OCULAR
BIOMETRY
Yonas Tilahun, MD
Objectives
 A-Scan Principles
 Steps in Biometry
 Source of Errors
 Minimizing Errors-good biometrist
 IOL calculation formula
Pre Test
 The usual frequency of A-scan biometry
probe is;
 A. 10Mhz
 B. 15Mhz
 C. 25Mhz
Pre test
2. Which technique of biometry has a
higher tendency of corneal compression
 A. Contact Biometry
 B. Immersion biometry
 C. Optical biometry
Pre Test
3. Which of the following information can
not be expected from A-scan biometry
 A. lens thickness
 B. Axial length.
 C. Keratometry
 D. AC depth
Pre Test
4. The junction between any two ocular
media of different densities and
velocities is called
 A. Gate
 B. Gain
 C. Interface
 D. Frequency
Pre test
5. How many gates do you expect in a
routine a scan measurement?
 A. 1
 B. 2
 C. 3
 D. 4
Pre Test
6. Which one of the following
measurement is considered the most
important in Iol power determination?
 A. Keratometer
 B. AC depth
 C. Axial length
 D. Lens thickness
Pre Test
7. Which one of the following uses optical
interferometry for measuring intraocular
distances?
 A. A-scan E. A&B
 B, B-scan F. C&D
 C. Lenstar
 D. IOL master
Pre Test
8. Ultrasound travels faster in lens than
in aqueous or vitreous
A. True
B. B. False
Pre Test
9. The IOL master is better than U/S for
measuring axial length for eyes with
dense cataract or media opacity
 A. True
 B. False
Pre Test
10. Optical Biometry measures axial
length from apex of the cornea to the
level of
 A. Internal limiting membrane
 B. Retinal nerve fiber level
 C. Photoreceptors
 D. Retinal Piegment Epithelium
Pre Test
11. Which one of the following is a more
reliable IOL calculation formula ?
 A. SRK T
 B. Holladay II
 C. Haigis
 D. Hoffer Q
Pre Test
12. The most common error in contact
biometry is
 A. Corneal Compression
 B. Misallignment
 C. Wrong IOL formula
 D. Wrong label
BIOMETRY
 Is a clinical procedure used to
 Meassure axial length for IOL
powercalculation
 Monitor congenital glaucoma, myopia,
nanophthalmos
 Meassure intraocular parameters like:
 AC depth
 Lens thickness
A-Scan Techniques
A-Scan Ultrasound
(PRINCIPLES)
 A-Scan what does A stand for?
 Sound wave-a vibration that propagates
as acoustic waves through Gas, liquid or
solid.
AMPLITUDE
A-Scan Ultrasound
(PRINCIPLES)
 Sound wave frequency ranges 20-
20,000hz
 Ultrasound (in audible sound) >20k hz
 A-Scan –Biometry uses ultrasound
(10mhz) to measure distances between
ocular structures using echoes of u/s
A-Scan parts
 1. Pulser
 2. Receiver
 3. Display system
Pulser and Receiver
 Comes in a probe
 Piezoelectric
Substance that
Generates US when stimulated by burst of
electricity.
The crystal converts the electric energy to
sound wave and Mechanical vibration from
echoes are converted to electrical energy and
plotted as spikes
BIOMETRY
 There is no machine brighter than a
good operator
A-Scan principle
 In A-scan biometry, one thin, parallel sound beam is emitted
from the probe tip at its given frequency of approximately 10
MHz, with an echo bouncing back into the probe tip as the sound
beam strikes each interface.
 An interface is the junction between any two media of different
densities and velocities, which, in the eye, include the anterior
corneal surface, the aqueous/anterior lens surface, the
posterior lens capsule/anterior vitreous, the posterior
vitreous/retinal surface, and the choroid/anterior scleral
surface.
 The echoes received back into the probe from each of these
interfaces are converted by the biometer to spikes arising from
baseline.
U/S Velocity
A-Scan Typical spikes
A-Scan typical spikes
Gates
 Electronic calipers on the
display..Biometers are programmed to
place..check correctness
 4 typical gaits..3 sections to be
