Nuclear Associates 76-903
                                                                        MRI Multi-Purpose Phantom




                                                                                   Users Manual




March 2005
Manual No. 76-903-1 Rev. 2
©2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies
Fluke Biomedical
Radiation Management Services

6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564

www.flukebiomedical.com/rms
Table of Contents
Section 1:   Introduction................................................................................................ 1-1
   1.1       Introduction .................................................................................................. 1-1
   1.2       Orientation – Imaging Method...................................................................... 1-1
   1.3       Choosing Echo Time.................................................................................... 1-1
   1.4       NMR Solution............................................................................................... 1-2

Section 2:   Specifications ............................................................................................ 2-1
   2.1       Phantom Design .......................................................................................... 2-1

Section 3:   Theory of Operation................................................................................... 3-1
   3.1       Omni Directional Application........................................................................ 3-1
   3.2       Concentric Conical Rings Section................................................................ 3-1
   3.2.1       Slice Thickness Uniformity ........................................................................ 3-1
   3.2.2       Aspect Ratio.............................................................................................. 3-1
   3.2.3       Magnetic Field Homogeneity..................................................................... 3-2
   3.3       Flood Section............................................................................................... 3-2
   3.3.1       RF Uniformity ............................................................................................ 3-3
   3.3.2       T1 and T2 Values ....................................................................................... 3-3
   3.4       Insert Section............................................................................................... 3-3
   3.4.1        Folded Step Ramps................................................................................ 3-4
   3.4.2        Fold Step Ramp Analysis ....................................................................... 3-4
   3.4.3        Measurement Method............................................................................. 3-4
   3.4.4        Slice Contiguity (Gap)............................................................................. 3-5
   3.4.5        Resolution Pattern .................................................................................. 3-5
   3.4.6        Star Pattern ............................................................................................ 3-6




                                                         i
(Blank Page)
Introduction
                                                                                            Introduction     1

                                                                                            Section 1
                                                                                   Introduction

1.1 Introduction
This phantom is designed for comprehensive evaluation of critical imaging parameters of magnetic
resonance imaging (MRI) in a time efficient manner. Particular attention was paid to obtaining all the
necessary image quality information with a minimum number of imaging sequences. It is designed to
study parameters unique to multi-slice two-dimensional and three-dimensional Fourier imaging.
The overall dimensions of the phantom adjust so that it can be positioned in each of the three major
imaging planes: axial, coronal and sagittal.
The phantom can be used for the measurement of absolute values for calibration purposes. However, its
design is optimized for time efficient daily quality assurance.
The performance parameters that can be evaluated using the phantom include: slice orientation, slice
thickness, interslice distance, aspect ratio, magnetic field uniformity, radio-frequency signal uniformity,
spatial resolution, T1 and T2 values, and modulation transfer function. The phantom can also be used to
evaluate such parameters as resonant frequency, radio frequency power setting, radio frequency
quadrature setting, magnetic field gradient calibration, slice off-set parameters, image off-set, and other
image artifacts peculiar to MRI.


1.2 Orientation Imaging Method
In a two-dimensional Fourier transform (2DFT) imaging of MRI, the data is collected in two orthogonal
directions, i.e., encoding and readout.
In some imaging sequences, the spatial resolution may be different in the two directions. The star pattern
is designed to study such asymmetric resolution as well as qualitative modulation transfer function (MTF)
in MR images. For this reason, it is important to properly orient the star pattern. Note four lines or dots on
the wall of the phantom.
Place the phantom in the imaging system SO that one is in either the phase encoding or frequency
readout direction.


1.3 Choosing Echo Time
Any clinical multi-slice imaging sequence can be used. Often, however, the imaging sequences with the
shortest echo time (TE), largest number of slices possible within the pulse repetition time (TR), and single
excitation (no averaging) will demonstrate the performance of an imaging system under one of the most
demanding situations. For a daily QA program, it is important to record all the important imaging
parameters. An example of such a table of imaging parameters can be found in Table 1-1.




                                                                                                           1-1
Nuclear Associates 76-903
Operators Manual


1.4 NMR Solution
Any NMR signal-producing solution can be used to fill the phantom. It is desirable to use solutions with T1
and T2 values around 500 - 1000 sec, because it can better simulate the clinical imaging conditions. One
should note that
T1 and T2 values depend on the magnetic field strength.
Filling the phantom requires special care to avoid any trapped air bubbles. Water can be degassed by first
boiling and cooling prior to filling. Use of a wetting agent and algicide can also prolong the useful life of
the solution. A fairly full phantom can be placed in a vacuum chamber to eliminate the bubbles. One way
to avoid the trapped bubbles, especially in small holes, is to fill the holes with solution using a small
syringe needle. This may require disassembling the phantom.
Any solution can be used; however, for higher field systems, solutions with chemical shift should be
avoided. Copper sulfite or manganese chloride were found to be useful.
A water solution of 0.1 millimolar manganese chloride can be used to fill the phantom. The T1 value of this
solution is approximately 500 msec. A shorter relaxation time is not suggested, as it may not correctly
represent the system performance in a realistic clinical environment.

