SlideShare une entreprise Scribd logo
1  sur  46
USG B SCAN




  Dr.Gyanendra Lamichhanae
Vitreo retinal Fellow, Gunma University ,JAPAN
       Lumbini Eye Institute, Bhairahawa
What is ultrasound
•   sound pressure with a frequency greater than the upper limit of
    human hearing.




•   Although this limit varies from person to person, it is approximately
    20 kilohertz (20,000 hertz) in healthy, young adults




                                                                            2
3
• B-scan ultrasonography is an important
  noninvasive technique for the clinical
  assessment of various ocular and orbital
  diseases




                                             4
HISTORY
• 1793: Lazzaro Spallanzani (Italy) discovered that bats orient
  themselves with the help of sound whistles while flying in
  darkness. This was the basis of modern ultrasound application




             Bats use ultrasounds to navigate in the
                            darkness
                                                                  5
History contd…….
•   World war II: a device based on piezoelectric effect developed
    by Paul Langevin (France) ,able of emitting & receiving ultrasound
    under water used as sonar.

•    1956: first documented use of ocular USG, Mundt and
    Hughes used A scan technique to detect intraocular tumour.

•    1972: First use of hand held B scan by Bronston &
    workers ,which was applied directly to the closed lid without a
    water bath




                                                                         6
•   Principles of ultrasound:
•   By definition, an ultrasound wave has a frequency greater than 20
    kHz (20,000 oscillations/ second)

•   As the frequency of USG increases, the wavelength decreases
    and wavelength of an ultrasound determines its depth of tissue
    penetration and resolution
             Wavelength α Depth of penetration of the ultrasound

•   So, Larger is the frequency of US = shorter is its wavelength =
    shallower is its penetration = better is the resolution of resultant
    echo graph.


                                                                           7
•   That’s why USG probes used for Ocular USG are of higher
    frequency(10MHz)as it needs much less tissue penetration (an
    eye is 23.5 mm long on average) & higher resolution.

•   In contrast, ultrasound probes used for purposes such as
    obstetrics, use lower frequencies (1-5Hz) for deeper
    penetration into the body, and, because the structures being
    imaged are larger, they do not require the same degree of resolution




                                                                       8
Ultrasound Principles and Physics


Ophthalmic ultrasonography uses high-frequency sound waves,

 transmitted from a probe into the eye.

As the sound waves strike intraocular structures,

they are reflected back to the probe and converted into an electric
   signal.

The signal is subsequently reconstructed as an image on a monitor,




                                                                      9
Velocity

•   The velocity of the sound wave is dependent on the density of the
    medium through which the sound travels.

•   Sound travels faster through solids than liquids, an important
    principle to understand since the eye is composed of both.



•    There are known velocities of different components of the eye, with
    sound traveling through both aqueous and vitreous at a speed of
    1,532 meters/second (m/s) and through the cornea and lens at
    an average speed of 1,641 m/s



                                                                       10
Reflectivity

•   When sound travels from one medium to another medium of
    different density, part of the sound is reflected from the interface
    between those media back into the probe.

•    This is known as an echo; the greater the density difference at that
    interface, the stronger the echo, or the higher the reflectivity

•   In A-scan ultrasonography, a thin, parallel sound beam is emitted,
    which passes through the eye and images one small axis of
    tissue; the echoes of which are represented as spikes arising
    from a baseline. The stronger the echo, the higher the spike



                                                                           11
12
•    In B-scan ultrasonography, an oscillating sound beam is
     emitted, passing through the eye and imaging a slice of tissue;
     the echoes of which are represented as a multitude of dots that
     together form an image on the screen.

    The stronger the echo, the brighter the dot.

    example, the dots that form the posterior vitreous hyaloid membrane
     are not as bright as the dots that form the retinal membrane.

