RSNA 2012
Clinical Cases Ingenia 3.0T
Ingenia 3.0T
             SmartBreast with the dS Breast 16ch coil

Background Information
Hospital:
•	Wellstar Health System, Marietta, USA
                                                                                                                              SmartBreast – dStream Breast 16ch coil
•	Group of five hospitals and seven imaging centers
                                                                                                                              Axial T2w 3D TSE (VISTA) and T1w 3D TFE, Resolution 0.8 x 0.8 x 2.0 mm (VISTA) and 0.6 x 0.6 x 2.0 mm (T1w),
•	Philips Achieva 1.5T & 3.0T, Panorama
                                                                                                                              Scan time 1:30 min (VISTA) and 1:22 min (T1w), Courtesy: Wellstar Health System, Marietta, USA
  HFO and Ingenia 3.0T
•	1000 Breast MR exams a year, 130 Breast MR biopsies
•	Referred primarily from breast surgeons in the area
•	New Women’s Center opened in fall 2011


Patient:
•	57 year old female with a new diagnosis of Ca.
•	Sample obtained by US biopsy:
  “Invasive Ductal Ca, grade 3”
•	MR requested to determine treatment


Case Highlights
•	70 cm bore to support patient comfort
•	dStream Breast 16ch coil designed for
  high spatial and temporal resolution with
  coverage deep into the axillae region
•	SmartBreast for consistent fat suppression due
  to unique patient-adaptive shimming method
•	MultiTransmit  adapts the RF transmission signals                                                                           SmartBreast – dStream Breast 16ch coil
  to the size and shape  of each individual patient                                                                           Axial T2w TSE with SPAIR and Diffusion (b800), Resolution 1.0 x 1.15 x 2.0 mm (T2w) and 2.2 x 2.2 x 1.8 mm (DWI),
                                                                                                                              Scan time 6:52 min (T2w) and 4:13 min (DWI), Courtesy: Wellstar Health System, Marietta, USA
Description
SmartBreast allows for consistent fat suppression and image
uniformity due to unique patient-adaptive shimming methods




                                                              SmartBreast – dStream Breast 16ch coil
                                                              Dynamic axial eTHRIVE
                                                              Resolution 0.8 x 0.8 x 2.0 mm. Scan time 1:22 min per dynamic   SmartBreast – dStream Breast 16ch coil
                                                              Courtesy: Wellstar Health System, Marietta, USA                 Dynamic axial eTHRIVE, Resolution 0.8 x 0.8 x 2.0 mm. Scan time 1:22 min per dynamic, Courtesy: Wellstar Health System, Marietta, USA


2    Clinical Cases Ingenia 3.0T                                                                                                                                                                                                                         Clinical Cases Ingenia 3.0T   3
High Resolution Knee imaging with Cartilage Assessment
                Ingenia 3.0T
                                                                                    Background Information
                                                                                    Hospital:                                                                        High Resolution Imaging
                                                                                    •	Palo Alto Medical Foundation, California, USA                                  – dStream Knee 8ch coil
                                                                                    •	Pioneer in the multispecialty group practice of medicine                       Coronal T1w TSE and
                                                                                    •	Provides Healthcare services in 4 California counties                          Sagittal PDw aTSE
                                                                                      including Alameda, San Mateo, Santa Clara and Santa Cruz                       Resolution 0.26 x 0.39 x
                                                                                    •	Sports Medicine is one of the hospitals specializations                        2.0 mm (T1w) and 0.19 x
                                                                                    •	Average time planned for patients is 45 minutes                                0.30 x 2.0 mm (PDw)
                                                                                                                                                                     Scan time 4:23 min (T1w)
                                                                                    Patient:                                                                         and 5:24 min (PDw)
                                                                                    •	Middle aged board surfer presents with chronic knee pain                       Courtesy: Palo Alto Medical
                                                                                    •	X-Ray shows no abnormalities                                                   Foundation, California, USA
                                                                                    •	Follow-up MR examination requested