meassured
 A. corneal spike
 B. ant lens surface spike
 C. Post lens surface spike
 D. Retinal surface spike
Measurement formula principles
 Summation of gates
 Cornea to ant lens surface (AC depth)
 Velocity through aqueous 1532m/s (D=VxT/2)
 Ant lens surface to Post lens surface (lens
thickness)
 Velocity 1641m/s
 Ant vit surface to ant retinal surface
 Velocity 1532m/s
Modes
 Phakic—3 gates displayed as above select
cataract, dens cataract etc to adjust
velocity)
 Phakic average..takes average spped of
1550m/s and 2 gates (cornea/retina) –gross
 Aphakic -2 gates (Cornea/Retina) V=1532
 Pseudophakic –lens options/ if not consider it
as PMMA
BIOMETRY
Gain
 Electrical amplification of signals
(Intensity)
 Gain knob
 Too high..picks signal fast and increases
amplitude of spikes but results in poor
resolution and poor accuracy
 Too low ..difficult to get spikes.
….Measurement
 Recommended gain 50-70
Source of Errors
 A 0.1 mm error in an average length eye will result in
about a 0.25 diopter (D) postoperative refractive
error.
 A 0.5 mm will result in approximately 1.25 D and an
error of 1.0 mm will result in approximately 2.50 D
 Longer eyes are more forgiving, with a 1.0 mm error in
an eye of 30 mm length result in 1.75 D.
 Small eyes are the least forgiving, an error of 1.0 mm
in an eye that is 22.0 mm long will result in a post-
operative error of about 3.75 D.
Source of errors
 Corneal compression-myopic shift
 Check for ac depth
 Misallignment (not perpendicular)- The
angle of incidence, which is determined
by the probe orientation to the visual
axis… hyperopic shift
 low Ant/post lens surface spikes
 Absent scleral spike
Source of Errors
Compression/Misallignments
WRONG
Source of Errors
 The shape and smoothness of each interface also
affects spike quality. Lubrication, osd Rx
 Macular pathology could adversely affect spike
quality. A perfect high, steeply rising retinal spike
may be impossible when macular pathology is present
(eg, macular edema, macular degeneration, epiretinal
membranes, posterior staphylomas).
Source of Errors
Source of errors
 Gates position..
 Not properly placed (adjust or repeat)
 Poor spikes repeat
 Dry eye, OSD
 Corneal opacity
 Squint
 AMD
 Poor patient and eye position
Pseudophakic biometry
 To check fellow eye power
 Iol exchange
 Type of iol pmma/foldable –
Reverberation artifact
Reverberation artifact
The longer chain of artifact spikes from
polymethyl methacrylate implants. The
image on the right demonstrates the
shorter chain of artifact in the vitreous
Steps in Biometry
 Calibrate and clean probe
 Patient should be seated looking straight ahead or at
the probe light (if could fix)
 Stand at the side of the patient and screen should be
placed where you can easily see it
 Apply anesthetic drop
 Align the probe to the optical axis and applanate at
the cornea apex
 Check variation in ACD, and select one with max value
 SD should be less than 0.3mm (ideally 0.06)
A Good Biometrist .must be
smarter than the machine!
 Must be able to recognize
 when readings appear abnormal
 standard dimensions of the eye.
 The average axial eye length is 23.5 mm, with a range of 22.0-24.5 mm.
 A patient can be myopic because of steep corneal curvature rather than long axial
length, and a patient can be hyperopic because of flat corneal curvature rather than
short axial length.
 Compare axial length to the precataract refractive error of the patient to ensure that
the readings appear accurate.
 The reference range of AL between the right eye and the left eye of the same patient
is within 0.3 mm, unless evidence suggests the contrary (eg, previous scleral buckling,
anisometropia, corneal transplantation, keratoconus, refractive surgery, hypotony).