This is very useful in differentiating a posterior vitreous detachment (a
  benign condition) from a more highly reflective retinal
  detachment (a blinding condition) because retina is more dense
  than vitreous.
                                                                        13
Angle of incidence

•   The angle of incidence of the probe is critical for both A-scan and B-
    scan ultrasonography.
•   When the probe is held perpendicular to the area of interest,
    more of the echo is reflected directly back into the probe tip
    and sent to the display screen.
•    When held oblique to the area imaged, part of the echo is
    reflected away from the probe tip and less is sent to the display
    screen.
•   The more oblique the probe is held from the area of interest, the
    weaker the returning echo and, thus, the more compromised the
    displayed image.
.


                                                                        14
•   On A-scan, the greater the perpendicularity, the more steeply
    rising the spike is from baseline and the higher the spike.




•   On B-scan, the greater the perpendicularity, the brighter the dots
    on the surface of the area of interest




                                                                     15
•   Because various parts of the eye and various pathologies are
    different in size and shape, understanding this concept and
    anticipating the best possible display for that eye are important.



•   Perpendicularity to the area of interest should be maintained to
    achieve the strongest echo possible for that structure




                                                                         16
Absorption

•   Ultrasound is absorbed by every medium through which it
    passes.



•   The more dense the medium, the greater the amount of
    absorption.



•   This means that the density of the solid lid structure results in
    absorption of part of the sound wave when B-scan is
    performed through the closed eye, thereby compromising the
    image of the posterior segment


                                                                        17
•   Therefore, B-scan should be performed on the open eye unless the
    patient is a small child or has an open wound   .

•   Likewise, when performing an ultrasound through a dense cataract
    as opposed to the normal crystalline lens, more of the sound is
    absorbed by the dense cataractous lens and less is able to pass
    through to the next medium, resulting in weaker echoes and images
    on both A-scan and B-scan. For this reason, the best images of
    the posterior segment are obtained when the probe is in
    contact with the sclera rather than the corneal surface,
    bypassing the crystalline lens or intraocular lens implant   .
                                                                     18
Instrumentation

•   Ophthalmic ultrasound instruments use what is known as a pulse-
    echo system, which consists of a series of emitted pulses of sound,
    each followed by a brief pause (microseconds) for the receiving of
    echoes and processing to the display screen.

•   The amplification of the display can be altered by adjusting the gain,
    which is measured in decibels (dB). Adjusting the gain in no way
    changes the frequency or velocity of the sound wave but acts to
    change the sensitivity of the instrument's display screen.
•    When the gain is high, weaker signals are displayed, such as
    vitreous opacities and posterior vitreous detachments.
•    When the gain is low, the weaker signals disappear, and only
    the stronger echoes, such as the retina, remain on the screen.

                                                                        19
•   Typically, all examinations begin on highest gain so that no
    weak signals are missed; then, the gain is reduced as necessary
    for good resolution of the stronger signals




                                                                      20
•   The probe face is usually oval in shape and when placed on the
    globe is represented by the initial white line on the left side of the
    display screen. The vitreous cavity is displayed in the center of
    the echogram, and the posterior pole is displayed on the right side
    of the echogram




                                                                             21
Indications
•   when direct visualization of intraocular structures is difficult or
    impossible.

•   Situations that prevent normal examination
    lid problems (eg, severe edema, partial or total tarsorrhaphy),
    corneal opacities (eg, scars, severe edema),
     hyphema, hypopyon,
     miosis, pupillary membranes
     dense cataracts
     vitreous opacities (eg, hemorrhage, inflammatory debris).




                                                                          22
Normal USG




             23
Clinical Applications
Differentiation between VH & asteroid Hyalosis:




•   AH is highly ecogenic,they are still visible when the gain setting is
    reduced upto 60dB whereas VH which usually disappears by 60 dB




                                                                        24
Asteriod Hyalosis   Vitreous Haemorrhage
                                           25
Vitreous Inflammation

USG is very helpful in assessing the severity and extent of intraocular
  inflammation in a patient suspected of having endophthalmitis.

