TI191C/07/en/07.08
71078186
Technical Information
Orbipac CPF81/CPF82 and CPF81D/CPF82D
Compact pH/ORP electrodes, analog or with digital Memosens
technology
For installation or immersion in industrial water and wastewater
Application
• Industrial and municipal wastewater treatment
• Water conditioning
• Condensate cleaning
Your benefits
• Suitable for flow and immersion installation
• Suitable for measurement within pH range 0 to 14 pH and
temperature range 0 to 110 °C (32 to 230 °F)
• Analog pH electrodes with or without integrated temperature
sensor, digital pH sensors with integrated temperature sensor
• With patented KNO3 electrolyte bridge for better protection
against electrode poisons like S2- or CN- ions
• Protection guard against damage
• Flat membrane suitable for high flow rates and fibrous media
• Optional integrated preamplifier for low-impedance and
noise-free measuring value transmission
• NPT ¾" threaded connection, top and bottom
Further benefits offered by Memosens technology
• Maximum process safety through contactless inductive signal
transmission
• Data safety through digital data transmission
• Easy handling thanks to storage of sensor-specific data in the
sensor
• Predictive maintenance possible thanks to registration of
sensor load data in the sensor
CPF81/82, CPF81D/82D
2 Endress+Hauser
Function and system design
Measuring principle pH measurement
The pH value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. The membrane
glass of the electrode supplies an electrochemical potential which is dependent upon the pH value of the
medium. This potential is generated by the selective penetration of H+ ions through the outer layer of the
membrane. An electrochemical boundary layer with an electric potential forms at this point. An integrated
Ag/AgCl reference system serves as reference electrode.
The transmitter converts the measured voltage into the corresponding pH value using the Nernst equation.
ORP measurement
The ORP potential is a unit of measurement for the state of equilibria between oxidizing and reducing
components of a medium. ORP potential is measured similarly to the pH value. A platinum electrode is used
instead of pH-sensitive membrane glass. Analog to the pH measurement, an integrated Ag/AgCl reference
system is used as a reference electrode.
Important properties of
CPF81D/CPF82D
Maximum process safety
The inductive and non-contacting measured value transmission of Memosens guarantees maximum process
safety and offers the following benefits:
• All problems caused by moisture are eliminated.
– The plug-in connection is non-contacting and therefore free from corrosion.
– Measured value distortion from moisture is not possible.
– The plug-in system can even be connected under water.
• The transmitter is galvanically decoupled from the medium. The result: No more need to ask about
"symmetrically high-impedance" or "unsymmetrical" (for pH/ORP measurement) or an impedance
converter.
• EMC safety is guaranteed by screening measures for the digital measured value transmission.
Data safety through digital data transfer
The Memosens technology digitalizes the measured values in the sensor and transfers them to the transmitter
contactlessly and free from interference potential. The result:
• An automatic error message is generated if the sensor fails or the connection between sensor and transmitter
is interrupted.
• The availability of the measuring point is dramatically increased by immediate error detection.
Easy handling
Sensors with Memosens technology have integrated electronics that allow for saving calibration data and
further information such as total hours of operation and operating hours under extreme measuring conditions.
When the sensor is mounted, the calibration data are automatically transferred to the transmitter and used to
calculate the current measured value. Storing the calibration data in the sensor allows for calibration away from
the measuring point. The result:
• Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive
maintenance is possible.
• The sensor history can be documented on external data carriers and evaluation programs at any time. Thus,
the current application of the sensors can be made to depend on their previous history.
• The transmitter does not need to be installed close to the measuring point but can be placed in the control
room.
Communication with the transmitter
Always connect digital sensors to a transmitter with Memosens technology. Data transmission to a transmitter
for analog sensors is not possible.
CPF81/82, CPF81D/82D
Endress+Hauser 3
Data storage CPF81D
The digital sensors are able to store the following system data in the sensor.
• Manufacturing data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Calibrated slope at 25 °C (77 °F)
– Calibrated zero point at 25 °C (77 °F)
– Temperature offset
– Number of calibrations
– Serial number of the transmitter used for the last calibration
