1. Regulated Bioanalysis:
Science and Art of Meeting the Requirements
India Technology Seminar
Boston, 29 October – November 1, 2007
Dr. Bhaswat S. Chakraborty
30.10.2007
2. Bioanalytical Guiding Principles
Used in quantitative determination of drugs and
their metabolites in biological fluids
BE, PK, TK and Meatbolomics studies
Often the Study is as good as the bioanalytical
Refinement of all aspects in last 20 years
Implemented as though regulations
More and more demanding
From 1980s to date
US FDA setting the standards
3. Key Guiding Documents & Publications
Shah VP et al. Analytical methods validation…. Eur J
Drug Metab Pharmacokinet. 1991
Draft Guidance on Bioanalytical: USFDA. 1999
Bioanalytical Workshop. 2000
Guidance for Industry: Bioanalytical: USFDA. 2001
DeSilva B et al. Validation of ligand-binding assays…
macromolecules. Pharm Res. 2003
Draft Guidance on Safety Testing of Drug Metabolites:
USFDA. 2005
FDA-AAPS Bioanalytical Workshop. 2006
4. Scientific Questions behind Guidance
Does the method measure precisely and accurately the
analyte(s)?
If so, how was it validated? How complete and accurate is
the validation?
Were adequate CC and QC used during incurred sample
runs? Was each and every batch acceptable?
Were the incurred samples (& repeats) accurately and
reproducibly measured and were they stable at all times?
Is the documentation comprehensive and reconstructive?
5. Non-Chromatographic Assays: Issues
E.g., RIA & other ligand binding assays
Selectivity
Specific
Non-specific or Matrix
Often analyte is also present endogenously
Quantification
Non-linear CC
More points than Chromatographic CC
Accuracy is questionable, so replicates required
Cross validation
Stability is a complex issue
Stabilty of receptor binding activity as well?
6. Non-Chromatographic Assays: Validation
Deplete the matrix of the analyte or employ a
“surrogate” matrix
5 or more validation concentrations
Intra- and interbatch variation
LLOQ, ~3 times LLOQ, mid [geometric mean], 75% of
ULOQ,and ULOQ
At least 2 independent determinations
Interbatch imprecision (%CV) and inaccuracy (%RE):
±20% (25% at LLOQ and ULOQ).
Total target error (imprecision and inaccuracy) ≤30%
(40% at LLOQ &ULOQ)
7. Non-Chromatographic Assays: Batch Runs
CCs
At least 75% of the standard points should be within 20% of
nominal concentrations (25% at LLOQ)
Does not apply to “ anchor calibrators, ” those outside the
anticipated validation range and improve “ sigmoidal ” fitting
QCs
LQC, MQC, HQC in duplicate
4 – 6-20 rule
4 of the 6 QCs must be within 20% of nominal
At least one QC per conc. meets this criterion
If additional QCs are used, then 50% of them be within 20%
of nominal
8. Non-Chromatographic Assays:
Cross Validation
E.g., RIA vs LC-MS
The comparisons can be done both ways.
Cross-validation with spiked matrix and subject
samples be conducted at each site or laboratory
Interlaboratory reliability
if analyses within a single study are conducted at more
than one site or lab
If data generated using different analytical techniques
9. Chromatographic and Hyphenated Assays
HPLC, UPLC, GC, LC-MS
More selective and accurate than LBAs
Used for small molecules
Drugs, metabolites, toxicants
Linear response
LC-MS: highly sensitive
High capacity
Highly reproducible response stability
10. LC-MS or LC-MS-MS
LC coupled with one or mass analyzers
Most popular, very accurate, high throughput
Quadrupole mass analyzers: most widely used
Aligns different m/z according to their retention times
Time-of-flight mass analyzers
Analyzes different m/z according to the different times taken
to traverse a fixed distance
FTICR mass analyzers
Applies a radiofrequency voltage at the same cyclotron
frequency of a m/z, the later latter is moved at a larger radius
than the RF
11. Validation Batches
Analyze at least 3 batches for accuracy and
precision
At least 1 validation batch should be made as
large as the largest anticipated sample analysis
batch
12. Calibration Samples & Acceptance Criteria
CC concentrations
A minimum of 6 non-zero standards
Matrix blank: Matrix sample without internal standard
Zero standard: Matrix sample with internal standard
Acceptance criteria
Intra- and inter-batch imprecision (%CV) and inaccuracy
(%RE) ≤15% except at LLOQ where up to 20% can be
allowed
14. QC Samples & Acceptance Criteria
QC concentrations
LLQC: About 3 times the LOQ; MQC: ~geometric
mean of LQC & HQC; HQC: ~70% to 85% of
ULOQ; Dilution QC: sufficient to cover highest
anticipated dilution
Acceptance criteria
Intra- and inter-batch imprecision (%CV) and
inaccuracy (%RE) ≤15% except at LLOQ where up
to 20% can be allowed
16. Selectivity (Non-interference from Matrix)
Review noninterference in at least 6 sources of
matrix for non-MS assays
For MS assays determine MFs in 6 sources if
the nonisotopically labeled IS is used
If isotopically labeled IS is used, demonstrate
that IS normalized MF is close to unity
Interference in blank matrix ≤20% of LOQ
17. Example
Matrix effect accuracy QC %nominal value 100.70
Hemolysis effect accuracy QC % nominal value
For LQC is 101.41 and
For HQC is is 100.74
Hemolysis effect precision QC Coefficient of variation
For LQC is 9.57 and for
HQC is 5.65
18. Reproducibility of the Method
Precision and accuracy:
Inter-run precision and accuracy of the QCs
Second column or instrument verification:
Reproducibility of the method on an alternate column or
instrument
A batch of precision and accuracy samples is analyzed on a
different column or instrument on one of the days of
validation. Good practice but not mandatory
Reproducibility using incurred samples:
Sample availability can be an issue
20. Analyte Recovery
Detector response of extracted vs direct input
Extent of recovery of analyte and IS
Demonstrate that it is
Consistent
Precise
Reproducible
Determine recovery at 3 concentration levels
Does not have to be close to100%
21. Example: Drug Recovery
Low QC
Extracted response Unextracted response
325712 605788
330450 611672
332840 621078
309710 607364
345282 606242
375431 602309
N 6 6
Mean 336570.83 609075.50
SD (±) 22257 6608
CV (%) 6.61 1.08
Mean recovery (%) 55.26
22. Stability of Samples
Stock solution:
Minimum of 6 hours at room temp. & Fridge temp. for 24 hr
Postpreparative (extracted samples/autosampler tray):
Longest time from preparation through analysis. Vs. fresh
standards
Benchtop:
At ambient temp. (or processing temp.) – for extraction duration
(typically ~4-24 hr)
Freeze-thaw:
QC samples at least 2 conc., 3 cycles, completely thawed,
refrozen for 24/12 hr, at anticipated storage temp.