                                                                                    Case Highlights
                                                                                    •	dStream Knee 8ch coil is designed for high resolution
                                                                                      visualization of fine cartilage structures and small ligaments
                                                                                                                                                                    Comprehensive Knee Imaging
                                                                                    •	Coil conforms to the anatomy for optimal signal-to-noise ratio
                                                                                                                                                                    – dStream Knee 8ch coil
Description                                                                         •	Coil can be slightly rotated relative to its base plate
                                                                                                                                                                    Axial Dual IR sequence
dStream Knee 8ch coil designed for high resolution                                    to ease coil setup and enhance patient comfort
                                                                                                                                                                    Resolution 0.5 x 0.5 x 3.0 mm
visualization of fine cartilage structures and small ligaments                      •	Split coil design and single handle for easy coil setup
                                                                                                                                                                    Scan time 4:28 min
                                                                                                                                                                    Courtesy: Palo Alto Medical
                                                                                                                                                                    Foundation, California, USA




                                                                                                                                                                    Comprehensive Knee Imaging
                                                                                                                                                                    – dStream Knee 8ch coil
                                                                                                                                                                    Axial T2 calculation sequence
                                                                                                                                                                    (multi-echo SE)
                                                                                                                                                                    Resolution 0.4 x 0.5 x 2.5 mm
                                                                                                                                                                    Scan time 5:24 min
                                                                                                                                                                    Courtesy: Palo Alto Medical
                                                                                                                                                                    Foundation, California, USA

    Homogeneous fat suppression – dStream Knee 8ch coil
    Axial, Coronal and Sagittal PDw aTSE with SPAIR fat suppression, Resolution 0.24 x 0.34 x 2.0 mm, Scan time 4:44 min (axial and sagittal), 5:15 min (coronal)
    Courtesy: Palo Alto Medical Foundation, California, USA



4       Clinical Cases Ingenia 3.0T                                                                                                                                                                 Clinical Cases Ingenia 3.0T   5
Lumbar Spine in obese patient
             Ingenia 3.0T


Background Information                                                                                                                 Lumbar Spine - dStream
Hospital:                                                                                                                              Posterior FlexCoverage coil
•	Krankenhaus der Barmherzigen Brüder, Trier Germany                                                                                   Sagittal T2w TSE and T1w
•	One of the largest hospitals within its region                                                                                       TSE pre and post Gd
  with 612 beds and 15 clinical specialties                                                                                            Resolution 0.8 x 1.0 x
                                                                                                                                       4.0 mm. Scan time 3:23
Patient:
                                                                                                                                       (T2w) and 2:19 (T1w)
•	32- year old male patient, 180 kg, with back pain
                                                                                                                                       Courtesy: Krankenhaus
•	Was first examined on CT but quality
                                                                                                                                       der Barmherzigen Brüder,
  was not sufficient for diagnosis
                                                                                                                                       Trier, Germany
•	MRI gave excellent images. Diagnosis: small
  protrusion of the disc L5/S1, no meningioma


Case Highlights
•	70 cm bore for increased patient comfort and easy positioning
•	Homogeneous signal intensity over the entire field of view
•	Integrated dStream Posterior FlexCoverage coil reduces patient set
  up time and provides high SNR and large signal penetration depth
•	MultiTransmit adapts the RF signals to each patient
  leading to superb image uniformity and faster imaging


Description:
Integrated dStream Posterior FlexCoverage coil reduces patient set                                                                     Lumbar Spine - dStream
up time and provides high SNR and large signal penetration depth                                                                       Posterior FlexCoverage coil
                                                                                                                                       Coronal STIR, radial M2D
                                                                                                                                       myelo, axial T2w and T1w TSE
                                                                                                                                       Resolution 0.9 x 1.3 x 4.0
                                                                                                                                       mm (STIR), 0.5 x 1.6 x
                                                                                                                                       40.0 mm (myelo), 0.7 x
                                                                                                                                       0.8 x 4.0 mm (axials).
                                                                                                                                       Scan time 3:23 (STIR), 2:40
                                                                                                                                       (myelo), 3:21 (axial T2w TSE)
                                                                                                                                       and 5:09 (axial T1w TSE)
                                                                                                                                       Courtesy: Krankenhaus
                                                                                                                                       der Barmherzigen Brüder,
                                                                                                                                       Trier, Germany
                                                                       Lumbar Spine - dStream Posterior FlexCoverage coil
                                                                       Comparison between CT image and MR images


