PRODUCTION OF X-RAYS
Dr. Khazi Mohammed Athar
X-Ray tube
KILOVOLTAGE,MILLIAMPERAGE
Two sources of electrical energy are required and are derived from AC
mains using transformer.
1) Filament heating voltage(10V), current (10A)
2) Accelerating voltage(typically 30-150kV) between anode and
cathode; this derives current of electrons(typically around 0.5-
1000mA).
The mA is controlled by adjusting filament voltage and current and
thus filament temperature.
1) Cathode: Negative terminal of tube is cathode.
It has filament which is the source of electrons made up of
Tungsten wire (0.2mm in diameter)coiled in vertical spiral
filament(0.2cm in diameter and 1cm or less in length).
Connecting wires which supply heating filament voltage (about 10V) and
amperage(average about 3-5A) which heats the filament and metal
focussing cup.
When metal is heated, atoms absorbs thermal energy adequate in amount
to escape from its nuclear attraction by a process called Thermionic
emission.
Electron cloud surrounding filament produced by thermionic emission is
termed as Edison effect. Pure tungsten must be heated to atleast to 2200°c
to emit useful number of electrons.
Electrons emitted from tungsten filament form a small cloud in
immediate vicinity of filament. This collection is called Space charge.
The tendency of space charge to limit emission of more electrons from
filament is called space charge effect.
High currents that can be produced by thermionic emission are possible
because large numbers of electrons can be accelerated from cathode to
anode.
The number of x rays produced depends on number of electrons that flow
from filament to target anode
The tube current measured in mA refers to number of electrons flowing
per second from filament to target.
Tungsten is not as efficient as it’s alloys in emitting electrons but it can
be drawn into thin wire and quite strong, has high melting
point(3370°)and little tendency to vaporise.
2) Glass enclosure: Two electrodes are sealed in Pyrex glass in a
vacuum. Vacuum allows the number and speed of accelerating electrons
to be controlled independently.
Pyrex can withstand high temperatures it is used in x ray tubes
Connecting wires must be sealed into glass wall.
3) Focusing cup: is maintained at negative potential.
So that it’s electrical forces causes electron stream to converge onto
target in required size and shape.
It is usually made up of Nickel.
4) Anode: Tungsten is chosen as target material because,
I) Has high atomic number
II) High melting point
III) Good material for absorption and rapid dissipation of heat.
Positive electrode can be stationary and rotating
Stationary: Small plate(square or rectangle, each side >1cm) of tungsten
embedded in large mass of copper. Anode angle is usually 15-20°.
Focal spot is area of anode bombarded by electrons from cathode. Most of
energy of electrons is lost in the form of heat, with less than 1% being
converted to x rays.
Line focus principle: Because of angulation of anode(6-20°) when focal spot
is viewed from direction in which x rays emerge, surface is foreshortened
and appears smaller.
Decrease in projected focal spot size is directly related to sine of angle
of anode.
Rotating: The purpose of rotating anode is to spread heat over large area
produced during exposure.
The rotating anode consists of large disc of tungsten or an alloy of
tungsten, which rotates at speed of 3600 rpm (theoretically)and disc has
beveled edge. Angle of bevel 6-20°
The power to effect rotation is provided by magnetic field produced by
stator coils that surround neck of x ray tube outside the envelope.
Magnetic field produced by coils induces current in copper rotor of
induction motor, this induced current provides power to rotate the
anode assembly.
Metallic lubricant like silver is used to reduce friction in bearings
Stem made up of Molybdenum is used to connect Tungsten target with
rest of anode assembly in order to avoid undesirable absorption of heat by
bearings of anode as Molybdenum is poor conductor of heat.
Length of Molybdenum is kept as short as possible to reduce load on
bearings.
Anode made up of alloy of Tungsten and Rhenium make it more
resistant to surface roughening.
GRID CONTROLLED X-RAY TUBE
A third electrode is used in focussing cup that surrounds that’s filament.
Negative bias voltage around 1500V is applied to cup relative to filament
to punch off flow of electrons.It act as on an off switch.
Intensity of x ray beam that leaves x ray tube is not uniform throughout
all portions of beam.
Intensity of beam toward
anode side is significantly
less than cathode side.
This factor can be used in
obtaining balanced densit
ies in radiographs of body
parts of different thickness
Thicker parts should be pl
aced towards cathode side
Tube shielding: Tube housing is lined with lead and serves to absorb
primary and secondary x rays that would otherwise produce needless
exposure to patient and personnel.
Off focus radiation: Produced when high speed electrons interact with
metal other than focal tract of anode usually from backscattering from
anode.It can partly controlled by collimators or lead diaphragm.
Characteristic Brehmsstrahlung
THANK YOU

