OCCLUSAL
CONSIDERATIONS
FOR IMPLANT
SUPPORTED
PROSTHESISwww.indiandentalacademy.com
INTRODUCTION
www.indiandentalacademy.com
The clinical success and longevity of endosteal dental
implants as load-bearing abutments are controlled largely by the
mechanical setting in which they function. The treatment plan is
responsible for the design, number and position of the implants.
After achievement of rigid fixation, proper crestal bone contour,
gingival health, mechanical stress, and/or strain beyond the physical
limits of hard tissues have been suggested as the primary cause of
initial bone loss around implants. After successful surgical and
prosthetic rehabilitation with a passive prosthesis, such noxious
stresses and loads applied to the implant and surrounding tissues
result primarily from occlusal contacts. Complications (prosthetic
and/or bony support) reported in follow-up studies underline
occlusion as a determining factor for success or failure.
www.indiandentalacademy.com
The choice of an occlusal scheme for implant-supported
prostheses is broad and often controversial. Almost all concepts are
based on those developed with natural teeth and are transposed to
implant support systems with almost no modification. No controlled
clinical studies have been published comparing the various implant
occlusal theories.
RISK FACTORS - Implant prostheses with extended
cantilevers have been successful, however, biomechanical factors
clearly demonstrate an increased risk.
Biomechanical parameters are excellent indicators of the
increased risk because they are objective and can be measured. One
can determine which condition presents greater risk, and by how
much the risk is increased. Hence the occlusal concepts developed in
this seminar stem from biomechanical risk factors.
www.indiandentalacademy.com
The prosthodontist has specific responsibilities to minimize
overload to the bone-to-implant interface. These include a proper
diagnosis leading to a treatment plan providing adequate support,
based on the patient’s individual force factors, a passive prosthesis of
adequate retention and form and progressive loading to improve the
amount and density of the adjacent bone and further reduce the risk
of stress beyond physiologic limits. The final element is the
development of an occlusal scheme that minimizes risk factors and
allows the restoration to function in harmony with the rest of the
stomatognathic system.
www.indiandentalacademy.com
TERMINOLOGY
(GPT 1999)
www.indiandentalacademy.com
Anterior protected articulation
A form of mutually protected articulation in which the vertical
and horizontal overlap of the anterior teeth disengage the posterior
teeth in all mandibular excursive movements.
Balanced articulation
The bilateral, simultaneous, anterior, and posterior occlusal
contact of teeth in centric and eccentric positions
Canine protected articulation
A form of mutually protected articulation in which the vertical
and horizontal overlap of the canine teeth disengage the posterior teeth
in the excursive movements of the mandible.
Implant Prosthodontics
The phase of Prosthodontics concerning the replacement of
missing teeth and / or associated structures by restorations that are
attached to dental implants.
www.indiandentalacademy.com
Lingualized occlusion
First described by S. Howard Payne, DDS, in 1941, this form
of denture occlusion articulates the maxillary lingual cusps with the
mandibular occlusal surfaces in centric working and nonworking
mandibular positions. The term is attributed to Earl Pound.
Payne SH. A posterior set up to meet individual requirements.
Dent. 1941; 47:20-2.
Pound E. Utilizing speech to simplify a personalized denture
service. J. Prosthet Dent. 1970;24:585-600.
Linguocclusion
An occlusion in which a tooth or group of teeth is located
lingual to its normal position.
www.indiandentalacademy.com
Mutually protected articulation
An occlusal scheme in which the posterior teeth prevent
excessive contact of the anterior teeth in maximum intercuspation,
and the anterior teeth disengage the posterior teeth in all mandibular
excursive movements.
Occlusion (1645)
1: The act or process of closure or of bring closed or shut off
2 : the static relationship between the incising or masticating surfaces
of the maxillary or mandibular teeth or tooth analogues
Spherical form of occlusion
An arrangement of teeth that places their occlusal surfaces on
the surfaces of an imaginary sphere (usually 8 inches in diameter)
with its center above the level of the teeth (GPT-4).
www.indiandentalacademy.com
IMPLANT PROTECTIVE OCCLUSION (IPO)
www.indiandentalacademy.com
A proper occlusal scheme is a primary requisite for long-term
survival, especially when parafunction or marginal foundations are
present. A poor occlusal scheme both increases the magnitude of
loads and intensifies mechanical stresses (and strain) at the crest of
the bone. Implant Protective Occlusion (IPO) was previously known
as medial positioned-lingualized occlusion. This occlusal concept
refers to an occlusal plane that is often unique and specifically
designed for the restoration of endosteal implant, providing an
environment for improved clinical longevity of both implant and
prosthesis.
www.indiandentalacademy.com
OCCLUSAL CONSIDERATIONS
www.indiandentalacademy.com
Natural Tooth vs. Implant Mobility:
In comparison to an implant, the support system of a natural tooth is
designed to reduce the forces distributed at the crestal bone. The fibrous
tissue interface (periodontal ligament) surrounding natural teeth acts as a
viscoelastic “shock absorber”, serving to both decrease the magnitude of
stress to the bone at the crest, as well as extend the time in which the load is
dissipated.
The presence of a periodontal membrane around natural teeth
significantly reduces the amount of stress transmitted to the bone,
especially at the crestal region. Compared
with a tooth the direct bone interface with an
implant is not as resilient, so the energy
imparted by an occlusal force is not partially
dissipated (the displacement of the
periodontal membrane dissipates energy), but
rather transmits a higher intensity to the
contiguous bone. An analogy of this is hitting
a nail with a steel hammer compared with a
rubber hammer.
www.indiandentalacademy.com
The mobility of a natural tooth can increase with occlusal
trauma. This movement dissipates stresses and strains otherwise
imposed on the adjacent bone interface or the prosthetic components.
After the occlusal trauma is eliminated, the tooth can return to its
original condition with respect to the magnitude of movement. Mobility
of an implant can also develop under occlusal trauma. However, after
the offending element is eliminated, an implant rarely returns to its
original rigid condition. Instead, its health is compromised, and failure
is usually eminent.
The width of almost every natural tooth is greater than the width
of the implant used to replace the tooth. The greater the width of a
transosteal structure (tooth or implant), the lesser magnitude of stress
transmitted to the surrounding bone. The cross-section shape of the
natural tooth at the crest is biomechanically optimized to resist lateral
(buccolingual) loads because of the tooth’s bending fracture resistance
(moment of inertia) and the direction of occlusal forces. Implants are
almost all round in cross-section, which is less effective in resisting
lateral bending loads and consequent stress concentration in the crestal
region in the jaws.
www.indiandentalacademy.com
The elastic modulus of a tooth is closer to bone than any of the
currently available dental implant biomaterials. The greater the
flexibility difference between two materials (metal and bone or tooth
and bone), the greater the potential relative motion generated between
the two surfaces at the transosteal region. Hence under similar
mechanical loading conditions, implants generate greater stresses and
strains at the crest of bone compared with a tooth.
The precursor signs of occlusal trauma on natural teeth are
usually reversible and include hyperemia and occlusal or cold
sensitivity. Condition often results with the patient seeking
professional treatment to reduce the sensitivity, usually by occlusal
adjustment and a reduction in force magnitude. If the patient does not
have an occlusal adjustment, the tooth often further increases in
mobility to dissipate the occlusal forces. If the patient still fails to seek
professional treatment for the increased mobility, the tooth may
orthodontically migrate away from the cause of the occlusal stress.
Even excess tongue or oral habits can cause tooth migration away
from the causative element.www.indiandentalacademy.com
The initial reversible signs and symptoms of trauma on
natural teeth do not occur with endosteal implants. The magnitude of
stress may cause bone microfractures, place the surrounding bone in
the pathologic loading zone causing bone loss, and lead to the
mechanical failure of prosthetic or implant components. Unlike the
reversible signs and symptoms exhibited by natural teeth, implant
bone loss or unsecured restorations most often occur without any
warning signs. Implant occlusal sensitivity is uncommon and
signifies more advanced complications. The loss of crestal bone
around the implant is not reversible without surgical intervention and
results in a decreased implant support and increased sulcus depth
around the abutment. As a result, unless the density of bone
increases or the amount or duration of force decreases, the condition
is likely to progress and even accelerate until implant loss. In
addition, implants cannot move orthodontically away from the
offending force.
www.indiandentalacademy.com
The tooth can show clinical signs of increased stress such as
enamel wear facets, stress lines, lines of Luder (in amalgam fillings),
cervical abfraction, and pits on the cusps of teeth. An implant crown
rarely shows clinical signs other than fatigue fracture. As a result,
fewer diagnostic signs are present to warn the prosthodontist to reduce
the stress on the support system.
www.indiandentalacademy.com
When teeth oppose each other, an interference perception is
approximately 20m. An implant opposing a natural tooth has an
interference perception of 48m, therefore more than twice as poor.
An implant opposing implant has an interference perception of 64m,
and when a tooth opposes an implant overdenture the awareness is
108m (5 times poorer than teeth opposing each other). As a result,
premature occlusal contacts on teeth are usually associated with a
modification of arc of closure and with a decreased force, before
centric occlusion or full interdigitation. In addition, the mandible may
close in a different position to avoid the premature contact and result
in centric occlusion different from centric relation occlusion.
Unfortunately, because of the decreased occlusal awareness of
implants, the premature contact does not trigger such as an adaptation.
In addition, premature contacts are often on smaller areas of load and
therefore result in greater stress (S=F/A). They are most often on
inclines of posterior teeth, which also generates an angled load of
greater stress to the implant bone interface.www.indiandentalacademy.com
www.indiandentalacademy.com
Implants and teeth also have different proprioceptive
information relayed by both entities. Teeth deliver a rapid, sharp pain
sensation under high pressure that triggers a protective mechanism. On
the other hand, implants deliver a slow, dull pain that triggers a
delayed reaction, if any.
Clinical evidence of occlusal trauma on teeth includes an
overall increase in the periodontal membrane thickness and an
increased radiopacity and thickness of the cribriform plate around the
tooth, observed on radiographs and not just localized at the crest.
There are no generalized radiographic signs around an implant under
excess occlusal force, except at the crestal region, which demonstrates
bone loss but may be misdiagnosed as periimplant disease from
bacteria.
www.indiandentalacademy.com
The tooth slowly erupts into occlusion and is present in the mouth
from childhood. The surrounding bone has developed in response to the
biomechanical loads. The permanent teeth are gradually introduced, while
others are present. Hence periodontal tissues have had time to organize in
order to sustain increasing loads, including those brought to bear by an
attached prosthesis. The only progressive bone loading around an implant
is performed by the prosthodontist, and in a much more rapid and intense
fashion.
A lateral force on a natural tooth is rapidly dissipated away from
the crest of bone toward the apex of the tooth. The healthy, natural tooth
moves almost immediately 56 to 108m and pivots two-thirds turn toward
the tapered apex with a lateral load. This action minimizes crestal loads to
the bone. An implant does not exhibit a primary immediate movement, but
a secondary movement 10 to 50m under similar lateral loads. In addition,
it does not pivot (as a tooth) toward the apex but instead concentrates
greater forces at he crest of surrounding bone. Therefore if an initial
angled load of equal magnitude and direction is placed on both an implant
and a natural tooth, the implant sustains a higher proportion of the load
that is not dissipated to the surrounding structures.
www.indiandentalacademy.com
The natural tooth, with its modulus of elasticity similar to
bone, periodontal ligament, and unique cross-sections and dimensions
constitutes a near perfect optimization system to handle stress. In fact,
the stress is handled so well, bacteria-related disease is the weak link.
An implant handles stress so poorly (capturing the stress at the crest
of the ridge), has an elastic modulus 5 to 10 times that of bone, and is
unable to increase mobility without failure that stress is the weakest
link in the system. As a result, ways to decrease stress are a constant
concern to minimize the risk of implants complications.
www.indiandentalacademy.com
www.indiandentalacademy.com
Occlusion on Natural Teeth and Implants:
There has been an ongoing controversy regarding whether a
rigidly fixated implant may remain successful when splinted to natural
teeth. Because the implant has no periodontal membrane, concerns
center around the potential for the “nonmobile” implant to bear the
total load of the prosthesis when joined to the “mobile” natural tooth.
The actual mobility of potential natural abutments may influence the
treatment more than any other factor. In the implant tooth fixed
prosthesis, four important components may contribute movement to
the system, the implant, bone, tooth, and prosthesis.
The sudden, initial tooth movement ranges from 8 to 28m in a
vertical direction under a 3 to 5 lb load, depending on the size,
number, and geometry of the roots and the time elapsed since the last
load application. Once the initial tooth movement occurs, the
secondary tooth movement reflects the property of the surrounding
bone and is very similar to the bone implant movement. The axial
movement of an implant has no initial, sudden movement and ranges
from 3 to 5 m with little correlation to the implant body length.www.indiandentalacademy.com
www.indiandentalacademy.com
When teeth oppose each other, the combined intrusive
movements of the contacting elements may be 56m (28m + 28m).
When a tooth opposes an implant, the combined intrusive movement is
33m (28m + 5m). When implant prostheses oppose each other, the
biomechanical mismatch between teeth in the rest of the mouth and
implants increase. The total combined movement may be 10m,
compared with 56m in the rest of the mouth, and contrary to the teeth
that move immediately, even with light loads, the implants only move
this amount under a heavy occlusal load. A lighter load may generate a
total implant movement of less than 3m.
www.indiandentalacademy.com
For difference in vertical movement of teeth and implants in
the same arch, the existing occlusion is evaluated before implant
reconstruction. Occlusal prematurities are ideally eliminated on teeth
before implant reconstruction. Thin articulating paper (less than 25m
thickness) is then used for the initial implant occlusal adjustment in
centric relation occlusion under a light tapping force. The implant
prosthesis should barely contact, and the adjacent teeth should exhibit
greater initial contacts. Only axial occlusal contacts should be present
on the implant crown. Once the equilibration with a light bite force is
completed, a heavier centric relation occlusal force is applied. The
contacts should remain axial over the implant body and may be of
similar intensity on the implant crown and the adjacent teeth under
greater bite force to allow all elements to react similar to the occlusal
load. Hence to harmonize the occlusal forces between implants and
teeth, a heavy bite force occlusal adjustment is used because it
depresses the natural teeth, positioning them closer to the depressed
implant position and equally sharing the load.www.indiandentalacademy.com
www.indiandentalacademy.com
If healthy anterior teeth and/or natural canines are present, the
occlusion allows those teeth to distribute horizontal loads in
excursions, while the posterior teeth disocclude during excursions.
Anterior, compared with posterior bite force measurements and
electromyographic studies provide evidence that the stomatognathic
system elicits significantly less force when the posterior segments are
not in contact. As a result, all lateral excursions of IPO opposing
fixed prostheses or natural teeth should disocclude the posterior
components. The resultant lateral forces are thus distributed only to
the anterior segments of the jaws, resulting in a decrease in overall
occlusal force magnitude because of diminished muscle firing and
recruitment.
This occlusal scheme should be followed whether or not
anterior implants are in the arch. However, if anterior implants must
disocclude the posterior teeth in excursion, two or more implants
splinted together should help dissipate the lateral forces.
www.indiandentalacademy.com
When anterior implants and teeth are not connected - The
initial lateral movement of healthy anterior teeth ranges from 68 to
108m before secondary tooth movement. Anterior implant
movements are not immediate and range from 10 to 50m. Because of
the greater discrepancies in lateral movement, the occlusal adjustment
in this direction is more critical to implant success and survival. Light
force and thin articulating paper(20m) are first used to ensure that no
implant crown contact occurs during the initial occlusal or lateral
movement of the teeth. A heavier force during centric occlusion and
excursions is then used to develop similar occlusal contacts on both
anterior implants and natural teeth.
www.indiandentalacademy.com
Unlike teeth, implants do not extrude, rotate, or migrate under
occlusal forces. Natural teeth exhibit mesial drift and slight changes
in occlusal position do occur over time. The proposed occlusal
adjustment does not encourage additional tooth movement because
regular occlusal contacts occur. The teeth opposing implants are not
taken out of occlusion. Brief occlusal contacts on a daily basis
maintain the tooth in its original position (similar to the rest of the
mouth). In addition, because most teeth occlude with two teeth, the
opposing teeth positions are even more likely to remain the same.
No occlusal scheme will prevent mesial drift and minor tooth
movement from occurring. An integral part of the IPO philosophy is
the regular evaluation and control of occlusal contacts at each
regularly scheduled hygiene appointment. This permits the correction
of minor variations occurring during long-term function and also
helps prevent porcelain fracture and other stress-related
complications on the remainder of the natural teeth.
www.indiandentalacademy.com
For implants joined to natural teeth a similar scenario is used
for the occlusal equilibration. A light force and thin articulating paper
are used, and the implant crown exhibits minimum contact compared
with the natural abutment crown. A gradient of force is designed on
the pontics. A heavy bite force is then used to establish equal occlusal
contacts for all abutments and the entire prosthesis, whether implant
or natural.
www.indiandentalacademy.com
Implant Orientation and Influence of Load Direction:
Forces acting on dental implants are referred to as vectors (defined
in both magnitude and direction). Occlusal forces are typically three-
dimensional, with components directed along one or more of the clinical
coordinate axes.
www.indiandentalacademy.com
Implants are designed for a long axis load to the implant
body. Stress contours were primarily concentrated at the transosteal
(crestal) region. An axial load over the long axis of an implant body
generates a greater proportion of compressive stress than tension or
shear forces.
Any load that is applied at an angle may be separated into
normal (compressive and tensile) and shear forces. The greater the
angle of loads to the implant long axis, the greater the compressive,
tensile and shear stresses. When FEA evaluates the direction of the
force changed to a more angled or horizontal load, the magnitude of
the stress is increased by 3 times or more. In addition, rather than a
compressive type of force primarily, greater tensile and shear forces
are also demonstrated and increase more than 10 times compared
with the amount found with an axial force. These stress contours
resemble the pattern of early crestal bone loss on implants.
www.indiandentalacademy.com
www.indiandentalacademy.com
BONE MECHANICS AND OCCLUSION
www.indiandentalacademy.com
The effect of offset or angled loads to bone is further
exacerbated because of the anisotropy of cortical bone. Anisotropy
refers to the character of bone, whereby its mechanical properties,
including ultimate strength, depend on the direction in which the bone
is loaded. Cortical bone of human long bones has been reported as
strongest in compression, 30% weaker in tension, and 65% weaker in
shear. Therefore IPO attempts to eliminate or reduce all shear loads to
the implant to bone interface.
www.indiandentalacademy.com
A force applied at a 30-degree angle decreased the bone
strength limits by 10% under compression and 25% with tension. A
60-degree force reduced the strength 30% under compression and
55% under tension. Therefore not only does the crestal bone load
increase around the implant with angled forces, but the amount of
stress the bone may withstand is also decreased. The greater the angle
of load, the lower the ultimate strength.
www.indiandentalacademy.com
The primary component of the occlusal force should therefore
be directed along the long axis of the implant body, not on an angle or
following an angled abutment post. Angled abutments are used only
to improve the path of insertion of the prosthesis or the final esthetic
result. The angled abutment, which is loaded along the abutment axis,
will transmit a significant moment load to both the implant crestal
region and abutment screw, proportional to its angle of inclination. In
addition, the angled implant often requires an angled abutment.
Angled abutments are fabricated in two pieces and are weaker in
design than a one-piece post. Furthermore, a larger transverse load
component develops at the crest as a result of angled loads. An angled
load to the implant long axis increases the compressive forces at the
crest of the ridge on the opposite side of the implant in which the
force is directed, increasing the tension component of force along the
same side. The greater the angle of force to the long axis of the
implant body, the greater the potentially damaging load at the crest of
the bone. www.indiandentalacademy.com
Hence the angled load increases the amount of crestal stresses
around the implant body, transforms a greater percentage of the force to
tensile and shear force, and reduces bone strength in compression and
tension. In contrast, the surrounding implant body stress magnitude is
least and the strength of bone is greatest under a load axial to the
implant body.
Premature occlusal contacts result in localized lateral loading of
the opposing contacting crowns. Because the surface area of a
premature contact is small, the magnitude of stress in the bone increases