                                                                       Courtesy: Krankenhaus der Barmherzigen Brüder, Trier, Germany




6    Clinical Cases Ingenia 3.0T                                                                                                                Clinical Cases Ingenia 3.0T   7
Imaging 2.0 – Transforming care,                                                                                    Philips Healthcare is partof
  together.                                                                                                           Royal Philips Electronics


  At Philips, we recognize that radiology is the                                                                      How to reach us
  cornerstone of diagnosis and treatment.                                                                             www.philips.com/healthcare
                                                                                                                      healthcare@philips.com
  And that clinical integration and
  collaboration are key to more personal care,                                                                        Asia
  better patient outcomes, and lower costs.                                                                           +49 7031 463 2254


  Imaging 2.0 signals our endless quest to raise                                                                      Europe, Middle East, Africa
  the bar on clinical excellence – by working                                                                         +49 7031 463 2254
  with you to develop innovations that
  enable you to collaborate freely, diagnose                                                                          Latin America
  confidently, and care passionately.                                                                                 +55 11 2125 0744


  Together, we will continue to spark a                                                                               North America
  revolution in imaging science with advanced                                                                         +1 425 487 7000
  technologies designed to deliver greater                                                                            800 285 5585 (toll free, US only)
  collaboration and integration, increased
  patient focus, and improved economic value.


  Learn more at www.philips.com/imaging2.0




Please visit www.philips.com/mri


                       © 2012 Koninklijke Philips Electronics N.V.
                       All rights are reserved.

                       Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or	
                       obligation and will not be liable for any consequences resulting from the use of this publication.

                       Printed in The Netherlands.
                       NOV 2012

Contenu connexe

En vedette

Uso de las radiaciones ionizantes en medicina
Uso de las radiaciones ionizantes en medicinaUso de las radiaciones ionizantes en medicina
Uso de las radiaciones ionizantes en medicinaMiguel Martín Landrove
 
Presentación2 rmn 2012
Presentación2 rmn 2012Presentación2 rmn 2012
Presentación2 rmn 2012Sonia M Lemus
 
Espectroscopia de resonancia magnética nuclear de sólidos aplicada a la carac...
Espectroscopia de resonancia magnética nuclear de sólidos aplicada a la carac...Espectroscopia de resonancia magnética nuclear de sólidos aplicada a la carac...
Espectroscopia de resonancia magnética nuclear de sólidos aplicada a la carac...Neffer Yánez Vanegas
 
Técnica de la imagen por resonancia magnética
Técnica de la imagen por resonancia magnéticaTécnica de la imagen por resonancia magnética
Técnica de la imagen por resonancia magnéticaT.E.R.T Samuel Ramirez
 
Introduccion a la resonancia magnetica musculo esqueletica
Introduccion a la resonancia magnetica musculo esqueleticaIntroduccion a la resonancia magnetica musculo esqueletica
Introduccion a la resonancia magnetica musculo esqueleticaYuri pe?
 
Resonancia nuclear magnetica 885
Resonancia nuclear magnetica 885Resonancia nuclear magnetica 885
Resonancia nuclear magnetica 885Facundo Gaston Meza
 
Clase 16 Resonancia Nuclear MagnéTica
Clase 16 Resonancia Nuclear MagnéTicaClase 16 Resonancia Nuclear MagnéTica
Clase 16 Resonancia Nuclear MagnéTicaDR. CARLOS Azañero
 
6.5 Tipos mas comunes de antenas
6.5 Tipos mas comunes de antenas6.5 Tipos mas comunes de antenas
6.5 Tipos mas comunes de antenasEdison Coimbra G.
 