Contenu connexe

Similaire à X-ray ppt.pptx

Production of x rays & generators
Production of x rays & generatorsProduction of x rays & generators
Production of x rays & generatorsRakesh Ca
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdfgametheory686sharech
 
Production of x rays chapter -1
Production  of  x   rays  chapter -1Production  of  x   rays  chapter -1
Production of x rays chapter -1sadhamhussain52
 
X ray tube and production of xrays
X ray tube and production of xraysX ray tube and production of xrays
X ray tube and production of xraysIndermeet Mangat
 
medical equipment lec 2 x-rays
medical equipment lec 2 x-raysmedical equipment lec 2 x-rays
medical equipment lec 2 x-rayscairo university
 
Components of X-Ray Tube
Components of X-Ray TubeComponents of X-Ray Tube
Components of X-Ray TubeAlisha Karmali
 
X ray tube production rakesh
X ray tube production rakeshX ray tube production rakesh
X ray tube production rakeshSylvia Diamond
 
ProductionOfXrays.pptx
ProductionOfXrays.pptxProductionOfXrays.pptx
ProductionOfXrays.pptxIslah Raoof
 
X-Ray Tube complete-1.pptx
X-Ray Tube complete-1.pptxX-Ray Tube complete-1.pptx
X-Ray Tube complete-1.pptxNMdcat2021
 
X RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxX RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxSrinath Chowdary
 
Production of x rays
Production of x raysProduction of x rays
Production of x rayspraveen kumar
 
Ch.2-Anatomy of x-ray machine.ppt
Ch.2-Anatomy of x-ray machine.pptCh.2-Anatomy of x-ray machine.ppt
Ch.2-Anatomy of x-ray machine.pptAbubakar Hammadama
 
Xray tubes .pptx
Xray tubes .pptxXray tubes .pptx
Xray tubes .pptxrose25eva
 
Basic principle of x ray
Basic principle of x rayBasic principle of x ray
Basic principle of x rayprahlad maurya
 
file_5e883cd7d3f5c.pdf
file_5e883cd7d3f5c.pdffile_5e883cd7d3f5c.pdf
file_5e883cd7d3f5c.pdfpradyumna42
 
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptxPRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptxManojMandal65
 

Similaire à X-ray ppt.pptx (20)

Production of x rays & generators
Production of x rays & generatorsProduction of x rays & generators
Production of x rays & generators
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdf
 
Production of x rays chapter -1
Production  of  x   rays  chapter -1Production  of  x   rays  chapter -1
Production of x rays chapter -1
 
X ray tube and production of xrays
X ray tube and production of xraysX ray tube and production of xrays
X ray tube and production of xrays
 
medical equipment lec 2 x-rays
medical equipment lec 2 x-raysmedical equipment lec 2 x-rays
medical equipment lec 2 x-rays
 
Components of X-Ray Tube
Components of X-Ray TubeComponents of X-Ray Tube
Components of X-Ray Tube
 
X ray tube production rakesh
X ray tube production rakeshX ray tube production rakesh
X ray tube production rakesh
 
ProductionOfXrays.pptx
ProductionOfXrays.pptxProductionOfXrays.pptx
ProductionOfXrays.pptx
 
X-Ray Tube complete-1.pptx
X-Ray Tube complete-1.pptxX-Ray Tube complete-1.pptx
X-Ray Tube complete-1.pptx
 
X RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxX RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptx
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
Ch.2-Anatomy of x-ray machine.ppt
Ch.2-Anatomy of x-ray machine.pptCh.2-Anatomy of x-ray machine.ppt
Ch.2-Anatomy of x-ray machine.ppt
 