Table 1-1. Summary of Imaging Parameters
Orientation:
Imaging Technique:
TE= Shortest possible
TR=
Number of Slices =
Slice Thickness =
Slice Sequence = Interleaved or others
Slice to Slice Gap =
Number of Encoding = 256, 128 or 64
Signal Averaging = No averaging, single
RF Tuning Parameters (i.e. attenuation settings):

                                                    NOTE
                         For daily QA, exactly same parameters using same
                         pulse sequence should be used.




1-2
Specifications
                                                                                      Phantom Design      2

                                                                                           Section 2
                                                                              Specifications

2.1 Phantom Design
All of the parts in the phantom are made with plastic and glass, free of magnetic material. The basic
design of the phantom is a 9” diameter plexiglass cylindrical container with various inserts filled with an
NMR signal producing solution. The phantom is designed to fit inside a head coil. The cylinder has three
sections, flood, cone, and insert section (Figure 2-1). The flood section is a typical flood phantom that can
be filled with a solution.




Figure 2-1.

The conic section has four concentric cones (Figure 2-2). The concentric conical grooves of 45° are cut
from a solid plexiglass block.




Figure 2-2.




                                                                                                        2-1
Nuclear Associates 76-903
Operators Manual

The insert section has six inserts, one 120° star pattern wedge assembly, two folding step wedges,
                                                      spatial resolution insert, and T1 and T2 sample vials
                                                      (Figure 2-3).




Figure 2-3.

                                                   NOTE

                         Addendum for Model 76-903: MRI Multi-Purpose
                         Phantom: The negative contrast resolution insert
                         (rods), as seen in Figure 2-3 has been discontinued
                         and replaced by a T1 and T2 sample vial insert, as
                         described in Figure 3-13.




2-2
Theory of Operation
                                                                         Omni Directional Application     3

                                                                                         Section 3
                                                               Theory Of Operation

3.1 Omni Directional Application
Since MR images are obtained not only in axial but also sagittal and coronal orientation, the phantom is
designed to be placed in all three directions. The selection of inserts, their sizes and arrangements are
made to obtain the maximum amount of imaging information in any orientation in a single scan.


3.2 Concentric Conical Rings Section
The concentric conical rings are machined by cutting out a circular groove at an angle of 45°, as shown
(Figure 3-1).




Figure 3-1. Outside diameter, 9, Number of segments, 4

The conical section can provide the following information by simple visual inspection:
 •    Slice thickness uniformity
 •    Aspect ratio
 •    Magnetic field homogeneity


3.2.1 Slice Thickness Uniformity
The dark and light concentric rings in the image represent the slice thickness multiplied by the square root
of 2.0 due to the 45° grooves. The width of the rings (bands) should be the same throughout the image
for uniform slice thickness. More careful analysis can be carried out by measuring the brightness profiles
with region of interest computer routines.

3.2.2 Aspect Ratio
Visual inspection followed by a measurement of the diameter of the outermost ring can provide
information on the aspect ratio. The video display should be checked independently to avoid any
distortions in the display screen.




                                                                                                        3-1
Nuclear Associates 76-903
Operators Manual

Figure 3-2 shows the cross section of the concentric conical part of the phantom. The aspect ratio is
correct, as is the uniformity of the slice thickness as represented by the uniform widths of dark and light
                                                  circular bands.




Figure 3-2. Cross Section Image

3.2.3 Magnetic Field Homogeneity
Any distortions in the roundness of the rings can be a result of non-uniformity in the main magnetic field,
as shown (Figure 3-3).




Figure 3-3.


3.3 Flood Section
This section is a typical flood phantom, 9" in diameter by 1.25" thick. It may be filled with any signal
producing solution (Figure 3-4).
  •   RF Signal Uniformity
  •   Single T1 and T2 Values




3-2
Theory of Operation
                                                                                        Flood Section     3




Figure 3-4. Diameter, 9", Depth, 1.25"

3.3.1 RF Uniformity
The flood section is used to measure RF uniformity and to visualize artifacts that can come from many
different sources (Figure 3-5).




Figure 3-5.

Radio-frequency (brightness) non-uniformity can come from various sources. Most RF coils have a
number of active wires and the regions near the wires can show increased signal intensity. It can also
come from magnetic field uniformity, eddy current problems and poor RF power setting. RF uniformity can
also be affected by internal motion of the solution. One should also look for any aliasing, fold-over or
ghosts.

3.3.2T1 and T2 Values
A single value of T1 and T2 can be measured using the flood section. It should be pointed out that T1 and
T2 values depend on magnetic field strength, temperature, imaging sequence, and they are very difficult to
measure precisely. Extreme caution should be taken for RF power setting.
An insert with 6 vials is available SO that various solutions can be imaged for a range of relaxation values.


3.4 Insert Section
This section has the following inserts:
  1.   Folded Step Ramps: Slice thickness and slice contiguity (gap), phantom orientation.
  2.   Resolution Pattern: Square hole pattern for positive contrast.



                                                                                                        3-3
Nuclear Associates 76-903
Operators Manual

  3.   Star Pattern: Modulation transfer function (MTF), horizontal and vertical spatial resolution,
       quadrature setting and baseline correction.