• Application data
– Temperature application range
– pH application range
– Date of first commissioning
– Maximum temperature value
– Total operating hours
– Operating hours at temperatures above 80 °C (176 °F) and 100 °C (212 °F)
– Operating hours at very low and very high pH values (Nernst voltage below -300 mV, above +300 mV)
– Glass membrane impedance
These system data can be displayed with Mycom S CPM153 and Liquiline M CM42 transmitters.
CPF82D
The digital sensors are able to store the following system data in the sensor.
• Manufacturing data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Calibrated offset (operating mode "mV")
– % slope (operating mode "%")
– Number of calibrations
– Serial number of the transmitter used for the last calibration
• Application data
– Temperature application range
– ORP application range
– Date of first commissioning
– Operating hours
These system data can be displayed with Mycom S CPM153 or Liquiline M CM42 transmitters.
CPF81/82, CPF81D/82D
4 Endress+Hauser
Input
Measured values
Measuring range
Caution!
Please note the process operating conditions.
Cable specification CPF81/82 with fixed cable
a0010182-en
Measuring cable
A Fixed cable for CPF81 without temperature sensor and CPF82
B Fixed cable for CPF81 with temperature sensor
* PM is only connected for sensor versions with internal solution ground (CPF81-xxx2xx).
CPF81/82 with TOP68 plug-in head
a0003303-en
CPK9 special measuring cable
CPF81D/82D
a0003350
CYK10 Memosens data cable
CPF81, CPF81D: pH value
Temperature
CPF82, CPF82D: ORP potential
CPF81/81D
Electrode version LH
pH: 0 to 14 pH
Temperature: 0 to 110 °C (32 to 230 °F)
Electrode version NN
pH: 0 to 14 pH (reduced accuracy for 11 to 14 pH)
Temperature: 0 to 80 °C (32 to 176 °F)
CPF82/82D
ORP potential: -1500 mV to +1500 mV
BK
BK
GN/YE
BN
BK
BK
GN
WH
YE
GN/YE
BN
GN
WH
YE
A B
Reference
pH/ORP
Temperature
Temperature
Temperature
Outer screen
PM / n. c.*
pH/ORP
+ Pt 100
PM Coax BK
WH
GN
BN
Screen connection
WH
YE
Ref.
pH
Temp
Temp.
Temp.
PM
CPF81/82, CPF81D/82D
Endress+Hauser 5
Installation
Installation instructions The electrode is suitable for flow and immersion installation.
Note!
Make sure to follow the installation instructions in the Operating Instructions of the used assembly.
Environment
Ambient temperature Caution!
Danger of frost damage
Do not use the electrodes at temperatures below 0 °C (32 °F.)
Storage temperature 0 to 50 °C (32 to 122 °F)
Ingress protection
Process
Process temperature
Process pressure 0 to 10 bar @ 80 °C (0 to 145 psi @ 176 °F)
a0010183
Flow installation
a0010184
Immersion installation with Flowfit CYA611
Fixed-cable version: IP 67
TOP68 plug-in head: IP 68 (1 m (3.28 ft) water column, 50 °C (122 °F), 168 h)
Memosens plug-in head: IP 68 (10 m (32.8 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)
CPF81/81D
Electrode version LH: 0 to 110 °C (32 to 230 °F)
Electrode version NN: 0 to 80 °C (32 to 176 °F)
CPF82/82D 0 to 80 °C (32 to 176 °F)
CPF81/82, CPF81D/82D
6 Endress+Hauser
Pressure temperature load
curve
a0010185
Glass impedance 150 MΩ @ 25 °C (77 °F )
Minimum conductivity min. 50 μS/cm
Mechanical construction
Design, dimensions
CPF81/82
60 80 100
20 40
0
2
4
6
8
10
140 180 220
60 100
32
40
120
°C
°F
bar
psi
50.8
176
110
230
80
0 0
3.5
145
a0010179
Fixed-cable version, short shaft,
protection guard
a0010180
CPF81 with TOP68 plug-in head,
short shaft, flat membrane
a0010181
Fixed-cable version, long shaft,
protection guard
1
2
3
4
Fixed cable
NPT ¾" thread
Across flats AF 26
Protection guard
1
2
3
4
TOP68 plug-in head
NPT ¾" thread
Across flats AF 26
Flat membrane
1
2
3
4
Fixed cable
NPT ¾" thread
Across flats AF 26
Protection guard
Ø 22 (0.87)
23
(0.91)
150
(5.91)
mm (inch)
1
2
3
4
2
Ø 22 (0.87)
12.5
(0.49)
140
(5.51)
mm (inch)
1
2
3
4
2
Ø 22 (0.87)
58.5
(2.3)
150
(5.91)
mm (inch)
1
2
3
4
2
CPF81/82, CPF81D/82D
Endress+Hauser 7
Design, dimensions
CPF81D/82D
Material
Process connection NPT ¾" thread
Integrated preamplifier
(if ordered)
Note!
With preamplifier versions, the sensor check function (SCS) of the transmitter is not available and should be
turned off.
a0010174
CPF82D, short shaft, protection guard
a0010171
CPF81D, flat membrane
a0010173
CPF81D, long shaft, protection guard
1
2
3
4
5
Memosens plug-in head
NPT ¾" thread
Across flats AF 26
Platinum ring
Protection guard
1
2
3
4
Memosens plug-in head
NPT ¾" thread
Across flats AF 26
Flat membrane
1
2
3
4
Memosens plug-in head
NPT ¾" thread
Across flats AF 26
Protection guard
23
(0.91)
150
(5.91)
Ø 22 (0.87)
35
(1.38)
mm (inch)
1
2
3
4
2
5
12.5
(0.49)
140
(5.51)
Ø 22 (0.87)
35
(1.38)
mm (inch)
1
2
3
4
2
58.5
(2.3)
150
(5.91)
Ø 22 (0.87)
35
(1.38)
mm (inch)
1
2
3
4
2
Housing, electrode shaft: PPS
pH electrode (in contact with medium): lead-free membrane glass, suitable for process applications
ORP electrode (in contact with medium): platinum ring
Double chamber reference system: KNO3 and KCl/AgCl
Design: cast in sensor body
