National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
1. National Ocular Biometry
Course 2015
An echoing presentation
Prepared by:
Anis Suzanna Binti Mohamad
Rohaila Binti Ariffin
Pegawai Optometri U41
Hospital Sultanah Bahiyah, Alor Setar
2. COURSE OBJECTIVES:
Understand the basic principal of biometry
Perform, define the discuss the pathology that would require the
biometry test
Interpret the result of biometry test
Measure axial length
Name four ocular structures that reflect ultrasound echoes
Able to perform biometry test
Able to perform topography, discuss corneal contours of normal,
pathological and post-surgical cornea
Take pre-operative testing and intraocular lens power calculation
method to the next level of accuracy and efficiency
Utilize standardize diagnostic ultrasonography as a valuable
adjuvant to current clinical evaluation.
3. Cataract surgery
• What is cataract surgery?
– Precise biometry is essential for accurate outcomes in cataract and
refractive surgeries.
– The measurement of axial length by ultrasound was the gold
standard for many years.
– With the introduction of optical biometry in the US in the year
2000, this technology has become more & more popular & is now
the most common method for the measurement of axial length.
– Optical biometry uses a partially coherent wave that has approx. 9x
the resolution of a 10 MHz sound wave, making the measurement
of axial length very precise.
– It helps to avoid operator variations in measurement.
– Also, to increase the accuracy as contact with cornea is not needed
(eliminates compression to cornea).
– Since Optical biometry measures to the centre of macula, it gives
the refractive AXL vs the anatomical AXL achieved with ulrasound.
9. 2. A-Scan Biometry
• Measurement of Axial Eye Length by Ultrasound
• Average Axial Length of Normal Eye 23.06 mm (Majority :
22.0 to 24.5 mm)
• Accuracy of AL measurement using A-scan ultrasound is +
0.1 mm
• Difference in AL measurement between both eyes + 0.3 mm
• Values are 0.14 to 0.36 mm longer with immersion
technique than with contact method
14. Ultrasound biometry
Ultrasound is
produced in the
ultrasonographic
probe by the
oscillation of a
pizoelectric
crystal
Converting
electrical energy
into mechanical
energy
The probe emits
and receives
pulsed signal
Reflectivity vs
time is displayed
for the single
direction in which
the probe is
pointing
This value then
be converted to
milimitres
22. Potential Sources of Error
with Contact Method
1. Corneal Compression
2. Fluid Excess
3. Misalignment of Sound Beam
4. Inappropriate Eye type
Error caused by 1 mm Corneal Compression:
Average eye 2.5 D
Long eye 1.75 D
Short eye 3.75 D
31. Gain Setting
• Initially high gain setting should be used to assess the overall
appearance of the echogram , then gain should be reduced to a
medium level to improve resolution of spikes .
• Error can occur when the gain is set too high or too low .
Very high gain : short reading
Very low gain : long reading
32.
33. Use of average sound velocity ,although
sufficient in normal phakic eye , may
result in slight error when the lens is
inordinately thin or thick or when the
eye is very short or very long .
34. • The use of individual sound velocity may provide
more consistent and accurate AL reading .
35.
36. • Manual measurement mode is better to help
ensure alignment of sound beam .
Aphakia & Pseudophakia
37. Posterior Staphyloma
• Posterior staphylomas often causes an irregular shape of the ocular wall
resulting in an inability to display a distinct , high retinal spike , leading to a
significant error in A-scan measurement .
• Deepest portion of the staphyloma may be located eccentric to macula thus the
measurement may be longer than true AL along the visual axis .
• B-scan can be used to demonstrate the shape of posterior ocular wall and the
relationship of macula to the staphyloma .
• Probes with fixation light are preferable
42. Edema
DMS
Tumor
Macular Lesions
RD
The presence of an elevated macular lesion may
prevent the display of a distinct retinal spike and
often causes a shortened AL measurement .
47. Dense Cataract
• Strong sound attenuation produced by a very dense cataract can
significantly impair the ability to display spikes from the various
interfaces along the visual axis .
• Maximum gain setting may be required to obtain spikes of
sufficient height from the posterior lens capsule and retina .
• Semiautomatic mode should be used in eyes with dense cataract.
48.
49. Silicone Oil
• Sound velocity in silicone oil
1040 m/s 5000 cs
980 m/s 1000cs
• This low sound velocity can result in
pronounced sound attenuation and difficulty
in identifying the retinal spikes .