VITRITIS appears in B-scan as scattered particle or large
  aggregates.

sometimes in absence of external inflammatory signs, it is important
  to differentiate between endophthalmitis and vitreous
  hemorrhage. VH is generally associated with PVD and layering
  of blood in inferior portion of the eye to produce sheet-like
  echoes



                                                                          26
27
PVD
•   In PVD with normal eye, the reflectivity is very low, high
    gain(90dB) setting is required the reflectivity disappears
    lowering the sensitivity,under 70 dB.

•   It should be kept in mind that PVD with hemorrhage shows
    extremely high reflectivity   .

•   Kinetic echography typically shows a very undulating movement
    that continue after the eye movements stops, which
    differentiates PVD from less mobile retinal and choroidal
    detachments


                                                                 28
RD vs PVD
In presence of opaque media,the differentiation between PVD and RD is
challenging. Few points are-
1) RD is usually uniformly high reflective and of even thickness whereas
    tilting of probe in different direction may reveal uneven thickness &
    reflectivity of membranes in PVD,

2) The image of PVD will disappear from the screen at higher gain setting
(70dB) than a RD(40-50dB)

4) PVD may appear as a line with multiple discontinuities or may be
    completely detached from ON.Rhegmatogenous RD:appear as a mobile
    membrane attached anterior to the ora serrata and posterior to the ON
    head



.                                                                           29
4) On kinetic Echography,a PVD has much more after
   movements when compared to RD.

The mobility of RD depends on
duration of the detachment. Recent bullous RD may be highly
    mobile,whereas chronic RD with proliferative Vitreoretinopathy
    appear stiff.




                                                                     30
31
32
•   Choroidal Detachment: CD appear as smooth, convex elevations
    from the posterior eye wall. In massive CD, choroids from opposite
    fundus areas may touch in the middle of the vitreous cavity-“Kissing
    Choroid”




                                                                       33
34
35
36
37
38
Posterior Scleritis
                  Subtenon fluid




                        OPTIC NERVE




      T-sign                       39
40
41
42
43
44
45
THANK   YOU




              46

Contenu connexe

Tendances

Retinoscopy @adi
Retinoscopy @adiRetinoscopy @adi
Retinoscopy @adi
Farhana Adi
 

Tendances (20)

Slit lamp techniques.pptx
Slit lamp techniques.pptxSlit lamp techniques.pptx
Slit lamp techniques.pptx
 
Keratometry
KeratometryKeratometry
Keratometry
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
 
Corneal Topography
Corneal TopographyCorneal Topography
Corneal Topography
 
A quick guide to Ophthalmic Ultrasound/ B-Scan interpretation
A quick guide to Ophthalmic Ultrasound/ B-Scan interpretation A quick guide to Ophthalmic Ultrasound/ B-Scan interpretation
A quick guide to Ophthalmic Ultrasound/ B-Scan interpretation
 
Measuring interpupillary distance
Measuring interpupillary distanceMeasuring interpupillary distance
Measuring interpupillary distance
 
Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Measuring interpupillary distance
Measuring interpupillary distanceMeasuring interpupillary distance
Measuring interpupillary distance
 
Non contact tonometer ppt
Non contact tonometer pptNon contact tonometer ppt
Non contact tonometer ppt
 
corneal Pachymetry
 corneal Pachymetry corneal Pachymetry
corneal Pachymetry
 
Slit Lamp Illumination Techniques
Slit Lamp Illumination TechniquesSlit Lamp Illumination Techniques
Slit Lamp Illumination Techniques
 
Optical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --maculaOptical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --macula
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
 
Retinoscopy and its principles
Retinoscopy and its principlesRetinoscopy and its principles
Retinoscopy and its principles
 