Power supply: via integrated coin cells
Reference potential: reference electrode
CPF81/82, CPF81D/82D
8 Endress+Hauser
Ordering information
Product structure CPF81
Product structure CPF82
Application range
LH pH 0 to 14, T = 0 to 110 °C (32 to 230 °F)
NN pH 0 to 11, T = 0 to 80 °C (32 to 176 °F)
Version
1 Insertion length 23 mm (0.91") with protection guard
2 Insertion length 12.5 mm (0.51") with flat membrane
3 Insertion length 58.5 mm (2.28") with protection guard
Process connection
1 NPT ¾"
2 NPT ¾" with integrated solution ground
Equipment
A No preamplifier, no Pt 100
B With preamplifier, no Pt 100
C No preamplifier, with Pt 100
D With preamplifier, with Pt 100
Cable connection
2 4.5 m (15 ft) fixed cable
3 9 m (30 ft) fixed cable
4 13 m (45 ft) fixed cable
8 TOP68/ESA plug-in head
CPF81- complete order code
Application range
PA ORP, platinum ring, 0 to 80 °C (32 to 176 °F)
Version
1 Insertion length 23 mm (0.91") with protection guard
3 Insertion length 58.5 mm (2.28") with protection guard
Process connection
1 NPT ¾"
Equipment
A No preamplifier
B With preamplifier
Cable connection
2 4.5 m (15 ft) fixed cable
3 9 m (30 ft) fixed cable
4 13 m (45 ft) fixed cable
8 TOP68/ESA plug-in head
CPF82- complete order code
CPF81/82, CPF81D/82D
Endress+Hauser 9
Product structure CPF81D
Product structure CPF82D
Version
7 Basic version
Application range
LH 0 to 14 pH; 0 to 110 °C (32 to 230 °F)
NN 0 to 11 pH; 0 to 80 °C (32 to 176 °F)
Insertion length
1 23 mm (0.91") + electrode guard
2 13 mm (0.51"); flat membrane
3 58 mm (2.28") + electrode guard
Approval
1 Non-hazardous area
CPF81D- complete order code
Version
7 Basic version
Application range
PA ORP, platinum, 0 to 80 °C (32 to 176 °F)
Insertion length
1 23 mm (0.91") + electrode guard
3 58 mm (2.28") + electrode guard
Approval
1 Non-hazardous area
CPF82D- complete order code
CPF81/82, CPF81D/82D
10 Endress+Hauser
Accessories
Note!
In the following sections, you find the accessories available at the time of issue of this documentation.
For information on accessories that are not listed here, please contact your local service.
Transmitters Liquiline M CM42
• Modular two-wire transmitter, stainless steel or plastic, field or panel instrument,
• various Ex approvals (ATEX, FM, CSA, Nepsi, TIIS),
• HART, PROFIBUS or FOUNDATION Fieldbus available
• Ordering acc. to product structure, see Technical Information (TI381C/07/en)
Liquisys M CPM223/253
• Transmitter for pH and ORP, field or panel-mounted housing,
• HART or PROFIBUS available
• Ordering acc. to product structure, see Technical Information (TI194C/07/en)
Mycom S CPM153
• Transmitter for pH and ORP, one or two channel version, Ex or non-Ex,
• HART or PROFIBUS available
• Ordering acc. to product structure, see Technical Information (TI233C/07/en)
Assembly Immersion assembly Dipfit W CYA611
• For sensor immersion in basins, open channels and tanks, PVC
• Ordering acc. to product structure, see Technical Information TI166C/07/en
Buffer solutions pH
High-quality buffer solutions of Endress+Hauser - CPY20
The secondary buffer solutions have been referenced to primary reference material of the PTB (German Federal
Physico-technical Institute) and to standard reference material of NIST (National Institute of Standards and
Technology) according to DIN 19266 by a DKD (German Calibration Service) accredited laboratory.
ORP
Technical buffer solutions for ORP electrodes
• +220 mV, pH 7.0, 100 ml (3.4 fl.oz.); order no. CPY3-0
• +468 mV, pH 0.1, 100 ml (3.4 fl.oz.); order no. CPY3-1
pH value
A pH 2.00 (accuracy ± 0.02 pH)
C pH 4.00 (accuracy ± 0.02 pH)
E pH 7.00 (accuracy ± 0.02 pH)
G pH 9.00 (accuracy ± 0.02 pH)
I pH 9.20 (accuracy ± 0.02 pH)
K pH 10.00 (accuracy ± 0.05 pH)
M pH 12.00 (accuracy ± 0.05 pH)
Quantity
01 20 x 18 ml (0.68 fl.oz) only buffer solutions pH 4.00 and 7.00
02 250 ml (8.45 fl.oz)
10 1000 ml (0.26 US gal)
50 5000 ml (1.32 US gal) canister for Topcal S
Certificates
A Buffer analysis certificate
Version
1 Standard
CPY20- complete order code
CPF81/82, CPF81D/82D
Endress+Hauser 11
Measuring cables CPK9 special measuring cable
• For sensors with TOP68 plug-in head, for high-temperature and high-pressure applications, IP 68
• Ordering acc. to product structure, see Technical Information (TI118C/07/en)
CYK10 Memosens data cable
• For digital sensors with Memosens technology
• Ordering according to product structure, see below
CYK10 product structure
Certificates
A Standard, non Ex
G ATEX II 1G EEx ia IIC T6/T4
Cable length
03 Cable length: 3 m (9.8 ft)
05 Cable length: 5 m (16 ft)
10 Cable length: 10 m (33 ft)
15 Cable length: 15 m (49 ft)
20 Cable length: 20 m (66 ft)
25 Cable length: 25 m (82 ft)
88 ... m length
89 ... ft length
Ready-made
1 Wire terminals
CYK10- complete order code
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
Tel.+41 61 715 81 00
Fax+41 61 715 25 00
www.endress.com
info@ii.endress.com
TI191C/07/en/07.08
Printed in Germany / FM+SGML 6.0 / DT
71078186