Componentes de un equipo de resonancia magnética
Componentes de un equipo de resonancia magnéticaComponentes de un equipo de resonancia magnética
Componentes de un equipo de resonancia magnéticamdpmadpmadp
 

En vedette (17)

Magnetom flash 45
Magnetom flash 45Magnetom flash 45
Magnetom flash 45
 
Uso de las radiaciones ionizantes en medicina
Uso de las radiaciones ionizantes en medicinaUso de las radiaciones ionizantes en medicina
Uso de las radiaciones ionizantes en medicina
 
Irm
IrmIrm
Irm
 
SAR
SARSAR
SAR
 
Presentación2 rmn 2012
Presentación2 rmn 2012Presentación2 rmn 2012
Presentación2 rmn 2012
 
Espectroscopia de resonancia magnética nuclear de sólidos aplicada a la carac...
Espectroscopia de resonancia magnética nuclear de sólidos aplicada a la carac...Espectroscopia de resonancia magnética nuclear de sólidos aplicada a la carac...
Espectroscopia de resonancia magnética nuclear de sólidos aplicada a la carac...
 
Técnica de la imagen por resonancia magnética
Técnica de la imagen por resonancia magnéticaTécnica de la imagen por resonancia magnética
Técnica de la imagen por resonancia magnética
 
hardware en resonacia magnetica UAP
hardware en resonacia magnetica UAPhardware en resonacia magnetica UAP
hardware en resonacia magnetica UAP
 
Im rmn-2012 13f
Im rmn-2012 13fIm rmn-2012 13f
Im rmn-2012 13f
 
Introduccion a la resonancia magnetica musculo esqueletica
Introduccion a la resonancia magnetica musculo esqueleticaIntroduccion a la resonancia magnetica musculo esqueletica
Introduccion a la resonancia magnetica musculo esqueletica
 
Resonancia nuclear magnetica 885
Resonancia nuclear magnetica 885Resonancia nuclear magnetica 885
Resonancia nuclear magnetica 885
 
CT Artifacts
CT ArtifactsCT Artifacts
CT Artifacts
 
Resonancia magnetica
Resonancia magneticaResonancia magnetica
Resonancia magnetica
 
Clase 16 Resonancia Nuclear MagnéTica
Clase 16 Resonancia Nuclear MagnéTicaClase 16 Resonancia Nuclear MagnéTica
Clase 16 Resonancia Nuclear MagnéTica
 
6.5 Tipos mas comunes de antenas
6.5 Tipos mas comunes de antenas6.5 Tipos mas comunes de antenas
6.5 Tipos mas comunes de antenas
 
Componentes de un equipo de resonancia magnética
Componentes de un equipo de resonancia magnéticaComponentes de un equipo de resonancia magnética
Componentes de un equipo de resonancia magnética
 
Transformadores
TransformadoresTransformadores
Transformadores
 

Plus de Jhon Arriaga Cordova (20)

Resonancia no ecoplanar en colesteatoma.pptx
Resonancia no ecoplanar en colesteatoma.pptxResonancia no ecoplanar en colesteatoma.pptx
Resonancia no ecoplanar en colesteatoma.pptx
 
Oído.pptx
Oído.pptxOído.pptx
Oído.pptx
 
Toshiba Visions 23
Toshiba Visions 23Toshiba Visions 23
Toshiba Visions 23
 
FieldStrength 49
FieldStrength 49FieldStrength 49
FieldStrength 49
 
FieldStrength 50 advanced neuro
FieldStrength 50 advanced neuroFieldStrength 50 advanced neuro
FieldStrength 50 advanced neuro
 
Ct somatom sessions-33
Ct somatom sessions-33Ct somatom sessions-33
Ct somatom sessions-33
 
Somatom sessions 32
Somatom sessions 32Somatom sessions 32
Somatom sessions 32
 
Somatom sessions 31
Somatom sessions 31Somatom sessions 31
Somatom sessions 31
 
Somatom sessions 30
Somatom sessions 30Somatom sessions 30
Somatom sessions 30
 
Somatom sessions 26
Somatom sessions 26Somatom sessions 26
Somatom sessions 26
 
Somatom sessions 25
Somatom sessions 25Somatom sessions 25
Somatom sessions 25
 
Somatom sessions 24
Somatom sessions 24Somatom sessions 24
Somatom sessions 24
 
Somatom session 29
Somatom session 29Somatom session 29
Somatom session 29
 
Somatom session 29
Somatom session 29Somatom session 29
Somatom session 29
 
Somatom session 28
Somatom session 28Somatom session 28
Somatom session 28
 
Somatom session 27
Somatom session 27Somatom session 27
Somatom session 27
 
Somatom session 27
Somatom session 27Somatom session 27
Somatom session 27
 
Neurography Supplement to Issue 50
Neurography Supplement to Issue 50Neurography Supplement to Issue 50
Neurography Supplement to Issue 50
 