proportionately (i.e., stress=force/area). All the occlusal force is applied
to one region rather than being shared by several abutments and/or
teeth. In addition, the premature contact is most often on an inclined
plane, therefore creating a greater horizontal component to the load and
increasing compressive and tensile crestal stresses. Therefore occlusal
evaluation and adjustment in partially edentulous implant patients are
more important than in the natural dentition because the premature
contacts can result in more damaging consequences on implants
compared with teeth.
www.indiandentalacademy.com
www.indiandentalacademy.com
The elimination of premature contacts is more important than
in natural teeth because the implant is less mobile and often cannot
effectively dissipate the forces. In addition, the teeth benefit from a
greater occlusal awareness (proprioception) or oral tactile function
than implants.
Once the natural teeth are removed, the bone remodels to the
height at or below the lowest level of the lateral cortical plates.
Hence the implant crown height is often greater than the original
natural anatomic crown, even in Division A bone. Crown height,
with a lateral load, is a magnifier of stress to an implant to bone
interface. The greater the crown height, the greater the resulting
crestal moment with any lateral component of force that develops as
a consequence of an angled load. Angled abutments loaded in the
direction of the abutment with an increase in crown height are
subject to even greater crestal moment loads because of both the
lateral load and the increased lever effect from the crown height.
www.indiandentalacademy.com
In the anterior maxilla, labial concavities may require that the
implant be angled away from the labial bone and the abutment toward the
facial crown contour. These implant bodies are more frequently loaded at an
angle, and an angled prosthetic abutment is required. As a result, larger
diameter implants or a greater number of implants are indicated to minimize
the crestal bone stress on each abutment. IPO aims at reducing the force of
occlusal contacts, increasing implant number, and/or increasing implant
diameter for implants subjected to angled loads or with an increased crown
height or on the cantilever portion of a prosthesis.
www.indiandentalacademy.com
OCCLUSAL SCHEMES
A primary goal of an occlusal scheme is to maintain the
occlusal load that has been transferred to the implant body within the
physiologic limits of each patient. These limits are not identical for all
patients or restorations. The forces generated by a patient are
influenced by parafunction, masticatory dynamics, tongue size,
implant arch position and location, and implant arch form and crown
height. The prosthodontist can best address these force factors by
selecting the proper implant size, number, and position, using stress-
relieving elements, increasing bone density by progressive loading
and selecting the appropriate occlusal scheme.
www.indiandentalacademy.com
CLASSIFICATION OF OSSEOINTEGRATED
PROSTHESIS
www.indiandentalacademy.com
Hobo et al
1. Fully bone anchored bridge
2. Overdenture
3. Freestanding bridges
a. Kennedy class I
b. Kennedy class II
c. Kennedy class III
d. Kennedy class IV
4. Bridge connected to the natural teeth.
5. Single tooth replacement.www.indiandentalacademy.com
Misch C.E et al
TYPE DEFINITION
FP-1 Fixed prosthesis, replaces only the crown, looks
like a natural tooth
FP-2 Fixed prosthesis, replaces the crown and a portion
of the root, crown contour appears normal in the
occlusal half is elongated or hypercontoured in the
gingival half
FP-3 Fixed prosthesis, replacing missing crowns and
gingival colour and a portion of the edentulous site,
prosthesis must often use denture teeth and acrylic
gingiva, but may be porcelain to metal
RP-4 Removable prosthesis, overdenture supported
completely by implant
RP-5 Removable prosthesis, overdenture supported by
both soft tissue and implants
www.indiandentalacademy.com
IMPLANT PROTECTIVE OCCLUSION(IPO)
www.indiandentalacademy.com
When teeth are present, the maxillary dentate posterior ridge is
positioned slightly more facial than its mandibular counterpart. Once
the maxillary teeth are lost, the edentulous ridge resorbs in a medial
direction as it evolves from Division A to B, Division B to C, and
Division C to D. As a result, the maxillary permucosal implant site
gradually shifts toward the midline as the ridge resorbs.
As a result of ridge resorption in
width the maxillary posterior implant
permucosal site may even be lingual to the
opposing natural mandibular tooth. The
posterior mandible also resorbs lingually as
the bone resorbs from Division A to B. As a
consequence, endosteal implants are also
more lingual than their natural tooth
predecessors.
www.indiandentalacademy.com
Occlusal Table Width:
A wide occlusal table favors offset contacts during mastication
or parafunction. Narrower implant bodies are even more vulnerable to
occlusal table width and offset loads. Wider root form implants can
accept a broader range of vertical occlusal contacts while still
transmitting lesser forces at the permucosal site under offset loads.
Therefore in IPO the width of the occlusal table is directly related to
the width of the implant body.
During mastication, the amount of force used to penetrate the
food bolus is also related to occlusal table width. For example, less
force is required to cut a piece of meat with a sharp knife (narrow
occlusal table), than with a dull knife (wider occlusal table). The
greater surface area requires greater force to achieve a similar result.
Hence the wider the occlusal table, the greater the force developed by
the biologic system to penetrate the bolus of food.
www.indiandentalacademy.com
The posterior narrow occlusal table also facilitates daily home
care. The laboratory technician often attempts to fabricate occlusal
facial and lingual contours similar to that of natural teeth. This often
results in ridge laps or porcelain extension at the facial gingival margin
of the implant, to create an occlusal table approximately 8 to 10 mm
wide. As a result, home care in the sulcular region of the implant is
impaired by the overcontoured crown design. On the contrary, a narrow
occlusal table combined with a reduced buccal contour (in the posterior
mandible) permits easier sulcular oral hygiene in manner similar to a
tooth and improves axial loading.
www.indiandentalacademy.com
The narrower occlusal contour also reduces the risk of
porcelain fracture. A facial profile similar to a natural tooth on the
smaller diameter implant results in cantilevered restorative materials.
The facial porcelain is most often not supported by a metal
substructure because the gingival region of the crown is also
porcelain. As a result, shear forces result on the buccal cusp on the
mandibular crown or lingual cusps in the maxillary crown, and are
more likely to increase the risk of porcelain fracture.
Restorations mimicking the occlusal anatomy of natural teeth
often result in offset loads (increased stress), complicated home care
and increased risk of porcelain fracture. As a result, in nonesthetic
regions of the mouth, the occlusal table should be reduced in width
compared with natural teeth.
www.indiandentalacademy.com
DIVISION OF AVAILABLE BONE:A,B,C&D
(MISCH)
www.indiandentalacademy.com
www.indiandentalacademy.com
www.indiandentalacademy.com
CROWN CONTOUR
www.indiandentalacademy.com
Division A Bone
The primary component of the occlusal force is evaluated during
the treatment-planning phase. In an edentulous ridge with abundant
height and width and little resorption, the implant may be placed in a
more ideal position for occlusion and esthetics.
Offset loads are used to describe cantilevered buccal or lingual
occlusal contacts, not directed along the long axis of the implant body.
When offset loads are generated at an angle, the distance between the
offset contact and the long axis acts as a moment arm that magnifies
the effect of the lateral force.
www.indiandentalacademy.com
The most common implant placement corresponds to a central
position in the residual ridge. The implant osteotomy begins in the center
of the crest and is gradually increased to the optimal width indicated in
relation to the recipient bone. Facial concavities are avoided, and the
thinner facial cortical bone is protected, to limit surgical complications
such as labial dehiscence. As a consequence, whether in the maxilla or
the mandible, the implant is frequently placed under the central fossa
region of the former natural tooth.
To load the implant body in an
axial direction, the primary occlusal
contact should therefore be the central
fossa region in Division A bone. Thus
for maxillary implant opposing
mandibular natural teeth, the
mandibular buccal cusp acts as the
primary tooth contact.
www.indiandentalacademy.com
Because bone loss occurs at the expense of the facial plate, a
modified buccal contour anatomy may need to be generated in Division
A or B mandibles. The occlusal table width is reduced to favor an axial
load on the implant in nonesthetic regions. The Division A mandibular
implant is placed under the central fossa region of the natural tooth.
When opposing a natural maxillary molar, the primary contacting cusp
becomes the maxillary lingual cusp opposing the mandibular implant
crown, with the mandibular buccal cusp of decreased height and width
over the implant body. Hence all contacts are situated medially
compared with those on natural teeth.
The lingual contour of the mandibular
implant crown is similar to the original natural
dentition in position, complete with horizontal
overlap to the maxillary lingual cusp to
prevent tongue biting during function. There
is no occlusal contact on the lingual cusp, so
offset loads during parafunction are
eliminated. www.indiandentalacademy.com
The posterior maxillary crown is reduced only from the lingual
aspect, compared with a natural maxillary molar, to reduce the occlusal table
width. such a reduction increases the lingual overjet when the teeth are in
occlusion. Narrower opposing mandibular occlusal tables are desirable to
direct occlusal forces over the maxillary implant body. As a result, when
opposing maxillary implants, the buccal cusps of natural mandibular teeth
(or crowns on implants) should be recontoured to minimize offset loads in
centric relation occlusion. The maxillary buccal cusp may then be retained
for esthetics, but the functional occlusal table is reduced.
When esthetics are not a concern the distal one half of the first
molar and / or the entire second molar is often restored in cross bite to
improve the direction of forces. In the posterior esthetic regions of the
maxilla with facial bone resorption and / or lingually placed implants, a
wider occlusal table is required to project the facial contours for ideal
esthetics. Bone grafting to increase bone width may be required in these
esthetics zones, so a larger diameter implant may be placed that permits
restorations of the buccal contours with maintenance of cervical contours
with emergence profiles, which permit proper hygiene of the sulcular
regions. www.indiandentalacademy.com
www.indiandentalacademy.com
Posterior implants opposing each other attempt to axially load
both entities. The facial cusp of the maxillary crown is required for
esthetics. The other contours of the opposing crowns are reduced in
width to minimize the occlusal table width and axially load the
implants.
WHENEVER POSSIBLE THE PORTIONS OF AN IMPLANT
CROWN THAT ARE NOT SUPPORTED BY AN AXIALLY
POSITIONED IMPLANT SHOULD BE RECONTOURED SO
THEY DO NOT RECEIVE OCCLUSAL LOADS.
ALTERNATIVELY, SEVERAL ADDITIONAL IMPLANTS
SHOULD BE USED TO DISSIPATE THE FORCE.
www.indiandentalacademy.com
DIVISION B BONE
Division B bone has maxillary and mandibular implants positioned
under the lingual cusp when compared with the original natural tooth
position. As a result, mandibular crowns require even more reduced buccal
contours to avoid offset occlusal contacts. The primary contact of occlusion
on an opposing natural posterior maxillary tooth is the lingual cusp, which
is reshaped to axially load the implant.
www.indiandentalacademy.com
The buccal cusp of the mandibular implant crown is located
near the original central fossa of the natural tooth. The medially
positioned Division B mandibular implant crown may have a central
fossa, but it is more lingual than the original position. The lingual
contour of the crown is similar to that of the original natural tooth and
has an overjet with the opposing natural tooth to prevent biting the
tongue during function. The mandibular posterior implant may, on
occasion, be angled medially because of the sub mandibular fossa.
As a result, an angled abutment
and a lingual straight emergence crown
profile to minimize the lingual volume of
the restoration are indicated.
Augmentation of the mandibular
Division B ridge is often required when
stress factors are moderate to improve
the implant position and prosthetic
guidelines. www.indiandentalacademy.com
A Division B maxillary implant is often placed under the palatal cusp
region of the original natural tooth. The maxillary occlusal table cannot
always be reduced from the facial aspect for esthetic reasons; therefore the
buccal cusp is offset facially but left completely out of occlusion (as with
natural teeth) in centric relation occlusion and during all mandibular
excursions. The buccal cusp of the opposing natural tooth is recontoured in
width and height to reduce offset loads to the opposing crown on the
maxillary implant. The primary
occlusal contact is centric relation
occlusion is the maxillary palatal
cusp over the implant body and
the central fossa region of the
mandibular natural tooth. Bone
augmentation for larger implant
width is more indicated in the
maxilla because of the less dense
bone and the prosthetic needs to
replace an esthetic buccal crown
contour.
www.indiandentalacademy.com
When Division B
implants are placed in both
arches, the maxillary and
mandibular prostheses are
similar to that described in the
previous scenario. However,
it is usually not possible to
load both arches with an axial
load, so the weakest implant
in bone density, width, or
prosthesis type (fixed vs.
removable) determines the
axial load, because it is the
most vulnerable arch.
www.indiandentalacademy.com
When further resorption occurs and the ridge evolves into
Division C or D, the maxillary palatal cusp becomes the
primary contact area, situated directly over the implant body.
Hence the occlusal contacts differ from those of a natural tooth
and may even be positioned more medial than the natural palatal
cusp when the implant is placed in Division C or D bone.
www.indiandentalacademy.com
Influence of Surface Area:
An important parameter in IPO is the adequate surface area to
sustain the load transmitted to the prosthesis. It is important to
remember that mechanical stress, in its simplest form, can be defined as
the force magnitude divided by the cross sectional area over which that
force is applied.
When implants of decreased surface area are subject to angled
or increased loads, the magnified stress and strain magnitudes in the
interfacial tissues can be minimized by placing an additional implant in
the region of concern.
Thus when narrow diameter implants are used in regions that
receive greater forces, additional splinted implants are even more
indicated to compensate for their narrow design and to help decrease
and distribute the load over a broader region. When forces are
increased in magnitude, direction or duration (e.g., parafunction), ridge
augmentation maybe required to improve implant placement, reduce
crown height, and increase implant width and number to compensate
for the increased loads.
www.indiandentalacademy.com
Maxillary Root surface area
(mm2)
CENTRAL 204
LATERAL 179
CANINE 273
FIRST
PREMOLAR
234
SECOND
PREMOLAR
220
FIRST MOLAR 433
SECOND
MOLAR
431
Mandibular Root surface area
(mm2)
CENTRAL 154
LATERAL 168
CANINE 268
FIRST
PREMOLAR
180
SECOND
PREMOLAR
207
FIRST MOLAR 431
SECOND
MOLAR
426
www.indiandentalacademy.com
Radius (mm) Length (mm) Surface Area (mm2)
1.75 10
12
14
110
132
154
2.00 10
12
14
126
151
176
2.25 10
12
14
141
169
197
2.50 10
12
14
157
188
220
2.75 10
12
14
172
207
242
3.00 10
12
14
188
226
264www.indiandentalacademy.com
www.indiandentalacademy.com
The prosthesis type may also be modified from a fixed
restoration (FP-1 to FP-3) to a removable prosthesis (RP-4). This is
most effective when nocturnal parafunction is present because the
restorations may then be removed while sleeping. In addition, stress
relieving elements may be included in the removable restoration, and
additional support may be gained from the soft tissue (RP-5
restorations).
Wider diameter root form implants have a greater area of bone
contact at the crest than narrow implants (resulting from their
increased circumferential bone contact areas). As a result, for a given
occlusal load, the mechanical stress at the crest is reduced with wider
implants compared with narrow ones.
Natural teeth follow similar principles of diameter and surface
area as just described. The anterior region of the mouth is
characterized by reduced bite force compared with the posterior
region. Consequently, the anterior tooth cross section is smaller, and
the surface area is reduced compared with the greater diameter and
surface area of posterior teeth.
www.indiandentalacademy.com
Design to the Weakest Arch:
Any complex engineering structure will typically fail at its
“Weakest link”, and dental implant structures are no exception.
The amount of force distributed to a system can be reduced by
stress relieving components that may dramatically reduce impact loads
to the implant support. The soft tissue of a traditional completely
removable prosthesis opposing implant prosthesis is displaced more
than 2 mm and is an efficient stress reducer. Lateral loads do not result
with as great a crestal load to the implants because the opposing
prosthesis is not rigid.
The most common implant treatment, which includes a
traditional soft tissue supported complete denture, is a maxillary denture
opposing a mandibular implant supported restoration. The occlusal
scheme for this condition raises the posterior occlusal plane, uses a
medial positioned lingualized occlusion, and has a bilateral balanced
scheme. Whether the mandibular restoration is FP-1, FP-2, FP-3, RP-4,
or RP-5, the maxillary denture follows these guidelines.www.indiandentalacademy.com
The mandibular implant supported restoration may exert
greater force on the premaxilla than a mandibular denture and cause
accelerated bone loss. Therefore modification of the occlusal scheme
aims at protecting the premaxilla under a maxillary denture by the
total elimination of anterior contacts with the mandibular anterior
teeth in centric occlusal relation.
Reduced occlusal forces with an absence of lateral contacts in
excursions are recommended on posterior cantilevers or anterior
offset pontics whenever possible. This minimizes the moment forces
on the abutments.
It is better for mandibular cantilever pontics to oppose
maxillary implants than the reverse situation.
www.indiandentalacademy.com
Full – Arch fixed prostheses
(FP-1 to RP-4)
Fixed prostheses on natural teeth opposing FP-1 to RP-4
implant restorations should follow mutually protected occlusal schemes
whenever possible. In protrusion, there should be total absence of
posterior contacts, especially for cantilevered posterior units. The
masticatory force generated during lateral excursions is decreased in
absence of posterior contacts. This assists in reducing the noxious
effect of lateral forces on the anterior implants. Two or more implants
should share any lateral force, and lateral excursions should occur as
far forward as is practical and include the canine.
Minimal occlusal contact in the cantilevered regions and the
total absence of posterior lateral contacts during excursions are
indicated when opposing the natural dentition or a fixed restorations.
Seven to eight implants to support a complete implant
prosthesis in two separate units are suggested in the mandible for a
fixed restoration opposing a fixed prosthesis or natural teeth with
inadequate to severe stress factors.
www.indiandentalacademy.com
In the edentulous maxilla, flexure of the bone is not a
concern. A full arch prosthesis may be fabricated in one section.
Eight to ten maxillary implants most often are required for a
twelve unit fixed prosthesis opposing a fixed dentition on teeth and /
or implants with moderate to severe stress factors. Posterior implants
are more critical in the maxilla, in order to eliminate cantilevers and
increase the anteroposterior implant distance, which further
decreases stress to the maxillary anterior implants.
www.indiandentalacademy.com
Fully bone anchored bridge
Mandibular edentulous case
for fully bone anchored
bridge
•Recommended to have a mutually
protected occlusion.
•In centric it is necessary to have 30m
clearance at the anterior region
•disocclusion should be employed.
•To avoid localization of stresses
anterior group function should be
employed.
•The anterior guidance should be made
slightly flatter than natural teeth to
avoid over stresses on the fixtures.
www.indiandentalacademy.com
Kennedy Class I
OCCLUSION FOR FREESTANDING
BRIDGES
o Clearance of the anterior teeth
should be smaller than the
natural teeth.
o Amount of disocclusion
required is same as natural teeth
since the anterior guidance is
provided by the remaining
anterior natural teeth.
Protrusive : 1.1mm
Non working side : 1.0 mm
Working Side : 0.5 mm
www.indiandentalacademy.com
Kennedy Class II
•In centric the posterior
osseointegrated bridge should
have 30m open contacts, while
anterior teeth also should have 30
m open contacts and begin to
contact under strong bite pressure.
•Amount of disocclusion required
is same as natural teeth since the
anterior guidance is provided by
the remaining anterior natural
teeth.
Protrusive : 1.1mm
Non working side : 1.0mm
Working Side : 0.5 mm
www.indiandentalacademy.com
Kennedy Class III
 Vertical dimension is maintained
by remaining natural teeth
 The osseointegrated bridge should
contact only under strong pressure.
 Amount of disocclusion required
is same as natural teeth since the
anterior guidance is provided by the
remaining anterior natural teeth.
Protrusive : 1.1mm
Non working side : 1.0 mm
Working Side : 0.5 mm
www.indiandentalacademy.com
Kennedy Class IV
 To minimize horizontal loads
group function occlusion is
recommended.
 During lateral movement
posterior teeth on working side can
bear the horizontal load while non
working side can be discluded.
Anterior guidance should be flatter than natural dentition
to minimize load induced on the fixture during protrusive
movement.
Amount of disclussion suggested is as follows
Protrusive : 0.8 mm
Non- working side : 0.4 mm
Working side : 0.0 mmwww.indiandentalacademy.com
IMPLANT AND SOFT TISSUE SUPPORTED OVERDENTURE
(RP-5)
Anterior Tooth Position
Centric stops or pressure from the tongue and muscle positions
usually prevent continued extrusion of anterior natural teeth. maxillary
anterior prosthetic teeth are positioned forward of the anterior
supporting bone to satisfy phonetic and esthetic requirements. Moment
forces result from contact with the anterior teeth, which may cause
instability of the maxillary prosthesis. Therefore the maxillary denture
usually does not have anterior incisal centric stops. This helps protect
the premaxilla from excess forces in centric occlusion relation and
initial excursions of the mandible, as the premaxilla is vulnerable to
resorption from external stresses.
www.indiandentalacademy.com
Posterior Tooth Position
The maxillary edentulous posterior ridge resorbs in a medial directions
it transforms from Division A to B, Division B to C, and Division C to D.
therefore the maxillary denture tooth gradually becomes more cantilevered off
the bone support, even when positioned in the same spatial location. The
mandibular edentulous posterior ridge also resorbs in a medial redirection as it
transforms from Division A to B, but then resorbs laterally from Division B to
C, and more lateral as it resorbs from Division C to D. In complete dentures,