 The average anterior chamber depth is 3.24 mm but varies greatly.
 The average lens thickness is 4.63 mm but this also varies, and, with cataractous
changes, the lens will increase in thickness to as much as 7.0 mm in extremely dense
cases.
A Good Biometrist
 Should realize;
 The average keratometry (K) reading is 43.0-44.0 D, with one
eye typically within a diopter of each other.
 If one eye is found to differ from the other by more than 1 D,
immediately begin researching the cause and alert the physician.
( refractive surgery, corneal transplantation, an injury with a
resultant corneal scar, or has keratoconus)
 If any of the above eye measurements is found to be unusual,
another technician should recheck the measurements and
immediately alert the physician.
Reviewing measurements
 SD of AL with in 0.06mm delete extremes
 Check corneal compression by variation of AC depth
 Ant and Post lens spikes should be nearly equal (post
slightly shorter)
 Retina spike straight and high
 Scleral spike should be seen separately from that of
the retina
 Do both eyes and if there is a difference of >0.3mm
in AL… review
IOL calculation formula
 2 variable formula ( AL and
keratometry)
 Using the correct IOL calculation formula is important for good
surgical outcomes.
 SRK Formula: P=A-2.5L-0.9K
 Current 2-variable formulas that are considered the most
accurate include the Hoffer Q, SRK/T, and Holladay I.
 Multivariable formulas have proven to be the most accurate due
to more of the eye anatomy being considered
IOL Calculation Formula
 The Haigis formula is a 3-variable equation, using not
only axial length and corneal curvature but also the
anterior chamber depth of the eye.
 The Holladay II formula is a 7-variable equation
widely thought to be the most accurate formula; it
takes into account axial length, corneal curvature,
horizontal white-to-white, anterior chamber depth,
lens thickness, precataract refractive error, and age
of the patient.
IOL Calculation Formula
 Predicting lens position is one of the most common
causes of a postoperative surprise; by taking more of
the eye anatomy into account, this can be more
accurately predicted. For average-length eyes with
average Ks, these formulas give almost identical
calculations. [3] However, when the eye is small,
formula selection is more critical. In eyes that are
less than 22 mm in length, the Hoffer Q and the
Holladay II IOL Consultant formulas are the most
accurate. For long eyes, the SRK/T and the Holladay
II IOL Consultant formulas are the most accurate.
Simple formula recommendation
Axial Length <22mm 22-24mm >24mm
Formula HofferQ SRKT,HofferQ,
HolladayII
Holladay II,
SRKII
VELOCITY CONVERSION
 Intra op You found that the patient is
aphakic hile iol was calculated with
phakic mode
 Velocity (correct)/Velocity (measured) X
Apparent Length = True Length
 E.g. 1532/1550 X 24.1 = 23.82 mm = true eye
length.
 Intraop you found pt has silicon oil
 980/1532 X Apparent Vitreous Length =
True Vitreous Length
Optical Biometers
 Current method for highly accurate axial length measurements
does not use ultrasound at all, but rather optical coherent light.
In this method, optical coherent light passes through the visual
axis and reflects back from the retinal pigment
epithelium.(internal limiting membrane as with
ultrasound/0.1mm)
 However, this method cannot be used in the event of significant
media opacity (eg, dense cataracts or corneal or vitreal opacity)
due to absorption of the light
Other Biometers
 Optical/Laser (near infra red..partial
coherence laser)
 IOL MASTER (Carl Zeiss)
 Lens star (Hagstreit)
 low coherence interferometry
 Alladin (Topcon)
Optical
Biometry Yonas.res.ppt