ISMRM Edition - Issue 49
ISMRM Edition - Issue 49ISMRM Edition - Issue 49
ISMRM Edition - Issue 49
 
Pediatric Imaging - Issue 47
Pediatric Imaging - Issue 47Pediatric Imaging - Issue 47
Pediatric Imaging - Issue 47
 

Mr ingenia clinical_cases_brochure_rsna_2012_v_final_7__30_oct_12 - copia

  • 2. Ingenia 3.0T SmartBreast with the dS Breast 16ch coil Background Information Hospital: • Wellstar Health System, Marietta, USA SmartBreast – dStream Breast 16ch coil • Group of five hospitals and seven imaging centers Axial T2w 3D TSE (VISTA) and T1w 3D TFE, Resolution 0.8 x 0.8 x 2.0 mm (VISTA) and 0.6 x 0.6 x 2.0 mm (T1w), • Philips Achieva 1.5T & 3.0T, Panorama Scan time 1:30 min (VISTA) and 1:22 min (T1w), Courtesy: Wellstar Health System, Marietta, USA HFO and Ingenia 3.0T • 1000 Breast MR exams a year, 130 Breast MR biopsies • Referred primarily from breast surgeons in the area • New Women’s Center opened in fall 2011 Patient: • 57 year old female with a new diagnosis of Ca. • Sample obtained by US biopsy: “Invasive Ductal Ca, grade 3” • MR requested to determine treatment Case Highlights • 70 cm bore to support patient comfort • dStream Breast 16ch coil designed for high spatial and temporal resolution with coverage deep into the axillae region • SmartBreast for consistent fat suppression due to unique patient-adaptive shimming method • MultiTransmit adapts the RF transmission signals SmartBreast – dStream Breast 16ch coil to the size and shape of each individual patient Axial T2w TSE with SPAIR and Diffusion (b800), Resolution 1.0 x 1.15 x 2.0 mm (T2w) and 2.2 x 2.2 x 1.8 mm (DWI), Scan time 6:52 min (T2w) and 4:13 min (DWI), Courtesy: Wellstar Health System, Marietta, USA Description SmartBreast allows for consistent fat suppression and image uniformity due to unique patient-adaptive shimming methods SmartBreast – dStream Breast 16ch coil Dynamic axial eTHRIVE Resolution 0.8 x 0.8 x 2.0 mm. Scan time 1:22 min per dynamic SmartBreast – dStream Breast 16ch coil Courtesy: Wellstar Health System, Marietta, USA Dynamic axial eTHRIVE, Resolution 0.8 x 0.8 x 2.0 mm. Scan time 1:22 min per dynamic, Courtesy: Wellstar Health System, Marietta, USA 2 Clinical Cases Ingenia 3.0T Clinical Cases Ingenia 3.0T 3
  • 3. High Resolution Knee imaging with Cartilage Assessment Ingenia 3.0T Background Information Hospital: High Resolution Imaging • Palo Alto Medical Foundation, California, USA – dStream Knee 8ch coil • Pioneer in the multispecialty group practice of medicine Coronal T1w TSE and • Provides Healthcare services in 4 California counties Sagittal PDw aTSE including Alameda, San Mateo, Santa Clara and Santa Cruz Resolution 0.26 x 0.39 x • Sports Medicine is one of the hospitals specializations 2.0 mm (T1w) and 0.19 x • Average time planned for patients is 45 minutes 0.30 x 2.0 mm (PDw) Scan time 4:23 min (T1w) Patient: and 5:24 min (PDw) • Middle aged board surfer presents with chronic knee pain Courtesy: Palo Alto Medical • X-Ray shows no abnormalities Foundation, California, USA • Follow-up MR examination requested Case Highlights • dStream Knee 8ch coil is designed for high resolution visualization of fine cartilage structures and small ligaments Comprehensive Knee Imaging • Coil conforms to the anatomy for optimal signal-to-noise ratio – dStream Knee 8ch coil Description • Coil can be slightly rotated relative to its base plate Axial Dual IR sequence dStream Knee 8ch coil designed for high resolution to ease coil setup and enhance patient comfort Resolution 0.5 x 0.5 x 3.0 mm visualization of fine cartilage structures and small ligaments • Split