the position of the mandibular posterior teeth is often determined first. Bone
support concepts of occlusion often position the mandibular teeth
perpendicular to the edentulous ridge. This positions the central fossa of the
posterior mandibular teeth more medial than that of their natural predecessors
in Division B, but more facial in Division C, and very facial in Division D
compared with the natural tooth placement.
The maxillary teeth are then situated farther facially than the original
teeth, if a normal cusp fossa relation is maintained. Consequently, maxillary
denture teeth are always placed lateral to the resorbing bony support, and the
condition is compounded in cases of advanced atrophy (Division C or D
bone). www.indiandentalacademy.com
The basic, concept of lingualized occlusion was first introduced
by Gysi. Later Payne suggested the maxillary buccal cusps of posterior
teeth should be reduced, so only the lingual cusps would be in contact.
Pound discussed a similar concept, but reduced the buccal cusp of the
mandible and introduced the term “lingualized” occlusion. Pound also
placed the lingual cusp of the mandibular posterior teeth between lines
drawn from the canine to each side of the retromolar pad. Consistent in
the Philosophy of Payne and Pound, was the belief that the palatal cusp
should be the only area of maxillary
tooth contact. These occlusal schemes
were designed to narrow the occlusal
table and improve mastication, reduce
forces to the underlying bone, and help
stabilize a lower denture. The
techniques of Payne and Pound may be
modified further to a medial positioned
lingualized occlusion, proposed by
Misch. www.indiandentalacademy.com
Medial Positioned Lingualized occlusion : Laboratory steps
1. Mount the upper cast using a face bow record. Mount the lower
cast using the centric relation record. Set the horizontal condylar
guidance according to the protrusive record.
2. Set the maxillary and mandibular anterior teeth for esthetics,
phonetics, and lip support.
3. Cut back the posterior flange of the lower record base to expose
the retromolar pad. Outline the retromolar pad in pencil. Draw a
line from the lingual border of the pad to the mesial aspect cuspid.
The central fossa of mandibular posterior teeth will be set along
this line.
www.indiandentalacademy.com
4. Using a flat plane or 10 degree mold, set the mandibular posterior
teeth in a compensating curve. The curve should have both a
mediolateral and anteroposterior dimension that progressively
develops as the teeth are set posteriorly. The curve starts with the
first premolar and becomes more accentuated in the molar region
(closer to the condyle) (i.e., first premolar 0 to 5 degrees, second
premolar 5 to 10 degrees, first molar 15 to 20 degrees, second
molar 20 to 25 degrees). The anteroposterior angle of the curve
is the second molar region and should ideally approximate the
horizontal condylar guidance (i.e., 20 to 25 degrees).
5. Drop the incisal pin of the articulator 1 to 2 mm.
6. Using a 30 to 33 degree mold, set the maxillary posterior teeth
with the buccal cusp tilted out facially. The lingual cusp should
contact the central groove of the mandibular teeth (this will be the
only tooth contact point). There should be no contact between the
mandibular buccal cusp and the opposing maxillary tooth.
www.indiandentalacademy.com
7. Return the incisal pin to the original position (up 1 to 2 mm).
8. Using articulating paper to check the contacts, grind a small fossa
in the mandibular tooth for the maxillary lingual cusp tip.
Continue to adjust the occlusion until the incisal pin touches the
incisal table. Check for balanced occlusion in excursions.
9. Festoon the set up for the try in appointment.
www.indiandentalacademy.com
Overdenture
Maxillary edentulous case for
overdenture
• Recommended occlusion for
overdenture is fully balanced
occlusion with lingualized
occlusion.
• Incase maxillary overdenture is
opposed by a mandibular fully
bone anchored bridge, in centric
a small clearance is
recommended in the anterior
teeth, while posterior contact
simultaneously.
• Disocclusion is not employed
here. www.indiandentalacademy.com
OCCLUSAL MATERIALS
www.indiandentalacademy.com
The materials selected for the occlusal surface of the prosthesis
affect the transmission of forces and the maintenance of occlusal
contacts. In addition, occlusal material fracture is one of the most
common complications for restorations on natural teeth or implants.
Therefore it is wise to consider the occlusal material for each
individual restoration. Occlusal materials maybe evaluated by
esthetics, impact force, a static load, chewing efficiency, fracture, wear,
interarch space requirements, and accuracy of castings. The three most
common groups of occlusal material are fixed prostheses on implants
are reviewed with relevance to the previous eight criteria.
www.indiandentalacademy.com
Esthetics:
Esthetics is a major concern for patients. The most esthetic
material available today is porcelain. Acrylic is acceptable for esthetics,
and metal is a poor choice of materials when esthetics is the chief
criterion. However, there are many situations in which esthetics is not
an important aspect of the restorations. For example, when a maxillary
second molar is restored, most patients do not expose this area when
smiling or laughing.
www.indiandentalacademy.com
Forces:
The materials on the occlusal aspect of the prostheses affect the
transmission of force to the bones. Impact loads give rise to brief episodes of
increases force, primarily related to the speed of closure and the dampening
effect of the occlusal material. The hardness of material is related to its ability
to absorb stress from impact loads. All porcelain occlusal surface exhibits a
hardness 2.5 times greater than that of natural teeth. Acrylic resin has a Knoop
hardness of 17 kg/mm2, and enamel has a 350 kg/mm2 hardness. A composite
resin may exhibit a hardness of 85% that of enamel. Therefore impact loads
are lowest with acrylic, increase with composite and metal occlusals, increase
even more with enamel, and further increase with porcelain. As a
consequence, it has been suggested to use resin because of its dampening
characteristics.
Clenching patients do not have a considerable amount of stress
reduction when acrylic versus porcelain materials are used on the occlusal
surfaces.
Progressive bone loading is performed with acrylic transitional
prostheses. This material may reduce the impact force on the early implant
bone interface. As the bone matures and its density increases, the need for
force reduction decreases.
www.indiandentalacademy.com
Chewing efficiency:
Fixed prostheses exhibit an improve efficiency compared with
removable soft tissue borne prostheses, regardless of the occlusal
material.
Acrylic was 30% less efficient than porcelain or metal,
whereas there was no difference between gold and porcelain.
www.indiandentalacademy.com
Wear:
The definition of wear is the deterioration, change or loss of a
surface caused by use. The factors affecting the amount of wear include
magnitude, angle, duration, speed, hardness, and surface finish of the
opposing force and surface, together with the lubricant, temperature,
and chemical natural of the surrounding environment. Most occlusal
wear occurs as a result of bruxism. An intuitive feeling is the harder the
occlusal material, the less the wear. However, surface hardness has
been shown to be a poor indicator of wear rate. Acrylic resin wears 7 to
30 times faster when opposing gold, resin, enamel, or polished
porcelain, compared with gold opposing gold, enamel or porcelain.
Gold occlusal surface exhibit less volume loss (sum of loss of opposing
occlusal surfaces) than any other combination of materials. Porcelain
opposing porcelain wears more than porcelain opposing gold or metal.
www.indiandentalacademy.com
The wear rate of occlusal materials, especially in the partially
edentulous patient with unrestored teeth, should be similar to enamel.
In this way, occlusal changes will not dramatically change the occlusal
scheme. Lambrechts et al in an in vivo investigation reported vertical
wear of premolar and molar tooth enamel to be 20 to 40 um per year
when opposing the enamel of natural teeth.
In principle, for the partially edentulous patient it would be
better to have more occlusal wear on implants, rather than less,
because the additional forces on the teeth are better tolerated than on
implant prostheses. As a result, total volume wear may favor porcelain
opposing enamel for the implant prosthesis opposing teeth in the
partially edentulous patient and metal opposing enamel in the other
regions of the mouth that require restorations of natural teeth.
The use of gold, regardless of the opposing combination,
always provides the least total volume loss.
www.indiandentalacademy.com
Adhesive wear occurs when one hard surface slides over a
surface of lesser hardness. As a result, wear fragments of one
material adhere to the other material. Gold occlusal surfaces are
observed to have gold particles adhered to enamel. This may account
for less total volume loss when opposing other materials.
For full arch implant supported prostheses, the restoring
doctor may consider metal occlusal to minimize wear and prolong
the accuracy of occlusal scheme long term. Porcelain in esthetic
regions opposing gold in the more nonesthetic area or metal occlusal
in both arches when parafunction or marginal interarch space is
present are the material most often selected as implant occlusal
materials.
www.indiandentalacademy.com
Materials fracture:
Materials fracture is one of the more common factors that lead
to refabrication of a prosthesis. Porcelain, acrylic and composite
fractures occur under excessive loads or even with a lesser load of
longer duration, angulation, or frequency. Acrylic or composite
materials fracture more easily. The compressive strength of acrylic
resin is 11,000 psi, compared with 40,000 psi for enamel. Composite
resin is 3 times stronger than acrylic.
Porcelain opposing porcelain is not suggested with extreme
parafunction, because it may fracture more often than porcelain
opposing metal. Metal occlusals do not easily fracture, provide good
wear resistance, and have minimum impact load compared with
porcelain.
www.indiandentalacademy.com
Accuracy:
Metal shrinkage is 10 times less than porcelain or acrylic and
therefore permits the fabrication of a more passive casting. When
accuracy of the casting is paramount, as with screw retained
restorations, the occlusal material may make a significant difference.
This is most important in regions of long spans and / or with a large
volume of materials.
Interarch space:
Acrylic restorations receive their strength from bulk and
therefore require greater interarch space. Metal occlusals require the
least amount of space. In addition, when increased retention of a
cement retained prosthesis is required, a high abutment and greater
retention may be accomplished with a metal occlusal. Porcelain is
intermediate in the interarch space requirement.www.indiandentalacademy.com
Therefore when all seven criteria are evaluated, metal is
an excellent occlusal material, with improved properties in
accuracy, wear, fracture resistance, abutment retention, and good
qualities for impact or static force. Esthetics is best satisfied with
porcelain, which has improved properties compared with acrylic
concerning fractures and retention.
www.indiandentalacademy.com
REVIEW OF LITERATURE
J.B. Brunskin and J.A. Hipp in 1984 studied the in vivo
forces on dental implants. Methods are presented for
measuring vertical force components or bridged titanium
dental implants in dog mandibles. These methods have
included custom made strain gauge transducers, plus
hard wiring and telemetric schemes for data collection.
The essential components of the measurements system
are described, and typical bite force data are illustrated
www.indiandentalacademy.com
Rangert et al in 1989 carried out a study on the
forces and moments on Brenamark implants. The
placement of fixture (implants) in relation to the geometry
of a prosthetic restoration has a great influence on the
mechanical loading of the implant. Based on Theoretic
consideration and clinical experiences with the Brenamark
system, this article gives simples guidelines for controlling
these loads. The emphasis is on design rules that can be
used in clinical practice. With the class I lever as a
reference. Various clinical implant prosthesis situations are
discussed and evaluated.
www.indiandentalacademy.com
Parker et al in 1991 reviewed the occlusal considerations in restorative
dentistry. The major topics include the assessment and treatment of
occlusal wear, the controversies surrounding treatment position of the
mandibular condyles, occlusal considerations in osseointegrated
prosthesis, the two way relationship between occlusal factors and
temporomandibular disorders, design criteria and longevity studies in
resin bonded, fixed partial denture.
Hobo et al in 1991 presented a case report on occlusion for
osseointegrated prosthesis and concluded that the concept of occlusion
suitable for osseointegrated prosthesis is basically the same as the
gnathological occlusion. However the natural tooth sinks about 30µm
during function, while an osseointegrated bridge which is supporte
only by the bone does not sink. Therefore it is necessary to adjust the
centric contacts of the osseointegrated fixed bridge slightly more open
than the natural teeth. During the eccentric movement, in order to
minimize horizontal loading, the concept of disocclusion is generally
used. www.indiandentalacademy.com
James et al in 1993 discussed the edentulous implants
an emphasized that the occlusal contacts of the final
fixed restoration are affected significantly by implant
position. Lateral occlusal forces, may lead to abutment
screw fracture. They may be due to either excessive
lateral occlusal pressure or a malposed implant that
requires non axial loading during normal function.
www.indiandentalacademy.com
C.E. Misch et al in 1994 discussed an implant protected
occlusal on a biomechanical rationale. The clinical
success and longevity of endosteal dental implants are
controlled, in a large part, by the mechanical milieu
within which they function. The occlusion is a critical
component of such a mechanical environment. “Implant
protected occlusion” refer to an occlusal scheme that is
often uniquely specific to the restoration of endosteal
implant prosthesis. Implant orientation and the
influence of load direction, the surface area of implants,
occlusal table width, and protecting the weakest area are
blended together from a biomechanical rationale to
provide support for a specific occlusal philosophy.
www.indiandentalacademy.com
Tashkandi et al in 1996 did an analysis of strain at
selected bone sites of a cantilevered implant supported
prosthesis. the results revealed that the maximum strain
occurred at the strain gauge positioned on cortical bone
over the apex of the most distal implant under 10 and 20 lb
loading conditions.
Osamu et al in 2002 did a study on influence of supra
structure materials on strain around an implant under two
loading conditions. The results showed under static and
non impact dynamic loading the three super structure
materials tested (highly filled composite resin, acrylic resin
and gold alloy), had the same influence on the strain
transmitted to the bone simulant that surrounded a single
implant. www.indiandentalacademy.com
Kent et al in 2004 did a photoelectric analysis of the effect
of palatal support on various implant supported overdenture
designs and concluded that at the removal of the palatal
support produces a greater effect and more concentrated
stress difference for maxillary overdenture than difference
between the attachment designs.
With palatal coverage Without palatal coverage
www.indiandentalacademy.com
Steven et al in 2004 did a study of stress transfer of four
mandibular implant overdenture cantilevered designs. His
results condluded that under load all prosthetic designs
demonstrated a low stress transfer to the ipsilateral
abutment and to the contralateral side of the arch. The
plunger retained prosthesis retained by two implants
demonstrated a more uniform stress trasnfer to the
ipsilateral terminal abutment than the clip retained
prosthesis retained by three implants and provided more
retention.
www.indiandentalacademy.com
Lucie et al in 2004 Did a finite element analysis on the influence of
implant length and diameter on stress distribution. Results showed
an increase in the implant diameter decreased the maximum von
Mises equivalent stress around the implant neck more than an
increase in the implant length, as a result of a more favorable
distribution of the simulated masticatory forces applied in this study.
www.indiandentalacademy.com
Osseointegrated supported prosthesis (ISP) have shown high standard
of success. This success rate depends not only on meticulous surgical
protocol but also on understanding concept of occlusion.
Occlusion should be a key factor to overall success rate. The concept
of occlusion suitable for implant supported prosthesis is basically the
same as gnathological occlusion.
In Centric contact of the Osseo integrated crown or fixed prosthesis
should be slightly more open than natural teeth.
In centric Osseo integrated crown or fixed prosthesis should not
contact with opposing teeth under the soft bite pressure (to avoid the
occlusal load on the I.S.P. which leads overload the bone structure)
Under the strong bite pressure Osseo integrated supported prosthesis
should contact after the natural tooth intrudes approximately 30µm.
SUMMARY
www.indiandentalacademy.com
• To avoid overloading of the occlusal surface, the I.S.P. should not
have plane-to-plane contact.
• Point contact especially cusp-to-fossa tripodal contact is preferred.
• During eccentric movement, in order to minimize horizontal
movement, the concept of disocclusion is generally recommended.
• Anterior segments of the osseointegrated prosthesis should guide the
mandible to produce posterior disocclusion.
• Canine guided occlusion is not recommend for the Osseo integrated
prosthesis to avoid excessive occlusal forces into the single implant
fixture which is placed in the canine area. Group function is
recommended to distribute the stress over the entire fixture.
• The ideal place to bear the horizontal load is the trapezoid area,
which is surrounded by the osseointegrated fixtures.
• Load transmitted to the fixture is not so destructive when extended
mesially in the anterior region, whereas more destructive when
extended distally.
www.indiandentalacademy.com
CONCLUSION
The local occlusal considerations in implant dentistry include
the transosteal forces, bone biomechanics, basic biomechanics,
differences in natural teeth and implants, muscles of mastication and
occlusal force, and bone resorption. The incorporation of all these
factors lead to an occlusal scheme (IPO) discussed in this seminar.
Occlusal schemes consider the weakest component, full or
partial edentulous arches, and posterior or anterior teeth and / or
implants. An IPO is a consistent approach for implant occlusal
schemes.
The material from which the occlusal region are fabricated
may affect implant loading and also affect implant reaction forces to
the opposing arch. These occlusal material also affect wear and
fracture, which affects the occlusal contacts, vertical occlusal
dimension, and esthetics.www.indiandentalacademy.com
References
1. Misch CE : Dentistry of bone and effect on treatment plans, surgical
approach, healing and progressive loading. Int. J. Oral implantol 6:23-31,
1990.
2. Misch CE : Progressive bone loading, Pract Periodontics Aesthet Dent.
2:27-30, 1990.
3. Misch CE : Progressive bone loading. In Misch CE, editor; Contemporary
implant dentistry, pp 623-650, St. Louis, 1993, Mosby.
4. Bidez MW, Misch CE : Force transfer in implant dentistry; basic concepts
and principles, Oral implant 18: 264-274, 1992.
5. Misch CE , Bidaz MW : Implant protected occlusion, a biomechanical
rationale, Comp Cont. Dent Educ. 15(11):1330-1343, 1994.
6. Misch CE : early crestal bone loss etiology and its effect on treatment
planning for implants, post Grad Dent. (2)3:3-17, 1995.
7. Jemt T, Linden B, Lekholm U : failures and complications in127
consecutively placed fixed partial prostheses supported by Branemark
implants ; from prosthetic treatments to first annual check up , Int. J. Oral
Maxillofac Impl. 7;40-44, 1992.
8. Isidor F : Loss of osteointegration caused by occlusal load of oral
implants, Clin Oral Implant Res.7:143-152, 1996.
www.indiandentalacademy.com
9. Misch CE : Medial positioned lingualized occlusion, Misch Institute
Manual, Birmingham, Mich, 1991.
10. Misch CE : Occlusal considerations for implant supported prostheses.
In Misch CE, editior: contemporary implant dentistry, pp 705-733, St.
Louis 1993, Mosby.
11. Muhlemann HR : Tooth mobility : review of clinical aspects and
research findings, J. Periodontal 38:686 1967.
12. Van Steenbergh D : A retrospective multicenter evaluation of the
survival rate of fixed prosthesis on four or six implants and modum
Branemark in full edentulism. J. Prosthet Dent. 61:217-223, 1989.
13. Chee WWL, Cho GC : A rationale for not connecting implants to
natural teeth, J. Prosthod.6(1) :7-10, 1997.
14. Goldstein GR : the relationship of canine protected occlusion to a
periodontal index, J. Prosthet. Dent. 41:277-283, 1979.
15. Williamson EH, Lundquist DO: Anterior guidance its effect on
electromyographic activity of the temporal and masseter muscles, J.
Prosthet Dent. 49:816-823, 1983.
16. Schupe RJ et al : effects of occlusal guidance on jaw muscles activity, J
Prosthet Dent 51:811-818, 1984.www.indiandentalacademy.com
17. Manns A, Chan C, Miralles R : influences of group function and canine
guidance on electromyographic activity of elevator muscles, J. Prosthet
Dent 57:494-501, 1987.
18. Ko CC DH, Hollister SJ : Micromechanics of implants /tissue
interfaces, J oral implantol 18: 220, 1992.
19. Misch CE: Three dimensional finite element analysis of two plate
form neck designs, Master’s thesis, University of Pittsburgh, 1989.
20. Clelland NL, Lee JK, Bimbenet OC et al : A three dimensional finite
element stress analysis of angled abutments for an implant placed in the
anterior maxilla, J. Prosthodont 4(2):95-100, 1995.
21. Papavasillou G, Kamposiora P et al : Three dimensional finite element
analysis of stress distribution around single tooth implants as a function
of bony support prosthesis type and loading during function, J. Prosthet
Dent 76: 633-640, 1996.
22. Reilly DT, Burstein AH : The elastic and ultimate properties of compact
bone tissue, J Biomech 80:393-405, 1975.
23. Cowin SC : Bone mechanics, Boca Raton , Fla 1989, CRC Press.
24. Parein AM, Eckert SE, Wollan PC et al : Implant reconstruction in the
posterior mandible : a long term retrospective study. J. Prosthet Dent
78:34-42, 1997.
www.indiandentalacademy.com
25. De Marco TL, Paine S : Mandibular flexure in opening and closure
movements, J Prosthet Dent. 31:482-485, 1974.
26. Chibirka RM, Razzoog ME, Lang BR et al : determining the force
absorption, quotient for restorative materials used in implant occlusal
surfaces, J. Prosthet Dent 67 (3): 361-364, 1992.
27. Naert I, Quirynen M, Van Steenberghe D et al : A six year prosthodontic
study of 509 consecutively inserted implants for the treatment of partial
edentulism. J. Prosthet Dent 67:236-245, 1992.
28. Shultz AW : comfort and chewing efficiency in dentures, J. Prosthet
Dent. 65:38-48, 1951.
29. Okesm JP : Management of Temporomandibular disorders and
occlusion, pp 259-260, St. Louis, 1989, Mosby.
30. Hudson JD, Goldstein GR, Georgescur M. Enamel wear caused by three
different restorative material, J. Prosthet 74:647-654, 1995.
31. Monasky GE, Tough DF: Studies of wear of porcelain, enamel and gold,
J Proshet Dent 25(3):299-306, 1971.
32. Krejci I,Lutz F,Reimer M et al: Wear of ceramic inlays, their enamel
antagonist and luting cements, J. Prosthet Dent 69: 425-431, 1993.
33. Seghi RR, Daher, Caputo A : Relative flexural strength of dental
restorative ceramics, Dent. Mater 6:181-184, 1990.
www.indiandentalacademy.com
THANK YOU
www.indiandentalacademy.com
For more details please visit
www.indiandentalacademy.com