SLIT LAMP
SLIT LAMPSLIT LAMP
SLIT LAMP
 
Retinoscopy @adi
Retinoscopy @adiRetinoscopy @adi
Retinoscopy @adi
 
Keratometry
KeratometryKeratometry
Keratometry
 
Retinoscopy
RetinoscopyRetinoscopy
Retinoscopy
 

En vedette

Presentation1.pptx, ultrasound examination of the uterus and ovaries.
Presentation1.pptx, ultrasound examination of the uterus and ovaries.Presentation1.pptx, ultrasound examination of the uterus and ovaries.
Presentation1.pptx, ultrasound examination of the uterus and ovaries.
Abdellah Nazeer
 
Presentation1.pptx, radilogical imaging of ovarian lesions.
Presentation1.pptx, radilogical imaging of ovarian lesions.Presentation1.pptx, radilogical imaging of ovarian lesions.
Presentation1.pptx, radilogical imaging of ovarian lesions.
Abdellah Nazeer
 
Presentation1.pptx, ultrasound examination of the 1st trimester pregnancy.
Presentation1.pptx, ultrasound examination of the 1st trimester pregnancy.Presentation1.pptx, ultrasound examination of the 1st trimester pregnancy.
Presentation1.pptx, ultrasound examination of the 1st trimester pregnancy.
Abdellah Nazeer
 
Atlas of opthalmology_tanta_university
Atlas of opthalmology_tanta_universityAtlas of opthalmology_tanta_university
Atlas of opthalmology_tanta_university
DrAfiqahMF
 
Presentation1.pptx, ultrasound examination of the 2nd & 3rd trimester pregnancy.
Presentation1.pptx, ultrasound examination of the 2nd & 3rd trimester pregnancy.Presentation1.pptx, ultrasound examination of the 2nd & 3rd trimester pregnancy.
Presentation1.pptx, ultrasound examination of the 2nd & 3rd trimester pregnancy.
Abdellah Nazeer
 
Presentation1.pptx, ultrasound examination of the orbit.
Presentation1.pptx, ultrasound examination of the orbit.Presentation1.pptx, ultrasound examination of the orbit.
Presentation1.pptx, ultrasound examination of the orbit.
Abdellah Nazeer
 
Ultrasonography in ophthalmology
Ultrasonography in ophthalmologyUltrasonography in ophthalmology
Ultrasonography in ophthalmology
Barun Garg
 
Presentation1.pptx, ultrasound examination of the adrenal glands and kidneys.
Presentation1.pptx, ultrasound examination of the adrenal glands and kidneys.Presentation1.pptx, ultrasound examination of the adrenal glands and kidneys.
Presentation1.pptx, ultrasound examination of the adrenal glands and kidneys.
Abdellah Nazeer
 

En vedette (14)

Ophthalmic ultrasound
Ophthalmic ultrasoundOphthalmic ultrasound
Ophthalmic ultrasound
 
Level II usg
Level II usgLevel II usg
Level II usg
 
Breast lecture
Breast  lectureBreast  lecture
Breast lecture
 
Ophthalmic USG
Ophthalmic USGOphthalmic USG
Ophthalmic USG
 
Presentation1.pptx, ultrasound examination of the uterus and ovaries.
Presentation1.pptx, ultrasound examination of the uterus and ovaries.Presentation1.pptx, ultrasound examination of the uterus and ovaries.
Presentation1.pptx, ultrasound examination of the uterus and ovaries.
 
Eye ultrasound
Eye ultrasoundEye ultrasound
Eye ultrasound
 
Presentation1.pptx, radilogical imaging of ovarian lesions.
Presentation1.pptx, radilogical imaging of ovarian lesions.Presentation1.pptx, radilogical imaging of ovarian lesions.
Presentation1.pptx, radilogical imaging of ovarian lesions.
 
Presentation1.pptx, ultrasound examination of the 1st trimester pregnancy.
Presentation1.pptx, ultrasound examination of the 1st trimester pregnancy.Presentation1.pptx, ultrasound examination of the 1st trimester pregnancy.
Presentation1.pptx, ultrasound examination of the 1st trimester pregnancy.
 