3.4.1 Folded Step Ramps
Accurate slice thickness measurement in MR is very difficult and various methods are under discussion.
The slice profile (shape of slice thickness) is also a very complex subject that needs further discussion
and definition by the physics community. This folded ramp approach is adopted in this phantom because
of its ease of use and efficiency in time. For a quick evaluation, a simple counting of the number of visible
steps will give an estimate of slice thickness in millimeters.
Slice parameters are studied using a pair of folding step wedges. The height of the steps range from 16
mm to
40 mm and the heights vary every 1.0 mm (Figure 3-6). Each step area has a dimension of 10 mm x 10
mm. The wedges were folded to accommodate a greater number of steps in the limited space available.
Two inserts of the folded steps are placed at opposite sides of the phantom so that slice orientation with
respect to the phantom can be studied.
  1.   Slice Thickness
  2.   Slice Orientation
  3.   Interslice Gap




Figure 3-6.       Step Interval, 1 mm Range, 36 mm


3.4.2 Fold Step Ramp Analysis
Quick Visual Evaluation
Slice thickness can be evaluated quickly by counting the number of steps that can be visualized at a
relatively wide window setting (Figure 3-7). The number of visible steps is equal to the slice thickness in
millimeters since the height of each step varies by 1 millimeter. Record the window setting for future
reference, for the visible steps vary depending on the window setting.
In Figure 3-7, five gray scale steps can be seen in the 10 o’clock and 5 o'clock positions, representing a
slice thickness of 6 mm.

3.4.3 Measurement Method
Better evaluation of slice thickness can be made by using the region of interest (ROI) routine in the imaging system.
The area of the region slightly smaller than the step size (10 mm x 10 mm) should be used to read out the signal
intensity (NMR number). Than take the difference of the "NMR" number of the adjacent steps. Plot a graph of the
difference in intensity on the vertical axis and step sequence numbers on the horizontal axis. The A line graph
connecting the dots (diff. vs. step number) represents slice profile and the width of the graph at half of its maximum
intensity (full width half maximum) is the measured slice thickness.




 3-4
Nuclear Associates 76-903
Operators Manual




Figure 3-7.

3.4.4 Slice Continuity (Gap)
The slice contiguity can be evaluated by quick visual inspection as well as in a more precise manner.
While visualizing the steps of one slice in the image, note the locations of the steps…especially the faint
steps. Then, repeat the same procedure viewing the next slice using a preset window. The number of
missing steps between the two adjacent images represents the gap in millimeters. It can be observed in
Figure 3-7 and Figure 3-8 that two steps are missing between the two slices, representing a 2 mm gap. A
more precise measurement can be made by plotting a graph of more than one slice on the same axis.




Figure 3-8

3.4.5 Resolution Pattern
Hole Pattern Solution Insert
The resolution insert (Figure 3-9) has 4 sets of square holes (0.5 mm, 0.75 mm, 1 mm and 2 mm
machined in an acrylic block. The spacing between the holes is the same as the size of the hole.
In filling the phantom with solution, care must be taken to avoid air bubbles becoming trapped in the
holes. This can be accomplished by squirting the solution into the holes using syringes with small
needles.




3-6
Nuclear Associates 76-903
Operators Manual

Figure 3-9.        0.5 mm x 0.5 mm2, 0.75 mm x 0.75 mm2, 1.0 mm x 1.0mm2, and 2.0 mm x 2.0mm2
              square holed.
              Five of each dimension. Hole depth, 0.5”.

T1 and T2 Solution Insert
This insert (Figure 3-10) holds six (6) 5 cc glass vials (15 mm diameter x 47 mm high) that can hold
solutions of varying T1 and T2 values. One should note that measured T1 and T2 values depend on many
parameters including field strength and imaging sequence used.




Figure 3-10.      T1, T2 Solution Insert. Six 5 cc refillable glass vials with caps. (15 mm diameter x 47 mm
height.)

3.4.6 Star Pattern
Modulation Transfer Function (MTF)
Modulation transfer function (MTF) is a mathematical concept that provides an objective measurement of
the combined effect of image sharpness and resolution. Direct measurement of MTF is often very difficult
and calculation of MTF based on measurable parameters such as point-spread function (PSF) or line
spread function (LSF) are clearly outside the scope of this manual. Difficult as it may be to have precise
mathematical definition, it is however relatively easy to present MTF in a qualitative manner. Similar
techniques are used to evaluate x-ray focal spot sizes using a star pattern.
The star pattern developed for this purpose is particularly useful in MR imaging where resolution in
vertical and horizontal axes can be different and certain image artifacts that are peculiar to Fourier
imaging may not be visible under any other test.
  •   MTF Evaluation
  •   Horizontal & Vertical Spatial Resolution
  •   Quadrature Setting
  •   Baseline Correction




Figure 3-11.       Wedge angle, 3°. Number of wedges, 9. Fan angle, 60°. Wedge length, 3". Width, 2".




3-6
Theory of Operation
                                                                                          Insert Section     3
The star pattern for MTF evaluation is made up of 3° wedges placed on a 60° fan angle. Each wedge is 3”
in the radial dimension and 2” in height. Two such sections are placed next to each other to cover an
overall fan angle of 120° (Figure 3-11).
A star pattern is a very useful way to measure the many subtle image artifacts that are unique to NMR
imaging. It may be difficult to pinpoint the exact source of the artifacts, but this is a powerful test to
demonstrate image artifacts that are otherwise difficult to detect.