Contenu connexe

Tendances

Utilizing topolyzer vario & oculyzer ii for accurate refractive
Utilizing topolyzer vario & oculyzer ii for accurate refractiveUtilizing topolyzer vario & oculyzer ii for accurate refractive
Utilizing topolyzer vario & oculyzer ii for accurate refractiveMichael Mrochen
 
A scan ultrasonography
A scan ultrasonographyA scan ultrasonography
A scan ultrasonographySamuel Ponraj
 
Superior limbic keratoconjunctivitis
Superior limbic keratoconjunctivitisSuperior limbic keratoconjunctivitis
Superior limbic keratoconjunctivitisPushpraj Singh
 
Tangent screen
Tangent screenTangent screen
Tangent screensana butt
 
Pentacam analysis
Pentacam analysisPentacam analysis
Pentacam analysisNikhil Rp
 
Subjective methods of Refraction
Subjective methods of Refraction Subjective methods of Refraction
Subjective methods of Refraction Harsh Jain
 
Achieving accurate iol power calculations
Achieving accurate iol power calculationsAchieving accurate iol power calculations
Achieving accurate iol power calculationsAbdelmonem Hamed
 
Anterior segment OCT & UBM
Anterior segment OCT & UBMAnterior segment OCT & UBM
Anterior segment OCT & UBMDinesh Madduri
 
IOL POWER CALCULATION IN DIFFICULT SITUATIONS
IOL POWER CALCULATION IN DIFFICULT SITUATIONSIOL POWER CALCULATION IN DIFFICULT SITUATIONS
IOL POWER CALCULATION IN DIFFICULT SITUATIONSDr vishwanath ankad
 
Biometry instruments & equipment
Biometry instruments & equipmentBiometry instruments & equipment
Biometry instruments & equipmentDevdutta Nayak
 
Iol power calculation normal and post lasik eyes
Iol power calculation normal and post lasik eyesIol power calculation normal and post lasik eyes
Iol power calculation normal and post lasik eyesDINESH and SONALEE
 
Minimally invasive Glaucoma surgery MIGS
Minimally invasive Glaucoma surgery MIGSMinimally invasive Glaucoma surgery MIGS
Minimally invasive Glaucoma surgery MIGSankita mahapatra
 
Iol power And IOL power calculation
Iol power And IOL power calculationIol power And IOL power calculation
Iol power And IOL power calculationmdalbanuddin
 
Biometry procedure
Biometry procedureBiometry procedure
Biometry procedureAsad Zaman
 
A scan anu101 updated
A scan anu101 updatedA scan anu101 updated
A scan anu101 updatedAnurag Shukla
 
Corneal topography
Corneal topographyCorneal topography
Corneal topographySatish Jeria
 

Tendances (20)

Utilizing topolyzer vario & oculyzer ii for accurate refractive
Utilizing topolyzer vario & oculyzer ii for accurate refractiveUtilizing topolyzer vario & oculyzer ii for accurate refractive
Utilizing topolyzer vario & oculyzer ii for accurate refractive
 
A scan ultrasonography
A scan ultrasonographyA scan ultrasonography
A scan ultrasonography
 
Superior limbic keratoconjunctivitis
Superior limbic keratoconjunctivitisSuperior limbic keratoconjunctivitis
Superior limbic keratoconjunctivitis
 
Tangent screen
Tangent screenTangent screen
Tangent screen
 
Pentacam analysis
Pentacam analysisPentacam analysis
Pentacam analysis
 
Subjective methods of Refraction
Subjective methods of Refraction Subjective methods of Refraction
Subjective methods of Refraction
 
Achieving accurate iol power calculations
Achieving accurate iol power calculationsAchieving accurate iol power calculations
Achieving accurate iol power calculations
 
Anterior segment OCT & UBM
Anterior segment OCT & UBMAnterior segment OCT & UBM
Anterior segment OCT & UBM
 
Biometry
BiometryBiometry
Biometry
 
IOL POWER CALCULATION IN DIFFICULT SITUATIONS
IOL POWER CALCULATION IN DIFFICULT SITUATIONSIOL POWER CALCULATION IN DIFFICULT SITUATIONS
IOL POWER CALCULATION IN DIFFICULT SITUATIONS
 
Biometry instruments & equipment
Biometry instruments & equipmentBiometry instruments & equipment
Biometry instruments & equipment
 
Iol power calculation normal and post lasik eyes
Iol power calculation normal and post lasik eyesIol power calculation normal and post lasik eyes
Iol power calculation normal and post lasik eyes
 
Minimally invasive Glaucoma surgery MIGS
Minimally invasive Glaucoma surgery MIGSMinimally invasive Glaucoma surgery MIGS
Minimally invasive Glaucoma surgery MIGS
 
Macular hole
Macular holeMacular hole
Macular hole
 
Biometry made easy
Biometry made easy Biometry made easy
Biometry made easy
 
Iol power And IOL power calculation
Iol power And IOL power calculationIol power And IOL power calculation
Iol power And IOL power calculation
 
RECENT ADVANCES IN INTRAOCULAR LENS
RECENT ADVANCES IN INTRAOCULAR LENSRECENT ADVANCES IN INTRAOCULAR LENS
RECENT ADVANCES IN INTRAOCULAR LENS
 