Atlas of opthalmology_tanta_university
Atlas of opthalmology_tanta_universityAtlas of opthalmology_tanta_university
Atlas of opthalmology_tanta_university
 
Ct orbit
Ct orbitCt orbit
Ct orbit
 
Presentation1.pptx, ultrasound examination of the 2nd & 3rd trimester pregnancy.
Presentation1.pptx, ultrasound examination of the 2nd & 3rd trimester pregnancy.Presentation1.pptx, ultrasound examination of the 2nd & 3rd trimester pregnancy.
Presentation1.pptx, ultrasound examination of the 2nd & 3rd trimester pregnancy.
 
Presentation1.pptx, ultrasound examination of the orbit.
Presentation1.pptx, ultrasound examination of the orbit.Presentation1.pptx, ultrasound examination of the orbit.
Presentation1.pptx, ultrasound examination of the orbit.
 
Ultrasonography in ophthalmology
Ultrasonography in ophthalmologyUltrasonography in ophthalmology
Ultrasonography in ophthalmology
 
Presentation1.pptx, ultrasound examination of the adrenal glands and kidneys.
Presentation1.pptx, ultrasound examination of the adrenal glands and kidneys.Presentation1.pptx, ultrasound examination of the adrenal glands and kidneys.
Presentation1.pptx, ultrasound examination of the adrenal glands and kidneys.
 

Similaire à USG B Scan

ULTRASONOGRAPHY (USG) AND ULTRASOUND BIOMICROSCOPY(UBM)
ULTRASONOGRAPHY (USG)  AND ULTRASOUND BIOMICROSCOPY(UBM)ULTRASONOGRAPHY (USG)  AND ULTRASOUND BIOMICROSCOPY(UBM)
ULTRASONOGRAPHY (USG) AND ULTRASOUND BIOMICROSCOPY(UBM)
Dr. Gaurav Shukla
 
ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT  ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT
PAWAN JARWAL
 
Ultrsonography Principle and application
Ultrsonography Principle and applicationUltrsonography Principle and application
Ultrsonography Principle and application
suniu
 
3_2018_12_16!11_22_57_PM.pptx
3_2018_12_16!11_22_57_PM.pptx3_2018_12_16!11_22_57_PM.pptx
3_2018_12_16!11_22_57_PM.pptx
KhaVuong2
 
Point of Care Ultrasound - Hyperechoic Future in Medical School?
Point of Care Ultrasound - Hyperechoic Future in Medical School?Point of Care Ultrasound - Hyperechoic Future in Medical School?
Point of Care Ultrasound - Hyperechoic Future in Medical School?
cbyrne2014
 

Similaire à USG B Scan (20)

BScan and Ascan in ophthalmology and eye field
BScan and Ascan  in ophthalmology and eye fieldBScan and Ascan  in ophthalmology and eye field
BScan and Ascan in ophthalmology and eye field
 
B-Scan.pptx
B-Scan.pptxB-Scan.pptx
B-Scan.pptx
 
B scan
B scanB scan
B scan
 
B scan ultrasonography
B scan ultrasonographyB scan ultrasonography
B scan ultrasonography
 
Ultrasound in Ophthalmology.pptx
Ultrasound in Ophthalmology.pptxUltrasound in Ophthalmology.pptx
Ultrasound in Ophthalmology.pptx
 
Ultrasonography of eye
Ultrasonography of eyeUltrasonography of eye
Ultrasonography of eye
 
ULTRASONOGRAPHY (USG) AND ULTRASOUND BIOMICROSCOPY(UBM)
ULTRASONOGRAPHY (USG)  AND ULTRASOUND BIOMICROSCOPY(UBM)ULTRASONOGRAPHY (USG)  AND ULTRASOUND BIOMICROSCOPY(UBM)
ULTRASONOGRAPHY (USG) AND ULTRASOUND BIOMICROSCOPY(UBM)
 
DOC-20230615-WA0003..pptx
DOC-20230615-WA0003..pptxDOC-20230615-WA0003..pptx
DOC-20230615-WA0003..pptx
 