Horizontal and Vertical Spatial Resolution
One can also observe asymmetric image resolution between the encoding and readout direction by
observing the resolving power in the star pattern (Figure 3-12). It can be seen that the vertical resolution
is superior to the horizontal resolution by observing that the narrow ends of the star pattern are better
                                               resolved in the vertical direction.




Figure 3-12.

Quadrature Setting
In Figure 3-13, one can visualize a faint star-like pattern in the central 12 o'clock to 5 o’clock position. The
star-like ghost pattern is in the opposite orientation of the real star pattern image. This is attributable to
the quadrature setting. In quadrature signal detection, which is widely used in MR imaging, the RF signal
is split into two orthogonal signals (real and imaginary) before digitization. If the phase angle between the
real and imaginary components is not 90°, a ghost of point symmetry appears along the diagonal axis
between the encoding and readout directions. This artifact also shows up as a "fold-over" in the star
pattern.




Figure 3-13.




                                                                                                           3-7
Fluke Biomedical
Radiation Management Services

6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564

www.flukebiomedical.com/rms

Contenu connexe

En vedette (6)

Modulo C - Radioprotezione per Odontoiatri
Modulo C - Radioprotezione per OdontoiatriModulo C - Radioprotezione per Odontoiatri
Modulo C - Radioprotezione per Odontoiatri
 
Modulo D - Radioprotezione per Odontoiatri
Modulo D - Radioprotezione per OdontoiatriModulo D - Radioprotezione per Odontoiatri
Modulo D - Radioprotezione per Odontoiatri
 
Modulo F - Radioprotezione per Odontoiatri
Modulo F - Radioprotezione per OdontoiatriModulo F - Radioprotezione per Odontoiatri
Modulo F - Radioprotezione per Odontoiatri
 
Modulo B - Radioprotezione per Odontoiatri
Modulo B - Radioprotezione per OdontoiatriModulo B - Radioprotezione per Odontoiatri
Modulo B - Radioprotezione per Odontoiatri
 
Modulo A - Radioprotezione per Odontoiatri
Modulo A - Radioprotezione per OdontoiatriModulo A - Radioprotezione per Odontoiatri
Modulo A - Radioprotezione per Odontoiatri
 
Modulo E - Radioprotezione per Odontoiatri
Modulo E - Radioprotezione per OdontoiatriModulo E - Radioprotezione per Odontoiatri
Modulo E - Radioprotezione per Odontoiatri
 

Similaire à Manuale fantoccio mri 76 903

APE_IM_Article_AppliedOptics_Vol47_No29_Oct10_2008_AsPublished
APE_IM_Article_AppliedOptics_Vol47_No29_Oct10_2008_AsPublishedAPE_IM_Article_AppliedOptics_Vol47_No29_Oct10_2008_AsPublished
APE_IM_Article_AppliedOptics_Vol47_No29_Oct10_2008_AsPublished
Rainer Wilhelm
 
Gp E Getting Started Din
Gp E Getting Started DinGp E Getting Started Din
Gp E Getting Started Din
guestbe3d76d
 
A new wavelet feature for fault diagnosis
A new wavelet feature for fault diagnosisA new wavelet feature for fault diagnosis
A new wavelet feature for fault diagnosis
IAEME Publication
 

Similaire à Manuale fantoccio mri 76 903 (20)

Design and implementation of microstrip rotman lens for ISM band applications
Design and implementation of microstrip rotman lens for ISM band applicationsDesign and implementation of microstrip rotman lens for ISM band applications
Design and implementation of microstrip rotman lens for ISM band applications
 
Rte0 mjetotk
Rte0 mjetotk Rte0 mjetotk
Rte0 mjetotk
 
IRJET-Design and Analysis of Manual Rice Planter
IRJET-Design and Analysis of Manual Rice PlanterIRJET-Design and Analysis of Manual Rice Planter
IRJET-Design and Analysis of Manual Rice Planter
 
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...
Comparative Analysis of Natural Frequency of Transverse Vibration of a Cantil...
 
Design Philosophy and Material Choice for a Tuner in an Electromagnetic Rever...
Design Philosophy and Material Choice for a Tuner in an Electromagnetic Rever...Design Philosophy and Material Choice for a Tuner in an Electromagnetic Rever...
Design Philosophy and Material Choice for a Tuner in an Electromagnetic Rever...
 
Evaluation of the Turbine Bladed Disc Design for Resonant Vibration Excitation
Evaluation of the Turbine Bladed Disc Design for Resonant Vibration ExcitationEvaluation of the Turbine Bladed Disc Design for Resonant Vibration Excitation
Evaluation of the Turbine Bladed Disc Design for Resonant Vibration Excitation
 
APE_IM_Article_AppliedOptics_Vol47_No29_Oct10_2008_AsPublished
APE_IM_Article_AppliedOptics_Vol47_No29_Oct10_2008_AsPublishedAPE_IM_Article_AppliedOptics_Vol47_No29_Oct10_2008_AsPublished
APE_IM_Article_AppliedOptics_Vol47_No29_Oct10_2008_AsPublished
 
Instruction Manual ATN ThOR Series Thermal Weapon Sights | Optics Trade
Instruction Manual ATN ThOR Series Thermal Weapon Sights | Optics TradeInstruction Manual ATN ThOR Series Thermal Weapon Sights | Optics Trade
Instruction Manual ATN ThOR Series Thermal Weapon Sights | Optics Trade
 