Contenu connexe

Tendances

02- Occlusion in prosthodontics. Balanced occlusion
02- Occlusion in prosthodontics. Balanced occlusion02- Occlusion in prosthodontics. Balanced occlusion
02- Occlusion in prosthodontics. Balanced occlusionAmal Kaddah
 
Implants bio mechanics /certified fixed orthodontic courses by Indian denta...
Implants bio mechanics   /certified fixed orthodontic courses by Indian denta...Implants bio mechanics   /certified fixed orthodontic courses by Indian denta...
Implants bio mechanics /certified fixed orthodontic courses by Indian denta...Indian dental academy
 
occlusion in implant
occlusion in implantocclusion in implant
occlusion in implants farrokhi
 
Stress & force factors in implants /certified fixed orthodontic courses by I...
Stress & force factors in implants  /certified fixed orthodontic courses by I...Stress & force factors in implants  /certified fixed orthodontic courses by I...
Stress & force factors in implants /certified fixed orthodontic courses by I...Indian dental academy
 
Functionally generated path occlusion final
Functionally generated path  occlusion finalFunctionally generated path  occlusion final
Functionally generated path occlusion finalDr Mujtaba Ashraf
 
Concepts of orthodontic bracket positioning techniques / fixed orthodontics c...
Concepts of orthodontic bracket positioning techniques / fixed orthodontics c...Concepts of orthodontic bracket positioning techniques / fixed orthodontics c...
Concepts of orthodontic bracket positioning techniques / fixed orthodontics c...Indian dental academy
 
Over dentures and its attachment
Over dentures and its attachmentOver dentures and its attachment
Over dentures and its attachmentChaithraPrabhu3
 
loading of dental implants/ dental courses
loading of dental implants/ dental coursesloading of dental implants/ dental courses
loading of dental implants/ dental coursesIndian dental academy
 
Luting cements/prosthodontic courses
Luting cements/prosthodontic coursesLuting cements/prosthodontic courses
Luting cements/prosthodontic coursesIndian dental academy
 
Dentalxp 10 tips for dental implant placement
Dentalxp 10 tips for dental  implant placementDentalxp 10 tips for dental  implant placement
Dentalxp 10 tips for dental implant placementMuaiyed Mahmoud Buzayan
 
Biology of tooth movement
Biology of tooth movementBiology of tooth movement
Biology of tooth movementranjits275
 
Canine retraction in pre adjusted edgewise technique /certified fixed orthodo...
Canine retraction in pre adjusted edgewise technique /certified fixed orthodo...Canine retraction in pre adjusted edgewise technique /certified fixed orthodo...
Canine retraction in pre adjusted edgewise technique /certified fixed orthodo...Indian dental academy
 
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...Indian dental academy
 
Stability in complete dentures
Stability in complete denturesStability in complete dentures
Stability in complete dentureszainab khan
 
Occlusal Considerations For Implant Supported Prostheses Implant Protectes O...
Occlusal Considerations For Implant Supported  Prostheses Implant Protectes O...Occlusal Considerations For Implant Supported  Prostheses Implant Protectes O...
Occlusal Considerations For Implant Supported Prostheses Implant Protectes O...Mohammed Alshehri
 
Removable partial denture _ Midterm
Removable partial denture _ MidtermRemovable partial denture _ Midterm
Removable partial denture _ MidtermMO'men AbuDaif
 

Tendances (20)

02- Occlusion in prosthodontics. Balanced occlusion
02- Occlusion in prosthodontics. Balanced occlusion02- Occlusion in prosthodontics. Balanced occlusion
02- Occlusion in prosthodontics. Balanced occlusion
 
Implants bio mechanics /certified fixed orthodontic courses by Indian denta...
Implants bio mechanics   /certified fixed orthodontic courses by Indian denta...Implants bio mechanics   /certified fixed orthodontic courses by Indian denta...
Implants bio mechanics /certified fixed orthodontic courses by Indian denta...
 
Implant prosthesis occlusion
Implant prosthesis occlusionImplant prosthesis occlusion
Implant prosthesis occlusion
 
occlusion in implant
occlusion in implantocclusion in implant
occlusion in implant
 
Stress & force factors in implants /certified fixed orthodontic courses by I...
Stress & force factors in implants  /certified fixed orthodontic courses by I...Stress & force factors in implants  /certified fixed orthodontic courses by I...
Stress & force factors in implants /certified fixed orthodontic courses by I...
 
Functionally generated path occlusion final
Functionally generated path  occlusion finalFunctionally generated path  occlusion final
Functionally generated path occlusion final
 
Concepts of orthodontic bracket positioning techniques / fixed orthodontics c...
Concepts of orthodontic bracket positioning techniques / fixed orthodontics c...Concepts of orthodontic bracket positioning techniques / fixed orthodontics c...
Concepts of orthodontic bracket positioning techniques / fixed orthodontics c...
 
Over dentures and its attachment
Over dentures and its attachmentOver dentures and its attachment
Over dentures and its attachment
 
loading of dental implants/ dental courses
loading of dental implants/ dental coursesloading of dental implants/ dental courses
loading of dental implants/ dental courses
 
Luting cements/prosthodontic courses
Luting cements/prosthodontic coursesLuting cements/prosthodontic courses
Luting cements/prosthodontic courses
 
Implant loading
Implant loading  Implant loading
Implant loading
 
Dentalxp 10 tips for dental implant placement
Dentalxp 10 tips for dental  implant placementDentalxp 10 tips for dental  implant placement
Dentalxp 10 tips for dental implant placement
 
Biology of tooth movement
Biology of tooth movementBiology of tooth movement
Biology of tooth movement
 
Canine retraction in pre adjusted edgewise technique /certified fixed orthodo...
Canine retraction in pre adjusted edgewise technique /certified fixed orthodo...Canine retraction in pre adjusted edgewise technique /certified fixed orthodo...
Canine retraction in pre adjusted edgewise technique /certified fixed orthodo...
 
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
 
Bonding and banding
Bonding and bandingBonding and banding
Bonding and banding
 
Implant occlusion
 Implant occlusion Implant occlusion
Implant occlusion
 
Stability in complete dentures
Stability in complete denturesStability in complete dentures
Stability in complete dentures
 
Occlusal Considerations For Implant Supported Prostheses Implant Protectes O...
Occlusal Considerations For Implant Supported  Prostheses Implant Protectes O...Occlusal Considerations For Implant Supported  Prostheses Implant Protectes O...
Occlusal Considerations For Implant Supported Prostheses Implant Protectes O...
 
Removable partial denture _ Midterm
Removable partial denture _ MidtermRemovable partial denture _ Midterm
Removable partial denture _ Midterm
 

En vedette

Biomechanics in fixed partial prosthodontics /certified fixed orthodontic cou...
Biomechanics in fixed partial prosthodontics /certified fixed orthodontic cou...Biomechanics in fixed partial prosthodontics /certified fixed orthodontic cou...
Biomechanics in fixed partial prosthodontics /certified fixed orthodontic cou...Indian dental academy
 
Occulasl consideration for implant supported prostehsi/ dentistry jobs
Occulasl consideration for implant supported prostehsi/ dentistry jobsOcculasl consideration for implant supported prostehsi/ dentistry jobs
Occulasl consideration for implant supported prostehsi/ dentistry jobsIndian dental academy
 
Seminario 10
Seminario 10Seminario 10
Seminario 10igorcorso
 
Implant supported fixed bridge for edentulous mandible
Implant supported fixed bridge for edentulous mandibleImplant supported fixed bridge for edentulous mandible
Implant supported fixed bridge for edentulous mandibleBotan Khafaf
 
Presentación Prótesis Fija Plura
Presentación Prótesis Fija PluraPresentación Prótesis Fija Plura
Presentación Prótesis Fija PluraAxelJara
 
occlusal considerations for Implant supported Prosthesis /certified fixed or...
occlusal considerations for Implant supported Prosthesis  /certified fixed or...occlusal considerations for Implant supported Prosthesis  /certified fixed or...
occlusal considerations for Implant supported Prosthesis /certified fixed or...Indian dental academy
 
Principles of crown preparation/ orthodontic seminars
Principles of crown preparation/ orthodontic seminarsPrinciples of crown preparation/ orthodontic seminars
Principles of crown preparation/ orthodontic seminarsIndian dental academy
 
INTRODUCTION TO ARTICULATORS
INTRODUCTION TO ARTICULATORSINTRODUCTION TO ARTICULATORS
INTRODUCTION TO ARTICULATORSPatel telecom
 
Properties of orthodontic wires /certified fixed orthodontic courses by India...
Properties of orthodontic wires /certified fixed orthodontic courses by India...Properties of orthodontic wires /certified fixed orthodontic courses by India...
Properties of orthodontic wires /certified fixed orthodontic courses by India...Indian dental academy
 
Abutment evaluation / cosmetic dentistry training
Abutment evaluation  / cosmetic dentistry trainingAbutment evaluation  / cosmetic dentistry training
Abutment evaluation / cosmetic dentistry trainingIndian dental academy
 
Pontics Design in fixed prosthodontics
Pontics Design in fixed prosthodonticsPontics Design in fixed prosthodontics
Pontics Design in fixed prosthodonticsIndian dental academy
 
Types of bridges/fixed orthodontics courses
Types of bridges/fixed orthodontics coursesTypes of bridges/fixed orthodontics courses
Types of bridges/fixed orthodontics coursesIndian dental academy
 
Different gingival finish lines (margins) of crowns and bridges
Different gingival finish lines (margins) of crowns and bridgesDifferent gingival finish lines (margins) of crowns and bridges
Different gingival finish lines (margins) of crowns and bridgesSana Mateen Munshi
 
Biomechanics of tooth movement /certified fixed orthodontic courses by Indian...
Biomechanics of tooth movement /certified fixed orthodontic courses by Indian...Biomechanics of tooth movement /certified fixed orthodontic courses by Indian...
Biomechanics of tooth movement /certified fixed orthodontic courses by Indian...Indian dental academy
 
Distal extension removable partial denture prosthesis /certified fixed orthod...
Distal extension removable partial denture prosthesis /certified fixed orthod...Distal extension removable partial denture prosthesis /certified fixed orthod...
Distal extension removable partial denture prosthesis /certified fixed orthod...Indian dental academy
 

En vedette (20)

Biomechanics in fixed partial prosthodontics /certified fixed orthodontic cou...
Biomechanics in fixed partial prosthodontics /certified fixed orthodontic cou...Biomechanics in fixed partial prosthodontics /certified fixed orthodontic cou...
Biomechanics in fixed partial prosthodontics /certified fixed orthodontic cou...
 
Occulasl consideration for implant supported prostehsi/ dentistry jobs
Occulasl consideration for implant supported prostehsi/ dentistry jobsOcculasl consideration for implant supported prostehsi/ dentistry jobs
Occulasl consideration for implant supported prostehsi/ dentistry jobs
 
Seminario 10
Seminario 10Seminario 10
Seminario 10
 
Implant supported fixed bridge for edentulous mandible
Implant supported fixed bridge for edentulous mandibleImplant supported fixed bridge for edentulous mandible
Implant supported fixed bridge for edentulous mandible
 
Edentulous Mandible - Fixed Prostheses
Edentulous Mandible - Fixed ProsthesesEdentulous Mandible - Fixed Prostheses
Edentulous Mandible - Fixed Prostheses
 
Pontics in fpd
Pontics in fpdPontics in fpd
Pontics in fpd
 
Presentación Prótesis Fija Plura
Presentación Prótesis Fija PluraPresentación Prótesis Fija Plura
Presentación Prótesis Fija Plura
 
occlusal considerations for Implant supported Prosthesis /certified fixed or...
occlusal considerations for Implant supported Prosthesis  /certified fixed or...occlusal considerations for Implant supported Prosthesis  /certified fixed or...
occlusal considerations for Implant supported Prosthesis /certified fixed or...
 
Principles of crown preparation/ orthodontic seminars
Principles of crown preparation/ orthodontic seminarsPrinciples of crown preparation/ orthodontic seminars
Principles of crown preparation/ orthodontic seminars
 
INTRODUCTION TO ARTICULATORS
INTRODUCTION TO ARTICULATORSINTRODUCTION TO ARTICULATORS
INTRODUCTION TO ARTICULATORS
 
Properties of orthodontic wires /certified fixed orthodontic courses by India...
Properties of orthodontic wires /certified fixed orthodontic courses by India...Properties of orthodontic wires /certified fixed orthodontic courses by India...
Properties of orthodontic wires /certified fixed orthodontic courses by India...
 