Biometry procedure
Biometry procedureBiometry procedure
Biometry procedure
 
A scan anu101 updated
A scan anu101 updatedA scan anu101 updated
A scan anu101 updated
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
 

Similaire à Biometry Yonas.res.ppt

biometry for ON.ppt
biometry for ON.pptbiometry for ON.ppt
biometry for ON.pptmikaelgirum
 
OCULAR BIOMETRY AND IOL.pptx
OCULAR BIOMETRY AND IOL.pptxOCULAR BIOMETRY AND IOL.pptx
OCULAR BIOMETRY AND IOL.pptxAyienwi Wilson
 
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation Anis Suzanna Mohamad
 
Optical Biometry Measurements For Future Iol’S
Optical Biometry Measurements For Future Iol’SOptical Biometry Measurements For Future Iol’S
Optical Biometry Measurements For Future Iol’Smeikocat
 
Ocular Biometry- IOL calculation methods
Ocular Biometry- IOL calculation methodsOcular Biometry- IOL calculation methods
Ocular Biometry- IOL calculation methodsDrMadhumita Prasad
 
introduction into Biometry and it's formulas.pptx
introduction into Biometry and it's formulas.pptxintroduction into Biometry and it's formulas.pptx
introduction into Biometry and it's formulas.pptxQusaiAbusleem1
 
Biometry explanation and the used formulas.pptx
Biometry explanation and the used formulas.pptxBiometry explanation and the used formulas.pptx
Biometry explanation and the used formulas.pptxQusaiAbusleem1
 
biometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfbiometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfssuser0f453c
 
Biometry & Iol calculations
Biometry & Iol calculationsBiometry & Iol calculations
Biometry & Iol calculationsrakesh jaiswal
 

Similaire à Biometry Yonas.res.ppt (20)

A scan biometry
A scan biometryA scan biometry
A scan biometry
 
biometry for ON.ppt
biometry for ON.pptbiometry for ON.ppt
biometry for ON.ppt
 
OCULAR BIOMETRY AND IOL.pptx
OCULAR BIOMETRY AND IOL.pptxOCULAR BIOMETRY AND IOL.pptx
OCULAR BIOMETRY AND IOL.pptx
 
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
 
Biometry
BiometryBiometry
Biometry
 
Optical Biometry Measurements For Future Iol’S
Optical Biometry Measurements For Future Iol’SOptical Biometry Measurements For Future Iol’S
Optical Biometry Measurements For Future Iol’S
 
Ocular Biometry- IOL calculation methods
Ocular Biometry- IOL calculation methodsOcular Biometry- IOL calculation methods
Ocular Biometry- IOL calculation methods
 
Biometry-1, rifat.pptx
Biometry-1, rifat.pptxBiometry-1, rifat.pptx
Biometry-1, rifat.pptx
 
Biometry
BiometryBiometry
Biometry
 
introduction into Biometry and it's formulas.pptx
introduction into Biometry and it's formulas.pptxintroduction into Biometry and it's formulas.pptx
introduction into Biometry and it's formulas.pptx
 
Iol
IolIol
Iol
 
Biometry explanation and the used formulas.pptx
Biometry explanation and the used formulas.pptxBiometry explanation and the used formulas.pptx
Biometry explanation and the used formulas.pptx
 
IOL Master
IOL MasterIOL Master
IOL Master
 
A project on
A project onA project on
A project on
 
Biometry
BiometryBiometry
Biometry
 
biometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfbiometry-160212135415 (1).pdf
biometry-160212135415 (1).pdf
 
Biometry & Iol calculations
Biometry & Iol calculationsBiometry & Iol calculations
Biometry & Iol calculations
 
High Precision Biometry
High Precision BiometryHigh Precision Biometry
High Precision Biometry
 
Biometry.pptx
Biometry.pptxBiometry.pptx
Biometry.pptx
 
Biometry
Biometry Biometry
Biometry
 

Dernier

General AI for Medical Educators April 2024
General AI for Medical Educators April 2024General AI for Medical Educators April 2024
General AI for Medical Educators April 2024Janet Corral
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajanpragatimahajan3
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Celine George
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introductionMaksud Ahmed
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfJayanti Pande
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Krashi Coaching
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxheathfieldcps1
 