L 2 ct physics
L 2  ct physics L 2  ct physics
L 2 ct physics
 
Xray tubes .pptx
Xray tubes .pptxXray tubes .pptx
Xray tubes .pptx
 
Basic principle of x ray
Basic principle of x rayBasic principle of x ray
Basic principle of x ray
 
Lecture-2.2.pdf
Lecture-2.2.pdfLecture-2.2.pdf
Lecture-2.2.pdf
 
Production of xray
Production  of  xrayProduction  of  xray
Production of xray
 
X-ray Tube (Cathode)
X-ray Tube (Cathode)X-ray Tube (Cathode)
X-ray Tube (Cathode)
 
file_5e883cd7d3f5c.pdf
file_5e883cd7d3f5c.pdffile_5e883cd7d3f5c.pdf
file_5e883cd7d3f5c.pdf
 
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptxPRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
 

Dernier

PT MANAGEMENT OF URINARY INCONTINENCE.pptx
PT MANAGEMENT OF URINARY INCONTINENCE.pptxPT MANAGEMENT OF URINARY INCONTINENCE.pptx
PT MANAGEMENT OF URINARY INCONTINENCE.pptxdrtabassum4
 
5CL-ADB powder supplier 5cl adb 5cladba 5cl raw materials vendor on sale now
5CL-ADB powder supplier 5cl adb 5cladba 5cl raw materials vendor on sale now5CL-ADB powder supplier 5cl adb 5cladba 5cl raw materials vendor on sale now
5CL-ADB powder supplier 5cl adb 5cladba 5cl raw materials vendor on sale nowSherrylee83
 
Cas 28578-16-7 PMK ethyl glycidate ( new PMK powder) best suppler
Cas 28578-16-7 PMK ethyl glycidate ( new PMK powder) best supplerCas 28578-16-7 PMK ethyl glycidate ( new PMK powder) best suppler
Cas 28578-16-7 PMK ethyl glycidate ( new PMK powder) best supplerSherrylee83
 
Relationship between vascular system disfunction, neurofluid flow and Alzheim...
Relationship between vascular system disfunction, neurofluid flow and Alzheim...Relationship between vascular system disfunction, neurofluid flow and Alzheim...
Relationship between vascular system disfunction, neurofluid flow and Alzheim...Catherine Liao
 
Antiplatelets in IHD, Dose Duration, DAPT vs SAPT
Antiplatelets in IHD, Dose Duration, DAPT vs SAPTAntiplatelets in IHD, Dose Duration, DAPT vs SAPT
Antiplatelets in IHD, Dose Duration, DAPT vs SAPTAkashGanganePatil1
 
Multiple sclerosis diet.230524.ppt3.pptx
Multiple sclerosis diet.230524.ppt3.pptxMultiple sclerosis diet.230524.ppt3.pptx
Multiple sclerosis diet.230524.ppt3.pptxMeenakshiGursamy
 
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...Catherine Liao
 
CURRENT HEALTH PROBLEMS AND ITS SOLUTION BY AYURVEDA.pptx
CURRENT HEALTH PROBLEMS AND ITS SOLUTION BY AYURVEDA.pptxCURRENT HEALTH PROBLEMS AND ITS SOLUTION BY AYURVEDA.pptx
CURRENT HEALTH PROBLEMS AND ITS SOLUTION BY AYURVEDA.pptxDr KHALID B.M
 
Creating Accessible Public Health Communications
Creating Accessible Public Health CommunicationsCreating Accessible Public Health Communications
Creating Accessible Public Health Communicationskatiequigley33
 
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1DR SETH JOTHAM
 
TUBERCULINUM-2.BHMS.MATERIA MEDICA.HOMOEOPATHY
TUBERCULINUM-2.BHMS.MATERIA MEDICA.HOMOEOPATHYTUBERCULINUM-2.BHMS.MATERIA MEDICA.HOMOEOPATHY
TUBERCULINUM-2.BHMS.MATERIA MEDICA.HOMOEOPATHYDRPREETHIJAMESP
 
linearity concept of significance, standard deviation, chi square test, stude...
linearity concept of significance, standard deviation, chi square test, stude...linearity concept of significance, standard deviation, chi square test, stude...
linearity concept of significance, standard deviation, chi square test, stude...KavyasriPuttamreddy
 