Abutment evaluation / cosmetic dentistry training
Abutment evaluation  / cosmetic dentistry trainingAbutment evaluation  / cosmetic dentistry training
Abutment evaluation / cosmetic dentistry training
 
Pontics Design in fixed prosthodontics
Pontics Design in fixed prosthodonticsPontics Design in fixed prosthodontics
Pontics Design in fixed prosthodontics
 
Types of bridges/fixed orthodontics courses
Types of bridges/fixed orthodontics coursesTypes of bridges/fixed orthodontics courses
Types of bridges/fixed orthodontics courses
 
Different gingival finish lines (margins) of crowns and bridges
Different gingival finish lines (margins) of crowns and bridgesDifferent gingival finish lines (margins) of crowns and bridges
Different gingival finish lines (margins) of crowns and bridges
 
Articulator
ArticulatorArticulator
Articulator
 
Biomechanics of tooth movement /certified fixed orthodontic courses by Indian...
Biomechanics of tooth movement /certified fixed orthodontic courses by Indian...Biomechanics of tooth movement /certified fixed orthodontic courses by Indian...
Biomechanics of tooth movement /certified fixed orthodontic courses by Indian...
 
Articulators
ArticulatorsArticulators
Articulators
 
Distal extension removable partial denture prosthesis /certified fixed orthod...
Distal extension removable partial denture prosthesis /certified fixed orthod...Distal extension removable partial denture prosthesis /certified fixed orthod...
Distal extension removable partial denture prosthesis /certified fixed orthod...
 
Pontics
PonticsPontics
Pontics
 

Similaire à Occulasl consideration for implant supported prostehsis

Splinting
Splinting Splinting
Splinting puri456
 
Occlusal considerations for implant / orthodontic seminars
Occlusal considerations for implant / orthodontic seminarsOcclusal considerations for implant / orthodontic seminars
Occlusal considerations for implant / orthodontic seminarsIndian dental academy
 
Stress & force factors /certified fixed orthodontic courses by Indian dental ...
Stress & force factors /certified fixed orthodontic courses by Indian dental ...Stress & force factors /certified fixed orthodontic courses by Indian dental ...
Stress & force factors /certified fixed orthodontic courses by Indian dental ...Indian dental academy
 
Support in complete denture /orthodontic courses by Indian dental academy 
Support in complete denture /orthodontic courses by Indian dental academy Support in complete denture /orthodontic courses by Indian dental academy 
Support in complete denture /orthodontic courses by Indian dental academy Indian dental academy
 
classification, Type of Fixtures Sterilization and Passivation/endodontic cou...
classification, Type of Fixtures Sterilization and Passivation/endodontic cou...classification, Type of Fixtures Sterilization and Passivation/endodontic cou...
classification, Type of Fixtures Sterilization and Passivation/endodontic cou...Indian dental academy
 
dr.neha srivastava ipo seminar.pptx
dr.neha srivastava ipo seminar.pptxdr.neha srivastava ipo seminar.pptx
dr.neha srivastava ipo seminar.pptxnehasrivastava643617
 
Single complete denture
Single complete dentureSingle complete denture
Single complete denturepriyanka konda
 
Implants /certified fixed orthodontic courses by Indian dental academy
Implants   /certified fixed orthodontic courses by Indian dental academy Implants   /certified fixed orthodontic courses by Indian dental academy
Implants /certified fixed orthodontic courses by Indian dental academy Indian dental academy
 
Tooth implant connection, fix partial dentur,
Tooth implant connection,  fix partial dentur,Tooth implant connection,  fix partial dentur,
Tooth implant connection, fix partial dentur,ali moazam
 
Splinting part i /certified fixed orthodontic courses by Indian dental academy
Splinting part i /certified fixed orthodontic courses by Indian dental academy Splinting part i /certified fixed orthodontic courses by Indian dental academy
Splinting part i /certified fixed orthodontic courses by Indian dental academy Indian dental academy
 
Single maxillary denture/ cosmetic dentistry training
Single maxillary denture/ cosmetic dentistry trainingSingle maxillary denture/ cosmetic dentistry training
Single maxillary denture/ cosmetic dentistry trainingIndian dental academy
 
Dental Implants / dental implants courses
Dental Implants / dental implants coursesDental Implants / dental implants courses
Dental Implants / dental implants coursesIndian dental academy
 
Single maxillary denture /invisible aligners
Single maxillary denture /invisible alignersSingle maxillary denture /invisible aligners
Single maxillary denture /invisible alignersIndian dental academy
 
Occlusal considerations in rpd/endodontic courses
Occlusal considerations in rpd/endodontic coursesOcclusal considerations in rpd/endodontic courses
Occlusal considerations in rpd/endodontic coursesIndian dental academy
 

Similaire à Occulasl consideration for implant supported prostehsis (20)

Splinting
Splinting Splinting
Splinting
 
Occlusal considerations for implant / orthodontic seminars
Occlusal considerations for implant / orthodontic seminarsOcclusal considerations for implant / orthodontic seminars
Occlusal considerations for implant / orthodontic seminars
 
Stress & force factors /certified fixed orthodontic courses by Indian dental ...
Stress & force factors /certified fixed orthodontic courses by Indian dental ...Stress & force factors /certified fixed orthodontic courses by Indian dental ...
Stress & force factors /certified fixed orthodontic courses by Indian dental ...
 
100008874.pptx
100008874.pptx100008874.pptx
100008874.pptx
 
100008874 (1).ppt
100008874 (1).ppt100008874 (1).ppt
100008874 (1).ppt
 
Over denture
Over dentureOver denture
Over denture
 
Support in complete denture /orthodontic courses by Indian dental academy 
Support in complete denture /orthodontic courses by Indian dental academy Support in complete denture /orthodontic courses by Indian dental academy 
Support in complete denture /orthodontic courses by Indian dental academy 
 
classification, Type of Fixtures Sterilization and Passivation/endodontic cou...
classification, Type of Fixtures Sterilization and Passivation/endodontic cou...classification, Type of Fixtures Sterilization and Passivation/endodontic cou...
classification, Type of Fixtures Sterilization and Passivation/endodontic cou...
 
dr.neha srivastava ipo seminar.pptx
dr.neha srivastava ipo seminar.pptxdr.neha srivastava ipo seminar.pptx
dr.neha srivastava ipo seminar.pptx
 
Single complete denture
Single complete dentureSingle complete denture
Single complete denture
 
Implants /certified fixed orthodontic courses by Indian dental academy
Implants   /certified fixed orthodontic courses by Indian dental academy Implants   /certified fixed orthodontic courses by Indian dental academy
Implants /certified fixed orthodontic courses by Indian dental academy
 
Tooth implant connection, fix partial dentur,
Tooth implant connection,  fix partial dentur,Tooth implant connection,  fix partial dentur,
Tooth implant connection, fix partial dentur,
 
Splinting part i /certified fixed orthodontic courses by Indian dental academy
Splinting part i /certified fixed orthodontic courses by Indian dental academy Splinting part i /certified fixed orthodontic courses by Indian dental academy
Splinting part i /certified fixed orthodontic courses by Indian dental academy
 
Single complete denture final ppt - copy
Single complete denture final ppt - copySingle complete denture final ppt - copy
Single complete denture final ppt - copy
 
Single maxillary denture/ cosmetic dentistry training
Single maxillary denture/ cosmetic dentistry trainingSingle maxillary denture/ cosmetic dentistry training
Single maxillary denture/ cosmetic dentistry training
 
Splinting
SplintingSplinting
Splinting
 
Dental Implants / dental implants courses
Dental Implants / dental implants coursesDental Implants / dental implants courses
Dental Implants / dental implants courses
 
13.natural vs. complete centure occlusion
13.natural vs. complete centure occlusion13.natural vs. complete centure occlusion
13.natural vs. complete centure occlusion
 
Single maxillary denture /invisible aligners
Single maxillary denture /invisible alignersSingle maxillary denture /invisible aligners
Single maxillary denture /invisible aligners
 
Occlusal considerations in rpd/endodontic courses
Occlusal considerations in rpd/endodontic coursesOcclusal considerations in rpd/endodontic courses
Occlusal considerations in rpd/endodontic courses
 

Plus de Indian dental academy

Indian Dentist - relocate to united kingdom
Indian Dentist - relocate to united kingdomIndian Dentist - relocate to united kingdom
Indian Dentist - relocate to united kingdomIndian dental academy
 
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...Indian dental academy
 
Invisalign -invisible aligners course in india
Invisalign -invisible aligners course in india Invisalign -invisible aligners course in india
Invisalign -invisible aligners course in india Indian dental academy
 
Invisible aligners for your orthodontics pratice
Invisible aligners for your orthodontics praticeInvisible aligners for your orthodontics pratice
Invisible aligners for your orthodontics praticeIndian dental academy
 
Development of muscles of mastication / dental implant courses
Development of muscles of mastication / dental implant coursesDevelopment of muscles of mastication / dental implant courses
Development of muscles of mastication / dental implant coursesIndian dental academy
 
Corticosteriods uses in dentistry/ oral surgery courses  
Corticosteriods uses in dentistry/ oral surgery courses  Corticosteriods uses in dentistry/ oral surgery courses  
Corticosteriods uses in dentistry/ oral surgery courses  Indian dental academy
 
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...Indian dental academy
 
Diagnosis and treatment planning in completely endntulous arches/dental courses
Diagnosis and treatment planning in completely endntulous arches/dental coursesDiagnosis and treatment planning in completely endntulous arches/dental courses
Diagnosis and treatment planning in completely endntulous arches/dental coursesIndian dental academy
 
Properties of Denture base materials /rotary endodontic courses
Properties of Denture base materials /rotary endodontic coursesProperties of Denture base materials /rotary endodontic courses
Properties of Denture base materials /rotary endodontic coursesIndian dental academy
 
Use of modified tooth forms in complete denture occlusion / dental implant...
Use of modified  tooth forms  in  complete denture occlusion / dental implant...Use of modified  tooth forms  in  complete denture occlusion / dental implant...
Use of modified tooth forms in complete denture occlusion / dental implant...Indian dental academy
 
Dental luting cements / oral surgery courses  
Dental   luting cements / oral surgery courses  Dental   luting cements / oral surgery courses  
Dental luting cements / oral surgery courses  Indian dental academy
 
Dental casting alloys/ oral surgery courses  
Dental casting alloys/ oral surgery courses  Dental casting alloys/ oral surgery courses  
Dental casting alloys/ oral surgery courses  Indian dental academy
 
Dental casting investment materials/endodontic courses
Dental casting investment materials/endodontic coursesDental casting investment materials/endodontic courses
Dental casting investment materials/endodontic coursesIndian dental academy
 
Dental casting waxes/ oral surgery courses  
Dental casting waxes/ oral surgery courses  Dental casting waxes/ oral surgery courses  
Dental casting waxes/ oral surgery courses  Indian dental academy
 
Dental ceramics/prosthodontic courses
Dental ceramics/prosthodontic coursesDental ceramics/prosthodontic courses
Dental ceramics/prosthodontic coursesIndian dental academy
 
Dental implant/ oral surgery courses  
Dental implant/ oral surgery courses  Dental implant/ oral surgery courses  
Dental implant/ oral surgery courses  Indian dental academy
 
Dental perspective/cosmetic dentistry courses
Dental perspective/cosmetic dentistry coursesDental perspective/cosmetic dentistry courses
Dental perspective/cosmetic dentistry coursesIndian dental academy
 
Dental tissues and their replacements/ oral surgery courses  
Dental tissues and their replacements/ oral surgery courses  Dental tissues and their replacements/ oral surgery courses  
Dental tissues and their replacements/ oral surgery courses  Indian dental academy
 

Plus de Indian dental academy (20)

Indian Dentist - relocate to united kingdom
Indian Dentist - relocate to united kingdomIndian Dentist - relocate to united kingdom
Indian Dentist - relocate to united kingdom
 
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
 
Invisalign -invisible aligners course in india
Invisalign -invisible aligners course in india Invisalign -invisible aligners course in india
Invisalign -invisible aligners course in india
 
Invisible aligners for your orthodontics pratice
Invisible aligners for your orthodontics praticeInvisible aligners for your orthodontics pratice
Invisible aligners for your orthodontics pratice
 
online fixed orthodontics course
online fixed orthodontics courseonline fixed orthodontics course
online fixed orthodontics course
 
online orthodontics course
online orthodontics courseonline orthodontics course
online orthodontics course
 
Development of muscles of mastication / dental implant courses
Development of muscles of mastication / dental implant coursesDevelopment of muscles of mastication / dental implant courses
Development of muscles of mastication / dental implant courses
 
Corticosteriods uses in dentistry/ oral surgery courses  
Corticosteriods uses in dentistry/ oral surgery courses  Corticosteriods uses in dentistry/ oral surgery courses  
Corticosteriods uses in dentistry/ oral surgery courses  
 
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
 
Diagnosis and treatment planning in completely endntulous arches/dental courses
Diagnosis and treatment planning in completely endntulous arches/dental coursesDiagnosis and treatment planning in completely endntulous arches/dental courses
Diagnosis and treatment planning in completely endntulous arches/dental courses
 
Properties of Denture base materials /rotary endodontic courses
Properties of Denture base materials /rotary endodontic coursesProperties of Denture base materials /rotary endodontic courses
Properties of Denture base materials /rotary endodontic courses
 
Use of modified tooth forms in complete denture occlusion / dental implant...
Use of modified  tooth forms  in  complete denture occlusion / dental implant...Use of modified  tooth forms  in  complete denture occlusion / dental implant...
Use of modified tooth forms in complete denture occlusion / dental implant...
 
Dental luting cements / oral surgery courses  
Dental   luting cements / oral surgery courses  Dental   luting cements / oral surgery courses  
Dental luting cements / oral surgery courses  
 
Dental casting alloys/ oral surgery courses  
Dental casting alloys/ oral surgery courses  Dental casting alloys/ oral surgery courses  
Dental casting alloys/ oral surgery courses  
 
Dental casting investment materials/endodontic courses
Dental casting investment materials/endodontic coursesDental casting investment materials/endodontic courses
Dental casting investment materials/endodontic courses
 
Dental casting waxes/ oral surgery courses  
Dental casting waxes/ oral surgery courses  Dental casting waxes/ oral surgery courses  
Dental casting waxes/ oral surgery courses  
 
Dental ceramics/prosthodontic courses
Dental ceramics/prosthodontic coursesDental ceramics/prosthodontic courses
Dental ceramics/prosthodontic courses
 
Dental implant/ oral surgery courses  
Dental implant/ oral surgery courses  Dental implant/ oral surgery courses  
Dental implant/ oral surgery courses  
 
Dental perspective/cosmetic dentistry courses
Dental perspective/cosmetic dentistry coursesDental perspective/cosmetic dentistry courses
Dental perspective/cosmetic dentistry courses
 
Dental tissues and their replacements/ oral surgery courses  
Dental tissues and their replacements/ oral surgery courses  Dental tissues and their replacements/ oral surgery courses  
Dental tissues and their replacements/ oral surgery courses  
 