coil design and single handle for easy coil setup Scan time 4:28 min Courtesy: Palo Alto Medical Foundation, California, USA Comprehensive Knee Imaging – dStream Knee 8ch coil Axial T2 calculation sequence (multi-echo SE) Resolution 0.4 x 0.5 x 2.5 mm Scan time 5:24 min Courtesy: Palo Alto Medical Foundation, California, USA Homogeneous fat suppression – dStream Knee 8ch coil Axial, Coronal and Sagittal PDw aTSE with SPAIR fat suppression, Resolution 0.24 x 0.34 x 2.0 mm, Scan time 4:44 min (axial and sagittal), 5:15 min (coronal) Courtesy: Palo Alto Medical Foundation, California, USA 4 Clinical Cases Ingenia 3.0T Clinical Cases Ingenia 3.0T 5
  • 4. Lumbar Spine in obese patient Ingenia 3.0T Background Information Lumbar Spine - dStream Hospital: Posterior FlexCoverage coil • Krankenhaus der Barmherzigen Brüder, Trier Germany Sagittal T2w TSE and T1w • One of the largest hospitals within its region TSE pre and post Gd with 612 beds and 15 clinical specialties Resolution 0.8 x 1.0 x 4.0 mm. Scan time 3:23 Patient: (T2w) and 2:19 (T1w) • 32- year old male patient, 180 kg, with back pain Courtesy: Krankenhaus • Was first examined on CT but quality der Barmherzigen Brüder, was not sufficient for diagnosis Trier, Germany • MRI gave excellent images. Diagnosis: small protrusion of the disc L5/S1, no meningioma Case Highlights • 70 cm bore for increased patient comfort and easy positioning • Homogeneous signal intensity over the entire field of view • Integrated dStream Posterior FlexCoverage coil reduces patient set up time and provides high SNR and large signal penetration depth • MultiTransmit adapts the RF signals to each patient leading to superb image uniformity and faster imaging Description: Integrated dStream Posterior FlexCoverage coil reduces patient set Lumbar Spine - dStream up time and provides high SNR and large signal penetration depth Posterior FlexCoverage coil Coronal STIR, radial M2D myelo, axial T2w and T1w TSE Resolution 0.9 x 1.3 x 4.0 mm (STIR), 0.5 x 1.6 x 40.0 mm (myelo), 0.7 x 0.8 x 4.0 mm (axials). Scan time 3:23 (STIR), 2:40 (myelo), 3:21 (axial T2w TSE) and 5:09 (axial T1w TSE) Courtesy: Krankenhaus der Barmherzigen Brüder, Trier, Germany Lumbar Spine - dStream Posterior FlexCoverage coil Comparison between CT image and MR images Courtesy: Krankenhaus der Barmherzigen Brüder, Trier, Germany 6 Clinical Cases Ingenia 3.0T Clinical Cases Ingenia 3.0T 7
  • 5. Imaging 2.0 – Transforming care, Philips Healthcare is partof together. Royal Philips Electronics At Philips, we recognize that radiology is the How to reach us cornerstone of diagnosis and treatment. www.philips.com/healthcare healthcare@philips.com And that clinical integration and collaboration are key to more personal care, Asia better patient outcomes, and lower costs. +49 7031 463 2254 Imaging 2.0 signals our endless quest to raise Europe, Middle East, Africa the bar on clinical excellence – by working +49 7031 463 2254 with you to develop innovations that enable you to collaborate freely, diagnose Latin America confidently, and care passionately. +55 11 2125 0744 Together, we will continue to spark a North America revolution in imaging science with advanced +1 425 487 7000 technologies designed to deliver greater 800 285 5585 (toll free, US only) collaboration and integration, increased patient focus, and improved economic value. Learn more at www.philips.com/imaging2.0 Please visit www.philips.com/mri © 2012 Koninklijke Philips Electronics N.V. All rights are reserved. Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. Printed in The Netherlands. NOV 2012