas the horizontal resolution of the analog camera is defined by its electronic band-pass.
  • 50. 50 For a 15-cm-diameter intensifier input, the limiting resolution of the CCD camera would be 358 lp/150 mm or 2.4 lp/mm. This result is about half the resolution of a photospot film. This resolution loss is made up for by the ability to digitally increase display contrast, reduce noise, and enhance the edges of digital images. There are several other advantages to digital photospot images. Mechanical devices are not needed for film transport. Film processing is not required. Images can be viewed immediately. The linear response of the digital system makes it very forgiving of under- or overexposure.
  • 51. 51 Fluoroscopic Equipment Configurations The basic configurations include radiography/fluoroscopy (R/F) tables with either an under-table or over-table x-ray tube and fixed C-arm, mobile C-arm, and mini C-arm R/F Units with Under-Table X-ray Tube: most common fluoroscopic equipment configuration The x-ray tube and collimator are mounted below the tabletop with the image intensifier tower mounted above the table on a carriage that can be panned over the patient. In addition to the standard fluoroscopic imaging chain, R/F systems include an overhead x-ray tube that can be used for regular radiography with a Bucky incorporated into the table. Other common features include a tilting table and image recording devices.
  • 52. 52 Under-table x-ray tube R/F system. Photograph shows an example of an Under-table x-ray tube R/F system. Photograph shows an example of an R/F table that includes a spot film device and side-mounted video camera. R/F table that includes a spot film device and side-mounted video camera.
  • 53. 53 R/F Units with Over-Table X-ray Tube R/F Units with Over-Table X-ray Tube  x-ray tube mounted over the table with the x-ray tube mounted over the table with the image intensifier below. image intensifier below.  this configuration results in increased patient this configuration results in increased patient access, which is helpful for interventional access, which is helpful for interventional procedures. procedures.  Radiography can be performed with the same x- Radiography can be performed with the same x- ray tube and a Bucky incorporated into the table. ray tube and a Bucky incorporated into the table.  The x-ray tube can be angled to acquire The x-ray tube can be angled to acquire angulated projections or tomograms. angulated projections or tomograms.
  • 54. 54 Over-table x-ray tube R/F system. Photograph shows a sample system Over-table x-ray tube R/F system. Photograph shows a sample system that can be controlled from within the procedure room with the pedestal that can be controlled from within the procedure room with the pedestal control panel (left) or from outside the room from the remote desk controls control panel (left) or from outside the room from the remote desk controls (right). (right).
  • 55. 55 SUMMARY:  Early fluroscopy was accomplished by radiologists looking directly at a fluoroscopic screen.  The image on the screen was only 0.0001 as bright as the image of a routinely viewed radiograph, so dark adaptation of eyes was required.  In 1950s the image intensifier alleviated this situation by producing an image bright enough to be viewed with cone vision.  The input phosphor of modern image intensifier is CsI; the output phosphor is zinc cadmium
  • 56. 56  Brightness gain is the product of minification gain and flux gain.  Imaging characteristics important in the evaluation of image intensifier fluoroscopy include contrast, lag and distortion.  Large field of view image intensifier tubes are available to fill special needs, such as digital and spot film angiography.  Most image intensifiers allow dual field or triple field imaging.
  • 57. 57  Output phospher image is processed by television Output phospher image is processed by television camera tube (vidicon, plumbicon, CCD) camera tube (vidicon, plumbicon, CCD)  The image is displayed on TV monitor The image is displayed on TV monitor  Standard x-ray closed-circuit television uses Standard x-ray closed-circuit television uses 525x525 format with 5-MHZ band pass 525x525 format with 5-MHZ band pass  Vertical resolution limited by scan line format, Vertical resolution limited by scan line format, horizontal resolution is a function of band pass horizontal resolution is a function of band pass  Light image is recorded by photo spot camera or Light image is recorded by photo spot camera or cine camera cine camera
  • 58. 58