Contenu connexe

Tendances

2008 Mitsubishi Lancer Evolution 10 Service Repair Manual
2008 Mitsubishi Lancer Evolution 10 Service Repair Manual2008 Mitsubishi Lancer Evolution 10 Service Repair Manual
2008 Mitsubishi Lancer Evolution 10 Service Repair Manual
jdksmemd
 
2007 Mitsubishi Eclipse Service Repair Manual
2007 Mitsubishi Eclipse Service Repair Manual2007 Mitsubishi Eclipse Service Repair Manual
2007 Mitsubishi Eclipse Service Repair Manual
jdksmemd
 

Tendances (6)

2008 Mitsubishi Lancer Evolution 10 Service Repair Manual
2008 Mitsubishi Lancer Evolution 10 Service Repair Manual2008 Mitsubishi Lancer Evolution 10 Service Repair Manual
2008 Mitsubishi Lancer Evolution 10 Service Repair Manual
 
Manual de caldera beretta ciao 24 csi
Manual de caldera beretta ciao 24 csiManual de caldera beretta ciao 24 csi
Manual de caldera beretta ciao 24 csi
 
2007 Mitsubishi Eclipse Service Repair Manual
2007 Mitsubishi Eclipse Service Repair Manual2007 Mitsubishi Eclipse Service Repair Manual
2007 Mitsubishi Eclipse Service Repair Manual
 
New holland tn95 fa tractor service repair manual
New holland tn95 fa tractor service repair manualNew holland tn95 fa tractor service repair manual
New holland tn95 fa tractor service repair manual
 
USG B SCAN OPHTHALMOLOGY
USG B SCAN OPHTHALMOLOGYUSG B SCAN OPHTHALMOLOGY
USG B SCAN OPHTHALMOLOGY
 
Interpretation of visual fields with special reference to octopus
Interpretation of visual fields with special reference to octopusInterpretation of visual fields with special reference to octopus
Interpretation of visual fields with special reference to octopus
 

Similaire à Orbipac cpf81 cpf81 d_cpf82_cpf82d-endress+hauser datasheet-ph sensor

Monicon brochure
Monicon brochureMonicon brochure
Monicon brochure
a1-cbiss
 
VEGA Pressure & Level Measurement - Renewable Energy Industry Applications
VEGA Pressure & Level Measurement  - Renewable Energy Industry ApplicationsVEGA Pressure & Level Measurement  - Renewable Energy Industry Applications
VEGA Pressure & Level Measurement - Renewable Energy Industry Applications
Thorne & Derrick UK
 

Similaire à Orbipac cpf81 cpf81 d_cpf82_cpf82d-endress+hauser datasheet-ph sensor (20)

Orbisint cps11 cps11 d-endress+hauser datasheet-ph electrodes
Orbisint cps11 cps11 d-endress+hauser datasheet-ph electrodesOrbisint cps11 cps11 d-endress+hauser datasheet-ph electrodes
Orbisint cps11 cps11 d-endress+hauser datasheet-ph electrodes
 
Combination of pH and ORP electrode
Combination of pH and ORP electrodeCombination of pH and ORP electrode
Combination of pH and ORP electrode
 
pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41
 
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensorsCondumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
 
pH.ORP Transmitter-4-wire-controller
pH.ORP Transmitter-4-wire-controllerpH.ORP Transmitter-4-wire-controller
pH.ORP Transmitter-4-wire-controller
 
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensorsCondumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
 
CAS74 stip scan cam74-cas74-endress+hauser datasheet-BOD sensor
CAS74 stip scan cam74-cas74-endress+hauser datasheet-BOD sensorCAS74 stip scan cam74-cas74-endress+hauser datasheet-BOD sensor
CAS74 stip scan cam74-cas74-endress+hauser datasheet-BOD sensor
 
Monicon brochure
Monicon brochureMonicon brochure
Monicon brochure
 
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygenOxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygen
 
Siemens Air velocity sensor
Siemens Air velocity sensorSiemens Air velocity sensor
Siemens Air velocity sensor
 
Dissolved Oxygen Sensors-Oxymax COS51D
Dissolved Oxygen Sensors-Oxymax COS51DDissolved Oxygen Sensors-Oxymax COS51D
Dissolved Oxygen Sensors-Oxymax COS51D
 
Vanjen Group IntelliSAW Overview 2018
Vanjen Group IntelliSAW Overview 2018Vanjen Group IntelliSAW Overview 2018
Vanjen Group IntelliSAW Overview 2018
 
Dissolved Oxygen Sensors-Oxymax COS22/22D
Dissolved Oxygen Sensors-Oxymax COS22/22DDissolved Oxygen Sensors-Oxymax COS22/22D
Dissolved Oxygen Sensors-Oxymax COS22/22D
 
Sensepoint_Pro_pdf.pdf
Sensepoint_Pro_pdf.pdfSensepoint_Pro_pdf.pdf
Sensepoint_Pro_pdf.pdf
 
CM14
CM14CM14
CM14
 
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...
 
Products
ProductsProducts
Products
 
VEGA Pressure & Level Measurement - Renewable Energy Industry Applications
VEGA Pressure & Level Measurement  - Renewable Energy Industry ApplicationsVEGA Pressure & Level Measurement  - Renewable Energy Industry Applications
VEGA Pressure & Level Measurement - Renewable Energy Industry Applications
 
IoT Connected Steam Flow Meter - Compact Orifice
IoT Connected Steam Flow Meter - Compact OrificeIoT Connected Steam Flow Meter - Compact Orifice
IoT Connected Steam Flow Meter - Compact Orifice
 
A Trusted Steam Flow Meter Manufacturer and Supplier- Manas
A Trusted Steam Flow Meter Manufacturer and Supplier- ManasA Trusted Steam Flow Meter Manufacturer and Supplier- Manas
A Trusted Steam Flow Meter Manufacturer and Supplier- Manas
 

Plus de ENVIMART

Plus de ENVIMART (20)

Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdfZinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
 
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdfNickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
 
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdfManganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
 
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdfLead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
 
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdfIron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
 
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdfIron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
 
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdfChromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdfCopper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdfCopper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdfCopper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
 
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdfCobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
 
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdfCadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
 
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdfArsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
 
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdfAluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
 
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdfATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
 
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdfATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
 
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdfATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
 
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdfATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
 
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdfATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
 
B12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdfB12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdf
 

Dernier

Principle of erosion control- Introduction to contouring,strip cropping,conto...
Principle of erosion control- Introduction to contouring,strip cropping,conto...Principle of erosion control- Introduction to contouring,strip cropping,conto...
Principle of erosion control- Introduction to contouring,strip cropping,conto...
ZAPPAC1
 