Long-term: Can be postvalidation
For longest time – collection to for any sample (3 aliquots; low
and high conc. with fresh standard curves); assess vs. nominal
23. Example
Auto sampler stability QC Mean % Change
(~49 hr) LQC is –4.28 and for
HQC is –2.35
Bench Top Stability in matrix at room temperature QC Mean Change for
(~ 25.5 Hrs.) LQC is 4.56 for
HQC is 3.73
Freeze and thaw stability QC Mean % Change for
At –30±5°C after 3 cycles LQC is 5.60 and for
HQC is 3.47
Dry extract stability QC Mean % Change for
At –30±5°C after (Approx 51 Hrs.) LQC is -6.75 and for
HQC is 3.16
Short-term stability in solution at room temperature Mean % change is -1.10
(~16.5 Hrs.)
Short-term stability in solution at refrigerator Mean % change is –0.48
(~ 96 Hrs.)
24. Dilution Integrity for Concentrations >ULQ
One or more additional QC >>ULQ prepared
and diluted with blank matrix to bring the
concentration to within the calibration range
and then analyzed
The acceptance criteria for the diluted QC are
the same as for other QCs
Intra- and inter-batch imprecision (%CV) and
inaccuracy (%RE) ≤15%
25. Example
Dilution integrity accuracy (as QC %nominal value)
For 1/5th dilution factor is 96.13
For 1/10th dilution factor is 111.49
Dilution integrity precision (as QC %CV)
For 1/5th dilution factor is 6.82 and
For 1/10th dilution factor is 4.76
26. MS Techniques: Matrix Factor
Ion Enhancement
Syringe
Pump MS
Drug ISTD
Autosampler
Ion Suppression
27. MS Techniques: Matrix Factor (MF)
A quantitative measure of the matrix effects due to
suppression or enhancement of ionization in an MS
detector
MFs can be determined for the analyte and the IS
Ratio is called IS-normalized MF for the analyte
IS-normalized MFs using stable isotope labeled IS
Usually close to unity for bioanalytical samples
MF or IS-normalized MF be determined in 6
independent lots of matrices with desirable CV <15%
(not for stable isotope labeled IS using methods)
28. PK Repeats
SOP
All PK repeats are
chosen before interim
analysis of data
Based on sound PK and
Bioanalytical principles
without any bias Ct
Original value is retained
if the repeat value is
within 15%
Repeats with one or
more “correct” estimates
t
30. Data
Conc. (ng/ml) Peak Area Ratio
1501 1.54
1002.5 0.91
703 0.77
249.5 0.25
99.5 0.13
30 0.04
10 0.01
5 0.01
31. Comparison: LS vs. MLE
Least Squares MLE
Intercept 0.009557334 0.009426942
Slope 0.00099529 0.001014858
Std. Error 0.023385281 0.029742873
df 6 6
…very similar in this case but can be very different in other cases
32. Least Squares vs. MLE
1.8
MLE
1.6
1.4
1.2
Peak Area Ratio
Least Squares
1
0.8
0.6
0.4
0.2
0
0 200 400 600 800 1000 1200 1400 1600
Conc. (ng/ml)
33. Comparison: LS vs. MLE
An Example where differences can be substantial
X Y Least Squares MLE
5.66 59
4.63 43 Intercept 3.192698019 10.36237254
5.21 41 Slope 7.424965403 6.621244357
6.01 47 Std. Error 0.157779577 3.300103342
6.17 53 df 10 10
4.82 43
5.08 46
5.51 45
4.95 41
5.13 42
4.66 39
4.74 45
36. Current Guidelines are based on
Expertise
Correct understanding of the isolation of the analyte from
matrix, metabolites & noise
Relevant approach to method development & validation, e.g.,
target bias, establishment of method characteristics
Ethics
Non-deviation from protocol, QA, QC, audit trail
Focus
Specificity, accuracy, precision, reproducibility & stability
37. You are Successful when Your Data
Enable the regulators know exactly what you
know & come to the same conclusions
Provide evidence of ‘correct’ determination of
unknown concentrations
Present validation and analytical report
Are traceable, accounted for and of proven
integrity
Are complete
38. In addition ..
Be aware of
CC algorithms, their strengths & pitfalls
Peak smoothing techniques & expertise
Peak instability, non-reproducibility of a specific assay
Having correct & enough SOPs
Long term sample storage
Assays where replication of clinical samples assay may be
necessary
Regulatory queries & how to respond to them