Student login on Anyboli platform.helpin
Student login on Anyboli platform.helpinStudent login on Anyboli platform.helpin
Student login on Anyboli platform.helpinRaunakKeshri1
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfchloefrazer622
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxiammrhaywood
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfsanyamsingh5019
 
IGNOU MSCCFT and PGDCFT Exam Question Pattern: MCFT003 Counselling and Family...
IGNOU MSCCFT and PGDCFT Exam Question Pattern: MCFT003 Counselling and Family...IGNOU MSCCFT and PGDCFT Exam Question Pattern: MCFT003 Counselling and Family...
IGNOU MSCCFT and PGDCFT Exam Question Pattern: MCFT003 Counselling and Family...PsychoTech Services
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfciinovamais
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...christianmathematics
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdfSoniaTolstoy
 

Dernier (20)

General AI for Medical Educators April 2024
General AI for Medical Educators April 2024General AI for Medical Educators April 2024
General AI for Medical Educators April 2024
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajan
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdf
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
Student login on Anyboli platform.helpin
Student login on Anyboli platform.helpinStudent login on Anyboli platform.helpin
Student login on Anyboli platform.helpin
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdf
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdf
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
IGNOU MSCCFT and PGDCFT Exam Question Pattern: MCFT003 Counselling and Family...
IGNOU MSCCFT and PGDCFT Exam Question Pattern: MCFT003 Counselling and Family...IGNOU MSCCFT and PGDCFT Exam Question Pattern: MCFT003 Counselling and Family...
IGNOU MSCCFT and PGDCFT Exam Question Pattern: MCFT003 Counselling and Family...
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
 