Scientificity and feasibility study of non-invasive central arterial pressure...
Scientificity and feasibility study of non-invasive central arterial pressure...Scientificity and feasibility study of non-invasive central arterial pressure...
Scientificity and feasibility study of non-invasive central arterial pressure...Catherine Liao
 
SURGICAL ANATOMY OF ORAL IMPLANTOLOGY.pptx
SURGICAL ANATOMY OF ORAL IMPLANTOLOGY.pptxSURGICAL ANATOMY OF ORAL IMPLANTOLOGY.pptx
SURGICAL ANATOMY OF ORAL IMPLANTOLOGY.pptxSuresh Kumar K
 
Circulation through Special Regions -characteristics and regulation
Circulation through Special Regions -characteristics and regulationCirculation through Special Regions -characteristics and regulation
Circulation through Special Regions -characteristics and regulationMedicoseAcademics
 
Factors Affecting child behavior in Pediatric Dentistry
Factors Affecting child behavior in Pediatric DentistryFactors Affecting child behavior in Pediatric Dentistry
Factors Affecting child behavior in Pediatric DentistryDr Simran Deepak Vangani
 
Vaccines: A Powerful and Cost-Effective Tool Protecting Americans Against Dis...
Vaccines: A Powerful and Cost-Effective Tool Protecting Americans Against Dis...Vaccines: A Powerful and Cost-Effective Tool Protecting Americans Against Dis...
Vaccines: A Powerful and Cost-Effective Tool Protecting Americans Against Dis...PhRMA
 
CT scan of penetrating abdominopelvic trauma
CT scan of penetrating abdominopelvic traumaCT scan of penetrating abdominopelvic trauma
CT scan of penetrating abdominopelvic traumassuser144901
 
BMK Glycidic Acid (sodium salt) CAS 5449-12-7 Pharmaceutical intermediates
BMK Glycidic Acid (sodium salt)  CAS 5449-12-7 Pharmaceutical intermediatesBMK Glycidic Acid (sodium salt)  CAS 5449-12-7 Pharmaceutical intermediates
BMK Glycidic Acid (sodium salt) CAS 5449-12-7 Pharmaceutical intermediatesdorademei
 

Dernier (20)

PT MANAGEMENT OF URINARY INCONTINENCE.pptx
PT MANAGEMENT OF URINARY INCONTINENCE.pptxPT MANAGEMENT OF URINARY INCONTINENCE.pptx
PT MANAGEMENT OF URINARY INCONTINENCE.pptx
 
5CL-ADB powder supplier 5cl adb 5cladba 5cl raw materials vendor on sale now
5CL-ADB powder supplier 5cl adb 5cladba 5cl raw materials vendor on sale now5CL-ADB powder supplier 5cl adb 5cladba 5cl raw materials vendor on sale now
5CL-ADB powder supplier 5cl adb 5cladba 5cl raw materials vendor on sale now
 
Cas 28578-16-7 PMK ethyl glycidate ( new PMK powder) best suppler
Cas 28578-16-7 PMK ethyl glycidate ( new PMK powder) best supplerCas 28578-16-7 PMK ethyl glycidate ( new PMK powder) best suppler
Cas 28578-16-7 PMK ethyl glycidate ( new PMK powder) best suppler
 
Relationship between vascular system disfunction, neurofluid flow and Alzheim...
Relationship between vascular system disfunction, neurofluid flow and Alzheim...Relationship between vascular system disfunction, neurofluid flow and Alzheim...
Relationship between vascular system disfunction, neurofluid flow and Alzheim...
 