Occulasl consideration for implant supported prostehsis

  • 1. INDIAN DENTAL ACADEMY Leader in continuing dental education www.indiandentalacademy.com www.indiandentalacademy.com
  • 4. The clinical success and longevity of endosteal dental implants as load-bearing abutments are controlled largely by the mechanical setting in which they function. The treatment plan is responsible for the design, number and position of the implants. After achievement of rigid fixation, proper crestal bone contour, gingival health, mechanical stress, and/or strain beyond the physical limits of hard tissues have been suggested as the primary cause of initial bone loss around implants. After successful surgical and prosthetic rehabilitation with a passive prosthesis, such noxious stresses and loads applied to the implant and surrounding tissues result primarily from occlusal contacts. Complications (prosthetic and/or bony support) reported in follow-up studies underline occlusion as a determining factor for success or failure. www.indiandentalacademy.com
  • 5. The choice of an occlusal scheme for implant-supported prostheses is broad and often controversial. Almost all concepts are based on those developed with natural teeth and are transposed to implant support systems with almost no modification. No controlled clinical studies have been published comparing the various implant occlusal theories. RISK FACTORS - Implant prostheses with extended cantilevers have been successful, however, biomechanical factors clearly demonstrate an increased risk. Biomechanical parameters are excellent indicators of the increased risk because they are objective and can be measured. One can determine which condition presents greater risk, and by how much the risk is increased. Hence the occlusal concepts developed in this seminar stem from biomechanical risk factors. www.indiandentalacademy.com
  • 6. The prosthodontist has specific responsibilities to minimize overload to the bone-to-implant interface. These include a proper diagnosis leading to a treatment plan providing adequate support, based on the patient’s individual force factors, a passive prosthesis of adequate retention and form and progressive loading to improve the amount and density of the adjacent bone and further reduce the risk of stress beyond physiologic limits. The final element is the development of an occlusal scheme that minimizes risk factors and allows the restoration to function in harmony with the rest of the stomatognathic system. www.indiandentalacademy.com
  • 8. Anterior protected articulation A form of mutually protected articulation in which the vertical and horizontal overlap of the anterior teeth disengage the posterior teeth in all mandibular excursive movements. Balanced articulation The bilateral, simultaneous, anterior, and posterior occlusal contact of teeth in centric and eccentric positions Canine protected articulation A form of mutually protected articulation in which the vertical and horizontal overlap of the canine teeth disengage the posterior teeth in the excursive movements of the mandible. Implant Prosthodontics The phase of Prosthodontics concerning the replacement of missing teeth and / or associated structures by restorations that are attached to dental implants. www.indiandentalacademy.com
  • 9. Lingualized occlusion First described by S. Howard Payne, DDS, in 1941, this form of denture occlusion articulates the maxillary lingual cusps with the mandibular occlusal surfaces in centric working and nonworking mandibular positions. The term is attributed to Earl Pound. Payne SH. A posterior set up to meet individual requirements. Dent. 1941; 47:20-2. Pound E. Utilizing speech to simplify a personalized denture service. J. Prosthet Dent. 1970;24:585-600. Linguocclusion An occlusion in which a tooth or group of teeth is located lingual to its normal position. www.indiandentalacademy.com
  • 10. Mutually protected articulation An occlusal scheme in which the posterior teeth prevent excessive contact of the anterior teeth in maximum intercuspation, and the anterior teeth disengage the posterior teeth in all mandibular excursive movements. Occlusion (1645) 1: The act or process of closure or of bring closed or shut off 2 : the static relationship between the incising or masticating surfaces of the maxillary or mandibular teeth or tooth analogues Spherical form of occlusion An arrangement of teeth that places their occlusal surfaces on the surfaces of an imaginary sphere (usually 8 inches in diameter) with its center above the level of the teeth (GPT-4). www.indiandentalacademy.com
  • 11. IMPLANT PROTECTIVE OCCLUSION (IPO) www.indiandentalacademy.com
  • 12. A proper occlusal scheme is a primary requisite for long-term survival, especially when parafunction or marginal foundations are present. A poor occlusal scheme both increases the magnitude of loads and intensifies mechanical stresses (and strain) at the crest of the bone. Implant Protective Occlusion (IPO) was previously known as medial positioned-lingualized occlusion. This occlusal concept refers to an occlusal plane that is often unique and specifically designed for the restoration of endosteal implant, providing an environment for improved clinical longevity of both implant and prosthesis. www.indiandentalacademy.com
  • 14. Natural Tooth vs. Implant Mobility: In comparison to an implant, the support system of a natural tooth is designed to reduce the forces distributed at the crestal bone. The fibrous tissue interface (periodontal ligament) surrounding natural teeth acts as a viscoelastic “shock absorber”, serving to both decrease the magnitude of stress to the bone at the crest, as well as extend the time in which the load is dissipated. The presence of a periodontal membrane around natural teeth significantly reduces the amount of stress transmitted to the bone, especially at the crestal region. Compared with a tooth the direct bone interface with an implant is not as resilient, so the energy imparted by an occlusal force is not partially dissipated (the displacement of the periodontal membrane dissipates energy), but rather transmits a higher intensity to the contiguous bone. An analogy of this is hitting a nail with a steel hammer compared with a rubber hammer. www.indiandentalacademy.com
  • 15. The mobility of a natural tooth can increase with occlusal trauma. This movement dissipates stresses and strains otherwise imposed on the adjacent bone interface or the prosthetic components. After the occlusal trauma is eliminated, the tooth can return to its original condition with respect to the magnitude of movement. Mobility of an implant can also develop under occlusal trauma. However, after the offending element is eliminated, an implant rarely returns to its original rigid condition. Instead, its health is compromised, and failure is usually eminent. The width of almost every natural tooth is greater than the width of the implant used to replace the tooth. The greater the width of a transosteal structure (tooth or implant), the lesser magnitude of stress transmitted to the surrounding bone. The cross-section shape of the natural tooth at the crest is biomechanically optimized to resist lateral (buccolingual) loads because of the tooth’s bending fracture resistance (moment of inertia) and the direction of occlusal forces. Implants are almost all round in cross-section, which is less effective in resisting lateral bending loads and consequent stress concentration in the crestal region in the jaws. www.indiandentalacademy.com
  • 16. The elastic modulus of a tooth is closer to bone than any of the currently available dental implant biomaterials. The greater the flexibility difference between two materials (metal and bone or tooth and bone), the greater the potential relative motion generated between the two surfaces at the transosteal region. Hence under similar mechanical loading conditions, implants generate greater stresses and strains at the crest of bone compared with a tooth. The precursor signs of occlusal trauma on natural teeth are usually reversible and include hyperemia and occlusal or cold sensitivity. Condition often results with the patient seeking professional treatment to reduce the sensitivity, usually by occlusal adjustment and a reduction in force magnitude. If the patient does not have an occlusal adjustment, the tooth often further increases in mobility to dissipate the occlusal forces. If the patient still fails to seek professional treatment for the increased mobility, the tooth may orthodontically migrate away from the cause of the occlusal stress. Even excess tongue or oral habits can cause tooth migration away from the causative element.www.indiandentalacademy.com
  • 17. The initial reversible signs and symptoms of trauma on natural teeth do not occur with endosteal implants. The magnitude of stress may cause bone microfractures, place the surrounding bone in the pathologic loading zone causing bone loss, and lead to the mechanical failure of prosthetic or implant components. Unlike the reversible signs and symptoms exhibited by natural teeth, implant bone loss or unsecured restorations most often occur without any warning signs. Implant occlusal sensitivity is uncommon and signifies more advanced complications. The loss of crestal bone around the implant is not reversible without surgical intervention and results in a decreased implant support and increased sulcus depth around the abutment. As a result, unless the density of bone increases or the amount or duration of force decreases, the condition is likely to progress and even accelerate until implant loss. In addition, implants cannot move orthodontically away from the offending force. www.indiandentalacademy.com
  • 18. The tooth can show clinical signs of increased stress such as enamel wear facets, stress lines, lines of Luder (in amalgam fillings), cervical abfraction, and pits on the cusps of teeth. An implant crown rarely shows clinical signs other than fatigue fracture. As a result, fewer diagnostic signs are present to warn the prosthodontist to reduce the stress on the support system. www.indiandentalacademy.com
  • 19. When teeth oppose each other, an interference perception is approximately 20m. An implant opposing a natural tooth has an interference perception of 48m, therefore more than twice as poor. An implant opposing implant has an interference perception of 64m, and when a tooth opposes an implant overdenture the awareness is 108m (5 times poorer than teeth opposing each other). As a result, premature occlusal contacts on teeth are usually associated with a modification of arc of closure and with a decreased force, before centric occlusion or full interdigitation. In addition, the mandible may close in a different position to avoid the premature contact and result in centric occlusion different from centric relation occlusion. Unfortunately, because of the decreased occlusal awareness of implants, the premature contact does not trigger such as an adaptation. In addition, premature contacts are often on smaller areas of load and therefore result in greater stress (S=F/A). They are most often on inclines of posterior teeth, which also generates an angled load of greater stress to the implant bone interface.www.indiandentalacademy.com
  • 21. Implants and teeth also have different proprioceptive information relayed by both entities. Teeth deliver a rapid, sharp pain sensation under high pressure that triggers a protective mechanism. On the other hand, implants deliver a slow, dull pain that triggers a delayed reaction, if any. Clinical evidence of occlusal trauma on teeth includes an overall increase in the periodontal membrane thickness and an increased radiopacity and thickness of the cribriform plate around the tooth, observed on radiographs and not just localized at the crest. There are no generalized radiographic signs around an implant under excess occlusal force, except at the crestal region, which demonstrates bone loss but may be misdiagnosed as periimplant disease from bacteria. www.indiandentalacademy.com
  • 22. The tooth slowly erupts into occlusion and is present in the mouth from childhood. The surrounding bone has developed in response to the biomechanical loads. The permanent teeth are gradually introduced, while others are present. Hence periodontal tissues have had time to organize in order to sustain increasing loads, including those brought to bear by an attached prosthesis. The only progressive bone loading around an implant is performed by the prosthodontist, and in a much more rapid and intense fashion. A lateral force on a natural tooth is rapidly dissipated away from the crest of bone toward the apex of the tooth. The healthy, natural tooth moves almost immediately 56 to 108m and pivots two-thirds turn toward the tapered apex with a lateral load. This action minimizes crestal loads to the bone. An implant does not exhibit a primary immediate movement, but a secondary movement 10 to 50m under similar lateral loads. In addition, it does not pivot (as a tooth) toward the apex but instead concentrates greater forces at he crest of surrounding bone. Therefore if an initial angled load of equal magnitude and direction is placed on both an implant and a natural tooth, the implant sustains a higher proportion of the load that is not dissipated to the surrounding structures. www.indiandentalacademy.com
  • 23. The natural tooth, with its modulus of elasticity similar to bone, periodontal ligament, and unique cross-sections and dimensions constitutes a near perfect optimization system to handle stress. In fact, the stress is handled so well, bacteria-related disease is the weak link. An implant handles stress so poorly (capturing the stress at the crest of the ridge), has an elastic modulus 5 to 10 times that of bone, and is unable to increase mobility without failure that stress is the weakest link in the system. As a result, ways to decrease stress are a constant concern to minimize the risk of implants complications. www.indiandentalacademy.com
  • 25. Occlusion on Natural Teeth and Implants: There has been an ongoing controversy regarding whether a rigidly fixated implant may remain successful when splinted to natural teeth. Because the implant has no periodontal membrane, concerns center around the potential for the “nonmobile” implant to bear the total load of the prosthesis when joined to the “mobile” natural tooth. The actual mobility of potential natural abutments may influence the treatment more than any other factor. In the implant tooth fixed prosthesis, four important components may contribute movement to the system, the implant, bone, tooth, and prosthesis. The sudden, initial tooth movement ranges from 8 to 28m in a vertical direction under a 3 to 5 lb load, depending on the size, number, and geometry of the roots and the time elapsed since the last load application. Once the initial tooth movement occurs, the secondary tooth movement reflects the property of the surrounding bone and is very similar to the bone implant movement. The axial movement of an implant has no initial, sudden movement and ranges from 3 to 5 m with little correlation to the implant body length.www.indiandentalacademy.com
  • 27. When teeth oppose each other, the combined intrusive movements of the contacting elements may be 56m (28m + 28m). When a tooth opposes an implant, the combined intrusive movement is 33m (28m + 5m). When implant prostheses oppose each other, the biomechanical mismatch between teeth in the rest of the mouth and implants increase. The total combined movement may be 10m, compared with 56m in the rest of the mouth, and contrary to the teeth that move immediately, even with light loads, the implants only move this amount under a heavy occlusal load. A lighter load may generate a total implant movement of less than 3m. www.indiandentalacademy.com
  • 28. For difference in vertical movement of teeth and implants in the same arch, the existing occlusion is evaluated before implant reconstruction. Occlusal prematurities are ideally eliminated on teeth before implant reconstruction. Thin articulating paper (less than 25m thickness) is then used for the initial implant occlusal adjustment in centric relation occlusion under a light tapping force. The implant prosthesis should barely contact, and the adjacent teeth should exhibit greater initial contacts. Only axial occlusal contacts should be present on the implant crown. Once the equilibration with a light bite force is completed, a heavier centric relation occlusal force is applied. The contacts should remain axial over the implant body and may be of similar intensity on the implant crown and the adjacent teeth under greater bite force to allow all elements to react similar to the occlusal load. Hence to harmonize the occlusal forces between implants and teeth, a heavy bite force occlusal adjustment is used because it depresses the natural teeth, positioning them closer to the depressed implant position and equally sharing the load.www.indiandentalacademy.com
  • 30. If healthy anterior teeth and/or natural canines are present, the occlusion allows those teeth to distribute horizontal loads in excursions, while the posterior teeth disocclude during excursions. Anterior, compared with posterior bite force measurements and electromyographic studies provide evidence that the stomatognathic system elicits significantly less force when the posterior segments are not in contact. As a result, all lateral excursions of IPO opposing fixed prostheses or natural teeth should disocclude the posterior components. The resultant lateral forces are thus distributed only to the anterior segments of the jaws, resulting in a decrease in overall occlusal force magnitude because of diminished muscle firing and recruitment. This occlusal scheme should be followed whether or not anterior implants are in the arch. However, if anterior implants must disocclude the posterior teeth in excursion, two or more implants splinted together should help dissipate the lateral forces. www.indiandentalacademy.com
  • 31. When anterior implants and teeth are not connected - The initial lateral movement of healthy anterior teeth ranges from 68 to 108m before secondary tooth movement. Anterior implant movements are not immediate and range from 10 to 50m. Because of the greater discrepancies in lateral movement, the occlusal adjustment in this direction is more critical to implant success and survival. Light force and thin articulating paper(20m) are first used to ensure that no implant crown contact occurs during the initial occlusal or lateral movement of the teeth. A heavier force during centric occlusion and excursions is then used to develop similar occlusal contacts on both anterior implants and natural teeth. www.indiandentalacademy.com
  • 32. Unlike teeth, implants do not extrude, rotate, or migrate under occlusal forces. Natural teeth exhibit mesial drift and slight changes in occlusal position do occur over time. The proposed occlusal adjustment does not encourage additional tooth movement because regular occlusal contacts occur. The teeth opposing implants are not taken out of occlusion. Brief occlusal contacts on a daily basis maintain the tooth in its original position (similar to the rest of the mouth). In addition, because most teeth occlude with two teeth, the opposing teeth positions are even more likely to remain the same. No occlusal scheme will prevent mesial drift and minor tooth movement from occurring. An integral part of the IPO philosophy is the regular evaluation and control of occlusal contacts at each regularly scheduled hygiene appointment. This permits the correction of minor variations occurring during long-term function and also helps prevent porcelain fracture and other stress-related complications on the remainder of the natural teeth. www.indiandentalacademy.com
  • 33. For implants joined to natural teeth a similar scenario is used for the occlusal equilibration. A light force and thin articulating paper are used, and the implant crown exhibits minimum contact compared with the natural abutment crown. A gradient of force is designed on the pontics. A heavy bite force is then used to establish equal occlusal contacts for all abutments and the entire prosthesis, whether implant or natural. www.indiandentalacademy.com
  • 34. Implant Orientation and Influence of Load Direction: Forces acting on dental implants are referred to as vectors (defined in both magnitude and direction). Occlusal forces are typically three- dimensional, with components directed along one or more of the clinical coordinate axes. www.indiandentalacademy.com
  • 35. Implants are designed for a long axis load to the implant body. Stress contours were primarily concentrated at the transosteal (crestal) region. An axial load over the long axis of an implant body generates a greater proportion of compressive stress than tension or shear forces. Any load that is applied at an angle may be separated into normal (compressive and tensile) and shear forces. The greater the angle of loads to the implant long axis, the greater the compressive, tensile and shear stresses. When FEA evaluates the direction of the force changed to a more angled or horizontal load, the magnitude of the stress is increased by 3 times or more. In addition, rather than a compressive type of force primarily, greater tensile and shear forces are also demonstrated and increase more than 10 times compared with the amount found with an axial force. These stress contours resemble the pattern of early crestal bone loss on implants. www.indiandentalacademy.com
  • 37. BONE MECHANICS AND OCCLUSION www.indiandentalacademy.com
  • 38. The effect of offset or angled loads to bone is further exacerbated because of the anisotropy of cortical bone. Anisotropy refers to the character of bone, whereby its mechanical properties, including ultimate strength, depend on the direction in which the bone is loaded. Cortical bone of human long bones has been reported as strongest in compression, 30% weaker in tension, and 65% weaker in shear. Therefore IPO attempts to eliminate or reduce all shear loads to the implant to bone interface. www.indiandentalacademy.com
  • 39. A force applied at a 30-degree angle decreased the bone strength limits by 10% under compression and 25% with tension. A 60-degree force reduced the strength 30% under compression and 55% under tension. Therefore not only does the crestal bone load increase around the implant with angled forces, but the amount of stress the bone may withstand is also decreased. The greater the angle of load, the lower the ultimate strength. www.indiandentalacademy.com
  • 40. The primary component of the occlusal force should therefore be directed along the long axis of the implant body, not on an angle or following an angled abutment post. Angled abutments are used only to improve the path of insertion of the prosthesis or the final esthetic result. The angled abutment, which is loaded along the abutment axis, will transmit a significant moment load to both the implant crestal region and abutment screw, proportional to its angle of inclination. In addition, the angled implant often requires an angled abutment. Angled abutments are fabricated in two pieces and are weaker in design than a one-piece post. Furthermore, a larger transverse load component develops at the crest as a result of angled loads. An angled load to the implant long axis increases the compressive forces at the crest of the ridge on the opposite side of the implant in which the force is directed, increasing the tension component of force along the same side. The greater the angle of force to the long axis of the implant body, the greater the potentially damaging load at the crest of the bone. www.indiandentalacademy.com