Corporate_Science-based_Target_Setting.pptx
Corporate_Science-based_Target_Setting.pptxCorporate_Science-based_Target_Setting.pptx
Corporate_Science-based_Target_Setting.pptx
arnab132
 
Training Of Trainers FAI Eng. Basel Tilapia Welfare.pdf
Training Of Trainers FAI Eng. Basel Tilapia Welfare.pdfTraining Of Trainers FAI Eng. Basel Tilapia Welfare.pdf
Training Of Trainers FAI Eng. Basel Tilapia Welfare.pdf
Basel Ahmed
 

Dernier (20)

Hertwich_EnvironmentalImpacts_BuildingsGRO.pptx
Hertwich_EnvironmentalImpacts_BuildingsGRO.pptxHertwich_EnvironmentalImpacts_BuildingsGRO.pptx
Hertwich_EnvironmentalImpacts_BuildingsGRO.pptx
 
Sustainable Recovery and Reconstruction Framework (SURRF)(1).pdf
Sustainable Recovery and Reconstruction Framework  (SURRF)(1).pdfSustainable Recovery and Reconstruction Framework  (SURRF)(1).pdf
Sustainable Recovery and Reconstruction Framework (SURRF)(1).pdf
 
Enhancing forest data transparency for climate action
Enhancing forest data transparency for climate actionEnhancing forest data transparency for climate action
Enhancing forest data transparency for climate action
 
Principle of erosion control- Introduction to contouring,strip cropping,conto...
Principle of erosion control- Introduction to contouring,strip cropping,conto...Principle of erosion control- Introduction to contouring,strip cropping,conto...
Principle of erosion control- Introduction to contouring,strip cropping,conto...
 
Call girl in Sharjah 0503464457 Sharjah Call girl
Call girl in Sharjah 0503464457 Sharjah Call girlCall girl in Sharjah 0503464457 Sharjah Call girl
Call girl in Sharjah 0503464457 Sharjah Call girl
 
Call Girls in Veraval - 8250092165 Our call girls are sure to provide you wit...
Call Girls in Veraval - 8250092165 Our call girls are sure to provide you wit...Call Girls in Veraval - 8250092165 Our call girls are sure to provide you wit...
Call Girls in Veraval - 8250092165 Our call girls are sure to provide you wit...
 
Call girl in Ajman 0503464457 Ajman Call girl services
Call girl in Ajman 0503464457 Ajman Call girl servicesCall girl in Ajman 0503464457 Ajman Call girl services
Call girl in Ajman 0503464457 Ajman Call girl services
 
Jumping Scales and Producing peripheries.pptx
Jumping Scales and Producing peripheries.pptxJumping Scales and Producing peripheries.pptx
Jumping Scales and Producing peripheries.pptx
 
Corporate_Science-based_Target_Setting.pptx
Corporate_Science-based_Target_Setting.pptxCorporate_Science-based_Target_Setting.pptx
Corporate_Science-based_Target_Setting.pptx
 
NO1 Google Kala Jadu Expert Specialist In Qatar Kala Jadu Expert Specialist I...
NO1 Google Kala Jadu Expert Specialist In Qatar Kala Jadu Expert Specialist I...NO1 Google Kala Jadu Expert Specialist In Qatar Kala Jadu Expert Specialist I...
NO1 Google Kala Jadu Expert Specialist In Qatar Kala Jadu Expert Specialist I...
 
Presentation: Farmer-led climate adaptation - Project launch and overview by ...
Presentation: Farmer-led climate adaptation - Project launch and overview by ...Presentation: Farmer-led climate adaptation - Project launch and overview by ...
Presentation: Farmer-led climate adaptation - Project launch and overview by ...
 
Training Of Trainers FAI Eng. Basel Tilapia Welfare.pdf
Training Of Trainers FAI Eng. Basel Tilapia Welfare.pdfTraining Of Trainers FAI Eng. Basel Tilapia Welfare.pdf
Training Of Trainers FAI Eng. Basel Tilapia Welfare.pdf
 
Sensual Call Girls in Surajpur { 9332606886 } VVIP NISHA Call Girls Near 5 St...
Sensual Call Girls in Surajpur { 9332606886 } VVIP NISHA Call Girls Near 5 St...Sensual Call Girls in Surajpur { 9332606886 } VVIP NISHA Call Girls Near 5 St...
Sensual Call Girls in Surajpur { 9332606886 } VVIP NISHA Call Girls Near 5 St...
 
Introduction to heat waves and Heatwaves in Bangladesh.pptx
Introduction to heat waves and Heatwaves in Bangladesh.pptxIntroduction to heat waves and Heatwaves in Bangladesh.pptx
Introduction to heat waves and Heatwaves in Bangladesh.pptx
 
Test bank for beckmann and ling s obstetrics and gynecology 8th edition by ro...
Test bank for beckmann and ling s obstetrics and gynecology 8th edition by ro...Test bank for beckmann and ling s obstetrics and gynecology 8th edition by ro...
Test bank for beckmann and ling s obstetrics and gynecology 8th edition by ro...
 
Yil Me Hu Spring 2024 - Nisqually Salmon Recovery Newsletter
Yil Me Hu Spring 2024 - Nisqually Salmon Recovery NewsletterYil Me Hu Spring 2024 - Nisqually Salmon Recovery Newsletter
Yil Me Hu Spring 2024 - Nisqually Salmon Recovery Newsletter
 
NO1 Google Best Black Magic Specialist Near Me Spiritual Healer Powerful Love...
NO1 Google Best Black Magic Specialist Near Me Spiritual Healer Powerful Love...NO1 Google Best Black Magic Specialist Near Me Spiritual Healer Powerful Love...
NO1 Google Best Black Magic Specialist Near Me Spiritual Healer Powerful Love...
 