Antiplatelets in IHD, Dose Duration, DAPT vs SAPT
Antiplatelets in IHD, Dose Duration, DAPT vs SAPTAntiplatelets in IHD, Dose Duration, DAPT vs SAPT
Antiplatelets in IHD, Dose Duration, DAPT vs SAPT
 
Multiple sclerosis diet.230524.ppt3.pptx
Multiple sclerosis diet.230524.ppt3.pptxMultiple sclerosis diet.230524.ppt3.pptx
Multiple sclerosis diet.230524.ppt3.pptx
 
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
 
CURRENT HEALTH PROBLEMS AND ITS SOLUTION BY AYURVEDA.pptx
CURRENT HEALTH PROBLEMS AND ITS SOLUTION BY AYURVEDA.pptxCURRENT HEALTH PROBLEMS AND ITS SOLUTION BY AYURVEDA.pptx
CURRENT HEALTH PROBLEMS AND ITS SOLUTION BY AYURVEDA.pptx
 
Creating Accessible Public Health Communications
Creating Accessible Public Health CommunicationsCreating Accessible Public Health Communications
Creating Accessible Public Health Communications
 
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
 
TUBERCULINUM-2.BHMS.MATERIA MEDICA.HOMOEOPATHY
TUBERCULINUM-2.BHMS.MATERIA MEDICA.HOMOEOPATHYTUBERCULINUM-2.BHMS.MATERIA MEDICA.HOMOEOPATHY
TUBERCULINUM-2.BHMS.MATERIA MEDICA.HOMOEOPATHY
 
linearity concept of significance, standard deviation, chi square test, stude...
linearity concept of significance, standard deviation, chi square test, stude...linearity concept of significance, standard deviation, chi square test, stude...
linearity concept of significance, standard deviation, chi square test, stude...
 
Scientificity and feasibility study of non-invasive central arterial pressure...
Scientificity and feasibility study of non-invasive central arterial pressure...Scientificity and feasibility study of non-invasive central arterial pressure...
Scientificity and feasibility study of non-invasive central arterial pressure...
 
SURGICAL ANATOMY OF ORAL IMPLANTOLOGY.pptx
SURGICAL ANATOMY OF ORAL IMPLANTOLOGY.pptxSURGICAL ANATOMY OF ORAL IMPLANTOLOGY.pptx
SURGICAL ANATOMY OF ORAL IMPLANTOLOGY.pptx
 
Circulation through Special Regions -characteristics and regulation
Circulation through Special Regions -characteristics and regulationCirculation through Special Regions -characteristics and regulation
Circulation through Special Regions -characteristics and regulation
 
Factors Affecting child behavior in Pediatric Dentistry
Factors Affecting child behavior in Pediatric DentistryFactors Affecting child behavior in Pediatric Dentistry
Factors Affecting child behavior in Pediatric Dentistry
 
Vaccines: A Powerful and Cost-Effective Tool Protecting Americans Against Dis...
Vaccines: A Powerful and Cost-Effective Tool Protecting Americans Against Dis...Vaccines: A Powerful and Cost-Effective Tool Protecting Americans Against Dis...
Vaccines: A Powerful and Cost-Effective Tool Protecting Americans Against Dis...
 
CT scan of penetrating abdominopelvic trauma
CT scan of penetrating abdominopelvic traumaCT scan of penetrating abdominopelvic trauma
CT scan of penetrating abdominopelvic trauma
 
BMK Glycidic Acid (sodium salt) CAS 5449-12-7 Pharmaceutical intermediates
BMK Glycidic Acid (sodium salt)  CAS 5449-12-7 Pharmaceutical intermediatesBMK Glycidic Acid (sodium salt)  CAS 5449-12-7 Pharmaceutical intermediates
BMK Glycidic Acid (sodium salt) CAS 5449-12-7 Pharmaceutical intermediates
 
Scleroderma: Treatment Options and a Look to the Future - Dr. Macklin
Scleroderma: Treatment Options and a Look to the Future - Dr. MacklinScleroderma: Treatment Options and a Look to the Future - Dr. Macklin
Scleroderma: Treatment Options and a Look to the Future - Dr. Macklin
 