  • 41. Hence the angled load increases the amount of crestal stresses around the implant body, transforms a greater percentage of the force to tensile and shear force, and reduces bone strength in compression and tension. In contrast, the surrounding implant body stress magnitude is least and the strength of bone is greatest under a load axial to the implant body. Premature occlusal contacts result in localized lateral loading of the opposing contacting crowns. Because the surface area of a premature contact is small, the magnitude of stress in the bone increases proportionately (i.e., stress=force/area). All the occlusal force is applied to one region rather than being shared by several abutments and/or teeth. In addition, the premature contact is most often on an inclined plane, therefore creating a greater horizontal component to the load and increasing compressive and tensile crestal stresses. Therefore occlusal evaluation and adjustment in partially edentulous implant patients are more important than in the natural dentition because the premature contacts can result in more damaging consequences on implants compared with teeth. www.indiandentalacademy.com
  • 43. The elimination of premature contacts is more important than in natural teeth because the implant is less mobile and often cannot effectively dissipate the forces. In addition, the teeth benefit from a greater occlusal awareness (proprioception) or oral tactile function than implants. Once the natural teeth are removed, the bone remodels to the height at or below the lowest level of the lateral cortical plates. Hence the implant crown height is often greater than the original natural anatomic crown, even in Division A bone. Crown height, with a lateral load, is a magnifier of stress to an implant to bone interface. The greater the crown height, the greater the resulting crestal moment with any lateral component of force that develops as a consequence of an angled load. Angled abutments loaded in the direction of the abutment with an increase in crown height are subject to even greater crestal moment loads because of both the lateral load and the increased lever effect from the crown height. www.indiandentalacademy.com
  • 44. In the anterior maxilla, labial concavities may require that the implant be angled away from the labial bone and the abutment toward the facial crown contour. These implant bodies are more frequently loaded at an angle, and an angled prosthetic abutment is required. As a result, larger diameter implants or a greater number of implants are indicated to minimize the crestal bone stress on each abutment. IPO aims at reducing the force of occlusal contacts, increasing implant number, and/or increasing implant diameter for implants subjected to angled loads or with an increased crown height or on the cantilever portion of a prosthesis. www.indiandentalacademy.com
  • 45. OCCLUSAL SCHEMES A primary goal of an occlusal scheme is to maintain the occlusal load that has been transferred to the implant body within the physiologic limits of each patient. These limits are not identical for all patients or restorations. The forces generated by a patient are influenced by parafunction, masticatory dynamics, tongue size, implant arch position and location, and implant arch form and crown height. The prosthodontist can best address these force factors by selecting the proper implant size, number, and position, using stress- relieving elements, increasing bone density by progressive loading and selecting the appropriate occlusal scheme. www.indiandentalacademy.com
  • 47. Hobo et al 1. Fully bone anchored bridge 2. Overdenture 3. Freestanding bridges a. Kennedy class I b. Kennedy class II c. Kennedy class III d. Kennedy class IV 4. Bridge connected to the natural teeth. 5. Single tooth replacement.www.indiandentalacademy.com
  • 48. Misch C.E et al TYPE DEFINITION FP-1 Fixed prosthesis, replaces only the crown, looks like a natural tooth FP-2 Fixed prosthesis, replaces the crown and a portion of the root, crown contour appears normal in the occlusal half is elongated or hypercontoured in the gingival half FP-3 Fixed prosthesis, replacing missing crowns and gingival colour and a portion of the edentulous site, prosthesis must often use denture teeth and acrylic gingiva, but may be porcelain to metal RP-4 Removable prosthesis, overdenture supported completely by implant RP-5 Removable prosthesis, overdenture supported by both soft tissue and implants www.indiandentalacademy.com
  • 50. When teeth are present, the maxillary dentate posterior ridge is positioned slightly more facial than its mandibular counterpart. Once the maxillary teeth are lost, the edentulous ridge resorbs in a medial direction as it evolves from Division A to B, Division B to C, and Division C to D. As a result, the maxillary permucosal implant site gradually shifts toward the midline as the ridge resorbs. As a result of ridge resorption in width the maxillary posterior implant permucosal site may even be lingual to the opposing natural mandibular tooth. The posterior mandible also resorbs lingually as the bone resorbs from Division A to B. As a consequence, endosteal implants are also more lingual than their natural tooth predecessors. www.indiandentalacademy.com
  • 51. Occlusal Table Width: A wide occlusal table favors offset contacts during mastication or parafunction. Narrower implant bodies are even more vulnerable to occlusal table width and offset loads. Wider root form implants can accept a broader range of vertical occlusal contacts while still transmitting lesser forces at the permucosal site under offset loads. Therefore in IPO the width of the occlusal table is directly related to the width of the implant body. During mastication, the amount of force used to penetrate the food bolus is also related to occlusal table width. For example, less force is required to cut a piece of meat with a sharp knife (narrow occlusal table), than with a dull knife (wider occlusal table). The greater surface area requires greater force to achieve a similar result. Hence the wider the occlusal table, the greater the force developed by the biologic system to penetrate the bolus of food. www.indiandentalacademy.com
  • 52. The posterior narrow occlusal table also facilitates daily home care. The laboratory technician often attempts to fabricate occlusal facial and lingual contours similar to that of natural teeth. This often results in ridge laps or porcelain extension at the facial gingival margin of the implant, to create an occlusal table approximately 8 to 10 mm wide. As a result, home care in the sulcular region of the implant is impaired by the overcontoured crown design. On the contrary, a narrow occlusal table combined with a reduced buccal contour (in the posterior mandible) permits easier sulcular oral hygiene in manner similar to a tooth and improves axial loading. www.indiandentalacademy.com
  • 53. The narrower occlusal contour also reduces the risk of porcelain fracture. A facial profile similar to a natural tooth on the smaller diameter implant results in cantilevered restorative materials. The facial porcelain is most often not supported by a metal substructure because the gingival region of the crown is also porcelain. As a result, shear forces result on the buccal cusp on the mandibular crown or lingual cusps in the maxillary crown, and are more likely to increase the risk of porcelain fracture. Restorations mimicking the occlusal anatomy of natural teeth often result in offset loads (increased stress), complicated home care and increased risk of porcelain fracture. As a result, in nonesthetic regions of the mouth, the occlusal table should be reduced in width compared with natural teeth. www.indiandentalacademy.com
  • 54. DIVISION OF AVAILABLE BONE:A,B,C&D (MISCH) www.indiandentalacademy.com
  • 58. Division A Bone The primary component of the occlusal force is evaluated during the treatment-planning phase. In an edentulous ridge with abundant height and width and little resorption, the implant may be placed in a more ideal position for occlusion and esthetics. Offset loads are used to describe cantilevered buccal or lingual occlusal contacts, not directed along the long axis of the implant body. When offset loads are generated at an angle, the distance between the offset contact and the long axis acts as a moment arm that magnifies the effect of the lateral force. www.indiandentalacademy.com
  • 59. The most common implant placement corresponds to a central position in the residual ridge. The implant osteotomy begins in the center of the crest and is gradually increased to the optimal width indicated in relation to the recipient bone. Facial concavities are avoided, and the thinner facial cortical bone is protected, to limit surgical complications such as labial dehiscence. As a consequence, whether in the maxilla or the mandible, the implant is frequently placed under the central fossa region of the former natural tooth. To load the implant body in an axial direction, the primary occlusal contact should therefore be the central fossa region in Division A bone. Thus for maxillary implant opposing mandibular natural teeth, the mandibular buccal cusp acts as the primary tooth contact. www.indiandentalacademy.com
  • 60. Because bone loss occurs at the expense of the facial plate, a modified buccal contour anatomy may need to be generated in Division A or B mandibles. The occlusal table width is reduced to favor an axial load on the implant in nonesthetic regions. The Division A mandibular implant is placed under the central fossa region of the natural tooth. When opposing a natural maxillary molar, the primary contacting cusp becomes the maxillary lingual cusp opposing the mandibular implant crown, with the mandibular buccal cusp of decreased height and width over the implant body. Hence all contacts are situated medially compared with those on natural teeth. The lingual contour of the mandibular implant crown is similar to the original natural dentition in position, complete with horizontal overlap to the maxillary lingual cusp to prevent tongue biting during function. There is no occlusal contact on the lingual cusp, so offset loads during parafunction are eliminated. www.indiandentalacademy.com
  • 61. The posterior maxillary crown is reduced only from the lingual aspect, compared with a natural maxillary molar, to reduce the occlusal table width. such a reduction increases the lingual overjet when the teeth are in occlusion. Narrower opposing mandibular occlusal tables are desirable to direct occlusal forces over the maxillary implant body. As a result, when opposing maxillary implants, the buccal cusps of natural mandibular teeth (or crowns on implants) should be recontoured to minimize offset loads in centric relation occlusion. The maxillary buccal cusp may then be retained for esthetics, but the functional occlusal table is reduced. When esthetics are not a concern the distal one half of the first molar and / or the entire second molar is often restored in cross bite to improve the direction of forces. In the posterior esthetic regions of the maxilla with facial bone resorption and / or lingually placed implants, a wider occlusal table is required to project the facial contours for ideal esthetics. Bone grafting to increase bone width may be required in these esthetics zones, so a larger diameter implant may be placed that permits restorations of the buccal contours with maintenance of cervical contours with emergence profiles, which permit proper hygiene of the sulcular regions. www.indiandentalacademy.com
  • 63. Posterior implants opposing each other attempt to axially load both entities. The facial cusp of the maxillary crown is required for esthetics. The other contours of the opposing crowns are reduced in width to minimize the occlusal table width and axially load the implants. WHENEVER POSSIBLE THE PORTIONS OF AN IMPLANT CROWN THAT ARE NOT SUPPORTED BY AN AXIALLY POSITIONED IMPLANT SHOULD BE RECONTOURED SO THEY DO NOT RECEIVE OCCLUSAL LOADS. ALTERNATIVELY, SEVERAL ADDITIONAL IMPLANTS SHOULD BE USED TO DISSIPATE THE FORCE. www.indiandentalacademy.com
  • 64. DIVISION B BONE Division B bone has maxillary and mandibular implants positioned under the lingual cusp when compared with the original natural tooth position. As a result, mandibular crowns require even more reduced buccal contours to avoid offset occlusal contacts. The primary contact of occlusion on an opposing natural posterior maxillary tooth is the lingual cusp, which is reshaped to axially load the implant. www.indiandentalacademy.com
  • 65. The buccal cusp of the mandibular implant crown is located near the original central fossa of the natural tooth. The medially positioned Division B mandibular implant crown may have a central fossa, but it is more lingual than the original position. The lingual contour of the crown is similar to that of the original natural tooth and has an overjet with the opposing natural tooth to prevent biting the tongue during function. The mandibular posterior implant may, on occasion, be angled medially because of the sub mandibular fossa. As a result, an angled abutment and a lingual straight emergence crown profile to minimize the lingual volume of the restoration are indicated. Augmentation of the mandibular Division B ridge is often required when stress factors are moderate to improve the implant position and prosthetic guidelines. www.indiandentalacademy.com
  • 66. A Division B maxillary implant is often placed under the palatal cusp region of the original natural tooth. The maxillary occlusal table cannot always be reduced from the facial aspect for esthetic reasons; therefore the buccal cusp is offset facially but left completely out of occlusion (as with natural teeth) in centric relation occlusion and during all mandibular excursions. The buccal cusp of the opposing natural tooth is recontoured in width and height to reduce offset loads to the opposing crown on the maxillary implant. The primary occlusal contact is centric relation occlusion is the maxillary palatal cusp over the implant body and the central fossa region of the mandibular natural tooth. Bone augmentation for larger implant width is more indicated in the maxilla because of the less dense bone and the prosthetic needs to replace an esthetic buccal crown contour. www.indiandentalacademy.com
  • 67. When Division B implants are placed in both arches, the maxillary and mandibular prostheses are similar to that described in the previous scenario. However, it is usually not possible to load both arches with an axial load, so the weakest implant in bone density, width, or prosthesis type (fixed vs. removable) determines the axial load, because it is the most vulnerable arch. www.indiandentalacademy.com
  • 68. When further resorption occurs and the ridge evolves into Division C or D, the maxillary palatal cusp becomes the primary contact area, situated directly over the implant body. Hence the occlusal contacts differ from those of a natural tooth and may even be positioned more medial than the natural palatal cusp when the implant is placed in Division C or D bone. www.indiandentalacademy.com
  • 69. Influence of Surface Area: An important parameter in IPO is the adequate surface area to sustain the load transmitted to the prosthesis. It is important to remember that mechanical stress, in its simplest form, can be defined as the force magnitude divided by the cross sectional area over which that force is applied. When implants of decreased surface area are subject to angled or increased loads, the magnified stress and strain magnitudes in the interfacial tissues can be minimized by placing an additional implant in the region of concern. Thus when narrow diameter implants are used in regions that receive greater forces, additional splinted implants are even more indicated to compensate for their narrow design and to help decrease and distribute the load over a broader region. When forces are increased in magnitude, direction or duration (e.g., parafunction), ridge augmentation maybe required to improve implant placement, reduce crown height, and increase implant width and number to compensate for the increased loads. www.indiandentalacademy.com
  • 70. Maxillary Root surface area (mm2) CENTRAL 204 LATERAL 179 CANINE 273 FIRST PREMOLAR 234 SECOND PREMOLAR 220 FIRST MOLAR 433 SECOND MOLAR 431 Mandibular Root surface area (mm2) CENTRAL 154 LATERAL 168 CANINE 268 FIRST PREMOLAR 180 SECOND PREMOLAR 207 FIRST MOLAR 431 SECOND MOLAR 426 www.indiandentalacademy.com
  • 71. Radius (mm) Length (mm) Surface Area (mm2) 1.75 10 12 14 110 132 154 2.00 10 12 14 126 151 176 2.25 10 12 14 141 169 197 2.50 10 12 14 157 188 220 2.75 10 12 14 172 207 242 3.00 10 12 14 188 226 264www.indiandentalacademy.com
  • 73. The prosthesis type may also be modified from a fixed restoration (FP-1 to FP-3) to a removable prosthesis (RP-4). This is most effective when nocturnal parafunction is present because the restorations may then be removed while sleeping. In addition, stress relieving elements may be included in the removable restoration, and additional support may be gained from the soft tissue (RP-5 restorations). Wider diameter root form implants have a greater area of bone contact at the crest than narrow implants (resulting from their increased circumferential bone contact areas). As a result, for a given occlusal load, the mechanical stress at the crest is reduced with wider implants compared with narrow ones. Natural teeth follow similar principles of diameter and surface area as just described. The anterior region of the mouth is characterized by reduced bite force compared with the posterior region. Consequently, the anterior tooth cross section is smaller, and the surface area is reduced compared with the greater diameter and surface area of posterior teeth. www.indiandentalacademy.com
  • 74. Design to the Weakest Arch: Any complex engineering structure will typically fail at its “Weakest link”, and dental implant structures are no exception. The amount of force distributed to a system can be reduced by stress relieving components that may dramatically reduce impact loads to the implant support. The soft tissue of a traditional completely removable prosthesis opposing implant prosthesis is displaced more than 2 mm and is an efficient stress reducer. Lateral loads do not result with as great a crestal load to the implants because the opposing prosthesis is not rigid. The most common implant treatment, which includes a traditional soft tissue supported complete denture, is a maxillary denture opposing a mandibular implant supported restoration. The occlusal scheme for this condition raises the posterior occlusal plane, uses a medial positioned lingualized occlusion, and has a bilateral balanced scheme. Whether the mandibular restoration is FP-1, FP-2, FP-3, RP-4, or RP-5, the maxillary denture follows these guidelines.www.indiandentalacademy.com
  • 75. The mandibular implant supported restoration may exert greater force on the premaxilla than a mandibular denture and cause accelerated bone loss. Therefore modification of the occlusal scheme aims at protecting the premaxilla under a maxillary denture by the total elimination of anterior contacts with the mandibular anterior teeth in centric occlusal relation. Reduced occlusal forces with an absence of lateral contacts in excursions are recommended on posterior cantilevers or anterior offset pontics whenever possible. This minimizes the moment forces on the abutments. It is better for mandibular cantilever pontics to oppose maxillary implants than the reverse situation. www.indiandentalacademy.com
  • 76. Full – Arch fixed prostheses (FP-1 to RP-4) Fixed prostheses on natural teeth opposing FP-1 to RP-4 implant restorations should follow mutually protected occlusal schemes whenever possible. In protrusion, there should be total absence of posterior contacts, especially for cantilevered posterior units. The masticatory force generated during lateral excursions is decreased in absence of posterior contacts. This assists in reducing the noxious effect of lateral forces on the anterior implants. Two or more implants should share any lateral force, and lateral excursions should occur as far forward as is practical and include the canine. Minimal occlusal contact in the cantilevered regions and the total absence of posterior lateral contacts during excursions are indicated when opposing the natural dentition or a fixed restorations. Seven to eight implants to support a complete implant prosthesis in two separate units are suggested in the mandible for a fixed restoration opposing a fixed prosthesis or natural teeth with inadequate to severe stress factors. www.indiandentalacademy.com
  • 77. In the edentulous maxilla, flexure of the bone is not a concern. A full arch prosthesis may be fabricated in one section. Eight to ten maxillary implants most often are required for a twelve unit fixed prosthesis opposing a fixed dentition on teeth and / or implants with moderate to severe stress factors. Posterior implants are more critical in the maxilla, in order to eliminate cantilevers and increase the anteroposterior implant distance, which further decreases stress to the maxillary anterior implants. www.indiandentalacademy.com
  • 78. Fully bone anchored bridge Mandibular edentulous case for fully bone anchored bridge •Recommended to have a mutually protected occlusion. •In centric it is necessary to have 30m clearance at the anterior region •disocclusion should be employed. •To avoid localization of stresses anterior group function should be employed. •The anterior guidance should be made slightly flatter than natural teeth to avoid over stresses on the fixtures. www.indiandentalacademy.com
  • 79. Kennedy Class I OCCLUSION FOR FREESTANDING BRIDGES o Clearance of the anterior teeth should be smaller than the natural teeth. o Amount of disocclusion required is same as natural teeth since the anterior guidance is provided by the remaining anterior natural teeth. Protrusive : 1.1mm Non working side : 1.0 mm Working Side : 0.5 mm www.indiandentalacademy.com
  • 80. Kennedy Class II •In centric the posterior osseointegrated bridge should have 30m open contacts, while anterior teeth also should have 30 m open contacts and begin to contact under strong bite pressure. •Amount of disocclusion required is same as natural teeth since the anterior guidance is provided by the remaining anterior natural teeth. Protrusive : 1.1mm Non working side : 1.0mm Working Side : 0.5 mm www.indiandentalacademy.com
  • 81. Kennedy Class III  Vertical dimension is maintained by remaining natural teeth  The osseointegrated bridge should contact only under strong pressure.  Amount of disocclusion required is same as natural teeth since the anterior guidance is provided by the remaining anterior natural teeth. Protrusive : 1.1mm Non working side : 1.0 mm Working Side : 0.5 mm www.indiandentalacademy.com
  • 82. Kennedy Class IV  To minimize horizontal loads group function occlusion is recommended.  During lateral movement posterior teeth on working side can bear the horizontal load while non working side can be discluded. Anterior guidance should be flatter than natural dentition to minimize load induced on the fixture during protrusive movement. Amount of disclussion suggested is as follows Protrusive : 0.8 mm Non- working side : 0.4 mm Working side : 0.0 mmwww.indiandentalacademy.com
  • 83. IMPLANT AND SOFT TISSUE SUPPORTED OVERDENTURE (RP-5) Anterior Tooth Position Centric stops or pressure from the tongue and muscle positions usually prevent continued extrusion of anterior natural teeth. maxillary anterior prosthetic teeth are positioned forward of the anterior supporting bone to satisfy phonetic and esthetic requirements. Moment forces result from contact with the anterior teeth, which may cause instability of the maxillary prosthesis. Therefore the maxillary denture usually does not have anterior incisal centric stops. This helps protect the premaxilla from excess forces in centric occlusion relation and initial excursions of the mandible, as the premaxilla is vulnerable to resorption from external stresses. www.indiandentalacademy.com
  • 84. Posterior Tooth Position The maxillary edentulous posterior ridge resorbs in a medial directions it transforms from Division A to B, Division B to C, and Division C to D. therefore the maxillary denture tooth gradually becomes more cantilevered off the bone support, even when positioned in the same spatial location. The mandibular edentulous posterior ridge also resorbs in a medial redirection as it transforms from Division A to B, but then resorbs laterally from Division B to C, and more lateral as it resorbs from Division C to D. In complete dentures, the position of the mandibular posterior teeth is often determined first. Bone support concepts of occlusion often position the mandibular teeth perpendicular to the edentulous ridge. This positions the central fossa of the posterior mandibular teeth more medial than that of their natural predecessors in Division B, but more facial in Division C, and very facial in Division D compared with the natural tooth placement. The maxillary teeth are then situated farther facially than the original teeth, if a normal cusp fossa relation is maintained. Consequently, maxillary denture teeth are always placed lateral to the resorbing bony support, and the condition is compounded in cases of advanced atrophy (Division C or D bone). www.indiandentalacademy.com