Call Girls in Dattatreya Nagar / 8250092165 Genuine Call girls with real Phot...
Call Girls in Dattatreya Nagar / 8250092165 Genuine Call girls with real Phot...Call Girls in Dattatreya Nagar / 8250092165 Genuine Call girls with real Phot...
Call Girls in Dattatreya Nagar / 8250092165 Genuine Call girls with real Phot...
 
Faridabad Call Girl ₹7.5k Pick Up & Drop With Cash Payment 8168257667 Badarpu...
Faridabad Call Girl ₹7.5k Pick Up & Drop With Cash Payment 8168257667 Badarpu...Faridabad Call Girl ₹7.5k Pick Up & Drop With Cash Payment 8168257667 Badarpu...
Faridabad Call Girl ₹7.5k Pick Up & Drop With Cash Payment 8168257667 Badarpu...
 
FYP1 FINAL PRESENTATION POLITEKNIK MALAYSIA
FYP1 FINAL PRESENTATION POLITEKNIK MALAYSIAFYP1 FINAL PRESENTATION POLITEKNIK MALAYSIA
FYP1 FINAL PRESENTATION POLITEKNIK MALAYSIA
 

Orbipac cpf81 cpf81 d_cpf82_cpf82d-endress+hauser datasheet-ph sensor

  • 1. TI191C/07/en/07.08 71078186 Technical Information Orbipac CPF81/CPF82 and CPF81D/CPF82D Compact pH/ORP electrodes, analog or with digital Memosens technology For installation or immersion in industrial water and wastewater Application • Industrial and municipal wastewater treatment • Water conditioning • Condensate cleaning Your benefits • Suitable for flow and immersion installation • Suitable for measurement within pH range 0 to 14 pH and temperature range 0 to 110 °C (32 to 230 °F) • Analog pH electrodes with or without integrated temperature sensor, digital pH sensors with integrated temperature sensor • With patented KNO3 electrolyte bridge for better protection against electrode poisons like S2- or CN- ions • Protection guard against damage • Flat membrane suitable for high flow rates and fibrous media • Optional integrated preamplifier for low-impedance and noise-free measuring value transmission • NPT ¾" threaded connection, top and bottom Further benefits offered by Memosens technology • Maximum process safety through contactless inductive signal transmission • Data safety through digital data transmission • Easy handling thanks to storage of sensor-specific data in the sensor • Predictive maintenance possible thanks to registration of sensor load data in the sensor
  • 2. CPF81/82, CPF81D/82D 2 Endress+Hauser Function and system design Measuring principle pH measurement The pH value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. The membrane glass of the electrode supplies an electrochemical potential which is dependent upon the pH value of the medium. This potential is generated by the selective penetration of H+ ions through the outer layer of the membrane. An electrochemical boundary layer with an electric potential forms at this point. An integrated Ag/AgCl reference system serves as reference electrode. The transmitter converts the measured voltage into the corresponding pH value using the Nernst equation. ORP measurement The ORP potential is a unit of measurement for the state of equilibria between oxidizing and reducing components of a medium. ORP potential is measured similarly to the pH value. A platinum electrode is used instead of pH-sensitive membrane glass. Analog to the pH measurement, an integrated Ag/AgCl reference system is used as a reference electrode. Important properties of CPF81D/CPF82D Maximum process safety The inductive and non-contacting measured value transmission of Memosens guarantees maximum process safety and offers the following benefits: • All problems caused by moisture are eliminated. – The plug-in connection is non-contacting and therefore free from corrosion. – Measured value distortion from moisture is not possible. – The plug-in system can even be connected under water. • The transmitter is galvanically decoupled from the medium. The result: No more need to ask about "symmetrically high-impedance" or "unsymmetrical" (for pH/ORP measurement) or an impedance converter. • EMC safety is guaranteed by screening measures for the digital measured value transmission. Data safety through digital data transfer The Memosens technology digitalizes the measured values in the sensor and transfers them to the transmitter contactlessly and free from interference potential. The result: • An automatic error message is generated if the sensor fails or the connection between sensor and transmitter is interrupted. • The availability of the measuring point is dramatically increased by immediate error detection. Easy handling Sensors with Memosens technology have integrated electronics that allow for saving calibration data and further information such as total hours of operation and operating hours under extreme measuring conditions. When the sensor is mounted, the calibration data are automatically transferred to the transmitter and used to calculate the current measured value. Storing the calibration data in the sensor allows for calibration away from the measuring point. The result: • Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive maintenance is possible. • The sensor history can be documented on external data carriers and evaluation programs at any time. Thus, the current application of the sensors can be made to depend on their previous history. • The transmitter does not need to be installed close to the measuring point but can be placed in the control room. Communication with the transmitter Always connect digital sensors to a transmitter with Memosens technology. Data transmission to a transmitter for analog sensors is not possible.