IRJET- Analysis of Fuel Droplet Parameter using Laser Diagnosis System
IRJET- Analysis of Fuel Droplet Parameter using Laser Diagnosis SystemIRJET- Analysis of Fuel Droplet Parameter using Laser Diagnosis System
IRJET- Analysis of Fuel Droplet Parameter using Laser Diagnosis System
 
Opd scan
Opd scanOpd scan
Opd scan
 
Opd scan
Opd scanOpd scan
Opd scan
 
Opd scan
Opd scanOpd scan
Opd scan
 
RM_Laser_Manual_FINAL
RM_Laser_Manual_FINALRM_Laser_Manual_FINAL
RM_Laser_Manual_FINAL
 
IRJET- Design of two Dimensional Photonic Crystal Based Pressure Sensor us...
IRJET- 	  Design of two Dimensional Photonic Crystal Based Pressure Sensor us...IRJET- 	  Design of two Dimensional Photonic Crystal Based Pressure Sensor us...
IRJET- Design of two Dimensional Photonic Crystal Based Pressure Sensor us...
 
Bt31482484
Bt31482484Bt31482484
Bt31482484
 
LG RAT375B
LG  RAT375BLG  RAT375B
LG RAT375B
 
Gp E Getting Started Din
Gp E Getting Started DinGp E Getting Started Din
Gp E Getting Started Din
 
A new wavelet feature for fault diagnosis
A new wavelet feature for fault diagnosisA new wavelet feature for fault diagnosis
A new wavelet feature for fault diagnosis
 
ManualYushi1678502511.pdf
ManualYushi1678502511.pdfManualYushi1678502511.pdf
ManualYushi1678502511.pdf
 
Lg+1512 s
Lg+1512 sLg+1512 s
Lg+1512 s
 

Plus de Luca Gastaldi

Profiler2_reference guide
Profiler2_reference guideProfiler2_reference guide
Profiler2_reference guide
Luca Gastaldi
 
Profiler2_getting started
Profiler2_getting startedProfiler2_getting started
Profiler2_getting started
Luca Gastaldi
 
Gaussmetro thm7025 user manual
Gaussmetro thm7025   user manualGaussmetro thm7025   user manual
Gaussmetro thm7025 user manual
Luca Gastaldi
 
Gaussmetro thm7025 data sheet
Gaussmetro thm7025   data sheetGaussmetro thm7025   data sheet
Gaussmetro thm7025 data sheet
Luca Gastaldi
 

Plus de Luca Gastaldi (16)

5.Test_uninovara
5.Test_uninovara5.Test_uninovara
5.Test_uninovara
 
4.Sistemi e procedure di sicurezza_uninovara
4.Sistemi e procedure di sicurezza_uninovara4.Sistemi e procedure di sicurezza_uninovara
4.Sistemi e procedure di sicurezza_uninovara
 
3.Sicurezza in RM_uninovara
3.Sicurezza in RM_uninovara3.Sicurezza in RM_uninovara
3.Sicurezza in RM_uninovara
 
1.Principi Fisici RM_uninovara
1.Principi Fisici RM_uninovara1.Principi Fisici RM_uninovara
1.Principi Fisici RM_uninovara
 
2.Effetti Fisici e Biologici RM_uninovara
2.Effetti Fisici e Biologici RM_uninovara2.Effetti Fisici e Biologici RM_uninovara
2.Effetti Fisici e Biologici RM_uninovara
 
Libretto radioterapia
Libretto radioterapiaLibretto radioterapia
Libretto radioterapia
 
Profiler2_reference guide
Profiler2_reference guideProfiler2_reference guide
Profiler2_reference guide
 
Profiler2_getting started
Profiler2_getting startedProfiler2_getting started
Profiler2_getting started
 
Gaussmetro thm7025 user manual
Gaussmetro thm7025   user manualGaussmetro thm7025   user manual
Gaussmetro thm7025 user manual
 
Gaussmetro thm7025 data sheet
Gaussmetro thm7025   data sheetGaussmetro thm7025   data sheet
Gaussmetro thm7025 data sheet
 
Corso mlc_gen03
Corso mlc_gen03Corso mlc_gen03
Corso mlc_gen03
 
Corso tsrm_dic01
Corso tsrm_dic01Corso tsrm_dic01
Corso tsrm_dic01
 
0.introduzione
0.introduzione0.introduzione
0.introduzione
 
2.effetti fisici e_biologici
2.effetti fisici e_biologici2.effetti fisici e_biologici
2.effetti fisici e_biologici
 
1.principi fisici
1.principi fisici1.principi fisici
1.principi fisici
 
3.normativa e sicurezza
3.normativa e sicurezza3.normativa e sicurezza
3.normativa e sicurezza
 

Dernier

Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
chetankumar9855
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
adilkhan87451
 
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
mahaiklolahd
 

Dernier (20)

Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
 
Call Girls Kurnool Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kurnool Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kurnool Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kurnool Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
 
Top Rated Bangalore Call Girls Majestic ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Majestic ⟟  9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Majestic ⟟  9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Majestic ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
 
Call Girls Kakinada Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kakinada Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kakinada Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kakinada Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
 
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
 
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Shimla Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service Available
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
 