B scan by Pushkar Dhir
B scan by Pushkar DhirB scan by Pushkar Dhir
B scan by Pushkar Dhir
 
B scan
B scanB scan
B scan
 
Usg
UsgUsg
Usg
 
ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT  ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT
 
Ultra sound by Dr nada alarb
Ultra sound by Dr nada alarbUltra sound by Dr nada alarb
Ultra sound by Dr nada alarb
 
Ultrsonography Principle and application
Ultrsonography Principle and applicationUltrsonography Principle and application
Ultrsonography Principle and application
 
3_2018_12_16!11_22_57_PM.pptx
3_2018_12_16!11_22_57_PM.pptx3_2018_12_16!11_22_57_PM.pptx
3_2018_12_16!11_22_57_PM.pptx
 
Ocular Ultrasound
Ocular UltrasoundOcular Ultrasound
Ocular Ultrasound
 
Ocular Ultrasonography/ Ophthalmic Ultrasonography (Ocular USG/ Ophthalmic US...
Ocular Ultrasonography/ Ophthalmic Ultrasonography (Ocular USG/ Ophthalmic US...Ocular Ultrasonography/ Ophthalmic Ultrasonography (Ocular USG/ Ophthalmic US...
Ocular Ultrasonography/ Ophthalmic Ultrasonography (Ocular USG/ Ophthalmic US...
 
Point of Care Ultrasound - Hyperechoic Future in Medical School?
Point of Care Ultrasound - Hyperechoic Future in Medical School?Point of Care Ultrasound - Hyperechoic Future in Medical School?
Point of Care Ultrasound - Hyperechoic Future in Medical School?
 
B SCAN BY DR.PUSHKAR DHIR, DHIR HOSPITAL BHIWANI.pptx
B SCAN BY DR.PUSHKAR DHIR, DHIR HOSPITAL BHIWANI.pptxB SCAN BY DR.PUSHKAR DHIR, DHIR HOSPITAL BHIWANI.pptx
B SCAN BY DR.PUSHKAR DHIR, DHIR HOSPITAL BHIWANI.pptx
 
OCT
OCTOCT
OCT
 

Dernier

Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 

Dernier (20)

Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
 
Coimbatore Call Girls in Coimbatore 7427069034 genuine Escort Service Girl 10...
Coimbatore Call Girls in Coimbatore 7427069034 genuine Escort Service Girl 10...Coimbatore Call Girls in Coimbatore 7427069034 genuine Escort Service Girl 10...
Coimbatore Call Girls in Coimbatore 7427069034 genuine Escort Service Girl 10...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
 