Biometry Yonas.res.ppt

  • 2. Objectives  A-Scan Principles  Steps in Biometry  Source of Errors  Minimizing Errors-good biometrist  IOL calculation formula
  • 3. Pre Test  The usual frequency of A-scan biometry probe is;  A. 10Mhz  B. 15Mhz  C. 25Mhz
  • 4. Pre test 2. Which technique of biometry has a higher tendency of corneal compression  A. Contact Biometry  B. Immersion biometry  C. Optical biometry
  • 5. Pre Test 3. Which of the following information can not be expected from A-scan biometry  A. lens thickness  B. Axial length.  C. Keratometry  D. AC depth
  • 6. Pre Test 4. The junction between any two ocular media of different densities and velocities is called  A. Gate  B. Gain  C. Interface  D. Frequency
  • 7. Pre test 5. How many gates do you expect in a routine a scan measurement?  A. 1  B. 2  C. 3  D. 4
  • 8. Pre Test 6. Which one of the following measurement is considered the most important in Iol power determination?  A. Keratometer  B. AC depth  C. Axial length  D. Lens thickness
  • 9. Pre Test 7. Which one of the following uses optical interferometry for measuring intraocular distances?  A. A-scan E. A&B  B, B-scan F. C&D  C. Lenstar  D. IOL master
  • 10. Pre Test 8. Ultrasound travels faster in lens than in aqueous or vitreous A. True B. B. False
  • 11. Pre Test 9. The IOL master is better than U/S for measuring axial length for eyes with dense cataract or media opacity  A. True  B. False
  • 12. Pre Test 10. Optical Biometry measures axial length from apex of the cornea to the level of  A. Internal limiting membrane  B. Retinal nerve fiber level  C. Photoreceptors  D. Retinal Piegment Epithelium
  • 13. Pre Test 11. Which one of the following is a more reliable IOL calculation formula ?  A. SRK T  B. Holladay II  C. Haigis  D. Hoffer Q
  • 14. Pre Test 12. The most common error in contact biometry is  A. Corneal Compression  B. Misallignment  C. Wrong IOL formula  D. Wrong label
  • 15. BIOMETRY  Is a clinical procedure used to  Meassure axial length for IOL powercalculation  Monitor congenital glaucoma, myopia, nanophthalmos  Meassure intraocular parameters like:  AC depth  Lens thickness
  • 17. A-Scan Ultrasound (PRINCIPLES)  A-Scan what does A stand for?  Sound wave-a vibration that propagates as acoustic waves through Gas, liquid or solid. AMPLITUDE
  • 18. A-Scan Ultrasound (PRINCIPLES)  Sound wave frequency ranges 20- 20,000hz  Ultrasound (in audible sound) >20k hz  A-Scan –Biometry uses ultrasound (10mhz) to measure distances between ocular structures using echoes of u/s
  • 19. A-Scan parts  1. Pulser  2. Receiver  3. Display system
  • 20. Pulser and Receiver  Comes in a probe  Piezoelectric Substance that Generates US when stimulated by burst of electricity. The crystal converts the electric energy to sound wave and Mechanical vibration from echoes are converted to electrical energy and plotted as spikes
  • 21. BIOMETRY  There is no machine brighter than a good operator
  • 22. A-Scan principle  In A-scan biometry, one thin, parallel sound beam is emitted from the probe tip at its given frequency of approximately 10 MHz, with an echo bouncing back into the probe tip as the sound beam strikes each interface.  An interface is the junction between any two media of different densities and velocities, which, in the eye, include the anterior corneal surface, the aqueous/anterior lens surface, the posterior lens capsule/anterior vitreous, the posterior vitreous/retinal surface, and the choroid/anterior scleral surface.  The echoes received back into the probe from each of these interfaces are converted by the biometer to spikes arising from baseline.
  • 26. Gates  Electronic calipers on the display..Biometers are programmed to place..check correctness  4 typical gaits..3 sections to be meassured  A. corneal spike  B. ant lens surface spike  C. Post lens surface spike  D. Retinal surface spike
  • 27. Measurement formula principles  Summation of gates  Cornea to ant lens surface (AC depth)  Velocity through aqueous 1532m/s (D=VxT/2)  Ant lens surface to Post lens surface (lens thickness)  Velocity 1641m/s  Ant vit surface to ant retinal surface  Velocity 1532m/s
  • 28. Modes  Phakic—3 gates displayed as above select cataract, dens cataract etc to adjust velocity)  Phakic average..takes average spped of 1550m/s and 2 gates (cornea/retina) –gross  Aphakic -2 gates (Cornea/Retina) V=1532  Pseudophakic –lens options/ if not consider it as PMMA
  • 30. Gain  Electrical amplification of signals (Intensity)  Gain knob  Too high..picks signal fast and increases amplitude of spikes but results in poor resolution and poor accuracy  Too low ..difficult to get spikes. ….Measurement  Recommended gain 50-70
  • 31. Source of Errors  A 0.1 mm error in an average length eye will result in about a 0.25 diopter (D) postoperative refractive error.  A 0.5 mm will result in approximately 1.25 D and an error of 1.0 mm will result in approximately 2.50 D  Longer eyes are more forgiving, with a 1.0 mm error in an eye of 30 mm length result in 1.75 D.  Small eyes are the least forgiving, an error of 1.0 mm in an eye that is 22.0 mm long will result in a post- operative error of about 3.75 D.
  • 32. Source of errors  Corneal compression-myopic shift  Check for ac depth  Misallignment (not perpendicular)- The angle of incidence, which is determined by the probe orientation to the visual axis… hyperopic shift  low Ant/post lens surface spikes  Absent scleral spike
  • 35.
  • 36.
  • 37. WRONG