Manuale fantoccio mri 76 903

  • 1. Nuclear Associates 76-903 MRI Multi-Purpose Phantom Users Manual March 2005 Manual No. 76-903-1 Rev. 2 ©2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies
  • 2. Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms
  • 3. Table of Contents Section 1: Introduction................................................................................................ 1-1 1.1 Introduction .................................................................................................. 1-1 1.2 Orientation – Imaging Method...................................................................... 1-1 1.3 Choosing Echo Time.................................................................................... 1-1 1.4 NMR Solution............................................................................................... 1-2 Section 2: Specifications ............................................................................................ 2-1 2.1 Phantom Design .......................................................................................... 2-1 Section 3: Theory of Operation................................................................................... 3-1 3.1 Omni Directional Application........................................................................ 3-1 3.2 Concentric Conical Rings Section................................................................ 3-1 3.2.1 Slice Thickness Uniformity ........................................................................ 3-1 3.2.2 Aspect Ratio.............................................................................................. 3-1 3.2.3 Magnetic Field Homogeneity..................................................................... 3-2 3.3 Flood Section............................................................................................... 3-2 3.3.1 RF Uniformity ............................................................................................ 3-3 3.3.2 T1 and T2 Values ....................................................................................... 3-3 3.4 Insert Section............................................................................................... 3-3 3.4.1 Folded Step Ramps................................................................................ 3-4 3.4.2 Fold Step Ramp Analysis ....................................................................... 3-4 3.4.3 Measurement Method............................................................................. 3-4 3.4.4 Slice Contiguity (Gap)............................................................................. 3-5 3.4.5 Resolution Pattern .................................................................................. 3-5 3.4.6 Star Pattern ............................................................................................ 3-6 i
  • 5. Introduction Introduction 1 Section 1 Introduction 1.1 Introduction This phantom is designed for comprehensive evaluation of critical imaging parameters of magnetic resonance imaging (MRI) in a time efficient manner. Particular attention was paid to obtaining all the necessary image quality information with a minimum number of imaging sequences. It is designed to study parameters unique to multi-slice two-dimensional and three-dimensional Fourier imaging. The overall dimensions of the phantom adjust so that it can be positioned in each of the three major imaging planes: axial, coronal and sagittal. The phantom can be used for the measurement of absolute values for calibration purposes. However, its design is optimized for time efficient daily quality assurance. The performance parameters that can be evaluated using the phantom include: slice orientation, slice thickness, interslice distance, aspect ratio, magnetic field uniformity, radio-frequency signal uniformity, spatial resolution, T1 and T2 values, and modulation transfer function. The phantom can also be used to evaluate such parameters as resonant frequency, radio frequency power setting, radio frequency quadrature setting, magnetic field gradient calibration, slice off-set parameters, image off-set, and other image artifacts peculiar to MRI. 1.2 Orientation Imaging Method In a two-dimensional Fourier transform (2DFT) imaging of MRI, the data is collected in two orthogonal directions, i.e., encoding and readout. In some imaging sequences, the spatial resolution may be different in the two directions. The star pattern is designed to study such asymmetric resolution as well as qualitative modulation transfer function (MTF) in MR images. For this reason, it is important to properly orient the star pattern. Note four lines or dots on the wall of the phantom. Place the phantom in the imaging system SO that one is in either the phase encoding or frequency readout direction. 1.3 Choosing Echo Time Any clinical multi-slice imaging sequence can be used. Often, however, the imaging sequences with the shortest echo time (TE), largest number of slices possible within the pulse repetition time (TR), and single excitation (no averaging) will demonstrate the performance of an imaging system under one of the most demanding situations. For a daily QA program, it is important to record all the important imaging parameters. An example of such a table of imaging parameters can be found in Table 1-1. 1-1
  • 6. Nuclear Associates 76-903 Operators Manual 1.4 NMR Solution Any NMR signal-producing solution can be used to fill the phantom. It is desirable to use solutions with T1 and T2 values around 500 - 1000 sec, because it can better simulate the clinical imaging conditions. One should note that T1 and T2 values depend on the magnetic field strength. Filling the phantom requires special care to avoid any trapped air bubbles. Water can be degassed by first boiling and cooling prior to filling. Use of a wetting agent and algicide can also prolong the useful life of the solution. A fairly full phantom can be placed in a vacuum chamber to eliminate the bubbles. One way to avoid the trapped bubbles, especially in small holes, is to fill the holes with solution using a small syringe needle. This may require disassembling the phantom. Any solution can be used; however, for higher field systems, solutions with chemical shift should be avoided. Copper sulfite or manganese chloride were found to be useful. A water solution of 0.1 millimolar manganese chloride can be used to fill the phantom. The T1 value of this solution is approximately 500 msec. A shorter relaxation time is not suggested, as it may not correctly represent the system performance in a realistic clinical environment. Table 1-1. Summary of Imaging Parameters Orientation: Imaging Technique: TE= Shortest possible TR= Number of Slices = Slice Thickness = Slice Sequence = Interleaved or others Slice to Slice Gap = Number of Encoding = 256, 128 or 64 Signal Averaging = No averaging, single RF Tuning Parameters (i.e. attenuation settings): NOTE For daily QA, exactly same parameters using same pulse sequence should be used. 1-2