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
 
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
 
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
 

USG B Scan

  • 1. USG B SCAN Dr.Gyanendra Lamichhanae Vitreo retinal Fellow, Gunma University ,JAPAN Lumbini Eye Institute, Bhairahawa
  • 2. What is ultrasound • sound pressure with a frequency greater than the upper limit of human hearing. • Although this limit varies from person to person, it is approximately 20 kilohertz (20,000 hertz) in healthy, young adults 2
  • 3. 3
  • 4. • B-scan ultrasonography is an important noninvasive technique for the clinical assessment of various ocular and orbital diseases 4
  • 5. HISTORY • 1793: Lazzaro Spallanzani (Italy) discovered that bats orient themselves with the help of sound whistles while flying in darkness. This was the basis of modern ultrasound application Bats use ultrasounds to navigate in the darkness 5
  • 6. History contd……. • World war II: a device based on piezoelectric effect developed by Paul Langevin (France) ,able of emitting & receiving ultrasound under water used as sonar. •  1956: first documented use of ocular USG, Mundt and Hughes used A scan technique to detect intraocular tumour. •  1972: First use of hand held B scan by Bronston & workers ,which was applied directly to the closed lid without a water bath 6
  • 7. Principles of ultrasound: • By definition, an ultrasound wave has a frequency greater than 20 kHz (20,000 oscillations/ second) • As the frequency of USG increases, the wavelength decreases and wavelength of an ultrasound determines its depth of tissue penetration and resolution Wavelength α Depth of penetration of the ultrasound • So, Larger is the frequency of US = shorter is its wavelength = shallower is its penetration = better is the resolution of resultant echo graph. 7
  • 8. That’s why USG probes used for Ocular USG are of higher frequency(10MHz)as it needs much less tissue penetration (an eye is 23.5 mm long on average) & higher resolution. • In contrast, ultrasound probes used for purposes such as obstetrics, use lower frequencies (1-5Hz) for deeper penetration into the body, and, because the structures being imaged are larger, they do not require the same degree of resolution 8
  • 9. Ultrasound Principles and Physics Ophthalmic ultrasonography uses high-frequency sound waves, transmitted from a probe into the eye. As the sound waves strike intraocular structures, they are reflected back to the probe and converted into an electric signal. The signal is subsequently reconstructed as an image on a monitor, 9
  • 10. Velocity • The velocity of the sound wave is dependent on the density of the medium through which the sound travels. • Sound travels faster through solids than liquids, an important principle to understand since the eye is composed of both. • There are known velocities of different components of the eye, with sound traveling through both aqueous and vitreous at a speed of 1,532 meters/second (m/s) and through the cornea and lens at an average speed of 1,641 m/s 10
  • 11. Reflectivity • When sound travels from one medium to another medium of different density, part of the sound is reflected from the interface between those media back into the probe. • This is known as an echo; the greater the density difference at that interface, the stronger the echo, or the higher the reflectivity • In A-scan ultrasonography, a thin, parallel sound beam is emitted, which passes through the eye and images one small axis of tissue; the echoes of which are represented as spikes arising from a baseline. The stronger the echo, the higher the spike 11
  • 12. 12
  • 13. In B-scan ultrasonography, an oscillating sound beam is emitted, passing through the eye and imaging a slice of tissue; the echoes of which are represented as a multitude of dots that together form an image on the screen. The stronger the echo, the brighter the dot. example, the dots that form the posterior vitreous hyaloid membrane are not as bright as the dots that form the retinal membrane. This is very useful in differentiating a posterior vitreous detachment (a benign condition) from a more highly reflective retinal detachment (a blinding condition) because retina is more dense than vitreous. 13
  • 14. Angle of incidence • The angle of incidence of the probe is critical for both A-scan and B- scan ultrasonography. • When the probe is held perpendicular to the area of interest, more of the echo is reflected directly back into the probe tip and sent to the display screen. • When held oblique to the area imaged, part of the echo is reflected away from the probe tip and less is sent to the display screen. • The more oblique the probe is held from the area of interest, the weaker the returning echo and, thus, the more compromised the displayed image. . 14
  • 15. On A-scan, the greater the perpendicularity, the more steeply rising the spike is from baseline and the higher the spike. • On B-scan, the greater the perpendicularity, the brighter the dots on the surface of the area of interest 15
  • 16. Because various parts of the eye and various pathologies are different in size and shape, understanding this concept and anticipating the best possible display for that eye are important. • Perpendicularity to the area of interest should be maintained to achieve the strongest echo possible for that structure 16