  • 3. CPF81/82, CPF81D/82D Endress+Hauser 3 Data storage CPF81D The digital sensors are able to store the following system data in the sensor. • Manufacturing data – Serial number – Order code – Date of manufacture • Calibration data – Calibration date – Calibrated slope at 25 °C (77 °F) – Calibrated zero point at 25 °C (77 °F) – Temperature offset – Number of calibrations – Serial number of the transmitter used for the last calibration • Application data – Temperature application range – pH application range – Date of first commissioning – Maximum temperature value – Total operating hours – Operating hours at temperatures above 80 °C (176 °F) and 100 °C (212 °F) – Operating hours at very low and very high pH values (Nernst voltage below -300 mV, above +300 mV) – Glass membrane impedance These system data can be displayed with Mycom S CPM153 and Liquiline M CM42 transmitters. CPF82D The digital sensors are able to store the following system data in the sensor. • Manufacturing data – Serial number – Order code – Date of manufacture • Calibration data – Calibration date – Calibrated offset (operating mode "mV") – % slope (operating mode "%") – Number of calibrations – Serial number of the transmitter used for the last calibration • Application data – Temperature application range – ORP application range – Date of first commissioning – Operating hours These system data can be displayed with Mycom S CPM153 or Liquiline M CM42 transmitters.
  • 4. CPF81/82, CPF81D/82D 4 Endress+Hauser Input Measured values Measuring range Caution! Please note the process operating conditions. Cable specification CPF81/82 with fixed cable a0010182-en Measuring cable A Fixed cable for CPF81 without temperature sensor and CPF82 B Fixed cable for CPF81 with temperature sensor * PM is only connected for sensor versions with internal solution ground (CPF81-xxx2xx). CPF81/82 with TOP68 plug-in head a0003303-en CPK9 special measuring cable CPF81D/82D a0003350 CYK10 Memosens data cable CPF81, CPF81D: pH value Temperature CPF82, CPF82D: ORP potential CPF81/81D Electrode version LH pH: 0 to 14 pH Temperature: 0 to 110 °C (32 to 230 °F) Electrode version NN pH: 0 to 14 pH (reduced accuracy for 11 to 14 pH) Temperature: 0 to 80 °C (32 to 176 °F) CPF82/82D ORP potential: -1500 mV to +1500 mV BK BK GN/YE BN BK BK GN WH YE GN/YE BN GN WH YE A B Reference pH/ORP Temperature Temperature Temperature Outer screen PM / n. c.* pH/ORP + Pt 100 PM Coax BK WH GN BN Screen connection WH YE Ref. pH Temp Temp. Temp. PM
  • 5. CPF81/82, CPF81D/82D Endress+Hauser 5 Installation Installation instructions The electrode is suitable for flow and immersion installation. Note! Make sure to follow the installation instructions in the Operating Instructions of the used assembly. Environment Ambient temperature Caution! Danger of frost damage Do not use the electrodes at temperatures below 0 °C (32 °F.) Storage temperature 0 to 50 °C (32 to 122 °F) Ingress protection Process Process temperature Process pressure 0 to 10 bar @ 80 °C (0 to 145 psi @ 176 °F) a0010183 Flow installation a0010184 Immersion installation with Flowfit CYA611 Fixed-cable version: IP 67 TOP68 plug-in head: IP 68 (1 m (3.28 ft) water column, 50 °C (122 °F), 168 h) Memosens plug-in head: IP 68 (10 m (32.8 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl) CPF81/81D Electrode version LH: 0 to 110 °C (32 to 230 °F) Electrode version NN: 0 to 80 °C (32 to 176 °F) CPF82/82D 0 to 80 °C (32 to 176 °F)
  • 6. CPF81/82, CPF81D/82D 6 Endress+Hauser Pressure temperature load curve a0010185 Glass impedance 150 MΩ @ 25 °C (77 °F ) Minimum conductivity min. 50 μS/cm Mechanical construction Design, dimensions CPF81/82 60 80 100 20 40 0 2 4 6 8 10 140 180 220 60 100 32 40 120 °C °F bar psi 50.8 176 110 230 80 0 0 3.5 145 a0010179 Fixed-cable version, short shaft, protection guard a0010180 CPF81 with TOP68 plug-in head, short shaft, flat membrane a0010181 Fixed-cable version, long shaft, protection guard 1 2 3 4 Fixed cable NPT ¾" thread Across flats AF 26 Protection guard 1 2 3 4 TOP68 plug-in head NPT ¾" thread Across flats AF 26 Flat membrane 1 2 3 4 Fixed cable NPT ¾" thread Across flats AF 26 Protection guard Ø 22 (0.87) 23 (0.91) 150 (5.91) mm (inch) 1 2 3 4 2 Ø 22 (0.87) 12.5 (0.49) 140 (5.51) mm (inch) 1 2 3 4 2 Ø 22 (0.87) 58.5 (2.3) 150 (5.91) mm (inch) 1 2 3 4 2
  • 7. CPF81/82, CPF81D/82D Endress+Hauser 7 Design, dimensions CPF81D/82D Material Process connection NPT ¾" thread Integrated preamplifier (if ordered) Note! With preamplifier versions, the sensor check function (SCS) of the transmitter is not available and should be turned off. a0010174 CPF82D, short shaft, protection guard a0010171 CPF81D, flat membrane a0010173 CPF81D, long shaft, protection guard 1 2 3 4 5 Memosens plug-in head NPT ¾" thread Across flats AF 26 Platinum ring Protection guard 1 2 3 4 Memosens plug-in head NPT ¾" thread Across flats AF 26 Flat membrane 1 2 3 4 Memosens plug-in head NPT ¾" thread Across flats AF 26 Protection guard 23 (0.91) 150 (5.91) Ø 22 (0.87) 35 (1.38) mm (inch) 1 2 3 4 2 5 12.5 (0.49) 140 (5.51) Ø 22 (0.87) 35 (1.38) mm (inch) 1 2 3 4 2 58.5 (2.3) 150 (5.91) Ø 22 (0.87) 35 (1.38) mm (inch) 1 2 3 4 2 Housing, electrode shaft: PPS pH electrode (in contact with medium): lead-free membrane glass, suitable for process applications ORP electrode (in contact with medium): platinum ring Double chamber reference system: KNO3 and KCl/AgCl Design: cast in sensor body Power supply: via integrated coin cells Reference potential: reference electrode