X-ray ppt.pptx

  • 1. PRODUCTION OF X-RAYS Dr. Khazi Mohammed Athar
  • 3. KILOVOLTAGE,MILLIAMPERAGE Two sources of electrical energy are required and are derived from AC mains using transformer. 1) Filament heating voltage(10V), current (10A) 2) Accelerating voltage(typically 30-150kV) between anode and cathode; this derives current of electrons(typically around 0.5- 1000mA). The mA is controlled by adjusting filament voltage and current and thus filament temperature.
  • 4. 1) Cathode: Negative terminal of tube is cathode. It has filament which is the source of electrons made up of Tungsten wire (0.2mm in diameter)coiled in vertical spiral filament(0.2cm in diameter and 1cm or less in length). Connecting wires which supply heating filament voltage (about 10V) and amperage(average about 3-5A) which heats the filament and metal focussing cup. When metal is heated, atoms absorbs thermal energy adequate in amount to escape from its nuclear attraction by a process called Thermionic emission. Electron cloud surrounding filament produced by thermionic emission is termed as Edison effect. Pure tungsten must be heated to atleast to 2200°c to emit useful number of electrons.
  • 5. Electrons emitted from tungsten filament form a small cloud in immediate vicinity of filament. This collection is called Space charge. The tendency of space charge to limit emission of more electrons from filament is called space charge effect. High currents that can be produced by thermionic emission are possible because large numbers of electrons can be accelerated from cathode to anode. The number of x rays produced depends on number of electrons that flow from filament to target anode The tube current measured in mA refers to number of electrons flowing per second from filament to target.
  • 6. Tungsten is not as efficient as it’s alloys in emitting electrons but it can be drawn into thin wire and quite strong, has high melting point(3370°)and little tendency to vaporise. 2) Glass enclosure: Two electrodes are sealed in Pyrex glass in a vacuum. Vacuum allows the number and speed of accelerating electrons to be controlled independently. Pyrex can withstand high temperatures it is used in x ray tubes Connecting wires must be sealed into glass wall. 3) Focusing cup: is maintained at negative potential. So that it’s electrical forces causes electron stream to converge onto target in required size and shape. It is usually made up of Nickel.
  • 7. 4) Anode: Tungsten is chosen as target material because, I) Has high atomic number II) High melting point III) Good material for absorption and rapid dissipation of heat. Positive electrode can be stationary and rotating Stationary: Small plate(square or rectangle, each side >1cm) of tungsten embedded in large mass of copper. Anode angle is usually 15-20°. Focal spot is area of anode bombarded by electrons from cathode. Most of energy of electrons is lost in the form of heat, with less than 1% being converted to x rays. Line focus principle: Because of angulation of anode(6-20°) when focal spot is viewed from direction in which x rays emerge, surface is foreshortened and appears smaller.
  • 8. Decrease in projected focal spot size is directly related to sine of angle of anode. Rotating: The purpose of rotating anode is to spread heat over large area produced during exposure. The rotating anode consists of large disc of tungsten or an alloy of tungsten, which rotates at speed of 3600 rpm (theoretically)and disc has beveled edge. Angle of bevel 6-20° The power to effect rotation is provided by magnetic field produced by stator coils that surround neck of x ray tube outside the envelope.
  • 9. Magnetic field produced by coils induces current in copper rotor of induction motor, this induced current provides power to rotate the anode assembly.
  • 10. Metallic lubricant like silver is used to reduce friction in bearings Stem made up of Molybdenum is used to connect Tungsten target with rest of anode assembly in order to avoid undesirable absorption of heat by bearings of anode as Molybdenum is poor conductor of heat. Length of Molybdenum is kept as short as possible to reduce load on bearings. Anode made up of alloy of Tungsten and Rhenium make it more resistant to surface roughening. GRID CONTROLLED X-RAY TUBE A third electrode is used in focussing cup that surrounds that’s filament. Negative bias voltage around 1500V is applied to cup relative to filament to punch off flow of electrons.It act as on an off switch.
  • 11. Intensity of x ray beam that leaves x ray tube is not uniform throughout all portions of beam. Intensity of beam toward anode side is significantly less than cathode side. This factor can be used in obtaining balanced densit ies in radiographs of body parts of different thickness Thicker parts should be pl aced towards cathode side
  • 12. Tube shielding: Tube housing is lined with lead and serves to absorb primary and secondary x rays that would otherwise produce needless exposure to patient and personnel. Off focus radiation: Produced when high speed electrons interact with metal other than focal tract of anode usually from backscattering from anode.It can partly controlled by collimators or lead diaphragm.
  • 14.
  • 15.