  • 85. The basic, concept of lingualized occlusion was first introduced by Gysi. Later Payne suggested the maxillary buccal cusps of posterior teeth should be reduced, so only the lingual cusps would be in contact. Pound discussed a similar concept, but reduced the buccal cusp of the mandible and introduced the term “lingualized” occlusion. Pound also placed the lingual cusp of the mandibular posterior teeth between lines drawn from the canine to each side of the retromolar pad. Consistent in the Philosophy of Payne and Pound, was the belief that the palatal cusp should be the only area of maxillary tooth contact. These occlusal schemes were designed to narrow the occlusal table and improve mastication, reduce forces to the underlying bone, and help stabilize a lower denture. The techniques of Payne and Pound may be modified further to a medial positioned lingualized occlusion, proposed by Misch. www.indiandentalacademy.com
  • 86. Medial Positioned Lingualized occlusion : Laboratory steps 1. Mount the upper cast using a face bow record. Mount the lower cast using the centric relation record. Set the horizontal condylar guidance according to the protrusive record. 2. Set the maxillary and mandibular anterior teeth for esthetics, phonetics, and lip support. 3. Cut back the posterior flange of the lower record base to expose the retromolar pad. Outline the retromolar pad in pencil. Draw a line from the lingual border of the pad to the mesial aspect cuspid. The central fossa of mandibular posterior teeth will be set along this line. www.indiandentalacademy.com
  • 87. 4. Using a flat plane or 10 degree mold, set the mandibular posterior teeth in a compensating curve. The curve should have both a mediolateral and anteroposterior dimension that progressively develops as the teeth are set posteriorly. The curve starts with the first premolar and becomes more accentuated in the molar region (closer to the condyle) (i.e., first premolar 0 to 5 degrees, second premolar 5 to 10 degrees, first molar 15 to 20 degrees, second molar 20 to 25 degrees). The anteroposterior angle of the curve is the second molar region and should ideally approximate the horizontal condylar guidance (i.e., 20 to 25 degrees). 5. Drop the incisal pin of the articulator 1 to 2 mm. 6. Using a 30 to 33 degree mold, set the maxillary posterior teeth with the buccal cusp tilted out facially. The lingual cusp should contact the central groove of the mandibular teeth (this will be the only tooth contact point). There should be no contact between the mandibular buccal cusp and the opposing maxillary tooth. www.indiandentalacademy.com
  • 88. 7. Return the incisal pin to the original position (up 1 to 2 mm). 8. Using articulating paper to check the contacts, grind a small fossa in the mandibular tooth for the maxillary lingual cusp tip. Continue to adjust the occlusion until the incisal pin touches the incisal table. Check for balanced occlusion in excursions. 9. Festoon the set up for the try in appointment. www.indiandentalacademy.com
  • 89. Overdenture Maxillary edentulous case for overdenture • Recommended occlusion for overdenture is fully balanced occlusion with lingualized occlusion. • Incase maxillary overdenture is opposed by a mandibular fully bone anchored bridge, in centric a small clearance is recommended in the anterior teeth, while posterior contact simultaneously. • Disocclusion is not employed here. www.indiandentalacademy.com
  • 91. The materials selected for the occlusal surface of the prosthesis affect the transmission of forces and the maintenance of occlusal contacts. In addition, occlusal material fracture is one of the most common complications for restorations on natural teeth or implants. Therefore it is wise to consider the occlusal material for each individual restoration. Occlusal materials maybe evaluated by esthetics, impact force, a static load, chewing efficiency, fracture, wear, interarch space requirements, and accuracy of castings. The three most common groups of occlusal material are fixed prostheses on implants are reviewed with relevance to the previous eight criteria. www.indiandentalacademy.com
  • 92. Esthetics: Esthetics is a major concern for patients. The most esthetic material available today is porcelain. Acrylic is acceptable for esthetics, and metal is a poor choice of materials when esthetics is the chief criterion. However, there are many situations in which esthetics is not an important aspect of the restorations. For example, when a maxillary second molar is restored, most patients do not expose this area when smiling or laughing. www.indiandentalacademy.com
  • 93. Forces: The materials on the occlusal aspect of the prostheses affect the transmission of force to the bones. Impact loads give rise to brief episodes of increases force, primarily related to the speed of closure and the dampening effect of the occlusal material. The hardness of material is related to its ability to absorb stress from impact loads. All porcelain occlusal surface exhibits a hardness 2.5 times greater than that of natural teeth. Acrylic resin has a Knoop hardness of 17 kg/mm2, and enamel has a 350 kg/mm2 hardness. A composite resin may exhibit a hardness of 85% that of enamel. Therefore impact loads are lowest with acrylic, increase with composite and metal occlusals, increase even more with enamel, and further increase with porcelain. As a consequence, it has been suggested to use resin because of its dampening characteristics. Clenching patients do not have a considerable amount of stress reduction when acrylic versus porcelain materials are used on the occlusal surfaces. Progressive bone loading is performed with acrylic transitional prostheses. This material may reduce the impact force on the early implant bone interface. As the bone matures and its density increases, the need for force reduction decreases. www.indiandentalacademy.com
  • 94. Chewing efficiency: Fixed prostheses exhibit an improve efficiency compared with removable soft tissue borne prostheses, regardless of the occlusal material. Acrylic was 30% less efficient than porcelain or metal, whereas there was no difference between gold and porcelain. www.indiandentalacademy.com
  • 95. Wear: The definition of wear is the deterioration, change or loss of a surface caused by use. The factors affecting the amount of wear include magnitude, angle, duration, speed, hardness, and surface finish of the opposing force and surface, together with the lubricant, temperature, and chemical natural of the surrounding environment. Most occlusal wear occurs as a result of bruxism. An intuitive feeling is the harder the occlusal material, the less the wear. However, surface hardness has been shown to be a poor indicator of wear rate. Acrylic resin wears 7 to 30 times faster when opposing gold, resin, enamel, or polished porcelain, compared with gold opposing gold, enamel or porcelain. Gold occlusal surface exhibit less volume loss (sum of loss of opposing occlusal surfaces) than any other combination of materials. Porcelain opposing porcelain wears more than porcelain opposing gold or metal. www.indiandentalacademy.com
  • 96. The wear rate of occlusal materials, especially in the partially edentulous patient with unrestored teeth, should be similar to enamel. In this way, occlusal changes will not dramatically change the occlusal scheme. Lambrechts et al in an in vivo investigation reported vertical wear of premolar and molar tooth enamel to be 20 to 40 um per year when opposing the enamel of natural teeth. In principle, for the partially edentulous patient it would be better to have more occlusal wear on implants, rather than less, because the additional forces on the teeth are better tolerated than on implant prostheses. As a result, total volume wear may favor porcelain opposing enamel for the implant prosthesis opposing teeth in the partially edentulous patient and metal opposing enamel in the other regions of the mouth that require restorations of natural teeth. The use of gold, regardless of the opposing combination, always provides the least total volume loss. www.indiandentalacademy.com
  • 97. Adhesive wear occurs when one hard surface slides over a surface of lesser hardness. As a result, wear fragments of one material adhere to the other material. Gold occlusal surfaces are observed to have gold particles adhered to enamel. This may account for less total volume loss when opposing other materials. For full arch implant supported prostheses, the restoring doctor may consider metal occlusal to minimize wear and prolong the accuracy of occlusal scheme long term. Porcelain in esthetic regions opposing gold in the more nonesthetic area or metal occlusal in both arches when parafunction or marginal interarch space is present are the material most often selected as implant occlusal materials. www.indiandentalacademy.com
  • 98. Materials fracture: Materials fracture is one of the more common factors that lead to refabrication of a prosthesis. Porcelain, acrylic and composite fractures occur under excessive loads or even with a lesser load of longer duration, angulation, or frequency. Acrylic or composite materials fracture more easily. The compressive strength of acrylic resin is 11,000 psi, compared with 40,000 psi for enamel. Composite resin is 3 times stronger than acrylic. Porcelain opposing porcelain is not suggested with extreme parafunction, because it may fracture more often than porcelain opposing metal. Metal occlusals do not easily fracture, provide good wear resistance, and have minimum impact load compared with porcelain. www.indiandentalacademy.com
  • 99. Accuracy: Metal shrinkage is 10 times less than porcelain or acrylic and therefore permits the fabrication of a more passive casting. When accuracy of the casting is paramount, as with screw retained restorations, the occlusal material may make a significant difference. This is most important in regions of long spans and / or with a large volume of materials. Interarch space: Acrylic restorations receive their strength from bulk and therefore require greater interarch space. Metal occlusals require the least amount of space. In addition, when increased retention of a cement retained prosthesis is required, a high abutment and greater retention may be accomplished with a metal occlusal. Porcelain is intermediate in the interarch space requirement.www.indiandentalacademy.com
  • 100. Therefore when all seven criteria are evaluated, metal is an excellent occlusal material, with improved properties in accuracy, wear, fracture resistance, abutment retention, and good qualities for impact or static force. Esthetics is best satisfied with porcelain, which has improved properties compared with acrylic concerning fractures and retention. www.indiandentalacademy.com
  • 101. REVIEW OF LITERATURE J.B. Brunskin and J.A. Hipp in 1984 studied the in vivo forces on dental implants. Methods are presented for measuring vertical force components or bridged titanium dental implants in dog mandibles. These methods have included custom made strain gauge transducers, plus hard wiring and telemetric schemes for data collection. The essential components of the measurements system are described, and typical bite force data are illustrated www.indiandentalacademy.com
  • 102. Rangert et al in 1989 carried out a study on the forces and moments on Brenamark implants. The placement of fixture (implants) in relation to the geometry of a prosthetic restoration has a great influence on the mechanical loading of the implant. Based on Theoretic consideration and clinical experiences with the Brenamark system, this article gives simples guidelines for controlling these loads. The emphasis is on design rules that can be used in clinical practice. With the class I lever as a reference. Various clinical implant prosthesis situations are discussed and evaluated. www.indiandentalacademy.com
  • 103. Parker et al in 1991 reviewed the occlusal considerations in restorative dentistry. The major topics include the assessment and treatment of occlusal wear, the controversies surrounding treatment position of the mandibular condyles, occlusal considerations in osseointegrated prosthesis, the two way relationship between occlusal factors and temporomandibular disorders, design criteria and longevity studies in resin bonded, fixed partial denture. Hobo et al in 1991 presented a case report on occlusion for osseointegrated prosthesis and concluded that the concept of occlusion suitable for osseointegrated prosthesis is basically the same as the gnathological occlusion. However the natural tooth sinks about 30µm during function, while an osseointegrated bridge which is supporte only by the bone does not sink. Therefore it is necessary to adjust the centric contacts of the osseointegrated fixed bridge slightly more open than the natural teeth. During the eccentric movement, in order to minimize horizontal loading, the concept of disocclusion is generally used. www.indiandentalacademy.com
  • 104. James et al in 1993 discussed the edentulous implants an emphasized that the occlusal contacts of the final fixed restoration are affected significantly by implant position. Lateral occlusal forces, may lead to abutment screw fracture. They may be due to either excessive lateral occlusal pressure or a malposed implant that requires non axial loading during normal function. www.indiandentalacademy.com
  • 105. C.E. Misch et al in 1994 discussed an implant protected occlusal on a biomechanical rationale. The clinical success and longevity of endosteal dental implants are controlled, in a large part, by the mechanical milieu within which they function. The occlusion is a critical component of such a mechanical environment. “Implant protected occlusion” refer to an occlusal scheme that is often uniquely specific to the restoration of endosteal implant prosthesis. Implant orientation and the influence of load direction, the surface area of implants, occlusal table width, and protecting the weakest area are blended together from a biomechanical rationale to provide support for a specific occlusal philosophy. www.indiandentalacademy.com
  • 106. Tashkandi et al in 1996 did an analysis of strain at selected bone sites of a cantilevered implant supported prosthesis. the results revealed that the maximum strain occurred at the strain gauge positioned on cortical bone over the apex of the most distal implant under 10 and 20 lb loading conditions. Osamu et al in 2002 did a study on influence of supra structure materials on strain around an implant under two loading conditions. The results showed under static and non impact dynamic loading the three super structure materials tested (highly filled composite resin, acrylic resin and gold alloy), had the same influence on the strain transmitted to the bone simulant that surrounded a single implant. www.indiandentalacademy.com
  • 107. Kent et al in 2004 did a photoelectric analysis of the effect of palatal support on various implant supported overdenture designs and concluded that at the removal of the palatal support produces a greater effect and more concentrated stress difference for maxillary overdenture than difference between the attachment designs. With palatal coverage Without palatal coverage www.indiandentalacademy.com
  • 108. Steven et al in 2004 did a study of stress transfer of four mandibular implant overdenture cantilevered designs. His results condluded that under load all prosthetic designs demonstrated a low stress transfer to the ipsilateral abutment and to the contralateral side of the arch. The plunger retained prosthesis retained by two implants demonstrated a more uniform stress trasnfer to the ipsilateral terminal abutment than the clip retained prosthesis retained by three implants and provided more retention. www.indiandentalacademy.com
  • 109. Lucie et al in 2004 Did a finite element analysis on the influence of implant length and diameter on stress distribution. Results showed an increase in the implant diameter decreased the maximum von Mises equivalent stress around the implant neck more than an increase in the implant length, as a result of a more favorable distribution of the simulated masticatory forces applied in this study. www.indiandentalacademy.com
  • 110. Osseointegrated supported prosthesis (ISP) have shown high standard of success. This success rate depends not only on meticulous surgical protocol but also on understanding concept of occlusion. Occlusion should be a key factor to overall success rate. The concept of occlusion suitable for implant supported prosthesis is basically the same as gnathological occlusion. In Centric contact of the Osseo integrated crown or fixed prosthesis should be slightly more open than natural teeth. In centric Osseo integrated crown or fixed prosthesis should not contact with opposing teeth under the soft bite pressure (to avoid the occlusal load on the I.S.P. which leads overload the bone structure) Under the strong bite pressure Osseo integrated supported prosthesis should contact after the natural tooth intrudes approximately 30µm. SUMMARY www.indiandentalacademy.com
  • 111. • To avoid overloading of the occlusal surface, the I.S.P. should not have plane-to-plane contact. • Point contact especially cusp-to-fossa tripodal contact is preferred. • During eccentric movement, in order to minimize horizontal movement, the concept of disocclusion is generally recommended. • Anterior segments of the osseointegrated prosthesis should guide the mandible to produce posterior disocclusion. • Canine guided occlusion is not recommend for the Osseo integrated prosthesis to avoid excessive occlusal forces into the single implant fixture which is placed in the canine area. Group function is recommended to distribute the stress over the entire fixture. • The ideal place to bear the horizontal load is the trapezoid area, which is surrounded by the osseointegrated fixtures. • Load transmitted to the fixture is not so destructive when extended mesially in the anterior region, whereas more destructive when extended distally. www.indiandentalacademy.com
  • 112. CONCLUSION The local occlusal considerations in implant dentistry include the transosteal forces, bone biomechanics, basic biomechanics, differences in natural teeth and implants, muscles of mastication and occlusal force, and bone resorption. The incorporation of all these factors lead to an occlusal scheme (IPO) discussed in this seminar. Occlusal schemes consider the weakest component, full or partial edentulous arches, and posterior or anterior teeth and / or implants. An IPO is a consistent approach for implant occlusal schemes. The material from which the occlusal region are fabricated may affect implant loading and also affect implant reaction forces to the opposing arch. These occlusal material also affect wear and fracture, which affects the occlusal contacts, vertical occlusal dimension, and esthetics.www.indiandentalacademy.com
  • 113. References 1. Misch CE : Dentistry of bone and effect on treatment plans, surgical approach, healing and progressive loading. Int. J. Oral implantol 6:23-31, 1990. 2. Misch CE : Progressive bone loading, Pract Periodontics Aesthet Dent. 2:27-30, 1990. 3. Misch CE : Progressive bone loading. In Misch CE, editor; Contemporary implant dentistry, pp 623-650, St. Louis, 1993, Mosby. 4. Bidez MW, Misch CE : Force transfer in implant dentistry; basic concepts and principles, Oral implant 18: 264-274, 1992. 5. Misch CE , Bidaz MW : Implant protected occlusion, a biomechanical rationale, Comp Cont. Dent Educ. 15(11):1330-1343, 1994. 6. Misch CE : early crestal bone loss etiology and its effect on treatment planning for implants, post Grad Dent. (2)3:3-17, 1995. 7. Jemt T, Linden B, Lekholm U : failures and complications in127 consecutively placed fixed partial prostheses supported by Branemark implants ; from prosthetic treatments to first annual check up , Int. J. Oral Maxillofac Impl. 7;40-44, 1992. 8. Isidor F : Loss of osteointegration caused by occlusal load of oral implants, Clin Oral Implant Res.7:143-152, 1996. www.indiandentalacademy.com
  • 114. 9. Misch CE : Medial positioned lingualized occlusion, Misch Institute Manual, Birmingham, Mich, 1991. 10. Misch CE : Occlusal considerations for implant supported prostheses. In Misch CE, editior: contemporary implant dentistry, pp 705-733, St. Louis 1993, Mosby. 11. Muhlemann HR : Tooth mobility : review of clinical aspects and research findings, J. Periodontal 38:686 1967. 12. Van Steenbergh D : A retrospective multicenter evaluation of the survival rate of fixed prosthesis on four or six implants and modum Branemark in full edentulism. J. Prosthet Dent. 61:217-223, 1989. 13. Chee WWL, Cho GC : A rationale for not connecting implants to natural teeth, J. Prosthod.6(1) :7-10, 1997. 14. Goldstein GR : the relationship of canine protected occlusion to a periodontal index, J. Prosthet. Dent. 41:277-283, 1979. 15. Williamson EH, Lundquist DO: Anterior guidance its effect on electromyographic activity of the temporal and masseter muscles, J. Prosthet Dent. 49:816-823, 1983. 16. Schupe RJ et al : effects of occlusal guidance on jaw muscles activity, J Prosthet Dent 51:811-818, 1984.www.indiandentalacademy.com
  • 115. 17. Manns A, Chan C, Miralles R : influences of group function and canine guidance on electromyographic activity of elevator muscles, J. Prosthet Dent 57:494-501, 1987. 18. Ko CC DH, Hollister SJ : Micromechanics of implants /tissue interfaces, J oral implantol 18: 220, 1992. 19. Misch CE: Three dimensional finite element analysis of two plate form neck designs, Master’s thesis, University of Pittsburgh, 1989. 20. Clelland NL, Lee JK, Bimbenet OC et al : A three dimensional finite element stress analysis of angled abutments for an implant placed in the anterior maxilla, J. Prosthodont 4(2):95-100, 1995. 21. Papavasillou G, Kamposiora P et al : Three dimensional finite element analysis of stress distribution around single tooth implants as a function of bony support prosthesis type and loading during function, J. Prosthet Dent 76: 633-640, 1996. 22. Reilly DT, Burstein AH : The elastic and ultimate properties of compact bone tissue, J Biomech 80:393-405, 1975. 23. Cowin SC : Bone mechanics, Boca Raton , Fla 1989, CRC Press. 24. Parein AM, Eckert SE, Wollan PC et al : Implant reconstruction in the posterior mandible : a long term retrospective study. J. Prosthet Dent 78:34-42, 1997. www.indiandentalacademy.com
  • 116. 25. De Marco TL, Paine S : Mandibular flexure in opening and closure movements, J Prosthet Dent. 31:482-485, 1974. 26. Chibirka RM, Razzoog ME, Lang BR et al : determining the force absorption, quotient for restorative materials used in implant occlusal surfaces, J. Prosthet Dent 67 (3): 361-364, 1992. 27. Naert I, Quirynen M, Van Steenberghe D et al : A six year prosthodontic study of 509 consecutively inserted implants for the treatment of partial edentulism. J. Prosthet Dent 67:236-245, 1992. 28. Shultz AW : comfort and chewing efficiency in dentures, J. Prosthet Dent. 65:38-48, 1951. 29. Okesm JP : Management of Temporomandibular disorders and occlusion, pp 259-260, St. Louis, 1989, Mosby. 30. Hudson JD, Goldstein GR, Georgescur M. Enamel wear caused by three different restorative material, J. Prosthet 74:647-654, 1995. 31. Monasky GE, Tough DF: Studies of wear of porcelain, enamel and gold, J Proshet Dent 25(3):299-306, 1971. 32. Krejci I,Lutz F,Reimer M et al: Wear of ceramic inlays, their enamel antagonist and luting cements, J. Prosthet Dent 69: 425-431, 1993. 33. Seghi RR, Daher, Caputo A : Relative flexural strength of dental restorative ceramics, Dent. Mater 6:181-184, 1990. www.indiandentalacademy.com