50.
51. • If proper sound velocity are not used , erroneously long AL
measurement will be obtained .
• For accurate AL measurement ,various ocular components
should be measured separately with appropriate sound
velocity .
• The least preferred method is use of average sound velocity.
Average sound velocity in eyes with average length (23.5 mm):
1,139 m/s phakic eye
1,052 m/s aphakic eye
• Due to strong sound attenuation AL measurement often can
not be obtained from an eye containing emulsified silicone oil .
58. Keratometry
A second person should confirm measurements prior to A-scan
ultrasonography if: The corneal power is less than 40.0 diopters, or
greater than 47.0 diopters.
If there has been prior keratorefractive surgery. In this case the corneal
power will need to be estimated by either the historical, or the contact lens
method.
The average corneal power difference between the two eyes is greater
than 1.00 diopter.
The patient cannot fixate, as seen with a mature cataract, or macular hole.
The amount of corneal astigmatism by keratometry, or topography,
correlates poorly with the amount of astigmatism on the most recent
manifest refraction.
The corneal diameter is less than 11.00 mm.
There is any problem with patient cooperation, or understanding.
59. Immersion A-scan Ultrasonography
A second person should re-measure both eyes if: The axial length is less
than 22.00 mm, or greater than 25.00 mm in either eye.
The axial length is greater than 26.0 mm, and there is a poor retinal spike,
or wide variability in the readings.
There is a difference in axial length between the two eyes of greater
than 0.33 mm that cannot be correlated with the patient's oldest
refraction.
Axial length measurements do not correlate with the patient's refractive
error. In general, myopes should have eyes longer than 24.0 mm and
hyperopes should have eyes shorter than 24.0 mm. Exceptions to this
rule involve steep, or flat corneas. Be sure to use the oldest refractive
data.
There is difficulty obtaining correctly positioned, high, steeply rising
echoes, or wide variability in individual axial length readings for either
eye.
60. Intraocular Lens Power
A second person should repeat the axial length measurements,
keratometry readings and re-run the IOL power calculations for both eyes
if: The IOL power for emmetropia is greater than 3.00 diopters different
than anticipated.
There is a difference in IOL power of greater than 1.00 diopter between the
two eyes.
If the patient has had prior keratorefractive surgery and the calculated IOL
power for standard phacoemulsification is less than +20.0 D or greater than
+23.0 D.
65. Theoretical Formula
These formulas contain many
assumptions including values of
postop ACD , refractive index of
cornea and ocular humors , retinal
thickness
73. SRK II
A1 = A + 3 AL < 20mm
A1 = A + 2 AL 20-21
A1 = A + 1 AL 21-22
A1 = A AL 22-24.5
A1 = A – 0.5 AL >24.5
74. SRK/T
• It is a nonlinear theoretical optical formula
empirically optimized for postop ACD, retinal
thickness , corneal refractive index . It
combines advantages of theoretical and
regression formulas .
79. There are currently three IOL constants in use: The
SRK/T formula uses an "A-constant."
The Holladay 1 formula uses a "Surgeon Factor."
The Holladay 2 formula, and the Hoffer Q formula,
both use an "Anterior Chamber Depth." aka: ACD.
80. d = the effective lens position, where ...
d = a0 + (a1 * ACD) + (a2 * AL)
Haigis Formula
81. * The a0 constant basically moves the curve up,
or down, in much the same way that the A-
constant, Surgeon Factor, or ACD does for the
Holladay 1, Holladay 2, Hoffer Q and SRK/T
formulas.
* The a1 constant is tied to the measured
anterior chamber depth.
* The a2 constant is tied to the measured axial
length. The way the a0, a1 and a2 constants are
derived is by generating a set of surgeon, and
IOL-specific
82. AL < 19 mm (<0.1%)
Holladay 2
AL 19-22 mm (8%)
Holladay 2 , Hoffer-Q
AL 22-24.5 mm (72%)
SRK II , Hoffer-Q ,Holladay 1
AL 24.5-26 mm (15%)
Holladay 1 , Hoffer-Q
AL > 26 mm ( 15%)
SRK/T
Formula Choice
A-scan ultrasound biometry, commonly referred to as an A-scan, is routine type of diagnostic test used in ophthalmology. The A-scan provides data on the length of the eye, which is a major determinant in common sight disorders.