  • 8. CPF81/82, CPF81D/82D 8 Endress+Hauser Ordering information Product structure CPF81 Product structure CPF82 Application range LH pH 0 to 14, T = 0 to 110 °C (32 to 230 °F) NN pH 0 to 11, T = 0 to 80 °C (32 to 176 °F) Version 1 Insertion length 23 mm (0.91") with protection guard 2 Insertion length 12.5 mm (0.51") with flat membrane 3 Insertion length 58.5 mm (2.28") with protection guard Process connection 1 NPT ¾" 2 NPT ¾" with integrated solution ground Equipment A No preamplifier, no Pt 100 B With preamplifier, no Pt 100 C No preamplifier, with Pt 100 D With preamplifier, with Pt 100 Cable connection 2 4.5 m (15 ft) fixed cable 3 9 m (30 ft) fixed cable 4 13 m (45 ft) fixed cable 8 TOP68/ESA plug-in head CPF81- complete order code Application range PA ORP, platinum ring, 0 to 80 °C (32 to 176 °F) Version 1 Insertion length 23 mm (0.91") with protection guard 3 Insertion length 58.5 mm (2.28") with protection guard Process connection 1 NPT ¾" Equipment A No preamplifier B With preamplifier Cable connection 2 4.5 m (15 ft) fixed cable 3 9 m (30 ft) fixed cable 4 13 m (45 ft) fixed cable 8 TOP68/ESA plug-in head CPF82- complete order code
  • 9. CPF81/82, CPF81D/82D Endress+Hauser 9 Product structure CPF81D Product structure CPF82D Version 7 Basic version Application range LH 0 to 14 pH; 0 to 110 °C (32 to 230 °F) NN 0 to 11 pH; 0 to 80 °C (32 to 176 °F) Insertion length 1 23 mm (0.91") + electrode guard 2 13 mm (0.51"); flat membrane 3 58 mm (2.28") + electrode guard Approval 1 Non-hazardous area CPF81D- complete order code Version 7 Basic version Application range PA ORP, platinum, 0 to 80 °C (32 to 176 °F) Insertion length 1 23 mm (0.91") + electrode guard 3 58 mm (2.28") + electrode guard Approval 1 Non-hazardous area CPF82D- complete order code
  • 10. CPF81/82, CPF81D/82D 10 Endress+Hauser Accessories Note! In the following sections, you find the accessories available at the time of issue of this documentation. For information on accessories that are not listed here, please contact your local service. Transmitters Liquiline M CM42 • Modular two-wire transmitter, stainless steel or plastic, field or panel instrument, • various Ex approvals (ATEX, FM, CSA, Nepsi, TIIS), • HART, PROFIBUS or FOUNDATION Fieldbus available • Ordering acc. to product structure, see Technical Information (TI381C/07/en) Liquisys M CPM223/253 • Transmitter for pH and ORP, field or panel-mounted housing, • HART or PROFIBUS available • Ordering acc. to product structure, see Technical Information (TI194C/07/en) Mycom S CPM153 • Transmitter for pH and ORP, one or two channel version, Ex or non-Ex, • HART or PROFIBUS available • Ordering acc. to product structure, see Technical Information (TI233C/07/en) Assembly Immersion assembly Dipfit W CYA611 • For sensor immersion in basins, open channels and tanks, PVC • Ordering acc. to product structure, see Technical Information TI166C/07/en Buffer solutions pH High-quality buffer solutions of Endress+Hauser - CPY20 The secondary buffer solutions have been referenced to primary reference material of the PTB (German Federal Physico-technical Institute) and to standard reference material of NIST (National Institute of Standards and Technology) according to DIN 19266 by a DKD (German Calibration Service) accredited laboratory. ORP Technical buffer solutions for ORP electrodes • +220 mV, pH 7.0, 100 ml (3.4 fl.oz.); order no. CPY3-0 • +468 mV, pH 0.1, 100 ml (3.4 fl.oz.); order no. CPY3-1 pH value A pH 2.00 (accuracy ± 0.02 pH) C pH 4.00 (accuracy ± 0.02 pH) E pH 7.00 (accuracy ± 0.02 pH) G pH 9.00 (accuracy ± 0.02 pH) I pH 9.20 (accuracy ± 0.02 pH) K pH 10.00 (accuracy ± 0.05 pH) M pH 12.00 (accuracy ± 0.05 pH) Quantity 01 20 x 18 ml (0.68 fl.oz) only buffer solutions pH 4.00 and 7.00 02 250 ml (8.45 fl.oz) 10 1000 ml (0.26 US gal) 50 5000 ml (1.32 US gal) canister for Topcal S Certificates A Buffer analysis certificate Version 1 Standard CPY20- complete order code
  • 11. CPF81/82, CPF81D/82D Endress+Hauser 11 Measuring cables CPK9 special measuring cable • For sensors with TOP68 plug-in head, for high-temperature and high-pressure applications, IP 68 • Ordering acc. to product structure, see Technical Information (TI118C/07/en) CYK10 Memosens data cable • For digital sensors with Memosens technology • Ordering according to product structure, see below CYK10 product structure Certificates A Standard, non Ex G ATEX II 1G EEx ia IIC T6/T4 Cable length 03 Cable length: 3 m (9.8 ft) 05 Cable length: 5 m (16 ft) 10 Cable length: 10 m (33 ft) 15 Cable length: 15 m (49 ft) 20 Cable length: 20 m (66 ft) 25 Cable length: 25 m (82 ft) 88 ... m length 89 ... ft length Ready-made 1 Wire terminals CYK10- complete order code
  • 12. Instruments International Endress+Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel.+41 61 715 81 00 Fax+41 61 715 25 00 www.endress.com info@ii.endress.com TI191C/07/en/07.08 Printed in Germany / FM+SGML 6.0 / DT 71078186