  • 38. Source of Errors  The shape and smoothness of each interface also affects spike quality. Lubrication, osd Rx  Macular pathology could adversely affect spike quality. A perfect high, steeply rising retinal spike may be impossible when macular pathology is present (eg, macular edema, macular degeneration, epiretinal membranes, posterior staphylomas).
  • 40. Source of errors  Gates position..  Not properly placed (adjust or repeat)  Poor spikes repeat  Dry eye, OSD  Corneal opacity  Squint  AMD  Poor patient and eye position
  • 41. Pseudophakic biometry  To check fellow eye power  Iol exchange  Type of iol pmma/foldable – Reverberation artifact
  • 42. Reverberation artifact The longer chain of artifact spikes from polymethyl methacrylate implants. The image on the right demonstrates the shorter chain of artifact in the vitreous
  • 43. Steps in Biometry  Calibrate and clean probe  Patient should be seated looking straight ahead or at the probe light (if could fix)  Stand at the side of the patient and screen should be placed where you can easily see it  Apply anesthetic drop  Align the probe to the optical axis and applanate at the cornea apex  Check variation in ACD, and select one with max value  SD should be less than 0.3mm (ideally 0.06)
  • 44. A Good Biometrist .must be smarter than the machine!  Must be able to recognize  when readings appear abnormal  standard dimensions of the eye.  The average axial eye length is 23.5 mm, with a range of 22.0-24.5 mm.  A patient can be myopic because of steep corneal curvature rather than long axial length, and a patient can be hyperopic because of flat corneal curvature rather than short axial length.  Compare axial length to the precataract refractive error of the patient to ensure that the readings appear accurate.  The reference range of AL between the right eye and the left eye of the same patient is within 0.3 mm, unless evidence suggests the contrary (eg, previous scleral buckling, anisometropia, corneal transplantation, keratoconus, refractive surgery, hypotony).  The average anterior chamber depth is 3.24 mm but varies greatly.  The average lens thickness is 4.63 mm but this also varies, and, with cataractous changes, the lens will increase in thickness to as much as 7.0 mm in extremely dense cases.
  • 45. A Good Biometrist  Should realize;  The average keratometry (K) reading is 43.0-44.0 D, with one eye typically within a diopter of each other.  If one eye is found to differ from the other by more than 1 D, immediately begin researching the cause and alert the physician. ( refractive surgery, corneal transplantation, an injury with a resultant corneal scar, or has keratoconus)  If any of the above eye measurements is found to be unusual, another technician should recheck the measurements and immediately alert the physician.
  • 46. Reviewing measurements  SD of AL with in 0.06mm delete extremes  Check corneal compression by variation of AC depth  Ant and Post lens spikes should be nearly equal (post slightly shorter)  Retina spike straight and high  Scleral spike should be seen separately from that of the retina  Do both eyes and if there is a difference of >0.3mm in AL… review
  • 47. IOL calculation formula  2 variable formula ( AL and keratometry)  Using the correct IOL calculation formula is important for good surgical outcomes.  SRK Formula: P=A-2.5L-0.9K  Current 2-variable formulas that are considered the most accurate include the Hoffer Q, SRK/T, and Holladay I.  Multivariable formulas have proven to be the most accurate due to more of the eye anatomy being considered
  • 48. IOL Calculation Formula  The Haigis formula is a 3-variable equation, using not only axial length and corneal curvature but also the anterior chamber depth of the eye.  The Holladay II formula is a 7-variable equation widely thought to be the most accurate formula; it takes into account axial length, corneal curvature, horizontal white-to-white, anterior chamber depth, lens thickness, precataract refractive error, and age of the patient.
  • 49. IOL Calculation Formula  Predicting lens position is one of the most common causes of a postoperative surprise; by taking more of the eye anatomy into account, this can be more accurately predicted. For average-length eyes with average Ks, these formulas give almost identical calculations. [3] However, when the eye is small, formula selection is more critical. In eyes that are less than 22 mm in length, the Hoffer Q and the Holladay II IOL Consultant formulas are the most accurate. For long eyes, the SRK/T and the Holladay II IOL Consultant formulas are the most accurate.
  • 50. Simple formula recommendation Axial Length <22mm 22-24mm >24mm Formula HofferQ SRKT,HofferQ, HolladayII Holladay II, SRKII
  • 51. VELOCITY CONVERSION  Intra op You found that the patient is aphakic hile iol was calculated with phakic mode  Velocity (correct)/Velocity (measured) X Apparent Length = True Length  E.g. 1532/1550 X 24.1 = 23.82 mm = true eye length.  Intraop you found pt has silicon oil  980/1532 X Apparent Vitreous Length = True Vitreous Length
  • 52. Optical Biometers  Current method for highly accurate axial length measurements does not use ultrasound at all, but rather optical coherent light. In this method, optical coherent light passes through the visual axis and reflects back from the retinal pigment epithelium.(internal limiting membrane as with ultrasound/0.1mm)  However, this method cannot be used in the event of significant media opacity (eg, dense cataracts or corneal or vitreal opacity) due to absorption of the light
  • 53. Other Biometers  Optical/Laser (near infra red..partial coherence laser)  IOL MASTER (Carl Zeiss)  Lens star (Hagstreit)  low coherence interferometry  Alladin (Topcon)