  • 7. Specifications Phantom Design 2 Section 2 Specifications 2.1 Phantom Design All of the parts in the phantom are made with plastic and glass, free of magnetic material. The basic design of the phantom is a 9” diameter plexiglass cylindrical container with various inserts filled with an NMR signal producing solution. The phantom is designed to fit inside a head coil. The cylinder has three sections, flood, cone, and insert section (Figure 2-1). The flood section is a typical flood phantom that can be filled with a solution. Figure 2-1. The conic section has four concentric cones (Figure 2-2). The concentric conical grooves of 45° are cut from a solid plexiglass block. Figure 2-2. 2-1
  • 8. Nuclear Associates 76-903 Operators Manual The insert section has six inserts, one 120° star pattern wedge assembly, two folding step wedges, spatial resolution insert, and T1 and T2 sample vials (Figure 2-3). Figure 2-3. NOTE Addendum for Model 76-903: MRI Multi-Purpose Phantom: The negative contrast resolution insert (rods), as seen in Figure 2-3 has been discontinued and replaced by a T1 and T2 sample vial insert, as described in Figure 3-13. 2-2
  • 9. Theory of Operation Omni Directional Application 3 Section 3 Theory Of Operation 3.1 Omni Directional Application Since MR images are obtained not only in axial but also sagittal and coronal orientation, the phantom is designed to be placed in all three directions. The selection of inserts, their sizes and arrangements are made to obtain the maximum amount of imaging information in any orientation in a single scan. 3.2 Concentric Conical Rings Section The concentric conical rings are machined by cutting out a circular groove at an angle of 45°, as shown (Figure 3-1). Figure 3-1. Outside diameter, 9, Number of segments, 4 The conical section can provide the following information by simple visual inspection: • Slice thickness uniformity • Aspect ratio • Magnetic field homogeneity 3.2.1 Slice Thickness Uniformity The dark and light concentric rings in the image represent the slice thickness multiplied by the square root of 2.0 due to the 45° grooves. The width of the rings (bands) should be the same throughout the image for uniform slice thickness. More careful analysis can be carried out by measuring the brightness profiles with region of interest computer routines. 3.2.2 Aspect Ratio Visual inspection followed by a measurement of the diameter of the outermost ring can provide information on the aspect ratio. The video display should be checked independently to avoid any distortions in the display screen. 3-1
  • 10. Nuclear Associates 76-903 Operators Manual Figure 3-2 shows the cross section of the concentric conical part of the phantom. The aspect ratio is correct, as is the uniformity of the slice thickness as represented by the uniform widths of dark and light circular bands. Figure 3-2. Cross Section Image 3.2.3 Magnetic Field Homogeneity Any distortions in the roundness of the rings can be a result of non-uniformity in the main magnetic field, as shown (Figure 3-3). Figure 3-3. 3.3 Flood Section This section is a typical flood phantom, 9" in diameter by 1.25" thick. It may be filled with any signal producing solution (Figure 3-4). • RF Signal Uniformity • Single T1 and T2 Values 3-2
  • 11. Theory of Operation Flood Section 3 Figure 3-4. Diameter, 9", Depth, 1.25" 3.3.1 RF Uniformity The flood section is used to measure RF uniformity and to visualize artifacts that can come from many different sources (Figure 3-5). Figure 3-5. Radio-frequency (brightness) non-uniformity can come from various sources. Most RF coils have a number of active wires and the regions near the wires can show increased signal intensity. It can also come from magnetic field uniformity, eddy current problems and poor RF power setting. RF uniformity can also be affected by internal motion of the solution. One should also look for any aliasing, fold-over or ghosts. 3.3.2T1 and T2 Values A single value of T1 and T2 can be measured using the flood section. It should be pointed out that T1 and T2 values depend on magnetic field strength, temperature, imaging sequence, and they are very difficult to measure precisely. Extreme caution should be taken for RF power setting. An insert with 6 vials is available SO that various solutions can be imaged for a range of relaxation values. 3.4 Insert Section This section has the following inserts: 1. Folded Step Ramps: Slice thickness and slice contiguity (gap), phantom orientation. 2. Resolution Pattern: Square hole pattern for positive contrast. 3-3