  • 17. Absorption • Ultrasound is absorbed by every medium through which it passes. • The more dense the medium, the greater the amount of absorption. • This means that the density of the solid lid structure results in absorption of part of the sound wave when B-scan is performed through the closed eye, thereby compromising the image of the posterior segment 17
  • 18. Therefore, B-scan should be performed on the open eye unless the patient is a small child or has an open wound . • Likewise, when performing an ultrasound through a dense cataract as opposed to the normal crystalline lens, more of the sound is absorbed by the dense cataractous lens and less is able to pass through to the next medium, resulting in weaker echoes and images on both A-scan and B-scan. For this reason, the best images of the posterior segment are obtained when the probe is in contact with the sclera rather than the corneal surface, bypassing the crystalline lens or intraocular lens implant . 18
  • 19. Instrumentation • Ophthalmic ultrasound instruments use what is known as a pulse- echo system, which consists of a series of emitted pulses of sound, each followed by a brief pause (microseconds) for the receiving of echoes and processing to the display screen. • The amplification of the display can be altered by adjusting the gain, which is measured in decibels (dB). Adjusting the gain in no way changes the frequency or velocity of the sound wave but acts to change the sensitivity of the instrument's display screen. • When the gain is high, weaker signals are displayed, such as vitreous opacities and posterior vitreous detachments. • When the gain is low, the weaker signals disappear, and only the stronger echoes, such as the retina, remain on the screen. 19
  • 20. Typically, all examinations begin on highest gain so that no weak signals are missed; then, the gain is reduced as necessary for good resolution of the stronger signals 20
  • 21. The probe face is usually oval in shape and when placed on the globe is represented by the initial white line on the left side of the display screen. The vitreous cavity is displayed in the center of the echogram, and the posterior pole is displayed on the right side of the echogram 21
  • 22. Indications • when direct visualization of intraocular structures is difficult or impossible. • Situations that prevent normal examination lid problems (eg, severe edema, partial or total tarsorrhaphy), corneal opacities (eg, scars, severe edema), hyphema, hypopyon, miosis, pupillary membranes dense cataracts vitreous opacities (eg, hemorrhage, inflammatory debris). 22
  • 24. Clinical Applications Differentiation between VH & asteroid Hyalosis: • AH is highly ecogenic,they are still visible when the gain setting is reduced upto 60dB whereas VH which usually disappears by 60 dB 24
  • 25. Asteriod Hyalosis Vitreous Haemorrhage 25
  • 26. Vitreous Inflammation USG is very helpful in assessing the severity and extent of intraocular inflammation in a patient suspected of having endophthalmitis. VITRITIS appears in B-scan as scattered particle or large aggregates. sometimes in absence of external inflammatory signs, it is important to differentiate between endophthalmitis and vitreous hemorrhage. VH is generally associated with PVD and layering of blood in inferior portion of the eye to produce sheet-like echoes 26
  • 27. 27
  • 28. PVD • In PVD with normal eye, the reflectivity is very low, high gain(90dB) setting is required the reflectivity disappears lowering the sensitivity,under 70 dB. • It should be kept in mind that PVD with hemorrhage shows extremely high reflectivity . • Kinetic echography typically shows a very undulating movement that continue after the eye movements stops, which differentiates PVD from less mobile retinal and choroidal detachments 28
  • 29. RD vs PVD In presence of opaque media,the differentiation between PVD and RD is challenging. Few points are- 1) RD is usually uniformly high reflective and of even thickness whereas tilting of probe in different direction may reveal uneven thickness & reflectivity of membranes in PVD, 2) The image of PVD will disappear from the screen at higher gain setting (70dB) than a RD(40-50dB) 4) PVD may appear as a line with multiple discontinuities or may be completely detached from ON.Rhegmatogenous RD:appear as a mobile membrane attached anterior to the ora serrata and posterior to the ON head . 29
  • 30. 4) On kinetic Echography,a PVD has much more after movements when compared to RD. The mobility of RD depends on duration of the detachment. Recent bullous RD may be highly mobile,whereas chronic RD with proliferative Vitreoretinopathy appear stiff. 30
  • 31. 31
  • 32. 32
  • 33. Choroidal Detachment: CD appear as smooth, convex elevations from the posterior eye wall. In massive CD, choroids from opposite fundus areas may touch in the middle of the vitreous cavity-“Kissing Choroid” 33
  • 34. 34
  • 35. 35
  • 36. 36
  • 37. 37
  • 38. 38
  • 39. Posterior Scleritis Subtenon fluid OPTIC NERVE T-sign 39
  • 40. 40
  • 41. 41
  • 42. 42
  • 43. 43
  • 44. 44
  • 45. 45
  • 46. THANK YOU 46