  • 12. Nuclear Associates 76-903 Operators Manual 3. Star Pattern: Modulation transfer function (MTF), horizontal and vertical spatial resolution, quadrature setting and baseline correction. 3.4.1 Folded Step Ramps Accurate slice thickness measurement in MR is very difficult and various methods are under discussion. The slice profile (shape of slice thickness) is also a very complex subject that needs further discussion and definition by the physics community. This folded ramp approach is adopted in this phantom because of its ease of use and efficiency in time. For a quick evaluation, a simple counting of the number of visible steps will give an estimate of slice thickness in millimeters. Slice parameters are studied using a pair of folding step wedges. The height of the steps range from 16 mm to 40 mm and the heights vary every 1.0 mm (Figure 3-6). Each step area has a dimension of 10 mm x 10 mm. The wedges were folded to accommodate a greater number of steps in the limited space available. Two inserts of the folded steps are placed at opposite sides of the phantom so that slice orientation with respect to the phantom can be studied. 1. Slice Thickness 2. Slice Orientation 3. Interslice Gap Figure 3-6. Step Interval, 1 mm Range, 36 mm 3.4.2 Fold Step Ramp Analysis Quick Visual Evaluation Slice thickness can be evaluated quickly by counting the number of steps that can be visualized at a relatively wide window setting (Figure 3-7). The number of visible steps is equal to the slice thickness in millimeters since the height of each step varies by 1 millimeter. Record the window setting for future reference, for the visible steps vary depending on the window setting. In Figure 3-7, five gray scale steps can be seen in the 10 o’clock and 5 o'clock positions, representing a slice thickness of 6 mm. 3.4.3 Measurement Method Better evaluation of slice thickness can be made by using the region of interest (ROI) routine in the imaging system. The area of the region slightly smaller than the step size (10 mm x 10 mm) should be used to read out the signal intensity (NMR number). Than take the difference of the "NMR" number of the adjacent steps. Plot a graph of the difference in intensity on the vertical axis and step sequence numbers on the horizontal axis. The A line graph connecting the dots (diff. vs. step number) represents slice profile and the width of the graph at half of its maximum intensity (full width half maximum) is the measured slice thickness. 3-4
  • 13. Nuclear Associates 76-903 Operators Manual Figure 3-7. 3.4.4 Slice Continuity (Gap) The slice contiguity can be evaluated by quick visual inspection as well as in a more precise manner. While visualizing the steps of one slice in the image, note the locations of the steps…especially the faint steps. Then, repeat the same procedure viewing the next slice using a preset window. The number of missing steps between the two adjacent images represents the gap in millimeters. It can be observed in Figure 3-7 and Figure 3-8 that two steps are missing between the two slices, representing a 2 mm gap. A more precise measurement can be made by plotting a graph of more than one slice on the same axis. Figure 3-8 3.4.5 Resolution Pattern Hole Pattern Solution Insert The resolution insert (Figure 3-9) has 4 sets of square holes (0.5 mm, 0.75 mm, 1 mm and 2 mm machined in an acrylic block. The spacing between the holes is the same as the size of the hole. In filling the phantom with solution, care must be taken to avoid air bubbles becoming trapped in the holes. This can be accomplished by squirting the solution into the holes using syringes with small needles. 3-6
  • 14. Nuclear Associates 76-903 Operators Manual Figure 3-9. 0.5 mm x 0.5 mm2, 0.75 mm x 0.75 mm2, 1.0 mm x 1.0mm2, and 2.0 mm x 2.0mm2 square holed. Five of each dimension. Hole depth, 0.5”. T1 and T2 Solution Insert This insert (Figure 3-10) holds six (6) 5 cc glass vials (15 mm diameter x 47 mm high) that can hold solutions of varying T1 and T2 values. One should note that measured T1 and T2 values depend on many parameters including field strength and imaging sequence used. Figure 3-10. T1, T2 Solution Insert. Six 5 cc refillable glass vials with caps. (15 mm diameter x 47 mm height.) 3.4.6 Star Pattern Modulation Transfer Function (MTF) Modulation transfer function (MTF) is a mathematical concept that provides an objective measurement of the combined effect of image sharpness and resolution. Direct measurement of MTF is often very difficult and calculation of MTF based on measurable parameters such as point-spread function (PSF) or line spread function (LSF) are clearly outside the scope of this manual. Difficult as it may be to have precise mathematical definition, it is however relatively easy to present MTF in a qualitative manner. Similar techniques are used to evaluate x-ray focal spot sizes using a star pattern. The star pattern developed for this purpose is particularly useful in MR imaging where resolution in vertical and horizontal axes can be different and certain image artifacts that are peculiar to Fourier imaging may not be visible under any other test. • MTF Evaluation • Horizontal & Vertical Spatial Resolution • Quadrature Setting • Baseline Correction Figure 3-11. Wedge angle, 3°. Number of wedges, 9. Fan angle, 60°. Wedge length, 3". Width, 2". 3-6
  • 15. Theory of Operation Insert Section 3 The star pattern for MTF evaluation is made up of 3° wedges placed on a 60° fan angle. Each wedge is 3” in the radial dimension and 2” in height. Two such sections are placed next to each other to cover an overall fan angle of 120° (Figure 3-11). A star pattern is a very useful way to measure the many subtle image artifacts that are unique to NMR imaging. It may be difficult to pinpoint the exact source of the artifacts, but this is a powerful test to demonstrate image artifacts that are otherwise difficult to detect. Horizontal and Vertical Spatial Resolution One can also observe asymmetric image resolution between the encoding and readout direction by observing the resolving power in the star pattern (Figure 3-12). It can be seen that the vertical resolution is superior to the horizontal resolution by observing that the narrow ends of the star pattern are better resolved in the vertical direction. Figure 3-12. Quadrature Setting In Figure 3-13, one can visualize a faint star-like pattern in the central 12 o'clock to 5 o’clock position. The star-like ghost pattern is in the opposite orientation of the real star pattern image. This is attributable to the quadrature setting. In quadrature signal detection, which is widely used in MR imaging, the RF signal is split into two orthogonal signals (real and imaginary) before digitization. If the phase angle between the real and imaginary components is not 90°, a ghost of point symmetry appears along the diagonal axis between the encoding and readout directions. This artifact also shows up as a "fold-over" in the star pattern. Figure 3-13. 3-7